1 Detection of endograft fractures with multidetector row computed tomography Justus E. Roos, MD, Jeffrey C. Hellinger, MD, Richard Hallet, MD, Dominik Fleischmann, MD, Christopher K. Zarins, MD, and Geoffrey D. Rubin, MD, Stanford, Calif Delayed endograft metallic strut failures detected in vivo with multidetector row computed tomography (MDCT) are reported in two patients who underwent endovascular abdominal aortic aneurysm repair with AneuRx and Talent endografts. In both instances, nitinol fractures were associated with proximal migration and type I endoleak. In both cases, the metallic strut fractures were detected with transverse sections from 16-channel MDCT angiograms and confirmed by using volume rendering. These cases highlight the previously unreported ability of thin-section, highresolution MDCT angiography to detect endograft strut fractures. (J Vasc Surg 2005;42: ) Successful endovascular aneurysm repair requires secure fixation and sealing in nonaneurysmal aortic and iliac arterial segments, proximal and distal to the abdominal aortic aneurysm (AAA), respectively. After deployment, patients require routine surveillance to assess for a variety of complications, including limb occlusion, 1 endoleaks, 2 endograft migration, 3 structural failure, 4-11 aneurysm sac enlargement, and AAA rupture. 12 In many centers, the assessment for metallic strut fractures is performed with abdominal radiographs, and either computed tomographic (CT) angiography or duplex sonography is performed for identifying endoleak and changes in aneurysm size. A spectrum of follow-up and imaging recommendations are available from the manufacturers. To our knowledge, data have not been published on the level of compliance with these recommendations to indicate the frequency with which abdominal radiographs are acquired routinely after endograft repair. Recent technical improvements in CT scanners allow the routine acquisition of thin, approximately 1-mm-thick sections, and this improves the possibility that CT would be suitable for detecting metallic strut fractures. We present two cases of endograft metallic strut fracture detected exclusively by CT. To our knowledge, these are the first reported cases of strut fracture detected prospectively by CT. CASE REPORTS Imaging technique. For both cases, CT angiograms were obtained by using a 16-channel multidetector row CT (MDCT) scanner (Lightspeed; GE Medical Systems, Milwaukee, Wis) with the following parameters: 1.25-mm detector width, pitch, 0.6-second gantry rotation speed, 325 ma and 120 kvp, 1.25-mm nominal reconstructed section thickness, and 0.8-mm reconstruction interval. A total of 120 ml of nonionic contrast medium with an iodine content of 350 mg/ml (Omnipaque-350; General From the Department of Radiology, Stanford Medical Center. Competition of interest: none. Reprint requests: Justus E. Roos, MD, Department of Radiology, Stanford Medical Center, 300 Pasteur Dr, Room S-072, Stanford, CA ( justus.roos@stanford.edu) /$30.00 Copyright 2005 by The Society for Vascular Surgery. doi: /j.jvs Electric Healthcare Biosciences, Princeton, NJ) was injected at 5 ml/s trough a 20-gauge antecubital intravenous catheter. Image interpretation. The CT data were transferred to a server-based online three-dimensional workstation (AquariusNet; TeraRecon, San Mateo, Calif). Transverse CT sections, multiplanar reformations, and volume renderings were assessed interactively and in real time by the interpreting radiologist in both cases. Interaction time required approximately 5 minutes per case. Patient 1. An 89-year-old woman with a 72-mm infrarenal AAA underwent successful exclusion with a bifurcated 26-mm AneuRx endograft (Medtronic, Minneapolis, Minn). The AneuRx endograft is a modular, bifurcated, self-expanding Dacron graft (DuPont, Wilmington, Del) supported with a nitinol exoskeleton. 13 At 25 months after deployment, 16-channel MDCT angiography demonstrated an opening of the fourth row of metallic struts anteriorly, disruption of at least four junctions between the superior aspect of the fourth row of metallic struts and the inferior aspect of the third strut row, and disruption of five junctions between the inferior aspect of the fourth row of metallic struts and the superior aspect of the fifth strut row (Fig 1, C-E). In comparison to the MDCT angiogram performed after endograft deployment, there was interval loss of the apposition between the endograft and the aortic wall at the proximal neck, development of a type Ia endoleak (Fig 1, C), expansion of the proximal neck from 21 to 27 mm, and a 5-mm distal migration of the anterior aspect of the proximal end of the endograft (Fig 1, A and B). The AAA sac had not enlarged compared with 25 months previously. No abdominal radiographs were acquired simultaneously with the MDCT angiogram that showed the endograft fracture. Patient 2. A 76-year-old man with a mm infrarenal AAA had a proximal infrarenal neck diameter and length of 30 mm each. The patient underwent endovascular aneurysm repair with a Talent endograft (34-mm proximal and 22-mm distal diameter; 155 mm long; Medtronic Vascular, Santa Rosa, Calif), which is a modular, bifurcated, self-expanding device consisting of one uncovered proximal stent, making suprarenal deployment possible, and multiple Dacron-covered nitinol stents connected by several longitudinal metallic bars. 14 Three days after deployment, MR angiography was performed because of an increased serum creatinine level and depicted an endoleak adjacent to the proximal aspect of the flow divider of the endograft. Subsequent MR angiograms 1 and 7 months after deployment of the endograft showed no

2 JOURNAL OF VASCULAR SURGERY Volume 42, Number 5 Roos et al 1003 Fig 1. Multidetector row computed tomography images of a patient treated with the AneuRx endograft. The first postdeployment MIP image (A) demonstrates anterior angulation of the proximal endograft; the MIP image 25 months after deployment (B) reveals a marked change in the angulation of the endograft, with caudal migration of the anterior rim of the endograft and no change in the position of the posterior rim of the endograft. This was associated with the development of a type I endoleak. A cross-sectional image (C) demonstrates the proximal type I endoleak with contrast in the aneurysm sac (arrowheads) and a break in the integrity of the stent ring (arrow). Volume-rendered right oblique (D) and left oblique (E) images demonstrate a nitinol stent ring fracture with disruption of the metallic framework (double arrowheads).

3 1004 Roos et al JOURNAL OF VASCULAR SURGERY November 2005 Fig 2. Transverse multidetector row computed tomography and volume-rendered (VR) images demonstrating the Talent endograft at 15 months after deployment. The transverse image (A) and VR image (B) demonstrate a proximal type I endoleak with contrast outside the lumen of the endograft in the posterior aspect of the aneurysm sac (arrowheads). Although an isolated metallic strut fracture is hard to recognize on the transverse image, a VR image (C) through the same view angle as (B), but rendered with an opacity transfer function that excludes the contrast-enhanced aortic lumen, allows precise localization of the metallic framework. Fracture of the uppermost stent ring is apparent (arrows). A VR image (D) isolates the proximal stent ring and best demonstrates the separation (curved arrow) of the metallic framework at the stent ring fracture. change in the AAA sac size or in the extent of endoleak. Fifteen months after deployment, the patient s serum creatinine normalized, and a CT angiograph was performed. Sixteen-channel MDCT angiography demonstrated a single fracture of a proximal strut in the first row on the posterior aspect of the endograft (Fig 2, C and D) with a new accompanying type Ia endoleak (Fig 2, A and B) at the proximal aspect of the endograft. The aneurysm diameter increased from 70 to 80 mm. Abdominal radiographs were not acquired with this MDCT angiogram. DISCUSSION Despite an improved understanding of the metallurgic and fabric properties of endografts, material failure continues to be a potential problem that can precipitate

4 JOURNAL OF VASCULAR SURGERY Volume 42, Number 5 Roos et al 1005 endoleaks, device migration, aneurysm sac expansion, and aneurysm rupture. 6-11,15-18 The primary failure modes of endografts are suture breakage, fabric fatigue, and metallic strut fractures. Metallic strut fractures are potentially caused by stress fatigue and metal corrosion. 6,7 A reduction in the structural integrity of the stent graft might be expected to be associated with a loss of device fixation and migration. Nevertheless, on the basis of the current literature, an explicit association between device failure and adverse outcomes has not been established. A recent study of 120 explanted AneuRx endografts reported that migration was associated with more stent strut fractures and that suture breaks were seen in most of the devices. Nevertheless, a causal relationship between device failure and clinical outcome was not established. 5 In an analysis of endograft fatigue in 686 patients, 60 patients had endograft fractures. 6 Most of these patients were asymptomatic and did not undergo intervention. While the true clinical relevance of endograft fracture detection is not known, until their significance is known, we should not dismiss their detection as being unimportant. CT scanning reveals endograft failure modes, and the detection of structural failures may lead to more robust designs in the future. In vivo visualization of strut fractures is challenging because of the thinness of the strut widths (approximating 0.25 mm) and the complex architecture of the metallic framework of the endograft. 6 Although conventional abdominal radiographs have sufficient spatial resolution to visualize the strut width, the depiction of strut fractures is often difficult. 10,19 To optimize these shortcomings, some manufacturers recommend oblique views in addition to standard anteroposterior and lateral projections. Abdominal radiographs were not obtained at the time of MDCT angiography; therefore, we do not know if abdominal radiographs would have detected the strut fractures in these two cases. Although the evaluation of the problems that can plague abdominal radiography, such as strut overlap, poor contrast in obese patients, and motion blurring, might result in CT s being the preferred method to assess metallic strut integrity, formal hypothesis testing is required before it is assumed that a single CT scan could replace the combination of abdominal radiographs and CT scans after endovascular aneurysm repair. To our knowledge, these are the first reported examples of prospective detection of stent-graft strut fracture with MDCT. The high volumetric spatial resolution of CT angiography affords unparalleled detail of the metallic exoskeleton in vivo. Although the strut fractures were initially detected from stacked transverse sections, interactive volume rendering provided a better depiction of the nature and number of the fractures and their relationship to the new type Ia endoleaks. The value of these visualization techniques is directly related to the quality of the CT acquisition. Had the CT scans been acquired with sections thicker than 1.5 mm and without at least a 50% overlap, detection of these strut fractures might not have been possible. In conclusion, this case report suggests that when properly performed and scrutinized, CT angiography can detect strut fractures that may precipitate endograft migration. However, establishing the diagnostic accuracy of CT for detecting strut fractures and determining their clinical relevance will require further investigation REFERENCES 1. Carpenter JP. Multicenter trial of the PowerLink bifurcated system for endovascular aortic aneurysm repair. J Vasc Surg 2002;36: Walschot LH, Laheij RJ, Verbeek AL. Outcome after endovascular abdominal aortic aneurysm repair: a meta-analysis. J Endovasc Ther 2002;9: Tonnessen BH, Sternbergh WC III, Money SR. Late problems at the proximal aortic neck: migration and dilation. Semin Vasc Surg 2004; 17: Najibi S, Steinberg J, Katzen BT, Zemel G, Lin PH, Weiss VJ, et al. Detection of isolated hook fractures 36 months after implantation of the Ancure endograft: a cautionary note. J Vasc Surg 2001;34: Zarins CK, Arko FR, Crabtree T, Bloch DA, Ouriel K, Allen RC, et al. Explant analysis of AneuRx stent grafts: relationship between structural findings and clinical outcome. J Vasc Surg 2004;40: Jacobs TS, Won J, Gravereaux EC, Faries PL, Morrissey N, Teodorescu VJ, et al. Mechanical failure of prosthetic human implants: a 10-year experience with aortic stent graft devices. J Vasc Surg 2003; 37: Heintz C, Riepe G, Birken L, Kaiser E, Chakfe N, Morlock M, et al. Corroded nitinol wires in explanted aortic endografts: an important mechanism of failure? J Endovasc Ther 2001;8: Chakfe N, Riepe G, Dieval F, Le Magnen JF, Wang L, Urban E, et al. Longitudinal ruptures of polyester knitted vascular prostheses. J Vasc Surg 2001;33: Beebe HG, Cronenwett JL, Katzen BT, Brewster DC, Green RM. Results of an aortic endograft trial: impact of device failure beyond 12 months. J Vasc Surg 2001;33:S Holzenbein TJ, Kretschmer G, Thurnher S, Schoder M, Aslim E, Lammer J, et al. Midterm durability of abdominal aortic aneurysm endograft repair: a word of caution. J Vasc Surg 2001;33:S Marin ML, Hollier LH, Ellozy SH, Spielvogel D, Mitty H, Griepp R, et al. Endovascular stent graft repair of abdominal and thoracic aortic aneurysms: a ten-year experience with 817 patients. Ann Surg 2003; 238:586-93; discussion Harris PL, Vallabhaneni SR, Desgranges P, Becquemin JP, van Marrewijk C, Laheij RJ. Incidence and risk factors of late rupture, conversion, and death after endovascular repair of infrarenal aortic aneurysms: the EUROSTAR experience. European Collaborators on Stent/graft techniques for aortic aneurysm repair. J Vasc Surg 2000;32: Zarins CK, White RA, Schwarten D, Kinney E, Diethrich EB, Hodgson KJ, et al. AneuRx stent graft versus open surgical repair of abdominal aortic aneurysms: multicenter prospective clinical trial. J Vasc Surg 1999;29: ; discussion May J, White GH, Harris JP. Devices for aortic aneurysm repair. Surg Clin North Am 1999;79: Carpenter JP, Baum RA, Barker CF, Golden MA, Velazquez OC, Mitchell ME, et al. Durability of benefits of endovascular versus conventional abdominal aortic aneurysm repair. J Vasc Surg 2002;35: Torsello G, Klenk E, Tessarek J. Aneurysm treatment with the Talent stent graft. An update after 4 years of experience [in German]. Chirurg 2002;73: Greenberg RK, Chuter TA, Sternbergh WC III, Fearnot NE. Zenith AAA endovascular graft: intermediate-term results of the US multicenter trial. J Vasc Surg 2004;39:

5 1006 Roos et al JOURNAL OF VASCULAR SURGERY November Sternbergh WC III, Money SR, Greenberg RK, Chuter TA. Influence of endograft oversizing on device migration, endoleak, aneurysm shrinkage, and aortic neck dilation: results from the Zenith Multicenter Trial. J Vasc Surg 2004;39: Gorich J, Rilinger N, Sokiranski R, Orend KH, Ermis C, Kramer SC, et al. Leakages after endovascular repair of aortic aneurysms: classification based on findings at CT, angiography, and radiography. Radiology 1999; 213: Hiatt MD, Rubin GD. Surveillance for endoleaks: how to detect all of them. Semin Vasc Surg 2004;17: Submitted Apr 1, 2005; accepted Jul 9, 2005.