IgG Purity/Heterogeneity and SDS-MW Assays with High- Speed Separation Method and High Throughput Tray Setup

Size: px
Start display at page:

Download "IgG Purity/Heterogeneity and SDS-MW Assays with High- Speed Separation Method and High Throughput Tray Setup"

Transcription

1 IgG Purity/Heterogeneity and SDS-MW Assays with High- Speed Separation Method and High Throughput Tray Setup High Throughput Methods to Maximize the Use of the PA 800 Plus system Jose-Luis Gallegos-Perez SCIEX Separations, Framingham, MA Introduction The PA 800 Plus Pharmaceutical System provides a comprehensive, automated, and quantitative solution for the characterization and analysis of therapeutic proteins. The application menu for the PA 800 Plus includes, among others, the IgG Purity/Heterogeneity assay and the SDS-gel molecular weight (SDS-MW) analysis.,2 The first method allows for the resolution of reduced and non-reduced immunoglobulins by size and to subsequently quantify the heterogeneity and impurities that may be present in IgG preparations. The SDS-MW analysis is designed to provide an estimation of the molecular weight of a protein in a sample. There are two types of analysis methods that can be used with these assays: ) The high resolution (HR) method with an inlet to detection window capillary length of 20.0 cm, using the capillary cartridge in a left-to-right configuration. 2) The high-speed (HS) method that uses the capillary cartridge in the right-to-left configuration with a sample introduction inlet to detection window capillary length of 0.0 cm. Both methods are used with a tray configuration that allows running up to 2 samples before having to change reagents. However, other tray setups are possible to optimize and maximize the number of sample runs. This report illustrates the use of the HS method and proposes a High Throughput Tray (HTT) setup to maximize the number of samples per run. We demonstrate that HS and HR analysis methods can be used with this optimized tray array. Advantages of the HS Analysis Method with the High Throughput Tray Setup The HS method provides faster separation time (-20 min vs. 0 min of the HR method); it represents 72 samples/day instead of 8/day. PA 800 Plus system The use of an HTT setup allows for analysis of up to 8 samples (2 using the conventional tray array) before changing chemical reagents. It represents savings in consumables and time. Using the HS method and high throughput tray setup saves time and reagents and helps maximize the use of the PA 800 Plus. Experimental Procedure Reagents: Separations using the IgG Purity Heterogeneity Assay Kit and SDS-MW Kit were performed as described in their respective manuals.,2 Iodoacetamide (IAA) and 2- mercaptoethanol were purchased from Sigma-Aldrich and used without any further purification. Ultrapure water (type ) was obtained from a Milli-Q Direct Ultrapure Water System. Bovine Serum Albumin (BSA) and Myoglobin (Myo) were purchased from Sigma-Aldrich. Capillary: A bare fused-silica capillary was used for separation of monoclonal antibody samples; 0 m i.d. x 0 m o.d. x 0.2 cm total length. Inlet-to-detection window equal to 20 cm for the HR method and 0 cm for the HS method. p

2 Sample Preparation: IgG control standard: A vial containing 9 L of the IgG control standard was combined with 2 L of the 0 kda internal standard and L of 2-mercaptoethanol in a fume hood. The vial was mixed thoroughly and centrifuged at 00g for min. The vial was sealed with Parafilm and heated to 70 C for 0 minutes, after which the vial was cooled for minutes in a room-temperature water bath. Non-reduced IgG sample preparation: A vial containing 9 L of the IgG control was combined with 2 L of the 0 kda internal standard and L of a 20 mm IAA solution. The vial was mixed thoroughly and centrifuged at 00g for minute. The vial was sealed with Parafilm and heated to 70 C for 0 minutes, after which the vial was cooled for minutes in a room-temperature water bath. HTT configuration: HR method: Figure 2 shows the position of the buffer vials in each inlet and outlet buffer tray using the HTT setup. In this configuration the buffer vials in each row are enough for 8 consecutive experiments, therefore it is possible to run up to 8 individual samples. (8 runs per row x rows). Method parameters are shown in Figure. SDS-MW Size Standard: 0 L of molecular weight size standard, 8 L of sample buffer, 2 L of internal standard, and L of 2-mercaptoethanol were combined in a vial. The mixture was heated to 00 C for minutes and placed into a room-temperature water bath to cool down for five minutes before injection. BSA and Myo were diluted with the SDS-MW sample buffer to a final concentration of 0. mg/ml each. For non-reduced protein preparation, a vial containing 9 L of sample solution was combined with 2 L of 0 kda internal standard and L of 20 mm IAA solution. The vial was mixed thoroughly and centrifuged at 00g for minute. The vial was sealed with Parafilm and heated to 70 C for min. For the reduced protein sample, 9 L of the sample was combined with 2 L of the 0 kda internal standard, and L of 2- mercaptoethanol. The mixture was heated to 00 C for min. HS method with conventional tray setups Because the HS method uses the capillary cartridge in the rightto-left configuration, the buffer tray on the right must be used as the inlet tray while the buffer tray on the left must be used as the outlet tray (see Figure ). The polarity for injection and separation should be normal. Using the conventional tray configuration, a maximum of 2 individual samples can be analyzed without reloading the trays with fresh solutions. Outlet Inlet Figure : Image of the inlet and outlet buffer vial tray set-up for the HS method in the PA800 Plus. Gel-R Gel-S NaOH HCl Gel-R Gel-S NaOH HCl Gel-R Gel-S NaOH HCl Gel-R Gel-S NaOH HCl 2 Gel-R Gel-S NaOH HCl Gel-R Gel-S NaOH HCl INLET BUFFER TRAY Method HR HS* Tray on the left Tray on the right 2 OUTLET BUFFER TRAY * Use only rows - (see text) Figure 2: Scheme of the inlet and outlet buffer vial tray setup for the HTT configuration using the HR or the HS method. Ensure the position of the buffer trays in the instrument according to the method being used. (HS method for no more than 0 individual samples, see text). Gel-S = Gel used for separation, Gel-R = Gel used for rinsing, both are of the same gel type. p 2

3 HS method: If less than 0 individual samples are being analyzed, buffer trays must be set up as shown in Figure 2 using rows from to only. IMPORTANT: In this case, the F-column in the SAMPLE TRAY must be kept empty, with no samples to avoid a tray collision during analysis. Follow the method parameters shown in Figure. HS method and HTT setup to maximize the number of sample analyses: To use the HTT configuration with the HS method for analysis of up to 8 samples requires switching reagents in positions C and D in the outlet and inlet buffer trays to avoid tray collision when samples are injected (see Figure ). All columns in the sample tray (from A to F) can be loaded with samples. For this specific case, use the method parameters show in Figure Sample Tray Sample Gel-R NaOH 2Gel-S HCl Sample Table indicates the volume of reagents used in each buffer vial and their number for all 8 runs using the HTT setup. Table. Volume in Buffer Vials. Reagents Vol (ml) Number of vials (8 runs) Water. 2 Gel. Water. 8 Waste (water) NaOH. HCl. Instrument: Separations were performed using a PA 800 Plus Pharmaceutical Analysis System from SCIEX. PDA detector settings were configured according to the application guides. The instrument was controlled using 2 Karat software v0. The sample and capillary storage temperatures were set at ºC and 2ºC respectively. CE Instrument setups for the HR and HS methods with the HTT configuration are shown in Figures, and. Waste Waste HGel-S 2 O Waste Waste Waste Waste Gel-S Waste Waste Waste Waste Gel-S Waste Waste Waste Waste Gel-S Waste Waste 2 Waste Waste Gel-S Waste Waste Waste Waste Gel-S Waste Waste 2 Waste Gel-R NaOH Gel-S HCl Waste Waste Gel-S Waste Outlet Buffer Tray Inlet Buffer Tray Figure : Scheme of the inlet and outlet buffer vial tray setup for the HTT configuration using the HS method to inject up to 8 individual samples. p

4 Figure : High Resolution (HR) Method Parameters with HTT configuration. p

5 Figure : High-Speed (HS) Method with HTT configuration for up to 0 individual samples p

6 Figure : High-Speed Method with HTT configuration for up to 8 individual samples p

7 Results and Discussion HR and HS methods using the HTT setup To demonstrate assay repeatability using the HTT setup, the SDS-MW Size Standard was used. For the HR and HS methods, 8 sample replicates were injected using the conventional (C)-tray array and 8 more samples for the HTT configuration. Figure 7 shows profiles obtained in the experiments with a) HR method and b) HS method. It is evident that both tray setups produce the same analyte profile and that no ghost peaks, artifacts, or any other spurious signal is present in the HTT configuration. This result demonstrates that the HTT setup can be implemented with high-sensitivity or high-resolution methods, therefore the IgGpurity/heterogeneity or the SDS-MW applications could be run under these experimental conditions. Application of the HR and HS Methods with the IgG/Purity Heterogeneity Assay Using the HTT configuration, separation reproducibility of the HR and HS methods was evaluated. Eight replicates of the IgG control standard were injected under non-reducing and reducing conditions. Under non-reducing conditions (Figure 8) the peaks observed correspond to the 0 kda internal standard, the intact antibody (IgG) and the non-glycosylated heavy chain (NG). In addition, impurities that are known to be present in the IgG control standard, such as heavy-heavy (HH) and 2 heavy light chain (2HL) are resolved from the whole antibody using either method. Under reducing conditions (Figure 9), the peaks observed correspond to the 0 kda internal standard (IS), the IgG light chain (LC), IgG non-glycosylated heavy chain (NG) and the IgG heavy chain (HC). As in the previous case, all peaks can be easily identified using either method. Figure 7: Repeatability of HR and HS Methods using Conventional-tray and HTT setups. Stacked data illustrating 8 separations per tray setup using the a) HR method or b) HS method. To facilitate easier analysis, electropherograms were offset by 0.02 AU. Samples were alternately injected using the conventional (C) or the HTT setup. The order in which the samples were injected is indicated by the number. Figure 8: Comparison of HR and HS methods with the HTT setup for IgG Control Standard under Non-reducing Conditions. Peak identification: 0 kda is the Internal Standard, IgG is the intact antibody, NG the non-glycosylated heavy chain, HH and 2HL are the heavy-heavy (HH) and 2 heavy light chain impurities. p 7

8 Table 2. Accuracy and Repeatability of the HS and HR Methods under Non-Reducing Conditions Parameter Migration Time Area Percent Method HR HS HR HS Injection NG IgG NG IgG NG IgG NG IgG Mean SD %CV Table. Accuracy and Repeatability of the HS and HR Methods under Reducing Conditions Figure 9: Comparison of HR and HS methods with the HTT setup for IgG Control Standard under Reducing Conditions. Peak identification: 0 kda is the Internal Standard, LC is the Light Chain, NG the Non-glycosylated heavy chain and HC the Heavy Chain. Migration times and area percent were determined for the main identified peaks (NG, intact IgG, LC, and HC) (Tables 2 and ). These results indicate that the evaluated parameters show high reproducibility for 8 separations using either the HR or the HS method. Reproducibility for migration time was always less 0. %CV and lower than.7 %CV for the peak area percent. As expected, resolution was slightly lower in the HS mode due to the somewhat shorter capillary length; however adjacent peak pairs with close migration times, such as NG/IgG or NG/HC, can be resolved and unequivocally identified and quantified. This demonstrates that the HS method can be used with complete confidence under standardized and validated conditions. Molecular Weight Estimation of Myoglobin (Myo) and Bovine Serum Albumin (BSA) The SDS-MW analysis kit is designed for the separation and estimation of molecular weight (MW) of a protein in a range between 0 and 22 kda. The use of the HS method, for this experiment, was compared to the conventional HR method. The HTT setup was used for all experiments. The molecular weight of an unknown protein can be estimated by making a comparison against a standard curve created using proteins of known sizes. For this reason, a protein sizing standard mixture was injected in parallel to the Myo and BSA samples. The sizing standard reagent contained six proteins: 0, 20,, 0, 00 and 22 kda. Sample mixtures containing the internal standard (IS), Myo and BSA were treated under reducing or non-reducing conditions and three replicates of each condition were analyzed using the PA800 Plus instrument (Figure 0). In all cases, analyte peaks were well resolved using either the high speed or the high resolution method. Molecular weights for Myo and BSA were estimated using the qualitative analysis feature in the 2 Karat p 8

9 Control and Analysis Software. Results listed in Table indicate high reproducibility (%CV < 0.2, n = ) for the MW estimation of Myo or BSA, under non-reducing and reducing conditions. Similarly, the difference of the MW values determined by either method (HS or HR) was minimal with an absolute value difference lower than 0.2 kda. Therefore, both methods can be used under optimized experimental conditions producing comparable results. Table. Determination of Molecular Weight for Myoglobin and Bovine Albumin Serum (in kda) Non-reduced Reduced HR method HS method HR method HS-Method Injection Myo BSA Myo BSA Myo BSA Myo BSA Mean SD %CV Conclusions The IgG Purity/Heterogeneity and SDS-MW assays can be performed using a high speed (HS) separation method with an inlet to detection window of 0.0 cm, using the capillary cartridge in the right to left configuration. In comparison to the more highly resolving conventional High Resolution (HR) method, results are achieved faster saving assay time. The use of a high throughput tray (HTT) setup in order to maximize the number of samples per run sequence does not cause deterioration in the separation. The combination of the HS method with the HTT setup is an alternative when in need of higher throughput sample analysis. References Figure 0: Determination of Protein Molecular Weight using HR and HS methods. Peak identification: 0 kda is the Internal Standard, Myo is myoglobin and BSA is Bovine Serum Albumin. Three injections under reduced and non-reduced conditions were analyzed. A size standard mixture was run before to inject the samples with the HS or HR-methods.. PA 800 Plus Pharmaceutical Analysis System-IgG Purity/Heterogeneity Assay Application Guide. SCIEX part number A97AD, January PA 800 Plus Pharmaceutical Analysis System-SDS-MW- Analysis Assay Application Guide. SCIEX part number A970AD, January AB SCIEX. For Research Use Only. Not for use in diagnostic procedures. AB SCIEX is doing business as SCIEX. The trademarks mentioned herein are the property of AB SCIEX Pte. Ltd. or their respective owners. AB SCIEX is being used under license. Document number: RUO-MKT-02-2-A p 9

Simplifying CE-SDS Data Processing

Simplifying CE-SDS Data Processing Simplifying CE-SDS Data Processing Approach for Mitigating Product Peak Migration Time Drift Samuel Shepherd 1, KieranKumar Mistry 1, Whitney Lane Smith 2, Nicholas Bond 1, Vivian Lindo 1, Sam Fox 3, Jim

More information

IgG Purity and Heterogeneity Assay Kit For the PA 800 Plus Pharmaceutical Analysis System. Application Guide

IgG Purity and Heterogeneity Assay Kit For the PA 800 Plus Pharmaceutical Analysis System. Application Guide Kit For the PA 800 Plus Pharmaceutical Analysis System RUO-IDV-05-6935-A April 2018 This document is provided to customers who have purchased SCIEX equipment to use in the operation of such SCIEX equipment.

More information

Capillary Electrophoresis of Proteins

Capillary Electrophoresis of Proteins Capillary Electrophoresis of Proteins SDS Capillary Gel Electrophoresis SDS-CGE Outline CE-SDS Gel Analysis Description of Technique Method Development Tips PA800 plus kits SDS-MW IgG Purity & Heterogeneity

More information

Application Note. Biopharm. Author. Abstract. Christian Wenz Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Biopharm. Author. Abstract. Christian Wenz Agilent Technologies, Inc. Waldbronn, Germany Performance of commercially available gels for protein characterization by capillary gel electrophoresis with UV detection on the Agilent 7100 CE System Application Note Biopharm Author Christian Wenz

More information

Unified Workflow for Monoclonal Antibody Charge Heterogeneity, Purity, and Molecular Weight Analysis

Unified Workflow for Monoclonal Antibody Charge Heterogeneity, Purity, and Molecular Weight Analysis Procedures Unified Workflow for Monoclonal Antibody Charge Heterogeneity, Purity, and Molecular Weight Analysis Separation and Online Detection of Intact mab Variants and Impurities using CESI-MS Bryan

More information

Analysis of Monoclonal Antibody Charge Variants by Capillary Zone Electrophoresis

Analysis of Monoclonal Antibody Charge Variants by Capillary Zone Electrophoresis Analysis of Monoclonal Antibody Charge Variants by Capillary Zone Electrophoresis Marcia R Santos Global Tactical Marketing, Beckman Coulter Life Sciences, Brea, CA USA INTRODUCTION Charge heterogeneity

More information

Impurity Detection with a New Light Emitting Diode Induced Fluorescence Detector Coupled to the Agilent 7100 Capillary Electrophoresis System

Impurity Detection with a New Light Emitting Diode Induced Fluorescence Detector Coupled to the Agilent 7100 Capillary Electrophoresis System Impurity Detection with a New Light Emitting Diode Induced Fluorescence Detector Coupled to the Agilent 7100 Capillary Electrophoresis System Application Note Biologics & Biosimilars, Proteomics & Protein

More information

Assay of IgG Purity and Heterogeneity using High-Resolution Sodium Dodecyl Sulfate Capillary Gel Electrophoresis

Assay of IgG Purity and Heterogeneity using High-Resolution Sodium Dodecyl Sulfate Capillary Gel Electrophoresis Assay of IgG Purity and Heterogeneity using High-Resolution Sodium Dodecyl Sulfate Capillary Gel Electrophoresis Lucy Y. Liu, Chitra Ratnayake, Jeff Chapman, Narasaiah Dontha, Sae Choo, and M. P. Reddy

More information

Optimization of cief to Resolve Challenging MAb Samples of Clinical Immunodiagnostic Relevance

Optimization of cief to Resolve Challenging MAb Samples of Clinical Immunodiagnostic Relevance Optimization of cief to Resolve Challenging MAb Samples of Clinical Immunodiagnostic Relevance Sushma Rampal, Sr. Application Scientist, Beckman Coulter Life Sciences, Brea, CA, USA Marcia Santos, Staff

More information

Repeatability of C100ht Biologics Analyzer, for High Throughput Glycan Screening

Repeatability of C100ht Biologics Analyzer, for High Throughput Glycan Screening Repeatability of C100ht Biologics Analyzer, for High Throughput Glycan Screening Marcia Santos, Tingting Li, Mervin Gutierrez, Anna Lou, and Clarence Lew SCIEX Separations, Brea, CA Introduction The ability

More information

Separation of Recombinant Human Erythropoietin (rhepo) using the European Pharmacopoeia Method on the PA 800 plus Pharmaceutical Analysis System

Separation of Recombinant Human Erythropoietin (rhepo) using the European Pharmacopoeia Method on the PA 800 plus Pharmaceutical Analysis System Separation of Recombinant Human Erythropoietin (rhepo) using the European Pharmacopoeia Method on the PA 800 plus Pharmaceutical Analysis System Marcia R Santos Global Tactical Marketing, Beckman Coulter

More information

deltadot ANTIBODY ANALYSIS Application Note: Antibodies using LABEL FREE INTRINSIC IMAGING LONDON United Kingdom

deltadot ANTIBODY ANALYSIS Application Note: Antibodies using LABEL FREE INTRINSIC IMAGING LONDON United Kingdom ANTIBODY ANALYSIS using LABEL FREE INTRINSIC IMAGING Application Note: Antibodies Ana l y s i s o f Im m u n o g l o b u l i n G u s i n g deltadot s PEREGRINE I 512-p i x e l La b e l Fr e e Ca p i l

More information

High-Throughput Glycan Screening to Remove Biopharmaceutical Development Bottlenecks

High-Throughput Glycan Screening to Remove Biopharmaceutical Development Bottlenecks High-Throughput Glycan Screening to Remove Biopharmaceutical Development Bottlenecks Mark Lies, Marcia Santos, and Tingting Li SCIEX Separations, Brea, CA Introduction Development of biopharmaceuticals

More information

LabChip GXII: Antibody Analysis

LabChip GXII: Antibody Analysis WHITE PAPER LabChip GXII: Antibody Analysis Antibody Analysis using microfluidic technology in high throughput Quality by Design Experiments Abstract Current initiatives in Process Analytical Technology

More information

SCIEX Anion Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide

SCIEX Anion Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems A49108AE May 2015 AB Sciex Pte. Ltd and its affiliates disclaims all warranties with respect to this document, expressed or implied, including

More information

Rediscover the Strengths of Your PA 800 Plus - A Multi-Attribute Protein Analyzer

Rediscover the Strengths of Your PA 800 Plus - A Multi-Attribute Protein Analyzer Rediscover the Strengths of Your PA 800 Plus - A Multi-Attribute Protein Analyzer Marcia R Santos SCIEX Separations, Brea, CA Introduction Capillary Electrophoresis is a well-accepted analytical separation

More information

The Importance of Temperature Control in Capillary Electrophoresis of Carbohydrates

The Importance of Temperature Control in Capillary Electrophoresis of Carbohydrates The Importance of Temperature Control in Capillary Electrophoresis of Carbohydrates Introducing Temperature Gradient Capillary Electrophoresis Andras Guttman, 1 Marton Szigeti 2 and Jeff Chapman 1 1 SCIEX

More information

Automated and quantitative analysis of biologics

Automated and quantitative analysis of biologics PA 800 PLUS PHARMACEUTICAL ANALYSIS SYSTEM Automated and quantitative analysis of biologics PA 800 PLUS PHARMACEUTIC AL ANALYSIS SYSTEM Designed for the needs of the biopharmaceutical industry Therapeutic

More information

Analysis of Intact Monoclonal Antibodies Using an M3 MicroLC with the TripleTOF 6600

Analysis of Intact Monoclonal Antibodies Using an M3 MicroLC with the TripleTOF 6600 Procedures Analysis of Intact Monoclonal Antibodies Using an M3 MicroLC with the TripleTOF 66 Robust and Sensitive Workflow for Qualitative and Quantitative Analysis of Biotherapeutic IgGs Khatereh Motamedchaboki,

More information

More Samples. Better Decisions Faster.

More Samples. Better Decisions Faster. More Samples. Better Decisions Faster. High Throughput Glycan Screening and Identification C100HT Biologics Analyzer Screening large sample cohorts is challenging in many ways. The neutral charge of glycans

More information

Automated and Quantitative Analysis of Biologics

Automated and Quantitative Analysis of Biologics PA 800 PLUS PHARMACEUTICAL ANALYSIS SYSTEM Automated and Quantitative Analysis of Biologics PA 800 PLUS PHARMACEUTICAL ANALYSIS SYSTEM For Research Use Only. Not for use in diagnostic procedures. Flexibility

More information

HT Pico Protein Express LabChip Kit, Frequently Asked Questions

HT Pico Protein Express LabChip Kit, Frequently Asked Questions Protein Detection 1. How does the GXII determine protein concentration? Protein samples contain a single lower marker. Alignment to a single marker does not provide enough constraints to align large proteins,

More information

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider Combining small-scale purification and analysis of monoclonal antibodies on one instrument Protein purification with high-volume injection using the Agilent 126 Infinity Bio-inert Quaternary LC System

More information

Comparison of Different Brands of Carrier Ampholytes for Monoclonal Antibody Charge Heterogeneity Analysis by Capillary Isoelectric Focusing

Comparison of Different Brands of Carrier Ampholytes for Monoclonal Antibody Charge Heterogeneity Analysis by Capillary Isoelectric Focusing Comparison of Different Brands of Carrier Ampholytes for Monoclonal Antibody Charge Heterogeneity Analysis by Capillary Isoelectric Focusing Application Note Biotherapeutics & Biosimilars Author Christian

More information

Knock! Knock! Who s there? Maurice Who? SungAe Suhr Park, Mee Ko, Eleanor Le, Janice Chen Amgen Inc.

Knock! Knock! Who s there? Maurice Who? SungAe Suhr Park, Mee Ko, Eleanor Le, Janice Chen Amgen Inc. Knock! Knock! Who s there? Maurice Who? SungAe Suhr Park, Mee Ko, Eleanor Le, Janice Chen Amgen Inc. Abstract SungAe Suhr Park, Mee Ko, Eleanor Le and Janice Chen Amgen Inc. Applications of Capillary Electrophoresis

More information

Answers, Simple and Streamlined. Solutions for Biotherapeutic Intact Mass Analysis

Answers, Simple and Streamlined. Solutions for Biotherapeutic Intact Mass Analysis Answers, Simple and Streamlined Solutions for Biotherapeutic Intact Mass Analysis Expert Functionality for Mass Spec Non-Experts SCIEX OS on the X500B QTOF system dramatically improves the user experience,

More information

CE Oligonucleotide Analysis Kit Instruction Manual

CE Oligonucleotide Analysis Kit Instruction Manual CE Oligonucleotide Analysis Kit Instruction Manual Catalog Number 148-4140 For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800-4BIORAD (1-800-424-6723) Table of Contents Section

More information

Streamlined and Complete LC-MS Workflow for MAM

Streamlined and Complete LC-MS Workflow for MAM Streamlined and Complete LC-MS Workflow for MAM Multiple Attribute Methodology (MAM) for Biotherapeutic Attribute Monitoring and Purity Testing BV MAM WORKFLOW Define Track Quantify Purity New Peaks Report

More information

E-Coli Cell Lysate By High Performance Capillary Electrophoresis

E-Coli Cell Lysate By High Performance Capillary Electrophoresis E-Coli Cell Lysate By High Performance Capillary Electrophoresis Analysis of E-Coli Lysate Label free analysis using the HPCE-512 ABSTRACT Analytical instruments with the capability of analyzing complex

More information

LIFluor EnhanCE. for use with dsdna 1000 and dsdna 20,000 Kits

LIFluor EnhanCE. for use with dsdna 1000 and dsdna 20,000 Kits LIFluor EnhanCE for use with dsdna 1000 and dsdna 20,000 Kits For the Separation of Double Stranded DNA by P/ACE Fluorescent Capillary Electrophoresis 725824 AB September 2006 Beckman Coulter, Inc. 4300

More information

Capillary Electrophoresis in Quality Control PART II: CE-SDS: Method Development and Robustness

Capillary Electrophoresis in Quality Control PART II: CE-SDS: Method Development and Robustness Capillary Electrophoresis in Quality Control PART II: CE-SDS: Method Development and Robustness Chantal Felten *1 and Oscar Salas Solano 2 1 Alpine Analytical Acadamy, Whistler, British Columbia 2 Seattle

More information

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC The Agilent 126 Infinity Bio-Inert Quaternary LC System and the AdvanceBio SEC 3Å, 2.7 µm Column Application Note Biologics

More information

Application Note # ET-20 BioPharma Compass: A fully Automated Solution for Characterization and QC of Intact and Digested Proteins

Application Note # ET-20 BioPharma Compass: A fully Automated Solution for Characterization and QC of Intact and Digested Proteins Application Note # ET-20 BioPharma Compass: A fully Automated Solution for Characterization and QC of Intact and Digested Proteins BioPharma Compass TM is a fully automated solution for the rapid characterization

More information

PA 800 plus Catalog. Systems/Software

PA 800 plus Catalog. Systems/Software PA 800 plus Catalog Built on 20 years of innovation and leadership in the field of capillary electrophoresis, Beckman Coulter s PA 800 plus provides solutions for protein purity analysis, charge heterogeneity

More information

TECHNICAL INFORMATION

TECHNICAL INFORMATION T-1860A TECHNICAL INFORMATION Capillary Electrophoresis PRECISION IN CAPILLARY ELECTROPHORESIS Harry Whatley and Jeff Chapman Beckman Coulter, Inc. Introduction Instrumentation for routine chemical analysis

More information

Universal Solution for Monoclonal Antibody Quantification in Biological Fluids Using Trap-Elute MicroLC-MS Method

Universal Solution for Monoclonal Antibody Quantification in Biological Fluids Using Trap-Elute MicroLC-MS Method Universal Solution for Monoclonal Antibody Quantification in Biological Fluids Using Trap-Elute MicroLC-MS Method Featuring the SCIEX QTRAP 6500+ LC-MS/MS System with OptiFlow Turbo V source and M5 MicroLC

More information

Fully Automated C100HT Biologics Analyzer Sample Preparation on a Biomek i5 Multichannel Workstation

Fully Automated C100HT Biologics Analyzer Sample Preparation on a Biomek i5 Multichannel Workstation Fully Automated C100HT Biologics Analyzer Sample Preparation on a Biomek i5 Multichannel Workstation Jane Luo 1, Clarence Lew 1, Alice Tanibata-Branham 2, Marcia Santos 1 1 SCIEX Separations, Brea, CA;

More information

Improving the Detection Limits for Highly Basic Neuropeptides Using CESI-MS

Improving the Detection Limits for Highly Basic Neuropeptides Using CESI-MS Improving the Detection Limits for Highly Basic Neuropeptides Using CESI-MS Stephen Lock SCIEX UK Overview Who Should Read This: Senior Scientists and Lab Directors with an Interest in Protein Analysis

More information

Improving Sensitivity for an Immunocapture LC-MS Assay of Infliximab in Rat Plasma Using Trap-and-Elute MicroLC-MS

Improving Sensitivity for an Immunocapture LC-MS Assay of Infliximab in Rat Plasma Using Trap-and-Elute MicroLC-MS Improving Sensitivity for an Immunocapture LC-MS Assay of in Rat Plasma Using Trap-and-Elute MicroLC-MS Using the SCIEX M3 MicroLC system for Increased Sensitivity in Antibody Quantitation Remco van Soest

More information

BioBA High Capacity Enrichment Sample Preparation Kit:

BioBA High Capacity Enrichment Sample Preparation Kit: BioBA High Capacity Enrichment Sample Preparation Kit: Biologics Bioanalysis Made Easy Ian Moore 1, Chang Liu 1, Gary Impey 1, Suma Ramagiri 1 1 Sciex, Concord, Canada Common Immunocapture Sample Extraction

More information

Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS

Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS Using the SCIEX M3 MicroLC system for Increased Sensitivity in Antibody Quantitation Remco van Soest and Lei Xiong SCIEX, Redwood City,

More information

Automating the C100ht Biologics Analyzer Sample Preparation on a Biomek i5 Liquid Handler Platform

Automating the C100ht Biologics Analyzer Sample Preparation on a Biomek i5 Liquid Handler Platform Automating the C100ht Biologics Analyzer Sample Preparation on a Biomek i5 Liquid Handler Platform Mervin Gutierrez, Tingting Li, and Marcia Santos SCIEX Separations, Brea, CA Introduction There is an

More information

Developing Quantitative UPLC Assays with UV

Developing Quantitative UPLC Assays with UV Developing Quantitative UPLC Assays with UV Detection for Antibodies & Other Proteins Steve Taylor 2011 Waters Corporation 1 Outline UPLC technology for RP protein separations Method development parameters

More information

Ultra-Sensitive Quantification of Trastuzumab Emtansine in Mouse Plasma using Trap-Elute MicroLC MS Method

Ultra-Sensitive Quantification of Trastuzumab Emtansine in Mouse Plasma using Trap-Elute MicroLC MS Method Ultra-Sensitive Quantification of Trastuzumab Emtansine in Mouse Plasma using Trap-Elute MicroLC MS Method Featuring SCIEX QTRAP 6500+ System with OptiFlow Turbo V source and M5 MicroLC system Lei Xiong,

More information

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography Agilent 1260 Infinity Bio-inert Quaternary LC System with Agilent Bio Columns

More information

Isotopic Resolution of Chromatographically Separated IdeS Subunits Using the X500B QTOF System

Isotopic Resolution of Chromatographically Separated IdeS Subunits Using the X500B QTOF System Isotopic Resolution of Chromatographically Separated IdeS Subunits Using the X500B QTOF System Fan Zhang 2, Sean McCarthy 1 1 SCIEX, MA, USA, 2 SCIEX, CA USA Analysis of protein subunits using high resolution

More information

ACCURACY THE FIRST TIME. Bring Your Esoteric Assays In-House Routinely SCIEX DIAGNOSTICS MASS SPECTROMETERS

ACCURACY THE FIRST TIME. Bring Your Esoteric Assays In-House Routinely SCIEX DIAGNOSTICS MASS SPECTROMETERS ACCURACY THE FIRST TIME Bring Your Esoteric Assays In-House Routinely SCIEX DIAGNOSTICS MASS SPECTROMETERS Your routine diagnostic testing lab requires accurate trace level quantitation of multiple compounds

More information

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution Clone Selection Using the Agilent 9 Infinity Online D-LC/MS Solution Application Note Biopharmaceuticals Authors Abstract Ravindra Gudihal and This Application Note describes the Agilent 9 Infinity online

More information

Biotherapeutic Non-Reduced Peptide Mapping

Biotherapeutic Non-Reduced Peptide Mapping Biotherapeutic Non-Reduced Peptide Mapping Routine non-reduced peptide mapping of biotherapeutics on the X500B QTOF System Method details for the routine non-reduced peptide mapping of a biotherapeutic

More information

"VUPNBUFE BOE RVBOUJUBUJWF

VUPNBUFE BOE RVBOUJUBUJWF "VUPNBUFEBOERVBOUJUBUJWF BOBMZTJTPGCJPMPHJDT 1" plus1ibsnbdfvujdbm"obmztjt4ztufn Genomics Cell Analysis Particle Characterization Capillary Electrophoresis Lab Automation Centrifugation Bioseparation Lab

More information

quantitate host cell DNA with high sensitivity and throughput resdnaseq Host Cell DNA Quantitation Systems: CHO, E.

quantitate host cell DNA with high sensitivity and throughput resdnaseq Host Cell DNA Quantitation Systems: CHO, E. quantitate host cell DNA with high sensitivity and throughput resdnaseq Host Cell DNA Quantitation Systems: CHO, E. coli, Vero, NS0 Precision counts Integrated real-time qpcr systems for quantitation of

More information

Journal of Chemical and Pharmaceutical Research, 2017, 9(3): Research Article

Journal of Chemical and Pharmaceutical Research, 2017, 9(3): Research Article Available online www.jocpr.com Journal of Chemical and Pharmaceutical Research, 2017, 9(3):390-400 Research Article ISSN : 0975-7384 CODEN(USA) : JCPRC5 Determination Of Clips Generated by Acid Hydrolysis

More information

One-stop cief and CE-SDS. One-day method dev. Results in a snap. FDA, PDQ.

One-stop cief and CE-SDS. One-day method dev. Results in a snap. FDA, PDQ. One-stop cief and CE-SDS. One-day method dev. Results in a snap. FDA, PDQ. Meet Maurice. He s a real problem solver. Need CE-SDS and cief info on mabs, ADCs and vaccines? Done. He ll even up you one and

More information

Contaminant bovine IgG assay

Contaminant bovine IgG assay ILA Application Note Contaminant bovine IgG assay INTRODUCTION Bovine IgG, a 16, Dalton protein, is one of the constituents of bovine serum. Fetal calf serum containing bovine IgG is commonly used in mammalian

More information

Microfluidic Assays for High Throughput Screening of Protein Quality in Bioprocess Development

Microfluidic Assays for High Throughput Screening of Protein Quality in Bioprocess Development 1 Microfluidic Assays for High Throughput Screening of Protein Quality in Bioprocess Development Bahram Fathollahi Microfluidics R&D CPAC Summer Institute 2009 2 What is LabChip Technology? Miniaturization

More information

OptiFlow Quant Solution for High-Sensitivity Quantitation

OptiFlow Quant Solution for High-Sensitivity Quantitation OptiFlow Quant Solution for High-Sensitivity Quantitation The Intelligent Union of Sensitivity, Ease-of-Use, and Robustness MicroFlow Sensitivity Analytical Flow Usability Transform Your Quantitation Microflow

More information

Application Guide. PA 800 plus Pharmaceutical Analysis System. SDS-MW Analysis. A51970AA April 2009

Application Guide. PA 800 plus Pharmaceutical Analysis System. SDS-MW Analysis. A51970AA April 2009 Application Guide PA 800 plus Pharmaceutical Analysis System SDS-MW Analysis April 2009 Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835 Application Guide PA 800 plus Pharmaceutical Analysis

More information

Biotherapeutic Peptide Mapping Information Dependent Acquisition (IDA) Method

Biotherapeutic Peptide Mapping Information Dependent Acquisition (IDA) Method Biotherapeutic Peptide Mapping Information Dependent Acquisition (IDA) Method Routine biotherapeutic accurate mass peptide mapping analysis on the X500B QTOF System Method details for the routine peptide

More information

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns Agilent AdvanceBio SEC 3 Å, 2.7 µm columns Application note Bio-Pharmaceutical Author

More information

Applications. Capillary Electrophoresis. Cell Lysate Analysis. Analysis of E. Coli Lysate. Technology

Applications. Capillary Electrophoresis. Cell Lysate Analysis. Analysis of E. Coli Lysate. Technology Applications Capillary Electrophoresis Cell Lysate Analysis Analysis of E. Coli Lysate withdeltadot Technology Abstract Analytical instruments with the capability of analysing complex mixtures of bio-molecules

More information

Mass Spectrometry Analysis of Liquid Chromatography Fractions using Ettan LC MS System

Mass Spectrometry Analysis of Liquid Chromatography Fractions using Ettan LC MS System GUIDE TO LC MS - December 21 1 Spectrometry Analysis of Liquid Chromatography Fractions using Ettan LC MS System Henrik Wadensten, Inger Salomonsson, Staffan Lindqvist, Staffan Renlund, Amersham Biosciences,

More information

Separation of Native Monoclonal Antibodies and Identification of Charge Variants:

Separation of Native Monoclonal Antibodies and Identification of Charge Variants: Separation of Native Monoclonal Antibodies and Identification of Charge Variants: Teamwork of the Agilent 31 OFFGEL Fractionator, Agilent 21 Bioanalyzer and Agilent LC/MS Systems Application Note Biosimilar

More information

Highly Resolved Separation of DNA Fragments on the Fragment Analyzer

Highly Resolved Separation of DNA Fragments on the Fragment Analyzer Highly Resolved Separation of DNA Fragments on the Fragment Analyzer Abstract Complete analysis of PCR amplicons requires an instrument capable of delivering reliable sizing and quality control parameters.

More information

electrophoresis tech Performance Comparison of the Experion Automated Electrophoresis System and SDS-PAGE for Protein Analysis

electrophoresis tech Performance Comparison of the Experion Automated Electrophoresis System and SDS-PAGE for Protein Analysis electrophoresis tech note 5299 Performance Comparison of the Experion Automated Electrophoresis System and for Protein Analysis Karen Zhu, Marie Nguyen, William Strong, and Christina Whitman-Guliaev, Bio-Rad

More information

TOYOPEARL GigaCap Series

TOYOPEARL GigaCap Series TOYOPEARL GigaCap Series INTRODUCTION Ion Exchange Chromatography (IEC) is one of the most frequently used chromatographic modes for the separation and purification of biomolecules. Compared with other

More information

Highly Resolved Separation of DNA Fragments on the Fragment Analyzer

Highly Resolved Separation of DNA Fragments on the Fragment Analyzer Highly Resolved Separation of DNA Fragments on the Fragment Analyzer Abstract Complete analysis of PCR amplicons requires an instrument capable of delivering reliable sizing and quality control parameters.

More information

A Robust cief Method: Intermediate Precision for the ph 5-7 Range

A Robust cief Method: Intermediate Precision for the ph 5-7 Range Robust cief Method: Intermediate Precision for the ph 5-7 Range Ingrid. Cruzado-Park, Scott Mack and Chitra K. Ratnayake SCIEX, Brea, C US Introduction Optimization of the numerous variables associated

More information

Development and Validation of the 3500 Series Genetic Analyzer for Human Identification

Development and Validation of the 3500 Series Genetic Analyzer for Human Identification Development and Validation of the 3500 Series Genetic Analyzer for Human Identification Jeff Sailus Forensic Science Applications Group 1 Development of a Next Generation Genetic Analyzer A Whole System

More information

Separation of SDS-Protein Complexes Using the CE-SDS Protein Kit ezid

Separation of SDS-Protein Complexes Using the CE-SDS Protein Kit ezid Separation of SDS-Protein Complexes Using the CE-SDS Protein Kit Catalog Numbers 148-4160 148-4161 ezid For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800-4BIORAD (1-800-424-6723)

More information

Column for High Performance,High-Binding Capacity Ion Exchange Chromatography:TSKgel SuperQ-5PW and Its Applications

Column for High Performance,High-Binding Capacity Ion Exchange Chromatography:TSKgel SuperQ-5PW and Its Applications ANALYSIS S e p a r a t i o n R e p o r t N o. 9 3 Column for High Performance,High-Binding Capacity Ion Exchange Chromatography:TSKgel SuperQ-5PW and Its Applications Table of Contents 1. Introduction

More information

Automated and quantitative

Automated and quantitative Automated and quantitative analysis of biologics. PA 800 plus Pharmaceutical Analysis System Genomics Cell Analysis Particle Characterization Capillary Electrophoresis Lab Automation Centrifugation Bioseparation

More information

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies Alternating Column Regeneration Using an Agilent 1200 Infinity Series Quick-Change Bio-inert 2-position/10 port Valve Application

More information

Sean M. McCarthy and Martin Gilar Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL RESULTS AND DISCUSSION

Sean M. McCarthy and Martin Gilar Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL RESULTS AND DISCUSSION UPLC Separation of DNA Duplexes Sean M. McCarthy and Martin Gilar Waters Corporation, Milford, MA, U.S. INTRODUCTION Over the past 2 years there has been a considerable amount of effort focused on the

More information

Encompassing Method Development, Small Scale Purification, and Post-Purification Analysis with the Investigator SFC System

Encompassing Method Development, Small Scale Purification, and Post-Purification Analysis with the Investigator SFC System Encompassing Method Development, Small Scale Purification, and Post-Purification Analysis with the Investigator SFC System Jacquelyn Runco and Rui Chen, Ph.D. Waters Corporation, Milford, MA, USA A P P

More information

Technical tips for ELISA and multiplex

Technical tips for ELISA and multiplex Technical tips for ELISA and multiplex Antibody pairs... 3 Protein standards... 4 Assay platform and format... 4 Titrate antibodies... 5 Optimize buffers and assay specificity... 6 Determine the optimal

More information

ssdna 100-R Kit Care and Use Instructions AG June 2015

ssdna 100-R Kit Care and Use Instructions AG June 2015 Care and Use Instructions 726479 AG June 2015 Beckman Coulter, Inc. 250 S. Kraemer Blvd., Brea, CA 92821 USA 2015 Beckman Coulter, Inc. Copyright, Licenses and Trademarks Copyright Beckman Coulter, Inc.,

More information

PRODUCT DATA SHEET. Carboxylated Fluorescent Gold Nanoparticles. Description. Characteristics

PRODUCT DATA SHEET. Carboxylated Fluorescent Gold Nanoparticles. Description. Characteristics PRODUCT DATA SHEET Carboxylated Fluorescent Gold Nanoparticles Description Cytodiagnostics carboxylated fluorescent gold nanoparticles is a unique product that combines our Cyto fluorescent dyes and gold

More information

Size Exclusion BioHPLC Columns

Size Exclusion BioHPLC Columns Size Exclusion BioHPLC Columns Size Exclusion Product Families Particle Porosity Functionalities Particle Pore Size Application Sizes Agilent Bio SEC- Silica Fully porous N/A um 00A, 0A, 00A High efficiency

More information

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column Disulfide Linkage Analysis of IgG1 using an Agilent 126 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column Application Note Biotherapeutics & Biosimilars Author M. Sundaram

More information

Contaminant bovine transferrin assay

Contaminant bovine transferrin assay ILA Application Note Contaminant bovine transferrin assay INTRODUCTION Bovine transferrin, a 76, Dalton glycoprotein, is one of the constituents of bovine serum. Transferrin found in serum can be associated

More information

Application Note. Abstract. Author. Biopharmaceutical

Application Note. Abstract. Author. Biopharmaceutical Innovator and Biosimilar Monoclonal Antibody-Peptide Map Comparison Using the Agilent 70 Capillary Electrophoresis System and Agilent MatchCompare Software Application Note Biopharmaceutical Author Arunkumar

More information

Automated Scouting of Stationary and Mobile Phases Using the Agilent 1290 Infinity II Method Development Solution

Automated Scouting of Stationary and Mobile Phases Using the Agilent 1290 Infinity II Method Development Solution Automated Scouting of Stationary and Mobile Phases Using the Agilent 129 Infinity II Method Development Solution Technical Overview Authors Edgar Naegele and Sonja Schneider Agilent Technologies, Inc.

More information

Contaminant human transferrin assay

Contaminant human transferrin assay ILA Application Note Contaminant human transferrin assay INTRODUCTION Human transferrin is an 8, Dalton glycoprotein found in human serum that facilitates transport of iron between cells. The iron-poor

More information

Mobile Phase Optimization in SEC Method Development

Mobile Phase Optimization in SEC Method Development Application Note Pharma & Biopharma Mobile Phase Optimization in SEC Method Development Author Richard Hurteau Agilent Technologies, Inc., Wilgton, DE, USA Abstract Aggregation of monoclonal antibody (mab)

More information

MultiQuant Software Version 3.0

MultiQuant Software Version 3.0 MultiQuant Software Version 3.0 Improving Data Quality and Processing Throughput with Better Peak Integration, Quantitative and Qualitative Compound Review for the Analysis of Food, Drinking Water, and

More information

Selecting the Appropriate Laboratory Equipment: The Coomassie (Bradford) Protein Assay Performed Manually Using Gilson PIPETMAN M Multichannel, and PIPETMAN L, and Automated Using GX-274 ASPEC and GX-241

More information

10 Reasons to Switch from SDS-PAGE to the Agilent 2100 Bioanalyzer

10 Reasons to Switch from SDS-PAGE to the Agilent 2100 Bioanalyzer Welcome to our E-Seminar: 10 Reasons to Switch from SDS-PAGE to the Agilent 2100 Bioanalyzer Presenter: Dr. Tanja Wulff Agilent Technologies 10 Reasons to Switch from SDS-PAGE to the Agilent 2100 Bioanalyzer

More information

This is not unique to CE. It took for instance about 25 years for HPLC to establish itself properly in the pharmaceutical industry.

This is not unique to CE. It took for instance about 25 years for HPLC to establish itself properly in the pharmaceutical industry. CE Pharm 2017 Troubleshooting Report: CZE and CE-SDS the Discussion Continues Workshop facilitators: Tim Blanc 1, Cari Sänger-van de Griend 2 Scribes: Nathan Lacher 3, David Michels 4, Zoran Sosic 5 1

More information

A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies

A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies Combining Size Exclusion Chromatography with Method Development and Light Scattering Application Note Biotherapeutics and Biosimilars

More information

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808 Reversed-Phase Optimization for Ultra Fast Profiling of Intact and Reduced Monoclonal Antibodies using Agilent ZORBAX Rapid Resolution High Definition 3SB-C3 Column Application Note Biopharma Authors James

More information

Proteomics PA 800 (Beckman) ProteomeLab System (Detect Instrument Picture to Jump to that section of SOP)

Proteomics PA 800 (Beckman) ProteomeLab System (Detect Instrument Picture to Jump to that section of SOP) Beckman PA800 Proteomics PA 800 (Beckman) ProteomeLab System (Detect Instrument Picture to Jump to that section of SOP) Windows Login: Username: Administrator Password: gold To Instrument 32 Karat sw:

More information

Human IgE ELISA Antibody Pair Kit

Human IgE ELISA Antibody Pair Kit For detection and measurement of human immunoglobulin E Catalog #01993 1 Kit for 6 Plates Product Description The Human Immunoglobulin E (IgE) ELISA Antibody Pair Kit is for customers who want the flexibility

More information

LabChip 90 System with DataViewer Software

LabChip 90 System with DataViewer Software Revolutionizing DNA and Protein Gel Electrophoresis LabChip 90 System with DataViewer Software The Proven Alternative to Slab Gels. The massive flow of information emerging from today s genomic and proteomic

More information

CSE Separation of Cancer Cell Lysate

CSE Separation of Cancer Cell Lysate Application Note: CSE Separation of Cancer Cell Lysate No. 9.200411 Rev. A 2004 Target Discovery, Inc. Target Discovery, Inc. 4015 Fabian Way Palo Alto, CA 94303 Tel: 650-812-8100 www.targetdiscovery.com/eotrol

More information

Pharmacokinetic Analysis of the mab Adalimumab by ELISA and Hybrid LBA/LC/MS: A Comparison Study

Pharmacokinetic Analysis of the mab Adalimumab by ELISA and Hybrid LBA/LC/MS: A Comparison Study Pharmacokinetic Analysis of the mab Adalimumab by ELISA and Hybrid LBA/LC/MS: A Comparison Study Featuring the SCIEX BioBA Solution Xi Qu 1, Shuyu Hou 1, Meghan McCann 1, Caroline Becker 1, Xun Wang 1,

More information

The Simple Western Approach to Vaccine and Clinical Protein Research

The Simple Western Approach to Vaccine and Clinical Protein Research The Simple Western Approach to Vaccine and Clinical Protein Research The Simple Western Approach to Vaccine and Clinical Protein Research Broadcast Date: Thursday, October 4, 2012 Time: 1 pm ET, 10 am

More information

Standard Sensitivity NGS Fragment Analysis Kit (1 bp 6000 bp) User Guide (DNF ) (DNF )

Standard Sensitivity NGS Fragment Analysis Kit (1 bp 6000 bp) User Guide (DNF ) (DNF ) New Software Version Refer to red highlighted text for additional updates Standard Sensitivity NGS Fragment Analysis Kit (1 bp 6000 bp) User Guide (DNF-473-0500) (DNF-473-1000) For use with the Fragment

More information

High Sensitivity NGS Fragment Analysis Kit (1 bp 6000 bp) User Guide (DNF ) (DNF )

High Sensitivity NGS Fragment Analysis Kit (1 bp 6000 bp) User Guide (DNF ) (DNF ) New Software Version Refer to red highlighted text for additional updates High Sensitivity NGS Fragment Analysis Kit (1 bp 6000 bp) User Guide (DNF-474-0500) (DNF-474-1000) For use with the Fragment Analyzer

More information

Rhinophase -AB, Zirconia-based Monoclonal Antibody Purification System

Rhinophase -AB, Zirconia-based Monoclonal Antibody Purification System Rhinophase -AB, Zirconia-based Monoclonal Antibody Purification System Welcome to the sixth issue of ZirChrom's electronic newsletter. This newsletter introduces a biocompatible stationary phase useful

More information

SDS-CGE Method Development for Purified Protein Vaccine Antigens

SDS-CGE Method Development for Purified Protein Vaccine Antigens SDS-CGE Method Development for Purified Protein Vaccine Antigens Michael Leach, Elena Newman and Bruce Carpick Sanofi Pasteur Analytical R&D, Biochemistry Overview SDS-CGE method development for purified

More information