Technical Note. An evaluation of the precision of QIAGEN Quantiplex HYres kit as a screening tool on low level touch DNA evidence

Size: px
Start display at page:

Download "Technical Note. An evaluation of the precision of QIAGEN Quantiplex HYres kit as a screening tool on low level touch DNA evidence"

Transcription

1 Technical Note European Journal of Forensic Sciences DOI: /ejfs An evaluation of the precision of QIAGEN Quantiplex HYres kit as a screening tool on low level touch DNA evidence Stephen Chi-Yuen Ip, Kam-Ming Lai Forensic Science Division, Government Laboratory, Homantin Government Offices, Kowloon, Hong Kong, China Address for correspondence: Stephen Chi-Yuen Ip, Forensic Science Division, Government Laboratory, Homantin Government Offices, Kowloon, Hong Kong, China. kmlai@ govtlab.gov.hk Received: July 15, 2015 Accepted: October 07, 2015 Published: November 02, 2015 ABSTRACT Objective: The primary aim of DNA quantification in forensic DNA analysis is to determine and optimize the amount of DNA for the downstream profiling analysis. In addition, it may be used as a parameter for the cut-off of the subsequent profiling, especially for touch DNA samples collected from property crime scenes, which often contain low amount of DNA and yield no profiles. The choice of quantification kit should have both high accuracy and precision. This work aims to examine the latter of five selected commonly used real-time quantitative polymerase chain reaction (qpcr) kits, namely Applied Biosystems Quantifiler kit, Quantifiler Y kit, Quantifiler Duo kit, Promega Plexor HY system, and QIAGEN Quantiplex HYres kit, and to seek for a qpcr system fitted for the purpose of initial screening of touched DNA evidence for profiling. Materials and Methods: In this work, we first tested the precision of the five selected qpcr kits using their respective DNA standard and human blood samples as substrates. Subsequently, we compared the capability of Quantiplex in predicting the presence of DNA with that of Quantifiler when applying a cut-off threshold of ng/μl and ng/μl, i.e. their respective lowest detection limit with reference to DNA typing analysis. Results: Our results demonstrated that Quantiplex exhibits a higher precision and consistency than the others, especially at low DNA concentration, i.e ng/μl. We also showed that, using a total of 332 simulated touch DNA samples, Quantiplex is a fairly reliable predictor of the AmpFlSTR Identifiler profiling results. Conclusion: Quantiplex can be employed as a cost-effective screening of samples of volume crime cases. KEY WORDS: Forensic sciences, forensic genetics, DNA quantification, DNA typing, Plexor, Quantifiler, Quantiplex INTRODUCTION Forensic DNA analysis of crime scene samples usually involves five steps, including DNA extraction, quantification of extracted DNA, polymerase chain reaction (PCR) amplification of short tandem repeats (STR), capillary electrophoresis of amplified PCR products, and data interpretation; among which DNA quantification is essential in the overall DNA profiling process as it allows both an assessment of DNA extraction efficiency and a determination of how much DNA template to be added into the downstream analysis [1]. Typical optimum range of total input DNA lies between 0.5 and 2 ng for most of the commercially available STR amplification kits. Too much DNA template in a PCR reaction results in overblown signals in the electropherogram leading to shift in molecular size of detected signal as well as artifacts, whereas too little DNA leads to allelic drop-out and loss of information owing to stochastic effect, which happens due to random sampling of template DNA during the first few PCR cycles. These unfavorable conditions could generate unreliable results which will significantly interfere the data interpretation and impede the overall efficacy of DNA analysis. Hence, quantification of extracted DNA is indispensable to ensure PCR amplifications performed under an efficient and optimized condition. Until now, a number of DNA quantification methods, including ultraviolet absorbance, slot blot, PicoGreen, and real-time quantitative PCR (qpcr), have been employed in the forensic field to estimate the concentration of the input DNA [1-4]. qpcr is the preferred technique among these methods for forensic samples collected from scenes, as it is highly sensitive, automatable, and most importantly in human-specific. It also offers a broad dynamic range of quantification. qpcr not only identifies the presence of PCR inhibitor(s) in the samples which will affect STR amplification, but also provides essential information on whether sufficient amount of human DNA can be recovered for subsequent downstream analysis. Although qpcr analysis allows a higher sensitivity over the other quantification methods, it is nevertheless strongly affected by Eur J Forensic Sci Apr-Jun 2016 Vol 3 Issue 2 61

2 the stochastic effect that is very common in most PCR-based methods. of low level DNA samples often yield false negative quantification results [5,6], i.e. samples with quantification values below the lowest standard but produce informative STR profiles. Hence, this prompts the needs to look for a DNA quantification kit with both high precision and reliability, allowing forensic scientists to have a more accurate prediction on the STR amplification results and to formulate a cut-off threshold from the quantification results to decide which samples will be proceeded for further testing. In this study, we have explored the precision of several commercially available qpcr kits commonly used in forensics, including Applied Biosystems Quantifiler (refer to as Quantifiler ) [5,7-8], Quantifiler Y (refer to as Quantifiler Y ) [7], Quantifiler Duo Human DNA kits (refer to as Quantifiler Duo ) [9,10], Promega Plexor HY System (refer to as Plexor ) [11], and QIAGEN Investigator Quantiplex HYres Kit (refer to as Quantiplex ) [12]. There are several differences between these qpcr quantification kits [Table 1]. First, the chemistry for detection is different. The Quantifiler series and Quantiplex utilize Taqman assay and Scorpion primerbased assay, respectively, whereas Plexor employs a primer quenching assay. This means that the fluorescence signal from the TaqMan probes or Scorpion primers increases as PCR proceeds, whereas signal from the Plexor primers drops as a result of quenching. Because Plexor measures the accumulation of product by a decrease in fluorescence signal, an extra tailor-made program by the manufacturer is required for the analysis of preliminary data collected, and hence lengthening the total processing time. Second, the size of their target regions and copy numbers in the human genome are different. The Quantifiler series target single copy genes in the genome, whereas both Plexor and Quantiplex target multi-copy genes. Quantifiler and Quantifiler Y amplify a 62 bp and a 64 bp region on chromosome 5 and the Y chromosome, respectively [13]. Since the targets for these two kits are smaller than those of the STR amplification kits in the subsequent step, i.e. usually between 100 and 450 bp, degraded samples with positive quantification result does not necessarily guarantee the generation of DNA profile. To tackle this problem, the target size for Quantifiler Duo, Plexor, and Quantiplex are all increased to the range similar to the STR Table 1: Details of the five real-time quantitative PCR kits tested in this study kit Detection Target copy no. Target size (Chr.) Reaction time Quantifier Taqman probes Single 62 bp (5) ~110 min Quantifiler Y Taqman probes Single 64 bp (Y) ~110 min Quantifier Duo Taqman pr obes Single 140 bp (14) 130 bp (Y) ~90 min Plexor Primer quenching Multi 99 bp (17) 133 bp (Y) ~60 min Quantiplex Scorpion primers Multi 146 bp (several) ~60 min 129 bp (Y) Chr. indicates the chromosome where the target is located. PCR: Polymerase chain reaction kits. Quantifiler Duo amplifies a 140 bp and a 130 bp target on chromosome 14 and the Y chromosome, respectively [14], whereas Plexor amplifies a 99 bp and a 133 bp multi-copy target on chromosome 17 and the Y chromosome, respectively [15]. Quantiplex, on the other hand, targets a proprietary region that is a 146 bp that are present in several autosomal regions of the human genome, and a 129 bp multi-copy fragment on the Y chromosome. Hence, with these adjustments, these kits are claimed to offer a greater sensitivity and a higher reliability within different individuals and populations [16]. In addition, Quantifiler Duo [9,17], Plexor [11], and Quantiplex [12,17] that are capable of quantifying autosomal and male DNA simultaneously, all offer further information on the relative quantities of human male to female DNA in a sample that will facilitate the selection and implementation of appropriate downstream methods for genotyping analysis. The first objective of this study was to compare the precision and consistency of the five selected qpcr kits, viz., Quantifiler, Quantifiler Y, Quantifiler Duo, Plexor, and Quantiplex. In addition, it was aimed to seek for a qpcr system fitted for the purpose of initial screening of touched DNA evidence for profiling. MATERIALS AND METHODS Preparation of DNA Standard Samples Series of DNA standards of each quantification kits were prepared freshly prior to the experiment according to the manufacturer s recommendations to avoid any deterioration of the quality of standards in the course of the experiments. The precision of a real-time qpcr method was examined by performing three runs on different days using the standard concentrations (refer to as Standard ; in ng/μl) in the range of , , and ng/ μl for autosomal DNA [Table 2], and , , , and ng/μl for male DNA [Table 3]. Each run contained 10 replicates of the selected standard concentrations (n = 30). Preparation of Whole Blood DNA Sample Whole blood samples were collected from two female and two male volunteers, respectively, with consent. The extraction of DNA from the whole blood samples were performed using the QIAamp DNA Blood Mini Kit (QIAGEN, Hilden, Germany) [18] following the manufacturer s recommendations. DNA was eluted in approximately 200 μl of Buffer AE. The extracted whole blood DNA samples were diluted to approximately 0.5 and ng/μl with Trisethylenediaminetetraacetic acid buffer as estimated with Quantifiler. These samples, which were defined as High and Low range, respectively, were stored at 20 C before analysis. The precision of a real-time qpcr method was examined by performing three runs on different days on the whole blood DNA. Each run contained 10 replicates of samples (n = 30). 62 Eur J Forensic Sci Apr-Jun 2016 Vol 3 Issue 2

3 Table 2: The relationship between the cycle threshold (C T ) and its standard deviation with the concentration of the DNA template when autosomal DNA was targeted (n=30) Standard concentration (ng/μl) kit Quantifiler (±1.2) (±0.6) (±1.9) NA NA NA Quantifier Duo (±1.3) (±1.0) (±2.8) NA NA NA Plexor (±2.2) (±2.1) (±3.0) (±3.2) Quantiplex (±0.8) (±0.8) (±0.9) * 33.2 (±3.4) NA indicates DNA quantification is not applicable for Quantifiler and Quantifiler Duo in the range of ng/μl. *Indicates the detection of samples with undetermined C T value. SD: Standard deviation Table 3: The relationship between the cycle threshold (C T ) and its standard deviation with the concentration of the DNA template when male DNA was targeted (n=30) Standard concentration (ng/μl) kit Quantifier Y (±1.1) (±1.0) (±3.1) NA NA NA Quantifiler Duo (±1.3) (±1.3) * 36.0 (±2.4) NA NA NA Plexor (±0.9) (±1.6) (±2.1) (±2.0) Quantiplex (±0.6) (±0.7) (±1.1) (±2.0) NA indicates DNA quantification is not applicable for Quantifiler Y and Quantifiler Duo in the range of ng/μl. *Indicates the detection of samples with undetermined C T value. SD: Standard deviation Preparation of Simulated Touch DNA Samples and STR DNA Typing Analysis A total of 332 simulated touch DNA samples, which were prepared as described previously [19], were used in this study. The extraction of DNA from these touch samples were performed with Chelex 100, Promega DNA IQ TM system, QIAGEN QIAamp DNA Investigator Kit, and QIAGEN QIAsymphony DNA Investigator Kit, respectively, as described [19]. Extracts were then amplified with AmpFlSTR Identifiler PCR Amplification Kit on a Applied Biosystems GeneAmp PCR System 9700 (Life Technologies TM, Carlsbad, CA, USA) and analyzed on the ABI PRISM 3100 Genetic Analyzer (Life Technologies TM, Carlsbad, CA, USA), and Genetic Analyzer, as described [19]. Real-time qpcr Methods and Statistical Analysis DNA quantification of DNA standard and whole blood samples were performed using the Applied Biosystems Quantifiler, Quantifiler Y, Quantifiler Duo Human DNA Kits (Life Technologies TM, Carlsbad, CA, USA), Promega Plexor HY System (Promega, Madison, WI, USA), and QIAGEN Investigator Quantiplex HYres Kit (QIAGEN, Hilden, Germany) in the ABI PRISM 7500 Sequence Detection System (Life Technologies TM, Carlsbad, CA, USA) according to the manufacturer s recommendations. DNA quantification of simulated touch DNA samples were performed using the Applied Biosystems Quantifiler Kit and QIAGEN Investigator Quantiplex HYres Kit in the ABI PRISM 7500 Sequence Detection System. The collected data were analyzed by Applied Biosystems SDS Software v1.0 (Life Technologies TM, Carlsbad, CA, USA). For the data obtained from Plexor, data analysis was accomplished by exporting the collected data to the Plexor Analysis Software. The variation in the concentration for the DNA standards and the whole blood DNA samples were obtained from various quantification methods, and the variation in the average number of loci obtained from the STR analysis were represented as standard deviation (±SD). RESULTS Quantiplex Demonstrates a Higher Precision than Quantifiler, Quantifiler Y, Quantifiler Duo, and Plexor for the of both Autosomal and Male DNA Similar to most PCR-based analysis, real-time PCR quantification is subjected to the stochastic or random sampling effects [20]. When a limited number of DNA target molecules are present in the amplification reaction, the PCR primers in excess will not be able to target the specific DNA regions for quantification, and therefore affecting the precision of the quantification results. In an attempt to compare the precision of the selected quantification kits, we first used their respective DNA standards supplied in the kit and diluted them to the manufacturer s recommended concentration in the following ranges: , , and ng/μl for autosomal DNA [Table 2], and , , and ng/μl for male DNA [Table 3]. Regarding the nature of these DNA standards, we believed the results could provide us a clear picture on the precision of each individual quantification kit. The DNA standards for Plexor and Quantiplex were further diluted to and ng/μl for autosomal Eur J Forensic Sci Apr-Jun 2016 Vol 3 Issue 2 63

4 and male DNA, respectively, according to manufacturer s recommendations. The diluted samples prepared were then used for DNA quantification, and the cycle threshold, C T values were recorded. Our results demonstrated that C T deviation obtained from Quantiplex was kept below 0.9% for , , and ng/μl [Table 2]. For the other three kits, the deviations were not varied that much between and ng/μl, and these were kept within %, %, and % for Quantifiler, Quantifiler Duo, and Plexor, respectively, indicating these kits offer a fair consistency in these ranges and were not affected much by stochastic amplification. However, the deviation obtained from these three kits reached a maximum of 1.9%, 2.8%, and 3.0%, respectively, as the DNA concentration dropped to ng/μl, indicating that these kits were subjected to stochastic effect at this range. The above data shows that Quantiplex offered the highest precision of quantification for autosomal DNA with concentration down to ng/μl. We further tested the precisions of both Plexor and Quantiplex at the range of ng/μl, i.e. their lowest quantification standards. Our results showed that both Plexor and Quantiplex were affected by stochastic amplification at such low concentration and offered comparable deviations of 3.2% and 3.4%, respectively. When male DNA was targeted, we observed a similar trend of variation in precision from Quantifiler Y, Quantifiler Duo, Plexor, and Quantiplex [Table 3]. Again, Quantiplex offered the highest precision among these kits for the quantification of male DNA, and its deviation was kept below 1.1% even when ng/μl of male DNA was tested as compared to 3.1%, 2.4%, and 2.1% for Quantifiler Y, Quantifiler Duo, and Plexor, respectively. To show the variation of the actual quantification values obtained from these kits, the C T values obtained were converted to concentration by comparing them to their standard curves. The standard curves prepared from each kit were consistent with their respective QC parameters (data not shown). For autosomal DNA, Quantiplex offered a deviation of 9% for both and ng/μl, i.e. the lowest deviation observed among the four tested kits [Figure 1]. On the other hand, Quantifiler, Quantifiler Duo, and Plexor all offered relatively higher deviations of 10-15%, 17-25%, and 15-36%, respectively. At ng/μl, Quantiplex demonstrated the lowest deviation of 18%, whereas Quantifiler, Quantifiler Duo, and Plexor yielded significantly higher deviations of 41%, 43%, and 54%, respectively. At ng/μl, Quantiplex deviation increased to 28%, but it was still significantly lower than that from Plexor, which was jumped up to 52%. For the performance of the quantification of male DNA, Quantiplex and Quantifiler Y offered a deviation of 8% only when and ng/μl of male DNA was targeted as compared to 17-31% and 14-32% for Quantifiler Duo and Plexor, respectively [Figure 2]. At ng/μl, Quantiplex continued to offer the highest precision, with a deviation of 19%, as compared to 56%, 62%, and 47% by Quantifiler Y, Quantifiler Duo, and Plexor, respectively. Collectively, our data showed that Quantiplex exhibits the highest precision for the quantification of both autosomal and male DNA. Blood Matrix has Little or No Effect on the Precision of Quantiplex We further compared the precision of these kits on DNA extracted from the whole blood samples, as it is one of the most commonly encountered biological evidence in serious crime and has a relatively complex matrix as compared to the DNA standards. Quantiplex yielded a deviation of 12% and 22% obtained from the high (0.5 ng/μl) and low (0.025 ng/μl) of autosomal DNA, respectively [Figure 3a]. In contrast, Quantifiler, Quantifiler Duo, and Plexor exhibited a lower precision ( high/low ) with deviations of 30/40%, 26/50%, and 19/47%, respectively. Similar performance were obtained when male DNA was targeted, where Quantifiler Y, Quantifiler Duo, Plexor, and Quantiplex exhibited deviations of 17/52%, 25/54%, 13/21%, and 7/17%, respectively [Figure 3b]. These data indicate that Quantiplex exhibits the highest precision not only for the clean DNA standard samples, but also for the whole blood DNA extracts. Quantiplex is a Better Predictor for the Typing Results of Touch DNA Samples as Compared to Quantifiler We have demonstrated from above that Quantiplex shows the highest precision among these kits for both quantifications of autosomal and male DNA. Here, we further explored the effectiveness of Quantiplex in analyzing DNA obtained from touch evidence in which these samples contribute to the majority of the property crime cases. A total of 332 simulated touch DNA samples were collected as described previously [19]. We compared the capability of Quantiplex in predicting the presence of DNA with that of Quantifiler when applying a cut-off threshold of ng/μl and ng/μl, i.e. their respective lowest detection limit with reference to DNA typing analysis. We also examined if a true zero value of quantification as STR cut-off could be established. by Quantifiler revealed 189 positive and 143 negative results from the 332 touch DNA samples (using a cut-off threshold of ng/μl) [Figure 4a]. 175 and 7 samples out of the 189 positive samples generated typing results with 11 to 15 and 6 to 10 loci, respectively. The remaining 7 samples yielded profiles with <6 loci, and these were referred as poor positive samples, i.e. samples with a concentration above the cut-off threshold yielded profiles with <6 loci. Among the 143 negative samples, only 70 samples i.e. 22 and 48 samples yielded profiles with <6 loci were true negative samples, which contribute to around 49% within the negative samples. The remaining 46 and 27 samples yielded DNA profiles with 6 or more loci were indeed false negative samples, which contributed to 51% of the negative samples. In contrast, Quantiplex yielded 282 positive and 50 negative results (using a cut-off threshold of ng/μl) [Figure 4b]. Among these 282 positive samples, 221 and 30 samples yielded profiles with and 6-10 loci, respectively, whereas the remaining 31 samples generated profiles with <6 loci. Among the 50 negative samples, 46 samples (19 and 27 samples with <6 loci) were true negative samples (~92%), whereas only 4 samples were false 64 Eur J Forensic Sci Apr-Jun 2016 Vol 3 Issue 2

5 Figure 1: Comparison of the precision of quantifi cation kits when autosomal DNA was targeted in their respective DNA standard samples (n = 30). Numerical values (in %) indicated the standard deviations obtained. The rectangular box shows median, upper, and lower quarters of the quantifi cation values. The error bar shows the maximum and minimum quantifi cation values in the particular quantifi cation method. *represents the lowest deviation obtained for that specifi c range of concentration Figure 2: Comparison of the precision of quantification kits when male DNA was targeted in their respective DNA standard samples (n = 30). Numerical values (in %) indicated the standard deviations obtained. The rectangular box shows median, upper, and lower quarters of the quantifi cation values. The error bar shows the maximum and minimum quantifi cation values in the particular quantifi cation method. *represents the lowest deviation obtained for that specifi c range of concentration negative (~8%). In summary, these findings indicate that Quantiplex (true negative: 92%) is a better predictor of DNA typing results than Quantifiler (true negative: ~49%), even though at the cost of yielding nearly three times (11% instead of 4%) more poor positive samples. To further evaluate the sensitivity of Quantiplex, the relationship between the DNA concentration of the tested samples and the number of loci obtained from Identifiler were examined [Table 4]. Our results showed that the average number of loci obtained increased with the quantification values for both Quantifiler and Quantiplex. The key difference between them lies in the effective quantification ranges they offered and the cut-off thresholds we applied. Importantly, Quantiplex with the cut-off threshold set at ng/μl rescued more samples with positive typing results, i.e. a total of 282 samples yielded positive results from at least an average of 9.9 loci and missed only 50 samples (~15%) with an average of 2.0 ± 2.4 loci with limited forensic value. On the other hand, Quantifiler missed a total of 143 samples (~43%) with positive typing results in an average of 6.6 ± 5.1 loci. All these data indicated that Quantiplex is a better predictor of DNA typing results, and hence allowing a better judgment on the subsequent processing of samples. Eur J Forensic Sci Apr-Jun 2016 Vol 3 Issue 2 65

6 a b Table 4: The relationship of the average number of loci with the quantification value obtained by (a) Quantifiler or (b) Quantiplex for the 332 simulated touch DNA samples value (ng/μl) Number of samples Average number of loci with results (±SD) Quantifier < (±5.1) (False ve: 73) (±3.2) (±2.5) Quantiplex < (±2.4) (False +ve: 4) (±4.4) (±2.8) (±2.0) The standard deviation is shown as ±SD. Bold indicates the cut-off threshold of Quantifiler (0.023 ng/μl) and Quantiplex ( ng/μl), respectively. False -ve indicates samples with concentration below the cut-off threshold yielded profiles with 6 or more loci Figure 3: Comparison of the precision of the quantifi cation kits when (a) Autosomal DNA, or (b) male DNA, was targeted in a whole blood sample (n = 30). Numerical values (in %) shows the standard deviations obtained. The rectangular box shows median, upper, and lower quarters of the quantifi cation values. The error bar shows the maximum and minimum quantifi cation values in the particular quantifi cation method. *represents the lowest deviation obtained for that specifi c range of concentration a b Figure 4: Distribution of DNA profi ling results for the 332 simulated touch DNA samples when quantified by, (a) Quantifiler, and (b) Quantiplex with cut-off threshold at ng/μl and ng/μl, respectively. Poor +ve indicates samples with concentration above the cut-off threshold yielded profi les with <6 loci. True ve and false ve indicate samples with concentration below the cut-off threshold yielded profi les with <6 loci and profi les with 6 or more loci, respectively DISCUSSION In the past decade, a number of studies have been conducted on the performance of these qpcr kits. It has been reported that the single copy Quantifiler Duo approach provides better accuracy, whereas the multi-copy Plexor approach offers a higher sensitivity [21], and the authors suggested the selection of kits depends upon the objective of the user. Thomas et al. reported in another study, using a tested half-volume reaction on buccal swabs, that Quantiplex may be useful for predicting STR success of which 66% of the undetected samples yielded no profile [12]. The findings were comparable with previously reported data on Quantifiler of which 73% of the undetected buccal swab samples yielded no profile [5]. Recently, Fregeau and Laurin reported that Quantiplex offers an enhanced performance over the Quantifiler Duo, a newer version of Quantifiler in terms of sensitivity, accuracy, consistency, reliability, and robustness [17]. In this work, in an attempt to obtain the precision of these qpcr kits, we made use of their respective DNA standard as quantification substrates and took advantage of their clean nature, instead of using buccal swabs as reported in other studies [5,12]. Our results demonstrated that Quantiplex displayed a higher precision of quantitation not only for its clean DNA standards, but also for DNA obtained from the complex blood extracts as compared to the other four kits. It is noteworthy that at the range of ng/μl autosomal DNA, Quantiplex offered a precision of ±18%, which is approximately 2-3 times better than those offered by the Quantifiler series (±41-43%) and Plexor (±54%), respectively. This high consistency results generated by Quantiplex, hence, can be used as an effective screening tool for subsequent profiling. Of note, Quantiplex is indeed affected by the stochastic amplification at a very low DNA concentration, i.e. ~10-fold lower as compared to the others. This observation highlights the fact that even quantification kits targeting multi-copy loci are subjected to a certain degree of variability, especially when the target DNA is close to its lower end of quantification. Using a total of 332 simulated touch samples, we were unable to establish a zero cut-off threshold using Quantiplex; however, it improves the consistency and sensitivity of DNA quantification, allowing a better judgment on the subsequent processing of samples. Most of the samples (92%) considered as negative using Quantiplex yielded profiles with limited forensic value. In contrast, only half of them (49%) were true negative using Quantifiler. Considering that Quantifiler (i.e. 62 bp) targets a relatively smaller amplicon than that of the downstream Identifiler STR amplification kit (i.e bp), we expected that Quantifiler will have yielded a greater number 66 Eur J Forensic Sci Apr-Jun 2016 Vol 3 Issue 2

7 of positive quantification results with poor quality than that of the Quantiplex, which produces a similar amplicon size as compared to the Identifiler (i.e. 146 bp). However, this was not the case as we observed only 7 (~4% of total positive samples) positive samples with poor quality from Quantifiler as compared to 31 (~11% of total positive samples) from that of the Quantiplex. Importantly, Quantiplex also yielded less false negative quantification results than that of the Quantifiler i.e. 4 (~8%) instead of 73 (~51%). This observation could be reconciled by the lower detection threshold limit in Quantiplex, which recovers more samples with low DNA concentration for amplification. To this, one may consider to lower the detection threshold limit of Quantifiler to the range of ng/μl as compared to the Quantiplex. However, our work [Figure 1 and Table 2] demonstrates that in such condition, Quantifiler generates a greater deviation of at least 41% as compared to 28% from Quantiplex. This relatively high inconsistency could affect the effectiveness of Quantifiler as a screening tool. Our study also shows that quantification of the same set of samples by these kits yield different values (data not shown), and such variation could be due to the differences of the copy number of target amplicons and the concentration of their DNA standards [21]. Hence, we strongly recommend that performing a complete validation for both DNA quantification and STR amplification when there is any upgrade of either system, as a new optimal range of input DNA has to be determined for the profiling. In addition, to make use of the DNA quantification kit as a cut-off for the subsequent amplification, it is recommended to choose one that offers a higher sensitivity and reproducibility than the amplification kit; however, choosing one with lower false positive rate may still be an option for those with resource limitation. ACKNOWLEDGMENTS We thank Dr. C. M. Lau, the Former Government Chemist of the Government Laboratory, Dr. F. C. Kwok, the Assistant Government Chemist, and Mr. B. K. K. Cheung, the Chief Chemist for their support and the approval of the publication of this work. We thank Mr. W. N. Cheng for his outstanding technical support for this work. We also thank other members of our laboratory for their inputs and discussions. REFERENCES 1. Lee SB, McCord B, Buel E. Advances in forensic DNA quantification: A review. Electrophoresis 2014;35: Niels en K, Mogensen HS, Hedman J, Niederstätter H, Parson W, Morling N. Comparison of five DNA quantification methods. Forensic Sci Int Genet 2008; Nickla s JA, Buel E. of DNA in forensic samples. Anal Bioanal Chem 2003;376: Alonso A, Martin P, Albarrán C, García P, Primorac D, García O, et al. Specific quantification of human genomes from low copy number DNA samples in forensic and ancient DNA studies. Croat Med J 2003;44: Cupples C M, Champagne JR, Lewis KE, Cruz TD. STR profiles from DNA samples with undetected or low quantifiler results. J Forensic Sci 2009;54: Ip SC, Li n SW, Li C, Lai KM. Forensic DNA typing strategy of degraded DNA on discarded cigarette ends using the AmpFlSTR Identifiler, Identifiler Plus and MiniFiler PCR amplification kits. Sci Justice 2014;54: Green RL, Ro inestad IC, Boland C, Hennessy LK. Developmental validation of the quantifiler real-time PCR kits for the quantification of human nuclear DNA samples. J Forensic Sci 2005;50: Seo SB, Lee HY, Zhang AH, Kim HY, Shin DH, Lee SD. Effects of humic acid on DNA quantification with Quantifiler Human DNA kit and short tandem repeat amplification efficiency. Int J Legal Med 2012;126: Barbisin M, F ang R, O Shea CE, Calandro LM, Furtado MR, Shewale JG. Developmental validation of the Quantifiler Duo DNA kit for simultaneous quantification of total human and human male DNA and detection of PCR inhibitors in biological samples. J Forensic Sci 2009;54: Grgicak CM, Urban ZM, Cotton RW. Investigation of reproducibility and error associated with qpcr methods using Quantifiler Duo DNA quantification kit. J Forensic Sci 2010;55: Krenke BE, Na ssif N, Sprecher CJ, Knox C, Schwandt M, Storts DR. Developmental validation of a real-time PCR assay for the simultaneous quantification of total human and male DNA. Forensic Sci Int Genet 2008;3: Thomas JT, Be rlin RM, Barker JM, Dawson Cruz T. Qiagen s investigator quantiplex Kit as a predictor of STR amplification success from low-yield DNA samples. J Forensic Sci 2013;58: Applied Biosystems. Quantifi ler Human and Y Human Male DNA Kits User Guide. Publication Number : Revision G; Applied Biosystems. Quantifil er Duo DNA Kit User s Manual. Publication Number : Revision E; Promega. Technical Manual: Plexor HY System for the Applied Biosystems 7500 and 7500 FAST Real-Time PCR Systems. TM293: 2/12; QIAGEN. Investigato r Quantiplex HYres Handbook Frégeau CJ, Laurin N. The Qiagen Investigator Quantiplex HYres as an alternative kit for DNA quantification. Forensic Sci Int Genet 2015;16: QIAGEN. QIAamp DNA Min i and Blood Mini Handbook. 3 rd ed Ip SC, Lin SW, Lai KM. An evaluation of the performance of five extraction methods: Chelex 100, QIAamp DNA Blood Mini Kit, QIAamp DNA Investigator Kit, QIAsymphony DNA Investigator Kit and DNA IQ. Sci Justice 2015;55: Timken MD, Klein SB, Buonc ristiani MR. Stochastic sampling effects in STR typing: Implications for analysis and interpretation. Forensic Sci Int Genet 2014;11: LaSalle HE, Duncan G, McCord B. An analysis of single and multicopy methods for DNA quantitation by real-time polymerase chain reaction. Forensic Sci Int Genet 2011;5: Source of Support: Nil, Conflict of Interest: None declared. Eur J Forensic Sci Apr-Jun 2016 Vol 3 Issue 2 67

Validation of a Multiplexed System for Quantification of Human DNA and Human Male DNA and Detection of PCR Inhibitors in Biological Samples

Validation of a Multiplexed System for Quantification of Human DNA and Human Male DNA and Detection of PCR Inhibitors in Biological Samples Validation of a Multiplexed System for Quantification of Human DNA and Human Male DNA and Detection of PCR Inhibitors in Biological Samples Maura Barbisin*, Rixun Fang, Cristin E. O Shea, Pius M. Brzoska,

More information

Quantifiler Human DNA Quantification Kit Quantifiler Y Human Male DNA Quantification Kit

Quantifiler Human DNA Quantification Kit Quantifiler Y Human Male DNA Quantification Kit Product Bulletin Human Identification Quantifiler Human DNA Quantification Kit Quantifiler Y Human Male DNA Quantification Kit The Quantifiler kits produce reliable and reproducible results, helping to

More information

Forensic News. FAS Corner April The Maximizing Data Quality Series Part 2: DNA Quantitation

Forensic News. FAS Corner April The Maximizing Data Quality Series Part 2: DNA Quantitation The Maximizing Data Quality Series Part 2: DNA Quantitation (Editor's Note: This is part of an ongoing series of articles focusing on maximizing data quality throughout the HID workflow. Part 1 on DNA

More information

Investigator Quantiplex HYres Kit

Investigator Quantiplex HYres Kit These specifications are for the procurement of a validation study at GBI Division of Forensic Sciences facilities of two commercially available kits used in forensic DNA analysis. The two kits to be validated

More information

Reliable extraction of DNA from Whatman FTA cards

Reliable extraction of DNA from Whatman FTA cards Sample collection Reliable extraction of DNA from Whatman FTA cards This study examined the yield and quality of DNA from samples applied to Whatman FTA cards, using five common methods of DNA extraction.

More information

PowerPlex. Y System Validation

PowerPlex. Y System Validation PowerPlex Y System Validation By Patricia M. Fulmer, Dawn Rabbach, Kimberly Huston, Curtis Knox and Cynthia Sprecher, Promega Corporation Abstract We have improved the manufacturing process for the PowerPlex

More information

User Bulletin. Veriti 96-Well Thermal Cycler AmpFlSTR Kit Validation. Overview

User Bulletin. Veriti 96-Well Thermal Cycler AmpFlSTR Kit Validation. Overview User Bulletin Veriti 96-Well Thermal Cycler AmpFlSTR Kit Validation June 2009 SUBJECT: AmpFlSTR PCR Amplification Kit validation on the Veriti 96-Well Thermal Cycler with 0.2 ml sample block format In

More information

Naming Convention and Laboratory Methods: PROVEDIt Database

Naming Convention and Laboratory Methods: PROVEDIt Database Naming Convention and Laborary Methods: PROVEDIt Database Sec 1. Naming Convention and Laborary Methods Single source sample names will follow the format below and are best explained through example: RD14-0003-21d1x-0.5IP-Q0.8_002.20sec

More information

DNA Typing Strategy for the Identification of Old Skeletal Remains

DNA Typing Strategy for the Identification of Old Skeletal Remains DNA Typing Strategy for the Identification of Old Skeletal Remains Hwan Young Lee, Ph.D. Dept. of Forensic Medicine Yonsei University College of Medicine Presentation Overview DNA Extraction Complete demineralization

More information

QIAGEN Validation Report

QIAGEN Validation Report QIAGEN Validation Report Developmental validation of the Investigator HDplex Kit The QIAGEN Investigator HDplex Kit is used for multiplex PCR in human identification and forensic casework. It was developed

More information

SIMULTANEOUS PURIFICATION AND QUANTITATION OF DNA FROM DATABASE BLOOD AND BUCCAL SWAB SAMPLES

SIMULTANEOUS PURIFICATION AND QUANTITATION OF DNA FROM DATABASE BLOOD AND BUCCAL SWAB SAMPLES SIMULTANEOUS PURIFICATION AND QUANTITATION OF DNA FROM DATABASE BLOOD AND BUCCAL SWAB SAMPLES Allan Tereba, Paraj V. Mandrekar, Susan Koller and Rex Bitner Promega Corporation, 2800 Woods Hollow Road,

More information

Internal Validation of the Promega PowerPlex Fusion System with the Applied Biosystems 3130xl Genetic Analyzer

Internal Validation of the Promega PowerPlex Fusion System with the Applied Biosystems 3130xl Genetic Analyzer Internal Validation of the Promega PowerPlex Fusion System with the Applied Biosystems 3130xl Genetic Analyzer Roy Al Ahmar, B.S. Marshall University Forensic Science Center DNA Laboratory Overview What

More information

GlobalFiler TM Extra Cycle Evaluation Report

GlobalFiler TM Extra Cycle Evaluation Report GlobalFiler TM Extra Cycle Evaluation Report Table of Content Chapter 1 Summary 1. Introduction... 2 2. Systems, Kits and Conditions... 3 3. Studies... 4 Chapter 2 Minimum Threshold and Background Study

More information

Testing Candidate DNA Quantitation Standards with Several Real-Time Quantitative PCR Methods*

Testing Candidate DNA Quantitation Standards with Several Real-Time Quantitative PCR Methods* Testing Candidate DNA Quantitation Standards with Several Real-Time Quantitative PCR Methods* Margaret C. Kline, Peter M. Vallone, Amy E. Decker, Janette W. Redman, David L. Duewer, and John M. Butler

More information

Evaluation of the PowerQuant System on the QuantStudio 5 Real-Time PCR System

Evaluation of the PowerQuant System on the QuantStudio 5 Real-Time PCR System Evaluation of the PowerQuant System on the QuantStudio Real-Time PCR System Erica K. Graham, Jon Drobac, Jonelle Thompson, Mary Loten and Anupama Gopalakrishnan Promega Corporation, 28 Woods Hollow Rd.,

More information

AmpF STR NGM PCR Amplification Kit - Overview

AmpF STR NGM PCR Amplification Kit - Overview AmpF STR NGM PCR Amplification Kit - Overview The most advanced STR kits optimized for analysis of forensic casework and database samples in Europe Data Quality Worth Sharing! 9/7/2011 Life Technologies

More information

Implement robotic liquid handlers in the performance of STR typing

Implement robotic liquid handlers in the performance of STR typing Development Areas! Automate the STR typing process Implement robotic liquid handlers in the performance of STR typing Replace the slot blot hybridization procedure with a real-time PCR based assay for

More information

DNA Quantification Results Obtained from Bone Samples Using the Investigator Quantiplex HYres Kit

DNA Quantification Results Obtained from Bone Samples Using the Investigator Quantiplex HYres Kit DNA Quantification Results Obtained from Bone Samples Using the Investigator Quantiplex HYres Kit Laura Gaydosh Combs, M.A. Doctoral Candidate Department of Molecular and Medical Genetics QIAGEN would

More information

An Efficient Method of Extracting DNA from Bone Remains from the Spanish Civil War A Comparative Study of Two Methods: PrepFiler BTA and DNAzol.

An Efficient Method of Extracting DNA from Bone Remains from the Spanish Civil War A Comparative Study of Two Methods: PrepFiler BTA and DNAzol. An Efficient Method of Extracting DNA from Bone Remains from the Spanish Civil War A Comparative Study of Two Methods: PrepFiler BTA and DNAzol. Betancor 1,2, E.; Fregel 2,3, R.; Suárez 2, N.M.; Cabrera

More information

Validation Using SDS Software Version on the Applied Biosystems 7500 Real-Time PCR System and the ABI PRISM 7000 Sequence Detection System

Validation Using SDS Software Version on the Applied Biosystems 7500 Real-Time PCR System and the ABI PRISM 7000 Sequence Detection System User Bulletin Quantifiler Kits April 2006 SUBJECT: Validation Using SDS Software Version 1.2.3 on the Applied Biosystems 7500 Real-Time PCR System and the ABI PRISM 7000 Sequence Detection System Note:

More information

Guidelines for Developing Robust and Reliable PCR Assays

Guidelines for Developing Robust and Reliable PCR Assays Guidelines for Developing Robust and Reliable PCR Assays Leta Steffen, PhD Applications Scientist Promega Corporation Outline 1) PCR reaction components What is in the reaction? How does it affect assay

More information

HID Real-Time PCR Analysis Software v1.1

HID Real-Time PCR Analysis Software v1.1 HID Real-Time PCR Analysis Software v1.1 USER BULLETIN NOVEMBER 2010 SUBJECT: Validation and new features In this user bulletin This user bulletin covers: Validation...........................................................

More information

Guidelines and Considerations for Evaluating or Validating Extraction Chemistry

Guidelines and Considerations for Evaluating or Validating Extraction Chemistry Guidelines and Considerations for Evaluating or Validating Extraction Chemistry Introduction In order to evaluate a DNA extraction method effectively, it is imperative to consider all aspects of chemistry

More information

DNA Recovery from Water-soluble Adhesive Tape Kawin Rasmeepaisarn Advisors: Thanit Kusamran Forensic science graduate programme, Mahidol University

DNA Recovery from Water-soluble Adhesive Tape Kawin Rasmeepaisarn Advisors: Thanit Kusamran Forensic science graduate programme, Mahidol University 20 th National Grad Research Conference 1 DNA Recovery from Water-soluble Adhesive Tape Kawin Rasmeepaisarn Advisors: Thanit Kusamran Forensic science graduate programme, Mahidol University Abstract O-BS004

More information

Investigation of the effects of sample degradation and inhibition in forensic DNA typing

Investigation of the effects of sample degradation and inhibition in forensic DNA typing Investigation of the effects of sample degradation and inhibition in forensic DNA typing Bruce McCord Professor Analytical/Forensic Chemistry FIU Department of Chemistry International Forensic Research

More information

PowerSeq 46GY System Eases the transition to MPS for forensic laboratories

PowerSeq 46GY System Eases the transition to MPS for forensic laboratories PowerSeq 46GY Eases the transition to MPS for forensic laboratories The Power to Solve...from Sample to Analysis Coming Soon! The Power to Solve... from Sample to Analysis The Promega PowerSeq s enable

More information

A comparative performance evaluation of illustra Ready-To-Go GenomiPhi V3 and illustra GenomiPhi V2 DNA amplification kits

A comparative performance evaluation of illustra Ready-To-Go GenomiPhi V3 and illustra GenomiPhi V2 DNA amplification kits GE Healthcare Life Sciences Application note 29-0245-46 AA Nucleic acid sample preparation A comparative performance evaluation of illustra Ready-To-Go GenomiPhi V3 and illustra GenomiPhi V2 DNA amplification

More information

AutoMate Express Forensic DNA Extraction System

AutoMate Express Forensic DNA Extraction System Product Bulletin Human Identification AutoMate Express Forensic DNA Extraction System Designed to improve the yield, reproducibility, and overall purity of DNA isolated from both routine and challenging

More information

DNA Typing for the Identification of Korean War Victims

DNA Typing for the Identification of Korean War Victims DNA Typing for the Identification of Korean War Victims Kyoung-Jin Shin, D.D.S., Ph.D. Department of Forensic Medicine Yonsei University College of Medicine Korean War (1950-1953) Military conflict between

More information

QIAGEN Validation Report

QIAGEN Validation Report QIAGEN Validation Report Developmental validation of the Investigator ESSplex SE Plus Kit The QIAGEN Investigator ESSplex SE Plus Kit is intended for molecular biology applications in forensic, human identity,

More information

DBS DNA Extraction, Validation & Quantitation

DBS DNA Extraction, Validation & Quantitation DBS DNA Extraction, Validation & Quantitation Suzanne Cordovado, PhD Molecular Quality Improvement Program NBS Molecular Training Class July 8, 2013 National Center for Environmental Health Centers for

More information

Validation of a Multiplexed System for Quantification of Human DNA and Human Male DNA and Detection of PCR Inhibitors in Biological Samples

Validation of a Multiplexed System for Quantification of Human DNA and Human Male DNA and Detection of PCR Inhibitors in Biological Samples 18 th International Symposium on Human Identification Validation of a Multiplexed System for Quantification of Human DNA and Human Male DNA and Detection of PCR Inhibitors in Biological Samples Maura Barbisin,

More information

Comparison of the Promega Power Y and AB AmpFlSTR YFiler kits and the Subsequent Validation of YFiler.

Comparison of the Promega Power Y and AB AmpFlSTR YFiler kits and the Subsequent Validation of YFiler. Comparison of the Promega Power Y and AB AmpFlSTR YFiler kits and the Subsequent Validation of YFiler. Dr. Timothy P McMahon Armed Forces DNA Identification Laboratory Rockville, Md 301-319 319-0238 Timothy.mcmahon@afip.osd.mil

More information

Data Quality Worth Sharing

Data Quality Worth Sharing Product Bulletin Human Identification AmpFlSTR NGM and NGM SElect PCR Amplification Kits Next generation amplification chemistries including the 5 new loci from the expanded European Standard Set Enhanced

More information

Characters of degraded DNA

Characters of degraded DNA STR Typing of Degraded DNA using a Repair Enzyme and Whole Genome Amplification Jeong Eun Sim 1, Hwan Young Lee 1, Seung Hwan Lee 2, Woo Ick Yang 1, Kyoung Jin Shin 1 1 Department of Forensic Medicine

More information

QIAGEN Validation Report

QIAGEN Validation Report QIAGEN Validation Report Developmental validation of the Investigator ESSplex Plus Kit The QIAGEN Investigator ESSplex Plus Kit is intended for molecular biology applications in forensic, human identity,

More information

Predicting the success of STR typing using the Plexor HY System and extraction of mock cases using the Differex System

Predicting the success of STR typing using the Plexor HY System and extraction of mock cases using the Differex System Predicting the success of STR typing using the Plexor HY System and extraction of mock cases using the Differex System Susan A. Greenspoon 1, Rebecca C. Bohr 2, Alexandra Pittock 3, Jenni Grubb 1, Katie

More information

Artifacts Identified Post-Developmental Validation: AmpFLSTR Identifiler Plus PCR Amplification Kit

Artifacts Identified Post-Developmental Validation: AmpFLSTR Identifiler Plus PCR Amplification Kit March 29, 2018 TECHNICAL NOTE Artifacts Identified Post-Developmental Validation: AmpFLSTR Identifiler Plus PCR Amplification Kit The purpose of this document is to assist with data interpretation by providing

More information

PowerPlex ESX & ESI Fast Systems

PowerPlex ESX & ESI Fast Systems PowerPlex ESX & ESI Fast Systems PowerPlex ESX & ESI Fast Systems The PowerPlex ESX and ESI Fast Systems meet recommendations of the European Network of Forensic Science Institutes (ENFSI) for DNA profile

More information

TECHNICAL MANUAL. SwabSolution Kit. Instructions for Use of Product DC8271. Revised 9/16 TMD037

TECHNICAL MANUAL. SwabSolution Kit. Instructions for Use of Product DC8271. Revised 9/16 TMD037 TECHNICAL MANUAL SwabSolution Kit Instructions for Use of Product DC8271 Revised 9/16 TMD037 SwabSolution Kit All technical literature is available at: www.promega.com/protocols/ Visit the web site to

More information

Automated Extraction of High Quality Genomic DNA from Forensic Evidence Samples Using a Cartridge-Based System

Automated Extraction of High Quality Genomic DNA from Forensic Evidence Samples Using a Cartridge-Based System Automated Extraction of High Quality Genomic DNA from Forensic Evidence Jason Yingjie Liu1, Ph.D.*, Maxim Brevnov1, Ph.D., Allison Holt1, Ph.D., James E. Stray1, Ph.D., Declan Donovan2, Ph.D., Alan B.

More information

Genetic Identity. Steve Harris SPASH - Biotechnology

Genetic Identity. Steve Harris SPASH - Biotechnology Genetic Identity Steve Harris SPASH - Biotechnology Comparison of Organisms ORGANISM GENES BASE PAIRS Lambda Phage 40 50,000 E.coli 400 5,000,000 Yeast 13,000 15,000,000 Human 20,000 3,000,000,000 (3 billion)

More information

Development and Validation of the 3500 Series Genetic Analyzer for Human Identification

Development and Validation of the 3500 Series Genetic Analyzer for Human Identification Development and Validation of the 3500 Series Genetic Analyzer for Human Identification Jeff Sailus Forensic Science Applications Group 1 Development of a Next Generation Genetic Analyzer A Whole System

More information

Y chromosomal STRs in forensics

Y chromosomal STRs in forensics Haplotype and mutation analysis for newly suggested Y-STRs in Korean father-son pairs Yu Na Oh, Hwan Young Lee, Eun Young Lee, Eun Hey Kim, Woo Ick Yang, Kyoung-Jin Shin Department of Forensic Medicine

More information

Plexor HY System for the Applied Biosystems 7500 and 7500 FAST Real-Time PCR Systems

Plexor HY System for the Applied Biosystems 7500 and 7500 FAST Real-Time PCR Systems TECHNICAL MANUAL Plexor HY System for the Applied Biosystems 7500 and 7500 FAST Real-Time PCR Systems Instructions for Use of Products DC1000, DC1001 and DC1500 Technical Manuals with instructions for

More information

AutoLys workstation: Validation of a fully automated forensic sample lysis solution

AutoLys workstation: Validation of a fully automated forensic sample lysis solution AutoLys workstation: Validation of a fully automated forensic sample lysis solution Application note Top 3 reasons for automation of this application Sample lysis and DNA extraction can be regarded as

More information

Developmental validation of the Investigator Argus X-12 QS Kit

Developmental validation of the Investigator Argus X-12 QS Kit Validation Report December 2016 Developmental validation of the Investigator Argus X-12 QS Kit The QIAGEN Investigator Argus X-12 QS Kit is intended for molecular biology applications in forensic, human

More information

!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"

!!!!!!!!!!!!!!!!!!!!!!!!!!! STR ANALYSIS BY CAPILLARY ELECTROPHORESIS: DEVELOPMENT OF INTERPRETATION GUIDELINES FOR THE PROFILER PLUS AND COFILER SYSTEMS FOR USE IN FORENSIC SCIENCE Deborah Hobson, Jill Smerick, and Jenifer Smith

More information

Efficient Method for Isolation of High Quality Concentrated Cellular RNA with Extremely Low Levels of Genomic DNA Contamination Application

Efficient Method for Isolation of High Quality Concentrated Cellular RNA with Extremely Low Levels of Genomic DNA Contamination Application Efficient Method for Isolation of High Quality Concentrated Cellular RNA with Extremely Low Levels of Genomic DNA Contamination Application Gene Expression Authors Ilgar Abbaszade, Claudia Robbins, John

More information

Direct PCR Amplification of Blood Samples on Copan NUCLEIC-CARD TM Systems Using the AmpFlSTR Identifiler Direct PCR Amplification Kit

Direct PCR Amplification of Blood Samples on Copan NUCLEIC-CARD TM Systems Using the AmpFlSTR Identifiler Direct PCR Amplification Kit Technical Focus Direct PCR Amplification of Blood Samples on Copan NUCLEIC-CARD TM Systems Using the AmpFlSTR Identifiler Direct PCR Amplification Kit Nicola Oldroyd, Jennifer Schroeder, Dennis Wang, Alessandro

More information

Validation of PowerPlex Y23 PCR Amplification System for Forensic Casework

Validation of PowerPlex Y23 PCR Amplification System for Forensic Casework Validation of PowerPlex Y23 PCR Amplification System for Forensic Casework Jessica Tokarz (B.S.) 1, Sarah Foster (B.S.) 2, Scott Milne (B.S.) 2, Pamela Staton (PhD) 1, Kathleen Press (M.S.F.S.) 3 1 Marshall

More information

Development of an RNA-based screening assay for forensic stain identification

Development of an RNA-based screening assay for forensic stain identification Development of an RNA-based screening assay for forensic stain identification 1 Trisha L. Noreault-Conti, Ph.D. and 1 Eric Buel, Ph.D. 1 Vermont Forensic Laboratory, Department of Public Safety, 103 South

More information

AUTOMATION OF DNA ISOLATION, QUANTITATION, AND PCR SETUP

AUTOMATION OF DNA ISOLATION, QUANTITATION, AND PCR SETUP AUTOMATION OF DNA ISOLATION, QUANTITATION, AND PCR SETUP Paraj Mandrekar, Laura Flanagan, Robert McLaren and Allan Tereba Promega Corporation, Madison, WI Abstract As a result of increased casework due

More information

Upgrading the Workflow: Enhancing Information Recovery from Forensic Evidence. The world leader in serving science

Upgrading the Workflow: Enhancing Information Recovery from Forensic Evidence. The world leader in serving science Upgrading the Workflow: Enhancing Information Recovery from Forensic Evidence The world leader in serving science The GlobalFiler Kit: Going Further for Casework Powered by 6-Dye Chemistry 2 What is the

More information

HID Validation: The 3500 Series Genetic Analyzers

HID Validation: The 3500 Series Genetic Analyzers Outline Scope Materials and Reagents Validation Elements Selection of Validation Data Additional Sources of Validation Information User Bulletin 2 Validation Scope Hardware and Reagents HID Validation:

More information

Reliable. answers, significant savings. Applied Biosystems QuantStudio 5 Real-Time PCR System for Human Identification

Reliable. answers, significant savings. Applied Biosystems QuantStudio 5 Real-Time PCR System for Human Identification Reliable answers, significant savings Applied Biosystems QuantStudio 5 Real-Time PCR System for Human Identification Discover the QuantStudio 5 Real-Time PCR System for Human Identification The modern

More information

A Recommended Procedure for Real-Time Quantitative TaqMan PCR for Roundup Ready Canola RT73 Monsanto Biotechnology Regulatory Sciences

A Recommended Procedure for Real-Time Quantitative TaqMan PCR for Roundup Ready Canola RT73 Monsanto Biotechnology Regulatory Sciences Page 1 of 7 Overview Purpose & Scope This procedure describes an event-specific real-time TaqMan PCR method for determination of the relative content of Roundup Ready canola RT73 (hereafter referred to

More information

STR Analysis Following Latent Blood Detection by Luminol, Fluorescein, and BlueStar

STR Analysis Following Latent Blood Detection by Luminol, Fluorescein, and BlueStar Technical Note STR Analysis Following Latent Blood Detection by Luminol, Fluorescein, and BlueStar Cathy J. Jakovich San Diego Sheriff s Crime Laboratory San Diego, CA Abstract: Luminol and f luorescein

More information

Executive Summary. clinical supply services

Executive Summary. clinical supply services clinical supply services case study Development and NDA-level validation of quantitative polymerase chain reaction (qpcr) procedure for detection and quantification of residual E.coli genomic DNA Executive

More information

Functional Genomics Research Stream. Research Meeting: June 19, 2012 SYBR Green qpcr, Research Update

Functional Genomics Research Stream. Research Meeting: June 19, 2012 SYBR Green qpcr, Research Update Functional Genomics Research Stream Research Meeting: June 19, 2012 SYBR Green qpcr, Research Update Updates Alternate Lab Meeting Fridays 11:30-1:00 WEL 4.224 Welcome to attend either one Lab Log thanks

More information

DNA Methylation Profiling Using SNaPshot Multiplex

DNA Methylation Profiling Using SNaPshot Multiplex DNA Methylation Profiling Using SNaPshot Multiplex This picture is the property of the Genetic Science Learning Center at the University of Utah How to set up DNA methylation profiling in your lab Introducing

More information

AmpFlSTR Identifiler PCR Amplification Kit

AmpFlSTR Identifiler PCR Amplification Kit Product Bulletin Human Identification AmpFlSTR Identifiler PCR Amplification Kit Fifteen STR (short tandem repeat) loci and Amelogenin co-amplified in a single tube Incorporation of the same proven, reliable

More information

Testing the Effectiveness of the PrepFiler Kit for DNA Extraction from Forensic Samples (An Overview of Test Site Data)

Testing the Effectiveness of the PrepFiler Kit for DNA Extraction from Forensic Samples (An Overview of Test Site Data) July 28 Testing the Effectiveness of the PrepFiler Kit for DNA Extraction from Forensic Samples (An Overview of Test Site Data) Introduction Optimizing the forensic sample processing workflow requires

More information

Evaluation of low trace DNA recovery techniques from ridged surfaces. Graham Williams*, Manohar Pandre, Waseeh Ahmed, Emma Beasley, and Emma.

Evaluation of low trace DNA recovery techniques from ridged surfaces. Graham Williams*, Manohar Pandre, Waseeh Ahmed, Emma Beasley, and Emma. Evaluation of low trace DNA recovery techniques from ridged surfaces Graham Williams*, Manohar Pandre, Waseeh Ahmed, Emma Beasley, and Emma Omelia Forensic Biology Group, School of Applied Sciences, University

More information

Developmental Validation of Amplicon Rx, a post-pcr Purification System

Developmental Validation of Amplicon Rx, a post-pcr Purification System Developmental Validation of Amplicon Rx, a post-pcr Purification System Introduction Allelic drop out and allele peaks below analytical threshold are common observation in forensic DNA analysis. These

More information

Technical Review. Real time PCR

Technical Review. Real time PCR Technical Review Real time PCR Normal PCR: Analyze with agarose gel Normal PCR vs Real time PCR Real-time PCR, also known as quantitative PCR (qpcr) or kinetic PCR Key feature: Used to amplify and simultaneously

More information

Accurately and reproducibly quantify as little as 20 fg of bacterial DNA using real-time PCR.

Accurately and reproducibly quantify as little as 20 fg of bacterial DNA using real-time PCR. INSTRUCTION MANUAL Femto Bacterial DNA Quantification Kit Catalog No. E2006 Highlights Accurately and reproducibly quantify as little as 20 fg of bacterial DNA using real-time PCR. High specificity and

More information

Precise quantification of Ion Torrent libraries on the QuantStudio 3D Digital PCR System

Precise quantification of Ion Torrent libraries on the QuantStudio 3D Digital PCR System APPLICATION NOTE QuantStudio 3D Digital PCR System Precise quantification of Ion Torrent libraries on the QuantStudio 3D Digital PCR System Introduction The template preparation step in the next-generation

More information

Functional DNA Quality Analysis Improves the Accuracy of Next Generation Sequencing from Clinical Specimens

Functional DNA Quality Analysis Improves the Accuracy of Next Generation Sequencing from Clinical Specimens Functional DNA Quality Analysis Improves the Accuracy of Next Generation Sequencing from Clinical Specimens Overview We have developed a novel QC, the SuraSeq DNA Quantitative Functional Index (QFI ).

More information

Male/female DNA mixtures: a challenge for Y-STR analysis

Male/female DNA mixtures: a challenge for Y-STR analysis International Congress Series 1239 (2003) 295 299 Male/female DNA mixtures: a challenge for Y-STR analysis Burkhard Berger, Harald Niederstätter, Silvano Köchl, Martin Steinlechner, Walther Parson * Institute

More information

The Recovery of Touch DNA from RDX-C4 Evidences: Case Study

The Recovery of Touch DNA from RDX-C4 Evidences: Case Study Page 1 of 6 The Recovery of Touch DNA from RDX-C4 Evidences: Case Study N. R. Al Snan* *Forensic Science Laboratory, Directorate of Forensic Science, Ministry of Interior, Kingdom of Bahrain Abstract.

More information

Fully Automated Library Quantification for Illumina Sequencing on the NGS STAR

Fully Automated Library Quantification for Illumina Sequencing on the NGS STAR Fully Automated Library Quantification for Illumina Sequencing on the NGS STAR Introduction Hamilton Robotics, an industry leader in liquid handling and laboratory automation equipment, has partnered with

More information

QIAGEN Whole Genome Amplification REPLI-g Eliminating Sample Limitations, Potential Use for Reference Material

QIAGEN Whole Genome Amplification REPLI-g Eliminating Sample Limitations, Potential Use for Reference Material QIAGEN Whole Genome Amplification REPLI-g Eliminating Sample Limitations, Potential Use for Reference Material MDA Technology Protocols Locus Bias Applications Kits & Service Summary QIAGEN REPLI-g WGA

More information

Use of Spike-ins for Sample Tracking in Agilent Array CGH

Use of Spike-ins for Sample Tracking in Agilent Array CGH Use of Spike-ins for Sample Tracking in Agilent Array CGH Application Note Authors Srirangan Sampath, Ph.D., FACMG and Heather Klein PreventionGenetics Iman Kishawi, Ph.D. Agilent Technologies, Inc. Abstract

More information

Module 7B DNA Collection, Extraction, and Analysis. Forensic Science Teacher Professional Development

Module 7B DNA Collection, Extraction, and Analysis. Forensic Science Teacher Professional Development Module 7B DNA Collection, Extraction, and Analysis Forensic Science Teacher Professional Development Unit 3 Sample Collection, DNA Extraction, and DNA Quantitation Part 1 Sample Collection DNA can be found

More information

Sus scrofa Pig genesig Advanced Speciation Kit

Sus scrofa Pig genesig Advanced Speciation Kit TM Primerdesign Ltd Pig genesig Advanced Speciation Kit 100 tests For general laboratory and research use only 1 Principles of the test Real-time PCR This kit provides a method for detecting mitochondrial

More information

Going MULTI How to Easily Achieve High Multiplexing in Real-Time PCR

Going MULTI How to Easily Achieve High Multiplexing in Real-Time PCR qpcr Symposium 2005 - Weihenstephan Going MULTI How to Easily Achieve High Multiplexing in Real-Time PCR Dr. Andreas Missel Associate Director Research & Development Overview Introduction Multiplex real-time

More information

ProNex DNA QC Assay Calibration Kit, 7500

ProNex DNA QC Assay Calibration Kit, 7500 TECHNICAL MANUAL ProNex DNA QC Assay Calibration Kit, 7500 Instructions for Use of Products NG1001 9/17 TM515 ProNex DNA QC Assay Calibration Kit, 7500 All technical literature is available at: www.promega.com/protocols/

More information

University of Huddersfield Repository

University of Huddersfield Repository University of Huddersfield Repository Williams, Graham, Pandre, Manohar, Ahmed, Waseeh, Beasley, Emma and Omelia, Emma J. Evaluation of low trace DNA recovery techniques from ridged surfaces Original Citation

More information

High specificity and sensitivity for fungal DNA allows reliable quantification in a background of nonfungal. Contents

High specificity and sensitivity for fungal DNA allows reliable quantification in a background of nonfungal. Contents INSTRUCTION MANUAL Femto Fungal DNA Quantification Kit Catalog No. E2007 Highlights Accurately and reproducibly quantify as little as 20 fg of fungal DNA from 1 µl of sample using realtime PCR. High specificity

More information

GlobalFiler Kit Casework Test Site Results and Other Development Updates

GlobalFiler Kit Casework Test Site Results and Other Development Updates GlobalFiler Kit Casework Test Site Results and Other Development Updates April Orbison Sr. Field Applications Specialist Human Identification 1 Agenda GlobalFiler Kit Update: Recent Casework Test Site

More information

DNA 101. DNA 101:Kate Philpott and Jennifer Friedman

DNA 101. DNA 101:Kate Philpott and Jennifer Friedman DNA 101 DNA 101:Kate Philpott and Jennifer Friedman DNA found in cells of your body cell nucleus DNA Found in cells once they leave your body too CONSIDER THE SOURCE Blood* Semen* Hair Bone Saliva* Urine*

More information

PRODUCT INFORMATION Thermo Scientific Luminaris Probe Low ROX qpcr Master Mix #K0944 For 5000 rxns Lot Expiry Date Store at -20 C in the dark CERTIFICATE OF ANALYSIS The absence of endo-, exodeoxyribonucleases

More information

Application Note Detecting low copy numbers. Introduction. Methods A08-005B

Application Note Detecting low copy numbers. Introduction. Methods A08-005B Application Note Detecting low copy numbers A08-005B Introduction Sensitivity of a qpcr assay is highly dependent on primer efficiency. Not all assays will be capable of detecting a single copy of template

More information

DNA Genotyping from Human FFPE Samples Reliable and Reproducible

DNA Genotyping from Human FFPE Samples Reliable and Reproducible DNA Genotyping from Human FFPE Samples Reliable and Reproducible Use this optimized protocol to obtain reliable and reproducible SNP Genotyping data from those difficult human FFPE samples. 1 FFPE DNA Isolation

More information

Sexual Assault Kit Screening Using the QIAcube HT & Investigator Quantiplex HYres!

Sexual Assault Kit Screening Using the QIAcube HT & Investigator Quantiplex HYres! Sexual Assault Kit Screening Using the QIAcube HT & Investigator Quantiplex HYres! Joshua Stewart, MSFS! QIAGEN North American Users Meeting! Houston, TX! July 30, 2014! QIAGEN would like to thank our

More information

Validation of the AmpF STR R MiniFiler TM PCR Amplification Kit for Use in Forensic Casework*

Validation of the AmpF STR R MiniFiler TM PCR Amplification Kit for Use in Forensic Casework* TECHNICAL NOTE J Forensic Sci, September 2009, Vol. 54, No. 5 doi: 10.1111/j.1556-4029.2009.01099.x Available online at: www.blackwell-synergy.com Coral Luce, 1 M.S.; Shawn Montpetit, 2 M.S.; David Gangitano,

More information

PowerPlex Y23 Developmental Validation

PowerPlex Y23 Developmental Validation PowerPlex Y23 Developmental Validation Jonelle Thompson, R&D Scientist September 2012 PowerPlex Y23 System Presentation outline Benefits of Y-STR Analysis Background information on PowerPlex Y23 Developmental

More information

TaqPath ProAmp Master Mixes

TaqPath ProAmp Master Mixes PRODUCT BULLETIN es es Applied Biosystems TaqPath ProAmp Master Mixes are versatile master mixes developed for high-throughput genotyping and copy number variation (CNV) analysis protocols that require

More information

Sus scrofa Pig genesig Speciation Kit

Sus scrofa Pig genesig Speciation Kit TM Primerdesign Ltd Sus scrofa genesig Speciation Kit 100 tests For general laboratory and research use only 1 Principles of the test Real-time PCR This kit provides a method for detecting Sus scrofa mitochondrial

More information

HID EVOlution Extraction System and HID EVOlution Combination System

HID EVOlution Extraction System and HID EVOlution Combination System HID EVOlution Extraction System and HID EVOlution Combination System USER BULLETIN JUNE 2010 SUBJECT: Validation of PrepFiler Wash Buffer B and the Related Modifications to Worktable Layout and Scripts

More information

ABSTRACT INTRODUCTION AND BACKGROUND

ABSTRACT INTRODUCTION AND BACKGROUND New DNA Extraction Methods for Casework Evidence Martha F. Burger, Todd W. Bille, and James W. Schumm The Bode Technology Group, Springfield, VA 22150 USA ABSTRACT To ease the national backlog of crime

More information

Pressure Cycling Technology (PCT) Applications for DNA Extractions from Challenging Forensic Samples

Pressure Cycling Technology (PCT) Applications for DNA Extractions from Challenging Forensic Samples Pressure Cycling Technology (PCT) Applications for DNA Extractions from Challenging Forensic Samples Suzanne Gonzalez, Elizabeth Feller, Dixie Peters, Bruce Budowle, and Arthur Eisenberg University of

More information

HID Real-Time PCR Analysis Software v1.3

HID Real-Time PCR Analysis Software v1.3 PRODUCT BULLETIN Human identification HID Real-Time PCR Analysis Software v1.3 Enhanced data analysis for QuantStudio 5 and 7500 Real-Time PCR Systems with Quantifiler kit Applied Biosystems Quantifiler

More information

Quantitative, multiplexed amplification with the Plexor qpcr Systems. Douglas R. Storts, Ph.D. Director R&D

Quantitative, multiplexed amplification with the Plexor qpcr Systems. Douglas R. Storts, Ph.D. Director R&D Quantitative, multiplexed amplification with the Plexor qpcr Systems Douglas R. Storts, Ph.D. Director R&D Novel Base-Pairing IsoC & IsoG dntps Recognized by DNA Polymerase Do not base pair with ACGTU

More information

Procedure for Casework DNA Interpretation

Procedure for Casework DNA Interpretation Procedure for Casework DNA Interpretation 1.0 Purpose The purpose of this document is to provide guidelines for the interpretation of autosomal DNA results when amplified with Identifiler Plus. 2.0 Scope

More information

Absolute Human Telomere Length and Mitochondrial DNA Copy Number Dual Quantification qpcr Assay Kit (AHDQ) Catalog # reactions

Absolute Human Telomere Length and Mitochondrial DNA Copy Number Dual Quantification qpcr Assay Kit (AHDQ) Catalog # reactions Absolute Human Telomere Length and Mitochondrial DNA Copy Number Dual Quantification qpcr Assay Kit (AHDQ) Catalog #8958 100 reactions Product Description Telomeres are repetitive nucleotide elements at

More information

Accurately and reproducibly quantify as little as 20 fg of human DNA using real-time PCR.

Accurately and reproducibly quantify as little as 20 fg of human DNA using real-time PCR. INSTRUCTION MANUAL Femto Human DNA Quantification Kit Catalog No. E2005 Highlights Accurately and reproducibly quantify as little as 20 fg of human DNA using real-time PCR. High specificity and sensitivity

More information

Application of mixsep software package: Performance verification of male-mixed DNA analysis

Application of mixsep software package: Performance verification of male-mixed DNA analysis MOLECULAR MEDICINE REPORTS 12: 2431-2442, 2015 Application of mixsep software package: Performance verification of male-mixed DNA analysis NA HU 1, BIN CONG 1, TAO GAO 2, YU CHEN 2, JUNYI SHEN 3, SHUJIN

More information

PAXgene Blood DNA PAXgene Blood DNA Tube (IVD) for clinical use PAXgene Blood DNA System (RUO) for research use. Explore more at

PAXgene Blood DNA PAXgene Blood DNA Tube (IVD) for clinical use PAXgene Blood DNA System (RUO) for research use. Explore more at PAXgene Blood DNA PAXgene Blood DNA Tube (IVD) for clinical use PAXgene Blood DNA System (RUO) for research use Explore more at www.preanalytix.com Situation The composition, amount, quality and integrity

More information