1 Computed Tomography: Optimization of acquisition protocols & Justification of clinical referrals Koos Geleijns, medical physicist

2 CT delivers excellent 3D image quality

3 CT delivers excellent 3D image quality

4 CT delivers excellent 3D image quality

5 CT delivers excellent 4D image quality

6 CT delivers excellent image quality, but...

7 Optimization of acquisition protocols Tube current optimization Tube current modulation Selective organ shielding Unexpected circumstances

8 Optimization Aspects of optimization not discussed here: Tube voltage (lower tube voltage in iodine contrast enhanced scans) Rotation time Pitch Reconstruction filter Slice thickness Overbeaming (penumbra effect): beam as wide as possible Overranging (extra rotatoins before and after the planned range): beam as small as possible Image reading, special postprocessing, image reformats, quality of the workstation

9 Tube current optimization (fixed ma) Children undergoing CT scans are routinely getting adultsize doses of x-rays (2000)

10 Tube current optimization (fixed ma) Depending on the age/size Adults vs children; compensation for the x-ray attenuation Normal, obese vs small sized adults; compensation for the composition (soft tissue vs fat tissue) Depending on the clinical question For example CT brain: orbita very low dose, ventricles low dose, brain (white and grey) higher dose. How? In cooperation with manufacturer and radiologists (consensus); scientific research; trial and error; legislation (European Guidelines).

11 Low dose acquisitions in coronary CT angiography Joemai, Geleijns, Veldkamp, European Radiology

12 Low dose acquisitions in coronary CT angiography 100 % 50 % 25 % 12.5 % Joemai, Geleijns, Veldkamp, European Radiology

13 Low dose acquisitions in CT brain 100% 50% 25% 12.5%

14 Low dose acquisitions in coronary CT angiography Clinical image quality, LUMC sub msv volumetric heart scan (dr. Kroft) Cx > 50 % stenosis 72 yom, 82 kg, 178 cm Single beat, 128 mm range, 100 kv, 320 ma, BMI 26 (overweight), 55 bpm 80 ml contrast, 0.35 s rot. time, 256 x 0.5 mm

15 Diagnostic reference level (European Guidelines) The diagnostic reference level is in The Netherlands supplemented by an achievable (optimized) dose level This is done to stimulate optimization of scan protocols Diagnostic reference levels and achievable dose levels for computed tomography, volume computed tomography dose index and dose -length product (average adult). Diagnostic reference level Achievable dose level CT abdomen/pelvis, acute abdomen CTDIvol100: 15 mgy DLP: 700 mgy cm CTDIvol100: 8 mgy DLP: 400 mgy cm CT chest angiography, lung embolism CTDIvol100: 10 mgy DLP: 350 mgy cm CTDIvol100: 6 mgy DLP: 200 mgy cm Pediatric CT head; trauma (age 0 (neonate), 1, 5, 10 year)

16 Helical scan, tube current modulation To assist the optimization of CT acquisitions, scanner manufacturers have developed tube current modulation (TCM) techniques Currently, these can be divided into three main categories: angular modulation in the xy-plane, longitudinal modulation along the z-axis; and volumetric (combined) modulation in the xyz-volume TCM should provide more uniform image quality with reduced interpatient and intrapatient variability. In addition, it has potential for dose reduction

17 Helical scan, tube current modulation Longitudinal modulation (along the z-axis)

18 Tube current modulation works good! Noise (HU) TCM Noise versus SD Level P75 Max Median Min P25 130mA DLP L 32SD L 30SD DLP DLP L 28SD L 26SD DLP DLP L 24SD C 32SD* C 30SD* C 28SD C 26SD C 24SD DLP DLP DLP DLP DLP DLP SD level Box-and-whisker plots of the noise variations along the z-axis in the Rando Alderson phantom abdomen for scans using fixed tube current and with longitudinal and combined tube current modulation with variable SD.

19 Review of 43 papers, 64 row Coronary CT Angiography 15 msv 12 msv 13.5 msv 4 msv < 1 msv??

20 In-plane bismuth shielding: eye lens, thyroid Eye shield in place. Hopper et al. AJNR 2001Clin Thyroid shield in position. McLaughlin et al. Clin. Radiol. 2004

21 In-plane bismuth shielding: breasts (pediatric CT)

22 In-plane bismuth shielding: dose and image quality Absorbed dose to superficial organs and tissues was reported to be reduced by applying bismuth shielding on the skin close to superficial organs and tissues. Deterioration of image quality by beam hardening artefacts but they did not to extend to the tissues of interest for diagnosis, No obvious change in image quality resulted from shielding The artefact produced by shielding was reported as slightly distracting but did not interfere with interpretation of the images.

23 In-plane bismuth shielding: frontal view

24 In-plane bismuth shielding: frontal and lateral view

25 Quantitative assessment of dose and image quality (2008) Sabrina Vollmar and Willi Kalender, Eur Radiol (2008) 18: Bismuth shielding may compromise image quality, increase noise level and introduce streak artifacts. We do not recommend the use of bismuth shielding because of their impact on image noise.

26 Quantitative assessment of dose and image quality (2006) The observed reduction of organ dose and total energy imparted could be achieved more efficiently by a reduction of tube current. The application of in-plane selective shielding is therefore discouraged.

27 Optimization Be alert for unexpected circumstances.

28 Badly performed AND repeated brain CT perfusion studies cause hair loss in three patients (MDCT) 2003 Radiation-induced temporary hair loss after cerebral perfusion studies with multi-detector CT, Imanishi et al., ECR patients, each one or more MDCT perfusion studies, some also DSA The combination of two or more MDCT perfusion studies, at a rather high tube current, and angiography resulted in hair loss in three patients

29 Badly performed brain CT perfusion studies in many patients 2009

30 Badly performed AND repeated CT brain study The toddler received 151 scans in a single imaging session Civil trial in a lawsuit filed by the parents Knickerbocker tried to start the examination, but the machine did one rotation before it stopped.

31 Image artifacts in a brain scan

32 Justification of clinical referrals Currently: mainly based on consensus Our aim: evidence-based CT referral

33 Need for justification of clinical referrals! In 2006, Americans were exposed to more than seven times as much ionizing radiation from medical procedures as was the case in the early 1980s. ± 3 msv/y

34 Need for justification of clinical referrals!

35 Need for justification of clinical referrals! Target: referring physicians and radiologists Currently practice in justification and referral: Individual decisions (referring physician) Local practices National guidelines European referral guidelines (and quality criteria) American appropriateness criteria Drawbacks in current referral practices: Mainly based on consensus, few evidence based practices No scientific consideration of all relevant aspects

36 Our aim: evidence-based CT referral A patient presents with signs and symptoms a sign is an objective symptom of a disease; a symptom is a subjective sign of disease Medical history & physical examination, (anamnesis) Clinical decision 1 Clinical examination Lab test, organ function Clinical decision 2 Wait and see no treatment Clinical decisions may be supported by objective clinical decision rules Medical imaging Clinical decision 3 Treatment

37 Our aim: evidence-based CT referral X-ray (plain radiograph, mammography) Computed tomography (CT) Ultrasound (US) and Doppler ultrasound Magnetic resonance (MR) imaging Nuclear medicine (SPECT, PET)

38 Radiation protection: conscious utilization of CT CT is accurate, rapid, reproducible, and convenient But. CT is associated with radiation exposure Our aim: evidence-based CT referral, by: Assessment of outcomes: disability weights (WHO) Evidence-based: systematic review, patient studies Model-based stepwise guideline development All relevant aspects can now be taken into account: Medical History and Physical Examination ; Lab Tests ; Possible Imaging Modalities ; consequences of false negatives and false positives (morbidity and mortality); adverse effects, including radiation induced effects (morbidity and mortality)

39 Bayesian network for probability of CAD based on the most important test results prior to imaging or intervention Decision analysis is based on the premise that humans are reasonably capable of framing a decision problem, listing possible decision options, determining relevant factors, and quantifying uncertainty and preferences, but are rather weak in combining this information into a rational decision. A Bayesian network, or belief network, shows conditional probability and causality relationships between variables.

40 Radiation protection: conscious utilization of CT Acute abdominal pain

41 Radiation protection: conscious utilization of CT Symptom: Appendicitis First: Suspected appendicitis scoring Next: Assess the optimal imaging policy

42 Radiology: ultrasound and CT of the abdomen CT (upper) and US (lower) of acute appendicitis in a 25-year-old man [Keyzer-ea Rad 2005]

43 Radiation protection: conscious utilization of CT treatment, follow-up, additional diagnosis

44 Radiation protection: conscious utilization of CT The biggest challenge in radiation protection (in diagnostic radiology) nowadays is to extend the scientific basis for conscious utilization of x-ray imaging, particularly for CT Cost effectiveness may be considered Proper justification is relevant for diagnosis and follow up Should CT be performed or not; when should CT be performed? Benefits of diagnosis and treatment Risks of diagnosis and treatment (including radiation exposure)

GENESIS Edition Transforming CT

GENESIS Edition Transforming CT GENESIS Edition Transforming CT 2 Transforming clinical confidence Transforming patient experience Transforming your workspace GENESIS Edition Transforming CT Brought to you by the leaders in area detector

More information

GENESIS Edition. Transforming CT

GENESIS Edition. Transforming CT GENESIS Edition Transforming CT Transforming clinical confidence Transforming patient experience Transforming your workspace GENESIS Edition Transforming CT Brought to you by the leaders in area detector

More information

Handzettel 1. CARE Right Computed Tomography, committed to the right dose Ivo Driesser Austin, July 22 nd 2014

Handzettel 1. CARE Right Computed Tomography, committed to the right dose Ivo Driesser Austin, July 22 nd 2014 CARE Right Computed Tomography, committed to the right dose Ivo Driesser Austin, July 22 nd 2014 Answers for life. Answers for life. Is 1 msv the right dose for every patient? Female, 54 Abdominal CT Male,

More information

Leading Canadian Pediatric Hospital Continues to Drive Dose Down for Sick Children

Leading Canadian Pediatric Hospital Continues to Drive Dose Down for Sick Children Leading Canadian Pediatric Hospital Continues to Drive Dose Down for Sick Children At Canada s premier pediatric hospital also recognized as one of the world s foremost pediatric healthcare institutions

More information

2017 ACR Computed Tomography Quality Control Manual FAQS

2017 ACR Computed Tomography Quality Control Manual FAQS Updated 11-15-2017 2017 ACR Computed Tomography Quality Control Manual FAQS Q. The updated 2017 ACR Computed Tomography Quality Control Manual has been released. (Visit www.acr.org/education/education-catalog.)

More information

GE Healthcare. Introducing Veo * on Discovery * CT750 HD Great care by design

GE Healthcare. Introducing Veo * on Discovery * CT750 HD Great care by design GE Healthcare Introducing Veo * on Discovery * CT750 HD Great care by design Radiologists and clinicians demand the ultimate levels of clarity and detail in their images. In making an informed and confident

More information

New Material for Shields in X-Ray Diagnostic Procedures

New Material for Shields in X-Ray Diagnostic Procedures Journal of US-China Medical Science 1 (016) 06-1 doi: 10.1765/1548-6648/016.04.004 D DAVID PUBLISHING Maria Anna Staniszewska 1, Marian Zaborski, Edyta Kusiak and Piotr Pankowski 1. Department of Medical

More information

A New Standard for the Future

A New Standard for the Future technology history For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

Optimization of Low-Dose CT Protocol in Pediatric Nuclear Medicine Imaging

Optimization of Low-Dose CT Protocol in Pediatric Nuclear Medicine Imaging Optimization of Low-Dose CT Protocol in Pediatric Nuclear Medicine Imaging Hanna Piwowarska-Bilska 1, Leszek J. Hahn 2, Bozena Birkenfeld 1, Katarzyna Cichon-Bankowska 1, Maria H. Listewnik 1, and Piotr

More information

IAEA-TECDOC Dose Reduction in CT while Maintaining Diagnostic Confidence: A Feasibility/Demonstration Study

IAEA-TECDOC Dose Reduction in CT while Maintaining Diagnostic Confidence: A Feasibility/Demonstration Study IAEA-TECDOC-1621 Dose Reduction in CT while Maintaining Diagnostic Confidence: A Feasibility/Demonstration Study IAEA-TECDOC-1621 Dose Reduction in CT while Maintaining Diagnostic Confidence: A Feasibility/Demonstration

More information

GE Healthcare. Optima * CT660 Second Edition

GE Healthcare. Optima * CT660 Second Edition GE Healthcare Optima * CT660 Second Edition Your vision of quality care made real The Optima CT660 system helps you deliver competent, personalized care that helps fulfill your mission and please your

More information

PERFORM Operating Document. Radiation Dose Estimates, Calculations and Tracking Methodology

PERFORM Operating Document. Radiation Dose Estimates, Calculations and Tracking Methodology PERFORM Operating Document Radiation Dose Estimates, Calculations and Tracking Methodology PC-POD-IM-003-v03 Revision History Version Reason for Revision Date 01 New POD Jul/27/2015 02 Update section 2.2

More information

www.aquilionprime.com WARNING: Any reference to x-ray exposure, intravenous contrast dosage, and other medication is intended as a reference guideline only. The guidelines in this document do not substitute

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May Dose Reference Levels

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May Dose Reference Levels 2033-7 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dose Reference Levels Peter Homolka EFOMP Training Course on Medical Physics in Diagnostic

More information

Motiva: a TV-based platform for remote patient management

Motiva: a TV-based platform for remote patient management Technology news New products Motiva: a TV-based platform for remote patient management a simplified user interface, a secure broadband connection and a standard set-top box. A nurse care manager, using

More information

Jonathon A. Nye, Ph.D Emory University School of Medicine Atlanta, GA

Jonathon A. Nye, Ph.D Emory University School of Medicine Atlanta, GA Artifacts related to CT Jonathon A. Nye, Ph.D Emory University School of Medicine Atlanta, GA Acknowledgements: Fabio Esteves MD 1, James Galt PhD 1, Perry Sprawls PhD 1, Robert Eisner PhD 2, James Hamill

More information

3/6/2017 TOMOSYNTHESIS GUIDED BREAST BIOPSY LEARNING OBJECTIVES DISCLOSURES

3/6/2017 TOMOSYNTHESIS GUIDED BREAST BIOPSY LEARNING OBJECTIVES DISCLOSURES TOMOSYNTHESIS GUIDED BREAST BIOPSY Amy Kerger, D.O. Assistant Professor The Ohio State Wexner Medical Center Stefanie Spielman Comprehensive Breast Center LEARNING OBJECTIVES What is the difference between

More information

Progress in X-Ray & MR

Progress in X-Ray & MR Progress in X-Ray & MR Michiel Manuel Analyst Meeting June 15 th, 2005 X-Ray & MR: Agenda Introduction General X-Ray Cardio/Vascular X-Ray Magnetic Resonance China growth opportunity Conclusion 2 X-Ray

More information

Low Dose, Lightning Speed, Latest Applications

Low Dose, Lightning Speed, Latest Applications RXL EDITION Low Dose, Lightning Speed, Latest Applications The New Standard for Radiology Toshiba Medical Systems is committed to the development of new technologies to minimize radiation dose while maintaining

More information

Clinical Applications. ImagingRite. Neuro Intervention. 1 ImagingRite

Clinical Applications. ImagingRite. Neuro Intervention. 1 ImagingRite Clinical Applications ImagingRite Neuro Intervention 1 ImagingRite ImagingRite, a comprehensive suite of imaging tools offered with Infinix -i angiographic systems, was designed to assist clinicians in

More information

IMV 2006 CT Market Summary Report Featuring results from 2,565 hospital and non-hospital CT imaging sites in the United States

IMV 2006 CT Market Summary Report Featuring results from 2,565 hospital and non-hospital CT imaging sites in the United States Medical Information Division 1400 East Touhy Ave., Suite 250 Des Plaines, Illinois 60018 Phone: 847-297-1404 Fax: 847-297-5010 www.imvinfo.com IMV 2006 CT Market Summary Report Featuring results from 2,565

More information

Radiography Curriculum Analysis

Radiography Curriculum Analysis Program Number Program Name Date / /20 Radiography Curriculum Analysis DIRECTIONS: Determine the course(s) in which each of the following content area is covered and enter the course number(s) and/or title(s).

More information

R/F. Experiences Using SONIALVISION safire and the Utility of Tomosynthesis. 1. Introduction. 2. Basics of Tomosynthesis.

R/F. Experiences Using SONIALVISION safire and the Utility of Tomosynthesis. 1. Introduction. 2. Basics of Tomosynthesis. R/F Experiences Using SONIALVISION safire and the Utility of Tomosynthesis Radiology Division, Dokkyo Medical University Koshigaya Hospital Masahiro Nakajima Mr. Masahiro Nakajima 1. Introduction The hospital

More information

Joint Commission Medical Imaging Requirements. Laura Gruber, MBA, RT(R), RDMS, RVT Sr. Director Department of Medical Imaging SPR Annual Meeting

Joint Commission Medical Imaging Requirements. Laura Gruber, MBA, RT(R), RDMS, RVT Sr. Director Department of Medical Imaging SPR Annual Meeting Joint Commission Medical Imaging Requirements Laura Gruber, MBA, RT(R), RDMS, RVT Sr. Director Department of Medical Imaging SPR Annual Meeting My team of professionals 2 Medical Physicist s 3 FTE s (Board

More information

The World s First Adaptive Scanner

The World s First Adaptive Scanner The World s First Adaptive Scanner SOMATOM Definition AS Excel Edition Answers for life. SOMATOM Definition AS Excel Edition 2 3 Siemens CT Vision Siemens CT Vision The justification for the existence

More information

Gastrointestinal Imaging Original Research

Gastrointestinal Imaging Original Research Gastrointestinal Imaging Original Research Mayer et al. Radiation Dose Savings in CT Gastrointestinal Imaging Original Research Caroline Mayer 1 Mathias Meyer 1 Christian Fink 1 Bernhard Schmidt 2 Martin

More information

Computed Tomography. Printed in Japan

Computed Tomography. Printed in Japan Computed Tomography WARNING: Any reference to x-ray exposure, intravenous contrast dosage, and other medication is intended as a reference guideline only. The guidelines in this document do not substitute

More information

MEDICAL PHYSICS (MED PHYS)

MEDICAL PHYSICS (MED PHYS) Medical Physics (MED PHYS) 1 MEDICAL PHYSICS (MED PHYS) MED PHYS/PHYSICS 265 INTRODUCTION TO MEDICAL PHYSICS Primarily for premeds and other students in the medical and biological sciences. Applications

More information

CT QA SOLUTIONS. Ensure Accurate Screening, Diagnosis and Monitoring

CT QA SOLUTIONS. Ensure Accurate Screening, Diagnosis and Monitoring CT QA SOLUTIONS Ensure Accurate Screening, Diagnosis and Monitoring ACCREDITATION ADVANCED CT IMAGE QUALITY QA AUTOMATIC EXPOSURE CONTROL CT CHARACTERIZATION MULTI-ENERGY CT CT PERFUSION QA COMPLETE CT

More information

Installation of a regional CT dose registry with a dose management software

Installation of a regional CT dose registry with a dose management software Installation of a regional CT dose registry wi a dose management software Poster No.: ESI-0036 Congress: EuroSafe Imaging 2016 Type: EuroSafe Imaging Auors: S. Schindera, A. Euler, A. Parakh ; Riehen/CH,

More information

Toward Large-Scale Process Control to Enable Consistent CT Radiation Dose Optimization

Toward Large-Scale Process Control to Enable Consistent CT Radiation Dose Optimization Pediatric Imaging Review Larson et al. CT Radiation Dose Optimization Pediatric Imaging Review FOCUS ON: David B. Larson 1 Keith J. Strauss 2 Daniel J. Podberesky 3 Larson DB, Strauss KJ, Podberesky DJ

More information

Experience With Patient Dosimetry and Quality Control Online for Diagnostic and Interventional Radiology Using DICOM Services

Experience With Patient Dosimetry and Quality Control Online for Diagnostic and Interventional Radiology Using DICOM Services Medical Physics and Informatics Review Vano et al. Online Quality Control for Diagnostic and Interventional Radiology Medical Physics and Informatics Review FOCUS ON: Eliseo Vano 1,2 Jose I. Ten 2,3 Jose

More information

Aquilion ONE ViSION EDITION

Aquilion ONE ViSION EDITION Aquilion ONE ViSION EDITION Table of Contents ViSIONPROTECT... 12 ViSIONDESIGN... 16 ViSIONCLINIC...20 Emergency ONE... 22 Pediatric ONE... 22 Cardio ONE... 24 Neuro ONE... 26 Oncology ONE... 27 Interventional

More information

New PET/CT from Siemens helps more patients benefit from premium technologies

New PET/CT from Siemens helps more patients benefit from premium technologies Press Healthcare Erlangen, October 9, 2015 EANM 2015, Congress Center Hamburg (CCH) New PET/CT from Siemens helps more patients benefit from premium technologies Versatile new PET/CT system addresses a

More information

Radiation Physics and Safety Considerations in the EP Lab

Radiation Physics and Safety Considerations in the EP Lab Cardiorhythm 2009 Radiation Physics and Safety Considerations in the EP Lab Joel A. Kirsh, MD, FRCPC, CCDS Cardiology and Critical Care, Hospital for Sick Children Associate Professor of Pediatrics, University

More information

Standardizing CT-Guided Biopsy Procedures: Patient Dose and Image Noise

Standardizing CT-Guided Biopsy Procedures: Patient Dose and Image Noise Vascular and Interventional Radiology Original Research Tam et al. Standardization of Image Acquisition Protocol for CT-Guided Biopsy Vascular and Interventional Radiology Original Research FOCUS ON: Alda

More information

Patient doses in lower limb angiography in a recently installed flat detector x-ray system

Patient doses in lower limb angiography in a recently installed flat detector x-ray system Patient doses in lower limb angiography in a recently installed flat detector x-ray system Poster No.: C-0939 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper V. Tsapaki, D. Papavdis,

More information

GE Healthcare. Optima NM/CT 640

GE Healthcare. Optima NM/CT 640 GE Healthcare Optima NM/CT 640 Great and lasting performance in a SPECT/CT system? Precisely. We invented hybrid SPECT/CT. And now we re perfecting it. Introducing the Optima* NM/CT 640 from GE Healthcare

More information

Types of Radiation Units

Types of Radiation Units RADIATION UNITS Types of Radiation Units Exposure in air - amount of ionization produced by or radiation in air. Unit - (R) or / (c/kg) Absorbed Dose energy transfer of / (unit of energy) of any. Used

More information

Update on Radiation Dose Analytics: Meaningful Use or Useless Meaning?

Update on Radiation Dose Analytics: Meaningful Use or Useless Meaning? Update on Radiation Dose Analytics: Meaningful Use or Useless Meaning? SPR 2015 Donald P. Frush, MD, FACR FAAP No Disclosures Duke University Medical Center 1 Speakers Keith Strauss, MS Steven Don, MD

More information

Extending quality care to more people. Brivo CT315

Extending quality care to more people. Brivo CT315 GE Healthcare Extending quality care to more people. Brivo CT315 Big impact. Improving the quality of healthcare around the world is a goal we all share. But this can be a daunting challenge, with patients

More information

2 / 1. SUMMARY public, should be well documented, and should be reviewed on a regular basis to determine whether it continues to meet the operational

2 / 1. SUMMARY public, should be well documented, and should be reviewed on a regular basis to determine whether it continues to meet the operational 1. Summary This Report is concerned with the protection of individuals who may be exposed to radiation emitted by x-ray equipment and both sealed and unsealed radioactive sources in the practice of veterinary

More information

Dual-Energy MDCT in Hypervascular Liver Tumors: Effect of Body Size on Selection of the Optimal Monochromatic Energy Level

Dual-Energy MDCT in Hypervascular Liver Tumors: Effect of Body Size on Selection of the Optimal Monochromatic Energy Level Medical Physics and Informatics Original Research Mileto et al. Effect of Body Size on Dual-Energy MDCT of Hypervascular Liver Tumors Medical Physics and Informatics Original Research Achille Mileto 1

More information

BIOMEDICAL SIGNAL AND IMAGE PROCESSING

BIOMEDICAL SIGNAL AND IMAGE PROCESSING BIOMEDICAL SIGNAL AND IMAGE PROCESSING EE 5390-001 SYLLABUS Instructor: Wei Qian, Ph.D. Professor of Electrical and Computer Engineering Medical Signal and Image Computerized Processing Scheme for Medical

More information

Introducing a new take on efficient workflow: Deeply integrated clinical applications

Introducing a new take on efficient workflow: Deeply integrated clinical applications Clinical Applications Introducing a new take on efficient workflow: Deeply integrated clinical applications The EXPERIENCE of TRUE CLINICAL APPLICATIONS While developments in image acquisition technology

More information

One exam. One workflow. Immeasurable versatility.

One exam. One workflow. Immeasurable versatility. One exam. One workflow. Immeasurable versatility. Symbia. There s only one like it. The first system to integrate state-of-the-art SPECT with diagnostic multislice-ct, Symbia combines superb image clarity

More information

CT of the Abdomen with Reduced Tube Voltage in Adults: A Practical Approach 1

CT of the Abdomen with Reduced Tube Voltage in Adults: A Practical Approach 1 GASTROINTESTINAL IMAGING 1922 CT of the Abdomen with Reduced Tube Voltage in Adults: A Practical Approach 1 Adeel R. Seyal, MD Atilla Arslanoglu, MD Samir F. Abboud, MD Azize Sahin, MD Jeanne M. Horowitz,

More information

Medical Imaging Technologies in Canada, 2006 Methodological Notes

Medical Imaging Technologies in Canada, 2006 Methodological Notes Medical Imaging Technologies in Canada, 2006 Methodological Notes Introduction The Canadian Institute for Health Information (CIHI) aims to provide accurate and timely data and information to help shape

More information

Department of Radio Diagnosis

Department of Radio Diagnosis Department of Radio Diagnosis S.No. Name of Equipment Specification Qty required Suggested Manufacturers Name Approx Unit cost (Rs.) Usage 1 Digital substraction angiography (D.S.A.) A single plane, ceiling

More information

Image Gently: Ten Steps You Can Take to Optimize Image Quality and Lower CT Dose for Pediatric Patients

Image Gently: Ten Steps You Can Take to Optimize Image Quality and Lower CT Dose for Pediatric Patients Medical Physics and Informatics Commentary Strauss et al. Image Quality and CT Dose for Pediatric Patients Medical Physics and Informatics Commentary FOCUS ON: Keith J. Strauss 1 Marilyn J. Goske 2 Sue

More information

Making the difference with Live Image Guidance Dose management in Allura systems

Making the difference with Live Image Guidance Dose management in Allura systems DoseWise Allura Making the difference with Live Image Guidance Dose management in Allura systems The importance of radiation dose management The harmful effects of X-rays on human tissues had already become

More information

Healthcare. Philips DoseWise Portal. Radiation exposure management software

Healthcare. Philips DoseWise Portal. Radiation exposure management software Healthcare Philips DoseWise Portal Radiation exposure management software The first step in building an effective radiation dose management strategy is collecting complete and accurate data so you can

More information

Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy. Jean-Pierre Bissonnette, PhD, MCCPM

Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy. Jean-Pierre Bissonnette, PhD, MCCPM Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy Jean-Pierre Bissonnette, PhD, MCCPM Disclosure Work supported, in part, by Elekta Oncology Systems Commercial Interest in Penta-Guide

More information

Quality Control Concerns for New Generations of CT, MRI, & PET

Quality Control Concerns for New Generations of CT, MRI, & PET Quality Control Concerns for New Generations of CT, MRI, & PET David W. Jordan, Ph.D. Sr. Medical Physicist, UH Case Medical Center Clinical Assistant Professor, CWRU Disclosures No financial disclosures

More information

Gastrointestinal Imaging Original Research

Gastrointestinal Imaging Original Research Gastrointestinal Imaging Original Research Schindera et al. Abdominal CT for Obese Patients Gastrointestinal Imaging Original Research Sebastian T. Schindera 1,2 Devang Odedra 1 Diego Mercer 1 Seng Thipphavong

More information

Investigation of lung nodule detectability in low-dose 320-slice computed tomography

Investigation of lung nodule detectability in low-dose 320-slice computed tomography Investigation of lung nodule detectability in low-dose 320-slice computed tomography J. D. Silverman Institute of Biomaterials and Biomedical Engineering, University of Toronto, Toronto, Ontario M5G 2M9,

More information

8/2/2017. Key attributes of scientific excellence: rigor, innovation, and relevance. Medical Physics 3.0. Key Attributes of Scientific Excellence

8/2/2017. Key attributes of scientific excellence: rigor, innovation, and relevance. Medical Physics 3.0. Key Attributes of Scientific Excellence Medical Physics 3.0 in Design Key attributes of scientific excellence: rigor, innovation, and relevance Maryellen Giger, Ph.D. A. N. Pritzker Professor of Radiology / Medical Physics The University of

More information

Biomedical Imaging Modalities

Biomedical Imaging Modalities 1 Biomedical Imaging Modalities The introduction of advanced imaging techniques has improved significantly the quality of medical care available to patients. Noninvasive imaging modalities allow a physician

More information

Disclaimer and Notices.

Disclaimer and Notices. Fluoroscopy Dose Reduction and Clinical Management Fluoroscopy Trajectories (Curves), Global Dose Monitoring and Tracking of Patient Dose. Pei Jan Paul Lin, Ph.D. Division of Radiation Physics and Biology

More information

SPECIFICATIONS FOR TENDER # SUPPLY OF ONE MULTI-SLICE CT SCANNER WESTERN MEMORIAL REGIONAL HOSPITAL AND SIR THOMAS RODDICK HOSPITAL

SPECIFICATIONS FOR TENDER # SUPPLY OF ONE MULTI-SLICE CT SCANNER WESTERN MEMORIAL REGIONAL HOSPITAL AND SIR THOMAS RODDICK HOSPITAL SPECIFICATIONS FOR TENDER #0171-0754 SUPPLY OF ONE MULTI-SLICE CT SCANNER WESTERN MEMORIAL REGIONAL HOSPITAL AND SUPPLY OF ONE MULTI-SLICE CT SCANNER SIR THOMAS RODDICK HOSPITAL CLOSING DATE: 14 September

More information

Original article Paediatrics Today 2016;12(1):75-80 DOI /p

Original article Paediatrics Today 2016;12(1):75-80 DOI /p Original article Paediatrics Today 2016;12(1):75-80 DOI 10.5457/p2005-114.138 THE EFFECT OF ANTI-SCATTER GRID ON RADIATION DOSE IN CHEST RADIOGRAPHY IN CHILDREN Igor ŠABIČ 1*, Damjana KLJUČEVŠEK 1, Martin

More information

Image Quality from High-Resolution CT of the Lung: Comparison of Axial Scans and of Sections Reconstructed from Volumetric Data Acquired Using MDCT

Image Quality from High-Resolution CT of the Lung: Comparison of Axial Scans and of Sections Reconstructed from Volumetric Data Acquired Using MDCT Studler et al. High- Resolution CT and MDCT of the Lung Chest Imaging Original Research Ueli Studler 1 Thomas Gluecker Georg Bongartz Jakob Roth Wolfgang Steinbrich Studler U, Gluecker T, Bongartz G, Roth

More information

Automated Breast Volumes. Simplified.

Automated Breast Volumes. Simplified. www.siemens.com/ultrasound Automated Breast Volumes. Simplified. ACUSON S2000 Automated Breast Volume Scanner (ABVS) Answers for life. Automated Breast Volumes. Simplified. ACUSON S2000 Automated Breast

More information

FLUOROSCOPIC SAFETY AND PATIENT AND OPERATOR RADIATION DOSE REDUCTION: CREATING A FOCUSED FLUOROSCOPIC CURRICULUM FOR RADIOLOGY RESIDENTS

FLUOROSCOPIC SAFETY AND PATIENT AND OPERATOR RADIATION DOSE REDUCTION: CREATING A FOCUSED FLUOROSCOPIC CURRICULUM FOR RADIOLOGY RESIDENTS FLUOROSCOPIC SAFETY AND PATIENT AND OPERATOR RADIATION DOSE REDUCTION: CREATING A FOCUSED FLUOROSCOPIC CURRICULUM FOR RADIOLOGY RESIDENTS Rutgers Robert Wood Johnson Medical School, New Brunswick, NJ Abstract

More information

Submillisievert Chest CT With Filtered Back Projection and Iterative Reconstruction Techniques

Submillisievert Chest CT With Filtered Back Projection and Iterative Reconstruction Techniques Cardiopulmonary Imaging Original Research Padole et al. Submillisievert Chest CT Cardiopulmonary Imaging Original Research Atul Padole 1 Sarabjeet Singh Jeanne B. Ackman Carol Wu Synho Do Sarvenaz Pourjabbar

More information

Join the World of CT SOMATOM Spirit

Join the World of CT SOMATOM Spirit Join the World of CT SOMATOM Spirit Katsumi Tadokoro, MD, Radiology Department, Medical Center Hospital, Saitama, Japan The SOMATOM Spirit is a compact CT that allows us to quickly perform routine exams.

More information

Optimization of anode-filter combinations in digital mammography: An experimental study

Optimization of anode-filter combinations in digital mammography: An experimental study Optimization of anode-filter combinations in digital mammography: An experimental study Poster No.: C-2915 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology Authors: A. Peruzzo Cornetto

More information

low Dose, Lightning Speed

low Dose, Lightning Speed CXL edition low Dose, Lightning Speed Maximize the potential of today s clinical resources. In today s demanding healthcare environment, you need a CT system that satisfies all your requirements on a day-to-day

More information

RADIATION EXPOSURE IN ENDOVASCULAR PROCEDURES - AN ASSESSMENT BASED ON EXAMINATIONS PERFORMED AT SAHLGRENSKA UNIVERSITY HOSPITAL

RADIATION EXPOSURE IN ENDOVASCULAR PROCEDURES - AN ASSESSMENT BASED ON EXAMINATIONS PERFORMED AT SAHLGRENSKA UNIVERSITY HOSPITAL SAHLGRENSKA ACADEMY DEPARTMENT OF RADIATION PHYSICS RADIATION EXPOSURE IN ENDOVASCULAR PROCEDURES - AN ASSESSMENT BASED ON EXAMINATIONS PERFORMED AT SAHLGRENSKA UNIVERSITY HOSPITAL M.Sc. thesis Roham D.

More information

Understanding The Provider Registration Scorecard. Powered By

Understanding The Provider Registration Scorecard. Powered By Understanding The Provider Registration Scorecard Powered By Understanding Your Scorecard The provider registration scorecard is divided into section scores based on the information that your facility

More information

ESTIMATION OF ORGAN DOSES OF PATIENT UNDERGOING HEPATIC CHEMOEMBOLIZATION PROCEDURES

ESTIMATION OF ORGAN DOSES OF PATIENT UNDERGOING HEPATIC CHEMOEMBOLIZATION PROCEDURES X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN

More information

Optimize efficiency with a universal system.

Optimize efficiency with a universal system. wins www.siemens.com/artis-one Optimize efficiency with a universal system.. Designed around you. Answers for life. Consistently efficient. Improve your competitiveness by mastering the balancing act between

More information

Medical Physics and Informatics Original Research

Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Jaffe et al. Radiation Dose for Body CT Medical Physics and Informatics Original Research Tracy A. Jaffe 1 Terry T. Yoshizumi Greta Toncheva Colin Anderson-Evans

More information

Evaluation of efficacy of metal artefact reduction technique using contrast media in Computed Tomography

Evaluation of efficacy of metal artefact reduction technique using contrast media in Computed Tomography Journal of Physics: Conference Series PAPER OPEN ACCESS Evaluation of efficacy of metal artefact reduction technique using contrast media in Computed Tomography To cite this article: Diana Yusob et al

More information

Automatic Bone Removal Technique in Whole-Body Dual-Energy CT Angiography: Performance and Image Quality

Automatic Bone Removal Technique in Whole-Body Dual-Energy CT Angiography: Performance and Image Quality Vascular and Interventional Radiology Original Research Schulz et al. Bone Removal in CT Angiography Vascular and Interventional Radiology Original Research Boris Schulz 1 Katharina Kuehling Wolfgang Kromen

More information

F R O S T & S U L L I V A N Positron Corporation: Recipient of the 2010 North American Molecular Imaging Systems New Product Innovation Award

F R O S T & S U L L I V A N Positron Corporation: Recipient of the 2010 North American Molecular Imaging Systems New Product Innovation Award F R O S T & S U L L I V A N 2010 Positron Corporation: Recipient of the 2010 North American Molecular Imaging Systems New Product Innovation Award We accelerate growth. 2010 North American Cardiac Molecular

More information

Evaluation of 1D, 2D and 3D nodule size estimation by radiologists for spherical and non-spherical nodules through CT thoracic phantom imaging

Evaluation of 1D, 2D and 3D nodule size estimation by radiologists for spherical and non-spherical nodules through CT thoracic phantom imaging Evaluation of 1D, 2D and 3D nodule size estimation by radiologists for spherical and non-spherical nodules through CT thoracic phantom imaging Nicholas Petrick * * a, Hyun J. Grace Kim b, David Clunie

More information

E2E SBRT Thorax Phantom

E2E SBRT Thorax Phantom E2E SBRT Thorax Phantom Model 036A-CVXX-xx SBRT END-TO-END SBRT TESTING SOLUTION SCAN PLAN LOCALIZE TREAT This product is available through: JRT Associates 800-22-0 2428 Almeda Avenue Suite 36 Norfolk,

More information

Registration of 3D Ultrasound to Computed Tomography Images of the Kidney

Registration of 3D Ultrasound to Computed Tomography Images of the Kidney Registration of 3D Ultrasound to Computed Tomography Images of the Kidney Jing Xiang Supervisors: Dr. Robert Rohling Dr. Purang Abolmaesumi Electrical and Computer Engineering University of British Columbia

More information

Act anything done, being done, or to be done; the process of doing. Synonymous with procedure and clinical services.

Act anything done, being done, or to be done; the process of doing. Synonymous with procedure and clinical services. Act anything done, being done, or to be done; the process of doing. Synonymous with procedure and clinical services. Action plan A program or method that explains the actions or steps to be taken. Advanced-practice

More information

BEFORE CT AND DSA 8/3/2016 THEN AND NOW: DEVELOPMENT AND STATE OF THE ART OF NEUROENDOVASCULAR THERAPY OVERVIEW

BEFORE CT AND DSA 8/3/2016 THEN AND NOW: DEVELOPMENT AND STATE OF THE ART OF NEUROENDOVASCULAR THERAPY OVERVIEW THEN AND NOW: DEVELOPMENT AND STATE OF THE ART OF NEUROENDOVASCULAR THERAPY A PERSONAL JOURNEY Charles M. Strother, MD University of Wisconsin, Madison OVERVIEW Endovascular Therapy for Cerebrovascular

More information

Vantage Titan 1.5T MR. The Intelligent Choice.

Vantage Titan 1.5T MR. The Intelligent Choice. Vantage Titan 1.5T MR. The Intelligent Choice. Scalable. M-Powered. Performance. Vantage Titan MR is designed to satisfy the demands of today s challenging clinical environment, addressing the requirements

More information

Prevention of radiation injury from medical imaging

Prevention of radiation injury from medical imaging Prevention of radiation injury from medical imaging Erica L. Mitchell, MD, a and Patricia Furey, MD, b Portland, Ore; and Manchester, NH Introduction: Concerns over radiation safety are valid. Understanding

More information

Stay Ahead of the Curve in Radiation Reduction Standards

Stay Ahead of the Curve in Radiation Reduction Standards Radiation mandates are coming... MR Suite CT Suite I/Dx Suite Nuclear Suite Stay Ahead of the Curve in Radiation Reduction Standards New NEXO Dose: Radiation aggregation technology that measures multiple

More information

New Product Innovation Leadership

New Product Innovation Leadership Philips Healthcare A Frost & Sullivan Position Paper Nadim Michel Daher TABLE OF CONTENTS SIGNIFICANCE OF THE NEW PRODUCT INNOVATION LEADERSHIP AWARD... 3 KEY INDUSTRY CHALLENGES ADDRESSED BY PHILIPS HEALTHCARE...

More information

Publication for the Philips MRI Community Issue 40 May 2010

Publication for the Philips MRI Community Issue 40 May 2010 FieldStrength Publication for the Philips MRI Community Issue 40 May 2010 MRI scanner at Manipal Hospital gets a new lease on life New coils and upgrade to release 12 bring Manipal Hospital s aging Intera

More information

Gastrointestinal Imaging Original Research

Gastrointestinal Imaging Original Research Gastrointestinal Imaging Original Research Han et al. CT of Patients With Metal Hip Prostheses Gastrointestinal Imaging Original Research Seung Chol Han 1 Yong Eun Chung 1 Young Han Lee 1 Kwan Kyu Park

More information

Reporting Results and Incidental Findings to Research Participants

Reporting Results and Incidental Findings to Research Participants Reporting Results and Incidental Findings to Research Participants Jeffrey R Botkin, MD, MPH Professor of Pediatrics Chief, Division of Medical Ethics and Humanities Associate Vice President for Research

More information

General Radiology Improvement Database

General Radiology Improvement Database The American College of Radiology General Radiology Improvement Database ABR PQI Project Description February 22, 2013 American College of Radiology 1891 Preston White Drive Reston, VA 20191-4397 Copyright

More information

MRI Systems Market to 2018

MRI Systems Market to 2018 MRI Systems Market to 2018 Technological advancements, Increasing Number of Applications and Advent of MRI Compatible Pacemakers to Drive Future Growth GBI Research Report Guidance GBI Research Report

More information

Intuitive 3D catheter guidance. Philips EP navigator and 3D rotational scan

Intuitive 3D catheter guidance. Philips EP navigator and 3D rotational scan Intuitive 3D catheter guidance Philips EP navigator and 3D rotational scan 3D - efficiently and accurately guide For patients with symptomatic paroxysmal atrial fibrillation (AF) who do not respond adequately

More information

1 Introduction. 2 Material and methods. 2.1 Available methods. Key words: dose conversion factor, organ dose, effective dose, uncertainty

1 Introduction. 2 Material and methods. 2.1 Available methods. Key words: dose conversion factor, organ dose, effective dose, uncertainty Interpretation of measured dose data in X-ray imaging Toroi P. 1, 2, Kelaranta A. 1, Vock P. 2, Siiskonen T. 1, Tapiovaara M. 1 and Kosunen A. 1 STUK Radiation and Nuclear Safety Authority, Laippatie 4,

More information

Authors: (see complete list in Appendix A) History and comments: (see complete details in Appendix B) Preface

Authors: (see complete list in Appendix A) History and comments: (see complete details in Appendix B) Preface Diagnostic Radiology Residents Physics Curriculum AAPM Subcommittee of the Medical Physics Education of Physicians Committee May 2009 Supported by: AAPM Education Council and the Academic Council of the

More information

Molecular Imaging: Definition, Overview and Goals

Molecular Imaging: Definition, Overview and Goals This tutorial will define what is currently considered molecular imaging. It will provide history and an overview, discuss the goals and the advantages of molecular imaging. It will clarify what is and

More information