INSTRUCTIONS FOR USE

Size: px
Start display at page:

Download "INSTRUCTIONS FOR USE"

Transcription

1 VITROS Chemistry Products Slides Rx ONLY Intended Use For in vitro diagnostic use only. VITROS Slides quantitatively measure creatine kinase () activity in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. Summary and Explanation of the Test Creatine kinase, also referred to as creatine phosphokinase, is a cellular enzyme with a wide tissue distribution. is found mainly in skeletal and cardiac muscle. s physiological role is associated with ATP generation for contractile or transport systems. Serum is almost always increased following acute myocardial infarction or skeletal muscle damage. The enzyme is commonly elevated in myocarditis of any cause, cerebrovascular accidents, rhabdomyolysis, polymyositis, and acute physical exertion. is also increased in the muscular dystrophies; in Duchenne s muscular dystrophy, elevations of times normal are common. Low may reflect decreased muscle mass or muscle wasting. Reference values for must consider the age, gender, and physical activity of the person. Low serum activities are common in the elderly, in the bedridden, and in patients with advanced malignancy. 1 Principles of the Procedure The VITROS Slide method is performed using the VITROS Slides and the VITROS Chemistry Products Calibrator Kit 3 on VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. The VITROS Slide is a multilayered, analytical element coated on a polyester support. A drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. This layer also contains N-acetylcysteine (NAC) to activate without pretreating the sample. When the sample is deposited on the slide, creatine kinase catalyzes the conversion of creatine phosphate and ADP to creatine and ATP. In the presence of glycerol kinase (GK), glycerol is phosphorylated to L-α-glycerophosphate by ATP. Oxidation of L-α-glycerophosphate to dihydroxyacetone phosphate and hydrogen peroxide occurs in the presence of L α glycerophosphate oxidase (α-gpo). Finally, leuco dye is oxidized by hydrogen peroxide in the presence of peroxidase to form a dye. Reflection densities are monitored during incubation. The rate of change in reflection density is then converted to enzyme activity. Test Type and Conditions Test Type Multiple-point rate VITROS System 5600, 4600, 5,1 FS, 950, 250/350 Not all products and systems are available in all countries. Approximate Incubation Time Temperature Wavelength Reaction Sample Volume 5 minutes 37 C (98.6 F) 670 nm 11 µl Version 10.0 Pub. No. MP2-42_EN 1 of 11

2 INSTRUCTIONS FOR USE Warnings and Precautions Reaction Scheme creatine phosphate + ADP NAC, Mg +2 creatine + ATP glycerol + ATP GK L-α-glycerophosphate + ADP L-α-glycerophosphate + O 2 α-gpo dihydroxyacetone phosphate + H 2 O 2 H 2 O 2 + leuco dye peroxidase dye + 2H 2 O Warnings and Precautions For in vitro diagnostic use only. Reagents WARNING: Take care when handling materials and samples of human origin. Since no test method can offer complete assurance that infectious agents are absent, consider all clinical specimens, controls, and calibrators potentially infectious. Handle specimens, solid and liquid waste, and test components in accordance with local regulations and CLSI Guideline M29 2 or other published biohazard safety guidelines. For specific warnings and precautions for calibrators, quality control materials, and other components, refer to the Instructions for Use for the appropriate VITROS product, or to other manufacturer s product literature. Slide Diagram Slide Ingredients 1. Upper slide mount 2. Spreading layer (TiO 2 ): Reactive Ingredients per cm 2 L-alpha-glycerophosphate oxidase (Aerococcus sp.) 0.4 U; peroxidase (horseradish root) 1.4 U; glycerol kinase (E.coli) 0.5 U; creatine phosphate 170 µg; N acetylcysteine 54 µg; magnesium acetate 20 µg; glycerol 20 µg; 2-(3,5-dimethoxy-4- hydroxyphenyl)-4,5-bis-(4-dimethylaminophenyl) imidazole (leuco dye) 20 µg and adenosine diphosphate 20 µg. Other Ingredients Pigment, binder, buffers, surfactants, inhibitors, stabilizers, cross linking agent, dye solubilizer, scavenger and chelator. N-acetylcysteine 3. Reagent layer buffer, ph 7.0 adenosine diphosphate glycerol, magnesium acetate glycerol kinase, leuco dye peroxidase glycerophosphate oxidase creatine phosphate 4. Support layer 5. Lower slide mouont Reagent Handling Caution: Do not use slide cartridges with damaged or incompletely sealed packaging. Inspect the packaging for signs of damage. Be careful when opening the outer packaging with a sharp instrument so as to avoid damage to the individual product packaging. Reagent Preparation IMPORTANT: The slide cartridge must reach room temperature, C (64 82 F), before it is unwrapped and loaded into the slide supply. 1. Remove the slide cartridges from storage. 2. Warm the wrapped cartridge at room temperature for 30 minutes when taken from the refrigerator or 60 minutes from the freezer. 3. Unwrap and load the cartridge into the slide supply. Note: Load the cartridges within 24 hours after they reach room temperature, C (64 82 F). 2 of 11 Pub. No. MP2-42_EN Version 10.0

3 Reagent Storage and Stability Reagent Storage and Stability VITROS Slides are stable until the expiration date on the carton when they are stored and handled as specified. Do not use beyond the expiration date. Reagent Storage Condition Stability Unopened Refrigerated 2 8 C (36 46 F) 4 weeks Frozen -18 C ( 0 F) Until expiration date Opened On-analyzer System turned on 1 week On-analyzer System turned off 2 hours Verify performance with quality control materials: If the system is turned off for more than 2 hours. After reloading cartridges that have been removed from the slide supply and stored for later use. Specimen Collection, Preparation and Storage Specimens Recommended Serum Plasma: 3 Heparin IMPORTANT: Certain collection devices have been reported to affect other analytes and tests. 4 Owing to the variety of specimen collection devices available, Ortho-Clinical Diagnostics is unable to provide a definitive statement on the performance of its products with these devices. Confirm that your collection devices are compatible with this test. Specimens Not Recommended Do not use grossly hemolyzed specimens. 5 Serum and Plasma Specimen Collection and Preparation Collect specimens using standard laboratory procedures. 6, 7 Note: For details on minimum fill volume requirements, refer to the operating instructions for your system. Patient Preparation No special patient preparation is necessary. Special Precautions is unstable in serum. Centrifuge specimens and remove the serum or plasma from the cellular material within 4 hours of collection. 8 Specimen Handling and Storage Handle and store specimens in stoppered containers to avoid contamination and evaporation. Mix samples by gentle inversion and bring to room temperature, C (64 82 F), prior to analysis. Specimen Storage and Stability 8 Storage Temperature Stability Room temperature C (64 82 F) 4 hours Refrigerated 2 8 C (36 46 F) 5 days Frozen -18 C ( 0 F) 1 month Testing Procedure Materials Provided VITROS Chemistry Products Slides Materials Required but Not Provided VITROS Chemistry Products Calibrator Kit 3 Version 10.0 Pub. No. MP2-42_EN 3 of 11

4 INSTRUCTIONS FOR USE Calibration Quality control materials, such as VITROS Chemistry Products Performance Verifier I and II or VITROS Chemistry Products Isoenzyme Performance Verifier I and II VITROS Chemistry Products 7% BSA VITROS Chemistry Products FS Diluent Pack 2 (BSA/Saline) (for on-analyzer dilution) Operating Instructions Check reagent inventories at least daily to ensure that quantities are sufficient for the planned workload. For additional information, refer to the operating instructions for your system. IMPORTANT: Sample Dilution Bring all fluids and samples to room temperature, C (64 82 F), prior to analysis. Serum and Plasma If creatine kinase activity exceeds the system s measuring (reportable or dynamic) range: Manual Sample Dilution 1. Dilute the sample with VITROS 7% BSA. 2. Reanalyze. 3. Multiply the results by the dilution factor to obtain an estimate of the original sample s creatine kinase activity. Sample dilution results in higher creatine kinase activities than expected. 10 On-Analyzer Sample Dilution (VITROS Integrated, VITROS 5,1 FS/4600 and VITROS 250/350 only) Refer to the operating instructions for your system for more information on the On-Analyzer Dilution Procedure. For VITROS Integrated and 5,1 FS/4600 Chemistry Systems, use VITROS Chemistry Products FS Diluent Pack 2 for the dilution. Calibration Required Calibrators VITROS Chemistry Products Calibrator Kit 3 Calibrator Preparation, Handling, and Storage Refer to the Instructions for Use for VITROS Calibrator Kit 3. Calibration Procedure Refer to the operating instructions for your system. When to Calibrate Calibrate: When the slide lot number changes. When critical system parts are replaced due to service or maintenance. When government regulations require. For example, in the USA, CLIA regulations require calibration or calibration verification at least once every six months. The VITROS test may also need to be calibrated: If quality control results are consistently outside acceptable range. After certain service procedures have been performed. For additional information, refer to the operating instructions for your system. Calculations Based on sequential readings of the slide s reflectance at 670 nm over the defined incubation period, a rate of change in reflectance is determined. This rate is used in the software-resident multi-point rate calibration model to compute enzyme activity. Once a calibration has been performed for each slide lot, creatine kinase activity in unknown samples can be determined from the rate of change in reflectance measured for each unknown test slide. Validity of a Calibration Calibration parameters are automatically assessed by the system against a set of quality parameters detailed in the Coefficients and Limits screen on VITROS 250/350/950 Systems (on the VITROS Integrated and VITROS 5,1 FS/4600 Systems, see the Review Assay Data screen). Failure to meet any of the pre-defined quality parameters results in a failed calibration. The calibration report should be used in conjunction with quality control results to determine the validity of a calibration. 4 of 11 Pub. No. MP2-42_EN Version 10.0

5 Quality Control Measuring (Reportable or Dynamic) Range Conventional and SI Units Alternate Units (U/L) (μkat/l) For out-of-range samples, refer to Sample Dilution. Traceability of Calibration Values assigned to the VITROS Chemistry Products Calibrator Kit 3 for creatine kinase () are traceable to a modification of the Scandinavian Committee on Enzymes recommended method for the determination of creatine kinase at 37 C (98.6 F). 11, 12 Quality Control Quality Control Material Selection IMPORTANT: VITROS Performance Verifiers are recommended for use with the VITROS Chemistry and Integrated Systems. Evaluate the performance of other commercial control fluids for compatibility with this test before using for quality control. Control materials other than VITROS Performance Verifiers may show a difference when compared with other creatine kinase methods if they: Depart from a true human matrix. Contain high concentrations of preservatives, stabilizers, or other nonphysiological additives. Do not use control materials stabilized with ethylene glycol. Quality Control Procedure Recommendations Choose control levels that check the clinically relevant range. Analyze quality control materials in the same manner as patient samples, before or during patient sample processing. To verify system performance, analyze control materials: After calibration. According to local regulations or at least once each day that the test is being performed. After specified service procedures are performed. Refer to the operating instructions for your system. If control results fall outside your acceptable range, investigate the cause before deciding whether to report patient results. For general quality control recommendations, refer to Statistical Quality Control for Quantitative Measurements: Principles and Definitions; Approved Guideline-Third Edition 9 or other published guidelines. For additional information, refer to the operating instructions for your system. Quality Control Material Preparation, Handling, and Storage Refer to the Instructions for Use for VITROS Chemistry Products Performance Verifier I and II or VITROS Chemistry Products Isoenzyme Performance Verifier I and II or to other manufacturer's product literature. Results Reporting Units and Unit Conversion The VITROS Chemistry and Integrated Systems may be programmed to report results in conventional, SI, or alternate units. Conventional and SI Units Alternate Units U/L µkat/l (U/L x ) Limitations of the Procedure Known Interferences The VITROS Slide method was screened for interfering substances following NCCLS Protocol EP7. 23, 24 Carbon dioxide at a level of 40 mmol/l may cause up to a 30% negative bias in creatine kinase. For substances that were tested and did not interfere, refer to Specificity. Version 10.0 Pub. No. MP2-42_EN 5 of 11

6 INSTRUCTIONS FOR USE Expected Values Other Limitations Certain drugs and clinical conditions are known to alter creatine kinase activity in vivo. For additional information, refer to one of the published summaries. 17, 18 Expected Values Reference Interval Reference Interval 13 Conventional and SI Units (U/L) Alternate Units (μkat/l) Females Males The upper limit of the reference interval is reported to be affected by population characteristics such as the degree of physical activity 14 and race. 15 Distributions of values from normal, healthy subjects often demonstrate a positive skew 16 which leads to variable upper reference limit estimates. Each laboratory should confirm the validity of these intervals for the population it serves. Performance Characteristics Method Comparison The plots and table show the results of a comparison of serum samples analyzed on the VITROS 750 System with those analyzed using a modification of the comparative method recommended by the Scandinavian Committee on Enzymes 11, 12 at 37 C (98.6 F). Testing followed NCCLS Protocol EP9. 19 The tables also show the results of comparisons of serum samples on the VITROS 250 and 950 Systems and the VITROS 750 System, and comparisons of serum samples on the VITROS 5,1 FS System with the VITROS 950 System. In addition, the table shows the results of a comparison of serum and plasma samples on the VITROS 5600 Integrated System and the VITROS 5,1 FS Chemistry System. Testing followed NCCLS Protocol EP9. 20 Conventional and SI Units Alternate Units VITROS 750 System (U/L) VITROS 750 System (µkat/l) Comparative Method: Modification-Scandinavian Committee on Enzymes (U/L) Comparative Method: Modification-Scandinavian Committee on Enzymes (μkat/l) 6 of 11 Pub. No. MP2-42_EN Version 10.0

7 Performance Characteristics n Slope Correlation Coefficient Conventional and SI Units (U/L) Range of Sample Activity Intercept Sy.x Alternate Units (µkat/l) Range of Sample Activity Intercept Sy.x 750 vs. comparative method vs vs ,1 FS vs vs. 5,1 FS Analytical processing hardware and software algorithms on the VITROS 4600 Chemistry System are designed to the same specifications as those applied to the VITROS 5,1 FS Chemistry System. Assay performance on the VITROS 4600 System has been demonstrated to be comparable to that on the VITROS 5,1 FS System. All performance characteristics for VITROS 5,1 FS System are therefore applicable to the VITROS 4600 System. Precision Precision was evaluated with quality control materials on VITROS 250, 750, 950, and 5,1 FS Systems following NCCLS Protocol EP5. 21 Precision was evaluated with quality control materials on VITROS 5600 Integrated System following NCCLS protocol EP5. 22 The data presented are a representation of test performance and are provided as a guideline. Variables such as sample handling and storage, reagent handling and storage, laboratory environment, and system maintenance can affect reproducibility of test results. Conventional and SI Units (U/L) Alternate Units (µkat/l) Within Mean Activity Within Day SD * Within Lab SD ** Mean Activity Within Day SD * Within Lab SD ** Lab CV % ** No. Observ ,1 FS Specificity * Within Day precision was determined using two runs/day with two to three replications. ** Within Lab precision was determined using a single lot of slides and calibrating weekly. No. Days Analytical processing hardware and software algorithms on the VITROS 4600 Chemistry System are designed to the same specifications as those applied to the VITROS 5,1 FS Chemistry System. Assay performance on the VITROS 4600 System has been demonstrated to be comparable to that on the VITROS 5,1 FS System. All performance characteristics for VITROS 5,1 FS System are therefore applicable to the VITROS 4600 System. Substances that Do Not Interfere The substances listed in the table were tested with VITROS Slides and found not to interfere, bias <38 U/L (0.63 µkat/l), for analyte activities 300 U/L or 12.8% activity>300 U/L, at the concentrations shown. Version 10.0 Pub. No. MP2-42_EN 7 of 11

8 INSTRUCTIONS FOR USE References Compound Concentration Compound Concentration Adenylate kinase 200 U/L 200 U/L Hemoglobin 150 mg/dl 1.5 g/l Ascorbic acid 3 mg/dl 170 µmol/l Intralipid 800 mg/dl 8 g/l Bilirubin 40 mg/dl 684 µmol/l Isoniazid 0.5 mg/dl 36 µmol/l Creatine 15 mg/dl 1 mmol/l L-dopa 0.6 mmol/l 0.6 mmol/l Ethanol 300 mg/dl 65 mmol/l 6-Mercaptopurine 1.5 mg/dl 99 µmol/l Gentisic acid 0.5 mg/dl 32 µmol/l N-Acetylcysteine mg/dl 10.2 mmol/l Glutathione 1 mg/dl 33 µmol/l Salicylic acid 35 mg/dl 3 mmol/l References 1. Tietz NW (ed). Fundamentals of Clinical Chemistry. ed. 3. Philadelphia: WB Saunders; ; CLSI. Protection of Laboratory Workers from Instrument Biohazards and Infectious Diseases Transmitted by Blood, Body Fluids and Tissue; Approved Guideline Fourth Edition; CLSI Document M29-A4, Wayne, PA: Clinical and Laboratory Standards Institute; Doumas BT, et al. Differences Between Values for Plasma and Serum in Tests Performed in the Ektachem 700 XR Analyzer, and Evaluation of Plasma Separator Tubes (PST). Clin. Chem. 35:1: ; Calam RR. Specimen Processing Separator Gels: An Update. J Clin Immunoassay. 11:86-90; Tietz NW (ed). Fundamentals of Clinical Chemistry. ed. 3. Philadelphia: WB Saunders; ; CLSI. Procedures for the Collection of Diagnostic Blood Specimens by Venipuncture; Approved Standard Sixth Edition. NCCLS document H3-A6 (ISBN ). CLSI, 940 West Valley Road, Suite 1400, Wayne, PA USA; NCCLS. Procedures and Devices for the Collection of Diagnostic Capillary Blood Specimens; Approved Standard Fifth Edition. NCCLS document H4-A5 [ISBN ]. CLSI, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania USA, Clinical Laboratory Handbook for Patient Preparation and Specimen Handling. Fascicle VI: Chemistry/Clinical Microscopy. Northfield, IL: College of American Pathologists: Statistical Quality Control for Quantitative Measurements: Principles and Definitions; Approved Guideline Third Edition. CLSI document C24-A3 (ISBN ). CLSI, 940 West Valley Road, Suite 1400, Wayne, PA USA; Tietz NW (ed). Textbook of Clinical Chemistry. ed. 3. Philadelphia: WB Saunders; 662; Scandinavian Committee on Enzymes: Recommended Method for the Determination of in Blood. Scand. J. Clin. Lab. Invest. 39:1-5; Scandinavian Committee on Enzymes: Recommended Method for the Determination of in Blood. Scand. J. Clin. Lab. Invest. 36: ; Henry JB. Clinical Diagnosis and Management by Laboratory Methods. Philadelphia: WB Saunders; 2088; Krahn J. Upper Reference Limit for. Clin. Chem. 31(1):158; Black HR. Quallich H-D, and Garlect CB. Racial Difference in Serum Levels. Amer. J. Med. 81: ; Miller WG. Chinchilli HD, Nance WD. Sampling from a Skewed Population Distribution as Exemplified by Estimation of the Upper Reference Limit. Clin. Chem. 30(1): 18 23; Young DS. Effects of Drugs on Clinical Laboratory Tests. ed. 4. Washington D.C.: AACC Press; Friedman RB. Young DS. Effects of Disease on Clinical Laboratory Tests. Washington, D.C.: AACC Press; CLSI. Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline. CLSI Document EP9. CLSI, 940 West Valley Road, Suite 1400, Wayne, PA USA; NCCLS. Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline. NCCLS document EP9-A2 [ISBN ]. CLSI, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania USA; CLSI. User Evaluation of Precision Performance with Clinical Chemistry Devices. CLSI Document EP5. CLSI, 940 West Valley Road, Suite 1400, Wayne, PA USA; NCCLS. Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline Second Edition. NCCLS document EP5-A2 [ISBN ]. CLSI, 940 West Valley Road, Suite 1400, Wayne, PA USA; NCCLS. Interference Testing In Clinical Chemistry; Proposed Guideline. NCCLS document EP7-P (ISBN ). NCCLS, 940 West Valley Road, Wayne, Pennsylvania , CLSI. Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition. CLSI Document EP7-A2. Wayne, PA: Clinical and Laboratory Standards Institute; of 11 Pub. No. MP2-42_EN Version 10.0

9 Glossary of Symbols Glossary of Symbols Revision History Date of Revision Version Description of Technical Changes* Known Interferences: updated references Specificity: updated references and added N-acetylcysteine References: updated Reactive Ingredients: removed enzyme EC codes and updated enzyme source References: updated M29 Glossary of Symbols: added Globally Harmonized Symbols to comply with the Classification, Labelling and Packaging (CLP) Regulations Prescription Use statement added. Updated EC Representative address. Added USA to legal manufacture address Glossary of Symbols: added Date of Manufacture Glossary of Symbols: updated Added information for the VITROS 4600 Chemistry System Version 10.0 Pub. No. MP2-42_EN 9 of 11

10 INSTRUCTIONS FOR USE Revision History Date of Revision Version Description of Technical Changes* Added information for the VITROS 5600 Integrated System Test Type and Conditions Added statement Method Comparison Added information on sample types References Updated Glossary of Symbols Updated Minor wording and formatting changes Added VITROS 5,1 FS Chemistry System Specimen Requirements, Special Precautions wording update Quality Control updated reference Specificity added Intralipid, updated Bilirubin References updated data Glossary of Symbols updated data New organization and sections consistent with IVD Directive Sample Dilution added that dilution results in higher activities than expected Method comparison updated data for all comparisons Precision updated data for 750 System References added 3, 5, 9, and APR New format, technically equivalent to 11/96. * The change bars indicate the position of a technical amendment to the text with respect to the previous version of the document. When this Instructions For Use is replaced, sign and date below and retain as specified by local regulations or laboratory policies, as appropriate. Signature Obsolete Date 10 of 11 Pub. No. MP2-42_EN Version 10.0

11 Revision History Ortho-Clinical Diagnostics Felindre Meadows Pencoed Bridgend CF35 5PZ United Kingdom Ortho-Clinical Diagnostics, Inc. 100 Indigo Creek Drive Rochester, NY USA VITROS is a registered trademark of Ortho-Clinical Diagnostics, Inc. Ortho-Clinical Diagnostics, Inc., Version 10.0 Pub. No. MP2-42_EN 11 of 11

INSTRUCTIONS FOR USE

INSTRUCTIONS FOR USE INSTRUCTIONS FOR USE VITROS Chemistry Products Fe Slides Fe 151 5808 192 4547 Caution: Rx ONLY Intended Use For in vitro diagnostic use only. VITROS Chemistry Products Fe Slides quantitatively measure

More information

INSTRUCTIONS FOR USE. VITROS Chemistry Products Performance Verifier I and II. Intended Use. Reagents. Warnings and Precautions.

INSTRUCTIONS FOR USE. VITROS Chemistry Products Performance Verifier I and II. Intended Use. Reagents. Warnings and Precautions. VITROS Chemistry Products 806 7324 823 1474 Rx ONLY Intended Use For in vitro diagnostic use only. VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS

More information

Date. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver and/or kidneys.

Date. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver and/or kidneys. SYNCHRON System(s) Chemistry Information Sheet Copyright 2013 Beckman Coulter, Inc. Albumin REF 442765 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by Date Reviewed by Date PRINCIPLE INTENDED USE

More information

CR-E PRINCIPLE REF A ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

CR-E PRINCIPLE REF A ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Creatinine REF A60298 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE INTENDED

More information

Date. Alkaline phosphatase measurements are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.

Date. Alkaline phosphatase measurements are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases. SYNCHRON CX System(s) Chemistry Information Sheet Copyright 2008 Beckman Coulter, Inc. Alkaline Phosphatase Kit Reorder # 442670 (200 tests/cartridge) Kit Reorder # 476821 (400 tests/cartridge) For In

More information

AAG PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

AAG PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE IMMAGE Immunochemistry Systems Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Alpha 1 -Acid Glycoprotein REF 447780 (150 tests) For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by:

More information

CKMB PRINCIPLE REF ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

CKMB PRINCIPLE REF ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Creatine Kinase - MB Isoenzyme REF 445375 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date

More information

For In Vitro Diagnostic Use

For In Vitro Diagnostic Use SYNCHRON System(s) Chemistry Information Sheet Lactate Dehydrogenase REF 442660 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE INTENDED USE reagent, when used in

More information

CREA PRINCIPLE SPECIMEN REF ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

CREA PRINCIPLE SPECIMEN REF ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Creatinine REF 442760 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE INTENDED

More information

PRINCIPLE. REF (200 tests/cartridge) REF (300 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

PRINCIPLE. REF (200 tests/cartridge) REF (300 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. LD Lactate Dehydrogenase REF 442655 (200 tests/cartridge) REF 476841 (300 tests/cartridge) For In Vitro Diagnostic Use

More information

Date. Alkaline phosphatase measurements are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.

Date. Alkaline phosphatase measurements are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases. SYNCHRON System(s) Chemistry Information Sheet Copyright 2013 Beckman Coulter, Inc. Alkaline Phosphatase REF 442670 (200 tests/cartridge) REF 476821 (400 tests/cartridge) For In Vitro Diagnostic Use ANNUAL

More information

APB PRINCIPLE SPECIMEN. REF (300 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

APB PRINCIPLE SPECIMEN. REF (300 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE IMMAGE Immunochemistry Systems Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Apolipoprotein B REF 447730 (300 tests) For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed

More information

PAB PRINCIPLE. Kit Reorder # ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

PAB PRINCIPLE. Kit Reorder # ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE SYNCHRON CX System(s) Chemistry Information Sheet Copyright 2008 Beckman Coulter, Inc. Prealbumin Kit Reorder # 475106 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE

More information

For In Vitro Diagnostic Use

For In Vitro Diagnostic Use SYNCHRON System(s) Chemistry Information Sheet Albumin REF 467858 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Reviewed by: Date Refer to coversheet in front of method PRINCIPLE INTENDED USE

More information

AAT PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

AAT PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE IMMAGE Immunochemistry Systems Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Alpha 1 -Antitrypsin REF 447740 (150 tests) For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed

More information

M-TP PRINCIPLE REF ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

M-TP PRINCIPLE REF ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Microprotein REF 445860 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE INTENDED

More information

PRINCIPLE SPECIMEN. REF (300 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

PRINCIPLE SPECIMEN. REF (300 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE IMMAGE Immunochemistry Systems Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Complement REF 446450 (300 tests) For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by:

More information

ALT PRINCIPLE. REF (200 tests/cartridge) REF (400 tests/cartridge) ANNUAL REVIEW Reviewed by: Date.

ALT PRINCIPLE. REF (200 tests/cartridge) REF (400 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Alanine Aminotransferase REF 442620 (200 tests/cartridge) REF 476826 (400 tests/cartridge) For In Vitro Diagnostic Use

More information

DBIL PRINCIPLE SPECIMEN. Kit Reorder # (200 tests/cartridge) Kit Reorder # (300 tests/cartridge) ANNUAL REVIEW Reviewed by: Date.

DBIL PRINCIPLE SPECIMEN. Kit Reorder # (200 tests/cartridge) Kit Reorder # (300 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. SYNCHRON CX System(s) Chemistry Information Sheet Copyright 2008 Beckman Coulter, Inc. Direct Bilirubin Kit Reorder # 439715 (200 tests/cartridge) Kit Reorder # 476856 (300 tests/cartridge) For In Vitro

More information

For In Vitro Diagnostic Use

For In Vitro Diagnostic Use SYNCHRON System(s) Chemistry Information Sheet TG Triglycerides GPO REF 445850 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Reviewed by: Date Refer to coversheet in front of method PRINCIPLE

More information

PRINCIPLE REF ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

PRINCIPLE REF ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. T-Uptake REF 445999 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE INTENDED

More information

For In Vitro Diagnostic Use

For In Vitro Diagnostic Use SYNCHRON System(s) Chemistry Information Sheet C-Reactive Protein REF (200 tests/cartridge) 465131 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE INTENDED USE reagent,

More information

PRINCIPLE SPECIMEN REF ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

PRINCIPLE SPECIMEN REF ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2013 Beckman Coulter, Inc. Microalbumin REF 475100 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by Date Reviewed by Date PRINCIPLE INTENDED

More information

IGALC PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

IGALC PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE IMMAGE Immunochemistry Systems Chemistry Information Sheet Low Concentration Immunoglobulin A Copyright 2010 Beckman Coulter, Inc. REF 447460 (150 tests) For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed

More information

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date SYNCHRON System(s) Chemistry Information Sheet 2017 Beckman Coulter, Inc. All rights reserved. TG Triglycerides GPO 445850 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by

More information

LAM PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

LAM PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE IMMAGE Immunochemistry Systems Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Lambda Light Chain REF 446470 (150 tests) For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed

More information

Fibrinogen Universal Kit

Fibrinogen Universal Kit Order references Reagents CONT FITUR-B00 Universal kit 1 x 50 ml R1 + 1 x 5 ml R2 FITUR-00 Universal kit 2 x 50 ml R1 + 1 x 15 ml R2 FITUR-C00 Universal kit 2 x 17 ml R1 + 1 x 4 ml R2 Other necessary products

More information

This package insert contains information to run the Albumin BCP assay on the ARCHITECT c Systems and the AEROSET System.

This package insert contains information to run the Albumin BCP assay on the ARCHITECT c Systems and the AEROSET System. ALBUMIN BCP 7D54-21 304320/R1 ALBUMIN BCP This package insert contains information to run the Albumin BCP assay on the ARCHITECT c Systems and the AEROSET System. NOTE: This package insert must be read

More information

This package insert contains information to run the Albumin BCG assay on the ARCHITECT c Systems and the AEROSET System.

This package insert contains information to run the Albumin BCG assay on the ARCHITECT c Systems and the AEROSET System. ALBUMIN BCG 7D53-20 30-3927/R6 ALBUMIN BCG This package insert contains information to run the Albumin BCG assay on the ARCHITECT c Systems and the AEROSET System. NOTE: Changes Highlighted NOTE: This

More information

Laboratory Procedure Manual

Laboratory Procedure Manual Laboratory Procedure Manual Analyte: Matrix: Method: Cholesterol Refrigerated Serum Beckman Synchron LX20 Method No.: Revised: as performed by: Collaborative Laboratory Services, L.L.C Contact: Important

More information

Fibrinogen Universal Kit

Fibrinogen Universal Kit Order references Reagents CONT FITUR-B00 Universal kit 1 x 50 ml R1 + 1 x 5 ml R2 FITUR-00 Universal kit 2 x 50 ml R1 + 1 x 15 ml R2 FITUR-C00 Universal kit 2 x 17 ml R1 + 1 x 4 ml R2 Other necessary products

More information

Complement C4 Universal Kit

Complement C4 Universal Kit Order references Reagents CONT C4TUR-B00 Universal kit 1 x 50 ml R1 + 1 x 5 ml R2 C4TUR-00 Universal kit 2 x 50 ml R1 + 1 x 15 ml R2 C4TUR-C00 Universal kit 1 x 12 ml R1 + 1 x 2 ml R2 Other necessary products

More information

FER PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

FER PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE IMMAGE Immunochemistry Systems Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Ferritin REF 447160 (150 tests) For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by:

More information

PRINCIPLE REF ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

PRINCIPLE REF ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2013 Beckman Coulter, Inc. T-Uptake REF 445999 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by Date Reviewed by Date PRINCIPLE INTENDED USE

More information

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date SYNCHRON System(s) Chemistry Information Sheet 2014 Beckman Coulter, Inc. All rights reserved. GLU Glucose 442640 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by Date PRINCIPLE

More information

WORTHAM LABORATORIES, INC. DRAFT Thrombin Reagent

WORTHAM LABORATORIES, INC. DRAFT Thrombin Reagent WORTHAM LABORATORIES, INC. DRAFT Thrombin Reagent Intended Use Wortham Laboratories Thrombin Reagent is intended for thrombin to convert fibrinogen in the quantitative determination of fibrinogen in plasma

More information

Immunoglobulin A Universal Kit

Immunoglobulin A Universal Kit Order references Reagents CONT IATUR-B00 Universal kit 1 x 50 ml R1 + 1 x 5 ml R2 IATUR-00 Universal kit 2 x 50 ml R1 + 1 x 15 ml R2 IATUR-C00 Universal kit 2 x 17 ml R1 + 1 x 4 ml R2 Other necessary products

More information

Application Note for Gentian Cystatin C Immunoassay on. ABX Pentra 400 1, Horiba Medical 1 v03en-june 2011

Application Note for Gentian Cystatin C Immunoassay on. ABX Pentra 400 1, Horiba Medical 1 v03en-june 2011 Page 1 of 5 Application Note for Gentian Cystatin C Immunoassay on For in vitro diagnostic use only ABX Pentra 400 1, Horiba Medical 1 v03en-june 2011 Intended Use The Gentian Cystatin C Immunoassay is

More information

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date IMMAGE Immunochemistry Systems Chemistry Information Sheet 2017 Beckman Coulter, Inc. All rights reserved. LAM Lambda Light Chain 446470 (150 tests) For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed

More information

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date SYNCHRON System(s) Chemistry Information Sheet 2017 Beckman Coulter, Inc. All rights reserved. TOB Tobramycin 467983 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by Date

More information

For In Vitro Diagnostic Use

For In Vitro Diagnostic Use SYNCHRON System(s) Chemistry Information Sheet Tobramycin 467983 For In Vitro Diagnostic Use PRINCIPLE INTENDED USE reagent, when used in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems

More information

HELICA BIOSYSTEMS, INC. HIGH SENSITIVITY HUMAN C-REACTIVE PROTEIN FOR RESEARCH USE ONLY (Not for in vitro diagnostic use)

HELICA BIOSYSTEMS, INC. HIGH SENSITIVITY HUMAN C-REACTIVE PROTEIN FOR RESEARCH USE ONLY (Not for in vitro diagnostic use) INTENDED USE HELICA BIOSYSTEMS, INC. HIGH SENSITIVITY HUMAN C-REACTIVE PROTEIN FOR RESEARCH USE ONLY (Not for in vitro diagnostic use) The Helica C-reactive protein assay is intended for the detection

More information

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date SYNCHRON System(s) Chemistry Information Sheet 2015 Beckman Coulter, Inc. All rights reserved. TBIL Total Bilirubin 442745 (300 tests/cartridge) 476861 (400 tests/cartridge) For In Vitro Diagnostic Use

More information

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date SYNCHRON System(s) Chemistry Information Sheet 2015 Beckman Coulter, Inc. All rights reserved. C3 Complement C3 988462 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by Date

More information

S1725 Method Validation and Verification: Case Studies and Laboratory Challenges

S1725 Method Validation and Verification: Case Studies and Laboratory Challenges S1725 Method Validation and Verification: Case Studies and Laboratory Challenges James H. Nichols, PhD, D(ABCC) Lauren N. Person, DO, MPH, FCAP Objectives Identify the difference between method validation

More information

For In Vitro Diagnostic Use

For In Vitro Diagnostic Use SYNCHRON System(s) Chemistry Information Sheet Gentamicin 469137 For In Vitro Diagnostic Use PRINCIPLE INTENDED USE reagent, when used in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems

More information

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date SYNCHRON System(s) Chemistry Information Sheet 2015 Beckman Coulter, Inc. All rights reserved. FE Iron 467910 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by Date PRINCIPLE

More information

HELICA BIOSYSTEMS, INC. MOUSE C-REACTIVE PROTEIN QUANTITATION BY ELISA FOR RESEARCH USE ONLY

HELICA BIOSYSTEMS, INC. MOUSE C-REACTIVE PROTEIN QUANTITATION BY ELISA FOR RESEARCH USE ONLY HELICA BIOSYSTEMS, INC. MOUSE C-REACTIVE PROTEIN QUANTITATION BY ELISA FOR RESEARCH USE ONLY INTENDED USE The Helica C-reactive protein assay is intended for the detection and quantification of mouse C-reactive

More information

MOUSE ALPHA 1-ANTITRYPSIN CATALOG NUMBER: OKIA00088

MOUSE ALPHA 1-ANTITRYPSIN CATALOG NUMBER: OKIA00088 MOUSE ALPHA 1-ANTITRYPSIN CATALOG NUMBER: OKIA00088 Immunoperoxidase Assay for Determination of Alpha 1-Antitrypsin in Mouse Samples DIRECTIONS FOR USE Version 4.0 Please Read this Package Insert Completely

More information

DNB PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

DNB PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE IMMAGE Immunochemistry Systems Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Antideoxyribonuclease B REF 447480 (150 tests) For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date

More information

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date SYNCHRON System(s) Chemistry Information Sheet 2015 Beckman Coulter, Inc. All rights reserved. VPA Valproic Acid 467995 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by Date

More information

For In Vitro Diagnostic Use

For In Vitro Diagnostic Use SYNCHRON System(s) Chemistry Information Sheet Vancomycin REF 474824 For In Vitro Diagnostic Use ANNUAL REVIEW Refer to coversheet in front of method PRINCIPLE INTENDED USE reagent, when used in conjunction

More information

INSTRUCTIONS FOR USE

INSTRUCTIONS FOR USE VITROS Immunodiagnostic Products CA 19-9 Reagent Pack VITROS Immunodiagnostic Products CA 19-9 Calibrators 680 2241 680 0035 Caution: Patients known to be genotypically negative for the Lewis blood group

More information

MOUSE SERUM AMYLOID P (SAP) CATALOG NUMBER: OKIA00113

MOUSE SERUM AMYLOID P (SAP) CATALOG NUMBER: OKIA00113 MOUSE SERUM AMYLOID P (SAP) CATALOG NUMBER: OKIA00113 Immunoperoxidase Assay for Determination of SAP in Mouse Samples DIRECTIONS FOR USE Version 4.0 Please Read this Package Insert Completely Before Using

More information

Guide to Fulfillment of Validation and Verification of Examination Requirements

Guide to Fulfillment of Validation and Verification of Examination Requirements DIAGNOSTIC ACCREDITATION PROGRAM College of Physicians and Surgeons of British Columbia 300 669 Howe Street Telephone: 604-733-7758 ext. 2635 Vancouver BC V6C 0B4 Toll Free: 1-800-461-3008 (in BC) www.cpsbc.ca

More information

TNF-α ELISA, Mouse. Cat. No. KT-129 K-ASSAY. For the quantitative determination of TNF-α in mouse serum and culture media. For research use only.

TNF-α ELISA, Mouse. Cat. No. KT-129 K-ASSAY. For the quantitative determination of TNF-α in mouse serum and culture media. For research use only. K-ASSAY TNF-α ELISA, Mouse For the quantitative determination of TNF-α in mouse serum and culture media Cat. No. KT-129 For research use only. 1 Rev. 052708 K-ASSAY PRODUCT INFORMATION TNF-α ELISA, Mouse

More information

CANINE NGAL CATALOG NUMBER: OKIA00031

CANINE NGAL CATALOG NUMBER: OKIA00031 CANINE NGAL CATALOG NUMBER: OKIA00031 Immunoperoxidase Assay for Determination of NGAL in Dog Samples DIRECTIONS FOR USE Version 4.0 Please Read this Package Insert Completely Before Using This Product

More information

CHICKEN IgY CATALOG NUMBER: OKIA00018

CHICKEN IgY CATALOG NUMBER: OKIA00018 CHICKEN IgY CATALOG NUMBER: OKIA00018 Immunoperoxidase Assay for Determination of IgY in Chicken Samples DIRECTIONS FOR USE Version 4.0 Please Read this Package Insert Completely Before Using This Product

More information

IV3-105E. High Sensitivity CRP (hscrp) ELISA Kit. Definitions LOT REF. English. For in-vitro diagnostic use only IVD. Instructions for use

IV3-105E. High Sensitivity CRP (hscrp) ELISA Kit. Definitions LOT REF. English. For in-vitro diagnostic use only IVD. Instructions for use IV3-105E Definitions Instructions for use REF Catalogue number English High Sensitivity CRP (hscrp) ELISA Kit For in-vitro diagnostic use only Σ 96 LOT IVD Σ N Use by Lot/Batch Code Storage temperature

More information

ImmuLux Human IL-6 Fluorescent ELISA Kit

ImmuLux Human IL-6 Fluorescent ELISA Kit INSTRUCTIONS ImmuLux Human IL- Fluorescent ELISA Kit 3747 N. Meridian Road P.O. Box 7 Rockford, IL 05 840 53w Product Description Number Description 840 ImmuLux Human IL- Fluorescent ELISA Kit The Pierce

More information

Cyfra 21-1 IRMA. Product information Information about other products is available at: Userś Manual DE52100

Cyfra 21-1 IRMA. Product information Information about other products is available at:  Userś Manual DE52100 Product information Information about other products is available at: www.demeditec.com Userś Manual Cyfra 21-1 IRMA The CYFRA 21.1 IRMA system provides a direct in vitro quantitative determination of

More information

Fentanyl (Human) ELISA Kit

Fentanyl (Human) ELISA Kit Fentanyl (Human) ELISA Kit Catalog Number KA0931 96 assays Version: 06 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of the

More information

For the quantitative determination of Complement Factor 3 in guinea pig serum and plasma

For the quantitative determination of Complement Factor 3 in guinea pig serum and plasma For the quantitative determination of Complement Factor 3 in guinea pig serum and plasma For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 41-CO3GP-E01 Size: 96 wells Version:

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k100499 B. Purpose for Submission: New Device C. Measurand: Rheumatoid Factors IgG, IgM and IgA and Rheumatoid

More information

IGE PRINCIPLE. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

IGE PRINCIPLE. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE IMMAGE Immunochemistry Systems Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Total Immunoglobulin E REF 474620 (150 tests) For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date

More information

For In Vitro Diagnostic Use

For In Vitro Diagnostic Use SYNCHRON System(s) Chemistry Information Sheet Digoxin REF 650182 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Reviewed by: Date Refer to coversheet in front of method PRINCIPLE INTENDED USE

More information

H-ferritin (Human) ELISA Kit

H-ferritin (Human) ELISA Kit H-ferritin (Human) ELISA Kit Catalog Number KA0211 96 assays Version: 04 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of

More information

HORSE IgG CATALOG NUMBER: OKIA00043

HORSE IgG CATALOG NUMBER: OKIA00043 HORSE IgG CATALOG NUMBER: OKIA00043 Immunoperoxidase Assay for Determination of IgG in Horse Samples DIRECTIONS FOR USE Version 4.0 Please Read this Package Insert Completely Before Using This Product

More information

Instructions for use. Anti TPO ELISA TF E-3600

Instructions for use. Anti TPO ELISA TF E-3600 Instructions for use Anti TPO ELISA TF E-3600 Anti-TPO ELISA 1. Intended use and principle of the test Enzyme Immunoassay for the quantitative determination of anti-thyroid Peroxidase antibodies in human

More information

Human Myoglobin ELISA Kit

Human Myoglobin ELISA Kit Human Myoglobin ELISA Kit Catalog No: IHMYGKT-1 Lot No: SAMPLE INTENDED USE The Myoglobin test kits are a highly sensitive two-site enzyme linked immunoassay (ELISA) for measuring Myoglobin in biological

More information

Human Beta 2-Microglobulin ELISA Kit

Human Beta 2-Microglobulin ELISA Kit Human Beta 2-Microglobulin ELISA Kit CATALOG NO: IRKTAH1125 LOT NO: SAMPLE INTENDED USE The Beta 2-Microglobulin test kit is a highly sensitive two-site enzyme linked immunoassay (ELISA) for measuring

More information

Immunoperoxidase Assay for Determination of C3 in Guinea Pig Sera Lot # 2F

Immunoperoxidase Assay for Determination of C3 in Guinea Pig Sera Lot # 2F GUINEA PIG C3 ELISA Immunoperoxidase Assay for Determination of C3 in Guinea Pig Sera 40-374-130032 Lot # 2F Directions for Use For Research Use Only For In Vitro Use Only GUINEA PIG COMPLEMENT FACTOR

More information

Luteinizing Hormone ELISA kit

Luteinizing Hormone ELISA kit 55R-IB19104 Luteinizing Hormone ELISA kit Enzyme Immunoassay for the determination of LH in human serum 1. Principle of the test: The LH ELISA kit is a solid phase direct sandwich method. The samples and

More information

IgG1 (Mouse) ELISA Kit

IgG1 (Mouse) ELISA Kit IgG1 (Mouse) ELISA Kit Catalog Number KA1011 96 assays Version: 05 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 Principle of the Assay... 3 General

More information

Sheep IgG ELISA Kit. Innovative Research, Inc. Catalog No: ISHIGGKT. Lot No: SAMPLE

Sheep IgG ELISA Kit. Innovative Research, Inc. Catalog No: ISHIGGKT. Lot No: SAMPLE Catalog No: ISHIGGKT Sheep IgG ELISA Kit Lot No: SAMPLE INTENDED USE The total Sheep IgG test kits are a highly sensitive two-site enzyme linked immunoassay (ELISA) for measuring IgG in sheep biological

More information

Laboratory Procedure Manual

Laboratory Procedure Manual Laboratory Procedure Manual Analyte: Matrix: Method: Total Iron Binding Capacity (TIBC) Serum Beckman Synchron LX20 Method No.: Revised: as performed by: Collaborative Laboratory Services Ottumwa, Iowa

More information

Troponin I (Rat) ELISA Kit

Troponin I (Rat) ELISA Kit Troponin I (Rat) ELISA Kit Catalog Number KA4341 96 assays Version: 01 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 General Information...

More information

Manual. Catalog number:

Manual. Catalog number: LLAMA IGG ELISA Quantitation Kit Manual Catalog number: 40-374-130033 Immunoperoxidase Assay for Determination of Llama IgG in Llama serum/plasma. This kit is for research use only, and is not for use

More information

AssayMax Human Transferrin ELISA Kit

AssayMax Human Transferrin ELISA Kit AssayMax Human Transferrin ELISA Kit Assaypro LLC 3400 Harry S Truman Blvd St. Charles, MO 63301 T (636) 447-9175 F (636) 395-7419 www.assaypro.com For any questions regarding troubleshooting or performing

More information

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date SYNCHRON System(s) Chemistry Information Sheet 2014 Beckman Coulter, Inc. All rights reserved. PHY Phenytoin 469188 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by Date PRINCIPLE

More information

Francesca Di Serio 1, G. Amodio 2, Mario Caputo 3, Martina Zaninotto 4, Cosimo Ottomano 5

Francesca Di Serio 1, G. Amodio 2, Mario Caputo 3, Martina Zaninotto 4, Cosimo Ottomano 5 Francesca Di Serio 1, G. Amodio 2, Mario Caputo 3, Martina Zaninotto 4, Cosimo Ottomano 5 1 Department of Clinical Pathology I, University Hospital, Bari, Italy 2 Emergency Cardiology Department, University

More information

Estradiol (Bovine) ELISA Kit

Estradiol (Bovine) ELISA Kit Estradiol (Bovine) ELISA Kit Catalog Number KA2276 96 assays Version: 04 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Principle of the Assay... 3

More information

Guinea Pig IgA ELISA Kit

Guinea Pig IgA ELISA Kit CATALOG NO: IGPIGAKT Guinea Pig IgA ELISA Kit LOT NO: SAMPLE INTENDED USE The IgA test kits are a highly sensitive two-site enzyme linked immunoassay (ELISA) for measuring IgA in biological samples of

More information

Human Collagen IV ELISA, Urine

Human Collagen IV ELISA, Urine K-ASSAY Human Collagen IV ELISA, Urine For the quantitative determination of human Type IV Collagen in urine Cat. No. KT-036 For Research Use Only. Not for Use in Diagnostic Procedures. - 1 - Rev. 11152016

More information

B2M (Human) ELISA Kit

B2M (Human) ELISA Kit B2M (Human) ELISA Kit Catalog Number KA0222 96 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of the Assay...

More information

-ECVAM Skin Irritation Validation Study-

-ECVAM Skin Irritation Validation Study- Version : 1.2 Page 1 of 16 September 2005 EPIDERMIS MODEL EPISKIN -ECVAM Skin Irritation Validation Study- VALIDATION OF THE EPISKIN SKIN IRRITATION TEST - 42 HOURS ASSAY FOR THE PREDICTION OF ACUTE SKIN

More information

HUMAN IgG. Immunoperoxidase Assay for Determination of IgG in Human Samples

HUMAN IgG. Immunoperoxidase Assay for Determination of IgG in Human Samples HUMAN IgG Immunoperoxidase Assay for Determination of IgG in Human Samples DIRECTIONS FOR USE Version3 L30A 31 For Research Use Only, NOT for Diagnostic Purposes Please Read this Package Insert Completely

More information

Qualitative determination of Anti-Treponema pallidum antibodies For in vitro use only Store at 2-8 C

Qualitative determination of Anti-Treponema pallidum antibodies For in vitro use only Store at 2-8 C Qualitative determination of Anti-Treponema pallidum antibodies For in vitro use only Store at 2-8 C INTRODUCTION Syphilis is a venereal disease caused by infection with T. pallidum. The microorganism

More information

Canine Fibrinogen ELISA

Canine Fibrinogen ELISA Canine Fibrinogen ELISA For the quantitative determination of Fibrinogen in canine serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 41-FIBCA-E01 Size: 96 wells

More information

DRG Anti-Tissue Transglutaminase

DRG Anti-Tissue Transglutaminase Please use only the valid version of the package insert provided with the kit. This kit is intended for Research Use Only. Not for use in diagnostic procedures. NAME AND INTENDED USE Anti-Tissue-Transglutaminase

More information

CANINE CRP ELISA KIT

CANINE CRP ELISA KIT CANINE CRP ELISA KIT CATALOG NO: IDGCRPKT LOT NO: SAMPLE INTENDED USE The CRP test kits are a highly sensitive two-site enzyme linked immunoassay (ELISA) for measuring CRP in Dog biological samples. INTRODUCTION

More information

TF (Human) ELISA Kit. Catalog Number KA assays Version: 08. Intended for research use only.

TF (Human) ELISA Kit. Catalog Number KA assays Version: 08. Intended for research use only. TF (Human) ELISA Kit Catalog Number KA0510 96 assays Version: 08 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 Principle of the Assay... 3 General Information...

More information

Fibrinogen ELISA. For the quantitative determination of fibrinogen in biological fluids, serum, and plasma.

Fibrinogen ELISA. For the quantitative determination of fibrinogen in biological fluids, serum, and plasma. Fibrinogen ELISA For the quantitative determination of fibrinogen in biological fluids, serum, and plasma. Please read carefully due to Critical Changes, e.g., Changes to preparation of standard volume

More information

Transferrin ELISA. For the quantitative determination of transferrin in human biological samples

Transferrin ELISA. For the quantitative determination of transferrin in human biological samples Transferrin ELISA For the quantitative determination of transferrin in human biological samples Please see Appendix A for Reference Serum information. For Research Use Only. Not For Use In Diagnostic Procedures.

More information

Mouse Collagen Type III ELISA

Mouse Collagen Type III ELISA Mouse Collagen Type III ELISA Catalog Number M046062 For the quantitative determination of Collagen Type III in mouse plasma and serum samples. For research use only. This product insert must be read in

More information

PictArray ToRCH ELISA Kit

PictArray ToRCH ELISA Kit TORCH Instructions PictArray ToRCH ELISA Kit Instructions for testing human serum samples on PictArray ToRCH Panel consisting of Toxoplasma gondii, Rubella, Cytomegalovirus (CMV), Herpes Simplex Virus

More information

Rat Hemoglobin ELISA

Rat Hemoglobin ELISA Rat Hemoglobin ELISA For the quantitative determination of Hemoglobin in rat serum and plasma For Research use Only. Not For Use In Diagnostic Procedures. Please See Appendix A for Reference Serum Information.

More information