1 Clin. Lab. 2014;60: Copyright SHORT COMMUNICATION Urinalysis: Comparison between Microscopic Analysis and a New Automated Microscopy Image-Based Urine Sediment Instrument PAULA V. BOTTINI, MAYARA H. M. MARTINEZ, CÉLIA R. GARLIPP Division of Clinical Pathology, University of Campinas/UNICAMP, Campinas, SP, Brazil SUMMARY Background: Urinalysis is a high demand procedure, with a large amount of manual labor and poorly standardized. Recently a new walk-away automated urine analyzer has been introduced. The aim of this study was to evaluate the performance of UriSed as an alternative to the microscopic analysis of urine samples. Methods: Four hundred mid-stream urine samples from patients with several clinical conditions were analyzed by bright field microscopy and by UriSed. The validation protocol included studies of precision, carryover, and comparison with the gold standard microscopy. Results: Our data showed that UriSed is a precise method. Repeatability and reproducibility ranged from 8 to 15%. Carryover was negligible. All the elements showed good agreement between both methods, with an extremely high correlation between the erythrocyte and leukocyte counts (r > 0.95) Conclusions: UriSed is a precise and accurate alternative to microscopy that allows a better workflow and may significantly improve turnaround time. (Clin. Lab. 2014;60: DOI: /Clin.Lab ) KEY WORDS urinalysis, microscopic examination, automated sediment analysis INTRODUCTION Urinalysis is a high demand procedure which includes physical, chemical, and microscopic examinations. Although the analysis of urine sediment provides essential information on the functional state of the kidneys, it is a labor intensive and time consuming procedure. Besides that, it is poorly standardized and features a wide interobserver variability as several factors may contribute to the inaccuracy of this method (e.g., centrifugation, differences in interpretation of cellular elements, cylinders, etc.) [1,2]. Automation seems to be the answer to this issue [3]. In an attempt to automate the microscopic analysis of urine, some urine analysis systems have been developed beginning in the late 80s. The great advantage of automated analysis over microscopic procedure is that the former reduces the inter-observer variability and increases the accuracy and the productivity. While microscopic analysis usually takes several minutes to be performed automation can release a result in about one minute [4]. A few years ago, a new walk-away automated urine sediment analyzer based on the KOVA method with on-screen review of the images was introduced. This analyzer is basically the automation of the traditional manual microscopy [5]. The aim of this study was to evaluate the performance of this image based automated sediment analyzer (UriSed, also called sedimax in some countries) as an alternative to the microscopic analysis of urine samples. MATERIALS AND METHODS We analyzed 400 mid-stream urine samples from patients (262 women and 138 men) with several clinical conditions. Their ages varied from 2 to 91 years (median = 41 years). All samples were analyzed by bright field microscopy Short Communication accepted July 24, 2013 Clin. Lab. 4/

2 P. V. BOTTINI et al. Table 1. Precision of UriSed. Intra-assay precision Inter-assay precision LEVEL 1 LEVEL 2 LEVEL 1 LEVEL 2 element/hpf mean ± SD CV mean ± SD CV mean ± SD CV mean ± SD CV RBC 25.7 ± 2.0 8% 83.6 ± 7.2 9% 25.5 ± 2.0 8% 95.8 ± 7.7 8% WBC 11.7 ± % 64.7 ± % 16.1 ± % ± % RBC - red blood cell, WBC - white blood cell hpf - high power field, SD - standard deviation, CV - coefficient of variation. Table 2. Diagnostic performance of UriSed. RBC WBC Hyaline Casts Path Casts Crystals Bacteria Sensitivity (%) Specificity (%) Positive Predictive value (%) Negative Predictive value (%) Agreement (%) p < < < 0.05 < < < RBC - red blood cell, WBC - white blood cell, Path - pathological. using the KOVA method with the results expressed in terms of quantitative elements (erythrocytes-rbc and leukocytes-wbc; arithmetic mean of 10 fields of 400x) and qualitative elements (casts, crystals, and bacteria, considering all fields), and by a fully automated sediment analyzer (UriSed - 77 Elektronika Kft, Budapest, Hungary). UriSed is based on microscopic examination of urine samples (KOVA method). The analyzer homogenizes, aspires, and transfers the sample to a special disposable cuvette. The cuvette is centrifuged and placed in microscopy position and whole-field images are obtained and evaluated by an auto image evaluation module (AIEM). Results were expressed as particle concentration/hpf (high power field). These high definition images are displayed on-screen and can be stored for further review and for educational purposes [5]. All 15 images of each sample were reviewed by an experienced analyst and manually edited, if necessary. The validation protocol was based on local regulation, on CLSI documents EP05-A2 [6] and EP15-A2 [7], and on ICSH guideline [8] and included studies of precision (repeatability/reproducibility) and carryover. Repeatability (intra assay precision) was obtained from the analysis of 20 replicates of two urine samples with different particle numbers. The inter assay precision (reproducibility) was performed by analyzing two levels of control samples (Liquichek Urinalysis Control - Biorad) on 20 consecutive days. Carryover effect was assessed by testing a high-level urine sample consecutively in triplicate (H1, H2 and H3) followed by a low-level sample (L1, L2 and L3). Carryover was calculated using the following formula: carryover (%) = [(L1 - L3)/(H1 - L3)] x 100. Accuracy of the automated method was evaluated by comparison with microscopic analysis. Simple linear regression (least square method) was used to evaluate the correlation between quantitative analysis (RBC and WBC) by both methods. The Chi-squared statistical method was used to determine the agreement between microscopic analysis and UriSed. RBC and WBC counts > 5/hpf were classified as abnormal. Qualitative elements were compared according to the detection of them by each method. RESULTS Intra assay precision (repeatability) and inter assay precision (reproducibility) ranged from 8 to 15%, depending on the considered parameter (Table 1). As expected, there was no carryover since UriSed uses single-use cuvette for each sample. The results of high to low carryover returned negative values: -3.2% (RBC), -1.8% (WBC) and -1.2% (bacteria). There was a good correlation between the erythrocyte 694 Clin. Lab. 4/2014

3 URINALYSIS: MICROSCOPIC AND AUTOMATED ANALYSIS Figure 1. Distribution of RBC and WBC counts. and leukocyte counts by both methods (r > 0.95). Figure 1 shows the distribution of RBC and WBC counts by manual and automated analysis. All the elements showed good agreement between microscopic analysis and UriSed, as seen in Table 2. The analyzer easily detected the most common crystals seen in urine samples. The major causes of misclassification are the presence of squamous epithelial cells, mucus or amorphous crystals, emphasizing the importance of a recently collected mid-stream urine sample. DISCUSSION There is no doubt that microscopy is an important part of urinalysis and may provide useful information. As it is a labor intensive and time consuming procedure many efforts have been made to develop a better, cost-effective strategy without loss in the diagnostic yield. The first attempt in reducing the number of urine microscopy analyses was based on urine dipsticks. This procedure led to conflicting results as studies showed that a significant percentage of urine samples with normal chemistry results on the dipsticks had abnormal findings on microscopic analysis of sediment [9-11]. Studies have demonstrated significant inter-observer variation in the identification and interpretation of urine structures [1,12,13]. Automated urine microscopy is able to classify, quantify, and report urine particles and it seems to be an efficient alternative to reduce the manual workload and improve inter-observer reliability. Microscopic examination of urine sediment involves classification and quantification of particles present in this fluid and automated devices were developed to facilitate and standardize this procedure. This goal can be achieved through three systems with different methodologies. One of them (Sysmex UF-100/UF-1000i) is based on the use of fluorescence flow cytometry to classify the elements present in urine. Results are displayed as scattergrams and numerical values are reported [2]. Samples whose results have been flagged need to be reviewed by microscopy. Another analytical system is based on flow cell digital imaging with automatic particle recognition software (Iris iq 200) to identify and classify isolated images based on the texture, contrast, size, and shape of each element. The images are displayed to the analyst, allowing edition and reclassification, if necessary [14]. These images are considerably different from those observed in microscopy requiring substantial training before using the equipment [15]. The third analyzer (UriSed) is based on microscopic urine sediment analysis with digital images and automatic recognition of particles. UriSed corresponds to the automation of manual microscopy. Due to its characteristics, UriSed eliminates the need of microscopic Clin. Lab. 4/

4 P. V. BOTTINI et al. review of most of the samples. Images are very similar to those observed by microscopy and can be easily reviewed by the analyst. It is noteworthy that automated urinalysis, enables better workflow and may significantly improve turnaround time, since its processing time is much faster than the manual method ( samples/hour versus around 10 samples hour/technologist, respectively). Our data showed that UriSed is a precise method. Interand intra-assay precision ranged from 8% to 15%. This rate is much lower than that observed in manual microscopy where several studies showed a variation of more than 50% depending on the particle counts [14-16]. Carryover was negligible, as described by Zaman et al. [17]. All the elements showed good agreement between microscopic and automated analysis. Our data were in accordance with those obtained by Akin et al. [14] and Zaman et al. [17]. Other authors, using automated urine analyzers based on flow cytometry (UF-100/UF-1000i) or digital image with particle recognition (iq 200) also observed good correlation with the gold standard microscopy [16,18,19]. Akin et al. compared the analytical performance of UriSed with another automated analyzer (iq 200) and observed a significant correlation between them [14]. Appropriate preanalytical procedures are the basic condition for obtaining an adequate sample that allows reliable results and a proper clinical interpretation. As stated before, one of the major causes of misclassification was the presence of contaminating elements such as squamous epithelial cells and mucus. Collecting a midstream sample is the most relevant step to prevent contamination. A recent multicenter study conducted by Manoni et al. [20] clearly confirmed mid-stream urine as the most suitable sample for routine urinalysis. In summary, UriSed is a precise and accurate alternative to the gold standard microscopy. The implementation of this automated routine allows a better workflow and may significantly improve turnaround time. Besides that, it has the potential to eliminate the need of a microscopic review of most of the samples as UriSed permits an on-screen review of the images. It is noteworthy that the use of a system that photographs and stores high-definition images provides an invaluable training tool for technologists and medical students. Acknowledgement: We thank José Ricardo Lauand, Luciane Franco-Fernandes, and Solange Gomes Lara Cioffi for helping us during the validation process. Declaration of Interest: The authors declare that there are no conflicts of interest. References: 1. Winkel P, Statland BE, Jorgenson J. Urine microscopy: an ill-defined method examined by a multifactorial technique. Clin Chem 1974;20: Delanghe JR, Kouri TT, Huber AR, et al. The role of automated urine particle flow cytometry in clinical practice Clinica Chimica Acta 2000;301: Langlois MR, Delanghe JR, Steyaert SR, Everaert KC, De Buyzere ML. Automated flow cytometry compared with an automated dipstick reader for urinalysis. Clinical Chemistry 1999;45: Block DR, Lieske JC. Automated urinalysis in the clinical lab. Medical Laboratory Observer 2012;44(10): Barta Z, Kránicz Tünde, Bayer G. UriSed technology - A standardised automatic method of urine sediment analysis. European Infectious Disease 2011;5(2): Clinical and Laboratory Standards Institute. Evaluation of precision performance of quantitative measurement methods. CLSI document EP05-A2. Wayne, PA, USA: CLSI; Clinical and Laboratory Standards Institute. User verification of performance for precision and trueness; approved guideline - second edition. CLSI document EP15-A2. Wayne, PA, USA: CLSI; International Council for Standardization in Haematology. Guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte counting and cell marker applications. Clin Lab Haemat 1994;16: Szwed JJ, Schaust C. The Importance of microscopic examination of the urinary sediment. American Journal of Medical Technology 1982;48(2): Shaw ST Jr, Poon SY, Wong ET. Routine urinalysis : is the dipstick enough? JAMA 1985;253: Resnick M. Comment on simplifying urinalysis. Clin Chem 1985;31: Wald R, Bell CM, Nisenbaum Ret al. Interobserver reliability of urine sediment interpretation. Clin J Am Soc Nephrol 2009;4: Tsai JJ, Yeun JY, Kumar VA, Don BR. Comparison and interpretation of urinalysis performed by a nephrologist versus a hospitalbased clinical laboratory. American Journal of Kidney Diseases 2005;46(5): Akin OK, Serdat MA, Cizmeci Z, et al. Comparison of LabUMatwith-UriSed and iq 200 fully automatic urine sediment analysers with manual urine analysis. Biotechnol Appl Biochem 2009:53: Block DR, Lieske JC. Automated urinalysis in the clinical lab. MLO Med Lab Obs 2012;44(10):8-10,12;quiz Mayo S, Acevedo D, Quiñones-Torrelo C, Canós I, Sancho M. Clinical laboratory automated urinalysis: comparison among automated microscopy, flow cytometry, two test strips analyzers, and manual microscopic examination of the urine sediments. J Clin Lab Anal 2008;22(4): Zaman Z, Fogazzi GB; Garigali G et al. Urine sediment analysis: Analytical and diagnostic performance of sedimax - A new automated microscopy image-based urine sediment analyser. Clinica Chimica Acta 2010;411: Clin. Lab. 4/2014

5 URINALYSIS: MICROSCOPIC AND AUTOMATED ANALYSIS 18. Jiang T, Chen P, Ouyang J, Zhang S, Cai D. Urine particles analysis: performance evaluation of Sysmex UF-1000i and comparison among urine flow cytometer, dipstick, and visual microscopic examination. Scand J Clin Lab Invest 2011;71(1): Manoni F, Tinello A, Fornasiero L, et al. Urine particle evaluation: a comparison between the UF-1000i and quantitative microscopy. Clin Chem Lab Med 2010;48(8): Manoni F, Gessoni G, Alessio MG, et al. Mid-stream vs. firstvoided urine collection by using automated analyzers for particle examination in healthy subjects: an Italian multicenter study. Clin Chem Lab Med 2011;50(4): Correspondence: Paula V. Bottini MD, PhD Divisão de Patologia Clínica/HC - UNICAMP Rua Vital Brasil 251- Cidade Universitária Zeferino Vaz CEP: , Campinas, S.P., Brasil Tel.: Fax: paula@hc.unicamp Clin. Lab. 4/

Cornelia Ottiger and Andreas R. Huber * Analytical Techniques. Clinical Chemistry 49: (2003) Automation and

Cornelia Ottiger and Andreas R. Huber * Analytical Techniques. Clinical Chemistry 49: (2003) Automation and Clinical Chemistry 49:4 617 623 (2003) Automation and Analytical Techniques Quantitative Urine Particle Analysis: Integrative Approach for the Optimal Combination of Automation with UF-100 and Microscopic

More information

True Walkaway Automated Urinalysis, Chemistry to Microscopy The CLINITEK AUWi System Answers for life.

True Walkaway Automated Urinalysis, Chemistry to Microscopy The CLINITEK AUWi System Answers for life. www.siemens.com/diagnostics True Walkaway Automated Urinalysis, Chemistry to Microscopy The CLINITEK AUWi System Answers for life. Can I achieve true unattended urinalysis testing in my laboratory? With

More information

Comparison between IRIS iq 200ELITE and microscopy for urinalysis and evaluation of performance in predicting outcome of urine cultures

Comparison between IRIS iq 200ELITE and microscopy for urinalysis and evaluation of performance in predicting outcome of urine cultures Article Reprint Comparison between IRIS iq 200ELITE and microscopy for urinalysis and evaluation of performance in predicting outcome of urine cultures S. Ledru, J.P. Canonne Laboratoire de microbiologie,

More information

Meets Standards AFTER REVIEW ( 2.5 pts) Fail to meet the standard safety precautions, counseled one time

Meets Standards AFTER REVIEW ( 2.5 pts) Fail to meet the standard safety precautions, counseled one time Morgan State University 2014 MDTC 411 CLINICAL CHEMISTRY AND URINALYSIS PRACTICUM SUGGESTED LABORATORY ROTATION SCORING RUBRIC Please attach the student s test record to this evaluation Clinical Chemistry

More information

Verification or validation, that is the question

Verification or validation, that is the question Perspective Page 1 of 7 Verification or validation, that is the question Giorgia Antonelli 1, Andrea Padoan 1, Ada Aita 1, Laura Sciacovelli 2, Mario Plebani 1,2 1 U.O.C. of Laboratory Medicine, Department

More information

Pushing limits in routine laboratory haematology with the XT-4000i

Pushing limits in routine laboratory haematology with the XT-4000i Pushing limits in routine laboratory haematology with the XT-4000i Sysmex Xtra Online December 2013 Fig. 1 The XT-4000i haematology system With the launch of the XE-5000, Sysmex set new standards in haematology.

More information

Automated Hematology Analyzer. Small Footprint. Big Difference.

Automated Hematology Analyzer. Small Footprint. Big Difference. XS-1000i Automated Hematology Analyzer Small Footprint. Big Difference. Small But Powerful: The XS-1000i Even with these challenges, the need for hematology testing has remained steady or continues to

More information

CONCURRENT 10: THE COMPLETE BLOOD COUNT AND BEYOND: QUALITY ISSUES AND REFERENCE METHODS

CONCURRENT 10: THE COMPLETE BLOOD COUNT AND BEYOND: QUALITY ISSUES AND REFERENCE METHODS CONCURRENT 10: THE COMPLETE BLOOD COUNT AND BEYOND: QUALITY ISSUES AND REFERENCE METHODS Saturday May 14 th 2016 1:30 3:00 PM (13.30 15.00hrs) International Society for Laboratory Hematology (ISLH) CONCURRENT

More information

Automated Hematology System. Your Complete Choice

Automated Hematology System. Your Complete Choice XE-5000 Automated Hematology System Your Complete Choice Advanced Technology Solutions to Meet Your Lab s Needs Even with these challenges, the need for hematology testing has remained steady or continued

More information

Automated Hematology Analyzer. Differentiate with the XT-4000i

Automated Hematology Analyzer. Differentiate with the XT-4000i XT-4000i Automated Hematology Analyzer Differentiate with the XT-4000i Advanced Technology Solutions to Meet Your Lab s Needs Even with these challenges, the need for hematology testing has remained steady

More information

BEYOND A BETTER BOX DI-60 INTEGRATED SLIDE PROCESSING SYSTEM. Seamless Integration

BEYOND A BETTER BOX DI-60 INTEGRATED SLIDE PROCESSING SYSTEM. Seamless Integration BEYOND A BETTER BOX DI-60 INTEGRATED SLIDE PROCESSING SYSTEM Seamless Integration RESHAPING HEMATOLOGY Scalable Automation and On-board Slide Processing The Sysmex DI-60 Integrated Cell Image Analyzer

More information

SAMPLE. Urinalysis; Approved Guideline Third Edition

SAMPLE. Urinalysis; Approved Guideline Third Edition February 2009 Urinalysis; Approved Guideline Third Edition This document addresses procedures for testing urine, including materials and equipment; macroscopic/physical evaluation; chemical analysis; and

More information

Evaluation of the Automated Immature Granulocyte Count (IG) on Sysmex XE-2100 Automated Haematology Analyser vs. Visual Microscopy (NCCLS H20-A)

Evaluation of the Automated Immature Granulocyte Count (IG) on Sysmex XE-2100 Automated Haematology Analyser vs. Visual Microscopy (NCCLS H20-A) Evaluation of the Automated Immature Granulocyte Count (IG) on Sysmex XE-21 Automated Haematology Analyser vs. Visual Microscopy (NCCLS H2-A) Th WEILAND *1, H KALKMAN *2, and H HEIHN *1 *1 Central Laboratory,

More information

BEYOND A BETTER BOX XN-V SERIES TM MULTISPECIES HEMATOLOGY ANALYZERS. Accelerate Your Multispecies Research Laboratory XN-V IS FOR ANIMAL USE ONLY

BEYOND A BETTER BOX XN-V SERIES TM MULTISPECIES HEMATOLOGY ANALYZERS. Accelerate Your Multispecies Research Laboratory XN-V IS FOR ANIMAL USE ONLY BEYOND A BETTER BOX XN-V SERIES TM MULTISPECIES HEMATOLOGY ANALYZERS Accelerate Your Multispecies Research Laboratory XN-V IS FOR ANIMAL USE ONLY MOVE YOUR MULTI-SPECIES LABORATORY BEYOND A BETTER BOX*

More information

Immature Granulocyte Enumeration Our Journey from Manual to Automated Reporting

Immature Granulocyte Enumeration Our Journey from Manual to Automated Reporting Immature Granulocyte Enumeration Our Journey from Manual to Automated Reporting Mary Capper MT(ASCP)SH Supervisor, Hematology/Hemostasis University of Iowa Hospitals and Clinics Objectives Describe the

More information

SAMPLE. Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard Second Edition

SAMPLE. Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard Second Edition June 2010 Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard Second Edition This document provides guidance for the validation, verification, calibration,

More information

MOLECULAR TESTING: VERIFYING/VALIDATING INSTRUMENTS, REAGENTS AND ASSAYS. Richard L. Hodinka, Ph.D.

MOLECULAR TESTING: VERIFYING/VALIDATING INSTRUMENTS, REAGENTS AND ASSAYS. Richard L. Hodinka, Ph.D. MOLECULAR TESTING: VERIFYING/VALIDATING INSTRUMENTS, REAGENTS AND ASSAYS Richard L. Hodinka, Ph.D. University of South Carolina School of Medicine Greenville Greenville Health System, Greenville, SC hodinka@greenvillemed.sc.edu

More information

Guide to Fulfillment of Validation and Verification of Examination Requirements

Guide to Fulfillment of Validation and Verification of Examination Requirements DIAGNOSTIC ACCREDITATION PROGRAM College of Physicians and Surgeons of British Columbia 300 669 Howe Street Telephone: 604-733-7758 ext. 2635 Vancouver BC V6C 0B4 Toll Free: 1-800-461-3008 (in BC) www.cpsbc.ca

More information

ICSH guidelines for the verification and performance of automated cell counters for body fluids

ICSH guidelines for the verification and performance of automated cell counters for body fluids International Journal of Laboratory Hematology ORIGINAL ARTICLE The Official journal of the International Society for Laboratory Hematology INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY ICSH guidelines

More information

Application Note for Gentian Cystatin C Immunoassay on. ABX Pentra 400 1, Horiba Medical 1 v03en-june 2011

Application Note for Gentian Cystatin C Immunoassay on. ABX Pentra 400 1, Horiba Medical 1 v03en-june 2011 Page 1 of 5 Application Note for Gentian Cystatin C Immunoassay on For in vitro diagnostic use only ABX Pentra 400 1, Horiba Medical 1 v03en-june 2011 Intended Use The Gentian Cystatin C Immunoassay is

More information

MANAGEMENT BRIEF. How Automation is Revolutionizing White Cell Differential Analysis.

MANAGEMENT BRIEF. How Automation is Revolutionizing White Cell Differential Analysis. MANAGEMENT BRIEF How Automation is Revolutionizing White Cell Differential Analysis Introduction Confidence building becomes a real issue when a laboratory technologist generalist must classify abnormal

More information

AUTOMATED HEMATOLOGY IN THE CLINICAL LAB

AUTOMATED HEMATOLOGY IN THE CLINICAL LAB AUTOMATED HEMATOLOGY IN THE CLINICAL LAB Diane Gaspari, SH(ASCP) Core Lab Manager York Hospital, York PA ggaspari@wellspan.org Objectives of Automation Increase efficiency; streamline workflow Increase

More information

M-TP PRINCIPLE REF ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

M-TP PRINCIPLE REF ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Microprotein REF 445860 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE INTENDED

More information

Clinical Chemistry Approach to Evaluation of Commutability

Clinical Chemistry Approach to Evaluation of Commutability Clinical Chemistry Approach to Evaluation of Commutability Hubert W. Vesper, Ph.D. Director, Clinical Standardization Programs Division of Laboratory Sciences Centers for Disease Control and Prevention,

More information

S1725 Method Validation and Verification: Case Studies and Laboratory Challenges

S1725 Method Validation and Verification: Case Studies and Laboratory Challenges S1725 Method Validation and Verification: Case Studies and Laboratory Challenges James H. Nichols, PhD, D(ABCC) Lauren N. Person, DO, MPH, FCAP Objectives Identify the difference between method validation

More information

Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer

Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer Journal of Automated Methods & Management in Chemistry, 5 (5), no. 4, 35 39 c 5 Hindawi Publishing Corporation Validation of Platelet ing Accuracy With the Celltac F Automated Hematology Analyzer Yutaka

More information

SAMPLE. Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline Third Edition

SAMPLE. Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline Third Edition October 2010 Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline Third Edition This document contains guidelines for determining reference values and

More information

Performance Evaluation of the Coulter LH 750 Hematology Analyzer

Performance Evaluation of the Coulter LH 750 Hematology Analyzer ISLH Laboratory Hematology 7:217 228 2001 Carden Jennings Publishing Co., Ltd. Official Publication Performance Evaluation of the Coulter LH 750 Hematology Analyzer TRACEY FERNANDEZ, LYNN BESSERT DOMACK,

More information

MORE SAMPLES, LESS MICROSCOPE

MORE SAMPLES, LESS MICROSCOPE MORE SAMPLES, LESS MICROSCOPE WE HAVE A SOLUTION TO YOUR NEEDS IN URINALYSIS IRIS: MOVING URINALYSIS FORWARD WE HAVE A SOLUTION TO YOUR NEEDS We can help solve your problems 1. Improved Workflow and Productivity

More information

Greater Possibilities By Design

Greater Possibilities By Design Greater Possibilities By Design January 23, 2013 Nilam Patel MT(ASCP)SH Sr. Product Manager, Automation Solutions Objectives Review current and new portfolio Technology and benefits Operational benefits

More information

SAMPLE. Risk Management Techniques to Identify and Control Laboratory Error Sources; Approved Guideline Second Edition

SAMPLE. Risk Management Techniques to Identify and Control Laboratory Error Sources; Approved Guideline Second Edition November 2009 Risk Management Techniques to Identify and Control Laboratory Error Sources; Approved Guideline Second Edition This guideline describes risk management techniques that will aid in identifying,

More information

Ready, Set, Test! AACC Conference Mass Spectrometry in the Clinical Lab: Best Practice and Current Applications September 17-18, 2013 St.

Ready, Set, Test! AACC Conference Mass Spectrometry in the Clinical Lab: Best Practice and Current Applications September 17-18, 2013 St. Ready, Set, Test! Ross Molinaro, PhD, MLS(ASCP) CM, DABCC, FACB Medical Director, Clinical Laboratories Emory University Hospital Midtown Emory Clinical Translational Research Laboratory AACC Conference

More information

WAIT! Ready, Set, Test! Financial Disclosure. Research/Educational grants/consulting/salary support

WAIT! Ready, Set, Test! Financial Disclosure. Research/Educational grants/consulting/salary support Ready, Set, Test! Ross Molinaro, PhD, MLS(ASCP) CM, DABCC, FACB Medical Director, Clinical Laboratories Emory University Hospital Midtown Emory Clinical Translational Research Laboratory AACC Conference

More information

Optimizing an internal quality control program of automated ANA IIF

Optimizing an internal quality control program of automated ANA IIF Optimizing an internal quality control program of automated ANA IIF Optimizing an iqc program of automated ANA IIF Laura Bogaert Supervisor: Apr. Biol. S. Schouwers & Apr. Biol. L. Van Hoovels Introduction

More information

9/7/17. Implementing the ARK Immunoassay for Therapeutic Drug Monitoring of High-Dose Methotrexate. Disclosure. Learning Objectives

9/7/17. Implementing the ARK Immunoassay for Therapeutic Drug Monitoring of High-Dose Methotrexate. Disclosure. Learning Objectives Implementing the ARK Immunoassay for Therapeutic Drug Monitoring of High-Dose Methotrexate Kristine R. Crews, PharmD, FCCP, BCPS St. Jude Children s Research Hospital September 14, 2017 Disclosure I declare

More information

5 Tips for Better Cell Counting

5 Tips for Better Cell Counting 5 Tips for Better Cell Counting Counting cells is a tedious job that needs to be done properly, be it in industry or academia, be it manually or automated, you want trustworthy results to continue your

More information

Date. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver and/or kidneys.

Date. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver and/or kidneys. SYNCHRON System(s) Chemistry Information Sheet Copyright 2013 Beckman Coulter, Inc. Albumin REF 442765 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by Date Reviewed by Date PRINCIPLE INTENDED USE

More information

BEYOND A BETTER BOX XN-SERIES TM AUTOMATED HEMATOLOGY SYSTEMS. Reshaping Compact Automation

BEYOND A BETTER BOX XN-SERIES TM AUTOMATED HEMATOLOGY SYSTEMS. Reshaping Compact Automation BEYOND A BETTER BOX XN-SERIES TM AUTOMATED HEMATOLOGY SYSTEMS Reshaping Compact Automation SMALL, SMART, COMPACT AUTOMATION It s an approach to hematology testing that s never been done before. An approach

More information

CLSI C60: Assay Validation & Post-Validation Monitoring

CLSI C60: Assay Validation & Post-Validation Monitoring CLSI C60: Assay Validation & Post-Validation Monitoring Ross J. Molinaro, MT(ASCP), PhD, DABCC, FACB Medical Director Core Laboratory, Emory University Hospital Midtown Emory Clinical Translational Research

More information

Upon completion of the Clinical Hematology rotation, the MLS student will be able to:

Upon completion of the Clinical Hematology rotation, the MLS student will be able to: Clinical Performance Objectives in Clinical Hematology Department of Medical and Research Technology University of Maryland School of Medicine Spring 2015 Upon completion of the Clinical Hematology rotation,

More information

XN-SERIES AUTOMATED HEMATOLOGY SYSTEMS

XN-SERIES AUTOMATED HEMATOLOGY SYSTEMS XN-SERIES AUTOMATED HEMATOLOGY SYSTEMS Reshaping Compact Automation BEYOND A BETTER BOX MOVE YOUR HEMATOLOGY LABORATORY BEYOND A BETTER BOX It s what makes Sysmex the smarter choice and sets our integrated

More information

AAG PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

AAG PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE IMMAGE Immunochemistry Systems Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Alpha 1 -Acid Glycoprotein REF 447780 (150 tests) For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by:

More information

Quality Control in Flow. Dr David Westerman Head of Haematopathology Peter MacCallum Cancer Centre

Quality Control in Flow. Dr David Westerman Head of Haematopathology Peter MacCallum Cancer Centre Quality Control in Flow Dr David Westerman Head of Haematopathology Peter MacCallum Cancer Centre Aims Quality Assurance Quality Control Literature In house competencies SHOT DATA 1996-2009 Ref: SHOT Annual

More information

XN-9100 AUTOMATED HEMATOLOGY SYSTEMS

XN-9100 AUTOMATED HEMATOLOGY SYSTEMS XN-9100 AUTOMATED HEMATOLOGY SYSTEMS Reshaping Scalable Automation BEYOND A BETTER BOX MOVE YOUR HEMATOLOGY LABORATORY BEYOND A BETTER BOX It s what makes Sysmex the smarter choice and sets our integrated

More information

TELE-HEMATOLOGY: CRITICAL DETERMINANTS FOR SUCCESSFUL IMPLEMENTATION

TELE-HEMATOLOGY: CRITICAL DETERMINANTS FOR SUCCESSFUL IMPLEMENTATION Blood First Edition From Paper, www.bloodjournal.org prepublished by online guest on September July 4, 2018. 11, For 2003; personal DOI use 10.1182/blood-2003-05-1615 only. TELE-HEMATOLOGY: CRITICAL DETERMINANTS

More information

COMPARISON OF PLATELET COUNTS BY SYSMEX XE 2100 AND LH 750 WITH THE INTERNATIONAL FLOW REFERENCE METHOD IN THROMBOCYTOPENIC PATIENTS

COMPARISON OF PLATELET COUNTS BY SYSMEX XE 2100 AND LH 750 WITH THE INTERNATIONAL FLOW REFERENCE METHOD IN THROMBOCYTOPENIC PATIENTS COMPARISON OF PLATELET COUNTS BY SYSMEX XE 2100 AND LH 750 WITH THE INTERNATIONAL FLOW REFERENCE METHOD IN THROMBOCYTOPENIC PATIENTS TINA DADU 1, KUNAL SEHGAL 2, ANJUM SHAIKH 2, SHANAZ KHODAIJI 2 1. BLK

More information

SF Cube. Exclusive Technology, Inclusive Approach. BC-6800 Auto Hematology Analyzer. Cell Analysis Technology

SF Cube. Exclusive Technology, Inclusive Approach. BC-6800 Auto Hematology Analyzer. Cell Analysis Technology Cell Analysis Technology Exclusive Technology, Inclusive Approach BC-6800 Auto Hematology Analyzer Cell Analysis Technology SF Cube SF Cube is a pathbreaking technology for reliable blood cell analysis,

More information

Examples of body fluid analysis with the XE-5000

Examples of body fluid analysis with the XE-5000 Examples of body fluid analysis with the XE-5000 Sysmex Xtra Online March 2011 If patients are to be treated rapidly and effectively, laboratory results must be both rapid and reliable if at all possible,

More information

APB PRINCIPLE SPECIMEN. REF (300 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

APB PRINCIPLE SPECIMEN. REF (300 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE IMMAGE Immunochemistry Systems Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Apolipoprotein B REF 447730 (300 tests) For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed

More information

Mouse Collagen Type III ELISA

Mouse Collagen Type III ELISA Mouse Collagen Type III ELISA Catalog Number M046062 For the quantitative determination of Collagen Type III in mouse plasma and serum samples. For research use only. This product insert must be read in

More information

Screening of urine samples by flow cytometry reduces the need for culture

Screening of urine samples by flow cytometry reduces the need for culture JCM Accepts, published online ahead of print on 30 June 2010 J. Clin. Microbiol. doi:10.1128/jcm.00617-10 Copyright 2010, American Society for Microbiology and/or the Listed Authors/Institutions. All Rights

More information

Mouse OVA IgG1 ELISA. Catalog Number M046072

Mouse OVA IgG1 ELISA. Catalog Number M046072 Mouse OVA IgG1 ELISA Catalog Number M046072 For the quantitative determination of Ovalbumin specific IgG1 in mouse plasma, serum, tissue and cell culture supernate samples. For research use only. This

More information

Molecular Diagnosis Challenges & Solutions. Using Molecular Kits or Laboratory Developed Tests (Home Brew), Emphasis on Validation

Molecular Diagnosis Challenges & Solutions. Using Molecular Kits or Laboratory Developed Tests (Home Brew), Emphasis on Validation Using Molecular Kits or Laboratory Developed Tests (Home Brew), Emphasis on Validation Molecular Diagnosis Challenges & Solutions Behzad Poopak, DCLS PhD Tehran Medical Branch- Islamic Azad University

More information

Urine Protein Electrophoresis and Immunoelectrophoresis Using Unconcentrated or Minimally Concentrated Urine Samples

Urine Protein Electrophoresis and Immunoelectrophoresis Using Unconcentrated or Minimally Concentrated Urine Samples Immunopathology / Electrophoresis of Unconcentrated Urine Samples Urine Protein Electrophoresis and Immunoelectrophoresis Using Unconcentrated or Minimally Concentrated Urine Samples Anja C. Roden, MD,

More information

BC Auto Hematology Analyzer. A CUTE 5-part

BC Auto Hematology Analyzer. A CUTE 5-part BC-5000 Auto Hematology Analyzer A CUTE 5-part Why do we need 5-part hematology analyzers? WBC differential: 3-part WBC differential: 5-part Lymphocyte Mid-size cell? Eosinophil Basophil Neutrophil DIFF

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Department of Haematology Pathology Centre C37 New Cross Hospital Wolverhampton Road Wolverhampton WV10 0QP United Kingdom Contact: Katy New

More information

The Future of Hematological Diagnosis

The Future of Hematological Diagnosis Interview The Future of Hematological Diagnosis Bruce H. Davis, M.D. Bruce H. Davis, M.D. is the Director of Analytical Cytometry in the Department of Clinical Pathology at the William Beaumont Hospital

More information

Prepared by Date Adopted Supersedes Procedure # Adapted from HPTN Policy. Review Date Revision Date Signature

Prepared by Date Adopted Supersedes Procedure # Adapted from HPTN Policy. Review Date Revision Date Signature APPENDIX IV: LABORATORY QUALITY CONTROL POLICY Prepared by Date Adopted Supersedes Procedure # Adapted from HPTN Policy N/A Review Date Revision Date Signature Distributed to # of Copies Distributed to

More information

BC Auto Hematology Analyzer. A CUTE 5-part

BC Auto Hematology Analyzer. A CUTE 5-part BC-5150 Auto Hematology Analyzer A CUTE 5-part Why do we need 5-part hematology analyzers? WBC differential: 3-part WBC differential: 5-part Lymphocyte Mid-size cell? Eosinophil Basophil Neutrophil DIFF

More information

Reshaping compact automation

Reshaping compact automation Automated Hematology Analyzer XN-SERIES Reshaping compact automation www.sysmex.com/us Optimize your laboratory As analyzers become increasingly sophisticated, the laboratory s contribution to the continuum

More information

METHOD VALIDATION TECHNIQUES PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER 2012

METHOD VALIDATION TECHNIQUES PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER 2012 METHOD VALIDATION PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE TECHNIQUES OCTOBER/NOVEMBER 2012 Prepared by for ENAO Assessor Calibration B SCOPE Introduction House Rules Central Tendency Statistics Population

More information

Reshaping compact automation

Reshaping compact automation Automated Hematology Analyzer XN-SERIES Reshaping compact automation www.sysmex.com/us Optimize your laboratory As analyzers become increasingly sophisticated, the laboratory s contribution to the continuum

More information

Detection of Significant Bacteriuria. by Use of the iq200 Automated Urine Microscope

Detection of Significant Bacteriuria. by Use of the iq200 Automated Urine Microscope JCM Accepts, published online ahead of print on 28 May 2014 J. Clin. Microbiol. doi:10.1128/jcm.00112-14 Copyright 2014, American Society for Microbiology. All Rights Reserved. 1 2 3 4 5 6 7 8 9 10 11

More information

SAMPLE. Evaluation of Total Analytical Error for Quantitative Medical Laboratory Measurement Procedures

SAMPLE. Evaluation of Total Analytical Error for Quantitative Medical Laboratory Measurement Procedures 2nd Edition EP21 Evaluation of Total Analytical Error for Quantitative Medical Laboratory Measurement Procedures This guideline provides manufacturers and end users with an understanding of concepts related

More information

Novel POC analysis for determination of total and 5-part differential WBC count among a US population, in comparison to Beckman Coulter LH750

Novel POC analysis for determination of total and 5-part differential WBC count among a US population, in comparison to Beckman Coulter LH750 Novel POC analysis for determination of total and -part differential WBC count among a US population, in comparison to Beckman Coulter LH7 Reed, Jeremy*, Johnsson, Elin**, Hockum, Sandra** * Quest Diagnostics,

More information

Cellometer. for Cell Counting & Analysis. Brewing Yeast Wine Yeast Platelets and Other Small Cells. Image Cytometer

Cellometer. for Cell Counting & Analysis. Brewing Yeast Wine Yeast Platelets and Other Small Cells. Image Cytometer Cellometer 2 Image Cytometer for Cell Counting & Analysis Brewing Yeast Wine Yeast and Other Small Cells Cellometer 2 Image Cytometer Optimized Analysis for Yeast and other Small Cells Features of the

More information

BIOEQUIVALENCE TRIAL INFORMATION FORM (Medicines and Allied Substances Act [No. 3] of 2013 Part V Section 39)

BIOEQUIVALENCE TRIAL INFORMATION FORM (Medicines and Allied Substances Act [No. 3] of 2013 Part V Section 39) ZAMRA BTIF BIOEQUIVALENCE TRIAL INFORMATION FORM (Medicines and Allied Substances Act [No. 3] of 2013 Part V Section 39) The Guidelines on Bioequivalence Studies to be consulted in completing this form.

More information

ABX Pentra XL 80. Hematology analyzer. Auto-sampling 5-part Differential Customized Dilution Ratio Integrated Validation Station

ABX Pentra XL 80. Hematology analyzer. Auto-sampling 5-part Differential Customized Dilution Ratio Integrated Validation Station ABX Pentra XL 80 Hematology analyzer Auto-sampling 5-part Differential Customized Dilution Ratio Integrated Validation Station Pentra XL 80 A new way to manage your hematology workflow With Pentra XL 80

More information

Designing QC Activities to More Precisely Manage Analytical Accuracy and Patient Risk

Designing QC Activities to More Precisely Manage Analytical Accuracy and Patient Risk Designing QC Activities to More Precisely Manage Analytical Accuracy and Patient Risk Curtis A. Parvin, Ph.D. Manager of Advanced Statistical Research Bio-Rad Laboratories, QSD curtis_parvin@bio-rad.com

More information

Improving the efficiency and efficacy of pre-analytical and analytical work-flow of urine cultures with urinary flow cytometry

Improving the efficiency and efficacy of pre-analytical and analytical work-flow of urine cultures with urinary flow cytometry NEW MICROBIOLOGICA, 31, 501-505, 2008 Improving the efficiency and efficacy of pre-analytical and analytical work-flow of urine cultures with urinary flow cytometry Shamanta Grosso, Graziano Bruschetta,

More information

Test Method Validation of In-Vitro Diagnostic Products. Tim Carr QA Manager, Beckman Coulter

Test Method Validation of In-Vitro Diagnostic Products. Tim Carr QA Manager, Beckman Coulter Test Method Validation of In-Vitro Diagnostic Products Tim Carr QA Manager, Beckman Coulter IVD Overview What is an In-Vitro Diagnostic Test IVDs are medical devices and accessories Used to perform tests

More information

School of Allied Medical Sciences. Course Description Guide Associate in Medical Laboratory Sciences. Page 1 of 15

School of Allied Medical Sciences. Course Description Guide Associate in Medical Laboratory Sciences. Page 1 of 15 School of Allied Medical Sciences Course Description Guide Associate in Medical Laboratory Sciences Page 1 of 15 Program Name & Definition: Associate in Medical Laboratory Sciences Laboratory sciences

More information

ABX Pentra DX120. Hematology analyzer. 45 parameters Integrated slide-maker Expert validation system

ABX Pentra DX120. Hematology analyzer. 45 parameters Integrated slide-maker Expert validation system ABX Pentra DX120 Hematology analyzer 45 parameters Integrated slide-maker Expert validation system ABX Pentra DX 120 In the heart of hematopoiesis Complete cytology platform 120 samples per hour Continuous

More information

Image Analysis. Tools for basic and translational research. visualize analyze interpret report archive customize

Image Analysis. Tools for basic and translational research. visualize analyze interpret report archive customize Image Analysis Tools for basic and translational research visualize analyze interpret report archive customize Unlock the benefits of image analysis. Image analysis is an integrated part of Aperio s digital

More information

Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer

Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer Automated Methods & Management in Chemistry, 25 (25), no. 4, 235 239 c 25 Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer Yutaka Nagai, 1 Hiroshi Kondo, 2 and

More information

Starrsed RS. Gold Standard ESR

Starrsed RS. Gold Standard ESR Starrsed RS Gold Standard ESR Starrsed RS Does your busy laboratory perform up to 135 erythrocyte sedimentation rate (ESR) tests per hour and are you looking to automate this process? Or did you already

More information

Highly Accurate, Reliable and Real-time Enumeration of Individual Bacteria for the Determination of Raw milk Hygienic Quality

Highly Accurate, Reliable and Real-time Enumeration of Individual Bacteria for the Determination of Raw milk Hygienic Quality Highly Accurate, Reliable and Real-time Enumeration of Individual Bacteria for the Determination of Raw milk Hygienic Quality Validated to be at least equivalent to the EN-ISO 4833-1:2013 and 4833-2:2013

More information

GOOD LABORATORY PRACTICE Agilent Cary 8454 UV-Visible Spectroscopy System

GOOD LABORATORY PRACTICE Agilent Cary 8454 UV-Visible Spectroscopy System GOOD LABORATORY PRACTICE Agilent Cary 8454 UV-Visible Spectroscopy System Introduction The objective of good laboratory practice (GLP) is to obtain accurate and precise results, using a measurement system

More information

ACCURACY THE FIRST TIME. Bring Your Esoteric Assays In-House Routinely SCIEX DIAGNOSTICS MASS SPECTROMETERS

ACCURACY THE FIRST TIME. Bring Your Esoteric Assays In-House Routinely SCIEX DIAGNOSTICS MASS SPECTROMETERS ACCURACY THE FIRST TIME Bring Your Esoteric Assays In-House Routinely SCIEX DIAGNOSTICS MASS SPECTROMETERS Your routine diagnostic testing lab requires accurate trace level quantitation of multiple compounds

More information

Carl Zeiss MicroImaging AIS. Digital Pathology The Future. Andrew Lesniak Director, Product Management

Carl Zeiss MicroImaging AIS. Digital Pathology The Future. Andrew Lesniak Director, Product Management Carl Zeiss MicroImaging AIS Digital Pathology The Future Andrew Lesniak Director, Product Management Defined Industry Needs Computer Aided Imaging and Analysis will play a key role in the increase of Diagnostics

More information

XN Series. What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems Sysmex America, Inc. All rights reserved.

XN Series. What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems Sysmex America, Inc. All rights reserved. XN Series What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems 2012 Sysmex America, Inc. All rights reserved. Agenda XN Series Product Overview A system to fit every lab

More information

LabGuide 13 How to Verify Performance Specifications

LabGuide 13 How to Verify Performance Specifications 1 LG 13 Comments? Feedback? Questions? Email us at info@cola.org or call us at 800 981-9883 LabGuide 13 How to Verify Performance Specifications The verification of performance specifications confirms

More information

Date. Alkaline phosphatase measurements are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.

Date. Alkaline phosphatase measurements are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases. SYNCHRON CX System(s) Chemistry Information Sheet Copyright 2008 Beckman Coulter, Inc. Alkaline Phosphatase Kit Reorder # 442670 (200 tests/cartridge) Kit Reorder # 476821 (400 tests/cartridge) For In

More information

PRINCIPLE SPECIMEN. REF (300 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

PRINCIPLE SPECIMEN. REF (300 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE IMMAGE Immunochemistry Systems Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Complement REF 446450 (300 tests) For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by:

More information

An Automated Method for the Determination of Intestinal Disaccharidase and Glucoamylase Activities

An Automated Method for the Determination of Intestinal Disaccharidase and Glucoamylase Activities Automated Methods and Management in Chemistry Volume 2006, Article ID 93947, Pages 1 4 DOI 10.1155/JAMMC/2006/93947 An Automated Method for the Determination of Intestinal Disaccharidase and Glucoamylase

More information

Measure-iT Thiol Assay Kit (M30550)

Measure-iT Thiol Assay Kit (M30550) Product Information Revised: 24 March 2008 Measure-iT Thiol Assay Kit (M30550) Quick Facts Storage upon receipt: Store at 20ºC Protect from light Desiccate Avoid freeze/thaw cycles thiol quantitation buffer

More information

Total Analytic Error From Concept to Application

Total Analytic Error From Concept to Application Página 1 de 5 Clinical Laboratory News Subscribe CLN Daily CLN Stat CLN Articles Total Analytic Error From Concept to Application Author: James O. Westgard, PhD, and Sten A. Westgard, MS // Date: SEP.1.2013

More information

Cellometer Auto 2000 Cell Viability Counter Optimized Analysis of Primary Cells Cellometer Features of the Cellometer Auto 2000

Cellometer Auto 2000 Cell Viability Counter Optimized Analysis of Primary Cells Cellometer Features of the Cellometer Auto 2000 Cellometer 2000 Cell Counter Optimized Analysis of Primary Features of the Cellometer 2000 Dual Fluorescence and Bright Field Imaging: staining of both live and dead cells in heterogeneous samples The

More information

CR-E PRINCIPLE REF A ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

CR-E PRINCIPLE REF A ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Creatinine REF A60298 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE INTENDED

More information

BEYOND A BETTER BOX XN-9000 AUTOMATED HEMATOLOGY SYSTEMS. Reshaping Scalable Automation

BEYOND A BETTER BOX XN-9000 AUTOMATED HEMATOLOGY SYSTEMS. Reshaping Scalable Automation BEYOND A BETTER BOX XN-9000 AUTOMATED HEMATOLOGY SYSTEMS Reshaping Scalable Automation 2 Sysmex XN-Series Automated Hematology Analyzers are designed to exceed the expectations of real people who work

More information

Successful performance in proficiency testing (PT) is required

Successful performance in proficiency testing (PT) is required Performance Characteristics of Several Rules for Self-interpretation of Proficiency Testing Data R. Neill Carey, PhD; George S. Cembrowski, MD; Carl C. Garber, PhD; Zohreh Zaki, MD Context. Proficiency

More information

The utility of qnmr to improve accuracy and precision of LC-MS bioanalysis

The utility of qnmr to improve accuracy and precision of LC-MS bioanalysis JOURNAL OF APPLIED BIOANALYSIS, January 2018, p. 26-31. http://dx.doi.org/10.17145/jab.18.005 (ISSN 2405-710X) Vol. 4, No. 1 METHODOLOGY The utility of qnmr to improve accuracy and precision of LC-MS bioanalysis

More information

HST-330 TM and Total Laboratory Automation

HST-330 TM and Total Laboratory Automation HST-330 TM and Total Laboratory Automation Hyun-Sook CHI, MD* * Department of Clinical Pathology, University of Ulsan College of Medicine and Asan Medical Center, Seoul, Korea. Total laboratory automation

More information

YOU Can Make a Difference!

YOU Can Make a Difference! YOU Can Make a Difference! Lucia M. Berte Laboratories Made Better! www.laboratoriesmadebetter.com 1 What is Quality? Quality is doing the right things and doing those things right. Philip Crosby 2 Do

More information

Automation In Life Sciences and Diagnostics Applications

Automation In Life Sciences and Diagnostics Applications Automation In Life Sciences and Diagnostics Applications Pedro Diaz Director of Research Beckman Coulter Life Sciences February 20-22, 2013 Orlando World Marriott Center Orlando, Florida USA Beckman Coulter

More information

WADA Technical Document TD2019BAR. Blood Analytical Requirements for the Athlete Biological Passport

WADA Technical Document TD2019BAR. Blood Analytical Requirements for the Athlete Biological Passport 1. Introduction Blood Analytical Requirements for the Athlete Biological Passport This Technical Document (TD) has been established to harmonize the analysis of blood Samples collected, both In-Competition

More information

Validation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals.

Validation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals. Validation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals. 1 Analytical Method Validation: 1..1 Philosophy: Method validation is

More information

The Journey Towards Standardization in Validation of Fluorescent Cell-Based Assays

The Journey Towards Standardization in Validation of Fluorescent Cell-Based Assays The Journey Towards Standardization in Validation of Fluorescent Cell-Based Assays Teri Oldaker, CLS, Ccy Genoptix Medical Laboratory Flow Cytometry Workshop ISCT 2015, Las Vegas, NV Objectives Background

More information

Critical appraisal checklist for papers on biological variation

Critical appraisal checklist for papers on biological variation Biological Variation Working Group Critical appraisal checklist for papers on biological variation Dr Bill Bartlett Biological Variation Working Group EFLM Biological Variation Working Group Member These

More information

Fibrinogen Universal Kit

Fibrinogen Universal Kit Order references Reagents CONT FITUR-B00 Universal kit 1 x 50 ml R1 + 1 x 5 ml R2 FITUR-00 Universal kit 2 x 50 ml R1 + 1 x 15 ml R2 FITUR-C00 Universal kit 2 x 17 ml R1 + 1 x 4 ml R2 Other necessary products

More information