Compensation: Fundamental Principles

Size: px
Start display at page:

Download "Compensation: Fundamental Principles"

Transcription

1 Flow Cytometry Seminar Series 2017 : Fundamental Principles Spillover correction in multicolor flow cytometry

2 Contents Fluorescence and its detection Absorption and emission Detection and spillover of spillover Theory and calculation Practical example Limitations of compensation

3 Fluorophores Synthetic dyes and fluorescent proteins pdb.org Usually have conjugated systems of π electron that can absorb and release energy in the form of photons.

4 Fluorescence Absorption Absorption: a molecule absorbs a photon and enters a higher energy, excited state. Excited state Transition probability E = h λ Transition energy Ground state 1) Transitions to different vibrational levels possible, but only in defined energies. 2) Some transitions are more likely than others. -> Absorption spectrum

5 Fluorescence Stokes Shift 1) Fast relaxation to lowest energy excited state level -> Energy released as heat. Excited state Heat E = h λ Photons Ground state 2) Relaxation to an energy level of the ground state -> Energy released as a photon.

6 Fluorescence Summary Fluorophores absorb and emit light over a range of wavelengths. An emitted photon is of lower energy (higher wavelength) than the absorbed one.

7 Fluorescence Detection Normal detectors are blind to the wavelength -> optical filters AHF.de

8 Fluorescence Detection Light filtering: 1) Long pass (LP) filter Reflects everything below its designated wavelength.

9 Fluorescence Detection Light filtering: 1) Long pass (LP) filter Reflects everything below its designated wavelength. 2) Band pass (BP) Lets only a narrow band of wavelengths through.

10 Fluorescence Detection Detection window (DW): Combined effect of LP and BP filter. Detected are all photons passing both filters. -> Only a part of the total dye emission is detected in the channel.

11 Fluorescence Detection Multiple dyes need multiple detection windows (and detectors).

12 Fluorescence Detection DW2 Multiple dyes need multiple detection windows. Problem: Emission of one dye is detected in multiple DWs.

13 Signal Flow Cytometry Facility Fluorescence Detection DW2 DW2

14 Signal Flow Cytometry Facility Fluorescence Detection DW2 Spillover DW2

15 Detection Summary DW2 We have to use optical filters because our detectors are colorblind. Each filter window only detects a part of the total fluorophore emission. Because emission spectra are broad, a single detection window can receive light from multiple fluorophores. -> Spillover

16 Signal Flow Cytometry Facility Spillover 1 Signal arriving on the detector is the correct signal + spillover: 1 S = + 2 S DW2 = + 2 DW2

17 Signal Signal Signal Flow Cytometry Facility Spillover 1 Spillover is a constant quantity: Can be determined from single stain controls DW2 DW2 DW2 s FITC = S 2 FITC s PE = S 1 PE

18 Signal Flow Cytometry Facility Spillover 1 With spillover constants s: S 1 PE = s PE S = + S 1 PE = + s PE 2 = s FITC 2 DW2 2 S DW2 = + = s FITC + 2 equations for 2 unknowns.

19 Signal Flow Cytometry Facility Spillover 1 Using linear algebra to solve the system of equations (much more convenient for computers): S = + s PE S DW2 = s FITC + S S DW2 = 1 s PE s FITC 1 s = 1 s PE s FITC 1 2 s 1 S S DW2 = s 1 s DW2 s 1 S S DW2 =

20 Signal Signal Flow Cytometry Facility Spillover One simple multiplication yields the corrected signals: 1 s 1 S S DW2 = 2 DW2 For s, we only need the single stain controls. Extension to more channels/colors is straightforward. DW2 s 1 = 1/(1 s PE s FITC ) (s PE /(1 s PE s FITC )) (s FITC /(1 s PE s FITC )) 1/(1 s PE s FITC )

21 Signal Flow Cytometry Facility Spillover Background bg 1 bg 2 a 1 a 2 1 What if we also have other contributions to the signal? Auto-fluorescence (a) Other background (bg) S = + s PE + a 1 + bg 1 S DW2 = s FITC + + a 2 + bg 2 2 s 1 S a1 + bg 1 S DW2 a 2 + bg 2 = DW2 Background can be subtracted from signal before compensation to obtain the correct signal.

22 Signal Signal Flow Cytometry Facility Spillover Background 1 bg 2 a 1 subtracted before bg 1 a 2 What if background is not compensation, like its usually the case with autofluorescence? The displayed absolute values will be wrong. 2 DW2 bg 1 bg 2 a 1 a 2 DW2 However, the difference between the compensated stained and compensated unstained population still yields the correct absolute signal.

23 Signal Flow Cytometry Facility Spillover Summary 1 Spillover can be completely compensated for. The only thing we need for compensation are single stain controls for each fluorophore used. Autofluorescence and background can be corrected for after compensation. 2 DW2 s 1 S S DW2 =

24 Practical Example (FlowJo) The difference between the positive and negative population medians is our fluorescent signal. Gates are only defined in main channel.

25 Practical Example (FlowJo) The ratios of the fluorescent signals in all channels for each fluorophore are then used to calculate the spillover matrix (s): Dyes Channels The compensation matrix (S 1 ) is the inverse of the spillover matrix (not shown by FlowJo).

26 A700 CD3 A488 CD32 A700 CD3 Flow Cytometry Facility Practical Example (FlowJo) - comp. + comp. - comp. + comp. PE CD86 APC-H7 CD comp. + comp. - comp. + comp. PE-Cy5.5 CD209 A647 2H2

27 Spillover Limitations 1) can only be as good as the single stain measurements. -> any error in the single stain measurements will propagate to your compensated data. 2) introduces artifacts in populations that would not need full compensation (e.g. unstained and single stained populations). 3) Mathematically, any amount of spillover can be compensated. In practice, the total signal level and the relative signal contributions in one channel become limiting: 1) Linear range of the detector 2) Signal broadening (spillover spreading)

28 1) Single Stain Measurements The ideal single stain control is: As bright as possible -> High signal to noise ratio reduces measurement error. The same thing you measure in your sample (minus all the other dyes) -> Dyes can change their emission properties depending on their environment. Potential problems with single stain cell controls: Sample is precious and you cannot afford to use any for single stain controls. Low antigen expression -> low signal Cells with your marker are rare -> low signal

29 1) Single Stain Measurements Beads as substitute for cells: Bright and low background. Does not consume actual sample. Different environment than cell. Other substitutions: Less expensive antibody with same fluorophore Test first! Less expensive antibody with different fluorophore for same channel Don t do that.

30 2) Artifacts of Without compensation With compensation Green: 50% -> Red Red: 10% -> Green Extreme example By applying compensation to signals that result not from the fluorophores (background, instrument noise, autofluorescence), some amount of distortion is introduced.

31 Signal Measured signal Flow Cytometry Facility 3) Linear Range of the Detectors Oversaturation bg 1 a 1 1 and background correction are linear operations. -> only works within the linear range. Dark counts Linear Range Actual signal -> Adjust gain voltages to have brightest signal within the linear range. -> For correct compensation, gain voltages must not be changed during the experiment.

32 3) Signal Broadening Error in the measurement Photon emission (our signal) is a stochastic process that is described by a Poisson distribution. P k, λ = λk e λ k! Mean: λ Standard deviation: λ Even under perfect conditions, that is the variation in the measurement. The relative variation ( λ/λ) decreases with higher signals, but the absolute variation ( λ) increases.

33 3) Signal Broadening Spillover Spreading Mean: 10 3 SD: 32 Mean: 10 5 SD: 315 Mean: 10 3 SD: 315 Populations shifted (compensated) from high to low intensities take their variation with them. -> Compensated populations have larger variations than an original population at that signal level would have.

34 Limitations Summary Single stain measurements are a compromise between best signal and a control as similar to your sample as possible. Non-fluorophore signals (e.g. autofluorescence) are distorted by compensation. Keep signal within linear range of detectors or compensation cannot be applied. Spillover leads to signal spreading that cannot be corrected for.

35 Summary The main points: Fluorophores have broad emission spectra and are detected in multiple channels (spillover). Spillover can be compensated for with single stain controls. quality depends directly on the quality and suitability of the controls. Spillover leads to signal spreading that cannot be corrected for.

36 Some Reading Material An Introduction to for Multicolor Assays on Digital Flow Cytometers. BD Technical Bulletin Bagwell CB, Adams EG. Fluorescence spectral overlap compensation for any number of flow cytometer parameters. Ann NY Acad Sci in Flow Cytometry. Roederer M. Current Protocols in Cytometry. 2002

Each question may have MULTIPLE correct answers. Select all that are correct.

Each question may have MULTIPLE correct answers. Select all that are correct. Knowledge Assessment Flow Cytometry Workshop, Part 1 April 20, 2015 Each question may have MULTIPLE correct answers. Select all that are correct. 1. Tandem dyes are a. highly stable fluorophores after

More information

Principles of flow cytometry: overview of flow cytometry and its uses for cell analysis and sorting. Shoreline Community College BIOL 288

Principles of flow cytometry: overview of flow cytometry and its uses for cell analysis and sorting. Shoreline Community College BIOL 288 Principles of flow cytometry: overview of flow cytometry and its uses for cell analysis and sorting Shoreline Community College BIOL 288 Flow Cytometry What is Flow Cytometry? Measurement of cells or particles

More information

Anatomy of a flow cytometer

Anatomy of a flow cytometer Anatomy of a flow cytometer Fluidics Optics Electronics Cells in suspension flow in single-file through an illuminated volume where they scatter light and emit fluorescence that is collected, filtered,

More information

Flow Cytometry - The Essentials

Flow Cytometry - The Essentials Flow Cytometry - The Essentials Pocket Guide to Flow Cytometry: 1. Know your Cytometer 2. Understanding Fluorescence and Fluorophores 3. Gating Process 4. Controls 5. Optimization 6. Panel Building 7.

More information

Designing and Implementing a High-Level Multicolor Flow Cytometry Assay. Brent Wood MD PhD Department of Laboratory Medicine University of Washington

Designing and Implementing a High-Level Multicolor Flow Cytometry Assay. Brent Wood MD PhD Department of Laboratory Medicine University of Washington Designing and Implementing a High-Level Multicolor Flow Cytometry Assay Brent Wood MD PhD Department of Laboratory Medicine University of Washington Define Purpose of Assay Most important question What

More information

Selecting Reagents for Multicolor Flow Cytometry

Selecting Reagents for Multicolor Flow Cytometry HotLines Platinum Edition f a l l 0 0 6 Selecting Reagents for Multicolor Flow Cytometry By Holden Maecker and Joe Trotter The availability of flow cytometers capable of detecting 6, 8, and more colors

More information

What you need to know before designing a panel

What you need to know before designing a panel Design What you need to know before designing a panel For Research Use Only. Not for use in diagnostic or therapeutic procedures. Alexa Fluor is a registered trademark of Life Technologies Corporation.

More information

Beyond Colorful Science. A Practical Approach to Setting Up Multicolor Panels

Beyond Colorful Science. A Practical Approach to Setting Up Multicolor Panels Beyond Colorful Science A Practical Approach to Setting Up Multicolor Panels ISCT Paris April 2014 Layout: Multicolor Panel Design Identify the pitfalls Have a look at issues with spectral overlap How

More information

Incorporating New, Bright Fluorochromes into Multicolor Panel Design

Incorporating New, Bright Fluorochromes into Multicolor Panel Design Incorporating New, Bright Fluorochromes into Multicolor Panel Design Maria C. Jaimes, MD Senior Staff Scientist BD Biosciences 23-14684-00 Overview Multicolor flow: successful application prerequisites

More information

Expanding Multicolor Options in Flow Cytometry with Novel Brilliant Violet TM Fluorophores. Carsten Wiethe Scientific Application Manager BioLegend

Expanding Multicolor Options in Flow Cytometry with Novel Brilliant Violet TM Fluorophores. Carsten Wiethe Scientific Application Manager BioLegend Expanding Multicolor Options in Flow Cytometry with Novel Brilliant Violet TM Fluorophores Carsten Wiethe Scientific Application Manager BioLegend Seminar Outline BrilliantViolet TM (BV)fluorophores Instrument

More information

Selecting Fluorochrome Conjugates for Maximum Sensitivity

Selecting Fluorochrome Conjugates for Maximum Sensitivity 2004 Wiley-Liss, Inc. Cytometry Part A 62A:169 173 (2004) Technical Primer Selecting Fluorochrome Conjugates for Maximum Sensitivity Holden T. Maecker,* Tom Frey, Laurel E. Nomura, and Joe Trotter BD Biosciences,

More information

Using Fluorescence Spillover to Advantage

Using Fluorescence Spillover to Advantage Using Fluorescence Spillover to Advantage Boston June 9, 2011 Compensation: Why and When is it Necessary? Clare Rogers, Marketing and Applications Accuri Cytometers, Inc. Ann Arbor/St.Ives 1 June 11 Summary

More information

AMREP FLOW CYTOMETRY CORE FACILITY. Flow Cytometry Data Analysis Workshop Friday 24 th October

AMREP FLOW CYTOMETRY CORE FACILITY. Flow Cytometry Data Analysis Workshop Friday 24 th October AMREP FLOW CYTOMETRY CORE FACILITY Flow Cytometry Data Analysis Workshop Friday 24 th October Polychromatic Flow Cytometry Data Analysis: Immunophenotyping Lymphocyte sub-populations and targeting their

More information

BASICS OF FLOW CYTOMETRY

BASICS OF FLOW CYTOMETRY BASICS OF FLOW CYTOMETRY AUTHOR: Ana Isabel Vieira APPROVAL: Henrique Veiga Fernandes Ana Sílvia Gonçalves SOP.UCF.002 03-09-2015 Pag. 1/9 Overview Flow: Fluid Cyto: Cell Metry: Measurement Flow cytometry

More information

11/19/2013. Janine Zankl FACS Core Facility 13. November Cellular Parameters. Cellular Parameters. Monocytes. Granulocytes.

11/19/2013. Janine Zankl FACS Core Facility 13. November Cellular Parameters. Cellular Parameters. Monocytes. Granulocytes. DEPARTEMENT BIOZENTRUM Janine Zankl FACS Core Facility 13. November 2013 Cellular Parameters Granulocytes Monocytes Basophils Neutrophils Lymphocytes Eosinophils Cellular Parameters 1 What Is Flow Cytometry?

More information

Principles of Multicolor Panel Design BD. BD, the BD Logo and all other trademarks are property of Becton, Dickinson and Company.

Principles of Multicolor Panel Design BD. BD, the BD Logo and all other trademarks are property of Becton, Dickinson and Company. 1 Principles of Multicolor Panel Design 2 Common Multicolor Applications Intracellular cytokine staining Regulatory T cells (Tregs) Protein phosphorylation (BD Phosflow) Leukemia and lymphoma phenotyping

More information

FACS Core Facility User Group Meeting. 25. April 2018 Janine Bögli

FACS Core Facility User Group Meeting. 25. April 2018 Janine Bögli FACS Core Facility User Group Meeting 25. April 2018 Janine Bögli Overview Basis: Forensic Flow Cytometry by Jennifer Wilshire, PhD, Stemcell Technologies You will suggest the crime in each case 8 Practical

More information

Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis

Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis Bilge Karaçalı, PhD Department of Electrical and Electronics Engineering Izmir Institute of Technology Outline Experimental design

More information

Flow Cytometry Facility The Panel Design Tool

Flow Cytometry Facility The Panel Design Tool The Automated design of multicolor flow cytometry panels. 01.06.2017 Contents The antibody fluorophore panel Parameters & desired properties The What it can do The input needed Input processing Panel creation

More information

A previous ICCS Module entitled Instrument optimization - Adjusting PMT voltages and compensation 1 should be read as a prerequisite to this module.

A previous ICCS Module entitled Instrument optimization - Adjusting PMT voltages and compensation 1 should be read as a prerequisite to this module. Sponsored and reviewed by ICCS Quality and Standards Committee Title: Compensation Tips for Beckman Coulter 10-Color Navios Platform Written by: Salima Janmohamed-Anastasakis Ph.D., Applications Scientist,

More information

Qdot nanocrystal. wide range of biological investigations, Qdot nanocrystals are powerful complements

Qdot nanocrystal. wide range of biological investigations, Qdot nanocrystals are powerful complements Feature nanocrystal conjugates for flow cytometry Take the easy route to multicolor flow cytometry. With applications across a wide range of biological investigations, nanocrystals are powerful complements

More information

Spherotech, Inc. 1. SPHERO TM Technical Note STN-8 Rev C

Spherotech, Inc. 1. SPHERO TM Technical Note STN-8 Rev C SPHERO TM Technical Note STN- Rev C. 0070 CALIBRATION AND PERFORMANCE TRACKING OF FLOW CYTOMETERS USING SPHERO TM CALIBRATION PARTICLES Introduction The SPHERO TM Calibration Particles are versatile, stable,

More information

INTRODUCTION TO FLOW CYTOMETRY

INTRODUCTION TO FLOW CYTOMETRY DEPARTEMENT BIOZENTRUM INTRODUCTION TO FLOW CYTOMETRY F ACS C ore F acility Janine Zankl FACS Core Facility 3. Dezember 2015, 4pm Cellular Parameters Granulocytes Monocytes Basophils Lymphocytes Neutrophils

More information

Flowcytometry Dirk Pacholsky

Flowcytometry Dirk Pacholsky Flowcytometry Dirk Pacholsky Flowcytometry: Overview 1 2 Overview Flow Cyto Metry Fluid Cell Measurement measuring cell properties of cells in suspension Most Flow Cytometer have Spatially separated Lasers

More information

Understanding Flow Cytometry

Understanding Flow Cytometry Understanding Flow Cytometry The Basic Concepts Maree Bagnara Products Sales Specialist/Account Manager Flow Cytometry PN775136 1 Successful Flow Cytometry data is driven by.. Understanding the Biology

More information

Application Note. Assay Portability on the BD FACSVerse System. Summary. Maria Jaimes, Yibing Wang, Catherine McIntyre, and Dev Mittar

Application Note. Assay Portability on the BD FACSVerse System. Summary. Maria Jaimes, Yibing Wang, Catherine McIntyre, and Dev Mittar September Assay Portability on the BD FACSVerse System Maria Jaimes, Yibing Wang, Catherine McIntyre, and Dev Mittar Contents Summary Introduction 3 Objective 4 Methods 6 Results Discussion Conclusions

More information

Definitions. What functions does a flow cytometer be able to do? The build of the Flow cytometry and sorting. Flow cytometry

Definitions. What functions does a flow cytometer be able to do? The build of the Flow cytometry and sorting. Flow cytometry The build of the Flow cytometry and sorting Flow cytometry Seminar Mónika Tóth 11-12-13.04.2011 Definitions Flow cytometry Process or measurement method can measure discrete properties physical, chemical,

More information

BD Multicolor CompBeads

BD Multicolor CompBeads 4/2015 23-9955-01 IVD BD Multicolor CompBeads 100 compensation setups Catalog No. 644204 BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company. 2015 BD Becton, Dickinson and

More information

ANAT 3231 Cell Biology Lab12 Stem Cell Analysis

ANAT 3231 Cell Biology Lab12 Stem Cell Analysis ANAT 3231 Cell Biology Lab12 Stem Cell Analysis 2 June 2010 Dr Antonio Lee Neuromuscular & Regenera9ve Medicine Unit School of Medical Sciences, UNSW Introduction to Flow Cytometry Contributed by Vittoria

More information

Attune TM Acoustic Focusing Cytometer Training. Manik Punj Attune Training

Attune TM Acoustic Focusing Cytometer Training. Manik Punj Attune Training Attune TM Acoustic Focusing Cytometer Training Manik Punj Attune Training Attune Training Agenda Section 1 An Introduction to Flow Cytometry Section 2 An Introduction to Acoustic Focusing Hydrodynamic

More information

Flow Cytometry. Flow Cytometry Basics Guide

Flow Cytometry. Flow Cytometry Basics Guide Flow Cytometry Flow Cytometry Basics Guide Table of Contents Chapter 1 Chapter 2 Chapter 3 Chapter 4 Chapter 5 Principles of the Flow Cytometer Fluidics System.... 3 Optics and Detection.... 4 Signal and

More information

seminar

seminar seminar 02.26.-28.2013. Flow cytometry Process or measurement method can measure discrete properties physical, chemical, biochemical, biological parameters of separate particles, e.g. biological cells

More information

Widefield Microscopy Bleed-Through

Widefield Microscopy Bleed-Through In widefield microscopy the excitation wavelengths which illuminate the sample, and the emission wavelengths which reach the CCD camera are selected throughout a filter cube. A filter cube consists of

More information

Fluorescence spectroscopy

Fluorescence spectroscopy Fluorescence spectroscopy The light: electromagnetic wave Zoltán Ujfalusi Biophysics seminar Dept. of Biophysics, University of Pécs 14-16 February 2011 Luminescence: light is not generated by high temperatures!!!

More information

Application Information Bulletin: DuraClone IM Tubes Compensation setup for high content DuraClone reagents

Application Information Bulletin: DuraClone IM Tubes Compensation setup for high content DuraClone reagents Application Information Bulletin: DuraClone IM Tubes Compensation setup for high content DuraClone reagents Compensation setup for high content DuraClone reagents INTRODUCTION High content flow cytometry

More information

Michael Waring Associate Director Ragon Institute Imaging Core Facility Cambridge, Massachusetts, USA

Michael Waring Associate Director Ragon Institute Imaging Core Facility Cambridge, Massachusetts, USA Michael Waring Associate Director Ragon Institute Imaging Core Facility Cambridge, Massachusetts, USA Ragon Institute Imaging Core Maintain and oversee operation of the Institute s flow cytometry equipment

More information

The analysis of fluorescence microscopy images for FRET detection

The analysis of fluorescence microscopy images for FRET detection The analysis of fluorescence microscopy images for FRET detection Ela Claridge, Dale J. Powner and Michael J.O. Wakelam School of Computer Science, The University of Birmingham B5 2TT Institute for Cancer

More information

High-throughput automation with the Attune NxT Autosampler: consistent results across all wells and across plates

High-throughput automation with the Attune NxT Autosampler: consistent results across all wells and across plates APPLICATION NOTE Attune NxT Flow Cytometer with Autosampler High-throughput automation with the Attune NxT Autosampler: consistent results across all wells and across plates Introduction The emerging field

More information

Fluorescence spectroscopy

Fluorescence spectroscopy Fluorescence spectroscopy The light: electromagnetic wave Tamás Huber Biophysics seminar Dept. of Biophysics, University of Pécs 05-06. February 2014. 1 Luminescence: light emission of an excited system.

More information

Spherotech, Inc. 1. SPHERO TM Technical Note STN-9 Rev D

Spherotech, Inc. 1. SPHERO TM Technical Note STN-9 Rev D SPHERO TM Technical Note STN-9 Rev D. 00 MEASURING MOLECULES OF EQUIVALENT FLUOROCHROME (MEF) USING SPHERO TM RAINBOW AND ULTRA RAINBOW CALIBRATION PARTICLES Introduction The Molecules of Equivalent Fluorochrome

More information

SPHERO TM Calibration Particles

SPHERO TM Calibration Particles SPHERO TM Calibration Particles SPHERO TM Calibration Particles are designed for routine calibration of flow cytometers. They are used extensively by many laboratories for QC and long term performance

More information

Designing and Validating a Multicolor Flow Cytometry Assay. Brent Wood MD PhD Department of Laboratory Medicine University of Washington

Designing and Validating a Multicolor Flow Cytometry Assay. Brent Wood MD PhD Department of Laboratory Medicine University of Washington Designing and Validating a Multicolor Flow Cytometry Assay Brent Wood MD PhD Department of Laboratory Medicine University of Washington Specimen Handling Sample Requirements 5 ml Peripheral blood (EDTA,

More information

Attune NxT Acoustic Focusing Cytometer The next generation in acoustic cytometry

Attune NxT Acoustic Focusing Cytometer The next generation in acoustic cytometry Attune NxT Acoustic Focusing Cytometer The next generation in acoustic cytometry Maybelline Giam Field Application Scientist The world leader in serving science Attune NxT Flow Cytometer Attune NxT Acoustic

More information

Visualization of digital data Interpretation of the visual

Visualization of digital data Interpretation of the visual Technical aspects of 8-color flow cytometry in the diagnosis and classification of hematopoietic malignancies Tomas Kalina!"#$%&'()*+,&$'+-./(0 *1 (2#34%-.(56(7&1+3+*&/( 8$#94&/(!:&3"(;&

More information

Challenges for Flow Cytometry in Regulated Bioanalysis

Challenges for Flow Cytometry in Regulated Bioanalysis Challenges for Flow Cytometry in Regulated Bioanalysis Minesh Patel Merck Millipore Discovery & Development Solutions. Oxford, UK. Overview Flow cytometry principles Current uses and regulatory environments

More information

Fluorescence spectroscopy

Fluorescence spectroscopy Fluorescence spectroscopy The light: electromagnetic wave Tamás Huber Biophysics seminar Dept. of Biophysics, University of Pécs 05-07. February 2013. Luminescence: light emission of an excited system.

More information

Spherotech, Inc Irma Lee Circle, Unit 101, Lake Forest, Illinois

Spherotech, Inc Irma Lee Circle, Unit 101, Lake Forest, Illinois 78 Irma Lee Circle, Unit, Lake Forest, Illinois 00 SPHERO TM Technical Note STN-9 Rev D. 0 MEASURING MOLECULES OF EQUIVALENT FLUOROCHROME (MEF) USING SPHERO TM RAINBOW AND ULTRA RAINBOW CALIBRATION PARTICLES

More information

Designing and executing a successful flow cytometry experiment

Designing and executing a successful flow cytometry experiment Designing and executing a successful flow cytometry experiment 1x1 0 5 Orange Fluorescent Protein 10000 1000 100 10 0 1000 2000 3000 4000 Forward Scatter This presentation is based on true events, however,

More information

Flow cytometry within reach.

Flow cytometry within reach. Affordable, but more than capable. The Accuri C6 Flow Cytometer is flow cytometry at its best. Its affordable price makes it accessible beyond compare. The C6 has a foot print small enough to fit most

More information

FLUORESCENCE. Matyas Molnar and Dirk Pacholsky

FLUORESCENCE. Matyas Molnar and Dirk Pacholsky FLUORESCENCE Matyas Molnar and Dirk Pacholsky 1 Information This lecture contains images and information from the following internet homepages http://micro.magnet.fsu.edu/primer/index.html http://www.microscopyu.com/

More information

Introduction to Flow Cytometry. -- BD FACSCanto II TM. Daisy Kuo Application Specialist BDBiosciences

Introduction to Flow Cytometry. -- BD FACSCanto II TM. Daisy Kuo Application Specialist   BDBiosciences Introduction to Flow Cytometry -- BD FACSCanto II TM Daisy Kuo Application Specialist E-mail: daisy_kuo@bd.com BDBiosciences Outline Basic Concept of Flow Cytometry FACSCanto II System Introduction Application

More information

Multiplexed 3D FRET imaging in deep tissue of live embryos Ming Zhao, Xiaoyang Wan, Yu Li, Weibin Zhou and Leilei Peng

Multiplexed 3D FRET imaging in deep tissue of live embryos Ming Zhao, Xiaoyang Wan, Yu Li, Weibin Zhou and Leilei Peng Scientific Reports Multiplexed 3D FRET imaging in deep tissue of live embryos Ming Zhao, Xiaoyang Wan, Yu Li, Weibin Zhou and Leilei Peng 1 Supplementary figures and notes Supplementary Figure S1 Volumetric

More information

Sapphire. Biomolecular Imager THE NEXT GENERATION OF LASER-BASED IMAGING

Sapphire. Biomolecular Imager THE NEXT GENERATION OF LASER-BASED IMAGING Sapphire Biomolecular Imager THE NEXT GENERATION OF LASER-BASED IMAGING Breakthrough image capture and analysis The Sapphire Biomolecular Imager is a next generation laser scanning system that provides

More information

determine optimum instrument settings for their own instruments and establish their own daily values.

determine optimum instrument settings for their own instruments and establish their own daily values. PC7 (770/488) SETUP KIT 6607121 PN 4299504-C FLOW CYTOMETER ALIGNMENT VERIFICATION FLUOROSPHERES FLOW CYTOMETER DETECTOR STANDARDIZATION FLUOROSPHERES INTENDED USE For Research Use Only. Not for use in

More information

a Beckman Coulter Life Sciences: White Paper

a Beckman Coulter Life Sciences: White Paper a Beckman Coulter Life Sciences: White Paper Long Term Stabilization of Tandem Dyes for Use in High Content, Multi Variant Flow Cytometry Authors: Snehita Sattiraju 1, Tewfik Miloud 2, Neha Girish 1, Murthy

More information

SUPPORTING INFORMATION

SUPPORTING INFORMATION Electronic Supplementary Material (ESI) for Dalton Transactions. This journal is The Royal Society of Chemistry 2015 Terbium-Based Time-Gated Förster Resonance Energy Transfer Imaging for Evaluating Protein-Protein

More information

Dino-Lite knowledge & education. Fluorescence Microscopes

Dino-Lite knowledge & education. Fluorescence Microscopes Dino-Lite knowledge & education Fluorescence Microscopes Dino-Lite Fluorescence models Smallest fluorescence microscope in the world Revolution to biomedical and educational applications Flexible Easy

More information

FLOW CYTOMETRY. CyAn ADP. Analyzer

FLOW CYTOMETRY. CyAn ADP. Analyzer FLOW CYTOMETRY CyAn ADP Analyzer Experience the Power of the CyAn ADP and its optimal performance The Power of Detection The Power of Speed The Power of Ease The CyAn ADP Analyzer is the next step in Advanced

More information

Welcome to the CORES Flow Cytometry Sorter Facility

Welcome to the CORES Flow Cytometry Sorter Facility Welcome to the CORES Flow Cytometry Sorter Facility The FACSAria I and FACSARIA III are twin laser high speed sorters capable of analyzing cells based on 7-8 distinct fluorescent properties (besides FSC

More information

SYNGENE APPLICATION TRAINING GUIDE

SYNGENE APPLICATION TRAINING GUIDE SYNGENE APPLICATION TRAINING GUIDE Contents Page Lighting and Filter combinations 3 Examples.. 6 Chemiluminescence versus Fluorescence Western blot detection. 8 Film versus digital imaging Chemiluminescence

More information

NovoCyte Flow Cytometer

NovoCyte Flow Cytometer NovoCyte Flow Cytometer The Flow Cytometer for Everyone 2 Experience the NovoCyte Advantage Focus on advancing your research. Let the flow cytometer do the rest. NovoCyte Flow Cytometer High Performance

More information

Institute of Science & Technology in Medicine Control of Substance Hazardous to Health COSHH assessment

Institute of Science & Technology in Medicine Control of Substance Hazardous to Health COSHH assessment Institute of Science & Technology in Medicine Control of Substance Hazardous to Health COSHH assessment Title of Experiment/Procedure: Cell counting and viability using FACS, counting beads and propidium

More information

Multicolor flow cytometry analysis of human pluripotent stem cell cultures

Multicolor flow cytometry analysis of human pluripotent stem cell cultures Multicolor flow cytometry analysis of human pluripotent stem cell cultures ackground Thorough characterization of pluripotency is mandatory during the establishment of newly derived human embryonic or

More information

Detecting human circulating endothelial cells using the Attune Acoustic Focusing Cytometer

Detecting human circulating endothelial cells using the Attune Acoustic Focusing Cytometer APPLICATION NOTE Attune Acoustic Focusing Cytometer Detecting human circulating endothelial cells using the Attune Acoustic Focusing Cytometer Circulating endothelial cells (CECs) are mature cells shed

More information

SA3800 Spectral Analyzer

SA3800 Spectral Analyzer SA38 Spectral Analyzer Sony Biotechnology Inc. SA38 Spectral Analyzer The Sony SA38 spectral cell analyzer incorporates advanced electronics, patented optical technologies and many points of automation

More information

Introduc)on to Flow Cytometry. Stephen De Rosa April 20, 2015

Introduc)on to Flow Cytometry. Stephen De Rosa April 20, 2015 Introduc)on to Flow Cytometry Stephen De Rosa April 20, 2015 Instrumenta)on Outline Fluidics Lasers, light detectors and op)cs Fluorescence Spectra Mul)- parameter flow cytometry Flow data display Basic

More information

a Beckman Coulter Life Sciences: White Paper

a Beckman Coulter Life Sciences: White Paper a Beckman Coulter Life Sciences: White Paper DuraClone improves standardization of compensation workflow for high content multicolor applications by flow cytometry Authors: Neha Girish 2, Sudharsan Sathyamurthy

More information

Review of techniques in imaging and cytometry. Peter Meeus Onze Lieve Vrouw Ziekenhuis, Aalst 7 may 2009, SCK/CEN Mol

Review of techniques in imaging and cytometry. Peter Meeus Onze Lieve Vrouw Ziekenhuis, Aalst 7 may 2009, SCK/CEN Mol Review of techniques in imaging and cytometry Peter Meeus Onze Lieve Vrouw Ziekenhuis, Aalst 7 may 2009, SCK/CEN Mol Cytometry: The counting and measuring of cells,... Mosby's Medical Dictionary, 8th edition.

More information

Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement

Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Andreas Spittler, MD, Associate Professor for Pathophysiology, Medical University

More information

Localization Microscopy

Localization Microscopy Localization Microscopy Theory, Sample Prep & Practical Considerations Patrina Pellett & Ann McEvoy Applications Scientist GE Healthcare, Cell Technologies May 27 th, 2015 Localization Microscopy Talk

More information

Your Research, Revolutionized

Your Research, Revolutionized Your Research, Revolutionized Drive Your Research Forward Your research needs are evolving and with the CytoFLEX flow cytometer you ll see just how far your data can take you. CytoFLEX has the advanced

More information

Best practices in panel design to optimize the isolation of cells of interest

Best practices in panel design to optimize the isolation of cells of interest Sort Best practices in panel design to optimize the isolation of cells of interest For Research Use Only. Not for use in diagnostic or therapeutic procedures. Alexa Fluor is a registered trademark of Life

More information

Concept review: Fluorescence

Concept review: Fluorescence 16 Concept review: Fluorescence Some definitions: Chromophore. The structural feature of a molecule responsible for the absorption of UV or visible light. Fluorophore. A chromophore that remits an absorbed

More information

efluor Organic Dyes 450/50 BP Fluorescence Intensity Wavelength (nm)

efluor Organic Dyes 450/50 BP Fluorescence Intensity Wavelength (nm) efluor Organic Dyes efluor Organic Dyes Catalog No. Description Clone Application Anti-Mouse efluor 450 Products 48-0042 Anti-mouse CD4 RM4-5 Flow Cytometry 48-0081 Anti-mouse CD8 53-6.7 Flow Cytometry

More information

Ruud Hulspas, Mike Keeney, Ben Hedley and Andrea Illingworth

Ruud Hulspas, Mike Keeney, Ben Hedley and Andrea Illingworth Sponsored and reviewed by ICCS Quality and Standards Committee Title: Quality of Reagents Monoclonal Antibodies Written by: Date: March 6, 2018 Ruud Hulspas, Mike Keeney, Ben Hedley and Andrea Illingworth

More information

Formalization of the MESF Unit of Fluorescence Intensity

Formalization of the MESF Unit of Fluorescence Intensity Cytometry Part B (Clinical Cytometry) 57B:1 6 (2004) Report Formalization of the MESF Unit of Fluorescence Intensity Abe Schwartz, 1 Adolfas K. Gaigalas, 2 Lili Wang, 2 Gerald E. Marti, 3 Robert F. Vogt,

More information

The Basics of Flow Cytometry

The Basics of Flow Cytometry The Basics of Flow Cytometry F ACS C ore F acility Janine Bögli, Biozentrum, 29. January 2018 The functions of the FACS Core Facility Centralization of equipment and expertise Train users Sorter operation

More information

Titration of Fluorochrome-Conjugated Antibodies for Labeling Cell Surface Markers on Live Cells

Titration of Fluorochrome-Conjugated Antibodies for Labeling Cell Surface Markers on Live Cells Titration of Fluorochrome-Conjugated Antibodies for Labeling Cell Surface Markers on Live Cells Ruud Hulspas 1 UNIT 6.29 1 Cytonome/ST, Boston, Massachusetts ABSTRACT Nonspecific antibody binding is best

More information

Quality Control in Flow. Dr David Westerman Head of Haematopathology Peter MacCallum Cancer Centre

Quality Control in Flow. Dr David Westerman Head of Haematopathology Peter MacCallum Cancer Centre Quality Control in Flow Dr David Westerman Head of Haematopathology Peter MacCallum Cancer Centre Aims Quality Assurance Quality Control Literature In house competencies SHOT DATA 1996-2009 Ref: SHOT Annual

More information

Lower spectral overlap. Optimized stain index. Innovative antibodies. Superior multicolor flow cytometry. Violet Laser Reagents: The efluor Solution

Lower spectral overlap. Optimized stain index. Innovative antibodies. Superior multicolor flow cytometry. Violet Laser Reagents: The efluor Solution Lower spectral overlap Optimized stain index Innovative antibodies Superior multicolor flow cytometry Violet Laser Reagents: The efluor Solution www.ebioscience.com ebioscience provides innovative high

More information

Review of techniques in flow cytometry. Peter Meeus Onze-Lieve-Vrouwziekenhuis, Aalst 5 may 2011, SCK/CEN Mol

Review of techniques in flow cytometry. Peter Meeus Onze-Lieve-Vrouwziekenhuis, Aalst 5 may 2011, SCK/CEN Mol Review of techniques in flow cytometry Peter Meeus Onze-Lieve-Vrouwziekenhuis, Aalst 5 may 2011, SCK/CEN Mol Cytometry definitions The counting and measuring of cells,... Mosby's Medical Dictionary, 8th

More information

Bioinstrumentation Light Sources Lasers or LEDs?

Bioinstrumentation Light Sources Lasers or LEDs? Bioinstrumentation Light Sources Lasers or LEDs? A comprehensive analysis of all the factors involved in designing and building life sciences instrumentation reveals that lasers provide superior performance

More information

Lab 1: Ensemble Fluorescence Basics

Lab 1: Ensemble Fluorescence Basics Lab 1: Ensemble Fluorescence Basics This laboratory module is divided into two sections. The first one is on organic fluorophores, and the second one is on ensemble measurement of FRET (Fluorescence Resonance

More information

Special Techniques 1. Mark Scott FILM Facility

Special Techniques 1. Mark Scott FILM Facility Special Techniques 1 Mark Scott FILM Facility SPECIAL TECHNIQUES Multi-photon microscopy Second Harmonic Generation FRAP FRET FLIM In-vivo imaging TWO-PHOTON MICROSCOPY Alternative to confocal and deconvolution

More information

Flow Cytometry. Flow Cytometry Basics Guide

Flow Cytometry. Flow Cytometry Basics Guide Flow Cytometry Flow Cytometry Basics Guide Table of Contents Chapter 1 Chapter 2 Chapter 3 Chapter 4 Principles of the Flow Cytometer Fluidics System.... 3 Optics and Detection.... 4 Signal and Pulse

More information

Reagent Titration for the ICS Assay

Reagent Titration for the ICS Assay Purpose This standard operating procedure (SOP) describes how to titrate new lots of fluorochromeconjugated antibody reagents and the cell viability marker for use in the intracellular cytokine staining

More information

Boundary-breaking acoustic focusing cytometry

Boundary-breaking acoustic focusing cytometry Boundary-breaking acoustic focusing cytometry Introducing the Attune NxT Acoustic Focusing Cytometer a high-performance system that s flexible enough for any lab One of the main projects in my laboratory

More information

Imaging facilities at WUR

Imaging facilities at WUR Imaging facilities at WUR Advanced light microscopy facilities at Wageningen UR Programme Thursday 13 June 2013 Lunch meeting organized by Cat-Agro Food 12.00 Welcome and sandwich lunch 12.10 Introduction

More information

PALM/STORM, BALM, STED

PALM/STORM, BALM, STED PALM/STORM, BALM, STED Last class 2-photon Intro to PALM/STORM Cyanine dyes/dronpa This class Finish localization super-res BALM STED Localization microscopy Intensity Bins = pixels xx 2 = ss2 + aa 2 /12

More information

Introduction to. BD FACSAria TM Cell Sorter. Flow = Fluid Cyto = Cell Metry = Measurement

Introduction to. BD FACSAria TM Cell Sorter. Flow = Fluid Cyto = Cell Metry = Measurement What is Flow Cytometry? Introduction to BD FACSAria TM Cell Sorter Flow = Fluid Cyto = Cell Metry = Measurement BD Biosciences Application Specialist 產品應用專員 Daisy Kuo 郭正佼 A variety of measurements are

More information

More on fluorescence

More on fluorescence More on fluorescence Last class Fluorescence Absorption emission Jablonski diagrams This class More on fluorescence Common fluorophores Jablonski diagrams to spectra Properties of fluorophores Excitation

More information

Design for Manufacturability (DFM) in the Life Sciences

Design for Manufacturability (DFM) in the Life Sciences T E C H N I C A L N O T E Design for Manufacturability (DFM) in the Life Sciences Fluorescence Spectroscopy Product Platform Realized with TracePro TM Suite of Opto-Mechanical Design Software Tools Authors:

More information

Flow Cytometry Immune Activation SOP

Flow Cytometry Immune Activation SOP Flow Cytometry Immune Activation SOP Purpose This SOP standardizes the procedure for measuring immune activation of T cells using flow cytometry in ACTG Immunology Laboratories. Materials 1. 12x75mm flow

More information

Time-resolved Measurements Using the Agilent Cary Eclipse Fluorescence Spectrophotometer A Versatile Instrument for Accurate Measurements

Time-resolved Measurements Using the Agilent Cary Eclipse Fluorescence Spectrophotometer A Versatile Instrument for Accurate Measurements Time-resolved Measurements Using the Agilent Cary Eclipse Fluorescence Spectrophotometer A Versatile Instrument for Accurate Measurements Technical Overview Authors Dr. Fabian Zieschang, Katherine MacNamara,

More information

TECHNICAL BULLETIN. QUANTUM FLUORESCENCE KITS FOR MESF UNITS OF FITC Product Numbers QMF-2 AND QMF-10 Storage Temperature 2-8 C.

TECHNICAL BULLETIN. QUANTUM FLUORESCENCE KITS FOR MESF UNITS OF FITC Product Numbers QMF-2 AND QMF-10 Storage Temperature 2-8 C. QUANTUM FLUORESCENCE KITS FOR MESF UNITS OF FITC Product Numbers QMF-2 AND QMF-10 Storage Temperature 2-8 C. Do Not Freeze TECHNICAL BULLETIN Product Description Quantum Fluorescence Kits for MESF Units

More information

Spectral Separation of Multifluorescence Labels with the LSM 510 META

Spectral Separation of Multifluorescence Labels with the LSM 510 META Microscopy from Carl Zeiss Spectral Separation of Multifluorescence Labels with the LSM 510 META Indians living in the South American rain forest can distinguish between almost 200 hues of green in their

More information

Introduction to Flow Cytometry: A Learning Guide

Introduction to Flow Cytometry: A Learning Guide 11-11032-03 rev. A December 2002 Introduction to Flow Cytometry: A Learning Guide Introduction to Flow Cytometry: A Learning Guide Copyright 2002 Becton, Dickinson and Company. All rights reserved. No

More information

SPHERO TM Fluorescent Particles

SPHERO TM Fluorescent Particles The SPHERO TM fluorescent microparticles are prepared by either staining polystyrene particles with a fluorophore solution or by polymerizing a fluorophore in styrene in the presence of polystyrene core

More information

Green Fiber Lasers: An Alternative to Traditional DPSS Green Lasers for Flow Cytometry

Green Fiber Lasers: An Alternative to Traditional DPSS Green Lasers for Flow Cytometry Technical Note Green Fiber Lasers: An Alternative to Traditional DPSS Green Lasers for Flow Cytometry William G. Telford, 1 * Sergey A. Babin, 2 Serge V. Khorev, 2 Stephen H. Rowe 2 1 Experimental Transplantation

More information

Theremino DNA Meter. Fluorometer for DNA Measurements. Ana Rodriguez (MUSE) - Lodovico Lappetito (Theremino)

Theremino DNA Meter. Fluorometer for DNA Measurements. Ana Rodriguez (MUSE) - Lodovico Lappetito (Theremino) Theremino DNA Meter Fluorometer for DNA Measurements Ana Rodriguez (MUSE) - Lodovico Lappetito (Theremino) Theremino_DNAMeter_Hardware_ENG - 11/01/2016 Pag. 1 Table of Contents Principle of Operation...

More information