Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016

Size: px
Start display at page:

Download "Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016"

Transcription

1 Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, Method background: This method employs full evaporation technique headspace analysis (FET HS) combined with GC MS to determine levels of residual solvents in cannabis concentrates. This method has undergone a partial validation, including precision/accuracy, lower limit of detection (LLOD)/ lower limit of quantification (LLOQ), repeatability, limited stability, and matrix effects. Interday/batch and ruggedness/robustness validation was not performed. Although a partial validation was performed on this method, this partial validation shall in no way affect the need for in house validation when implementing this method. This method was validated for the following levels of the following residual solvents: Compound LLOQ ULOQ Propane Isobutane Butane Methanol Pentane Ethanol Acetone Isopropanol Hexane Chloroform Benzene Heptane Toluene m/p xylene (sum of isomers) o xylene Differing ranges were chosen for different analytes based on the dynamic range of the MS analyzer as well as expected/acceptable levels of each solvent. Solvents with higher toxicological significance are quantified in a lower concentration range and vice versa. The USP Class 1 and 2 residual solvents included in this list have been validated for quantification at or below the PDE for these solvents (assuming a 2g/day dose) as listed in the current version of USP <467> (USP 38 NF 33). FID detection may be employed for residual solvents analysis, however this method has not been validated using FID detection.

2 Page 2 The FET approach was chosen for this method because it eliminates matrix effects present when using traditional (e.g. dilution approach) headspace methods. The theory behind FET HS is that a single, non partitioning gas phase system is established in the headspace vial through the use of very small sample sizes 1. The establishment of a non adsorptive system was verified using multiple headspace extraction (MHE) experiments during method validation, however additional MHE experiments must be performed to evaluate matrix effects if any from cannabis concentrate matrices. 2.0 Standard, Sample, and QC Preparation: This method employs a six point standard curve. Due to the differing validation ranges for different analytes, each analyte will be quantified against a five point standard curve (i.e. lowlevel analytes will exclude the high calibrator and high level analytes will exclude the low calibrator). Internal standard (IS) normalization is employed in order to improve quantitative accuracy. For this preliminary work, ethylbenzene was used as the internal standard, however due to the possible presence of ethylbenzene in samples containing xylenes, ethylbenzene will not be an acceptable IS for established laboratory methods. Chlorobenzene is preliminarily recommended as a suitable replacement internal standard. Batch accuracy is validated through duplicate matrix spikes at low, mid, and high levels. Due to the different validation ranges for the analytes being quantified, two QC levels will be included for each analyte: low level analytes will be evaluated against QC low (QCL) and QC mid (QCM), while high level analytes will be evaluated against QCM and QC high (QCH). All dilutions may be performed in dimethylsulfoxide (DMSO) or dimethylacetamide (DMA) of sufficient purity to ensure no co eluting contaminant peaks. A blank of the dilution solvent is prepared for each batch to verify dilution solvent cleanliness. It is strongly recommended that DMA be used as a dilution solvent since it does not degrade when subjected to heat during sample incubation and remains liquid at refrigerator temperatures. Single and double blanks are prepared to evaluate carryover. Due to the high volatility of the gas standards, the calibration curve should be prepared just prior to injection on the instrument. Gases present in the samples are more stable than gases in dilution solvent, so stability of gases in samples is less of a concern. The order of operations for sample/standard/qc preparation should be as follows: Prepare internal standard solution Prepare samples Prepare blanks Prepare dilutions for calibration curve and QCs Prepare calibration curve Prepare QCs Analyze immediately

3 Page Internal Standard Preparation*: The stock solution for the internal standard is ethylbenzene at 1840µg/mL in DMSO (Restek cat# 36296). Prior to preparing the internal standard, allow the stock standard to thaw at room temperature. Gentle (< 40 C) warming may be employed to thaw internal standard stock solution. Deliver 920µL of dilution solvent to an autosampler vial. Deliver 80µL of the stock solution to the autosampler vial, then cap and shake to mix. Expiration of the diluted internal standard (IS) will be set at 2 months. Concentration/commodity levels for IS are listed below: stock level Final conc Level in Compound Amt used (µl) Amt in vial (µg) (µg/ml) (µg/ml) commodity Ethylbenzene * It is highly recommended to use chlorobenzene or similar compound as the internal standard in place of ethylbenzene, due to the potential of ethylbenzene to be present in samples contaminated with xylenes. If chlorobenzene is used as internal standard, an 1840µg/ml solution may be prepared in dimethylacetamide and used in place of the ethylbenzene standard listed above. Chlorobenzene elutes between toluene and m/pxylene under the instrument conditions listed in this protocol. 2.2 Sample Preparation: Weigh 20mg of each sample into a headspace vial Add 5µL of internal standard solution and cap immediately Weigh out six additional 20mg aliquots from one sample, and cap loosely. These will be the QC samples. 2.3 Blank Preparation: Prepare three double blanks by capping empty headspace vials Prepare three single blanks by adding 5µL of internal standard solution to headspace vials and cap immediately. Prepare one dilution solvent blank by adding 15µL dilution solvent and 5µL of internal standard solution to a headspace vial and cap immediately. 2.4 Calibration Curve Preparation: Work with all stocks and dilutions cold. Use a cold block or other suitable method (e.g. frozen peas) to keep the temperature of all vials as low as possible. Do not place solutions comprised of DMSO in cooling apparatus, since this will cause the contents to freeze. The instructions below assume the use of one mixed gas standard with each gas at 1000ppm and one mixed solvent standard with each component at 1000ppm in dimethylacetamide.

4 Page 4 Stock solutions: It is highly recommended to use a separate gas stock standard containing gases (e.g. propane butane) only, since the gases are highly unstable. To ensure quantitative accuracy, gas stock solutions should be considered single use only. Immediately after opening stock ampules, transfer contents to empty autosampler vials using a Pasteur pipet and cap. Do not mix stock standards. Dilution 1: Add 800µL of dilution solvent (preferably DMA) to an autosampler vial. Add 100 µl of each stock solution. Cap and shake to mix. Dilution 2: Add 950µL of dilution solvent (preferably DMA) to an autosampler vial. Add 50µL of Dilution 1. Cap and shake to mix. Diluton 3: Add 800µL of dilution solvent to an autosampler vial. Add 200µL of Dilution 2. Concentrations of stock and dilutions are listed below. Stock: 1000µg/mL Dilution 1: 100µg/mL Dilution 2: 5µg/mL Dilution 3: 1µg/mL After all dilutions have been prepared, prepare the calibration curve as shown in the table below, adding 5µL of internal standard to each vial. Note that only one aliquot is required from the dilutions, but separate aliquots from the gas and solvents stock solutions are required for the 250 and 500ppm levels. Level (ppm in commodity) µg in vial Solution Used µl reqd Dilution Dilution Dilution Dilution Stocks Stocks QC Sample Preparation: Prepare duplicate low, mid, and high QC samples using the stocks and dilutions used for the calibration curve as shown in the table below, preferably using second source standards. Also add 5µL of internal standard to each QC sample and cap immediately after delivering the solutions: Level (ppm in commodity) µg in vial Solution Used µl reqd LQC Dilution 2 8 MQC Dilution 1 7 HQC Stocks 4

5 Page Instrumental Analysis For the validation work for this method, analysis was carried out using a GC MS instrument coupled to a loop style headspace autosampler. If an FID instrument is used, all parameters may be kept the same, and the FID should be operated at 310 C. 3.1 Instrument Conditions GC Conditions: Column: Rxi 624Sil MS 30m x 0.25mm x 1.4µm (cat# 13868) Carrier: 80cm/sec (4.95mL/min, C), constant linear velocity Inlet: Shimadzu HS 20 transfer line/valve, 10:1 split Oven Program: 35 C (hold 1.5 min) to C/min (hold 10 min) Headspace Conditions: Oven temp: 140 C Sample line temp: 160 C Transfer line temp: 160 C Shaking: off Multi injection count: 1 Pressurizing Gas Pressure: 25.0psi Equilibrating Time: min Pressurizing time: 2.00 min Pressure Equilib Time: 0.10 min Load Time: 0.20 min Load Equilib Time: 0.10 min Injection Time: 1.00 min Needle Flush Time: 2.00 min GC Cycle Time: min MS Parameters: Source Temp: 280 C Interface Temp: 310 C Solvent Cut Time: 0 min Micro scan width: 0µ Detector voltage (relative to tuning result): 0kV Threshold: 0 Scan program: min, m/z, event time= 0.20 sec 3.2 Batch Setup The order of samples in the batch can be similar to this, although any order is acceptable.

6 Page 6 Sample Name Double blank Single blank Dilution solvent blank 0.5 ppm standard 2.5 ppm standard 25 ppm standard 50 ppm standard 250 ppm standard 500 ppm standard LQC1 MQC1 HQC 1 Single blank Samples Single blank Double blank LQC2 MQC2 HQC2 Double blank 4.0 Quantification All samples and QCs shall be quantified against the calibration curve that was run in the same batch. Analyte confirmation is performed using ion ratios with one target and one reference ion. For FID analyses, it is strongly recommended to perform confirmatory analyses on a separate column with orthogonal selectivity. Based on preliminary work not yet published, the 30m x 0.25mm x 1.00µm Rtx 1 (Restek cat# 10153) is recommended for confirmatory analyses. 5.0 System Suitability All system suitability parameters shall be met prior to reporting results. 5.1 System suitability criteria Linearity: r^2 values for all calibration curves are 0.97 Accuracy: Quantified values for all analytes in QC samples fall within 20% of nominal concentration. If the sample used for QC is shown to contain any solvents 10% of the LLOQ, these analytes will not be included in the pass/fail for QC sample accuracy. Precision: Quantified values for duplicate QCs vary by 20% Carryover: Single blanks do not contain any peaks that quantify 10% of the LLOQ

7 Page 7 Dilution solvent cleanliness: Dilution solvent does not contain any peaks that quantify 10% of the LLOQ All analyte ion ratios must fall within ±30 % of reference ratios from calibrators. Due to the lack of usable fragments for many analytes, only one qualifier ion was used during validation. 6.0 Report Samples Report only values that fall within the calibration curve. If an analyte quantifies above/below the calibration curve, it will be reported as either above (> ULOQ) or below (< LLOQ) the limit of quantification. If any other solvent is identified with a spectral similarity of > 90%, it may be reported qualitatively for informational purposes only. References: 1. Kolb, B; Ettre, L. Static Headspace Gas Chromatography: Theory and Practice, 2 nd ed.; John Wiley & Sons, Inc: Hoboken, NJ, 2006; p For demonstration only. All parameters are suggestions.

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools Application Note Abstract The Commonwealth of Massachusetts Department of Environmental Protection

More information

Residual Solvents in medical cannabis analysis using the SRI 8610C FID GC June 2014

Residual Solvents in medical cannabis analysis using the SRI 8610C FID GC June 2014 June 2014 The SRI Medical Herb Potency 8610C GC is shown at right. This GC can also be used to test for residual solvents (i.e. butane, acetone, gasoline residue, etc.) in medical cannabis. These solvents

More information

Determination of Volatile Organic Compounds (VOCs) present in the interiors of car by using GCMS/MS with static and dynamic headspace

Determination of Volatile Organic Compounds (VOCs) present in the interiors of car by using GCMS/MS with static and dynamic headspace PO-CON44E Determination of Volatile Organic Compounds (VOCs) present in the interiors of car by using GCMS/MS ASMS 20 ThP 138 Sanket Chiplunkar, Ankush Bhone, Durvesh Sawant, Dheeraj Handique, Prashant

More information

Analysis of styrene leached from polystyrene cups using GCMS coupled with Headspace (HS) sampler

Analysis of styrene leached from polystyrene cups using GCMS coupled with Headspace (HS) sampler PO-CON464E Analysis of styrene leached from polystyrene cups using GCMS coupled with Headspace (HS) sampler ASMS 204 TP763 Ankush Bhone (), Dheeraj Handique (), Prashant Hase (), Sanket Chiplunkar (),

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis Cannabigerol (CBG), Primary Standard C-141 Page 1 of 7 Catalog Number: C-141 Lot: Retest: Description: October 2015 Cannabigerol in Methanol Packaging:

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis S(-)-Nicotine, Primary Standard N-008 Page 1 of 7 Catalog Number: N-008 H Lot: Expiration: October 2019 N Description: S(-)-Nicotine in Methanol.

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Page 1 of 6 Certified Reference Material - Certificate of Analysis (±)--Tocopherol (Vitamin E), Primary Standard Catalog Number: Lot: Retest: Description: Packaging: Storage: Shipping: Intended Use: March

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis T-32 Page 1 of 7 (-)- 8 -THC, Primary Standard Catalog Number: T-32 Lot: Expiration: Description: October 22 (-)- 8 -THC in Methanol. Packaging: Solution

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis T-005 Page 1 of 7 Catalog Number: T-005 Lot: (-)- 9 -THC, Primary Standard CH 3 OH Expiration: Description: April 2019 (-)- 9 -THC in Methanol. H

More information

Agilent Consumables Improve Performance with Shimadzu GC/MS

Agilent Consumables Improve Performance with Shimadzu GC/MS Agilent Consumables Improve Performance with Shimadzu GC/MS Application Note Environmental & Food Safety Author Ken Lynam Agilent Technologies, Inc. Abstract Leak-free performance over 150 heat cycles

More information

CORESTA Recommended Method No. 57

CORESTA Recommended Method No. 57 Cooperation Centre for Scientific Research Relative to Tobacco Tobacco and Tobacco Products Analytes Sub-Group CORESTA Recommended Method No. 57 DETERMINATION OF WATER IN TOBACCO AND TOBACCO PRODUCTS BY

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis S(+)-Methamphetamine, Primary Standard M-020 Page 1 of 7 Catalog Number: M-020 Lot: Expiration: Description: Packaging: Storage: Shipping: Intended

More information

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph Analysis of USP Residual Solvents using the Agilent 797A Headspace Sampler with the Agilent 79B Gas Chromatograph Application Note Pharmaceuticals Authors Bart Tienpont, Frank David, and Pat Sandra

More information

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE.

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE Authors: DANI Instruments Inc 257 Simarano Drive Marlborough, MA 01752 Ph: 508-281-6232

More information

Volatile Organic Compounds in Solids PBM

Volatile Organic Compounds in Solids PBM Organics Revision Date: Aug 15, 2014 Volatile Organic Compounds in Solids PBM Parameter Analytical Method Introduction Method Summary Parameter Applicability, MDLs Volatile Organic Compounds (VOCs) in

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis Cannabigerolic acid (CBGA), Primary Standard Page 1 of 6 Catalog Number: Lot: Expiration: Description: November 2015 Cannabigerolic acid in Acetonitrile.

More information

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications TECHNICAL BRIEF Gas Chromatography/ Mass Spectrometry The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications C 14 scans Introduction As gas chromatography

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis Lorazepam, Primary Standard Page 1 of 6 Catalog Number: Lot: Expiration: Description: October 2020 Lorazepam in Acetonitrile. NH Packaging: Solution

More information

Using Alternative Carrier Gas for US EPA Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Using Alternative Carrier Gas for US EPA Application Note. Abstract. Introduction. Experimental-Instrument Conditions Using Alternative Carrier Gas for US EPA 524.3 Application Note Abstract Due to regional shortages and increasing costs of helium, the preferred carrier gas in gas chromatography, alternative carrier gas

More information

Analytical Method of Limit Test for Hexachlorobenzene. in Picloram TC and Method Validation Data*

Analytical Method of Limit Test for Hexachlorobenzene. in Picloram TC and Method Validation Data* Analytical Method of Limit Test for Hexachlorobenzene in Picloram TC and Method Validation Data* Test Method (ABCTM-2010-01-07) 1. Apparatus Electronic Balance GC/MS System with NCI Ultrasonic Water Bath

More information

When optimization is discussed throughout this talk, it s in terms of performing more samples, in less time, while minimizing investment in capital

When optimization is discussed throughout this talk, it s in terms of performing more samples, in less time, while minimizing investment in capital 1 2 When optimization is discussed throughout this talk, it s in terms of performing more samples, in less time, while minimizing investment in capital instruments. 3 Efficiency is a measure of peak width.

More information

Two Headspace Solutions for the Analysis of Dissolved Gases in Water by Method RSKSOP-175

Two Headspace Solutions for the Analysis of Dissolved Gases in Water by Method RSKSOP-175 Two Headspace Solutions for the Analysis of Dissolved Gases in Water by Method RSKSOP-175 Application Note Abstract Dissolved gas analysis has been performed using a simplified headspace method following

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis Page 1 of 9 Fluconazole- 13 C 3, Primary Standard Catalog Number: Lot: Retest: July 2015 Description: Fluconazole- 13 C 3 in Methanol Packaging: Solution

More information

2.1 This procedure provides steps and conditions for the determination of Vapor Space Organic Compounds in ground waters, and industrial waste waters.

2.1 This procedure provides steps and conditions for the determination of Vapor Space Organic Compounds in ground waters, and industrial waste waters. 1. TITLE: MSD Analytical Method for Vapor Space Organics 2. PURPOSE: 2.1 This procedure provides steps and conditions for the determination of Vapor Space Organic Compounds in ground waters, and industrial

More information

US EPA Method Atomx XYZ and Agilent 7890B GC/5977A MS

US EPA Method Atomx XYZ and Agilent 7890B GC/5977A MS Application Note Environmental US EPA Method 524.3 Atomx XYZ and Agilent 7890B GC/5977A MS Author Amy Nutter Applications Chemist, Teledyne Tekmar Abstract US EPA Method 524.3 was used to determine the

More information

EXTRACTION OF BIOLOGICAL TISSUES FOR AROMATIC AND CHLORINATED HYDROCARBONS

EXTRACTION OF BIOLOGICAL TISSUES FOR AROMATIC AND CHLORINATED HYDROCARBONS EXTRACTION OF BIOLOGICAL TISSUES FOR AROMATIC AND CHLORINATED HYDROCARBONS Donell. S. Frank, Bo Wang, Juan A. Ramirez, Susanne J. McDonald, Rebecca Price, and James M. Brooks TDI-Brooks International/B&B

More information

2-Propanol Standard. Catalog # Lot # A

2-Propanol Standard. Catalog # Lot # A Analytical Reference Materials 2-Propanol Standard Catalog # 30473 Lot # A0106317 110 Benner Circle Bellefonte, PA 16823-8812 1-814-353-1300 1-800-356-1168 www.restek.com FOR LABORATORY USE ONLY. READ

More information

a,a,a-trifluorotoluene Mix

a,a,a-trifluorotoluene Mix Analytical Reference Materials a,a,a-trifluorotoluene Mix Catalog # 30048 Lot # A0102664 110 Benner Circle Bellefonte, PA 16823-8812 1-814-353-1300 1-800-356-1168 www.restek.com FOR LABORATORY USE ONLY.

More information

Chromatography 2015, 36, Technical Note

Chromatography 2015, 36, Technical Note Technical Note Development and Validation of an Analytical Method for Simultaneous Quantitation of Organic Volatile Impurities in Technical-Grade Active Ingredients of Pesticides and Agrochemicals by Headspace

More information

Surrogate Standard. Catalog # Lot # A

Surrogate Standard. Catalog # Lot # A Analytical Reference Materials Surrogate Standard Catalog # 30484 Lot # A0106703 110 Benner Circle Bellefonte, PA 16823-8812 1-814-353-1300 1-800-356-1168 www.restek.com FOR LABORATORY USE ONLY. READ MSDS

More information

Residual Solvents n-hexane Standard

Residual Solvents n-hexane Standard Analytical Reference Materials Residual Solvents n-hexane Standard Catalog # 36299 Lot # A088531 110 Benner Circle Bellefonte, PA 16823-8812 (814) 353-1300 FOR LABORATORY USE ONLY. READ MSDS PRIOR TO USE.

More information

Unleaded Gasoline Standard (Unweathered)

Unleaded Gasoline Standard (Unweathered) Analytical Reference Materials Unleaded Gasoline Standard (Unweathered) Catalog # 30096 Lot # A0101583 110 Benner Circle Bellefonte, PA 16823-8812 1-814-353-1300 1-800-356-1168 www.restek.com FOR LABORATORY

More information

Standard Operation Procedure (SOP) for the Extraction and Instrument Analysis of Siloxanes from Cartridges

Standard Operation Procedure (SOP) for the Extraction and Instrument Analysis of Siloxanes from Cartridges Standard Operation Procedure (SOP) for the Extraction and Instrument Analysis of Siloxanes from Cartridges Created March 17, 2016 Updated April 14, 2016 Dr. Rachel Marek (Asst. Research Scientist) 4105

More information

Ethylene Oxide Standard

Ethylene Oxide Standard Analytical Reference Materials Ethylene Oxide Standard Catalog # 36005 Lot # A092280 110 Benner Circle Bellefonte, PA 16823-8812 (814) 353-1300 FOR LABORATORY USE ONLY. READ MSDS PRIOR TO USE. RAW MATERIAL

More information

Acetone Standard. Catalog # Lot # A094008

Acetone Standard. Catalog # Lot # A094008 Analytical Reference Materials Acetone Standard Catalog # 30245 Lot # A094008 110 Benner Circle Bellefonte, PA 16823-8812 (814) 353-1300 FOR LABORATORY USE ONLY. READ MSDS PRIOR TO USE. RAW MATERIAL TEST

More information

Determination of Table 3 Compounds by LC/MS/MS Chemours Fluoroproducts Analytical Method Revision date: 4/26/2018

Determination of Table 3 Compounds by LC/MS/MS Chemours Fluoroproducts Analytical Method Revision date: 4/26/2018 Determination of Table 3 Compounds by LC/MS/MS Chemours Fluoroproducts Analytical Method Revision date: 4/26/2018 Instrument Setup (Note: Trapping column installed on Line A. Delay column installed between

More information

alpha-bhc Standard Catalog # Lot # A099219

alpha-bhc Standard Catalog # Lot # A099219 Analytical Reference Materials alpha-bhc Standard Catalog # 32206 Lot # A099219 110 Benner Circle Bellefonte, PA 16823-8812 1-814-353-1300 1-800-356-1168 www.restek.com FOR LABORATORY USE ONLY. READ MSDS

More information

NWTPH-Gx. Volatile Petroleum Products Method for Soil and Water

NWTPH-Gx. Volatile Petroleum Products Method for Soil and Water NWTPH-Gx Volatile Petroleum Products Method for Soil and Water Summary The NWTPH-Gx method is intended to replace the Oregon s TPH-G and Washington s WTPH-G methods and to present a more comprehensive

More information

Formaldehyde Oxazolidine Standard

Formaldehyde Oxazolidine Standard Analytical Reference Materials Formaldehyde Oxazolidine Standard Catalog # 33004 Lot # A099348 110 Benner Circle Bellefonte, PA 16823-8812 1-814-353-1300 1-800-356-1168 www.restek.com FOR LABORATORY USE

More information

Gas Chromatography SSI-GC-003

Gas Chromatography SSI-GC-003 Gas Chromatography SSI-GC-003 Trace Oxygenated Hydrocarbons in Liquid Hydrocarbon Streams by Multidimensional GC Introduction It has been widely accepted that oxygenates are related to corrosion and fouling

More information

Course Outline. in Soil/Sediment&WasteSediment&Waste by. Overview of VOCs Analysis Sample Collection VOCs Analysis

Course Outline. in Soil/Sediment&WasteSediment&Waste by. Overview of VOCs Analysis Sample Collection VOCs Analysis Determination of Volatile Organic Compounds in Soil/Sediment&WasteSediment&Waste by Headspace Technique.( ) MS (Environmental Toxicology) The University of Michigan, USA Phaitoon_m@yahoo.com Mobile: 0861300055

More information

Chloromethane Standard

Chloromethane Standard Analytical Reference Materials Chloromethane Standard Catalog # 30267 Lot # A0101059 110 Benner Circle Bellefonte, PA 16823-8812 1-814-353-1300 1-800-356-1168 www.restek.com FOR LABORATORY USE ONLY. READ

More information

Standard Mixture Stock Solution

Standard Mixture Stock Solution Analytical Reference Materials Standard Mixture Stock Solution Catalog # 31630 Lot # A0104212 110 Benner Circle Bellefonte, PA 16823-8812 1-814-353-1300 1-800-356-1168 www.restek.com FOR LABORATORY USE

More information

Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D*

Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D* Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D* UCT Products: ECUNIPAH (ENVIRO-CLEAN Universal PAH/DRO 2000 mg/83 ml cartridge) Method Summary Method 8015D may be used

More information

USP Class 1 Residual Solvent Mixture

USP Class 1 Residual Solvent Mixture Analytical Reference Materials USP Class 1 Residual Solvent Mixture Catalog # 36279 Lot # A093744 110 Benner Circle Bellefonte, PA 16823-8812 (814) 353-1300 FOR LABORATORY USE ONLY. READ MSDS PRIOR TO

More information

EPA Method 524 for Determination of VOCs in Drinking Water Using Agilent 5975T LTM GC/MSD with Static Headspace

EPA Method 524 for Determination of VOCs in Drinking Water Using Agilent 5975T LTM GC/MSD with Static Headspace EPA Method 524 for Determination of VOCs in Drinking Water Using Agilent 5975T LTM GC/MSD with Static Headspace Application Note Environmental Authors Xiaohua Li Agilent Technologies (Shanghai) Co., Ltd.

More information

A New Horizon of Performance Agilent G1888 Network Headspace Sampler

A New Horizon of Performance Agilent G1888 Network Headspace Sampler A New Horizon of Performance Agilent G1888 Network Headspace Sampler Full system integration Extended sample capacity Built-in ease-of-use Uncompromised reliability The Steps To Confidence From sample

More information

Naphthalene Standard. Catalog # Lot # A093142

Naphthalene Standard. Catalog # Lot # A093142 Analytical Reference Materials Naphthalene Standard Catalog # 31280 Lot # A093142 110 Benner Circle Bellefonte, PA 16823-8812 (814) 353-1300 FOR LABORATORY USE ONLY. READ MSDS PRIOR TO USE. RAW MATERIAL

More information

SOM01.0 SIM Analysis Standard

SOM01.0 SIM Analysis Standard Analytical Reference Materials SOM01.0 SIM Analysis Standard Catalog # 33913 Lot # A0106668 110 Benner Circle Bellefonte, PA 16823-8812 1-814-353-1300 1-800-356-1168 www.restek.com FOR LABORATORY USE ONLY.

More information

Methyl Methacrylate Standard

Methyl Methacrylate Standard Analytical Reference Materials Methyl Methacrylate Standard Catalog # 30299 Lot # A093861 110 Benner Circle Bellefonte, PA 16823-8812 (814) 353-1300 FOR LABORATORY USE ONLY. READ MSDS PRIOR TO USE. RAW

More information

Ethylene Oxide Standard

Ethylene Oxide Standard Analytical Reference Materials Ethylene Oxide Standard Catalog # 30548 Lot # A0102719 110 Benner Circle Bellefonte, PA 16823-8812 1-814-353-1300 1-800-356-1168 www.restek.com FOR LABORATORY USE ONLY. READ

More information

of Methane, Ethylene, Acetylene and Ethane in Water by Headspace-Gas Chromatography (HS-GC) with

of Methane, Ethylene, Acetylene and Ethane in Water by Headspace-Gas Chromatography (HS-GC) with Optimizing RSK 175 for the analysis of Methane, Ethylene, Acetylene and Ethane in Water by Headspace-Gas Chromatography (HS-GC) with Flame Ionization Detection (FID) Lee Marotta, Product Specialist, PerkinElmer

More information

Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e 1

Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e 1 Application Note US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e 1 Abstract US EPA Method 8260 in conjunction

More information

Residual Solvents Ethylene Glycol Standard

Residual Solvents Ethylene Glycol Standard Analytical Reference Materials Residual Solvents Ethylene Glycol Standard Catalog # 36297 Lot # A086740 110 Benner Circle Bellefonte, PA 16823-8812 (814) 353-1300 FOR LABORATORY USE ONLY. READ MSDS PRIOR

More information

Corn Syrup Analysis E-1a- 1. ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

Corn Syrup Analysis E-1a- 1. ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) Corn Syrup Analysis E-1a- 1 ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) PRINCIPLE Isovaleraldehyde (IVA) and acetaldehyde are released from the syrup with the aid of dilute phosphoric acid and

More information

C146-E096 TD-20. Shimadzu Thermal Desorption System

C146-E096 TD-20. Shimadzu Thermal Desorption System C146-E096 TD-20 Shimadzu Thermal Desorption System Thermal Desorption Thermal desorption systems trap target substances carried in mobile phase by adsorption in a sample tube packed with adsorbent. Then,

More information

EURL-SRM - Analytical Observations Report

EURL-SRM - Analytical Observations Report concerning the following EURL-SRM - Analytical Observations Report o Compound(s): Sulfuryl fluoride o Commodities: cereals, nuts, pulses and other dry materials o Extraction Method(s): none o Instrumental

More information

Butyltin Chlorides Calibration Mixture

Butyltin Chlorides Calibration Mixture Analytical Reference Materials Butyltin Chlorides Calibration Mixture Catalog # 31472 Lot # A0101755 110 Benner Circle Bellefonte, PA 16823-8812 1-814-353-1300 1-800-356-1168 www.restek.com FOR LABORATORY

More information

Improved Gas Chromatograph Method for the Analysis of Trace Hydrocarbon Impurities in 1, 3-Butadiene

Improved Gas Chromatograph Method for the Analysis of Trace Hydrocarbon Impurities in 1, 3-Butadiene Improved Gas Chromatograph Method for the Analysis of Trace Hydrocarbon Impurities in 1, 3-Butadiene Application Note Hydrocarbon processing industry Author Chunxiao Wang Agilent Technologies (Shanghai)

More information

Atrazine-d5 Standard. Catalog # Lot # A & A090569

Atrazine-d5 Standard. Catalog # Lot # A & A090569 Analytical Reference Materials Atrazine-d5 Standard Catalog # 31984 Lot # A091473 & A090569 110 Benner Circle Bellefonte, PA 16823-8812 (814) 353-1300 FOR LABORATORY USE ONLY. READ MSDS PRIOR TO USE. RAW

More information

Analysis of Semivolatiles by GC/FID using the Ultra Inert Inlet Liners with Wool

Analysis of Semivolatiles by GC/FID using the Ultra Inert Inlet Liners with Wool Analysis of Semivolatiles by GC/FID using the Ultra Inert Inlet Liners with Wool Application Note Environmental Authors Limian Zhao and Bill Wilson Agilent Technologies, Inc. 2850 Centerville Road Wilmington,

More information

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water Application Note Environmental Authors Min Cai and Yun Zou Agilent Technologies Co. Ltd, 412 Ying Lun Road Waigaoqiao

More information

Organohalide Pesticides in Water by GC/µECD with Agilent J&W DB-CLP1 and DB-CLP2

Organohalide Pesticides in Water by GC/µECD with Agilent J&W DB-CLP1 and DB-CLP2 Organohalide Pesticides in Water by GC/µECD with Agilent J&W DB-CLP and DB-CLP Application Note Environmental Authors Doris Smith and Ken Lynam Agilent Technologies, Inc. Abstract Organohalide pesticides

More information

[ APPLICATION NOTE ] APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS

[ APPLICATION NOTE ] APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS Quantitative Method for the Determination of Potentially Mutagenic Impurities of Ondansetron Using UPC 2 Coupled with a Xevo TQ-S micro Jennifer Simeone, Paula Hong, and Patricia R. McConville Waters Corporation,

More information

Additional guidance on: TNsG on Data Requirements, Part A, Chapter 2, Point 4. Analytical Methods for Detection and Identification.

Additional guidance on: TNsG on Data Requirements, Part A, Chapter 2, Point 4. Analytical Methods for Detection and Identification. Additional guidance on: TNsG on Data Requirements, Part A, Chapter 2, Point 4 Analytical Methods for Detection and Identification and Part B, Chapter 2, Point 4 Methods of Identification and Analysis These

More information

Quantitatitive Analysis of Phosphorothioate Oligonucleotide in Human Plasma Using LC-MS/MS with On-Line Extraction

Quantitatitive Analysis of Phosphorothioate Oligonucleotide in Human Plasma Using LC-MS/MS with On-Line Extraction Laixin Wang, Sherry Liu, Qiuying Zhu, Scott Reuschel and Min Meng Tandem Labs Quantitatitive Analysis of Phosphorothioate Oligonucleotide in Human Plasma Using LC-MS/MS with On-Line Extraction Introduction

More information

US EPA Method 8260 with the Atomx XYZ and the Thermo Scientific TRACE 1310 GC and ISQ MS

US EPA Method 8260 with the Atomx XYZ and the Thermo Scientific TRACE 1310 GC and ISQ MS APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN1803 P+T VOC ENV US EPA Method 8260 with the Atomx XYZ and the Thermo Scientific TRACE 1310 GC and ISQ MS Amy Nutter, Applications Chemist; Teledyne Tekmar Page

More information

Optimizing Productivity and Reliability for Monocyclic Aromatic Hydrocarbon Purity Analysis According to ASTM D7504 on the Agilent 8890 GC System

Optimizing Productivity and Reliability for Monocyclic Aromatic Hydrocarbon Purity Analysis According to ASTM D7504 on the Agilent 8890 GC System Application Note Petrochemicals Optimizing Productivity and Reliability for Monocyclic Aromatic Hydrocarbon Purity Analysis According to ASTM D50 on the Agilent 0 GC System Authors Jie Pan, Lukas Wieder,

More information

VOA Calibration Mix #1

VOA Calibration Mix #1 Analytical Reference Materials VOA Calibration Mix #1 Catalog # 30006 Lot # A090981 & A090791 110 Benner Circle Bellefonte, PA 16823-8812 (814) 353-1300 FOR LABORATORY USE ONLY. READ MSDS PRIOR TO USE.

More information

A Sub-picogram Quantification Method for Desmopressin in Plasma using the AB SCIEX Triple Quad 6500 System

A Sub-picogram Quantification Method for Desmopressin in Plasma using the AB SCIEX Triple Quad 6500 System A Sub-picogram Quantification Method for Desmopressin in Plasma using the AB SCIEX Triple Quad 6500 System A high-throughput method for detecting ultra-low levels (0.5 pg/ml) of a therapeutic peptide in

More information

ADVANCES IN MICRO GAS CHROMATOGRAPHY

ADVANCES IN MICRO GAS CHROMATOGRAPHY ADVANCES IN MICRO GAS CHROMATOGRAPHY Fast Analysis of Hydrogen Sulfide in a Natural Gas Matrix using a Temperature Programmable Micro GC Fusion Gulf Coast Conference 2015 Debbie Alcorn INFICON, Inc. 2

More information

Olefin Analyzer Solutions for Polymer Grade Ethylene and Propylene

Olefin Analyzer Solutions for Polymer Grade Ethylene and Propylene Olefin Analyzer Solutions for Polymer Grade Ethylene and Propylene Angela Smith Henry Kelly Beard George Gonzalez 1 Impurity Monitoring in Olefin Production Analysis of contaminants in ethylene and propylene

More information

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1 Application Note US EPA Method 524.3 with the Tekmar Lumin P & T Concentrator and Agilent 7890B GC and 5977A MSD System Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1 Abstract US EPA Method

More information

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Application Note US EPA Method 524.4 with the Tekmar Lumin P & T Concentrator and Agilent 7890B GC and 5977A MSD System Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1 Abstract US EPA Method

More information

Quantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy

Quantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy Advances in Bioscience and Biotechnology, 2010, 1, 367-371 doi:10.4236/abb.2010.15049 Published Online December 2010 (http://www.scirp.org/journal/abb/). Quantitative determination of residual 2-(2-chloroethoxy)

More information

Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application

Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application Food Authors Frank David Research Institute for Chromatography Pres. Kennedypark

More information

Improving Sensitivity for an Immunocapture LC-MS Assay of Infliximab in Rat Plasma Using Trap-and-Elute MicroLC-MS

Improving Sensitivity for an Immunocapture LC-MS Assay of Infliximab in Rat Plasma Using Trap-and-Elute MicroLC-MS Improving Sensitivity for an Immunocapture LC-MS Assay of in Rat Plasma Using Trap-and-Elute MicroLC-MS Using the SCIEX M3 MicroLC system for Increased Sensitivity in Antibody Quantitation Remco van Soest

More information

METHOD TRANSLATION OF HJ FOR INTUVO

METHOD TRANSLATION OF HJ FOR INTUVO METHOD TRANSLATION OF HJ679-2013 FOR INTUVO Translating a 530 µm id Column Method to 250 µm or 320 µm column Methods for the Agilent Intuvo 9000 GC Introduction HJ method 679-2013 describes the determination

More information

Method Revised Chlorinated Pesticides Mix #2

Method Revised Chlorinated Pesticides Mix #2 Analytical Reference Materials Method 525.2 Revised Chlorinated Pesticides Mix #2 Catalog # 33011 Lot # A088244 110 Benner Circle Bellefonte, PA 16823-8812 (814) 353-1300 FOR LABORATORY USE ONLY. READ

More information

Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS

Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS Using the SCIEX M3 MicroLC system for Increased Sensitivity in Antibody Quantitation Remco van Soest and Lei Xiong SCIEX, Redwood City,

More information

Maximizing Sample Recovery on the ACCQPrep HP125

Maximizing Sample Recovery on the ACCQPrep HP125 Maximizing Sample Recovery on the ACCQPrep HP125 Chromatography Technical Note TN42 Overview This technical note is to investigate the sample recovery and injection efficiency of different methods of injection

More information

Determination of Volatile Organic Compounds (VOCs) in Various Drinking Water Sources by GC/MS

Determination of Volatile Organic Compounds (VOCs) in Various Drinking Water Sources by GC/MS Determination of Volatile Organic Compounds (VOCs) in Various Drinking Water Sources by GC/MS Application Note Abstract Consumers have become increasingly concerned with the quality of their drinking water.

More information

HYDROCARBONS, BP C 1500

HYDROCARBONS, BP C 1500 HYDROCARBONS, BP 36 216 C 1500 FORMULA: Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1 METHOD: 1500, Issue 3 EVALUATION: PARTIAL Issue 1: 15 August 1990 Issue 3: 15 March 2003 OSHA : Table 2 NIOSH: Table

More information

Performance characteristics of the 1260 Infinity Quaternary LC system

Performance characteristics of the 1260 Infinity Quaternary LC system Performance characteristics of the 1260 Infinity Quaternary LC system The new standard in HPLC Technical Overview Introduction The Agilent 1260 Infinity LC system consists of modular units that operate

More information

Comprehensive screening of environmental contaminants by flow-modulated GC GC

Comprehensive screening of environmental contaminants by flow-modulated GC GC Comprehensive screening of environmental contaminants by flow-modulated GC GC Matthew Edwards, Separation Science Product Manager NEMC, Fri 11 th Aug 2017 GC GC modulation Over 20 years of innovation Giddings

More information

A Sub-picogram Quantification Method for Desmopressin in Plasma using the SCIEX Triple Quad 6500 System

A Sub-picogram Quantification Method for Desmopressin in Plasma using the SCIEX Triple Quad 6500 System A Sub-picogram Quantification Method for Desmopressin in Plasma using the SCIEX Triple Quad 6500 System A high-throughput method for detecting ultra-low levels (0.5 pg/ml) of a therapeutic peptide in human

More information

SSENTIALS RESTEK. Must-Have Solutions for Your Lab Leverage 3-in-1 Technology for Outstanding Inertness Low Bleed High Reproducibility

SSENTIALS RESTEK. Must-Have Solutions for Your Lab Leverage 3-in-1 Technology for Outstanding Inertness Low Bleed High Reproducibility RESTEK SSENTIALS Must-Have Solutions for Your Lab 2014.2 Leverage 3-in-1 Technology for Outstanding Inertness Low Bleed High Reproducibility ALSO IN THIS ISSUE: Raptor Columns: Biphenyl & ARC-18 Sky Inlet

More information

Propylene Glycol Standard

Propylene Glycol Standard Analytical Reference Materials Propylene Glycol Standard Catalog # 31892 Lot # A095654 110 Benner Circle Bellefonte, PA 16823-8812 (814) 353-1300 FOR LABORATORY USE ONLY. READ MSDS PRIOR TO USE. RAW MATERIAL

More information

Method 0040 for Volatile Total Organics as Unspeciated Mass in Stack Gas Emissions

Method 0040 for Volatile Total Organics as Unspeciated Mass in Stack Gas Emissions Method 0040 for Volatile Total Organics as Unspeciated Mass in Stack Gas Emissions Sampling and Field Procedure for Volatile Unspeciated Mass from Method 0040 Samples in Stack Gas Sample Name: Sampler:

More information

Application Note. Abstract. Authors. Introduction. Environmental

Application Note. Abstract. Authors. Introduction. Environmental EPA Method 538: Determination of Selected Organic Contaminants in Drinking Water by Direct Aqueous Injection with the Agilent 646 Triple Quadrupole LC/MS System Application Note Environmental Authors Imma

More information

ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER

ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER ENVIRONMENTAL ANALYSIS ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER Solutions for Your Analytical Business

More information

ITSP Extraction of Cortisol from Plasma

ITSP Extraction of Cortisol from Plasma ITSP Extraction of Cortisol from Plasma Introduction Solid Phase Extraction (SPE) is well established as a preferred tool for extracting small organic molecules from complex biological matrices (e.g. plasma)

More information

9 THC CRM 6aR,7,8,10aR tetrahydro 6,6,9 trimethyl 3 pentyl 6H dibenzo[b,d]pyran 1 ol SAMPLE

9 THC CRM 6aR,7,8,10aR tetrahydro 6,6,9 trimethyl 3 pentyl 6H dibenzo[b,d]pyran 1 ol SAMPLE Item #: ISO60157 Batch #: 0515871 CAS Registry Number: 1972-08-03 9 THC CRM 6aR,7,8,10aR tetrahydro 6,6,9 trimethyl 3 pentyl 6H dibenzo[b,d]pyran 1 ol Expiry: 13OCT2020 (valid from date of certification)

More information

Analysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS

Analysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS Analysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS Authors: Conor Smith, Doug Fryer United Science Corp. 15911 Furuby Rd. Center City, MN 55012 USA Abstract

More information

Validation Data for the Analysis of Chlorotalonil in Leek Using Mini-Luke Method Followed by GC-QqQ-MS/MS

Validation Data for the Analysis of Chlorotalonil in Leek Using Mini-Luke Method Followed by GC-QqQ-MS/MS Validation Data for the Analysis of Chlorotalonil in Leek Using Mini-Luke Method Followed by GC-QqQ-MS/MS CONTENTS 1. Aim and Scope 2 2. Short Description 2 3. Apparatus and Consumables 2 4. Chemicals

More information

USEPA Methods and 8021B using the Eclipse Purge-and-Trap (P&T) Sample Concentrator and PID/ELCD Tandem Detectors

USEPA Methods and 8021B using the Eclipse Purge-and-Trap (P&T) Sample Concentrator and PID/ELCD Tandem Detectors Application Note 31630508 USEPA Methods 502.2 and 8021B using the Eclipse Purge-and-Trap (P&T) Sample Concentrator and PID/ELCD Tandem Detectors Keywords Volatile Organic Compounds (VOCs) Purge-and-Trap

More information

Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run

Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run Application Note Author Jessica Westland Agilent Technologies, Inc. Abstract A selected target compound list of 195 various pesticides

More information

PAL SPME Fibers Optimized for Automation

PAL SPME Fibers Optimized for Automation PAL SPME Fibers Optimized for Automation PAL SPME Fiber Since its introduction by Pawliszyn et al. (ref. 1) Solid Phase Micro Extraction (SPME) has seen a tremendous development. SPME is a very effective

More information

NISTmAb characterization with a high-performance RP chromatography column

NISTmAb characterization with a high-performance RP chromatography column APPLICATION NOTE 21848 NISTmAb characterization with a high-performance RP chromatography column Author Xin Zhang Thermo Fisher Scientific, Sunnyvale, CA, USA Keywords MAbPac RP column, inter-column reproducibility,

More information

USEPA Methods and 8021B using the Eclipse Purge-and-Trap (P&T) Sample Concentrator and PID/ELCD Tandem Detectors

USEPA Methods and 8021B using the Eclipse Purge-and-Trap (P&T) Sample Concentrator and PID/ELCD Tandem Detectors Application Note 31630710 USEPA Methods 502.2 and 8021B using the Eclipse Purge-and-Trap (P&T) Sample Concentrator and PID/ELCD Tandem Detectors Keywords Volatile Organic Compounds (VOCs) Purge-and-Trap

More information