1 1 Applying Cleaning Validation Principles in the Manufacture of Medical Devices and Diagnostics August 20, 2014 IVT San Diego Destin A. LeBlanc Cleaning Validation Technologies 2014 Destin A. LeBlanc
2 Focus Initial device manufacture Three areas we will cover In vitro diagnostics Implantables Non-implantables 2
3 3 Won t cover Cleaning of reusable devices Reprocessing of single use devices In vivo diagnostics
4 In vitro diagnostics Product types Liquids Solids Solid on substrate Production issues are similar to standard pharmaceuticals Liquid doses Tablets/Powder doses Patches 4
5 Differences? Scale of manufacture may be much smaller Larger number of products on same equipment Limits are not set on dosing and toxicity, but rather on possible interferences with subsequent diagnostic performance 5
6 How set limits? What effect of carryover on - Performance of next diagnostic Wrong values False positives False negatives Stability of next diagnostic Others? 6
7 Carryover study? Spiking study in lab Diagnostic A into Diagnostic B at different levels Evaluate Performance of B Stability of B (performance at end of approved shelf life) 7
8 Acceptable carryover? Highest level with no effect (NOAEL?) Consider additional safety factor (particularly if are able to achieve values significantly below the NOAEL)
9 Every combination? Depends on validation approach Best to use grouping approach Diagnostic most difficult to clean Diagnostic most sensitive to interferences (based on R&D work or scientific judgment) Consider types of interferences and determine whether most critical ones are components of other diagnostics 9
10 Also consider Effects of residues of cleaning agents Effects of bioburden residues Finally, for small scale consider disposables (such as tank liners) 10
11 Non-implantable med devices Examples Blood pressure cuffs IV supports Oxygen masks 11
12 12 Residues Bioburden Cleaning agents Extractables Processing Chemicals
13 Risk Assessment Bioburden Skin Irritation Functionality Concerns Materials compatibility Lower risk than implantables Some of same issues for analysis and sampling as for implantables 13
14 Implantable med devices Examples Knees Hips Bone screws Artificial hearts Focus for this talk will be metallic implantables 14
15 Higher risk Issues Implantables Biocompatibility Functional compatibility Bioburden/endotoxin Systemic toxicity Materials compatibility 15
16 How different from pharma? Pharma: clean equipment used to manufacture drug Implantable med device: clean the device itself May also clean the equipment used to manufacture device To minimize residue on device To maximize efficiency 16
17 Possible target residues Manufacturing aids Cutting fluids Coolants Burnishing agents Cleaning agent Bioburden Endotoxin Particles 17
18 Acceptable level? Based on potential effects of residue on device use Possible effects & issues Biocompatibility Other safety/toxicity issues Functionality May utilize a safety factor 18
19 19 Possible types of limits Visual Chemical Bioburden Endotoxin Particles
20 20 Limit for endotoxin Will test devices for endotoxin because of FDA concern Limit based on current standard for medical devices 20 EU per device
21 Limit for bioburden Will generally test for bioburden to conform to sterilization validation For established device, bioburden limit based on historical data Mean plus 3 standard deviations Default such as 10 CFU per device For new device, base on data for similar device, or on educated estimate With good cleaning, bioburden should be low 21
22 Limits for process chemicals Option 1 for limits For established product- Assume current production is acceptable Measure residue on current production (3 lots) Set limit at mean plus 3 S.D. 22
23 Example - Option 1 Measure TOC (μg C/device) on 5 devices each from 3 lots Lot 1: 12, 15, 11, 18, 8 Lot 2: 7, 14, 20, 19, 13 Lot 3: 21, 16, 10, 12, 15 Mean = 14 S.D. = 4 Limit is mean + 3 S.D. = 26 μg C per device 23
24 24 Option 1 limit For protocol, each device must meet this requirement May also consider a limit for each lot average (e.g., mean + 1 S.D.) Why not more stringent? If changing the cleaning process for better cleaning, may make it more stringent
25 Process chemicals Option 2 for limits For new product- Determine possible residues Spike clean devices with one or more levels of those residues Test for biocompatibility Lowest level with acceptable results is limit May consider safety factor May consider combination spike (multiple residues) 25
26 Example - Option 2 Only one process cutting fluid used Spike devices at levels of (for example) 10, 20, 40 and 80 μg of cutting fluid per device Test by appropriate biocompatibility procedure(s) Lowest level with acceptable response is limit 26
27 Example - Option 2 (cont.) Results 10 μg pass 20 μg pass 40 μg pass 80 μg - fail Limit would be set at 40 μg cutting fluid/device Could add safety factor & use 20 μg 27
28 28 Example - Option 2 (cont.) Alternative: If know expected levels, test only one level (above expected level) by biocompatibility tests If it passes, it is acceptance limit Example Expect data of μg cutting fluid per device Spike only at 50 μg and test for biocompatibility If 50 μg passes, it is set as limit
29 Issues in limits Effect of residue based on Residue per device? Number of devices implanted or used? One time use or multiple frequency? Is effect based on total residue or residue per surface area? 29
30 30 Cleaning agent limits Use same principles as for process chemicals with additional option Measure baseline for setting limits Spike with known amount(s) to determine limit based on biocompatibility Determine systemic toxicity of cleaning agent (ADI) ADI estimated based on LD50 Same route of administration (IV for systemic effects)
31 31 Cleaning agent systemic toxicity Determine from short term LD 50 values Calculate ADI (acceptable daily intake) from short term toxicity by IV route and body weight Base limit on number of devices implanted/used at one time May use additional safety factor
32 32 Particulates Difficult to establish limits based on spiking and biocompatibility testing Which particles? What sizes? Better approach is to base on baseline data from previous production or for similar product with good track record
33 Why limits for particulates? It is a good measure of overall cleanliness Like TOC can t tell you what particle is due to, but is good indicator of consistency Need to combine with other measures 33
34 Target chemical residues Analytical method dependent upon type of residue Select target residues first Processing chemicals Cleaning (or passivation) agent Bioburden Endotoxin Particles 34
35 35 ASTM method ASTM subcommittee F Extraction of metallic implants Solvent extraction (ultrasonic or reflux) measure dissolved residue gravimetrically Aqueous extraction (ultrasonic) measures dissolved residue gravimetrically Debris (particulate) - insoluble residue by filtration with gravimetric measurement
36 36 ASTM method (2) What measures Total extractable residues on device by that extraction procedure Is it a good method? Yes, for intended purpose
37 37 ASTM method (3) Should it be used for cleaning validation? May not be a good measure of cleaning effectiveness Key is before and after results May be a sledgehammer, when a rubber mallet is needed Note: This is my opinion and is definitely not shared by all
38 Processing chemicals Most are mixture; most are organic Does one use method for a specific component of a possible residue? HPLC FTIR UV Does one use method for a general characteristic of a possible residue? TOC Non-volatile residue (NVR) Conductivity 38
39 39 Other methods Auger electron spectroscopy ESCA GC-MS or HPLC-MS May be good for investigation of problems to narrow nature and cause but probably should be avoided for validation purposes
40 40 Cleaning agents Generally will measure cleaner Two limit approaches Limit of cleaner as whole (most common) Limit for each component (less common) Two analytical approaches Measure individual species (representative) within cleaner Measure gross property (e.g., TOC)
41 41 Bioburden testing Extraction and count Focus on total aerobic count Same method as for bioburden for sterilization validation
42 42 Endotoxin testing Extraction and measure Gel clot Kinetic USP / AAMI
43 43 Particulate testing Sample and measure USP <788> Extract or flush sampling Usually extract with water Particle sizes μ >25 μ
44 44 Sampling methods Direct surface (fiber optics) Swab Extraction
45 45 Direct surface Advantages Immediate results Simplified concept Can focus on worst case locations Avoid recovery issues Issues May requires flat surfaces Calibration Interferences
46 46 Swabs Advantages Direct surface sampling Can focus on worst case locations Mechanical means of removing substances Issues Swab must release analyte Care in swab handling procedures Interferences from swab Swabbing is a manual procedure
47 47 Extraction Advantages Samples all surfaces, including hard to reach Can add sonication/shaking Can vary extraction fluid Consistency Issues Must assure conditions/fluid are adequate to dissolve residue Volume of fluid may limit detection limit
48 Exhaustive extraction Want to make sure extraction is adequate Time Temperature Extracting solvent Sonication/shaking 48
49 Is extraction adequate? Option 1: Extract different samples for different times and compare Sample A: 10, Sample B: 20, Sample C: 30 Assumes initial samples are the same Option 2: Extract same sample multiple times Sample A: 10 and test; additional 10 and test; additional 10 and test 49
50 50 Extraction example #1 Results are average of three devices at each time under defined extraction conditions. Time Results for device A Repeat Results 10 minutes 15 μg 13 μg 20 minutes 20 μg 19 μg 30 minutes 20 μg 19 μg
51 Extraction example #2 Results are average of three devices at each time under defined extraction conditions. Time 1 st 10 minutes 2 nd 10 minutes 3 rd 10 minutes Results for device A Repeat Results 15 μg 13 μg 5 μg 6 μg <1 μg <1 μg 51
52 52 Recovery studies Procedure Spike clean device with known amount of residue Allow to dry Remove in sampling procedure Analyze Compare to amount spiked Done at acceptance limit on device
53 Recovery problems Possible causes of low recovery Volatility of spiked material Residue adhesion to surface Interferences Possible cause of high recovery Interferences 53
54 54 Recovery issues For different sizes In general, recovery may decrease as size increases provided extraction volume constant Worst case = largest device For different geometries In general, recovery may decrease as complexity increases provided extraction volume constant Worst case = more cuts, cannulations, etc.
55 55 Validation protocols Do not reinvent the wheel Follow your company s practices for other process validation protocols - cleaning is another process!!
56 Critical regulatory document GHTF SG3/N99-10 (2004): Process Validation Guidance nical-docs/ghtf-sg3-n qmsprocess-guidance pdf Global Harmonization Task Force replaced by International Medical Device Regulators Forum (IMDRF), but GHTF document still used Also note that cleaning is just a process and covered by process validation guidance 56
57 Validation vs. verification Verification Establish by examination and provision of objective evidence that the specified requirements have been fulfilled (verify device) Validation Establishing by objective evidence that process consistently produces product meeting requirements (validate process) 57
58 58 Fully verifying for cleaning How can fully verify results of cleaning process? Test every device May be possible with nondestructive testing May be more feasible with small number of extremely high value devices
59 GHTF decision tree Is output verifiable? Yes, and verification is sufficient, then fully verify No, then either validate or redesign process GHTF SG3/N99, section
60 How apply to cleaning? Cleaning generally viewed as swing process If can test 100% by non-destructive (and non-contaminating) test procedure (such as FTIR), may not need to validate However, for many applications, have to or will want to validate 60
61 New ASTM guide? Working on new guideline WK33660 Guide for Validating Cleaning Processes Used During the Manufacture of Medical Devices Not yet finalized and approved 61
Designing an effective cleaning procedure for medical devices through laboratory studies
Designing an effective cleaning procedure for medical devices through laboratory studies Elizabeth Rivera Technical Service Specialist STERIS Corporation elizabeth_rivera@steris.com October 2012 Agenda
More informationMicrobiological Cleaning Method Validation
Microbiological Cleaning Method Validation The purpose of cleaning procedures should never be to reduce bioburden to an acceptable level! Fergus O Connell QA Manager Eurofins ams Laboratories www.eurofins.com
More informationPDA Technical Report #26:
PDA Technical Report #26: Implications on Liquid Filter Validation Maik W. Jornitz, Sartorius Corp. Filter Qualification Tests for internal Validation Guides Physical Tests flow rates/delta p throughput
More informationPrecision. Proficiency. Proximity.
Precision. Proficiency. Proximity. Precision Accuracy in testing is paramount to ensuring product safety and efficacy. And fast turnaround times are critical to keeping your project on schedule. With
More information3M Purification Inc. Technical and Scientific Services Global Support for the Life Science Industry. Global Expertise delivered locally
3M Purification Inc. Technical and Scientific Services Global Support for the Life Science Industry Global Expertise delivered locally Global expertise delivered locally As a global leader in measuring,
More informationCLEANING VALIDATION FOOD & BEVERAGE WASTEWATER LABORATORY PHARMACEUTICAL & MEDICAL DEVICE
CLEANING VALIDATION FOOD & BEVERAGE WASTEWATER LABORATORY PHARMACEUTICAL & MEDICAL DEVICE Company Overview International Products Corporation, incorporated in 1923, manufactures specialty cleaners and
More informationMethod Development and Validation for Online UV-Dissolution Methods Using Fiber-Optic Technology
Technical Overview Method Development and Validation for Online UV-Dissolution Methods Using Fiber-Optic Technology Introduction Online fiber-optic and multicell UV-dissolution systems have become increasingly
More informationThe Parenteral Drug Association
C O N F E R E N C E R E P O R T Equipment Cleaning Microbial Control Issues By Destin A. LeBlanc, M.A. Cleaning Validation Technologies Validation: The Parenteral Drug Association (PDA) spring conference
More informationsterility assurance Prove the power and your processes
sterility assurance Prove the power and efficacy of your processes sterility assurance programs: an introduction YOUR CHALLENGES Contamination events cost a lot of time and money. Sterility problems can
More informationUSP Chapter 823 USP 32 (old) vs. USP 35 (new)
USP Chapter 823 USP 32 (old) vs. USP 35 (new) Sally W. Schwarz, MS, BCNP Research Associate Professor of Radiology Washington University School of Medicine St. Louis, MO Why USP Chapter ? FDA has
More informationBest Practices for a Robust Cleaning Validation
Best Practices for a Robust Cleaning Validation Beth Kroeger Technical Services Specialist STERIS Life Sciences & Amanda Deal Senior Technical Services Associate STERIS Life Sciences 1/ Agenda Variables
More informationFast Facts. Trinseo MAGNUM ABS Resins (Acrylonitrile Butadiene Styrene) for Medical Applications
Fast Facts Trinseo MAGNUM ABS Resins (Acrylonitrile Butadiene Styrene) for Medical Applications Fast Facts Unique Advantages for the Medical Market As the demand for attractive medical equipment grows,
More informationValidation of Sterilizing Grade Filters
Validation of Sterilizing Grade Filters Presented by Laura Okhio-Seaman Sartorius Corporation 1 Sterilizing Grade Filters The definition of a sterilizing grade filter is one that will produce a sterile
More informationDerivation and Justification of Safety Thresholds
Derivation and Justification of Safety Thresholds Douglas J. Ball, MS, DABT Chair, PQRI L&E Toxicology Subgroup Research Fellow, Safety Sciences - Pfizer, Inc. Agenda Basic Definitions Current Regulatory
More informationCLEANING VALIDATION WITH RISK ASSESSMENT
CLEANING VALIDATION WITH RISK ASSESSMENT Bangkok, Thailand July 26 2017 Jairaj (Jai) Mehta, Consultant, Promoting the Quality of Medicines 2016. ALL RIGHTS RESERVED. Presentation Outline: Cleaning Validation
More informationUSP Perspective on Pharmaceutical Waters. Antonio Hernandez-Cardoso, M.Sc. Senior Scientific Liaison September 7, 2017
USP Perspective on Pharmaceutical Waters Antonio Hernandez-Cardoso, M.Sc. Senior Scientific Liaison September 7, 2017 Why is Water Important? Raw material Solvent Ingredient Reagent Cleaning agent (hot
More informationCleaning Validation for Medical Device Manufacturinguring
alconox.com Critical Cleaning Experts SM AQUEOUS CRITICAL CLEANING: A WHITE PAPER What Is Cleaning Validation? 1 Definitions Production and Process Controls Inspection, Measuring, and Test Equipment 2
More informationSTERILIZATION VALIDATION IN THIS SECTION REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES. Radiation Sterilization Validation
STERILIZATION VALIDATION All sterilization processes require validation of the efficacy and reproducibility of the process. Depending on the type of sterilization, this may be accomplished by partial,
More informationfor IND and RDRC Regulated PET Compounding
Overview of USP Chapter for IND and RDRC Regulated PET Compounding Distributed Manufacturing of PET Radiopharmaceuticals for Multi-Center Clinical Trials SNM, Annual Meeting Toronto, Ontario, Canada
More informationPrimary Packaging change: Switch from a lyophilisate in vial to a lyophilisate in double chamber cartridge
Primary Packaging change: Switch from a lyophilisate in vial to a lyophilisate in double chamber cartridge Christoph Stark Head Pharmaceutical Development Novartis Pharma AG, Novartis Biologics CASSS CMC
More informationAramus Single-Use 2D Bag Assemblies
LIFE SCIENCES VALIDATION PACKAGE Aramus Single-Use 2D Bag Assemblies High-grade, gamma-stable fluoropolymer film providing higher purity and greater reliability TABLE OF CONTENTS Section 1: Introduction...
More informationSterile Compounding elearning Course Curriculum 28 lessons with 33 hours of CE. Fundamentals of Sterile Compounding (8 lessons/8 hours CE)
Sterile Compounding elearning Course Curriculum 28 lessons with 33 hours of CE CriticalPoint s Sterile Compounding elearning curriculum is written by industry experts and covers both Chapter and
More informationa solution for reducing the
Outsourcing the precision cleaning, a solution for reducing the environmental impact Trade fair Hanover 2015 Heiko ZSCHIEDRICH Hanover April 15, 2015 Summary 1. Introduction and presentation of ECP 2.
More informationOverview of a sterility assurance program for PET drugs
Coalition for PET Drug Approval Radiopharmaceutical Sciences Council Overview of a sterility assurance program for PET drugs Eric Webster, PETNET Solutions Disclosures Employee of PETNET Solutions, a Siemens
More informationMicrobiological Consideration for Non-Sterile Pharmaceutical
May 1-3, 2012 Javits Center New York, NY Microbiological Consideration for Non-Sterile Pharmaceutical Dr. Leonard W. Mestrandrea Principal Consultant MESTRANDREA CONSULTING LLC Title Date Microbial Control
More information10 Things You Should Do Before You Validate Your Next Package
sponsored by 10 Things You Should Do Before You Validate Your Next Package Before you set out to validate a medical device package design or process, you ve got to do your homework. You need to understand
More informationPHARMACEUTICAL TESTING
WHITEHOUSE, NJ PHARMACEUTICAL TESTING Pharmaceutical Expertise for GMP & CMC Testing Our Pharmaceutical Expertise With more than 20 years of experience in a variety of industries, our Whitehouse, New Jersey
More informationHot Topics in Drug Product Process Validation: A Reviewer s Perspective
Hot Topics in Drug Product Process Validation: A Reviewer s Perspective Colleen Thomas, Ph.D. Quality Assessment Lead (Acting) FDA/CDER/OPQ/OPF Division of Microbiology Assessment CASSS CMC Strategy Forum
More informationHeating, ventilation, and air system encompasses heating, ventilation, and air conditioning, which is integral component of
REVIEW ARTICLE Demand of pharmaceutical facility functionality: Validation and qualification of HVAC system Anamika Singh, Sapna Malviya, Anil Kharia Department of Quality Assurance of Modern Institute
More informationErin Patton, MS Senior Product Specialist Charles River Labs, Microbial Solutions
Validation of a Rapid Microbial Method Approach to Validation Case Studies: Equivalence Verification of an Alternate Assay for Microbial Limits Screening and Sterility Testing of Pharmaceutical Products
More informationSingle-use components such as
S i n g l e - U s e APPLICATIONS Development of a Single-Use Filling Needle Implementation in a Single-Use System for Aseptic Filling by Jean-Pascal Zambaux Single-use components such as tubing, connectors,
More informationResponses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar
Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar MONITORING THE SYSTEM 1E PROCESSOR Q: I want to confirm that there is no need for a Biological Indicator (BI) and that use
More informationInitiation of Validation
Initiation of Validation Design Review Who Validates? What Process to Validate? Process Validation Decision Tree When to Validate? Prospective Validation Concurrent Validation Retrospective Validation
More informationCleaning Validation in a Biologics Facility Case Study. CBE Pty Ltd
Cleaning Validation in a Biologics Facility Case Study CBE Pty Ltd This training program is copyright to CBE Pty Ltd and may not be modified, reproduced, sold, loaned, hired or traded in any form without
More informationGuiding Principles for the implementation of fluid management technologies for modern single use aseptic processing
Guiding Principles for the implementation of fluid management technologies for modern single use aseptic processing Jean-Marc Cappia, Vice President Marketing Fluid Management Technologies Agenda 1. Aseptic
More informationConducting Supplier Audits: Ensure Validation Compliance
Conducting Supplier Audits: Ensure Validation Compliance Presented by: Gamal Amer, Ph. D. Principal Premier Compliance Services, Inc. All rights reserved. Do not copy without permission. 1 Outline Why
More informationCleaning Procedures. These guidelines are intended for Bulk Pharmaceutical Chemicals but almost every word applies to Drug Products
CLEANING DEVELOPMENT Cleaning Procedures as the FDA See It These guidelines are intended for Bulk Pharmaceutical Chemicals but almost every word applies to Drug Products BPC CLEANING (& Drug Products)
More informationREUSABLE MEDICAL DEVICE VALIDATION
All sterilization processes require validation of the efficacy and reproducibility of the process. Depending on the type of sterilization, this may be accomplished by partial, sub-lethal, or repetitive
More informationThe Right Analytical Method for the Right Application: TOC Analysis for Cleaning Validation
GE Water & Process Technologies Analytical Instruments Paper The Right Analytical Method for the Right Application: TOC Analysis for Cleaning Validation Advances in technologies and mounting manufacturing
More informationFDA s Guidance for Industry
Sterile Drug Products Produced by Aseptic Processing - CGMPs Midwest FDC Conference November 4 2005 Susan Bruederle, Investigator FDA / ORA / Central Region Chicago District / Hinsdale IL FDA s Guidance
More informationFlexsafe 3D Pre-Designed Solutions for Storage and Shipping Take Your Facility Into The Future
Flexsafe 3D Pre-Designed Solutions for Storage and Shipping Take Your Facility Into The Future 2 Flexsafe 3D Pre-designed Solutions Table Please click on headline to go to the selected page Digital Selection
More informationExtractables and leachables: An Introduction
Extractables and leachables: An Introduction Tim Hulme Smithers Rapra THulme@smithers.com 44(0)1939 252 418 1 Extractables and leachables: An Introduction Tim Hulme Smithers Rapra thulme@smithers.com 2
More informationISPE S GUIDES AND HOW THEY APPLY TO CLEANING AND CLEANING VALIDATION
ISPE S GUIDES AND HOW THEY APPLY TO CLEANING AND CLEANING VALIDATION Stephanie A. Wilkins, PE ISPE NORDIC CLEANING VALIDATION CONFERENCE 20 April 2016 Objectives > ISPE s Risk-MaPP Guide and its impact
More informationcleaning validation criteria
cleaning validation criteria http://www.pharmweb.net/forum/0068/2002/msg01782.html Page 1 of 1 cleaning validation criteria Subject: cleaning validation criteria From: "Hiroaki Matsumoto"
More informationGUIDE TO INSPECTIONS OF STERILE DRUG SUBSTANCE MANUFACTURERS
GUIDE TO INSPECTIONS OF STERILE DRUG SUBSTANCE MANUFACTURERS Note: This document is reference material for investigators and other FDA personnel. The document does not bind FDA, and does no confer any
More informationWHO PUBLIC INSPECTION REPORT (WHOPIR) Quality Control Laboratory
SOP 408.4 Annex D 20, avenue Appia CH-1211 Geneva 27 Switzerland Tel central +41 22 791 2111 Fax central +41 22 791 3111 www.who.int WHO PUBLIC INSPECTION REPORT (WHOPIR) Quality Control Laboratory Part
More informationEvaluating single-use systems
B i o p r o c e s s Validation Extractables and Leachables from Single-Use Disposables Denise Bestwick and Raymond Colton Evaluating single-use systems for extractables and leachables is new territory
More informationCritical Environment Products and Services
Critical Environment Products and Services For over two decades, EP Scientific products and services have defined clean for environmental sampling containers and services. Our proprietary cleaning methods
More informationIMDRF Project: List of international standards recognized by IMDRF management committee members. Progress Report
List of international standards recognized by IMDRF management committee members Progress Report Dr. Matthias Neumann Federal Ministry of Health, Germany Nice, 19 th of March 213 Mandate: 2 Steps IMDRF
More informationPerforming Particle Counting on Opaque and Heavily Contaminated Samples
Performing Particle Counting on Opaque and Heavily Contaminated Samples Author Bill Quesnel / CINRG Systems Inc., Burlington, ON, CANADA Co-Author Alistair Geach / WearCheck Canada Inc., Burlington, ON,
More informationValidation Strategies for Multiproduct Manufacturing
Cleaning and Cleaning Validation Strategies for Multiproduct Manufacturing Facilities John M. Hyde Chairman and Founder Hyde Engineering + Consulting, Inc. Hyde Engineering g + Consulting Europe Limited
More informationProcess Filtration From Pure to Sterile
From Pure to Sterile MAIN FEATURES & BENEFITS: Excellent material resistance towards all aggressive media Inherently hydrophobic PTFE membrane Absolute rating at 0,2 µm (HIMA/ASTM) High flow rates Biologically
More informationAnalysis of elemental impurities in drug products using the Thermo Scientific icap 7600 ICP-OES Duo
Analysis of elemental impurities in drug products using the Thermo Scientific icap 7600 ICP-OES Duo Grégory Lecornet, Applications Specialist, Thermo Fisher Scientific Application Note 43149 Key Words
More informationFlexboy 2D Pre-designed Solutions for Storage Best Performance, Assurance of Supply and Closure Integrity for All Process Steps
Storage Best Performance, Assurance of Supply and Closure Integrity for All Process Steps 2 Flexboy 2D Pre-designed Solutions Table Please click on the headline to go to the selected topic Digital Selection
More informationISO : A Risked Based Approach to Biological Safety
ISO 10993 : A Risked Based Approach to Biological Safety Dr Arthur Brandwood Australian Delegation Leader to ISO TC 194 Biocompatibility and Clinical Trials Previous Director Devices Registration and Assessment
More informationAccredited Laboratory
Accredited Laboratory A2LA has accredited St. Paul, MN for technical competence in the field of Biological Testing This laboratory is accredited in accordance with the recognized International Standard
More informationCritical Analytical Measurements for Bioreactor Optimization. controlling an organism s chemical environment leads to consistent and
Critical Analytical Measurements for Bioreactor Optimization Mettler-Toledo Ingold, Inc., Bedford, MA Abstract Most bioreactor processes share a basic principle; optimizing and controlling an organism
More informationCleaning Validation References Contents
alconox.com Critical Cleaning Experts SM AQUEOUS CRITICAL CLEANING: A WHITE PAPER Cleaning Validation References Contents Pharmaceutical Cleaning Validation...2 Identifying Residues...2 Selecting a Residue
More informationADVANCED TESTING LABORATORY
RESOLVING ISSUES. REDUCING RISK. ADVANCED TESTING LABORATORY PROVIDING PEACE-OF-MIND. ADVANCED TESTING LABORATORY 6954 Cornell Road Cincinnati, OH 45242 513.489.8447 513.489.9291 fax ClientRelations@advancedtesting.net
More informationDisinfectant Qualification Bacteria, Fungi, Viruses and Atypical Organisms A Step by Step Workshop
Microrite, Inc. brings you this unique advanced course in ; part of Microrite s Step by Step Workshop Series Participate in the only workshop which will guide you through the process of choosing a cost
More informationBiological evaluation of medical devices --
Translated English of Chinese Standard: GB/T16886.1-2011 Translated by: www.chinesestandard.net Wayne Zheng et al. Email: Sales@ChineseStandard.net ICS 11.040.01 C 30 NATIONAL STANDARD OF THE PEOPLE S
More informationEfficient Validation Strategies and VMPs
FDA Medical Device Industry Coalition Efficient Validation Strategies and VMPs Anne Holland Wilson Agenda History and Recent Trends Why Validate 21 CFR 820.75 Process Validation Standards and Guidance
More informationWATER & PARTICULATE CONTAMINATION IN OIL & LUBRICATED SYSTEMS
IN OIL & LUBRICATED SYSTEMS Author: Tony Arenas, Y2K Fluid Power, Tony.Arenas@y2kfluidpower.com Abstract or Executive Summary Every major sustainable market in the world from Power Generation, AG, and
More informationPDA: A Global. Association. Matrix Approach to Media Fills. (c) 2012 Catalent Pharma Solutions. All rights reserved.
PDA: A Global Matrix Approach to Media Fills Association (c) 2012 Catalent Pharma Solutions. All rights reserved. Overview Guidance Overview Matrix Introduction Why Matrix Media Fills Sample Matrix New
More informationVisible Particles: Regulatory and Compendial Requirements
Visible Particles: Regulatory and Compendial Requirements John G. Shabushnig, Ph.D. Insight Pharma Consulting, LLC June 2014 Agenda Visible Particle Definitions US FDA US Pharmacopeia (USP) EC GMP s European
More informationFiltration in Preparation of Cell Culture Media and Buffers
Filtration in Preparation of Cell Culture Media and Buffers Cell Culture Media and Buffers in Biopharmaceutical Production Cell culture media and process buffers are used in all biopharmaceutical operations.
More informationAdvancements on implementation of single use technology in vaccine manufacturing
Advancements on implementation of single use technology in vaccine manufacturing October 25 th, 2013 DCVMN Meeting Rio de Janeiro, Brazil Outline Overview of single use products Applications of single
More informationMICROBIAL LIMIT TESTS FOR RAW MATERIALS AND NON-STERILE PRODUCTS REVIEW-DISCUSSIONS-INTERPRETATIONS
Microrite, Inc. brings you this unique learning experience in microbial limit tests for raw materials and non-sterile products-review-discussions-interpretations Part of Microrite s step-by-step webinar
More informationProcess Filtration From Pure to Sterile
From Pure to Sterile MAIN FEATURES & BENEFITS: Extremely durable polypropylene construction Absolute rating at 0,2 µm (HIMA/ASTM) High flow rates Biologically inert Approved for Food Contact Use acc. to
More information3M Purification Inc. Filter Systems for Small Molecule Pharmaceutical Purification
3M Purification Inc. Filter Systems for Small Molecule Pharmaceutical Purification 2 Filter Systems for Pharmaceutical Separations 3M Purification Inc. Core Filtration Applications 3M Purification Inc.
More informationMarch 6, Division of Dockets Management (HFA-305) Food and Drug Administration 5630 Fishers Lane, Room 1061 Rockville, MD 20852
701 Pennsylvania Avenue, NW Suite 800 Washington, D.C. 20004 2654 Tel: 202 783 8700 Fax: 202 783 8750 www.advamed.org March 6, 2017 Division of Dockets Management (HFA-305) Food and Drug Administration
More informationMedidée Services SA. Nano-Tera.ch. 05 February 2015 part 8. PMA, 510k, IDE. Pierre-Alain Sommer
Nano-Tera.ch 05 February 2015 part 8 PMA, 510k, IDE Pierre-Alain Sommer Pierre-alain.sommer@medidee.com www.medidee.com Nano-Tera 2015 05.02.2015 USA/FDA Pre Market Approval System - PMA, Pre Market Notifcation
More informationThreshold of Toxicological Concern Initiatives for Medical Devices Richard Hutchinson, DVM, PhD, DABT J&J Global Surgery Group.
Threshold of Toxicological Concern Initiatives for Medical Devices Richard Hutchinson, DVM, PhD, DABT J&J Global Surgery Group Abstract The Threshold of Toxicological Concern (TTC) is a concept used to
More informationIntroduction to TFF. Sengyong Lee Ph.D. Professor/ Program Chair Biotechnology/ Biology Ivy Tech Community College Bloomington, Indiana
Introduction to TFF Sengyong Lee Ph.D. Professor/ Program Chair Biotechnology/ Biology Ivy Tech Community College Bloomington, Indiana Main Agenda Biomanufacturing and Filtration Filtration Principles
More informationA2LA. R231 Specific Requirements: Threat Agent Testing Laboratory Accreditation Program. December 6, 2017
Laboratory Page 1 of 17 Laboratory December 6, 2017 2017 by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means without the prior written permission of A2LA.
More informationInternational Journal of Pharma and Bio Sciences DEVELOPMENT OF ACCELERATED STABILITY PROTOCOL FOR SILDENAFIL TABLETS A EUROPEAN PERSPECTIVE REVIEW
International Journal of Pharma and Bio Sciences DEVELOPMENT OF ACCELERATED STABILITY PROTOCOL FOR SILDENAFIL TABLETS A EUROPEAN PERSPECTIVE REVIEW SUKHDEV SINGH *1 AND JASBIR SINGH 2 1 Rayat Institute
More informationAAMI Quality Systems White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 1
AAMI s White Paper Comparison of 21 CFR Part 820 to ISO 13485:2016 February 2017 AUTHORS Seb Clerkin, GMP Advisory Services Nicola Martin, Owner, Nicola Martin Consulting Jack Ward, Owner, Ward Sciences
More informationRecent FDA Guidance For Industry; BCS Class 1 and 3 August 2015
Recent FDA Guidance For Industry; BCS Class 1 and 3 August 2015 Bryan Crist Scientific Affairs Manager, Agilent Technologies, Dissolution Systems Dissolution Exchange WebEx Bryan.crist@agilent.com August,
More informationTechnical Guidance on Clinical Evaluation of Medical Devices
Annex Technical Guidance on Clinical Evaluation of Medical Devices I. Purpose The clinical evaluation of medical devices is the assessment procedure conducted by registration applicants to validate whether
More informationUpdate on the IVDR. Sue Spencer
Update on the IVDR Sue Spencer Caution The new regulations are draft the principles have now been agreed but the Annexes are subject to minor changes Further details will be added later pre and post application
More informationPROPOSED DOCUMENT. Global Harmonization Task Force
PROPOSED DOCUMENT Global Harmonization Task Force Title: Quality management system Medical devices - Nonconformity Grading System for Regulatory Purposes and Information Exchange Authoring Group: Study
More informationMetrics in Microbiology Monitoring Practices
Metrics in Microbiology Monitoring Practices Crystal Booth, M.M. Masters of Microbiology from North Carolina State University Former Associate Director of Microbiology at Novartis Metrics in Microbiology
More informationMS Certified Vials. Pre-cleaned and certified vials for mass spectrometry
Pre-cleaned and certified vials for mass spectrometry The FIRST and ONLY pre-cleaned, low particle, low background chromatography vial When your instrumentation, sample handling and methodology is pushing
More informationCleaning Validation: A Timely Solution for Improving Quality and Containing Cost
: A Timely Solution for Improving Quality and Containing Cost by Christopher Crone This article presents an economic case for the use of on-line Total Organic Carbon (TOC) and conductivity analysis for
More informationConsiderations for Ophthalmic Drug Products in Semi-Permeable Packaging
Considerations for Ophthalmic Drug Products in Semi-Permeable Packaging Bausch + Lomb Thresholds and Best Practices for Parenteral and Ophthalmic Drug Products (PODP) February 22-23, 2011 Bethesda 23/02/2011
More informationAppNote 5/2006 ABSTRACT
AppNote 5/26 Stir Bar Sorptive Extraction Based on Restricted Access Material for the Direct Extraction of Drugs and Metabolites in Biological Fluids Wayne M. Mullett and Elizabeth Kwong Merck Frosst,
More informationAvoid recurrent microbial contamination using trending of historical data. El Azab Walid Technical Service Manager STERIS
Avoid recurrent microbial contamination using trending of historical data El Azab Walid Technical Service Manager STERIS Agenda Factors influencing the cleaning and disinfection performance Recurring microbial
More informationTPP Techno Plastic Products AG
1. Identification of the substance / preparation and of the company / undertaking Name of product: Filter device for sterilization of aqueous solutions Material use: Polyether sulfone (PES) membrane material
More informationOne such problem is blood residual inside the channels of a robotic arm.
SUBJECT: Robotic Arm Testing for residue blood DEPARTMENT: Central Service APPROVED BY: EFFECTIVE: REVISED: 07/2016 PURPOSE: To test for detection of blood residue inside and outside of a robotic arm and
More informationMedical Devices. The technical requirements also apply when the submitter is not the legal manufacturer 1 (i.e: a distribution company).
UNFPA Technical requirements for medical devices 1. Introduction The following document provides UNFPA s technical requirements in the procurement of medical devices (medical equipment, renewable medical
More informationQuantos Solutions. Gravimetric Sample Preparation (GSP) Automated Weighing
Quantos Solutions Gravimetric Sample Preparation (GSP) Automated Weighing Innovation Weighing Liquids and Solids to Reach Perfect Concentrations Do you spend a lot of time on sample preparation? Do you
More informationApplication of PAT for Tablet Analysis. Case examples from Novartis Lorenz Liesum, Lead PAT Hamburg, 19 th of April 2013
Application of PAT for Tablet Analysis Case examples from Novartis Lorenz Liesum, Lead PAT Hamburg, 19 th of April 2013 Agenda PAT@Novartis Organization Business Drivers and Cases NIR Spectroscopy for
More informationElemental Impurities Regulations View from a CRO
White Paper Elemental Impurities Regulations View from a CRO Author: Alan Cross, Technical Specialist November 2016 1 www.rssl.com Abstract Regulatory control of elemental impurities in pharmaceutical
More informationPROPYLUX HS rods are exclusively produced by Westlake Plastics.
PROPYLUX HS rods are exclusively produced by Westlake Plastics. Launched in 2004 by Westlake Plastics Europe, PROPYLUX HS has gained wide acceptance by orthopedic device manufacturers in Europe. PROPYLUX
More informationHighly Accurate, Reliable and Real-time Enumeration of Individual Bacteria for the Determination of Raw milk Hygienic Quality
Highly Accurate, Reliable and Real-time Enumeration of Individual Bacteria for the Determination of Raw milk Hygienic Quality Validated to be at least equivalent to the EN-ISO 4833-1:2013 and 4833-2:2013
More informationMedical Device Regulatory Framework 9 SEPTEMBER 2015 FUNDISA CONFERENCE JANE ROGERS
Medical Device Regulatory Framework 9 SEPTEMBER 2015 FUNDISA CONFERENCE JANE ROGERS Key Topics Definitions Essential Principles Classification Conformity Assessment Framework License to Manufacture, Import,
More informationplastics flexible or rigid, insulating or conductive, hydrophilic or hydrophobic, transparent or opaque and chemically and sterilization resistant.
TO VALIDATE OR NOT TO VALIDATE? THAT IS THE QUESTION FOR PLASTICS PROCESSORS IN THE MEDICAL DEVICE INDUSTRY. By Vinny Sastri, PhD, Winovia LLC SPE European Conference on Medical Polymers, Belfast, Northern
More informationPerformance Evaluation and Cleanability Study using Pellicon 3 Cassettes with 30 kd Biomax and Ultracel Ultrafiltration Membranes
Application Note Performance Evaluation Cleanability Study using Pellicon Cassettes with kd Biomax Ultracel Ultrafiltration Membranes Consistent, high performance ultrafiltration throughout multiple process
More information