MEDIUM WATER PURIFICATION SYSTEM

Size: px
Start display at page:

Download "MEDIUM WATER PURIFICATION SYSTEM"

Transcription

1 PRODUCT CATALOG MEDIUM WATER PURIFICATION SYSTEM

2 MEDIUM WATER PURIFICATION SYSTEM This medium series is an excellent choice of high-performance water purifying system. The self-flushing of the reverse osmosis membrane, extend the life of RO membrane. These systems are Level II password protected for all the parameters setting, and forbid any unauthorized settings change also It has multiple alarm functions. 100 SERIES MEDIUM WATER PURIFICATION SYSTEM Automatic microcomputer controlling system. Self-flushing of the reverse osmosis membrane, extend the life of RO membrane. Multiple alarm functions: such as no water, full water, disqualification of feed water, RO water, deionized water or ultrapure water, cartridges life-span ends. The cartridge's life-span can be set, the time used and left can be displayed, replacing auto-reminding, avoiding the decline of water quality. Level II password, protect all the parameters setting, and prohibit any unauthorized settings change. RS 232/USB communication port(optional), at least store 1 years water quality data. 3 way on-line sensor, detect the quality of feed water, RO water, or ultrapure water respectively. Pretreatment cartridges, RO module, ultrapure cartridges, all designed to modularization independently. Pipeline and fast-plug adaptor with NSF authorization, assure high quality ultrapure water. KDF pretreating cartridge, replace the ordinary active carbon, prolong the lifespan to 12 months, reduce the running cost. DOW s RO membrane, ensure stable operation and high desalinization rate. Double wavelength (185&254nm) ultraviolet lamp module, restrain bacteria s increase and reduce TOC. MWCO 5000D ultrafiltration module, effectively eliminate endotoxin precise cell cultivating and IVF. ( )μm double layer PES terminal disinfection filter, assure the quality absolutely axenic. SPECIFICATIONS Model BDPS-101 BDPS-102 BDPS-103 BDPS-104 Flow Procedure** RO water Quality Ion rejection rate Organic rejection rate Particles and bacteria rejection rate Deionized water quality Resistivity Heavy metal ion Tap water: TDS<200 ppm (Extra pretreatment filter is recommended, if TDS>200ppm) PF+KDF+AC+RO+AC+ DI PF+KDF+AC+RO+AC+DI+U V+TF Kgf/cm² PF+KDF+AC+RO+AC+ DI 96%-99% (New RO membrane) >99%, when MW>200 Dalton >99% > MΩ.cm <0.1ppb PF+KDF+AC+RO+AC+DI+U V+TF Bacteria - <0.1cfu/ml - <0.1cfu/ml Particle(>0.2µm) - <1/ml - <1/ml Output(25 C)**** 15 L/hr 30 L/hr 2

3 Flow rate 2.0 L/min (with pressure tank) RO water and Deionized water Weight (Kg) 22 Standard configuration Power Consumption 200 SERIES MEDIUM WATER PURIFICATION SYSTEM Main body (Including 1 set of cartridges)+15 liters tank+ accessory bag *The feed water quality will influence the pure water s quality and cartridges life-span. **PF:polypropylene spun fiber, KDF:kinetic degradation fluxion, AC:active carbon, RO:reverse osmosis, DI:ion exchange, UV:ultraviolet, TF:terminal microfiltration. ***All the specifications are tested under the situation:feed water s TDS=200ppm, 25 C, 50psi and 15% recovery rate. Automatic microcomputer controlling system. 3 way on-line sensor, detect the quality of feed water, RO water, or ultrapure water respectively. System circulation function, circulate water when the system stops working, to keep water quality. Multiple alarm functions: such as no water, full water, disqualification of feed water, RO water, deionized water or ultrapure water, cartridges life-span ends. The cartridge's life-span can be set, the time used and left can be displayed, replacing auto-reminding, avoiding the decline of water quality. Level II password, protect all the parameters setting, and prohibit any unauthorized settings change. RS 232/USB communication port(optional), at least store 1 years water quality data. Pipeline and fast-plug adaptor with NSF authorization, assure high quality ultrapure water. KDF pretreating cartridge, replace the ordinary active carbon, prolong the life-span to 12 months, reduce the running cost. DOW s RO membrane, ensure stable operation and high desalinization rate. Double wavelength (185&254nm) ultraviolet lamp module, restrain bacteria s increase and reduce TOC. MWCO 5000D ultrafiltration module, effectively eliminate endotoxin precise cell cultivating and IVF. ( )μm double layer PES terminal disinfection filter, assure the quality absolutely axenic. MEDIUM WATER PURIFICATION SYSTEM SPECIFICATIONS Model BDPS-201 BDPS-202 BDPS-203 BDPS-204 Flow Procedure** RO water Quality Ion rejection rate Organic rejection rate Particles and bacteria rejection rate Tap water: TDS<200 ppm (Extra pretreatment filter is recommended, if TDS>200ppm) PF+KDF+AC+RO+DI+ TF PF+KDF+AC+RO+DI+UF+T F Kgf/cm² PF+KDF+AC+RO+UV+ DI+TF 96%-99% (New RO membrane) >99%, when MW>200 Dalton >99% PF+KDF+AC+RO+UV+DI+UF +TF 3

4 Ultrapure Water Quality Resistivity(25 C) Heavy Metal Ion 18.2MΩ.cm <0.1ppb TOC*** <10ppb <3ppb Bacteria Particle (>0.2µm) <0.1cfu/ml <1/ml Endotoxin - <0.001Eu/ml - <0.001Eu/ml Rnases - <0.01ng/ml - <0.01ng/ml Dnases - <4pg/µl - <4pg/µl Output(25 C)**** Flow rate 15 L/hr 2.0 L/min (with pressure tank) RO water and Ultrapure water Weight (Kg) 25 Standard configuration Power Consumption Main body (Including 1 set of cartridges)+15 liters tank+ accessory bag *The feed water quality will influence the pure water s quality and cartridges life-span. **PF:polypropylene spun fiber, KDF:kinetic degradation fluxion, AC:active carbon, RO:reverse osmosis, DI:ion exchange, UV:ultraviolet, UF:ultrafiltration, TF:terminal microfiltration. ***Value of number will be influenced by temperature and feed water quality. ****All the specifications are tested under the situation:feed water s TDS=200ppm, 25 C, 50psi and 15% recovery rate. Model BDPS-205 BDPS-206 BDPS-207 BDPS-208 Flow Procedure** RO water Quality Ion rejection rate Organic rejection rate Particles and bacteria rejection rate Ultrapure Water Quality Resistivity(25 C) Heavy Metal Ion Tap water: TDS<200 ppm (Extra pretreatment filter is recommended, if TDS>200ppm) PF+KDF+AC+RO+DI+ TF PF+KDF+AC+RO+DI+UF+T F Kgf/cm² PF+KDF+AC+RO+UV+ DI+TF 96%-99% (New RO membrane) >99%, when MW>200 Dalton >99% 18.2MΩ.cm <0.1ppb TOC*** <10ppb <3ppb Bacteria Particle (>0.2µm) <0.1cfu/ml <1/ml PF+KDF+AC+RO+UV+DI+UF +TF Endotoxin - <0.001Eu/ml - <0.001Eu/ml Rnases - <0.01ng/ml - <0.01ng/ml 4

5 Dnases - <4pg/µl - <4pg/µl Output(25 C)**** Flow rate 30 L/hr 2.0 L/min (with pressure tank) RO water and Ultrapure water Weight (Kg) 25 Standard configuration Power Consumption 300 SERIES MEDIUM WATER PURIFICATION SYSTEM Main body (Including 1 set of cartridges)+15 liters tank+ accessory bag *The feed water quality will influence the pure water s quality and cartridges life-span. **PF:polypropylene spun fiber, KDF:kinetic degradation fluxion, AC:active carbon, RO:reverse osmosis, DI:ion exchange, UV:ultraviolet, UF:ultrafiltration, TF:terminal microfiltration. ***Value of number will be influenced by temperature and feed water quality. ****All the specifications are tested under the situation:feed water s TDS=200ppm, 25 C, 50psi and 15% recovery rate. Automatic microcomputer controlling system. 3 way on-line sensor, detect the quality of feed water, RO water, or ultrapure water respectively. System circulation function, circulate water when the system stops working, to keep water quality. Multiple alarm functions: such as no water, full water, disqualification of feed water, RO water, deionized water or ultrapure water, cartridges life-span ends. The cartridge's life-span can be set, the time used and left can be displayed, replacing auto-reminding, avoiding the decline of water quality. Level II password, protect all the parameters setting, and prohibit any unauthorized settings change. RS 232/USB communication port(optional), at least store 1 years water quality data. Pretreatment cartridges, RO module, ultrapure cartridges, all designed to modularization independently. Easy to maintenance and replacement. KDF pretreating cartridge, replace the ordinary active carbon, prolong the life-span to 12 months, reduce the running cost. DOW s RO membrane, ensure stable operation and high desalinization rate. Double wavelength (185&254nm) ultraviolet lamp module, restrain bacteria s increase and reduce TOC. MWCO 5000D ultrafiltration module, effectively eliminate endotoxin precise cell cultivating and IVF. ( )μm double layer PES terminal disinfection filter, assure the quality absolutely axenic. MEDIUM WATER PURIFICATION SYSTEM 5

6 SPECIFICATIONS Model BDPS-301 BDPS-302 BDPS-303 Tap water: TDS<200 ppm (Extra pretreatment filter is recommended, if TDS>200ppm) Kgf/cm² Flow Procedure** PF+KDF+AC+RO+RO+AC PF+KDF+AC+RO+RO+AC+DI+TF RO Water Quality 1st stage RO water's TDS 2nd stage RO water's conductivity Output(25 C)**** Flow rate TDS (ppm,mg/l) < TDS of tap water x 5% PF+KDF+AC+RO+RO+(UV)+AC+DI+ (UF)+TF 1-5µs/cm, Organic rejection rate>99% (when MW>200 Dalton), Particles and bacteria rejection rate>99% 1st and 2nd stage RO water L/hr 2.0 L/min (with pressure tank) 2nd stage RO and Deionized water Weight (Kg) 25 Standard configuration Power Consumption Deionized water quality 2nd stage RO and Ultrapure water Main body (Including 1 set of cartridges)+ assorted tank (built-in tank+15 liters external tank)+ accessory bag *The feed water quality will influence the pure water s quality and cartridges life-span. ** PF:polypropylene spun fiber, KDF:kinetic degradation fluxion, AC:active carbon, RO:reverse osmosis, DI:ion exchange, UV:ultraviolet, UF:ultrafiltration, TF:terminal microfiltration. ***Value of number will be influenced by temperature and feed water quality. ****All the specifications are tested under the situation:feed water s TDS=200ppm, 25 C, 50psi and 15% recovery rate. Resistivity - > MΩ.cm - Heavy metal ion - <0.1ppb - Bacteria - <0.1cfu/ml - Particle(>0.2µm) - <1/ml - Ultrapure Water Quality Resistivity(25 C) MΩ.cm Heavy Metal Ion - - <0.1ppb TOC*** - - <10ppb Bacteria - - <0.1cfu/ml Particle (>0.2µm) - - <1/ml Endotoxin Rnases Dnases

7 Model BDPS-304 BDPS-305 BDPS-306 Flow Procedure** RO Water Quality 1st stage RO water's TDS 2nd stage RO water's conductivity Output(25 C)**** Flow rate Tap water: TDS<200 ppm (Extra pretreatment filter is recommended, if TDS>200ppm) Kgf/cm² PF+KDF+AC+RO+RO+(UV)+AC+DI+(UF)+TF TDS (ppm,mg/l) < TDS of tap water x 5% 1-5µs/cm, Organic rejection rate>99% (when MW>200 Dalton), Particles and bacteria rejection rate>99% L/hr 2.0 L/min (with pressure tank) 2nd stage RO and Ultrapure water Weight (Kg) 25 Standard configuration Power Consumption Deionized water quality Main body (Including 1 set of cartridges)+ assorted tank (built-in tank+15 liters external tank)+ accessory bag *The feed water quality will influence the pure water s quality and cartridges life-span. ** PF:polypropylene spun fiber, KDF:kinetic degradation fluxion, AC:active carbon, RO:reverse osmosis, DI:ion exchange, UV:ultraviolet, UF:ultrafiltration, TF:terminal microfiltration. ***Value of number will be influenced by temperature and feed water quality. ****All the specifications are tested under the situation:feed water s TDS=200ppm, 25 C, 50psi and 15% recovery rate. Resistivity Heavy metal ion Bacteria Particle(>0.2µm) Ultrapure Water Quality Resistivity(25 C) Heavy Metal Ion 18.2MΩ.cm <0.1ppb TOC*** <10ppb <3ppb Bacteria Particle (>0.2µm) <0.1cfu/ml <1/ml Endotoxin <0.001Eu/ml - <0.001Eu/ml Rnases <0.01ng/ml - <0.01ng/ml Dnases <4pg/µl - <4pg/µl MEDIUM WATER PURIFICATION SYSTEM 7

8 400 SERIES MEDIUM WATER PURIFICATION SYSTEM Automatic microcomputer controlling system. 3 way on-line sensor, detect the quality of feed water, RO water, or ultrapure water respectively. System circulation function, circulate water when the system stops working, to keep water quality. Multiple alarm functions: such as no water, full water, disqualification of feed water, RO water, deionized water or ultrapure water, cartridges life-span ends. The cartridge's life-span can be set, the time used and left can be displayed, replacing auto-reminding, avoiding the decline of water quality. Level II password, protect all the parameters setting, and prohibit any unauthorized settings change. RS 232/USB communication port(optional), at least store 1 years water quality data. Pipeline and fast-plug adaptor with NSF authorization, assure high quality ultrapure water. KDF pretreating cartridge, replace the ordinary active carbon, prolong the life-span to 12 months, reduce the running cost. DOW s RO membrane, ensure stable operation and high desalinization rate. Double wavelength (185&254nm) ultraviolet lamp module, restrain bacteria s increase and reduce TOC. MWCO 5000D ultrafiltration module, effectively eliminate endotoxin precise cell cultivating and IVF. ( )μm double layer PES terminal disinfection filter, assure the quality absolutely axenic. SPECIFICATIONS Model BDPS-401 BDPS-402 BDPS-403 BDPS-404 RO water, Distilled water, Deionized water 1 atm* Flow Procedure** AC+DI+TF AC+DI+UF+TF UV+AC+DI+TF UV+AC+DI+UF+TF Ultrapure Water Quality Resistivity(25 C) Heavy Metal Ion 18.2MΩ.cm <0.1ppb TOC*** <10ppb <3ppb Bacteria Particle (>0.2µm) <0.1cfu/ml <1/ml Endotoxin - <0.001Eu/ml - <0.001Eu/ml Rnases - <0.01ng/ml - <0.01ng/ml Dnases - <4pg/µl - <4pg/µl Deionized water quality Resistivity Output(25 C)**** >5MΩ.cm Utmost up to 2.0 L/min (less output with UF cartridge) Deionized water and Ultrapure water Weight (Kg) 20 Standard configuration Main body (Including 1 set of cartridges)+ accessory bag 8

9 Power Consumption *The feed water quality will influence the pure water s quality and cartridges life-span. ** AC:active carbon,di:ion exchange,uv:ultraviolet,uf:ultrafiltration,tf:terminal microfiltration. ***Value of number will be influenced by feed water quality. ****The output will decrease with terminal filter or UF cartridge. MEDIUM WATER PURIFICATION SYSTEM 9

10 Biolab Scientific Limited 3660 Midland Avenue, Suite 300, Toronto, Ontario M1V 0B8, Canada Phone: Website: