Quality Control Concerns for New Generations of CT, MRI, & PET
|
|
- Gillian Harmon
- 6 years ago
- Views:
Transcription
1 Quality Control Concerns for New Generations of CT, MRI, & PET David W. Jordan, Ph.D. Sr. Medical Physicist, UH Case Medical Center Clinical Assistant Professor, CWRU
2 Disclosures No financial disclosures or industry relationships Member of organizations discussed: AAPM TG-126 (PET) AAPM WGMRQA (MRI)
3 Objectives Understand the typical elements of a QC program for CT, MRI, and PET equipment. Anticipate the QC challenges created by new imaging technology advances.
4 Outline Overview of current QC procedures for CT, MRI, and PET. Discuss new CT technology : wide/cone beams and iterative reconstruction Discuss new MRI technology: parallel imaging and massively multichannel coils Discuss new PET technology: time of flight and alternatives to PMT detectors
5 QC Perspectives Medical Physicist Imaging Technologist
6 QC Goals Obtain data to characterize performance (and dose) Obtain data to optimize performance Monitor constancy of function/performance
7 Meta QC: What is the Quality of your Quality Control? Does each test contribute to characterization, and/or optimization, and/or constancy?
8 Meta QC: What is the Quality of your Quality Control? Is each test: Clinically relevant? Feasible? (time, cost) Sensitive? Specific?
9 Meta QC: What is the Quality of your Quality Control? Is each/every test mandated by: ACR? AAPM TG # such and such? Joint Commission? ACRIN / CRO? Ohio Department of Health? Etc.?
10 CT
11 Current QC: CT QC phantom measurements: CT # accuracy/linearity Uniformity Noise / CNR / SNR High & low contrast Scanner alignment
12 Current QC: CT Dose: CTDI & 32 cm dia. cylinders (10-15 cm thick) Measure scanner output
13 CT Advances: Iterative Recon M.J. Willemink et al., Iterative reconstruction techniques for computed tomography Part 1: technical principles., Eur. Radiol. 23(6), (2013).
14 CT IR QC Challenges IR techniques manipulate image noise Radiologists report texture concerns with reconstructed images Details of reconstruction proprietary Is CT # manipulated / better / worse?
15 CT IR QC: What and How To Test? Characterization? Optimization? Constancy? Clinically relevant? Feasible? Sensitive? Specific?
16 CT Advances: Wide Beams 100-mm CTDI pencil chamber Xoran Technologies, Toshiba Medical Systems, anguage=en
17 Wide Beam QC Challenges CTDI 100 inadequate for dose (scanner output) measurements Proposed alternatives (TG-111) cumbersome: Large/long phantom Non-standard equipment & metrics TG 200 working on feasible implementation
18 C. Descamps, M. Gonzalez, E. Garrigo, A. Germanier, and D. Venencia, Measurements of the dose delivered during CT exams using AAPM Task Group Report No. 111., J. Appl. Clin. Med. Phys. 13(6), 3934 (2012).
19 Wide/Cone Beam CT Dose: How To Test? Characterization? Optimization? Constancy? Clinically relevant? Feasible? Sensitive? Specific?
20 CT Coming Attractions CT performance assessment including iterative reconstructions: AAPM Task Group 223 Established 2012 Ehsan Samei, Chair Development of wide/cone beam dosimetry: AAPM Task Group 200 Established 2010 Donovan Bakalyar, Chair
21 MRI
22 Magnet homogeneity Scanner performance Slice position/thickness Geometric accuracy Resolution Artifacts Performance of all individual coils SNR, uniformity Current QC: MRI
23 MRI Advances: Parallel Imaging R. Hashemi, W.G. Bradley, and C.J. Lisanti, MRI: The Basics (Lippincott Williams & Wilkins, 2010).
24 Parallel MRI QC Challenges
25 Parallel MRI: What/How To Test? Characterization? Optimization? Constancy? Clinically relevant? Feasible? Sensitive? Specific?
26 MRI Advances: Coils T. Niendorf et al., Progress and promises of human cardiac magnetic resonance at ultrahigh fields: a physics perspective., J. Magn. Reson. 229, (2013).
27 Coil QC Challenges State of the art coil arrays have up to 128 channels Many MORE individual elements in some arrays! Existing methods: Simple/aggregated scan may miss problems Thorough testing can take hours per coil, typical scanner has coils
28 MRI Coil Arrays: What/How To Test? Characterization? Optimization? Constancy? Clinically relevant? Feasible? Sensitive? Specific?
29 MRI Coming Attractions Parallel Imaging Technology & QC: AAPM Task Group 118 Established 2005 Nathan Yanasak, Chair RF Coil QC (and other topics): AAPM WG on MR Testing and QA Established 2005 Joseph Och, Chair
30 PET
31 Phantom images SUV accuracy Contrast cold/ warm /hot objects Resolution Count rate response vs. activity Sensitivity Spatial resolution Corrections Current QC: PET
32 PET Advances: Time of Flight Diagram of line of response vs segment of response
33 TOF-PET: What/How To Test? Characterization? Optimization? Constancy? Clinically relevant? Feasible? Sensitive? Specific?
34 PET Advances: New Detectors
35 Non-PMT Detector QC Challenges Air-cooled arrays performance depends on ambient temperature Detection efficiency and speed affect sensitivity, count rate performance Enable or require use of different scintillators? Effect of timing resolution?
36 Post-PMT PET: What/How To Test? Characterization? Optimization? Constancy? Clinically relevant? Feasible? Sensitive? Specific?
37 PET Coming Attractions PET-CT Acceptance Testing & QC: AAPM Task Group 126 Established 2005 Osama Mawlawi, Chair
38 Review of Objectives New technology developments: CT: iterative recon, wide beams, dose reduction technology MRI: parallel imaging and multichannel coils PET: time of flight and non-pmt detectors New technologies may or may not be adequately addressed by existing QC methods
39 Questions? Thank you!
2017 ACR Computed Tomography Quality Control Manual FAQS
Updated 11-15-2017 2017 ACR Computed Tomography Quality Control Manual FAQS Q. The updated 2017 ACR Computed Tomography Quality Control Manual has been released. (Visit www.acr.org/education/education-catalog.)
More informationCT QA SOLUTIONS. Ensure Accurate Screening, Diagnosis and Monitoring
CT QA SOLUTIONS Ensure Accurate Screening, Diagnosis and Monitoring ACCREDITATION ADVANCED CT IMAGE QUALITY QA AUTOMATIC EXPOSURE CONTROL CT CHARACTERIZATION MULTI-ENERGY CT CT PERFUSION QA COMPLETE CT
More informationJoint Commission Medical Imaging Requirements. Laura Gruber, MBA, RT(R), RDMS, RVT Sr. Director Department of Medical Imaging SPR Annual Meeting
Joint Commission Medical Imaging Requirements Laura Gruber, MBA, RT(R), RDMS, RVT Sr. Director Department of Medical Imaging SPR Annual Meeting My team of professionals 2 Medical Physicist s 3 FTE s (Board
More informationQuality Assurance for particle beam therapy
Quality Assurance for particle beam therapy PTCOG Educational Workshop, Essen 2013 Heidelberg Ion Beam Therapy Center at the University Hospital German Cancer Research Center, Heidelberg, Germany Outline
More informationTreatment planning for scanned proton beams Francesca Albertini, Center for Proton Therapy, Paul Scherrer Institut Switzerland
Treatment planning for scanned proton beams Francesca Albertini, Center for Proton Therapy, Paul Scherrer Institut Switzerland Outlook 1. Planning with active scanning protons in practice 2. Single Field
More informationTreatment Quality Assurance Cone Beam Image Guided Radiation Therapy. Jean-Pierre Bissonnette, PhD, MCCPM
Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy Jean-Pierre Bissonnette, PhD, MCCPM Disclosure Work supported, in part, by Elekta Oncology Systems Commercial Interest in Penta-Guide
More informationHandzettel 1. CARE Right Computed Tomography, committed to the right dose Ivo Driesser Austin, July 22 nd 2014
CARE Right Computed Tomography, committed to the right dose Ivo Driesser Austin, July 22 nd 2014 Answers for life. Answers for life. Is 1 msv the right dose for every patient? Female, 54 Abdominal CT Male,
More informationNew PET/CT from Siemens helps more patients benefit from premium technologies
Press Healthcare Erlangen, October 9, 2015 EANM 2015, Congress Center Hamburg (CCH) New PET/CT from Siemens helps more patients benefit from premium technologies Versatile new PET/CT system addresses a
More informationVantage Titan 1.5T MR. The Intelligent Choice.
Vantage Titan 1.5T MR. The Intelligent Choice. Scalable. M-Powered. Performance. Vantage Titan MR is designed to satisfy the demands of today s challenging clinical environment, addressing the requirements
More informationOptimization of Low-Dose CT Protocol in Pediatric Nuclear Medicine Imaging
Optimization of Low-Dose CT Protocol in Pediatric Nuclear Medicine Imaging Hanna Piwowarska-Bilska 1, Leszek J. Hahn 2, Bozena Birkenfeld 1, Katarzyna Cichon-Bankowska 1, Maria H. Listewnik 1, and Piotr
More informationMEDICAL PHYSICS (MED PHYS)
Medical Physics (MED PHYS) 1 MEDICAL PHYSICS (MED PHYS) MED PHYS/PHYSICS 265 INTRODUCTION TO MEDICAL PHYSICS Primarily for premeds and other students in the medical and biological sciences. Applications
More informationGENESIS Edition. Transforming CT
GENESIS Edition Transforming CT Transforming clinical confidence Transforming patient experience Transforming your workspace GENESIS Edition Transforming CT Brought to you by the leaders in area detector
More informationLow Dose, Lightning Speed, Latest Applications
RXL EDITION Low Dose, Lightning Speed, Latest Applications The New Standard for Radiology Toshiba Medical Systems is committed to the development of new technologies to minimize radiation dose while maintaining
More informationBIOMEDICAL SIGNAL AND IMAGE PROCESSING
BIOMEDICAL SIGNAL AND IMAGE PROCESSING EE 5390-001 SYLLABUS Instructor: Wei Qian, Ph.D. Professor of Electrical and Computer Engineering Medical Signal and Image Computerized Processing Scheme for Medical
More informationPET/MR: A COMMERCIAL SUCCESS STORY?
PET/MR: A COMMERCIAL SUCCESS STORY? ANTONIS KALEMIS P H D M B A Philips Healthcare Snr Manager, Clinical Science - Advanced Molecular Imaging AIPES (Association of Imaging Producers & Equipment Suppliers)
More informationIngenia MR-RT. MR Systems. The comprehensive MR-sim solution to fit your planning
Ingenia MR-RT MR Systems The comprehensive MR-sim solution to fit your planning Table of contents Experience the difference MRI makes 3 A comprehensive MR-sim solution 4 Position with precision 6 See clearly
More informationGENESIS Edition Transforming CT
GENESIS Edition Transforming CT 2 Transforming clinical confidence Transforming patient experience Transforming your workspace GENESIS Edition Transforming CT Brought to you by the leaders in area detector
More informationA New Standard for the Future
technology history For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.
More informationProgress in X-Ray & MR
Progress in X-Ray & MR Michiel Manuel Analyst Meeting June 15 th, 2005 X-Ray & MR: Agenda Introduction General X-Ray Cardio/Vascular X-Ray Magnetic Resonance China growth opportunity Conclusion 2 X-Ray
More information8/2/2017. Key attributes of scientific excellence: rigor, innovation, and relevance. Medical Physics 3.0. Key Attributes of Scientific Excellence
Medical Physics 3.0 in Design Key attributes of scientific excellence: rigor, innovation, and relevance Maryellen Giger, Ph.D. A. N. Pritzker Professor of Radiology / Medical Physics The University of
More informationMRI Systems Market to 2018
MRI Systems Market to 2018 Technological advancements, Increasing Number of Applications and Advent of MRI Compatible Pacemakers to Drive Future Growth GBI Research Report Guidance GBI Research Report
More informationT.W.I.S.T. IDEAL Radiological Imaging Services Presented by Adrienne Coya & Sandra Marcelin
T.W.I.S.T. IDEAL Radiological Imaging Services Presented by Adrienne Coya & Sandra Marcelin Evan O Donovan July 20, 2016 Education Program Manager, Quality Assurance Unit JCTO.WEILL.CORNELL.EDU Objectives
More informationReport on Human Resources Surveys in Medical Imaging & Radiation Therapy for the Canadian Association of Medical Radiation Technologists (CAMRT)
Report on Human Resources Surveys in Medical Imaging & Radiation Therapy for the Canadian Association of Medical Radiation Technologists (CAMRT) Keith Christopher PhD, CMC Principal Consultant, KC Surveys
More informationE2E SBRT Thorax Phantom
E2E SBRT Thorax Phantom Model 036A-CVXX-xx SBRT END-TO-END SBRT TESTING SOLUTION SCAN PLAN LOCALIZE TREAT This product is available through: JRT Associates 800-22-0 2428 Almeda Avenue Suite 36 Norfolk,
More informationGE Healthcare. Optima * CT660 Second Edition
GE Healthcare Optima * CT660 Second Edition Your vision of quality care made real The Optima CT660 system helps you deliver competent, personalized care that helps fulfill your mission and please your
More information03/08/2017. Advances and Challenges in Contour QA for Adaptive RT. Objective. Disclosures
Advances and Challenges in Contour QA for Adaptive RT Kristy K Brock, PhD, DABR, FAAPM Professor, Department of Imaging Physics Director, Image Guided Cancer Therapy Program University of Texas MD Anderson
More information7/31/2017. Implementing a Clinical Practice Guideline; Lessons From An Early Adopter of MPPG 5.a. Disclosures
Implementing a Clinical Practice Guideline; Lessons From An Early Adopter of MPPG 5.a Dustin Jacqmin, PhD Assistant Professor, Human Oncology University of Wisconsin Madison None Disclosures 1 Outline
More informationInvestigation of lung nodule detectability in low-dose 320-slice computed tomography
Investigation of lung nodule detectability in low-dose 320-slice computed tomography J. D. Silverman Institute of Biomaterials and Biomedical Engineering, University of Toronto, Toronto, Ontario M5G 2M9,
More informationSiemens Ct Scanner Manual
Siemens Ct Scanner Manual ZettaMed offers CT and MRI service for used GE, - ZettaMed offers CT service & MRI service including replacement parts; coils, new & used CT tubes, coldheads & compressors' replacements,
More informationNCI Research Networks Model for Collaboration with Bioelectronics Round Table
NCI Research Networks Model for Collaboration with Bioelectronics Round Table Larry Clarke, Cancer Imaging Program DCDT, NCI Detail: NIBIB Guest Scientist: NIST 1 Bioelectronics Round Table Research Triangle
More informationIMV 2006 CT Market Summary Report Featuring results from 2,565 hospital and non-hospital CT imaging sites in the United States
Medical Information Division 1400 East Touhy Ave., Suite 250 Des Plaines, Illinois 60018 Phone: 847-297-1404 Fax: 847-297-5010 www.imvinfo.com IMV 2006 CT Market Summary Report Featuring results from 2,565
More informationThe Department of Radiation Oncology. In loving Memory of Penny Slattery ( ) Medical Physics Journal Manager ( )
The Department of Radiation Oncology In loving Memory of Penny Slattery (1962-2015) Medical Physics Journal Manager (1996-2015) The Department of Radiation Oncology Medical Physics Workshop: Editorial
More informationGoing Clinical with the Accuray Radixact System. The Montefiore Experience
Going Clinical with the Accuray Radixact System The Montefiore Experience Disclaimer The views expressed in the presentation are those of the presenters and do not necessarily reflect the views or policies
More informationsiemens.com/mct-flow Biograph mct Flow Moving the standard to personalized care.
siemens.com/mct-flow Biograph mct Flow Moving the standard to personalized care. Every patient is unique. Every organ is different. Every disease is specific. So a one-size-fits-all exam protocol is not
More informationQIBA Profile: Diffusion-Weighted Magnetic Resonance Imaging (DWI)
QIBA Profile: Diffusion-Weighted Magnetic Resonance Imaging (DWI) Stage: A. Initial Draft Notation in this Template Template Element Appears as Instructions Boilerplate text Plain black text Don't change.
More informationThe Roles and Task of Medical Physicists in Clinical Trial (Therapy Trials: RTOG and NRG)
1 The Roles and Task of Medical Physicists in Clinical Trial (Therapy Trials: RTOG and NRG) YING XIAO, PHD AAPM Annual Meeting, 2017 NCI National Clinical Trials Network The NCTN structure includes five
More informationMRI-Guided On-line Adaptive Radiotherapy The UCLA Physics Experience. Disclosures 8/3/2016
MRI-Guided On-line Adaptive Radiotherapy The UCLA Physics Experience James Lamb, Nzhde Agazaryan, Minsong Cao, Mitch Kamrava, Percy Lee, Daniel Low, David Thomas, Yingli Yang Department of Radiation Oncology
More informationDuoFLEX. MRI Coil Suite. Flexible coils that contour to your patient and your budget.
DuoFLEX MRI Coil Suite Flexible coils that contour to your patient and your budget. DuoFLEX MRI Coil Suite contours to your patient and your budget while meeting or exceeding imaging needs. The DuoFLEX
More informationLeading Canadian Pediatric Hospital Continues to Drive Dose Down for Sick Children
Leading Canadian Pediatric Hospital Continues to Drive Dose Down for Sick Children At Canada s premier pediatric hospital also recognized as one of the world s foremost pediatric healthcare institutions
More informationThe World s First Adaptive Scanner
The World s First Adaptive Scanner SOMATOM Definition AS Excel Edition Answers for life. SOMATOM Definition AS Excel Edition 2 3 Siemens CT Vision Siemens CT Vision The justification for the existence
More informationUNTAPPED POTENTIAL PART 1 OF 3 How Radiology Administrators made the most of their CT investments
UNTAPPED POTENTIAL PART 1 OF 3 How Radiology Administrators made the most of their CT investments Automated. Accurate. Accessible. How Salem Regional Medical Center increased the speed and quality of CT
More informationDelivery uncertainties and their consequences
Planning CT CT after 5 weeks Beam stops at distal edge Beam overshoot Tony Lomax, Centre for Proton Radiotherapy, Paul Scherrer Institute, Switzerland 1. Density heterogeneities 2. Range uncertainty 3.
More informationI-class. All together a better way to work. MAGNETOM Avanto.
I-class. All together a better way to work. MAGNETOM Avanto www.siemens.com/medical Introducing I-class Innovative applications. Innovative workflow. Welcome to the next generation of MRI. Powered by innovative
More informationTake control of dose management
DoseWise Portal Take control of dose management across your organization Key benefits Make data-informed decisions to improve patient care with support from dose tracking and advanced analytics Enhance
More informationClinical Applications. ImagingRite. Neuro Intervention. 1 ImagingRite
Clinical Applications ImagingRite Neuro Intervention 1 ImagingRite ImagingRite, a comprehensive suite of imaging tools offered with Infinix -i angiographic systems, was designed to assist clinicians in
More informationThe Most Popular CT in the World *
www.siemens.com/healthcare The Most Popular CT in the World * SOMATOM Emotion Excel Edition *Based on the number of systems sold worldwide. Answers for life. 2 Emotion SOMATOM Emotion Excel Edition 3 CT
More informationIntroducing a new take on efficient workflow: Deeply integrated clinical applications
Clinical Applications Introducing a new take on efficient workflow: Deeply integrated clinical applications The EXPERIENCE of TRUE CLINICAL APPLICATIONS While developments in image acquisition technology
More informationAlternative Career Opportunities for Medical Physicists
Alternative Career Opportunities for Medical Physicists Robert Jeraj Professor of Medical Physics, Human Oncology, Radiology and Biomedical Engineering Director of Translational Imaging Research Program
More informationInterventional CT: More Procedures, Reduced Dose, Improved Speed at Minimal Cost
www.usa.siemens.com/healthcare Interventional CT: More Procedures, Reduced Dose, Improved Speed at Minimal Cost St. Nicholas Hospital Answers for life. Key Benefits Clinical: Improved image quality including
More informationX-ray Tomography in Industry: Current Status and Future Trends
X-ray Tomography in Industry: Current Status and Future Trends Dr. Simon Zabler Fraunhofer Magnetic Resonance- and X-ray Imaging Outline Industrial CT Current examples and limits of ict Future applications
More informationIntroduction: what do we mean by Quantitative MRI?
Introduction: what do we mean by Quantitative MRI? Paul Tofts PhD formerly Chair in Imaging Physics Brighton and Sussex Medical School, UK ISMRM 2010 Quantitative MRI slide 1 Overview 1. Quantification:
More informationUSP Chapter 823 USP 32 (old) vs. USP 35 (new)
USP Chapter 823 USP 32 (old) vs. USP 35 (new) Sally W. Schwarz, MS, BCNP Research Associate Professor of Radiology Washington University School of Medicine St. Louis, MO Why USP Chapter ? FDA has
More informationlow Dose, Lightning Speed
CXL edition low Dose, Lightning Speed Maximize the potential of today s clinical resources. In today s demanding healthcare environment, you need a CT system that satisfies all your requirements on a day-to-day
More informationwww.aquilionprime.com WARNING: Any reference to x-ray exposure, intravenous contrast dosage, and other medication is intended as a reference guideline only. The guidelines in this document do not substitute
More informationComputed Tomography. Printed in Japan
Computed Tomography WARNING: Any reference to x-ray exposure, intravenous contrast dosage, and other medication is intended as a reference guideline only. The guidelines in this document do not substitute
More informationJoint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May Dose Reference Levels
2033-7 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dose Reference Levels Peter Homolka EFOMP Training Course on Medical Physics in Diagnostic
More information3/6/2017 TOMOSYNTHESIS GUIDED BREAST BIOPSY LEARNING OBJECTIVES DISCLOSURES
TOMOSYNTHESIS GUIDED BREAST BIOPSY Amy Kerger, D.O. Assistant Professor The Ohio State Wexner Medical Center Stefanie Spielman Comprehensive Breast Center LEARNING OBJECTIVES What is the difference between
More informationPower-class. Expect more performance. MAGNETOM Symphony.
Power-class. Expect more performance. MAGNETOM Symphony www.siemens.com/healthcare MAGNETOM Symphony with Power-class: Our MRI does more to boost your ROI. It s already one of the world s most in-demand
More informationVibration effects in healthcare facilities
Vibration effects in healthcare facilities A. Nash Charles M. Salter Associates, 130 Sutter, 5th Floor, San Francisco, CA 94104, USA anthony.nash@cmsalter.com 1633 In the USA, the principal rating system
More informationTreatment planning for scanned proton beams and IMPT
Treatment planning for scanned proton SFUD IMPT Tony Lomax, Centre for Proton Radiotherapy, Paul Scherrer Institute, Switzerland Treatment planning for scanning 1. Principles of treatment planning for
More informationEric W. Abelquist, Ph.D., CHP President, Health Physics Society. August 2017
Careers in Health Physics Eric W. Abelquist, Ph.D., CHP President, Health Physics Society August 2017 What Is Health Physics? Study of radiation and its effects on people Multidisciplinary Physics Biology
More information1. Buying Imaging Equipment for Interoperability and Compliance: Beyond DICOM 2. Informatics for the Acceptance Testing of Imaging Equipment
TU-A-134-1 Tuesday 8:00AM - 8:55AM Room: 134 1. Buying Imaging Equipment for Interoperability and Compliance: Beyond DICOM 2. Informatics for the Acceptance Testing of Imaging Equipment For patient service,
More informationEvaluation of 1D, 2D and 3D nodule size estimation by radiologists for spherical and non-spherical nodules through CT thoracic phantom imaging
Evaluation of 1D, 2D and 3D nodule size estimation by radiologists for spherical and non-spherical nodules through CT thoracic phantom imaging Nicholas Petrick * * a, Hyun J. Grace Kim b, David Clunie
More informationJonathon A. Nye, Ph.D Emory University School of Medicine Atlanta, GA
Artifacts related to CT Jonathon A. Nye, Ph.D Emory University School of Medicine Atlanta, GA Acknowledgements: Fabio Esteves MD 1, James Galt PhD 1, Perry Sprawls PhD 1, Robert Eisner PhD 2, James Hamill
More informationIAEA HumAn HEAltH SErIES
IAEA HumAn HEAltH SErIES no. 1 Quality Assurance for PET and PET/CT Systems @ QUALITY ASSURANCE FOR PET AND PET/CT SYSTEMS The following States are Members of the International Atomic Energy Agency: AFGHANISTAN
More informationMulticenter oncology clinical trials are increasingly using
Quantitative PET/CT Scanner Performance Characterization Based Upon the Society of Nuclear Medicine and Molecular Imaging Clinical Trials Network Oncology Clinical Simulator Phantom John J. Sunderland
More informationUnderstanding The Provider Registration Scorecard. Powered By
Understanding The Provider Registration Scorecard Powered By Understanding Your Scorecard The provider registration scorecard is divided into section scores based on the information that your facility
More informationHHS Public Access Author manuscript Abdom Imaging. Author manuscript; available in PMC 2016 February 01.
Practical guide for implementing hybrid PET/MR clinical service: lessons learned from our experience Nainesh Parikh 1, Kent P. Friedman 1, Shetal N. Shah 2, and Hersh Chandarana 1 1 Department of Radiology,
More informationKnowledge-based treatment planning: fundamentally different, or more of the same?
Knowledge-based treatment planning: fundamentally different, or more of the same? Kevin L. Moore, Ph.D., DABR Associate Director of Physics Manager of Dosimetry Group Disclosure Statement Patent for dosimetric
More informationR/F. Experiences Using SONIALVISION safire and the Utility of Tomosynthesis. 1. Introduction. 2. Basics of Tomosynthesis.
R/F Experiences Using SONIALVISION safire and the Utility of Tomosynthesis Radiology Division, Dokkyo Medical University Koshigaya Hospital Masahiro Nakajima Mr. Masahiro Nakajima 1. Introduction The hospital
More informationOutline AAPM -- TG-127. TG and ABR. TG Scope Current General Status of MOC in eyes of TG-127 Continuing Education SAMs PQI Summary
Outline AAPM -- TG-127 Michael Yester, Ph.D. University of Alabama at Birmingham TG - 127 Scope Current General Status of MOC in eyes of TG-127 Continuing Education SAMs PQI Summary TG - 127 Scope The
More informationProton Treatment Planning Systems Daniel Yeung
New Developments in Proton Treatment Planning Systems Daniel Yeung Statement of Disclosure Funded Research & Development: Philips Medical Systems IBA Proton Planning Systems Commercial Systems Academic
More informationImage Gently: Ten Steps You Can Take to Optimize Image Quality and Lower CT Dose for Pediatric Patients
Medical Physics and Informatics Commentary Strauss et al. Image Quality and CT Dose for Pediatric Patients Medical Physics and Informatics Commentary FOCUS ON: Keith J. Strauss 1 Marilyn J. Goske 2 Sue
More informationStrategies for Adaptive RT
Strategies for Adaptive RT Olga L. Green Disclosures Honoraria and travel grants from ViewRay, Inc. 1 Learning Objectives What is ART? What is needed to implement real-time, online ART in the clinic? Example
More informationDual-Energy MDCT in Hypervascular Liver Tumors: Effect of Body Size on Selection of the Optimal Monochromatic Energy Level
Medical Physics and Informatics Original Research Mileto et al. Effect of Body Size on Dual-Energy MDCT of Hypervascular Liver Tumors Medical Physics and Informatics Original Research Achille Mileto 1
More informationAvailable online at ScienceDirect. Materials Today: Proceedings 2 (2015 )
Available online at www.sciencedirect.com ScienceDirect Materials Today: Proceedings 2 (2015 ) 5582 5586 International Conference on Solid State Physics 2013 (ICSSP 13) Thickness dependent optimization
More informationNew Product Innovation Leadership
Philips Healthcare A Frost & Sullivan Position Paper Nadim Michel Daher TABLE OF CONTENTS SIGNIFICANCE OF THE NEW PRODUCT INNOVATION LEADERSHIP AWARD... 3 KEY INDUSTRY CHALLENGES ADDRESSED BY PHILIPS HEALTHCARE...
More informationGastrointestinal Imaging Original Research
Gastrointestinal Imaging Original Research Mayer et al. Radiation Dose Savings in CT Gastrointestinal Imaging Original Research Caroline Mayer 1 Mathias Meyer 1 Christian Fink 1 Bernhard Schmidt 2 Martin
More informationToward Large-Scale Process Control to Enable Consistent CT Radiation Dose Optimization
Pediatric Imaging Review Larson et al. CT Radiation Dose Optimization Pediatric Imaging Review FOCUS ON: David B. Larson 1 Keith J. Strauss 2 Daniel J. Podberesky 3 Larson DB, Strauss KJ, Podberesky DJ
More informationTECHNOLOGY SPECIFIC REVIEW CRITERIA FOR EURO-BIOIMAGING NODES
MEDICAL IMAGING TECHNOLOGY SPECIFIC REVIEW CRITERIA FOR EURO-BIOIMAGING NODES January 31 st, 2013 TABLE OF CONTENTS Introduction... 3 Technology Review Criteria for Medical Imaging Euro-BioImaging Nodes...
More informationFATIGUE LIFE ASSESSMENT FOR COMPOSITE MATERIALS
18 TH INTERNATIONAL CONFERENCE ON COMPOSITE MATERIALS FATIGUE LIFE ASSESSMENT FOR COMPOSITE MATERIALS Y. Nikishkov*, A. Makeev School of Aerospace Engineering, Georgia Institute of Technology, Atlanta,
More informationRAI NEWS. Innovators in Translational & Quantitative Imaging
RAI NEWS A Newsletter of the Carl E. Ravin Advanced Imaging Laboratories Department of Radiology Duke University Medical Center Volume 8, Spring 2013 Innovators in Translational & Quantitative Imaging
More informationMedical Imaging Technologies in Canada, 2006 Methodological Notes
Medical Imaging Technologies in Canada, 2006 Methodological Notes Introduction The Canadian Institute for Health Information (CIHI) aims to provide accurate and timely data and information to help shape
More informationMotiva: a TV-based platform for remote patient management
Technology news New products Motiva: a TV-based platform for remote patient management a simplified user interface, a secure broadband connection and a standard set-top box. A nurse care manager, using
More informationHeavy Ion Therapy- The search for the Holy Grail of Radiation Therapy
Heavy Ion Therapy- The search for the Holy Grail of Radiation Therapy Presented by George Coutrakon,PhD Loma Linda University Medical Center Loma Linda, CA Energy deposited (on film) per unit length vs.
More informationCODING GUIDELINES. Radiation Therapy. Effective January 1, 2018
CODING GUIDELINES Radiation Therapy Effective January 1, 2018 Coding guidelines for medical necessity review of radiation therapy services. 2018 evicore healthcare. All rights reserved. Please note the
More informationSBRT LUNG CANCER CLINICAL PATHWAY
SBRT LUNG CANCER CLINICAL PATHWAY Final Draft March 2015 Cancer Clinical Performance Group Radiation Oncology SBRT Workgroup Membership: Rex Hoffman, MD, Clinical Lead, Disney Family Cancer Center (Burbank,
More informationEVERGREEN: IMPROVING THE LIFETIME OF PLASTIC SCINTILLATORS
EVERGREEN: IMPROVING THE LIFETIME OF PLASTIC SCINTILLATORS Mike Kusner and Peter Menge Saint-Gobain Crystals, Hiram Ohio USA BACKGROUND: Thousands of plastic scintillators are deployed outdoors where they
More informationMAGNETOM Skyra: The Mannheim Perspective
MAGNETOM Skyra: The Mannheim Perspective Henrik J. Michaely; Stefan O. Schoenberg Institute of Clinical Radiology and Nuclear Medicine, University Medical Center Mannheim, University of Heidelberg, Germany
More informationMark van der Zande April 2 nd 2015 Sense of Contact 17. Magnetic Resonance Technology A new concept for Multiphase Flow Measurement
Mark van der Zande April 2 nd 2015 Sense of Contact 17 Magnetic Resonance Technology A new concept for Multiphase Flow Measurement 1. KROHNE 2. Multiphase flow in the oil industry 3. Magnetic Resonance
More informationThermoelectric Detection of Spherical Tin Inclusions in Copper by Magnetic Sensing
Thermoelectric Detection of Spherical Tin Inclusions in Copper by Magnetic Sensing Hector Carreon, Prof. Peter B. Nagy, and Prof. Adnan H. Nayfeh Department of Aerospace Engineering and Engineering Mechanics
More informationLight dosimetry for Low-Level Laser therapy: Accounting for differences in tissue and depth
Light dosimetry for Low-Level Laser therapy: Accounting for differences in tissue and depth Robert Weersink a, Roger White b, Lothar Lilge c a Laboratory for Applied Biophotonics, University Health Network,
More information2009 Strategic Planning Meeting SNM Clinical Trials Network Chicago, IL
2009 Strategic Planning Meeting SNM Clinical Trials Network Chicago, IL Wednesday, September 30, 2009 AGENDA Item BREAKFAST Welcome and Introductions Time 8:00am 8:15 Discussion Leaders M. Graham 2009
More information