1 Enhancing your HPLC Separations using Temperature Programming Brian Jones, Stephanie J. Marin, Jody Clark, and Dale Felix Selerity Technologies, Inc W. Custer Road Salt Lake City, UT
2 The High Temperature HPLC Advantage Speed Efficiency Selectivity Gradient elution
3 Faster and More Efficient Separations Speed Flatter van Deemter curves allow operation at flow rates many times optimal velocity Higher efficiency - better resolution Increased diffusion rates provide lower plate heights at higher flow rates Lower viscosity and back pressure permits higher flow rates with smaller particle size packings
4 Better Chromatography with Temperature Gradient Programming Change retention through temperature gradient programming Replace solvent gradients with temperature gradients Water less polar and more organic-like so less organic modifier needed
5 Obstacles to High Temperature HPLC Need a fully programmable easy-to-use oven capable of fast response Must eliminate thermal mismatch to reduce band-broadening Need columns stable at elevated temperatures Must correct for refractive index differences when mobile phase temperature is increased
6 The Selerity Series 9000 Total Temperature Controller Small footprint Sub-zero to 200 C Mobile phase preheating or precooling Precision mobile phase outlet temperature control Flammable vapor sensor Variable low voltage mobile phase preheater Integral high speed microcomputer control
7 Small and Versatile Temperature Programmable Column Heater/Chiller
8 Vapor Sensor Provides operator safety when working with flammable or toxic mobile phases at elevated temperatures Automatically cools and ventilates oven and sends a stop signal to the HPLC pump
9 Multiple Point Mobile Phase Temperature Monitoring Preheater (integral) Outlet temperature controller Auxiliary for optional user placement Electrically isolated thermocouples and RTD All can be internally calibrated to the column oven, which is itself calibrated with a NIST traceable reference standard
10 Cryo Capability with Mobile Phase Pre-Cooling Solid state Peltier electronic cooling Cryo fluid option for both mobile phase and oven (nitrogen and CO 2 )
11 Precision Mobile Phase Temperature Conditioning The hazards and detector instability problems of hot exiting mobile phases are eliminated
12 Fast Temperature Programming, Fast Cool-down Temperature gradients to +30 C/min Cool-down from 200 C in 2 minutes --High Analysis Throughput--
13 Solvent Pre-heater Design (1) connection tubing from injector, (2) heater, (3) temperature sensor, (4) microprocessor controller, (5) column, (6) mobile phase flow
14 Mobile Phase Pre-heater Very fast response and non-invasive Low-mass and low-volume, <1 µl 0.005, and ID available Adds no dead volume
15 Axial Heat Profile
16 Variable Low Voltage Pre-heater Automatically steps power to the preheater in 60 mv DC increments Safe low voltage operation Typical voltage feeding the heater at 1 ml/min mobile phase flow brought to 150 C is only 6 volts
17 Variable Low Voltage Pre-heater Can lead, match, or lag oven temperature program for analyte focusing Automatically responds to changing mobile phase heat capacity through composition gradients or flow programming Convenient plug-in access in the front door
18 Why is Solvent Pre-heating so Important? No Pre-heating Flow Flow With Pre-heating
19 Separation of Barbiturates Mobile Phase Pre-heating Improves Chromatography Barbital Pre-heater Off mv 20 Butabarbital Carbromal Secobarbital 10 Pre-heater On Zirchrom PBD, 80 C Minutes
20 Separation of Steroids Using Water as a Mobile Phase Flow Rate: Mobile Phase: Temperature: ml/min Water 200 o C Testosterone Mix 1. Uracil 2. Androstadienedione 3. Androstenedione 4. Epitestosterone mv Flow Rate: 3.0 ml/min Mobile Phase: 25:75 acetonitrile:water Temperature: 50 o C Minutes Zirchrom PBD 100 X 4.6 mm column, UV 254 nm
21 Traditional Silica Columns Can t Take the Heat Water attacks siloxane bond or behind point of phase attachment Backside Attack H 2O: R Degradation of the Silica Support CH 3 Si + OH OH HO :OH O CH 3 Si O Si O Si H O + Si
22 The Selerity Blaze Silica Column Can Take the Heat Selerity polydentate phase protects the silanol groups Polymer attaches to backbone at several points R R R Si Si Si O OH O O Si O Si O Si O Si O
23 Separation of Analgesics on a Selerity Blaze C8 Using a Thermal Gradient Acetaminophen Caffeine Salicylamide Aspirin Salicylic acid Ibuprofen Naproxen mv Minutes Column: Selerity Blaze C8, 3 µm, 100 x 4.6 mm Mobile Phase: 40:60 acetonitrile:water with 0.1%TFA Flow Rate: 1.5 ml/min Detection: UV 220 nm Temperature Program: hold at 50 C for one minute, ramp to 100 C at 30 C/min, hold six min.
24 Other Columns That Can Take the Heat Columns stable at high temperatures Hamilton PRP-1 columns Thermo Hypersil-Keystone Hypercarb ZirChrom TM stationary phases Other columns under evaluation
25 Blank Thermal Gradients 50:50 Acetonitrile:Water at 254 nm Thermo Hypersil-Keystone Hypercarb Selerity Blaze C8 Hamilton PRP-1 Zirchrom Diamondbond Zirchrom CARB Zirchrom PBD 40 to 200 C at 15 /min hold 5 min 40 to 100 C at 15 /min hold 5 min 40 to 150 C at 15 /min hold 5 min 40 to 200 C at 15 /min hold 5 min 40 to 200 C at 15 /min hold 5 min 40 to 150 C at 15 /min hold 5 min mv Minutes
26 Separation of Analgesics using a Hypercarb Column and a Thermal Gradient mv (1)Caffeine (2)Aspirin (3)Salicylic Acid (4)Ibuprofen (5)Phenacetin (6)Acetaminophen (7)Naproxen Minutes Column: Thermo Hypersil-Keystone Hypercarb, 7 µm, 100 x 4.6 mm Mobile Phase: 35:65 acetonitrile:water with 0.1% TFA Flow Rate: 4.0 ml/min Detection: UV 220 nm Temperature Program: thermal gradient from 125 to 200 C at 30 /min, hold five min.
27 Separation of Narcotics on Hypercarb Column Using a Thermal Gradient 300 mv Morphine Hydromorphone Codeine Methadone Hydrocodone Oxycodone Diacetylmorphine Minutes Column: Thermo Hypersil-Keystone Hypercarb, 7 µm, 100 x 4.6 mm Mobile Phase: 50:50 acetonitrile:50 mm ammonium acetate ph 9.0 Flow Rate: 1.0 ml/min Detection: UV 220 nm Temperature Program: hold at 50 C for two minutes, ramp to 150 C at 30 /min, hold ten min.
28 Aromatic Hydrocarbons at Three Temperatures Uracil Benzene Toluene Ethylbenzene Naphthalene Biphenyl Tetrahydronaphthalene Anthracene Phenanthrene Chrysene mv o C o C 0 50 o C Chrysene elutes at 22 min Minutes ZirChrom PBD 100 x 4.6 mm, UV 254 nm, 35:65 Acetonitrile:Water, 2.0 ml/min
29 Aromatic Hydrocarbons on PBD Column Using a Temperature Gradient Uracil Benzene Toluene Ethylbenzene Naphthalene Biphenyl Tetrahydronaphthalene Anthracene Phenanthrene Chrysene (+impurity) mv 5 -baseline corrected Minutes Column: Zirchrom PBD 100 x 4.6 mm Mobile Phase: 35:65 Acetonitrile:Water Flow Rate: 2.0 ml/min Detection: UV 254 nm Temperature Program: 50 C to 150 C at 20 /min
30 Conclusions Temperature programming can be used successfully with 4.6 mm ID columns Independent mobile phase pre-heater control allows for improvement in peak efficiency The benefits of elevated temperature HPLC are now available for routine use
31 Future Work Explore the new capability of precision mobile phase temperature matching in subambient HPLC Determine optimal preheater program profiles as a function of column dimensions and oven ramp rates
32 Acknowledgements Thermo Hypersil-Keystone Hamilton Company ZirChrom Separations
33 Turn up the Heat Bring on the Cold Selerity Technologies Inc W. Custer Rd. Salt Lake City, UT Patent applications have been filed relative to the new technologies presented in this work.
HPLC Peak Focusing Facilitated by Independent Mobile Phase Pre-heating
HPLC Peak Focusing Facilitated by Independent Mobile Phase Pre-heating Brian Jones, Stephanie J. Marin, Jody Clark, Dale Felix Selerity Technologies, Inc. 2484 W. Custer Road Salt Lake City, UT 84104 www.selerity.com
More informationUsing High Temperature HPLC for Improved Analysis in the Pharmaceutical Industry
Using High Temperature HPLC for Improved Analysis in the Pharmaceutical Industry Stephanie J. Marin Brian Jones, Dale Felix, Jody Clark Selerity Technologies, Inc. Salt Lake City, UT www.selerity.com The
More informationReplacing Solvent Gradient HPLC Methods with Temperature Programmed HTLC
Replacing Solvent Gradient HPLC Methods with Temperature Programmed HTLC Stephanie J. Marin Brian Jones, Dale Felix, Jody Clark Selerity Technologies, Inc. Salt Lake City, UT www.selerity.com Introduction
More informationUsing High Temperature HPLC for Improved Analysis in the Pharmaceutical Industry
Using High Temperature HPLC for Improved Analysis in the Pharmaceutical Industry Stephanie J. Marin Brian Jones, Dale Felix, Jody Clark Selerity Technologies, Inc. Salt Lake City, UT www.selerity.com Abstract
More informationWater Only HPLC With FID Detection
Water Only HPLC With FID Detection W. Dale Felix, Stephanie J. Marin Brian A. Jones Selerity Technologies, Inc. 2484 W. Custer Road Salt Lake City, UT 84104 www.selerity.com 1 Introduction Utilizing elevated
More informationTemperature and ph Stability of a New High Temperature C18 Polydentate Silica Column
Temperature and ph Stability of a New High Temperature C18 Polydentate lica Column Stephanie J. Marin 1, Brian A. Jones 1 Jody Clark 1, Nathan L. Porter 1 J. Andreas Lippert 2, and Todd M. Johnson 2 1.
More informationNatural Product Analysis Using Temperature Programmed HPLC: Sub-Ambient through 200 ºC
Natural Product Analysis Using Temperature Programmed HPLC: Sub-Ambient through 200 ºC Dale Felix, Stephanie J. Marin Brian Jones, Jody Clark Selerity Technologies, Inc. Salt Lake City, UT www.selerity.com
More informationUse and Practicality of Water- Only HPLC Separations
Use and Practicality of Water- Only HPLC Separations Jody Clark, Stephanie J. Marin, Brian A. Jones and W. Dale Felix Selerity Technologies, Inc. 2484 W. Custer Road Salt Lake City, UT 84104 www.selerity.com
More informationHigh Temperature Liquid Chromatography
High Temperature Liquid Chromatography C.V. Mceff 1*, B. Yan 1, D. R. Stoll 2, R.A. Henry 3 1 ZirChrom Separations, Inc., 617 Pierce Street, Anoka, M 55303 2 Department of Chemistry, University of Minnesota,
More informationPart III Improving Throughput Through the Use of Elevated Column Temperature
Part III Improving Throughput Through the Use of Elevated Column Temperature Review potential approaches to improving the speed of HPLC Review the concepts of Ultra-Fast High Temperature Liquid Chromatography
More informationNew Stable Chemically Bonded Carbon Stationary Phases for HPLC
New Stable Chemically Bonded Carbon Stationary Phases for HPLC Angelos Kyrlidis, Lynn Toomey, and Elena Khmelnitskaia, Cabot Corporation Dwight Stoll, Clayton V. McNeff, and Peter W. Carr ZirChrom Separations,
More informationDwight Stoll and Peter W. Carr. University of Minnesota Department of Chemistry
High Speed Two-Dimensional Liquid Chromatography Through the Use of Ultra-Fast High Temperature Liquid Chromatography as the Second Separation Dimension Dwight Stoll and Peter W. Carr University of Minnesota
More informationCombining High Temperature and Small Particles: The Advantages of Zirconia
Combining High Temperature and Small Particles: The Advantages of Zirconia EAS 9 Dan owlan 1, Bingwen Yan 1, Clayton V. Mceff 1, R.A. Henry 1 ZirChrom Separations, Inc. 617 Pierce St., Anoka, M 5533 Independent
More informationFundamentals and Techniques of Preparative HPLC. Parto Zist Behboud Tel: 42108
Fundamentals and Techniques of Preparative HPLC Parto Zist Behboud www.partoz.com Tel: 42108 Today s Overview 1 Basic know-how for up-scaling 2 Practical examples of up-scaling 3 Recycle preparation 4
More informationEvaluation of Sub-2µm Zirconia-PBD Particles for Multi-Modal UHPLC
Evaluation of Sub-2µm Zirconia-PBD Particles for Multi-Modal UHPLC Dan Nowlan, B. Yan, C. V. McNeff and R. A. Henry ZirChrom Separations, Inc., 617 Pierce St., Anoka, MN 55303 Horvath 2009 1 Introduction
More informationAPPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology
Determination of Impurities and Related Substances for (Ph. Eur. Monograph 8): Increased Sensitivity, Improved Resolution and Faster Analysis Using Kinetex.6 µm Core-Shell LC Columns Ellie Abbasi, Jeff
More informationA Brief Overview of HPLC & UHPLC Method Development and Optimization. Dr. Chris Message UHPLC/HPLC Product Specialist Phenomenex
A Brief Overview of HPLC & UHPLC Method Development and Optimization Dr. Chris Message UHPLC/HPLC Product Specialist Phenomenex Poll Question 1 What learning objectives most attracted you to this webinar?
More informationReversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 300SB-C8 1.
Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 3SB-C8 1.8 µm Column Application Note Biopharmaceuticals Authors James Martosella and Phu
More informationThe Role of Temperature in HPLC How Column Thermostatting and Mobile Phase Pre-Conditioning Details enable to take full Advantage of It
The Role of Temperature in HPLC How Column Thermostatting and Mobile Phase Pre-Conditioning Details enable to take full Advantage of It Frank Steiner Thermo Fisher Scientific, Germering, Germany The world
More informationKromasil Eternity Designed for long life
1 Kromasil Eternity Designed for long life 2 Outline of the presentation Introduction of the Kromasil Eternity p Modifying the silica p Functionalizing the silica: C18 and PhenylHexyl Product characteristics/assortment
More informationApplication Note. Authors. Abstract. Introduction. Pharmaceutical
Scalability of Agilent s Across HPLC and UHPLC Instruments Application Note Pharmaceutical Authors Anne E. Mack, William J. Long Agilent Technologies, Inc. 2850 Centerville Road, Wilmington, DE 19808 USA
More informationNovel Wide-Pore Superficially Porous Particles for Biomacromolecular Separations
Novel Wide-Pore Superficially Porous Particles for Biomacromolecular Separations S. A. Schuster, B. M. Wagner, B. E. Boyes and J. J. Kirkland Advanced Materials Technology, Inc. 3521 Silverside Rd., Quillen
More informationIntroducition of a new product from Inertsil Series
Introducition of a new product from Inertsil Series is a C8 column developed focusing on superb inertness to basic, acidic, neutral and chelating compounds, which can virtually provide perfect peak shapes
More informationAPPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology
TN-7 Determination of Impurities and Related Substances for Glibenclamide (EP Monograph 78). Increased Sensitivity, Improved Resolution and Faster Analysis Using Kinetex.6 µm Core-Shell LC Columns Elli
More informationFaster Separations Using Agilent Weak Cation Exchange Columns
Faster Separations Using Agilent Weak Cation Exchange Columns Application Note BioPharma Author Andrew Coffey Agilent Technologies, Inc. Abstract Ion exchange is a commonly used technique for the separation
More informationInertsil ODS-3 2 µm. Base Silica Physical Properties and Chemical Modification. Base Silica: High Purity Silica Gel % Surface Area: 450 m 2 /g
0 High-Speed Separation Inertsil ODS-3 2 µm Base Silica Physical Properties and Chemical Modification Base Silica: High Purity Silica Gel 99.999% Surface Area: 450 m 2 /g Pore Size: 100 Å Pore Volume:
More informationSFC Säulen für analytische und preparative Anwendungen
SFC Säulen für analytische und preparative Anwendungen Technology Symposium Vienna, 25 th October 2016 DI Verena Schmid 2016 Waters Corporation 1 CHIRAL SEPARATIONS ACQUITY UPC 2 Trefoil 2016 Waters Corporation
More informationRomain Laverriere Tatiana Pachova HPLC
Romain Laverriere Tatiana Pachova HPLC Table of content: 1 Abstract... 3 2 Introduction... 3 3 Methodology... 3 4 Results... 3 4.1 Peak identification... 4 4.2 Determination of the optimum flow rate...
More informationBivalirudin Purification:
Bivalirudin Purification: Sorbent Screening and Overload Experiments Marc Jacob, Joshua Heng, and Tivadar Farkas Phenomenex, Inc., 411 Madrid Ave., Torrance, CA 90501 USA PO94190412_W Abstract In this
More informationOptimizing Productivity and Reliability for Monocyclic Aromatic Hydrocarbon Purity Analysis According to ASTM D7504 on the Agilent 8890 GC System
Application Note Petrochemicals Optimizing Productivity and Reliability for Monocyclic Aromatic Hydrocarbon Purity Analysis According to ASTM D50 on the Agilent 0 GC System Authors Jie Pan, Lukas Wieder,
More informationHPLC to UPLC Method Migration: An Overview of Key Considerations and Available Tools
HPLC to UPLC Method Migration: An Overview of Key Considerations and Available Tools Dr. Michael Swartz, Ph. D. Principal Consulting Scientist Worldwide Pharmaceutical Business Operations Waters Corporation
More informationOverview of preparative HPLC. Analytical Technologies Limited
Overview of preparative HPLC Analytical Technologies Limited 1.Review of liquid phase chromatography origin develop prospect Mobile phase Tewett 1903 Stationa ry phase origin Separation result Liquid-stationary
More informationAnalytical Instrument Qualification According to USP <1058>: Requirements and Examples for the Agilent 1290 Infinity LC System
Analytical Instrument Qualification According to USP : Requirements and Examples for the Agilent 1290 Infinity LC System 1) USP : Scope, approach, requirements 2) Qualification Examples for
More informationFast Detailed Hydrocarbon Analysis by Modified ASTM Method D6730
application Note Gas Chromatography Author Terry Osenbach PerkinElmer, Inc. Shelton, CT 06484 USA Fast Detailed Hydrocarbon Analysis by Modified ASTM Method D6730 Introduction Detailed hydrocarbon analysis
More informationACQUITY Arc TM System
ACQUITY Arc TM System Introducing the Newest Member of the Waters LC Portfolio 2015 Waters Corporation 1 Adopting Modern LC Technology in a Global Economy Highly competitive, regulated business environment
More informationChem 321 Lecture 23 - Liquid Chromatography 11/19/13
Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Student Learning Objectives High Performance Liquid Chromatography With the advent of relatively inexpensive and reliable pumps, the development of
More informationPerformance characteristics of the 1260 Infinity Quaternary LC system
Performance characteristics of the 1260 Infinity Quaternary LC system The new standard in HPLC Technical Overview Introduction The Agilent 1260 Infinity LC system consists of modular units that operate
More informationA Comparison of Compendia Normal Phase LC Methods Run Under Supercritical Fluid Chromatography Conditions
A Comparison of Compendia Normal Phase LC Methods Run Under Supercritical Fluid Chromatography Conditions Stéphane Dubant Waters ESAC April 5th 2011 2011 Waters Corporation 1 Agenda Introduction and Benefits
More informationHPLC Columns HIC PH-814 MANUAL. Shodex HPLC Columns Europe, Middle East, Africa, Russia. For technical support please use contact details shown below:
HPLC Columns MANUAL HIC PH-814 Shodex HPLC Columns Europe, Middle East, Africa, Russia For technical support please use contact details shown below: SHOWA DENKO EUROPE GmbH Shodex Business Konrad-Zuse-Platz
More informationfor Acclaim Carbamate Column
for Acclaim Carbamate Column Product Manual for Acclaim Carbamate Page 1 of 13 Product Manual for Acclaim Carbamate Analytical Columns 5 µm, 4.6 x 250 mm, P/N 072924 3 µm, 4.6 x 150 mm, P/N 072925 3 µm,
More informationSpecial media HPLC Columns -
Special media HPLC Columns - With the partnership of Teknokroma Zirconia (ZrO2) Columns SPECIAL MEDIA HPLC COLUMNS Zirconia or zirconium dioxide (Zr02) is a metal oxide that exists in a variety of crystalline
More informationAgilent SD-1 Purification System. Purify your way SD-1
Agilent SD-1 Purification System Purify your way SD-1 AGILENT SD-1 PURIFICATION SYSTEM PURIFY YOUR WAY WITH HIGH QUALITY SEPARATIONS AT ANY SCALE The Agilent SD-1 Purification System achieves better gradient
More informationPerformance Characteristics of the Agilent 1220 Infinity Gradient LC system
Performance Characteristics of the Agilent 122 Infinity Gradient LC system An integrated LC system for conventional LC and UHPLC Technical Overview 7 5 4 3 2 1.5 1 1.5 2 2.5 3 Introduction The Agilent
More informationGradient Elution. Slide 2
Gradient Elution Why choose this separation mode? What HPLC parameters affect a gradient separation? How can I use these parameters to improve my gradient separation? Slide 2 Is Gradient or Isocratic Elution
More informationAgilent 218 Purification System. Purify your way
Agilent 218 Purification System Purify your way 218 AGILENT 218 PURIFICATION SYSTEM PURIFY YOUR WAY WITH A FLEXIBLE SYSTEM WITHIN REACH OF YOUR BUDGET The Agilent 218 Purification System has the flexibility
More informationTemperature Effect on Peak Width and Column Efficiency in Subcritical Water Chromatography
Temperature Effect on Peak Width and Column Efficiency in Subcritical Water Chromatography Yu Yang 1, *, Lori J. Lamm 1, Ping He 1, and Toru Kondo 2 1 Department of Chemistry, East Carolina University,
More information10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor
229 10. Validated Normal Phase HPLC Method for the Determination of Fulvestrant in Pharmaceutical Dosage Forms 10.1 Introduction Fulvestrant is primarily used in the treatment of hormone receptor positive
More informationThe Suite for Environmental GC Analysis
The Suite for Environmental GC Analysis SGE Environmental GC Columns Performance Selectivity Delivery Promise www.sge.com SGE Environmental GC Columns The Suite for Environmental GC Analysis SGE GC Columns
More informationTransferring a Method from a Conventional 3.5 µm or 5 µm to an Agilent Poroshell 120 Column to Improve Analytical Speed and Resolution
Video Notes Transferring a Method from a Conventional 3.5 µm or 5 µm to an Agilent Poroshell 120 Column to Improve Analytical Speed and Resolution Ron Majors is Senior Scientist for Agilent s Columns and
More informationAccucore. Ultimate Core Performance LC Column Technology to Maximize Your Investment. Dave Jarzinski. October 2011
Accucore Ultimate Core Performance LC Column Technology to Maximize Your Investment Dave Jarzinski Thermo Fisher Scientific Account Manager Greater Boston/Cambridge & North Phone: 978-408-1576 Email: dave.jarzinski@thermofisher.com
More informationUltra High Performance Liquid Chromatograph. Nexera XR C196-E082C
Ultra High Performance Liquid Chromatograph Nexera XR C196-E82C The Most Accessible UHPLC Available Developed with expandability and compatibility in mind, the Nexera XR ultra high performance liquid chromatograph
More informationComparison Guide. Comparison Data on Commonly Used C18 Phases TO C18 REVERSED P HASE HPLC COLUMNS. Stationary Phase Specifications
Comparison Guide TO C8 REVERSED P HASE HPLC COLUMNS Comparison Data on Commonly Used C8 Phases Stationary Phase Specifications Phases Compared According to Relative Hydrophobicity Comparison of Column
More informationBÜCHI Labortechnik AG
BÜCHI Labortechnik AG Purification BUCHI Chromatography History 1977: BUCHI Fraction Collector 1978: 1 st preparative chromatography system 2008: automated chromatography system 2014: automated high performance
More informationZwitterion Chromatography ZIC
Zwitterion Chromatography ZIC A novel technique, with unique selectivity, suitable for preparative scale separations? PhD Einar Pontén What is Zwitterion Chromatography? Our definition: Liquid chromatography
More informationApplication Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808
Reversed-Phase Optimization for Ultra Fast Profiling of Intact and Reduced Monoclonal Antibodies using Agilent ZORBAX Rapid Resolution High Definition 3SB-C3 Column Application Note Biopharma Authors James
More informationGC Method Development
GC Method Development Simon Jones Application Engineer April 13, 2010 What to Consider The Sample Method of injection Inlet Detector Carrier Gas Column COMPOUND REQUIREMENTS FOR GC Only 10-20% of all compounds
More informationPatricia McConville Manager Product Positioning Laboratory. Next Generation Instrument Design
Patricia McConville Manager Product Positioning Laboratory Next Generation Instrument Design Patricia McConville Manager Product Positioning Laboratory 2005 Waters Corporation Next Generation Instrument
More informationBiotherapeutic Method Development Guide
Biotherapeutic Method Development Guide HALO BIOCLASS 1000 Å PROTEIN SELECTIVITY KIT Fused-Core Particle Technology BIOCLASS Strategy for Optimizing Protein Separations Using Reversed-phase Liquid Chromatography
More informationRapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column
Rapid UHP Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 3SB-C8 Column Application Note BioPharma Authors James Martosella, Phu Duong, Susanne
More informationultra why micro-lc? product note introducing the ExpressLC -Ultra system: UHPLC with all the advantages of micro-lc
product note introducing the ExpressLC -Ultra system: UHPLC with all the advantages of micro-lc With the new ExpressLC-Ultra, Eksigent offers an LC system that delivers all the advantages of micro-lc at
More informationUltra High Performance Liquid Chromatograph. Nexera XR C196-E082D
Ultra High Performance Liquid Chromatograph Nexera XR C196-E82D The Most Accessible UHPLC Available Developed with expandability and compatibility in mind, the Nexera XR ultra high performance liquid chromatograph
More informationFast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides
Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides High-pH-stable, superficially porous particle columns for LC/UV and LC/MS Application Note Biologics and Biosimilars Authors
More informationRAM DIRECT INJECTION (RESTRICTED ACCESS MEDIA)
RAM CHROMATOGRAPHY RAM DIRECT INJECTION (RESTRICTED ACCESS MEDIA) A Tool for the Separation of Small Molecules in the Presence of Large Biomolecules Conventional Sample Precipitate proteins Centrifuge
More informationEvaluation of Acclaim HPLC Columns using the National Institute of Standards and Technology Standard Reference Material 870
pplication Update 6 Evaluation of cclaim HPL olumns using the National Institute of Standards and Technology Standard Reference Material 870 INTRODUTION The National Institute of Standards and Technology
More informationReversed-Phase HPLC Columns
Reversed-Phase HPLC Columns Hamilton reversed-phase HPLC columns combine the best characteristics of silica-based and polymeric columns to arrive at a product that is highly inert and long-lasting. Hamilton
More informationBiotherapeutic Method Development Guide
Biotherapeutic Method Development Guide HALO BIOCLASS 1000 Å PROTEIN SELECTIVITY KIT Fused-Core Particle Technology BIOCLASS Strategy for Optimizing Protein Separations Using Reversed-phase Liquid Chromatography
More informationADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS
Agilent AdvanceBio Columns ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS with Agilent ZORBAX RRHD 3Å 1.8 µm columns ns Rapid resolution high definition columns for UHPLC protein and peptide separations
More informationION EXCHANGE KIT FOR MAB SEPARATIONS
ION EXCHANGE KIT FOR MAB SEPARATIONS Sepax Technologies, Inc. 5 Innovation Way Newark, Delaware, USA Tel: (32) 366-111 Fax: (32) 366-1151 Toll free: www.sepax-tech.com Content Introduction... 1 Technical
More informationOptimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools
Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools Application Note Abstract The Commonwealth of Massachusetts Department of Environmental Protection
More informationPreparative HPLC: Factors and Parameters that Directly Affect Recovery of Collected Fractions
Preparative HPLC: Factors and Parameters that Directly Affect Recovery of Collected Fractions Gary Scharrer and Joan Stevens, Ph.D. Gilson, Inc Middleton, WI www.gilson.com Abstract Purification of synthetic,
More informationFast, Low Pressure Analysis of Food and Beverage Additives Using a Superficially Porous Agilent Poroshell 120 EC-C18 Column
Fast, Low Pressure Analysis of Food and Beverage Additives Using a Superficially Porous Agilent Poroshell 12 EC-C18 Column Application ote Food and Beverage Authors Anne E. Mack and William J. Long Agilent
More informationPreparative, Pilot & Process Columns
Preparative, Pilot & Process Columns 321 Preparative, Pilot & Process Columns Contents Prep Columns... 322-323 Dynamic Axial Compression Columns (DAC)... 324-325 Pilot Columns... 326-327 Chromatography
More informationSCION GC Series. The Gas Chromatographers Choice for Separations
SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Scion Instrument s tradition of innovation and product reliability have combined to create the next generation
More informationSCION GC Series. The Gas Chromatographers Choice for Separations
SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Scion Instrument s tradition of innovation and product reliability have combined to create the next generation
More informationApplication Note. Author. Abstract. Pharmaceuticals. Detlef Wilhelm ANATOX GmbH & Co. KG. Fuerstenwalde, Germany mau
Development, validation, and comparison of an HPLC method to analyze paracetamol and related impurities according to the European Pharmacopoeia (EP) and USP using the Agilent 1120 Compact LC and the Agilent
More informationProtein-Pak Hi Res HIC Column and HIC Protein Standard
Protein-Pak Hi Res HIC Column and HIC Protein Standard CONTENTS I. INTRODUCTION II. a. Mobile Phase b. Flow Direction CONNECTING COLUMN TO LC SYSTEM I. INTRODUCTION This offering contains non-porous, polymethacrylate-based
More informationRAM Direct Injection. (Restricted Access Media) A Tool for the Separation of Small Molecules in the Presence of Large Biomolecules
RAM Direct Injection (Restricted Access Media) A Tool for the Separation of Small Molecules in the Presence of Large Biomolecules HPLC analysis of small molecules contained within a protein matrix can
More informationFast and Comprehensive 2-Dimensional Liquid Chromatography of Block Copolymers through Temperature Gradient Driven Trapping
Fast and Comprehensive 2-Dimensional Liquid Chromatography of Block Copolymers through Temperature Gradient Driven Trapping International Symposium on GPC/SEC and related techniques Bastiaan Staal & Joshua
More informationHPLC Column Troubleshooting: Is It Really The Column?
HPLC Column Troubleshooting: Is It Really The Column? Agilent Technologies, Inc. Rita Steed Application Engineer January 29, 2009 Page 1 Troubleshooting in HPLC ma U 20 00 15 00 10 005 0 0 0 1 0 5 0 1
More informationCLEAN SCREEN FASt FILTER AND SHOOT UCT FORENSICS. CLEAN SCREEN FASt SPE COLUMNS
FORENSICS UCT CLEAN SCREEN FASt FILTER AND SHOOT CLEAN SCREEN FASt SPE COLUMNS CLEAN SCREEN FASt SPE COLUMNS Since the introduction and implementation of LC/MS as a staple analytical tool in forensic laboratories,
More informationHPLC 920-LC ANALY TIC AL LIQUID CHROMATOGR APHY SOLUTIONS.
HPLC varian, inc. 92-LC ANALY TIC AL LIQUID CHROMATOGR APHY SOLUTIONS www.varianinc.com VARIAN, INC. Integrating Performance and Reliability Fast HPLC The Varian 92-LC analytical high performance liquid
More informationSimple Techniques for Improving the Isolation of Synthetic Peptides Jo-Ann Jablonski Principal Scientist Waters Corporation
Simple Techniques for Improving the Isolation of Synthetic Peptides Jo-Ann Jablonski Principal Scientist Waters Corporation 2016 Waters Corporation 1 Agenda Background Techniques Scaling a separation Focusing
More informationMethod Translation in Liquid Chromatography
Method Translation in Liquid Chromatography Technical Overview Abstract Ronald E. Majors Agilent Technologies, Inc. 2850 Centerville Rd Wilmington, DE 19808 USA With the recent emphasis on high performance
More informationFast Analysis of Environmental Phenols with Agilent Poroshell 120 EC-C18 Columns
Fast Analysis of Environmental Phenols with Agilent Poroshell 120 EC-C18 Columns Application Note Environmental Authors William J. Long and Anne E. Mack Agilent Technologies, Inc. 2850 Centerville Road
More informationDIDANOSINE ORAL POWDER Final text for addition to The International Pharmacopoeia
Document QAS/06.175(A)FINAL August 2007 DIDANOSINE ORAL POWDER Final text for addition to The International Pharmacopoeia This monograph was adopted at the 41 st WHO Expert Committee on Specifications
More informationThermo Scientific Dionex UltiMate 3000 RSLC Systems Rapid separation systems high-end performance
Thermo Scientific Dionex UltiMate 3 RSLC Systems Rapid separation systems high-end performance Speed Resolution Versatility Rapid separation LC accelerate your analysis The Thermo Scientific Dionex UltiMate
More informationHigh-Performance Liquid Chromatography
High-Performance Liquid Chromatography ULTRON VX STR Silica Columns VX / STR Series VX-ODS VX-Octyl VX-SIL STR ODS-II STR ODS-M ULTRON VX-ODS ULTRON VX-ODS columns are reversed phase columns employing
More informationA Guide To Speeding Up Your Separation
A Guide To Speeding Up Your Separation Page 1 Choosing The Column Overview Reducing The Column Length Changing The Column Id Adjusting The Injection Volume Transferring to 1.8um Columns Speeding It Up
More informationEASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime
EASY-Spray Columns Guidance for column set up and installation Tips to maximize column lifetime EASY-Spray Column Tips and Tricks This document provides guidance for Thermo Scientific EASY-Spray column
More informationUpdating Solid Phase Extraction Methods: Tips and Tricks for Improving Existing and New SPE Methods using Method Development
Updating Solid Phase Extraction Methods: Tips and Tricks for Improving Existing and New SPE Methods using Method Development Golnar Javadi Application Engineer Sample Preparation Products Spp-support@agilent.com
More informationSmart HPLC LC800. Ph
Smart HPLC LC800 Distributed in Australia & New Zealand WINLAB PTY LTD PO Box 5007, Brendale,Winlab Queensland, Australia Pty Ltd 4500 Ph: +61 (0)7 3205 1209 Email: sales@winlab.com.au Web: www.winlab.com.au
More informationUltrahydrogel Columns
CONTENTS I. INTRODUCTION II. INSTALLATION a. Solvents b. Column c. Organic solvents d. Aqueous salt and buffer solution e. Flow rate f. Solvent and sample preparation g. Guard columns h. Column installation
More informationNISTmAb characterization with a high-performance RP chromatography column
APPLICATION NOTE 21848 NISTmAb characterization with a high-performance RP chromatography column Author Xin Zhang Thermo Fisher Scientific, Sunnyvale, CA, USA Keywords MAbPac RP column, inter-column reproducibility,
More informationSmartube Benchmark of Nano-LC
Smartube Benchmark of Nano-LC The state of the art Smartube is a high performance chromatographic column product line. It is made for high efficiency, high selectivity microscale separations. Compared
More informationProduct Guide for LudgerSep TM N2 High Resolution Amide HPLC Columns for Glycan Analysis
Product Guide for LudgerSep TM N2 High Resolution Amide HPLC Columns for Glycan Analysis (Ludger Product Codes: LS-N2-4.6x150 and LS-N2-2.0x150) Ludger Document # LS-N2-Guide-v4.0 Ludger Ltd Culham Science
More informationConcentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography
Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography Carl Fisher, Monika Verma, and Pranathi Perati; Thermo Fisher Scientific,
More informationIs there a Preparative Chromatography Green, Economic and Productive?
Is there a Preparative Chromatography Green, Economic and Productive? Granada, 23 Febrero 2012 Alberto Méndez Applications Specialist alberto_mendez@waters.com 2012 Waters Corporation 1 Why CO 2? CO 2
More informationHPLC Columns. Protein LW-803 MANUAL. Shodex HPLC Columns Europe, Middle East, Africa, Russia
HPLC Columns MANUAL Protein LW-803 Shodex HPLC Columns Europe, Middle East, Africa, Russia For technical support please use contact details shown below: SHOWA DENKO EUROPE GmbH Shodex Business Konrad-Zuse-Platz
More informationEmulation of the Agilent 1100 Series LC Through Waters Empower Software Analysis of an Analgesic Mixture
Agilent 129 Infinity II LC with ISET Emulation of the Agilent 11 Series LC Through Waters Empower Software Analysis of an Analgesic Mixture Application Note Small Molecule Pharmaceuticals Author Melanie
More informationPreparative HPLC is still the
B I O P R O C E S S TECHNICAL Optimizing Sample Load Capacity and Separation Through a Series of Short Prep Columns Mark Crawford, Joan Stevens, and Luke Roenneburg Preparative HPLC is still the dominate
More information