Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0

Size: px
Start display at page:

Download "Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0"

Transcription

1 User Bulletin Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 September 9, 2010 SUBJECT: Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products In This User Bulletin Overview Audience This user bulletin includes the following topics: Safety How to Obtain Support New Features and Design Changes Reagents, Kits, and Run Module for HID Analysis Section 1 Performing HID Analysis Section 2 Examples of DNA Profiles Generated on the 3130/3130xl Genetic Analyzers Appendix A, Maintenance This document describes: The design changes and new features that differentiate the Applied Biosystems 3130/3130xl Genetic Analyzers from the ABI PRISM 3100/3100-Avant Genetic Analyzers. The tasks required to perform Human Identification (HID) analysis with the 3130/3130xl Genetic Analyzers, using 3130/3130xl Genetic Analyzer Data Collection Software v3.0 and GeneMapper ID Software v3.2. Examples of DNA profiles generated on the 3130/3130xl Genetic Analyzers. Maintenance schedule and maintenance procedures for the 3130/3130xl Genetic Analyzers. This user bulletin is intended for customers who will be using the 3130/3130xl Genetic Analyzers to perform HID analysis.

2 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Related Documentation The following documents are referenced in this user bulletin. Be sure to have these on hand when you are performing experiments with the 3130/3130xl Genetic Analyzers. Applied Biosystems 3130/3130xl Genetic Analyzers Getting Started Guide (PN , Rev. B). GeneMapper ID Software v3.1 and v3.2: Human Identification Analysis Tutorial (PN , Rev. A) Applied Biosystems 3130/3130xl Genetic Analyzers Maintenance, Troubleshooting, and Reference Guide (PN , Rev. B) Safety Complete Safety information for the 3130/3130xl Genetic Analyzers may be found in the Applied Biosystems 3130/3130xl Genetic Analyzers Getting Started Guide. How to Obtain Support For the latest services and support information for all locations, go to then click the link for Support. At the Support page, you can: Search through frequently asked questions (FAQs) Submit a question directly to Technical Support Order Applied Biosystems user documents, MSDSs, certificates of analysis, and other related documents Download PDF documents Obtain information about customer training Download software updates and patches In addition, the Support page provides access to worldwide telephone and fax numbers to contact Applied Biosystems Technical Support and Sales facilities. 2 User Bulletin

3 New Features and Design Changes New Features and Design Changes Instrument Modifications Polymer delivery system (PDS) Applied Biosystems has taken a proven PDS technology from the Applied Biosystems 3730 DNA Analyzer and applied it to the 4- and 16-capillary instruments. The PDS improves instrument performance and maintenance by reducing manual intervention. The primary benefit lies in the automation of the polymer delivery process. The automated PDS of the 3130/3130xl Genetic Analyzers does not use polymer syringes. Eliminating the manual steps for filling, priming, and loading polymer provides faster, easier, and more reliable instrument setup. You simply load the polymer bottle onto the instrument, then perform setup using the software wizards (see Data Collection Software Modifications on page 4). Note: The positions of the waste and rinse reservoirs are reversed on the 3130/3130xl Genetic Analyzers (as compared to the 3100/3100-Avant Genetic Analyzers). See step 7 on page 64. Oven door Applied Biosystems has changed the composition of the gasket material used to seal the oven door. CCD camera shutter This modification affects the CCD camera shutter only. It does not alter mechanically or introduce variation to the performance of the CCD camera or optical path components. Detection cell heater The detection cell heater is only activated when using run modules created for the POP-7 polymer. Applied Biosystems does not recommend activation of the detection cell heater for HID applications. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 3

4 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Data Collection Software Modifications Data Collection Software v3.0 represents Applied Biosystems latest controlling software for the capillary electrophoresis instruments. As in Data Collection Software v2.0, there are software wizards to guide you through maintenance procedures: Several of these wizards are new in Data Collection Software v3.0 and support the new pump-based PDS of the 3130/3130xl Genetic Analyzers: Change Polymer Type Wizard (to change to a different polymer type) Replenish Polymer Wizard (to replenish with the same polymer, from the same or different lot) Bubble Remove Wizard Water Wash Wizard Instrument Shutdown Wizard 4 User Bulletin

5 Reagents, Kits, and Run Module for HID Analysis Reagents, Kits, and Run Module for HID Analysis The reagents, AmpFlSTR kits, and run module listed in this section are available for performing HID analysis using the 3130/3130xl Genetic Analyzers. Reagents Part Number Name 3130 Genetic Analyzer 3130xl Genetic Analyzer Matrix Standard Set DS-32 (Dye Set F) Note: Matrix Standard Set DS-32 is for use on multicapillary instruments (that is, the 3100/3100-Avant Genetic Analyzers and the 3130/3130xl Genetic Analyzers) Matrix Standard Set DS-33 (Dye Set G5) Note: Matrix Standard Set DS-33 is for use on multicapillary instruments (that is, the 3100/3100-Avant Genetic Analyzers and the 3130/3130xl Genetic Analyzers) POP-4 polymer Hi-Di formamide Genetic Analyzer Buffer with EDTA Note: Use 10 Genetic Analyzer Buffer with EDTA to prepare 1 Genetic Analyzer Buffer with EDTA (1 running buffer) and 3100-Avant Capillary Array, 36 cm xl and 3100 Capillary Array, 36 cm Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 5

6 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 AmpFlSTR Kits Kit Dye Set Matrix Standard Set AmpFlSTR Identifiler PCR Amplification Kit AmpFlSTR Yfiler PCR Amplification Kit AmpFlSTR SEfiler PCR Amplification Kit G5 DS-33 AmpFlSTR Profiler Plus PCR Amplification Kit AmpFlSTR COfiler PCR Amplification Kit AmpFlSTR Profiler Plus ID PCR Amplification Kit AmpFlSTR SGM Plus PCR Amplification Kit Other four-dye AmpFlSTR kits F DS-32 Run Module Run Module Capillary Length (cm) Polymer Type Run Time (min) 24-hr Throughput (GT a ) 3130 Analyzer 3130xl Analyzer Resolution (bp) Specification (SD) b HIDFragment Analysis36_POP4_1 36 POP ,560 10, a. 20 GT (genotypes)/capillary/injection. b. 1-bp resolution at 99.99% accuracy. 6 User Bulletin

7 Section 1 Performing HID Analysis Section 1 Performing HID Analysis This section covers the following topics: How to Perform the Tasks in This Section Preparing the Instrument Performing a Spatial Calibration Performing a Spectral Calibration Performing a Sample Run How to Perform the Tasks in This Section In this section, the tasks required to perform HID analysis on the 3130/130xl Genetic Analyzers are grouped under four major steps: 1. Preparing the Instrument on page 8 2. Performing a Spatial Calibration on page Performing a Spectral Calibration on page Performing a Sample Run on page 25 Many of the tasks listed under each of these major steps are described in detail in the Applied Biosystems 3130/3130xl Genetic Analyzers Getting Started Guide (3130/3130xl Genetic Analyzers GSG). Consequently, this user bulletin is meant to be used in conjunction with the 3130/3130xl Genetic Analyzers GSG: When tasks required to perform HID analysis are the same as those described in the 3130/3130xl Genetic Analyzers GSG, you are referred to the appropriate page number in the 3130/3130xl Genetic Analyzers GSG. IMPORTANT! Be sure to use the correct revision of the 3130/3130xl Genetic Analyzers GSG: Part Number , Revision B (PN , Rev. B). If you do not use PN , Rev. B, the referenced page numbers may be incorrect. If you do not have this revision, you can download it from Applied Biosystems website; see How to Obtain Support on page 2. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 7

8 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 When tasks required to perform HID analysis are different from those described in the 3130/3130xl Genetic Analyzers GSG, you are referred to the appropriate page number in this user bulletin. The user bulletin references are highlighted in blue text (for *.pdf files) or bold text (for printed documents). Preparing the Instrument Workflow To prepare the instrument for a run: Step Task See... 1 Review the instrument parts and functions. page 10 of this user bulletin. 2 Start the computer workstation, then start the 3130/3130xl Genetic Analyzer. 3 Start the 3130/3130xl Genetic Analyzer Data Collection Software v Inspect the instrument, then perform any necessary maintenance tasks. IMPORTANT! As part of instrument maintenance, you must change or replenish polymer weekly. For HID analysis, be sure to use 3130 POP-4 polymer (PN ). page 7 of the 3130/3130xl Genetic Analyzers GSG. page 9 of the 3130/3130xl Genetic Analyzers GSG. pages 12 to15 of the 3130/3130xl Genetic Analyzers GSG. Appendix A on page 55 of this user bulletin. CHEMICAL HAZARD. POP- 4 polymer causes eye, skin, and respiratory tract irritation. Read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. 8 User Bulletin

9 Preparing the Instrument Step Task See... 5 Use 10 Genetic Analyzer Buffer with EDTA (PN ) to prepare 1 Genetic Analyzer Buffer with EDTA (1 running buffer), then fill the instrument reservoirs. page 16 of the 3130/3130xl Genetic Analyzers GSG. CHEMICAL HAZARD. 10 Genetic Analyzer Buffer with EDTA may cause eye, skin, and respiratory tract irritation. Read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. CHEMICAL HAZARD. 1 Genetic Analyzer Buffer with EDTA may cause eye, skin, and respiratory tract irritation. Read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 9

10 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Instrument Layout Pump block Polymer delivery pump (PDP) Oven Detection cell block and heater Capillary array Buffer and water reservoirs GR2460 Autosampler Lower polymer block Anode buffer reservoir Part Anode buffer reservoir Buffer and water reservoirs (four) Autosampler Capillary array Detection cell block and heater Function Contains 16 ml of 1 running buffer. Each contain 16 ml of 1 running buffer or water. Holds the sample plates and reservoirs and moves to align the samples, water, or buffer with the capillaries. Enables the separation of the fluorescentlabeled DNA fragments by electrophoresis. It is a replaceable unit composed of 4 or 16 silica capillaries. Holds the capillaries in place for laser detection. 10 User Bulletin

11 Performing a Spatial Calibration Lower polymer block Oven Part Polymer delivery pump (PDP) Pump block Function Contains the buffer valve, anode electrode, and anode buffer reservoir. Maintains uniform capillary array temperature. Pumps polymer into the array and performs maintenance procedures. Includes the displacement pump chamber, piston water seal, array attachment point (array port), and connection to the lower polymer block through the interconnect tube. Performing a Spatial Calibration The Data Collection software uses images collected during the spatial calibration to establish a relationship between the signal emitted by each capillary and the position where that signal falls on and is detected by the CCD camera. When to Perform the Calibration Workflow You are required to perform a spatial calibration when you: Install or replace a capillary array Temporarily remove the capillary array from the detection block Move the instrument To perform a spatial calibration: Step Task See... 1 Create a spatial calibration file. page 22 of the 3130/3130xl Genetic Analyzers GSG. 2 Evaluate the spatial calibration file. page 23 of the 3130/3130xl Genetic Analyzers GSG. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 11

12 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Performing a Spectral Calibration A spectral calibration creates a matrix that corrects for the overlapping fluorescence emission spectra of the dyes. Although each of these dyes emits its maximum fluorescence at a different wavelength, there is some overlap in the emission spectra between the dyes. The goal of multicomponent analysis is to effectively correct for spectral overlap. Performing a spectral calibration is similar to performing a sample run, except that calibration standards are run in place of samples and a spectral calibration module is used in place of a run module. When to Perform the Calibration Perform a spectral calibration: When you use a new dye set on the instrument When you change capillary array length or polymer type Note: For every fragment analysis dye set, you must create a separate spectral calibration for each capillary array length and polymer type combination you use. After the laser or CCD camera has been realigned/replaced by a service engineer If you begin to see a decrease in spectral separation (pull-up and/or pull-down peaks) Workflow To perform a spectral calibration: Step Task See... 1 Choose the correct dye set and matrix standard set for the AmpFlSTR kit you are using. page 14 of this user bulletin. Note: New matrix standard sets are available for all validated multi-capillary instruments (that is, the 3100/3100-Avant Genetic Analyzers and the 3130/3130xl Genetic Analyzers). 12 User Bulletin

13 Performing a Spectral Calibration Step Task See... 2 Select a spectral calibration protocol for Dye Set G5 or Dye Set F and create a plate record. IMPORTANT! For HID applications, the default condition number boundaries for the Dye Set G5 spectral calibration protocol should be edited in Data Collection Software v3.0. You will only need to perform this edit once. See Edit the Dye Set G5 Spectral Calibration Protocol on page 15 of this user bulletin. page 15 of this user bulletin. 3 Prepare the spectral calibration chemistry. page 21 of this user bulletin. 4 Load the spectral calibration samples. page 32 of the 3130/3130xl Genetic Analyzers GSG. 5 Place the plate assembly into the instrument. page 39 of the 3130/3130xl Genetic Analyzers GSG. 6 Run the spectral calibration plate, which includes: Linking the plate Starting the run Viewing the pass/fail status 7 Evaluate the spectral calibration data. IMPORTANT! If you obtain off-scale peaks in a 3130/3130xl Genetic Analyzer spectral calibration when using the new matrix standard sets, see Troubleshooting: Off-Scale Spectral Calibration Peak Height on page 23 of this user bulletin. pages 40 to 42 of the 3130/3130xl Genetic Analyzers GSG. page 43 of the 3130/3130xl Genetic Analyzers GSG. 8 Activate the spectral calibration. page 47 of the 3130/3130xl Genetic Analyzers GSG. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 13

14 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Choose the Dye Set and Matrix Standards Choose the appropriate dye set and matrix standard for the AmpFlSTR PCR Amplification Kit you are using, as shown in the table below. For AmpFlSTR kits that use a... Use... And use... AmpFlSTR Kit Examples five-dye system, which includes the following dyes: 6FAM VIC NED PET LIZ Dye Set G5 Matrix Standard Set DS-33 AmpFlSTR Identifiler PCR Amplification Kit AmpFlSTR Yfiler PCR Amplification Kit AmpFlSTR SEfiler PCR Amplification Kit four-dye system, which includes the following dyes: 5FAM JOE NED ROX Dye Set F Matrix Standard Set DS-32 AmpFlSTR Profiler Plus PCR Amplification Kit AmpFlSTR COfiler PCR Amplification Kit AmpFlSTR Profiler Plus ID PCR Amplification Kit AmpFlSTR SGM Plus PCR Amplification Kit Other four-dye AmpFlSTR kits 14 User Bulletin

15 Performing a Spectral Calibration Select a Spectral Calibration Protocol and Create a Plate Record For Dye Set G5 Follow this procedure if you are performing a spectral calibration for kits using a five-dye system, which includes the 6FAM, VIC, NED, PET, and LIZ dyes. IMPORTANT! For HID applications, the default condition number boundaries for the Dye Set G5 spectral calibration protocol should be edited in Data Collection Software v3.0. Instructions for editing the condition number boundaries are provided below; you will only need to perform this edit once. Edit the Dye Set G5 Spectral Calibration Protocol 1. In the tree pane of the Data Collection software, click GA Instruments > ga3130xl or ga3130 > Protocol Manager to open the Protocol Manager window. 2. In the Instrument Protocols pane, highlight the spectral calibration protocol for Dye Set G5: Spectral36_POP4_G5. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 15

16 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v Click Edit to open the Protocol Editor. 4. Change the condition number boundaries: a. Click Edit Param. The Edit Spectral Params window opens. b. Change the condition number boundary in the Lower field from 8.5 to 7.0. c. Change the condition number boundary in the Upper field from 14.5 to d. Click OK to return to the Protocol Editor. 16 User Bulletin

17 Performing a Spectral Calibration 5. Click OK to save your changes and exit the Protocol Editor. Note: Once you have saved your changes, you will not need to edit the Dye Set G5 spectral calibration protocol again. The lower and upper condition number boundaries are now set at 7.0 and 12.0, respectively. 6. Continue with Select the Dye Set G5 Spectral Calibration Protocol and Create a Plate Record below. Select the Dye Set G5 Spectral Calibration Protocol and Create a Plate Record 1. In the tree pane of the Data Collection software, click GA Instruments > ga3130xl or ga3130 > Plate Manager to open the Plate Manager window. 2. Click New to open the New Plate dialog box. 3. Complete the New Plate dialog box: a. Name: Type a name for the plate. b. Description: If desired, type a description for the plate. c. Application: Select Spectral Calibration from the dropdown list. d. Plate Type: Select 96-Well. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 17

18 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 e. Owner Name: Type a name for the owner. f. Operator Name: Type a name for the operator. 4. Click OK to open the Spectral Calibration Plate Editor. 5. Complete the Spectral Calibration Plate Editor. a. In the Sample Name column, enter a sample name, then click the next cell. The value 100 automatically displays in the Priority column. b. Optional. In the Comments column, enter any additional comments or notations for the sample at the corresponding position of the plate. c. In the Instrument Protocol 1 column, select Spectral36_POP4_G5. 6. Highlight the entire row, then select Edit > Fill Down Special. The software automatically fills in the appropriate well numbers for a single run. 18 User Bulletin

19 Performing a Spectral Calibration 7. Click OK to save your changes, then close the plate record. IMPORTANT! After clicking OK within the Plate Editor, the completed plate record is stored in the Plate Manager database. Once in the Plate Manager database, the plate record can be searched for, edited, exported, or deleted. For Dye Set F Follow this procedure if you are performing a spectral calibration for kits using a four-dye system, which includes the 5FAM, JOE, NED, and ROX dyes. 1. In the tree pane of the Data Collection software, click GA Instruments > ga3130xl or ga3130 > Plate Manager to open the Plate Manager window. 2. Click New to open the New Plate dialog box. 3. Complete the New Plate dialog box. a. Name: Type a name for the plate. b. Description: If desired, type a description for the plate. c. Application: Select Spectral Calibration from the dropdown list. d. Plate Type: Select 96-Well. e. Owner Name: Type a name for the owner. f. Operator Name: Type a name for the operator. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 19

20 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v Click OK to open the Spectral Calibration Plate Editor. 5. Complete the Spectral Calibration Plate Editor. a. In the Sample Name column, enter a sample name, then click the next cell. The value 100 automatically displays in the Priority column. b. Optional. In the Comments column, enter any additional comments or notations for the sample at the corresponding position of the plate. c. In the Instrument Protocol 1 column, select Spectral36_POP4_F. 6. Highlight the entire row, then select Edit > Fill Down Special. Note: The software automatically fills in the appropriate well numbers for a single run. 7. Click OK to save your changes, then close the plate record. IMPORTANT! After clicking OK within the Plate Editor, the completed plate record is stored in the Plate Manager database. Once in the Plate Manager database, the plate record can be searched for, edited, exported, or deleted. 20 User Bulletin

21 Performing a Spectral Calibration Prepare the Spectral Calibration Chemistry Prepare the Matrix Standards for Dye Set G5 Follow this procedure if you are setting up spectral (matrix) calibration standards for kits using a five-dye system, which includes the 6-FAM, VIC, NED, PET, and LIZ dyes. 1. Thoroughly vortex the Matrix Standard Set DS-33 tube for Dye Set G5. 2. Spin the tube briefly in a microcentrifuge. 3. Prepare Matrix Standard Set DS-33 for Dye Set G5 by combining the following in a labeled 1.5-mL microcentrifuge tube: Reagent Volume (µl) 3130 System Volume (µl) 3130xl System Matrix Standard Set DS Hi-Di Formamide Final Volume CHEMICAL HAZARD. Hi-Di Formamide. Exposure causes eye, skin, and respiratory tract irritation. It is a possible developmental and birth defect hazard. Read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. 4. Vortex thoroughly. 5. Spin the mixture briefly in a microcentrifuge. 6. Heat the tube at 95 C for 5 min to denature the sample. 7. Immediately place the tube on ice for 3 min. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 21

22 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Prepare the Matrix Standards for Dye Set F Follow this procedure if you are setting up spectral (matrix) calibration standards for kits using a four-dye system, which includes the 5-FAM, JOE, NED, and ROX dyes. 1. Thoroughly vortex the Matrix Standard Set DS-32 tube for Dye Set F. 2. Spin the tube briefly in a microcentrifuge. 3. Prepare Matrix Standard Set DS-32 for Dye Set F by combining the following in a labeled 1.5-mL microcentrifuge tube: Reagent Volume (µl) 3130 System Volume (µl) 3130xl System Matrix Standard Set DS Hi-Di Formamide Final Volume CHEMICAL HAZARD. Hi-Di Formamide. Exposure causes eye, skin, and respiratory tract irritation. It is a possible developmental and birth defect hazard. Read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. 4. Vortex thoroughly. 5. Spin the mixture briefly in a microcentrifuge. 6. Heat the tube at 95 C for 5 min to denature the sample. 7. Immediately place the tube on ice for 3 min. 22 User Bulletin

23 Performing a Spectral Calibration Troubleshooting: Off-Scale Spectral Calibration Peak Height New universal matrix standards are available for the multi-capillary electrophoresis instruments (that is, the 3100/3100-Avant Genetic Analyzers and the 3130/3130xl Genetic Analyzers): Matrix Standard Set DS-32 (Dye Set F), PN Matrix Standard Set DS-33 (Dye Set G5), PN The new matrix standard sets were designed with improved purity, signal balance, and signal strength. If you obtain off-scale peaks when using the new matrix standards for spectral calibration, follow the procedure below. Troubleshoot Off-Scale Peak Heights If the peak height of any one of the matrix standard fragments is saturated (off-scale), re-run the spectral calibration, but use one-half the amount of the matrix standards. The mixture will be twice as dilute as before. For Dye Set G5 1. Follow the procedure on page 21, Prepare the Matrix Standards for Dye Set G5. 2. In step 3 on page 21, use the following volumes: Reagent Volume (µl) 3130 System Volume (µl) 3130xl System Matrix Standard Set DS Hi-Di Formamide Final Volume CHEMICAL HAZARD. Formamide. Exposure causes eye, skin, and respiratory tract irritation. It is a possible developmental and birth defect hazard. Read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. Note: These same instructions are present on the Matrix Standard Set DS-33 product insert. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 23

24 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 For Dye Set F 1. Follow the procedure on page 22, Prepare the Matrix Standards for Dye Set F. 2. In step 3 on page 22, use the following volumes: Reagent Volume (µl) 3130 System Volume (µl) 3130xl System Matrix Standard Set DS Hi-Di Formamide Final Volume CHEMICAL HAZARD. Formamide. Exposure causes eye, skin, and respiratory tract irritation. It is a possible developmental and birth defect hazard. Read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. Note: These same instructions are present on the Matrix Standard Set DS-32 product insert. 24 User Bulletin

25 Performing a Sample Run Performing a Sample Run In order for data collection and autoanalysis to be successful, each run of samples must have an instrument protocol and results group assigned within a plate record. Workflow To perform a run: Step Task See... 1 Set up GeneMapper ID Software v3.2 (that is, import panels and bin sets, create analysis methods, etc.). Note: This step is only required for autoanalysis GeneMapper ID Software v3.1 and v3.2: Human Identification Analysis Tutorial (PN , Rev. A) 2 Create an instrument protocol for HID analysis. page 26 of this user bulletin. 3 Create a results group. page 115 of the 3130/3130xl Genetic Analyzers GSG. 4 Create a plate record in GeneMapper ID Software v3.2. page 28 of this user bulletin. 5 Prepare your sample chemistry. page 32 of this user bulletin. 6 Load your samples. page 33 of this user bulletin. 7 Place the plate assembly into the instrument. page 39 of the 3130/3130xl Genetic Analyzers GSG. 8 Run your sample plate, which includes: Linking/unlinking a plate Starting the run Controlling/monitoring the run Viewing data pages 128 to 160 of the 3130/3130xl Genetic Analyzers GSG. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 25

26 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Create an Instrument Protocol An instrument protocol contains all the settings necessary to run the instrument: Protocol name Type of run Run module Dye set Create an Instrument Protocol 1. In the tree pane of the Data Collection software, click GA Instruments > ga3130xl or ga3130 > Protocol Manager to open the Protocol Manager window. Create instrument protocols in this pane. Create analysis protocols in this pane. Click New here. 2. In the Instrument Protocols pane, click New to open the Protocol Editor. 26 User Bulletin

27 Performing a Sample Run 3. Complete the Protocol Editor fields: a. Name: Type a name for the protocol. b. Description: If desired, type a description for the protocol. c. Type: Select REGULAR from the drop-down list. d. Run Module: Select HIDFragmentAnalysis36_POP4_1. e. Dye Set: Per the table below, select the correct dye set for your run. Kit AmpFlSTR Identifiler PCR Amplification Kit AmpFlSTR Yfiler PCR Amplification Kit AmpFlSTR SEfiler PCR Amplification Kit Dye Set G5 Matrix Standard Set DS-33 Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 27

28 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Kit AmpFlSTR Profiler Plus PCR Amplification Kit AmpFlSTR COfiler PCR Amplification Kit AmpFlSTR Profiler Plus ID PCR Amplification Kit AmpFlSTR SGM Plus PCR Amplification Kit Other four-dye AmpFlSTR kits Dye Set F Matrix Standard Set DS Click OK to save your changes and exit the Protocol Editor. Create a Plate Record in GeneMapper ID Software v3.2 Create a Plate Record 1. In the tree pane of the Data Collection software, click GA Instruments > ga3130xl or ga3130 > Plate Manager to open the Plate Manager window. 2. Click New to open the New Plate dialog box. 3. Complete the New Plate dialog box: 28 User Bulletin

29 Performing a Sample Run a. Name: Type a name for the plate. b. Description: If desired, type a description for the plate. c. In the Application drop-down list, select: GeneMapper-Generic or GeneMapper-<Computer Name> When you are performing autoanalysis, you must select the GeneMapper-<Computer Name> application. d. Plate Type: Select 96-Well. e. Owner Name: Type a name for the owner. f. Operator Name: Type a name for the operator. 4. Click OK. The GeneMapper Plate Editor opens. 5. Continue with one of the following procedures, as appropriate: Complete the Plate Record for Manual Analysis on page 30 Complete the Plate Record for Autoanalysis on page 30 Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 29

30 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Complete the Plate Record for Manual Analysis Complete the Plate Record for Autoanalysis 1. In the Sample Name column of a row, enter a sample name, then click the next cell. The value 100 is automatically displayed in the Priority column. 2. Optional. In the Comment column, enter any additional comments or notations for the sample. 3. In the Results Group 1 column, select a group from the dropdown list. 4. In the Instrument Protocol 1 column, select a module from the drop-down list. When you use the 3130/3130xl Genetic Analyzers with Data Collection Software v3.0 and the AmpFlSTR kits, Applied Biosystems recommends that you select the HIDFragmentAnalysis36_POP4_1 module. 5. If you want to perform more than one run: a. Select Edit > Add Sample Run. Additional Results Group, Instrument Protocol and Analysis Protocol columns are added to the right end of the plate record. b. Complete the Results Group and Instrument Protocol columns for the additional runs. 6. Click OK to save your changes, then close the plate record. IMPORTANT! After clicking OK within the Plate Editor, the completed plate record is stored in the Plate Manager database. Once in the Plate Manager database, the plate record can be searched for, edited, exported, or deleted. 1. In the Sample Name column of a row, enter a sample name, then click the next cell. The value 100 is automatically displayed in the Priority column. 2. In the Comment column, enter any additional comments or notations for the sample. 3. In the Priority column, change the priority value, if desired. 4. In the Sample Type column, select a sample type from the dropdown list that corresponds to the sample in that well. 5. In the Size Standard column, select a size standard from the drop-down list: For four-dye AmpFlSTR kits, select GS500 ROX. 30 User Bulletin

31 Performing a Sample Run For five-dye AmpFlSTR kits, select GS500 LIZ. 6. In the Panel column, select the appropriate AmpFlSTR panel from the drop-down list. 7. In the Analysis Method column, select a method from the dropdown list. 8. Leave the Snp Set column blank. 9. Enter text for User-Defined columns 1 to In the Results Group 1 column, select a group from the dropdown list. 11. In the Instrument Protocol 1 column, select a module from the drop-down list. When you use the 3130/3130xl Genetic Analyzers with Data Collection Software v3.0 and the AmpFlSTR kits, Applied Biosystems recommends that you select the HIDFragmentAnalysis36_POP4_1 module. 12. To complete the rest of the plate record based on the samples loaded in your plate, do one of the following: If the column contains... the same sample types and the same protocols different sample types and different protocols Then... highlight the entire column, then select Edit > Fill Down Special. Based on the plate type (96- well) and capillary array (16 or 4 capillaries) you are using, the software automatically fills in the appropriate well numbers for a single run. complete the plate record manually. 13. If you want to perform more than one run: a. Select Edit > Add Sample Run. Additional Results Group, Instrument Protocol and Analysis Protocol columns are added to the right end of the plate record. b. Complete the columns for the additional runs. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 31

32 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v Click OK to save your changes, then close the plate record. IMPORTANT! After clicking OK within the Plate Editor, the completed plate record is stored in the Plate Manager database. Once in the Plate Manager database, the plate record can be searched for, edited, exported, or deleted. Prepare Your Sample Chemistry Perform PCR Prepare the Formamide:Size Standard Mixture To prepare your DNA samples and perform PCR, follow the instructions in the appropriate AmpFlSTR PCR Amplification Kit user manual. You can prepare the formamide:size standard mixture for each individual sample or prepare a bulk for all samples in the run. 1. Combine the following in a single microcentrifuge tube: Preparation Type Individual sample preparation: For each sample, combine... Reagent GeneScan -500 ROX Size Standard Dye Set F Volume (µl) Dye Set G5 0.5 GeneScan -500 LIZ Size Standard 0.3 Hi-Di Formamide Bulk sample preparation: For 16 samples, combine... GeneScan-500 ROX Size Standard 10 GeneScan-500 LIZ Size Standard 6 Hi-Di Formamide Note: Prepare the appropriate size standard formulation for your dye set. CHEMICAL HAZARD. Hi-Di Formamide. Exposure causes eye, skin, and respiratory tract irritation. It is a possible developmental and birth defect hazard. Read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. 2. Vortex the tube to mix, then spin briefly in a microcentrifuge. 32 User Bulletin

33 Performing a Sample Run Load Your Samples Load the Samples and Allelic Ladder 1. In a 96-well reaction plate: Dispense 9 µl of the formamide:size standard mixture into each well that will contain sample or allelic ladder. Add 10 µl of the formamide to each blank well. CHEMICAL HAZARD. Formamide. Exposure causes eye, skin, and respiratory tract irritation. It is a possible developmental and birth defect hazard. Read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. 2. Load 1 µl of the sample or allelic ladder into the wells. 3. Cover the reaction plate with an appropriate septa. Use: Reservoir septa, or 96-well plate septa if samples for more than one run were prepared 4. Briefly spin the reaction plate in a centrifuge to ensure that the contents of each well are mixed and collected at the bottom. 5. To denature, heat the reaction plate in a thermal cycler at 95 C for 3 min. 6. Place the reaction plate immediately on ice for 3 min. Prepare the Plate Assembly 1. Assemble the plate assembly: a. Place the sample plate into the plate base. b. Snap the plate retainer onto the plate and plate base. 2. Verify that the holes of the plate retainer and the septa strip are aligned. If not, re-assemble the plate assembly. IMPORTANT! Damage to the array tips will occur if the plate retainer and septa strip holes do not align correctly. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 33

34 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v User Bulletin

35 Section 2 Examples of DNA Profiles Generated on the 3130/3130xl Genetic Analyzers Section 2 Examples of DNA Profiles Generated on the 3130/3130xl Genetic Analyzers This section covers the following topics: Appendix, Examples of DNA Profiles Generated on the 3130 Genetic Analyzer Examples of DNA Profiles Generated on the 3130xl Genetic Analyzer Genotype Concordance Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 35

36 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Examples of DNA Profiles Generated on the 3130 Genetic Analyzer Overview This section contains examples of data generated on the 3130 Genetic Analyzer during the Verification Study. The experiments conducted on the 3130 Genetic Analyzer are as follows: Reproducibility (page 36) Mixture Studies (page 38) Resolution (page 40) Sensitivity (page 41) All studies were performed using Data Collection Software v3.0 and GeneMapper ID Software v3.2. All DNA samples used in the studies were quantitated using the Quantifiler Human DNA Quantification Kit. Reproducibility Figure 1 shows a comparison of signal intensity across a 3130 instrument capillary array using AmpFlSTR Control DNA 9948 (1 ng) amplified using the AmpFlSTRYfiler PCR Amplification Kit. Capillary 1 Capillary 2 Capillary 3 Capillary 4 Figure 1 DNA samples amplified using the Yfiler kit 36 User Bulletin

37 Examples of DNA Profiles Generated on the 3130 Genetic Analyzer Figure 2 shows a comparison of signal intensity across a 3130 instrument capillary array using AmpFlSTR Control DNA 9948 (2 ng) amplified using the AmpFlSTR Profiler Plus PCR Amplification Kit. Capillary 1 Capillary 2 Capillary 3 Capillary 4 Figure 2 DNA samples amplified using the Profiler Plus kit Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 37

38 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Mixture Studies In Figure 3, the first and last panels display profiles of each DNA sample amplified individually with a total DNA input of 1 ng using the AmpFlSTR Identifiler PCR Amplification Kit. (The female AmpFlSTR Control DNA 9947a sample is in the top panel and the male AmpFlSTR Control DNA 9948 sample is in the bottom panel). The other three panels display the mixtures of DNA at approximate ratios of 1:1, 1:5, and 1:10. 1:0 1:1 1:5 1:10 0:1 Figure 3 DNA samples amplified using the Identifiler kit 38 User Bulletin

39 Examples of DNA Profiles Generated on the 3130 Genetic Analyzer In Figure 4, the first and last panels display profiles of each DNA sample amplified individually with a total DNA input of 2 ng using the AmpFlSTR Profiler Plus PCR Amplification Kit. (The female AmpFlSTR Control DNA 9947a sample is in the top panel and the male AmpFlSTR Control DNA 9948 sample is in the bottom panel). The other three panels display the mixtures of DNA at approximate ratios of 1:1, 1:5, and 1:10. 1:0 1:1 1:5 1:10 0:1 Figure 4 DNA samples amplified using the Profiler Plus kit Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 39

40 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Resolution Figure 5 shows three capillaries of a 3130 instrument 4-capillary array displaying single nucleotide resolution. As shown below, the 9.3 and 10 alleles were resolved and individually detected in all capillaries in which they were run. Capillary 1 Capillary 2 Capillary 3 Figure 5 A 4-capillary array displaying single nucleotide resolution 40 User Bulletin

41 Examples of DNA Profiles Generated on the 3130 Genetic Analyzer Sensitivity Figure 6 shows dilutions for the AmpFlSTR Control DNA 9948 sample quantified using the Quantifiler Human DNA Quantification Kit and amplified using the AmpFlSTR Identifiler PCR Amplification Kit: 1.0 ng, 0.5 ng, 0.25 ng, and ng of input DNA. The Y-axis is magnified for lower input DNA amounts. With 125 pg of input DNA quantified using the Quantifiler kit, it was observed at the FGA locus that one of the alleles was not detected at a peak amplitude threshold of 50 RFU. 1.0 ng 0.5 ng 0.25 ng ng Figure 6 DNA sample amplified using the Identifiler kit Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 41

42 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Figure 7 shows dilutions for the AmpFlSTR Control DNA 9948 sample amplified using the Yfiler kit: 1.0 ng, 0.5 ng, 0.25 ng, and ng of input DNA. The Y-axis is magnified for lower input DNA amounts. A full profile is presented in each panel. 1.0 ng 0.5 ng 0.25 ng ng Figure 7 DNA sample amplified using the Yfiler kit 42 User Bulletin

43 Examples of DNA Profiles Generated on the 3130 Genetic Analyzer Figure 8 shows dilutions for the AmpFlSTR Control DNA 9948 sample amplified using the Profiler Plus kit: 2.0 ng, 1.0 ng, 0.5 ng, and 0.25 ng of input DNA. The Y-axis is magnified for lower input DNA amounts. A full profile is presented in each panel. 2.0 ng 1.0 ng 0.5 ng 0.25 ng Figure 8 DNA sample amplified using the Profiler Plus kit Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 43

44 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Examples of DNA Profiles Generated on the 3130xl Genetic Analyzer Overview Reproducibility This section contains examples of data generated on the 3130xl Genetic Analyzer during the Verification Study. The experiments conducted on the 3130xl Genetic Analyzer are as follows: Reproducibility (page 44) Mixture Studies (page 46) Resolution (page 49) Sensitivity (page 50) All studies were performed using Data Collection Software v3.0 and GeneMapper ID Software v3.2. All DNA samples used in the studies were quantitated using the Quantifiler Human DNA Quantification Kit. Figure 9 shows a comparison of signal intensity across a 3130xl instrument capillary array using AmpFlSTR Control DNA 9948 (1 ng) amplified using the AmpFlSTRYfiler PCR Amplification Kit. Capillary 4 Capillary 10 Capillary 12 Capillary 14 Figure 9 DNA samples amplified using the Yfiler kit 44 User Bulletin

45 Examples of DNA Profiles Generated on the 3130xl Genetic Analyzer Figure 10 shows a comparison of signal intensity across a 3130xl instrument capillary array using AmpFlSTR Control DNA 9948 (2 ng) amplified using the AmpFlSTR Profiler Plus PCR Amplification Kit. Capillary 4 Capillary 10 Capillary 12 Capillary 14 Figure 10 DNA samples amplified using the Profiler Plus kit Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 45

46 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Mixture Studies In Figure 11 on page 47, the first and last panels display profiles of each DNA sample amplified individually with a total DNA input of 1 ng using the AmpFlSTR Identifiler PCR Amplification Kit. (The female AmpFlSTR Control DNA 9947a sample is in the top panel and the male AmpFlSTR Control DNA 9948 sample is in the bottom panel). The other three panels display the mixtures of DNA at approximate ratios of 1:1, 1:5, and 1:10. The panel inset displays the expanded view of the DNA samples mixed at an approximate ratio of 1: User Bulletin

47 Examples of DNA Profiles Generated on the 3130xl Genetic Analyzer 1:0 1:1 1:5 1:10 0:1 Figure 11 DNA samples amplified using the Identifiler kit Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 47

48 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 In Figure 12, the first and last panels display profiles of each DNA sample amplified individually with a total DNA input of 2 ng using the AmpFlSTR Profiler Plus PCR Amplification Kit. (The female AmpFlSTR Control DNA 9947a sample is in the top panel and the male AmpFlSTR Control DNA 9948 sample is in the bottom panel). The other three panels display the mixtures of DNA at approximate ratios of 1:1, 1:5, and 1:10. 1:0 1:1 1:5 1:10 0:1 Figure 12 DNA samples amplified using the Profiler Plus kit 48 User Bulletin

49 Examples of DNA Profiles Generated on the 3130xl Genetic Analyzer Resolution Figure 13 shows three capillaries of a 3130xl instrument 16-capillary array displaying single nucleotide resolution. As shown below, the 9.3 and 10 alleles were resolved and individually detected in all capillaries in which they were run. Capillary 1 Capillary 3 Capillary 5 Figure 13 resolution A 16-capillary array displaying single nucleotide Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 49

50 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Sensitivity Figure 14 shows dilutions for the AmpFlSTR Control DNA 9948 sample amplified using the AmpFlSTR Identifiler PCR Amplification Kit: 1.0 ng, 0.5 ng, 0.25 ng, and ng of input DNA. The Y-axis is magnified for lower input DNA amounts. A full profile is presented in each panel. 1.0 ng 0.5 ng 0.25 ng ng Figure 14 DNA sample amplified using the Identifiler kit 50 User Bulletin

51 Examples of DNA Profiles Generated on the 3130xl Genetic Analyzer Figure 15 shows dilutions for the AmpFlSTR Control DNA 9948 sample amplified using the Yfiler kit: 1.0 ng, 0.5 ng, 0.25 ng, and ng of input DNA. The Y-axis is magnified for lower input DNA amounts. A full profile is presented in each panel. 1.0 ng 0.5 ng 0.25 ng ng Figure 15 DNA sample amplified using the Yfiler kit Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 51

52 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Figure 16 shows dilutions for the AmpFlSTR Control DNA 9948 sample amplified using the Profiler Plus kit: 2.0 ng, 1.0 ng, 0.5 ng, and 0.25 ng of input DNA. The Y-axis is magnified for lower input DNA amounts. A full profile is presented in each panel. 2.0 ng 1.0 ng 0.5 ng 0.25 ng Figure 16 DNA sample amplified using the Profiler Plus kit 52 User Bulletin

53 Genotype Concordance Genotype Concordance The data summarized in Table 1 below were obtained by amplifying the positive Control DNA 007, 9947a, and 9948 samples using the AmpFlSTR Identifiler and Profiler Plus PCR Amplification Kits. Additionally, Control DNA 007 and 9948 samples were amplified with the AmpFlSTR Yfiler PCR Amplification Kit. The samples were prepared for electrophoresis by combining the following: For the four-dye kit: 1 µl of PCR product with 8.5 µl of Hi-Di formamide and 0.5 µl of GeneScan -500 ROX Size Standard For the five-dye kits: 1 µl of PCR product with 8.7 µl of Hi-Di formamide and 0.3 µl of GeneScan -500 LIZ Size Standard Each DNA sample was replicated six times on the plate and injected twice on one 3130 Genetic Analyzer and two 3130xl Genetic Analyzers, for a total of 36 datapoints per allele. The Data Collection Software v3.0 run module for fragment analysis (HIDFragmentAnalysis36_POP4_1) was used to run the samples. Data were analyzed using GeneMapper ID Software v3.2 to evaluate genotype concordance. Table 1 Genotype concordance Control DNA Sample AmpFlSTR Kit Total Number of Alleles Evaluated Concordant Alleles 007 Identifiler kit % 007 Yfiler kit % 007 Profiler Plus kit % 9947a Identifiler kit % 9947a Profiler Plus kit % 9948 Identifiler kit % 9948 Yfiler kit % 9948 Profiler Plus kit % Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 53

54 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v User Bulletin

55 Genotype Concordance Appendix A Maintenance Note: For your convenience, this appendix is reproduced from the Applied Biosystems 3130/3130xl Genetic Analyzers Maintenance, Troubleshooting, and Reference Guide (PN , Rev. B). This appendix covers the following topics: Polymer Delivery Pump Performing Maintenance Tasks Routine Cleaning Resetting the Instrument Moving and Leveling the Instrument Shutting Down the Instrument Wizards Flushing and Filling the Water Trap Fluids and Waste Capillary Array Storing Capillary Arrays Autosampler Calibration Manual Control Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 55

56 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Polymer Delivery Pump Components of the polymer delivery pump (PDP) are identified in the drawing below. Mounting pin Syringe fitting PDP motor PDP motor cover Water seal Waste fitting Water trap Mounting pin Piston Pump chamber Pump block Check valve Capillary array tip Capillary array Capillary array knob Double-tapered ferrule Array port Interconnect tube Polymer supply tube Polymer supply bottle cap with hole Polymer Electrode GR2459 Mounting pin O-ring Buffer valve pin Lower polymer block Mounting pin Overflow hole Buffer fill-line Buffer jar (16-mL anode reservoir) 56 User Bulletin

57 Performing Maintenance Tasks Performing Maintenance Tasks This section lists common tasks required to maintain your Applied Biosystems 3130/3130xl Genetic Analyzers in good working condition. The tasks are divided into tables based on how often you should perform each task. Wear appropriate protection, including gloves, laboratory goggles, and coat whenever you work with the fluids used on this instrument, or parts that may come into contact with these fluids. Daily Tasks Perform these tasks at least once per day. Maintenance Task Ensure adequate levels of buffer and water in reservoirs. Ensure the plate assemblies are properly assembled. IMPORTANT! The holes in the plate retainer must align with the holes in the septa, or the capillary tips will be damaged. Ensure the plate assemblies are positioned on the plate deck properly. Plates should sit snugly on the deck. IMPORTANT! Never use warped plates. Check the level of buffer in the buffer jar. Ensure that the overflow hole faces the front of the instrument and is not occluded. Replace the water and 1 running buffer in the reservoirs on the instrument and make sure that the outside of the assemblies are dry. Check for bubbles in the pump block, lower polymer block, interconnect tube, polymer supply tube, and channels. Remove all bubbles with the Bubble Remove wizard. Check the loading-end header to ensure the capillary tips are not crushed or damaged. Check the level of polymer in the bottle to ensure sufficient volume for runs. Frequency Before each run Before each run Before each run Before each run Every 48 hours Daily or before each run Daily or before each run Daily or before each run Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 57

58 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Maintenance Task Check the pump block and the lower polymer block to ensure they fit securely on the instrument. Clean the instrument surfaces. Check for leaks around the array knob, interconnecting tube nuts, and check valve. Frequency Daily Daily Daily Weekly Tasks Perform these tasks at least once per week. Maintenance Task Replace the polymer using the Replenish Polymer Wizard. Check the storage conditions of the used arrays. Restart the computer and instrument. Frequency Weekly or as needed Weekly Weekly Monthly Tasks Perform these tasks at least once per month. Maintenance Task Run the Water Wash Wizard. Flush the array port during this wizard, whether or not bubbles are present in the array port. Flush the water trap. See Flushing and Filling the Water Trap on page 68. Defragment the hard drive. Frequency Monthly or as needed Monthly or as needed Monthly As-Needed Tasks Perform these tasks as needed. Clean the drip tray. Change the array. Maintenance Task Remove any dried polymer from the capillary tips. Use a lint-free wipe moistened with deionized water. Frequency As needed As needed As needed 58 User Bulletin

59 Routine Cleaning Routine Cleaning General Cleaning 1. Ensure the oven and instrument doors are closed. 2. Press the Tray button on the front of the instrument to move the autosampler to the forward position. IMPORTANT! Never use organic solvents to clean the instrument. Tray button 3. Wipe off any liquid on or around the autosampler using a lintfree tissue. 4. Clean off any polymer build-up (crystals) on the instrument including the capillary tips and the stripper plate with deionized water and lint-free tissue. 5. Clean the array port knob, plug, or opening threads of these parts with moistened lab wipes. 6. Clean out the drip trays with deionized water and lint-free tissue. Resetting the Instrument Reset the instrument when: A fatal error as indicated by the red status light The instrument does not respond to the Data Collection software Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 59

60 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Two procedures can reset the instrument: Press the reset button through the pin hole on the front of the instrument to dump and reload the firmware and to reset the electronics. Try this method first. Shut down and restart the computer and the instrument. Resetting With the Reset Button 1. Close the instrument doors. 2. Using a long narrow implement, such as a straightened paper clip, press the reset button on the front of the instrument. Reset button Resetting by Powering Down 1. Close the instrument doors. 2. Power off the instrument by pressing the on/off button on the front of the instrument. On/Off button 3. Restart the computer. a. Select Start > Turn off Computer. 60 User Bulletin

61 Moving and Leveling the Instrument b. In the dialog box, select Restart, then click OK. IMPORTANT! Wait until the computer has completely restarted before proceeding. 4. Turn on the instrument, then wait for the solid green light. 5. Launch the Data Collection software (Service Console applications start automatically). Moving and Leveling the Instrument PHYSICAL INJURY HAZARD. Do not attempt to lift the instrument or any other heavy objects unless you have received related training. Incorrect lifting can cause painful and sometimes permanent back injury. Use proper lifting techniques when lifting or moving the instrument. Two or three people are required to lift the instrument, depending upon instrument weight. 6. Remove the following components from the instrument: Any plate assemblies from the autosampler. Water and buffer reservoirs from the autosampler. Capillary array, by selecting Instrument Shutdown Wizard. (See Performing a Long-Term Shutdown on page 65.) Anode buffer reservoir. 7. Switch off the breaker on the back of the instrument. 8. Disconnect the power cord and the Ethernet cable. IMPORTANT! While moving the instrument, avoid any shock or vibration. 9. Move the instrument. 10. Place the bubble level on the autosampler deck. 11. Turn the instrument legs to level the instrument. To move the instrument corner... up down Turn the leg... right (clockwise) left (counterclockwise) Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 61

62 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Shutting Down the Instrument Perform the appropriate shutdown procedure based on the information in the following table: If the instrument will be unattended for... no more than 1 week with a full buffer reservoir Perform this shutdown procedure... Short-term IMPORTANT! The key to a successful shortterm shutdown is keeping the capillary array in 1 running buffer. This prevents the polymer from drying in the capillaries. CHEMICAL HAZARD. 1 Genetic Analyzer Buffer with EDTA may cause eye, skin, and respiratory tract irritation. Read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. for more than 1 week Long-term 62 User Bulletin

63 Shutting Down the Instrument Performing a Short-Term Shutdown Fill the Capillary With Fresh Polymer Using Manual Control 1. Ensure the oven and instrument doors are closed. 2. Collect polymer waste: a. Click GA Instruments > ga3130 or ga3130xl> instrument name> Manual Control. b. In the Send Defined Command drop-down menu, select Autosampler. c. In the Command Name drop-down menu, select Move autosampler to site. d. In the Value menu, select Waste. e. Click Send Command. Wait for the autosampler to stop moving and Send Command becomes active, before proceeding. 3. Fill the capillaries: a. In the Send Defined Command for drop-down menu, select Polymer Delivery Pump. b. In the Command Name, select the appropriate Fill <length> cm capillary array length. c. Click Send Command.The array fill is finished when Send Command becomes active. d. Return the buffer reservoir to the capillaries. Cleaning the Reservoirs 1. Press the Tray button to move the autosampler forward. 2. Open the doors, then remove the: Plates Cathode buffer reservoir and water reservoirs 3. Dispose of remaining fluids and rinse out the reservoirs with deionized water. Note: Follow your company s waste disposal practices for appropriate disposal procedures. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 63

64 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v Rinse the cathode reservoir with 1 running buffer, and then fill to the line with 1 running buffer (about 16 ml). CHEMICAL HAZARD. 1 Genetic Analyzer Buffer with EDTA may cause eye, skin, and respiratory tract irritation. Read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. Fill line GR1857b 5. Fill the three water reservoirs to the line with quality deionized water (about 16 ml). Be sure that the septa fit snugly and flush on the tops of the reservoirs in order to prevent damaging the capillary tips. 6. Place a clean reservoir septa on each reservoir, and dry the outside of the reservoirs using a lint-free wipe. 7. Place the reservoirs into position on the autosampler as shown below. Water reservoir (waste) Cathode reservoir (1 running buffer) Water reservoir (rinse) Water reservoir (spare) Home position 8. Close the instrument doors. Note: Closing the doors returns the autosampler to the home position, placing the tips of the capillaries in buffer. 9. Shut down the computer and turn off the instrument. 64 User Bulletin

65 Shutting Down the Instrument Performing a Long-Term Shutdown IMPORTANT! Make sure all parts are completely dry before longterm storage. Select Instrument Shutdown Wizard and follow the prompts. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 65

66 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Wizards Accessing Wizards In the tree pane of the Data Collection software, click GA Instruments > ga3130 or ga3130xl > instrument name or any topic name below instrument name to see Wizards in the menu bar. The wizards in the Data Collection software guide you through several maintenance procedures. If plates are linked in the Run Scheduler and you complete a wizard, the plates automatically unlink. You will get a warning dialog box. Click, and then relink the plate if applicable. General Use Guidelines The following table lists the wizards and when to use them. Wizard Use to... Install Array Install a capillary array: On a new instrument To reactivate an instrument that has been shut down Replace an installed capillary array with another capillary array 66 User Bulletin

67 Wizards Wizard Change Polymer Type Replenish Polymer Bubble Remove Use to... Change to a different polymer type than the one presently being used Replenish the polymer supply Replace the polymer in the PDP with polymer of the same or different lot Enter polymer information when Data Collection software is installed or upgraded Remove bubbles in the PDP chamber, channels, and tubing Water Wash Wash the PDP chamber, lower polymer block a, channels, and tubing with water: As part of a monthly maintenance protocol To remove any suspected contaminants in the PDP To remove persistent bubbles (followed by the Bubble Remove Wizard, if needed) To replace old polymer in the PDP Instrument Shutdown Autosampler Calibration Update Cap Array Info Prepare the instrument for a period of disuse of greater than one week Calibrate the autosampler positions Update the capillary array information and the serial number Correct an entry mistake after using a wizard a. The lower polymer block is cleaned on the instrument using this wizard and should not be removed. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 67

68 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Flushing and Filling the Water Trap The PDP water trap should be flushed with either distilled or deionized water at least once per month to wash out any diluted polymer and to clear bubbles. Leave the trap filled with either distilled or deionized water. To flush the water seal trap: 1. Fill the supplied 20 ml, all-plastic Luer lock syringe (PN ) with distilled or deionized water. Expel any bubbles from the syringe. IMPORTANT! Do not use a syringe smaller than 20 ml. Doing so may generate excessive pressure within the trap. 2. Attach the syringe to the forward-facing Luer fitting at the top of the pump block. Hold the fitting with one hand while threading the syringe onto the fitting with the other hand. 3. Open the Luer fitting by grasping the body of the fitting and turning it and the attached syringe approximately one-half turn counterclockwise. 4. Open the exit fitting at the top left side of the pump block by turning it approximately one-half turn counterclockwise. 5. Hold an empty tube or beaker under the exit fitting to receive approximately 5 ml of waste. Flush the trap by pushing steadily on the syringe plunger. IMPORTANT! DO NOT USE EXCESSIVE FORCE when you push the syringe plunger as this may damage the trap seals. Take approximately 30 seconds to flush 5 ml of either distilled or deionized water through the trap. Note: Because the water trap volume is approximately 325 µl, a relatively small volume of water is adequate for complete flushing. However, a larger volume only improves flushing as long as force and flow rate are kept within the limits given above. 68 User Bulletin

69 Flushing and Filling the Water Trap 6. Close the fittings in this order by turning each clockwise until the fittings seal against the block: a. Luer fitting b. Exit fitting IMPORTANT! Do not over-tighten the fittings. Very little pressure develops within the trap during pump operation, so the fittings require only enough tightening to prevent water leaks. Excessive tightening can damage the fittings. c. Remove the syringe from the Luer fitting. Hold the fitting with one hand while turning the syringe counterclockwise with the other hand. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 69

70 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Fluids and Waste Buffer When to Change the Buffer Making Buffer for a Single Run CHEMICAL HAZARD. 10 Genetic Analyzer Buffer with EDTA may cause eye, skin, and respiratory tract irritation. Read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. Change the buffer before each batch of runs or at least every 24 hours. To prepare 50 ml of 1 running buffer: 1. Into a graduated cylinder, add: a. 5 ml of 10 Genetic Analyzer buffer b. 45 ml of deionized water to bring the total volume up to 50 ml. 50 ml 45 ml of DI H 2 O 50-mL graduated cylinder 5 ml of 10 Genetic Analyzer buffer 2. Mix well. 70 User Bulletin

71 Fluids and Waste Storing the Buffer The 1 running buffer can be stored at: 2 to 8 C for up to 1 month Room temperature for 1 week Buffer Storage Conditions Option A 2 C to 8 C Option B 20 C to 25 C Su M T W Th F S Su M T W Th F S 1 month 7 days Polymer Storing Polymer Store any remaining POP polymer at 2 to 8 C until the expiration date printed on the jar. Note: Excessively hot environments may shorten the working life of the polymer. When to Change the Polymer Change the polymer weekly. The polymer is good at 25 C for about 7 days. POP-4 or POP-6 Polymer Storage Conditions Su M T W Th F S 20 C to 25 C 7 days Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 71

72 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Before Using the Polymer Replenishing the Polymer 1. Remove the polymer from 4 C storage. 2. Loosen the cap and bring the polymer to room temperature. 3. To dissolve deposits, tighten the cap and gently swirl the polymer. CHEMICAL HAZARD. POP polymers cause eye, skin, and respiratory tract irritation. Read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. IMPORTANT! Wear gloves when you handle the polymer. 1. Select Wizards > Replenish Polymer Wizard. 2. Follow the directions given in the wizard to load fresh polymer on the instrument. 3. Relink plate(s), if applicable. 72 User Bulletin

73 Fluids and Waste Changing to a Different Polymer Type CHEMICAL HAZARD. POP polymers cause eye, skin, and respiratory tract irritation. Read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. IMPORTANT! Wear gloves when you handle the polymer. 1. Select Wizards > Change Polymer Type Wizard. 2. Follow the wizard prompts. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 73

74 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Capillary Array When to Change a Capillary Array Checking the Cathode Bar A capillary array should last approximately 100 runs. The following problems may indicate that a new capillary array is required: Poor sizing precision or allele calling Poor resolution and/or decreased signal intensity ELECTRICAL SHOCK/FIRE HAZARD. Do not leave liquid on the cathode header. This can lead to electric shock or even fire if not properly maintained. When placing a used array back on the instrument, be sure that the cathode bar is dry (see figure below). A wet bar could lead to arcing. Back view: Ensure the cathode header is dry especially in the center GR xl instrument capillary array 3130 instrument capillary array 74 User Bulletin

75 Capillary Array Installing, Removing, or Replacing a Capillary Array IMPORTANT! Wear gloves when you handle the polymer blocks. CHEMICAL HAZARD. POP polymers cause eye, skin, and respiratory tract irritation. Read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. 1. Close the oven and instrument doors, and then press the Tray button. 2. Select Wizards > Install Array Wizard. IMPORTANT! The capillary array length defined in the wizard must match the array length you are using. 3. Open instrument and oven doors. 4. Follow the directions given in the wizard to install or replace an array. 5. Click when done. 6. Close and lock the oven door, then close the instrument doors. IMPORTANT! If you installed or replaced an array that is a different length than the one you were using, you must reset the active spectral calibration or create a new spectral calibration for the dye set and array length combination (see Activating a Spectral Calibration in the Applied Biosystems 3130/3130xl Genetic Analyzers Getting Started Guide). 7. Relink plate(s), if applicable. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 75

76 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Updating Capillary Array Information Caring for the Capillary Array and Work Area Use the Update Cap Array Info wizard to: Update the capillary array length and serial number information into the database Correct an entry error after using another wizard IMPORTANT! The capillary array length defined in the wizard must match the array length you are using. Follow these guidelines to properly care for the capillary array and area: Wear gloves and handle the capillary array gently. Do not touch the detection cell. Keep the ends of the capillary array wet at all times. Do not overtighten the capillary array knob. Clean off any polymer buildup (crystals) on the instrument, including the capillary electrodes and the stripper plate, with deionized water and lint-free tissue. Keep clean CHEMICAL HAZARD. POP polymers cause eye, skin, and respiratory tract irritation. Read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. Note: When cleaning the capillary electrodes, be careful not to bend them out of position. If the electrodes do get bent, follow the procedure Storing Capillary Arrays on page 77. IMPORTANT! Never use organic solvents to clean the instrument. 76 User Bulletin

77 Storing Capillary Arrays Filling the Capillary Array Using Manual Control See Fill the Capillary With Fresh Polymer Using Manual Control on page 63. Storing Capillary Arrays Storing a Capillary Array on the Instrument Storing a Capillary Array off the Instrument Store the capillary array on the instrument only when the capillary array will be unused for less than 1 week. To store the capillary array on the instrument, follow the instructions to perform a short-term shutdown described on page 63. Store the capillary array off of the instrument when the capillary array will be unused for longer than 1 week. IMPORTANT! Before storing the capillary array for long periods, fill the capillaries with fresh polymer. IMPORTANT! Wear gloves while performing the following procedure, and any other time you handle the capillary array, septa, or buffer reservoirs. CHEMICAL HAZARD. POP polymers cause eye, skin, and respiratory tract irritation. Read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. CHEMICAL HAZARD. 1 Genetic Analyzer Buffer with EDTA may cause eye, skin, and respiratory tract irritation. Read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 77

78 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v Remove the capillary array from the instrument by selecting Install Array Wizard. 2. Select Store Array and follow the prompts. 3. Replace the cover over the detection cell. 4. Fill a buffer reservoir with fresh 1 running buffer and cover with a septa strip. Insert the capillary tips into the buffer. 5. Fill the shipping vial with fresh 1 running buffer and insert the detection end of the capillary array. 6. Store the capillary array upright. 7. Check the 1 running buffer level in the reservoir and tube weekly and replenish the buffer as needed. 78 User Bulletin

79 Storing Capillary Arrays Removing Bubbles from the Polymer Blocks Bubbles may be found in the polymer system, especially after a polymer change or array installation. Remove the bubbles from all parts of the polymer system including the pump chamber, the pump block channel, polymer supply and interconnect tubing and the lower polymer block channel. To clear bubbles: 1. Select Wizards > Bubble Remove Wizard to clear bubbles. IMPORTANT! Remove bubbles from the interconnect tubing and the channel of the lower polymer block. These areas are part of the electrophoresis current path. Absence of bubbles in the current path is important for problem-free electrophoresis. 2. Replace the 1 running buffer if excess polymer is expelled into the anode buffer jar during bubble removal. CHEMICAL HAZARD. 1 Genetic Analyzer Buffer with EDTA may cause eye, skin, and respiratory tract irritation. Read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 79

80 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Autosampler Calibration When to Calibrate the Autosampler Calibrating the Autosampler Calibrate the autosampler only as needed. Symptoms of autosampler alignment problems may include: Poor injection for a small number of capillaries Low signal strength No evidence of sample 1. Close the oven and instrument doors. 2. Select Wizards > Autosampler Calibration Wizard. 3. If plates are linked in the Run Scheduler, the plates automatically are unlinked. In the Warning dialog box, click. 4. Follow the directions given in the wizard to calibrate the autosampler. 80 User Bulletin

81 Autosampler Calibration 5. Click. 6. Turn the instrument power off for 10 sec, then on. Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 81

82 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Manual Control Note: Manual control is active only if the oven and instrument doors are closed. Table of Commands The following table displays the manual control options in the Data Collection software. Command Function Command Options Value Electrophoresis Turn On/Off power supply Set voltage Read voltage Read current On Off A number between 0.0 and 15 kv N/A Laser Set laser state Idle On Off Set laser power Read laser power setting A number between 0 and 25 mw N/A Read laser power Read laser current Open/Close shutter Open Closed Oven Set state On Off Set temperature A number between 18 and 65 C 82 User Bulletin

83 Manual Control Command Function Command Options Value Autosampler Polymer Deliver Pump Buffer Valve Detection Cell Heater Initialize autosampler Bring autosampler to forward position Initialize and return to previous position Move autosampler up/down Move autosampler to site Initialize polymer delivery pump Home piston Read piston position Move piston down Move piston up Fill capillary array Close /Open buffer valve Turn On/Off detection cell heater Read detection cell heater temperature N/A A number between 500 and 500 steps Buffer (left, front for 1 running buffer), home position Water1 (left, rear for deionized water) Water2 (right, front for deionized water) Waste (right, rear for deionized water) N/A 1 to counts 1 to counts 1 to counts Close Open On Off N/A Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products 83

84 Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 Command Function Command Options Value Oven Turn On/Off oven On Off Set oven temperature Read oven temperature A number between 18 and 70 C N/A Using Manual Control Note: Manual control functions cannot be used during a run. 1. In the tree pane of the Data Collection software, click GA Instruments > ga3130 or ga3130xl> instrument name> Manual Control. 2. In the Send Defined Command For drop-down list, select a function. 84 User Bulletin

Development and Validation of the 3500 Series Genetic Analyzer for Human Identification

Development and Validation of the 3500 Series Genetic Analyzer for Human Identification Development and Validation of the 3500 Series Genetic Analyzer for Human Identification Jeff Sailus Forensic Science Applications Group 1 Development of a Next Generation Genetic Analyzer A Whole System

More information

HID Validation: The 3500 Series Genetic Analyzers

HID Validation: The 3500 Series Genetic Analyzers Outline Scope Materials and Reagents Validation Elements Selection of Validation Data Additional Sources of Validation Information User Bulletin 2 Validation Scope Hardware and Reagents HID Validation:

More information

PowerPlex S5 System INSTRUCTIONS FOR USE OF PRODUCTS DC6951 AND DC6950.

PowerPlex S5 System INSTRUCTIONS FOR USE OF PRODUCTS DC6951 AND DC6950. Technical Manual PowerPlex S5 System INSTRUCTIONS FOR USE OF PRODUCTS DC6951 AND DC6950. PRINTED IN USA. PowerPlex S5 System All technical literature is available on the Internet at: www.promega.com/protocols/

More information

Procedure for Use of the 3130XL Genetic Analyzer

Procedure for Use of the 3130XL Genetic Analyzer Procedure for Use of the 3130XL Genetic Analyzer 1.0 Purpose- This procedure specifies the steps for daily maintenance, preparing and running samples on the 3130XL ABI Genetic Analyzer. 2.0 Scope- This

More information

User Bulletin. Veriti 96-Well Thermal Cycler AmpFlSTR Kit Validation. Overview

User Bulletin. Veriti 96-Well Thermal Cycler AmpFlSTR Kit Validation. Overview User Bulletin Veriti 96-Well Thermal Cycler AmpFlSTR Kit Validation June 2009 SUBJECT: AmpFlSTR PCR Amplification Kit validation on the Veriti 96-Well Thermal Cycler with 0.2 ml sample block format In

More information

ABI PRISM 3100 Genetic Analyzer. ABI PRISM cm Capillary Array for High Throughput Microsatellite and SNP Genotyping

ABI PRISM 3100 Genetic Analyzer. ABI PRISM cm Capillary Array for High Throughput Microsatellite and SNP Genotyping User Bulletin ABI PRISM 3100 Genetic Analyzer November 12, 2001 SUBJECT: ABI PRISM 3100 22-cm Capillary Array for High Throughput Microsatellite and SNP Genotyping About This Bulletin High throughput Microsatellite

More information

PowerPlex 4C Matrix Standard

PowerPlex 4C Matrix Standard TECHNICAL MANUAL PowerPlex 4C Matrix Standard Instruc ons for Use of Product DG4800 Printed 10/15 TMD048 PowerPlex 4C Matrix Standard All technical literature is available at: www.promega.com/protocols/

More information

Use of the GenePrint 10 System with the Applied Biosystems 3500 and 3500xL Genetic Analyzers and POP-7 Polymer

Use of the GenePrint 10 System with the Applied Biosystems 3500 and 3500xL Genetic Analyzers and POP-7 Polymer Use of the GenePrint 10 System with the Applied Biosystems 3500 and 3500xL Genetic Analyzers and POP-7 Polymer GenePrint 10 System Application Note The GenePrint 10 System is Not For Medical Diagnostic

More information

PowerPlex 5-Dye Matrix Standards, 3100/3130

PowerPlex 5-Dye Matrix Standards, 3100/3130 TECHNICAL BULLETIN PowerPlex 5-Dye Matrix Standards, 3100/3130 Instructions for Use of Product DG4700 Note: These matrix fragments are compatible with the ABI PRISM 3100 and 3100-Avant and Applied Biosystems

More information

AmpFlSTR NGM SElect PCR Amplification Kit. User Guide

AmpFlSTR NGM SElect PCR Amplification Kit. User Guide AmpFlSTR NGM SElect PCR Amplification Kit User Guide Information in this document is subject to change without notice. Applied Biosystems assumes no responsibility for any errors that may appear in this

More information

PowerPlex Matrix Standards, 3100/3130

PowerPlex Matrix Standards, 3100/3130 TECHNICAL BULLETIN PowerPlex Matrix Standards, 3100/3130 Instruc ons for use of Product DG4650 Note: The PowerPlex Matrix Standards, 3100/3130, can be used to perform spectral calibra on on the Applied

More information

Use of the MSI Analysis System with the Applied Biosystems 3500 Genetic Analyzer and POP-7 Polymer

Use of the MSI Analysis System with the Applied Biosystems 3500 Genetic Analyzer and POP-7 Polymer Use of the MSI Analysis System with the Applied Biosystems 3500 Genetic Analyzer and POP-7 Polymer MSI Analysis System Application Note Promega Corporation Notes: The MSI Analysis System is For Research

More information

Polymer Block Cleaning Kit

Polymer Block Cleaning Kit Polymer Block Cleaning Kit For Applied Biosystems 3730/3730xl DNA Analyzers Protocol April 29, 2003 11:17 am, Title.fm Copyright 2003, Applied Biosystems. All rights reserved. For Research Use Only. Not

More information

AmpFlSTR NGM PCR Amplification Kit

AmpFlSTR NGM PCR Amplification Kit USER GUIDE AmpFlSTR NGM PCR Amplification Kit Publication Part Number 4466844 Rev. A Revision Date April 2011 AmpFlSTR NGM PCR Amplification Kit User s Guide Copyright 2011 Life Technologies Corporation.

More information

AmpFlSTR NGM PCR Amplification Kit. User s Guide

AmpFlSTR NGM PCR Amplification Kit. User s Guide AmpFlSTR NGM PCR Amplification Kit User s Guide AmpFlSTR NGM PCR Amplification Kit User s Guide Copyright 2009, 2010 Life Technologies Corporation. All rights reserved. For Research, Forensic, or Paternity

More information

PowerPlex ESI 16 System

PowerPlex ESI 16 System TECHNICAL MANUAL PowerPlex ESI 16 System Instructions for use of Products DC6770 and DC6771 TMD027 PowerPlex ESI 16 System All technical literature is available on the Internet at: www.promega.com/protocols/

More information

PowerPlex ESI 16 System

PowerPlex ESI 16 System TECHNICAL MANUAL PowerPlex ESI 16 System Instruc ons for use of Products DC6770 and DC6771 TMD027 PowerPlex ESI 16 System All technical literature is available at: www.promega.com/protocols/ Please visit

More information

TECHNICAL MANUAL. PowerPlex 16 System. Instructions for Use of Products DC6530 and DC6531. Revised 5/16 TMD012

TECHNICAL MANUAL. PowerPlex 16 System. Instructions for Use of Products DC6530 and DC6531. Revised 5/16 TMD012 TECHNICAL MANUAL PowerPlex 16 System Instructions for Use of Products DC6530 and DC6531 TMD012 PowerPlex 16 System All technical literature is available on the Internet at: www.promega.com/protocols/ Please

More information

PowerPlex 5-Dye Matrix Standards, 310

PowerPlex 5-Dye Matrix Standards, 310 TECHNICAL BULLETIN PowerPlex 5-Dye Matrix Standards, 310 Instructions for Use of Product DG4600 Revised 3/16 TBD023 PowerPlex 5-Dye Matrix Standards, 310 All technical literature is available at: www.promega.com/protocols/

More information

PowerPlex ESX 17 System

PowerPlex ESX 17 System TECHNICAL MANUAL PowerPlex ESX 17 System Instructions for use of Products DC6720 and DC6721 TMD024 PowerPlex ESX 17 System All technical literature is available at: www.promega.com/protocols/ Please visit

More information

PowerPlex ESX 16 System

PowerPlex ESX 16 System TECHNICAL MANUAL PowerPlex ESX 16 System Instructions for use of Products DC6710 and DC6711 TMD023 PowerPlex ESX 16 System All technical literature is available on the Internet at: www.promega.com/protocols/

More information

PowerPlex ESX 16 System

PowerPlex ESX 16 System TECHNICAL MANUAL PowerPlex ESX 16 System Instruc ons for use of Products DC6710 and DC6711 TMD023 PowerPlex ESX 16 System All technical literature is available on the Internet at: www.promega.com/protocols/

More information

AmpFlSTR Identifiler Plus

AmpFlSTR Identifiler Plus AmpFlSTR Identifiler Plus PCR Amplification Kit User s Guide Copyright 2009, 2010, 2011 Life Technologies. All rights reserved. For Research, Forensic, or Paternity Use Only. Not for use in diagnostic

More information

PowerPlex 5C Matrix Standards, 310

PowerPlex 5C Matrix Standards, 310 TECHNICAL MANUAL PowerPlex 5C Matrix Standards, 310 Instruc ons for Use of Product DG5640 Printed 10/15 TMD050 PowerPlex 5C Matrix Standards, 310 All technical literature is available at: www.promega.com/protocols/

More information

AmpFlSTR Identifiler PCR Amplification Kit

AmpFlSTR Identifiler PCR Amplification Kit Product Bulletin Human Identification AmpFlSTR Identifiler PCR Amplification Kit Fifteen STR (short tandem repeat) loci and Amelogenin co-amplified in a single tube Incorporation of the same proven, reliable

More information

PowerPlex CS7 System

PowerPlex CS7 System TECHNICAL MANUAL PowerPlex CS7 System Instructions for Use of Product DC6613 TMD042 PowerPlex CS7 System All technical literature is available on the Internet at: www.promega.com/protocols/ Please visit

More information

PowerPlex ESI 17 Pro System

PowerPlex ESI 17 Pro System TECHNICAL MANUAL PowerPlex ESI 17 Pro System Instructions for Use of Products DC7780 and DC7781 Revised 5/17 TMD036 PowerPlex ESI 17 Pro System All technical literature is available at: /protocols/ Visit

More information

Maximizing the Performance of Capillary Electrophoresis Systems The Maximizing Data Quality Series Part 4

Maximizing the Performance of Capillary Electrophoresis Systems The Maximizing Data Quality Series Part 4 Maximizing the Performance of Capillary Electrophoresis Systems The Maximizing Data Quality Series Part 4 Introduction Since the introduction of the ABI PRISM 310 Genetic Analyzer in 1996, capillary electrophoresis

More information

GlobalFiler TM Extra Cycle Evaluation Report

GlobalFiler TM Extra Cycle Evaluation Report GlobalFiler TM Extra Cycle Evaluation Report Table of Content Chapter 1 Summary 1. Introduction... 2 2. Systems, Kits and Conditions... 3 3. Studies... 4 Chapter 2 Minimum Threshold and Background Study

More information

TECHNICAL MANUAL. GenePrint 24 System. Instructions for Use of Product B1870 6/16 TM465

TECHNICAL MANUAL. GenePrint 24 System. Instructions for Use of Product B1870 6/16 TM465 TECHNICAL MANUAL GenePrint 24 System Instructions for Use of Product B1870 6/16 TM465 GenePrint 24 System All technical literature is available at: www.promega.com/protocols/ Visit the web site to verify

More information

Multiplex analysis of 10 STR loci plus Amelogenin

Multiplex analysis of 10 STR loci plus Amelogenin Multiplex analysis of 10 STR loci plus Amelogenin Page 1 of 15 Table of Content 1 Product Information.3 1.1 Product Description 3 1.2 Ordering Information and Kit Components 5 1.3 Storage Conditions..5

More information

Applied Biosystems 7900HT Fast Real-Time PCR System. Performing Fast Gene Quantitation with 384-Well Plates DRAFT

Applied Biosystems 7900HT Fast Real-Time PCR System. Performing Fast Gene Quantitation with 384-Well Plates DRAFT User Bulletin Applied Biosystems 7900HT Fast Real-Time PCR System October 3, 2005 SUBJECT: Performing Fast Gene Quantitation with 384-Well Plates In This User Bulletin System Overview System Requirements

More information

Investigator HDplex Handbook

Investigator HDplex Handbook November 2012 Investigator HDplex Handbook For multiplex amplification of 12 STR loci with high discriminatory power, plus Amelogenin Sample & Assay Technologies QIAGEN Sample and Assay Technologies QIAGEN

More information

TaqMan SNP Genotyping Protocol

TaqMan SNP Genotyping Protocol Precipio, Inc. TaqMan SNP Genotyping Protocol ICE COLD-PCR Product Analysis with TaqMan SNP Genotyping Assays using the ABI 7900HT System Table of Contents TaqMan SNP Genotyping Guidelines (ABI 7900HT

More information

AmpFlSTR SEfiler. User s Manual. PCR Amplification Kit. DRAFT March 22, :04 pm, SEfiler_TitlePage.fm

AmpFlSTR SEfiler. User s Manual. PCR Amplification Kit. DRAFT March 22, :04 pm, SEfiler_TitlePage.fm AmpFlSTR SEfiler PCR Amplification Kit User s Manual March 22, 2012 2:04 pm, SEfiler_TitlePage.fm 2012 Life Technologies Corporation. All rights reserved. Printed in the U.S.A. TRADEMARKS: AB (Design),

More information

PowerPlex Y23 System for Use on the Applied Biosystems Genetic Analyzers

PowerPlex Y23 System for Use on the Applied Biosystems Genetic Analyzers TECHNICAL MANUAL PowerPlex Y23 System for Use on the Applied Biosystems Genetic Analyzers Instructions for Use of Products DC2305 and DC2320 Revised 4/17 TMD035 PowerPlex Y23 System for Use on the Applied

More information

High Resolution Melting Protocol

High Resolution Melting Protocol Precipio, Inc. High Resolution Melting Protocol High Resolution Melting of ICE COLD-PCR Products Table of Contents High Resolution Melting of the ICE COLD-PCR products 2 Mutation Confirmation of HRM Positive

More information

Investigator Quantiplex HYres Kit

Investigator Quantiplex HYres Kit These specifications are for the procurement of a validation study at GBI Division of Forensic Sciences facilities of two commercially available kits used in forensic DNA analysis. The two kits to be validated

More information

Standard Sensitivity NGS Fragment Analysis Kit (1 bp 6000 bp) User Guide (DNF ) (DNF )

Standard Sensitivity NGS Fragment Analysis Kit (1 bp 6000 bp) User Guide (DNF ) (DNF ) New Software Version Refer to red highlighted text for additional updates Standard Sensitivity NGS Fragment Analysis Kit (1 bp 6000 bp) User Guide (DNF-473-0500) (DNF-473-1000) For use with the Fragment

More information

Relative Fluorescent Quantitation on Capillary Electrophoresis Systems:

Relative Fluorescent Quantitation on Capillary Electrophoresis Systems: Application Note Relative Fluorescent Quantitation on the 3130 Series Systems Relative Fluorescent Quantitation on Capillary Electrophoresis Systems: Screening for Loss of Heterozygosity in Tumor Samples

More information

High Sensitivity NGS Fragment Analysis Kit (1 bp 6000 bp) User Guide (DNF ) (DNF )

High Sensitivity NGS Fragment Analysis Kit (1 bp 6000 bp) User Guide (DNF ) (DNF ) New Software Version Refer to red highlighted text for additional updates High Sensitivity NGS Fragment Analysis Kit (1 bp 6000 bp) User Guide (DNF-474-0500) (DNF-474-1000) For use with the Fragment Analyzer

More information

The injection of the 28 and 29 PCR cycle data was performed at 3kV for 10 and 5 seconds.

The injection of the 28 and 29 PCR cycle data was performed at 3kV for 10 and 5 seconds. TECHNICAL NOTE Evaluation of a decreased cycle number (28 cycles) and decreased injection time (5 seconds) when the GlobalFiler PCR Amplification Kit is run on a 3130xl Genetic Analyzer The performance

More information

Sanger Sequencing Portfolio

Sanger Sequencing Portfolio Molecular Cloning Laboratories Sanger Sequencing Portfolio Significant Cost Savings vs ABI Uncompromised Performance Longer Read Length Longer Shelf Life No Recalibrations Necessary PAGE 1 BrightDye Terminator

More information

Volume of Each Kit Reagent (ml) XTerminator Solution. 2-mL mL 1, mL 2,

Volume of Each Kit Reagent (ml) XTerminator Solution. 2-mL mL 1, mL 2, BigDye XTerminator Purification Kit Quick Reference Card For safety and biohazard guidelines, refer to the Safety section in the Applied Biosystems BigDye XTerminator Purification Kit Protocol (PN 4374408).

More information

PowerPlex ESI 16 Fast System for Use on the Applied Biosystems Genetic Analyzers

PowerPlex ESI 16 Fast System for Use on the Applied Biosystems Genetic Analyzers TECHNICAL MANUAL PowerPlex ESI 16 Fast System for Use on the Applied Biosystems Genetic Analyzers Instructions for Use of Products DC1620 and DC1621 Revised 4/17 TMD044 PowerPlex ESI 16 Fast System for

More information

Capillary Electrophoresis in DNA Analysis

Capillary Electrophoresis in DNA Analysis Capillary Electrophoresis in DNA Analysis john.butler@nist.gov Troubleshooting the ABI 310 NEAFS Workshop Mystic, CT September 29-30, 2004 Dr. John M. Butler Dr. Bruce R. McCord mccordb@fiu.edu Outline

More information

PowerPlex ESX 17 Fast System for Use on the Applied Biosystems Genetic Analyzers

PowerPlex ESX 17 Fast System for Use on the Applied Biosystems Genetic Analyzers TECHNICAL MANUAL PowerPlex ESX 17 Fast System for Use on the Applied Biosystems Genetic Analyzers Instructions for Use of Products DC1710 and DC1711 Revised 8/17 TMD040 PowerPlex ESX 17 Fast System for

More information

Procedure for PCR Amplification for Casework

Procedure for PCR Amplification for Casework Procedure for PCR Amplification for Casework 1.0 Purpose - This procedure specifies the steps for performing PCR amplification with AmpFlSTR Identifiler Plus. 2.0 Scope - This procedure applies to casework

More information

NRAS Mutation Analysis Reagents (Codons 12 and 13)

NRAS Mutation Analysis Reagents (Codons 12 and 13) NRAS Mutation Analysis Reagents (Codons 12 and 13) User Manual V1.1 Cat No. GP18 32 reactions 1 CONTENTS Introduction 4 Overview of Mutector TM Assay 5 Materials Provided 6 Materials Required 7 Equipment

More information

Data Quality Worth Sharing

Data Quality Worth Sharing Product Bulletin Human Identification AmpFlSTR NGM and NGM SElect PCR Amplification Kits Next generation amplification chemistries including the 5 new loci from the expanded European Standard Set Enhanced

More information

MSI Analysis System, Version 1.2

MSI Analysis System, Version 1.2 TECHNICAL MANUAL MSI Analysis System, Version 1.2 Instructions for Use of Product MD1641 For Research Use Only. Not for use in diagnostic procedures. Revised 11/17 TM255 MSI Analysis System, Version 1.2

More information

Quantifiler Human DNA Quantification Kit Quantifiler Y Human Male DNA Quantification Kit

Quantifiler Human DNA Quantification Kit Quantifiler Y Human Male DNA Quantification Kit Product Bulletin Human Identification Quantifiler Human DNA Quantification Kit Quantifiler Y Human Male DNA Quantification Kit The Quantifiler kits produce reliable and reproducible results, helping to

More information

GenPlex HID Training Class I

GenPlex HID Training Class I Rixun Fang GenPlex HID Training Class I Outline of Presentation Introduction GenPlex HID kit Experimental plan Class schedule Forensic SNP Analysis GenPlex HID Training Class I 2 Potential Forensic Applications

More information

NRAS Codon 61 Mutation Analysis Reagents

NRAS Codon 61 Mutation Analysis Reagents NRAS Codon 61 Mutation Analysis Reagents User Manual V1.0 Cat No. GP19 32 reactions 1 CONTENTS Introduction 4 Overview of Mutector TM Assay 5 Materials Provided 6 Materials Required 7 Equipment Required

More information

Mentype Nonaplex I PCR Amplification Kit

Mentype Nonaplex I PCR Amplification Kit Mentype Nonaplex I PCR Amplification Kit Product description The Mentype Nonaplex I PCR Amplification Kit is a multiplex application for the Short Tandem Repeat (STR) loci, which belong to the German Forensic

More information

BRCA MAQ USER GUIDE Version 1.0

BRCA MAQ USER GUIDE Version 1.0 BRCA MAQ USER GUIDE Version 1.0 For Copy Number Analysis Assays IFU604 v170797 www.multiplicom.com 2012 Multiplicom NV, all rights reserved Table of Contents Introduction to Multiplex Amplicon Quantification...2

More information

Mentype Pentaplex ESS PCR Amplification Kit

Mentype Pentaplex ESS PCR Amplification Kit Mentype Pentaplex ESS PCR Amplification Kit Product description The Mentype Pentaplex ESS PCR Amplification Kit was developed for multiplex amplification of new ministr markers and Amelogenin which are

More information

Procedure for GeneMapper ID-X for Casework

Procedure for GeneMapper ID-X for Casework Procedure for GeneMapper ID-X for Casework 1.0 Purpose-This procedure specifies the steps for performing analysis on DNA samples amplified with AmpFlSTR Identifiler Plus using the GeneMapper ID-X (GMID-X)

More information

SeCore. GSSP Kits. Instructions for Use. 1 Research Use Only

SeCore. GSSP Kits. Instructions for Use. 1 Research Use Only SeCore GSSP Kits Instructions for Use 1 SeCore GSSP Kits Instructions for Use Invitrogen SeCore GSSP (Group Specific Sequencing Products) are designed to resolve ambiguous HLA heterozygous combinations,

More information

DNF-915 dsdna Reagent Kit User Guide (DNF-915-K0500) (DNF-915-K1000)

DNF-915 dsdna Reagent Kit User Guide (DNF-915-K0500) (DNF-915-K1000) New Software Version Refer to red highlighted text for additional updates DNF-915 dsdna Reagent Kit User Guide (DNF-915-K0500) (DNF-915-K1000) For use with the Fragment Analyzer Automated CE System Fragment

More information

Georgia Bureau of Investigation - Division of Forensic Sciences DNA Procedures Manual DNA QUANTIFICATION USING QUANTIFILER HUMAN / Y HUMAN MALE KIT

Georgia Bureau of Investigation - Division of Forensic Sciences DNA Procedures Manual DNA QUANTIFICATION USING QUANTIFILER HUMAN / Y HUMAN MALE KIT Purpose: To obtain quantitative value of human (higher primate) nuclear DNA isolated from samples. Materials: Previously extracted DNA 96 well optical reaction plate Optical adhesive cover Compression

More information

Procedure for GeneMapper ID-X for Casework

Procedure for GeneMapper ID-X for Casework Procedure for GeneMapper ID-X for Casework 1.0 Purpose-This procedure specifies the steps for performing analysis on DNA samples amplified with AmpFlSTR Identifiler Plus using the GeneMapper ID-X (GMID-X)

More information

Forensic News. FAS Corner April The Maximizing Data Quality Series Part 2: DNA Quantitation

Forensic News. FAS Corner April The Maximizing Data Quality Series Part 2: DNA Quantitation The Maximizing Data Quality Series Part 2: DNA Quantitation (Editor's Note: This is part of an ongoing series of articles focusing on maximizing data quality throughout the HID workflow. Part 1 on DNA

More information

Standard Sensitivity RNA Analysis Kit User Guide (DNF ) (DNF )

Standard Sensitivity RNA Analysis Kit User Guide (DNF ) (DNF ) New Diluent Marker Storage Condition and Handling Refer to red highlighted text for additional updates Standard Sensitivity RNA Analysis Kit User Guide (DNF-489-0500) (DNF-489-1000) For use with the Fragment

More information

Mentype Nonaplex I PCR Amplification Kit

Mentype Nonaplex I PCR Amplification Kit Mentype Nonaplex I PCR Amplification Kit Instructions for use The Mentype Nonaplex I PCR Amplification Kit is a multiplex application for the Short Tandem Repeat (STR) loci, which belong to the German

More information

Applied Biosystems TaqMan Array Microfluidic Cards

Applied Biosystems TaqMan Array Microfluidic Cards Performing Rapid Cycling Gene Quantitation on TaqMan Array Microfluidic Cards Applied Biosystems TaqMan Array Microfluidic Cards User Bulletin JULY 2010 SUBJECT: In this user bulletin Performing Rapid

More information

CYP 2C19 Genotyping Reagents

CYP 2C19 Genotyping Reagents CONTENTS Introduction 4 Overview of Mutector TM Assay 5 Materials Provided 6 Materials Required 7 CYP 2C19 Genotyping Reagents Equipment Required 7 DNA Sample Preparation 8 Sequencer Setup 8 Thermal Cycling

More information

CALR Mutation Analysis Reagents

CALR Mutation Analysis Reagents CALR Mutation Analysis Reagents User Manual V1.0 Cat No. GP20 32 reactions 1 CONTENTS Introduction 4 Materials Provided 4 Materials Required 5 Equipment Required 5 DNA Sample Preparation 6 Sequencer Setup

More information

dsdna 905 Reagent Kit User Guide (DNF-905-K0500) (DNF-905-K1000)

dsdna 905 Reagent Kit User Guide (DNF-905-K0500) (DNF-905-K1000) New Software Version Refer to red highlighted text for additional updates or important information dsdna 905 Reagent Kit User Guide (DNF-905-K0500) (DNF-905-K1000) For use with the Fragment Analyzer Automated

More information

BigDye Direct Cycle Sequencing Kit

BigDye Direct Cycle Sequencing Kit BigDye Direct Cycle Sequencing Kit QUICK REFERENCE CARD Note: For safety and biohazard guidelines, refer to the Safety section in the BigDye Direct Cycle Sequencing Kit Protocol (PN 4458040). For every

More information

PowerPlex 16 System INSTRUCTIONS FOR USE OF PRODUCTS DC6530 AND DC6531

PowerPlex 16 System INSTRUCTIONS FOR USE OF PRODUCTS DC6530 AND DC6531 Technical Manual PowerPlex 16 System INSTRUCTIONS FOR USE OF PRODUCTS DC6530 AND DC6531 This Technical Manual has been revised. Please read it carefully. The PowerPlex 16 System shipped with this Technical

More information

Perform an HRM Methylation Study

Perform an HRM Methylation Study Quick Reference Card Perform an HRM Methylation Study This quick reference card provides brief procedures for performing an HRM methylation study using MeltDoctor HRM Master Mix and High Resolution Melting

More information

Supported Protocol: Getting Started with the QuantStudio 3D Digital PCR System

Supported Protocol: Getting Started with the QuantStudio 3D Digital PCR System USER BULLETIN Supported Protocol: Getting Started with the QuantStudio 3D Digital PCR System Publication Number MAN0009821 Revision A.0 Overview... 1 Reuired materials... 2 Workflow... 3 Protocol... 3

More information

Biotype Template Files for GeneMapper ID Software

Biotype Template Files for GeneMapper ID Software Biotype Template Files for GeneMapper ID Software Introduction Biotype Template Files are software sets for the GeneMapper ID Software in order to simplify the evaluation of analysis data. Biotype Template

More information

Investigator Quantiplex Handbook

Investigator Quantiplex Handbook November 2014 Investigator Quantiplex Handbook For quantification of human and male DNA in forensic samples Sample & Assay Technologies QIAGEN Sample and Assay Technologies QIAGEN is the leading provider

More information

Protocol. Path-ID Multiplex One-Step RT-PCR Kit Protocol TaqMan probe-based multiplex one-step real-time RT-PCR detection of RNA targets

Protocol. Path-ID Multiplex One-Step RT-PCR Kit Protocol TaqMan probe-based multiplex one-step real-time RT-PCR detection of RNA targets Protocol TaqMan probe-based multiplex one-step real-time RT-PCR detection of RNA targets For Research Use Only. Not for use in diagnostic procedures. Information in this document is subject to change without

More information

Applied Biosystems 3130 and 3130xl DNA Analyzers

Applied Biosystems 3130 and 3130xl DNA Analyzers www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Applied Biosystems 3130 and 3130xl DNA Analyzers Key Features 16 capillaries

More information

Multiplex RT for TaqMan Array Human MicroRNA Panel

Multiplex RT for TaqMan Array Human MicroRNA Panel f Multiplex RT for TaqMan Low Density Array Multiplex RT for TaqMan Array Human MicroRNA Panel Quick Reference Card For safety and biohazard guidelines, refer to the Safety section in the Multiplex RT

More information

PowerPlex. Y System Validation

PowerPlex. Y System Validation PowerPlex Y System Validation By Patricia M. Fulmer, Dawn Rabbach, Kimberly Huston, Curtis Knox and Cynthia Sprecher, Promega Corporation Abstract We have improved the manufacturing process for the PowerPlex

More information

resdnaseq Quantitative DNA Kits

resdnaseq Quantitative DNA Kits resdnaseq Quantitative DNA Kits QUICK REFERENCE Note: For safety and biohazard guidelines, refer to the Safety section in the resdnaseq Quantitative DNA Kits User Guide (Part no. 4460662). For every chemical,

More information

Mentype Chimera Manual

Mentype Chimera Manual Mentype Chimera Manual The new standard for chimerism analysis In-Vitro-Diagnostics 25 100 400 1000 Version January 2013 45-13210-0025 45-13210-0100 45-13210-0400 45-13210-1000 Batch Code Biotype Diagnostic

More information

Mentype DIPscreen Manual

Mentype DIPscreen Manual Mentype DIPscreen Manual The entry to a quantitative chimerism analysis In-Vitro-Diagnostics 25 100 400 Version July 2015 45-45410-0025 45-45410-0100 45-45410-0400 Batch Code Biotype Diagnostic GmbH Moritzburger

More information

Artifacts Identified Post-Developmental Validation: AmpFLSTR Identifiler Plus PCR Amplification Kit

Artifacts Identified Post-Developmental Validation: AmpFLSTR Identifiler Plus PCR Amplification Kit March 29, 2018 TECHNICAL NOTE Artifacts Identified Post-Developmental Validation: AmpFLSTR Identifiler Plus PCR Amplification Kit The purpose of this document is to assist with data interpretation by providing

More information

Direct PCR Amplification of Blood Samples on Copan NUCLEIC-CARD TM Systems Using the AmpFlSTR Identifiler Direct PCR Amplification Kit

Direct PCR Amplification of Blood Samples on Copan NUCLEIC-CARD TM Systems Using the AmpFlSTR Identifiler Direct PCR Amplification Kit Technical Focus Direct PCR Amplification of Blood Samples on Copan NUCLEIC-CARD TM Systems Using the AmpFlSTR Identifiler Direct PCR Amplification Kit Nicola Oldroyd, Jennifer Schroeder, Dennis Wang, Alessandro

More information

MLPA assays on the SeqStudio Genetic Analyzer

MLPA assays on the SeqStudio Genetic Analyzer APPLICATION NOTE MLPA assays on the SeqStudio Genetic Analyzer SeqStudio Genetic Analyzer In this application note, we demonstrate that: The new Applied Biosystems SeqStudio Genetic Analyzer generates

More information

TaqMan PreAmp Master Mix Kit. Protocol

TaqMan PreAmp Master Mix Kit. Protocol TaqMan PreAmp Master Mix Kit Protocol Copyright 2007, 2010 Applied Biosystems. All rights reserved. For Research Use Only. Not for use in diagnostic procedures. Information in this document is subject

More information

HID EVOlution Extraction System and HID EVOlution Combination System

HID EVOlution Extraction System and HID EVOlution Combination System HID EVOlution Extraction System and HID EVOlution Combination System USER BULLETIN JUNE 2010 SUBJECT: Validation of PrepFiler Wash Buffer B and the Related Modifications to Worktable Layout and Scripts

More information

PowerSeq Quant MS System

PowerSeq Quant MS System TECHNICAL MANUAL PowerSeq Quant MS System Instructions for Use of Product PS5000 3/18 TM511 PowerSeq Quant MS System All technical literature is available at: www.promega.com/protocols/ Visit the web site

More information

Multiplex analysis of 19 STR loci plus Amelogenin. User manual

Multiplex analysis of 19 STR loci plus Amelogenin. User manual Multiplex analysis of 19 STR loci plus Amelogenin Page 1 of 18 Table of Content 1 Product Information.3 1.1 Product Description 3 1.2 Ordering Information and Kit Components 5 1.3 Storage Conditions..5

More information

Axiom TM 2.0 Automated Target Prep Protocol

Axiom TM 2.0 Automated Target Prep Protocol Quick Reference Card Axiom TM 2.0 Automated Target Prep Protocol Stage 1. DNA Amplification Introduction Running the Axiom 2.0 Assay requires the following sets of steps: 1. Genomic DNA Prep, described

More information

Thermo Scientific Bovine Genotypes Panel 2.2

Thermo Scientific Bovine Genotypes Panel 2.2 Thermo Scientific Bovine Genotypes Panel 2.2 F- 847S 100 reactions F- 847L 500 reactions Technical Manual Product Description Parentage testing and individual identification using short tandem repeat (STR)

More information

Procedure for DNA Quantitation with Quantifiler Duo

Procedure for DNA Quantitation with Quantifiler Duo Procedure for DNA Quantitation with Quantifiler Duo 1.0 Purpose This procedure specifies the steps for performing quantification of DNA extracts with the Quantifiler Duo kit on the ABI 7500 instruments.

More information

Canine ISAG STR Parentage Kit (2014)

Canine ISAG STR Parentage Kit (2014) Canine ISAG STR Parentage Kit (2014) USER GUIDE Parentage testing and individual identification using multiplex PCR of short tandem repeat loci Catalog Numbers A36138, A36139 Publication Number MAN0017047

More information

Plexor HY System for the Applied Biosystems 7500 and 7500 FAST Real-Time PCR Systems

Plexor HY System for the Applied Biosystems 7500 and 7500 FAST Real-Time PCR Systems TECHNICAL MANUAL Plexor HY System for the Applied Biosystems 7500 and 7500 FAST Real-Time PCR Systems Instructions for Use of Products DC1000, DC1001 and DC1500 Technical Manuals with instructions for

More information

Thermo Scientific Equine Genotypes Panel 1.1

Thermo Scientific Equine Genotypes Panel 1.1 Thermo Scientific Equine Genotypes Panel 1.1 F- 850S 100 reactions F- 850L 500 reactions Technical Manual Product Description Parentage testing and individual identification using short tandem repeat (STR)

More information

High Sensitivity Genomic DNA Analysis Kit User Guide (DNF ) (DNF )

High Sensitivity Genomic DNA Analysis Kit User Guide (DNF ) (DNF ) Updated Information and Software Refer to red highlighted text for updates High Sensitivity Genomic DNA Analysis Kit User Guide (DNF-488-0500) (DNF-488-1000) For use with the Fragment Analyzer Automated

More information

PulseNet MLVA protocols General overview

PulseNet MLVA protocols General overview PulseNet MLVA protocols General overview Eija Trees, Ph.D., D.V.M. PulseNet Methods Development and Reference Unit Enteric Diseases Laboratory Branch CDC, Atlanta, GA 1 Multiple-Locus VNTR Analysis (MLVA)

More information

Naming Convention and Laboratory Methods: PROVEDIt Database

Naming Convention and Laboratory Methods: PROVEDIt Database Naming Convention and Laborary Methods: PROVEDIt Database Sec 1. Naming Convention and Laborary Methods Single source sample names will follow the format below and are best explained through example: RD14-0003-21d1x-0.5IP-Q0.8_002.20sec

More information

Applied Biosystems 7500 Fast Real-Time PCR System

Applied Biosystems 7500 Fast Real-Time PCR System User Bulletin Applied Biosystems 7500 Fast Real-Time PCR System July 2006 SUBJECT: Using Expert Mode on the Applied Biosystems 7500 Fast Real-Time PCR System with the Applied Biosystems 7500 Fast System

More information

AmpF STR NGM PCR Amplification Kit - Overview

AmpF STR NGM PCR Amplification Kit - Overview AmpF STR NGM PCR Amplification Kit - Overview The most advanced STR kits optimized for analysis of forensic casework and database samples in Europe Data Quality Worth Sharing! 9/7/2011 Life Technologies

More information