Test record CFU measurement. Test ID Site/Clinic Spijkenisse

Size: px
Start display at page:

Download "Test record CFU measurement. Test ID Site/Clinic Spijkenisse"

Transcription

1 Test record CFU measurement Test ID Site/Clinic Spijkenisse Objective of the test: To verify that there are < 10 CFU/m3 in the clean air flow area. Type of test: CFU test, Microbial precipitation test Description of test method: Sampling Air sampling CFU-measurement and sedimentation HYCON Agar Strips TC en YM. for reference. Equipment used: RCS High Flow Touch Microbial Air Sampler - Validated according to ISO Air measurements are performed with a device known as an air sampler. This device samples a previously determined quantity of air of between 10 and 1000 liters. By inserting various sterile sampling strips into the device the air can be accurately sampled. The strips are specially prepared in our own laboratories to identify 8 species of bacteria and 18 species of fungi and yeasts. In addition to species assessment, these air samplers and subsequent laboratory testing allow for the determination of amounts of bacteria, fungi and yeasts. On the day of the investigation the air is sampled at a number of locations. At each specified measurement point two samples of air are extracted. The first air sample is drawn over a TC agar strip and the second air sample over an YM agar strip. The first strip is used to measure the number of aerobic (oxygen-demanding) and facultativeanaerobic (oxygen-preferring) bacteria. The second strip is used to measure the number of fungi and yeasts. All plates shall be marked with location of the plate, as A1-C9. The measuring points are in a straight angle from the middle part of the filter at different distances. Altogether 3 x 7 tests shall be done in the following situations: A: At rest - SurgiCube is off, drape is on. B: At rest - SurgiCube is on but no activity, drape is on. C: Operational - SurgiCube is on with simulated surgery, drape is on. In order to get a reference of ambient bacteria conditions, 2 agar strips shall be placed at single locations outside the clean air flow area. The strips shall be exposed during the same time as the ongoing CFU measurements. In addition, one agar strip shall be unexposed and left at the microbiological laboratory for batch control. 1/5 Clinical info

2 Analysis Both in our laboratory and during the measurement itself the Lab applies relevant norms and standards, as stipulated by the certification council. For the analysis of bacteria an incubation period of 48 hours is required. The incubation of fungi and yeasts takes up to 7 days for initial analysis. If a determination of the samples is required two additional days are needed. LOD (limit of detection) The limit of detection of this sampling and analysis method is 1 CFU/m3 at a required sampling volume of a 1000 Liters. Accreditation (ISO 17025) Bioconsult B.V. is accredited as a Test Laboratory by the Netherlands Accreditation Council (Raad voor Accreditatie) under number L 219 for the qualitative and quantitative investigation of air. Air samples are obtained by using fully certified air samplers able to assess the presence of species, genera or families of micro-organisms as listed in the appendix. This accreditation means that all the criteria of NEN-EN-ISO/IEC 17025:2000 are met. The scope of activities can be found the website of the Accreditation Council ( ) Setup conditions: Width of filter: 220cm Ventilation: None Temperature: In the winter 20-24ºC and in the summer 23-26ºC are recommended. Relative humidity: Target is between 30% and 70% Pass / Fail criteria: Pass: 10 CFU/m3 in the clean air flow area. Fail: > 10 CFU/m3 in the clean air flow area. Study endpoint: Significance, after 7 samples + 2 ambient reference plates at realistic /simulated surgical procedures. 2/5 Clinical info

3 Test date 07 august 2015 Test carried out by R.van Elk M. Bot (PhD) Device SurgiCube Number of door openings 3 per sample at Operational Clinic In house People in OR 3 Number of agar strips Type 9 x Incubation time 2 days 7 days HYCON Agar Strips TC en YM Incubation temp 36ºC for bacteria 25ºC for fungi and yeasts Size of agar plates -- Laboratory Bioconsult Measurement RCS High Flow Duration method Touch Microbial Air Pass/Fail Setup: Sampler Fail > Mean over 10 in clean area 27 samples 3 Batch control samples 3/5 Clinical info

4 Results summary: A1 Centre 55 cm 69,00 A2 Right 55 cm 80,00 A3 Right 75 cm 94,00 A4 Left 55 cm 75,00 A5 Left 75 cm 64,00 A6 Right 85 cm 70,00 A7 Left 85 cm 63,00 A8 Reference: Post periphery centre m 15,00 Ambient area A9 Reference: Pre periphery centre 1.2 m 24,00 Ambient area Mean A1-A5 76,40 Batch control A 0,00 B1 Centre 55 cm 0,00 B2 Right 55 cm 0,00 B3 Right 75 cm 8,00 B4 Left 55 cm 1,00 B5 Left 75 cm 4,00 B6 Right 85 cm 3,00 B7 Left 85 cm 1,00 B8 Reference: Post periphery centre m 15,00 Ambient area B9 Reference: Pre periphery centre 1.2 m 24,00 Ambient area Mean B1-B5 2,60 Batch control B 0,00 C1 Centre 55 cm 1,00 C2 Right 55 cm 1,00 C3 Right 75 cm 0,00 C4 Left 55 cm 1,00 C5 Left 75 cm 0,00 C6 Right 85 cm 1,00 C7 Left 85 cm 2,00 C8 Reference: Post periphery centre m 15,00 Ambient area C9 Reference: Pre periphery centre 1.2 m 24,00 Ambient area C1-C5 0,60 Batch control C 0,00 4/5 Clinical info

5 Conclusion: Mean of the SurgiCube: At Rest - shows that there are 2,60 CFU s in the clean air flow area. Mean of the SurgiCube - Operational - shows that there are 0,60 CFU s in the clean air flow area. The SurgiCube filters the air efficiently, passing the criteria for this test, having 10 CFU/m3 in the clean air flow area, in both situations (At Rest and Operational). The clean air flow area is 220cm wide and up to 75cm in depth. The SurgiCube is located in a standard room, that has no other means of ventilation or air conditioning. Measurements at 85cm depth, so 10cm outside the clean air flow area show that the pass criteria are still passed: At Rest: 2 CFU s / Operational: 1,50 CFU s The SurgiCube provides clean air at the clean air flow area and is effectively filtering out airborne particulate matter that can carry microbiological contaminants. We trust that this report has been of service to you, Saferspaces BV Test leader for SurgiCube International Martijn van Amelsfort General Manager Ramón Hilberink 5/5 Clinical info

UNDER SECRETARY FOR HEALTH S INFORMATION LETTER MICROBIOLOGICAL ASSESSMENT OF PHARMACEUTICAL CLEANROOMS

UNDER SECRETARY FOR HEALTH S INFORMATION LETTER MICROBIOLOGICAL ASSESSMENT OF PHARMACEUTICAL CLEANROOMS DEPARTMENT OF VETERANS AFFAIRS Veterans Health Administration Washington DC 20420 IL 10-2006-008 In Reply Refer To: 10NB UNDER SECRETARY FOR HEALTH S INFORMATION LETTER MICROBIOLOGICAL ASSESSMENT OF PHARMACEUTICAL

More information

Environmental Monitoring of Aseptic Processing Areas - 2

Environmental Monitoring of Aseptic Processing Areas - 2 Environmental Monitoring of Aseptic Processing Areas - 2 A war against an invisible enemy DCVMN - Víctor Maqueda - May 30, 31, June 1 2016 1 Active Air Monitoring (Viables) Air Sieve Samplers Draw air

More information

Raad voor Accreditatie (Dutch Accreditation Council RvA) Explanatory document on microbiology

Raad voor Accreditatie (Dutch Accreditation Council RvA) Explanatory document on microbiology Raad voor Accreditatie (Dutch Accreditation Council RvA) Explanatory document on microbiology Document code: A RvA-Explanatory note describes the policy and/or the procedures of the RvA concerning a specific

More information

ISO INTERNATIONAL STANDARD. Compressed air Part 7: Test method for viable microbiological contaminant content

ISO INTERNATIONAL STANDARD. Compressed air Part 7: Test method for viable microbiological contaminant content INTERNATIONAL STANDARD ISO 8573-7 First edition 2003-05-01 Compressed air Part 7: Test method for viable microbiological contaminant content Air comprimé Partie 7: Méthode d'essai pour la détermination

More information

Public Health England (PHE) Certified Reference Materials for Microbiology. Certificate of Analysis

Public Health England (PHE) Certified Reference Materials for Microbiology. Certificate of Analysis Public Health England (PHE) Certified Reference Materials for Microbiology Certificate of Analysis This certificate should not be parted from the PHE certified reference material with batch number: 1070-14

More information

Public Health England (PHE) Certified Reference Materials for Microbiology. Certificate of Analysis Clostridium perfringens CRM13170L

Public Health England (PHE) Certified Reference Materials for Microbiology. Certificate of Analysis Clostridium perfringens CRM13170L Public Health England (PHE) Certified Reference Materials for Microbiology Certificate of Analysis Clostridium perfringens CRM13170L This certificate should not be parted from the PHE certified reference

More information

Why Do I Test, What Do I Test & When Do I Test It? Ross Caputo, PhD Chief Technical Officer Eagle Analytical Services

Why Do I Test, What Do I Test & When Do I Test It? Ross Caputo, PhD Chief Technical Officer Eagle Analytical Services Why Do I Test, What Do I Test & When Do I Test It? Ross Caputo, PhD Chief Technical Officer Eagle Analytical Services Disclosures I, Ross Caputo, declare no conflicts of interest, real or apparent, and

More information

MICROBIAL MEASUREMENTS AT VENTILATION SYSTEMS SUPPLY VENTS

MICROBIAL MEASUREMENTS AT VENTILATION SYSTEMS SUPPLY VENTS MICROBIAL MEASUREMENTS AT VENTILATION SYSTEMS SUPPLY VENTS M.Hinker * Allgemeine Unfallversicherungsanstalt, Wien, Österreich (General Workers Compensation Board, Vienna, Austria) ABSTRACT The number of

More information

FDA s Guidance for Industry

FDA s Guidance for Industry Sterile Drug Products Produced by Aseptic Processing - CGMPs Midwest FDC Conference November 4 2005 Susan Bruederle, Investigator FDA / ORA / Central Region Chicago District / Hinsdale IL FDA s Guidance

More information

Tryckluft Del 7: Metod att bestämma halten levande mikrobiologiska föroreningar

Tryckluft Del 7: Metod att bestämma halten levande mikrobiologiska föroreningar SVENSK STANDARD SS-ISO 8573-7 Fastställd 2003-06-06 Utgåva 1 Tryckluft Del 7: Metod att bestämma halten levande mikrobiologiska föroreningar Compressed air Part 7: Test method for viable microbiological

More information

CERTIFICATE OF ACCREDITATION

CERTIFICATE OF ACCREDITATION CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Dynatec Scientific Laboratories, Inc. 11940

More information

CERTIFICATE OF ACCREDITATION

CERTIFICATE OF ACCREDITATION CERTIFICATE OF ACCREDITATION ANSI National Accreditation Board 11617 Coldwater Road, Fort Wayne, IN 46845 USA This is to certify that Dynatec Scientific Laboratories, Inc. 11940 Golden Gate Road El Paso,

More information

This document is a preview generated by EVS

This document is a preview generated by EVS INTERNATIONAL STANDARD ISO 11930 First edition 2012-04-01 Cosmetics Microbiology Evaluation of the antimicrobial protection of a cosmetic product Cosmétiques Microbiologie Évaluation de la protection antimicrobienne

More information

Accugen Laboratories, Inc.

Accugen Laboratories, Inc. Lab# 97478-2 www.accugenlabs.com Page 1 of 7 Accugen Laboratories, Inc. FINAL REPORT ASTM G22 ASTM Designation: G22-76(1996) "Standard Practice for determining Resistance of Plastics to Bacteria (Withdrawn

More information

MODULE 2F FOOD LABORATORY ACCREDITATION PROGRAM (FOODLAP) ADDITIONAL REQUIREMENTS

MODULE 2F FOOD LABORATORY ACCREDITATION PROGRAM (FOODLAP) ADDITIONAL REQUIREMENTS 2F.1 SCOPE MODULE 2F FOOD LABORATORY ACCREDITATION PROGRAM (FOODLAP) ADDITIONAL REQUIREMENTS The AIHA-Laboratory Accreditation Programs (AIHA-LAP), LLC s Food Laboratory Accreditation Program (FoodLAP)

More information

NCCEH Mould Investigation Toolkit

NCCEH Mould Investigation Toolkit NCCEH Mould Investigation Toolkit Overview of Microbial Sampling Methods NOVEMBER 2014 Table of Contents Air Sampling... 2 Liquid Impinger... 2 Fungal Spore Sampling/Spore Traps... 3 Impactor Samplers

More information

Tests to Support Sterility Claim. Imtiaz Ahmed

Tests to Support Sterility Claim. Imtiaz Ahmed Tests to Support Sterility Claim Imtiaz Ahmed Sterile Product As per TGO 77, a sterile product must comply with the requirements of the following tests: Sterility Test Bacterial Endotoxins Test Appendix

More information

Dutch Accreditation Council (RvA) Explanatory notes on EN-ISO/IEC 17025

Dutch Accreditation Council (RvA) Explanatory notes on EN-ISO/IEC 17025 Dutch Accreditation Council (RvA) Explanatory notes on EN-ISO/IEC 17025 Document code: RvA-T015-UK Version 6, 13-4-2018 Rv A explanatory notes describe the policy and/or the procedures of the Rv A concerning

More information

Halton Vita OR Space Ultraclean conditions and operational flexibility through Controlled Dilution

Halton Vita OR Space Ultraclean conditions and operational flexibility through Controlled Dilution Halton Vita OR Space Ultraclean conditions and operational flexibility through Controlled Dilution Dr Kim Hagström, Halton Kim Hagström Introduction to OR ventilation Operating Room Ventilation Users Needs

More information

HYGIENE. Scheme Description. Environmental Hygiene Monitoring PT Scheme. LGC Standards S.L.U. C/Salvador Espriu 59 2º Barcelona Spain

HYGIENE. Scheme Description. Environmental Hygiene Monitoring PT Scheme. LGC Standards S.L.U. C/Salvador Espriu 59 2º Barcelona Spain HYGIENE Environmental Hygiene Monitoring PT Scheme Scheme Description LGC Standards S.L.U. C/Salvador Espriu 59 2º 08005 Barcelona Spain Tel: +34 93 308 41 81 Fax: +34 93 307 36 12 Email: es@lgcgroup.com

More information

HYGIENE. Scheme Description. Environmental Hygiene Monitoring PT Scheme

HYGIENE. Scheme Description. Environmental Hygiene Monitoring PT Scheme HYGIENE Environmental Hygiene Monitoring PT Scheme Scheme Description LGC Standards Proficiency Testing 1 Chamberhall Business Park Chamberhall Green Bury, BL9 0AP UK. Telephone: +44 (0) 161 762 2500 Fax:

More information

NHS QA Symposium, September 2013

NHS QA Symposium, September 2013 1 www.pmtgb.com Germany - Benelux T o g e t h e r W e c r e a t e S o l u t i o n s France - Great Britain NHS QA Symposium, September 2013 John Cobb Microbiology Support Manager 2 www.pmtgb.com Germany

More information

IQC In Microbiology Testing

IQC In Microbiology Testing IQC In Microbiology Testing Christina Oscroft Campden Technology Ltd Chipping Campden Glos UK Tel: 00 44 1386 842000 email: c.oscroft@campden.co.uk Microbiological results help appraise safety, quality

More information

HYGIENE. Scheme Description. Environmental Hygiene Monitoring PT Scheme

HYGIENE. Scheme Description. Environmental Hygiene Monitoring PT Scheme HYGIENE Environmental Hygiene Monitoring PT Scheme Scheme Description LGC Standards Proficiency Testing 1 Chamberhall Business Park Chamberhall Green Bury, BL9 0AP UK. Telephone: +44 (0) 161 762 2500 Fax:

More information

Approval Block. Prepared by: Signature Date. Evan Parnell 18 JUN Reviewed by: Signature Date. Brian Flynn 18 JUN 2014

Approval Block. Prepared by: Signature Date. Evan Parnell 18 JUN Reviewed by: Signature Date. Brian Flynn 18 JUN 2014 ATS-SOI-5805 Page: 1 of 5 Approval Block Prepared by: Signature Date Evan Parnell 18 JUN 2014 Reviewed by: Signature Date Brian Flynn 18 JUN 2014 Approved by: Signature Date Kristal Jewell 18 JUN 2014

More information

NCCEH Mould Investigation Toolkit

NCCEH Mould Investigation Toolkit NCCEH Mould Investigation Toolkit Interpretation of Microbial Laboratory Reports NOVEMBER 2014 Table of Contents Introduction... 2 Generic Information All Microbial Laboratory Reports... 3 Direct Examination

More information

A2LA. R231 Specific Requirements: Threat Agent Testing Laboratory Accreditation Program. December 6, 2017

A2LA. R231 Specific Requirements: Threat Agent Testing Laboratory Accreditation Program. December 6, 2017 Laboratory Page 1 of 17 Laboratory December 6, 2017 2017 by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means without the prior written permission of A2LA.

More information

HYGIENE. Scheme Description. Environmental Hygiene Monitoring PT Scheme

HYGIENE. Scheme Description. Environmental Hygiene Monitoring PT Scheme HYGIENE Environmental Hygiene Monitoring PT Scheme Scheme Description LGC Standards Proficiency Testing 1 Chamberhall Business Park Chamberhall Green Bury, BL9 0AP UK. Telephone: +44 (0) 161 762 2500 Fax:

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 14729 First edition 2001-04-15 Ophthalmic optics Contact lens care products Microbiological requirements and test methods for products and regimens for hygienic management of

More information

MACRO- AND MICRO-EVALUATION OF AIR INTAKE - A DEMONSTRATION OF THE NEED FOR MORE OPTIMAL TOOLS

MACRO- AND MICRO-EVALUATION OF AIR INTAKE - A DEMONSTRATION OF THE NEED FOR MORE OPTIMAL TOOLS MACRO- AND MICRO-EVALUATION OF AIR INTAKE - A DEMONSTRATION OF THE NEED FOR MORE OPTIMAL TOOLS F Frydenlund, EN Haugen *, C Ahlén, I Bryn, H Davidsen, SO Hanssen and GT Møgedal SINTEF Energy Research,

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 18593 First edition 2004-06-01 Microbiology of food and animal feeding stuffs Horizontal methods for sampling techniques from surfaces using contact plates and swabs Microbiologie

More information

PERFORMANCE SPECIFICATIONS

PERFORMANCE SPECIFICATIONS Recycler TM Product and Performance Specifications The Recycler TM family of specialty water purification systems has two models specifically designed to ether provide significant amounts of purified drinking

More information

HYGIENE. Scheme Description. Environmental Hygiene Monitoring PT Scheme

HYGIENE. Scheme Description. Environmental Hygiene Monitoring PT Scheme HYGIENE Environmental Hygiene Monitoring PT Scheme Scheme Description LGC Standards Proficiency Testing 1 Chamberhall Business Park Chamberhall Green Bury, BL9 0AP UK. Telephone: +44 (0) 161 762 2500 Fax:

More information

PERFORMANCE QUALIFICATION PROTOCOL HVAC SYSTEM

PERFORMANCE QUALIFICATION PROTOCOL HVAC SYSTEM Page 1 of 24 PERFORMANCE QUALIFICATION PROTOCOL FOR HVAC SYSTEM Signing of this Performance Qualification Protocol indicates agreement with the Validation Master Plan approach of the equipment. Further

More information

Orthophenylphenol in healthcare environments: a trial related to a new administration method and a review of the literature*

Orthophenylphenol in healthcare environments: a trial related to a new administration method and a review of the literature* Turkish Journal of Medical Sciences http://journals.tubitak.gov.tr/medical/ Research Article Turk J Med Sci (2013) 43: 805-809 TÜBİTAK doi:10.3906/sag-1208-4 Orthophenylphenol in healthcare environments:

More information

How to Handle Excursions in Environmental Monitoring (EM) and Personnel Monitoring (PM) in an Aseptic Processing Plant

How to Handle Excursions in Environmental Monitoring (EM) and Personnel Monitoring (PM) in an Aseptic Processing Plant How to Handle Excursions in Environmental Monitoring (EM) and Personnel Monitoring (PM) in an Aseptic Processing Plant Randy Hutt, Ph.D. Associate Director Sterility Assurance and Microbiology Luitpold

More information

Synergy Health Ireland Ltd.

Synergy Health Ireland Ltd. IDA Business & Technology Park, Tullamore, Co.Offaly Testing Laboratory Registration number: 164T is accredited by the Irish National Board (INAB) to undertake testing as detailed in the Schedule bearing

More information

Official Controls for Salmonella spp. in Bivalve Molluscs, live echinoderms, gastropods and tunicates

Official Controls for Salmonella spp. in Bivalve Molluscs, live echinoderms, gastropods and tunicates European Union Reference Laboratory for monitoring bacteriological and viral contamination of bivalve molluscs Use of proprietary Salmonella detection kits for detection and identification of Salmonella

More information

PAPER TEMPLATE FOR THE 34 TH AIVC-3 RD TIGHTVENT-2 ND COOL ROOFS - 1 ST VENTICOOL CONFERENCE, 2013

PAPER TEMPLATE FOR THE 34 TH AIVC-3 RD TIGHTVENT-2 ND COOL ROOFS - 1 ST VENTICOOL CONFERENCE, 2013 PAPER TEMPLATE FOR THE 34 TH AIVC-3 RD TIGHTVENT-2 ND COOL ROOFS - 1 ST VENTICOOL CONFERENCE, 2013 Sanitary aspects of domestic ventilation systems: an in situ study Joris Van Herreweghe*, Samuel Caillou,

More information

Microgen and Airocide Technology as a Method for Sterilization in ICU: A Comparative Study

Microgen and Airocide Technology as a Method for Sterilization in ICU: A Comparative Study ISSN: 2319-7706 Special Issue-1 (2015) pp. 281-285 http://www.ijcmas.com Original Research Article Microgen and Airocide Technology as a Method for Sterilization in ICU: A Comparative Study Syed Moied

More information

A MICROBIAL RESISTANCE EVALUATION OF INDOOR MATERIALS AIR KRETE INSULATION SAMPLE. prepared for AIR KRETE

A MICROBIAL RESISTANCE EVALUATION OF INDOOR MATERIALS AIR KRETE INSULATION SAMPLE. prepared for AIR KRETE A MICROBIAL RESISTANCE EVALUATION OF INDOOR MATERIALS AIR KRETE INSULATION SAMPLE prepared for AIR KRETE May 16, 2003 AQS Report No. 10681-02 TABLE OF CONTENTS Page Number Executive Summary Project Description...

More information

CASE STUDIES OF DUST AMS CALIBRATION AT EXTREME CONDITIONS

CASE STUDIES OF DUST AMS CALIBRATION AT EXTREME CONDITIONS CASE STUDIES OF DUST AMS CALIBRATION AT EXTREME CONDITIONS Jurij ČRETNIK RACI d.o.o., Tehnološki park 24, SI-1000 LJUBLJANA, SLOVENIA jurij.cretnik@raci.si Matjaž ŠLIBAR RACI d.o.o., Tehnološki park 24,

More information

IN THIS SECTION MICROBIOLOGY TESTING EXPERT SOLUTIONS FOR PRODUCT DEVELOPMENT. Bacterial Endotoxin (LAL) Testing

IN THIS SECTION MICROBIOLOGY TESTING EXPERT SOLUTIONS FOR PRODUCT DEVELOPMENT. Bacterial Endotoxin (LAL) Testing EXPERT SOLUTIONS FOR PRODUCT DEVELOPMENT IN THIS SECTION MICROBIOLOGY TESTING Microbial assays involve a variety of tests, from the determination of the numbers and types of organisms naturally present

More information

Public Health England (PHE) Certified Reference Materials for Microbiology. Certificate of Analysis Salmonella Nottingham CRM07832L

Public Health England (PHE) Certified Reference Materials for Microbiology. Certificate of Analysis Salmonella Nottingham CRM07832L Public Health England (PHE) Certified Reference Materials for Microbiology Certificate of Analysis Salmonella Nottingham CRM07832L This certificate should not be parted from the PHE certified reference

More information

Food Microbiological Examination: General Guidelines

Food Microbiological Examination: General Guidelines Translated English of Chinese Standard: GB4789.1-2010 Translated by: www.chinesestandard.net Wayne Zheng et al. Email: Sales@ChineseStandard.net NATIONAL STANDARD GB OF THE PEOPLE S REPUBLIC OF CHINA GB

More information

Final Report sub-project (1)

Final Report sub-project (1) FACULTY OF BIOSCIENCE ENGINEERING Dhr. Dirk de Meester Dhr. Steven Hagens EBI FOOD SAFETY Johan v. Oldenbarneveltlaan 9 2582 NE Den Haag NEDERLAND Your reference Offer 13/11/2006 Our reference LV 07rep-03

More information

Public Health England (PHE) Certified Reference Materials for Microbiology. Certificate of Analysis Legionella pneumophila CRM12821M

Public Health England (PHE) Certified Reference Materials for Microbiology. Certificate of Analysis Legionella pneumophila CRM12821M Public Health England (PHE) Certified Reference Materials for Microbiology Certificate of Analysis Legionella pneumophila CRM12821M This certificate should not be parted from the PHE certified reference

More information

AIR QUALITY ASSESSMENT FOR TAYLOR STANLEY MARKETING PTE LTD 71, TANNERY LANE #-7-07, CITY INDUSTRIAL BUILDING SINGAPORE

AIR QUALITY ASSESSMENT FOR TAYLOR STANLEY MARKETING PTE LTD 71, TANNERY LANE #-7-07, CITY INDUSTRIAL BUILDING SINGAPORE Our Ref: ATS/IH/84/05KKL Date: 26 August, 2005 Page 1 of 33 AIR QUALITY ASSESSMENT FOR TAYLOR STANLEY MARKETING PTE LTD 71, TANNERY LANE #-7-07, CITY INDUSTRIAL BUILDING SINGAPORE 347807 Date of Survey

More information

Program OK 10. Bio products degradation in soil

Program OK 10. Bio products degradation in soil Application Date : 2012-03-01 Page : 1 / 5 Program OK 10 Bio products degradation in soil 1. Scope 4 The OK biodegradable SOIL certificate can be granted to the following materials or products: - All raw

More information

Environmental Monitoring of Aseptic Processing Areas - 1

Environmental Monitoring of Aseptic Processing Areas - 1 Environmental Monitoring of Aseptic Processing Areas - 1 A war against an invisible enemy DCVMN - Víctor Maqueda - May 30, 31, June 1 2016 1 Training Course Agenda Overview of Environmental Monitoring

More information

Overview of a sterility assurance program for PET drugs

Overview of a sterility assurance program for PET drugs Coalition for PET Drug Approval Radiopharmaceutical Sciences Council Overview of a sterility assurance program for PET drugs Eric Webster, PETNET Solutions Disclosures Employee of PETNET Solutions, a Siemens

More information

HYGIENE. Scheme Description. Environmental Hygiene Monitoring PT Scheme

HYGIENE. Scheme Description. Environmental Hygiene Monitoring PT Scheme HYGIENE Environmental Hygiene Monitoring PT Scheme Scheme Description LGC Standards Proficiency Testing 1 Chamberhall Business Park Chamberhall Green Bury, BL9 0AP UK. Telephone: +44 (0) 161 762 2500 Fax:

More information

Airborne contamination in the operating theatre is one of the causes of healthcare-related infections.

Airborne contamination in the operating theatre is one of the causes of healthcare-related infections. Airborne contamination in the operating theatre is one of the causes of healthcare-related infections. Toul Meditech guarantee ultra-clean air of 5cfu/m 3 in critical areas, such as the wound site as well

More information

Validation of cleaning and sterilization of reusable medical devices

Validation of cleaning and sterilization of reusable medical devices Synergy makes sence Validation of cleaning and sterilization of reusable medical devices MG-FSI72-119 Last revision: June 2012 Aurélien BIGNON 5, Chemin du Catupolan - 69120 Vaulx en Velin - France - Tel.

More information

The Microbiological Requirements of a Stability Study. Ngoc Anh-Thu Phan 19 th June 2012

The Microbiological Requirements of a Stability Study. Ngoc Anh-Thu Phan 19 th June 2012 The Microbiological Requirements of a Stability Study Ngoc Anh-Thu Phan 19 th June 2012 Contents Introduction Harmonized Tests Acceptance Criteria Selection of Tests Why PE Testing? Selection of Category

More information

ISO INTERNATIONAL STANDARD. Cleanrooms and associated controlled environments Biocontamination control Part 1: General principles and methods

ISO INTERNATIONAL STANDARD. Cleanrooms and associated controlled environments Biocontamination control Part 1: General principles and methods INTERNATIONAL STANDARD ISO 14698-1 First edition 2003-09-01 Cleanrooms and associated controlled environments Biocontamination control Part 1: General principles and methods Salles propres et environnements

More information

Media Fill A Process Simution. Presented By Shikha Chauhan

Media Fill A Process Simution. Presented By Shikha Chauhan Media Fill A Process Simution Presented By Shikha Chauhan 21 CFR 211.113 Validation of Aseptic Processing and Sterilization Process Simulation / Media Fill Filtration Efficacy Sterilization of Equipment,

More information

Biological Consulting Services of North Florida, Inc.

Biological Consulting Services of North Florida, Inc. Biological Consulting Services of North Florida, Inc. June 29, 2015 Village Water Filters 1258 Rainbow Drive #2839 Silverthrone, CO 80498 RE: Biological filtration efficacy testing of the filters provided

More information

ISO/TS TECHNICAL SPECIFICATION

ISO/TS TECHNICAL SPECIFICATION TECHNICAL SPECIFICATION ISO/TS 11133-2 First edition 2003-12-15 Microbiology of food and animal feeding stuffs Guidelines on preparation and production of culture media Part 2: Practical guidelines on

More information

SURFACE WATER SAMPLING PROCEDURES

SURFACE WATER SAMPLING PROCEDURES SURFACE WATER SAMPLING PROCEDURES Section 5 September 20, 2017 Status & Trend Networks Sampling Manual Carolyn Voyles WMS Project Manager Major Topics Determination of When / When Not to Collect Samples

More information

Kit Information 4 Introduction. 4 Kit Contents, Storage, and Testing Conditions. 4 Principle 4 Applicability. 5 Precautions Sample Preparation 6

Kit Information 4 Introduction. 4 Kit Contents, Storage, and Testing Conditions. 4 Principle 4 Applicability. 5 Precautions Sample Preparation 6 Contents 3. Kit Information 4 Introduction. 4 Kit Contents, Storage, and Testing Conditions. 4 Principle 4 Applicability. 5 Precautions.......................... 5 Sample Preparation 6 Test Procedure 7

More information

Discussion Topics. Some process engineering considerations

Discussion Topics. Some process engineering considerations Discussion Topics Definition of verification Introduction of ICMSF sampling terms Purposes of microbiological testing Maximizing the value of verification data Some process engineering considerations Definition:

More information

AN INVESTIGATION ON THE OCCURRENCE OF FUNGI AND BACTERIA IN THE MVAC SYSTEM IN AN OFFICE PREMISE

AN INVESTIGATION ON THE OCCURRENCE OF FUNGI AND BACTERIA IN THE MVAC SYSTEM IN AN OFFICE PREMISE AN INVESTIGATION ON THE OCCURRENCE OF FUNGI AND BACTERIA IN THE MVAC SYSTEM IN AN OFFICE PREMISE PK Chow, WY Chan *, LLP Vrijmoed Department of Biology and Chemistry, City University of Hong Kong, Tat

More information

Bacterial Counts - Quantitative Analysis of Microbes

Bacterial Counts - Quantitative Analysis of Microbes Bacterial Counts - Quantitative Analysis of Microbes Introduction: It is often important to know not only what types of bacteria are in a sample but also how many of them are present. Food manufacturers

More information

A FIELD COMPARISON OF METHODS FOR ENUMERATING AIRBORNE FUNGAL BIOAEROSOLS

A FIELD COMPARISON OF METHODS FOR ENUMERATING AIRBORNE FUNGAL BIOAEROSOLS A FIELD COMPARISON OF METHODS FOR ENUMERATING AIRBORNE FUNGAL BIOAEROSOLS KS Lee 1*, W Black 2, M Brauer 1, G Stephens 2, K Teschke 1, J Hsieh 1 and K Bartlett 1 1 School of Occupational and Environmental

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK aboratory locations: Schedule of Accreditation 2 Pine Trees, Chertsey ane, Staines-upon-Thames, TW18 3HR, UK Royal iverpool and Broadgreen University Hospitals NHS Trust, trading as Quality Control North

More information

Biological Consulting Services of North Florida, Inc.

Biological Consulting Services of North Florida, Inc. Biological Consulting Services of North Florida, Inc. September 09, 2013 Travis Merrigan Grayl 610 Maple Heights Rd Camano Island, WA 98282 Re: Bacterial, viral, and protozoan parasite filtration efficacy

More information

Laminar flow in orthopaedic operations - Essential if you can afford it. Peter Hoffman Consultant Clinical Scientist Public Health England

Laminar flow in orthopaedic operations - Essential if you can afford it. Peter Hoffman Consultant Clinical Scientist Public Health England Laminar flow in orthopaedic operations - Essential if you can afford it Peter Hoffman Consultant Clinical Scientist Public Health England Up until the 1950s in the UK (and elsewhere) Surgical instruments

More information

Surface Fungicidal Activity of Clinell Universal Wipes ~ Project Report Prepared for Gama Healthcare Ltd.

Surface Fungicidal Activity of Clinell Universal Wipes ~ Project Report Prepared for Gama Healthcare Ltd. Surface Fungicidal Activity of Clinell Universal Wipes ~ Project Report Prepared for Gama Healthcare Ltd. Inspiring tomorrow s professionals Surface Fungicidal Activity of Clinell Universal Wipes Author:

More information

USP <797> and Environmental Sampling

USP <797> and Environmental Sampling USP and Environmental Sampling Participants will be in listen only mode. 9 a.m. (PT) Download the PDF: https://www.emlab.com/m/media/usp797-webinar.pdf Presented by: Michael Berg, Ph.D. Continuing

More information

Specific Accreditation Criteria Materials ISO/IEC Annex. Cement testing

Specific Accreditation Criteria Materials ISO/IEC Annex. Cement testing Specific Accreditation Criteria Materials ISO/IEC 17025 Annex Cement testing January 2018 Copyright National Association of Testing Authorities, Australia 2013 This publication is protected by copyright

More information

Biological Consulting Services of North Florida, Inc.

Biological Consulting Services of North Florida, Inc. Biological Consulting Services of North Florida, Inc. June 8, 2016 Sagan, LLC 11035 Technology Place, Suite 100 San Diego, CA 92127 858-675-7017 ext. 2000 RE: Biological filtration efficacy test study

More information

School of Environmental & Resource, Inner Mongolia University, Huhhot, China

School of Environmental & Resource, Inner Mongolia University, Huhhot, China Advanced Materials Research Online: 2013-08-16 ISSN: 1662-8985, Vols. 726-731, pp 1373-1377 doi:10.4028/www.scientific.net/amr.726-731.1373 2013 Trans Tech Publications, Switzerland Spatial and temporal

More information

QCS. Scheme Description. Quality in Chocolate Scheme

QCS. Scheme Description. Quality in Chocolate Scheme QCS Quality in Chocolate Scheme Scheme Description LGC Standards Proficiency Testing 1 Chamberhall Business Park Chamberhall Green Bury Lancashire BL9 0AP United Kingdom Telephone: +44 (0) 161 762 2500

More information

TECHNICAL DATASHEET. in compliance with: USP, ISO and all appropriate subsections FQ 106 v3.0 August, 2015

TECHNICAL DATASHEET. in compliance with: USP, ISO and all appropriate subsections FQ 106 v3.0 August, 2015 ETIGAM TECHNICAL DATASHEET ETIGAM NSS biological indicator spore strips for radiation Bacillus pumilus This technical report provides relevant data and instructions for use of the NSS biological indicator

More information

Plasma Low Temperature Sterilization. With vaporized hydrogen peroxide. PlazMax Line

Plasma Low Temperature Sterilization. With vaporized hydrogen peroxide. PlazMax Line Plasma Low Temperature Sterilization With vaporized hydrogen peroxide PlazMax Line Low Temperature Sterilization PlazMax Line For heat and moisture sensitive equipment PlazMax provides an efficient sterilization

More information

Indoor Air Quality Report

Indoor Air Quality Report Indoor Air Quality Report Shawnee City Hall 222 N. Broadway, Shawnee, Oklahoma 74801 INDOOR AIR QUALITY SAMPLING REPORT Project Locations: Shawnee City Hall 222 N. Broadway, Shawnee, Oklahoma 74801 Prepared

More information

Biological Consulting Services of North Florida, Inc.

Biological Consulting Services of North Florida, Inc. Biological Consulting Services of North Florida, Inc. May 16, 2014 Mr. Dennis Brown Aquamira Technologies, Inc. 917 West 600 North Ste 105 Logan, Utah 84321 P: (360) 306-5586 Re: Microbial cyst filtration

More information

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (16) 86 R2 MANAGEMENT OF ENVIRONMENTAL CONDITIONS Full document title and reference Document type Management of Environmental

More information

MICROBIOLOGICAL VALIDATION PROGRAMME FOR SRCL CLINICAL WASTE TREATMENT AUTOCLAVE PROCESSING METALS (SINGLE USE DEVICES) AT LEEDS

MICROBIOLOGICAL VALIDATION PROGRAMME FOR SRCL CLINICAL WASTE TREATMENT AUTOCLAVE PROCESSING METALS (SINGLE USE DEVICES) AT LEEDS MICROBIOLOGICAL VALIDATION PROGRAMME FOR PROCESSING METALS (SINGLE USE DEVICES) AT LEEDS M G Holliday MBA, PhD, MSc, CSci, FIBMS. MCIWM. F InstLM. Microbiology Department Freeman Hospital Newcastle upon

More information

THE EFFECT OF DUCT CLEANING ON INDOOR AIR QUALITY IN OFFICE BUILDINGS

THE EFFECT OF DUCT CLEANING ON INDOOR AIR QUALITY IN OFFICE BUILDINGS THE EFFECT OF DUCT CLEANING ON INDOOR AIR QUALITY IN OFFICE BUILDINGS S Kolari 1*, M Luoma 1, M Ikäheimo 2 and P Pasanen 2 1 VTT Technical Research Centre of Finland, Espoo, Finland 2 University of Kuopio,

More information

Biocontamination control in pharmaceutical production

Biocontamination control in pharmaceutical production White Paper Data Sheet Biocontamination control in pharmaceutical production and compliance to ISO 14698 with validated active microbial air samplers Tim Sandle, Ph.D., Head of Microbiology at Bio Products

More information

INTRODUCTION. History RAYPA was founded by Mr. Ramón Espinar nd generation take over New makeover

INTRODUCTION. History RAYPA was founded by Mr. Ramón Espinar nd generation take over New makeover INTRODUCTION INTRODUCTION History 1970 RAYPA was founded by Mr. Ramón Espinar 2008 2 nd generation take over 2016 New makeover 2018 30 employees Turn over: 3.6 M 85% Sales on Export Sales in more than

More information

Exercise 13 DETERMINATION OF MICROBIAL NUMBERS

Exercise 13 DETERMINATION OF MICROBIAL NUMBERS Exercise 13 DETERMINATION OF MICROBIAL NUMBERS Introduction When biologists discuss the growth of microorganisms (microbial growth), they are actually referring to population size rather than to the size

More information

BIO & PHARMA ANALYTICAL TECHNIQUES

BIO & PHARMA ANALYTICAL TECHNIQUES BIO & PHARMA ANALYTICAL TECHNIQUES Chapter 11 by Dr. Siti Umairah Mokhtar Faculty of Engineering Technology umairah@ump.edu.my Chapter Description Aims Discuss theory, principles and application of analytical

More information

The attached report is the result of a mold and bacterial assessment conducted at th Street, Franklin, Ohio on April 24 and 25, 2008.

The attached report is the result of a mold and bacterial assessment conducted at th Street, Franklin, Ohio on April 24 and 25, 2008. May 8, 2008 Jon Lampros President Clean Air Systems 701 Rockland Rd. Lake Bluff, IL 60044 Re: Assessment of Ozone Treatment on Microbial Populations Dear Mr. Lampros, The attached report is the result

More information

PHARMACY IV ADMIXTURE. Pharmacy 483 January 17, 2006 Kim Donnelly Affiliate Associate Professor

PHARMACY IV ADMIXTURE. Pharmacy 483 January 17, 2006 Kim Donnelly Affiliate Associate Professor PHARMACY IV ADMIXTURE Pharmacy 483 January 17, 2006 Kim Donnelly Affiliate Associate Professor Pharmacist is responsible for ensuring that compounded sterile preparations are properly prepared, labeled,

More information

Methods of microbiological examination for air in public places - Determination of aerobic bacterial count

Methods of microbiological examination for air in public places - Determination of aerobic bacterial count Translated English of Chinese Standard: GB/T18204.1-2000 Translated by: www.chinesestandard.net Wayne Zheng et al. Email: Sales@ChineseStandard.net NATIONAL STANDARD OF THE PEOPLE S REPUBLIC OF CHINA GB

More information

MARCH BIOAEROSOL IMPACT ASSESSMENT AT KILMAINHAMWOOD

MARCH BIOAEROSOL IMPACT ASSESSMENT AT KILMAINHAMWOOD ODOUR MONITORING IRELAND LTD Unit 32 De Granville Court, Dublin Rd, Trim, Co. Meath Tel: +353 46 9437922 Mobile: +353 86 8550401 E-mail: info@odourireland.com www.odourireland.com MARCH 2016 - BIOAEROSOL

More information

RapidChek SELECT Salmonella Test Kit

RapidChek SELECT Salmonella Test Kit RapidChek SELECT Salmonella Test Kit Part #: 7000188-7000197 AOAC Approved Protocols: This test kit s performance was reviewed by AOAC Research Institute and was found to perform to the manufacturer s

More information

Evaluation of the performance of the Eudemon International Freshlight CFL

Evaluation of the performance of the Eudemon International Freshlight CFL Evaluation of the performance of the Eudemon International Freshlight CFL Test 2 Bench Scale Agar Plate Test A report prepared by Dr Louise Fletcher Institute of Pathogen Control Engineering (PaCE) School

More information

AirClean Food Industry Mar. 12

AirClean Food Industry Mar. 12 Micro organisms after incubation of a sampler of fresh air inlet in an air handling unit. Colonies usually live in the air conditioning system. Contamination on cheese during ripening. Sanitation is a

More information

Microbiology for Oral and Topical Products - The basics Scott Colbourne Business Manager NSW ALS Food & Pharmaceutical

Microbiology for Oral and Topical Products - The basics Scott Colbourne Business Manager NSW ALS Food & Pharmaceutical Microbiology for Oral and Topical Products - The basics Scott Colbourne Business Manager NSW ALS Food & Pharmaceutical RIGHT S O L U T I O N S RIGHT PARTNER Contents TGO 77 - Introduction Tests Performed

More information

Stem Cell Laboratory 8/F, Li Ka Shing Institute of Health Sciences

Stem Cell Laboratory 8/F, Li Ka Shing Institute of Health Sciences Stem Cell Laboratory 8/F, Li Ka Shing Institute of Health Sciences Constructed : Australian Therapeutic Goods Administration (TGA) standard (Orthocell Proprietary Limited Company, consultant for GMP setup)

More information

PHARMACY IV ADMIXTURE

PHARMACY IV ADMIXTURE PHARMACY IV ADMIXTURE Pharmacy 483 January 18, 2007 Kim Donnelly Affiliate Assistant Professor kimdon@u.washington.edu Pharmacist is responsible for ensuring that compounded sterile preparations are properly

More information

Biological Monitoring

Biological Monitoring Biological Monitoring Prof. Duygu PERÇİN, MD Department of Clinical Microbiology Erciyes University Faculty of Medicine, Kayseri-TURKEY duygu.percin@hotmail.com Learning Objectives The attandee will be

More information

Annex A2. Guidance on Process Validation Scheme for Aseptically Processed Products

Annex A2. Guidance on Process Validation Scheme for Aseptically Processed Products Annex A2 Guidance on Process Validation Scheme for Aseptically Processed Products 1 Table of content 1 PURPOSE... 3 2 SCOPE... 3 3 GENERAL INFORMATION... 3 4 INFORMATION NEEDED FOR ASEPTIC PROCESSES VALIDATION...

More information

Air Quality Monitoring of Operation Theaters in Government Medical College and Hospital, Aurangabad, India

Air Quality Monitoring of Operation Theaters in Government Medical College and Hospital, Aurangabad, India International Journal of Current Microbiology and Applied Sciences ISSN: 2319-7706 Volume 5 Number 6 (2016) pp. 42-49 Journal homepage: http://www.ijcmas.com Original Research Article http://dx.doi.org/10.20546/ijcmas.2016.506.005

More information

Review Validation of aseptic processes for pharmaceuticals

Review Validation of aseptic processes for pharmaceuticals OPEM www.opem.org Oriental Pharmacy and Experimental Medicine 2010 10(4), 231-238 DOI 10.3742/OPEM.2010.10.4.231 Review Validation of aseptic processes for pharmaceuticals Lincy Joseph*, Mathew George

More information

Understanding new Acceptance Criteria of Alternative Assemblies Thermal and Ignition Barriers over SPF

Understanding new Acceptance Criteria of Alternative Assemblies Thermal and Ignition Barriers over SPF Understanding new Acceptance Criteria of Alternative Assemblies Thermal and Ignition Barriers over SPF Why do we need to cover the SPF Spray Polyurethane Foam (SPF) is combustible As the SPF burns and

More information