1 A Sub-picogram Quantification Method for Desmopressin in Plasma using the SCIEX Triple Quad 6500 System A high-throughput method for detecting ultra-low levels (0.5 pg/ml) of a therapeutic peptide in human plasma using an SCIEX Triple Quad 6500 LC-MS/MS System and UHPLC Chromatography Rahul Baghla 1, Swati Guttikar 2, Dharmesh Patel 2, Abhishek Gandhi 2, Anoop Kumar 1, and Manoj Pillai 1 1 SCIEX, 121, Udyog Vihar, Phase IV, Gurgaon, Haryana, India 2 Veeda Clinical Research India, Ahmadabad, India Key challenges of desmopressin quantitation Impaired sensitivity in complex matrices Very low-level peptide detection (sub-pg/ml) can be suppressed by high background and competing ions in biological samples. Poor data quality Precision and accuracy can be compromised at low peptide levels, giving results below accepted bioanalytical standards. Key benefits of peptide quantitation on the SCIEX Triple Quad 6500 LC-MS/MS System Technology High sensitivity Very low level peptide detection in human plasma (at sub pg/ml concentrations) is enabled by IonDrive System Technology. Unique features of the SCIEX Triple Quad 6500 System for low-level peptide detection IonDrive Turbo V Source Increased ionization efficiency and heat transfer contribute to sensitivity enhancements, including improved signal-to-noise. IonDrive QJet Ion Guide Increased ion sampling improves method efficiency and ruggedness. IonDrive High Energy Detector Innovative detector technology boosts dynamic range and sensitivity. Excellent precision and accuracy at the LOQ level Data quality (for LOQ, LQC, MQC and HQC levels) met or exceeded USFDA bioanalytical method validation criteria. High throughput High sensitivity was achieved under high-flow conditions (0.750 ml/min), optimal for multi-sample analysis. Figure 2: Unique features of SCIEX Triple Quad 6500 System. Figure 1: SCIEX Triple Quad 6500 System. Introduction Low-level peptide detection has a number of applications in clinical studies and in the pharmaceutical discovery and development processes, highlighting the increasing relevance of sensitive and selective mass spectrometric platforms in the bioanalytical laboratory. Regulatory requirements demand intensive and rigorous quantitation of therapeutic peptides during pharmacokinetic, bioequivalence, and metabolic studies. In addition, drug discovery and development strategies seek to monitor and quantitate peptide biomarkers in complex biological samples, necessitating highly-selective separations of low concentration analytes from high background noise and prominent levels of competing ions. The SCIEX Triple Quad 6500 LC-MS/MS System, equipped with IonDrive System Technology for enhanced detector performance, has demonstrated particular strength in the detection of low-

2 level amounts of small molecules, and in this study, we extend the augmented signal-to-noise, broad dynamic range, and the efficient method development capacities of the SCIEX Triple Quad 6500 System to the detection of sub-picogram levels of a therapeutic peptide under high-throughput conditions. We have developed a reliable, fast, and sensitive method for the detection of a nine-amino-acid peptide, desmopressin (1-desamino-8-D-arginine, vasopressin), which is structurally similar to the hormone arginine vasopressin, but contains a deaminated first amino acid and dextro-arginine (rather than levo-) in the eighth position. Therapeutically, desmopressin reduces urine production, restricting water elimination from the kidneys by binding to the V2 receptors in renal-collecting ducts, thereby facilitating increased reabsorption. The longer half-life of desmopressin over vasopressin offers some therapeutic advantages, and typical doses of desmopressin to treat diabetes insipidus and bedwetting range between to 1.20 mg per day, resulting in very low plasma concentrations. In this bioanalytical study, we have established a sensitive and selective LC-MS/MS method for the quantitation of desmopressin in human plasma, detecting peptide levels as low as pg/ml with excellent accuracy and precision. This technique should facilitate additional mass spectrometric method development for accurate quantitation of a range of therapeutic peptides in biological matrices on the SCIEX Triple Quad 6500 System. Materials and Methods Sample Preparation Plasma samples (1000 µl) containing 2% desmopressin standard and 50µL internal standard were vortexed and spiked with 50µL of orthophosphoric acid (OPA). Samples were extracted on weak cation exchange cartridges conditioned with methanol followed by 100mM ammonium acetate solution. After loading, samples were washed in three steps: 1) 2% OPA:methanol (80:20 v/v); 2) 2% NaOH:Methanol (60:40 v/v); and 3) water:methanol (60:40 v/v). Analytes were eluted with 5% acetic acid in methanol, dried under nitrogen at 40 ºC, and reconstituted with 0.1% acetic acid (150 µl) prior to analysis by mass spectrometry. Chromatography LC System: GL Sciences LC 800 System Column: Agilent 300 Extend C 18 (150 x 2.1 mm, 3.5 μm) Column Temp.: 40 C Injection: 50 μl Flow Rate: ml/min Mobile phase: A) water, 0.1 % acetic acid B) acetonitrile, 0.1% acetic acid Gradient: Time/min A% B% Mass Spectrometry Analysis of desmopressin and desmopressin-d 5 required different mass spectrometry settings (Table 1). The MRM transition monitored for desmopressin was m/z 525.4/328.0 and 537.9/328.0 at a dwell time of 50 msec. Five replicate injections were performed at all concentrations. Data System: Interface: SCIEX Triple Quad 6500 System IonDrive Turbo V Source in positive ion mode Figure 3. Structure of desmopressin. Table 1. Compound-dependent parameters for desmopressin anddesmopressin-d 5 on the SCIEX Triple Quad 6500 System Instrument Parameter Desmopressin Desmopressin-d5 CUR TEM 600 C 600 C ISV GS GS CAD DP EP CE CXP Figure 4. Structure of internal standard, desmopressin-d 5.

3 Data processing All SCIEX Triple Quad 6500 System data was processed using MultiQuant Software.. The concentration curves were analyzed using a linear fit with a 1/x 2 weighting. Data acquired on the SCIEX Triple Quad 6500 System was processed using the quantitation tools within Analyst 1.6 Software. Results and Discussion Method analysis and data quality The desmopressin quantitative assay was validated by generating an internal standard curve using standards alone and standards spiked into human plasma. Left side pane of Figure 5 shows representative peaks for A) blank extract, B) plasma spiked with 0.5 pg/ml desmopressin and the right side pane of Figure 5 shows the MRM response from the internal standard. Standard concentrations varied from 0.5 to 100 pg/ml generating an LLOQ in plasma of 0.5 pg/ml resulting in a signal to noise ratio of 60.7 (Figure 6). Reproducibility of the assay was assessed by multiple technical replicates of the same sample (n = 6, Figure 7) on an LLOQ quality control sample of 0.5 pg/ml. The calibration curve for desmopressin in plasma shows excellent linearity over 2.5 orders of magnitude concentration range with an r value of >0.99 (Figure 8). Figure 6: High signal-to-noise ratio for desmopressin.. The signalto-noise ratio was calculated for desmopressin extracted from plasma at LLOQ level (0.500 pg/ml in plasma, S/N = 60.7). Figure 5. Desmopressin MRM signal (shown in left side ) panes for multiple concentrations and Desmopressin D5 MRM signal (Shown in right side panes). Figure 7: Desmopressin technical replicates. Chromatograms of six LLOQ quality control samples (0.502 pg/ml) for precision and accuracy calculations are shown (Table 2).

4 The data collected for a single calibration curve are presented in Table 2. Analyte retention time and internal standard peak retention times were consistent, with both eluting at approximately 2.6 min. The calculated concentration correlates well with the actual spikedanalyte concentration in plasma matrix with a percent accuracy of the standard curve very close to 100% for all concentrations of standard, and the quality control samples had a percent accuracy of 110%. Table 3 shows the individual statistics for three separate batch runs of desmopressin. Data from Table 2 are taken from Batch 3. Table 4 shows the mean values for the percent accuracy and % CV for three separate batch runs. For the LLOQ quality control, the mean accuracy was calculated to be 108% with a %CV of 10.5%. The percent recovery and plasma matrix effect were evaluated by comparing the peak areas for standard curve samples with and without plasma (Table 5). The mean percent recovery was calculated to be 93%. The recovery of the internal standard was calculated to be 78% (Table 6). Figure 8: Calibration curve of desmopressin in plasma from pg/ml to pg/ml. The method has shown excellent linearity over the concentration range with r = Table 2. Data quality analysis of desmopressin from human plasma samples Sample ID Sample Type Analyte Retention Time (min) Analyte Peak Area IS Retention Time (min) IS Peak Area Area Ratio Analyte Conc. (pg/ml) Calculated Conc. (pg/ml) % Accuracy AQM Unknown , N/A N/A BLANK 01 Blank #DIV/0! N/A #DIV/0! N/A BLANK+IS 01 Unknown ,418 0 N/A No Peak N/A STD A 01 Standard , , STD B 01 Standard , , STD C 01 Standard , , STD D 01 Standard , , STD E 01 Standard , , STD F 01 Standard , , STD G 01 Standard , , STD H 01 Standard , , LLOQ QC 01 Qual. Control , , LQC 01 Qual. Control , , MQC 01 Qual. Control , , HQC 01 Qual. Control , , LLOQ QC 02 Qual. Control , , LQC 02 Qual. Control 2.6 5, , MQC 02 Qual. Control , , HQC 02 Qual. Control , , LLOQ QC 03 Qual. Control , , LQC 03 Qual. Control , , MQC 03 Qual. Control , , HQC 03 Qual. Control , , LLOQ QC 04 Qual. Control , , LQC 04 Qual. Control , , MQC 04 Qual. Control , , HQC 04 Qual. Control , , LLOQ QC 05 Qual. Control , , LQC 05 Qual. Control , , MQC 05 Qual. Control , , HQC 05 Qual. Control , , LLOQ QC 06 Qual. Control , , LQC 06 Qual. Control , , MQC 06 Qual. Control , , HQC 06 Qual. Control , , Table 2. Full analysis of precision and accuracy measurements for desmopressin (batch 1 samples) in human plasma.

5 Table 3: Precision and accuracy for multi-batch desmopressin technical replicates DESMOPRESSIN Nominal Concentration (pg/ml) LLOQ QC LQC MQC HQC PA BATCH Mean S.D (+/-) C.V. (%) % Nominal N PA BATCH Mean S.D (+/-) C.V. (%) % Nominal N PA BATCH Mean S.D (+/-) C.V. (%) % Nominal N Table 3: Precision and accuracy calculations for individual batches of desmopressin samples. Table 4. Precision and accuracy of desmopressin technical replicates measured as a single group DESMOPRESSIN Nominal Concentration (pg/ml) LLOQ QC LQC MQC HQC Mean S.D (+/-) C.V. (%) % Nominal N Table 4: Mean precision and accuracy calculations for desmopressin for three batches of measurements from different days. Table 5. Recovery of desmopressin from human plasma, batch 03 LQC RESPONSE MQC RESPONSE HQC RESPONSE Sample ID Extracted Unextracted Extracted Unextracted Extracted Unextracted 01, 013 5,923 6, , , , ,315 02, 014 6,645 6, , , , ,138 03, 015 6,572 6, , , , ,730 04, 016 5,879 6, , , , ,201 05, 017 5,823 6, , , , ,383 06, 018 5,263 5, , , , ,647 Mean 6, , , , ,735.7 S.D , , , C.V N % Recovery Mean SD (+/-) CV(%) N 3 Table 5: Recovery of desmopressin from plasma at three different concentrations, LQC, MQC and HQC, was 92.72%.

6 Conclusions A highly sensitive and high-throughput bioanalytical method was developed and validated for the detection of ultra-low-levels of the therapeutic peptide, desmopressin, in human plasma on the SCIEX Triple Quad 6500 LC-MS/MS System. Method sensitivity for desmopressin detection was exceptional (0.5 pg/ml or 2.5 fg on column), and demonstrated high-reproducibility and cost effectiveness with good precision and accuracy. Analyte recovery is 92.7%, even under high-throughput conditions. Total run time for each sample was only 5 min, using a flow rate rapid enough for high-throughput analysis in the bioanalytical laboratory. References Friedman, F and Weiss JP. Desmopressin in the treatment of nocturia: clinical evidence and experience. Therapeutic Advances in Urology. 2013; 5(6): Neudert, L, Zaugg, M, Wood, S, Struwe, P. A high sensitivity dual solid phase extraction LC-MS/MS assay for the determination of the therapeutic peptide desmopressin in human plasma. Celerion white paper. Acknowledgements The authors are indebted to Dr.Venu Madhav, Chief Operating Officer (COO), Veeda Clinical Research, India, for his encouragement and support for the successful completion of the work. AB Sciex is doing business as SCIEX AB Sciex. For research use only. Not for use in diagnostic procedures. The trademarks mentioned herein are the property of the AB Sciex Pte. Ltd. or their respective owners. AB SCIEX is being used under license. RUO-MKT A Headquarters 500 Old Connecticut Path Framingham, MA USA Phone sciex.com International Sales For our office locations please call the division headquarters or refer to our website at sciex.com/offices

High Levels of Sensitivity and Robustness for Complex and Demanding Matrices SCIEX TRIPLE QUAD 5500 SYSTEM

High Levels of Sensitivity and Robustness for Complex and Demanding Matrices SCIEX TRIPLE QUAD 5500 SYSTEM High Levels of Sensitivity and Robustness for Complex and Demanding Matrices SCIEX TRIPLE QUAD 5500 SYSTEM The SCIEX Triple Quad 5500 System is designed to deliver a high level of sensitivity and robustness

More information

The LC-MS/MS Workhorse. SCIEX 4500 Series Mass Spectrometers

The LC-MS/MS Workhorse. SCIEX 4500 Series Mass Spectrometers The LC-MS/MS Workhorse SCIEX 4500 Series Mass Spectrometers The SCIEX 4500 Series takes the legendary API 4000 platform and intelligently re-engineers it to set a new benchmark for reliable quantitation.

More information

Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS

Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS Using the SCIEX M3 MicroLC system for Increased Sensitivity in Antibody Quantitation Remco van Soest and Lei Xiong SCIEX, Redwood City,

More information

Made for Routine Food, Environment, and Forensic Testing. The X500R QTOF System

Made for Routine Food, Environment, and Forensic Testing. The X500R QTOF System Made for Routine Food, Environment, and Forensic Testing. The X500R QTOF System RUO-MKT-03-2843-A We knew what you were looking for in a mass spec system, and we delivered on all fronts, starting with

More information

MultiQuant Software Version 3.0

MultiQuant Software Version 3.0 MultiQuant Software Version 3.0 Improving Data Quality and Processing Throughput with Better Peak Integration, Quantitative and Qualitative Compound Review for the Analysis of Food, Drinking Water, and

More information

Rapid Peptide Catabolite ID using the SCIEX Routine Biotransform Solution

Rapid Peptide Catabolite ID using the SCIEX Routine Biotransform Solution Rapid Peptide Catabolite ID using the SCIEX Routine Biotransform Solution Rapidly Identify Major Catabolites with the SCIEX X500R QTOF System and MetabolitePilot TM 2.0 Software Ian Moore and Jinal Patel

More information

AB SCIEX TripleTOF 5600 SYSTEM. High-resolution quant and qual

AB SCIEX TripleTOF 5600 SYSTEM. High-resolution quant and qual AB SCIEX TripleTOF 5600 SYSTEM High-resolution quant and qual AB SCIEX TripleTOF 5600 SYSTEM AB SCIEX is a global leader in the development of life science analytical technologies that help answer complex

More information

On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water

On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water application note On-line SPE-LC/MS/MS of Pesticides On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water API 3200 LC/MS/MS System Overview The feasibility of

More information

Quantitation of Microcystins and Nodularins in Water Samples Using LC-MS/MS

Quantitation of Microcystins and Nodularins in Water Samples Using LC-MS/MS Quantitation of Microcystins and Nodularins in Water Samples Using LC-MS/MS Craig Butt1, April Quinn-Paquet1, and Christopher Borton2 SCIEX (Framingham, MA) 2 SCIEX (Redwood City, CA) 1 Overview This application

More information

Pharmacokinetic Analysis of the mab Adalimumab by ELISA and Hybrid LBA/LC/MS: A Comparison Study

Pharmacokinetic Analysis of the mab Adalimumab by ELISA and Hybrid LBA/LC/MS: A Comparison Study Pharmacokinetic Analysis of the mab Adalimumab by ELISA and Hybrid LBA/LC/MS: A Comparison Study Featuring the SCIEX BioBA Solution Xi Qu 1, Shuyu Hou 1, Meghan McCann 1, Caroline Becker 1, Xun Wang 1,

More information

Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma

Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma Yun Wang Alelyunas, Mark D. Wrona, Jim Murphy, Angela Doneanu, Gregory

More information

Separation of Recombinant Human Erythropoietin (rhepo) using the European Pharmacopoeia Method on the PA 800 plus Pharmaceutical Analysis System

Separation of Recombinant Human Erythropoietin (rhepo) using the European Pharmacopoeia Method on the PA 800 plus Pharmaceutical Analysis System Separation of Recombinant Human Erythropoietin (rhepo) using the European Pharmacopoeia Method on the PA 800 plus Pharmaceutical Analysis System Marcia R Santos Global Tactical Marketing, Beckman Coulter

More information

Rapid Soft Spot Analysis using the SCIEX Routine Biotransform Solution

Rapid Soft Spot Analysis using the SCIEX Routine Biotransform Solution Rapid Soft Spot Analysis using the SCIEX Routine Biotransform Solution Rapidly Identify Top Metabolites with the SCIEX X500R QTOF System and MetabolitePilot TM 2.0 Software Shaokun Pang and Ian Moore SCIEX,

More information

Tyler Trent, Applications Sales Specialist; Teledyne Tekmar Page 1

Tyler Trent, Applications Sales Specialist; Teledyne Tekmar Page 1 Application Note Method Validation of the AutoMate-Q40 for Pesticide Residues in Grapes Tyler Trent, Applications Sales Specialist; Teledyne Tekmar Page 1 Abstract The QuEChERS (Quick-Easy-Cheap-Effective-Rugged-Safe)

More information

Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental

Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental Samples Michael S.Young, Jeremy C. Shia, Kim vantran, Kevin M. Jenkins and Masayo Yabo Waters Corporation 34 Maple Street,

More information

Above and beyond the extraordinary

Above and beyond the extraordinary AB SCIEX QTRAP 5500 System Above and beyond the extraordinary Above and beyond the competition speed and sensitivity that solve problems other systems cannot Powerful synergy results when the world s most

More information

MasterView Software. Turn data into answers reliable data processing made easy ONE TOUCH PRODUCTIVITY

MasterView Software. Turn data into answers reliable data processing made easy ONE TOUCH PRODUCTIVITY MasterView Software Turn data into answers reliable data processing made easy ONE TOUCH PRODUCTIVITY Mass spectrometers especially high-resolution, accurate mass instruments produce a significant amount

More information

Biologics Bioanalysis

Biologics Bioanalysis Biologics Bioanalysis GUIDE TO INNOVATION Biologics Bioanalysis Guide to Innovation Pharmaceutical companies have leveraged advancements in basic science perhaps more than any other industry. With the

More information

Monitoring for Microcystins in Raw Water Supply Reservoirs Using the Agilent 6410 Triple Quadrupole LC/MS

Monitoring for Microcystins in Raw Water Supply Reservoirs Using the Agilent 6410 Triple Quadrupole LC/MS Monitoring for Microcystins in Raw Water Supply Reservoirs Using the Agilent 641 Triple Quadrupole LC/MS Application Note Environmental Author Toni Hall Wessex Water Bath, UK Abstract A method for the

More information

DISCOVERY AND VALIDATION OF TARGETS AND BIOMARKERS BY MASS SPECTROMETRY-BASED PROTEOMICS. September, 2011

DISCOVERY AND VALIDATION OF TARGETS AND BIOMARKERS BY MASS SPECTROMETRY-BASED PROTEOMICS. September, 2011 DISCOVERY AND VALIDATION OF TARGETS AND BIOMARKERS BY MASS SPECTROMETRY-BASED PROTEOMICS September, 2011 1 CAPRION PROTEOMICS Leading proteomics-based service provider - Biomarker and target discovery

More information

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column Application Note Bio-Pharmaceutical Authors M. Sundaram Palaniswamy and Andrew Coffey Agilent Technologies,

More information

IgG Purity/Heterogeneity and SDS-MW Assays with High- Speed Separation Method and High Throughput Tray Setup

IgG Purity/Heterogeneity and SDS-MW Assays with High- Speed Separation Method and High Throughput Tray Setup IgG Purity/Heterogeneity and SDS-MW Assays with High- Speed Separation Method and High Throughput Tray Setup High Throughput Methods to Maximize the Use of the PA 800 Plus system Jose-Luis Gallegos-Perez

More information

Application Note. Abstract. Authors. Introduction. Environmental

Application Note. Abstract. Authors. Introduction. Environmental EPA Method 538: Determination of Selected Organic Contaminants in Drinking Water by Direct Aqueous Injection with the Agilent 646 Triple Quadrupole LC/MS System Application Note Environmental Authors Imma

More information

Thermo Scientific Mass Spectrometric Immunoassay (MSIA) Pipette Tips. Next generation immunoaffinity. Robust quantitative platform

Thermo Scientific Mass Spectrometric Immunoassay (MSIA) Pipette Tips. Next generation immunoaffinity. Robust quantitative platform Thermo Scientific Mass Spectrometric Immunoassay (MSIA) Pipette Tips Next generation immunoaffinity Robust quantitative platform Immunoaffinity sample preparation Thermo Scientific Mass Spectrometric Immunoassay

More information

Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016

Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016 Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016 1.0 Method background: This method employs full evaporation technique headspace

More information

Enabling Systems Biology Driven Proteome Wide Quantitation of Mycobacterium Tuberculosis

Enabling Systems Biology Driven Proteome Wide Quantitation of Mycobacterium Tuberculosis Enabling Systems Biology Driven Proteome Wide Quantitation of Mycobacterium Tuberculosis SWATH Acquisition on the TripleTOF 5600+ System Samuel L. Bader, Robert L. Moritz Institute of Systems Biology,

More information

Puzzled About LCMS? Sample. Sensitivity in Mass Spec Analysis. Prep. Adapting LC-UV. Optimizing and Maintaining Your Mass Spec LCMS

Puzzled About LCMS? Sample. Sensitivity in Mass Spec Analysis. Prep. Adapting LC-UV. Optimizing and Maintaining Your Mass Spec LCMS Puzzled About LCMS? Sensitivity in Mass Spec Analysis Sample Prep Adapting LC-UV Optimizing and Maintaining Your Mass Spec To LCMS Paul Altiero Alex Ucci Applications Phone Support Chemistry and Supplies

More information

New Approaches to Quantitative Proteomics Analysis

New Approaches to Quantitative Proteomics Analysis New Approaches to Quantitative Proteomics Analysis Chris Hodgkins, Market Development Manager, SCIEX ANZ 2 nd November, 2017 Who is SCIEX? Founded by Dr. Barry French & others: University of Toronto Introduced

More information

ichemistry SOLUTIONS Integrated chemistries to boost mass spectrometry workflows

ichemistry SOLUTIONS Integrated chemistries to boost mass spectrometry workflows ichemistry SOLUTIONS Integrated chemistries to boost mass spectrometry workflows ichemistry SOLUTIONS ichemistry solutions ichemistry solutions are the world s only reagents and consumables that are customdesigned

More information

Quantitation of cyanotoxins in drinking water according to EPA 544 guidelines

Quantitation of cyanotoxins in drinking water according to EPA 544 guidelines APPLICATION NOTE 64968 Quantitation of cyanotoxins in drinking water according to EPA 544 guidelines Authors Ali Haghani, 1 Andy Eaton, 1 Neloni Wijeratne, Claudia Martins 1 Eurofins Eaton Analytical,

More information

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions Screening and Confirmation Chemical Analysis Solution Unit November, 2009 Page 1 Agilent s New 6540 Ultra High

More information

Key Words Q Exactive Focus, SIEVE Software, Biomarker, Discovery, Metabolomics

Key Words Q Exactive Focus, SIEVE Software, Biomarker, Discovery, Metabolomics Metabolomic Profiling in Drug Discovery: Understanding the Factors that Influence a Metabolomics Study and Strategies to Reduce Biochemical and Chemical Noise Mark Sanders 1, Serhiy Hnatyshyn 2, Don Robertson

More information

[ product solution ] Waters Oasis µ Elution PlateS. Patented Innovation. Elution volume as low as 25 μl. No evaporation and reconstitution

[ product solution ] Waters Oasis µ Elution PlateS. Patented Innovation. Elution volume as low as 25 μl. No evaporation and reconstitution [ product solution ] Waters µ Elution PlateS Elution volume as low as 25 μl No evaporation and reconstitution Ideal for small sample volumes Up to a 15x increase in concentration Patented Innovation Now

More information

Rapid Extraction of Therapeutic Oligonucleotides from Primary Tissues for LC/ MS Analysis Using Clarity OTX, an Oligonucleotide Extraction Cartridge

Rapid Extraction of Therapeutic Oligonucleotides from Primary Tissues for LC/ MS Analysis Using Clarity OTX, an Oligonucleotide Extraction Cartridge Rapid Extraction of Therapeutic Oligonucleotides from Primary Tissues for LC/ MS Analysis Using Clarity OTX, an Oligonucleotide Extraction Cartridge G. Scott*, H. Gaus #, B. Rivera*, and M. McGinley* *Phenomenex,

More information

Agilent 6430 Triple Quadrupole LC/MS System

Agilent 6430 Triple Quadrupole LC/MS System Agilent 6430 Triple Quadrupole LC/MS System Ideal Quantitative LC/MS for UHPLC with Dynamic MRM and Fast Polarity Switching, plus Unsurpassed Sensitivity with Chip LC Summary The Agilent 6430 Triple Quadrupole

More information

Detecting Challenging Post Translational Modifications (PTMs) using CESI-MS

Detecting Challenging Post Translational Modifications (PTMs) using CESI-MS Detecting Challenging Post Translational Modifications (PTMs) using CESI-MS Sensitive workflow for the detection of PTMs in biological samples from small sample volumes Klaus Faserl, 1 Bettina Sarg, 1

More information

Converting a Liquid-Liquid Extraction Method for Vitamin D to a 96-Well Plate Supported Liquid Extraction Format

Converting a Liquid-Liquid Extraction Method for Vitamin D to a 96-Well Plate Supported Liquid Extraction Format Case Study: PPS367 Converting a Liquid-Liquid Extraction Method for Vitamin D to a 96-Well Plate Page 1 Converting a Liquid-Liquid Extraction Method for Vitamin D to a 96-Well Plate Supported Liquid Extraction

More information

HPLC method development and validation for the estimation of axitinib in rabbit plasma

HPLC method development and validation for the estimation of axitinib in rabbit plasma Brazilian Journal of Pharmaceutical Sciences http://dx.doi.org/10.1590/s2175-97902017000300012 Article HPLC method development and validation for the estimation of axitinib in rabbit plasma Achanta Suneetha

More information

Application Note. Authors. Abstract

Application Note. Authors. Abstract Automated, High Precision Tryptic Digestion and SISCAPA-MS Quantification of Human Plasma Proteins Using the Agilent Bravo Automated Liquid Handling Platform Application Note Authors Morteza Razavi, N.

More information

Agilent 6490 Triple Quadrupole LC/MS System REACH ULTIMATE SENSITIVITY WITH BREAKTHROUGH IFUNNEL TECHNOLOGY

Agilent 6490 Triple Quadrupole LC/MS System REACH ULTIMATE SENSITIVITY WITH BREAKTHROUGH IFUNNEL TECHNOLOGY Agilent 6490 Triple Quadrupole LC/MS System REACH ULTIMATE SENSITIVITY WITH BREAKTHROUGH IFUNNEL TECHNOLOGY AGILENT 6490 TRIPLE QUADRUPOLE LC/MS SYSTEM WITH IFUNNEL TECHNOLOGY CLEARLY BETTER SENSITIVITY

More information

BIOANALYTICAL METHOD DEVELOPMENT AND ITS VALIDATION

BIOANALYTICAL METHOD DEVELOPMENT AND ITS VALIDATION 321 P a g e International Standard Serial Number (ISSN): 2319-8141 International Journal of Universal Pharmacy and Bio Sciences 3(4): July-August 2014 INTERNATIONAL JOURNAL OF UNIVERSAL PHARMACY AND BIO

More information

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube Application Note Environmental Authors Don Noot, Ralph Hindle Vogon Laboratory Services Cochrane, AB

More information

Analysis of Illegal Dyes in Food Matrices Using Automated Online Sample Preparation with Liquid Chromatography-Mass Spectrometry

Analysis of Illegal Dyes in Food Matrices Using Automated Online Sample Preparation with Liquid Chromatography-Mass Spectrometry Analysis of Illegal Dyes in Food Matrices Using Automated Online Sample Preparation with Liquid Chromatography-Mass Spectrometry Yang Shi, Catherine Lafontaine, and François A. Espourteille Thermo Fisher

More information

Liquid Chromatograph Liquid Chromatograph Mass Spectrometer. CL Series. for In Vitro Diagnostic Use

Liquid Chromatograph Liquid Chromatograph Mass Spectrometer. CL Series. for In Vitro Diagnostic Use Liquid Chromatograph Liquid Chromatograph Mass Spectrometer CL Series for In Vitro Diagnostic Use Consistency. Performance. Reliability. Shimadzu s CL Series LC, LC-MS, and LC-MS/MS for in vitro diagnostic

More information

quantitate host cell DNA with high sensitivity and throughput resdnaseq Host Cell DNA Quantitation Systems: CHO, E.

quantitate host cell DNA with high sensitivity and throughput resdnaseq Host Cell DNA Quantitation Systems: CHO, E. quantitate host cell DNA with high sensitivity and throughput resdnaseq Host Cell DNA Quantitation Systems: CHO, E. coli, Vero, NS0 Precision counts Integrated real-time qpcr systems for quantitation of

More information

Bioanalytical method validation: An updated review

Bioanalytical method validation: An updated review Review Article www.phmethods.org Bioanalytical method validation: An updated review Abstract The development of sound bioanalytical method(s) is of paramount importance during the process of drug discovery

More information

Featuring Analyst software under Windows NT for enhanced performance and ease of use. API 150EX. LC/MS System

Featuring Analyst software under Windows NT for enhanced performance and ease of use. API 150EX. LC/MS System Featuring Analyst software under Windows NT for enhanced performance and ease of use. API 15EX LC/MS System compact, rugged, The API 15EX LC/MS System is the most rugged and reliable single quadrupole

More information

A Robust and Generic Method for Quantification of Monoclonal Antibody Therapeutics in Biological Matrices

A Robust and Generic Method for Quantification of Monoclonal Antibody Therapeutics in Biological Matrices A Robust and Generic Method for Quantification of Monoclonal Antibody Therapeutics in Biological Matrices Biologics analysis simplified with the BioBA kit, a hybrid strategy for therapeutic antibody assays

More information

A RAPID METHOD FOR THE DETERMINATION OF CHLORAMPHENICOL RESIDUES IN BLACK TIGER SHRIMP

A RAPID METHOD FOR THE DETERMINATION OF CHLORAMPHENICOL RESIDUES IN BLACK TIGER SHRIMP INTRODUCTION The global production of fish, as a food item, has greatly increased over the last few decades and is now at the highest levels on record. Currently, more than 5 of world protein consumption

More information

ISSN India; g,secunderabad. Abstractt. a flow rate. of 1ml/min. di hydrogen. which acts. and chronic. including minimize (5) Figure

ISSN India; g,secunderabad. Abstractt. a flow rate. of 1ml/min. di hydrogen. which acts. and chronic. including minimize (5) Figure B.Lakshmi et al SPJTS,2013,1(1),015-023 RP-HPLC METHOD FOR THE QUANTIFICATION OF ROFLUMILAST IN FORMULATIONS ISSN 2321-4597 B..Lakshmi 1, Prof. T.V.Reddy 2 1.Kallam Haranadha Reddy Institute of Technology,NH-5,

More information

Development of a Fast and Cost-Effective Multi-Residue Method to Determine Pesticides in Tobacco by LC/MS/MS

Development of a Fast and Cost-Effective Multi-Residue Method to Determine Pesticides in Tobacco by LC/MS/MS TECHNICAL NOTE Pesticides in Tobacco Development of a Fast and Cost-Effective Multi-Residue Method to Determine Pesticides in Tobacco by LC/MS/MS API 4000 LC/MS/MS System Overview This application note

More information

N. Tamilselvi *, Dona Sara Kurian. Department of Pharmaceutical Analysis. KMCH college of pharmacy Coimbatore

N. Tamilselvi *, Dona Sara Kurian. Department of Pharmaceutical Analysis. KMCH college of pharmacy Coimbatore BIOANALYTICAL METHOD DEVELOPMENT AND VALIDATION OF PIRFENIDONE BY RPHPLC METHOD AND ITS APPLICATION TO THE DETERMINATION OF DRUG FOOD INTERACTION STUDY IN WISTER RATS ABSTRACT N. Tamilselvi *, Dona Sara

More information

N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF

N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF Application Note Authors Xianming Liu, Wei Zhang, Yi Du, Sheng Yin, Hong Que, and Weichang Zhou WuXi AppTec iopharmaceuticals

More information

Extraction and Analysis of Neonicotinoid Pesticides using QuEChERS Approach

Extraction and Analysis of Neonicotinoid Pesticides using QuEChERS Approach Extraction and Analysis of Neonicotinoid Pesticides using QuEChERS Approach Katherine K. Stenerson 1, Olga Shimelis 1, and Christine Dumas 2 1 Supelco, Div. of Sigma-Aldrich Bellefonte, PA 16823 USA 2

More information

Validation of a concentration assay using Biacore C

Validation of a concentration assay using Biacore C GE Healthcare Application Note 48 Biacore systems Validation of a concentration assay using Biacore C Guideline for development of a GxP - compliant concentration assay Support for informed decision-making

More information

Preparative Purification of Corticosteroids by HPLC; Scalability and Loadability Using Agilent Prep C18 HPLC Columns Application

Preparative Purification of Corticosteroids by HPLC; Scalability and Loadability Using Agilent Prep C18 HPLC Columns Application Preparative Purification of Corticosteroids by PLC; Scalability and Loadability Using Agilent Prep C18 PLC Columns Application Pharmaceuticals Authors Cliff Woodward and Ronald Majors Agilent Technologies,

More information

Application Note LCMS-87 Automated Acquisition and Analysis of Data for Monitoring Protein Conjugation by LC-MS Using BioPharma Compass

Application Note LCMS-87 Automated Acquisition and Analysis of Data for Monitoring Protein Conjugation by LC-MS Using BioPharma Compass Application Note LCMS-87 Automated Acquisition and Analysis of Data for Monitoring Protein Conjugation by LC-MS Using BioPharma Compass Abstract Here we describe the development and use of an automated

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment QTRAP 5500

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment QTRAP 5500 www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment QTRAP 5500 Powerful synergy results when the world s most sensitive triple

More information

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography Carl Fisher, Monika Verma, and Pranathi Perati; Thermo Fisher Scientific,

More information

Genotoxicity is the property of a compound

Genotoxicity is the property of a compound Impurities Analysis in Pharmaceuticals: Genotoxicity is the property of a compound known to have irreversible effects on the structure and functionality of the DNA in cells and cause DNA loss, DNA replication

More information

Agilent 6400 Series Triple Quadrupole LC/MS. Breakthrough sensitivity now routine for your lab

Agilent 6400 Series Triple Quadrupole LC/MS. Breakthrough sensitivity now routine for your lab Agilent 6400 Series Triple Quadrupole LC/MS Breakthrough sensitivity now routine for your lab Agilent 6400 Series Triple Quadrupole LC/MS Unprecedented sensitivity to improve your results every day Whether

More information

Investigating Biological Variation of Liver Enzymes in Human Hepatocytes

Investigating Biological Variation of Liver Enzymes in Human Hepatocytes Investigating Biological Variation of Liver Enzymes in Human Hepatocytes SWATH Acquisition on the TripleTOF Systems Xu Wang 1, Hui Zhang 2, Christie Hunter 1 1 SCIEX, USA, 2 Pfizer, USA MS/MS ALL with

More information

Simultaneous in vivo Quantification and Metabolite Identification of Plasma Samples Using High Resolution QTof and Routine MS E Data Analysis

Simultaneous in vivo Quantification and Metabolite Identification of Plasma Samples Using High Resolution QTof and Routine MS E Data Analysis Simultaneous in vivo Quantification and Metabolite Identification of Plasma Samples Using High Resolution QTof and Routine MS E Data Analysis Mark Wrona, 1 Paul Rainville, 1 Eric Langlois, 2 Nigel Ewing,

More information

Pharmaceutical LC/MS Solutions from Agilent Technologies

Pharmaceutical LC/MS Solutions from Agilent Technologies Pharmaceutical LC/MS Solutions from Agilent Technologies Application Compendium Overview LC/MS plays a key role in the drug discovery and drug development process. Since the introduction of electrospray

More information

Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass

Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass Application Note: 529 Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass Xiang He, Marta Kozak; Thermo Fisher Scientific, San Jose,

More information

A Fast and Effective Approach for Running EPA Method 539: Determination of Hormones in Drinking Water using SPE and LC/MS/MS

A Fast and Effective Approach for Running EPA Method 539: Determination of Hormones in Drinking Water using SPE and LC/MS/MS A Fast and Effective Approach for Running EPA Method 539: Determination of Hormones in Drinking Water using SPE and LC/MS/MS Xianrong (Jenny) Wei 1, Matthew Trass 1, Ali Haghani 2, Andy Eaton 2 1 Phenomenex,

More information

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus Robert Birdsall, Thomas Wheat, and Weibin Chen Waters Corporation,

More information

Application Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany Quantification of trace-level herbicides in drinking water by online enrichment with the Agilent Infinity Series Online-SPE Solution and Triple Quadrupole MS Detection Application Note Environmental Author

More information

A High-Resolution, Accurate-Mass Approach for Ultra-High Throughput Screening Plasma Protein Binding

A High-Resolution, Accurate-Mass Approach for Ultra-High Throughput Screening Plasma Protein Binding A High-Resolution, Accurate-Mass Approach for Ultra-High Throughput Screening Plasma Protein Binding Keeley Murphy, 1 Patrick Bennett, 1 Francois Espourteille, 2 Maciej Bromirski 3 1 Thermo Fisher Scientific,

More information

Ligand Binding Assays: Summary and Consensus from the Bioanalytical Workshop (CC V)

Ligand Binding Assays: Summary and Consensus from the Bioanalytical Workshop (CC V) Ligand Binding Assays: Summary and Consensus from the Bioanalytical Workshop (CC V) Binodh DeSilva Executive Director Immunochemistry and Biomarker Development Bristol-Myers Squibb Contributors Lauren

More information

Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module

Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module Application Note Environmental Authors Shane Snyder Department of

More information

Using HRAM Survey Analysis Combined with Rapid MS2 Data to Develop a Fragmentation Based Detection Workflow for Structure ID Acquisition

Using HRAM Survey Analysis Combined with Rapid MS2 Data to Develop a Fragmentation Based Detection Workflow for Structure ID Acquisition Using HRAM Survey Analysis Combined with Rapid MS Data to Develop a Fragmentation Based Detection Workflow for Structure ID Acquisition Tim Stratton, 1 Caroline Ding, 1 Christoph Henrich, Hans Grensemann,

More information

Investigation of Solid Phase Microextraction as an Alternative to Dried Blood Spot

Investigation of Solid Phase Microextraction as an Alternative to Dried Blood Spot Investigation of Solid Phase Microextraction as an Alternative to Dried Blood Spot Craig Aurand and Robert Shirey Supelco, Div. of Sigma-Aldrich Bellefonte, PA 16823 USA T414068 sigma-aldrich.com/analytical

More information

Method Translation in Liquid Chromatography

Method Translation in Liquid Chromatography Method Translation in Liquid Chromatography Technical Overview Abstract Ronald E. Majors Agilent Technologies, Inc. 2850 Centerville Rd Wilmington, DE 19808 USA With the recent emphasis on high performance

More information

Current Best Practices in Commercial Kit Evaluation and Validation for Biomarker Assays

Current Best Practices in Commercial Kit Evaluation and Validation for Biomarker Assays Current Best Practices in Commercial Kit Evaluation and Validation for Biomarker Assays 10th Workshop on Recent Issues in Bioanalysis, 2016 Paul Rhyne, Ph.D. - Director of Immunoassay Services Copyright

More information

TurboMass GC/MS Software

TurboMass GC/MS Software TurboMass GC/MS Software GAS CHROMATOGRAPHY / MASS SPECTROMETRY P R O D U C T N O T E Today s demanding laboratory requires software that is easy to learn yet offers sophisticated instrument control, robust

More information

Available Online through (or) IJPBS Volume 2 Issue 3 JULY-SEPT Research Article Pharmaceutical Sciences

Available Online through  (or)  IJPBS Volume 2 Issue 3 JULY-SEPT Research Article Pharmaceutical Sciences Page1 Research Article Pharmaceutical Sciences RP-HPLC DETERMINATION OF RELATED SUBSTANCES OF TAPENTADOL IN BULK AND PHARMACEUTICAL DOSAGE FORM EDIGA SASI KIRAN GOUD* 1, V.KRISHNA REDDY 2 * 1,2 Department

More information

Experimental. Instrumentation

Experimental. Instrumentation Efficient Identification of Deoxynivalenol in the Global Wheat Food Chain Testing Unprocessed, Processed, and Consumer Wheat Products using Automated Solid Phase Extraction Cleanup and Quick High Performance

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis S(-)-Nicotine, Primary Standard N-008 Page 1 of 7 Catalog Number: N-008 H Lot: Expiration: October 2019 N Description: S(-)-Nicotine in Methanol.

More information

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider Combining small-scale purification and analysis of monoclonal antibodies on one instrument Protein purification with high-volume injection using the Agilent 126 Infinity Bio-inert Quaternary LC System

More information

Development of an Immunoprecipitation and LC-MS/MS based Method for Quantifying the 105 kda Recombinant Protein SXN in Plasma.

Development of an Immunoprecipitation and LC-MS/MS based Method for Quantifying the 105 kda Recombinant Protein SXN in Plasma. Development of an Immunoprecipitation and LC-MS/MS based Method for Quantifying the 105 kda Recombinant Protein SXN101959 in Plasma. Richard Kay Principal Scientist, Bioanalytical Sciences, Quotient Bioresearch

More information

Immunoassays and Mass Spectrometry: Transitioning Technology

Immunoassays and Mass Spectrometry: Transitioning Technology TECHNICAL BRIEF Immunoassays and Mass Spectrometry: Transitioning Technology This technical brief presents data obtained when measuring eicosanoids by immunoassay and mass spectrometry. It briefly considers

More information

Quantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy

Quantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy Advances in Bioscience and Biotechnology, 2010, 1, 367-371 doi:10.4236/abb.2010.15049 Published Online December 2010 (http://www.scirp.org/journal/abb/). Quantitative determination of residual 2-(2-chloroethoxy)

More information

Strategies for Developing Sensitive and Automated LC-MS/MS Assays of a Pharmaceutical Compound and Its Metabolite from Whole Blood Matrix

Strategies for Developing Sensitive and Automated LC-MS/MS Assays of a Pharmaceutical Compound and Its Metabolite from Whole Blood Matrix Pharmaceutics 2010, 2, 159 170; doi:10.3390/pharmaceutics2020159 Article OPEN ACCESS pharmaceutics ISSN 1999-4923 www.mdpi.com/journal/pharmaceutics Strategies for Developing Sensitive and Automated LC-MS/MS

More information

Technical Note CYP Inhibition Assays: Versatility in the ADDA High Speed Dual Arm Autosampling System

Technical Note CYP Inhibition Assays: Versatility in the ADDA High Speed Dual Arm Autosampling System Technical Note CYP Inhibition Assays: Versatility in the ADDA High Speed Dual Arm Autosampling System Data provided by Mike West and John Janiszewski, Pfizer Inc., Groton, CT Introduction ADDA System Features:

More information

Thermo Scientific TSQ Quantum Access MAX

Thermo Scientific TSQ Quantum Access MAX www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Scientific TSQ Quantum Access MAX UNSURPASSED PRICE TO PERFORMANCE Superior

More information

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study Application note Biotherapeutics and Biologics Authors Suresh Babu C.V., Anne

More information

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC Quality-by-Design-Based Method Development Using an Agilent 129 Infinity II LC An Efficient Method Development Workflow Combined with ISET-mediated Method Transfer Under Waters Empower 3 CDS Control Application

More information

Application Note # ET-20 BioPharma Compass: A fully Automated Solution for Characterization and QC of Intact and Digested Proteins

Application Note # ET-20 BioPharma Compass: A fully Automated Solution for Characterization and QC of Intact and Digested Proteins Application Note # ET-20 BioPharma Compass: A fully Automated Solution for Characterization and QC of Intact and Digested Proteins BioPharma Compass TM is a fully automated solution for the rapid characterization

More information

Automated Real-Time Determination of Bromate in Drinking Water Using LC-ICP-MS and EPA Method Application

Automated Real-Time Determination of Bromate in Drinking Water Using LC-ICP-MS and EPA Method Application Automated Real-Time Determination of Bromate in Drinking Water Using LC-ICP-MS and EPA Method 321.8 Application ICP-MS Author Jason A. Day, Anne Vonderheide, Joseph A. Caruso University of Cincinnati Cincinnati,

More information

Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run

Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run Application Note Author Jessica Westland Agilent Technologies, Inc. Abstract A selected target compound list of 195 various pesticides

More information

NOTICE OF INTENDED DECISION TO ENTER INTO A SINGLE SOURCE CONTRACT (PUR 7778)

NOTICE OF INTENDED DECISION TO ENTER INTO A SINGLE SOURCE CONTRACT (PUR 7778) NOTICE OF INTENDED DECISION TO ENTER INTO A SINGLE SOURCE CONTRACT (PUR 7778) This Notice of Intended Decision to Enter into a Single Source Contract is posted in accordance with section 287.057(3), Florida

More information

A highly sensitive and robust 150 µm column to enable high-throughput proteomics

A highly sensitive and robust 150 µm column to enable high-throughput proteomics APPLICATION NOTE 21744 Robust LC Separation Optimized MS Acquisition Comprehensive Data Informatics A highly sensitive and robust 15 µm column to enable high-throughput proteomics Authors Xin Zhang, 1

More information

Enhanced biodegradation of carbamazepine after UV/H 2 O 2 advanced. oxidation

Enhanced biodegradation of carbamazepine after UV/H 2 O 2 advanced. oxidation Supporting Information for the Environmental Science and Technology article: Enhanced biodegradation of carbamazepine after UV/H 2 O 2 advanced oxidation Prepared for submission to Environmental Science

More information

DETERMINATION OF PROTEIN BINDING BY UPLC/MS/MS

DETERMINATION OF PROTEIN BINDING BY UPLC/MS/MS DETERMINATION OF PROTEIN BINDING BY UPLC/MS/MS Darcy Shave and Pete Alden Waters Corporation, Milford, MA, U.S.A. INTRODUCTION Plasma protein binding (PPB) can significantly affect the therapeutic action

More information

Research Article Development and Validation of Acyclovir HPLC External Standard Method in Human Plasma: Application to Pharmacokinetic Studies

Research Article Development and Validation of Acyclovir HPLC External Standard Method in Human Plasma: Application to Pharmacokinetic Studies Advances in Pharmaceutics, Article ID 284652, 5 pages http://dx.doi.org/10.1155/2014/284652 Research Article Development and Validation of Acyclovir HPLC External Standard Method in Human Plasma: Application

More information

Quantification of Drug Metabolites in Early-Stage Drug Discovery Testing

Quantification of Drug Metabolites in Early-Stage Drug Discovery Testing Quantification of Drug Metabolites in Early-Stage Drug Discovery Testing Christopher Crafts, Bruce Bailey, John Waraska, Ian Acworth Thermo Fisher Scientific, Chelmsford, MA, USA Overview Purpose: The

More information

ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER

ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER ENVIRONMENTAL ANALYSIS ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER Solutions for Your Analytical Business

More information

Method of determination of phthalates by gas chromatography / mass spectrometry in wines

Method of determination of phthalates by gas chromatography / mass spectrometry in wines Method OIV-MA-AS323-10 Type IV methods 1. SCOPE This method applies to the detection and assay of phthalates in wines. 2. PRINCIPLE The sample is extracted using isohexane. The extract is concentrated

More information

Revision of 30 April 2013 draft, 4 November 2013

Revision of 30 April 2013 draft, 4 November 2013 GUIDANCE DOCUMENT FOR SINGLE LABORATORY VALIDATION OF QUANTITATIVE ANALYTICAL METHODS USED IN SUPPORT OF PRE- AND POST-REGISTRATION DATA REQUIREMENTS FOR PLANT PROTECTION AND BIOCIDAL PRODUCTS INTRODUCTION

More information