Technical Report Electronic Cigarette Aerosol Parameters Study

Size: px
Start display at page:

Download "Technical Report Electronic Cigarette Aerosol Parameters Study"

Transcription

1 E-Cigarette Task Force Technical Report 2014 Electronic Cigarette Aerosol Parameters Study March 2015 Task Force Coordinator: Chuck Garner, Ph.D. R. J. Reynolds, U.S.A. Task Force Secretary: Robert D. Stevens, Ph.D. Lorillard Tobacco Company, U.S.A. Author: Rana Tayyarah Lorillard Tobacco Company, U.S.A

2 Table of Contents 1. Summary 3 2. Introduction 3 3. Organisation Participants Protocol 4 4. Data - Raw 6 5. Data - Statistical Analysis Exclusion of Outliers Calculation of Repeatability and Reproducibility Calculation of Z-Scores 7 6. Data Interpretation Machine or product performance issues Impact of regime on overall yields Impact on relative yields of analytes Impact of regime on cumulative yield or puff trending Trend with flow rate among the test parameters Recommendations Additional Considerations Acknowledgements 12 APPENDIX A: Study Protocol and Data Template 13 APPENDIX D: Total Yield Graphs 80 APPENDIX E: Puff Set Graphs 84 APPENDIX F: ACM Variability - Percent Relative Standard Deviation for Puff Sets 88 ECIG-CTR Aerosol Parameters March /89

3 1. Summary In 2014, the CORESTA E-Cigarette Task Force (TF) conducted a collaborative study that included testing of commercial electronic cigarette products. This study was referred to as E- Cigarette Aerosol Parameters Study. The laboratory phase of this study was conducted from late 2013 to early 2014 and the results were presented to the Task Force in May The study included determinations of puff count, trapped aerosol, propylene glycol, nicotine, water, glycerin, and product weight loss. Four laboratories participated in the study. Each laboratory tested two products of choice. The goal of this work was to compare results from cigarette-machine puffing of e-cigarettes under a range of puffing parameters to make a recommendation for standardized parameters. From the previous TF meeting, parameters of 55mL volume taken over 3s every 30s using a square profile shape were favored. Thus, the more specific objective was to determine if this regime should be recommended as the standard or if any of the other regimes included in this study should be recommended. The study results demonstrate that any of the regimes evaluated would be suitable for machinetesting of e-cigarettes of a conventional design 1 for the determination of major constituents in trapped aerosol. Parameters of 55mL/3s/30s/square will be used for an upcoming proficiency test. These parameters will serve as a recommended machine regime for e-cigarette aerosol collection. 2. Introduction Though electronic cigarettes do not specifically contain tobacco, this is a relatively new product category that is of interest to many CORESTA members. A task force related to electronic cigarettes was formed by the Scientific Commission after a November 2012 workshop. One of the short term objectives of the Task Force is to gather and share preliminary data on analysis relevant to e-cigarettes worldwide with a view to making recommendations for product testing. As yet, there is no standardized method for puffing electronic cigarettes. Thus, evaluation of published literature and conductance of collaborative studies is hindered. At the September 2013 Task Force meeting, there was extensive discussion regarding puffing parameters 2. From this discussion it was clear that parameters of 55mL volume taken over 3s every 30s using a square profile shape were favored by the majority of the TF members but that there may be some benefit to an alternate volume (70mL) or an alternate puff duration (4s). Though typical for cigarette smoking, 35 ml puff volume, bell-shaped profiles, and 2s puff duration were each recommended against by researchers as being problematic from some e-cigarette products. The primary concerns discussed related to insufficient product trigger (time/heat), insufficient overlap between the machine (bell) profile and/or machine duration and the start and end times for some products puff time. Puff durations of 5s or longer were recommended against as being too long from some products. In order to prepare for a future proficiency study, the Task Force commissioned a study to determine the effect of puff profile on the yield of the major constituents in e-cigarette aerosols in order to recommend puffing parameters for aerosol collection. The study involved comparison of puffing parameters combinations including 55mL and 70mL puff volumes of 3s or 4s duration. Each participating laboratory tested two products of choice for glycerin, propylene glycol, water, and nicotine using six regimes. Since laboratories did not necessarily test the same products, no comparisons were made between any of the products 1 Self-actuating electronic cigarettes, approximately 25mm circumference with a cylindrical mouth-end 2 See Task Force meeting minutes from the September 2013 meeting for additional detail. ECIG-CTR Aerosol Parameters March /89

4 results. Comparisons were limited to understanding the impact of the regimes on yield, consistency of results, and reliability of the testing equipment for each product sample. Essentially, the study data were treated as eight individual studies of the impact of puffing parameters. 3. Organisation 3.1 Participants A list of the participating laboratories is provided in Table 1. The laboratories are listed in alphabetical order. The numerical laboratory codes (1-4) used in this report do not necessarily correspond to the same order as the list below. Table 1: List of Participating Laboratories Laboratory Altria Client Services, Inc. Arista Laboratories Enthalpy Analytical, Inc. Lorillard Tobacco Company 3.2 Protocol Participating laboratories supplied their own samples for testing. White Cloud and Red Kiwi supplied two of the laboratories with products. It was requested that two models, one disposable and one rechargeable, be tested. Sample identification was withheld with data reporting; samples were only identified as rechargeable or disposable. Participants were directed to store products at room temperature and that batteries be fully charged prior to use for the rechargeable products. Each laboratory conducted five replicate determinations for each product (Table 2) under each of the six regimes (Table 3). Test 1 (ISO) and Test 2 (Health Canada) were included in the study for informational purposes only. Products were tested for one battery charge to a maximum 300 puffs. Puffs were collected in 25 puff increments in a manner typical for each laboratory for e-cigarette aerosol collection. e-cigarettes were weighed between each set of 25 puffs for calculation of weight loss. Each laboratory tested the samples for the specified analytes using methods of choice. See Appendix A for the study protocol. ECIG-CTR Aerosol Parameters March /89

5 Table 2: Sample Identification Sample ID Description Lab1-Rechargeable Rechargeable product tested by Laboratory 1 Lab1-3 piece Rechargeable 3 piece-style product tested by Laboratory 1 Lab2-Rechargeable Rechargeable product tested by Laboratory 2 Lab2-Disposable Disposable product tested by Laboratory 2 Lab3-Rechargeable Rechargeable product tested by Laboratory 3 Lab3-Disposable Disposable product tested by Laboratory 3 Lab4-Rechargeable Rechargeable product tested by Laboratory 4 Lab4-Disposable Disposable product tested by Laboratory 4 Table 3: Collection Puffing Parameters Regime Code Puff Vol (ml) Duration (s) Shape Frequency Mean Flow (ml/s) Test 1 * 35/2/60/B 35 2 Bell Every 60s 17.5 Test 2 * 55/2/30/B 55 2 Bell Every 30s 27.5 Test 3 55/3/30/Sq 55 3 Square Every 30s 18.3 Test 4 55/4/30/Sq 55 4 Square Every 30s 13.8 Test 5 70/3/30/Sq 70 3 Square Every 30s 23.3 Test 6 70/4/30/Sq 70 4 Square Every 30s 17.5 *Note: Regime 1 and Regime 2 were not under evaluation because of previously reported issues with some products, but were included for informational purposes because they are based on standardized methods (ISO 3308 and HC-T-115) for traditional cigarette smoking. Where a bell shape profile has a non-constant flow and a square profile has constant flow ECIG-CTR Aerosol Parameters March /89

6 The study included determination of trapped aerosol, propylene glycol, nicotine, water, glycerin, and product weight loss. The aerosol collected mass (ACM) which is the measure of total particulates collected and the weight loss of the e-cigarette were determined gravimetrically. Analytes were determined using the method of choice by each laboratory. A comparison of the methods is shown below in Table 4. See Appendix B for summaries of each lab s methods. Table 4: Analytical Methods Overview Collection Sample Prep Analysis In Common Cerulean SM450 Standard Puff Engines ( 70mL puff) 44mm CFP, Conditioned to ISO std Room conditioned to ISO smoking std Organic Solvent 10-20mL Shake to extract Possible additional dilution for Gly/PG GC-TCD for water Packed column (most) GC-FID for other analytes Capillary column DB-Wax columns for most 6 12 calibration standards Linear (IS) calibration (most) r 2 >0.99 Unique Procedures Lab 4 used smaller labyrinth seals Lab 1 IPA (10mL) 30min Lab 2 EtOAc (20mL) 30min Lab 3 IPA (10mL) 20min Lab 4 n-propanol (20mL) 30min Lab 1 Internal Standard: EtOH (water), heptadecane (other analytes) Stabilwax column (nicotine, glycerin, PG) Lab 2 Internal Standard: 1,3-butanediol DBWax column for water Lab 3 Internal Standard: EtOH (water), anethole (other analytes), 3 GC runs, Linear (Ext) calibration for glycerin, PG Lab 4 Internal Standard: quinoline (nic); 1,4- butanediol (other analytes), quadratic (IS) calibration for water, glycerin 4. Data - Raw The summary descriptive statistics data are listed in Appendix C where each mean represents five replicates. Raw data plots are given in Appendix D. The data are displayed as mean with error bars representing one standard deviation. 5. Data - Statistical Analysis The purpose of the study was to gather information in preparation for a proficiency study. This study was not a formal proficiency test; there were four participating laboratories testing samples which were not in common using methodology which were not in common. Thus, many of the statistical analyses typically conducted following the recommendations given in ISO (1994) were not applicable to this data set. ECIG-CTR Aerosol Parameters March /89

7 5.1 Exclusion of Outliers Traditional outlier tests were not applied as discussed above. Data for devices that clearly did not function were removed by visual observation. When collecting puff-by-puff (or puff sets, in this case), puffs near the end of the device lifetime, as indicated by significant decrease in ACM yields or from a physical signal (such as a blinking LED) from the product, are typically of a very high variability. End-of-life puff sets yields were included in total yield calculations but the variability information was not included in the overall assessment of the product s variability under a given regime. In general, for puff set collection if a product or device had a lower than typical yield for a puff set the next puff set would have a higher than typical yield. Therefore, even though there may be puff set to puff set variability for a device, the cumulative yields may be consistent with the other replicates. This must be taken into account when reviewing the results. Also, analytes tend to track with total ACM; if the ACM is low for a measurement all other analytes are low by a similar magnitude. Thus, data were normalized to ACM in order to note any values that did not track with ACM for outlier investigation. Sample Lab1-3 piece was more variable overall compared to the other samples. The laboratory reported an experimental issue that was likely to be the cause of the apparent lower reliability of this sample. The sample s raw data are included here but the sample s results were not as heavily considered with regard to the overall conclusions drawn from the study. 5.2 Calculation of Repeatability and Reproducibility Each laboratory tested unique samples using non-standardized methodology. Thus, repeatability (r) and reproducibility (R) results were not calculated for this data set. 5.3 Calculation of Z-Scores Z-score values cannot be calculated for the data set. ECIG-CTR Aerosol Parameters March /89

8 6. Data Interpretation The objective of the study was to determine if 55mL of a 3s taken every 30s in a square profile (55/3/30/Sq) should be recommended as the standard OR if any of the other regimes should be recommended. To that end, we posed several questions to answer as we evaluated the data. 6.1 ISO and Canadian Intense Testing using ISO (Test 1) and Canadian Intense (Test 2) puffing parameters were included in the study for informational purposes but were not included in the evaluation for recommending a regime. Exclusion of these regimes was based on discussion at the September 2013 task force meeting during which multiple task force members discussed experience with device failure 3 when using bell shape puff profiles. In this study, there were no issues with product performance with either of these regimes. Simply based on the results of this study, either of these regimes would be viable choices if appropriate for a particular study design and product type. The 25 puff block yields were lower for these regimes. Thus, life-time puffing experiments with these parameters would require more puffs (longer collection times) than would be required for testing with the other regimes evaluated in this study. 6.2 Machine or product performance issues At least until e-cigarette-specific aerosol collection machines become more in use, it is important to ensure that the recommended regime will be compatible with conventional smoking machines. Indeed in this study, each participant used conventional 20-port linear smoking machines equipped with standard puff engines capable of a maximum puff volume of 70mL. No technical issues were noted with 55mL puff volume or with either duration used. There were a few technical issues noted for tests with 70mL puff volumes possibly because this is the maximum volume for the puff engines used. One lab reported issues with machine failure with 70mL puff volumes. Two laboratories reported that smoking machine modifications, such as air pressure modifications, were necessary for testing using 70mL puff volumes. No product failures were attributed to any of the smoking regimes. Based on machine performance, 55mL puff volume appears to be more favorable than 70mL. 6.3 Impact of regime on overall yields The aerosol collected mass (ACM) is an indication of overall performance. Results were compared for each sample type between the test regimes. Results were not compared between samples, as there is no expectation that any of the samples were in common between the laboratories. Overall yields were similar for the four regimes under consideration (Tests 3-6 Table 3); no practical or statistically significant differences were noted for any of the products. Precision for overall yields, as displayed in Appendix D, was similar for each of the regimes, as well. Additionally, a comparison between e-cigarette mass loss and ACM indicated that results were similar between the regimes tested. Ideally, the mass loss of the e-cigarette and the weight gain of the particulate trap would be virtually the same. As displayed below, the ratio of mass loss to ACM values were near 1 in most cases. 3 No or slow activation ECIG-CTR Aerosol Parameters March /89

9 Table 5: Ratio of e-cigarette Mass Loss to ACM Lab1- Rechargeable Lab2- Disposable Lab2- Rechargeable Lab3- Disposable Lab3- Rechargeable Lab4- Disposable Lab4- Rechargeable 35/2/60/B /2/30/B /3/30/Sq /4/30/Sq /3/30/Sq /4/30/Sq Sample Lab1-3 piece was excluded from this comparison due to experimental issues; raw data are shown in Appendix C Thus, there is no advantage for any of the alternative regimes versus Test 3, 55/3/30/Sq, for overall (ACM) yields. ECIG-CTR Aerosol Parameters March /89

10 6.4 Impact on relative yields of analytes Another consideration is whether different regimes would give a different proportion of the analytes tested. Thus the percent of analyte in the ACM was calculated as Analyte (mg) / ACM (mg) x 100. As shown in the table below as the range and the percent difference in composition for each regime, in most cases the percent composition varied by no more than 10% among the regimes for a given sample except for water. For example, for sample Lab1- Rechargeable, the percent composition of nicotine was 1.22%, 1.24%, 1.21%, and 1.24% for Tests 3, 4, 5, and 6, respectively. This is a range of 0.03pt (as an absolute value) or 2% (calculated as percent difference). Thus, there appears to be no impact on the composition of the ACM based on regime choice for the regimes tested. Table 6: Range in Analyte Percent Composition across the Regimes for each Sample As % Composition Total PG Total Nicotine Total Glycerin Total Water Lab1-Rechargeable Lab2-Disposable Lab2-Rechargeable Lab3-Disposable NA Lab3-Rechargeable NA Lab4-Disposable Lab4-Rechargeable As % Difference* in % Composition Total PG Total Nicotine Total Glycerin Total Water Lab1-Rechargeable 3% 2% 4% 21% Lab2-Disposable 4% 4% 5% 33% Lab2-Rechargeable 3% 3% 4% 32% Lab3-Disposable NA 1% 6% 3% Lab3-Rechargeable NA 2% 7% 9% Lab4-Disposable 10% 22% 16% 21% Lab4-Rechargeable 19% 8% 6% 17% *Calculated as (2(Max %Composition-Min %Composition))/(Max+Min)x100 Sample Lab1-3 piece was excluded from this comparison due to experimental issues; raw data are shown in Appendix C ECIG-CTR Aerosol Parameters March /89

11 6.5 Impact of regime on cumulative yield or puff trending The ACM was collected discretely rather than in total; 25 puffs were collected per pad up to 300 puffs. Thus delivery profile trends may be assessed. While the overall yields were similar among the regimes, the delivery profiles were impacted by puff duration (2s vs 3s), but not by puff volume. This trending is exemplified below for one of the samples. Trend lines for the regimes using 3s duration, 55/3/30/Sq and 70/3/30/Sq, are virtually identical. While trend lines for the 4s duration regimes, 55/4/30/Sq and 70/4/30/Sq, also group together. Note also that the 4s trends have higher yields and fewer puff sets than the 3s trends. All trend lines are displayed in Appendix E. While there appears to be a pattern to the puff set profiling based on duration, there is no advantage for any of the alternative regimes versus Test 3, 55/3/30/Sq, for cumulative yield. ACM (mg/pad) Test 3-55/3/30/Sq Test 4-55/4/30/Sq Test 5-70/3/30/Sq Test 6-70/4/30/Sq Puff Number Figure 1: Puff Set Trending for Sample Lab 2 Rechargeable ACM (mg) Yield per pad. 6.6 Difference in yield variability The relative standard deviation (%RSD) was calculated for each pad set to determine if any of the regimes appeared to have improved precision. End-of-unit life puffs would be expected to be highly variable no matter the regime. Thus partial last puff sets were not considered in the evaluation of variability. Collection variability results calculated as the percent relative standard deviation (%RSD) of the ACM delivery for each pad set are shown in Appendix F. Puff set precision was relatively high and generally similar among and between the sets. For all products tested, 80% of the RSD values were less than 15% regardless of regime. There is no advantage for any of the alternative regimes versus Test 3, 55/3/30/Sq, for variability. 6.7 Trend with flow rate among the test parameters One of the questions in the case of product failure for machine vaping relates to whether the flow rate is sufficient to activate the device. At the September 2013 meeting, TF participants discussed device failure issues for some products if the flow rate was not sufficient for device activation. A flowrate of 13.8 ml/s (55mL/4s) was specifically discussed as problematic. In this study, flow rates ranged from 13.8 ml/s to 23.3 ml/s and there were no apparent product failures due to regime (i.e. flow rate). There was no clear trending in yield when compared based on flow rate. Note that the number and range of products tested was limited. Thus, flow rate may still be a concern for less conventional products. Thus, the task force would be hesitant to recommend a regime using 55mL with 4s without additional testing. ECIG-CTR Aerosol Parameters March /89

12 7. Recommendations Based on discussions during TF meetings: A 70-mL puff volume may be problematic for smoking machine designs with 70mL as the maximum volume. Parameters which result in a low flow rate (i.e. 55mL with a 4s duration) may result in device failure depending on the product being tested. Puffing parameters using bell profile shapes may be problematic for some product types due to air flow issues. Based on results from this study: Overall, yields were similar for the four test regimes. Delivery profile trends were dependent on puff duration and not on puff volume. Precision was acceptable and similar for all regimes. There were no technical advantages among the regimes tested. Thus, based on TF discussion and on the results of this study, Test Regime 3 (a square 55mL puff, 3s in duration, taken twice a minute), was recommended for adoption as an interim standard vaping regime for e-cigarette testing. Choice of regime parameters may need to be refined as additional product types are tested. 8. Additional Considerations Limitations to the work that should be considered include: Only products of a conventional design were included in the study. Puff volume and combinations of volume and duration were limited by capabilities of conventional smoking machines in standard configuration. Reference products and standardized analytical methodology do not exist for e- cigarette aerosol collection at this time. Reporting units that may be useful for fostering a collective understanding of the research include mg/pad for incremental collection, mg/ total puffs for total collection on a single pad. In addition to analytes of interest collection and reporting of ACM, puff counts, and e- cigarette weight change are useful quality measures and would aid collaborative understanding of results. 9. Acknowledgements The study team, Drs. Rob Stevens, Narendra Meruva, and Gene Gillman; the participating laboratories; and White Cloud and Red Kiwi that supplied some of the e-cigarettes for testing are greatly appreciated. ECIG-CTR Aerosol Parameters March /89

13 APPENDIX A: Study Protocol and Data Template CORESTA E-CIGARETTE ANALYTICAL TESTING SUB TEAM (Team Leader Dr. Rob Stevens, Lorillard Tobacco) Project Title: 2013 Major constituents in e-cigarette aerosol Date: Nov 11, 2013 Written by: Gene Gillman, Ph.D. (Enthalpy Analytical) Phone: Confidentiality Notice: All data generated in the course of this analysis should be handled in the strictest of confidence by all participating laboratories and other taskforce members. ECIG-CTR Aerosol Parameters March /89

14 Study Protocol to Determine the Effect of Puff Profile on the Yield of the Major Constituents in E-Cigarette Aerosol 1. Objective The main objective of this study to determine the effect of puff profile on the yield of the major constituents in e-cigarette aerosol. This study will include six puff profiles or test conditions with various puff volumes, durations and puff shapes. 2. Study coordinator Dr. Robert Stevens Director, R&D Testing Services Lorillard Tobacco Company, R&D 420 N. English Street P.O. Box Greensboro, N.C Bus. (336) FAX (336) Analytes and Methods 3.1 Puff Count Puff count for each puffing increment (each pad) will be recorded. For most data points the puff count will be 25. For products which extinguish prior to 300 puffs there may be pads with fewer than 25 puffs. The actual puff count should be recorded. Record as count/pad in the data template. 3.2 Trapped Aerosol The total collected aerosol mass (ACM) for each sample will be determined by measuring the trapping system before and after each collection event. The total amount of aerosol collected will be determined by difference in mass (Pad weight after collection pad weight before collection). Record as mg/pad in the data template. This test measurement is akin to total particulate matter (TPM) for conventional cigarette testing. 3.3 Weight loss The amount of material lost from each e-cigarette will be determined by measuring the e- cigarette before and after each collection event. The total amount of weight loss will be determined by difference in mass (e-cigarette weight after collection e- cigarette weight before collection). Record as mg in the data template. 3.4 Propylene Glycol The amount of propylene glycol per sample will be determined by the method currently in use in each laboratory. Record as mg/pad in the data template. 3.6 Glycerin The amount of glycerin per sample will be determined by the method currently in use in each laboratory. Record as mg/pad in the data template. 3.7 Water The amount of water per sample will be determined by the method currently in use in each laboratory. Record as mg/pad in the data template. ECIG-CTR Aerosol Parameters March /89

15 4. Samples 4.1 Required materials Each participating laboratory shall supply product for their own testing. Two models shall be tested. These should include both a rechargeable model and disposable model. These two brands do not have to be products from the same manufacturer. A minimum of 30 e-cigarettes for each brand is required to fulfill the study design. 4.2 Sample Handling All samples should be stored at room temperature. Rechargeable devices should be charged before use. A single battery charge will be used for each new cartomizer. Note: Only one battery charge will be used for each cartomizer. 5. Experimental plan Five replicates for each e-cigarette under each test condition is required. If possible, it is requested that all samples for a test condition be analyzed on a single day. A series of discreet samples will be collected from each e-cigarette (disposable or cartomizer). For each e-cigarette, samples will be collected in 25 puffs increments (1-25, and etc.) until the battery is depleted. Five e-cigarettes should be tested under each test condition. For each aerosol sample, laboratories are asked to determine weight lost from the device, amount of ACM trapped, nicotine, propylene glycol, glycerin, and water. Data will be reported on a per mg basis. Aerosol should be generated from each type of e-cigarette using the test conditions given in Table 1. Note that Tests 1 and 2 use puffing parameters specified in for ISO smoking (ISO 3308) and Canadian Intense smoking (HC-T-114) of conventional cigarettes. This protocol does not attempt to make any other specifications with regard to these two standard smoking regimes (i.e. room conditions). Note: Five e-cigarettes should be tested under each test condition. Table 1: Test Conditions to be used for this study Test Puff Volume Shape Frequency Duration Not 1 35 Bell* Every 60 Sec 2 Sec ISO 3308 Puffing Parameters 2 55 Bell* Every 30 Sec 2 Sec HC-T-114 Puffing Parameters (without tape) 3 55 Square # Every 30 Sec 3 Sec 4 55 Square # Every 30 Sec 4 Sec 5 70 Square # Every 30 Sec 3 Sec 6 70 Square # Every 30 Sec 4 Sec * Bell Shape is an ISO 3308 type puff profile # Square Shape is a constant flow rate profile Note: Rechargeable devices must be tested using a new cartomizer and fully charged battery. Samples will be collected until the battery is discharged. Only one battery charge is to be used per cartomizer. ECIG-CTR Aerosol Parameters March /89

16 6. Data Submission The attached templates should be used for data submission. Please supply data in the requested format without creating new cells or rows in the spreadsheet. Results should be reported back to Rana Tayyarah on or before Feb 14 th, 2014 in order to allow sufficient time to evaluate the data. 7. Data Analysis The data will be analyzed statistically according to ISO Timescale First week of December 2013 All documents will be distributed to participating laboratories o Protocol of the study o Report form December 2013/January 2014 Laboratories will perform the study Mid February The results must be sent to the data coordinator on or before February 14 th 2014 in order to give to the statistician sufficient time to evaluate the data. Data received after February 14 th will not be included in the draft report discussed at the May meeting. March/April 2014 The statistician will perform the data analysis and the Task Force coordinators will format the data for presentation. May 2014 The taskforce member will assess the results of the collaborative study during the next taskforce meeting. ECIG-CTR Aerosol Parameters March /89

17 Data Sheet for 2013 Major Constituents in e-cigarette Aerosol Parameters Study Laboratory name Company Representative (Title, Name) Address Tel Fax address Note: Please don't modify formatted data sheet for the convenience of data analyses. Note: Laboratory names will not be used for reporting. All labs will be coded. Each lab will know their own code. Note: In this spreadsheet, cells that are shaded blue require input. Cells that are not shaded are calculations or information and should not be edited. Note: Please return this completed spreadsheet to by February 14, ECIG-CTR Aerosol Parameters March /89

18 0 Method Synopsis: please type an overall summary of your method in this box and provide specific details below Method Details (example) ACM Nicotine Water Glycerin PG Sample Handling Storage prior to testing Additional Details Type of Aerosol Collection Machine Style (Include a description of features if non-standard equipment was used) unopened at 65% RH 72F Linear Smoking Machine Make Vaping was Concurent for Which analytes? Room Conditions during aerosol collection Cerulean SM450 All were concurrent 65 %RH, 72F Additional Details for Aerosol Collection Trapping System Filter Pad / Diameter Were filter pads conditioned? Trap Type + Number of Traps CFP 44 mm Yes, 62+/-%RH, 75F Fritted Impinger Solvent Extraction Additional Filtration Additional sample processing prior to analysis Additional Details 20 ml MeOH (sub-ambient) wrist-action shaker, 30min PTFE: 0.45um extracts were diluted 1:2 for Gly/PG extracts analyzed same day for Nic/Water next day for Gly/PG Analytical Equipment Description Agilent GC/MS, EI in SIM mode using detuerated IS; D-5 pyridine (84), pyridine(79) & D-7 quinoline(136) & quinoline(129); Detection Systems + Conditions Column: DBWax 30mX0.25mmX0.25Um, 50oC for 6min, 4oC/min to 160oc, 30oC to 250oC for 5.5min: 10:1 split, flow 15mL/min, total flow 20ml/min, 1ul injection extracts were stored 4C prior to analysis Additional Details Reference -- Quantitation Information Limit of Quantitation (mg/pad) Limit of Detection (mg/pad) Number of standards Concentration of each Standard Internal Standard Calibration Type r 2 Additional Details Ethanol (water), Anethole (others) linear, with IS *Note: the example listed is not an e-cig method. It is just meant to show the requested level of detail for you to list for your aerosol method. ECIG-CTR Aerosol Parameters March /89

19 0 Please paste in a picture of an example chromatogram for a calibration standard and for one or more samples for chromatographic methods Please paste pictures relevant to any unusual observations noted during testing. Please include text explanation as necessary. ECIG-CTR Aerosol Parameters March /89

20 0 Condition Volume Shape Frequency Puff Duration NOTES Test 1 35 Bell Every 60 Sec 2 Sec ISO 3308 Puffing Parameters Test 2 55 Bell Every 30 Sec 2 Sec HC-T-114 Puffing Parameters (without tape) Test 3 55 Square Every 30 Sec 3 Sec Test 4 55 Square Every 30 Sec 4 Sec Test 5 70 Square Every 30 Sec 3 Sec Test 6 70 Square Every 30 Sec 4 Sec ECIG-CTR Aerosol Parameters March /89

21 0 Sample 1 - Test 1 Sample Description Note: e-cigarette brand names are not needed. At a minimum, indicate "rechargable" or "disposable" as a sample description. Test Conditions Puff Volume (ml) Shape Frequency Duration Test 1 35 Bell Every 60 Sec 2 Sec Testing Dates Test 1 Sample 1 Collection Extraction Analyses Observations Note: Please include under "observations" information about anything unusual in the behavior, etc of the samples. For example, unusual colors, smells, "spitting" (droplets inside pad holder and/or on pad). Note: If photograph(s) would be helpful to demonstrate an unusual observation, please paste in on the chromatogram and photos tab with text to explain. Testing Summary Total Puffs Total ACM Total PG Total Nicotine Total Glycerin Total Water mg mg mg mg mg ECIG-CTR Aerosol Parameters March /89

22 0 Sample 1 - Test 1 Test Conditions Puff Volume Shape Frequency Duration Test 1 35 Bell Every 60 Sec 2 Sec Note: Puffing should be ceased after one battery discharged OR after 300 puffs, whichever comes first. Puff Count Puffs Replicate 1 Replicate 2 Replicate 3 Replicate 4 Replicate Note: Later puff sets may not be a full 25 puffs. Record the actual puff counts for these pads. Weight Loss Puffs Replicate 1 Replicate 2 Replicate 3 Replicate 4 Replicate 5 (mg/device) ECIG-CTR Aerosol Parameters March /89

23 0 Sample 1 - Test 1 Test Conditions Puff Volume (ml) Shape Frequency Duration Test 1 35 Bell Every 60 Sec 2 Sec ACM Puffs Replicate 1 Replicate 2 Replicate 3 Replicate 4 Replicate 5 (mg/pad) PG Puffs Replicate 1 Replicate 2 Replicate 3 Replicate 4 Replicate 5 (mg/pad) ECIG-CTR Aerosol Parameters March /89

24 0 Sample 1 - Test 1 Test Conditions Puff Volume (ml) Shape Frequency Duration Test 1 35 Bell Every 60 Sec 2 Sec Nicotine Puffs Replicate 1 Replicate 2 Replicate 3 Replicate 4 Replicate 5 (mg/pad) Glycerin Puffs Replicate 1 Replicate 2 Replicate 3 Replicate 4 Replicate 5 (mg/pad) Water Puffs Replicate 1 Replicate 2 Replicate 3 Replicate 4 Replicate 5 (mg/pad) ECIG-CTR Aerosol Parameters March /89

25 APPENDIX B: Laboratory Method Summaries and Example Chromatograms Laboratory 1 Method Summary Nicotine Water Glycerin PG Sample Handling Storage prior to testing unopened at 65% RH 72F unopened at 65% RH 72F unopened at 65% RH 72F unopened at 65% RH 72F Additional Details Type of Aerosol Collection Machine Style (Include a description of features if non-standard equipment was used) Linear Smoking Machine Linear Smoking Machine Linear Smoking Machine Linear Smoking Machine Make Cerulean SM450 Cerulean SM450 Cerulean SM450 Cerulean SM450 Vaping was Concurrent for Which analytes? All were concurrent All were concurrent All were concurrent All were concurrent Room Conditions during aerosol collection 65 %RH, 72F 65 %RH, 72F 65 %RH, 72F 65 %RH, 72F Additional Details for Aerosol Collection ECIG-CTR Aerosol Parameters March /89

26 Laboratory 1 Method Summary (continued) Trapping System Nicotine Water Glycerin PG Filter Pad / Diameter CFP 44 mm CFP 44 mm CFP 44 mm CFP 44 mm Were filter pads conditioned? Yes, 62+/-%RH, 75F Yes, 62+/-%RH, 75F Yes, 62+/-%RH, 75F Yes, 62+/-%RH, 75F Trap Type + Number of Traps Solvent 10 ml 2-Propanol 10 ml 2- Propanol 10 ml 2-Propanol 10 ml 2-Propanol Extraction wrist-action shaker, 30min wrist-action shaker, 30min wrist-action shaker, 30min wrist-action shaker, 30min Additional Filtration Additional sample processing prior to analysis Additional Details extracts analyzed same day for Nic/Water /Gly/PG extracts analyzed same day for Nic/Water /Gly/PG extracts analyzed same day for Nic/Water /Gly/PG extracts analyzed same day for Nic/Water /Gly/PG ECIG-CTR Aerosol Parameters March /89

27 Laboratory 1 Method Summary (continued) Analytical Equipment Nicotine Water Glycerin PG Description Agilent 58/90 GC Series II FID/TCD Agilent 5890 GC Series II FID/TCD Agilent 5890 GC Series II FID/TCD Agilent 5890 GC Series II FID/TCD Detection Systems + Conditions Column: Restek Stabilwax 30mX0.32mmX1.0Um, 1ul injection Column: Supelco Porapak R 6ft. X 1/8 inch ID Column: Restek Stabilwax 30mX0.32mmX1.0Um, 1ul injection Column: Restek Stabilwax 30mX0.32mmX1.0Um, 1ul injection Additional Details Reference ECIG-CTR Aerosol Parameters March /89

28 Laboratory 1 Method Summary (continued) Quantitation Information Nicotine Water Glycerin PG Limit of Quantitation (mg/pad) Limit of Detection (mg/pad) Number of standards Concentration of each Standard , 0.103, 0.257, 0.514, 0.771, 1.03, 1.23 mg/ml 0.495, 1.24, 2.48, 3.71, 4.95, 5.94 mg/ml 0.331, 0.662, 1.66, 3.31, 4.97, 6.62, 7.95, 9.93, 13.2, 33.1 mg/ml 0.255, 0.510, 1.27, 2.55, 3.82, 5.10, 6.11, 7.64, 10.2, 25.5 mg/ml Internal Standard Ethanol (water), Heptadecane (others) Ethanol (water), Heptadecane (others) Ethanol (water), Heptadecane (others) Ethanol (water), Heptadecane (others) Calibration Type linear, with IS linear, with IS linear, with IS linear, with IS r Additional Details ECIG-CTR Aerosol Parameters March /89

29 Figure 1: Laboratory 1 Example Chromatogram Aerosol Sample Figure 2: Laboratory 1 Example Chromatogram Aerosol Sample ECIG-CTR Aerosol Parameters March /89

30 Figure 3: Laboratory 1 Example Chromatogram Calibration Standard Figure 4: Laboratory 1 Example Chromatogram Calibration Standard ECIG-CTR Aerosol Parameters March /89

31 Laboratory 2 Method Summary Nicotine Water Glycerin PG Sample Handling Storage prior to testing Unopened; room temp. Unopened; room temp. Unopened; room temp. Unopened; room temp. Additional Details N/A N/A N/A N/A Type of Aerosol Collection Machine Style (Include a description of features if non-standard equipment was used) Linear Smoking Machine Linear Smoking Machine Linear Smoking Machine Linear Smoking Machine Make Cerulean SM450 Cerulean SM450 Cerulean SM450 Cerulean SM450 Vaping was Concurrent for Which analytes? All were concurrent All were concurrent All were concurrent All were concurrent Room Conditions during aerosol collection 65 %RH, 72F 65 %RH, 72F 65 %RH, 72F 65 %RH, 72F Additional Details for Aerosol Collection N/A N/A N/A N/A ECIG-CTR Aerosol Parameters March /89

32 Laboratory 2 Method Summary (continued) Trapping System Nicotine Water Glycerin PG Filter Pad / Diameter CFP 44 mm CFP 44 mm CFP 44 mm CFP 44 mm Were filter pads conditioned? Yes, 60±2% RH; 22±2 C Yes, 60±2% RH; 22±2 C Yes, 60±2% RH; 22±2 C Yes, 60±2% RH; 22±2 C Trap Type + Number of Traps N/A N/A N/A N/A Solvent 20 ml EtOAc 20 ml EtOAc 20 ml EtOAc 20 ml EtOAc Extraction wrist-action shaker, 30min wrist-action shaker, 30min wrist-action shaker, 30min wrist-action shaker, 30min Additional Filtration N/A N/A N/A N/A Additional sample processing prior to analysis N/A N/A N/A N/A Additional Details N/A N/A N/A N/A ECIG-CTR Aerosol Parameters March /89

33 Laboratory 2 Method Summary (continued) Analytical Equipment Nicotine Water Glycerin PG Description Agilent 7890 GC-FID Agilent 6890 GC-TCD Agilent 7890 GC-FID Agilent 7890 GC-FID Detection Systems + Conditions Column: RTXWax (30mx0.32mmx1µm); Split 20:1; 120 C to 250 C (11 C/min ramp); 1 minute hold; 1 µl injection volume Column: DBWax (30mx0.53mmx1µm); 50 C to 240 C (20 C/min ramp); 1 minute hold; 0.5 µl injection volume Column: RTXWax (30mx0.32mmx1µm); Split 20:1; 120 C to 250 C (11 C/min ramp); 1 minute hold; 1 µl injection volume Column: RTXWax (30mx0.32mmx1µm); Split 20:1; 120 C to 250 C (11 C/min ramp); 1 minute hold; 1 µl injection volume Additional Details Reference ECIG-CTR Aerosol Parameters March /89

34 Laboratory 2 Method Summary (continued) Quantitation Information Nicotine Water Glycerin PG Limit of Quantitation (mg/pad) Limit of Detection (mg/pad) N/A N/A N/A N/A Number of standards Concentration of each Standard µg/ml µg/ml µg/ml µg/ml Internal Standard 1,3-butanediol 1,3-butanediol 1,3-butanediol 1,3-butanediol Calibration Type Linear; 1/x2 Linear; 1/x2 Linear; 1/x2 Linear; 1/x2 r Additional Details N/A N/A N/A N/A ECIG-CTR Aerosol Parameters March /89

35 Figure 1: Laboratory 2 Example Chromatogram Aerosol Sample Figure 2: Laboratory 2 Example Chromatogram Aerosol Sample ECIG-CTR Aerosol Parameters March /89

36 Figure 3: Laboratory 2 Example Chromatogram Calibration Standard Figure 4: Laboratory 2 Example Chromatogram Calibration Standard ECIG-CTR Aerosol Parameters March /89

37 Laboratory 3 Method Summary Nicotine Water Glycerin PG Sample Handling Storage prior to testing unopened at 65% RH 72F unopened at 65% RH 72F unopened at 65% RH 72F unopened at 65% RH 72F Additional Details Type of Aerosol Collection Machine Style (Include a description of features if non-standard equipment was used) Linear Smoking Machine Linear Smoking Machine Linear Smoking Machine Linear Smoking Machine Make Cerulean SM450 Cerulean SM450 Cerulean SM450 Cerulean SM450 Vaping was Concurrent for Which analytes? All were concurrent All were concurrent All were concurrent All were concurrent Room Conditions during aerosol collection 65 %RH, 72F 65 %RH, 72F 65 %RH, 72F 65 %RH, 72F Additional Details for Aerosol Collection One vape run, nic water concurrent, gly/pg concurrent One vape run, nic water concurrent, gly/pg concurrent One vape run, nic water concurrent, gly/pg concurrent One vape run, nic water concurrent, gly/pg concurrent ECIG-CTR Aerosol Parameters March /89

38 Laboratory 3 Method Summary (continued) Trapping System Nicotine Water Glycerin PG Filter Pad / Diameter CFP 44 mm CFP 44 mm CFP 44 mm CFP 44 mm Were filter pads conditioned? Yes, 62+/-%RH, 75F Yes, 62+/-%RH, 75F Yes, 62+/-%RH, 75F Yes, 62+/-%RH, 75F Trap Type + Number of Traps NA NA NA NA Solvent 10 ml IPA 10 ml IPA Extraction 20min shake 20min shake 20min shake 20min shake Additional Filtration NA NA NA NA Additional sample processing prior to analysis Additional Details extracts analyzed same day for Nic/Water next day for Gly/PG extracts analyzed same day for Nic/Water next day for Gly/PG Additional dilution for Gly based on ACM Samples were diluted if nec for gly Samples were diluted if nec for gly Laboratory 3 Method Summary (continued) Analytical Equipment Nicotine Water Glycerin PG Description GC-FID GC-TCD (dual column with Nic analysis) GC-FID GC-FID Detection Systems + Conditions Agilent DB-Wax (10 m x 0.18 mm x 0.3 µm film) Splitless (purge time 0.5 min, 50mL/min) psi 40 ml/min 400 ml/min 160 C hold for 0.5 minutes 80 C/minute to 200 C 50 C/ minute to 225 C hold 2.0 min. Stainless steel Packed column Haye Sep-Q 60/80 mesh; 3' x 1/4", or equivalent); on-column injection; He pressure 9psi; initial temp 160C DB-Wax Analytical GC column (15m X 0.25mm ID)(0.25 µm film thickness) DB-Wax Analytical GC column (15m X 0.25mm ID)(0.25 µm film thickness) Additional Details Reference ECIG-CTR Aerosol Parameters March /89

39 Laboratory 3 Method Summary (continued) Quantitation Information Nicotine Water Glycerin PG Limit of Quantitation (mg/pad) Limit of Detection (mg/pad) Number of standards Concentration of each Standard (mg/ml) 0, 0.05, 1.15, 0.20, 0.25, 0.50, 1.0 (mg/ml) 0, 0.25, 0.75, 1.25, 2.5, 5.0, 1.0 (mg/ml) 0.025, 0.099, 0.25, 1.2, 2.5 (mg/ml) , , 0.04, 0.16, 0.40, 2.0, 4.0 Internal Standard Ethanol (water), Anethole (others) Ethanol (water), Anethole (others) Ethanol (water), Anethole (others) Ethanol (water), Anethole (others) Calibration Type linear, with IS linear, with IS linear, with IS linear, with IS r 2 > > > > Additional Details ECIG-CTR Aerosol Parameters March /89

40 Figure 1: Laboratory 3 Example Chromatogram Calibration Standard Figure 2: Laboratory 3 Example Chromatogram Calibration Standard ECIG-CTR Aerosol Parameters March /89

41 Figure 3: Laboratory 3 Example Chromatogram Calibration Standard ECIG-CTR Aerosol Parameters March /89

42 Laboratory 4 Method Summary Nicotine Water Glycerin PG Sample Handling Storage prior to testing Samples stored under ambient conditions Samples stored under ambient conditions Samples stored under ambient conditions Samples stored under ambient conditions Additional Details Type of Aerosol Collection Machine Style (Include a description of features if non-standard equipment was used) Linear smoking machine Linear smoking machine Linear smoking machine Linear smoking machine Make Cerulean Cerulean Cerulean Cerulean Vaping was Concurrent for Which analytes? All were concurrent All were concurrent All were concurrent All were concurrent Room Conditions during aerosol collection 60 ± 4 %RH, 72 ± 5 F 60 ± 4 %RH, 72 ± 5 F 60 ± 4 %RH, 72 ± 5 F 60 ± 4 %RH, 72 ± 5 F Additional Details for Aerosol Collection Based on the collected aerosol mass, the amount of extraction solution to be added is adjusted Based on the collected aerosol mass, the amount of extraction solution to be added is adjusted Based on the collected aerosol mass, the amount of extraction solution to be added is adjusted Based on the collected aerosol mass, the amount of extraction solution to be added is adjusted ECIG-CTR Aerosol Parameters March /89

43 Laboratory 4 Method Summary (continued) Trapping System Nicotine Water Glycerin PG Filter Pad / Diameter CFP 44mm CFP 44mm CFP 44mm CFP 44mm Were filter pads conditioned? 60 ± 3 %RH, 72 ± 2 F 60 ± 3 %RH, 72 ± 2 F 60 ± 3 %RH, 72 ± 2 F 60 ± 3 %RH, 72 ± 2 F Trap Type + Number of Traps N/A N/A N/A N/A Solvent 20 ml of dry n-propanol extraction solution 20 ml of dry n-propanol extraction solution 20 ml of dry n- propanol extraction solution 20 ml of dry n- propanol extraction solution Extraction 30 min mixing time on an orbital shaker 30 min mixing time on an orbital shaker 30 min mixing time on an orbital shaker 30 min mixing time on an orbital shaker Additional Filtration N/A N/A N/A N/A Additional sample processing prior to analysis Additional Details Samples were diluted 1:10 with extraction solution for PG and Glycerin Care must be taken to ensure all glassware used for standard and sample preparation is free of residual water. Samples were diluted 1:10 with extraction solution for PG and Glycerin Care must be taken to ensure all glassware used for standard and sample preparation is free of residual water. Samples were diluted 1:10 with extraction solution for PG and Glycerin Care must be taken to ensure all glassware used for standard and sample preparation is free of residual water. Samples were diluted 1:10 with extraction solution for PG and Glycerin Care must be taken to ensure all glassware used for standard and sample preparation is free of residual water. ECIG-CTR Aerosol Parameters March /89

44 Laboratory 4 Method Summary (continued) Description Analytical Equipment Nicotine Water Glycerin PG Detection Systems + Conditions Additional Details Agilent GC equipped with dual hot split/splitless injection ports and dual detectors (FID and TCD). Columns: Two DB- ALC1 capillary columns (front and back), 30m x 0.32 mm ID, 1.8 µm film thickness. Oven Programming: 90 C for 1min, 15 C/min to 120 c, 40 C to 280 C for 2min, Bake time 2 min, total run time 9 min, Injection port temperatures of 250 C, GC-FID - 1µL injection at 25:1 split and carrier gas flow of 3mL/min. n-propanol extraction solvent is stored dried with molecular sieves (5 g per liter of extraction solution) to remove any trace levels of water contamination. Agilent GC equipped with dual hot split/splitless injection ports and dual detectors (FID and TCD). Columns: Two DB-ALC1 capillary columns (front and back), 30m x 0.32 mm ID, 1.8 µm film thickness. Oven Programming: 90 C for 1min, 15 C/min to 120 c, 40 C to 280 C for 2min, Bake time 2 min, total run time 9 min, Injection port temperatures of 225 C, GC-TCD - 2µL injection at 10:1 split and carrier gas flow of 1.5mL/min. n-propanol extraction solvent is stored dried with molecular sieves (5 g per liter of extraction solution) to remove any trace levels of water contamination. Agilent GC equipped with dual hot split/splitless injection ports and dual detectors (FID and TCD). Columns: Two DB- ALC1 capillary columns (front and back), 30m x 0.32 mm ID, 1.8 µm film thickness. Oven Programming: 90 C for 1min, 15 C/min to 120 c, 40 C to 280 C for 2min, Bake time 2 min, total run time 9 min, Injection port temperatures of 250 C, GC-FID - 1µL injection at 25:1 split and carrier gas flow of 3mL/min. n-propanol extraction solvent is stored dried with molecular sieves (5 g per liter of extraction solution) to remove any trace levels of water contamination. Agilent GC equipped with dual hot split/splitless injection ports and dual detectors (FID and TCD). Columns: Two DB- ALC1 capillary columns (front and back), 30m x 0.32 mm ID, 1.8 µm film thickness. Oven Programming: 90 C for 1min, 15 C/min to 120 c, 40 C to 280 C for 2min, Bake time 2 min, total run time 9 min, Injection port temperatures of 250 C, GC-FID - 1µL injection at 25:1 split and carrier gas flow of 3mL/min. n-propanol extraction solvent is stored dried with molecular sieves (5 g per liter of extraction solution) to remove any trace levels of water contamination. Reference N/A N/A N/A N/A ECIG-CTR Aerosol Parameters March /89

45 Laboratory 4 Method Summary (continued) Quantitation Information Nicotine Water Glycerin PG Limit of Quantitation (mg/pad) 51.1 µg/ml 516 µg/ml 50.7 µg/ml 49.3 µg/ml Limit of Detection (mg/pad) 16.9 µg/ml 170 µg/ml 16.7 µg/ml 16.3 µg/ml Number of standards Concentration of each Standard 50, 100, 250, 500, 1000, , 1000, 2500, 5000, 10000, , 100, 250, 500, 1000, , 100, 250, 500, 1000, 2000 Internal Standard Quinoline (FID) 1,4-Butanediol (TCD) 1,4-Butanediol (FID) 1,4-Butanediol (FID) Calibration Type Linear, with IS Quadratic, with IS Quadratic, with IS Linear, with IS r 2 >0.990 >0.990 >0.990 >0.990 Additional Details N/A N/A N/A N/A ECIG-CTR Aerosol Parameters March /89

46 Figure 1: Laboratory 4 Example Chromatogram Aerosol Sample Figure 2: Laboratory 4 Example Chromatogram Aerosol Sample ECIG-CTR Aerosol Parameters March /89

47 Figure 3: Laboratory 4 Example Chromatogram Calibration Standard Figure 4: Laboratory 4 Example Chromatogram Calibration Standard ECIG-CTR Aerosol Parameters March /89

48 APPENDIX C: Raw Data Means Raw data are presented as means ± 1 standard deviation for puff sets and for total yields Results for puff sets with average ACM values <6mg are not reported. Lab1-3 piece Puff Set Data Puff Count Lab1-3 piece Puff Set Data Weight Loss (mg) ECIG-CTR Aerosol Parameters March /89

49 Lab1-3 piece Puff Set Data ACM (mg) Lab1-3 piece Puff Set Data Propylene Glycol (mg) ECIG-CTR Aerosol Parameters March /89

50 Lab1-3 piece Puff Set Data Glycerin (mg) Lab1-3 piece Puff Set Data Nicotine (mg) ECIG-CTR Aerosol Parameters March /89

51 Lab1-3 piece Puff Set Data Water (mg) Lab1-Rechargeable Puff Set Data Puff Count ECIG-CTR Aerosol Parameters March /89

52 Lab1-R Puff Set Data Weight Loss (mg) Lab1-Rechargeable Puff Set Data ACM (mg) ECIG-CTR Aerosol Parameters March /89

53 Lab1-Rechargeable Puff Set Data Propylene Glycol (mg) Lab1-Rechargeable Puff Set Data Glycerin (mg) ECIG-CTR Aerosol Parameters March /89

54 Lab1-Rechargeable Puff Set Data Nicotine (mg) Lab1-Rechargeable Puff Set Data Water (mg) ECIG-CTR Aerosol Parameters March /89

55 Lab2-Disposable Puff Set Data Puff Count Lab 2-D Puff Set Data Weight Loss (mg) ECIG-CTR Aerosol Parameters March /89

56 Lab2-Disposable Puff Set Data ACM (mg) Lab2-Disposable Puff Set Data Propylene Glycol (mg) ECIG-CTR Aerosol Parameters March /89

57 Lab2-Disposable Puff Set Data Glycerin (mg) Lab2-Disposable Puff Set Data Nicotine (mg) ECIG-CTR Aerosol Parameters March /89

58 Lab2-Disposable Puff Set Data Water (mg) Lab2-Rechargeable Puff Set Data Puff Count ECIG-CTR Aerosol Parameters March /89

59 Lab 2-R Puff Set Data Weight Loss (mg) Lab2-Rechargeable Puff Set Data ACM (mg) ECIG-CTR Aerosol Parameters March /89

60 Lab2-Rechargeable Puff Set Data Propylene Glycol (mg) Lab2-Rechargeable Puff Set Data Glycerin (mg) ECIG-CTR Aerosol Parameters March /89

61 Lab2-Rechargeable Puff Set Data Nicotine (mg) Lab2-Rechargeable Puff Set Data Water (mg) ECIG-CTR Aerosol Parameters March /89

62 Lab3-Disposable Puff Set Data Puff Count Lab 3-D Puff Set Data Weight Loss (mg) ECIG-CTR Aerosol Parameters March /89

63 Lab3-Disposable Puff Set Data ACM (mg) Lab3-Disposable Puff Set Data Propylene Glycol (mg) ECIG-CTR Aerosol Parameters March /89

64 Lab3-Disposable Puff Set Data Glycerin (mg) Lab3-Disposable Puff Set Data Nicotine (mg) ECIG-CTR Aerosol Parameters March /89

65 Lab3-Disposable Puff Set Data Water (mg) Lab3-Rechargeable Puff Set Data Puff Count ECIG-CTR Aerosol Parameters March /89

66 Lab 3-R Puff Set Data Weight Loss (mg) Lab3-Rechargeable Puff Set Data ACM (mg) ECIG-CTR Aerosol Parameters March /89

67 Lab3-Rechargeable Puff Set Data Propylene Glycol (mg) Lab3-Rechargeable Puff Set Data Glycerin (mg) ECIG-CTR Aerosol Parameters March /89

68 Lab3-Rechargeable Puff Set Data Nicotine (mg) Lab3-Rechargeable Puff Set Data Water (mg) ECIG-CTR Aerosol Parameters March /89

69 Lab4-Disposable Puff Set Data Puff Count Lab 4-D Puff Set Data Weight Loss (mg) ECIG-CTR Aerosol Parameters March /89

70 Lab4-Disposable Puff Set Data ACM (mg) Lab4-Disposable Puff Set Data Propylene Glycol (mg) ECIG-CTR Aerosol Parameters March /89

71 Lab4-Disposable Puff Set Data Glycerin (mg) Lab4-Disposable Puff Set Data Nicotine (mg) ECIG-CTR Aerosol Parameters March /89

72 Lab4-Disposable Puff Set Data Water (mg) Lab4-Rechargeable Puff Set Data Puff Count ECIG-CTR Aerosol Parameters March /89

73 Lab 4-R Puff Set Data Weight Loss (mg) Lab4-Rechargeable Puff Set Data ACM (mg) ECIG-CTR Aerosol Parameters March /89

74 Lab4-Rechargeable Puff Set Data Propylene Glycol (mg) Lab4-Rechargeable Puff Set Data Glycerin (mg) ECIG-CTR Aerosol Parameters March /89

75 Lab4-Rechargeable Puff Set Data Nicotine (mg) Lab4-Rechargeable Puff Set Data Water (mg) ECIG-CTR Aerosol Parameters March /89

76 Lab1-3 piece Total Yields Puff Count Weight Loss (mg) ACM (mg) PG (mg) Glycerin (mg) Nicotine (mg) Water (mg) Test Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev 35/2/60/B /2/30/B /3/30/Sq /4/30/Sq 150 NA NA NA NA 3.61 NA 64.5 NA 70/3/30/Sq /4/30/Sq Lab1-Rechargeable Total Yields Puff Count Weight Loss (mg) ACM (mg) PG (mg) Glycerin (mg) Nicotine (mg) Water (mg) Test Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev 35/2/60/B /2/30/B /3/30/Sq /4/30/Sq /3/30/Sq /4/30/Sq ECIG-CTR Aerosol Parameters March /89

77 Lab2-Disposable Total Yields Puff Count Weight Loss (mg) ACM (mg) PG (mg) Glycerin (mg) Nicotine (mg) Water (mg) Test Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev 35/2/60/B /2/30/B /3/30/Sq /4/30/Sq /3/30/Sq /4/30/Sq Lab2-Rechargeable Total Yields Puff Count Weight Loss (mg) ACM (mg) PG (mg) Glycerin (mg) Nicotine (mg) Water (mg) Test Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev 35/2/60/B /2/30/B /3/30/Sq /4/30/Sq /3/30/Sq /4/30/Sq ECIG-CTR Aerosol Parameters March /89

78 Lab3-Disposable Total Yields Puff Count Weight Loss (mg) ACM (mg) PG (mg) Glycerin (mg) Nicotine (mg) Water (mg) Test Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev 35/2/60/B /2/30/B /3/30/Sq /4/30/Sq /3/30/Sq /4/30/Sq Lab3-Rechargeable Total Yields Puff Count Weight Loss (mg) ACM (mg) PG (mg) Glycerin (mg) Nicotine (mg) Water (mg) Test Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev 35/2/60/B /2/30/B /3/30/Sq /4/30/Sq /3/30/Sq /4/30/Sq ECIG-CTR Aerosol Parameters March /89

79 Lab4-Disposable Total Yields Puff Count Weight Loss (mg) ACM (mg) PG (mg) Glycerin (mg) Nicotine (mg) Water (mg) Test Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev 35/2/60/B /2/30/B /3/30/Sq /4/30/Sq /3/30/Sq /4/30/Sq Lab4-Rechargeable Total Yields Puff Count Weight Loss (mg) ACM (mg) PG (mg) Glycerin (mg) Nicotine (mg) Water (mg) Test Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev Average Stdev 35/2/60/B /2/30/B /3/30/Sq /4/30/Sq /3/30/Sq /4/30/Sq ECIG-CTR Aerosol Parameters March /89

80 APPENDIX D: Total Yield Graphs Total ACM (mg) Test 3 Test 4 Test 5 Test 6 Figure 1: Lab1-3 piece Total Yields Total ACM (mg) Test 3 Test 4 Test 5 Test 6 Figure 2: Lab1-Rechargeable Total Yields Regime Test 3 Test 4 Test 5 Test 6 Code 55/3/30/Sq 55/4/30/Sq 70/3/30/Sq 70/4/30/Sq ECIG-CTR Aerosol Parameters March /89

81 Total ACM (mg) Test 3 Test 4 Test 5 Test 6 Figure 3: Lab2-Disposable Total Yields Total ACM (mg) Test 3 Test 4 Test 5 Test 6 Figure 4: Lab2-Rechargeable Total Yields Regime Test 3 Test 4 Test 5 Test 6 Code 55/3/30/Sq 55/4/30/Sq 70/3/30/Sq 70/4/30/Sq ECIG-CTR Aerosol Parameters March /89

82 Total ACM (mg) Test 3 Test 4 Test 5 Test 6 Figure 5: Lab3-Disposable Total Yields Total ACM (mg) Test 3 Test 4 Test 5 Test 6 Figure 6: Lab3-Rechargeable Total Yields Regime Test 3 Test 4 Test 5 Test 6 Code 55/3/30/Sq 55/4/30/Sq 70/3/30/Sq 70/4/30/Sq ECIG-CTR Aerosol Parameters March /89

83 Total ACM (mg) Test 3 Test 4 Test 5 Test 6 Figure 7: Lab4-Disposable Total Yields Total ACM (mg) Test 3 Test 4 Test 5 Test 6 Figure 8: Lab4-Disposable Total Yields Regime Test 3 Test 4 Test 5 Test 6 Code 55/3/30/Sq 55/4/30/Sq 70/3/30/Sq 70/4/30/Sq ECIG-CTR Aerosol Parameters March /89

84 APPENDIX E: Puff Set Graphs ACM (mg/pad) Test 3 Test 4 Test 5 Test 6 Puff Number Figure 1: Lab1-3 piece Total Yields ACM (mg/pad) Test 3 Test 4 Test 5 Test 6 Figure 2: Lab1-Rechargeable Total Yields Puff Number Regime Test 3 Test 4 Test 5 Test 6 Code 55/3/30/Sq 55/4/30/Sq 70/3/30/Sq 70/4/30/Sq ECIG-CTR Aerosol Parameters March /89

85 ACM (mg/pad) Test 3 Test 4 Test 5 Test 6 Puff Number Figure 3: Lab2-Disposable Total Yields ACM (mg/pad) Test 3 Test 4 Test 5 Test 6 Figure 4: Lab2-Rechargeable Total Yields Puff Number Regime Test 3 Test 4 Test 5 Test 6 Code 55/3/30/Sq 55/4/30/Sq 70/3/30/Sq 70/4/30/Sq ECIG-CTR Aerosol Parameters March /89

86 ACM (mg/pad) Test 3 Test 4 Test 5 Test 6 Puff Number Figure 5: Lab3-Disposable Total Yields ACM (mg/pad) Test 3 Test 4 Test 5 Test 6 Figure 6: Lab3-Rechargeable Total Yields Puff Number Regime Test 3 Test 4 Test 5 Test 6 Code 55/3/30/Sq 55/4/30/Sq 70/3/30/Sq 70/4/30/Sq ECIG-CTR Aerosol Parameters March /89

87 ACM (mg/pad) Test 3 Test 4 Test 5 Test 6 Puff Number Figure 7: Lab4-Disposable Total Yields ACM (mg/pad) Test 3 Test 4 Test 5 Test 6 Puff Number Figure 8: Lab5-R Total Yields Regime Test 3 Test 4 Test 5 Test 6 Code 55/3/30/Sq 55/4/30/Sq 70/3/30/Sq 70/4/30/Sq ECIG-CTR Aerosol Parameters March /89

88 APPENDIX F: ACM Variability - Percent Relative Standard Deviation for Puff Sets Rechargeable Products Results n=5 ECIG-CTR Aerosol Parameters March /89

Gas Chromatography Assignment Chem 543/443

Gas Chromatography Assignment Chem 543/443 Gas Chromatography Assignment Chem 543/443 1. Introduction Capillary gas chromatography (GC) is one of the most popular analytical techniques used in today s research. Its popularity is mainly due to efficient

More information

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools Application Note Abstract The Commonwealth of Massachusetts Department of Environmental Protection

More information

Capillary GC Troubleshooting: a Practical Approach

Capillary GC Troubleshooting: a Practical Approach Capillary GC Troubleshooting: a Practical Approach Katherine Stenerson Gas Separations R&D Supelco, Supelco Park, Bellefonte, PA 16823 1999 Sigma-Aldrich Co. Outline Basic Troubleshooting Strategy Preventing

More information

Using the Pulsed Flame Photometric Detector (PFPD) for Low-level Analysis of Organophosphorus Pesticides

Using the Pulsed Flame Photometric Detector (PFPD) for Low-level Analysis of Organophosphorus Pesticides Application Note 25390206 Using the Pulsed Flame Photometric Detector (PFPD) for Low-level Analysis of Organophosphorus Pesticides Keywords Model 5380 Organophosphorus Pesticide PFPD PFPDView Phosphorus

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis S(-)-Nicotine, Primary Standard N-008 Page 1 of 7 Catalog Number: N-008 H Lot: Expiration: October 2019 N Description: S(-)-Nicotine in Methanol.

More information

MS Certified Vials. Pre-cleaned and certified vials for mass spectrometry

MS Certified Vials. Pre-cleaned and certified vials for mass spectrometry Pre-cleaned and certified vials for mass spectrometry The FIRST and ONLY pre-cleaned, low particle, low background chromatography vial When your instrumentation, sample handling and methodology is pushing

More information

Low-level Detection of Ethanol, 1,4-Dioxane, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator

Low-level Detection of Ethanol, 1,4-Dioxane, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator Application Note 25370206 Low-level Detection of,, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator Keywords Eclipse Fuel Oxygenates Methanol Model 4552 Purge and Trap Sample Concentrator

More information

System Overview. Valve oven. Restrictor. Valve. Sample in and out

System Overview. Valve oven. Restrictor. Valve. Sample in and out System Overview The GC system is designed to provide a continous reading of total hydrocarbons in a gas stream, while periodically performing a chromatographic separation of the sample to determine the

More information

Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis

Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis Atis Chakrabarti* and Roy Eksteen + Tosoh Bioscience LLC, King of Prussia, PA 19406 *Corresponding Author. + Current address:

More information

SCION GC Series. The Gas Chromatographers Choice for Separations

SCION GC Series. The Gas Chromatographers Choice for Separations SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Bruker s Scion Instrument s long tradition long of tradition innovation of innovation and product and reliability

More information

Fast Detailed Hydrocarbon Analysis by Modified ASTM Method D6730

Fast Detailed Hydrocarbon Analysis by Modified ASTM Method D6730 application Note Gas Chromatography Author Terry Osenbach PerkinElmer, Inc. Shelton, CT 06484 USA Fast Detailed Hydrocarbon Analysis by Modified ASTM Method D6730 Introduction Detailed hydrocarbon analysis

More information

Agilent 7693A Series Automatic Liquid Sampler. Inject new performance into your gas chromotography

Agilent 7693A Series Automatic Liquid Sampler. Inject new performance into your gas chromotography Agilent 7693A Series Automatic Liquid Sampler Inject new performance into your gas chromotography Agilent 7693A Series Automatic Liquid Sampler A powerful combination of productivity and flexibility How

More information

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Student Learning Objectives High Performance Liquid Chromatography With the advent of relatively inexpensive and reliable pumps, the development of

More information

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph Analysis of USP Residual Solvents using the Agilent 797A Headspace Sampler with the Agilent 79B Gas Chromatograph Application Note Pharmaceuticals Authors Bart Tienpont, Frank David, and Pat Sandra

More information

Quantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy

Quantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy Advances in Bioscience and Biotechnology, 2010, 1, 367-371 doi:10.4236/abb.2010.15049 Published Online December 2010 (http://www.scirp.org/journal/abb/). Quantitative determination of residual 2-(2-chloroethoxy)

More information

C146-E096 TD-20. Shimadzu Thermal Desorption System

C146-E096 TD-20. Shimadzu Thermal Desorption System C146-E096 TD-20 Shimadzu Thermal Desorption System Thermal Desorption Thermal desorption systems trap target substances carried in mobile phase by adsorption in a sample tube packed with adsorbent. Then,

More information

Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY

Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY AGILENT 7693A SERIES AUTOMATIC LIQUID SAMPLER A powerful combination of productivity and flexibility Autosampler

More information

Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY

Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY AGILENT 7693A SERIES AUTOMATIC LIQUID SAMPLER A powerful combination of productivity and flexibility It

More information

Gas Chromatography (GC) Operation, Calibration, Troubleshooting and Practice Maintenance

Gas Chromatography (GC) Operation, Calibration, Troubleshooting and Practice Maintenance Training Title Gas Chromatography (GC) Operation, Calibration, Troubleshooting and Practice Maintenance Training Duration 5 days Training Venue and Dates Gas Chromatography (GC) Operation, Calibration,

More information

Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016

Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016 Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016 1.0 Method background: This method employs full evaporation technique headspace

More information

Determination of Monocrotophos, Diazinon, Malathion, EPN, and Methamidaphos from Aqueous Samples Using Atlantic HLB SPE Disks

Determination of Monocrotophos, Diazinon, Malathion, EPN, and Methamidaphos from Aqueous Samples Using Atlantic HLB SPE Disks Determination of Monocrotophos, Diazinon, Malathion, EPN, and Methamidaphos from Aqueous Samples Using Jim Fenster, Julie McGettrick, Horizon Technology, Inc., Salem, NH Introduction Monocrotophos, Diazinon,

More information

Agilent Ultimate Plus Fused Silica Tubing

Agilent Ultimate Plus Fused Silica Tubing Agilent Ultimate Plus Fused Silica Tubing Technical Overview Introduction Deactivated fused silica tubing is widely used for guard columns, transfer lines, or long retention gaps in GC and GC/MS applications

More information

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website: Published in November 2013

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website:  Published in November 2013 5-100 Innovation Way Newark, DE 19711, USA Tel:302-3661101 Fax:302-3661151 Website: www.sepax-tech.com Published in November 2013 P/N SECKIT-7830 These Phases are developed based on innovative surface

More information

Preparative HPLC: Factors and Parameters that Directly Affect Recovery of Collected Fractions

Preparative HPLC: Factors and Parameters that Directly Affect Recovery of Collected Fractions Preparative HPLC: Factors and Parameters that Directly Affect Recovery of Collected Fractions Gary Scharrer and Joan Stevens, Ph.D. Gilson, Inc Middleton, WI www.gilson.com Abstract Purification of synthetic,

More information

METHOD 25D - DETERMINATION OF THE VOLATILE ORGANIC CONCENTRATION OF WASTE SAMPLES. NOTE: Performance of this method should not be

METHOD 25D - DETERMINATION OF THE VOLATILE ORGANIC CONCENTRATION OF WASTE SAMPLES. NOTE: Performance of this method should not be 1285 METHOD 25D - DETERMINATION OF THE VOLATILE ORGANIC CONCENTRATION OF WASTE SAMPLES NOTE: Performance of this method should not be attempted by persons unfamiliar with the operation of a flame ionization

More information

HYDROGEN AS A CARRIER GAS. Gas Chromatography new frontiers: can Hydrogen substitute Helium?

HYDROGEN AS A CARRIER GAS. Gas Chromatography new frontiers: can Hydrogen substitute Helium? HYDROGEN AS A CARRIER GAS Gas Chromatography new frontiers: can Hydrogen substitute Helium? Manufacturing UHP gas generators for lab analysis since 1979, CLAIND is seating in Lenno, on Como Lake, where

More information

Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES

Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES THE AGILENT 7820A GC SYSTEM FORWARD-THINKING INNOVATION, EVERYDAY RELIABILITY AND VALUE Is your lab looking for better, more reproducible results

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 R. J. REYNOLDS TOBACCO COMPANY 1 Product Services Analytical Testing Laboratories 950 Reynolds Blvd. Winston Salem, NC 27105 Jannell Rowe Phone: 336 741 4121

More information

PerkinElmer Supplies/Parts. GC Accessories. Autosampler Syringes

PerkinElmer Supplies/Parts. GC Accessories. Autosampler Syringes Autosampler Syringes Volume, Code, Gauge, Length, Point Style QTY OEM No. PerkinElmer No. Part No. Hamilton Syringes 5μL, ASN, 23s, 70mm, #3 1 88035 N6101390 24-88035 5μL, ASN, 26s, 70mm, #3 1 88040 N6101380

More information

Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run

Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run Application Note Author Jessica Westland Agilent Technologies, Inc. Abstract A selected target compound list of 195 various pesticides

More information

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications TECHNICAL BRIEF Gas Chromatography/ Mass Spectrometry The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications C 14 scans Introduction As gas chromatography

More information

Agilent MetaCarb Ca PLUS Carbohydrate Column Ca ++ Form User Manual Part No. A5205

Agilent MetaCarb Ca PLUS Carbohydrate Column Ca ++ Form User Manual Part No. A5205 Agilent MetaCarb Ca PLUS Carbohydrate Column Ca ++ Form User Manual Part No. A5205 WARNING: The Agilent MetaCarb Ca PLUS calcium column is packed with a polymeric material that requires special care. Introduction

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 5 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column Application Note Bio-Pharmaceutical Authors M. Sundaram Palaniswamy and Andrew Coffey Agilent Technologies,

More information

DOWNSTREAM DEFENDER TESTING USING MAINE DEPARTMENT OF ENVIRONMENTAL PROTECTION PROTOCOL UTILIZING OK-110 FEED SAND

DOWNSTREAM DEFENDER TESTING USING MAINE DEPARTMENT OF ENVIRONMENTAL PROTECTION PROTOCOL UTILIZING OK-110 FEED SAND DOWNSTREAM DEFENDER TESTING USING MAINE DEPARTMENT OF ENVIRONMENTAL PROTECTION PROTOCOL UTILIZING OK-110 FEED SAND JULY 2002 Copyright 2001 H.I.L. Technology, Inc. All rights reserved. This report was

More information

Preparative, Pilot & Process Columns

Preparative, Pilot & Process Columns Preparative, Pilot & Process Columns 321 Preparative, Pilot & Process Columns Contents Prep Columns... 322-323 Dynamic Axial Compression Columns (DAC)... 324-325 Pilot Columns... 326-327 Chromatography

More information

Why filter your samples before analysis? Benefits of Using Chromfilter Syringe Filters Daily: Designed for your application:

Why filter your samples before analysis? Benefits of Using Chromfilter Syringe Filters Daily: Designed for your application: Why filter your samples before analysis? Prefiltering your sample and mobile phase solutions for particulates and microbial growth prior to analysis is critical to preventing column and frit blockage,

More information

Provläsningsexemplar / Preview. First edition

Provläsningsexemplar / Preview. First edition Provläsningsexemplar / Preview TECHNICAL SPECIFICATION ISO/TS 16189 First edition 2013-07-15 Footwear Critical substances potentially present in footwear and footwear components Test method to quantitatively

More information

Validation Guide USTR 2114 (2) Validation Guide for Pall Emflon PFR Filter Cartridges

Validation Guide USTR 2114 (2) Validation Guide for Pall Emflon PFR Filter Cartridges Validation Guide USTR 2114 (2) Validation Guide for Pall Emflon PFR Filter Cartridges CONTENTS Part I. Overview 4 1. Introduction 4 2. Summary of Conclusions 4 Part II. Studies on Removal Efficiency 6

More information

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover columns ProSwift Reversed-Phase Monolith Columns for Protein Analysis ProSwift reversed-phase columns use a unique monolith technology for fast, high-resolution HPLC and LC/MS separations of proteins.

More information

ISO/TC 238. Secretariat: SIS Solid biofuels Determination of particle size distribution for uncompressed fuels

ISO/TC 238. Secretariat: SIS Solid biofuels Determination of particle size distribution for uncompressed fuels DRAFT INTERNATIONAL STANDARD ISO/DIS 17827-2 Solid biofuels Determination of particle size distribution for uncompressed fuels Part 2: Vibrating screen method using sieves with aperture of 3,15 mm and

More information

Formulas: See Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1. METHOD: 2555, Issue 1 EVALUATION: PARTIAL Issue 1: 15 March 2003

Formulas: See Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1. METHOD: 2555, Issue 1 EVALUATION: PARTIAL Issue 1: 15 March 2003 KETONES I 2555 Formulas: See Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1 METHOD: 2555, Issue 1 EVALUATION: PARTIAL Issue 1: 15 March 2003 OSHA : See Table 2 NIOSH: See Table 2 ACGIH: See Table 2 PROPERTIES:

More information

Fast Preparative Column Liquid Chromatography (PCLC)

Fast Preparative Column Liquid Chromatography (PCLC) Fast Preparative Column Liquid Chromatography (PCLC) Application Note 224 Joan Stevens, Ph.D. and Gary Scharrer Introduction Preparative HPLC is recognized as a prime method for obtaining pure compounds

More information

Analysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS

Analysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS Analysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS Authors: Conor Smith, Doug Fryer United Science Corp. 15911 Furuby Rd. Center City, MN 55012 USA Abstract

More information

Microencapsulation Studies for Mass Production of IFE Targets

Microencapsulation Studies for Mass Production of IFE Targets Microencapsulation Studies for Mass Production of IFE Targets B. Vermillion, G. Besenbruch, L. Brown, D. Goodin, B. McQuillan, M. Takagi, T.Woo Second US/Japan Workshop on Target Fabrication, Injection,

More information

Econo-Pac High Capacity Ion Exchange Cartridges, 5 ml Instruction Manual. Catalog Numbers , ,

Econo-Pac High Capacity Ion Exchange Cartridges, 5 ml Instruction Manual. Catalog Numbers , , Econo-Pac High Capacity Ion Exchange Cartridges, 5 ml Instruction Manual Catalog Numbers 732-0026, 732-0027 732-0066, 732-0067 Table of Contents Section 1 Introduction... 1 Section 2 Connection to Bio-Rad

More information

Volatile Organic Compounds in Solids PBM

Volatile Organic Compounds in Solids PBM Organics Revision Date: Aug 15, 2014 Volatile Organic Compounds in Solids PBM Parameter Analytical Method Introduction Method Summary Parameter Applicability, MDLs Volatile Organic Compounds (VOCs) in

More information

Method Translation in Liquid Chromatography

Method Translation in Liquid Chromatography Method Translation in Liquid Chromatography Technical Overview Abstract Ronald E. Majors Agilent Technologies, Inc. 2850 Centerville Rd Wilmington, DE 19808 USA With the recent emphasis on high performance

More information

Gas Chromatograph. GC-8A Series C184-E016B

Gas Chromatograph. GC-8A Series C184-E016B Gas Chromatograph GC-8A Series C184-E016B GC-8A Series Dedicated Single Detector Gas Chromatograph GC that meets increasing analytical requirements GC that can analyze instantaneously GC that is tough

More information

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC Quality-by-Design-Based Method Development Using an Agilent 129 Infinity II LC An Efficient Method Development Workflow Combined with ISET-mediated Method Transfer Under Waters Empower 3 CDS Control Application

More information

Genotoxicity is the property of a compound

Genotoxicity is the property of a compound Impurities Analysis in Pharmaceuticals: Genotoxicity is the property of a compound known to have irreversible effects on the structure and functionality of the DNA in cells and cause DNA loss, DNA replication

More information

Preparative HPLC is still the

Preparative HPLC is still the B I O P R O C E S S TECHNICAL Optimizing Sample Load Capacity and Separation Through a Series of Short Prep Columns Mark Crawford, Joan Stevens, and Luke Roenneburg Preparative HPLC is still the dominate

More information

Fundamentals and Techniques of Preparative HPLC. Parto Zist Behboud Tel: 42108

Fundamentals and Techniques of Preparative HPLC. Parto Zist Behboud  Tel: 42108 Fundamentals and Techniques of Preparative HPLC Parto Zist Behboud www.partoz.com Tel: 42108 Today s Overview 1 Basic know-how for up-scaling 2 Practical examples of up-scaling 3 Recycle preparation 4

More information

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus Robert Birdsall, Thomas Wheat, and Weibin Chen Waters Corporation,

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis ISO GUIDE 34 ANAB Cert# AR-1470 ISO/IEC 17025 ANAB Cert# AT-1467 SORBITOL CERTIFIED REFERENCE MATERIAL CERTIFIED PURITY: 98.4 %, U crm = ±0.1 % k = 2 (Mass Balance/as is basis)

More information

Available Laboratory Methods for Soil Gas Analysis

Available Laboratory Methods for Soil Gas Analysis Available Laboratory Methods for Soil Gas Analysis US EPA Soil Gas Sample Collection & Analysis Workshop San Diego, CA March 21-22, 2007 Gina M. Plantz NewFields Environmental Forensics Practice, LLC gplantz@newfields.com

More information

Agilent InfinityLab Preparative Flow Cells for MWD & DAD

Agilent InfinityLab Preparative Flow Cells for MWD & DAD Agilent InfinityLab Preparative Flow Cells for MWD & DAD Technical Note This technical note describes the preparative flow cells for the Agilent 1260 Infinity II DAD WR (G7115A) and Agilent 1260 Infinity

More information

Protein-Pak Hi Res HIC Column and HIC Protein Standard

Protein-Pak Hi Res HIC Column and HIC Protein Standard Protein-Pak Hi Res HIC Column and HIC Protein Standard CONTENTS I. INTRODUCTION II. a. Mobile Phase b. Flow Direction CONNECTING COLUMN TO LC SYSTEM I. INTRODUCTION This offering contains non-porous, polymethacrylate-based

More information

TVOC Heated FID VE 7

TVOC Heated FID VE 7 Rack Mount/Table Top THC Analyzer TVOC Heated FID VE 7 The VE7 19"/5PU rack mount and table top heated emission analyzer for the continuous determination of the mass concentration of total gaseous organic

More information

A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts

A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts Andrew Aubin and Ronan Cleary Waters Corporation, Milford, MA, USA APPLICATIN BENEFITS A modular preparative HPLC

More information

Extraction of POPs from Environmental Samples using Accelerated Solvent Extraction (ASE )

Extraction of POPs from Environmental Samples using Accelerated Solvent Extraction (ASE ) Extraction of POPs from Environmental Samples using Accelerated Solvent Extraction (ASE ) Frank Höfler* and Khalil Divan** *Dionex (Europe) Management AG Solothurner Strasse 259 4600 Olten Switzerland

More information

Can your unit pass a Particulate Emission Compliance Test?

Can your unit pass a Particulate Emission Compliance Test? Source Emissions Testing and Emissions Specialists Can your unit pass a Particulate Emission Compliance Test? Kevin Crosby The Avogadro Group, LLC California - Oregon - Arizona McIlvaine Hot Topic Hour,

More information

Impact of GC Parameters on The Separation Part 4: Choice of Film Thickness

Impact of GC Parameters on The Separation Part 4: Choice of Film Thickness Cd The Chrom Doctor Impact of GC Parameters on The Separation Part 4: Choice of Film Thickness Jaap de Zeeuw, Restek Corporation, Middelburg, The Netherlands. In part 1, 2 and 3 of this series we focused

More information

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany New Features of the Agilent Method Scouting Wizard for Automated Method Development of Complex Samples Analysis of Large Data Sets by Method Scouting Reports and Automated Adjustment of Flow Rates and

More information

The Analysis of 4-Vinyl-Cyclo-Hexene (VCH) in Butadiene with the DVLS LGI Injector

The Analysis of 4-Vinyl-Cyclo-Hexene (VCH) in Butadiene with the DVLS LGI Injector The Analysis of 4-Vinyl-Cyclo-Hexene (VCH) in Butadiene with the DVLS LGI Injector Introduction In synthetic rubber manufacturing, 4-Vinylcyclohexene (VCH) can be found as a byproduct from the chemical

More information

INTERNATIONAL PHARMACOPOEIA MONOGRAPH ON ARTEMETHER AND LUMEFANTRINE CAPSULES REVISED DRAFT FOR DISCUSSION

INTERNATIONAL PHARMACOPOEIA MONOGRAPH ON ARTEMETHER AND LUMEFANTRINE CAPSULES REVISED DRAFT FOR DISCUSSION October 2007 RESTRICTED ` INTERNATIONAL PHARMACOPOEIA MONOGRAPH ON ARTEMETHER AND LUMEFANTRINE CAPSULES REVISED DRAFT FOR DISCUSSION World Health Organization 2007 All rights reserved. This draft is intended

More information

Aqua Clear Aqua Clear multi-lobal cross section of individual filaments has more micro-voids giving higher dirt holding capacity and lower resistance

Aqua Clear Aqua Clear multi-lobal cross section of individual filaments has more micro-voids giving higher dirt holding capacity and lower resistance Aqua Clear STRING-WOUND FILTER CARTRIDGES Free of extractables No leaching Twice the life of conventional cartridges Made from continuous filaments No media migration % Polypropylene Wide chemical compatibility

More information

Fusion UV/Persulfate TOC Analyzer

Fusion UV/Persulfate TOC Analyzer Fusion UV/Persulfate TOC Analyzer The Fusion Total Organic Carbon (TOC) Analyzer utilizes powerful Ultra Violet (UV) Persulfate oxidation allowing superior carbon liberation from even the most challenging

More information

Cadmium Testing and On Site Calibration for Water Testing Detection Range: ppm

Cadmium Testing and On Site Calibration for Water Testing Detection Range: ppm Document: AND016 04 2014 Cadmium Testing and On Site Calibration for Water Testing Detection Range: 0.1 1.0ppm April, 2014 Edition 3 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of

More information

Southwest Clean Air Agency NE 99 th St, Suite 1294, Vancouver, WA 98682

Southwest Clean Air Agency NE 99 th St, Suite 1294, Vancouver, WA 98682 Source Test Guidance Test plans are required to be submitted to SWCAA prior to testing. Generally the plans must be submitted 2 weeks prior to testing but may be different based on individual facility

More information

MACRO/MICRO/NANO POROMETER MMN-1500A. Not just products... Solutions!

MACRO/MICRO/NANO POROMETER MMN-1500A. Not just products... Solutions! MACRO/MICRO/NANO POROMETER MMN-1500A Not just products... Solutions! Macro/Micro/Nano Porometer MMN-1500A The new Macro-Micro-Nano Porometer combines the benefits of our Capillary Flow Porometer and Liquid-Liquid

More information

Joint Committee Issue Paper

Joint Committee Issue Paper NSF Standard(s) Impacted: Background: Provide a brief background statement indicating the cause and nature of concern, the impacts identified relevant to public health, public understanding, etc, and any

More information

Proficiency Testing Program User Guide

Proficiency Testing Program User Guide Version 1.1 October 1, 2017 Powered by Service Area: LABORATORY QCS-LAB-PTUSERGUIDE Origination Date: 2017-08-01 Last Revision: 2017-10-01 Effective Date: 2017-10-01 Version Number: 1.1 Next Review: 2018-01-02

More information

APE SOFTWARE TOOLS FOR MANAGEMENT SYSTEMS. Calibration Control. Calibration Management Software

APE SOFTWARE TOOLS FOR MANAGEMENT SYSTEMS. Calibration Control. Calibration Management Software APE SOFTWARE TOOLS FOR MANAGEMENT SYSTEMS Calibration Control Calibration Management Software MAJOR FEATURES CREATED FROM THE MANUFACTURING USER S PERSPECTIVE WINDOWS 7 & 8 21 CFR PART 11 COMPLIANT MS

More information

Fast Analysis of Extended Natural Gas using the New Transportable INFICON Micro GC Fusion

Fast Analysis of Extended Natural Gas using the New Transportable INFICON Micro GC Fusion Fast Analysis of Extended Natural Gas using the New Transportable INFICON Micro GC Fusion Debbie Hutt PITTCON 2013 Philadelphia, Pa Wednesday March 20th, 2012 Cologne, Germany Shanghai, China 325 employees

More information

Inter-laboratory tests of the methodology for filtration efficiency tests in different filter media against nanoparticles

Inter-laboratory tests of the methodology for filtration efficiency tests in different filter media against nanoparticles Inter-laboratory tests of the methodology for filtration efficiency tests in different filter media against nanoparticles Panagiota Sachinidou, Shawn S.C. Chen, David Y.H. Pui, Paolo Tronville, Thomas

More information

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider Combining small-scale purification and analysis of monoclonal antibodies on one instrument Protein purification with high-volume injection using the Agilent 126 Infinity Bio-inert Quaternary LC System

More information

Automatic Determination of Greenhouse Gases by GC

Automatic Determination of Greenhouse Gases by GC Automatic Determination of Greenhouse Gases by GC Massimo Santoro, Thermo Fisher Scientific Cristiane de Oliveira Silva, Henrique Franciscato Melo Danilo Vinicius Pierone, NovaAnalitica, Brazil 1 The world

More information

It works as hard as you do.

It works as hard as you do. It works as hard as you do. Agilent 6890N Network GC Rugged and reliable for unsurpassed uptime Advanced EPC ensures precision and accuracy of results Retention Time Locking for consistent results Accomplish

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis ISO GUIDE 34 AAB Cert# AR-1470 ISO/IEC 17025 AAB Cert# AT-1467 FLUCOAZOLE CERTIFIED REFERECE MATERIAL F OH F CERTIFIED PURITY: 99.6%, U crm = ±0.2% k = 2 (Mass Balance/as is basis)

More information

Automated Mercury/Non-mercury Intrusion Porosimeter AMP-60K-A-1

Automated Mercury/Non-mercury Intrusion Porosimeter AMP-60K-A-1 Automated Mercury/Non-mercury Intrusion Porosimeter AMP-60K-A-1 Not just products...solutions! Description The PMI Automated Mercury Intrusion Concrete Porosimeter is a versatile and accurate instrument

More information

MonTech RP 3000 Rapid Plastimeter for Polymers and Natural Rubber

MonTech RP 3000 Rapid Plastimeter for Polymers and Natural Rubber Plasticity 02 MonTech RP 3000 Rapid Plastimeter for Polymers and Natural Rubber The MonTech Rapid Plastimeter measures the plasticity or viscosity of polymers and unvulcanised rubbers. The test is simple,

More information

CHROMATOGRAPHY/MASS SPECTROMETRY

CHROMATOGRAPHY/MASS SPECTROMETRY DETERMINATION OF 1,4-DIOXANE IN DRINKING WATER BY SOLID PHASE EXTRACTION (SPE) AND GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS) WITH SELECTED ION MONITORING (SIM)* EPA Method 522 Part Number: EU52112M6

More information

Generating Automated and Efficient LC/MS Peptide Mapping Results with the Biopharmaceutical Platform Solution with UNIFI

Generating Automated and Efficient LC/MS Peptide Mapping Results with the Biopharmaceutical Platform Solution with UNIFI with the Biopharmaceutical Platform Solution with UNIFI Vera B. Ivleva, Ying Qing Yu, Scott Berger, and Weibin Chen Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S The ability to

More information

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column Disulfide Linkage Analysis of IgG1 using an Agilent 126 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column Application Note Biotherapeutics & Biosimilars Author M. Sundaram

More information

Dräger Gas Filtration SUPER CLEAN GAS FILTERS. Dräger: Technology for Life

Dräger Gas Filtration SUPER CLEAN GAS FILTERS. Dräger: Technology for Life Dräger Gas Filtration SUPER CLEAN GAS FILTERS Dräger: Technology for Life 2 DRäGER GAS FILTRATION About Super Clean Gas Filters Cartridge systems make changing gas filters quick and easy. A base plate

More information

Super Clean Gas Filters Simply the best

Super Clean Gas Filters Simply the best Super Clean Gas Filters TM Simply the best Cartridge Type rev. 1.4 Features EP-B-0606960 U.S. Patent 5.478.378 Ultra-high capacity High-purity output ensures 99.9999% Pure gas Quick connect fittings for

More information

Soil - Plasticity 2017 (72) PROFICIENCY TESTING PROGRAM REPORT

Soil - Plasticity 2017 (72) PROFICIENCY TESTING PROGRAM REPORT www.labsmartservices.com.au Soil - Plasticity 2017 (72) PROFICIENCY TESTING PROGRAM REPORT Accredited for compliance with ISO/IEC 17043 Copyright: LabSmart Services Pty Ltd Copyright: LabSmart Services

More information

The Determination of Trace Elements in Stainless Steel by Forked Platform GFAAS

The Determination of Trace Elements in Stainless Steel by Forked Platform GFAAS The Determination of Trace Elements in Stainless Steel by Forked Platform GFAAS Application Note Atomic Absorption Author Lucinda M. Beach Introduction The metallurgical behavior of stainless steel is

More information

) and it s ideal van t Hoff factor is 4. Note that polyatomic ions do not break up into their constituent elements.

) and it s ideal van t Hoff factor is 4. Note that polyatomic ions do not break up into their constituent elements. Freezing Point Depression: Determining CaCl2 Van t Hoff Factor Minneapolis Community and Technical College C1152 v.12.15 I. Introduction The physical properties of solutions that depend on the number of

More information

GUS Assays. 1. Label all tubes. Prepare solutions and have ready at hand.

GUS Assays. 1. Label all tubes. Prepare solutions and have ready at hand. GUS Assays I. Protein isolation A. Method for ~1g or more of tissue. 1. Label all tubes. Prepare solutions and have ready at hand. 2. Remove the tissue from the 80 o C freezer and thaw on ice. If the tissue

More information

Abcite 585 EF. Technical Data and Application Guide. Product description. Typical applications. Product range. Product certifications.

Abcite 585 EF. Technical Data and Application Guide. Product description. Typical applications. Product range. Product certifications. Product description is a single layer, primer-free, halogen-free, high strength adhesive thermoplastic powder coating which provides shattering protection for glass pieces and preserves the original glass

More information

Guide to using the Bio Rad CFX96 Real Time PCR Machine

Guide to using the Bio Rad CFX96 Real Time PCR Machine Guide to using the Bio Rad CFX96 Real Time PCR Machine Kyle Dobbs and Peter Hansen Table of Contents Overview..3 Setup Reaction Guidelines 4 Starting up the Software 5 Setup Protocol on Software 6 Setup

More information

Guidelines on Quality Assurance/Quality Control for The Environmental Measurements of Dioxins

Guidelines on Quality Assurance/Quality Control for The Environmental Measurements of Dioxins PAGE 1 Guidelines on Quality Assurance/Quality Control for The Environmental Measurements of Dioxins November 14, 2000 The Environment Agency These Guidelines set forth measures to be voluntarily taken

More information

ACQUITY UPLC Protein BEH SEC Columns and Standards

ACQUITY UPLC Protein BEH SEC Columns and Standards and Standards CONTENTS I. INTRODUCTION II. CONFIGURING AN ACQUITY UPLC SYSTEM FOR USE IN SEC PROTEIN SEPARATIONS a. Calibrators III. GETTING STARTED a. ecord installation b. Column connectors c. Column

More information

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime EASY-Spray Columns Guidance for column set up and installation Tips to maximize column lifetime EASY-Spray Column Tips and Tricks This document provides guidance for Thermo Scientific EASY-Spray column

More information

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube Application Note Environmental Authors Don Noot, Ralph Hindle Vogon Laboratory Services Cochrane, AB

More information

OVERVIEW OF PROCEDURE 2B

OVERVIEW OF PROCEDURE 2B Packaged-Products Over 150 lb (68 kg) 2B 2 008 ISTA 2 Series Partial Simulation Performance Test Procedure VERSION DATE Last TECHNICAL Change: JANUARY 2008 Last EDITORIAL Change: JANUARY 2008 For complete

More information

Saltwater Analysis Using A High Temperature Total Organic Carbon (TOC) Analyzer with Static Pressure Concentration

Saltwater Analysis Using A High Temperature Total Organic Carbon (TOC) Analyzer with Static Pressure Concentration Saltwater Analysis Using A High Temperature Total Organic Carbon (TOC) Analyzer with Static Pressure Concentration Application Note Abstract Analysis of organic carbon in sea water can be challenging and

More information

Motivational and Attitudinal Patterns of Effective Leaders

Motivational and Attitudinal Patterns of Effective Leaders of Carl L. Harshman, Ph.D. This article identifies seven strong (high scores) and seven weak (low scores) motivational and attitudinal patterns (MAPs) and how they impact leadership effectiveness. In considering

More information