2 E1 Tapered Liners Design Rationale
3 Disclaimer This publication and all content, artwork, photographs, names, logos and marks contained in it are protected by copyright, trademarks and other intellectual property rights owned by Biomet or its affiliates or licensed to Biomet or its affiliates. This brochure must not be used, copied or reproduced in whole or in part for any purposes other than marketing by Biomet or its authorised representatives. For other purposes is prohibited. Biomet does not practice medicine and does not recommend any particular orthopaedic implant or surgical technique for use on a specific patient. The surgeon who performs any implant procedure is responsible for determining and utilising the appropriate techniques for implanting prosthesis in each individual patient. Biomet is not responsible for selection of the appropriate orthopaedic implant or surgical technique, nor does it advocate a particular technique to be utilised on an individual patient. Biomet 2011 E1 is a registered trademark of Biomet UK Ltd.
4 E1 Tapered Liners Design Rationale Contents Ultra Low Wear 2 Tri Bearing Options 3 Easier Revisions 4 Reduced Cup Inventory 4 Maximised Bearing Diameter 4 Optimal Mobility and Limited Risk of Dislocation 5 Stability 5 Fatigue Resistance 5 Liner Options 6 Ordering Information 7 References 9
5 E1 Tapered Liners Design Rationale The E1 tapered Liners complement the existing family of low wear hard bearing liner options used with the Exceed ABT taperfit cups. E1 Antioxidant Infused Technology prevents oxidative degradation of the polyethylene, and offers an alternative ultra low wear soft bearing option. The tri-bearing Exceed ABT acetabular platform addresses patients and practice needs by combining technological development in large diameter bearings with clinically proven fixation methods. Ultra Low Wear E1 Antioxidant Infused Technology surpasses the limitations of first generation highly crosslinked polyethylene. This unique technology is the first vitamin E stabilized highly crosslinked polyethylene bearing that has been shown in testing to provide high mechanical strength and true oxidative stability: Ultra low wear rates with large heads: 95 percent wear reduction when comparing 40mm E1 liners with 36mm ArCom liners and 99 percent wear reduction with smaller femoral heads when compared with already low wear rates of ArComXL polyethylene (figure 1 & 2) Volumetric Wear Rate mm 3 /10 6 Cycles Clean Conditions Volumetric Wear 5 million cycles on a hip simulator 36mm and 40mm head size 55.4 Fig Abrasive Conditions 40mm E1 36mm ArCom* Fatigue strength greater than first generation re-melted HXLPE 1,2,3 Stronger than sequentially crosslinked and annealed materials after an environmental stress cracking (ESC) study Small Diameter Volumetric Wear 1 5 million cycles on a hip simulator 28mm head size 53.3 True oxidative protection without re-melting: no evidence of measurable oxidation 1 Volumetric Wear Rate mm 3 /10 6 Cycles ArCom Polyethylene* ArComXL 0.24 E1 2 Fig 2
6 E1 Tapered Liners Design Rationale Tri-Bearing Options The Exceed-ABT acetabular system is a highly versatile multi-bearing platform capable of addressing the needs of both primary and revision hip surgery. The facility to select intra-operatively between C2A- Delta (ceramic), M2A (CoCr) and E1 large diameter bearings optimises implant function in line with clinical requirements and patient activity. Exceed ABT TaperFit Exceed ABT Ringloc-X Standard TaperFit Shell 15 TaperFit Shell 3-Hole Shell Multi-Hole Shell C2A Ceramic Liner E1 Liner M2A Metal Liner E1 Standard Liner E1 High-Wall Liner E1 10 Liner C2A Ceramic Head M2A CoCrMo Head 3
7 E1 Tapered Liners Design Rationale Easier Revisions The straightforward replacement of cobalt chromium or Biolox Delta ceramic liners with the E1 tapered liners makes revision cases easier, eliminating the need for an acetabular cup replacement. In addition, the 10-degree E1 Liner can be an ideal revision option should the acetabular cup be too vertically positioned. Reduced Cup Inventory If they wish, surgeons willing to use a soft bearing liner can use the Exceed-ABT Taperfit cup instead of the Exceed-ABT Ringloc- X Cup. The E1 line extension will permit hospitals to reduce their Exceed-ABT cup inventory, without having to compromise their choice of optimal bearing options for their patients. Maximised Bearing Diameter E1 tapered liners are made with E1 polyethylene inserts pressed into 1.5mm thick titanium tapered sleeves. A built-in offset (Fig 3) helps maximise the bearing diameter (28, 32, 36mm) without compromising the minimum polyethylene thickness (5.7mm minimum nominal thickness), and therefore without compromising the polyethylene mechanical properties. Fig 3 Fig 3 45 Thickness 4
8 E1 Tapered Liners Design Rationale Optimal Mobility and Limited Risk of Dislocation The E1 tapered liners have been developed to maximise the patient range of motion (Fig 4). The most critical head/liner/stem implant combinations have been tested and fully satisfy the range of motion performance requirements Internal/External Rotation 115 Abduction/Adduction Angle 135 Flexion/Extension Angle The range of motion has been maximised whilst preserving the femoral head coverage. Between 179 and 181 degrees of head coverage is achieved across the range 5 in order to limit the risk of implant dislocation. Stability Fig 4 Like the Biolox delta ceramic and cobalt chromium liners, the titanium sleeve of the E1 tapered liners solidly locks into the shell thanks to the Ceralock taper interface, technology from Ceramtec. The titanium sleeve has a circumferential retaining groove which prevents micro-motion of the inner E1 liners. The surface finish of the metal back combined with the tight press-fit of the inner E1 liner address risks of component rotation. Fig 5 The E1 tapered liner underwent series of levelout, push-out and torsion-out mechanical tests. It has been demonstrated that the E1 tapered liner provides excellent resistance to mechanical forces and minimises the risk of polyethylene disengagement from its titanium sleeve 6. Fatigue Resistance Rim impingement test (Fig 6) were carried-out by an independent institute. The 2 million cycle test unveiled the absence of cracks, deformation, delamination of the polyethylene subsequent to loading 7. Fig 6 Biolox and Ceralock are Ceramtec Brands 5
9 E1 Tapered Liners Design Rationale Liner Options Additional Lateralisation The E1 tapered liners are available in bearing sizes 28, 32 and 36mm in standard and 10-degree configurations. The 10-degree liners address vertically positioned acetabular shells, and help restore the biomechanics of the hip. Compared with the standard liner, It provides an additional 10-degree inclination/version depending upon placement of the liner in the cup. The Femoral head is therefore lateralised in comparison to the standard E1 tapered liners (Fig 7). The 10-degree liner has a radiolucent pin to check the correct rotational position of the liner inside the acetabular shell. Note: The 10-degree liner cannot be used in conjunction with the Exceed ABT Taperfit 15-degree shell. Standard E1 Tapered Liner 10 E1 Tapered Liner Fig 7 6
10 Ordering Information Implants Size (mm) Ø46 Ø48 Ø50 Ø52 Ø54 Ø56 Ø58 Ø60 Ø62 Ø64 Ø66 Ø68 Ø70 E1 tapered Liners* Ø28 Ø32 Ø36 Std 10 Std 10 Std 10 E E E E E E E E E E *The colours correspond to those on the shell, liner labels and related trial liners. Exceed ABT Taperfit Acetabular Components Size (mm) Standard TaperFit Shells PC HA/PC BM/PC Size (mm) 15 Degree TaperFit Shells PC HA/PC BM/PC Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM Ø HA BM *The colours correspond to those on the shell and liner labels. 7
11 Ordering Information Instruments Product Description Order Code 1 Exceed-ABT E1 Tapered Trial Liners Complete Set E Exceed-ABT E1 Tapered Trial Liners Empty Case E E1 Tapered Trial Liner 0-28/37 E T 1 E1 Tapered Trial Liner 0-32/39 E T 1 E1 Tapered Trial Liner 0-36/44 E T 1 E1 Tapered Trial Liner 0-36/48 E T 1 E1 Tapered Trial Liner 0-36/52 E T 1 E1 Tapered Trial Liner 10-28/37 E T 1 E1 Tapered Trial Liner 10-32/39 E T 1 E1 Tapered Trial Liner 10-36/44 E T 1 E1 Tapered Trial Liner 10-36/48 E T 1 E1 Tapered Trial Liner 10-36/52 E T 8
12 References 1. Data on file at Biomet. Bench test results not necessarily indicative of clinical performance. 2. Wannomae, K. Environmental Stress Cracking of Two-Tocopherol Doped, Irradiated UHMWPEs and Two Contemporary UHMWPEs. Report Provided by the Orthopaedic Biomechanics and Biomaterials Laboratory at Massachusetts General Hospital. January 12, Bhambri S., et al. The effect of aging on mechanical properties of melt-annealed highly crosslinked UHMWPE. Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene for Joint Replacements , Data on file at Biomet; Report E1 tapered Ensayo Rango Movimiento Angular Data on file at Biomet; Report E1 tapered Head coverage angle Data on file at Biomet; Report E1 tapered Level-out, Push-out, torsional test LE-061SO/ 7. Data on file at Biomet: Report E1 tapered liner Rim Impingement test Instituto de Biomecanica de Valencia 9
13 10 Notes
14 Notes 11
15
16
References. One Surgeon. One Patient. FLM /08
FLM 196 06/08 References 1. Biomet Biomaterials Laboratory The revolutionary second generation vitamin E stabilsed highlycrosslinked UHMWPE Jan 2007. 2. Wannomnnomae, K. Environmental Stress Cracking of
More informationProven 1-4, Safe Performance. Product overview
Proven 1-4, Safe Performance Product overview DePuy Polyethylene: proven 1-4, safe performance Polyethylene has been used as an orthopaedic bearing material for more than 4 years. Over this period of time
More informationReCap Technical Design Rationale
ReCap Technical Design Rationale ReCap Resurfacing Hip System System Description The ReCap Metal-Metal Resurfacing System removes a minimal amount of bone from the femur and acetabulum and reproduces the
More informationVitamin E technology Overview. Locking in Life *
Vitamin E technology Overview Locking in Life * Ultra High Molecular Weight Polyethylene (UHMWPE) has proven itself over many decades as the gold standard bearing choice in total joint arthroplasty 1.
More informationDuraCer. Ceramic Components. The Strong Choice In Ceramics
The Strong Choice In Ceramics is a magnesia-stabilized zirconia (MgPSZ) ceramic material whose combination of physical properties is attractive for use in orthopaedic surgery. Mg-PSZ ceramic often is referred
More informationJoint Replacement Implants Hip Joint Prostheses
Translated English of Chinese Standard: YY0118-2016 www.chinesestandard.net Sales@ChineseStandard.net PHARMACEUTICAL INDUSTRY STANDARD YY OF THE PEOPLE S REPUBLIC OF CHINA ICS 11.040.40 C 35 YY 0118-2016
More informationLet s Talk. Science... The Science of: > > > Motion13. Wear1. Triathlon. Knee System
Let s Talk Science... The Science of: > Motion13 > > Fit6 Wear1 Triathlon Knee System The Science of Increased Motion High flexion knee system designed for mobility with stability through 150 degrees of
More informationCeramics in Orthopaedic and Neurosurgery. B. Sonny Bal, MD MBA JD PhD University of Missouri-Columbia Amedica Corporation
Ceramics in Orthopaedic and Neurosurgery B. Sonny Bal, MD MBA JD PhD University of Missouri-Columbia Amedica Corporation Historic Concern: Wear Bearing Failures of the Past The Rationale for Alternatives
More informationTrabecular Metal Revision Shell. Surgical Technique IMAGE TO COME. The Best Thing Next to Bone
Trabecular Metal Revision Shell Surgical Technique IMAGE TO COME The Best Thing Next to Bone Trabecular Metal Revision Shell Surgical Technique Initial and Long-Term Stability Fully-interconnected trabecular
More informationThe information contained in this document is intended for healthcare professionals only.
The information contained in this document is intended for healthcare professionals only. Orthopaedics Modular Taper Junction and Neck Strength in Total Hip Arthroplasty Re s tora ti on Modular Revi s
More informationThe Required Mechanical Properties. of Hip and Knee Components
The Required Mechanical Properties of Hip and Knee Components S. M. Kurtz, Ph.D. Drexel University and Exponent, Inc., Philadelphia PA, USA Correspondence addresses: Implant Research Center, School of
More informationBiodynamics medical technology is the centre of excellence in the area of endoprosthetics and is specialised in joint implants and instruments.
2 3 Biodynamics medical technology is the centre of excellence in the area of endoprosthetics and is specialised in joint implants and instruments. CUSTOMER ORIENTATION AS A MATTER OF PRINCIPLE We stand
More informationInfluence of the diameter and the material type on the contact pressure in the acetabular component of a total hip prosthesis.
Influence of the diameter and the material type on the contact pressure in the acetabular component of a total hip prosthesis. Diogo Moreira Campos Ferreira de Almeida Instituto Superior Técnico Lisboa,
More informationvitamys The E-Factor Makes the Difference
vitamys The E-Factor Makes the Difference vitamys Mathys new highly cross-linked polyethylene Introduction vitamys is the new and innovative highly cross-linked polyethylene (HXLPE) from Mathys. Designed
More informationTHE TRUTH ABOUT HIGHLY CROSSLINKED POLYETHYLENE
THE TRUTH ABOUT HIGHLY CROSSLINKED POLYETHYLENE BIOMET BROCHURE ANALYSIS HIGHLY CROSSLINKED POLYETHYLENE, A STEP FORWARD OR A STEP BACK Biomet recently produced a brochure entitled Highly Crosslinked Polyethylene,
More informationCeramic on XLPE My Choice. Javad Parvizi MD, FRCS Professor Orthopedic Surgery Rothman Institute at Thomas Jefferson Uuiversity
Ceramic on XLPE My Choice Javad Parvizi MD, FRCS Professor Orthopedic Surgery Rothman Institute at Thomas Jefferson Uuiversity Research support: NIH OREF Stryker Orthopedics Depuy Zimmer Baxter 3M Biomemetics
More informationALTERNATIVE BEARING SURFACES: THE GOOD, BAD & UGLY
ALTERNATIVE BEARING SURFACES: THE GOOD, BAD & UGLY AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS 69th Annual Meeting February 13-17, 2002 Dallas, Texas COMMITTEE ON BIOMEDICAL ENGINEERING COMMITTEE ON HIP AND
More informationThe perfect equation for hips
The perfect equation for hips OXINIUM Oxidized Zirconium (proprietary ceramicised metal) + = XLPE (Smith & Nephew unique formula) VERILAST for Hips (Smith & Nephew Technology) Ceramicised metal has unmatched
More informationPlatelet Concentrate in Total Knees BIOLOGICS. This brochure is for International use only. It is not for distribution in the United States.
Platelet Concentrate in Total Knees BIOLOGICS This brochure is for International use only. It is not for distribution in the United States. Platelet Concentrate in Total Knees Total Knee Arthroplasty Total
More informationSETTING THE STANDARD FOR ORTHOPAEDIC PLASTICS
SETTING THE STANDARD FOR ORTHOPAEDIC PLASTICS PREMIUM ORTHOPAEDIC COMPONENTS THAT TRANSFORM LIVES A WORLD LEADER IN IMPLANTABLE PLASTIC MANUFACTURING Orthoplastics Limited is dedicated to the medical industry
More informationHighly Crosslinked Polyethylene for Total Hip Arthroplasty. William H Harris, MD, D.Sc
Highly Crosslinked Polyethylene for Total Hip Arthroplasty William H Harris, MD, D.Sc Alan Gerry Clinical Professor of Orthopædic Surgery Harvard Medical School Massachusetts General Hospital Reduction
More informationShort Aramid Fibre Reinforced Ultra-High Molecular Weight Polyethylene Composite. A new hip-prosthesis Hofsté, Joanna Maria
University of Groningen Short Aramid Fibre Reinforced Ultra-High Molecular Weight Polyethylene Composite. A new hip-prosthesis Hofsté, Joanna Maria IMPORTANT NOTE: You are advised to consult the publisher's
More informationCoAxial Spray Kit And Applicator Tips. Simplifying the Application of Platelet and Plasma Concentrate
CoAxial Spray Kit And Applicator Tips Simplifying the Application of Platelet and Plasma Concentrate Intended for international use only. Indications for use of products and /or therapies contained herein
More informationPOLYETHYLENE [UHMW PE]
POLYETHYLENE [UHMW PE] Charnley originally chose polytetrafluoroethylene (PTFE) as a bearing material sometime between 1956 and 1958, based on its general chemical inertness and low coefficient of friction.
More informationVERILAST Oxidized Zirconium with XLPE
Figure HT25: Cumulative Percent Revision of Primary Total Conventional Hip Replacement by Bearing Surface (Primary Diagnosis OA) Cumulative Percent Revision Ceramic/Ceramic Ceramic/Non Cross-Linked Polyethylene
More informationHas anyone really thought about the requirements for drilling living human bone tissue until now? One simple solution.
Has anyone really thought about the requirements for drilling living human bone tissue until now? One simple solution One easy decision Patents Pending Analysing the Needs For decades, orthopaedic surgery
More informationOverview of ASTM Subcommittee F04.15 on Material Test Methods Terry O. Woods, Ph.D. FDA Center for Devices & Radiological Health
Overview of ASTM Subcommittee F04.15 on Material Test Methods Terry O. Woods, Ph.D. FDA Center for Devices & Radiological Health PERU Workshop on Medical Device Regulation and Standards: Policy and Technical
More informationAdvanced Surface Technology
Advanced Surface Technology Why Take a Chance With Anything Else? Aesculap Orthopaedics 2 By 2030 the demand for total knee arthroplasty (TKA) in the United States is predicted to grow by 673%. 1, 2 As
More informationDate July 16, 2015 Court Intellectual Property High Court Case number 2015 (Gyo-Ke) 10002
Date July 16, 2015 Court Intellectual Property High Court Case number 2015 (Gyo-Ke) 10002 Fourth Division A case in which, with respect to a trademark consisting of a combination of a figure and a color
More informationX-LINKed UHMWPE. The new material from LINK. Materials
X-LINKed The new material from LINK Materials Introduction/Specifications The new material from LINK X-LINKed is a highly crosslinked material combining maximum durability with optimised properties. Looking
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. ELEMENT CINCINNATI 3701 Port Union Road Fairfield, Ohio Justin Riebesel Phone:
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ELEMENT CINCINNATI 3701 Port Union Road Fairfield, Ohio 45014 Justin Riebesel Phone: 513 771 2536 MECHANICAL Valid To: June 30, 2018 Certificate Number: 2422.01
More informationWe report two cases of surface deterioration of a
Phase transformation of a zirconia ceramic head after total hip arthroplasty K. Haraguchi, N. Sugano, T. Nishii, H. Miki, K. Oka, H. Yoshikawa From Osaka University Medical School, Japan We report two
More informationMEDPOR. Orbital Floor
MEDPOR Orbital Floor Orbital Fracture Repair Available in a variety of shapes, sizes, and configurations, MEDPOR and MEDPOR TITAN Implants provide surgeons with an expanding range of options for reconstruction
More informationCONSERVE Total A-CLASS ADVANCED METAL WITH BFH TECHNOLOGY HIP SYSTEM
CONSERVE Total A-CLASS ADVANCED METAL WITH BFH TECHNOLOGY HIP SYSTEM T E C H N I C A L M O N O G R A P H 2.50 Wear (mm 3 ) 2.00 1.50 1.00 0.50 Second-Generation: 54mm Metal-Metal Third Generation: A-CLASS
More informationVolumetric Wear Analysis of Hip Joint Implants by Optical Methods. Matúš Ranuša. Ph.D. Defense Brno, 8th January Doc. Ing. Martin Vrbka, Ph.D.
Volumetric Wear Analysis of Hip Joint Implants by Optical Methods Matúš Ranuša Doc. Ing. Martin Vrbka, Ph.D. Ph.D. Defense Brno, 8th January 2018 Motivation 101 651 Surgeries / year 2016 NJR for England,
More informationINDUSTRY OVERVIEW SOURCE OF INFORMATION
The information presented in this section, including certain facts, statistics and data, is derived from various official government publications and other publications and from the market research report
More informationHydroSet. Injectable HA Bone Substitute
HydroSet Injectable HA Bone Substitute The Latest Advancement in Bone Substitute Technology Fast Setting Injectable or Manual Implantation Osteoconductive Excellent Wet-field Characteristics Isothermic
More informationSafe Harbor Statement
Safe Harbor Statement This presentation may contain forward-looking statements as defined in the Private Securities Litigation Reform Act of 1995. These statements are based on current expectations of
More informationCeramic Implant Evaluations and OXINIUM Technology Joint Replacement with Oxidized Zirconium Bearing Components
Gordon Hunter, Ph.D. April 2017 Ceramic Implant Evaluations and OXINIUM Technology Joint Replacement with Oxidized Zirconium Bearing Components Trademark of Smith & Nephew. Patient Trends Younger and living
More informationcompany products development production innovation service customized products co-operation
Dear Ladies and Gentlemen, since our founding in the year 1988, implantcast GmbH can look back on many successful years. Constant changes both within the company and the medical sector have smoothed our
More informationEasyClip Xpress Sterile staple solution. Operative technique
EasyClip Xpress Sterile staple solution Operative technique Sterile staple solution Table of contents Indications 3 Operative technique 4 This publication sets forth detailed recommended procedures for
More informationThe Biomechanics of ProLayer Acellular Dermal Matrix: suture retention strength
The Biomechanics of ProLayer Acellular Dermal Matrix: suture retention strength Reginald Stilwell, B.S., C.T.B.S., Ryan Delaney, M.S. ProLayer, Centennial, CO Basic science volume 2 ProLayer Acellular
More informationBiomaterials in Medical device design
Biomaterials in Medical device design 1 Dr Joseph Buhagiar B. E n g ( H o n s. ) P h D ( B h a m ) D e p a r t m e n t o f M e t a l l u r g y a n d M a t e r i a l s E n g i n e e r i n g U n i v e r
More informationHANDLING OF WRIGHT MEDICAL DISPOSABLE PROPHECY TM ANKLE INSTRUMENTS
HANDLING OF WRIGHT MEDICAL DISPOSABLE PROPHECY TM ANKLE INSTRUMENTS 153290-0 The following languages are included in this packet: English (en) For additional languages, visit our website www.wright.com
More informationURGENT: Field Safety Notice. ABGII Modular Stems and ABGII Modular Necks Rejuvenate Modular Stems and Rejuvenate Modular Necks
Field Safety Notice RA2012-067 April XX, 2012 URGENT: Field Safety Notice FSCA identifier: Type of Action: Description: Catalog #: Lot Code: Product Field Action RA2012-067 Field Safety Corrective Action:
More informationCHAPTER 5 RETRIEVAL STUDY
101 CHAPTER 5 RETRIEVAL STUDY 5.1 Introduction Retrievals obtained after revision surgery provided valuable information with respect to the causes of wear that take place in-vivo. Note: For better clarity,
More informationExactech Investor Update
Exactech Investor Update Helping surgeons worldwide make patients more mobile Q1 2017 Investment Highlights Targeting $47 billion global biomedical device market Consistent long term growth 2x as fast
More informationCharles W. Jewett, P.E.
Charles W. Jewett, P.E. Senior Engineer Materials & Corrosion Engineering 149 Commonwealth Drive Menlo Park, CA 94025 (650) 688-7088 tel cjewett@exponent.com Professional Profile Mr. Jewett has over 30
More informationThe Orthopaedic Research Laboratory (ORL) performs research in the field of orthopaedics.
The Orthopaedic Research Laboratory (ORL) performs research in the field of orthopaedics. The biomechanical section focusses on bone, soft tissues and on pre-clinical testing of implants. We are specialized
More informationTrabecular Metal Technology
Trabecular Metal Technology Alexandre Nehme Departement of Orthopedic Surgery. Saint Georges University Hospital. Beirut. The cellular structure of Tantallum (Trabecular Metal) resembles bone and approximates
More informationTrabecular Metal Acetabular Revision System Cup-Cage Construct
Trabecular Metal Acetabular Revision System Cup-Cage Construct Surgical Technique The Best Thing Next to Bone Trabecular Metal Acetabular Revision System Cup-Cage Construct Trabecular Metal Acetabular
More informationCOATING TECHNOLOGIES REVIEW:
COATING TECHNOLOGIES REVIEW: How Orthopedic Implants Get Coated April, 2018 By: Tony Crivella Senior Global Commercial Manager, Coatings Parimal Bapat, Ph. D Research Engineer Shilesh Jani Engineering
More informationThis publication is not intended for distribution in the USA. PRODUCT RATIONALE
This publication is not intended for distribution in the USA. PRODUCT RATIONALE CONTENTS Introduction 2 Proven Clinical Heritage of POROCOAT 4 Built on the foundations of POROCOAT 5 Super Textured Asperity
More informationSaikko, Vesa & Kostamo, Jari RandomPOD - A new method and device for advanced wear simulation of orthopaedic
Powered by TCPDF (www.tcpdf.org) This is an electronic reprint of the original article. This reprint may differ from the original in pagination and typographic detail. Author(s): Title: Saikko, Vesa &
More informationHip Joint Tribology. Wear and Wear Mechanisms. Friction and Lubrication. Tribology Testing Using Hip Simulators. Amir Kamali
4 Hip Joint Tribology Amir Kamali According to the ASM International handbook, tribology is defined as the science and technology of interacting surfaces in relative motion and all practices related thereto.
More informationZIMMER 5-POINT P O LY E T H Y L E N E. One Source Addresses All The Poly Wear Issues.
ZIMMER 5-POINT P O LY E T H Y L E N E One Source Addresses All The Poly Wear Issues. Performance Proven By M M a t e r i a l The polyethylene used in orthopaedic implants today is not all of equal quality.
More informationREVA s Bioresorbable Scaffold Program Update. Alexandre Abizaid, MD, PhD, FACC Instituto Dante Pazzanese de Cardiologia Sao Paulo, Brazil
REVA s Bioresorbable Scaffold Program Update Alexandre Abizaid, MD, PhD, FACC Instituto Dante Pazzanese de Cardiologia Sao Paulo, Brazil REVA s Technology Transition to the Fantom Scaffold 2011: ReZolve
More informationAn Overview of Silicon Nitride as a Novel Biomaterial
An Overview of Silicon Nitride as a Novel Biomaterial Bryan J. McEntire, MBA, Alan Lakshminarayanan, Ph.D., B. Sonny Bal, MD, MBA, JD, and Thomas J. Webster, Ph.D. Si 3 N 4 has many industrial uses Bearings
More informationAssessment of Bore-Cone Taper Junctions in Explanted Modular Total Hip Replacements
Clemson University TigerPrints All Theses Theses 5-2015 Assessment of Bore-Cone Taper Junctions in Explanted Modular Total Hip Replacements Christine Margaret Stamer Clemson University Follow this and
More informationMechanical Properties of Hip Joint Heads Made of the Dispersion Ceramic - Alumina Toughened Zirconia
Key Engineering Materials Vols. 309-311 (2006) pp 1257-1260 Online: 2006-05-15 (2006) Trans Tech Publications, Switzerland doi:10.4028/www.scientific.net/kem.309-311.1257 Mechanical Properties of Hip Joint
More informationIntroduction to Materials Science & Engineering
Introduction to Materials Science & Engineering Course Objective... Introduce fundamental concepts in Materials Science You will learn about: material structure how structure dictates properties how processing
More informationSilForce* SL6625. Technical Data Sheet. SilForce* SL6625 Release Coating
Technical Data Sheet SilForce* SL6625 SilForce* SL6625 Release Coating Description SL6625 release coating is a pre-blended version of the SL6100-D1 addition curing, thermal solventless release coating
More informationMore than surface deep
Tritanium PL Posterior Lumbar Cage More than surface deep Featuring Tritanium In-Growth Technology: 1 TM Built to fuse Tritanium In-Growth Technology1 Stryker s proprietary Tritanium In-Growth Technology,
More informationStructural Optimization of a Total Replacement Hip Prosthesis
Structural Optimization of a Total Replacement Hip Prosthesis M. Grasso, F. Penta, G.P. Pucillo, V. Rosiello, A. Piscopo Abstract Planted articular joints replace bone articulation allowing the transfer
More informationSilForce* SL7562S Solventless Coating
Technical Data Sheet SilForce* SL7562S Solventless Coating Description SilForce SL7562S thermal solventless release coating is an excellent candidate to consider for glassine papers as well as for filmic
More informationBioresorbable Scaffold Program Update. Alexandre Abizaid, MD, PhD, FACC Instituto Dante Pazzanese de Cardiologia Sao Paulo, Brazil
Bioresorbable Scaffold Program Update Alexandre Abizaid, MD, PhD, FACC Instituto Dante Pazzanese de Cardiologia Sao Paulo, Brazil Disclosure Statement of Financial Interest Within the past 12 months, I
More informationMEDICAL HERMETIC DEVICES. Electrical Feedthrus and Windows for Hermetic Medical Applications
MEDICAL HERMETIC DEVICES Electrical Feedthrus and Windows for Hermetic Medical Applications Hermetic Medical Devices Overview Petrarottova - Fotolia.com Windows Electrical Feedthrus Overview Presentation
More informationSLAM* Technical Data Sheet
Technical Data Sheet SLAM* 3000 Description SLAM* 3000 silicone anti-mist additive is an excellent candidate to consider to help suppress the misting of solventless thermally cured silicone release coatings
More informationOK mill The most reliable and efficient mill for raw and cement grinding
mill The most reliable and efficient mill for raw and cement grinding WE DISCOVER POTENTIAL Based on proven technology FLSmidth has supplied the global cement industry with leading Vertical Roller Mill
More informationChoice of Test Machines
Choice of Test Machines 2003 George Plint Wear & Failure Mechanisms The more we can characterize the full scale problem the easier it becomes to ensure that the bench tests we run will provide useful information.
More informationMold Design. 12. Mold Materials. Bong-Kee Lee School of Mechanical Engineering Chonnam National University
12. Mold Materials Bong-Kee Lee Chonnam National University Mold Materials easy toolmaking good performance during production good machining properties ease of hear treatment where hardening is required
More informationA Study of Surface Modification on Current Orthopaedic Bioceramics after In Vivo Exposure to the Physiological Environment
Clemson University TigerPrints All Theses Theses 12-2014 A Study of Surface Modification on Current Orthopaedic Bioceramics after In Vivo Exposure to the Physiological Environment Alexander William Wusylko
More informationAdvancing the Open Ventral Hernia Repair Experience
Advancing the Open Ventral Hernia Repair Experience SOFT TISSUE REPAIR Right Procedure. Right Product. Right Outcome. Hernia Repair Fixation Absorbable Fixation System Optimized Design for Open Ventral
More informationNew creative textile construction using innovative braiding technology and materials
White Paper New creative textile construction using innovative braiding technology and materials DSM and Meister Provide Foundation for Medical Device Design Philippe Gédet, Engineer, Meister & Cie AG
More informationLeksell. Vantage Stereotactic System. Advancing stereotactic neurosurgery
Leksell Vantage Stereotactic System Advancing stereotactic neurosurgery Innovation built on strong foundations Since the founding of our company by Lars Leksell more than 45 years ago, Elekta has delivered
More informationMetallurgy 101 (by popular request)
Metallurgy 101 (by popular request) Metals are crystalline materials Although electrons are not shared between neighboring atoms in the lattice, the atoms of a metal are effectively covalently bonded.
More informationCrystallinity in Polymers. Polymers. Polymer Crystallinity. Outline. Crystallinity in Polymers. Introduction. % crystallinity 100
Outline Polymers Introduction Crystallinity Stress relaxation Advanced polymers - applications Crystallinity in Polymers Although it may at first seem surprising, polymers can form crystal structures (all
More informationTitle:Implant Based Differences in Adverse Local Tissue Reaction in Failed Total Hip Arthroplasties: A Morphological and Immunohistochemical Study
Author's response to reviews Title:Implant Based Differences in Adverse Local Tissue Reaction in Failed Total Hip Arthroplasties: A Morphological and Immunohistochemical Study Authors: Giorgio Perino (perinog@hss.edu)
More informationmediatek mediatek Library and archive shelving systems Taking shelving to the next level BC004.03
mediatek Library and archive shelving systems Taking shelving to the next level BC004.03 mediatek Library and archive shelving systems The Mediatek range of shelving is designed to offer space efficient
More informationMedian Sternotomy Closure Utilizing Platelet Concentrate BIOLOGICS
Median Sternotomy Closure Utilizing Platelet Concentrate BIOLOGICS This brochure is for International use only. It is not for distribution in the United States. The Surgical Procedure: Median Sternotomy
More informationCOOLING OF ROLLS USED IN HOT ROLLING OF LONG PRODUCTS
COOLING OF ROLLS USED IN HOT ROLLING OF LONG PRODUCTS M. Raudensky, J. Horsky, P. Kotrbacek, M. Pohanka Brno University of Technology, Heat Transfer and Fluid Flow Laboratory, Technicka 2896/2, 616 69
More informationSTRING THEORY: AN EXAMINATION OF THE PROPERTIES OF HIGH STRENGTH SUTURE MATERIALS
STRING THEORY: AN EXAMINATION OF THE PROPERTIES OF HIGH STRENGTH SUTURE MATERIALS A. Jhamb, J. Goldberg, W. Harper, A. Butler, P.J. Smitham, W.R. Walsh Surgical & Orthopaedic Research Laboratories University
More informationYOUR VALVE SPECIALIST
YOUR VALVE SPECIALIST SAFE AND ECONOMIC Low and high temperature Vacuum and high-pressure Min and Max Flow Control Various applications Abrasive media and corrosive media Small and large valve sizes SHUT-OFF
More informationCoatingImplants. ResearchEngineer OrchidOrthopedicSolutions
2014 CoatingImplants ParimalBapat,Ph.D. ResearchEngineer OrchidOrthopedicSolutions OMTEC 2014 Technical Session: Coating Implants June 11 th 2014 1:30 PM 2:30 PM Presenter: Dr. Parimal V. Bapat Research
More informationUnderfloor Heating System
More than just an Underfloor Heating System including More than just an Underfloor Heating System underfloor heating + acoustic flooring + integrated control technology Performance Technology Integrated
More informationPharmaCote. Durability, Productivity, Longevity. MKT-01 Issue 8
PharmaCote Durability, Productivity, Longevity. MKT-01 Issue 8 Contents Description Page Introduction 4 PharmaCote HC (Hard Chromium) 5 PharmaCote HC+ (Hard Chromium Plus) 6 PharmaCote EC (E-Chrome) 7
More informationStrength Characterization of Ceramic Femoral Heads of Total Hip Arthroplasties using a
Strength Characterization of Ceramic Femoral Heads of Total Hip Arthroplasties using a Fluid Pressure Controlled Destructive Proof Test and Finite Element Analysis A Thesis Submitted to the Faculty of
More informationInterline The next generation chemical tank coating
Interline 9001 The next generation chemical tank coating The smarter solution to Introducing new Interline 9001, the next generation in chemical tank coating technology. With enhanced cargo resistance,
More informationWeir Specialty Pumps. WEMCO Torque-Flow
Weir Specialty Pumps WEMCO Torque-Flow HIGH LOW OIL LEVEL WEMCO Torque-Flow Model C WEMCO Torque-Flow Model C Materials of Construction Ni-Hard, High Chrome Configurations Horizontal Sealing Multiple options
More informationZirconia and alumina ceramics in comparison with stainless-steel heads
Zirconia and alumina ceramics in comparison with stainless-steel heads POLYETHYLENE WEAR AFTER A MINIMUM TEN-YEAR FOLLOW-UP P. Hernigou, T. Bahrami From Hôpital Henri Mondor, Creteil, France A lthough
More informationTECHNOLOGY FOR THE WELDER S WORLD. ROBO Compact W600 Rugged, powerful & reliable.
TECHNOLOGY FOR THE WELDER S WORLD. ROBO Compact W600 Rugged, powerful & reliable. www.binzel-abicor.com W600 Three strong performers! Together with the two torch neck types ABIROB W600 and ROBO WH W600,
More informationStructure of Biological Materials
ELEC ENG 3BA3: Structure of Biological Materials Notes for Lecture #8 Tuesday, October 3, 2006 3.3 Mechanics of biomaterials Mechanical issues for implanted biomaterials include: failure (e.g., fracture,
More informationExperiment 4: Heat Engine Cycle
TD-8572A Experiment 4: Heat Engine Cycle Experiment 4: Heat Engine Cycle Equipment Needed Heat Engine/Gas Law Apparatus PASCO Interface and Data Collection Software PASCO Dual Pressure Sensor PASCO Rotary
More informationSoft Solids Research
Soft Solids Research Hydrogels for load-bearing applications Gavin Braithwaite Cambridge Polymer Group, 56 Roland Street, Suite 310 Boston, MA 02129 Cambridge Polymer Group, Inc. Testing, Consultation,
More informationUncompromised accuracy in neuroimaging and treatment
Uncompromised accuracy in neuroimaging and treatment 1 Helping clinicians improve patients lives. 2 Innovation built on strong foundations The tools used by the surgeon must be adapted to the task and,
More informationResearch Article Can Pin-on-Disk Testing Be Used to Assess the Wear Performance of Retrieved UHMWPE Components for Total Joint Arthroplasty?
BioMed Research International, Article ID 581812, 6 pages http://dx.doi.org/10.1155/2014/581812 Research Article Can Pin-on-Disk Testing Be Used to Assess the Wear Performance of Retrieved UHMWPE Components
More informationWear of Mixed Ceramic-on-Ceramic Hip Replacement Bearings under Adverse Edge Loading Conditions Due to Translational Mal-Positioning
Wear of Mixed Ceramic-on-Ceramic Hip Replacement Bearings under Adverse Edge Loading Conditions Due to Translational Mal-Positioning Mazen Al-Hajjar 1, Silvia Carbone 1, Louise M. Jennings 1, Sabine Begand
More information35, rue des Bas-Trévois CS Troyes Cedex - France Tél : +33 (0) Fax : +33 (0) web :
STEEL INDUSTRY 35, rue des Bas-Trévois CS 90104-10003 Troyes Cedex - France Tél : +33 (0)3 25 76 29 50 - Fax : +33 (0)3 25 80 31 62 web : www.sedis.com - e-mail : https://sedis.com/contact Finland Sweden
More informationCONNECTOR B L O C K S H E A R. Design for Composite Structural Action BLOCK SHEAR CONNECTOR APPLICATION. Airspace
B L O C K S H E A R CONNECTOR BLOCK SHEAR CONNECTOR APPLICATION Airspace Design for Composite Structural Action Exterior ashlar-type masonry veneer, typically of clay brick or concrete block masonry, are
More informationHIP IMPLANT ANALYSIS WITH CATIA
99 HIP IMPLANT ANALYSIS WITH CATIA Lecturer PhD. eng. Florin CIOFU, Constantin Brâncuşi University, florincristian33@gmail.com ABSTRACT: CATIA V5R16 is a product of Dassault Systemes, one of the most widely
More information