1 CE update [cytology hematology generalist] The Principles of Flow Cytometry Antony C. Bakke, PhD From the Department of Pathology, Oregon Health Sciences University, Portland, OR On completion of this article, the reader should be familiar with the basic components of a flow cytometer, the principles of its operation, specimen analysis, data presentation, and quality control. Cytology exam 0102 questions and the corresponding answer form are located after the Your Lab Focus section, p 213. The uses of flow cytometry The basic components of flow cytometers Flow cytometry specimen analysis Flow cytometry quality control Flow cytometry is defined as the simultaneous measurement of multiple physical characteristics of a single cell as the cell flows in suspension through a measuring device. It is a subset of the field of cytometry based on the use of specialized instruments. Flow Cytometry Compared With Other Techniques The original flow cytometer was the Coulter counter (Beckman Coulter M) invented by Wallace Coulter in the 1950s. All current automated, non slide-based hematology instruments are flow cytometers. These instruments require cells in suspension and the ability to move cells through the instrument. Blood is the ideal tissue because the cells are already in suspension and can be easily analyzed. The Coulter principle uses the resistance of a cell in an electrical current as it passes through an orifice to determine numbers of particles and the size of the particle. Flow cytometry has specifically come to denote the use of fluorescence measurement, usually with a laser light source. In laser flow cytometers, light scatter is used to measure the intrinsic size and granularity of the cell. In addition, fluorescence can be used to measure extrinsic features such as specific protein expression and nucleic acid content using added reagents, such as fluorescent stains and antibodies. Flow cytometers may be compared with fluorescence microscopes. With a microscope, the light source is a lamp, the objective lens focuses the image, and the eye is the detector. With a flow cytometer, the light source is a laser, an objective lens focuses the light, and the detector is a photomultiplier tube [F1]. With a fluorescence microscope, the observer views cells and recognizes their different sizes. If a fluorescent probe is bound to the cell, a specific-colored fluorescence is emitted. Flow cytometry does not provide a focused image as does a microscope, but it does detect light and transform the signal into statistical data that can be presented in various graphic formats. Components of a Flow Cytometer Optics A laser beam is used in most flow cytometers as the light source to excite a fluorescent tag (eg, a fluorescent antibody) bound to the cells in the specimen. An argon laser, which produces blue light, is the most commonly used, although some instruments add a red helium-neon laser, and occasionally a mercury arc lamp is substituted. Before 207
2 your lab focus 208 striking the cells, the light from either source passes through focusing and shaping optics to form an elliptical beam with a broad, maximal intensity in the center. The beam then passes through the center of a flow chamber, which is a quartz cuvette with an approximately 200-µm inner diameter. The specimen is also injected through the middle of this flow chamber, where the laser beam strikes the cell [F2]. When the laser strikes the cell, light is diffracted around the edges of the cell, producing a diffraction pattern along the path of the laser beam. This scattered light is approximately equivalent to the cell circumference and is the same wavelength as the exciting laser light (usually blue). Termed forward scatter (FSc) or forward angle light scatter (FALS), it is collected along the same axis as the laser beam. Because this is a very strong signal, a simple photodiode can convert the forward-scattered light into an electrical pulse. However, the direct laser beam could enter this photodiode, giving an overwhelming signal. A physical barrier eliminates the direct laser beam by blocking about a 2-degree angle of light along the path of the laser, allowing only the light diffracted at greater angles to pass. In addition to the forward-scattered light, blue laser light is also reflected off the cell and off internal structures in the cell. This is termed side scatter (SSc), right-angle light scatter (RALS), or 90- degree light scatter (90LS) because it is detected by a collection objective set at 90 degrees to the incident laser light. It is a measure of granularity of the cell. Granulocytes have a much larger sidescattered light signal than do lymphocytes ([F1]; see also [F5]). After being focused by the objective lenses, the blue light is reflected by a dichroic mirror. These mirrors direct specific wavelengths of light into the correct detector. The detector is a photomultiplier tube, which converts the light signal into an electrical signal. Fluorescent probes can also be added to the cell for detecting specific molecules in the cell. These probes are typically antibodies to cellular antigens with a covalently attached fluorochrome. The fluorochrome, excited by the laser 1. Labor intensive; rates of 100 cells/min 2. Total number analyzed usually low, 1000 or fewer 3. Morphology and intracellular probe distribution; tissue sections 4. Quantilation of fluorescence limited; eye sensitivity approximately equal to 10 change light, will fluoresce at a longer wavelength. Fluorescein gives a green fluorescence, phycoerythrin (an algal protein) gives orange, and peridinin chlorophyll protein or phycoerythrin coupled to cyanin 5 produces red fluorescence. Some flow cytometers have a second laser that can excite other fluorochromes. Like the side-scattered blue 1. Rapid analysis; rates of 50-4,000 cells/s 2. Total number analyzed can be very high; 5K- 100K+ (for rare events and accurate statistics) 3. Lack of information about intracellular probe distribution 4. Excellent quantitation of fluorescence per cell; range of 10 5 change 5. Single cell suspension critical 6. Correlated multiparameter analysis [F1] Light microscopy compared with flow cytometry for the detection of cells and cellular probes. Reproduced with permission from Becton-Dickinson M. [F2] Light scatter measurements in a flow cytometer. The blue laser beam intersects the cell in the middle of the flow chamber. Forward scatter or forward angle light scatter (FALS) is measured along the axis of the laser beam. Blue laser light reflected off the cell and off internal structures in the cell is termed side scatter or 90-degree light scatter (90LS). Reproduced and altered with permission from Purdue University Cytometry Laboratories. The copyright for the original figure cannot be sold or transferred. light, all of these fluorescent signals pass through the objective set at 90 degrees to the incident laser light. Additional dichroic mirrors reflect specific wavelengths of light, allowing separation of the signals from each fluorochrome with minimal overlap. In front of each photomultiplier, a final bandpass filter precisely defines the wavelength of light entering that detector.
3 your lab focus [F3] Flow cell showing hydrodynamic focusing of the specimen by the surrounding sheath fluid. Reproduced with permission from Becton- Dickinson M. Fluidics The flow cytometer has 2 sources of fluid passing through it: a sheath fluid that is usually a buffered saline, and the specimen containing the cells to be analyzed. The specimen forms a core, which is injected into the middle of the sheath fluid, and both pass through the flow chamber [F3]. The outer sheath fluid causes hydrodynamic focusing, which compresses the specimen core into a small area. Cells are forced to pass singly through the focal point of the laser, which is crucial for collecting accurate information about each cell in the specimen. Electronics The photodiode detectors or photomultipler tubes first collect the light signal from the cells. In both cases, the photons produce a pulse of electrical current. Photomultiplier tubes have a voltage applied across a series of plates. This increases their sensitivity so that fewer photons are required to produce an electrical current. The initial small current requires amplification. The amplified current ranges from 0 to 10 V depending on the strength of the signal. The voltage signal is an analog signal and cannot be recognized by the computer. It is converted by an analog-todigital converter into a digital signal, called a channel, which can range from 0 to 1,023 channels. The channel is a counter, and all signals of a specific voltage are counted in a channel corresponding to the strength of the signal. The amplification of the initial signal can be either linear or logarithmic. Linear amplification provides a direct visual relationship and is useful for scatter signals and fluorescent measurements of DNA. Linear amplification allows a dynamic range of approximately 200- to 1,000-fold. Logarithmic amplification is used for most other biologic signals, including immunofluorescence, owing to the extreme dynamic range of these signals. Logarithmic amplification gives a dynamic range of 2,000- to 10,000-fold. The fluorescent signals have been discussed above. Although these signals appear mainly as a single color under the microscope, actually the emission from any fluorochrome is broad. Although fluorescein isothiocyanate (FITC) appears apple-green, it also has lower-level emissions in the orange area of the spectrum. Hence, any green FITC-stained cell will also generate a signal in the orange channel, making the cell appear to be labeled by 2 fluorochromes. To allow multicolor fluorescence, this apparent overlap must be eliminated. This is done by fluorescence compensation. A percentage of any bright signal, such as FITC, is subtracted from any other associated signal. The instrument is set up at the beginning of a run with pure signals of each fluorochrome, and any associated fluorescence at the other wavelengths is compensated [F4]. a b c 209 [F4] Electronic compensation corrects spectral emission overlaps. Two-parameter contour plots of fluorescein vs phycoerythrin fluorescence. a, No compensation produces results with many cells apparently expressing both fluorochromes, although these are mutually exclusive stains (CD4 and CD8). b, Overcompensation leaves many cells on the axes and may decrease dual expression when present. c, Correct compensation produces results in which mutually exclusive stains appear to be pure green or orange and few dual-labeling cells are present. FL1-H, x-axis, green; FL2-H, y-axis, orange. Specimen Analysis Specimen Preparation Specimens are often whole blood but may be other tissues. For analysis, the specimen must contain greater than 95% single cells for accurate interpretation of the data. Whole blood is a single cell suspension, but tissues must be disaggregated into single cells before analylaboratorymedicine> april 2001> number 4> volume 32
4 your lab focus a b c Ungated total particles M1 0% 24% Gated lymphocytes 11% 65% FL1-H [F5] Data can be displayed on several plots and electronically gated to analyze pure populations. a, Dot plot of lysed whole blood showing forward scatter vs side scatter with a gate surrounding lymphocytes. b, Overlaid histograms of fluorescein fluorescence showing the difference with total and gated population analysis. The positive cells within the markers are 1% of the total particles, including debris, but 89% of the gated lymphocytes. c, Dot plot of fluorescein vs phycoerythrin fluorescence showing quadrant statistics. FL1-H, x-axis, green, fluorescein; FL2-H, y-axis, orange, phycoerythrin. 210 sis. Often antibodies are used to stain cells and select subpopulations or evaluate protein expression. In most cases, these antibodies are commercial products with robust staining properties. However, the staining patterns should be confirmed in the laboratory on each new lot of antibody using appropriate control materials. Unknown antibodies must be carefully evaluated for specificity and efficacy of staining. This is best done under a microscope initially because nonspecific staining or unusual patterns can be better distinguished with a microscope than with a flow cytometer. Once approximate staining characteristics are determined, the procedure can be fine-tuned for the flow cytometer. Data Processing The initial light signal is first converted by the detectors into an electric current, which is amplified and measured as an analog quantity, such as 5 V of a maximum of 10 V. Because this signal is analog, it must be converted into a digital signal for computer processing. The signal then becomes an event with a specific digital channel value such 512 on a 1,024-channel scale. The digital data can be displayed as 1-parameter histograms [F5b] or as 2-parameter plots [F4, F5a, F5c]. The information on a 1-parameter histogram is described either as the percentage of cells within a set of markers or as the mean fluorescence intensity of a population. For logarithmically amplified signals, the geometric mean fluorescence is used, since it is less influenced by the extreme ends of the distribution. A 2- parameter plot is usually divided into 4 quadrants, each containing a percentage of the total population [F5c]. This division is used to distinguish between fluorescent and nonfluorescent cells. It also defines expression and nonexpression of a cell molecule marked by a fluorescent antibody or other fluorochrome. Because the specimen that is analyzed may be complex with several populations plus debris, such as lysed whole blood, analysis of the total specimen would give variable and confusing results. Therefore, it is often necessary to analyze a single population, such as lymphocytes. Electronic gating allows separation of the total population into parts, which can be individually analyzed [F5a]. Quality Control There are several kinds of quality controls. First the flow cytometer itself must be evaluated for proper function. This is usually accomplished with standardized fluorescent beads. These give very precise, reproducible patterns, which quickly assess instrument function. In addition, the beads are fairly stable and can be used daily with little change in values. The bead values for instrument settings and signal intensity are used to generate Levy-Jennings plots and determine instrument errors. A second quality control material is used to set up the appropriate instrument settings for the type of staining used. These can be beads or antibody-stained cells. One of the primary functions of this second control is to establish the compensation settings for the instrument; therefore, either beads or cells should mimic the staining that will be used on actual specimens. The third level of quality control is a control substance that mimics actual specimens. These controls are available commercially and usually consist of stabilized blood, sometimes with added tissue-culture cells that mimic a specific cancer cell. These controls can be used to assess the sensitivity of the instrument. They will have published ranges for many of the antibodies used in the laboratory. Because they usually contain normal, stabilized blood cells, some proteins characteristic of specific cancer cells will not be present. For this reason, some tissue-culture cells may be included because they express specific proteins that can be used to evaluate other antibodies.
5 Conclusion Flow cytometry is used in the clinical laboratory for a variety of analyses including diagnostic immunophenotyping of lymphomas, leukemias, and congenital immunodeficiencies; detection of minimal residual disease in leukemias; monitoring CD4+ T-lymphocyte counts in HIV patients; DNA content analysis for prognosis of malignancies; screening of hematologic disorders such as paroxysmal nocturnal hemoglobinuria and idiopathic thrombocytopenic purpura; evaluation of peripheral stem cell products for transplantation; monitoring monoclonal antibody therapy; and determination of the extent of maternal/fetal bleeding by detection of hemoglobin F. Flow cytometry remains a laborintensive area of laboratory medicine requiring extensive training. Learning the fundamental principles of flow cytometry enhances both patient care and work satisfaction. In addition, these principles apply to other areas using cytometry techniques, such as hematology. For more details and theoretical discussion, the reader is referred to several books 1-4 and Internet sites. 5,6 1. Givan AL. Flow Cytometry: First Principles. New York, NY: John Wiley & Sons; Bauer KD, Duque RE, Shankey TV, eds. Clinical Flow Cytometry: Principles and Application. Baltimore, MD: Williams and Wilkins, Owens MA, Loken MR. Flow Cytometry Principles for Clinical Laboratory Practice, Quality Assurance for Quantitative Immunophenotyping. New York, NY: Wiley-Liss; Melamed MR, Lindmo T, Mendelsohn ML. Flow Cytometry and Sorting. 2nd ed. New York, NY: Wiley-Liss; Purdue University Cytometry Laboratories. Accessed January 8, Hematopathologic phenotypes made mockingly simple. Available at: ~uthman/cdphobia/cdphobia.html. Accessed January 8, PostScript Picture 211
Anatomy of a flow cytometer
Anatomy of a flow cytometer Fluidics Optics Electronics Cells in suspension flow in single-file through an illuminated volume where they scatter light and emit fluorescence that is collected, filtered,
More informationIntroduction to Flow Cytometry:
Introduction to Flow Cytometry: A Learning Guide Manual Part Number: 11-11032-01 April, 2000 BD Biosciences 2350 Qume Drive San Jose, CA 95131-1807 1-800-448-2347 Introduction to Flow Cytometry: A Learning
More informationIntroduction to Flow Cytometry: A Learning Guide
11-11032-03 rev. A December 2002 Introduction to Flow Cytometry: A Learning Guide Introduction to Flow Cytometry: A Learning Guide Copyright 2002 Becton, Dickinson and Company. All rights reserved. No
More informationDefinitions. What functions does a flow cytometer be able to do? The build of the Flow cytometry and sorting. Flow cytometry
The build of the Flow cytometry and sorting Flow cytometry Seminar Mónika Tóth 11-12-13.04.2011 Definitions Flow cytometry Process or measurement method can measure discrete properties physical, chemical,
More information11/19/2013. Janine Zankl FACS Core Facility 13. November Cellular Parameters. Cellular Parameters. Monocytes. Granulocytes.
DEPARTEMENT BIOZENTRUM Janine Zankl FACS Core Facility 13. November 2013 Cellular Parameters Granulocytes Monocytes Basophils Neutrophils Lymphocytes Eosinophils Cellular Parameters 1 What Is Flow Cytometry?
More informationPrinciples of flow cytometry: overview of flow cytometry and its uses for cell analysis and sorting. Shoreline Community College BIOL 288
Principles of flow cytometry: overview of flow cytometry and its uses for cell analysis and sorting Shoreline Community College BIOL 288 Flow Cytometry What is Flow Cytometry? Measurement of cells or particles
More informationEach question may have MULTIPLE correct answers. Select all that are correct.
Knowledge Assessment Flow Cytometry Workshop, Part 1 April 20, 2015 Each question may have MULTIPLE correct answers. Select all that are correct. 1. Tandem dyes are a. highly stable fluorophores after
More informationANAT 3231 Cell Biology Lab12 Stem Cell Analysis
ANAT 3231 Cell Biology Lab12 Stem Cell Analysis 2 June 2010 Dr Antonio Lee Neuromuscular & Regenera9ve Medicine Unit School of Medical Sciences, UNSW Introduction to Flow Cytometry Contributed by Vittoria
More informationFlow Cytometry - The Essentials
Flow Cytometry - The Essentials Pocket Guide to Flow Cytometry: 1. Know your Cytometer 2. Understanding Fluorescence and Fluorophores 3. Gating Process 4. Controls 5. Optimization 6. Panel Building 7.
More informationBasic Principles in Flow Cytometry. Flow Cytometry
Basic Principles in Flow Cytometry Flow Cytometry» Flow Cytometry is the technological process that allows for the individual measurements of cell fluorescence and light scattering. This process is performed
More informationINTRODUCTION TO FLOW CYTOMETRY
DEPARTEMENT BIOZENTRUM INTRODUCTION TO FLOW CYTOMETRY F ACS C ore F acility Janine Zankl FACS Core Facility 3. Dezember 2015, 4pm Cellular Parameters Granulocytes Monocytes Basophils Lymphocytes Neutrophils
More informationseminar
seminar 02.26.-28.2013. Flow cytometry Process or measurement method can measure discrete properties physical, chemical, biochemical, biological parameters of separate particles, e.g. biological cells
More informationBASICS OF FLOW CYTOMETRY
BASICS OF FLOW CYTOMETRY AUTHOR: Ana Isabel Vieira APPROVAL: Henrique Veiga Fernandes Ana Sílvia Gonçalves SOP.UCF.002 03-09-2015 Pag. 1/9 Overview Flow: Fluid Cyto: Cell Metry: Measurement Flow cytometry
More informationALL INFECTOLOGY. and now. Gucker, 1947 Wallace Coulter Coulter Orifice, Kamentsky, 1965 Mack Fulwyler sorter BLOOD:
ALL BLOOD: RBC, PLT, WBC ( /-/ ) (+ LDH, carbamide, ESR, periphery blasts, liver function deviation, renal function dev.) Classification: Morphology + genetics+ immunology! B-cell: T-cell: Dedifferentiated:
More informationALL INFECTOLOGY COMMERCIALLY AVAILABLE DEVICES. and now BLOOD:
ALL BLOOD: RBC, PLT, WBC ( /-/ ) (+ LDH, carbamide, ESR, periphery blasts, liver function deviation, renal function dev.) Classification: Morphology + genetics+ immunology! B-cell: T-cell: Dedifferentiated:
More informationImmunophenotyping of Peripheral Blood and Bone Marrow Cells by Flow Cytometry *Akanni EO and # Palini A.
Immunophenotyping of Peripheral Blood and Bone Marrow Cells by Flow Cytometry *Akanni EO and # Palini A. * Department of Haematology & Blood Transfusion,College of Health Science, Ladoke Akintola University
More informationSelected Topics in Electrical Engineering: Flow Cytometry Data Analysis
Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis Bilge Karaçalı, PhD Department of Electrical and Electronics Engineering Izmir Institute of Technology Outline Experimental design
More informationJ. Philip McCoy, Jr., Ph.D., H.C.L.D. J. Philip McCoy, Jr., PhD
J. Philip McCoy, Jr., Ph.D., H.C.L.D. Email: mccoyjp@mail.nih.gov 1 Introduction 2 How It All Works A very brief description of how a flow cytometer works. Please read it it will help you design your experiments.
More informationdetermine optimum instrument settings for their own instruments and establish their own daily values.
PC7 (770/488) SETUP KIT 6607121 PN 4299504-C FLOW CYTOMETER ALIGNMENT VERIFICATION FLUOROSPHERES FLOW CYTOMETER DETECTOR STANDARDIZATION FLUOROSPHERES INTENDED USE For Research Use Only. Not for use in
More informationFlow Cytometry. Marta Argenti, PhD student. Department of Biomedical Sciences Padua
Flow Cytometry Marta Argenti, PhD student Department of Biomedical Sciences Padua 14.12.12 Flow ~ cells in motion Cyto ~ cell Metry ~ measure Physical properties: Flow Cytometry is the measurement of cells
More informationDesigning and Implementing a High-Level Multicolor Flow Cytometry Assay. Brent Wood MD PhD Department of Laboratory Medicine University of Washington
Designing and Implementing a High-Level Multicolor Flow Cytometry Assay Brent Wood MD PhD Department of Laboratory Medicine University of Washington Define Purpose of Assay Most important question What
More informationFlow Cytometry. Flow Cytometry Basics Guide
Flow Cytometry Flow Cytometry Basics Guide Table of Contents Chapter 1 Chapter 2 Chapter 3 Chapter 4 Chapter 5 Principles of the Flow Cytometer Fluidics System.... 3 Optics and Detection.... 4 Signal and
More informationThe Basics of Flow Cytometry
The Basics of Flow Cytometry F ACS C ore F acility Janine Bögli, Biozentrum, 29. January 2018 The functions of the FACS Core Facility Centralization of equipment and expertise Train users Sorter operation
More informationFlowcytometry Dirk Pacholsky
Flowcytometry Dirk Pacholsky Flowcytometry: Overview 1 2 Overview Flow Cyto Metry Fluid Cell Measurement measuring cell properties of cells in suspension Most Flow Cytometer have Spatially separated Lasers
More informationSpherotech, Inc. 1. SPHERO TM Technical Note STN-8 Rev C
SPHERO TM Technical Note STN- Rev C. 0070 CALIBRATION AND PERFORMANCE TRACKING OF FLOW CYTOMETERS USING SPHERO TM CALIBRATION PARTICLES Introduction The SPHERO TM Calibration Particles are versatile, stable,
More informationHigh-throughput automation with the Attune NxT Autosampler: consistent results across all wells and across plates
APPLICATION NOTE Attune NxT Flow Cytometer with Autosampler High-throughput automation with the Attune NxT Autosampler: consistent results across all wells and across plates Introduction The emerging field
More informationA previous ICCS Module entitled Instrument optimization - Adjusting PMT voltages and compensation 1 should be read as a prerequisite to this module.
Sponsored and reviewed by ICCS Quality and Standards Committee Title: Compensation Tips for Beckman Coulter 10-Color Navios Platform Written by: Salima Janmohamed-Anastasakis Ph.D., Applications Scientist,
More informationFlow Cytometry in the Diagnosis of Hematopoietic Neoplasia. Brent Wood MD, PhD Professor, Laboratory Medicine University of Washington, Seattle
Flow Cytometry in the Diagnosis of Hematopoietic Neoplasia Brent Wood MD, PhD Professor, Laboratory Medicine University of Washington, Seattle 1 Flow Cytometer 2 The Power of Flow Cytometry Single cell
More informationIntroduction to Flow Cytometry. -- BD FACSCanto II TM. Daisy Kuo Application Specialist BDBiosciences
Introduction to Flow Cytometry -- BD FACSCanto II TM Daisy Kuo Application Specialist E-mail: daisy_kuo@bd.com BDBiosciences Outline Basic Concept of Flow Cytometry FACSCanto II System Introduction Application
More informationNavios EX FLOW CYTOMETER POWERFUL, DEPENDABLE CLINICAL FLOW CYTOMETRY
Navios EX FLOW CYTOMETER POWERFUL, DEPENDABLE CLINICAL FLOW CYTOMETRY BECAUSE EVERY EVENT MATTERS The Navios EX flow cytometer offers a solution for advanced cytometry applications optimized for the clinical
More informationFlow Cytometry For New PhDs
Flow Cytometry For New PhDs 2012 Simon Monard SCRM and CIR smonard@staffmail.ed.ac.uk Derek Davies CRUK derek.davies@cancer.org.uk Monday 20th and Tues 21st Feb 2012 Program Day 1: Day 2: Basics of flow
More informationIncorporating New, Bright Fluorochromes into Multicolor Panel Design
Incorporating New, Bright Fluorochromes into Multicolor Panel Design Maria C. Jaimes, MD Senior Staff Scientist BD Biosciences 23-14684-00 Overview Multicolor flow: successful application prerequisites
More informationIntroduction to. BD FACSAria TM Cell Sorter. Flow = Fluid Cyto = Cell Metry = Measurement
What is Flow Cytometry? Introduction to BD FACSAria TM Cell Sorter Flow = Fluid Cyto = Cell Metry = Measurement BD Biosciences Application Specialist 產品應用專員 Daisy Kuo 郭正佼 A variety of measurements are
More informationNovoCyte Flow Cytometer
NovoCyte Flow Cytometer The Flow Cytometer for Everyone 2 Experience the NovoCyte Advantage Focus on advancing your research. Let the flow cytometer do the rest. NovoCyte Flow Cytometer High Performance
More informationMinimum Information about a Flow Cytometry Experiment (MIFlowCyt) Annotation
Minimum Information about a Flow Cytometry Experiment (MIFlowCyt) Annotation 1. Experiment Overview 1.1 Purpose The purpose of these sets of experiments is to develop a methodology of identifying and quantifying
More informationINTRODUCTION TO FLOW CYTOMETRY
INTRODUCTION TO FLOW CYTOMETRY VERSION 5.1 COPYRIGHT 2004 GROFF M. SCHROEDER DOUGLAS E. SWARTZENDRUBER 1 PREFACE Flow cytometry is a growing field of instrumental analysis with a range of applications
More informationApplication Note. Assay Portability on the BD FACSVerse System. Summary. Maria Jaimes, Yibing Wang, Catherine McIntyre, and Dev Mittar
September Assay Portability on the BD FACSVerse System Maria Jaimes, Yibing Wang, Catherine McIntyre, and Dev Mittar Contents Summary Introduction 3 Objective 4 Methods 6 Results Discussion Conclusions
More informationFlow Cytometry. Principles and Applications. Edited by. Marion G. Macey, PhD
Flow Cytometry Flow Cytometry Principles and Applications Edited by Marion G. Macey, PhD Department of Haematology The Royal London Hospital London, United Kingdom 2007 Humana Press Inc. 999 Riverview
More informationAttune NxT Acoustic Focusing Cytometer The next generation in acoustic cytometry
Attune NxT Acoustic Focusing Cytometer The next generation in acoustic cytometry Maybelline Giam Field Application Scientist The world leader in serving science Attune NxT Flow Cytometer Attune NxT Acoustic
More informationTheory and Applications for Cell Sorter
Theory and Applications for Cell Sorter Tony Liu 劉聖德 Product Manager BD Biosciences Taiwan Sep. 07, 2016 1 Outline Basic Concept of Flow Cytometry FACSAria II System Introduction Cell Sorting Theory Application
More informationNavios EX FLOW CYTOMETER POWERFUL, DEPENDABLE FLOW CYTOMETRY
Navios EX FLOW CYTOMETER POWERFUL, DEPENDABLE FLOW CYTOMETRY BECAUSE EVERY EVENT MATTERS The Navios EX flow cytometer offers a solution for advanced cytometry applications with optimized workflows for
More informationFlow Cytometry. Flow Cytometry Basics Guide
Flow Cytometry Flow Cytometry Basics Guide Table of Contents Chapter 1 Chapter 2 Chapter 3 Chapter 4 Principles of the Flow Cytometer Fluidics System.... 3 Optics and Detection.... 4 Signal and Pulse
More informationAttune TM Acoustic Focusing Cytometer Training. Manik Punj Attune Training
Attune TM Acoustic Focusing Cytometer Training Manik Punj Attune Training Attune Training Agenda Section 1 An Introduction to Flow Cytometry Section 2 An Introduction to Acoustic Focusing Hydrodynamic
More informationBest practices in panel design to optimize the isolation of cells of interest
Sort Best practices in panel design to optimize the isolation of cells of interest For Research Use Only. Not for use in diagnostic or therapeutic procedures. Alexa Fluor is a registered trademark of Life
More informationPowERFUL. INTUITIvE. customizable. ThE NEw STAR of. BENchToP FLow cytometry
PowERFUL. INTUITIvE. customizable. ThE NEw STAR of BENchToP FLow cytometry ABoUT AcEA Established in 2002, ACEA Biosciences, Inc. develops cutting-edge cell analysis platforms for life science research.
More informationDEPArray Technology. Sorting and Recovery of Rare Cells
DEPArray Technology Sorting and Recovery of Rare Cells Delivering pure, single, viable cells The DEPArray system from Silicon Biosystems is the only automated instrument that can identify, quantify, and
More informationCell analysis and bioimaging technology illustrated
Cell analysis and bioimaging technology illustrated The Cell Analysis Center Scientific Bulletin Part 1 Sysmex has been studying and exploring principles of automated haematology analysers, making full
More informationBoundary-breaking acoustic focusing cytometry
Boundary-breaking acoustic focusing cytometry Introducing the Attune NxT Acoustic Focusing Cytometer a high-performance system that s flexible enough for any lab One of the main projects in my laboratory
More informationTitration of Fluorochrome-Conjugated Antibodies for Labeling Cell Surface Markers on Live Cells
Titration of Fluorochrome-Conjugated Antibodies for Labeling Cell Surface Markers on Live Cells Ruud Hulspas 1 UNIT 6.29 1 Cytonome/ST, Boston, Massachusetts ABSTRACT Nonspecific antibody binding is best
More informationApplication Note. Yeast Analysis with a life - dead staining kit (Yeast control - viability)
Application Note Yeast Analysis with a life - dead staining kit (Yeast control - viability) (CyFlow Cube 8 and CyFlow Cube 6 with CyView TM version 1.5 CyFlow Space) Introduction The Partec Yeast control
More informationIdentification of red and white blood cells from whole blood samples using the Agilent 2100 bioanalyzer. Application Note
Identification of red and white blood cells from whole blood samples using the Agilent 2100 bioanalyzer Application Note Sylvie Veriac Valérie Perrone Madeleine Avon Abstract Agilent Equipment: 2100 bioanalyzer
More informationFlow cytometry has become an increasingly important
9-Color and 10-Color Flow Cytometry in the Clinical Laboratory Brent Wood, MD, PhD Context. The development of commercial flow cytometers capable of detecting more than 10 simultaneous fluorescent signals
More informationThe Cell Cycle, DNA Replication, and Mitosis
Chapter 19 The Cell Cycle, DNA Replication, and Mitosis 首頁 OUTLINE Overview of the Cell Cycle DNA Replication DNA Damage and Repair Nuclear and Cell Division Regulation of the Cell Cycle Growth Factors
More informationAutomation In Life Sciences and Diagnostics Applications
Automation In Life Sciences and Diagnostics Applications Pedro Diaz Director of Research Beckman Coulter Life Sciences February 20-22, 2013 Orlando World Marriott Center Orlando, Florida USA Beckman Coulter
More informationApplication Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement
Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Andreas Spittler, MD, Associate Professor for Pathophysiology, Medical University
More informationTECHNICAL BULLETIN. QUANTUM FLUORESCENCE KITS FOR MESF UNITS OF FITC Product Numbers QMF-2 AND QMF-10 Storage Temperature 2-8 C.
QUANTUM FLUORESCENCE KITS FOR MESF UNITS OF FITC Product Numbers QMF-2 AND QMF-10 Storage Temperature 2-8 C. Do Not Freeze TECHNICAL BULLETIN Product Description Quantum Fluorescence Kits for MESF Units
More informationCompensation: Fundamental Principles
Flow Cytometry Seminar Series 2017 : Fundamental Principles Spillover correction in multicolor flow cytometry 28.02.2017 http://www.cytometry.uzh.ch Contents Fluorescence and its detection Absorption and
More informationACTG Laboratory Technologist Committee Revised Version 2.0 ACTG Lab Man CD38 Quantitation 07 May 2004
1. Principle Flow Cytometric quantitation of CD38 expression on CD8+ T lymphocytes The purpose of this assay is to quantitate the surface expression of the CD38 molecule on CD8 positive T lymphocytes.
More informationMicro-fluidic Chip for Flow Cytometry Jeff Wang
Micro-fluidic Chip for Flow Cytometry Jeff Wang September, 2015 ABSTRACT This lab course is intended to give students hands-on experience with microfabrication. The project is to make a micro-fluidic chip
More informationBeyond Colorful Science. A Practical Approach to Setting Up Multicolor Panels
Beyond Colorful Science A Practical Approach to Setting Up Multicolor Panels ISCT Paris April 2014 Layout: Multicolor Panel Design Identify the pitfalls Have a look at issues with spectral overlap How
More informationWhat you need to know before designing a panel
Design What you need to know before designing a panel For Research Use Only. Not for use in diagnostic or therapeutic procedures. Alexa Fluor is a registered trademark of Life Technologies Corporation.
More informationDetecting human circulating endothelial cells using the Attune Acoustic Focusing Cytometer
APPLICATION NOTE Attune Acoustic Focusing Cytometer Detecting human circulating endothelial cells using the Attune Acoustic Focusing Cytometer Circulating endothelial cells (CECs) are mature cells shed
More informationCELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE
CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each
More informationTECHNICAL BULLETIN. QUANTUM FLUORESCENCE KITS FOR MESF UNITS OF FITC Product Numbers QMF-2 AND QMF-10 Storage Temperature 2-8 C.
QUANTUM FLUORESCENCE KITS FOR MESF UNITS OF FITC Product Numbers QMF-2 AND QMF-10 Storage Temperature 2-8 C. Do Not Freeze TECHNICAL BULLETIN Product Description Quantum Fluorescence Kits for MESF Units
More informationFLOW CYTOMETRY. CyAn ADP. Analyzer
FLOW CYTOMETRY CyAn ADP Analyzer Experience the Power of the CyAn ADP and its optimal performance The Power of Detection The Power of Speed The Power of Ease The CyAn ADP Analyzer is the next step in Advanced
More informationBIMM18 Dec 20 th - Flow cytometry in clinical diagnostics
BIMM18 Dec 20 th - Flow cytometry in clinical diagnostics I. B cell leukemia and lymphomas Immunophenotyping as part of the diagnostic work- up of hematologic malignancies offers a rapid and effective
More informationFormalization of the MESF Unit of Fluorescence Intensity
Cytometry Part B (Clinical Cytometry) 57B:1 6 (2004) Report Formalization of the MESF Unit of Fluorescence Intensity Abe Schwartz, 1 Adolfas K. Gaigalas, 2 Lili Wang, 2 Gerald E. Marti, 3 Robert F. Vogt,
More informationLab 2. Isolation of mononuclear cells from peripheral blood and separation into subpopulations.
Lab 2 Isolation of mononuclear cells from peripheral blood and separation into subpopulations. Supervisors: Ulrika Andreasson ulrika.andreasson@immun.lth.se tel: 222 9264 Niclas Olsson niclas.olsson@immun.lth.se
More informationPhagocytosis Assay Kit (IgG PE)
Phagocytosis Assay Kit (IgG PE) Item No. 600540 www.caymanchem.com Customer Service 800.364.9897 Technical Support 888.526.5351 1180 E. Ellsworth Rd Ann Arbor, MI USA TABLE OF CONTENTS GENERAL INFORMATION
More informationReview of techniques in imaging and cytometry. Peter Meeus Onze Lieve Vrouw Ziekenhuis, Aalst 7 may 2009, SCK/CEN Mol
Review of techniques in imaging and cytometry Peter Meeus Onze Lieve Vrouw Ziekenhuis, Aalst 7 may 2009, SCK/CEN Mol Cytometry: The counting and measuring of cells,... Mosby's Medical Dictionary, 8th edition.
More informationNavios Now Because Every Event Matters
Navios Now Because Every Event Matters Powerful, Dependable 10-Color Flow Cytometry Because we understand the power inside every cell, we designed the Navios flow cytometer to capture the information you
More informationCRC Flow Cytometry Core Facility Knut och Alice Wallenbergs Stiftelse
CRC Flow Cytometry Core Facility CRC Flow Cytometry Core Facility Knut och Alice Wallenbergs Stiftelse CRC Flow Cytometry Core facility Instruments: 1. FACSCalibur: analyser, 6 parameters, 4 colors (BD),
More informationReview of techniques in flow cytometry. Peter Meeus Onze-Lieve-Vrouwziekenhuis, Aalst 5 may 2011, SCK/CEN Mol
Review of techniques in flow cytometry Peter Meeus Onze-Lieve-Vrouwziekenhuis, Aalst 5 may 2011, SCK/CEN Mol Cytometry definitions The counting and measuring of cells,... Mosby's Medical Dictionary, 8th
More informationAntigen-antibody reactions with labeled reagents
Antigen-antibody reactions with labeled reagents Department of Immunology Faculty of Medicine University of Belgrade ANTIGEN ANTIBODY REACTIONS WITH LABELED REAGENTS Enzyme immunoassay Radioimmunoassay
More informationCLEARLLAB LS LYMPHOID SCREEN REAGENT
CLEARLLAB LS LYMPHOID SCREEN REAGENT CE MARKED ANTIBODY COMBINATION FOR LEUKEMIA / LYMPHOMA ANALYSIS Because Your Patient is Her Everything BECAUSE YOUR PATIENT IS HER EVERYTHING ClearLLab LS Lymphoid
More informationQdot nanocrystal. wide range of biological investigations, Qdot nanocrystals are powerful complements
Feature nanocrystal conjugates for flow cytometry Take the easy route to multicolor flow cytometry. With applications across a wide range of biological investigations, nanocrystals are powerful complements
More informationPERFECT-COUNT MICROSPHERES
PERFECT-COUNT MICROSPHERES Perfect-Count Microspheres-Product code PCB-100 for 100 tests Introduction In recent years, the determination of absolute cell counts has been shown to be relevant in different
More informationSelecting Fluorochrome Conjugates for Maximum Sensitivity
2004 Wiley-Liss, Inc. Cytometry Part A 62A:169 173 (2004) Technical Primer Selecting Fluorochrome Conjugates for Maximum Sensitivity Holden T. Maecker,* Tom Frey, Laurel E. Nomura, and Joe Trotter BD Biosciences,
More informationBD FACSMelody. Cell Sorter. The simple solution for consistent, quality results
BD FACSMelody Cell Sorter The simple solution for consistent, quality results The right instrument is needed for the best results Simplicity, affordability and quality are key Cell sorting is fast becoming
More informationab Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis
ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis Instructions for Use To determine cell cycle status in tissue culture cell lines by measuring DNA content using a flow cytometer. This
More informationab Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis
ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis Instructions for Use To determine cell cycle status in tissue culture cell lines by measuring DNA content using a flow cytometer. This
More informationRuud Hulspas, Mike Keeney, Ben Hedley and Andrea Illingworth
Sponsored and reviewed by ICCS Quality and Standards Committee Title: Quality of Reagents Monoclonal Antibodies Written by: Date: March 6, 2018 Ruud Hulspas, Mike Keeney, Ben Hedley and Andrea Illingworth
More informationImmunological Techniques in Research and Clinical Medicine. Philip L. Cohen, M.D. Chief of Rheumatology, LKSOM 10 March 2016
Immunological Techniques in Research and Clinical Medicine Philip L. Cohen, M.D. Chief of Rheumatology, LKSOM 10 March 2016 Antibodies Remarkable Tools for Research and Diagnosis You can make an antibody
More informationThe NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee
January, 1999 From: To: Subject: The NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee NIAID DAIDS Flow Cytometry Laboratories Comparison study information for labs wishing to switch
More informationInstitute of Science & Technology in Medicine Control of Substance Hazardous to Health COSHH assessment
Institute of Science & Technology in Medicine Control of Substance Hazardous to Health COSHH assessment Title of Experiment/Procedure: Cell counting and viability using FACS, counting beads and propidium
More information18/08/2011. Principles of Flow Cytometry (Practised in a Clinical Laboratory) Cytometer Components. Noel Williams Immunobiology Division of Immunology
Optical Measurement Principles Principles of Flow Cytometry (Practised in a Clinical Laboratory) Noel Williams Immunobiology Division of Immunology Cytometer Components Reagents Cytometer Setup Cytometer
More informationLive and Dead Cell Assay
ab115347 Live and Dead Cell Assay Instructions for Use Differential fluorescent labeling of live and dead cells This product is for research use only and is not intended for diagnostic use. Last Updated
More informationPrinciples of Multicolor Panel Design BD. BD, the BD Logo and all other trademarks are property of Becton, Dickinson and Company.
1 Principles of Multicolor Panel Design 2 Common Multicolor Applications Intracellular cytokine staining Regulatory T cells (Tregs) Protein phosphorylation (BD Phosflow) Leukemia and lymphoma phenotyping
More informationApplication Information Bulletin: DuraClone IM Tubes Compensation setup for high content DuraClone reagents
Application Information Bulletin: DuraClone IM Tubes Compensation setup for high content DuraClone reagents Compensation setup for high content DuraClone reagents INTRODUCTION High content flow cytometry
More informationBD FACSCanto II. A proven research platform for maximum reliability and the highest quality results
BD FACSCanto II A proven research platform for maximum reliability and the highest quality results A proven platform for maximum reliability and the highest quality results Built on more than 25 years
More informationUsing Fluorescence Spillover to Advantage
Using Fluorescence Spillover to Advantage Boston June 9, 2011 Compensation: Why and When is it Necessary? Clare Rogers, Marketing and Applications Accuri Cytometers, Inc. Ann Arbor/St.Ives 1 June 11 Summary
More informationAnalysis of repetitive DNA in chromosomes by flow cytometry
Nature Methods Analysis of repetitive DNA in chromosomes by flow cytometry Julie Brind Amour & Peter Lansdorp Supplementary Figure 1 Supplementary Figure 2 Supplementary Figure 3 Supplementary Figure 4
More informationTECHNICAL BULLETIN. QUANTUM SIMPLY CELLULAR KIT Product Numbers QSC-20 AND QSC-100 Storage Temperature 2-8 C. Do Not Freeze
QUANTUM SIMPLY CELLULAR KIT Product Numbers QSC-20 AND QSC-100 Storage Temperature 2-8 C. Do Not Freeze TECHNICAL BULLETIN Product Description The Quantum Simply Cellular Kit provides a convenient method
More informationa Beckman Coulter Life Sciences: White Paper
a Beckman Coulter Life Sciences: White Paper DuraClone improves standardization of compensation workflow for high content multicolor applications by flow cytometry Authors: Neha Girish 2, Sudharsan Sathyamurthy
More informationSingle Tube, Six-Color Flow Cytometric Analysis Is a Sensitive and Cost-Effective Technique for Assaying Clonal Plasma Cells
Hematopathology / Flow Cytometric Analysis of Clonal Plasma Cells Single Tube, Six-Color Flow Cytometric Analysis Is a Sensitive and Cost-Effective Technique for Assaying Clonal Plasma Cells Derek K. Marsee,
More informationBIO 315 Lab Exam I. Section #: Name:
Section #: Name: Also provide this information on the computer grid sheet given to you. (Section # in special code box) BIO 315 Lab Exam I 1. In labeling the parts of a standard compound light microscope
More informationBD LSRFortessa. Performance without peer, choice without compromise
BD LSRFortessa Performance without peer, choice without compromise Performance without peer, choice without compromise The BD LSRFortessa cell analyzer offers the ultimate in choice for flow cytometry,
More informationFluorescence Microscopy. Terms and concepts to know: 10/11/2011. Visible spectrum (of light) and energy
Fluorescence Microscopy Louisiana Tech University Ruston, Louisiana Microscopy Workshop Dr. Mark DeCoster Associate Professor Biomedical Engineering 1 Terms and concepts to know: Signal to Noise Excitation
More informationab CFSE Fluorescent Cell Labeling Kit
ab113853 CFSE Fluorescent Cell Labeling Kit Instructions for Use For the durable fluorescent labeling of live cells for fluorescent microscopy and flow cytometry, population growth studies and within sample
More informationFlow Cytometric Devices Draft Guidance for Industry and Food and Drug Administration Staff DRAFT GUIDANCE
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 36 Flow Cytometric Devices Draft Guidance for Industry and Food and Drug Administration Staff DRAFT GUIDANCE This guidance document
More information