1 Human ICAM-2 / CD102 ELISA Pair Set Catalog Number : SEK10332 To achieve the best assay results, this manual must be read carefully before using this product and the assay is run as summarized in the General ELISA protocol. 1

2 BACKGROUND Intercellular adhesion molecule 2 (ICAM2, CD102), belongs to the ICAM family consisting of three members identified as ligands for integrin receptors. It is a type I transmembrane glycoprotein with two Ig-like C2-type domains, and binds to the leukocyte integrins LFA-1 (CD11a/CD18) and Mac-1 (CD11b/CD18). ICAM2 is mainly expressed on vascular endothelial and hematopoietic cells. Interactions of ICAM2 and the integrin receptors mediate cell adhesion in a wide range of lymphocyte, monocyte, natural killer cell, and granulocytewith other cells, and play important roles in many adhesion-dependent immune and inflammation responses, such as T cell aggregation, NK-cell cytotoxicity and migration, lymphocyte recirculation, etc. This protein can be encoded by several transcript variants. 2

3 PRINCIPLE OF THE TEST The Sino Biological ELISA Pair Set is a solid phase sandwich ELISA (Enzyme-Linked Immunosorbent Assay). It utilizes a monoclonal antibody specific for ICAM-2 / CD102 coated on a 96-well plate. Standards and samples are added to the wells, and any ICAM-2 / CD102 present binds to the immobilized antibody. The wells are washed and a horseradish peroxidase conjugated rabbit anti-icam-2 / CD102 monoclonal antibody is then added, producing an antibody-antigenantibody sandwich. The wells are again washed and TMB substrate solution is loaded, which produces color in proportion to the amount of ICAM-2 / CD102 present in the sample. To end the enzyme reaction, the stop solution is added and absorbances of the microwell are read at 450 nm. INTENDED USE The Human ICAM-2 / CD102 ELISA Pair Set is for the quantitative determination of Human ICAM-2 / CD102. This ELISA Pair Set contains the basic components required for the development of sandwich ELISAs. ASSAY PROCEDURE SUMMARY Capture Antibody Analyte HRP-conjugated Detection Antibody This Pair Set has been configured for research use only and is not to be used in diagnostic procedures. 3

4 MATERIALS PROVIDED Bring all reagents to room temperature before use. Capture Antibody 0.5 mg/ml of rabbit anti-icam2 monoclonal antibody. Dilute to a working concentration of 2.0 μg/ml in CBS before coating. (Catalog: # R022) Detection Antibody 0.5 mg/ml rabbit anti-icam2 monoclonal antibody conjugated to horseradish-peroxidase (HRP). Dilute to working concentration of 0.5 μg/ml in detection antibody dilution buffer before use. (Catalog: # R015) Standard Each vial contains 35 ng of recombinant ICAM2. Reconstitute standard powder with 1mL detection antibody dilution buffer. After reconstitution, store at -20 to -70 in a manual defrost freezer. A seven-point standard curve using 2-fold serial dilutions in sample dilution buffer, and a high standard of 3000 pg/ml is recommended. SOLUTIONS REQUIRED CBS M Na 2 CO 3, 0.05M NaHCO 3, ph 9.6, 0.2 μm filtered TBS - 20 mm Tris, 150 mm NaCl, ph 7.4 Wash Buffer % Tween20 in TBS, ph Blocking Buffer - 2% BSA in Wash Buffer Sample dilution buffer - 0.1% BSA in wash buffer, ph , 0.2 μm filtered Detection antibody dilution buffer - 0.5% BSA in wash buffer, ph , 0.2 μm filtered Substrate Solution : To achieve best assay results, fresh substrate solution is recommended Substrate stock solution - 10mg / ml TMB ( Tetramethylbenzidine ) in DMSO Substrate dilution buffer M Na 2 HPO 4 and 0.025M citric acid ; adjust ph to 5.5 Substrate working solution - For each plate dilute 250 μl substrate stock solution in 25ml substrate dilution buffer and then add 80 μl 0.75% H 2 O 2, mix it well Stop Solution - 2 N H 2 SO 4 PRECAUTION The Stop Solution suggested for use with this Pair Set is an acid solution. Wear eye, hand, face, and clothing protection when using this material. STORAGE Detection Antibody should be protected from prolonged exposure to light. Aliquot the reagents and store at -20 to -70 in a manual defrost freezer. 4

5 GENERAL ELISA PROTOCOL Plate Preparation 1. Dilute the capture antibody to the working concentration in CBS. Immediately coat a 96-well microplate with 100μL per well of the diluted capture antibody. Seal the plate and incubate overnight at Aspirate each well and wash with at least 300μl wash buffer, repeating the process two times for a total of three washes. Complete removal of liquid at each step is essential for good performance. After the last wash, remove any remaining wash buffer by inverting the plate and blotting it against clean paper towels. 3. Block plates by adding 300 μl of blocking buffer to each well. Incubate at room temperature for a minimum of 1 hour. 4. Repeat the aspiration/wash as in step 2. The plates are now ready for sample addition. Assay Procedure 1. Add 100 μl of sample or standards in sample dilution buffer per well. Seal the plate and incubate 2 hours at room temperature. 2. Repeat the aspiration/wash as in step 2 of plate preparation. 3. Add 100 μl of the detection antibody, diluted in antibody dilution buffer, to each well. Seal the plate and incubate 1 hour at room temperature. 4. Repeat the aspiration/wash as in step 2 of plate preparation. 5. Add 200 μl of substrate solution to each well. Incubate for 20 minutes at room temperature ( if substrate solution is not as requested, the incubation time should be optimized ). Avoid placing the plate in direct light. 6. Add 50 μl of stop solution to each well. Gently tap the plate to ensure thorough mixing. 7. Determine the optical density of each well immediately, using a microplate reader set to 450 nm. CALCULATION OF RESULTS Calculate the mean absorbance for each set of duplicate standards, controls and samples. Subtract the mean zero standard absorbance from each. Construct a standard curve by plotting the mean absorbance for each standard on the y-axis against the concentration on the x-axis and draw a best fit curve through the points on the graph. To determine the concentration of the unknowns, find the unknowns mean absorbance value on the y-axis and draw a horizontal line to the standard curve. At the point of intersection, draw a vertical line to the x-axis and read the concentration. If samples have been diluted, the concentration read from the standard curve must be multiplied by the dilution factor. Alternatively, computer-based curve-fitting statistical software may also be employed to calculate the concentration of the sample. 5

6 optical density (OD450nm) TYPICAL DATA This standard curve is only for demonstration purposes. A standard curve should be generated for each assay ICAM2 concentration(pg/ml) Concentration (pg/ml) Zero standard subtracted OD PERFORMANCE CHARACTERISTIC SENSITIVITY The minimum detectable dose of Human ICAM-2 / CD102 was determined to be approximately pg/ml. This is defined as at least three times standard deviations above the mean optical density of 10 replicates of the zero standard. 6

7 TROUBLE SHOOTING Problems Possible Sources Solutions Incorrect or no Detection Antibody was added Add appropriate Detection Antibody and continue No signal Substrate solution was not added Add substrate solution and continue Incorrect storage condition Check if the kit is stored at recommended condition and used before expiration date Standard was incompletely reconstituted or was inappropriately stored Aliquot reconstituted standard and store at -70 Poor Standard Curve Poor detection value High Background Imprecise / inaccurate pipetting Incubations done at inappropriate temperature, timing or agitation Background wells were contaminated The concentration of antigen in samples was too low Samples were ineffective Insufficient washes TMB Substrate Solution was contaminated Check / calibrate pipettes Follow the general ELISA protocol Avoid cross contamination by using the sealer appropriately Enriching samples to increase the concentration of antigen Check if the samples are stored at cold environment. Detect samples in timely manner Use multichannel pipettes without touching the reagents on the plate Increase cycles of washes and soaking time between washes TMB Substrate Solution should be clear and colorless prior to addition to wells Materials were contaminated. Use clean plates, tubes and pipettes tips Non-specificity Samples were contaminated The concentration of samples was too high Avoid cross contamination of samples Try higher dilution rate of samples 7

8 Human ICAM-2 / CD102 ELISA Pair Set