How is Sterilization Validation Performed by the Medical Device Manufacturer?

Size: px
Start display at page:

Download "How is Sterilization Validation Performed by the Medical Device Manufacturer?"

Transcription

1 How is Sterilization Validation Performed by the Medical Device Manufacturer? Rod Parker, Ph.D. Sr. Principal Scientist Stryker Instruments Division May 9, 2016

2 DISCLOSURE I am an employee of Stryker Corporation. I am involved with the manufacture of medical devices. No product preferences or names are listed or recommended by this presentation. No compensation has been received for this presentation or for travel to and from the seminar. All opinions are those of the presenter.

3 Topics of Interest So how do manufacturers validate their surgical equipment? o Cleaning Validations o Chemicals o Sterilization Validations

4 Who s driving the boat? How bad is the storm? Do you believe the Weatherman or what you see.

5 What does a storm have to do with sterilization of medical devices? The picture shows a front moving on the open ocean and the captain can either steer to the clear side and hope to outrun the storm, or turn into the storm and ride out the rough seas. In the market today with the current focus on healthcare acquired infections and antibiotic resistant micro-organisms, we at the a similar point in directing what is required of reprocessing medical devices. Turning to beat the storm may only work for the short run Riding it out will involve full cooperation of all crew members to insure the ship stay safe. But often.

6 We think we are playing the same game? But we are using the same board but have different strategies.

7 Definitions Sterilization Validated process used to render a product free from viable microorganisms. NOTE: In a sterilization process, the nature of microbiological inactivation is described by an exponential function. Therefore, the presence of a viable microorganism on any individual item can be expressed in terms of probability. While this probability can be reduced to a very low number, it can never be reduced to zero. See also sterility assurance level.

8 Definitions Validation Documented procedure for obtaining, recording, and interpreting the results required to establish that a process will consistently yield product complying with predetermined specifications. Verification (User) Documented procedures, performed in the user environment, for obtaining, recording, and interpreting the results required to establish that predetermined specifications have been met.

9 What Sterilization Really Looks Like. Before Gas Sterilization After Gas Sterilization

10 Challenge Devices Pure culture Spores Spores coated with 5% serum and hard water

11 The process starts with cleaning the used devices. We hear repeatedly If it is not clean, it can not be sterilized I would expand on that to state: If the level of bioburden on the device is greater than expected from the device s cleaning validation, then the ability to reach the level of sterilization expected from the sterilization validation on the device is not assured. This results in a device that is not at the sterility assurance level statistically required by the definition of sterilization. The manufacturer s process must then, first, know the expected level of bioburden resulting from the outcome of cleaning.

12 Initial Cleaning Evaluation Product information provided by engineer o Intended clinical application o Product design (i.e. diameter and length of lumens) o Materials of construction o Disassembly instructions o Cleaning methods Conduct product adoption or cleaning validation

13 Cleaning Validation Protocol (Example) Test Organisms (population of at least 1 x 10 6 ) Geobacillus stearothermophilus - spore) Escherichia coli, Staphylococcus aureus, Pseudomonas aeruginosa vegetative bacteria) Inoculation Method Simulated and predetermined most challenging locations Digital Images Device prior to inoculation Device post inoculation illustrating most challenging inoculation site Automated washer configuration

14 Cleaning Validation Protocol (Example) Test Assays Bioburden (microbiological log reduction) Uses the known most resistant spore for the sterilization process considered Carbohydrate (tissue simulation) Used for orthopedics since bone is higher physiologically in carbohydrate than protein or hemoglobin soil markers Protein (blood simulation) Used since protein is the marker most often associated with potential of disease transmission not related to infection.

15 Digital Image Illustrations

16 Digital Image Illustrations

17 Digital Image Illustrations

18

19 Digital Image Illustrations

20 Manual Cleaning Methods Gross removal of soil Cleaning of devices with a cleaner and suitable equipment (i.e. brushes, syringes) Rinsing of devices using filter/deionized water Drying of devices with (low lint) soft cloth and/or filtered pressurized air Visual inspection

21 Automated Cleaning Methods Gross removal of soil Arrangement of devices in washer basket Washer Disinfector Cycle (i.e. Steris 444) Pre-rinse for 2 minutes - cold water Enzyme wash 2 minutes - hot water Wash 2 minutes - hot tap water (60 C set point) Rinse for 2 minutes - hot tap water (60 C set point) Dry for 15 minutes at 115 C Visual inspection of devices

22 Current testing format

23 Digital Image Illustrations

24 Automatic cleaning in washer disinfector:

25 Acceptance Criteria (Example) Bioburden >= 3 log reduction (1 log = 900,000, 2 log + 90,000, 3 log + 9,000 so 3 log reduction is 999,000 or less that 1,000 remains, prior to sterilization.) Carbohydrate Lower than the limit of quantitation (<1.1 µg/ml), less than we can detect on sample. Protein Lower than the limit of quantitation (<6.4 µg/ml) less that we can find on the sample.

26 Levels of Cleanliness (cont.) Disinfection manual for personnel safety in handling o Indiscriminate use of any disinfectant can have fixative properties on critical surgical instruments and may be counter to the individual disinfectants claims. Thermal Disinfection using the A0 concept in automated washers o Must have sufficient rinsing to prevent detergent build-up resulting in potential corrosive actions

27 Recommendations, Decisions, and Options. Use pre-cleaning and moist environments to maintain loosen soil. Use enzymes for devices that can be soaked. Use detergents with high rinsing formulas. Use detergents when cleaning non-soaked devices. Rinse thoroughly with high quality water. Reduce complex chemistries whenever possible. One product for all device is not best practice. If you use a disinfectant remember to rinse it off before further use or processing.

28 The Sterilization Validation Process As the process follows: o Contamination (simulated justified worst case) o Exposure to the minimum cycle time at the minimum sterilization temperature. o For most processes the overkill method is used. o this involved the exposure of the worst case contamination to half the minimum cycle time. Subsequently in a linear model this validates a two time log reduction.

29 Linear Log Reduction Illustrations

30 Biological Indicators (Many Types) Typical for hospital use. Various types for industry. These come as purchased products These come in many sizes and materials.

31 Biological Indicator Placement

32 Hard to reach Worse-Case testing.

33 Testing - Biological Indicator Placement Table

34 Set up

35 Functional Testing All products tested to the new product quality inspection procedure (this is the procedure for the device on the manufacturing floor for as built new). o After lifetime sterilization exposures (i.e cycles) o After lifetime washing exposures (i.e cycles) o After all repairs Functional tests are routinely done on the worst case know exposure cycle. For global company releases that sell in France this would be 18 minutes at 134 C or as high as 137 C depending on testing equipment.

36 Other issues in reprocessing Dry Time (not a part of sterility but a part of sterility maintenance these times are not an extended cycle) Immediate Use Sterilization (Flash) Plastic Weight Moisture Cleaner Residues - Rust inhibitors (rust present?), lubricants (oils?), enzymes (proteins?), concentrates (too much or too little?)

37 Other methods of reprocessing Ethylene Oxide (EO) Vaporized Hydrogen Peroxide processes Dry heat Use the same overkill type and half cycle approach but may vary target organism dependent on method.

38 Terminal Sterilization - SUD Radiation Gamma, E-beam Ethylene Oxide (EO) All products tested statistically to 1 in a million o (Label means 10-6 validated.) Tested products fully submerged in test media.

39 Everything in the jar - covered or contacted. Test is 14 days - visual turbidity is a positive.

40 Conclusion Every product or family of products has a methodology to which it is validated for cleaning and sterilization. All chemicals used for have consequences. Water is a powerful positive and negative. Instructions are getting better but slowly and maybe not easier. Communication is still the key to success. People remain the focus. Patient s first, but include users, processors, designers, and regulators.

41 Questions? (269) North Break wall Lighthouse Ludington, Michigan USA (my hometown)

applying a unique combination of disciplines consulting, preclinical, toxicology,

applying a unique combination of disciplines consulting, preclinical, toxicology, NAMSA is a Medical Research Organization (MRO), accelerating product development through integrated laboratory, clinical and consulting services. Driven by our regulatory expertise, NAMSA's MRO Approach

More information

STERILIZATION VALIDATION IN THIS SECTION REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES. Radiation Sterilization Validation

STERILIZATION VALIDATION IN THIS SECTION REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES. Radiation Sterilization Validation STERILIZATION VALIDATION All sterilization processes require validation of the efficacy and reproducibility of the process. Depending on the type of sterilization, this may be accomplished by partial,

More information

Loaner Program Guidance

Loaner Program Guidance Loaner Program Guidance This document is valid only on the date printed. If unsure of the print date, please re-print to ensure use of the latest revision of this document (available at www.depuysynthes.com).

More information

Monitoring Flexible Endoscope Channels to Assure Cleaning

Monitoring Flexible Endoscope Channels to Assure Cleaning Monitoring Flexible Endoscope Channels to Assure Cleaning Dr. Michelle J. Alfa, Ph.D., FCCM Medical Director, Clinical Microbiology, Diagnostic Services of Manitoba, Winnipeg, Canada Disclaimers: Dr. M.

More information

Loaner Programme Guidance

Loaner Programme Guidance Loaner Programme Guidance This document is valid only on the date of printing. In case of doubt about the date of printing, please re-print to ensure that the latest revised version of this document is

More information

Loaner Programme Guidance

Loaner Programme Guidance Loaner Programme Guidance This document is valid only on the date of printing. In case of doubt about the date of printing, please re-print to ensure that the latest revised version of this document is

More information

Validation of cleaning and sterilization of reusable medical devices

Validation of cleaning and sterilization of reusable medical devices Synergy makes sence Validation of cleaning and sterilization of reusable medical devices MG-FSI72-119 Last revision: June 2012 Aurélien BIGNON 5, Chemin du Catupolan - 69120 Vaulx en Velin - France - Tel.

More information

CDWA3 / CDWA4 CDWAH. Page 4. BASKET ZONE: water pressure, spray system performance, detergent chemistry and concentration, temperature.

CDWA3 / CDWA4 CDWAH. Page 4. BASKET ZONE: water pressure, spray system performance, detergent chemistry and concentration, temperature. Rev. 2 March 2017 Control of cleaning and washing processes of medical instruments and materials is highly important since its outcome affects the success of subsequent disinfection and/or sterilization

More information

Low Temperature Processing. An Update on Guidelines, Standards and Requirements

Low Temperature Processing. An Update on Guidelines, Standards and Requirements Low Temperature Processing An Update on Guidelines, Standards and Requirements Continuing Education Contact Hours Participants must complete the entire presentation/seminar to achieve successful completion

More information

Technical Data Monograph. The Instrument Protection Properties of Prolystica 2X Concentrate Neutral Detergent Versus Other Neutral Detergent Products

Technical Data Monograph. The Instrument Protection Properties of Prolystica 2X Concentrate Neutral Detergent Versus Other Neutral Detergent Products Technical Data Monograph The Instrument Protection Properties of Prolystica 2X Concentrate Neutral Detergent Versus Other Neutral Detergent Products Table of Contents 1. Background.....................................................1

More information

Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy!

Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy! Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy! Dr. Michelle J. Alfa, Ph.D., FCCM Medical Director, Clinical Microbiology, Diagnostic Services of Manitoba, Winnipeg, Canada Disclaimers:

More information

Ideal Sterilization Method. Sterilization. Sterilization Practices in Healthcare Facilities

Ideal Sterilization Method. Sterilization. Sterilization Practices in Healthcare Facilities Sterilization Practices in Healthcare Facilities Virtual Tour of Central Processing William A. Rutala, Ph.D., M.P.H. Director, Hospital Epidemiology, Occupational Health and Safety Program, UNC Health

More information

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces HemoCheck TM Sample Policy SUBJECT: Detection of blood residue on various surfaces DEPARTMENT: Central Service APPROVED BY: EFFECTIVE: REVISED: 05/2018 PURPOSE: To test for detection of blood residue on

More information

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces HemoCheck TM Sample Policy SUBJECT: Detection of blood residue on various surfaces DEPARTMENT: Central Service APPROVED BY: EFFECTIVE: REVISED: 07/2016 PURPOSE: To test for detection of blood residue on

More information

Workshop on Medical Device Cleanliness: How Clean is Clean Enough? Sponsored by ASTM Committee F04 on Medical and Surgical Materials and Devices

Workshop on Medical Device Cleanliness: How Clean is Clean Enough? Sponsored by ASTM Committee F04 on Medical and Surgical Materials and Devices Breaking the Myth Case Medical Page 1 Our Position Caustic cleaners are unnecessary for instrument processing and present significant hazards to health and safety Properly formulated ph neutral instrument

More information

THE TERMS VALIDATION AND VERIFICATION ARE FREQUENTLY

THE TERMS VALIDATION AND VERIFICATION ARE FREQUENTLY Lesson No. CRCST 162 (Technical Continuing Education - TCE) Sponsored by: by Susan Klacik, BS, CRCST, ACE, CIS, FCS IAHCSMM Clinical Educator Validation and Verification LEARNING OBJECTIVES 1. Describe

More information

Manner of Assessment Code (check all that apply) & Surveyor Notes. Yes N/A. Yes. Yes N/A N/A. Yes N/A. N/A Yes N/A

Manner of Assessment Code (check all that apply) & Surveyor Notes. Yes N/A. Yes. Yes N/A N/A. Yes N/A. N/A Yes N/A Section 4. L Reprocessing of Semi-Critical Equipment Elements to be assessed Manner of Assessment Code (check all that apply) & Surveyor Notes Manner of Assessment Code (check all that apply) & Surveyor

More information

Bio-Burden Reduction in Biological Laboratories

Bio-Burden Reduction in Biological Laboratories Bio-Burden Reduction in Biological Laboratories In biological laboratories samples worked with can harbor microorganisms that could be pathogens. Work is also done with microorganisms that are known pathogens.

More information

SUBJECT: Detection of Blood residue inside the biopsy channel of a scope

SUBJECT: Detection of Blood residue inside the biopsy channel of a scope SUBJECT: Detection of Blood residue inside the biopsy channel of a scope DEPARTMENT: Central Service/ Endoscopy APPROVED BY: EFFECTIVE: REVISED: 2/2012 PURPOSE: To test for detection of blood residue inside

More information

Endoscope Microbial Surveillance Testing Made Easy

Endoscope Microbial Surveillance Testing Made Easy Endoscope Microbial Surveillance Testing Made Easy Alpa Patel, B.S.RM (NRCM) Senior Scientist Nelson Laboratories Kaumudi Kulkarni, M.S. Manager of Research and Development Healthmark Industries Disclosure

More information

Definitions. BIOL 3702: Chapter 8. Control of Microbes in the Environment. Mechanical Removal Methods. Pattern of Microbial Death

Definitions. BIOL 3702: Chapter 8. Control of Microbes in the Environment. Mechanical Removal Methods. Pattern of Microbial Death Definitions Control of Microbes in the Environment u Sterilization - destruction or removal of all viable organisms from an object or environment (agent = sterilant) u Disinfection - killing, inhibition,

More information

Medical Device Reprocessing Assessment - General (Cleaning, Disinfection, and/or Sterilization)

Medical Device Reprocessing Assessment - General (Cleaning, Disinfection, and/or Sterilization) Medical Device Reprocessing Assessment - General (Cleaning, Disinfection, and/or Sterilization) Date: Office Name: Office Number: Responsible Physician: Assessor: M.1. General Yes No N/A If No Please Comment

More information

Test Method for the Continuous Reduction of Bacterial Contamination on Copper Alloy Surfaces

Test Method for the Continuous Reduction of Bacterial Contamination on Copper Alloy Surfaces Test Method for the Continuous Reduction of Bacterial Contamination on Copper Alloy Surfaces Test Organisms: Staphylococcus aureus (ATCC 6538) Enterobacter aerogenes (ATCC 13048) Pseudomonas aeruginosa

More information

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of Contents Glossary of equivalent standards... ix Committee representation... xi Acknowledgments...xiii Foreword... xiv Introduction: Need for the recommended practice...1 1 Scope...5 1.1 General...5 1.2

More information

Definitions. BIOL 3702: Chapter 8. Control of Microbes in the Environment. Mechanical Removal Methods. Pattern of Microbial Death

Definitions. BIOL 3702: Chapter 8. Control of Microbes in the Environment. Mechanical Removal Methods. Pattern of Microbial Death Definitions Control of Microbes in the Environment u Sterilization - destruction or removal of all viable organisms from an object or environment (agent = sterilant) u Disinfection - killing, inhibition,

More information

Test Method for Efficacy of Copper Alloy Surfaces as a Sanitizer

Test Method for Efficacy of Copper Alloy Surfaces as a Sanitizer Test Method for Efficacy of Copper Alloy Surfaces as a Sanitizer Test Organisms: Staphylococcus aureus (ATCC 6538) Enterobacter aerogenes (ATCC 13048) Pseudomonas aeruginosa (ATCC 15442) Methicillin Resistant

More information

SUBJECT: Detection of Blood (hemoglobin) residue inside the biopsy channel of an endoendoscope

SUBJECT: Detection of Blood (hemoglobin) residue inside the biopsy channel of an endoendoscope SUBJECT: Detection of Blood (hemoglobin) residue inside the biopsy channel of an endoendoscope DEPARTMENT: Central Service/ Endoscopy APPROVED BY: EFFECTIVE: REVISED: 6/2016 PURPOSE: To test for detection

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 14161 First edition 2000-10-01 Sterilization of health care products Biological indicators Guidance for the selection, use and interpretation of results Stérilisation des produits

More information

Objectives. The Perioperative Nurse s Role 6/24/2016. Instrument Processing & Sterilization Chesapeake Perioperative Consortium 2016

Objectives. The Perioperative Nurse s Role 6/24/2016. Instrument Processing & Sterilization Chesapeake Perioperative Consortium 2016 Instrument Processing & Sterilization Chesapeake Perioperative Consortium 2016 Ensuring our patients get clean, sterile instruments every time. Objectives Define the perioperative nurse and surgical tech

More information

Tissue cleaning and sterilization provide an additional level of safety

Tissue cleaning and sterilization provide an additional level of safety Managing biologics Understanding tissue processing Third in a series on managing bone allografts. In the October issue, articles included were Allografts: Overview of the process; and Donor screening:

More information

Lotus. Reprocessing instructions for reusable instrumentation. French/Swiss protocol

Lotus. Reprocessing instructions for reusable instrumentation. French/Swiss protocol Lotus Reprocessing instructions for reusable instrumentation French/Swiss protocol Reprocessing instructions These instructions relate to the Lotus Transducers & Waveguides only. Following use remove the

More information

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of Contents Glossary of equivalent standards... ix Committee representation... xi Acknowledgments...xiii Background on Amendment 1...xiii Foreword... xiv Introduction: Need for the recommended practice...1

More information

One such problem is blood residual inside the channels of a robotic arm.

One such problem is blood residual inside the channels of a robotic arm. SUBJECT: Robotic Arm Testing for residue blood DEPARTMENT: Central Service APPROVED BY: EFFECTIVE: REVISED: 07/2016 PURPOSE: To test for detection of blood residue inside and outside of a robotic arm and

More information

BIOTECH-GERMANDE ENDOSCOPE REPROCESSING WITH BRUSHLESS CLEANING PROCESS CONDITIONS. Lionel PINEAU PhD

BIOTECH-GERMANDE ENDOSCOPE REPROCESSING WITH BRUSHLESS CLEANING PROCESS CONDITIONS. Lionel PINEAU PhD ENDOSCOPE REPROCESSING WITH BRUSHLESS CLEANING PROCESS CONDITIONS Lionel PINEAU PhD CONTENT General considerations Endoscope reprocessing, channel brushing, cleaning claims Description of the 2013 study

More information

Administrative Policies and Procedures. Cross Reference: Date Issued: 2/14 Date Reviewed: 9/14 Date: Revised: 12/14 Attachment: None Page of 1 of 6

Administrative Policies and Procedures. Cross Reference: Date Issued: 2/14 Date Reviewed: 9/14 Date: Revised: 12/14 Attachment: None Page of 1 of 6 Administrative Policies and Procedures Originating Venue: Infection Control Policy No.: IC 2306 Title: Cystoscope Reprocessing Policy & Procedure Cross Reference: Date Issued: 2/14 Date Reviewed: 9/14

More information

Best Practices for Reprocessing Robotic Instruments

Best Practices for Reprocessing Robotic Instruments Best Practices for Reprocessing Robotic Instruments Nupur Jain Manager, Validation Engineering Services Intuitive Surgical, US Agenda 1. Background on robotic instruments 2. Reprocessing procedure for

More information

The Significance of Validation to US CSSD Managers

The Significance of Validation to US CSSD Managers The Significance of Validation to US CSSD Managers Jonathan A. Wilder, Ph.D. President, H & W Technology, LLC/Stericert Co. Managing Director, Quality Processing Resource Group, LLC 1 Introduction This

More information

Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar

Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar MONITORING THE SYSTEM 1E PROCESSOR Q: I want to confirm that there is no need for a Biological Indicator (BI) and that use

More information

Competency Guide: Care and Handling of Rigid Endoscopes

Competency Guide: Care and Handling of Rigid Endoscopes Competency Guide: Care and Handling of Rigid Endoscopes This document details the basic steps to correctly process and maintain a rigid endoscope. This document can be used to perform a competency review

More information

Guidance for Industry

Guidance for Industry Guidance for Industry for the Submission Documentation for Sterilization Process Validation in Applications for Human and Veterinary Drug Products Center for Drug Evaluation and Research (CDER) Center

More information

TEST CODE DEFINITION STERRAD H202. Test Code Test Name Testing Description. Cycle. Efficacy. SE-200 STERRAD 200 Efficacy. Cycle. STD Cycle Efficacy

TEST CODE DEFINITION STERRAD H202. Test Code Test Name Testing Description. Cycle. Efficacy. SE-200 STERRAD 200 Efficacy. Cycle. STD Cycle Efficacy Code Name ing Description MC-100S SE-100S STERRAD 100S STERRAD 100S TEST CODE DEFINITION STERRAD H202 125 Highpower Road. Rochester, NY 14623 (888) -722-1529 www.highpowervtls.com Materials Compatibility

More information

sterility assurance Prove the power and your processes

sterility assurance Prove the power and your processes sterility assurance Prove the power and efficacy of your processes sterility assurance programs: an introduction YOUR CHALLENGES Contamination events cost a lot of time and money. Sterility problems can

More information

Lotus. Reprocessing instructions for reusable instrumentation

Lotus. Reprocessing instructions for reusable instrumentation Lotus Reprocessing instructions for reusable instrumentation Reprocessing instructions These instructions relate to the Lotus Transducers & Waveguides only. Following use remove the Single Use Handpiece

More information

Appendix 2. Revised 510(k) Summary

Appendix 2. Revised 510(k) Summary Appendix 2 Revised 510(k) Summary April 1, 2010 Appendix 2: Page 1 of 11 510(k) Summary For SYSTEM 1E Liquid Chemical Sterilant Processing System STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone:

More information

6/12/2016. Objectives. Manufacturers Instructions for Use: Impact on Patient Safety. Background

6/12/2016. Objectives. Manufacturers Instructions for Use: Impact on Patient Safety. Background Manufacturers Instructions for Use: Impact on Patient Safety N A N C Y C H O B I N, R N, A A S, A C S P, C S P M, C F E R P R E S I D E N T, S T E R I L E P R O C E S S I N G U N I V E R S I T Y, L L C

More information

MDRAO ONLINE COURSE FOURTH EDITION 2018

MDRAO ONLINE COURSE FOURTH EDITION 2018 MDRAO ONLINE COURSE FOURTH EDITION 2018 Introduction MDRD Function and Structure Virtual tour of MDRD Standardized workflow Organizational Structure Chapter 1: Quality Systems Creating a Culture of quality

More information

Inhibiting of Microbial Growth in vitro CLS 212

Inhibiting of Microbial Growth in vitro CLS 212 Inhibiting of Microbial Growth in vitro CLS 212 Microbicidal Microbicidal is the process or an agent that kills the microorganism. The suffix -cidal or cide means??( See chapter 8 page 131) Microbistatic

More information

Standards of Infection Prevention in Reprocessing Flexible Gastrointestinal Endoscopes Webinar FAQ

Standards of Infection Prevention in Reprocessing Flexible Gastrointestinal Endoscopes Webinar FAQ Standards of Infection Prevention in Reprocessing Flexible Gastrointestinal Endoscopes Webinar FAQ Disclaimer: The Society of Gastroenterology Nurses and Associates, Inc. assume no responsibility for the

More information

Disinfection and sterilisation

Disinfection and sterilisation Disinfection and sterilisation Mongolia 2011 Prof. Dr. Walter Popp Hospital Hygiene, University Clinics Essen, Germany 1 Term Definition Reduction factor of germs Cleaning Remove dirt including 10-100

More information

ISO/TC 198. Secretariat: ANSI

ISO/TC 198. Secretariat: ANSI DRAFT INTERNATIONAL STANDARD ISO/DIS 17664 ISO/TC 198 Secretariat: ANSI Voting begins on: Voting terminates on: 2016-04-28 2016-07-27 Processing of health care products Information to be provided by the

More information

Decontamination Effectiveness of Esco Celsafe CO2 Incubator Sterilization Feature using High Heat Temperature By Bekti Tri Sumaryati

Decontamination Effectiveness of Esco Celsafe CO2 Incubator Sterilization Feature using High Heat Temperature By Bekti Tri Sumaryati Decontamination Effectiveness of Esco Celsafe CO2 Incubator Sterilization Feature using High Heat Temperature By Bekti Tri Sumaryati 1. Summary Esco Celsafe CO 2 Incubator sterilization feature, which

More information

Hot Topics in Drug Product Process Validation: A Reviewer s Perspective

Hot Topics in Drug Product Process Validation: A Reviewer s Perspective Hot Topics in Drug Product Process Validation: A Reviewer s Perspective Colleen Thomas, Ph.D. Quality Assessment Lead (Acting) FDA/CDER/OPQ/OPF Division of Microbiology Assessment CASSS CMC Strategy Forum

More information

TRAINING PROGRAMME INTRODUCTION:

TRAINING PROGRAMME INTRODUCTION: INTRODUCTION: To provide a detailed core programme to member countries who wish to provide training to decontamination personnel based on European and International Standards. 1. Aims: Indicate the depth

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 14160 First edition 1998-03-15 Sterilization of single-use medical devices incorporating materials of animal origin Validation and routine control of sterilization by liquid

More information

AS/NZS 4187 RELEVANT STANDARDS. Robyn Williams NUM CSSD LBHHS & Standards Australia Chair HE-023

AS/NZS 4187 RELEVANT STANDARDS. Robyn Williams NUM CSSD LBHHS & Standards Australia Chair HE-023 AS/NZS 4187 RELEVANT STANDARDS Robyn Williams NUM CSSD LBHHS & Standards Australia Chair HE-023 Relevant Standards AS 2514 AS 2774 AS 2773 AS 2773.1 Drying cabinets for medical equipment Drying cabinets

More information

Understanding ANSI/AAMI ST 91. Flexible and Semi-Rigid Endoscope Reprocessing in Healthcare Facilities

Understanding ANSI/AAMI ST 91. Flexible and Semi-Rigid Endoscope Reprocessing in Healthcare Facilities Understanding ANSI/AAMI ST 91 Flexible and Semi-Rigid Endoscope Reprocessing in Healthcare Facilities Continuing Education Contact Hours Participants must complete the entire presentation/seminar to achieve

More information

Guide for Cleaning, Sterilization and Storage of Reusable Introducers

Guide for Cleaning, Sterilization and Storage of Reusable Introducers Guide for Cleaning, Sterilization and Storage of Reusable Introducers M Manufactured by: Caldera Medical, Inc. 5171 Clareton Drive Agoura Hills, CA 91301 U.S. Toll Free: 866-4-CALDERA Telephone: 818-879-6555

More information

ENDOSCOPY PRACTICE Manual cleaning and Disinfection of flexible endoscopes

ENDOSCOPY PRACTICE Manual cleaning and Disinfection of flexible endoscopes 1.0 Purpose Every patient undergoing endoscopy should be examined with clean, disinfected equipment. All endoscopic equipment will be cleaned and disinfected / sterilized according to current recommended

More information

5.1.1 & Changes started in 2008 and were mainly driven under the Chairmanship of Prof Hans van Doorne 11/10/2017

5.1.1 & Changes started in 2008 and were mainly driven under the Chairmanship of Prof Hans van Doorne 11/10/2017 5.1.1 & 5.1.2 V Iain Fenton-May Chairman Group 1 Methods of Preparation of Sterile Products Changes started in 2008 and were mainly driven under the Chairmanship of Prof Hans van Doorne 1 Methods of Preparation

More information

Inhibiting Microbial Growth in vitro. CLS 212: Medical Microbiology Zeina Alkudmani

Inhibiting Microbial Growth in vitro. CLS 212: Medical Microbiology Zeina Alkudmani Inhibiting Microbial Growth in vitro CLS 212: Medical Microbiology Zeina Alkudmani Microbicidal or Microbistatic? Microbicidal Microbicidal is the process or an agent that kills the microorganism. The

More information

NO 2 Gas Sterilization and Decontamination: Keeping Pace, from Start to Finish

NO 2 Gas Sterilization and Decontamination: Keeping Pace, from Start to Finish NO 2 Gas Sterilization and Decontamination: Keeping Pace, from Start to Finish Evan Goulet, Ph.D., VP of Microbiology Applications Laboratory PDA Midwest: Sterility Assurance and Risk Mitigation October

More information

CAN STERILE REALLY BE STERILE?: EFFECTIVE AUTOCLAVING METHODS

CAN STERILE REALLY BE STERILE?: EFFECTIVE AUTOCLAVING METHODS Vet Times The website for the veterinary profession https://www.vettimes.co.uk CAN STERILE REALLY BE STERILE?: EFFECTIVE AUTOCLAVING METHODS Author : James Gasson Categories : Vets Date : February 2, 2009

More information

CLEANING AND HANDLING OF MICROPORT INSTRUMENTS The following languages are included in this packet:

CLEANING AND HANDLING OF MICROPORT INSTRUMENTS The following languages are included in this packet: EN CLEANING AND HANDLING OF MICROPORT INSTRUMENTS 150802-1 The following languages are included in this packet: English (en) Deutsch (de) Nederlands (nl) Français (fr) Español (es) Italiano (it) Português

More information

A Sterile Processing User s Guide

A Sterile Processing User s Guide Product Manual A Sterile Processing User s Guide Safety, Assurance, Innovation, & Quality enabling perioperative professionals to achieve higher standards of performance and enhanced patient safety Phone:866-ONE-TRAY

More information

Overview: High Level Disinfection

Overview: High Level Disinfection High Level Disinfection Who you gonna call?. Bug Busters! Dr. Michelle J. Alfa, FCCM St. Boniface General Hospital, Winnipeg, MB Disclaimers: Dr. M. Alfa - Sponsored to give invited presentations at various

More information

VHP Sterilization and Prion Inactivation

VHP Sterilization and Prion Inactivation WFHSS Workshop Opatija Croatia March 27 th 2009 VHP Sterilization and Prion Inactivation Dr. Georgia Alevizopoulou Clinical Specialist Eastern Europe, Middle East & Africa VHP Sterilization and Prion Inactivation

More information

3.03 Understand support services

3.03 Understand support services 3.03 Understand support services Clean and Decontaminate the Healthcare Environment Manage Hazardous Materials and Wastes Manage and Store Materials 3.03 Understand support services 2 Clean and Decontaminate

More information

Getinge da Vinci Solution

Getinge da Vinci Solution Getinge da Vinci Solution A comprehensive cleaning solution for da Vinci EndoWrist instruments This document is intended to provide information to an international audience outside of the US. Validated,

More information

Dissecting Outbreaks of Multi-resistant Bacteria in Gastrointestinal Endoscopy

Dissecting Outbreaks of Multi-resistant Bacteria in Gastrointestinal Endoscopy Dissecting Outbreaks of Multi-resistant Bacteria in Gastrointestinal Endoscopy Dr. Michelle J. Alfa, Ph.D., FCCM Professor, Dept of Medical Microbiology, University of Manitoba, Winnipeg, Canada St Boniface

More information

2016. All rights reserved.

2016. All rights reserved. DOCUMENT: Future AAMI/ISO 17664, 12 May 2017 Final Draft Standard Sterilization of health care products Information to be provided by the device manufacturer for the processing of medical devices Public

More information

*Slides updated to include newer guidelines in February 2016 by Shelby Lassiter, BSN, RN, CPHQ, CIC

*Slides updated to include newer guidelines in February 2016 by Shelby Lassiter, BSN, RN, CPHQ, CIC *Slides updated to include newer guidelines in February 2016 by Shelby Lassiter, BSN, RN, CPHQ, CIC 1 2 3 4 * 5 6 7 If the instrument is not clean, the organic soil can harbor embedded microorganisms.

More information

REUSABLE MEDICAL DEVICE VALIDATION

REUSABLE MEDICAL DEVICE VALIDATION All sterilization processes require validation of the efficacy and reproducibility of the process. Depending on the type of sterilization, this may be accomplished by partial, sub-lethal, or repetitive

More information

Disinfection of stainless steel in hospitals

Disinfection of stainless steel in hospitals Disinfection of stainless steel in hospitals Summary of a report by Manchester Metropolitan University and AgroParisTech www.teamstainless.org 2 Introduction Thanks to its unique properties, stainless

More information

Not to be reproduced - AGPAL & QIP

Not to be reproduced - AGPAL & QIP Keep it Clean! Practical methods for implementing correct sterilisation processes advanced level Presented by: Lin Lochead, B. App. Sc Validation A Documented Procedure for obtaining, recording and interpreting

More information

Sterilization - validation, qualification requirements. Sterilization - Overview

Sterilization - validation, qualification requirements. Sterilization - Overview Sterilization - validation, qualification requirements Dawn Tavalsky 1 ISPE Boston Chapter Framingham MA 19Sep2013 Sterilization - Overview Objectives Discuss definition of Sterile Briefly describe sterilization

More information

The Effect of Air Pockets on the Efficiency of Disinfection of Respiratory Equipment by Pasteurization

The Effect of Air Pockets on the Efficiency of Disinfection of Respiratory Equipment by Pasteurization CENORIN MEDICAL INFORMATION SERIES The Effect of Air Pockets on the Efficiency of Disinfection of Respiratory Equipment by Pasteurization Performed by: Bio Research Laboratories, Inc. Cenorin, LLC 6324

More information

PRODUCT D-VALUE STUDIES: A CRITICAL TOOL WHEN DEVELOPING A STERILIZATION PROCESS

PRODUCT D-VALUE STUDIES: A CRITICAL TOOL WHEN DEVELOPING A STERILIZATION PROCESS Volume 1, Number 3 June 2004 PRODUCT D-VALUE STUDIES: A CRITICAL TOOL WHEN DEVELOPING A STERILIZATION PROCESS Kurt McCauley R & D Lab Manager The objective of a sterilization process is to kill the naturally

More information

What Dentistry Can Learn from AAMI ST79. Best Practices for device and instrument reprocessing

What Dentistry Can Learn from AAMI ST79. Best Practices for device and instrument reprocessing What Dentistry Can Learn from AAMI ST79 Best Practices for device and instrument reprocessing Learning goals: Purpose and goals of AAMI organization. Overview of ST79. Review selected topics: IFUs, tools

More information

SANITATION CLEANING AND DISINFECTANTS

SANITATION CLEANING AND DISINFECTANTS SANITATION CLEANING AND DISINFECTANTS Although the subject of sanitation procedures, as related to disease management, is frequently discussed, it is helpful to review basic principles and review facts

More information

Clostridium difficile Update Dr. Michelle Alfa, Winnipeg Sponsored by ARJO

Clostridium difficile Update Dr. Michelle Alfa, Winnipeg Sponsored by ARJO Clostridium difficile: Update Dr. Michelle Alfa Ph.D, FCCM Diagnostic Services of Manitoba St. Boniface General Hospital Site Hosted by Paul Webber paul@webbertraining.com Sponsored by www.arjo.com www.webbertraining.com

More information

Guide for Cleaning, Sterilization and Storage of Introducers for Caldera Medical Implants

Guide for Cleaning, Sterilization and Storage of Introducers for Caldera Medical Implants Guide for Cleaning, Sterilization and Storage of Introducers for Caldera Medical Implants M Manufactured by: Caldera Medical, Inc. 5171 Clareton Drive Agoura Hills, CA 91301 U.S. Toll Free: 866-4-CALDERA

More information

TOTAL SOLUTION FOR ENDOSCOPE REPROCESSING

TOTAL SOLUTION FOR ENDOSCOPE REPROCESSING TOTAL SOLUTION FOR ENDOSCOPE REPROCESSING A NEED FOR A COMPLETE PROCESS Lionel PINEAU PhD France PATHOGEN TRANSMISSION RELATED TO ENDOSCOPE WHAT IS THE CONTAMINATION RISK? Rates of bacteremia and infectious

More information

IN THIS SECTION MICROBIOLOGY TESTING EXPERT SOLUTIONS FOR PRODUCT DEVELOPMENT. Bacterial Endotoxin (LAL) Testing

IN THIS SECTION MICROBIOLOGY TESTING EXPERT SOLUTIONS FOR PRODUCT DEVELOPMENT. Bacterial Endotoxin (LAL) Testing EXPERT SOLUTIONS FOR PRODUCT DEVELOPMENT IN THIS SECTION MICROBIOLOGY TESTING Microbial assays involve a variety of tests, from the determination of the numbers and types of organisms naturally present

More information

Tests to Support Sterility Claim. Imtiaz Ahmed

Tests to Support Sterility Claim. Imtiaz Ahmed Tests to Support Sterility Claim Imtiaz Ahmed Sterile Product As per TGO 77, a sterile product must comply with the requirements of the following tests: Sterility Test Bacterial Endotoxins Test Appendix

More information

Biological Monitoring

Biological Monitoring Biological Monitoring Prof. Duygu PERÇİN, MD Department of Clinical Microbiology Erciyes University Faculty of Medicine, Kayseri-TURKEY duygu.percin@hotmail.com Learning Objectives The attandee will be

More information

POLICY FOR THE DECONTAMINATION OF FLEXIBLE ENDOSCOPES

POLICY FOR THE DECONTAMINATION OF FLEXIBLE ENDOSCOPES POLICY FOR THE DECONTAMINATION OF FLEXIBLE ENDOSCOPES Version 2.0 August 2012 Name of Policy: Purpose of Policy: Policy for the Decontamination of Flexible Endoscopes To provide guidance on the decontamination

More information

Products made in Switzerland

Products made in Switzerland Products made in Switzerland HYGIENIC AND MEDICALLY CLEAR DIFFERENT PRODUCTS BUT THE SAME QUALITY LABEL HYGIENIC AND MEDICALLY CLEAR 2 MDD TECHNOLOGIES FOR: 1. Cleaning 2. Disinfection 3. High Level Disinfection

More information

THE BASICS OF STERILIZATION

THE BASICS OF STERILIZATION THE BASICS OF STERILIZATION Objectives State the importance of sterilization to patient care Review three key essentials of the sterilization process Describe sterilization methods used, application, limitations,

More information

Ethicon, Inc All rights reserved. EVOTECH and the EVOTECH logo are registered trademarks owned by Advanced Sterilization Products.

Ethicon, Inc All rights reserved. EVOTECH and the EVOTECH logo are registered trademarks owned by Advanced Sterilization Products. Ethicon, Inc. 2009. All rights reserved. EVOTECH and the EVOTECH logo are registered trademarks owned by Advanced Sterilization Products. Consistent Cleaning with No Brushing. Now, the labor-intensive

More information

Entellus Medical Sinuscope INSTRUCTIONS FOR USE

Entellus Medical Sinuscope INSTRUCTIONS FOR USE Entellus Medical Sinuscope INSTRUCTIONS FOR USE Table of contents Entellus Medical Sinuscope Instructions for Use 1. About this document... 1 1.1 Purpose... 1 1.2 Symbols used... 1 2. Intended use... 2

More information

Is contamination of bronchoscopes really a problem?

Is contamination of bronchoscopes really a problem? Is contamination of bronchoscopes really a problem? 1 Is contamination of bronchoscopes really a problem? High Level Disinfection does not methodically clean bronchoscopes What we know 210 More than 210

More information

Best Practices for Reprocessing Endoscopes

Best Practices for Reprocessing Endoscopes Best Practices for Reprocessing Endoscopes Reprocessing of Endoscopes Reprocessing Room Standards The resources referenced include: Canadian Standards Association Provincial Infectious Disease Advisory

More information

Sterilization & Disinfection

Sterilization & Disinfection Sterilization & Disinfection Prof. Hanan Habib College of Medicine-KSU Objectives 1- Define the terms sterilization, disinfectant and antiseptic. 2- Classify the different methods of sterilization (physical

More information

ANSI/AAMI ST91: 2015

ANSI/AAMI ST91: 2015 ANSI/AAMI ST91: 2015 Flexible Endoscope Reprocessing and the Importance of AAMI ST91 April 21, 2017 Fred Alston, CSPDT Director of Clinical Sales and Services Healthmark Industries Objectives Discuss the

More information

Prevention of Transmission of the Superbug Carbapenem-Resistant Enterobacteriaceae (CRE) during Gastrointestinal Endoscopy

Prevention of Transmission of the Superbug Carbapenem-Resistant Enterobacteriaceae (CRE) during Gastrointestinal Endoscopy Prevention of Transmission of the Superbug Carbapenem-Resistant Enterobacteriaceae (CRE) during Gastrointestinal Endoscopy Sponsored by: Presentation by: Lawrence F. Muscarella, Ph.D. President LFM Healthcare

More information

Log Reduction & UV Deceptive Advertising

Log Reduction & UV Deceptive Advertising Log Reduction & UV Deceptive Advertising Sellers of both UV-C and PX-UV (pulsed UV) light room treatment products are falsely claiming that their products can disinfect, sterilize, or decontaminate, hospital

More information

ANSI/AAMI ST79 A2:2009 Key Changes in the 2009 Amendment

ANSI/AAMI ST79 A2:2009 Key Changes in the 2009 Amendment 3M Attest TM TM Products Sterile U Network T U T O R I A L S ANSI/AAMI ST79 A2:2009 Key Changes in the 2009 Amendment Background: A second amendment (A2:2009) to the Association for the Advancement of

More information

Sterilization, disinfection and reprocessing

Sterilization, disinfection and reprocessing Sterilization, disinfection and reprocessing David R. Woodard, MSc, CIC, FSHEA Definitions Cleaning the removal of visible soil from the surfaces, joints and lumen of instruments or devices. This is done

More information

Xerafy Autoclavable XS and XXS User Guide

Xerafy Autoclavable XS and XXS User Guide Xerafy Autoclavable XS and XXS User Guide General Information This user guide provides instructions for attaching Xerafy autoclavable XS and XXS series tags onto surgical instruments. This document is

More information