1 ANALYSIS of MYCOTOXINS Monitoring and control Wim Traag, Theo de Rijk and Paul Zomer
2 CONTENT Applicable methods Which method should I use?? Methods used at RIKILT Monitoring program Method development
3 ANALYTICAL METHODS Screening methods: TLC ELISA In between methods: HPLC (with IAC) GC-ECD (after deriv.) Confirmatory methods: LC-MS (MS or TOF) GC-MS (after deriv.)
4 Screening methods Elisa kits
5 In between methods
6 Pro s and con s of single compound analyis Pro Low detection limits Simple methods Con Labour intensive Only one mycotoxin detected
7 IN BETWEEN METHODS PRINCIPLE of IAC SAMPLE WASHING ELUTION HPLC-UV Detection
8 IAC, HOW IS IT PERFORMED
9 Single compound, HPLC equipment Switching valve Injection loop Pump HPLC-column UV or Fluorescence detector
10 DEOXYNIVALENOL Reference solution 0.4 µg/ml Negative sample animal feed H 3 C O OH O O OH OH Positive sample animal feed DON (Type B Trichothecene)
11 Single compound, HPLC equipment Switching valve Injection loop Pump HPLC-column Kobra-cel Fluorescence detector
12 Aflatoxin analysis: the Kobra-cell Without Kobra-cell With Kobra-cell
13 G2 G1 B2 B1 AFLATOXINS
14 Chromatogram of spiked maize after cleanup with IAC. Vicam versus R-Biopharm Area 5,00 4,50 4,00 3,50 3,00 2,50 2,00 1,50 1,00 0,50 0,00 3-Acetyl DON DON 15-Acetyl DON 0,0 1,0 2,0 3,0 4,0 5,0 6,0 7,0 8,0 9,0 10,0 11,0 12,0 13,0 14,0 Rt (min) Vicam Chromatogram of spiked maize after cleanup with IAC. R-Biopharm
15 Single compound methods: limits of detection HPLC - IAC analyse Aflatoxin B1: 0.10 µg/kg Aflatoxin M1: µg/kg Ochratoxin: 0.10 µg/kg Zearalenone: 5.0 µg/kg Deoxynivalenol: 0.10 µg/kg
16 Confirmatory methods
17 Mycotoxins, the LC-MS (MS) approach Pro High selectivity General applicable Confirmatory analysis No need for base-line separation Con High training level required Expensive system needed
18 LC-MS versus LC-MS/MS LC-MS techniques generally yield only one molecular ion, little or no fragmentation so structural information from fragmentation is lacking Obtaining structural information requires inducing of fragmentation
19 Fragmentation Why? Fragment ions are characteristic for the molecular structure of the ion and therefore yield information suitable for confirmatory analysis : In order to induce fragmentation LC-MS-MS is required
20 Mass analyzer: triple quadrupole
21 Fragmentation Prod. scan: 236; 10 µg/ml NPAOZ, 5 µl/min, coll. 17 ev Daughters of 236ES+ 3.41e % m/z
22 LC-MSMS, Limits of Detection Aflatoxin B1, B2, G1, G2: 5 µg/kg Fumonisin B1, B2, B3: 0.05 mg/kg T2: 0.10 mg/kg HT2: 0.10 mg/kg Ochratoxin: 0.01 mg/kg Zearalenon: mg/kg Deoxynivalenol: 0.10 mg/kg
23 Example: Mycotoxins in animal feed Reference solution mycotoxins Standaard myc_u1_030131_35 DON FB1 FB2 AB2 AG2 AG1 DAS AB1 HT2 bzl azl T2 OTA ZON STER AcDON % Time
24 WHICH METHOD SHOULD I USE????????
25 Possible approach (1) EIISA kits if interested in one mycotoxin and required answer is negative or suspected If suspected confirmation is needed for legal purposes Result can be quantified (semi) HPLC UV/FLU if interested in one (group) mycotoxin(s) and quantification is required. If above a certain level, result has to be confirmed
26 Possible approach (2) LC MS/MS if interested in limited number of target compounds ( n= < 25) and a quantitative result is required LC TOF- MS if interested in unknown/metabolites, target compounds (unlimited number) can be included
27 Mycotoxins at RIKILT National Plan Animal Feed: 300 samples National Plan Residues Animal Products: 50 samples Private companies :? samples Special projects :? samples NRL mycotoxins in feed Method development
28 Routine Mycotoxin Analysis in Feed IAC HPLC KOBRA-cell Fluorescence (Aflatoxins, validated, accredited) LC-MS-MS (for other mycotoxins) Validated and accredited for FB1/2/3, T2, HT2, OTA, ZON, DON and AFLA B1 (for Afla only screening)
29 IAC-method Extraction with chloroform Aliquot redissolved in MeOH/water IAC cleanup HPLC Kobra-cell + Fluorescence detection Detection Limit: 0.10 μg/kg Expanded Measurement Uncertainty U: 10% at 5 μg/kg, 25% at 1 μg/kg
30 LC-MSMS method Double extraction: - Acidified water + Acetonitril/water (84/16) No cleanup! HPLC MSMS detection: ESI+ (ZON: ESI-) Standard addition quantification - Addition at reporting level - Quantification max x2, otherwise re-analysis
31 Number of samples Total? Other Feed Afla IAC Total? Other Feed Multi
32 Percentage mycotoxins reporting limit Fumonisin DON ZON CORN 48% 48% 49% SOY 21% SILAGE 18% COMPOUND FEED FOR PIGS 34% 13% 15% WHEAT 20% SORGHUM 63%
33 AFLA B1 monitoring result Total number of samples = 1858 Number of samples < 1 ug/kg = 1750 Number of samples > 1 < 5 ug/kg = 85 Number of samples > 5 < 10 ug/kg = 15 Number of samples >10 < 20 ug/kg = 5 Remaining samples have levels of 33; 40 and 140 ug/kg Resultaten Aflatoxine B tm 2007.xls
34 Method development UPLC-MSMS Sharp peaks better s/n ratio = lower LOD s Shorter Rt s, at least factor 2 Stable isotope dillution Multi IAC cleanup Two manufacturers Lowering ionization suppression
35 Wageningen UR Thank you for your attention!
[ ONGOING DEDICATION TO FOOD SAFETY ]
PRODUCT GUIDE 2015 [ ONGOING DEDICATION TO FOOD SAFETY ] VICAM s mission is rooted in the fundamental human need for safe and abundant food supply. As awareness of the dangers of naturally occurring agricultural
More information[ ONGOING DEDICATION TO FOOD SAFETY ]
PRODUCT GUIDE 2017 [ ONGOING DEDICATION TO FOOD SAFETY ] VICAM s mission is rooted in the fundamental human need for safe and abundant food supply. As awareness of the dangers of naturally occurring agricultural
More informationMultiQuant Software Version 3.0
MultiQuant Software Version 3.0 Improving Data Quality and Processing Throughput with Better Peak Integration, Quantitative and Qualitative Compound Review for the Analysis of Food, Drinking Water, and
More informationApplication Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany
Quantification of trace-level herbicides in drinking water by online enrichment with the Agilent Infinity Series Online-SPE Solution and Triple Quadrupole MS Detection Application Note Environmental Author
More informationExploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma
Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma Yun Wang Alelyunas, Mark D. Wrona, Jim Murphy, Angela Doneanu, Gregory
More informationImproving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS
Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS Using the SCIEX M3 MicroLC system for Increased Sensitivity in Antibody Quantitation Remco van Soest and Lei Xiong SCIEX, Redwood City,
More informationConfident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions
Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions Screening and Confirmation Chemical Analysis Solution Unit November, 2009 Page 1 Agilent s New 6540 Ultra High
More informationThermo Scientific TSQ Quantum Access MAX
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Scientific TSQ Quantum Access MAX UNSURPASSED PRICE TO PERFORMANCE Superior
More informationOn-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water
application note On-line SPE-LC/MS/MS of Pesticides On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water API 3200 LC/MS/MS System Overview The feasibility of
More informationR-Biopharm AG. Product catalogue. Food & Feed Analysis. R-Biopharm for the safety of your analysis.
R-Biopharm AG Product catalogue Food & Feed Analysis 2017 R-Biopharm for the safety of your analysis. Product catalogue 2017 Food & Feed Analysis R-Biopharm AG Product catalogue 2017 Content Overview
More informationProteins. Patrick Boyce Biopharmaceutical Marketing Manager Waters Corporation 1
Routine Characterization of mabs and Other Proteins Patrick Boyce Biopharmaceutical Marketing Manager Europe and India 2011 Waters Corporation 1 Agenda Why? What scientific challenges? Technology Example
More informationAflatoxin & Ochratoxin Analysis. A Completely Automated Solution with the Freestyle, ThermELUTE, and SMART Immunoaffinity Cleanup Columns
Aflatoxin & Ochratoxin Analysis A Completely Automated Solution with the Freestyle, ThermELUTE, and SMART Immunoaffinity Cleanup Columns Visit us at www.lctech.de! Background Pickering Labs primarily makes
More informationApplication Note. Abstract. Authors. Introduction. Environmental
EPA Method 538: Determination of Selected Organic Contaminants in Drinking Water by Direct Aqueous Injection with the Agilent 646 Triple Quadrupole LC/MS System Application Note Environmental Authors Imma
More informationApplication of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study
Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study Application note Biotherapeutics and Biologics Authors Suresh Babu C.V., Anne
More informationTrastuzumab Glycan Batch-to-Batch Profiling using a UPLC/FLR/Mass Spectrometry Platform
Trastuzumab Glycan Batch-to-Batch Profiling using a UPLC/FLR/Mass Spectrometry Platform Ying Qing Yu, Joomi Ahn, and Martin Gilar Waters Corporation, Milford, MA, U.S. APPLICATION BENEFITS Used together,
More informationApplication Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany
D-LC/MS Characterization of Charge Variants Using Ion Exchange and Reversed-Phase Chromatography Multiple Heart-Cutting D-LC Analysis of Innovator versus Biosimilar Monoclonal Antibodies Application Note
More informationCAMAG TLC-MS INTERFACE 2
CAMAG TLC-MS INTERFACE 2 WORLD LEADER IN PLANAR CHROMATOGRAPHY IDENTIFICATION AND CONFIRMATION OF UNKNOWN SUBSTANCES The elution-based CAMAG TLC-MS Interface 2 is a highly convenient and versatile instrument
More informationR-Biopharm AG Product catalogue. Food & Feed Analysis. R-Biopharm dedicated to food safety.
R-Biopharm AG 1988 2018 Product catalogue Food & Feed Analysis 2018 R-Biopharm dedicated to food safety. Product catalogue 2018 Food & Feed Analysis R-Biopharm AG Product catalogue 2018 Content Overview
More informationThe lab approval in the framework of BNN is a modular system consisting of two columns: Column A) Parameter groups, and Column B) Food groups.
Guidelines for laboratory approval by Bundesverband Naturkost Naturwaren (BNN) e.v. (Federal Association for Natural Foods and Natural Products inc. soc.) Editorial revision 01 September 2015 The results
More informationRapid Soft Spot Analysis using the SCIEX Routine Biotransform Solution
Rapid Soft Spot Analysis using the SCIEX Routine Biotransform Solution Rapidly Identify Top Metabolites with the SCIEX X500R QTOF System and MetabolitePilot TM 2.0 Software Shaokun Pang and Ian Moore SCIEX,
More informationDETERMINATION OF MYCOTOXINE DEOXYNIVALENOL IN WHEAT FLOUR AND WHEAT BRAN
Revue de Cytologie et Biologie végétales-le Botaniste 2005-28 (s.i.): 64-70 DETERMINATION OF MYCOTOXINE DEOXYNIVALENOL IN WHEAT FLOUR AND WHEAT BRAN by F. ISRAEL-ROMING (1) M. AVRAM (2) V. GAGIU (2) G.
More informationVyužití cílené proteomiky pro kontrolu falšování potravin: identifikace peptidových markerů v mase pomocí LC- Q Exactive MS/MS
Využití cílené proteomiky pro kontrolu falšování potravin: identifikace peptidových markerů v mase pomocí LC- Q Exactive MS/MS Michal Godula Ph.D. Thermo Fisher Scientific The world leader in serving science
More informationMassPREP On-Line Desalting Cartridge
CONTENTS I. INTRODUCTION II. INSTALLING THE MASSPREP ON-LINE DESALTING CARTRIDGE INTO THE SENTRY.1 X 1 MM GUARD COLUMN HOLDER III. RECOMMENDED LC/MS SYSTEM CONFIGURATION TO MINIMIZE MS SOURCE CONTAMINATION
More informationĽUBOMÍR DAŠKO, DRAHOMÍRA RAUOVÁ, ELENA BELAJOVÁ and MILAN KOVÁČ. Food Research Institute, Bratislava, Slovak Republic
Determination of Fumonisins B 1 in Beer ĽUBOMÍR DAŠKO, DRAHOMÍRA RAUOVÁ, ELENA BELAJOVÁ and MILAN KOVÁČ Food Research Institute, Bratislava, Slovak Republic Abstract DAŠKO Ľ., RAUOVÁ D., BELAJOVÁ E., KOVÁČ
More informationHigh Resolution Accurate Mass Peptide Quantitation on Thermo Scientific Q Exactive Mass Spectrometers. The world leader in serving science
High Resolution Accurate Mass Peptide Quantitation on Thermo Scientific Q Exactive Mass Spectrometers The world leader in serving science Goals Explore the capabilities of High Resolution Accurate Mass
More informationCSIR-National Chemical Laboratory. Analytical/Testing Charges. Table of contents
CSIR-National Chemical Laboratory Analytical/Testing Charges Table of contents Serial no. Page no. 01 Biochemical Characterization 02 02 Elemental Analysis 02 03 ESCA Analysis 02 04 Magnetic measurements
More informationGenotoxicity is the property of a compound
Impurities Analysis in Pharmaceuticals: Genotoxicity is the property of a compound known to have irreversible effects on the structure and functionality of the DNA in cells and cause DNA loss, DNA replication
More informationCertainty within minutes. rapidust mycotoxins.
Certainty within minutes. rapidust mycotoxins. The challenge: The rapidust solution: Mycotoxins occur heterogeneously distributed throughout grain lots, thus complicating the correct assessment of contaminations
More informationDevelopment of a Fast and Cost-Effective Multi-Residue Method to Determine Pesticides in Tobacco by LC/MS/MS
TECHNICAL NOTE Pesticides in Tobacco Development of a Fast and Cost-Effective Multi-Residue Method to Determine Pesticides in Tobacco by LC/MS/MS API 4000 LC/MS/MS System Overview This application note
More informationA Complete Workflow Solution for Intact Monoclonal Antibody Characterization Using a New High-Performance Benchtop Quadrupole- Orbitrap LC-MS/MS
A Complete Workflow Solution for Intact Monoclonal Antibody Characterization Using a New High-Performance Benchtop Quadrupole- Orbitrap LC-MS/MS Zhiqi Hao, 1 Yi Zhang, 1 David Horn, 1 Seema Sharma, 1 Shiaw-Lin
More informationEPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube
EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube Application Note Environmental Authors Don Noot, Ralph Hindle Vogon Laboratory Services Cochrane, AB
More informationHDX-MS at Waters Waters Corporation 1
HDX-MS at Waters 2011 Waters Corporation 1 Hydrogen Deuterium Exchange MS systems Dedicated standards and UPLC BEH separation chemistries nanoacquity UPLC with HDX technology Xevo G2-S QTof MS and Synapt
More informationWith the globalization of trade, food production
November 2008 Comprehensive Screening, Confirmation, and Quantification of Organic Pesticides in Foods by GC MS and LC MS The global focus on keeping food safe has intensified, resulting in major changes
More informationSample Clean-Up with Immunoaffinity Columns and Gel Permeation Chromatography: A Product Overview
Sample Clean-Up with Immunoaffinity Columns and Gel Permeation Chromatography: A Product Overview Western Canada Trace Organic Workshop April 27, 2010 Wendy Rasmussen Technical Sales Manager wendyr@pickeringlabs.com
More informationCHROMATOGRAPHY/MASS SPECTROMETRY
DETERMINATION OF 1,4-DIOXANE IN DRINKING WATER BY SOLID PHASE EXTRACTION (SPE) AND GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS) WITH SELECTED ION MONITORING (SIM)* EPA Method 522 Part Number: EU52112M6
More informationLiquid Chromatograph Liquid Chromatograph Mass Spectrometer. CL Series. for In Vitro Diagnostic Use
Liquid Chromatograph Liquid Chromatograph Mass Spectrometer CL Series for In Vitro Diagnostic Use Consistency. Performance. Reliability. Shimadzu s CL Series LC, LC-MS, and LC-MS/MS for in vitro diagnostic
More informationTyler Trent, Applications Sales Specialist; Teledyne Tekmar Page 1
Application Note Method Validation of the AutoMate-Q40 for Pesticide Residues in Grapes Tyler Trent, Applications Sales Specialist; Teledyne Tekmar Page 1 Abstract The QuEChERS (Quick-Easy-Cheap-Effective-Rugged-Safe)
More informationMaximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns
Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns Hongwei Xie, Martin Gilar, and Jeff Mazzeo Waters Corporation, Milford, MA U.S. APPLICATION BENEFITS The ACQUITY UPLC
More informationAutomated Real-Time Determination of Bromate in Drinking Water Using LC-ICP-MS and EPA Method Application
Automated Real-Time Determination of Bromate in Drinking Water Using LC-ICP-MS and EPA Method 321.8 Application ICP-MS Author Jason A. Day, Anne Vonderheide, Joseph A. Caruso University of Cincinnati Cincinnati,
More informationState of the art in chromatographic analysis (LC MS/MS) of steroidal estrogens in surface water Status and Outlook
Schweizerisches Zentrum für angewandte Ökotoxikologie Centre Suisse d écotoxicologie appliquée Eawag-EPFL State of the art in chromatographic analysis (LC MS/MS) of steroidal estrogens in surface water
More informationDETERMINATION OF PROTEIN BINDING BY UPLC/MS/MS
DETERMINATION OF PROTEIN BINDING BY UPLC/MS/MS Darcy Shave and Pete Alden Waters Corporation, Milford, MA, U.S.A. INTRODUCTION Plasma protein binding (PPB) can significantly affect the therapeutic action
More informationA Unique LC-MS Assay for Host Cell Proteins(HCPs) ) in Biologics
A Unique LC-MS Assay for Host Cell Proteins(HCPs) ) in Biologics Catalin Doneanu,, Ph.D. Biopharmaceutical Sciences, Waters September 16, 2009 Mass Spec 2009 2009 Waters Corporation Host Cell Proteins
More informationAB SCIEX TripleTOF 5600 SYSTEM. High-resolution quant and qual
AB SCIEX TripleTOF 5600 SYSTEM High-resolution quant and qual AB SCIEX TripleTOF 5600 SYSTEM AB SCIEX is a global leader in the development of life science analytical technologies that help answer complex
More informationCertified Reference Material - Certificate of Analysis
Certified Reference Material - Certificate of Analysis S(-)-Nicotine, Primary Standard N-008 Page 1 of 7 Catalog Number: N-008 H Lot: Expiration: October 2019 N Description: S(-)-Nicotine in Methanol.
More informationImproved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental
Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental Samples Michael S.Young, Jeremy C. Shia, Kim vantran, Kevin M. Jenkins and Masayo Yabo Waters Corporation 34 Maple Street,
More informationTechnical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany
New Features of the Agilent Method Scouting Wizard for Automated Method Development of Complex Samples Analysis of Large Data Sets by Method Scouting Reports and Automated Adjustment of Flow Rates and
More informationCurrent practice and future vision on metabolite profiling and quantification in Drug development
Current practice and future vision on metabolite profiling and quantification in Drug development Geert Mannens EBF Focus workshop Metabolite Profiling and Quantification Strategies in Drug R&D Sept 25,
More informationConcentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography
Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography Carl Fisher, Monika Verma, and Pranathi Perati; Thermo Fisher Scientific,
More informationHigh-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns
High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns Agilent AdvanceBio SEC 3 Å, 2.7 µm columns Application note Bio-Pharmaceutical Author
More informationOn-Line. User s Guide SPE CARTRIDGES. for Rapid Cleanup and Extraction of Analytes
On-Line SPE CARTRIDGES for Rapid Cleanup and Extraction of Analytes User s Guide What is the strata-x on-line SPE cartridge? The strata-x on-line cartridge combines the revolutionary benefits of the patent
More informationKOBRA CELL Art. No. K01
R-Biopharm Rhône Ltd KOBRA CELL Art. No. K01 Instruction Manual R-Biopharm for the safety of your analysis. R-Biopharm Rhone Ltd 2 Contents 1. Introduction to the KOBRA CELL... 4 1.1 Intended Use... 4
More informationSAFEFOODERA-EMTOX project 7 FP Dr. E. Ioannou Kakouri, Chief Chemist, EFSA focal point
Effect of Climate Change on Emerging Natural Toxins in Plants and Seafood Production Επιδπάζειρ ηων Κλιμαηικών Αλλαγών ζηη Δημιοςπγία Αναδςόμενων Φςζικών Τοξινών και ζηην Παπαγωγή Φςηών και Θαλαζζινών
More informationRapid Peptide Catabolite ID using the SCIEX Routine Biotransform Solution
Rapid Peptide Catabolite ID using the SCIEX Routine Biotransform Solution Rapidly Identify Major Catabolites with the SCIEX X500R QTOF System and MetabolitePilot TM 2.0 Software Ian Moore and Jinal Patel
More informationHeart-cut 2D-LC/MS approach for pharmaceutical impurity identification using an Agilent 6540 Q-TOF LC/MS System
Heart-cut 2D-LC/M approach for pharmaceutical impurity identification using an Agilent 640 Q-TF LC/M ystem Application ote Pharmaceuticals Author iji Joseph Agilent Technologies, Inc. Bangalore, India
More informationEtG / EtS Analysis UCT INNOVATION THROUGH CHEMISTRY FORENSICS
FORENSICS UCT EtG / EtS nalysis INNOVTION THROUGH CHEISTRY EtG and EtS nalysis UCT has developed a rapid solution to combat the significant ion suppression EtG/EtS urine analysis typically suffers from
More informationProtocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016
Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016 1.0 Method background: This method employs full evaporation technique headspace
More informationQuantitation of cyanotoxins in drinking water according to EPA 544 guidelines
APPLICATION NOTE 64968 Quantitation of cyanotoxins in drinking water according to EPA 544 guidelines Authors Ali Haghani, 1 Andy Eaton, 1 Neloni Wijeratne, Claudia Martins 1 Eurofins Eaton Analytical,
More informationAutomated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module
Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module Application Note Environmental Authors Shane Snyder Department of
More informationLC/MS. Why is it the fastest growing analytical technique?
LC/MS Why is it the fastest growing analytical technique? Discussion topics Evolution of LC/MS Advantages of API Why should I use LC/MS? LC/MS markets Evolution of LC/MS interfaces 1970s to Present Moving
More informationDevelopment of a multiplex flow cytometric microsphere immunoassay for mycotoxins and evaluation of its application in feed
Mycotox Res (2011) 27:63 72 DOI 10.1007/s12550-010-0077-0 ORIGINAL PAPER Development of a multiplex flow cytometric microsphere immunoassay for mycotoxins and evaluation of its application in feed Jeroen
More informationThe Use of LC-MS in Systematic Toxicology Analysis
CAT Meeting Phoenix, AZ June 8, 7 Presentation outline Methods used in Systematic Toxicology Analysis The Use of LC-MS in Systematic Toxicology Analysis LC/MS approach ChromaLynx software Chromatogram
More informationValidation Report 19
EURL for Cereals and Feeding stuff National Food Institute Technical University of Denmark Validation Report 19 Determination of pesticide residues in oat, rye and wheat by GC-MS/MS and LC-MS/MS (QuEChERS
More information[ product solution ] Waters Oasis µ Elution PlateS. Patented Innovation. Elution volume as low as 25 μl. No evaporation and reconstitution
[ product solution ] Waters µ Elution PlateS Elution volume as low as 25 μl No evaporation and reconstitution Ideal for small sample volumes Up to a 15x increase in concentration Patented Innovation Now
More informationAgilent 6430 Triple Quadrupole LC/MS System
Agilent 6430 Triple Quadrupole LC/MS System Ideal Quantitative LC/MS for UHPLC with Dynamic MRM and Fast Polarity Switching, plus Unsurpassed Sensitivity with Chip LC Summary The Agilent 6430 Triple Quadrupole
More informationAnalysis of Targeted and Non-Targeted Identified Contaminants in Storm Water Retention Ponds Using LC-HRMS With Online Solid Phase Extraction
Analysis of Targeted and Non-Targeted Identified Contaminants in Storm Water Retention Ponds Using LC-HRMS With Online Solid Phase Extraction Gordon Getzinger, 1 P. Lee Ferguson, 1 Jonathan Beck, 2 Charles
More informationMeasurement of Hepcidin-25 in Serum Using the Agilent 6490 Triple Quadrupole LC/MS with ifunnel Technology
Measurement of Hepcidin-25 in Serum Using the Agilent 6490 Triple Quadrupole LC/MS with ifunnel Technology Application Note Clinical Research Authors Christophe Hirtz, Constance Delaby, Pauline Bros, Jérôme
More informationDetermination of organochlorine pesticide residues in eggs by gel-permeation chromatography (GPC) cleanup and GC-ECD
Determination of organochlorine pesticide residues in eggs by gel-permeation chromatography (GPC) cleanup and GC-ECD 1. Experimental Section LabTech, Inc. 1.1 Instruments and reagents AutoClean automatic
More informationAdvantages of Ion Mobility QTOF for Characterization of BioPharma Molecules
Advantages of Ion Mobility QTOF for Characterization of BioPharma Molecules Add a New Dimension to your Research Capability with Agilent s New Drift Ion Mobility QTOF System David Wong, Ph.D. Sr. Application
More informationAccurately Identify and Quantify One Hundred Pesticides in a Single GC Run
Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run Application Note Author Jessica Westland Agilent Technologies, Inc. Abstract A selected target compound list of 195 various pesticides
More informationImplement High Quality, Routine GC-MS/MS Multi-Residue Pesticides Analysis with Ease Paul Silcock & Jason Cole
Implement High Quality, Routine GC-MS/MS Multi-Residue Pesticides Analysis with Ease Paul Silcock & Jason Cole 1 The world leader in serving science Some discussion points for today.. Trend towards GC
More informationGenerating Automated and Efficient LC/MS Peptide Mapping Results with the Biopharmaceutical Platform Solution with UNIFI
with the Biopharmaceutical Platform Solution with UNIFI Vera B. Ivleva, Ying Qing Yu, Scott Berger, and Weibin Chen Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S The ability to
More informationThe residue then is redissolved in HPLC solvent, e.g. in 400 µ l in example calculation see below, and a aliquot is finally injected into the system.
Clean -up of Milk of Milk Products containing Aflatoxin M 1 (AFM1) Via Immunoaffinity Chromatography and Subsequent HPLC Analysis with Postcolumn Derivatization with Perbromide (for joint determination
More informationTechnicalNOTE METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES INTRODUCTION
METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES Jose M. Castro-Perez 1, Jane Kirby 1, Laurent Leclercq 2, Claude Delatour 2, Jean-Michel Mineur 2,
More informationAnalysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application
Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application Food Authors Frank David Research Institute for Chromatography Pres. Kennedypark
More informationAgilent 6500 Series Q-TOF LC/MS System
Agilent 6500 Series Q-TOF LC/MS System Quick Start Guide What s New in Data Acquisition 3 Where to find information 7 Getting Started 9 Step 1. Start the Data Acquisition software 10 Step 2. Prepare the
More informationChanging the Game in LC-MS/MS. New Tools for Unprecedented Performance in Residue Analysis. Dr. Thomas Glauner
Changing the Game in LC-MS/MS New Tools for Unprecedented Performance in Residue Analysis Dr. Thomas Glauner EMEA LC/MS Food Segment Scientist Agilent Technologies Waldbronn The New 6490 Triple Quadrupole
More informationBÜCHI Labortechnik AG
BÜCHI Labortechnik AG Purification BUCHI Chromatography History 1977: BUCHI Fraction Collector 1978: 1 st preparative chromatography system 2008: automated chromatography system 2014: automated high performance
More informationMetabolomics in Systems Biology
Metabolomics in Systems Biology Basil J. Nikolau W.M. Keck Metabolomics Research Laboratory Iowa State University February 7, 2008 Outline What is metabolomics? Why is metabolomics important? What are
More informationHigh-Throughput, Simplified Protein Precipitation: A Novel 96-Well Format for Precipitation and Filtration EAS 2003
Kimvan Tran 1, Diane M. Diehl 1, Tim Schultz 2, Wilf Wixwat 2, Brian P. Murphy 1, Jeff Kane 2, Emily Berlin 2, Kevin Seeley 2, Ziling Lu 1 1 Waters Corporation, Milford, MA 2 Pall Life Sciences, Ann Arbor,
More informationPuzzled About LCMS? Sample. Sensitivity in Mass Spec Analysis. Prep. Adapting LC-UV. Optimizing and Maintaining Your Mass Spec LCMS
Puzzled About LCMS? Sensitivity in Mass Spec Analysis Sample Prep Adapting LC-UV Optimizing and Maintaining Your Mass Spec To LCMS Paul Altiero Alex Ucci Applications Phone Support Chemistry and Supplies
More informationSCION GC Series. The Gas Chromatographers Choice for Separations
SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Bruker s Scion Instrument s long tradition long of tradition innovation of innovation and product and reliability
More informationThe LC-MS/MS Workhorse. SCIEX 4500 Series Mass Spectrometers
The LC-MS/MS Workhorse SCIEX 4500 Series Mass Spectrometers The SCIEX 4500 Series takes the legendary API 4000 platform and intelligently re-engineers it to set a new benchmark for reliable quantitation.
More informationPreparative SFC QDa Applications in Natural Products Research Judith M. Rollinger Ulrike Grienke
Preparative SFC QDa Applications in Natural Products Research Judith M. Rollinger Ulrike Grienke ÖSTERREICHISCHES TECHNOLOGIE SYMPOSIUM 2016, October 25, 2016 Pharmacognosy Group, UNIVIE Plants, fungi,
More informationDisulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column
Disulfide Linkage Analysis of IgG1 using an Agilent 126 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column Application Note Biotherapeutics & Biosimilars Author M. Sundaram
More informationChem 321 Lecture 23 - Liquid Chromatography 11/19/13
Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Student Learning Objectives High Performance Liquid Chromatography With the advent of relatively inexpensive and reliable pumps, the development of
More informationAgilent 6400 Series. System. Concepts Guide. The Big Picture. Agilent Technologies
Agilent 6400 Series Triple Quad LC/MS System Concepts Guide The Big Picture Agilent Technologies Notices Agilent Technologies, Inc. 2010 No part of this manual may be reproduced in any form or by any means
More informationLC/MS/MS Solutions for Biomarker Discovery QSTAR. Elite Hybrid LC/MS/MS System. More performance, more reliability, more answers
LC/MS/MS Solutions for Biomarker Discovery QSTAR Elite Hybrid LC/MS/MS System More performance, more reliability, more answers More is better and the QSTAR Elite LC/MS/MS system has more to offer. More
More informationPESTICIDE SCREENING APPLICATION SOLUTION
PESTICIDE SCREENING APPLICATION SOLUTION The most innovative pesticide screening solution ever built for food, beverage, and water testing. As the global food trade continues to expand, so does the challenge
More informationValidated Stability-indicating assay method for determination of Ilaprazole in bulk drug and tablets by high performance liquid chromatography
Validated Stability-indicating assay method for determination of Ilaprazole in bulk drug and tablets by high performance liquid chromatography Pradeep G. Shelke a*, Anil V. Chandewar a, Anil P. Dewani
More informationElectronic Supplementary Information. Synthesis and crystal structure of a rare square-planar Co (II) complex of a hydroxyamidinate ligand.
Electronic Supplementary Information Synthesis and crystal structure of a rare square-planar Co (II) complex of a hydroxyamidinate ligand. Mihaela Cibian, a Sofia Derossi, a and Garry S. Hanan* a Département
More informationOverview of preparative HPLC. Analytical Technologies Limited
Overview of preparative HPLC Analytical Technologies Limited 1.Review of liquid phase chromatography origin develop prospect Mobile phase Tewett 1903 Stationa ry phase origin Separation result Liquid-stationary
More informationValidation Report 16
EURL for Cereals and Feeding stuff National Food Institute Technical University of Denmark Validation Report 16 Determination of pesticide residues in wheat, barley and rice by GC-MS/MS and LC-MS/MS (QuEChERS
More informationN-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF
N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF Application Note Authors Xianming Liu, Wei Zhang, Yi Du, Sheng Yin, Hong Que, and Weichang Zhou WuXi AppTec iopharmaceuticals
More informationThe clearly better choice
Agilent Accurate-Mass 6500 Series Q-TOF and 600 Series TOF LC/MS Systems The clearly better choice in TOF and Q-TOF LC/MS Our measure is your success. products applications soft ware services The unmatched
More informationMonitoring for Microcystins in Raw Water Supply Reservoirs Using the Agilent 6410 Triple Quadrupole LC/MS
Monitoring for Microcystins in Raw Water Supply Reservoirs Using the Agilent 641 Triple Quadrupole LC/MS Application Note Environmental Author Toni Hall Wessex Water Bath, UK Abstract A method for the
More informationRecent Advancements in Analytical Methods of Drug Detection
Recent Advancements in Analytical Methods of Drug Detection Mario Thevis Gene and Cell Doping Symposium Beijing, June 5/6, 2013 www.wada-ama.org Challenges at the time accepted by anti-doping authorities
More informationHIGHER IMPACT LC/MS DIAGNOSTICS AGILENT IN VITRO DIAGNOSTIC MEDICAL DEVICES FOR CLINICAL LABORATORIES
HIGHER IMPACT LC/MS DIAGNOSTICS AGILENT IN VITRO DIAGNOSTIC MEDICAL DEVICES FOR CLINICAL LABORATORIES 2 AGILENT IN THE CLINIC THE PARTNER YOU TRUST Agilent s commitment to quality When a diagnosis is on
More informationFast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides
Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides High-pH-stable, superficially porous particle columns for LC/UV and LC/MS Application Note Biologics and Biosimilars Authors
More informationPharmaceutical LC/MS Solutions from Agilent Technologies
Pharmaceutical LC/MS Solutions from Agilent Technologies Application Compendium Overview LC/MS plays a key role in the drug discovery and drug development process. Since the introduction of electrospray
More informationUsing HRAM Survey Analysis Combined with Rapid MS2 Data to Develop a Fragmentation Based Detection Workflow for Structure ID Acquisition
Using HRAM Survey Analysis Combined with Rapid MS Data to Develop a Fragmentation Based Detection Workflow for Structure ID Acquisition Tim Stratton, 1 Caroline Ding, 1 Christoph Henrich, Hans Grensemann,
More information