For most biopharmaceuticals,

Size: px
Start display at page:

Download "For most biopharmaceuticals,"

Transcription

1 B i o P r o c e s s TECHNICAL Author Insights Online Exclusive Assay Acceptance Criteria for Multiwell-Plate Based Biological Potency Assays Draft for Consultation C. Jane Robinson, Michael Sadick, Stanley N. Deming, Sian Estdale, Svetlana Bergelson, and Laureen Little For most biopharmaceuticals, potency is assessed in a bioassay by comparing dose response curves of the test material and a reference standard. As with all analytical techniques, such assays require criteria by which their execution can be judged objectively to be valid, regardless of whether the desired or expected result is obtained for the test sample. The purpose of this paper is to provide guidance on setting assay acceptance criteria (AAC) for potency assays based on multiwell plates. Multiple components of the overall assay system from instruments to incubation media need to be within defined limits to permit execution of a valid assay, so they are tested for suitability either before or during the assay. Because of the complexity of bioassay systems, not every relevant factor can be controlled. Thus, it is Product Focus: Biologics Process Focus: R&D, QC Who Should Read: Product and process development, analytical, QC Keywords: Immunoassays, potency assays, product release, reference standards, control samples, sample acceptance Level: Advanced 30 BioProcess International 12(1) January 2014 necessary to rely strongly on analysis of data produced by each individual assay to determine whether that assay was executed correctly. This paper discusses criteria that can be applied to the assay results and the importance of assay design in selecting useful acceptance criteria. Scope Why Multiwell-Plate Based Assays Specifically? Analytic dilution assays using multiwell plates constitute the most common platform for measurement of biological activity by in vitro bioassay or measurement of immunoreactivity by immunoassay. The multiwell format provides a convenient means of handling the necessary number of doses and replicates, and it is supported by the availability of a wide range of plate types with a standardized footprint and supporting equipment and measurement systems. Multiwell-plate formats introduce specific artifacts to the measured responses. Thus, assay design especially the positioning of individual samples within and between plates is crucial to obtaining valid results and thus needs to be considered when setting acceptance criteria for an assay. Much of the following discussion is based on cell-based assays in 96-well plates, which reflects their wide use in Abbreviations and Terms 4PL: four-parameter logistic nonlinear regression model AAC: assay acceptance criteria CV: coefficient of variation : the dose of a drug that induces a response that is 50% of the maximum response PLA: commercial software package designed for the analysis of bioassay and immunoassay dose-response curves (Stegmann Systems GmbH) R 2 : coefficient of determination RSS: the residual sum of squares SAC: sample acceptance criteria Shewhart SPC: control chart named after Walter A. Shewhart SPC: statistical process control StatLIA: commercial software package designed for the analysis of bioassay and immunoassay dose-response curves (Brendan Technologies, Inc.)

2 Summary This paper is restricted to discussion of tests applied to the responses of a bioassay system to reference standards, controls and test samples obtained during performance of a potency assay. Assay acceptance criteria are based primarily on comparison of dose response curves of control samples with a reference standard, all of which should be well characterized in the assay system. At least one control sample should be known to behave similarly to the reference standard in the assay system. Both this control and the reference standard should be known to behave similarly to the expected behavior of test samples. We propose the name assay control sample for the control material that behaves similarly to the reference standard and the expected behavior of the test samples. The origin of the assay control sample should be as independent of the reference standard and test samples as is possible within the constraint that all behave similarly in the assay system. For plate-based biological potency assays, we propose the following two separate sets of acceptance criteria: assay acceptance criteria (AAC) and sample acceptance criteria (SAC). We propose a two-level, sequential assessment of acceptance criteria. First, AAC are assessed. Failure means that the entire plate is invalid. There is no processing of test sample data. Passing AAC allows processing of test sample data. Second, each test sample potency determination is then subjected to its own SAC. If it passes, then that test sample potency measurement is valid. If it fails, then that particular test sample potency quantification fails. Other test sample determinations on the same plate may still be valid. Similarity of dose response curves of reference standards and assay control samples is an essential AAC. Similarity of dose-response curves of the reference standard and test sample is an essential SAC. AAC and SAC applied to an assay should be demonstrated to be useful in judging the validity of the assay. Both the criteria and the limits set on their values should be reviewed and modified, if appropriate as more assays are performed and further data are accumulated. potency testing. With some adaptation, however, many of the arguments can be applied to other assay systems and plate formats for plate-based potency assays. What Types of Assay? Considerations for setting AAC apply to a number of different assay types, including functional assays (in which a biological response is measured) biochemical assays (such as clotting-factor activity assays) considered as functional assays or as a separate class binding assays (that measure binding of a ligand, receptor, cofactor, and so on), which may be potency assays, in which binding is the mode of action (MOA), or surrogate potency assays immunoassays such as enzymelinked immunosorbent assays (ELISAs) that measure binding of an antibody preparation (monoclonal or polyclonal) to one or more epitopes distinct from functional assays in which binding of an antibody induces a functional biological response hybrid assays (e.g., immunoassaycofactor binding assays). Many assays consist of multiple steps. For example, a functional assay may involve stimulation of cytokine secretion followed by immunoassay measurement of the secreted protein. AAC are generally set for the overall output, but may, in addition, be set for some individual steps. Potency assays for some types of products e.g., advanced therapy medicinal products (ATMPs) may require atypical assay designs and, consequently, atypical acceptance criteria. AAC or System Suitability? Guidelines for testing a system to demonstrate its suitability for an analytical procedure have been developed primarily for physicochemical techniques. ICH Q2(R1) states that a series of system suitability parameters (e.g., resolution test) is established to ensure that the validity of the analytical procedure is maintained whenever used and then provides specific examples for liquid and gas chromatography (1). USP chapter <621> on chromatography states, To ascertain the effectiveness of the final operating system, it should be subjected to a suitability test prior to use. The essence of such a test is the concept that the electronics, the equipment, the specimens and the analytical operations constitute a single analytical system, which is amenable to an overall test of system function (2). A useful definition, also from the field of liquid chromatography, is this: System suitability is the checking of a system to ensure system performance before or during the analysis of unknowns (3). So a system suitability test determines whether an analytical system is fit for use. In physicochemical analyses, making that decision may be possible before test samples have been committed to analysis. As with physicochemical analyses, multiple components of the overall assay system need to be within defined limits to permit execution of a valid bioassay. For example, instruments, media, and environmental conditions are tested for suitability either before or during an assay. However, guidelines and examples designed for physicochemical techniques are not necessarily appropriate or sufficient for bioassays. Bioassays and (generally to a lesser extent) immunoassays tend to be susceptible to a greater number of factors than are most physicochemical analytical techniques. Some of those factors may be poorly controlled or not identified. For example, biological media may contain unidentified or undetected components that vary from batch to batch and affect the response of an assay system. Because such uncontrollable or unidentified sources of variability can cause assay-to-assay variability, it is necessary to rely strongly on analyzing data produced by each individual assay to determine whether that assay was executed correctly. Which tests of those data will be useful in determining the validity of the assay should be investigated during assay development and characterization. The general term acceptance criteria has been defined to be conditions which must be fulfilled before an operation, process or item, such as a piece of equipment, is considered to be satisfactory or to have been completed in a satisfactory way (4). 32 BioProcess International 12(1) January 2014

3 Submission of Comments The authors invite comments on this draft best-practice article from interested parties to assist in completion of a guidance on assay acceptance criteria for multiwell-plate based biological potency assays. Send comments by to the Biopharmaceutical Emerging Best Practices Association at bebpa@ surewest.net. Please start the subject line with AAC. Comments should be received by 31 May All comments must be accompanied by a name, affiliation, and address and will be given due consideration by the authors. Any comment submitted could be posted on the BEBPA website ( white-papers). Comments may be posted in their entirety, or extracts may be posted instead. In the latter case, it will be noted that only part of the comment is posted. If you would like your name and affiliation to be posted along with your comments, please state that BEBPA has your permission to do so. What is the distinction between AAC and system suitability tests? The former are often considered to be a subclass of the latter. Both are set based on data acquired during assay development, characterization, and validation. System suitability may be checked either before or during the performance in which a test sample is assayed. The term assay acceptance criteria is generally used to describe conditions that must be met by data derived from an actual assay in which a test sample is assayed. Because biological assay systems are more variable than physicochemical systems, it is generally necessary for more system testing to be run simultaneously with (rather than prior to) sample testing. For potency assays using multiwell plates, some tests are generally considered to measure system suitability and others are generally considered to assess assay acceptance criteria. The important consideration for both system suitability tests and AAC is that they should be appropriate to the specific assay system, the precise purpose of the particular assay being performed, and the assay design. Classification of a particular test as a system suitability test or AAC may not be necessary or particularly helpful. This paper is restricted to discussion of tests applied to data derived from the responses of test samples, controls, and reference standards obtained during performance of an assay intended to yield a potency value for a test sample. Pharmacopeial Requirements: With pharmacopeial assays, all specified AAC will need to be met for those performed under relevant quality systems. Those criteria can provide useful guidance in setting criteria for similar assays and similar products. However, pharmacopeial assays generally specify fewer AAC than would be considered necessary for assays used for product release, so manufacturers would probably need to set additional AAC for release assays and develop appropriate criteria for novel assays and products. Assay Acceptance and Sample Acceptance For plate-based biological potency assays, we propose two separate sets of acceptance criteria: AAC and sample acceptance criteria (SAC). The former are based on responses of control samples and reference standards, and the latter are based on the responses of each separate test sample. Making this distinction permits the validity of an assay (or assay subsection) to be judged separately from that of each separate test sample. If an assay fails, then there is no processing of its test sample data. If analysis of data from one test sample fails the SAC, then that particular test sample potency quantification fails. Other test sample determinations on the same plate should be assessed separately and may still be valid. Assay Acceptance: AAC are based on control samples and a reference standard. These are materials that are well characterized in the assay system and are independent of the test samples. At least one control sample should be known to behave similarly to the reference standard in this assay system, and both that control and the reference standard should be known to behave similarly to the expected behavior of the test samples. (See the Similarity of Dose Response Curves section below.) The control material that behaves similarly to the reference standard and test samples is often called the assay control, control material, or control sample. We propose adopting the term assay control sample. Other types of controls include those designed to exclude responses of a system that are not caused by specific actions of the test samples and those designed to demonstrate a positive response by a different mode of action. The former are often called negative controls and include materials such as excipients of the test sample or antibodies of a different specificity. Whereas the assay control sample and reference standard must be tested at multiple dilutions, other controls may be tested at one concentration or a few concentrations. One feature of multiwell-plate based assays that imposes constraints on assay design is plate-to-plate variability. Because of that, AAC should be applied to each plate independently, which means that dilution curves for an assay control sample and a reference standard should be run on every plate. If assay design constraints dictate that an exception be made to this rule, then it must have been clearly demonstrated during assay characterization and validation that plate-to plate variability does not affect the assay s validity. AAC are based primarily on comparing dose response curves of the assay control sample(s) with a reference standard, both of which should be wellcharacterized preparations. The origin of the assay control sample should be as independent of the reference standard and test samples as is possible within the constraint that all should behave similarly in the assay. Different lots of material from the same production method are commonly used. Two independent dilutions of the reference standard material (one serving as the reference and one as the assay control sample) are not sufficient; using two dilution series of the same material is testing only the dilution procedure and subsequent handling. In early assay 34 BioProcess International 12(1) January 2014

4 development, however, an independent assay control sample may not be available, so using two samples of the reference standard could be the only option. In this case, the two samples should be processed as independently as possible, ideally starting from separate vials and if possible, using alternative intermediate dilution schemes. Ideally, the assay control sample and reference standard should be as similar as possible to the test samples in excipient formulation, concentration, aliquot volume, container, and so on. The assay control sample, reference standard, and test samples should be prepared and tested in exactly the same way. However, an initial step such as reconstitution of a lyophilized reference standard or dilution of a more concentrated solution may be necessary. Commonly, the reference standard may be presented in a formulation buffer different from that of the test samples to permit its storage as a frozen solution. It then requires dilution (concentration permitting) in the formulation buffer of the test samples to render it as similar as possible to the test samples. Similarly, the formulation of the assay control sample and other control samples should be brought as close as possible to that of the test samples. The assay control sample must be compared with the reference standard and evaluated in terms of curve shape and potency. Assay control sample potency is evaluated against a statistically generated assay control chart. The concept of a control chart for the assay control sample is essentially simple: Multiple independent executions of an assay are performed in which the assay control sample is included as a test sample, with no preexisting expectations other than curve similarity to the associated reference standard. After a certain number of assay repetitions (e.g., n = 25 or n = 50), the arithmetic average and associated standard deviation (SD) of that value are determined for the assay control sample potency. A control chart acceptance criterion is then established using the mean assay control sample potency, with plus or minus some multiple of the SD defining the upper and lower limits of acceptability. Although it is common practice to set statistical process control (SPC) limits at plus or minus three times the estimated SD ( Shewhart limits ), in many bioassay applications the estimated SD is based on only a few measurements and is thus a very uncertain estimate. An approach to setting SPC limits that takes this uncertainty into account is to use tolerance intervals (5). These intervals are wider when based on only a few measurements and (usually) become narrower as more and more data are acquired until, with an infinite amount of data, they become the same as the ±3 SD limits associated with typical Shewhart SPC charts. In the article Specification Setting: Setting Acceptance Criteria from Statistics of the Data (6), upper and lower control limits (tolerance limits) are provided for as few as five assays (with a corresponding multiplier of ±10.75) to as many as 199 assays (with a corresponding multiplier of ±3.03). Using the table provided in this reference, one can establish a reasonable initial control chart with 25 assays (with a corresponding multiplier of ±4.05), and then revise those upper and lower control limits as the number of collected values for the assay control sample relative potency increases. Sample Acceptance: SAC are tests applied separately to each test sample. Their passing or failing is independent of other test samples. Similarity of sample curve shape to that of the reference curve is a standard test. Other SAC (e.g., variability of replicates) are usually similar to those for reference and assay control samples, but some may be different. For example, with some assay systems (particularly those that may not use the whole dose response curve), the SAC may allow for fewer doses to be used in the curve fit than for the reference standard or assay control sample. Limits may be set for the of test samples to ensure that the dose response curve falls within the validated dose range. The reportable value is defined in the testing protocol. It is typically the mean of a defined minimum number of assay results from valid independent assays, with independent being defined in the protocol. Most protocols specify that failure of the assay to meet AAC or of a sample to meet SAC more than a specified maximum number or percentage of times will trigger specified actions. Assay Design Use of multiwell plates introduces particular sources of potential artifacts in the measured responses of an assay. Wells in the corners, on the edges, and in the center of each plate have different environments, and the individual plates may be subject to slightly different conditions, any of which differences may affect responses. These effects are observed so commonly that the terms well effects, edge effects, and plate effects are widely used in the bioassay field. In many assays particularly cell-based assays with long incubation times the edge effects are so significant that often only the 60 center wells of each 96-well plate are used for measurements and the edge wells are filled with media or other solutions. Evaporation of medium from external wells, even in a humidified atmosphere, is a major contributor to edge well effects. Recently developed 96-well plates with reservoir troughs around the edge can reduce the edge effect and, depending on the assay, may permit use of all 96 wells of each plate. During assay development, use of plate uniformity (the same dose applied to all wells, including the edge wells) can facilitate assay optimization by permitting assessment and possibly reduction or elimination of platepositional effects, including edge anomalies. Nonrandom distribution of samples and doses for example, placing all dilution curves for a particular sample in edge rows of the plates or reference standard in one plate and test sample in another can introduce bias to measured relative potencies. Good assay design should reduce such bias. Completely random distribution of samples and doses is generally infeasible, so block or other structured designs can be used (7). January (1) BioProcess International 35

5 Reference standard and assay control sample dilution curves should be included on every plate to permit the AAC to be applied to each plate. In the past, some assay designs have included no assay control sample. In such cases, assay acceptance would be based solely on the reference standard dilution curve. That is a less rigorous test of assay validity, so this design should be avoided. In some designs, several plates are treated as one unit, which passes or fails the AAC as a single unit. Usually this is done to reduce the proportion of wells required for reference standard and assay control sample and to increase the number of wells available for test samples. Treating several plates as one unit is not recommended, however: Plate-to-plate variability can limit the use of this approach. Within plates, replicate dilution series of the reference standard, assay control sample, and test samples should be placed in nonequivalent and nonsimilar positions (avoiding locating replicates of the dilution series all in edge wells, or all in center wells, in adjacent rows, all in the top of the plate, all in the bottom of the plate, and so on). Most assays are constrained by the number of wells and plates that can be used, so assay design is a compromise between the ideal and the feasible. The adequacy of a proposed design should be investigated during assay development and characterization, and the AAC should be selected and set accordingly. Commonly Used AAC and SAC Similarity of dose response curves is an absolute requirement for determining relative potency. In current practice, some form of assessment is almost universally applied. It is recommended that at least one control sample known to behave similarly to both the reference standard and the expected behavior of test samples in the particular assay system (the assay control sample) be run as a full dose response curve. Similarity of dose response curves for the reference standard and assay control sample is an essential assay acceptance criterion, and similarity of dose response curves of the reference standard and test sample is an essential sample acceptance criterion. Depending on the assay, whole or partial response curves (when doses do not permit an asymptote to be reached) and linear responses may be measured, and various combinations of criteria can be applied (see below). 4PL is commonly used for curve analysis. In-house programs and commercial software (e.g., PLA and StatLIA) are used. In some cases, the F-test is used initially to judge similarity and then, as historical data are acquired, it is replaced by equivalence testing. Against the recommendations of this paper, some assays do not include an assay control sample or other controls. In such cases, AAC are based on the reference standard dose response curve or comparison of test sample and reference standard. Curve Slope: Slope is an acceptance criterion applied in most assays, but in a few cases it is used for trending only. Most often, the ratio of slope is used, forming part of curve similarity testing. Limits on the ratio can be defined as confidence intervals or ranges (e.g., ). In some cases, absolute values are defined for the slopes and used either in addition to the ratios or alone. Absolute values are based on validation or control chart data (e.g., mean ±3 SD from historical data). When setting limits of ±3 SD, care should be taken to use a sufficiently large data set. AAC limits can differ from those for SAC. Lower asymptote is a useful AAC and SAC for some assays. Limits can be set as ratios, forming part of similarity testing, and as absolute values based on historical data. Lower asymptote values may be used only for trending. Upper Asymptote: The same comments apply as for the lower asymptote, with the additional point that an upper limit may be placed on the value to account for instrument limitations, particularly when optical density (OD) read-outs are used. In the case of OD, when the read-out is the difference between OD at two wavelengths, care should be taken that neither exceeds the limit of the instrument. Ratio of Upper Asymptote to Lower Asymptote: For some assays, an absolute value with limits can be assigned to this ratio, or limits can be set on the ratio between a reference and assay control sample or test sample of values for this ratio (ratio of ratios). The ratio of upper to lower asymptote can be useful for trending (e.g., indicating changes with cell passage number). Inflection Point: The inflection point is not commonly used as an independent AAC or SAC, but proves useful in a few cases. In 4PL, this corresponds to the. : The is an absolute value. is widely used for trending but not commonly used as an AAC or SAC. The ratio of of reference to of assay control sample or test sample is sometimes used to assess potency. Goodness of Fit: Goodness of fit is an essential AAC and SAC and is almost universally applied. R 2 is widely used, with limits commonly set from R to R Testing R 2 is relatively simple, but R 2 may not provide a very sensitive indicator of goodness of fit. During assay development, it can be useful for selecting the most appropriate model to plot residuals against dose to reveal whether there is a random or a systematic deviation from the model. Residual sum of squares (RSS) and chi squared (χ 2 ) are also used, the latter primarily for quantal assays. The most appropriate model and test depend on the individual assay and should, as with all acceptance criteria, be checked during routine use. Potency of Control: With the exception of a few unusual assays, an assay control sample (run as full dose response curve) should be included in every assay usually on every plate. At present, some assays do not include any control samples and base the AAC on the reference standard alone or in comparison with the test sample. Other assays include a positive control at only one dose level or a few dose levels. Limited space on assay plates is one reason for the latter case. Limits ranging from % to % are typical, with % being very common. Limits are often set based on historical data (e.g., mean ± 3 SD). When setting limits of ±3 SD, care should be taken to use a sufficiently large data set. And care should be taken to define assay blanks and negative controls. Blanks usually refer to the absence of the active pharmaceutical ingredient (API), corresponding to the addition of a sample of the formulation buffer. Negative controls usually consist of relevant substances without a specific action in the particular assay (e.g., in the assay of a therapeutic antibody, a negative control might be antibody without specific binding to the ligand used in the assay). (continued) 36 BioProcess International 12(1) January 2014

6 Selection of assay and sample acceptance criteria is inextricably linked to assay design. For example, if replicate curves are made on a single plate, then acceptance criteria might include limits on the coefficient of variation (CV) for the slopes of the replicate curves. If separate potency determinations are made from each plate in a multiplate assay, then acceptance criteria limits might be set on the CV of potency determinations. Commonly used acceptance criteria and some typical assigned values are listed in the Commonly Used AAC and SAC boxes. Monitoring and Trending Values observed for assay (and sample) acceptance criteria can be used for assay (and sample) monitoring and trending. For the simplest case of assay monitoring, visual inspection of run charts (trend charts) displaying observed data in a time sequence can reveal aberrant results and indicate whether an assay is stable or drifting. Comparison with the results of system suitability tests (e.g., cell density before harvest for an assay or degree of cell confluence before dosing) and records of critical reagents and equipment can help reveal the cause of drift or aberrant results. In addition to assay and sample acceptance criteria, some parameters are reported for information only. Those parameters do not determine the acceptance or failure of an assay or Commonly Used AAC and SAC (continued) Ratio of Values for Different Control Samples: This ratio is not commonly used. Use of more than one positive control is not common, but the ratio of high to medium or high to low positive controls, for example, can be useful AAC. The value of the ratio of positive to negative control responses will vary greatly with changes in response of the negative control if the negative control is very small. Minimum Number of Doses Used in Curve Fitting: Usually set as an AAC and SAC, the value depends on the particular assay and whether it is a linear or full curve fit. Usually all doses are required, but in some cases exclusion of one or more points is permitted (see Maximum Number of Statistical Outliers Excluded section below). Depending on assay design, this generally results in a minimum of 6 10 doses, with eight being very common. For some assays (e.g., those in which wide ranges of test sample potencies may be encountered) a smaller number of doses, specified as being consecutive, may be set as a SAC. Minimum Number of Doses in Linear Part of Dose Response Curve: Usually set as an AAC and SAC, both for linear and full curve fits; values generally range from three to six, with three and four being most common. Minimum Number of Doses in Upper and/or Lower Asymptote: This criterion is used in some assays with the most common value being two. This criterion is sometimes used during development and subsequently discarded. Minimum Dose Range Used in Curve Fit: Sometimes set explicitly as an assay acceptance and sample acceptance criterion, this commonly specifies that a minimum dose range (e.g., %) should lie within the linear range. In many assays, the minimum dose range is set implicitly as an acceptance criterion because the protocol specifies doses tested and the minimum number of doses used in curve fitting. Maximum Number of Statistical Outliers Excluded: For many assays, a maximum number of statistical outliers that can be excluded is set as an assay acceptance and sample acceptance criterion. This can be defined as the number of doses (all replicates) or individual points that can be excluded per curve, or the number of replicates per dose, or a combination of these. The statistical test used to identify an outlier should be defined and have been shown to be suitable for the assay. Limits on the number of permitted exclusions depend strongly on the assay. Typical examples are one dose or one point per eight-point curve or per sample. Many assays allow no exclusion of outliers; others allow exclusion only if an anomaly can be attributed to experimental error. To prevent bias, any point suspected of being subject to experimental error normally should be excluded before its value is determined. However, it is recognized that experimental error may sometimes be identified only after a point is observed to be anomalous. Variability of Replicates: Assessing the variability of replicates is essential, with limits normally set as AAC and SAC. Replicates commonly consist of the following: Replicate wells containing the same solution (aliquots of the same dilution point in a single dilution series) Replicate dilution series from a common starting solution Replicate aliquots, vials, ampules, and so on. The data may be analyzed from one multiwell plate or across several plates. When variability of replicates is quoted, what is replicated must be defined exactly. In the first case above, the variability measures the errors in pipetting solution into the wells and errors introduced in subsequent steps. It does not assess errors in preparation of the starting solution or the dilution series. For measuring relative potency, ideally the handling of each sample should be the same, and each replicate should capture independently as many steps as possible from the handling process. Variability depends on the assay. It is commonly expressed as %CV. For enzyme-linked immunosorbent assays (ELISAs) and cell-based potency assays for products such as cytokines or MAbs, AAC and SAC are generally set at %CV 10 30%. For vaccine assays, in which dilution series steps of log 10 may be used, considerably broader limits may be acceptable depending on the protocol and requirements of a given assay. An alternative option is to set a SAC as the relative 95% confidence interval around the mean value for test sample relative potency (e.g., at % or %). The requirements for the limits on assay variability are often subject to requirements for the precision of the assay result. If greater precision is required, one approach can be to modify the assay protocol to include more replicates. Although variability between replicate wells of one dose (measured by CV) is the most widely used assessment of replicate variability, other assessments may be used (e.g., variability between replicate potency determinations from single dose curves) depending on the assay design. Variability of replicates is one of the most important tools for trending. 38 BioProcess International 12(1) January 2014

7 sample, but they can be useful in monitoring and trending. Monitoring and trending of as wide a range as feasible of assay response characteristics is useful, particularly in the early stages of assay development and use. As data accumulate over a large number of assays, incorporating variability in operating conditions (such as reagent batches and operators), it can become evident which characteristics may be useful for continued monitoring and trending and which should be adopted or retained as acceptance criteria. Inclusion of an assay response characteristic in monitoring or trending should not carry an automatic expectation that it should be adopted as an acceptance criterion and have limits set. Such an expectation can discourage the investigation of response characteristics and eventual selection of optimum sets of acceptance criteria. When statistical analysis is applied to observed values, a statistical process control (SPC) chart can be obtained, permitting an objective analysis of the variation in assay performance and allowing limits to be set to indicate when action must be taken to prevent an assay drifting into failure. The use of statistical control charts to map the ongoing performance and stability of reference material during routine assays can be a useful quality control tool allowing for early detection of adverse trends (8) and SPC charts are a powerful tool for showing auditors the continual validation of an assay (9). Evolution of Acceptance Criteria During Development Assay and sample acceptance criteria (and their assigned values) will change during assay and product development processes as a result of improvement in assay performance, accumulation of more data on that performance, and stricter requirements at later stages of product development. Even when an established assay platform is applied to an additional product, it should be expected that there will be a development process for establishing assay and sample acceptance criteria. During assay development and on accumulation of performance data, it would normally be expected that limits on some criteria would be tightened. However, it may become apparent that initial data collected over a short period of time did not reflect the full variation in assay conditions (e.g., variation in reagent batches) that will be encountered over a longer period. The criteria may therefore be initially set too tightly and subsequently need to be widened. Additional criteria may be included later on. Conversely, with accumulation of data, it may become evident that some criteria initially set do not reflect the assay s validity. Those criteria should be removed as they do not contribute to assessing assay validity, and their inclusion could result in rejection of assays that are fit for purpose. Their values can be recorded for the purposes of monitoring and trending. Acceptance Criteria Absolute and Relative Values: A potency assay is comparative, measuring the potency of a test sample or control sample relative to that of a reference standard. A comparative assay is based on the assumption that a factor that affects the response of a system to a test sample should affect the response of that system to a standard sample equally. Therefore, absolute values for characteristics of a response curve should not be critical. In practice, however, most assay systems function adequately over only a limited range of conditions reflected by a limited range of acceptable values for some of the dose response curve characteristics. An unusually high or low value for the, for example, can indicate that an assay system is not behaving as usual. Similarity of Dose Response Curves: The response of a bioassay system to a sample can be affected by a variety of external factors, some of which may not be controlled, so potency measurement cannot be an absolute value. Bioassays are comparative, with the biological activity of a test material measured relative to that of a reference preparation (10). If two preparations are sufficiently similar, then their responses should be affected equally by any variation in the system, and relative potency should remain constant. It is a fundamental requirement for obtaining a valid relative potency that the reference standard and test sample must behave similarly in an assay system and hence their dose response curves must have the same mathematical form. Any displacement between those curves along the log-concentration axis must be constant at all responses. This constant displacement is used to estimate relative potency. Nonsimilarity of two preparations may lead to dose response curves of different mathematical forms with variation in the magnitude of displacement between the curves. In this case, any attempted measurement of relative potency would vary depending on the response level at which it was measured. Similarity of dose response is thus an essential assay acceptance criterion. To determine whether a reference standard and test sample demonstrate the same dose response relationship in a given bioassay, it is necessary to measure the response of each at several doses spread over an appropriate range. Assessing the similarity of dose response curves depends on statistical analysis of the data. For some pharmacopeial assays, the method of analysis may be specified precisely. In many cases, however, only general guidance can be obtained from pharmacopeial sources. It is beyond the scope of this paper to discuss the relative merits of different statistical methods, such as the F-test and equivalence testing, for assessing similarity in various circumstances. Further information can be found in the literature (11 15). Consultation with a statistician who is experienced specifically in analysis of bioassay data is strongly recommended during assay development and validation to assist in establishing methods of data analysis appropriate to a particular assay and setting appropriate AAC. For many well-characterized bioassays, a mathematical transformation of the response is selected to give a linear relation (over a range of doses) with log dose. This is the parallel-line assay and, in this case, dose response curve similarity is 40 BioProcess International 12(1) January 2014

8 referred to as parallelism. Even when parallel-line analysis is used, it may be appropriate to use asymptotic values as additional assay acceptance criteria. With the increasing availability of software packages, four-parameter logistic (4PL) curve fitting is widely used. The four parameters are minimum asymptote (A), Hill slope (curve steepness, B), inflection point (C) and maximum asymptote (D). Each of these parameters may provide an assay acceptance criterion for determining similarity of the dose response curves of test samples and reference standards and/or as an absolute value. 5PL fits are used less commonly but can provide a better fit for dose response curves that are not symmetrical around the inflection point. The Commonly Used AAC and SAC boxes list commonly used AAC and SAC, including comments on their utility and some typical values. AAC are applied to reference standard and control samples, whereas SAC are applied to each individual test sample. The list incorporates information provided by delegates at the 2013 BEBPA bioassay conference (16). Common Practices That Cause Problems: Certain common practices in setting acceptance criteria are the cause of frequent problems. One such practice is acceptance being made too dependent on an individual data point. A protocol illustrating this point involves standard, assay control sample and test samples tested in duplicate dose-response curves on a plate. Upper asymptotes are determined as the mean of duplicate responses at maximum dose with a maximum acceptable percentage difference between upper asymptotes on one plate. The reportable value is the mean of the potencies determined on two replicate plates, with a maximum acceptable percentage difference between the two potencies. One aberrant well at the maximum dose of the assay control sample on one plate results in the upper asymptote of the assay control sample being aberrant compared with the reference standard so the plate is rejected. No reportable value is obtained because of one aberrant well out of 120. Another common cause of problems is failure to ensure that each of the acceptance criteria is valuable in judging the validity of the assay as established during the assay characterization and continued monitoring. Inclusion of unnecessary criteria may result in the rejection of assays that are in fact fit for purpose. For example, in some systems, may vary widely between assays which are fit for purpose, so setting tight limits on an absolute value for could result in pointless rejection of an assay. References 1 ICH Q2(R1): Validation of Analytical Test Procedures: Text and Methodology. International Conference on Harmonisation of Technical Requirements for Registration of Pharmaceuticals for Human Use: Geneva, Switzerland, October 1994/November 1996; ICH_Products/Guidelines/Quality/Q2_R1/ Step4/Q2_R1 Guideline.pdf. 2 Chapter <621> Chromatography. USP29 NF24. US Pharmacopeial Convention: Rockville, MD. 3 Shabir GA. Validation of High- Performance Liquid Chromatography Methods for Pharmaceutical Analysis: Understanding the Differences and Similarities Between Validation Requirements of the US Food and Drug Administration, the US Pharmacopeia, and the International Conference on Harmonization. J. Chromatography A 987, 2003: United Nations Office on Drugs and Crime. Guidance for the Validation of Analytical Methodology and Calibration of Equipment Used for Testing of Illicit Drugs in Seized Materials and Biological Specimens. United Nations: New York, NY 2009; scientific/validation_e.pdf. 5 Hahn GJ, Meeker, WQ. Statistical Intervals: A Guide for Practitioners. Wiley: New York, NY, Orchard T. Specification Setting: Setting Acceptance Criteria from Statistics of the Data. BioPharm Int. 19(11) 2006: 22 29; Article/Specification-Setting-Setting- Acceptance-Criteria-/ArticleStandard/Article/ detail/ Lansky D. Strip-Plot Designs, Mixed Models, and Comparisons Between Linear and Nonlinear Models for Microtitre Plate Bioassays in the Design and Analysis of Potency Assays. Dev. Biol. 107, 2002: CBER. Guidance for Industry: Potency Tests for Cellular and Gene Therapy Products (draft guidance). US Food and Drug Administration: Rockville, MD, 2011; www. fda.gov/downloads/biologicsbloodvaccines/ GuidanceComplianceRegulatoryInformation/ Guidances/CellularandGeneTherapy/ ucm pdf. 9 Deming S. Trending, Statistical Process Control and Continual Validation. NEPDA Newsletter 3(1) 2008: 1 3; Chapters/North-America-cont/New-England/ Chapter-News/January-2008-Newsletter.aspx. 10 Lightbown JW. Biological Standardization and the Analyst: A Review. J. Soc. Anal. Chem. 86, 1961: Chapter 5.3: Statistical Analysis of Results of Biological Assays and Tests. European Pharmacopoeia 8. European Directorate for the Quality of Medicines and Health Care: Strasbourg, France. 12 Chapter <1034> Analysis of Biological Assays. USP36 NF31. US Pharmacopeial Convention: Rockville, MD. 13 Hauck WW, et al. Assessing Parallelism Prior to Determining Relative Potency. PDA J. Pharmaceut. Sci. Technol. 59, 2005: Callahan JD, Sajjadi NC. Testing the Null Hypothesis for a Specified Difference: The Right Way to Test for Parallelism. BioProcessing J. March April Plikaytis BD, et al. Determination of Parallelism and Nonparallelism in Bioassay Dilution Curves. J. Clin. Microbiol. 32, 1994: BEBPA Bioassay Conference, September 2013, Basel, Switzerland. BioPharmaceutical Emerging Best Practices Association: Citrus Heights, CA; org. Corresponding author C. Jane Robinson is a principal scientist at the National Institute for Biological Standards and Control in South Mimms, UK (jane.robinson@nibsc. org). Michael Sadick is senior manager of biopharmaceutical characterization at Catalent Pharma Solutions in Kansas City, MO (mike.sadick@catalent.com). Stanley N. Deming is president of Statistical Designs in Houston, TX (standeming@statisticaldesigns. com). Sian Estdale is a principal scientist in large-molecule biopharm CMC at Covance Laboratories Limited in Harrogate, UK (Sian. Estdale@covance.com). Svetlana Bergelson is director of analytical development at Biogen Idec in Cambridge, MA (svetlana.bergelson@biogenidec.com). And Laureen Little is principal consultant at Quality Services in Seattle, WA; biotech@ ix.netcom.com For electronic or printed reprints, contact Rhonda Brown of Foster Printing Service, rhondab@fosterprinting.com, x194. Download personal-use only PDFs online at January (1) BioProcess International 41

Assay Acceptance Criteria for Multiwell-Plate Based Biological Potency Assays

Assay Acceptance Criteria for Multiwell-Plate Based Biological Potency Assays Biopharmaceutical Best Practices Association Assay Acceptance Criteria for Multiwell-Plate Based Biological Potency Assays C. Jane Robinson, Michael Sadick, Stanley N. Deming, Sian Estdale, Svetlana Bergelson,

More information

Potency is a critical quality

Potency is a critical quality B i o P r o c e s s TECHNICAL Assessing Similarity with Parallel- Line and Parallel-Curve Models Implementing the USP Development/Validation Approach to a Relative Potency Assay Erica Bortolotto, Rejane

More information

Validation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals.

Validation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals. Validation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals. 1 Analytical Method Validation: 1..1 Philosophy: Method validation is

More information

Perspectives on Method Validation: Importance of Adequate Method Validation

Perspectives on Method Validation: Importance of Adequate Method Validation Perspectives on Method Validation: Importance of Adequate Method Validation Heather Bridwell, Vikas Dhingra, Daniel Peckman, Jennifer Roark and Thomas Lehman The appropriate validation of analytical methods

More information

WHO GUIDELINE Stability testing of active pharmaceutical ingredients and finished pharmaceutical products

WHO GUIDELINE Stability testing of active pharmaceutical ingredients and finished pharmaceutical products WHO GUIDELINE Stability testing of active pharmaceutical ingredients and finished pharmaceutical products 1. Introduction 1.1 Objectives of these guidelines 1.2 Scope of these guidelines 1.3 General principles

More information

Improving Bioassay Performance by Optimizing Plate Layout and Data Analysis

Improving Bioassay Performance by Optimizing Plate Layout and Data Analysis CASSS Bioassays 2011 Improving Bioassay Performance by Optimizing Plate Layout and Data Analysis Wei Zhang, Ph.D. Biogen Idec 01NOV11 Presentation Outline Introduction Plate layout Data analysis Case studies.

More information

á1225ñ VALIDATION OF COMPENDIAL PROCEDURES

á1225ñ VALIDATION OF COMPENDIAL PROCEDURES 1640 á1224ñ Transfer of Analytical Procedures / General Information USP 39 THE ANALYTICAL PROCEDURE The procedure should be written with sufficient detail and explicit instructions, so that a trained analyst

More information

Estimation of Relative Potency from Bioassay Data that Include Values below the Limit of Quantitation

Estimation of Relative Potency from Bioassay Data that Include Values below the Limit of Quantitation Estimation of Relative Potency from Bioassay Data that Include Values below the Limit of Quantitation FRANCIS BURSA, KELLY J. FLEETWOOD, KARIE J. HIRST, AND ANN YELLOWLEES Experimental data may include

More information

VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY *)

VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY *) VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY *) Guideline Title Validation of Analytical Procedures: Methodology Legislative basis Directive 75/318/EEC as amended Date of first adoption December 1996

More information

Analysis of Bioassays: Personal reflections on the approach of the European Pharmacopoeia

Analysis of Bioassays: Personal reflections on the approach of the European Pharmacopoeia Analysis of Bioassays: Personal reflections on the approach of the European Pharmacopoeia Rose Gaines Das Consultant in Biostatistics, UK Formerly Head of Biostatistics, National Institute for Biological

More information

Changes to a Potency Bioassay for Biotechnology Products: a Regulatory Perspective Kathleen A. Clouse, Ph.D., Director Division of Monoclonal Antibodi

Changes to a Potency Bioassay for Biotechnology Products: a Regulatory Perspective Kathleen A. Clouse, Ph.D., Director Division of Monoclonal Antibodi Changes to a Potency Bioassay for Biotechnology Products: a Regulatory Perspective Kathleen A. Clouse, Ph.D., Director Division of Monoclonal Antibodies Office of Biotechnology Products Center for Drug

More information

VICH Topic GL2 (Validation: Methodology) GUIDELINE ON VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY

VICH Topic GL2 (Validation: Methodology) GUIDELINE ON VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY The European Agency for the Evaluation of Medicinal Products Veterinary Medicines Evaluation Unit CVMP/VICH/591/98-FINAL London, 10 December 1998 VICH Topic GL2 (Validation: Methodology) Step 7 Consensus

More information

Process Characterization Essentials Part I: Process

Process Characterization Essentials Part I: Process Process Characterization Essentials Part I: Process Understanding and Health Authorities Guidance Thomas A. Little Ph.D. 2/16/2017 President, Thomas A. Little Consulting, Bioassay Sciences 12401 N Wildflower

More information

STABILITY TESTING OF NEW DRUG SUBSTANCES AND PRODUCTS

STABILITY TESTING OF NEW DRUG SUBSTANCES AND PRODUCTS INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE STABILITY TESTING OF NEW DRUG SUBSTANCES AND PRODUCTS

More information

Understanding and accounting for product

Understanding and accounting for product Understanding and Modeling Product and Process Variation Variation understanding and modeling is a core component of modern drug development. Understanding and accounting for product and process variation

More information

Project leader: David Wood (1) ; Statisticians: Dawn Sands (1), Alan Heath (1) BRPs for inactivated adsorbed hepatitis A vaccines 1.

Project leader: David Wood (1) ; Statisticians: Dawn Sands (1), Alan Heath (1) BRPs for inactivated adsorbed hepatitis A vaccines 1. COLLABORATIVE STUDY FOR THE ESTABLISHME OF THREE PRODUCT SPECIFIC EUROPEAN PHARMACOPOEIA BIOLOGICAL REFERENCE PREPARATIO FOR INACTIVATED ADSORBED HEPATITIS A VACCINES Project leader: David Wood (1) ; Statisticians:

More information

How to Obtain Reproducible Quantitative ELISA Results

How to Obtain Reproducible Quantitative ELISA Results TM How to Obtain Reproducible Quantitative ELISA Results Diane Sasaki^ and Robert A. Mitchell* ^Oxford Biomedical Research Oxford MI 48371 *Department of Biochemistry & Molecular Biology, Wayne State University,

More information

Validation of a concentration assay using Biacore C

Validation of a concentration assay using Biacore C GE Healthcare Application Note 48 Biacore systems Validation of a concentration assay using Biacore C Guideline for development of a GxP - compliant concentration assay Support for informed decision-making

More information

Establishment and optimization of a potency assay for an effector mab using "ADCC Reporter Bioassay

Establishment and optimization of a potency assay for an effector mab using ADCC Reporter Bioassay Establishment and optimization of a potency assay for an effector mab using "ADCC Reporter Bioassay Mike Sadick, Ph.D. Catalent Pharma Solutions Senior Manager Biopharmaceutical Characterization Scenario/Outline

More information

How to overcome the challenges presented by bioassays

How to overcome the challenges presented by bioassays How to overcome the challenges presented by bioassays Tony Mire-Sluis Vice President, North America, Singapore, Abingdon, Contract and Product Quality Amgen Inc Why are Bioassays Considered so Difficult

More information

Bioassays are required for a

Bioassays are required for a B I O P R O C E S S TECHNICAL Bioassay Survey 006 007 Cell-Based Bioassays in the Biopharmaceutical Industry C. Jane Robinson, Laureen E. Little, and Hans-Joachim Wallny Bioassays are required for a variety

More information

Draft regional guidelines on stability testing of active substances and pharmaceutical products

Draft regional guidelines on stability testing of active substances and pharmaceutical products Regional Committee for the EM/RC53/12 Eastern Mediterranean August 2006 Fifty-third Session Original: Arabic Agenda item 17 Draft regional guidelines on stability testing of active substances and pharmaceutical

More information

Overview of Statistics used in QbD Throughout the Product Lifecycle

Overview of Statistics used in QbD Throughout the Product Lifecycle Overview of Statistics used in QbD Throughout the Product Lifecycle August 2014 The Windshire Group, LLC Comprehensive CMC Consulting Presentation format and purpose Method name What it is used for and/or

More information

STIMULI TO THE REVISION PROCESS. Bioassay Glossary

STIMULI TO THE REVISION PROCESS. Bioassay Glossary Vol. 32(4) [July Aug. 2006] of the USPC or the USP Council of Experts 1359 Bioassay Glossary Robert Singer, PDL Biopharma; David M. Lansky, Lansky Consulting; and Walter W. Hauck, Thomas Jefferson University

More information

Guidance for Industry

Guidance for Industry Guidance for Industry Submission of Abbreviated Reports and Synopses in Support of Marketing Applications U.S. Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation

More information

COMMITTEE FOR PROPRIETARY MEDICINAL PRODUCTS (CPMP)

COMMITTEE FOR PROPRIETARY MEDICINAL PRODUCTS (CPMP) The European Agency for the Evaluation of Medicinal Products Evaluation of Medicines for Human Use London, 17 December 2003 CPMP/QWP/122/02, rev 1 COMMITTEE FOR PROPRIETARY MEDICINAL PRODUCTS (CPMP) GUIDELINE

More information

Thank You to United States Pharmacopeia for Their Sponsorship and Assistance

Thank You to United States Pharmacopeia for Their Sponsorship and Assistance Use of Measurement Uncertainty in Testing a Drug Substance M.L. Jane Weitzel Quality Analysis Consultants mljweitzel@msn.com Eurachem/CITAC Workshop June 2011 Thank You to United States Pharmacopeia for

More information

For the quantitative detection of human IL6 in serum, plasma, cell culture supernatants and urine.

For the quantitative detection of human IL6 in serum, plasma, cell culture supernatants and urine. m andw da a For the quantitative detection of human IL6 in serum, plasma, cell culture supernatants and urine. general information Catalogue Number Product Name Species cross-reactivity Range (calibration

More information

ACE2 (Human) ELISA Kit

ACE2 (Human) ELISA Kit ACE2 (Human) ELISA Kit Catalog Number KA4223 96 assays Version: 10 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of the Assay...

More information

HUMAN MYELOPEROXIDASE (MPO) ELISA KIT

HUMAN MYELOPEROXIDASE (MPO) ELISA KIT HUMAN MYELOPEROXIDASE ELISA KIT Page 1 HUMAN MYELOPEROXIDASE (MPO) ELISA KIT FOR THE QUANTITATIVE DETERMINATION OF HUMAN MYOELOPEROXIDASE CONCENTRATIONS IN CELL CULTURE SUPERNATES, EDTA PLASMA AND SERUM.

More information

Human breast cancer susceptibility protein 2, BRCA-2 ELISA Kit

Human breast cancer susceptibility protein 2, BRCA-2 ELISA Kit Human breast cancer susceptibility protein 2, BRCA-2 ELISA Kit Catalog No: E1364h 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE

More information

Human D-Dimer, D2D ELISA Kit

Human D-Dimer, D2D ELISA Kit Human D-Dimer, D2D ELISA Kit Catalog No: E0506h 96 Tests Operating instruction www.eiaab.cn FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH ENTIRE PROCEDURE BEFORE

More information

EVALUATION FOR STABILITY DATA

EVALUATION FOR STABILITY DATA INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE EVALUATION FOR STABILITY DATA Q1E Recommended for

More information

Manual 058 Out of Specification Results Investigation

Manual 058 Out of Specification Results Investigation 1 Purpose The purpose of this Guideline is to provide guidance for the investigation and response to Out of Specification (OOS) laboratory test results. 2 Scope and Applicability This Guideline applies

More information

Rat Caspase 3 (Casp-3) ELISA Kit

Rat Caspase 3 (Casp-3) ELISA Kit Rat Caspase 3 (Casp-3) ELISA Kit Catalog No. CSB-E08857r (96T) This immunoassay kit allows for the in vitro quantitative determination of rat Casp-3 concentrations in serum, plasma and other biological

More information

colorimetric sandwich ELISA kit datasheet

colorimetric sandwich ELISA kit datasheet colorimetric sandwich ELISA kit datasheet For the quantitative detection of human IL6 in serum, plasma, cell culture supernatants and urine. general information Catalogue Number Product Name Species cross-reactivity

More information

Regulatory Challenges for the Licensure of Future Vaccines

Regulatory Challenges for the Licensure of Future Vaccines Regulatory Challenges for the Licensure of Future Vaccines Tong Wu, Ph.D. Bacterial & Combination Vaccine Division, BGTD, Health Canada June 26-29, 2018, Seoul, Korea, the Global Bio Conference 1 Disclaimer

More information

Biotherapeutic medicines generated by living

Biotherapeutic medicines generated by living Biotherapeutic medicines generated by living cells or organisms are larger and more complex than chemically synthesized medicines, and feature varied mechanisms of action (MOA s; Box, p. TK). As their

More information

Transfer of Methods Supporting Biologics and Vaccines

Transfer of Methods Supporting Biologics and Vaccines Transfer of Methods Supporting Biologics and Vaccines Timothy Schofield Arlenda Inc. tim.schofield@arlenda.com Presented at the BPD event on Case Studies in Tech Transfer esearch Triangle Park, NC, December

More information

Human placenta lactogen (HPL) ELISA Kit

Human placenta lactogen (HPL) ELISA Kit Human placenta lactogen (HPL) ELISA Kit Catalog Number. CSB-E09665h For the quantitative determination of human placenta lactogen (HPL) concentrations in serum, plasma. This package insert must be read

More information

HUMAN OSTEOPROTEGERIN (OPG) ELISA KIT

HUMAN OSTEOPROTEGERIN (OPG) ELISA KIT HUMAN OSTEOPROTEGERIN (OPG) ELISA KIT P a g e 1 HUMAN OSTEOPROTEGERIN (OPG) ELISA KIT For the quantitative determination of Human OPG concentrations in cell culture supernates, serum, and plasma. PURCHASE

More information

ADCC Reporter Bioassay: A Novel, Bioluminescent Cell-Based Assay for Quantifying Fc Effector Function of Antibodies

ADCC Reporter Bioassay: A Novel, Bioluminescent Cell-Based Assay for Quantifying Fc Effector Function of Antibodies ADCC Reporter Bioassay: A Novel, Bioluminescent Cell-Based Assay for Quantifying Fc Effector Function of Antibodies Richard Somberg, Ph.D. October 2012 Outline Introduction to ADCC Problem with classic

More information

A Parametric Tolerance Interval Test for Improved Control of Delivered Dose Uniformity of Orally Inhaled and Nasal Drug Products

A Parametric Tolerance Interval Test for Improved Control of Delivered Dose Uniformity of Orally Inhaled and Nasal Drug Products International Pharmaceutical Aerosol Consortium on Regulation and Science 1 2 3 4 5 6 7 8 9 1 11 12 13 14 15 16 17 18 19 2 21 22 23 24 25 26 27 28 29 3 31 32 33 34 35 36 37 38 39 4 41 A Parametric Tolerance

More information

GSI Equine IL-1β ELISA Kit- Plasma/Serum DataSheet

GSI Equine IL-1β ELISA Kit- Plasma/Serum DataSheet Interleukin-1 beta (IL-1β) also known as catabolin, is a member of the interleukin 1 cytokine family. IL-1β precursor is cleaved by caspase 1 (interleukin 1 beta convertase). Cytosolic thiol protease cleaves

More information

Human procollagen type III (HPCⅢ) ELISA Kit

Human procollagen type III (HPCⅢ) ELISA Kit Human procollagen type III (HPCⅢ) ELISA Kit Catalog Number. CSB-E11989h For the quantitative determination of human procollagen type III (HPCⅢ) concentrations in serum. This package insert must be read

More information

GUIDANCE NOTES ON ANALYTICAL METHOD VALIDATION

GUIDANCE NOTES ON ANALYTICAL METHOD VALIDATION ON ANALYTICAL METHOD VALIDATION HSA September 2004 Reproduction prohibited for commercial purposes. Reproduction for internal use is authorised, provided the source is acknowledged. MQA Dir: DISK1\GUIDE-MQA-012A-004.doc

More information

(PⅢP) ELISA Kit. Catalog No. CSB-E08096r (96T)

(PⅢP) ELISA Kit. Catalog No. CSB-E08096r (96T) Rat procollagen Ⅲ propeptide (PⅢP) ELISA Kit Catalog No. CSB-E08096r (96T) This immunoassay kit allows for the in vitro quantitative determination of rat PⅢP concentrations in cell culture supernates,

More information

Monkey apolipoprotein A1 (Apo-A1) ELISA Kit. For the quantitative determination of monkey apolipoprotein A1 (Apo-A1) concentrations in serum, plasma.

Monkey apolipoprotein A1 (Apo-A1) ELISA Kit. For the quantitative determination of monkey apolipoprotein A1 (Apo-A1) concentrations in serum, plasma. Monkey apolipoprotein A1 (Apo-A1) ELISA Kit Catalog Number. For the quantitative determination of monkey apolipoprotein A1 (Apo-A1) concentrations in serum, plasma. This package insert must be read in

More information

Rat Cluster of Differentiation 68 (CD68) ELISA kit. Catalog No. MBS (96 T)

Rat Cluster of Differentiation 68 (CD68) ELISA kit. Catalog No. MBS (96 T) Rat Cluster of Differentiation 68 (CD68) ELISA kit Catalog No. MBS705029 (96 T) This immunoassay kit allows for the in vitro quantitative determination of rat CD68 concentrations in serum, plasma and other

More information

Rat Prostate Specific Antigen (PSA) ELISA Kit

Rat Prostate Specific Antigen (PSA) ELISA Kit Rat Prostate Specific Antigen (PSA) ELISA Kit Catalog No. CSB-E08275r (96T) This immunoassay kit allows for the in vitro quantitative determination of rat PSA concentrations in cell culture supernates,

More information

Human Leukotriene E4 (LTE4) ELISA Kit

Human Leukotriene E4 (LTE4) ELISA Kit Human Leukotriene E4 (LTE4) ELISA Kit Catalog No. CSB-E05176h (96T) This immunoassay kit allows for the in vitro quantitative determination of human LTE4 concentrations in serum, plasma and other biological

More information

Rat α-melanocyte stimulating hormone (α-msh) ELISA Kit

Rat α-melanocyte stimulating hormone (α-msh) ELISA Kit Rat α-melanocyte stimulating hormone (α-msh) ELISA Kit For the quantitative determination of rat α-melanocyte stimulating hormone (α-msh) concentrations in serum, plasma, tissue homogenates. This package

More information

Fully Automated EIA System. True open flexibility.

Fully Automated EIA System. True open flexibility. True open flexibility www.triturus.com The immunoassay analyzer of choice: completely open, fully automated, multi-test and multi-batch. TRUE OPEN FLEXIBILITY What is Triturus? Triturus is a completely

More information

Mouse Luteinizing Hormone (LH) ELISA

Mouse Luteinizing Hormone (LH) ELISA Mouse Luteinizing Hormone (LH) ELISA For the quantitative determination of mouse LH in serum, plasma and tissue homogenates Cat. No. KU-222 For Research Use Only. Not for use in diagnostic procedures.

More information

Human myelin basic protein(mbp) antibody ELISA Kit

Human myelin basic protein(mbp) antibody ELISA Kit Human myelin basic protein(mbp) antibody ELISA Kit Catalog Number.... For the quantitative determination of human myelin basic protein (MBP) antibody concentrations in serum, cerebrospinal fluid (CSF).

More information

Human von Willebrand Factor cleaving protease, ADAMTS-13/vWF-cp. ELISA Kit

Human von Willebrand Factor cleaving protease, ADAMTS-13/vWF-cp. ELISA Kit Human von Willebrand Factor cleaving protease, ADAMTS-13/vWF-cp ELISA Kit Catalog No. CSB-E13487h (96 T) This immunoassay kit allows for the in vitro quantitative determination of human ADAMTS-13/vWF-cp

More information

Reference Standards for Bioassays

Reference Standards for Bioassays Reference Standards for Bioassays C Jane Robinson Principal Scientist Biotherapeutics National Institute for Biological Standards & Control, Health Protection Agency, UK Jane.Robinson@nibsc.hpa.org.uk

More information

Automating HPLC and GC Analytical Method Validation Fusion AE Software Program White Paper

Automating HPLC and GC Analytical Method Validation Fusion AE Software Program White Paper Automating HPLC and GC Analytical Method Validation Fusion AE Software Program White Paper Varian, Inc. 2700 Mitchell Drive Walnut Creek, California 94598, USA www.varianinc.com Introduction Chromatographic

More information

Analytical Methods Development and Validation

Analytical Methods Development and Validation Understanding and Implementing Efficient Analytical Methods Development and Validation Jay Breaux, Kevin Jones, and Pierre Boulas Analytical methods development and validation play important roles in the

More information

Bovine prolactin/luteotropic hormone (PRL/LTH) ELISA Kit

Bovine prolactin/luteotropic hormone (PRL/LTH) ELISA Kit Bovine prolactin/luteotropic hormone (PRL/LTH) ELISA Kit Catalog Number. MBS703224 For the quantitative determination of bovine prolactin/luteotropic hormone (PRL/LTH) concentrations in serum, plasma.

More information

FOOD AND DRUGS AUTHORITY

FOOD AND DRUGS AUTHORITY G H A N A FOOD AND DRUGS AUTHORITY GUIDELINES FOR STABILITY TESTING OF ACTIVE PHARMACEUTICAL INGREDIENTS AND FINISHED PHARMACEUTICAL PRODUCTS Document No: FDA/DRI/DER/GL-STP/2013/07 Date of First Adoption:

More information

Capillary Electrophoresis Compendial Applications

Capillary Electrophoresis Compendial Applications Capillary Electrophoresis Compendial Applications Anita Szajek, Ph.D. Principal Scientific Liaison, Biologics & Biotechnology 14 th Symposium on the Practical Applications for the Analysis of Proteins,

More information

Human nuclear matrix protein 22 (NMP-22) ELISA Kit. MyBioSource.com

Human nuclear matrix protein 22 (NMP-22) ELISA Kit. MyBioSource.com Human nuclear matrix protein 22 (NMP-22) ELISA Kit Catalog Number. For the quantitative determination of human nuclear matrix protein 22 (NMP-22) concentrations in serum, plasma and urine. This package

More information

Total Analytic Error From Concept to Application

Total Analytic Error From Concept to Application Página 1 de 5 Clinical Laboratory News Subscribe CLN Daily CLN Stat CLN Articles Total Analytic Error From Concept to Application Author: James O. Westgard, PhD, and Sten A. Westgard, MS // Date: SEP.1.2013

More information

Understanding the Use of Bioassays in a Biomanufacturing Facility. Bernardo E. Cáceres Associate Principal Scientist Biotech Development

Understanding the Use of Bioassays in a Biomanufacturing Facility. Bernardo E. Cáceres Associate Principal Scientist Biotech Development Understanding the Use of Bioassays in a Biomanufacturing Facility Bernardo E. Cáceres Associate Principal Scientist Biotech Development 1 Definitions What is a Bioassay? A bioassay is an analytical procedure

More information

The GCC Guidelines for Stability Testing of Drug Substances and Pharmaceutical Products EDITION TWO 1428 H 2007 G

The GCC Guidelines for Stability Testing of Drug Substances and Pharmaceutical Products EDITION TWO 1428 H 2007 G The GCC Guidelines for Stability Testing of Drug Substances and Pharmaceutical Products EDITION TWO 1428 H 2007 G (1) INTRODUCTION The following guideline defines the stability data package for a drug

More information

Hamster follicle stimulating hormone (FSH) ELISA Kit

Hamster follicle stimulating hormone (FSH) ELISA Kit Hamster follicle stimulating hormone (FSH) ELISA Kit Catalog Number. CSB-EQ027176HA For the quantitative determination of hamster follicle stimulating hormone (FSH) concentrations in serum, plasma. This

More information

Proposed New USP General Chapter: The Analytical Procedure Lifecycle 1220

Proposed New USP General Chapter: The Analytical Procedure Lifecycle 1220 Page 1 of 9 STIMULI TO THE REVISION PROCESS Stimuli articles do not necessarily reflect the policies of the USPC or the USP Council of Experts Proposed New USP General Chapter: The Analytical Procedure

More information

Revision of 30 April 2013 draft, 4 November 2013

Revision of 30 April 2013 draft, 4 November 2013 GUIDANCE DOCUMENT FOR SINGLE LABORATORY VALIDATION OF QUANTITATIVE ANALYTICAL METHODS USED IN SUPPORT OF PRE- AND POST-REGISTRATION DATA REQUIREMENTS FOR PLANT PROTECTION AND BIOCIDAL PRODUCTS INTRODUCTION

More information

Deer insulin-like growth factors 1 (IGF-1) ELISA Kit. This package insert must be read in its entirety before using this product.

Deer insulin-like growth factors 1 (IGF-1) ELISA Kit. This package insert must be read in its entirety before using this product. Deer insulin-like growth factors 1 (IGF-1) ELISA Kit Catalog Number. For the quantitative determination of deer insulin-like growth factors 1 (IGF-1) concentrations in serum, plasma, tissue homogenates.

More information

Replacing Analytical Methods for Release and Stability Testing CBER Perspective

Replacing Analytical Methods for Release and Stability Testing CBER Perspective Replacing Analytical Methods for Release and Stability Testing CBER Perspective Presentation at the CMC Strategy Forum January 27, 2014 Lokesh Bhattacharyya Chief, Lab of Analytical Chemistry and Blood

More information

Human connective tissue growth factor (CTGF) ELISA Kit. MyBioSource.com. This package insert must be read in its entirety before using this product.

Human connective tissue growth factor (CTGF) ELISA Kit. MyBioSource.com. This package insert must be read in its entirety before using this product. Human connective tissue growth factor (CTGF) ELISA Kit Catalog Number. For the quantitative determination of human connective tissue growth factor (CTGF) concentrations in serum, plasma, tissue homogenates.

More information

Analytical QbD Applied to Analytical Methods for Therapeutic Proteins Opportunities & Challenges

Analytical QbD Applied to Analytical Methods for Therapeutic Proteins Opportunities & Challenges Analytical QbD Applied to Analytical Methods for Therapeutic Proteins Opportunities & Challenges Annick GERVAIS, PhD Analytical Sciences Biologicals, UCB CASSS CMC Strategy Forum, Copenhagen, 05 May 2015

More information

Guide. recombinant DNA proteins. for the elaboration of monographs on synthetic peptides and. European Pharmacopoeia

Guide. recombinant DNA proteins. for the elaboration of monographs on synthetic peptides and. European Pharmacopoeia Guide for the elaboration of monographs on synthetic peptides and recombinant DNA proteins European Pharmacopoeia European Directorate for the Quality of Medicines & HealthCare Edition Council of Europe,

More information

Human Alpha-2-Macroglobulin ELISA

Human Alpha-2-Macroglobulin ELISA Human Alpha-2-Macroglobulin ELISA Catalog Number M046034 For the quantitative determination of alpha-2-macroglobulin in human serum and plasma samples. For research use only. This product insert must be

More information

Human BNP Immunoassay Kit

Human BNP Immunoassay Kit Antibody and Immunoassay Services Li Ka Shing Faculty of Medicine, The University of Hong Kong Human BNP Immunoassay Kit Catalogue Number: 31660 For the quantitative determination of human BNP concentrations

More information

HUMAN OSTEOPROTEGERIN (OPG) ELISA KIT

HUMAN OSTEOPROTEGERIN (OPG) ELISA KIT HUMAN OSTEOPROTEGERIN (OPG) ELISA KIT P a g e 1 HUMAN OSTEOPROTEGERIN (OPG) ELISA KIT PURCHASE INFORMATION: FOR THE QUANTITATIVE DETERMINATION OF HUMAN OPG CONCENTRATIONS IN CELL CULTURE SUPERNATES, SERUM,

More information

Human bovine serum albumin (BSA) ELISA Kit

Human bovine serum albumin (BSA) ELISA Kit Human bovine serum albumin (BSA) ELISA Kit Catalog Number. CSB-E08633h For the quantitative determination of human bovine serum albumin (BSA) concentrations in serum, plasma. This package insert must be

More information

JPMA 06, Dec., Case Studies of Established Conditions in Analytical Procedures. Biopharmaceutical Committee Technology Working Committee

JPMA 06, Dec., Case Studies of Established Conditions in Analytical Procedures. Biopharmaceutical Committee Technology Working Committee CMC Strategy Forum Japan 2016 Case Studies of Established Conditions in Analytical Procedures Biopharmaceutical Committee Technology Working Committee JPMA 06, Dec., 2016 Akira Nonoyama Santen Pharmaceutical

More information

Guidance for Industry Q3B(R2) Impurities in New Drug Products

Guidance for Industry Q3B(R2) Impurities in New Drug Products Guidance for Industry Q3B(R2) Impurities in New Drug Products U.S. Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation and Research (CDER) Center for Biologics

More information

GSI Sheep IL-1β ELISA Kit- 2 Plates DataSheet

GSI Sheep IL-1β ELISA Kit- 2 Plates DataSheet Interleukin-1 beta (IL-1β) also known as catabolin, is a member of the interleukin 1 cytokine family. IL-1β precursor is cleaved by caspase 1 (interleukin 1 beta convertase). Cytosolic thiol protease cleaves

More information

Guidance for Industry

Guidance for Industry Guidance for Industry Investigating Out-of-Specification (OOS) Test Results for Pharmaceutical Production U.S. Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation

More information

Automating HPLC and GC Analytical Method Validation Fusion AE Software Program White Paper

Automating HPLC and GC Analytical Method Validation Fusion AE Software Program White Paper Automating HPLC and GC Analytical Method Validation Fusion AE Software Program White Paper S-Matrix Corporation 1594 Myrtle Avenue Eureka, CA 95501, USA www.smatrix.com Introduction Chromatographic analytical

More information

Rat Free Thyroxine (FT4) ELISA Kit

Rat Free Thyroxine (FT4) ELISA Kit Rat Free Thyroxine (FT4) ELISA Kit Catalog No. CSB-E05079r (96 tests) This immunoassay kit allows for the in vitro quantitative determination of rat FT4 concentrations in serum, plasma and other biological

More information

Verification of Method XXXXXX

Verification of Method XXXXXX Reviewed By: Date: Technical Supervisor Reviewed By: Date: General Supervisor Reviewed By: Date: Quality Assurance Coordinator Approved By: Date: Director Page 1 of 11 Table of Contents Purpose and Principle:...

More information

PicoKine ELISA. Catalog number: EK1326

PicoKine ELISA. Catalog number: EK1326 PicoKine ELISA Catalog number: EK1326 For the quantitation of Human IGF2R concentrations in cell culture supernates, cell lysates, serum and plasma (heparin, EDTA). This package insert must be read in

More information

Statistical Evaluation Of Stability Data

Statistical Evaluation Of Stability Data Reprinted from FDA s website by #219 GUIDANCE FOR INDUSTRY Statistical Evaluation Of Stability Data VICH GL51 DRAFT GUIDANCE This guidance document is being distributed for comment purposes only Submit

More information

HUMAN SOLUBLE NEUROPILIN 2 (snrp2) ELISA KIT

HUMAN SOLUBLE NEUROPILIN 2 (snrp2) ELISA KIT Page 1 HUMAN SOLUBLE NEUROPILIN 2 (snrp2) ELISA KIT PURCHASE INFORMATION: FOR THE QUANTITATIVE DETERMINATION OF HUMAN SOLUBLE NEUROPILIN 2 CONCENTRATIONS IN PLASMA AND SERUM. ELISA NAME Catalog No. Lot

More information

colorimetric sandwich ELISA kit datasheet

colorimetric sandwich ELISA kit datasheet colorimetric sandwich ELISA kit datasheet For the quantitative detection of mouse Interleukin 6 (IL6) concentrations in serum, plasma and cell culture supernatants. general information Catalogue Number

More information

Sheep oxytocin (OT) ELISA Kit

Sheep oxytocin (OT) ELISA Kit Sheep oxytocin (OT) ELISA Kit Catalog Number. CSB-E13236Sh For the quantitative determination of sheep oxytocin (OT) concentrations in serum, plasma. This package insert must be read in its entirety before

More information

biosensis Neurotrophin 3 (NT3) Rapid TM ELISA Kit: Human, Rat and Mouse (2 Plates) Catalogue Number: BEK P

biosensis Neurotrophin 3 (NT3) Rapid TM ELISA Kit: Human, Rat and Mouse (2 Plates) Catalogue Number: BEK P biosensis Neurotrophin 3 (NT3) Rapid TM ELISA Kit: Human, Rat and Mouse (2 Plates) Catalogue Number: BEK-2221-2P For the quantitative determination of human NT3 in cell culture supernatants, serum and

More information

Porcine Transferrin Receptor(TFR) ELISA Kit

Porcine Transferrin Receptor(TFR) ELISA Kit Porcine Transferrin Receptor(TFR) ELISA Kit Catalog No. CSB-E13481p (96T) This immunoassay kit allows for the in vitro quantitative determination of porcine TFR concentrations in serum, plasma. Expiration

More information

Successful transfer of ligand binding assays between different laboratories

Successful transfer of ligand binding assays between different laboratories Successful transfer of ligand binding assays between different laboratories Debbie McManus Team Leader, Department of Biomarkers, Bioanalysis and Clinical Sciences 10 th EBF Open Symposium Hesperia Tower,

More information

Rabbit immunoglobulin M(IgM) ELISA Kit

Rabbit immunoglobulin M(IgM) ELISA Kit Rabbit immunoglobulin M(IgM) ELISA Kit Catalog Number. CSB-E06950Rb For the quantitative determination of rabbit immunoglobulin M (IgM) concentrations in serum, plasma. This package insert must be read

More information

Lifecycle Management of Commercially Approved QC Potency Assay for a Biotech Product. A Case Study

Lifecycle Management of Commercially Approved QC Potency Assay for a Biotech Product. A Case Study Lifecycle Management of Commercially Approved QC Potency Assay for a Biotech Product. A Case Study Heather Runes, Ph.D., MMTech Genentech, A Member of the Roche Group Wei-Meng Zhao and Dieter Schmalzing

More information

HUMAN TRANSFORMING GROWTH FACTOR BETA 1 (TGF- 2) ELISA KIT

HUMAN TRANSFORMING GROWTH FACTOR BETA 1 (TGF- 2) ELISA KIT P a g e 1 HUMAN TRANSFORMING GROWTH FACTOR BETA 1 (TGF- 2) ELISA KIT FOR THE QUANTITATIVE DETERMINATION OF HUMAN TGF- 2 CONCENTRATIONS IN CELL CULTURE SUPERNATES, SERUM, AND PLASMA. Purchase Information:

More information

Pig vasoactive intestinal peptide (VIP) ELISA Kit

Pig vasoactive intestinal peptide (VIP) ELISA Kit Pig vasoactive intestinal peptide (VIP) ELISA Kit MBS706159 For the quantitative determination of pig vasoactive intestinal peptide (VIP) concentrations in serum, plasma. This package insert must be read

More information

Mouse immunoglobulin M(IgM) ELISA Kit

Mouse immunoglobulin M(IgM) ELISA Kit Mouse immunoglobulin M(IgM) ELISA Kit Catalog Number. CSB-E07977m For the quantitative determination of mouse immunoglobulin M (IgM) concentrations in serum, plasma. This package insert must be read in

More information

BIOEQUIVALENCE TRIAL INFORMATION FORM (Medicines and Allied Substances Act [No. 3] of 2013 Part V Section 39)

BIOEQUIVALENCE TRIAL INFORMATION FORM (Medicines and Allied Substances Act [No. 3] of 2013 Part V Section 39) ZAMRA BTIF BIOEQUIVALENCE TRIAL INFORMATION FORM (Medicines and Allied Substances Act [No. 3] of 2013 Part V Section 39) The Guidelines on Bioequivalence Studies to be consulted in completing this form.

More information