Laboratory Quality Control Requirements Should be Based on Risk Management Principles

Size: px
Start display at page:

Download "Laboratory Quality Control Requirements Should be Based on Risk Management Principles"

Transcription

1 Received Revisions Received Accepted Laboratory Quality Control Requirements Should be Based on Risk Management Principles Donald M. Powers, PhD (Powers Consulting Group, Pittsford, NY) DOI: /GBU6UH7Q3TFVPLJ7 Industry experience has demonstrated the value of a formal risk management process in the context of a quality system. It supports systematic, informed decision making, provides greater insight into patient safety risks, fosters quality by design and continuous improvement, and leads to greater predictability of results. It is time to question why manufacturers and laboratories are following such divergent paths to the same end. Risk management should offer clinical laboratories the same advantages it offers manufacturers. Proposed initiatives by the Centers for Medicare and Medicaid Services (CMS) would allow clinical laboratories to reduce quality control testing based on the extent of internal monitoring performed by a commercial laboratory analytical system. 1 To qualify for reduced testing, the proposal requires a laboratory to demonstrate that these internal safeguards are equivalent to traditional statistical quality control methods. While unnecessary testing should be eliminated, equivalence to traditional statistical quality control (QC) is not the appropriate benchmark. Quality control is only a tool to detect process changes with predetermined statistical power, typically 95% for clinical laboratories. Robust analytical systems designed to avoid or prevent incorrect results are superior to a quality control process that catch errors after they occur. Criteria for the appropriate level of QC should be based on the risks that incorrect results would present to patients. The current debate over how to demonstrate Equivalent QC (EQC) highlights the different approaches to quality taken by manufacturers and laboratories. United States manufacturers moved beyond statistical quality control in the 1970s to focus on total quality concepts, following the example of Japanese industry. By the 1990s, quality management systems and risk management had taken hold in the United States as the preferred approach. Risk as used here is the combination of severity of harm and the probability of that harm occurring. Revamped FDA regulations in 1996 gave in vitro diagnostic (IVD) and other medical device manufacturers the responsibility to decide the appropriate amount of quality control testing based on risk assessment. The United States Quality System Regulation (QSR) 2 and its European regulatory counterpart, the IVD Medical Device Directive 3 require manufacturers to integrate a formal risk management process into the quality management system. Although the QSR only mentions risk analysis under design controls, the preamble clearly describes an expectation that risk management is a key management responsibility. Australia, Canada, Japan, and the Global Harmonization Task Force have also embraced or are embracing risk management as part of the quality system. 4 While the trend in the medical device manufacturing industry has been away from prescriptive regulation, clinical laboratory regulations in the 1990s 5 prescribed the number of QC tests that must be performed daily regardless of the clinical significance of an erroneous result or the likelihood of occurrence, thus removing an incentive to seek inherently safer IVD medical devices. The revised CLIA regulations retained the prescriptive requirements. 6 Risk Management in Industry The risk management process for medical device manufacturers is described in an international standard, ISO The stepwise process is shown in Figure 1. A brief description of the main elements will reveal its applicability to clinical laboratories. Hazard Analysis The risk management process starts with systematic identification of known and foreseeable hazards based on the anticipated uses of the device. This includes not only the analytical performance required for medical use, but also any foreseeable use errors associated with the device or its results. Any failures to meet explicit and implied claims are also evaluated as potential hazards. In the case of a laboratory method, the main hazard to patients is an incorrect result, which in the hands of a physician could lead to a dangerous misdiagnosis or treatment decision. Another hazard might be a delay in availability of a critical result, which for certain analytes could lead to a harmful delay in treatment. Figure 1_Life-cycle risk management. by guest labmedicine.com October 2005 Volume 36 Number 10 LABMEDICINE 633

2 Risk Analysis Among several possible risk models applicable to in vitro diagnostic medical devices, the one shown in Figure 2 depicts a sequence of events that starts with a failure in a manufacturer s quality system that results in a defective device. 8 In this example, a laboratory using the device generates an incorrect result and reports it to a physician, who relies on it and reaches the wrong diagnosis. This creates a hazardous situation for the patient, who may be harmed by inappropriate treatment. Once a list of potential hazards is constructed, possible harm scenarios can be identified. Error grid analysis was developed by Clarke and colleagues 9 to classify incorrect glucose results based on the degree of error and the physiological status of the patient. Parkes and colleagues 10 have published an improved procedure for developing an error grid based on the consensus of a large number of medical practitioners. Their approach can be applied to many clinical analytes. An error grid overlaying hypothetical method comparison data is shown in Figure 3. Each of the zones is characterized by the potential of a result in that zone to cause harm to a patient, with severity increasing from Zone A (clinically acceptable performance) to Zone E (dangerously incorrect treatment). An error grid provides a logical basis for ranking the severity of harm on a scale of 1 (Zone A) to 5 (Zone E). In a typical risk analysis, each hazard is analyzed to estimate the probability that it will occur in actual use. When failure data is available from experience, semi-quantitative estimates of failure rates can be projected (eg, 1/1,000, 1/10,000, 1/100,000, etc). When failure data is not available, perhaps when a new assay, technology, or application is being evaluated, expert judgment is relied upon to rank the likelihood on a qualitative scale, using terms such as frequent, occasional, rare, and theoretical. Since the purpose of a preliminary hazard analysis is to identify the risks early so they can be systematically eliminated during the design phase of a product or process, or at least reduced to an acceptable level, knowing the relative risk is good enough. Subsequent experience will be used to verify and refine the risk estimates. The probabilities identified in a hazard analysis usually apply to the occurrence of the hazard, even though not all hazards develop into hazardous situations and not all hazardous situations result in harm. Risk management decisions are based on the probability of harm. If the incorrect result can be detected first by the laboratory or the physician, harm can be avoided. Therefore, the probability that an incorrect result will be detected by laboratory QC may be factored into the probability of harm. The way physicians use test results may also reduce the hazard from incorrect results. It is reasonable to assume that improbable results would be challenged and implausible results disregarded by physicians. Risk Evaluation The next step is to evaluate the risks against acceptability criteria that have been pre-determined in a risk management plan. Each manufacturer must decide the acceptable degree of risk, weighing the benefits to the patients and taking into account the current values of society. A risk acceptability chart can be constructed to distinguish between acceptable and unacceptable combinations of probability and severity. An example of a typical chart is given in Figure 4. Risk acceptability decisions are made initially for each individual risk before mitigation, again for each residual risk after Figure 2_IVD risk model. Figure 3_Error grid analysis. mitigation, and finally for the overall residual risk after all risk controls have been implemented. Risk Controls Individual risks that fall in the broadly acceptable zone are considered negligible and require no further attention. Risks that fall in the unacceptable zone must be reduced to an acceptable level. Risks that fall in between may also require mitigation. Although not required by the standard, many manufacturers establish a policy of reducing all risks even those in the acceptable zone to a level as low as practicable, which means taking into account both technical and economic feasibility. Figure 4_Risk acceptability. by guest 634 LABMEDICINE Volume 36 Number 10 October 2005 labmedicine.com

3 Built-in monitoring systems that inspired the EQC term are examples of risk control measures. The need for monitoring would be identified during a risk analysis when the instrument or assay is being designed, where potential failure modes are identified through a Failure Modes and Effects Analysis (FMEA). Systems to monitor the critical instrument parameters associated with potential failure causes (incubator temperature, reagent absorbance, detector voltage stability, cutoff point, sample level, etc) would be incorporated into the assay design to avoid or prevent reporting incorrect results. These internal monitoring systems are directed at the potential failure causes to prevent incorrect results from being produced. As a result of the manufacturer s risk assessments, risk controls may also be implemented in the manufacturing process to prevent the production of defective units. According to the ISO standard, risk controls must be implemented in the following priority order. First, an attempt must be made to eliminate significant risks from the device. Second, if inherently safe design is not practicable, protective measures must be incorporated in the device itself (eg, monitoring sensors, fail-safe alarms, fault-tolerant systems) or in the manufacturing process (eg, bar-coded materials, acceptance testing, inprocess monitoring). Third, if hazards cannot be avoided or prevented, the user must be provided with the information to understand and address residual risks. This is called information for safety in ISO 14971, which includes instructions for use, warnings and precautions, contraindications and limitations, performance characteristics and quality control recommendations. Risk Monitoring Risk management according to ISO is a product life-cycle process, which means it continues as long as the product is being produced and is still in active use. For some workhorse laboratory analyzers, this could be a decade or more. During this time, manufacturers monitor experience with the device to look out for the occurrence of previously unidentified hazards or indications that known risks might be considered no longer acceptable. In addition to external sources of user feedback, such as performance evaluations, customer complaints and adverse event reports, internal data sources are also monitored. Any newly identified hazards loop back through the risk assessment process as shown in Figure 1. Application to Clinical Laboratories Manufacturers are required to implement a risk management process in their quality system to protect patients from Figure 5_Clinical lab process map. hazardous results. Unfortunately, this addresses only a fraction of the possible failures shown in Figure 2. Many other opportunities to produce hazardous results are under the control of the laboratory. The same ISO risk management process can be used to analyze, evaluate, and control these risks. It would also fulfill the JCAHO requirement to define and implement an ongoing, proactive program for identifying risks to patient safety and reducing medical/health care errors. 11 One approach that can easily be adapted for use in a clinical laboratory is Hazard Analysis and Critical Control Points HACCP. It applies common risk analysis tools, such as Fault Tree Analysis (FTA) and FMEA to identify the activities in the testing process that require controls. Then the existing controls are evaluated, their ability to maintain patient risk within a predefined acceptable level is verified, and additional controls are added where the analysis shows they are needed. The following example for a typical analyte will illustrate how it works. The hazards incorrect results and any potential harm would be identified as described above, so this example will begin with identifying potential failure modes in the testing process. Map the Process To see where failures can occur, the laboratory starts by creating a detailed flowchart of the entire testing process. Most analytical systems are made up of several components specimens, instrument, reagents, accessories, calibrators, and control materials and each involves several sub-processes that taken together produce a reportable test result. For example, specimens must be collected, transported, stored, and prepared before being added to the measuring instrument, which must have been properly installed, functionally qualified, calibrated, maintained, and setup for the analysis. The reagents, calibrators, and controls must be acquired, qualified, stored, and prepared specifically for the analysis. Lastly, results are calculated from the analytical measurements, verified in some manner and uploaded into the laboratory s information management system for eventual communication to the physician. The process map for a clinical laboratory can be complex, as the high-level map in Figure 5 illustrates. Every step must be broken down to a level that enables the critical quality attributes of the activity to be analyzed and potential failure modes and suitable controls identified. For complex processes, a FTA can be used to focus the FMEA on significant failures. Brainstorm Potential Failure Modes and Likely Causes This step requires the participation of persons knowledgeable about the analytical technology, reagent chemistry, specimens/analytes, personnel, methods, and other components of the system. These subject matter experts meet to discuss the possibilities from their different perspectives, weigh the evidence and reach consensus on potential failures and their effects. A balanced, cross-functional team can complete an objective risk assessment in a short time. A table is created to document the team s decisions. A scribe is designated to record each process failure (hazard), failure cause, effect (harm), severity, existing process controls (to prevent the failure), probability of occurrence (of the failure), detectability (prior to harm), and comments explaining rationale. A format similar to that shown in Table 1 results in a valuable reference document for use subsequently in troubleshooting, root cause analysis and laboratory improvement programs. It is a living document that is updated whenever new risk information is obtained. by guest labmedicine.com October 2005 Volume 36 Number 10 LABMEDICINE 635

4 The team addresses potential failures of the analytical phase, as well as potential failures during the pre- and post-analytic phases. Ideally, manufacturers would disclose significant residual risks from the device in the labeling in a form that laboratories could incorporate directly into their risk assessment, but there are no standards for what risks to communicate and how to convey the information. These decisions are left to the manufacturers, and although historically manufacturers have not published expected failure rates, it will be helpful to provide them to their customers upon request. Laboratories can also identify potential failure modes and estimate the probability of their occurrence from their own experience with the instrument or with similar instruments. Guidelines are found in CLSI EP18-A, 12 which was written for unit-use devices but is generally applicable to all IVD medical devices, and Annex H of the new version of ISO 14971, 8 which should be completed in Laboratories should not hesitate to ask their suppliers for residual risk information as well as advice and technical assistance for performing a risk assessment involving their device. Evaluate Risks Next, the risks need to be evaluated against criteria approved by the laboratory director. To identify the process activities that must be controlled to maintain acceptable risk, the risk assessment team makes decisions based on the severity, occurrence, and detectability values. For simplicity, the 3 scales can be set up as shown in Table 2 so that low numbers are good and values 6 and above must be addressed. Note that the detectability scale has an inverse relationship to the probability of detection. Identify Essential Control Points The ratings assigned by the team are used to determine whether a process or subprocess requires an Essential Control Point (ECP) and whether existing controls are adequate or further controls are needed to reduce the risk. An ECP is defined as a process activity where control to maintain risk at an acceptable level can be applied. The following rules dictate the actions. All processes with SEV 6 require an ECP. Process activities with OCC 6 require an ECP to ensure control. The ECP must be an effective method of detection, since the OCC score means a significant number of process failures are expected to occur. The ECP is placed either at the point of failure or at a subsequent step. Process activities with DET 6 require an ECP to ensure control. In this case, the ECP must be a process control that prevents failures, since the DET score means the probability that failures will be detected if they occur is not high enough. If both OCC 6 and DET 6, the process activity lacks adequate controls and corrective action must be initiated, either to reduce the failure rate or to increase the ability to detect a failure, or both. Probability of occurrence and DET will be reanalyzed after the changes are implemented to determine if an ECP is needed. Additional rules may be established to identify and prioritize improvement opportunities based on combinations of SEV, OCC, and DET. Examples of possible outcomes are given in Table 3. Note that all steps have the same severity score because the medical Table 1_Risk Assessment # Component Potential failure Effect Failure cause SEV Existing controls OCC DET Comments/ mode Rationale 1 Reagent Stability not meeting Incorrect results/ Reagent deterioration due 8 SOP (validated storage conditions), 6 2 Manufacturer s claim (negative drift) misdiagnosis to improper storage trained personnel; weekly QC instructions 2 Reagent Large bias shift at lot Incorrect results/ Reagent lot-lot differences 8 QC acceptance testing, supplier 6 2 change misdiagnosis qualification 3 Instrument Increased imprecision Incorrect results/ Lamp aging 6 Preventive maintenance 2 8 Manufacturer s at high analyte misdiagnosis program / SOP (lamp instructions concentrations replacement schedule) 4 Instrument Sporadic outlier Incorrect results/ Unstable power source 8 Voltage regulator, installation 4 10 Observed with readings misdiagnosis in lab qualification similar instruments 5 Calibrator Large bias shift after Incorrect results/ Incorrect calibrator value 8 Certificate of traceability, post- 6 2 calibration misdiagnosis assigned by manufacturer calibration QC, proficiency testing 6 Calibrator Large bias shift after Incorrect results/ Calibrator reconstitution 8 Qualified personnel, SOP, training, 4 2 calibration misdiagnosis error post-calibration QC, proficiency testing 7 Sample Sporadic outlier Incorrect results/ Drug interference (known 8 Specimen requisition form; hospital 4 10 Observed in results misdiagnosis interferent) pharmacy drug alert system method verification study 8 Sample Unsuitable sample No result/delayed Improper specimen 4 SOP (sample preparation); training/ 6 4 Requires (hemolyzed) treatment preparation personnel qualification re-draw Table 2_Severity, Occurrence, and Detection Scales Score Severity of Harm (SEV) Probability of Occurrence (OCC) Detectability Prior to Harm (DET) 10 Catastrophic patient death Frequent ( 10-3) Almost impossible to detect 8 Critical permanent impairment or life-threatening injury Probable (< 10-3 and 10-4) Low probability of detection 6 Serious injury or impairment requiring medical intervention Occasional (< 10-4 and 10-5) Medium probability of detection 4 Minor temporary injury or impairment not requiring medical intervention Remote (< 10-5 and 10-6) High probability of detection 2 Negligible inconvenience or temporary discomfort Improbable/theoretical (< 10-6) Almost certain to be detected by guest 636 LABMEDICINE Volume 36 Number 10 October 2005 labmedicine.com

5 consequence of an incorrect result is often the same regardless of its cause. The medical actions and severity of harm may depend on the direction and magnitude of error. Verify and Monitor ECPs Once ECPs are identified, objective evidence is required to show that the controls in place are effective. The team reviews the evidence to satisfy itself that the controls will maintain risk at an acceptable level. In addition, because of the importance placed on these controls to prevent incorrect results from being reported, their continued effectiveness over time must be monitored. Any indication that a failure has led to incorrect results being reported must be investigated to determine where the controls failed and root causes must be addressed to prevent recurrence. The risk assessment is periodically reviewed and updated based on the laboratory s failure investigations, and confidence in the risk control measures continues to increase over time. Determining the Appropriate Amount of QC The CMS initiative to allow reduced frequency of QC for certain analytical systems has sparked a healthy debate over the evidence required to justify a laboratory s QC strategy. Arbitrary options to implement the revised CLIA rule could not be defended on scientific grounds and a consensus has emerged that QC alternatives must be validated. Proponents of Equivalent QC are looking to manufacturers to prove the effectiveness of their internal monitoring systems, which they are already required to do under design control regulations, and to FDA to certify that systems are EQC ready based on review of the manufacturer s risk assessment and design validation data. Unfortunately, this solution is too simplistic. The laboratory s entire QC process must have a solid scientific footing. Furthermore, the new proposals focus on controls immediately related to failure of commercial test systems, which are generally acknowledged to be low compared to the frequency of pre-analytic and post-analytic failures. The real question the laboratory must answer is this: What are the appropriate laboratory controls to minimize patient risk? The ISO risk management process offers a systematic way to answer that question. Whenever possible, control measures should be targeted specifically to events that might lead to failures, as identified in the risk assessment. Table 4 illustrates various types of events that are associated with failures leading to incorrect results and some possible risk controls. There are several hurdles that must be overcome before an effective solution can be realized. Cooperation among regulatory authorities (CMS and FDA), manufacturers and laboratories is needed to overcome them. CMS should harmonize its regulatory philosophy with the quality system approach adopted by FDA 10 years ago. The current CLIA regulations contain elements of a quality system, but they retain overly prescriptive requirements. The lack of agreement about medical requirements makes it difficult to determine when an incorrect result is a significant hazard. The laboratory community should own the task of defining the performance required for medical utility, perhaps aided by the CLSI consensus process. Forcing each manufacturer to figure out the medical requirements for assays they manufacture is highly inefficient and promotes inconsistency. Manufacturers need to know (1) what accuracy is needed for diagnosing, classifying, or treating patients and (2) what harm could occur from failure to meet the requirements. Consensus error grids could be constructed for many analytes using the Parkes approach. 10 Table 3_Assignment of ECPs Step SEV OCC DET ECP None Process Add control Improve Detection Table 4_Examples of controls targeted to specific events Laboratory Controls Event Prevention Detection Instrument drift Preventive maintenance program; calibration schedule QC samples run at predetermined intervals Reagent deterioration Storage per manufacturer s instructions; standard operating procedure; training QC samples run at predetermined intervals New lot of reagents, Material handling procedures; material specifications QC acceptance testing of materials upon receipt controls or calibrators New shipment of materials Validated shipping conditions (manufacturer) QC acceptance testing upon receipt System calibration Calibrate per manufacturer s instructions; standard operating procedure; training QC samples run post calibration; proficiency testing New calibrator Certificate of traceability QC acceptance testing upon receipt; proficiency testing Maintenance and repair Perform per manufacturer s instructions; personnel qualification, SOPs, training QC samples run afterward Method modification Validation or verification QC samples run after implementation Equipment relocation Requalification QC samples run afterward Sample preparation Prepare per manufacturer s instructions; personnel qualification, SOPs, training Plausibility checks Specimen identification Bar codes; personnel qualification, SOP, training Verification protocol; redundant checks Drug interference Sample acceptability requirements; drug alert system (pharmacy) None Instrument failure Preventive maintenance per manufacturer's instructions Check critical functions periodically Operator error Personnel qualification; training on manufacturer's instructions for use & SOPs QC samples run after operator sensitive procedures; verify critical steps Method imprecision (within Validation; improve/replace method if not meeting medical requirements QC samples run (mean and range charts) manufacturer's claims) Software update Software validation (manufacturer); verify critical functions (laboratory) QC samples run afterward by guest labmedicine.com October 2005 Volume 36 Number 10 LABMEDICINE 637

6 Laboratories should do a better job of qualifying their suppliers, not only specifying their performance requirements but also insisting on the information they need to control patient risk. Greater value should be placed on a cooperative relationship in the interest of patient safety. Manufacturers need to communicate the residual risk in a format that is useful to laboratories. The common practice of defaulting to a recommendation to implement traditional QC procedures prevents laboratories from designing a more appropriate and cost-effective control system. FDA can help level the playing field among manufacturers by providing clear guidance regarding risk management expectations and strictly enforcing compliance. For the system to work, laboratories need to have confidence that manufacturers are doing a good job in designing and validating their devices. It has been almost 10 years since the QSR was enacted, so by now every manufacturer should have implemented design controls and a comprehensive risk management program. CLSI should broker a consensus on the risk information that manufacturers need to provide users and on the supporting evidence that might be reviewed by the FDA to certify that internal device controls are effective. ISO/TC212 should develop an international standard on the application of the ISO risk management principles to clinical laboratories as a companion to the successful quality management standard, ISO There is a perception that risk assessments are beyond the capability of most laboratories, but that is not the case. There are plenty of textbooks, 14,15 training programs, and other resources to help get started, and there are examples of laboratories that have successfully applied FMEA methodology Nevertheless, a set of practical risk assessment guidelines targeted to the clinical laboratory would be an obvious extension of the Option 4 proposal. The side debate over whether the analytical performance of present methods is good enough for medical use should continue. 19 If analytes such as calcium and creatinine are not being measured with sufficient precision, then they must be producing some percentage of hazardous results in normal use, which would be identified and evaluated in a properly executed risk assessment if medical requirements were defined. Such a finding should stimulate improved methods under present design control regulations. However, no amount of QC testing will improve the poor performance of these assays, and this gap should not distract from a focus on identifying and implementing appropriate laboratory controls for risks that we do know. LM 1. Clinical Laboratory Improvement Amendments (CLIA) Equivalent Quality Control Procedures, Centers for Medicare and Medicaid Services, Brochure #4, Baltimore: 2003; Available at: Accessed January 27, Quality System Regulation, US Code of Federal Regulations, 21 CFR Part Council Directive 98/79/EC of the European Parliament and of the Council of 27 October 1998 on In Vitro Diagnostic Medical Devices," Official Journal of the European Union L331 (December 7, 1998). 4. Global Harmonization Task Force, Risk Management as an Integral Part of the Quality Management System, Proposed Draft SG3/N15R6, 5. U.S. Department of Health and Human Services. Medicare, Medicaid and CLIA programs: Regulations implementing the Clinical Laboratory Improvement Amendments of 1988 (CLIA). Final rule. Fed Regist 1992; 57: CLIA regulations, 42 CFR Part 493, Available at: Accessed January 27, ISO 14971:2000, Medical Devices Application of risk management to medical devices 8. ISO/DIS 14971:2005, Medical Devices Application of risk management to medical devices 9. Clarke WL. Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care. 1987;10: Parkes JL. A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose. Diabetes Care. 2000;23: Joint Commission on the Accreditation of Healthcare Organizations (JCAHO). Effective July 1, Revisions to Joint Commission Standards in Support of Patient Safety and Medical/Health Care Error Reduction. Chicago, Joint Commission on Accreditation of Healthcare Organizations. 12. Clinical and Laboratory Standards Institute (formerly NCCLS). Quality Management for Unit-Use Testing; Approved Guideline. NCCLS document EP18-A (ISBN ). NCCLS, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania USA, ISO 15189:2003; Medical laboratories Particular requirements for quality and competence. Geneva: ISO, Failure Mode and Effect Analysis: FMEA From Theory to Execution, Second Edition, Stamatis, D. H. 15. American Institute Of Chemical Engineers, Guidelines for Hazard Evaluation Procedures - With Worked Examples (2nd Edition). Center for Chemical Process Safety (CCPS) Burgmeier J. Failure mode and effect analysis: an application in reducing risk in blood transfusion. J Quality Improvement. 2002;28: Woodhouse S. Engineering for safety: Use of failure mode and effects analysis in the laboratory. Lab Med. 2005;36: Astion M. Interview: Error reduction and risk assessment with Shirley Weber, MHA. Laboratory Errors & Patient Safety. 2004; Clinical and Laboratory Standards Institute (CLSI). Proceedings from the QC for the Future Workshop; A Report. CLSI document X6-R (ISBN ). Clinical and Laboratory Standards Institute, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania USA, by guest 638 LABMEDICINE Volume 36 Number 10 October 2005 labmedicine.com

Evolving Clinical Laboratory Management Through Implementation of a Risk Assessment Plan

Evolving Clinical Laboratory Management Through Implementation of a Risk Assessment Plan 1 Evolving Clinical Laboratory Management Through Implementation of a Risk Assessment Plan SEDEF YENICE Session: Laboratory Management analytica conference 2016 May 10-12, 2016, Munich 2 Objectives Review

More information

Risk Management in IVD Producer Relation between manufacturer and user. S.M.Boutorabi DCLS, PhD

Risk Management in IVD Producer Relation between manufacturer and user. S.M.Boutorabi DCLS, PhD Risk Management in IVD Producer Relation between manufacturer and user S.M.Boutorabi DCLS, PhD IVD Manufacturer Requirements International Standard ISO 13485-2016 Medical Devices Quality Management Systems

More information

SAMPLE. Risk Management Techniques to Identify and Control Laboratory Error Sources; Approved Guideline Second Edition

SAMPLE. Risk Management Techniques to Identify and Control Laboratory Error Sources; Approved Guideline Second Edition November 2009 Risk Management Techniques to Identify and Control Laboratory Error Sources; Approved Guideline Second Edition This guideline describes risk management techniques that will aid in identifying,

More information

'How-To' Implement an Effective Proactive Risk Management Strategy

'How-To' Implement an Effective Proactive Risk Management Strategy 'How-To' Implement an Effective Proactive Management Strategy Sedef YENICE Satellite Educational Workshop on Intelligent Clinical Laboratory Management: Impacts on Quality System Improvement Hilton Durban

More information

IQCP for Commercially Prepared CLSI-Exempt Media

IQCP for Commercially Prepared CLSI-Exempt Media Please note that some references to protocol, publications, performance data etc. are fictitious in this EXAMPLE. Please use your own DATA for your IQCP. The following represents one example of how you

More information

Developing QC Plans Part 3. Performing a Risk Assessment to Identify Hazards and Target Failure Modes

Developing QC Plans Part 3. Performing a Risk Assessment to Identify Hazards and Target Failure Modes Webinar Series sponsored by Thermo Fisher Scientific Developing QC Plans Part 3. Performing a Risk Assessment to Identify Hazards and Target Failure Modes James O. Westgard Sten A. Westgard Westgard QC,

More information

SAMPLE. Evaluation of Total Analytical Error for Quantitative Medical Laboratory Measurement Procedures

SAMPLE. Evaluation of Total Analytical Error for Quantitative Medical Laboratory Measurement Procedures 2nd Edition EP21 Evaluation of Total Analytical Error for Quantitative Medical Laboratory Measurement Procedures This guideline provides manufacturers and end users with an understanding of concepts related

More information

HUMAN FACTORS ENGINEERING: DESIGN OF MEDICAL DEVICES

HUMAN FACTORS ENGINEERING: DESIGN OF MEDICAL DEVICES HUMAN FACTORS ENGINEERING: DESIGN OF MEDICAL DEVICES Chapter 5 Prof. Yasser Mostafa Kadah Managing the risk of use error Reports compiled by the Food and Drug Administration (FDA) indicate that as many

More information

Guidance for Industry and FDA Staff. Replacement Reagent and Instrument Family Policy

Guidance for Industry and FDA Staff. Replacement Reagent and Instrument Family Policy Guidance for Industry and FDA Staff Replacement Reagent and Instrument Family Policy Document issued on: December 11, 2003 This document supersedes Data for Commercialization of Original Equipment Manufacturers

More information

Verification or validation, that is the question

Verification or validation, that is the question Perspective Page 1 of 7 Verification or validation, that is the question Giorgia Antonelli 1, Andrea Padoan 1, Ada Aita 1, Laura Sciacovelli 2, Mario Plebani 1,2 1 U.O.C. of Laboratory Medicine, Department

More information

9/8/2014. Answers to All Your IQCP Questions. The History Lesson. Some repeat/review. Let s talk IQCPs. Disclaimers

9/8/2014. Answers to All Your IQCP Questions. The History Lesson. Some repeat/review. Let s talk IQCPs. Disclaimers Disclaimers I work for Alere Answers to All Your IQCP Questions Thomas Koshy, Ph.D. Sr. Director, Scientific Affairs Alere produces testing devices for use at the POC so I have more data on the testing

More information

Establishing Quality Control Target Values and Standard Deviations for Hematology Instrumentation

Establishing Quality Control Target Values and Standard Deviations for Hematology Instrumentation Establishing Quality Control Target Values and Standard Deviations for Hematology Instrumentation For informational purposes only. WHO REQUIRES IT AND WHAT THEY REQUIRE Regulatory agencies require the

More information

Guide to Fulfillment of Validation and Verification of Examination Requirements

Guide to Fulfillment of Validation and Verification of Examination Requirements DIAGNOSTIC ACCREDITATION PROGRAM College of Physicians and Surgeons of British Columbia 300 669 Howe Street Telephone: 604-733-7758 ext. 2635 Vancouver BC V6C 0B4 Toll Free: 1-800-461-3008 (in BC) www.cpsbc.ca

More information

Marcia L. Zucker, Ph.D. ZIVD LLC

Marcia L. Zucker, Ph.D. ZIVD LLC Marcia L. Zucker, Ph.D. ZIVD LLC 1 Two levels of liquid QC available from manufacturers Haphazard implementation 1990 s - POCT awareness increased Increased implementation of QC programs Expense of POCT

More information

IQCP for Commercial Antimicrobial Susceptibility Testing (AST) System XYZ

IQCP for Commercial Antimicrobial Susceptibility Testing (AST) System XYZ Please note that some references to protocol, publications, performance data etc. are fictitious in this EXAMPLE. Please use your own DATA for your IQCP. The following represents one example of how you

More information

IQCP for Streamlined ID QC of Vitek 2

IQCP for Streamlined ID QC of Vitek 2 IQCP for Streamlined ID QC of Vitek 2 Facility: EXAMPLE Test System: Vitek 2 Commercial Identification System Test System Primary SOPs include: GENERAL PROCEDURES: Quality Control for AST Cards and Identification

More information

Developing an Individualized Quality Control Plan (IQCP) For Cepheid s GeneXpert Diagnostic Systems

Developing an Individualized Quality Control Plan (IQCP) For Cepheid s GeneXpert Diagnostic Systems Developing an Individualized Quality Control Plan (IQCP) For BACKGROUND On January 1, 2014, the Center for Medicare and Medicaid Services (CMS) adopted an alternative Quality Control (QC) procedure that

More information

FINAL DOCUMENT. Global Harmonization Task Force. Title: Clinical Evidence for IVD medical devices Key Definitions and Concepts

FINAL DOCUMENT. Global Harmonization Task Force. Title: Clinical Evidence for IVD medical devices Key Definitions and Concepts GHTF/SG5/N6:2012 FINAL DOCUMENT Global Harmonization Task Force Title: Clinical Evidence for IVD medical devices Key Definitions and Concepts Authoring Group: Study Group 5 of the Global Harmonization

More information

IQCP for the Akers Bio PIFA PlussPF4 Rapid Assay

IQCP for the Akers Bio PIFA PlussPF4 Rapid Assay On December 31, 2015 the ability to use Equivalent Quality Control (EQC) policies will expire. Laboratories will be required to either follow the quality control regulations as outlined in CLIA 88 or develop

More information

9/2/15 (replaces 5/31/15 version) Page 1

9/2/15 (replaces 5/31/15 version) Page 1 Please note that some references to protocol, publications, performance data etc. are fictitious in this EXAMPLE. Please use your own DATA for your IQCP. The following represents one example of how you

More information

Plan Subject Index Number Section Subsection Category Contact Last Revised References Applicable To Detail PRINCIPLE:

Plan Subject Index Number Section Subsection Category Contact Last Revised References Applicable To Detail PRINCIPLE: Subject Quality Assurance for Laboratory Testing Index Number Lab-0135 Section Laboratory Subsection Laboratory - General Category Departmental Contact Nancy Ekern Last Revised 7/27/2017 References Required

More information

March 19, Division of Dockets Management (HFA-305) Food and Drug Administration 5630 Fishers Lane Room 1061 Rockville, MD 20852

March 19, Division of Dockets Management (HFA-305) Food and Drug Administration 5630 Fishers Lane Room 1061 Rockville, MD 20852 March 19, 2018 Division of Dockets Management (HFA-305) Food and Drug Administration 5630 Fishers Lane Room 1061 Rockville, MD 20852 RE: Docket No. FDA-2017-D-6765; Draft Guidance for Industry and Food

More information

Avoiding the Most Frequent Joint Commission Laboratory Survey Findings using Robust Process Improvement Tools (RPI) AVOIDING THE MOST

Avoiding the Most Frequent Joint Commission Laboratory Survey Findings using Robust Process Improvement Tools (RPI) AVOIDING THE MOST Avoiding the Most Frequent Joint Commission Laboratory Survey Findings using Robust Process Improvement Tools (RPI) AVOIDING THE MOST FREQUENT JOINT COMMISSION LABORATORY SURVEY FINDINGS USING ROBUST PROCESS

More information

Current Topics in NYS Clinical Laboratory Oversight

Current Topics in NYS Clinical Laboratory Oversight June 13, 2018 1 Current Topics in NYS Clinical Laboratory Oversight Derek Symula, Ph.D. and Stephanie Shulman, MPH, MS, MT (ASCP) June 13, 2018 2 (i) Clinical Laboratory Reference System (CLRS) to monitor,

More information

The procurement, qualification and calibration of lab instruments: An Overview

The procurement, qualification and calibration of lab instruments: An Overview The procurement, qualification and calibration of lab instruments: An Overview The reliability of analytical data generated from chemical and physical analyses is critically dependent on three factors:

More information

IQCP for the Akers Bio PIFA Heparin Platelet Factor 4 Rapid Assay

IQCP for the Akers Bio PIFA Heparin Platelet Factor 4 Rapid Assay On December 31, 2015 the ability to use Equivalent Quality Control (EQC) policies will expire. Laboratories will be required to either follow the quality control regulations as outlined in CLIA 88 or develop

More information

DRAFT MEDICAL DEVICE GUIDANCE DOCUMENT

DRAFT MEDICAL DEVICE GUIDANCE DOCUMENT November 2015 DRAFT DRAFT MEDICAL DEVICE GUIDANCE DOCUMENT REQUIREMENTS FOR LABELLING OF MEDICAL DEVICES Medical Device Authority MINISTRY OF HEALTH MALAYSIA Contents Page Preface... iii 1 Introduction.

More information

Disclosures. Laboratory Stakeholders. IVD vs. LDT. FDA Regulation of Laboratory Developed Tests 10/2/2015. FDA Regulation of LDTs

Disclosures. Laboratory Stakeholders. IVD vs. LDT. FDA Regulation of Laboratory Developed Tests 10/2/2015. FDA Regulation of LDTs Disclosures FDA Regulation of Laboratory Developed Tests Beaumont Health System, 24 th Annual Symposium on Molecular Pathology September 16, 2015 Roger D. Klein, MD JD Director, Molecular Pathology Clinical

More information

WHO post-market surveillance of in vitro diagnostics

WHO post-market surveillance of in vitro diagnostics WHO post-market surveillance of in vitro diagnostics Geneva, 27 November 2015 Anita Sands Prequalification Team World Health Organization About post-market surveillance of IVDs > Some quality, safety and

More information

A New Era in Quality Control: Developing an IQCP for Your Laboratory

A New Era in Quality Control: Developing an IQCP for Your Laboratory A New Era in Quality Control: Developing an IQCP for Your Laboratory Gerald A. Capraro, Ph.D., D(ABMM) Medical Director, Clinical Microbiology & Diagnostic Virology Laboratories LSU Health Sciences Center

More information

Office for Human Subject Protection. University of Rochester

Office for Human Subject Protection. University of Rochester POLICY 1. Purpose Outline the responsibilities and regulatory requirements when conducting human subject research that involves the use of drugs, agents, biological products, or nutritional products (e.g.,

More information

Example # 2. IQCP for Antimicrobial Susceptibility Testing (AST) Test System: Beckman Coulter Microscan Walkaway 96. Facility: Written by: Date:

Example # 2. IQCP for Antimicrobial Susceptibility Testing (AST) Test System: Beckman Coulter Microscan Walkaway 96. Facility: Written by: Date: IQCP for Antimicrobial Susceptibility Testing (AST) Test System: Beckman Coulter Microscan Walkaway 96 Facility: Written by: Date: Implementation Date: This risk assessment and IQCP plan has been approved

More information

RISK MANAGEMENT AND DESIGN CONTROL. PRACTICAL SOLUTION F. Akelewicz 03/06

RISK MANAGEMENT AND DESIGN CONTROL. PRACTICAL SOLUTION F. Akelewicz 03/06 RISK MANAGEMENT AND DESIGN CONTROL All rights reserved. No part of this work covered by the copyright hereon may be reproduced or used in any form or by any means including graphic, electronic, or mechanical,

More information

Re: Docket No. FDA-2017-D ; Draft Guidance on Recommendations for Dual 510(k) and CLIA Waiver by Application Studies

Re: Docket No. FDA-2017-D ; Draft Guidance on Recommendations for Dual 510(k) and CLIA Waiver by Application Studies Division of Dockets Management (HFA-305) Food and Drug Administration 5630 Fishers Lane, Room 1061 Rockville, MD 20852 Re: ; Draft Guidance on Recommendations for Dual 510(k) and CLIA Waiver by Application

More information

Quality Assurance by Liz McChlery

Quality Assurance by Liz McChlery Quality Assurance by Liz McChlery Define QA and Total Quality Management Describe Quality Control Define and distinguish Accuracy and Precision Define a Biological Reference Interval Describe ISO15189

More information

MEDICAL DEVICE CLINICAL INVESTIGATIONS AND ISO 14155

MEDICAL DEVICE CLINICAL INVESTIGATIONS AND ISO 14155 MEDICAL DEVICE CLINICAL INVESTIGATIONS AND ISO 14155 EXECUTIVE SUMMARY Medical device regulations around the world generally require manufacturers of most types of medical devices to supply data as part

More information

Systematic Risk Management: An Overview of ICH Q-9

Systematic Risk Management: An Overview of ICH Q-9 Systematic Risk Management: An Overview of ICH Q-9 March 12, 2014 12014 ParagonRx International LLC Today s Session Systematic Risk Management: An Overview of ICH Q-9 Speakers: Jeff Fetterman, President,

More information

External Defibrillator Improvement Initiative

External Defibrillator Improvement Initiative External Defibrillator Improvement Initiative November 2010 Center for Devices and Radiological Health U.S. Food and Drug Administration External Defibrillator Improvement Initiative Table of Contents

More information

QUALITY MANUAL DISTRIBUTION. Your Logo Here. Page 1 1 of 57. Rev.: A

QUALITY MANUAL DISTRIBUTION. Your Logo Here. Page 1 1 of 57. Rev.: A Page 1 1 of 57 President Design Engineering Production Production Engineering Materials Control Purchasing Service Marketing Sales Contracts Human Resources Quality Assurance Quality Control Production

More information

Project Proposal Form

Project Proposal Form Project Proposal Form Date: Submitter contact information: Name: Organization: Phone number: E-mail: Project Submission Proposed Title: Proposed product will be (check one): (See Part 3 for a detailed

More information

IVDR Breakout. Copyright 2017 BSI. All rights reserved.

IVDR Breakout. Copyright 2017 BSI. All rights reserved. IVDR Breakout 1 IVDR Annex I (SPRs) & Annex II (Tech documentation) 2 IVDR Annex I General Safety & performance requirements 3 Overview Context of the General Safety & Performance Requirements (SPRs) [our

More information

A Risk-based Approach for In Vitro Companion Diagnostics Device FDA Approval Process Associated with Therapies that have Breakthrough Designation

A Risk-based Approach for In Vitro Companion Diagnostics Device FDA Approval Process Associated with Therapies that have Breakthrough Designation A Risk-based Approach for In Vitro Companion Diagnostics Device FDA Approval Process Associated with Therapies that have Breakthrough Designation A Risk-based Approach for In Vitro Companion Diagnostics

More information

Urinalysis and Body Fluids CRg. Laboratory Regulation. Laboratory Regulation for Quality Assessment. Unit 1 B. Quality Assessment

Urinalysis and Body Fluids CRg. Laboratory Regulation. Laboratory Regulation for Quality Assessment. Unit 1 B. Quality Assessment Urinalysis and Body Fluids CRg Unit 1 B Laboratory Regulation Federal Regulations / Regulatory Organizations Laboratory structure / operation Quality test performed by qualified personnel to obtain quality

More information

Application of Quality Risk Management Tools for Cell Therapy Manufacturing

Application of Quality Risk Management Tools for Cell Therapy Manufacturing Application of Quality Risk Management Tools for Cell Therapy Manufacturing 17 th ISCT Annual Meeting May 20, 2011 Jean Stanton Associate Director, Product Quality Management Janssen Supply Chain Conflict/Disclaimer

More information

SAMPLE. User Evaluation of Between-Reagent Lot Variation; Approved Guideline

SAMPLE. User Evaluation of Between-Reagent Lot Variation; Approved Guideline September 2013 User Evaluation of Between-Reagent Lot Variation; Approved Guideline This document provides guidance for laboratories on the evaluation of a new reagent lot, including a protocol using patient

More information

2. Risk-based compliance was also a key component in the FDA's new approach for dealing with electronic records and signatures: 21 CFR Part 11.

2. Risk-based compliance was also a key component in the FDA's new approach for dealing with electronic records and signatures: 21 CFR Part 11. LITERATURE REVIEW: Risk management is well known and practiced in many industries and several industry task forces have developed guidance documents that facilitate risk management. Risk management for

More information

What is an LDT? The Do s s And Don ts of Validating Laboratory Developed Tests

What is an LDT? The Do s s And Don ts of Validating Laboratory Developed Tests The Do s s And Don ts of Validating Laboratory Developed Tests Deirdre Astin, MS, MT (ASCP) Lab Quality Confab 2011 November 2011 San Antonio Texas What is an LDT? Laboratory developed tests (LDTs) are

More information

LIFE CYCLE FACILITY ASSET MANAGEMENT. Presented by Pedro Dominguez Managing Principal, The Invenio Group

LIFE CYCLE FACILITY ASSET MANAGEMENT. Presented by Pedro Dominguez Managing Principal, The Invenio Group LIFE CYCLE FACILITY ASSET MANAGEMENT Presented by Pedro Dominguez Managing Principal, The Invenio Group LEARNING OBJECTIVES Correlate the current business environment to the demands placed on facility

More information

The Risk Management + Design Controls Connection: What Device Makers Need to Know

The Risk Management + Design Controls Connection: What Device Makers Need to Know !!! The Risk Management + Design Controls Connection: What Device Makers Need to Know Jon Speer Founder & VP of QA/RA greenlight.guru Table of Contents 1 Intended Use & User Needs 6 Verification, Validation,

More information

Risk Based Validation. Why, How and with what tools?

Risk Based Validation. Why, How and with what tools? Risk Based Validation Why, How and with what tools? Tech Talk Agenda History of FDA GMP initiative for the 21 st Century. Industry response to FDA initiative. Harmonisation through ICH. ASTM Standard on

More information

International Journal of Industrial Engineering Research and Development (IJIERD), ISSN 0976 INTERNATIONAL JOURNAL OF INDUSTRIAL ENGINEERING

International Journal of Industrial Engineering Research and Development (IJIERD), ISSN 0976 INTERNATIONAL JOURNAL OF INDUSTRIAL ENGINEERING INTERNATIONAL JOURNAL OF INDUSTRIAL ENGINEERING RESEARCH AND DEVELOPMENT (IJIERD) ISSN 0976 6979 (Print) ISSN 0976 6987 (Online) Volume 4, Issue 3, September - December (2013), pp. 61-72 IAEME: www.iaeme.com/ijierd.asp

More information

AAG PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

AAG PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE IMMAGE Immunochemistry Systems Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Alpha 1 -Acid Glycoprotein REF 447780 (150 tests) For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by:

More information

Perspective: Convergence of CLIA and FDA Requirements A Rational Shift in the Regulatory Paradigm

Perspective: Convergence of CLIA and FDA Requirements A Rational Shift in the Regulatory Paradigm Perspective: Convergence of CLIA and FDA Requirements A Rational Shift in the Regulatory Paradigm Planning for Efficiencies of Data, Resources, and Timelines A PRECISION BRIEF Introduction As precision

More information

SAMPLE. Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline Second Edition

SAMPLE. Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline Second Edition June 2012 Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline Second Edition This document provides guidance for evaluation and documentation of the detection

More information

Incorporating FDA Regulatory Thinking into Plans for Medical Devices

Incorporating FDA Regulatory Thinking into Plans for Medical Devices Incorporating FDA Regulatory Thinking into Plans for Medical Devices Ken Holroyd, MD, MBA Medical Director, Center for Technology Transfer and Commercialization Assistant Vice Chancellor for Research June

More information

HUMANITARIAN USE DEVICES 3/8/2016

HUMANITARIAN USE DEVICES 3/8/2016 DUKE UNIVERSITY HEALTH SYSTEM Human Research Protection Program HUMANITARIAN USE DEVICES 3/8/2016 It is the policy of the Duke University Health System Institutional Review Board (DUHS IRB) to review and

More information

General Accreditation Guidance. ISO/IEC 17025:2017 Gap analysis. April 2018

General Accreditation Guidance. ISO/IEC 17025:2017 Gap analysis. April 2018 General Accreditation Guidance Gap analysis April 2018 Copyright National Association of Testing Authorities, Australia 2018 This publication is protected by copyright under the Commonwealth of Australia

More information

The Integration and Application of ISO 9001 and the NIAHO Requirements In Medical Equipment Management

The Integration and Application of ISO 9001 and the NIAHO Requirements In Medical Equipment Management BUSINESS ASSURANCE The Integration and Application of ISO 9001 and the NIAHO Requirements In Medical Equipment Management Randall Snelling 20 October 2017 1 DNV GL 2017 20 October 2017 SAFER, SMARTER,

More information

SAMPLE. Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline Third Edition

SAMPLE. Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline Third Edition October 2014 Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline Third Edition This document provides guidance for evaluating the precision performance of quantitative measurement

More information

LG SOURCING VENDOR QUALITY MANAGEMENT SYSTEM REQUIREMENTS

LG SOURCING VENDOR QUALITY MANAGEMENT SYSTEM REQUIREMENTS LG SOURCING VENDOR QUALITY MANAGEMENT SYSTEM REQUIREMENTS LG Sourcing QMS ver01 03.29.2010 Page 1 of 16 PAGE INTENTIONALLY LEFT BLANK Copyright 2010, LF Corporation - All rights reserved No part of this

More information

ISO/IEC 17025:2017 Working Document Instruction Page

ISO/IEC 17025:2017 Working Document Instruction Page Instruction Page NOTES: 1. This working document is intended as a checklist for the assessor when conducting Testing and Calibration Laboratory and Sampling Organization Assessments according to ISO/IEC

More information

Draft Guidance for Industry, Clinical Laboratories, and FDA Staff. In Vitro Diagnostic Multivariate Index Assays

Draft Guidance for Industry, Clinical Laboratories, and FDA Staff. In Vitro Diagnostic Multivariate Index Assays Draft Guidance for Industry, Clinical Laboratories, and FDA Staff In Vitro Diagnostic Multivariate Index Assays DRAFT GUIDANCE This guidance document is being distributed for comment purposes only. Document

More information

ISO INTERNATIONAL STANDARD. Statistical methods for use in proficiency testing by interlaboratory comparisons

ISO INTERNATIONAL STANDARD. Statistical methods for use in proficiency testing by interlaboratory comparisons INTERNATIONAL STANDARD ISO 13528 First edition 2005-09-01 Statistical methods for use in proficiency testing by interlaboratory comparisons Méthodes statistiques utilisées dans les essais d'aptitude par

More information

The Intersection of Genomics Research and the IDE Regulation

The Intersection of Genomics Research and the IDE Regulation The Intersection of Genomics Research and the IDE Regulation Katherine Donigan, Ph.D. Personalized Medicine Staff FDA/CDRH/OIR October 19, 2017 1 In Vitro Diagnostic (IVD) Regulation Through the 1976 medical

More information

ISO 9001:2015 Expectations

ISO 9001:2015 Expectations The ISO 9001:2015 Standard was published September 15, 2015. The Standard is written such that it may be used by any organization type for the purpose of implementing and maintaining a Quality Management

More information

FDA Medical Device HFE Guidance

FDA Medical Device HFE Guidance W H I T E P A P E R www.makrocare.com U S FDA has established a new Draft Guidance; Applying Human Factors and Usability Engineering to Medical Devices to Optimize Safety and Effectiveness in Design. Manufacturers

More information

Common Deficiencies

Common Deficiencies Question text Deficiency #12: Does the laboratory have a documented system to ensure consistency of morphologic observations among all personnel performing (U/A, Micro, Hem) (sample) Suggested methods

More information

ASEAN Consultative Committee for Standards & Quality. Medical Device Product Working Group. Common Submission Dossier Template

ASEAN Consultative Committee for Standards & Quality. Medical Device Product Working Group. Common Submission Dossier Template ASEAN Consultative Committee for Standards & Quality Medical Device Product Working Group Common Submission Dossier Template 1.0 Introduction The document is intended to provide guidance for submission

More information

Bioresearch Monitoring Inspections in Vitro Diagnostics Devices

Bioresearch Monitoring Inspections in Vitro Diagnostics Devices Seite 1 von 7 U.S. Food and Drug Administration Protecting and Promoting Your Health Bioresearch Monitoring Inspections in Vitro Diagnostics Devices TABLE OF CONTENTS Introduction Nature, Scope, & Purpose

More information

LabGuide 13 How to Verify Performance Specifications

LabGuide 13 How to Verify Performance Specifications 1 LG 13 Comments? Feedback? Questions? Email us at info@cola.org or call us at 800 981-9883 LabGuide 13 How to Verify Performance Specifications The verification of performance specifications confirms

More information

MEDICAL DEVICE GUIDANCE DOCUMENT REQUIREMENTS FOR LABELLING OF MEDICAL DEVICES

MEDICAL DEVICE GUIDANCE DOCUMENT REQUIREMENTS FOR LABELLING OF MEDICAL DEVICES November 2018 Third Edition MEDICAL DEVICE GUIDANCE DOCUMENT REQUIREMENTS FOR LABELLING OF MEDICAL DEVICES Medical Device Authority MINISTRY OF HEALTH MALAYSIA Contents Page Preface.... iii 1 Introduction....

More information

Designing QC Activities to More Precisely Manage Analytical Accuracy and Patient Risk

Designing QC Activities to More Precisely Manage Analytical Accuracy and Patient Risk Designing QC Activities to More Precisely Manage Analytical Accuracy and Patient Risk Curtis A. Parvin, Ph.D. Manager of Advanced Statistical Research Bio-Rad Laboratories, QSD curtis_parvin@bio-rad.com

More information

Total Analytic Error From Concept to Application

Total Analytic Error From Concept to Application Página 1 de 5 Clinical Laboratory News Subscribe CLN Daily CLN Stat CLN Articles Total Analytic Error From Concept to Application Author: James O. Westgard, PhD, and Sten A. Westgard, MS // Date: SEP.1.2013

More information

11.0 FDA-Regulated Research Research Involving Investigational Drugs and Biologics

11.0 FDA-Regulated Research Research Involving Investigational Drugs and Biologics 11.0 FDA-Regulated Research The IRB evaluates the safety or efficacy of all drugs and devices used in research. Studies involving unapproved or investigational drugs or devices will be reviewed to ensure

More information

Choices IEC rd Edition and Component Selection

Choices IEC rd Edition and Component Selection Choices IEC 60601-1 3rd Edition and Component Selection Choices IEC 60601-1 3rd Edition and Component Selection Abstract When the 3rd edition of IEC 60601-1 was published, it marked the beginning of a

More information

CLSI Quality System Model. To manage quality you must measure it. Purpose of Measurement. Why Measure?

CLSI Quality System Model. To manage quality you must measure it. Purpose of Measurement. Why Measure? How Do You Know How Well You re Doing? (Quality Indicators for Measuring Laboratory ) Lucia M. Berte www.laboratoriesmadebetter.com HS01-A2: A Quality Management System for Health Care GP35-P (new!): Development

More information

1. Introduction to quality

1. Introduction to quality 1. Introduction to quality Definition of quality Level of accuracy required Negative consequences of laboratory error 1-1: The importance of laboratory quality Laboratory quality can be defi ned as accuracy,

More information

Laboratory OOS Investigations The Missing Link

Laboratory OOS Investigations The Missing Link Laboratory OOS Investigations The Missing Link India Pharmaceutical Alliance Carmelo Rosa, Psy.D Director Division of Drug Quality I November 6-17, 2017 Indian Pharmaceutical Alliance DISCLAIMER: The views

More information

Laboratory Accreditation Test Validation: A Brave New World for Anatomic Pathology

Laboratory Accreditation Test Validation: A Brave New World for Anatomic Pathology Laboratory Accreditation Test Validation: A Brave New World for Anatomic Pathology Francis E. Sharkey, MD, FCAP University of Texas Health Science Center, San Antonio, TX Richard W. Brown, MD, FCAP Memorial

More information

TERMS AND DEFINITIONS

TERMS AND DEFINITIONS TERMS AND DEFINITIONS Advisory notice A notice issued by the organization, subsequent to delivery of the medical device, to provide supplementary information and/or to advise what action should be taken

More information

Quality Manual ISO 9001:2008 ISO 9001:2015

Quality Manual ISO 9001:2008 ISO 9001:2015 Quality Manual ISO 9001:2008 ISO 9001:2015 SAE CIRCUITS, INC. 4820 63 rd Street Suite 100 Boulder, CO 80301 USA www.saecircuits.com Table of Contents 1. Company Information 3 2. QMS Scope and Exclusions

More information

Implementation Guide. 3a. Skills Checklist. 3b. Competency Exam. 3c. Answer Key: Competency Exam. 3d. Competency Assessment Checklist

Implementation Guide. 3a. Skills Checklist. 3b. Competency Exam. 3c. Answer Key: Competency Exam. 3d. Competency Assessment Checklist Implementation Guide TABLE OF CONTENTS SECTION 1 Introduction SECTION 2 Pre-Implementation Checklist SECTION 3 Introduction: Operator Training and Competency Certification 3a. Skills Checklist 3b. Competency

More information

Document issued on: January 30, The draft of this guidance document was released on September 7, 2005.

Document issued on: January 30, The draft of this guidance document was released on September 7, 2005. Guidance for Industry and FDA Staff: Recommendations for Clinical Laboratory Improvement Amendments of 1988 (CLIA) Waiver Applications for Manufacturers of In Vitro Diagnostic Devices Document issued on:

More information

Prevent Quality System Deficiencies by Conducting Effective Internal Audits. Whitepaper

Prevent Quality System Deficiencies by Conducting Effective Internal Audits. Whitepaper Prevent Quality System Deficiencies by Conducting Effective Internal Audits Whitepaper An internal audit system is one of the most effective ways to monitor, analyze, control, and improve quality management

More information

SOFTWARE APPS AS MEDICAL DEVICES. The Regulatory Landscape 2015 WHITE PAPER WHITE PAPER PRODUCED BY MAETRICS

SOFTWARE APPS AS MEDICAL DEVICES. The Regulatory Landscape 2015 WHITE PAPER WHITE PAPER PRODUCED BY MAETRICS WHITE PAPER SOFTWARE APPS AS MEDICAL DEVICES The Regulatory Landscape 2015 WHITE PAPER PRODUCED BY MAETRICS For more information, please contact: USA Office: +1 317 706 1493 UK Office: +44 115 921 6200

More information

COMMISSION RECOMMENDATION. of XXX. on the audits and assessments performed by notified bodies in the field of medical devices

COMMISSION RECOMMENDATION. of XXX. on the audits and assessments performed by notified bodies in the field of medical devices EUROPEAN COMMISSION Brussels, XXX [ ](2013) XXX draft COMMISSION RECOMMENDATION of XXX on the audits and assessments performed by notified bodies in the field of medical devices (Text with EEA relevance)

More information

Leveraging ISO s Quality Management System to Achieve 24/7 Inspection Readiness. Randy Querry Accreditation Manager Clinical

Leveraging ISO s Quality Management System to Achieve 24/7 Inspection Readiness. Randy Querry Accreditation Manager Clinical Leveraging ISO 15189 s Quality Management System to Achieve 24/7 Inspection Readiness Randy Querry Accreditation Manager Clinical Goals Provide brief background on A2LA Differentiate between CLIA and ISO

More information

11.0 FDA-Regulated Research Research Involving Investigational Drugs and Biologics

11.0 FDA-Regulated Research Research Involving Investigational Drugs and Biologics 11.0 FDA-Regulated Research The IRB evaluates the safety or efficacy of all drugs and devices used in research. Studies involving unapproved or investigational drugs or devices will be reviewed to ensure

More information

FDA s Proposed Regulatory Changes to Laboratory-Developed Tests (LDTs) By Roz Sweeney, Ph.D., Nerac Analyst

FDA s Proposed Regulatory Changes to Laboratory-Developed Tests (LDTs) By Roz Sweeney, Ph.D., Nerac Analyst FDA s Proposed Regulatory Changes to Laboratory-Developed Tests (LDTs) By Roz Sweeney, Ph.D., Nerac Analyst Introduction The FDA recently issued updated draft guidance for proposed regulatory changes affecting

More information

WHO guidelines on quality risk management

WHO guidelines on quality risk management Annex 2 WHO guidelines on quality risk management 1. Introduction 62 1.1 Background and scope 62 1.2 Principles of quality risk management 64 2. Glossary 67 3. Quality risk management process 70 3.1 Initiating

More information

ROLE OF ICH GUIDELINES IN PHARMACEUTICAL QUALITY RISK MANGAMENT- REVIEW

ROLE OF ICH GUIDELINES IN PHARMACEUTICAL QUALITY RISK MANGAMENT- REVIEW ejpmr, 2016,3(7), 161-169. SJIF Impact Factor 3.628 Review Article EUROPEAN JOURNAL European OF PHARMACEUTICAL Journal of Pharmaceutical and Medical Research AND MEDICAL RESEARCH ISSN 2394-3211 www.ejpmr.com

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD FDIS ISO/IEC 17025 General requirements for the competence of testing and calibration laboratories Exigences générales concernant la compétence des laboratoires d'étalonnages et

More information

UNIT 10 CLAUSE-WISE INTERPRETATION OF ISO 22000: 2005

UNIT 10 CLAUSE-WISE INTERPRETATION OF ISO 22000: 2005 ISO 22000:2005 UNIT 10 CLAUSE-WISE INTERPRETATION OF ISO 22000: 2005 Structure 10.0 Objectives 10.1 Introduction 10.2 Clause-wise Explanation of the Standard 10.2.1 Clause 1: Scope 10.2.2 Clause 2: Normative

More information

Risk Management in Clinical Trials in Today s ICH-GCP(R2) Framework

Risk Management in Clinical Trials in Today s ICH-GCP(R2) Framework Clinical Research in Resource Limited Settings: Mission Impossible or Role Model for Future Drug Development? Risk Management in Clinical Trials in Today s ICH-GCP(R2) Framework Ingrid Klingmann, MD, PhD,

More information

Clarifying digital health and software regulation: FDA releases three new guidance documents

Clarifying digital health and software regulation: FDA releases three new guidance documents Clarifying digital health and software regulation: FDA releases three new guidance documents December 15, 2017 On December 7, 2017, the Food and Drug Administration (FDA or the Agency) released three guidance

More information

Software Recalls Issues and Challenges

Software Recalls Issues and Challenges ASQ: Meeting the Compliance Challenge Perspectives from Government, Industry and Quality Professionals: April 28, 2010 Software Recalls Issues and Challenges Jeme Wallace Director Regulatory Affairs GE

More information

ISO 22000:2005 Standard INTERNATIONAL STANDARDS REGISTRATIONS

ISO 22000:2005 Standard INTERNATIONAL STANDARDS REGISTRATIONS ISO 22000:2005 Standard Food Safety Management System INTERNATIONAL STANDARDS REGISTRATIONS 3.1 FOOD SAFETY concept that food will not cause harm to the consumer when it is prepared and/or eaten according

More information

GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES

GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES Foreword This guidance document is informative and advisory and has no legal authority. Individual national enforcement authorities are bound

More information

IQCP : Where Do I Start? Developing a Successful IQCP: Let s Keep it Simple. Objectives 7/4/2015

IQCP : Where Do I Start? Developing a Successful IQCP: Let s Keep it Simple. Objectives 7/4/2015 Developing a Successful IQCP: Let s Keep it Simple Elia Mears, MHA MT(ASCP)SM Field Surveyor The Joint Commission Stacy Olea, MBA, MT(ASCP), FACHE Executive Director Laboratory Accreditation The Joint

More information

1/23/ CLIA CMS 2004 brochure on how to complete the initial performance verification. p.2

1/23/ CLIA CMS 2004 brochure on how to complete the initial performance verification. p.2 Disclaimers I work for Alere and Alere produces testing devices for use at the POC so I have more data on the testing areas my company covers. The handouts. The Point of Care Quality Control Debate Thomas

More information