Validation of cleaning and sterilization of reusable medical devices

Size: px
Start display at page:

Download "Validation of cleaning and sterilization of reusable medical devices"

Transcription

1 Synergy makes sence Validation of cleaning and sterilization of reusable medical devices MG-FSI Last revision: June 2012 Aurélien BIGNON 5, Chemin du Catupolan Vaulx en Velin - France - Tel. 33 (0) Fax : 33 (0)

2 MEDICAL LAB Established in 1989, MedicalGroup presently has five subsidiary companies specialising in the biomedical field: MedicalCoating, MedicalPackaging, MedicalBiomat, MedicalManufacturing and MedicalLab. All of these organisations are EN ISO certified. MedicalLab is a lab specialized in tests on medical devices: microbiological tests, composition and structure analysis of implant materials and coatings, static and dynamic mechanical tests on implants and coatings, cleaning, packaging and sterilization validation VALIDATED CLEANING AND STERILIZATION INSTRUCTIONS FOR REUASABLE MEDICAL DEVICES Reusable medical devices come in various shape and size. The design of these instruments, which are more and more complex, aims at helping the surgeons in their task, but cleaning and sterilization of these device should also be taken into account. Manufacturers of reusable medical devices shall provide validated cleaning and sterilization instructions with their product to ensure that their product can be properly cleaned and sterilized in health care facilities. Cleaning a device is the critical first step in reprocessing any device after it has been used on a patient. Failure to remove foreign material (e.g., soil, lubricants, microorganisms, organic and inorganic materials) from both the outside and the inside of the device can interfere with the effectiveness of subsequent disinfection and/or sterilization. Contamination of a variety of reusable devices and instruments (e.g., surgical instruments, ophthalmic instruments, laryngeal mask airways, endoscopes) are causes of concern for patient safety from cross-contamination between patients. Cleaning is normally accomplished by manual wiping, brushing, or flushing or by using mechanical aids (e.g., ultrasonic cleaners, washer decontaminators, washer sterilizers) in conjunction with water and detergents to remove foreign material. In general, reprocessing reusable medical devices involves three steps: 1. First, reprocessing begins at the point of use, which includes initial cleaning and measures to prevent drying of soil and contaminants in and on the device. 2. Second, the device is thoroughly cleaned in the dedicated cleaning area. 3. Third, depending on the intended use of the device, the device will be disinfected or sterilized, and routed back into use. There are few tests that can be used to verify cleaning. To verify cleaning of a given device, one must have a test soil and a quantitative test method for detecting residual soil after cleaning. In addition, manufacturers must consider: a) that exposure to chemicals, such as cleaning agents, could alter the material used in the device; b) whether the materials of construction will absorb or adsorb chemical agents, which could then gradually leach from the material over time; c) how cleaning processes could affect the function of the device; and d) that cleaning processes and tools must be able to contact all areas of the device that could become contaminated.

3 STANDARDS AND REGULATIONS Recently, the U.S. Food and Drug Administration (FDA) began requiring additional information on 510(k) concerning validation of cleaning and sterilisation of reusable medical device. A draft guidance was published on this subject in May It states that: Manufacturers of reusable medical devices should provide adequate labeling that includes instructions for reprocessing and reusing devices safely. The labeling should provide sufficient instructions on how to prepare the device for the next patient use. The user should be able to follow these instructions because they should refer to materials and equipment available to device users. The manufacturer should maintain in the Device Master Record and/or design history file as appropriate, documentation of tests that were performed to demonstrate that the instructions are complete and understandable and can reasonably be executed by the user. The users should ensure that they have the facilities, equipment, and easy access to manufacturer-specified sterilization/disinfection agents to execute the instructions, and that the instructions are followed for every patient use of the device. In Europe, 93/42/EEC directive states that: Where appropriate, the instructions for use must contain the following particulars: [ ] if the device is reusable, information on the appropriate processes to allow reuse, including cleaning, disinfection, packaging and, where appropriate, the method of sterilization of the device to be resterilized, and any restriction on the number of reuses Below is a list of applicable documents for validation of cleaning and sterilization of reusable devices: FDA Draft Guidance: Processing/Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling (March 2, 2011) AAMI TIR30 (2003): A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices AAMI TIR12 (2010): Designing, testing, and labelling reusable medical devices for reprocessing in health care facilities: A guide for medical device manufacturers

4 VALIDATION OF THE CLEANING PROCESS FDA admits that although many FDA-recognized consensus standards related to medical device sterilization are currently available, no standards and no dedicated guidance is currently available related to medical device cleaning. FDA instructions for this validation are: - To use soils that are clinically relevant to the type of device. - To use the worst case (least rigorous) implementation of the cleaning process, - To use medical devices that represent worst case (most contaminated) test devices. Predetermined cleaning test endpoints shall be specified. Validation protocols should be designed to establish that the most inaccessible locations on the devices can be adequately cleaned during routine processing. For all testing, a statistically significant number of replicate samples shall be chosen. This last recommendation of FDA, is not always easy to follow when validation is done on complex and expensive reusable medical devices. 1. PRODUCT FAMILIES Before starting validation, product families shall be carefully established taking into account: - The device shape and size - The materials constituting the device - The soils in contact with the device - The cleaning cycle(s) to be validated - The sterilization cycle(s) to be validated 2. ARTIFICIAL SOIL - INNOCULATION The manufacturer should select an artificial test soil composed of a formulation that includes or accurately represents all materials that the device would likely be subjected to during an actual clinical use, and would create the greatest (worst case) challenge to the cleaning process. FDA does not recommend the use of spore log reduction testing as a method to determine the effectiveness of the cleaning methodology. It is unclear whether or not the removal of bacterial spores directly correlates to the removal of clinical organic soil from the devices. Such testing only indicates how well a process reduces spore count, and provides no information on any other component of organic soil. AAMI TIR 12 provides different type of possible soils. Eg: Hucker s soil: Peanut butter 10g, butter10g, milk 15ml, printer s ink 20 drops, blood 1g, flour 10g, lard 10g, saline 3ml, egg 10g. This soil presents the advantage to represent all the different kind of product to clean (protein, blood, etc ) Soil inoculations should mimic worst case clinical use conditions. - FDA recommends using the artificial soil to inoculate the device in all locations likely to contact patient materials, including all locations that are difficult to clean.

5 - Simulated use conditions for the validation studies should be device specific. Simulated use conditions should account for real world use conditions, to mimic worst case clinical use, for example, the worst case duration of clinical exposure. - All functional procedures (articulations, flexures, manipulations) for which the device is intended to soil the device shall be conducted to represent worst case conditions. If the device is likely to be repeatedly subjected to "pushing" soil into a hard to reach area during use, validation soiling should include repeated soiling to adequately reproduce the "worst case" use situation. - If after use of the device, drying of soil might occur and cleaning might not be performed immediately after use, the validation methods should allow soils to dry for a length of time that simulates worst case (longest duration). 3. CLEANING PROCESS The cleaning validation protocols should specify the shortest times, lowest temperatures, weakest dilutions, etc., for each step of the cleaning instructions. Examples of worst-case scenarios: If the cleaning instructions recommend a 10 to 20 minute pre-soak, the validation protocols should specify 10 minutes or less. If the cleaning instructions advise the user to manually clean at 45ºC ± 5ºC, the validation protocols should specify cleaning at 40ºC or less. If a device consists of several components to be disassembled for adequate cleaning, examine what will result if the device is not disassembled or is incompletely disassembled. Enzymatic Detergents: In general, worst case implies shortest times, lowest temperatures, etc. A noteworthy exception to validation at lowest temperature would be the enzymatic detergents, which typically have optimally effective temperature ranges. Validation protocols should adequately address the temperature range specified in the cleaning instructions for enzymatic detergents. Medical Washers/Disinfectors: If your process validation uses automated medical washers/washer disinfectors or ultrasonic cleaners, your worst-case should include the worst extremes of the intended cycle parameters. 4. TESTING AFTER CLEANING ACCEPTANCE CRITERIA FDA recommends that you use a quantitative test method capable of measuring meaningful levels of clinically relevant soil to meet a related, predetermined cleaning endpoint. Many potential test methods exist for the evaluation of organic soil contamination, and the effectiveness of cleaning processes. AAMI TIR 30 provides a summary of what is available in the literature. Examples of possible tests are: Hydrocarbon oil index as per ISO : Water quality Determination of hydrocarbon oil index Part 2: Method using solvent extraction and gas chromatography Bioburden as per ISO : Sterilization of medical devices -- Microbiological methods -- Part 1: Determination of a population of microorganisms on products Protein rate as per ISO : Washer-disinfectors -- Part 1: General requirements, terms and definitions and tests Endotoxine LAL test as per European Pharmacopoeia Devices should be subjected to a validated method of extraction for testing. Exhaustive extraction is a commonly used method, and extraction using a known quantity of inoculum is also common (but has some procedural limitations). Extractions and calculations should address all hard to

6 clean internal surfaces (such as lumens), and worst case soiled surfaces (such as mated surfaces that have been soiled repeatedly or under pressure). Acceptance criteria shall be defined. Examples of acceptance criteria are: HCT rate < 1.8µg/cm² (AAMI TIR 30) Reduction of at least a 3log10 in recoverable bacteria (total aerobic bacterial count) (AAMI TIR 30) Protein rate 3 µg/device (ISO ) Endotoxin 2.2 EU/cm² (AAMI TIR 30) No visible soil after cleaning

7 VALIDATION OF THE STERILIZATION PROCESS Sterilization shall be validated according to available standards. The most used sterilization method in healthcare facilities is moist heat. In this case, validation shall be conducted as per: ISO : Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices The goal of this validation is to ensure that the sterilization process can inactivate the standard number of 10 6 thermo-resistant micro-organisms Geobacillus stearothermophilus on a group of 3 sterilized devices (inoculation in the critical location of the device). Then the group is exposed to the sterilization agent during a specified period and sterility tests are performed on the devices. This treatment is repeated three times in order to validate the sterilization reproducibility. After this validation, the official duration of the sterilization process will be doubled in order to ensure the sterilization process.

8 ? DESIGN: QUESTIONS TO ASK What are the guidelines and recognized standards in the countries targeted for product sales? Will my reusable medical device be easy to clean? What are my product families? How many samples will be needed for the validation? What type of soil will be in contact with the device? What part of the device will be in contact with the soil in real life clinical use? What are the worst case conditions for clinical use? What are the worst case conditions for cleaning? What type of cleaning agent will be available in healthcare facilities? What type of sterilization will be available in healthcare facilities? Will the materials of the device resist to repeated cleaning and sterilization? What is my recommended cleaning procedure? What is my recommended sterilization procedure?

9 REASONS FOR CHOOSING MEDICAL LAB Acknowledged expertise in medical device testing, and French, European, Chinese and US standards and regulations A full service partner offering training, assistance in establishing validation protocols and testing services An efficient ISO 13485:2003 quality control system ISO accredited lab (by COFRAC, A multilingual sales team (English, German, Italian, Spanish, Arabic, Chinese ) Our experts can assist you with your regulatory processes, and help you determine validation protocols Our collaborative interaction with the MEDICAL GROUP helps you limit your suppliers, save time in having your batches released and reduce your transport costs Fast communication of results (by or on our website)

10 CONTACTS Aurélien BIGNON MedicalLab General Manager Alice MARTINHO Manager of test department Régine GRIES Quality Control Director OUR SALES TEAM Alain BRUNEL Sales Manager Rachid ZEROUALI Export Sales Manager Richard PUZENAT Sales Manager (France)

11 MedicalGroup World Wide

12 5, Chemin du Catupolan Vaulx en Velin - France - Tel. 33 (0) Fax : 33 (0)

STERILIZATION VALIDATION IN THIS SECTION REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES. Radiation Sterilization Validation

STERILIZATION VALIDATION IN THIS SECTION REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES. Radiation Sterilization Validation STERILIZATION VALIDATION All sterilization processes require validation of the efficacy and reproducibility of the process. Depending on the type of sterilization, this may be accomplished by partial,

More information

Best Practices for Reprocessing Robotic Instruments

Best Practices for Reprocessing Robotic Instruments Best Practices for Reprocessing Robotic Instruments Nupur Jain Manager, Validation Engineering Services Intuitive Surgical, US Agenda 1. Background on robotic instruments 2. Reprocessing procedure for

More information

Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy!

Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy! Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy! Dr. Michelle J. Alfa, Ph.D., FCCM Medical Director, Clinical Microbiology, Diagnostic Services of Manitoba, Winnipeg, Canada Disclaimers:

More information

CSA Z314 Series Standards Understanding the Standards

CSA Z314 Series Standards Understanding the Standards CSA Z314 Series Standards Understanding the Standards September 10,2013 Ian Pequegnat Where do standards come from? What influences their development The International Organization for Standardization

More information

REUSABLE MEDICAL DEVICE VALIDATION

REUSABLE MEDICAL DEVICE VALIDATION All sterilization processes require validation of the efficacy and reproducibility of the process. Depending on the type of sterilization, this may be accomplished by partial, sub-lethal, or repetitive

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 14161 First edition 2000-10-01 Sterilization of health care products Biological indicators Guidance for the selection, use and interpretation of results Stérilisation des produits

More information

GENERAL SURGERY DEVICES

GENERAL SURGERY DEVICES GENERAL SURGERY DEVICES LAPAROSCOPIC INSTRUMENTS SUTURE PASSERS TROCARS ULTRASONIC SCALPELS REPROCESSED SURGICAL DEVICES STRYKER S PORTFOLIO OF REPROCESSED SINGLE-USE DEVICES Laparoscopic devices reprocessed

More information

AS/NZS 4187 RELEVANT STANDARDS. Robyn Williams NUM CSSD LBHHS & Standards Australia Chair HE-023

AS/NZS 4187 RELEVANT STANDARDS. Robyn Williams NUM CSSD LBHHS & Standards Australia Chair HE-023 AS/NZS 4187 RELEVANT STANDARDS Robyn Williams NUM CSSD LBHHS & Standards Australia Chair HE-023 Relevant Standards AS 2514 AS 2774 AS 2773 AS 2773.1 Drying cabinets for medical equipment Drying cabinets

More information

Hot Topics in Drug Product Process Validation: A Reviewer s Perspective

Hot Topics in Drug Product Process Validation: A Reviewer s Perspective Hot Topics in Drug Product Process Validation: A Reviewer s Perspective Colleen Thomas, Ph.D. Quality Assessment Lead (Acting) FDA/CDER/OPQ/OPF Division of Microbiology Assessment CASSS CMC Strategy Forum

More information

Christina Bradley Endoscope Decontamination A Webber Training Teleclass. Christina Bradley

Christina Bradley Endoscope Decontamination A Webber Training Teleclass. Christina Bradley ENDOSCOPE DECONTAMINATION Christina Bradley Hospital Infection Research Laboratory City Hospital NHS Trust Dudley Road, Birmingham B18 7QH Hosted by Paul Webber paul@webbertraining.com www.webbertraining.com

More information

What Dentistry Can Learn from AAMI ST79. Best Practices for device and instrument reprocessing

What Dentistry Can Learn from AAMI ST79. Best Practices for device and instrument reprocessing What Dentistry Can Learn from AAMI ST79 Best Practices for device and instrument reprocessing Learning goals: Purpose and goals of AAMI organization. Overview of ST79. Review selected topics: IFUs, tools

More information

Validation Strategy for an Automated Endoscope Reprocessor

Validation Strategy for an Automated Endoscope Reprocessor Validation Strategy for an Automated Endoscope Reprocessor November 16, 2010 Bradley Catalone AER Validation Strategy System Approach page 1 AER Validation Strategy AER Scope-specific connectors Requires

More information

Getinge da Vinci Solution

Getinge da Vinci Solution Getinge da Vinci Solution A comprehensive cleaning solution for da Vinci EndoWrist instruments This document is intended to provide information to an international audience outside of the US. Validated,

More information

One such problem is blood residual inside the channels of a robotic arm.

One such problem is blood residual inside the channels of a robotic arm. SUBJECT: Robotic Arm Testing for residue blood DEPARTMENT: Central Service APPROVED BY: EFFECTIVE: REVISED: 07/2016 PURPOSE: To test for detection of blood residue inside and outside of a robotic arm and

More information

2016. All rights reserved.

2016. All rights reserved. DOCUMENT: Future AAMI/ISO 17664, 12 May 2017 Final Draft Standard Sterilization of health care products Information to be provided by the device manufacturer for the processing of medical devices Public

More information

Draft Guidance for Industry and FDA Staff Processing/Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling

Draft Guidance for Industry and FDA Staff Processing/Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Draft Guidance for Industry and FDA Staff Processing/Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling DRAFT GUIDANCE This guidance document is being distributed for

More information

Guide for Cleaning, Sterilization and Storage of Introducers for Caldera Medical Implants

Guide for Cleaning, Sterilization and Storage of Introducers for Caldera Medical Implants Guide for Cleaning, Sterilization and Storage of Introducers for Caldera Medical Implants M Manufactured by: Caldera Medical, Inc. 5171 Clareton Drive Agoura Hills, CA 91301 U.S. Toll Free: 866-4-CALDERA

More information

Water for Instrument Processing

Water for Instrument Processing Water for Instrument Processing by Marcia Frieze, Case Medical Water, the universal solvent Water can dissolve more substances than any other liquid. It is essentially nonionic or neutral. While alkaline

More information

Best Practices for High Level Disinfection Part I

Best Practices for High Level Disinfection Part I Best Practices for High Level Disinfection Part I Nancy Chobin, RN, AAS, ACSP, CSPM, CFER 2017 Sterile Processing University, LLC. **This in-service has been Approved by the CBSPD for 1.5 CEUs. The Bible

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 14161 Second edition 2009-09-15 Sterilization of health care products Biological indicators Guidance for the selection, use and interpretation of results Stérilisation des produits

More information

ISO/TC 198. Secretariat: ANSI

ISO/TC 198. Secretariat: ANSI DRAFT INTERNATIONAL STANDARD ISO/DIS 17664 ISO/TC 198 Secretariat: ANSI Voting begins on: Voting terminates on: 2016-04-28 2016-07-27 Processing of health care products Information to be provided by the

More information

Best Practices for Reprocessing Endoscopes

Best Practices for Reprocessing Endoscopes Best Practices for Reprocessing Endoscopes Reprocessing of Endoscopes Reprocessing Room Standards The resources referenced include: Canadian Standards Association Provincial Infectious Disease Advisory

More information

March 6, Division of Dockets Management (HFA-305) Food and Drug Administration 5630 Fishers Lane, Room 1061 Rockville, MD 20852

March 6, Division of Dockets Management (HFA-305) Food and Drug Administration 5630 Fishers Lane, Room 1061 Rockville, MD 20852 701 Pennsylvania Avenue, NW Suite 800 Washington, D.C. 20004 2654 Tel: 202 783 8700 Fax: 202 783 8750 www.advamed.org March 6, 2017 Division of Dockets Management (HFA-305) Food and Drug Administration

More information

If All Else Fails Read the Directions. Disclosure. Manufacturer s Instructions. Presenter: Sue Lafferty CHICA Canada Conference, May 2010

If All Else Fails Read the Directions. Disclosure. Manufacturer s Instructions. Presenter: Sue Lafferty CHICA Canada Conference, May 2010 If All Else Fails Read the Directions The Importance of Medical Device Manufacturer s Instructions Presenter: Sue Lafferty CHICA Canada Conference, May 2010 Disclosure Member of 3M Canada Speakers Bureau:

More information

In use test of a Washer-disinfector reprocessing of flexible endoscopes

In use test of a Washer-disinfector reprocessing of flexible endoscopes of a Washer-disinfector reprocessing of flexible endoscopes BHT INNOVA E4 Clean Endoscope Copenhagen Hospital Corporation Denmark of INNOVA E4 Washer Disinfector from BHT, Germany. An aldehyde free thermo-chemical

More information

Efficacy Report Summarization for SoClean 2

Efficacy Report Summarization for SoClean 2 Efficacy Report Summarization for SoClean 2 October 2017 SoClean Inc 36 Town Forest Road Oxford, Massachusetts 01540 USA Tel. 508-363-0418 Email info@soclean.com SoClean 2 is USA FDA Registered 3009534409

More information

Your Sterile Processing Partner. Duraholder IPS

Your Sterile Processing Partner. Duraholder IPS Your Sterile Processing Partner Table of Contents Item Page Duraholder Case Study 17 AAMI ST79 Reference 20 Steam Sterilization Efficacy Testing 21 PRODUCT INDEX Product Page Standard Pouches DURA050/100

More information

ISO Part 1 and Part 2 Compliance Requirements. Cathriona O Neill

ISO Part 1 and Part 2 Compliance Requirements. Cathriona O Neill ISO 11607 Part 1 and Part 2 Compliance Requirements Cathriona O Neill I.S. EN 11607 Introduction ISO 11607 is the principal guidance document. Packaging for terminally sterilised medical devices - Part

More information

C. difficile as a Cleaning Problem New Deep Cleaning Process Beats Disinfection

C. difficile as a Cleaning Problem New Deep Cleaning Process Beats Disinfection C. difficile as a Cleaning Problem New Deep Cleaning Process Beats Disinfection PCS is proud to introduce our new MicroClean Deep Cleaning Process, which promises to change the industry s view on cleaning

More information

Biological Monitoring

Biological Monitoring Biological Monitoring Prof. Duygu PERÇİN, MD Department of Clinical Microbiology Erciyes University Faculty of Medicine, Kayseri-TURKEY duygu.percin@hotmail.com Learning Objectives The attandee will be

More information

Is contamination of bronchoscopes really a problem?

Is contamination of bronchoscopes really a problem? Is contamination of bronchoscopes really a problem? 1 Is contamination of bronchoscopes really a problem? High Level Disinfection does not methodically clean bronchoscopes What we know 210 More than 210

More information

Standards of Infection Prevention in Reprocessing Flexible Gastrointestinal Endoscopes Webinar FAQ

Standards of Infection Prevention in Reprocessing Flexible Gastrointestinal Endoscopes Webinar FAQ Standards of Infection Prevention in Reprocessing Flexible Gastrointestinal Endoscopes Webinar FAQ Disclaimer: The Society of Gastroenterology Nurses and Associates, Inc. assume no responsibility for the

More information

Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar

Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar MONITORING THE SYSTEM 1E PROCESSOR Q: I want to confirm that there is no need for a Biological Indicator (BI) and that use

More information

Case 6. Fight Against Nosocomial Infections

Case 6. Fight Against Nosocomial Infections Case 6 Fight Against Nosocomial Infections Decontamination and microbiological surveillance of endoscopes Prof. dr. Isabel Leroux-Roels Laboratory of Medical Microbiology & Infection Control Team UZ Gent

More information

Safe to Handle? Comparing Manually and Machine-Washed Medical Devices

Safe to Handle? Comparing Manually and Machine-Washed Medical Devices Safe to Handle? Comparing Manually and Machine-Washed Medical Devices Kaumudi Kulkarni, Dzelma Kaczorowski, Alex Bonkowski, Stephen Kovach, and Ralph Basile About the Authors Kaumudi Kulkarni, MS, is manager

More information

FREQUENTLY ASKED QUESTIONS IPAC Lapses

FREQUENTLY ASKED QUESTIONS IPAC Lapses FREQUENTLY ASKED QUESTIONS IPAC Lapses Updated: December 15, 2017 Reprocessing in Community Health Care Settings Medication Administration Laboratory Office design/facilities Reprocessing in Community

More information

sterility assurance Prove the power and your processes

sterility assurance Prove the power and your processes sterility assurance Prove the power and efficacy of your processes sterility assurance programs: an introduction YOUR CHALLENGES Contamination events cost a lot of time and money. Sterility problems can

More information

Validation of Sterilizing Grade Filters

Validation of Sterilizing Grade Filters Validation of Sterilizing Grade Filters Presented by Laura Okhio-Seaman Sartorius Corporation 1 Sterilizing Grade Filters The definition of a sterilizing grade filter is one that will produce a sterile

More information

Endoscope Washer Disinfector Cleaning Process Challenge Devices Research & Validation

Endoscope Washer Disinfector Cleaning Process Challenge Devices Research & Validation Endoscope Washer Disinfector Cleaning Process Challenge Devices Research & Validation Richard Bancroft, B.Sc (Hons), FRSB Science & Technical Director STERIS Corporation Registered Authorising Engineer

More information

Updates to AAMI s ST79 Steam Sterilization Standard Tuesday, January 15, :00 noon 1:30 pm Eastern

Updates to AAMI s ST79 Steam Sterilization Standard Tuesday, January 15, :00 noon 1:30 pm Eastern Updates to AAMI s ST79 Steam Sterilization Standard Tuesday, January 15, 2013 12:00 noon 1:30 pm Eastern Martha Young, MS, BS, CSPDT Martha L. Young, LLC marthalyoung1@aol.com Updates to AAMI s ST79 Steam

More information

Professional standard for Loaner instrumentation

Professional standard for Loaner instrumentation Sterilisatie Vereniging Nederland Professional standard for Loaner instrumentation Standard for loaning or renting of surgical instruments Version 04 November, 2010 Index Index... 2 Preface... 3 The Project

More information

Products made in Switzerland

Products made in Switzerland Products made in Switzerland HYGIENIC AND MEDICALLY CLEAR DIFFERENT PRODUCTS BUT THE SAME QUALITY LABEL HYGIENIC AND MEDICALLY CLEAR 2 MDD TECHNOLOGIES FOR: 1. Cleaning 2. Disinfection 3. High Level Disinfection

More information

AUTOMATED ENDOSCOPE REPROCESSOR. OER-Pro Easy. Fast. Reliable.

AUTOMATED ENDOSCOPE REPROCESSOR. OER-Pro Easy. Fast. Reliable. AUTOMATED ENDOSCOPE REPROCESSOR OER-Pro Easy. Fast. Reliable. Olympus Endoscope Reprocessor The only reprocessor designed by an endoscope manufacturer. The OER-Pro provides high-level disinfection of Olympus

More information

POSITION PAPER Moving from the MDD to the MDR

POSITION PAPER Moving from the MDD to the MDR A summary of Key Changes regarding Sterile Packaging and considerations on recommended changes to standards Introduction EN ISO 11607 specifies the requirements and test methods for materials, preformed

More information

Annex A2. Guidance on Process Validation Scheme for Aseptically Processed Products

Annex A2. Guidance on Process Validation Scheme for Aseptically Processed Products Annex A2 Guidance on Process Validation Scheme for Aseptically Processed Products 1 Table of content 1 PURPOSE... 3 2 SCOPE... 3 3 GENERAL INFORMATION... 3 4 INFORMATION NEEDED FOR ASEPTIC PROCESSES VALIDATION...

More information

USP Chapter 823 USP 32 (old) vs. USP 35 (new)

USP Chapter 823 USP 32 (old) vs. USP 35 (new) USP Chapter 823 USP 32 (old) vs. USP 35 (new) Sally W. Schwarz, MS, BCNP Research Associate Professor of Radiology Washington University School of Medicine St. Louis, MO Why USP Chapter ? FDA has

More information

FDA s Guidance for Industry

FDA s Guidance for Industry Sterile Drug Products Produced by Aseptic Processing - CGMPs Midwest FDC Conference November 4 2005 Susan Bruederle, Investigator FDA / ORA / Central Region Chicago District / Hinsdale IL FDA s Guidance

More information

European Packaging Regulations What to expect from a notified body audit of your packaging

European Packaging Regulations What to expect from a notified body audit of your packaging What to expect from a notified body audit of your packaging Dr. Ralf Walther TÜV Süd Product Service GmbH March 3-5, 2009 www.healthpack.net Content 1. Short introduction of TÜV Süd Product Service 2.

More information

Summary of Proposed Legislative Provisions

Summary of Proposed Legislative Provisions Summary of Proposed Legislative Provisions The evidence obtained by the Committee on Oversight and Government Reform during its investigation of duodenoscope-related patient infections identified significant

More information

Colorants in Medical Devices:

Colorants in Medical Devices: Colorants in Medical Devices: The Spectrum of Current Regulatory Expectations John Iannone Program Manager/ Technical Specialist Overview» Company Introduction» Why use Colorants in Devices?» Regulatory

More information

FREQUENTLY ASKED QUESTIONS FOR THE ADVANTAGE

FREQUENTLY ASKED QUESTIONS FOR THE ADVANTAGE FREQUENTLY ASKED QUESTIONS FOR THE ADVANTAGE What is an Automated Endoscope Reprocessor? An Automated Endoscopes Reprocessor (AER) performs all the cycle steps (fresh water preflush, rinsing, disinfection,

More information

Cleaning Procedures. These guidelines are intended for Bulk Pharmaceutical Chemicals but almost every word applies to Drug Products

Cleaning Procedures. These guidelines are intended for Bulk Pharmaceutical Chemicals but almost every word applies to Drug Products CLEANING DEVELOPMENT Cleaning Procedures as the FDA See It These guidelines are intended for Bulk Pharmaceutical Chemicals but almost every word applies to Drug Products BPC CLEANING (& Drug Products)

More information

Disclaimer. Healthcare facilities are required to have written policies approved by their governing body concerning the services that they provide.

Disclaimer. Healthcare facilities are required to have written policies approved by their governing body concerning the services that they provide. Disclaimer Healthcare facilities are required to have written policies approved by their governing body concerning the services that they provide. NxStage Medical, Inc. has prepared this document to assist

More information

Competency Guide: Care and Handling of Rigid Endoscopes

Competency Guide: Care and Handling of Rigid Endoscopes Competency Guide: Care and Handling of Rigid Endoscopes This document details the basic steps to correctly process and maintain a rigid endoscope. This document can be used to perform a competency review

More information

Are you on track for success?

Are you on track for success? Transitioning from a manual process to automated instrument management Are you on track for success? an ebook from Censis Technologies, Inc. Introduction It is well known that the HITECH Act (Health Information

More information

Sterilization Service Revolution Service Enhancement in HA Hospitals

Sterilization Service Revolution Service Enhancement in HA Hospitals Sterilization Service Revolution Service Enhancement in HA Hospitals Hong Kong Hospital Authority Convention 2014 7 May 2014 LAW Tat Hong Samuel Cluster Operations Manager Central Sterile Supplies Department,

More information

Instructions for use. Reusable surgical instruments. Hersteller:

Instructions for use. Reusable surgical instruments. Hersteller: Hersteller: ROLAN Instruments GmbH Ernst-Thälmann-Str. 50 15859 Storkow / Germany Tel.: 0049 33678 449433 E-Mail: info@rolan-instruments.de Item codes 01-02- 03-04- 05-06- 07-08- 09-10- 11-12- 13-14- 15-16-

More information

Precision. Proficiency. Proximity.

Precision. Proficiency. Proximity. Precision. Proficiency. Proximity. Precision Accuracy in testing is paramount to ensuring product safety and efficacy. And fast turnaround times are critical to keeping your project on schedule. With

More information

Sterile Compounding elearning Course Curriculum 28 lessons with 33 hours of CE. Fundamentals of Sterile Compounding (8 lessons/8 hours CE)

Sterile Compounding elearning Course Curriculum 28 lessons with 33 hours of CE. Fundamentals of Sterile Compounding (8 lessons/8 hours CE) Sterile Compounding elearning Course Curriculum 28 lessons with 33 hours of CE CriticalPoint s Sterile Compounding elearning curriculum is written by industry experts and covers both Chapter and

More information

Test record CFU measurement. Test ID Site/Clinic Spijkenisse

Test record CFU measurement. Test ID Site/Clinic Spijkenisse Test record CFU measurement Test ID 1130036 Site/Clinic Spijkenisse Objective of the test: To verify that there are < 10 CFU/m3 in the clean air flow area. Type of test: CFU test, Microbial precipitation

More information

Microbiological Cleaning Method Validation

Microbiological Cleaning Method Validation Microbiological Cleaning Method Validation The purpose of cleaning procedures should never be to reduce bioburden to an acceptable level! Fergus O Connell QA Manager Eurofins ams Laboratories www.eurofins.com

More information

Cleaning Validation for Medical Device Manufacturinguring

Cleaning Validation for Medical Device Manufacturinguring alconox.com Critical Cleaning Experts SM AQUEOUS CRITICAL CLEANING: A WHITE PAPER What Is Cleaning Validation? 1 Definitions Production and Process Controls Inspection, Measuring, and Test Equipment 2

More information

Tests to Support Sterility Claim. Imtiaz Ahmed

Tests to Support Sterility Claim. Imtiaz Ahmed Tests to Support Sterility Claim Imtiaz Ahmed Sterile Product As per TGO 77, a sterile product must comply with the requirements of the following tests: Sterility Test Bacterial Endotoxins Test Appendix

More information

SERIE 4. rapid, reliable, efficient

SERIE 4. rapid, reliable, efficient SERIE 4 rapid, reliable, efficient SERIE 4 Intuitive an machine 12/07/2014 15:23:00 1 35 C The power at your fingertips A NEW GENERATION OF INTERFACE The Serie 4 PA brings the benefit of a new intuitive

More information

Guidance. Media Fills for Validation of Aseptic Preparations for Positron Emission Tomography (PET) Drugs DRAFT GUIDANCE

Guidance. Media Fills for Validation of Aseptic Preparations for Positron Emission Tomography (PET) Drugs DRAFT GUIDANCE Guidance Media Fills for Validation of Aseptic Preparations for Positron Emission Tomography (PET) Drugs DRAFT GUIDANCE This guidance document is being distributed for comment purposes only. Comments and

More information

Guideline on requirements for the production and control of immunological veterinary medicinal products

Guideline on requirements for the production and control of immunological veterinary medicinal products 14 June 2012 EMA/CVMP/IWP/206555/2010 Committee for Medicinal Products for Veterinary Use (CVMP) Guideline on requirements for the production and control of immunological veterinary Draft agreed by Immunologicals

More information

Designing an effective cleaning procedure for medical devices through laboratory studies

Designing an effective cleaning procedure for medical devices through laboratory studies Designing an effective cleaning procedure for medical devices through laboratory studies Elizabeth Rivera Technical Service Specialist STERIS Corporation elizabeth_rivera@steris.com October 2012 Agenda

More information

pump pressure conveying quantity chamber geometry flow conditions Washing arms trays water quantity dosing temperature foam performance

pump pressure conveying quantity chamber geometry flow conditions Washing arms trays water quantity dosing temperature foam performance Validation of cleaning and disinfecting processes in WDs: Implementation of the ÖGSV- Guideline M. Gehrer, T. Miorini, V. Buchrieser, A. Gruber, G. Maierl Processing of instruments preferably automated

More information

In the Know about.appropriate Use of 3M Attest Rapid Readout Biological Indicators

In the Know about.appropriate Use of 3M Attest Rapid Readout Biological Indicators 3M Attest Sterile U Web Meeting Jan 22, 2009 In the Know about.appropriate Use of 3M Attest Rapid Readout Biological Indicators ww.3m.com/atteststerileuonline 7/15 Modified per Inma s input. SDF Prepared

More information

Entry Level Assessment Blueprint Biotechnology

Entry Level Assessment Blueprint Biotechnology Entry Level Assessment Blueprint Biotechnology Test Code: 4075 / Version: 01 Specific Competencies and Skills Tested in this Assessment: Work Habits Demonstrate professional work habits Demonstrate the

More information

VHP Sterilization and Prion Inactivation

VHP Sterilization and Prion Inactivation WFHSS Workshop Opatija Croatia March 27 th 2009 VHP Sterilization and Prion Inactivation Dr. Georgia Alevizopoulou Clinical Specialist Eastern Europe, Middle East & Africa VHP Sterilization and Prion Inactivation

More information

Bacterial Counts - Quantitative Analysis of Microbes

Bacterial Counts - Quantitative Analysis of Microbes Bacterial Counts - Quantitative Analysis of Microbes Introduction: It is often important to know not only what types of bacteria are in a sample but also how many of them are present. Food manufacturers

More information

Doc No.: MHD 12(0333)/ISO :2006 TITLE: Sterilization of health care products Biological indicators Part 1: General requirements

Doc No.: MHD 12(0333)/ISO :2006 TITLE: Sterilization of health care products Biological indicators Part 1: General requirements DRAFTS IN WIDE CIRCULATION TECHNICAL COMMITTEE: MHD 12 DOCUMENT DESPATCH ADVICE Ref Date MHD 12/T- 64, 65, 66, 67, 68, 69 & 70 26-03-2014 ADDRESSED TO: 1. All members of Hospital Equipment, Sectional Committee,

More information

for IND and RDRC Regulated PET Compounding

for IND and RDRC Regulated PET Compounding Overview of USP Chapter for IND and RDRC Regulated PET Compounding Distributed Manufacturing of PET Radiopharmaceuticals for Multi-Center Clinical Trials SNM, Annual Meeting Toronto, Ontario, Canada

More information

t-doc cycle machine data logging digitize your cycle processes

t-doc cycle machine data logging digitize your cycle processes t-doc cycle machine data logging digitize your cycle processes 2 T-DOC Cycle Digital Documentation monitor your cycle parameters in real-time Increasing demands to document sterile production processes

More information

Overview of a sterility assurance program for PET drugs

Overview of a sterility assurance program for PET drugs Coalition for PET Drug Approval Radiopharmaceutical Sciences Council Overview of a sterility assurance program for PET drugs Eric Webster, PETNET Solutions Disclosures Employee of PETNET Solutions, a Siemens

More information

Process validation in medical devices

Process validation in medical devices Process validation in medical devices Fulfil requirements with expert regulatory guidance 1TÜV SÜD Contents INTRODUCTION 4 VALIDATION PLANNING 5 INSTALLATION QUALIFICATION 7 OPERATIONAL QUALIFICATION 9

More information

Microbiological Consideration for Non-Sterile Pharmaceutical

Microbiological Consideration for Non-Sterile Pharmaceutical May 1-3, 2012 Javits Center New York, NY Microbiological Consideration for Non-Sterile Pharmaceutical Dr. Leonard W. Mestrandrea Principal Consultant MESTRANDREA CONSULTING LLC Title Date Microbial Control

More information

a solution for reducing the

a solution for reducing the Outsourcing the precision cleaning, a solution for reducing the environmental impact Trade fair Hanover 2015 Heiko ZSCHIEDRICH Hanover April 15, 2015 Summary 1. Introduction and presentation of ECP 2.

More information

QUALITY ASSURANCE IN AN MDRD

QUALITY ASSURANCE IN AN MDRD QUALITY ASSURANCE IN AN MDRD MAINTENANCE AND PREVENTATIVE MAINTENANCE MDRD is a machine dependent department washers, pasteurizers, cart washer, sterilizer, ultrasonic, heat sealers, RO water systems All

More information

QUALITY CONTROL/QUALITY ASSURANCE IN THE MOLECULAR MICROBIOLOGY LABORATORY

QUALITY CONTROL/QUALITY ASSURANCE IN THE MOLECULAR MICROBIOLOGY LABORATORY QUALITY CONTROL/QUALITY ASSURANCE IN THE MOLECULAR MICROBIOLOGY LABORATORY Richard L. Hodinka, Ph.D. University of South Carolina School of Medicine Greenville Greenville Health System, Greenville, SC

More information

The Most Probable Number Method and Its Use in QC Microbiology

The Most Probable Number Method and Its Use in QC Microbiology The Most Probable Number Method and Its Use in QC Microbiology Scott Sutton "Microbiology Topics" discusses various topics in microbiology of practical use in validation and compliance. We intend this

More information

Chemical Sterilants and High-Level Disinfectants: Managing Practical, Safe and Effective Use

Chemical Sterilants and High-Level Disinfectants: Managing Practical, Safe and Effective Use Chemical Sterilants and High-Level Disinfectants: Managing Practical, Safe and Effective Use by Donald P. Satchell, Ph.D. Objectives After completion of this self-study activity, the learner will be able

More information

Aseptic Process Validation

Aseptic Process Validation Aseptic Process Validation IMB GMP Information Seminar 27 th September 2012 Gerard Sheridan, Inspector Date Insert on Master Slide Slide 1 Overview Guidance Best Practices Common Deficiencies Slide 2 Aseptic

More information

New Requirements of ISO 17664/AAMI ST 81

New Requirements of ISO 17664/AAMI ST 81 New Requirements of ISO 17664/AAMI ST 81 How the instrument manufacturer validate the cleaning process Klaus Roth SMP GmbH Tübingen Germany 21.09.2017 SMP GmbH 2017 1 Operation, which have been cancelled

More information

CAUTION: FEDERAL LAW (USA) RESTRICTS THESE DEVICES TO SALE BY OR ON THE ORDER OF A PHYSICIAN.

CAUTION: FEDERAL LAW (USA) RESTRICTS THESE DEVICES TO SALE BY OR ON THE ORDER OF A PHYSICIAN. Manufacturer Silony Medical GmbH Leinfelder Strasse 60 70771 Leinfelden-Echterdingen Germany US Representative: Silony Medical Corp. 8200 NW 27TH STR, STE#104, DORAL, FL 33122 USA Phone: +1 305 916 0016

More information

Clostridium difficile Update Dr. Michelle Alfa, Winnipeg Sponsored by ARJO

Clostridium difficile Update Dr. Michelle Alfa, Winnipeg Sponsored by ARJO Clostridium difficile: Update Dr. Michelle Alfa Ph.D, FCCM Diagnostic Services of Manitoba St. Boniface General Hospital Site Hosted by Paul Webber paul@webbertraining.com Sponsored by www.arjo.com www.webbertraining.com

More information

December 2009 CDC-NIH

December 2009 CDC-NIH December 00 CDC-NIH Guidelines for Biosafety Laboratory Competency, CDC and the Assocation of Public Health Laboratories, CDC MMWR Supplement/Vol. 60 April, 0 PI: BSL BMBL th Edition CDC-NIH Dec. 00 N/A

More information

DRAFT UGANDA STANDARD

DRAFT UGANDA STANDARD DRAFT UGANDA STANDARD DUS 1659 First Edition 2016-11-09 Materials in Contact with Food Requirements for Packaging materials Reference number DUS 1659: 2016 UNBS 2016 DUS 1659: 2016 Compliance with this

More information

ISO EP, USP ISO EP, USP ISO EP, USP ISO EP, USP ISO MSDA ISO MSDA

ISO EP, USP ISO EP, USP ISO EP, USP ISO EP, USP ISO MSDA ISO MSDA Laboratory Price List 2018 In addition to the tests described in the list below, Medistri SA offers complete validations (examples: verification of cleaning processes with markers, search for residues

More information

This document is a preview generated by EVS

This document is a preview generated by EVS INTERNATIONAL STANDARD ISO 22196 Second edition 2011-08-01 Measurement of antibacterial activity on plastics and other non-porous surfaces Mesurage de l'action antibactérienne sur les surfaces en plastique

More information

Biomedical Waste Handling and Disposal

Biomedical Waste Handling and Disposal Approved by: Biomedical Waste Handling and Disposal Corporate Director, Environmental Supports Environmental Services Operating Standards Manual Number: Date Approved Next Review May 3, 2018 Purpose Applicability

More information

Microbial Growth and The Control of Microbial Growth (Chapter 6 & 7)

Microbial Growth and The Control of Microbial Growth (Chapter 6 & 7) Microbial Growth and The Control of Microbial Growth (Chapter 6 & 7) Lecture Materials for Amy Warenda Czura, Ph.D. Suffolk County Community College Eastern Campus Primary Source for figures and content:

More information

Chuck Reed at Weiler Engineering, Inc compares aseptic blow-fill-seal technology with traditional aseptic processing of pharmaceutical liquids

Chuck Reed at Weiler Engineering, Inc compares aseptic blow-fill-seal technology with traditional aseptic processing of pharmaceutical liquids Manufacturing Aseptic Advances Chuck Reed at Weiler Engineering, Inc compares aseptic blow-fill-seal technology with traditional aseptic processing of pharmaceutical liquids Acknowledged by the FDA as

More information

Reprocessing Summit. Priority Issues from the AAMI/FDA Medical Device Reprocessing Summit

Reprocessing Summit. Priority Issues from the AAMI/FDA Medical Device Reprocessing Summit Reprocessing 2011 Summit Priority Issues from the AAMI/FDA Medical Device Reprocessing Summit Medical Device Reprocessing Summit Conveners AAMI The Association for the Advancement of Medical Instrumentation

More information

ISO/TC 198. Secretariat: ANSI Sterilization of health care products Microbiological methods

ISO/TC 198. Secretariat: ANSI Sterilization of health care products Microbiological methods DRAFT INTERNATIONAL STANDARD ISO/DIS 11737-1 ISO/TC 198 Sterilization of health care products Microbiological methods Part 1: Determination of a population of microorganisms on product Stérilisation des

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 11737-1 Second edition 2006-04-01 Sterilization of medical devices Microbiological methods Part 1: Determination of a population of microorganisms on products Stérilisation des

More information