Assessment of MRI Issues for a 3-T Immune Programmable CSF Shunt Valve

Size: px
Start display at page:

Download "Assessment of MRI Issues for a 3-T Immune Programmable CSF Shunt Valve"

Transcription

1 Neuroradiology/Head and Neck Imaging Original Research Shellock et al. Neuroradiology/Head and Neck Imaging Original Research Frank G. Shellock 1 Allison Bedwinek 2 Morgan Oliver-Allen 3 Stephen F. Wilson 2 Shellock FG, Bedwinek A, Oliver-Allen M, Wilson SF Keywords: artifacts, cerebrospinal fluid, implants, intracranial pressure, MRI, specific absorption rate DOI: /AJR Received October 7, 2010; accepted after revision December 22, Supported by Codman, a Johnson & Johnson Company, Raynham, MA. 1 Institute for Magnetic Resonance Safety, Education, and Research, Keck School of Medicine, University of Southern California, 7511 McConnell Ave, Los Angeles, CA Address correspondence to F. G. Shellock (Frank.Shellock@gte.net). 2 Product Development and Support Group, Codman, a Johnson & Johnson Company, Raynham, MA. 3 Loyola Marymount University, Los Angeles, CA AJR 2011; 197: X/11/ American Roentgen Ray Society Assessment of MRI Issues for a 3-T Immune Programmable CSF Shunt Valve OBJECTIVE. A newly developed CSF shunt valve that incorporates a magnetically adjustable mechanism designed to resist unintended setting changes was evaluated for problems during 3-T MRI. MATERIALS AND METHODS. Standardized protocols were used to assess magnetic field interactions, MRI-related heating, artifacts, and functional changes related to multiple exposures and various MRI conditions in nine different samples at 3 T. RESULTS. The magnetic field interactions were not excessive. MRI-related heating, which was studied at a relatively high, MRI system reported whole body averaged specific absorption rate (2.9 W/kg), was at a level that should not pose a hazard to a patient. Although artifacts were large in relation to the dimensions of this programmable CSF shunt valve, the results were consistent with similar devices containing permanent magnets. Multiple exposures and various MRI conditions at 3 T did not damage or affect the functional aspects of the devices, and no unintentional changes to the valve setting were observed. CONCLUSION. In consideration of the test results, this new programmable CSF shunt valve is not adversely affected by the 3-T MRI environment and is acceptable for a patient undergoing MRI at 3 T or less when specific guidelines are followed, including verifying the valve setting according to manufacturer recommendations immediately after the MRI procedure. A CSF shunt is an implanted device that directs CSF in a controlled manner away from CNS fluid compartments to an internal site, providing an alternative pathway when CSF absorption is compromised or impaired owing to obstruction or other conditions [1, 2]. Controlling CSF volume and intracranial pressure with a shunt alleviates the buildup of excess fluid that results in hydrocephalus [1, 2]. The basic components of a shunt include a flexible inflow catheter used to drain CSF from the cerebral ventricles and subarachnoid spaces, a valve mechanism that regulates differential pressure or controls flow through the shunt catheters, and a flexible outflow catheter that directs CSF from the valve to a suitable drainage site, such as the peritoneum [1, 2]. Modern shunts commonly incorporate a programmable valve, which regulates pressure at a setting selected by the clinician according to the needs of the patient. The resistance of the programmable CSF shunt valve is typically adjusted, or programmed, with a magnetic field transmitted over the valve with an externally applied device. The ability to use a programmable CSF shunt valve as opposed to a fixed or monopressure valve allows noninvasive adjustments of the shunt system before and after implantation, optimizing patient care. Several magnetically programmable CSF shunt valves have been evaluated in association with MRI [3 17]. The 3-T MRI environment has been reported to pose problems that do not exist at 1.5 T. The difference is likely due to the nature of the mechanism of action involved in the magnetic components of a given shunt valve relative to this powerful static magnetic field [3 15]. One of the most troublesome performance problems is the possibility of inadvertent or unrecognized change in the valve setting, which can have serious consequences for the patient [9 11]. Because each magnetically programmable CSF shunt valve has unique operational components and features, it is necessary to characterize MRI problems case by case. Inherent differences can be hazardous in association with MRI, particularly in the 3-T MRI environment [7 14]. To ensure safe use 202 AJR:197, July 2011

2 of a newly developed programmable CSF shunt valve in patients referred for MRI, we conducted this investigation to evaluate factors that may affect this implant in the 3-T MRI environment (i.e., the worst case for a clinical MRI system). The factors included magnetic field interactions (translational attraction and torque), MRI-related heating, artifacts, and functional changes associated with multiple exposures and various MRI conditions [12 14]. Materials and Methods Programmable CSF Shunt Valve A magnetically programmable CSF shunt valve (Codman Certas Programmable Valve, Codman, a Johnson & Johnson Company), which is in regulatory review, was evaluated in a 3-T MRI environment (Fig. 1). This device is used to regulate intraventricular pressure by means of a ball-and-cone mechanism under the control of a calibrated flat spring and can be noninvasively adjusted with an external tool whereby permanent magnet elements are used to alter the force applied to the ball. Hard magnets fixed within a moveable cam react to the presence of the magnetic field of the tool by first translating vertically to an unlocked position and subsequently rotating in response to the movement of the external tool. Removal of the tool from the moveable cam returns the cam to a locked position in which rotation from one setting to another is prevented. Owing to its design, this valve inherently does not respond to large magnetic fields, such as those associated with MRI systems. The requirement that the cam first be attracted upward to allow its rotation and subsequently be rotated to a different setting is not allowed by the combined influence of the orientation of the magnets in the cam or the limited freedom of cam movement imposed by the fixed axle on which the cam translates and rotates. Exposure of the cam to a large magnetic gradient while it is in the locked, or down, position results in a rotational influence that is prevented by the constraining features of the mechanism. Furthermore, exposure of the cam to a strong static magnetic field causes the cam to bind on its axle because it is being drawn to rotate out of the permissible plane. This process hinders both translation and rotation of the cam around the axle. The adjustable CSF shunt valve accommodates eight settings by which the moveable cam provides an incrementally different offset to the flat spring, changing the performance of the valve (i.e., the valve setting). Metallic materials for the programmable valve include 316L stainless steel, unalloyed titanium, tantalum, and neodymium (Nd-Fe-B magnets) (Fig. 1). Inflow and outflow catheters made from silicone are used with this CSF shunt system. Fig. 1 Schematic shows programmable CSF shunt valve subjected to 3-T MRI. Silicone housing and upper casing of adjustment mechanism are translucent to show following components: two neodymium magnets (1), 316L stainless steel retention spring (2), 316L stainless steel spring (3), tantalum bead (4), titanium axle (5). Magnetic Field Interactions Previously described test procedures [12, 13, 18] were used to evaluate the programmable CSF shunt valve with two silicone catheters attached for translational attraction and torque in association with a 3-T active shielded horizontal-field MRI system (Excite, software G B, GE Healthcare). Translational Attraction The deflection angle test was used to assess translational attraction [12, 13, 18]. The programmable CSF shunt valve was attached to a special test fixture to measure the deflection angle in the MRI system. The test fixture consisted of a sturdy structure capable of holding the device in position without movement and contained a protractor with 1 graduated markings rigidly mounted to the structure. The 0 indicator on the protractor was oriented vertically. The programmable CSF shunt valve was suspended from a thin lightweight (weight less than 1% of the device) string attached at the 0 indicator position on the protractor. The test apparatus was positioned in the 3-T MRI system at the point of the highest spatial gradient magnetic field. For the 3-T unit used in this investigation, the highest spatial gradient magnetic field was 720 gauss/cm [18]. Sources of forced air movement within the MRI system bore were turned off during the measurements. The maximum deflection angle from vertical to the nearest 1 was measured three times for the valve, and an average value was calculated [18, 19]. Qualitative Assessment of Torque A previously described qualitative assessment technique [12, 13, 18] was used to determine the presence of magnetic field induced torque on the programmable CSF shunt valve. The test apparatus was a flat plastic device with a millimeter grid. The programmable CSF shunt valve was placed on the device in an orientation 45 relative to the static magnetic field of the 3-T MRI system. The apparatus was then positioned in the center of the MRI unit, where the effect of torque is the greatest, and the valve was observed for possible alignment or rotation relative to the 3-T static magnetic field. The programmable CSF shunt valve was then moved 45 relative to its previous position and observed for alignment or rotation. This process was repeated to encompass a full 360 rotation of positions for the valve. The following qualitative scale was applied to the results [12, 13, 18]: 0, no torque; +1, mild or low torque, the device slightly changed in orientation but did not align to the magnetic field; +2, moderate torque, the device aligned gradually to the magnetic field; +3, strong torque, the device showed rapid and forceful alignment to the magnetic field; +4, very strong torque, the device exhibited very rapid and very forceful alignment to the magnetic field. Phantom and experimental setup MRI-related heating was assessed for the programmable CSF shunt valve with a plastic phantom (length, 61 cm; width, 43 cm) filled to a depth of 10 cm with semisolid, gelled saline solution (0.8 g/l NaCl plus 5.85 g/l polyacrylic acid in distilled water) prepared to simulate human tissue according to American Society for Testing and Materials (ASTM) International designation F [12, 13, 18, 20]. The valve was placed in the phantom at a position midline on the left side, slightly (5 mm) below mid depth (vertical orientation) of the gelled saline solution to create a worst case for MRI-related heating of an implant [18], according to ASTM International designation F [20]. For the particular 3-T 128-MHz MRI system and experimental setup used in this investigation, the left side of the phantom was found to be associated with a greater temperature increase than the right side of the head-torso phantom for a given implant or device. Therefore, the programmable AJR:197, July

3 Shellock et al. CSF shunt was placed on the left side to produce a worst-case temperature increase for the described measurement conditions on the basis of results of a previous analysis of device heating for this particular MRI system (i.e., due to asymmetry in heating patterns for this phantom and the MRI system) [18]. Because this phantom and experimental setup lack blood flow, they simulate an extreme condition used to assess MRI-related heating for this device. Temperature recording system and placement of thermometry probes Temperature measurements were obtained with a fluoroptic thermometry system (model 3100, LumaSense Technologies). The fluoroptic thermometry probes (diameter, 0.5 mm) were positioned as follows on the programmable CSF shunt valve to record representative temperatures: probe 1, sensor portion of the probe placed in contact with one end of the device; probe 2, sensor portion of the probe placed in contact with opposite end of the device; probe 3, sensor portion of the probe placed in contact with middle portion of the device. The positions of the thermometric probes were inspected and verified immediately before and after the MRI-related heating assessment. MRI conditions MRI was performed at 3 T and 128 MHz with the aforementioned system. The body radiofrequency coil was used to transmit radiofrequency energy. MRI parameters were selected to generate a relatively high level of radiofrequency energy [18], producing an MRI system reported whole body averaged specific absorption rate of 2.9 W/kg for 15 minutes of image acquisition. The landmarking position (center position or anatomic region for the MRI procedure) and section locations were selected to encompass the entire area of the programmable CSF shunt valve. Experimental protocol The programmable CSF shunt valve was positioned in the plastic phantom by use of a grid and small plastic post setup, as previously described [12, 13, 18]. The fluoroptic thermometric system was calibrated, and the fluoroptic thermometric probes were applied. The phantom was filled with the gelled saline solution and allowed to equilibrate to environmental temperature for more than 24 hours. The MRI system fan was not on during the MRI-related heating assessment. The room and MRI system bore temperatures were at constant levels throughout the experimental session. After recording of baseline temperature (5 minutes), MRI was performed for 15 minutes and temperature was recorded at 10-second intervals. The highest temperature changes recorded with the fluoroptic thermometric probes were reported for the programmable CSF shunt valve. The background temperature in the phantom also was recorded. Thus the temperature change was recorded at the phantom in association with MRI-related heating of the gelled saline filled phantom without the device present. For recording of background temperature, a fluoroptic thermometric probe was placed in the phantom at a midline position on the left side slightly (5 mm) below the mid depth (vertical orientation) of the gelled saline solution [18, 20]. Artifacts MRI artifacts were evaluated at 3 T for the programmable CSF shunt valve by acquisition of MR images with the device attached to a plastic frame and then placed in a gadolinium-treated, saline solution filled plastic phantom [12, 13, 18]. MRI was performed with the aforementioned 3-T system, a transmit-receive radiofrequency head coil, and the following two sequences: T1-weighted spin-echo pulse sequence (TR/TE, 500/20; matrix size, ; section thickness, 10 mm; field of view, 28 cm 2 ; number of signals acquired, 2; bandwidth, 16 khz) and gradient-recalled echo pulse sequence (100/15; flip angle, 30 ; matrix size, ; section thickness, 10 mm; field of view, 28 cm 2 ; number of signals acquired, 2; bandwidth, 16 khz). The imaging planes were oriented to encompass the long and short axes of the programmable CSF shunt valve. The frequency-encoding direction was parallel to the plane of imaging. The image locations obtained through the programmable CSF shunt valve represented the largest, or worst-case, artifacts on the basis of review of multiple section locations in each imaging plane for this device, and these images were selected for evaluation. Planimetric software was used to measure (accuracy and resolution, ± 10%) the cross-sectional area of the largest artifact size for each pulse sequence and for each orientation of the section location. The image display settings (e.g., window and level, magnification) were carefully selected and used in a consistent manner for acquisition of valid measurements of the artifacts. This method has been used in many previous reports involving the characterization of artifacts for metallic implants [12, 13, 18]. Evaluation of Effects of 3-T MRI on Function Experiments were performed to assess the effects of the 3-T static magnetic field of the MRI system and various MRI conditions selected to be representative of typical techniques used for clinical MRI examinations to determine whether nine different samples of the programmable CSF shunt valve were damaged or whether there was a change in function [12, 13]. The different MRI exposures, conditions, and orientations of the devices were used to cover a range of possible scenarios. Conditions for 3-T static magnetic field exposure Nine samples of the valve were attached three per orthogonal orientation (axial, sagittal, and coronal) to a plastic copper sulfate filled phantom (length, 30 cm; width, 20 cm; height, 10 cm). The samples were carefully attached to the phantom, placed on the patient table of the MRI system, and inserted in (past isocenter and out the back of the unit to the farthest point) and out (approximately 0.5 m past the opening of the bore of the MRI system) of the unit 10 times [12, 13]. Immediately before and after each exposure, each programmable CSF shunt valve was subjected to functional testing for the ability both to indicate and to adjust the valve setting. Setup and conditions for MRI exposure Nine samples of the programmable CSF shunt valve were placed on the plastic copper sulfate filled phantom in a manner similar to that used for the 3-T static magnetic field exposure tests. MRI of the phantom with the nine valves was performed at 3 T with a transmit-receive body radiofrequency coil and eight different pulse sequences, running sequentially, for approximately 1 minute per pulse sequence with a 10-second delay between sequences [12, 13] (Table 1). The landmarking position (center position or anatomic region for the MRI procedure) and multiple section locations were selected to encompass all samples to ensure thorough exposure to these MRI conditions. After this exposure, each programmable CSF shunt valve was subjected to functional testing, which included confirming the ability to indicate the valve setting and the ability to adjust the valves to other settings. Results Magnetic Field Interactions The results for assessment of magnetic field interactions for the programmable CSF shunt valve indicated that the mean deflection angle was 16 and the torque was +1 (mild or low torque, the device slightly changed orientation but did not align to the magnetic field). The highest temperature changes recorded with the thermometric probes were 1.8 C (probe 1), 1.8 C (probe 2), and 1.6 C (probe 3) for the programmable CSF shunt valve. The highest background temperature change was 1.5 C. Artifacts Artifact test results are summarized in Table 2. The artifacts were predominantly losses of signal intensity larger than the size and shape of the programmable CSF shunt valve. The gradient-echo pulse sequence produced larger artifacts than the T1-weighted spin-echo pulse sequence. Figure 2 shows examples of artifacts for the gradient pulse sequence in long-axis and short-axis orientations relative to the device. 204 AJR:197, July 2011

4 TABLE 1: Sequences and Parameters Used to Expose Programmable CSF Shunt Valves to 3-T MRI Conditions Sequence Echo-Planar Imaging Parameter TR (ms) TE (ms) Flip angle NA NA NA NA 25 NA 25 NA Field of view (cm) Matrix size Section thickness (mm) Section gap (mm) Imaging plane Axial Axial Axial Axial Volume Volume Axial Axial Imaging time (min) Specific absorption rate, whole body (W/kg) GRE Magnetization Transfer Contrast T2-Weighted Fast Spin Echo 3D GRE 3D Fast GRE T1-Weighted Fast Spin Echo T2-Weighted Spin Echo T1-Weighted Spin Echo Note GRE = gradient-recalled echo, NA = not applicable. TABLE 2: Summary of Artifacts at 3-T MRI of Programmable CSF Shunt Valve Pulse Sequence Signal Void Size (mm 2 ) Imaging Plane T1-weighted spin echo 3568 Long axis T1-weighted spin echo 2367 Short axis Gradient-recalled echo 6487 Long axis Gradient-recalled echo 10,193 Short axis Effects of MRI on Function The settings of the nine programmable CSF shunt valves did not change, and the valves were not damaged in multiple exposures to the static magnetic field of the 3-T MRI system. With regard to exposure in the eight 3-T MRI conditions, the nine valves withstood appropriate adjustments, indicating that there was no damage and that the functional aspects of these devices were unaffected. No unintentional setting changes were found after any of the test scenarios relative to the 3-T MRI environment. Discussion Magnetic Field Interactions The average deflection angle was 16 at 3 T for the new programmable CSF shunt valve. This information should be considered in accordance with the following information provided by ASTM International [19], as follows: If the implant deflects less than 45, then the magnetically induced deflection force is less than the force on the implant due to gravity (its weight). Therefore, this valve passed the aforementioned acceptance criterion for a 3-T MRI system. The qualitatively measured torque was relatively low (+1). This valve therefore does not cause a hazard to a patient in a 3-T MRI environment with respect to rotational alignment to the static magnetic field. The magnetic field interactions (translational attraction and torque) found at 3 T for this programmable CSF shunt valve were comparable to levels reported for other magnetically activated valves [12, 13]. Of further note is that besides possible magnetic field problems, the intended in situ application of this valve may be taken into consideration [14]. Counterforces acting on this programmable CSF shunt valve that result from the use of sutures and attached inflow and outflow catheters associated with implantation in a patient will help prevent movement of this device during exposure to the MRI environment. In view of the findings on magnetic field interactions for this new valve, there are no safety issues for patients undergoing MRI on systems operating at 3 T or less. Furthermore, a patient with this programmable CSF shunt valve would be allowed to undergo an MRI procedure at 3 T or less immediately after implantation because of the relatively minor magnetic field interactions that are present. The highest temperature change recorded for the programmable CSF shunt valve during MRI performed at a relatively high MRI system reported whole body averaged specific absorption rate (2.9 W/kg) was 1.8 C. By comparison, the amount of heating that occurred at the same position in the phantom under the same MRI conditions without this implant was 1.5 C (i.e., the background temperature increase). This temperature level will not cause a hazard for a patient with this implant with the MRI parameters used in the heating test. In addition, because a static medium (i.e., no perfusion) was used for the evaluation, an acceptable margin of safety may be presumed, as has been the case for other programmable CSF shunt valves [12, 13]. Although it may be possible to use MRI parameters to reach a whole body averaged specific absorption rate of 4 W/ kg, this is an unlikely value to achieve in performance of a clinical MRI examination involving the area where this valve is implanted (i.e., brain and head). Substantial implant heating may be generated during MRI but only in an object made from conducting material of a certain length or in the shape of a closed loop with a relatively large diameter [14]. For the programmable CSF shunt valve used in this study, all metallic components are relatively small and encased in nonmetallic AJR:197, July

5 Shellock et al. A Fig. 2 Gradient-recalled echo 3-T MR images of programmable CSF shunt valve show artifacts. A, Section location oriented to long axis. B, Section location oriented to short axis. nonmagnetic material that effectively insulates and isolates them from the patient (Fig. 1). Therefore, as found in the evaluation of MRI-related heating, the temperature increase for this valve will not be substantially higher than the background temperature, even under extreme or worst-case experimental conditions at 3 T. Artifacts Artifacts associated with the programmable CSF shunt valve were relatively large in relation to the dimensions of the device. This finding is not surprising because the device incorporates two small neodymium magnets (Fig. 1). It is well known that implants that incorporate magnets produce large to very large artifacts on MR images, as has been reported for other magnetically activated CSF shunt valves and implants used for other applications [11 14, 16, 21, 22]. The extent of artifact for a magnetically activated shunt valve depends on the specific device, the magnetic susceptibility of the materials, and the imaging parameters used [10 14, 16]. If the anatomic area of interest is close to the programmable CSF shunt valve, as with other such devices [10 14, 16], the diagnostic use of MRI can be compromised, possibly concealing brain abnormalities on MR images [16]. This issue may influence the choice of valve implantation site and the type of valve selected for use [16]. Although it is possible to use imaging parameters to minimize the extent of the artifacts, the parameters have be chosen carefully to allow adequate signal-to-noise and contrast-to-noise ratios to obtain diagnostically useful MR images. The sizes of the MRI artifacts associated with the CSF shunt valve used in this study were consistent with those of similar devices containing permanent magnets [11 13, 16, 21, 22]. With regard to possibly reducing artifacts associated with such implants, use of fast spin-echo sequences, increasing the readout bandwidth, and decreasing the voxel size may be considered. Effects of MRI on Function A malfunctioning CSF shunt valve can cause serious complications due to overdrainage and elevated intracranial pressure [1, 2, 9]. It is vital to identify possible functional disturbances of magnetically programmable CSF shunt valves because the setting can be changed or the mechanism damaged as a result of exposure to a strong magnetic field, such as that used during MRI [3 15, 22 26]. Previous MRI-based investigations have documented both minor and substantial unintentional changes in valve settings related to exposure to magnetic fields [3 5, 7 9, 15]. There is evidence that even weaker magnets, including those found in toys, hair dryers, telephone speakers, and home use magnetic induction therapy, can alter the settings of programmable CSF shunt valves [23 25], causing some device manufacturers to recommend checking the valve setting on a regular basis [24, 25]. The findings of the tests designed to assess the effects of multiple exposures to the static magnetic field and to different MRI conditions at 3 T showed that there were no changes in the valve settings or damage to the valve. These results are consistent with the intent of this new magnetically adjustable shunt design, but the limitations of this study and the critical importance of the valve setting in hydrocephalic patients would suggest that in clinical use, confirming the unchanged status B of the setting after MRI examinations is prudent. As a precaution to ensure patient safety, the manufacturer of the programmable CSF shunt valve recommends checking the valve setting after the patient undergoes MRI. Conclusion The MRI findings on the new programmable CSF shunt valve indicate there are no significant concerns related to the 3-T MRI environment under the test conditions used for this evaluation, and as such, the device is not adversely affected by this powerful static magnetic field. The effect of higher-strength static magnetic fields on this device are unknown. The following guidelines are recommended [27, 28]. Nonclinical testing has shown that the Codman Certas Programmable Valve is MRI conditional in accordance with ASTM designation F [27]. A patient with this implant can safely undergo imaging immediately after implantation under the following conditions. Static Magnetic Field The static magnetic field must be 3 T or less. The highest spatial gradient magnetic field should be 720 gauss/cm or less. In nonclinical testing, the valve studied produced the highest temperature change of +1.8 C during MRI performed for 15 minutes of image acquisition (i.e., per pulse sequence) in the 3-T MRI system (128-MHz, Excite, HDx, software 14.M5, GE Healthcare). Therefore, the MRI-related heating experiments for the programmable valve at 3 T with a transmit-receive radiofrequency body coil at an MRI system reported whole body averaged specific absorption rate of 2.9 W/ kg (associated with a calorimetry-measured whole body averaged value of 2.7-W/kg) indicated that the greatest amount of heating that occurred in association with these specific conditions was +1.8 C or less. MRI Artifacts Artifacts were assessed with T1-weighted spin-echo and gradient-recalled echo pulse sequences. The details of these pulse sequences and the corresponding artifact sizes are shown in Table 2. MR image quality can be compromised if the area of interest is in the exact location of or relatively close to the location of the programmable valve. MRI artifacts can be minimized with careful selection of pulse sequence parameters. 206 AJR:197, July 2011

6 Specific Guidelines The valve setting must be verified immediately after MRI, according to the manufacturer s recommendations. 9. Watanabe A, Seguchi T, Hongo K. Overdrainage of cerebrospinal fluid caused by detachment of the pressure control cam in a programmable valve after 3-tesla magnetic resonance imaging. J Neurosurg displacement force on passive implants in the magnetic resonance environment. West Conshohocken, PA: ASTM International, American Society for Testing and Materials Inter- Acknowledgment We thank Sam Valencerina for assistance with this investigation. References 1. Vanneste JA. Diagnosis and management of normal-pressure hydrocephalus. J Neurol 2000; 247: Czosnyka Z, Czosnyka M, Richards H, Pickard JD. Hydrodynamic properties of hydrocephalus shunts. Acta Neurochir Suppl 1998; 71: Ortler M, Kostron H, Felber S. Transcutaneous pressure-adjustable valves and magnetic resonance imaging: an ex vivo examination of the Codman-Medos programmable valve and the Sophy adjustable pressure valve. Neurosurgery 1997; 40: Fransen P, Dooms G, Thauvoy C. Safety of the adjustable pressure ventricular valve in magnetic resonance imaging: problems and solutions. Neuroradiology 1992; 34: Fransen P. Transcutaneous pressure-adjustable valves and magnetic resonance imaging: an ex vivo examination of the Codman-Medos programmable valve and the Sophy adjustable pressure valve. Neurosurgery 1998; 42: Lüdemann W, Rosahl SK, Kaminsky J, Samii M. Reliability of a new adjustable shunt device without the need for readjustment following 3-Tesla MRI. Childs Nerv Syst 2005; 21: Lindner D, Preul C, Trantakis C, Moeller H, Meixensberger J. Effect of 3T MRI on the function of shunt valves: evaluation of Paedi GAV, Dual Switch and progav. Eur J Radiol 2005; 56: Inoue T, Kuzu Y, Ogasawara K, Ogawa A. Effect of 3-tesla magnetic resonance imaging on various pressure programmable shunt valves. J Neurosurg 2005; 103(suppl): ; 112: Akbar M, Aschoff A, Georgi JC, et al. Adjustable cerebrospinal fluid shunt valves in 3.0-Tesla MRI: a phantom study using explanted devices. Rofo 2010; 182: Lavinio A, Harding S, Van Der Boogaard F, et al. Magnetic field interactions in adjustable hydrocephalus shunts. J Neurosurg Pediatr 2008; 2: Shellock FG, Habibi R, Knebel J. Programmable CSF shunt valve: in vitro assessment of MRI safety at 3-Tesla. AJNR 2006; 27: Shellock FG, Wilson SF, Mauge CP. Magnetically programmable shunt valve: MRI at 3-Tesla. Magn Reson Imaging 2007; 25: Shellock FG. Reference manual for magnetic resonance safety, implants, and devices: 2011 edition. Los Angeles, CA: Biomedical Research Publishing Group, Miwa K, Kondo H, Sakai N. Pressure changes observed in Codman-Medos programmable valves following magnetic exposure and flipping. Childs Nerv Syst 2001; 17: Toma AK, Tarnaris A, Grieve JP, Watkins LD, Kitchen ND. Adjustable shunt valve-induced magnetic resonance imaging artifact: a comparative study. J Neurosurg 2010; 113: Lollis SS, Mamourian AC, Vaccaro TJ, Duhaime AC. Programmable CSF shunt valves: radiographic identification and interpretation. AJNR 2010; 31: Shellock FG, Valencerina S. In vitro evaluation of MR imaging issues at 3-T for aneurysm clips made from MP35N: findings and information applied to 155 additional aneurysm clips. AJNR 2010; 31: American Society for Testing and Materials International. Designation F : standard test method for measurement of magnetically induced national. Designation F : standard test method for measurement of radio frequency induced heating near passive implants during magnetic resonance imaging. West Conshohocken, PA: ASTM International, Majdani O, Rau TS, Götz F, et al. Artifacts caused by cochlear implants with non-removable magnets in 3T MRI: phantom and cadaveric studies. Eur Arch Otorhinolaryngol 2009; 266: Blankenstein FH, Truong B, Thomas A, Schröder RJ, Naumann M. Signal loss in magnetic resonance imaging caused by intraoral anchored dental magnetic materials. Rofo 2006; 178: Nakashima K, Oishi A, Itokawa H, Fujimoto M. Effect of magnetic fields from home-use magnetic induction therapy apparatuses on adjustable cerebrospinal fluid shunt valves. No Shinkei Geka 2010; 38: Anderson RC, Walker ML, Viner JM, et al. Adjustment and malfunction of a programmable valve after exposure to toy magnets. J Neurosurg 2004; 101: Schneider T, Knauff U, Nitsch J, et al. Electromagnetic field hazards involving adjustable shunt valves in hydrocephalus. J Neurosurg 2002; 96: Zemack G, Romner B. Seven years of clinical experience with the programmable Codman Hakim valve: a retrospective study of 583 patients. J Neurosurg 2000; 92: American Society for Testing and Materials (ASTM) International. Designation F : standard practice for marking medical devices and other items for safety in the magnetic resonance environment. West Conshohocken, PA: ASTM International, Shellock FG, Woods TO, Crues JV. MRI labeling information for implants and devices: explanation of terminology. Radiology 2009; 253:26 30 AJR:197, July