NO 2 Gas Sterilization and Decontamination: Keeping Pace, from Start to Finish

Size: px
Start display at page:

Download "NO 2 Gas Sterilization and Decontamination: Keeping Pace, from Start to Finish"

Transcription

1 NO 2 Gas Sterilization and Decontamination: Keeping Pace, from Start to Finish Evan Goulet, Ph.D., VP of Microbiology Applications Laboratory PDA Midwest: Sterility Assurance and Risk Mitigation October 6, 2016 Page 1

2 Agenda Noxilizer Company Overview About Nitrogen Dioxide (NO 2 ) Current Product Portfolio & Solutions From Start: Biodecontamination of RTF Syringe Tubs To Finish: Secondary Sterilization of Prefilled Syringes Page 2

3 Noxilizer Company Overview Providing Decontamination & Sterilization Solutions Ultra-low temperature decontamination & sterilization using Nitrogen Dioxide (NO 2 ) Terminal sterilization of medical devices Secondary sterilization of drug delivery devices administered in the sterile field Decontamination of biopharmaceutical equipment and containers Page 3

4 About Nitrogen Dioxide (NO 2 ) Chemical Decontamination & Sterilization Complementary to H 2 O 2 & Ethylene Oxide, but with important benefits: Ultra-low temperature (10 C 30 C) Works with or without vacuum for many products Excellent small lumen efficacy & complex geometries Rapid process, typically 2 hours (sterilization) Safe To Use In-House, Inline with production How Is NO 2 Delivered And Used? Decontamination equipment Sterilization equipment Consumables or On Demand Page 4

5 Applications Bio-Pharma Production Page 5

6 NO 2 Sterilization Solutions From Start Page 6

7 Bio-Pharma Production Component Preparation / Infeed Moduline TM -BioDecontamination: Rapid Biodecontamination Process SCF / RTF Tubs (syringes or other containers) Transfer into RABS or Class B / A filling area Reduces Risk Eliminates manual wipe down: spray and pray Eliminates risk of shadowing with e-beam NO 2 diffuses into the seams of tub lids Key Benefits Achieves 10 6 biodecontamination of tubs Keeps pace with low and medium speed filling lines 20 tubs in 20 minutes Room temperature process No hot equipment to handle or HVAC issues Only electrical power required (no drain) Page 7

8 Concentration (ppm) Biodecontamination With Nitrogen Dioxide Faster Aeration is the Key to Matching Filling Line Output The H 2 O 2 level reaches a plateau Chart Title FTIR Data EC Cell Draeger Tube 10 HPV Peroxide Plateau Time (Minutes) Aeration of NO 2 is significantly faster, allows for rapid processing. Page 8

9 NO 2 Diffuses into Narrow Gaps, Minimizing Risk in Seams Head-to-Head with HPV Page 9

10 NO 2 Diffuses into Narrow Gaps, Minimizing Risk in Seams Test Fixture and Gas Flow This test is meant to show diffusion into small spaces 3-mm gap between two plates This simulates closely positioned components in a filling line Gas enters down-wind side Gas Flow Box sealed on 3 sides Air Lock Page 10

11 NO 2 Diffuses into Narrow Gaps, Minimizing Risk in Seams Test Fixture and Gas Flow Hydrogen Peroxide Vapor Exposure BI Time Carrier 3 cm 8 cm 2 min. SS N P Q N N 4 min. SS P P Q N P 6 min. SS P P Q N P 8 min. SS N P Q N P 16 min. SS P P Q N P 24 min. SS P P Q N P NO 2 Gas Processing Exposure BI Data Carrier 3 cm 8 cm 10 mg/l, 4 min. SS N N 75% RH Q N N 10 mg/l, 10 min. SS N N 75% RH Q N N 10 mg/l, 16 min. SS N N 75% RH Q N N 10 mg/l, 24 min. SS N N 75% RH Q N N Page 11

12 NO 2 Sterilization Solutions To Finish Page 12

13 NO 2 Sterilization Solutions Critical Process Parameters and Their Ranges Page 13 Process Parameter Temperature 10 C to 30 C Vacuum (Optional) NO 2 Concentration Relative Humidity Dwell Time Exposure Pulses 1 to 6 Aeration PFS: 700 Torr to Devices: 20 Torr 5 mg/l to 20 mg/l Range PFS: 80% RH to Sensitive Devices: 40% RH PFS: Typically 60 min to Devices: 10 min to 240 min 60 to 120 min

14 NO 2 Sterilization Solutions Vacuum Applications Typical cycle for devices that can withstand pressure changes Page 14

15 NO 2 Sterilization Solutions Minimal-Vacuum Applications Typical prefilled syringe applications Page 15

16 Sterilization of Prefilled Syringes Regulatory Driving Force FDA has requested sterilization of the external surfaces of prefilled syringes in Ophthalmics or Sterile Field Applications Even external surfaces of vials for surgical applications 2004 Guidance for Industry, Sterile Drug Products Produced by Aseptic Processing Current Good Manufacturing Practices Sensitive drugs/biologics prevent terminal sterilization Design of delivery devices can also prevent terminal sterilization FDA recommends that manufacturers employ an adjunct sterilization process to mitigate risk to the patient Page 16

17 Validation of the NO 2 Sterilization Process Follows ISO 14937:2009 for Medical Devices ISO is a general sterilization standard for med device Closely aligned with ISO Standard governing EO sterilization Governs validation of sterilization process: Product Definition Bioburden Estimation (typically vary low for parenteral products) Includes packaging, fixturing, load size, etc. Process Definition Identify the extent of treatment that provides a 6-log reduction Performance Qualification Demonstrate repeatability, product quality and safety Also provides guidance for routine monitoring Page 17 ISO 13485:2003 FM

18 Validation of the NO 2 Sterilization Process Product Definition Prefilled Syringe Glass, 1-mL long syringe Staked needle Rigid needle shield Pharma rubber plunger PP Plunger rod Tyvek-Mylar pouch Configuration for clinical trials BI in the most challenging location Next to the plunger, inside of the syringe barrel 10 6 G. stearothermophilus spores Page 18

19 Validation of the NO 2 Sterilization Process Process Definition Prefilled Syringe BIs in syringes located throughout the load 10 BIs in total (Follow guidance in ISO 11135) 10 samples for product tests of sterility (PToS) Dunnage was used for the balance of the load Simulated a load of 300 syringes Clinical trial-sized batch Load volume of 360 liters Follows Overkill Approach Conservative method Annex D of ISO SAL of 10-6 Page 19

20 Validation of the NO 2 Sterilization Process Process Definition 6-Log Reduction Was Achieved NO 2 dose was 5 mg/l NO 2, RH was 70% RH Dwell Time was the process variable Prefer to add a safety factor into the process Page 20

21 Validation of the NO 2 Sterilization Process Sterilization Process Specification Parameter Load Size Set Point 300 syringes Temperature 20 C (+/- 2) Thermal Equilibration Time 120 min (+/- 4) BIs per load 10 Relative Humidity 75% RH (+/- 10) NO 2 Dose (2X Safety Factor) 10 mg/l (+/- 1) Dwell Time 60:00 (+/- 00:10) Aeration Time 90:00 (+/- 00:10) Page 21

22 Validation of the NO 2 Sterilization Process Process Definition Prefilled Syringe Follows a very conservative approach SAL of Half-cycle BIs and PToS samples are required Follow the USP <71> guidance for sampling plans Full-cycle BIs only are required Product quality samples are exposed to both cycles Meet lot release criteria Page 22

23 Single Batch Release with NO 2 Sterilization Ideal for Clinical Trials Follows a very conservative approach: SAL Follow guidance in AAMI TIR16 and USP <71> Half-cycle is performed first BIs and product test of sterility samples external surfaces Full-cycle is performed next Fresh BIs are the microbiological challenge Product and samples for release testing see 1.5 cycles Product stability, NO 2 ingress, sterility (contents), etc. Page 23

24 NO 2 Sterilization of Prefilled Syringes Top Three Questions Asked 1. What happens when NO 2 diffuses into water? 2 NO 2 (g) + H 2 O (l) HNO 2 (aq) + HNO 3 (aq) HNO 2 (aq) H + (aq) + NO 2 (aq) HNO 3 (aq) H + (aq) + NO 3 (aq) 2. If NO 2 ingress does occur, what will happen to my product? Spiking of product with HNO 3 or NO 2 to provide a signal for quality testing what to look for? 3. How can you prove that NO 2 ingress does not occur? Page 24

25 NO 2 Sterilization of Prefilled Syringes Detection of NO 2 Ingress Noxilizer employs a colorimetric assay for nitrate. Page 25

26 NO 2 Sterilization of Prefilled Syringes Detection of NO 2 Ingress Limit of Detection ppm NO 3 - Limit of Quantification ppm NO 3 - The LOD is 8X lower than European Pharmacopoeia limit of 0.2 ppm NO 3 - COP Syringes Staked needles Rubber needle shield NO 3- < LOD Glass Syringes Staked needles Rigid needle shield NO 3- < LOD This data can be coupled with stability and shelf-life tests to remove doubt about the quality of the product. Page 26

27 Contact Information North America Evan Goulet, PhD VP Microbiology Applications Lab Office: Mike Valentine Vice President, North American Sales Office: Angie Kearns Midwest Sales Representative Office: Page 27

STERILIZATION VALIDATION IN THIS SECTION REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES. Radiation Sterilization Validation

STERILIZATION VALIDATION IN THIS SECTION REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES. Radiation Sterilization Validation STERILIZATION VALIDATION All sterilization processes require validation of the efficacy and reproducibility of the process. Depending on the type of sterilization, this may be accomplished by partial,

More information

REUSABLE MEDICAL DEVICE VALIDATION

REUSABLE MEDICAL DEVICE VALIDATION All sterilization processes require validation of the efficacy and reproducibility of the process. Depending on the type of sterilization, this may be accomplished by partial, sub-lethal, or repetitive

More information

Technical Data Monograph. STERIS VHP LTS-V - Low Temperature Surfaces Terminal Sterilization

Technical Data Monograph. STERIS VHP LTS-V - Low Temperature Surfaces Terminal Sterilization Technical Data Monograph STERIS VHP LTS-V - Low Temperature Surfaces Terminal Sterilization STERIS VHP LTS-V Low Temperature Surfaces Terminal Sterilization Table of Contents 1. Abstract 3 2. VHP LTS-V

More information

Secure. Tested. Certified.

Secure. Tested. Certified. Secure. Tested. Certified. We solve problems with confidence and creativity. Monitoring Products for Dry Heat Steam-Formaldehyde FORM Chlorine Dioxide Table of Contents Crosstex Industrial is a division

More information

Medline Gemini Sterilization Wrap. Test Studies

Medline Gemini Sterilization Wrap. Test Studies Medline Gemini Sterilization Wrap Test Studies Steam Efficacy : Steam PreVac Steam Efficacy: Steam Gravity Shelf Life: 180 Days Plasma Efficacy: STERRAD 100NX Plasma Efficacy: STERRAD NX Plasma Efficacy:

More information

gke Steri-Record Testsets and PCDs for validation, process and batch monitoring of sterilization processes

gke Steri-Record Testsets and PCDs for validation, process and batch monitoring of sterilization processes gke Steri-Record Testsets and PCDs for validation, process and batch monitoring of sterilization processes Application Process Monitoring Systems (PMS) are process challenge devices that are used to test

More information

PRODUCT CATALOGUE. for gke Steri-Record Biological Indicators, Process Challenge Devices and Accessories

PRODUCT CATALOGUE. for gke Steri-Record Biological Indicators, Process Challenge Devices and Accessories PRODUCT CATALOGUE for gke Steri-Record Biological Indicators, Process Challenge Devices and Accessories Spore strips, Ampoules, Suspensions, Self-contained Mini-Bio-Plus Indicators to monitor steam, formaldehyde,

More information

Advancements on implementation of single use technology in vaccine manufacturing

Advancements on implementation of single use technology in vaccine manufacturing Advancements on implementation of single use technology in vaccine manufacturing October 25 th, 2013 DCVMN Meeting Rio de Janeiro, Brazil Outline Overview of single use products Applications of single

More information

DIFFERENTIATION WITHOUT DISRUPTION

DIFFERENTIATION WITHOUT DISRUPTION DIFFERENTIATION WITHOUT DISRUPTION In this article, Alan Shortall, Chief Executive Officer, Unilife Corporation, describes a new paradigm now driving the pharmaceutical market for advanced drug delivery

More information

Annex A2. Guidance on Process Validation Scheme for Aseptically Processed Products

Annex A2. Guidance on Process Validation Scheme for Aseptically Processed Products Annex A2 Guidance on Process Validation Scheme for Aseptically Processed Products 1 Table of content 1 PURPOSE... 3 2 SCOPE... 3 3 GENERAL INFORMATION... 3 4 INFORMATION NEEDED FOR ASEPTIC PROCESSES VALIDATION...

More information

PRODUCT CATALOGUE. for gke Steri-Record Biological Indicators, Process Challenge Devices and Accessories

PRODUCT CATALOGUE. for gke Steri-Record Biological Indicators, Process Challenge Devices and Accessories PRODUCT CATALOGUE for gke Steri-Record Biological Indicators, Process Challenge Devices and Accessories Spore strips, Ampoules, Suspensions, Self-contained Mini-Bio-Plus Indicators to monitor steam, formaldehyde,

More information

PRE-FILLABLE GLASS SYRINGES

PRE-FILLABLE GLASS SYRINGES PRE-FILLABLE GLASS SYRINGES FINE-TUNED TO YOUR NEEDS PRE-FILLABLE GLASS SYRINGES FINE-TUNED TO YOUR NEEDS The pre-fillable syringe market has grown significantly in the past years. Pharmaceutical companies

More information

8/27/2015. Learning Objectives. Purpose of Sterilization. To destroy all living pathogenic microorganisms

8/27/2015. Learning Objectives. Purpose of Sterilization. To destroy all living pathogenic microorganisms Learning Objectives Compare and contrast the available methods of sterilization for the healthcare setting Describe the appropriate use and monitoring for sterilization methods Purpose of Sterilization

More information

Innovative. Smart. Trusted.

Innovative. Smart. Trusted. Innovative. Smart. Trusted. We re your partner for sterilization products. Monitoring Products for Vaporized Hydrogen Peroxide Table of Contents Crosstex Industrial is a division of Crosstex International,

More information

Principles of Chlorine Dioxide Gas as a Decontamination Method

Principles of Chlorine Dioxide Gas as a Decontamination Method Principles of Chlorine Dioxide Gas as a Decontamination Method In order for any decontamination method to be effective, the following points must be satisfied The decontamination method must: Be able to

More information

Biological Monitoring

Biological Monitoring Biological Monitoring Prof. Duygu PERÇİN, MD Department of Clinical Microbiology Erciyes University Faculty of Medicine, Kayseri-TURKEY duygu.percin@hotmail.com Learning Objectives The attandee will be

More information

Critical Environment Products and Services

Critical Environment Products and Services Critical Environment Products and Services For over two decades, EP Scientific products and services have defined clean for environmental sampling containers and services. Our proprietary cleaning methods

More information

USP Chapter 823 USP 32 (old) vs. USP 35 (new)

USP Chapter 823 USP 32 (old) vs. USP 35 (new) USP Chapter 823 USP 32 (old) vs. USP 35 (new) Sally W. Schwarz, MS, BCNP Research Associate Professor of Radiology Washington University School of Medicine St. Louis, MO Why USP Chapter ? FDA has

More information

Hot Topics in Drug Product Process Validation: A Reviewer s Perspective

Hot Topics in Drug Product Process Validation: A Reviewer s Perspective Hot Topics in Drug Product Process Validation: A Reviewer s Perspective Colleen Thomas, Ph.D. Quality Assessment Lead (Acting) FDA/CDER/OPQ/OPF Division of Microbiology Assessment CASSS CMC Strategy Forum

More information

ISPE Nordic, Denmark Yves Scholler SKAN AG

ISPE Nordic, Denmark Yves Scholler SKAN AG Risk of contamination through pinholes in gloves and how to prove the integrity ISPE Nordic, Denmark Yves Scholler SKAN AG 07.10.2015 1 Agenda 1 Regulatory Requirements 2 Overview of Different Testing

More information

Primary Packaging change: Switch from a lyophilisate in vial to a lyophilisate in double chamber cartridge

Primary Packaging change: Switch from a lyophilisate in vial to a lyophilisate in double chamber cartridge Primary Packaging change: Switch from a lyophilisate in vial to a lyophilisate in double chamber cartridge Christoph Stark Head Pharmaceutical Development Novartis Pharma AG, Novartis Biologics CASSS CMC

More information

sterilisation monitoring products

sterilisation monitoring products sterilisation monitoring products Known for their consistent, reliable performance, Excelsior Scientific sterilisation monitoring products are used by pharmaceutical and medical device manufacturers, contract

More information

Environmental Monitoring of Aseptic Processing Areas - 1

Environmental Monitoring of Aseptic Processing Areas - 1 Environmental Monitoring of Aseptic Processing Areas - 1 A war against an invisible enemy DCVMN - Víctor Maqueda - May 30, 31, June 1 2016 1 Training Course Agenda Overview of Environmental Monitoring

More information

Product Catalogue. for gke Steri-Record Biological Indicators, Process Challenge Devices and Accessories

Product Catalogue. for gke Steri-Record Biological Indicators, Process Challenge Devices and Accessories Product Catalogue for gke Steri-Record Biological Indicators, Process Challenge Devices and Accessories Spore Strips, Ampoules, Suspensions, Self-contained Mini-Bio-Plus Indicators to monitor steam, formaldehyde,

More information

Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar

Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar MONITORING THE SYSTEM 1E PROCESSOR Q: I want to confirm that there is no need for a Biological Indicator (BI) and that use

More information

PDA: A Global. Association. Proposed Revisions to USP <1207> STERILE PRODUCT -

PDA: A Global. Association. Proposed Revisions to USP <1207> STERILE PRODUCT - Proposed Revisions to USP STERILE PRODUCT - PDA: A Global PACKAGE INTEGRITY EVALUATION Now in the Sep/Oct 2014 USP Pharmacopeial Forum For Public Comment Association Prepared by: Dana M. Guazzo

More information

Sterilization & Chemical Resistance of Healthcare Polymers

Sterilization & Chemical Resistance of Healthcare Polymers Sterilization & Chemical Resistance of Healthcare Polymers Key Terms & Definitions Sterilization - A process that eliminates or kills all forms of life, including transmissible agents, on a medical device;

More information

with Nemera s extensive experience in developing and manufacturing parenteral drug delivery devices

with Nemera s extensive experience in developing and manufacturing parenteral drug delivery devices Co-Development INJECTING NEW IDEAS Empower patients through good design Parenteral administration of a drug exposes users (patients and healthcare professionals) to numerous hazards. In designing a medical

More information

for IND and RDRC Regulated PET Compounding

for IND and RDRC Regulated PET Compounding Overview of USP Chapter for IND and RDRC Regulated PET Compounding Distributed Manufacturing of PET Radiopharmaceuticals for Multi-Center Clinical Trials SNM, Annual Meeting Toronto, Ontario, Canada

More information

DISINFECTION QUALIFI CATION TESTING CONSIDERATIONS FOR THE ASEPTIC AND CLEANROOM MANUFACTURING ENVIRONMENT

DISINFECTION QUALIFI CATION TESTING CONSIDERATIONS FOR THE ASEPTIC AND CLEANROOM MANUFACTURING ENVIRONMENT DISINFECTION QUALIFI CATION TESTING CONSIDERATIONS FOR THE ASEPTIC AND CLEANROOM MANUFACTURING ENVIRONMENT by Dave Rottjakob, M.T. (ASCP) December, 2013 INTRODUCTION Obtaining the highest confidence that

More information

Biological Indicators and H 2 O 2 Decontamination of Isolator Systems. Volker Sigwarth Skan AG Switzerland. Process Development

Biological Indicators and H 2 O 2 Decontamination of Isolator Systems. Volker Sigwarth Skan AG Switzerland. Process Development Biological Indicators and H 2 O 2 Decontamination of Isolator Systems Volker Sigwarth Skan AG Switzerland Process Development Biological Indicators and H 2 O 2 Decontamination Process Development Overview

More information

Bioburden Contamination Control: A Holistic Overview

Bioburden Contamination Control: A Holistic Overview Bioburden Contamination Control: A Holistic Overview Scott Sutton, PhD Microbiology Network, Inc. Scott Sutton, PhD is the Principal of Microbiology Network, Inc (http:// microbiologynetwork.com/) a company

More information

The Parenteral Drug Association

The Parenteral Drug Association C O N F E R E N C E R E P O R T Equipment Cleaning Microbial Control Issues By Destin A. LeBlanc, M.A. Cleaning Validation Technologies Validation: The Parenteral Drug Association (PDA) spring conference

More information

GUIDELINE FOR THE STABILITY TESTING

GUIDELINE FOR THE STABILITY TESTING 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 GUIDELINE FOR THE STABILITY TESTING OF NON-PRESCRIPTION (OTC)

More information

PACKAGE TESTING & VALIDATION IN THIS SECTION EXPERT SOLUTIONS FOR PRODUCT DEVELOPMENT

PACKAGE TESTING & VALIDATION IN THIS SECTION EXPERT SOLUTIONS FOR PRODUCT DEVELOPMENT EXPERT SOLUTIONS FOR PRODUCT DEVELOPMENT PACKAGE TESTING & VALIDATION A sterile medical device, which includes the packaging, must function as labeled for the shelf-life claimed for the device. The primary

More information

Validation of cleaning and sterilization of reusable medical devices

Validation of cleaning and sterilization of reusable medical devices Synergy makes sence Validation of cleaning and sterilization of reusable medical devices MG-FSI72-119 Last revision: June 2012 Aurélien BIGNON 5, Chemin du Catupolan - 69120 Vaulx en Velin - France - Tel.

More information

Compendial Effort to Revise Packaging Material Standards: Glass, Plastic, Elastomer, and Other

Compendial Effort to Revise Packaging Material Standards: Glass, Plastic, Elastomer, and Other Compendial Effort to Revise Packaging Material Standards: Glass, Plastic, Elastomer, and Other Desmond Hunt, Ph.D. PNP Stakeholder Forum October 19, 2017 1 2017 USP USP current efforts in revision of packaging

More information

Completing. By the end of this chapter, you will be able to: Introduction. Chapter 16

Completing. By the end of this chapter, you will be able to: Introduction. Chapter 16 Chapter 16 Completing Product Processing By the end of this chapter, you will be able to: l Explain the term cold chain l Describe the type of fridge used for correct storage of refrigerated products l

More information

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces HemoCheck TM Sample Policy SUBJECT: Detection of blood residue on various surfaces DEPARTMENT: Central Service APPROVED BY: EFFECTIVE: REVISED: 07/2016 PURPOSE: To test for detection of blood residue on

More information

Preparing Your Aseptic Processing Facility for an FDA Inspection. Valerie Welter, Director Quality Bayer HealthCare March 10, 2014

Preparing Your Aseptic Processing Facility for an FDA Inspection. Valerie Welter, Director Quality Bayer HealthCare March 10, 2014 Preparing Your Aseptic Processing Facility for an FDA Inspection Valerie Welter, Director Quality Bayer HealthCare March 10, 2014 Agenda Regulatory Requirements Establishing your Approach Aseptic Controls

More information

Aseptic Process Validation

Aseptic Process Validation Aseptic Process Validation IMB GMP Information Seminar 27 th September 2012 Gerard Sheridan, Inspector Date Insert on Master Slide Slide 1 Overview Guidance Best Practices Common Deficiencies Slide 2 Aseptic

More information

QbD In Drug Development. Mathew Cherian Ph.D. Director & Senior Fellow Pfizer, USA

QbD In Drug Development. Mathew Cherian Ph.D. Director & Senior Fellow Pfizer, USA QbD In Drug Development Mathew Cherian Ph.D. Director & Senior Fellow Pfizer, USA The Origin of QbD The concept of Quality by Design (QbD) was introduced by Romanian born US engineer Joseph Juran QbD was

More information

Decon 2.0: Emerging Decontamination Technologies

Decon 2.0: Emerging Decontamination Technologies Decon 2.0: Emerging Decontamination Technologies Vaporized Hydrogen Peroxide (VHP) and Chlorine Dioxide Gas Decontamination Case Studies: Peter Harris Director of Operations B & V Testing, INC. 800.851.9081

More information

PharmaED s. Featuring Case Studies and Lessons Learned from Industry Experts! Including Special Coverage On: Featuring Representation From:

PharmaED s. Featuring Case Studies and Lessons Learned from Industry Experts! Including Special Coverage On: Featuring Representation From: P PharmaED s JULY 26-27, 2012, SHERATON LA JOLLA HOTEL, LA JOLLA, CA Register by June 1 st and receive a $300 Discount! re-filled Syringes Forum 2012 Strategic Development, Inspection, Safety & Regulatory

More information

Visible Particles: Regulatory and Compendial Requirements

Visible Particles: Regulatory and Compendial Requirements Visible Particles: Regulatory and Compendial Requirements John G. Shabushnig, Ph.D. Insight Pharma Consulting, LLC June 2014 Agenda Visible Particle Definitions US FDA US Pharmacopeia (USP) EC GMP s European

More information

Precision. Proficiency. Proximity.

Precision. Proficiency. Proximity. Precision. Proficiency. Proximity. Precision Accuracy in testing is paramount to ensuring product safety and efficacy. And fast turnaround times are critical to keeping your project on schedule. With

More information

FlexAct BT (Bag Tester) Flexboy 2D November, 2015

FlexAct BT (Bag Tester) Flexboy 2D November, 2015 FlexAct BT (Bag Tester) Flexboy 2D November, 2015 Agenda 1 2 3 4 Container Closure Integrity for SUS FlexAct BT Value Proposition FlexAct BT Instruments FlexAct BT SU Flexboy 2D 5 FlexAct BT Validation

More information

Actavis Italy. Nerviano Plant

Actavis Italy. Nerviano Plant Actavis Italy Nerviano Plant Starting out in 1901 in Jerusalem The company known today as Teva was established as a small wholesale drug business by Chaim Salomon, Moshe Levin and Yitschak Elstein. (They

More information

Best Practice Monitoring for Sterile Processing Professionals

Best Practice Monitoring for Sterile Processing Professionals Best Practice Monitoring for Sterile Processing Professionals Sterile processing departments are faced with greater challenges than ever. Increasing demands for quick turnarounds. Pressure to streamline

More information

Hygiene Monitoring. Biological Indicators for Monitoring Sterilization Processes. Product Information. BI_07_en

Hygiene Monitoring. Biological Indicators for Monitoring Sterilization Processes. Product Information. BI_07_en Hygiene Monitoring Biological Indicators for Monitoring Sterilization Processes Product Information BI_07_en BAG-BioStrip Biological Indicators for Monitoring all Types of Sterilization Processes STEAM

More information

FDA s Guidance for Industry

FDA s Guidance for Industry Sterile Drug Products Produced by Aseptic Processing - CGMPs Midwest FDC Conference November 4 2005 Susan Bruederle, Investigator FDA / ORA / Central Region Chicago District / Hinsdale IL FDA s Guidance

More information

PACKAGE TESTING AND VALIDATION SERVICES

PACKAGE TESTING AND VALIDATION SERVICES PACKAGE TESTING AND VALIDATION SERVICES A sterile medical device, which includes the packaging, must function as labeled for the shelf-life claimed for the device. The primary package must maintain its

More information

ASTM D10-F02 Workshop FDA Regulatory Perspective. Patrick Weixel FDA Center for Devices and Radiological Health

ASTM D10-F02 Workshop FDA Regulatory Perspective. Patrick Weixel FDA Center for Devices and Radiological Health ASTM D10-F02 Workshop FDA Regulatory Perspective Patrick Weixel FDA Center for Devices and Radiological Health 1 FDA Regulatory Perspective Focus of talk: Use of Standards Auditing Packaging Process Deficiencies

More information

PDA Technical Report #26:

PDA Technical Report #26: PDA Technical Report #26: Implications on Liquid Filter Validation Maik W. Jornitz, Sartorius Corp. Filter Qualification Tests for internal Validation Guides Physical Tests flow rates/delta p throughput

More information

Study Summary. Results: All tested media-filled vials were negative for growth of any microorganisms.

Study Summary. Results: All tested media-filled vials were negative for growth of any microorganisms. A 7 Days Microbial Ingress Test of Single Use Vials Utilizing the EQUASHIELD Closed System Drug Transfer Device Study by Nelson Laboratories (Salt Lake City, UT) 2013, updated in 2014* Study Summary Abstract:

More information

One such problem is blood residual inside the channels of a robotic arm.

One such problem is blood residual inside the channels of a robotic arm. SUBJECT: Robotic Arm Testing for residue blood DEPARTMENT: Central Service APPROVED BY: EFFECTIVE: REVISED: 07/2016 PURPOSE: To test for detection of blood residue inside and outside of a robotic arm and

More information

What Dentistry Can Learn from AAMI ST79. Best Practices for device and instrument reprocessing

What Dentistry Can Learn from AAMI ST79. Best Practices for device and instrument reprocessing What Dentistry Can Learn from AAMI ST79 Best Practices for device and instrument reprocessing Learning goals: Purpose and goals of AAMI organization. Overview of ST79. Review selected topics: IFUs, tools

More information

SCHOTT Pharmaceutical Packaging. Dedicated to quality pharmaceutical packaging

SCHOTT Pharmaceutical Packaging. Dedicated to quality pharmaceutical packaging SCHOTT Pharmaceutical Packaging Dedicated to quality pharmaceutical packaging Worldwide presence SCHOTT forma vitrum is one of the world s leading suppliers of parenteral packaging for the pharmaceutical

More information

Trends in Sterile Manufacturing Technologies

Trends in Sterile Manufacturing Technologies Trends in Sterile Manufacturing Technologies ISPE Thailand Annual Meeting Charlotte Enghave Fruergaard 2013.07.18 Where we come from 1930s Danish Novo and Nordisk Gentofte (later Novo Nordisk) employed

More information

They can be used on washer-disinfectors, ultrasonic cleaners, surgical instruments and endoscopes. Page 5

They can be used on washer-disinfectors, ultrasonic cleaners, surgical instruments and endoscopes. Page 5 Rev. 0 October 2017 Terragene offers a wide range of simple and effective products to precisely control the washing, hygiene and sterilization processes in veterinary clinics. Available in a variety of

More information

ISO Part 1 and Part 2 Compliance Requirements. Cathriona O Neill

ISO Part 1 and Part 2 Compliance Requirements. Cathriona O Neill ISO 11607 Part 1 and Part 2 Compliance Requirements Cathriona O Neill I.S. EN 11607 Introduction ISO 11607 is the principal guidance document. Packaging for terminally sterilised medical devices - Part

More information

Doc No.: MHD 12(0333)/ISO :2006 TITLE: Sterilization of health care products Biological indicators Part 1: General requirements

Doc No.: MHD 12(0333)/ISO :2006 TITLE: Sterilization of health care products Biological indicators Part 1: General requirements DRAFTS IN WIDE CIRCULATION TECHNICAL COMMITTEE: MHD 12 DOCUMENT DESPATCH ADVICE Ref Date MHD 12/T- 64, 65, 66, 67, 68, 69 & 70 26-03-2014 ADDRESSED TO: 1. All members of Hospital Equipment, Sectional Committee,

More information

Testing of routinely sterilized products should

Testing of routinely sterilized products should Product Quality Assurance Testing: One Hospital s Experience Testing of routinely sterilized products should be part of a healthcare facility s quality control program, and ANSI/AAMI ST79:2006, Comprehensive

More information

Microbiological Consideration for Non-Sterile Pharmaceutical

Microbiological Consideration for Non-Sterile Pharmaceutical May 1-3, 2012 Javits Center New York, NY Microbiological Consideration for Non-Sterile Pharmaceutical Dr. Leonard W. Mestrandrea Principal Consultant MESTRANDREA CONSULTING LLC Title Date Microbial Control

More information

MedAccred Sterilization (STN)

MedAccred Sterilization (STN) MedAccred Sterilization (STN) Radiation and Ethylene Oxide Sterilization for Medical Device Manufacturing Critical Manufacturing Processes for the Medical Device Industry Mark Aubele Staff Engineer STN

More information

Vetter Development Service

Vetter Development Service A unique blend of expertise for clinical manufacturing Vetter Development Service Supporting your compound s success Answers that work 1 Your molecule s future starts here. By the time your new injectable

More information

GLOBAL QUALITY SOLUTIONS PACKAGE INTEGRITY SEAL INTEGRITY LEAK DETECTION

GLOBAL QUALITY SOLUTIONS PACKAGE INTEGRITY SEAL INTEGRITY LEAK DETECTION GLOBAL QUALITY SOLUTIONS PACKAGE INTEGRITY SEAL INTEGRITY LEAK DETECTION Redefining the standards for accuracy and reliability. About PTI PTI Packaging Technologies & Inspection is headquartered in Tuckahoe,

More information

Particulate Matter and Visual Inspection: Industry Trends 2015

Particulate Matter and Visual Inspection: Industry Trends 2015 Particulate Matter and Visual Inspection: Industry Trends 2015 John G. Shabushnig, Ph.D. Insight Pharma Consulting, LLC November 2015 Agenda FDA Recalls and 483 Observations Industry Benchmarking Survey

More information

Specialty Testing of Steam Sterilizers

Specialty Testing of Steam Sterilizers Specialty Testing of Steam Sterilizers Nancy Chobin, RN, AAS, ACSP, CSPM, CFER 2017 Sterile Processing University, LLC. **This in-service has been Approved by the CBSPD for 1.5 CEUs. Objectives 1. To describe

More information

Review Validation of aseptic processes for pharmaceuticals

Review Validation of aseptic processes for pharmaceuticals OPEM www.opem.org Oriental Pharmacy and Experimental Medicine 2010 10(4), 231-238 DOI 10.3742/OPEM.2010.10.4.231 Review Validation of aseptic processes for pharmaceuticals Lincy Joseph*, Mathew George

More information

Infection Control. Low Temperature Sterilization (LTS) Critical Aspects on Selection and Validation of LTS-Processes

Infection Control. Low Temperature Sterilization (LTS) Critical Aspects on Selection and Validation of LTS-Processes Infection Control Low Temperature Sterilization (LTS) Critical Aspects on Selection and Validation of LTS-Processes Dr. Tronje von dem Hagen tronje@vondemhagen.com Lübeck (Germany) T. v. dem Hagen, Lübeck/Germany

More information

Vaporized Peracetic Acid (VPA) Verified Effective for Sterilizing Bioabsorbable Medical Devices

Vaporized Peracetic Acid (VPA) Verified Effective for Sterilizing Bioabsorbable Medical Devices Vaporized Peracetic Acid (VPA) Verified Effective for Sterilizing Bioabsorbable Medical Devices Contents Introduction 1 Background 2 VPA Process 4 Test Results 5 Conclusion 9 ABSTRACT This report explains

More information

ISO INTERNATIONAL STANDARD. Sterilization of health care products Chemical indicators Part 1: General requirements

ISO INTERNATIONAL STANDARD. Sterilization of health care products Chemical indicators Part 1: General requirements INTERNATIONAL STANDARD ISO 11140-1 Second edition 2005-07-15 Sterilization of health care products Chemical indicators Part 1: General requirements Stérilisation des produits de santé Indicateurs chimiques

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 14161 Second edition 2009-09-15 Sterilization of health care products Biological indicators Guidance for the selection, use and interpretation of results Stérilisation des produits

More information

Overview of a sterility assurance program for PET drugs

Overview of a sterility assurance program for PET drugs Coalition for PET Drug Approval Radiopharmaceutical Sciences Council Overview of a sterility assurance program for PET drugs Eric Webster, PETNET Solutions Disclosures Employee of PETNET Solutions, a Siemens

More information

GENERAL SURGERY DEVICES

GENERAL SURGERY DEVICES GENERAL SURGERY DEVICES LAPAROSCOPIC INSTRUMENTS SUTURE PASSERS TROCARS ULTRASONIC SCALPELS REPROCESSED SURGICAL DEVICES STRYKER S PORTFOLIO OF REPROCESSED SINGLE-USE DEVICES Laparoscopic devices reprocessed

More information

re-filled Syringes Forum 2012

re-filled Syringes Forum 2012 P PharmaED s Register by January 30 th and receive a $300 Discount! re-filled Syringes Forum 2012 Strategic Development, Inspection, Safety & Regulatory Compliance and Commercialization of Pre-Filled Syringes

More information

510(k) Summary. 510(k) Number: K Summary Prepared: August 31, 2017

510(k) Summary. 510(k) Number: K Summary Prepared: August 31, 2017 510(k) Summary 510(k) Number: K162613 Summary Prepared: August 31, 2017 Submitter: Dr. Fred Ma Repro Med System, Inc., D/B/A RMS Medical Products 24 Carpenter Road, Chester, NY 10918 Tel: 845-469-2042

More information

Whitepaper. High temperature HEPA filtration. Dr.-Ing. Marc Schmidt, Dr.-Ing. Lothar Gail, Hugo Hemel MSc. Preview

Whitepaper. High temperature HEPA filtration. Dr.-Ing. Marc Schmidt, Dr.-Ing. Lothar Gail, Hugo Hemel MSc. Preview High temperature HEPA filtration Air filtration challenges and answers for dry heat sterilization tunnels Whitepaper Preview Dr.-Ing. Marc Schmidt, Dr.-Ing. Lothar Gail, Hugo Hemel MSc. Air filtration

More information

Type I Glass for Pharmaceutical Containers: Technical Requirements and regulatory update. Daniele Zuccato, Core Team Leader

Type I Glass for Pharmaceutical Containers: Technical Requirements and regulatory update. Daniele Zuccato, Core Team Leader Type I Glass for Pharmaceutical Containers: Technical Requirements and regulatory update Daniele Zuccato, Core Team Leader Glass Containers for Pharmaceutical Use Glass Containers Production Topics Delamination

More information

Sterile Compounding elearning Course Curriculum 28 lessons with 33 hours of CE. Fundamentals of Sterile Compounding (8 lessons/8 hours CE)

Sterile Compounding elearning Course Curriculum 28 lessons with 33 hours of CE. Fundamentals of Sterile Compounding (8 lessons/8 hours CE) Sterile Compounding elearning Course Curriculum 28 lessons with 33 hours of CE CriticalPoint s Sterile Compounding elearning curriculum is written by industry experts and covers both Chapter and

More information

GUIDE TO INSPECTIONS OF STERILE DRUG SUBSTANCE MANUFACTURERS

GUIDE TO INSPECTIONS OF STERILE DRUG SUBSTANCE MANUFACTURERS GUIDE TO INSPECTIONS OF STERILE DRUG SUBSTANCE MANUFACTURERS Note: This document is reference material for investigators and other FDA personnel. The document does not bind FDA, and does no confer any

More information

Welcome! How do I get a CE Certificate?

Welcome! How do I get a CE Certificate? M Sterile U Network M Sterile U Web Meeting November 0, 0 Today s meeting times: 9:00 a.m., :00 a.m., and :00 p.m. Central Standard Time To hear audio, call 800-97-00 and enter access code 7 Phone lines

More information

Competency Guide: Care and Handling of Rigid Endoscopes

Competency Guide: Care and Handling of Rigid Endoscopes Competency Guide: Care and Handling of Rigid Endoscopes This document details the basic steps to correctly process and maintain a rigid endoscope. This document can be used to perform a competency review

More information

New methods to monitor steam penetration into complex medical devices (MD) using Medical Device Simulators (MDS) and Batch Monitoring Systems (BMS)

New methods to monitor steam penetration into complex medical devices (MD) using Medical Device Simulators (MDS) and Batch Monitoring Systems (BMS) New methods to monitor steam penetration into complex medical devices (MD) using Medical Device Simulators (MDS) and Batch Monitoring Systems (BMS) WFHSS Workshop Bydgoszcz, Friday, December 7th, 2007

More information

PERFUSION TECHNOLOFISTS OF GREATER CHICAGO, INC. POLICY AND PROCEDURE MEDTRONIC HEMOTEC ACTIVATED CLOTTING TIME (ACT) OPERATING PROCEDURE

PERFUSION TECHNOLOFISTS OF GREATER CHICAGO, INC. POLICY AND PROCEDURE MEDTRONIC HEMOTEC ACTIVATED CLOTTING TIME (ACT) OPERATING PROCEDURE DATE OF REVISION Page 1 of 13 (ACT) OPERATING PROCEDURE PRINCIPLE: The ACT test is a definitive test used to monitor the anticoagulant effect of heparin. It measures the clotting time of fresh whole blood

More information

The Complete Solution for Clean Room Aerosol-Based Disinfection MINNCARE DRY FOG SYSTEM

The Complete Solution for Clean Room Aerosol-Based Disinfection MINNCARE DRY FOG SYSTEM The Complete Solution for Clean Room Aerosol-Based Disinfection Current Methods of Room Disinfection Surface Wiping Manual Spraying (with Spray Bottles) Heating Process (Vaporization) Cold Process - Wet

More information

BD PhaSeal System. Product Catalog

BD PhaSeal System. Product Catalog BD PhaSeal System Product Catalog BD PhaSeal is a clinically proven closed-system drug transfer device that prevents hazardous drug exposure. This product catalog contains information regarding the different

More information

Prequalification Team WHO PUBLIC INSPECTION REPORT Vaccine Manufacturer

Prequalification Team WHO PUBLIC INSPECTION REPORT Vaccine Manufacturer Part 1: General information Name of Manufacturer Production Block Physical address Contact address Prequalification Team WHO PUBLIC INSPECTION REPORT Vaccine Manufacturer. Clean Utilities in the Basement.

More information

The Opta SFT-D is available as individual device for end-user assembly with TPE tubing and autoclave sterilization.

The Opta SFT-D is available as individual device for end-user assembly with TPE tubing and autoclave sterilization. Fluid Management 259 Connections 260 Flexboy Bags 264 Flexel 3D Bags 276 Palletank 304 Instruments 314 Single-Use Mixing 318 Freeze-Thaw Systems 354 Aseptic Transfer System 380 Configurable Solutions FlexAct

More information

ISO/TC 198. Secretariat: ANSI

ISO/TC 198. Secretariat: ANSI DRAFT INTERNATIONAL STANDARD ISO/DIS 17664 ISO/TC 198 Secretariat: ANSI Voting begins on: Voting terminates on: 2016-04-28 2016-07-27 Processing of health care products Information to be provided by the

More information

Chuck Reed at Weiler Engineering, Inc compares aseptic blow-fill-seal technology with traditional aseptic processing of pharmaceutical liquids

Chuck Reed at Weiler Engineering, Inc compares aseptic blow-fill-seal technology with traditional aseptic processing of pharmaceutical liquids Manufacturing Aseptic Advances Chuck Reed at Weiler Engineering, Inc compares aseptic blow-fill-seal technology with traditional aseptic processing of pharmaceutical liquids Acknowledged by the FDA as

More information

Dimensional Stability of a Multicavity Injection Molded Article

Dimensional Stability of a Multicavity Injection Molded Article Technical Report Dimensional Stability of a Multicavity Injection Molded Article Technical Report Series 2016-008 Keywords: dimensions, syringe, blood tube, vacutainer, cavity, injection, molding, diameter,

More information

Validation of Sterilizing Grade Filters

Validation of Sterilizing Grade Filters Validation of Sterilizing Grade Filters Presented by Laura Okhio-Seaman Sartorius Corporation 1 Sterilizing Grade Filters The definition of a sterilizing grade filter is one that will produce a sterile

More information

Experience with Adaptive Dose-Ranging Studies in Early Clinical Development

Experience with Adaptive Dose-Ranging Studies in Early Clinical Development Experience with Adaptive Dose-Ranging Studies in Early Clinical Development Judith Quinlan MSc Vice President Adaptive Trials Cytel Inc. judith.quinlan@cytel.com Thanks to members of the PhRMA Adaptive

More information

Media Fill A Process Simution. Presented By Shikha Chauhan

Media Fill A Process Simution. Presented By Shikha Chauhan Media Fill A Process Simution Presented By Shikha Chauhan 21 CFR 211.113 Validation of Aseptic Processing and Sterilization Process Simulation / Media Fill Filtration Efficacy Sterilization of Equipment,

More information

The biopharmaceutical industry

The biopharmaceutical industry S i n g l e - U s e APPLICATIONS Understanding Particulates in Single-Use Bags Their Relationship to USP Chapter by Michael W. Johnson The biopharmaceutical industry is facing many challenges. Global

More information

Aseptic Processing Current Issues & Trends

Aseptic Processing Current Issues & Trends Aseptic Processing Current Issues & Trends 1 Introduction Richard M. Johnson Member, PDA for 20+ years President & CEO since 2009 Ladies and Gentlemen, I am happy to be here with you. Senhoras e Senhores,

More information

Driving Value through Innovation in Biotech Manufacturing. Agenda

Driving Value through Innovation in Biotech Manufacturing. Agenda Driving Value through Innovation in Biotech Manufacturing Jorg Thommes, PhD Senior Vice President Operations Technology and Innovation Biogen 7 October 2015 Agenda Biopharma Legacy, Trends, & Challenges

More information