Performance of the Sievers 500 RL On-Line TOC Analyzer

Size: px
Start display at page:

Download "Performance of the Sievers 500 RL On-Line TOC Analyzer"

Transcription

1 GE Water & Process Technologies Analytical Instruments Performance of the Sievers 500 RL On-Line TOC Analyzer imagination at work

2 Page 2 of 10 Introduction To determine the performance characteristics of the Sievers* 500 RL On-Line TOC Analyzer, a series of measurements of certified TOC standards, conductivity standards and on-line measurements on a low TOC water system were performed. The TOC measurements were performed on six analyzers from the beta build, five of which were equipped with the Super ios. * The on-line measurements were performed to estimate the limit of detection (LOD) using three base model analyzers, one with the standard ios and one with a Super ios. The conductivity measurements were performed on six different analyzers from the pilot production since the tubing in these instruments was passivated using the procedures that will be employed in production units. The conductivity linearity test was performed during the project's validation stage using two of the pre-production alpha units. The TOC testing involved running replicates of the analyzers system protocols which are listed in Table 1. All the standards used for these tests were prepared by our standards laboratory except for the high level standards of 2 and 2.5 ppm potassium hydrogen phthalate (KHP), which were prepared by one of the test participants. The TOC standards for the Model 500 RL are stored in glass vials and acidified to ph 3 with phosphoric acid to minimize absorption of atmospheric CO 2. To determine if there is any difference in the system suitability test using acidified standards, standards for the Sievers 400 TOC Analyzers, which are not acidified and stored in plastic vials, were also measured on the Sievers 500 RL. Table 1 Summary of TOC system protocols used for evaluation of TOC measurement performance of the Sievers 500 RL TOC Analyzer Standard Set Composition Replicates Reagent Water 4 Verification 500 ppb NIST Sucrose 12 for pilot production units 25 μs/cm HCl Acidified System Suitability Reagent Water 500 ppb Sucrose 500 ppb Benzoquinone Rinse Non-acidified Same (no rinse) 3 System Suits Linearity High Level Standards 750 ppb KHP The six 500 RL analyzers were connected to a small, online water system and the analyzers built-in rinse mode initiated. The analyzers were then allowed to operate on-line for two days prior to this study. The analyzers were calibrated using either the recommended singlepoint calibration at 1.5 ppm TOC using KHP, or the optional multi-point calibration which employs calibration with KHP at 0.5, 1.0 and 1.5 ppm. After calibration, a verification was performed, then analyzers returned to on-line measurements of the water system overnight. Over a two week period, the tests in Table 1 were performed on the analyzers, with a daily system suitability test and typically two other system protocols per day. Other than the daily system suitability test and the high levels tests which were performed at the end of the two-week period, the other tests were performed in random order. For the instruments with the Super ppb KHP 6 5 for S/N ppb KHP Reagent Water Calibration Blank 1000 ppb KHP 1500 ppb KHP 2 2 measurements of Blank 2500 ppb on S/N ppb KHP 2500 ppb KHP

3 Page 3 of 10 ios, the auto-restart options were turned on so that the instrument would return to on-line monitoring of the water system once the protocols were completed. The analyzer with the standard ios was not always returned to on-line monitoring of the water system overnight. To determine the limit of detection of the Sievers 500 RL, five analyzers were connected to a low TOC water system which was operated with the dissolved oxygen concentration at ~4 ppb. The analyzers and three Sievers PPT Analyzers were operated for two weeks and the measurement of the water system inspected to identify periods of stable TOC levels. Seven such periods encompassing 5 22 hours of measurements were identified and used for the LOD calculations. When the pilot production units became available, these analyzers were calibrated, then 2 3 verification protocols were run each day over the course of a week. Only the conductivity data from these measurements are included in this report. In a separate study prior to the TOC and conductivity measurements in this report, tests of the linearity of conductivity measurements using a larger water system and a standard addition device were performed. The standard addition device consists of a syringe pump and mixing system that allows on-line production of conductivity standards. By varying the concentration of the solution used to fill the syringe and the syringe flow rate, accurate standards over a range of concentrations can be produced. For the conductivity linearity tests, stock solutions of NaCl were prepared and the standard addition device used to generate conductivity standards from to 35 µs/cm. On-line measurements from two Sievers 500 RL analyzers were obtained during these tests, along with a conventional on-line conductivity sensor (Thornton T-1000). Outliers Several outliers were identified in the accuracy and precision data set. In eight of the standards sets (four system suitability tests, one linearity, one verification and two high level standards sets), unexpectedly high TOC (> 100 ppb) was measured in the reagent water blank. The average TOC from 156 measurements of reagent water was 29 ppb, suggesting that there was some form of TOC contamination in those reagent water samples that gave high TOC values. Since accuracy and precision are calculated from blank-corrected values, these samples with high reagent water TOC were excluded from the data set. Two of the non-acidified system suitability measurements gave unexpectedly low values for benzoquinone (~ 20 ppb), possibly due to mislabeling of reagent water blank vials. A Grubb's test was applied to the remaining 340 TOC standards, and seven outliers for accuracy (due to higher than expected TOC) and one outlier for poor precision were identified and removed from the data set. Figure 1 shows a histogram and box plot for accuracy measurements obtained from the six Sievers 500 RL TOC Analyzers. For the box plot, the center line is the median, the ends of the box are the second and third quartiles. The width of the box is the interquartile range. The means diamond shows the mean and 95% confidence interval of the mean. The bracket along the edge of the plot shows the most dense 50% of the data. The whiskers extend from the box to the outermost data points within ±1.5 times the interquartile distance. Possible outliers are shown outside the whiskers. TOC Accuracy and Precision For each of the TOC standards analyzed in this study (356 total measurements), the accuracy of the measurement was calculated from the blank-corrected average of the three replicate measurement as (expected-measured)/expected. The precision was calculated from the standard deviation of the three replicate measurements divided by the blankcorrected average. Figure 1. Histogram of Accuracy Measurement using six Sievers 500 RL Analyzers along with ± 5% Specification Limits

4 Page 4 of 10 Table 2 Statistics from Accuracy Measurements Statistic Value Mean -0.3% Std Dev 2.9% Std Err Mean 0.16% N 333 Skewness 0.25 Kurtosis 0.77 Statistics for the data set are shown in Table 2. The data are not normally distributed (Shapiro-Wilk W Test) and the kurtosis indicates the distribution is peaked, with more values in the center than a normal distribution. A total of 9.9% of the measurements were outside the ± 5% accuracy specification and a capability analysis indicates that in the long run, one would expect 9% of the measurements to be outside the ± 5% limits. A plot of the absolute value of accuracy versus concentration along with a least-squares regression line is shown in Figure 2. While slightly poorer accuracy was observed at the lower concentrations, the correlation is not statistically significant (p > 0.05). The regression is also strongly influenced by the large number of measurements at 500 ppb (200) versus the 16 to 34 measurements at the other concentrations A histogram and box plot for the precision of replicate measurements of the TOC standards along with the 1% specification is shown in Figure 3. A total of 1.8% of the measurements were outside the 1% precision specification, with a mean precision of 0.2% and a standard deviation of 0.24%. A plot of precision versus the concentration of the TOC standards is shown in Figure 4, with poorer precision observed at the lower concentrations (p < 0.05). Figure 5 shows box plots of the accuracy and precision of the TOC measurements by serial number of the Sievers 500 RL TOC analyzers. Serial number 26 showed the poorest accuracy with the largest variation in accuracy. A means comparison of accuracy using the Tukey-Kramer HSD test indicates that serial number 26 is significantly different from the other Sievers 500 RL used in this study. The poorest precision was observed for serial numbers 18 and 23, although only serial number 18 had a significantly different mean precision. Figure 2. Absolute value of Accuracy measurements versus concentration of TOC standard. Figure 3. Histogram of Precision Measurements using six Sievers 500 RL TOC Analyzers along with 1% specification limit Figure 4. Precision versus concentration of the TOC standards

5 Page 5 of 10 Figure 5. Box plots of Accuracy and Precision by Analyzer Figure 6. Box plots of Accuracy and Precision by Calibration Serial numbers 23, 26, and 27 were the analyzers that were calibrated using the optional multi-point calibration. As shown in Figure 6, there was a slightly poorer mean accuracy for the analyzers calibrated using multi-point versus single-point due to serial number 26. There was no difference in the mean precision between single-point versus multi-point calibration. System Suitability The system suitability test is part of the US Pharmacopoeia Method <643> for TOC measurement of Purified Water and Water for Injection. The test involves analysis of a reagent water blank, a 500 ppb solution of sucrose (viewed as a easy to oxidize compound) and a 500 ppb solution of 1,4-benzoquinone (viewed as a more difficult to oxidize compound ). To determine the suitability of a TOC analyzer for this method, a response efficiency is calculated from the ratio of the blank corrected TOC value for the benzoquinone solution divided by the blank corrected TOC of the sucrose solution. An analyzer is deemed suitable if the response efficiency is between 85% and 115%. A histogram and box plot of 54 measurements of the response efficiency on the six Sievers 500 RL s is shown in Figure 7 along with the USP specification limits. The response efficiencies range from 98% to 107% with a mean of 101% and a standard deviation of 1.8%. A capability analysis indicates that in the long run, no response efficiencies outside the USP limits would be expected using the Sievers 500 RL. The ability of the Sievers 500 RL to pass the system suitability test can also be expressed in terms of the process capability (C p ) or the Process Capability Index (C pk ). C p is

6 Page 6 of 10 simply the difference between the upper and lower specification limits divided by six times the standard deviation of the measurements. Since the C p does not not account for processes that are not centered between the limits, the process capability index is often used and is defined as the the minimum distance between the specification limits and the process mean, divided by three times the standard deviation. For a capable measurements system, values of C p and C pk should be > 1 and a value of 1.33 is often used an a minimum acceptable value. For these system suitability measurements on the Sievers 500 RL analyzers, the C p was 2.8 and the C pk was 2.5, indicating the analyzer is capable of passing the system suitability test. For the Sievers 500 RL, standards solutions are stored in glass vials and acidified to ph 3 with phosphoric acid to minimize absorption of atmospheric CO 2. Acidification may also help to stabilize the photochemically reactive 1,4-benzoquinone. While a stability test of the acidified standards was not part of this study, a comparison was made of the 500 RL standards and non-acidified standards stored in plastic vials. Figure 8 shows box plots of the response efficiency calculated from measurement of 37 acidified sucrose and benzoquinone standards and 16 non-acidified solutions. There is no significant difference in the means or in the standard deviations of the response efficiency using acidified versus non-acidified standards. TOC Linearity The linearity portion of this study consisted of the analysis of two sets of standards. The normal linearity system protocol on the Sievers 500 RL consists of a blank, a 250 ppb standard, a 500 ppb standard, and a 750 ppb standard (sucrose). From the measurement of these standards, the analyzer will calculate the square of the Pearson's Correlation Coefficient along with a limit of detection (LOD) based on the intercept of a plot of the standard deviation of the measurements versus concentration. For this study we also prepared standards at higher concentrations (1,000 ppb, 1,500 ppb, 2,000 ppb, and 2,500 ppb). Combined with the lower concentration standards, these higher concentration standards cover the upper range of the TOC analyzer. Figure 9 shows a plot of measured versus expected concentration for the 324 linearity standards analyzed on the six Sievers 500 RL analyzers. The R 2 for this combined data was Figure 10 shows the results from two individual analyzers (best and worst based on R 2 ) Figure 7. Histogram and box plot from 54 measurements of response efficiency using the Sievers 500 RL TOC Analyzer Figure 8. Box Plot of response efficiency from 37 measurements of acidified standards and 16 measurements with non-acidified standards Figure 9. Measured versus Expected TOC for Linearity Standards on Sievers 500 RL Analyzers

7 Page 7 of 10 along with least-squares regression lines for each of the five linearity sets. For S/N 21, the slopes ranged from 0.96 to 1.01 with R 2 for all five regression lines equal to 1.0, while for S/N 18, the slopes ranged from 0.93 to 0.99 with an R 2 for one regression line equal to The linearity protocol also estimates the limit of detection from the standard deviation of the three replicate measurements of each TOC standard. Figure 11 shows the histogram and box plot for the 35 LODs calculation in this study. The average LOD was 2.5 ppb, with a range from 0.12 to 6 ppb. Detection limits calculated from these TOC standards is always higher than LOD's determined from the standard deviation of on-line measurements of low TOC water. To determine if the TOC response was linear over the entire range of the Sievers 500 RL Analyzer, higher concentration standards of KHP were analyzed. Figure 12 shows a plot of measured versus expected for the complete linearity data set (315 measurements of low level standards plus 213 measurements of higher level standards). The response of the six Sievers 500 RL's was linear over the range from 250 to 2500 ppb with an R 2 of Figure 10. Linearity Data for two individual Sievers 500 RL TOC Analyzers. Regression lines for each of the 5 sets of linearity standards are shown. Figure 11. Histogram and Box plot of Limit of Detection calculated from Linearity Protocol for six Sievers 500 RL TOC Analyzers Figure 12. Plot of Measured versus Expected for 528 TOC standards on the Sievers 500 RL Analyzers.

8 Page 8 of 10 Detection Limit of the Sievers 500 RL The limit of detection for TOC measurements using the Sievers 500 RL was estimated from on-line measurements of a low TOC (< 1.5 ppb), low dissolved oxygen water system. Measurements were performed for two weeks and seven periods of stable TOC were identified. To compensate for water system variability during these periods (primarily due to operation of the reverse osmosis system), the average TOC from the five Sievers 500 RL analyzers was calculated for each measurement. This average was then subtracted from the measured TOC from each analyzer and the standard deviation of the adjusted measurements determined. The LOD is calculated as three times the standard deviation of the adjusted TOC measurements. Figure 13 shows a histogram and box plot of the calculated LOD from 20 measurements of the Sievers 500 RL base model. The average LOD was ppb with a standard deviation of 0.01 ppb. Figure 14 shows a box plot of the LOD data by serial number (along with the results from three PPT analyzers). The three 500 RL base model analyzers (with the sample inlet block) showed lower LOD's compared to the standard and Super ios instruments (S/N 24 and 25). There was no difference in the mean LOD and the standard deviation of the LOD measurements between the three base model instruments and the three PPT analyzers. On the basis of these measurements, we estimate the LOD for the Sievers 500 RL to be 0.03 ppb. Conductivity The results from the measurement of the 25 µs/cm HCl standards on the six Sievers 500 RL analyzers are shown in Figure 15. This variability plot shows the range, measurement means, and individual measurements, along with a separate standard deviation plot. Inspection of the data suggests the presence of several outliers. A Grubb's test was performed and indicated that the circled measurements in Figure 15 are outliers and these measurements were removed from the data set. Statistics for the data set are shown in Table 3, and a plot of the distribution of the remaining data is shown in Figure 16, along with limits corresponding to ± 1% accuracy. The data are not normally distributed (Shapiro-Wilk W Test), with the kurtosis indicating the distribution is peaked with a positive skew. Six of the measurements were less than and one measurement was greater than A capability analysis Figure 13. Histogram and box plot of Limit of Detection (3* standard deviation) from adjusted TOC measurements of a low TOC, low-dissolved oxygen water system. Figure 14. Box Plot of Limit of Detection Data by serial number. There is no difference in the mean LOD between the PPT's and the Sievers 500's except for S/N 24. Table 3 Statistics from Accuracy Measurements Statistic Value Mean Std Dev Std Err Mean N 207 Skewness Kurtosis 1.96

9 Page 9 of 10 Figure 15. Variability chart of measurements of 25 µs/cm conductivity standards on six Sievers 500 RL TOC Analyzers. Circled measurements are statistical outliers (Grubb's Test) and were removed from the data set before accuracy and precision calculations. Figure 16. Histogram (with normal distribution) and box plot for 207 measurements of a 25 µs/cm conductivity standard with specification limits at ± 1% Accuracy Figure 17. Histogram and box plot of precision of conductivity measurements (3 replicates per measurement, 207 measurements with upper specification limit at 0.25% using the ± 1% conductivity accuracy indicates that one could expect 3% of measurements of this 25 µs/cm standard would be outside the limits (1.97% below, 1.04% above). The precision was calculated from the mean and standard deviation of the three measurements, and plot of the distribution of the precision is shown in Figure 17 along with a 0.25% precision specification. The mean precision was 0.04% with a range from 0% to 0.1%. A capability analysis from these data indicates that the precision of replicate measurements of this 25 µs/cm standard should always be within this specification. The results from conductivity measurements of deionized water and NaCl conductivity standards from 0.1 to 35 µs/cm are shown in Figure 18. The response of the two 500 RL's was linear across this conductivity range.

10 Page 10 of 10 Plots of the conductivity measurements' accuracy and precision from this data set, along with the results from an in-line conductivity sensor, are shown in Figure 19. Both the accuracy and precision were independent of the concentration of the NaCl standards (R 2 = p=0.11 for S/N 14, R 2 = p=0.43 for S/N 15). The analyzer shows excellent performance for the online measurement of conductivity, with an accuracy of ±1%, a precision for replicate conductivity measurements of < 0.25% and a linear response from µs/cm to 35 µs/cm. Conclusions The results from these tests demonstrate that the Sievers 500 RL TOC Analyzer is capable of measuring total organic carbon in water with an accuracy of ± 5%, with a precision for replicate measurements of < 1% and a linear response over the concentration range from ppb. The analyzers also demonstrated excellent capability for the USP system suitability test. On-line measurements of low-toc, deionized water indicate the detection limit of the analyzers is 0.03 ppb. Figure 18. Linearity of Sievers 500 RL Conductivity Measurements Figure 19. Accuracy and precision for two Sievers 500 RL's and an in-line conductivity sensor over a range of conductivity concentrations (x = Thornton, + = 500 RL S/N 14, = 500 RL S/N 15) * Trademark of General Electric Company; may be registered in one or more countries. For more information, visit Find a sales partner near you through the Contact Us Section. USA GE Analytical Instruments 6060 Spine Road Boulder, CO USA T T F geai@ge.com , General Electric Company. All rights reserved. Europe Unit 3 Mercury Way Urmston, Manchester, M41 7LY United Kingdom T +44 (0) F +44 (0) generaluk.instruments@ge.com Rev A MC06-019

Sievers 500 RL On-Line TOC Analyzer

Sievers 500 RL On-Line TOC Analyzer GE Water & Process Technologies Analytical Instruments Sievers 500 RL On-Line TOC Analyzer Standard ios Model Super ios Model Overview The drive to improve quality and operating efficiencies demands analytical

More information

imagination at work Sievers Certified Reference Materials and Consumables Catalog GE Power & Water Water & Process Technologies

imagination at work Sievers Certified Reference Materials and Consumables Catalog GE Power & Water Water & Process Technologies GE Power & Water Water & Process Technologies Analytical Instruments Sievers Certified Reference Materials and Consumables Catalog imagination at work imagination at work www.geinstruments.com Sievers

More information

TOC600 ANALYZER S TANDARD OPERATING P ROCEDURES

TOC600 ANALYZER S TANDARD OPERATING P ROCEDURES TOC600 ANALYZER S TANDARD OPERATING P ROCEDURES Hach Ultra Analytics 5600 Lindbergh Drive Loveland, Colorado 80538 U.S.A. Voice: 1 970 663-9760 1 800 373 0531 Fax: 1 970 663-9761 Support Hot Line: 1 877

More information

imagination at work Sievers Certified Reference Materials and Consumables Catalog GE Power & Water Water & Process Technologies

imagination at work Sievers Certified Reference Materials and Consumables Catalog GE Power & Water Water & Process Technologies GE Power & Water Water & Process Technologies Analytical Instruments Sievers Certified Reference Materials and Consumables Catalog imagination at work imagination at work www.geinstruments.com Sievers

More information

Standard Operating Procedure Title: Operation and Calibration of Sievers 820 TOC Analyser

Standard Operating Procedure Title: Operation and Calibration of Sievers 820 TOC Analyser Table of Contents 1. 25 ppm IC as Sodium Carbonate - standard preparation procedure 2 2. 25 ppm TOC as KHP standard preparation procedure 3 3. Calibration Procedure 4 4. Performing Calibration - Prepare

More information

The Right Analytical Method for the Right Application: TOC Analysis for Cleaning Validation

The Right Analytical Method for the Right Application: TOC Analysis for Cleaning Validation GE Water & Process Technologies Analytical Instruments Paper The Right Analytical Method for the Right Application: TOC Analysis for Cleaning Validation Advances in technologies and mounting manufacturing

More information

New! Sievers M9 TOC Analyzers

New! Sievers M9 TOC Analyzers Now with Sample Conductivity! New! Sievers M9 TOC Analyzers Reach the peak of productivity Tools that boost productivity are essential to managing your chaotic workday. The new Sievers* M9 Total Organic

More information

QbD1200 TOTAL ORGANIC CARBON ANALYZER RELIABLE MEASUREMENT MADE SIMPLE.

QbD1200 TOTAL ORGANIC CARBON ANALYZER RELIABLE MEASUREMENT MADE SIMPLE. QbD1200 TOTAL ORGANIC CARBON ANALYZER RELIABLE MEASUREMENT MADE SIMPLE. The NEW QbD1200 delivers the highest performance with unparalleled ease of operation and maintenance Reliable data Stop throwing

More information

Lean Strategies for Laboratory Testing. Jenny Watson Global Pharmaceutical Applications Manager GE Analytical Instruments

Lean Strategies for Laboratory Testing. Jenny Watson Global Pharmaceutical Applications Manager GE Analytical Instruments Lean Strategies for Laboratory Testing Jenny Watson Global Pharmaceutical Applications Manager GE Analytical Instruments Agenda Operational Excellence & Lean Case Study The Balancing Act Time Quality Cost

More information

Sievers M9 TOC Analyzers

Sievers M9 TOC Analyzers www.sieversinstruments.com Water Technologies & Solutions Sievers M9 TOC Analyzers * m-power your team to be compliant and efficient dependability and flexibility: Sievers legacy As the world s leading

More information

Application Note AN1501

Application Note AN1501 Application Note AN1501 Ultra-Trace Mercury Determination in Bottled Water, EPA Method 1631, Using the QuickTrace M-8000 CVAFS Introduction Mercury determination in water and foodstuffs is a crucial part

More information

Validation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals.

Validation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals. Validation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals. 1 Analytical Method Validation: 1..1 Philosophy: Method validation is

More information

Functionality. Normal Operation

Functionality. Normal Operation Application Note Automating EPA Method 200.8 using the SDXHPLD High Performance Liquid Dilution System Jim Maximovich, Teledyne CETAC Technologies, Omaha, NE, USA, Jim.Maximovich@Teledyne.com The SDX HPLD

More information

Lab 6 Measurement of Ozone

Lab 6 Measurement of Ozone Georgia Institute of Technology School of Earth and Atmospheric Sciences EAS 4641 Spring 2008 Lab 6 Measurement of Ozone Purpose of Lab 6: In this lab you will measure the ambient concentration of ozone

More information

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement Method 8000 DR/4000 PROCEDURE Reactor Digestion Method* (0 to 150.0 mg/l COD) Scope and Application: For water, wastewater and seawater; digestion is required; USEPA Approved** for wastewater analyses.

More information

Simplicity. Accuracy. Durability. Lotix. Automated High Temperature Combustion TOC Analyzer

Simplicity. Accuracy. Durability. Lotix. Automated High Temperature Combustion TOC Analyzer Simplicity Accuracy Durability Lotix Automated High Temperature Combustion TOC Analyzer The Lotix TOC Combustion analyzer is designed to accurately measure carbon content in aqueous matrices down to the

More information

rat cortex data: all 5 experiments Friday, June 15, :04:07 AM 1

rat cortex data: all 5 experiments Friday, June 15, :04:07 AM 1 rat cortex data: all 5 experiments Friday, June 15, 218 1:4:7 AM 1 Obs experiment stimulated notstimulated difference 1 1 689 657 32 2 1 656 623 33 3 1 668 652 16 4 1 66 654 6 5 1 679 658 21 6 1 663 646

More information

Rev Experiment 10

Rev Experiment 10 Experiment 10 SPECTROPHOTOMETRIC DETERMINATION OF IRON IN DRINKING WATER 2 lab periods Reading: 1) Chapter 17, pg 393-403, Quantitative Chemical Analysis, 8 h Edition, Daniel C. Harris (7 th Edition: Chapter

More information

Corn Syrup Analysis E-1a- 1. ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

Corn Syrup Analysis E-1a- 1. ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) Corn Syrup Analysis E-1a- 1 ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) PRINCIPLE Isovaleraldehyde (IVA) and acetaldehyde are released from the syrup with the aid of dilute phosphoric acid and

More information

STAT 2300: Unit 1 Learning Objectives Spring 2019

STAT 2300: Unit 1 Learning Objectives Spring 2019 STAT 2300: Unit 1 Learning Objectives Spring 2019 Unit tests are written to evaluate student comprehension, acquisition, and synthesis of these skills. The problems listed as Assigned MyStatLab Problems

More information

Evaluation of the Mercury Concentration Accessory for US EPA Methodology

Evaluation of the Mercury Concentration Accessory for US EPA Methodology Evaluation of the Mercury Concentration Accessory for US EPA Methodology Application Note Atomic Absorption Author Lucinda M. Beach Introduction The Agilent Mercury Concentration Accessory (MCA-90) was

More information

5000TOC Sensor Standard Operating Procedure

5000TOC Sensor Standard Operating Procedure Part No. 84460 5000TOC Sensor Standard Operating Procedure for System Suitability Testing to meet USP 643 and EP 2.2.44 This document contains proprietary information, which is protected by copyright.

More information

QAM-I-113 Operation and Calibration of the Dissolved Oxygen Meter

QAM-I-113 Operation and Calibration of the Dissolved Oxygen Meter 1.0 Applicability and Purpose i. This procedure applies to the operation and calibration of the YSI Model 5100 dissolved oxygen meter or equivalent and associated probe. By performing the calibration procedure

More information

5000TOCi with ISM. Fast, Continuous and Reliable Total Organic Carbon Measurement

5000TOCi with ISM. Fast, Continuous and Reliable Total Organic Carbon Measurement 5000TOCi with ISM 5000TOCi Sensor Fast, continuous measurement Reliable operation Reagent and chemical free USP , , Ch P, EP 2.2.44, and JP compliant Compact sensor design ISM diagnostics Fast,

More information

Fusion Analytical Method Validation

Fusion Analytical Method Validation Fusion QbD Software Platform Fusion Analytical Method Validation The Only Software That Has It All! 100% aligned with FDA/ICH Quality by Design (QbD) guidances! Can be used for LC and Non-LC methods (e.g.

More information

VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY *)

VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY *) VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY *) Guideline Title Validation of Analytical Procedures: Methodology Legislative basis Directive 75/318/EEC as amended Date of first adoption December 1996

More information

Fusion UV/Persulfate TOC Analyzer

Fusion UV/Persulfate TOC Analyzer Fusion UV/Persulfate TOC Analyzer The Fusion Total Organic Carbon (TOC) Analyzer utilizes powerful Ultra Violet (UV) Persulfate oxidation allowing superior carbon liberation from even the most challenging

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical Oxygen Demand, Chemical DOC316.53.01101 Manganese III Reactor Digestion Method (with chloride removal) Method 10067 30 to 1000 mg/l COD Mn Test N Tube Vials Scope and application: For water and wastewater

More information

energy usage summary (both house designs) Friday, June 15, :51:26 PM 1

energy usage summary (both house designs) Friday, June 15, :51:26 PM 1 energy usage summary (both house designs) Friday, June 15, 18 02:51:26 PM 1 The UNIVARIATE Procedure type = Basic Statistical Measures Location Variability Mean 13.87143 Std Deviation 2.36364 Median 13.70000

More information

JAI RESEARCH FOUNDATION A Global Contract Research Organisation

JAI RESEARCH FOUNDATION A Global Contract Research Organisation JAI RESEARCH FOUNDATION A Global Contract Research Organisation METHOD FOR THE DETERMINATION OF MANCOZEB RESIDUES IN TEST MEDIA FOR ECOTOX STUDIES Author : S.Y. Pandey and P.K. Dubey Jai Research Foundation

More information

Importance of TOC measurement in WFI in light of European Pharmacopoeia change

Importance of TOC measurement in WFI in light of European Pharmacopoeia change WITE PAPER Importance of TOC measurement in WFI in light of European Pharmacopoeia change Tony arrison, Applications Specialist Beckman Coulter, Inc. Abstract The new revision of the European Pharmacopoeia

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical without chloride removal, 10067 Manganese III Reactor Digestion Method (without chloride removal) 1 30 to 1000 mg/l COD Mn Scope and Application: For water and wastewater 1 U.S. Patent 5,556,787 DOC316.53.01102

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical , TNTplus 823, 10212 DOC316.53.01104 USEPA Reactor Digestion Method Method 10212 UHR (250 15,000 mg/l COD) TNTplus 823 Scope and Application: For wastewater and process waters; digestion is required Test

More information

Microbial Detection. Continuous At-Line Microbial Monitoring For Pharmaceutical Waters

Microbial Detection. Continuous At-Line Microbial Monitoring For Pharmaceutical Waters Microbial Detection 7000RMS Microbial Detection Analyzer Continuous, real-time analysis Results in seconds No sample preparation No reagents required Process/batch sampling modes Continuous At-Line Microbial

More information

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor 229 10. Validated Normal Phase HPLC Method for the Determination of Fulvestrant in Pharmaceutical Dosage Forms 10.1 Introduction Fulvestrant is primarily used in the treatment of hormone receptor positive

More information

Fusion UV/Persulfate TOC Analyzer

Fusion UV/Persulfate TOC Analyzer Fusion UV/Persulfate TOC Analyzer Unparalleled Results with the Fusion UV/Persulfate TOC Analyzer Description The Fusion Total Organic Carbon (TOC) Analyzer utilizes powerful Ultra Violet (UV) Persulfate

More information

Statistical Data Assessment for the Ross Incinerator Site U.S. Environmental Protection Agency Region 8 Work Authorization 013

Statistical Data Assessment for the Ross Incinerator Site U.S. Environmental Protection Agency Region 8 Work Authorization 013 February 9, 4 Mr. Dan Powell U.S. Environmental Protection Agency Office of Superfund Remediation and Technology Innovation (51G) 1 Pennsylvania Avenue NW Washington, DC 46 Subject: Statistical Data Assessment

More information

Lab 7 Measurement of Ozone

Lab 7 Measurement of Ozone Georgia Institute of Technology School of Earth and Atmospheric Sciences EAS 4641 Spring 2007 Lab 7 Measurement of Ozone Purpose of Lab 7: In this lab you will measure the ambient concentration of ozone

More information

RELIABILITY OF TOC RESULTS IN ULTRAPURE WATER ANALYSIS

RELIABILITY OF TOC RESULTS IN ULTRAPURE WATER ANALYSIS Tech Notes: 030101 RELIABILITY OF TOC RESULTS IN ULTRAPURE WATER ANALYSIS The intent of this technical note is to discuss current methods of TOC (Total Organic Carbon) analysis, including their advantages

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical , TNTplus 820, 10211 DOC316.53.01103 USEPA Reactor Digestion Method Method 10211 ULR (1 60 mg/l COD) TNTplus 820 Scope and Application: For wastewater, process water, surface water, and cooling water;

More information

z /zr/r/ lnitials: tv4'wl z"z{-t L 3 /, /tt Operation and Calibration of the Dissolved Oxygen Meter Revision 5 Approval: Laboratory Man ager/lqao/rso

z /zr/r/ lnitials: tv4'wl zz{-t L 3 /, /tt Operation and Calibration of the Dissolved Oxygen Meter Revision 5 Approval: Laboratory Man ager/lqao/rso QAM.I-1 13 Operation and Calibration of the Dissolved Oxygen Meter Revision 5 Approval: Laboratory Man ager/lqao/rso Date z /zr/r/ Date z"z{-t L Effective date: Renewal date: * 1r i l: 3 /, /tt lnitials:

More information

Portable Multi-Parameter Water Quality Meter LMPWM-A1 Series

Portable Multi-Parameter Water Quality Meter LMPWM-A1 Series Portable MultiParameter Water Quality Meter LMPWMA1 Series www.labtron.com info@labtron.com Portable Multiparameter Water Quality Meter LMPWMA1 series Portable Multiparameter Water Quality Meter LMPWMA1

More information

Orono Spectral Solutions, Inc.

Orono Spectral Solutions, Inc. Page 1 of 14 Orono Spectral Solutions, Inc. STANDARD OPERATING PROCEDURE for ASTM Method DXXXX, Hydrocarbon Contamination in Urea Liquor REVISION # 1 Effective date: April 5, 2012 APPROVED BY Signature

More information

NPOC, VOC, TC-IC and TDN using Shimadzu TOC-V Protocols. Water Quality Analysis Laboratory at the University of New Hampshire

NPOC, VOC, TC-IC and TDN using Shimadzu TOC-V Protocols. Water Quality Analysis Laboratory at the University of New Hampshire NPOC, VOC, TC-IC and TDN using Shimadzu TOC-V Protocols Water Quality Analysis Laboratory at the University of New Hampshire Prepared by: Jeffrey Merriam Date of Last Revision: 10/26/2004 Signature of

More information

Validation of On-line Total Organic Carbon Analysers for Release Testing Using ICH Q2

Validation of On-line Total Organic Carbon Analysers for Release Testing Using ICH Q2 WITEPAPER Validation of On-line Total Organic arbon Analysers for Release Testing Using I Q2 Introduction As the pharmaceutical industry seeks to reduce the number and associated costs of quality control

More information

GOOD LABORATORY PRACTICE Agilent Cary 8454 UV-Visible Spectroscopy System

GOOD LABORATORY PRACTICE Agilent Cary 8454 UV-Visible Spectroscopy System GOOD LABORATORY PRACTICE Agilent Cary 8454 UV-Visible Spectroscopy System Introduction The objective of good laboratory practice (GLP) is to obtain accurate and precise results, using a measurement system

More information

Fusion Analytical Method Validation

Fusion Analytical Method Validation Fusion QbD Software System Fusion Analytical Method Validation The Only Software That Has It All! 100% aligned with FDA/ICH Quality by Design (QbD) guidances! Can be used for LC and Non-LC methods (e.g.

More information

VICH Topic GL2 (Validation: Methodology) GUIDELINE ON VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY

VICH Topic GL2 (Validation: Methodology) GUIDELINE ON VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY The European Agency for the Evaluation of Medicinal Products Veterinary Medicines Evaluation Unit CVMP/VICH/591/98-FINAL London, 10 December 1998 VICH Topic GL2 (Validation: Methodology) Step 7 Consensus

More information

Automating EPA 6020 Compliant Analysis with the Agilent 7900 ICP-MS and ESI prepfast Autodilution System

Automating EPA 6020 Compliant Analysis with the Agilent 7900 ICP-MS and ESI prepfast Autodilution System Automating EPA 6020 Compliant Analysis with the Agilent 7900 ICP-MS and ESI prepfast Autodilution System Application Note Environmental Authors Austin Schultz and Jake Unnerstall, Elemental Scientific,

More information

CORESTA RECOMMENDED METHOD N 61

CORESTA RECOMMENDED METHOD N 61 CORESTA RECOMMENDED METHOD N 61 DETERMINATION OF 1,2-PROPYLENE GLYCOL, GLYCEROL AND SORBITOL IN TOBACCO AND TOBACCO PRODUCTS BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC) (July 2015) 0. INTRODUCTION

More information

Method Development and Validation for Online UV-Dissolution Methods Using Fiber-Optic Technology

Method Development and Validation for Online UV-Dissolution Methods Using Fiber-Optic Technology Technical Overview Method Development and Validation for Online UV-Dissolution Methods Using Fiber-Optic Technology Introduction Online fiber-optic and multicell UV-dissolution systems have become increasingly

More information

Validation of a concentration assay using Biacore C

Validation of a concentration assay using Biacore C GE Healthcare Application Note 48 Biacore systems Validation of a concentration assay using Biacore C Guideline for development of a GxP - compliant concentration assay Support for informed decision-making

More information

Best Practices for Filtration of Dissolved Metals

Best Practices for Filtration of Dissolved Metals Best Practices for Filtration of Dissolved Metals Tim Crowther, Regional Client Services Manager ALS Environmental, Canada Right Solutions Right Partner www.alsglobal.com 1 Right Solutions Right Partner

More information

Materials Required for Lab Report Resubmissions: Original graded report with all required components

Materials Required for Lab Report Resubmissions: Original graded report with all required components Appendix III: What to Turn in for Each Lab and Resubmission Policies Lab Report Guidelines All lab reports are due in lecture. Graded reports will be handed back in the laboratory section. If you miss

More information

SOP-C-102 Determination of Chemical Oxygen Demand

SOP-C-102 Determination of Chemical Oxygen Demand i. Identification of the method a. Hach 8000, Chemical Oxygen Demand, with additions to meet project requirements ii. Applicable matrix or matrices a. Nonpotable water and aqueous solutions iii. iv. Limits

More information

Using Empower SystemsQT Qualification Tool

Using Empower SystemsQT Qualification Tool Using Empower SystemsQT Qualification Tool for Waters Modular HPLC Systems A summary of Empower SystemsQT Qualification Tool features and benefits, with emphasis on the technical attributes of the qualification

More information

Validation Using SDS Software Version on the Applied Biosystems 7500 Real-Time PCR System and the ABI PRISM 7000 Sequence Detection System

Validation Using SDS Software Version on the Applied Biosystems 7500 Real-Time PCR System and the ABI PRISM 7000 Sequence Detection System User Bulletin Quantifiler Kits April 2006 SUBJECT: Validation Using SDS Software Version 1.2.3 on the Applied Biosystems 7500 Real-Time PCR System and the ABI PRISM 7000 Sequence Detection System Note:

More information

Super-marketing. A Data Investigation. A note to teachers:

Super-marketing. A Data Investigation. A note to teachers: Super-marketing A Data Investigation A note to teachers: This is a simple data investigation requiring interpretation of data, completion of stem and leaf plots, generation of box plots and analysis of

More information

Understanding and accounting for product

Understanding and accounting for product Understanding and Modeling Product and Process Variation Variation understanding and modeling is a core component of modern drug development. Understanding and accounting for product and process variation

More information

Validation of Sulfur Dioxide Using. SKC UME x 200 Passive Sampler Cat. No Page 1 of 14 Publication 1781 Rev 1607

Validation of Sulfur Dioxide Using. SKC UME x 200 Passive Sampler Cat. No Page 1 of 14 Publication 1781 Rev 1607 Validation of Sulfur Dioxide Using SKC UME x 200 Passive Sampler Cat. No. 500-200 Page 1 of 14 Publication 1781 Rev 1607 Research Report Validation of Sulfur Dioxide Using SKC UME x 200 Passive Sampler

More information

CHAPTER 4. Labeling Methods for Identifying Outliers

CHAPTER 4. Labeling Methods for Identifying Outliers CHAPTER 4 Labeling Methods for Identifying Outliers 4.1 Introduction Data mining is the extraction of hidden predictive knowledge s from large databases. Outlier detection is one of the powerful techniques

More information

ACETALDEHYDE. The procedure is applicable to corn syrups including those containing fructose, and to most other starch hydrolyzates (Note 1).

ACETALDEHYDE. The procedure is applicable to corn syrups including those containing fructose, and to most other starch hydrolyzates (Note 1). ACETA.01-1 ACETALDEHYDE PRINCIPLE SCOPE Acetaldehyde is released from the syrup by treatment with dilute phosphoric acid and heat. The liberated acetaldehyde is then measured using a head space gas chromatographic

More information

Test Method Validation of In-Vitro Diagnostic Products. Tim Carr QA Manager, Beckman Coulter

Test Method Validation of In-Vitro Diagnostic Products. Tim Carr QA Manager, Beckman Coulter Test Method Validation of In-Vitro Diagnostic Products Tim Carr QA Manager, Beckman Coulter IVD Overview What is an In-Vitro Diagnostic Test IVDs are medical devices and accessories Used to perform tests

More information

Cleaning validation of cleanrooms and preparation equipments

Cleaning validation of cleanrooms and preparation equipments Cleaning validation of cleanrooms and preparation equipments Dr Farshid SADEGHIPOUR Head of production Central Pharmacy, Geneva University Hospitals EAHP Foundation Seminar: Patient Safety; More About

More information

Fast, Continuous and Reliable Total Organic Carbon Measurement

Fast, Continuous and Reliable Total Organic Carbon Measurement 5000TOCi Sensor Fast, continuous measurement Reliable operation Reagent and chemical free USP , , Ch P, EP 2.2.44, and JP compliant Compact sensor design ISM diagnostics Fast, Continuous and

More information

Trace Metals in Waters by GFAAS, in Accordance with U.S. EPA and Health Canada Requirements

Trace Metals in Waters by GFAAS, in Accordance with U.S. EPA and Health Canada Requirements application Note Atomic Absorption Author: Randy L. Hergenreder PerkinElmer, Inc. Shelton, CT 06484 USA Trace Metals in Waters by GFAAS, in Accordance with U.S. EPA and Health Canada Requirements Introduction

More information

Supplemental Information Development of a particulate mass measurement system for quantification of ambient reactive mercury

Supplemental Information Development of a particulate mass measurement system for quantification of ambient reactive mercury Supplemental Information Development of a particulate mass measurement system for quantification of ambient reactive mercury Ashley M. Pierce*, Mae Sexauer Gustin Department of Natural Resources and Environmental

More information

Process for comparing certified values of the same measurand in multiple reference materials (CRMs)

Process for comparing certified values of the same measurand in multiple reference materials (CRMs) Process for comparing certified values of the same measurand in multiple 1. Purpose This document describes JCTLM process which can be followed by the producers of Certified Reference Materials (CRMs)

More information

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Student Learning Objectives High Performance Liquid Chromatography With the advent of relatively inexpensive and reliable pumps, the development of

More information

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By the Authority Vested By Part 5 of the United States Code 552(a) and Part 1 of the Code of Regulations 51 the attached document has

More information

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement, 0 to 40 mg/l COD

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement, 0 to 40 mg/l COD Method 8000 DR/4000 PROCEDURE Reactor Digestion Method* (0 to 40.0** mg/l COD) Scope and Application: For water, wastewater and seawater; digestion is required. The estimated detection limit for program

More information

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters

More information

INGOLD. Leading Process Analytics THORNTON. MT Service. Leading Pure Water Analytics. A Partner You Can Trust to Get It Right from the Start

INGOLD. Leading Process Analytics THORNTON. MT Service. Leading Pure Water Analytics. A Partner You Can Trust to Get It Right from the Start MT Service INGOLD Leading Process Analytics THORNTON Leading Pure Water Analytics A Partner You Can Trust to Get It Right from the Start Contents Page Introduction 3 Bioburden 7000RMS Analyzer 4 Conductivity

More information

Outliers and Their Effect on Distribution Assessment

Outliers and Their Effect on Distribution Assessment Outliers and Their Effect on Distribution Assessment Larry Bartkus September 2016 Topics of Discussion What is an Outlier? (Definition) How can we use outliers Analysis of Outlying Observation The Standards

More information

Torch Combustion TOC Analyzer

Torch Combustion TOC Analyzer Torch Combustion TOC Analyzer The Torch Combustion TOC Analyzer utilizes patent pending Static Pressure Concentration (SPC) for the analysis of TOC using high temperature combustion with an optional TN

More information

FUNDAMENTALS OF QUALITY CONTROL AND IMPROVEMENT. Fourth Edition. AMITAVA MITRA Auburn University College of Business Auburn, Alabama.

FUNDAMENTALS OF QUALITY CONTROL AND IMPROVEMENT. Fourth Edition. AMITAVA MITRA Auburn University College of Business Auburn, Alabama. FUNDAMENTALS OF QUALITY CONTROL AND IMPROVEMENT Fourth Edition AMITAVA MITRA Auburn University College of Business Auburn, Alabama WlLEY CONTENTS PREFACE ABOUT THE COMPANION WEBSITE PART I PHILOSOPHY AND

More information

Analysis of wear metals and contaminants in engine oils using the 4100 MP-AES

Analysis of wear metals and contaminants in engine oils using the 4100 MP-AES Analysis of wear metals and contaminants in engine oils using the 4100 MP-AES Application note Energy and fuels Authors Phil Lowenstern and Elizabeth Reisman Agilent Technologies Melbourne, Australia Introduction

More information

Trace Moisture Contamination in Ultra-High Purity Phosphine: Techniques for Measurement and Control

Trace Moisture Contamination in Ultra-High Purity Phosphine: Techniques for Measurement and Control Trace Moisture Contamination in Ultra-High Purity Phosphine: Techniques for Measurement and Control M. Raynor, H. Funke, J. Yao, T. Watanabe and R. Torres, Matheson Tri-Gas, Advanced Technology Center,

More information

oil standards TRUE NORTH AND ATLANTIS

oil standards TRUE NORTH AND ATLANTIS oil standards TRUE NORTH AND ATLANTIS Accuracy Matters Accurate lab data is critical to sound decision making. Running certified oil standards is the ideal way to ensure the accuracy of your DGA results.

More information

Multi-element determination in pharmaceutical preparations using the Thermo Scientific icap Q ICP-MS

Multi-element determination in pharmaceutical preparations using the Thermo Scientific icap Q ICP-MS Multi-element determination in pharmaceutical preparations using the Thermo Scientific icap Q ICP-MS Dr. Simon Nelms, Thermo Fisher Scientific, UK Application Note 43100 Key Words icap Q, Pharmaceutical

More information

Determination of Manganese in Brine ( 30% NaCl)

Determination of Manganese in Brine ( 30% NaCl) Application Update 8 Determination of Manganese in Brine ( 0% NaCl INTRODUCTION Sodium chloride brines are used in chlor-alkali cells and must have low concentrations of alkali and alkaline earth metals

More information

Evaluating Microplate Detection Instruments

Evaluating Microplate Detection Instruments AppNote Evaluating Microplate Detection Instruments Introduction Important features to consider when evaluating a microplate reader are the following: sensitivity, dynamic range, throughput, integration

More information

APPENDIX III SAMPLE LAB REPORT. Experiment 1. High Performance Liquid Chromatography. Alfred E. Neuman

APPENDIX III SAMPLE LAB REPORT. Experiment 1. High Performance Liquid Chromatography. Alfred E. Neuman APPENDIX III SAMPLE LAB REPORT Experiment 1 High Performance Liquid Chromatography Alfred E. Neuman Chemistry 436 September 15, 1999 INTRODUCTION In this experiment, high performance liquid chromatography

More information

Moisture in Materials work at NPL - the first steps Presentation to TMAN meeting at NPL

Moisture in Materials work at NPL - the first steps Presentation to TMAN meeting at NPL Moisture in Materials work at NPL - the first steps Presentation to TMAN meeting at NPL Paul Carroll 4 th February 2009 Overview : Study completed : findings / recommendations Gaining expertise develop

More information

IMPROVE Carbon Analysis

IMPROVE Carbon Analysis IMPROVE Carbon Analysis Judith C. Chow (judith.chow@dri.edu) John G. Watson Jerome A. Robles Xiaoliang Wang Dana L. Trimble Desert Research Institute, Reno, NV Presented at the IMPROVE Steering Committee

More information

Utility of Variable Bandwidth Monochromators for Quantification of Fluorescent Probes in Produced Effluent Water

Utility of Variable Bandwidth Monochromators for Quantification of Fluorescent Probes in Produced Effluent Water A p p l i c a t i o n N o t e Utility of Variable Bandwidth Monochromators for Quantification of Fluorescent Probes in Produced Effluent Water Brad Larson and Peter Banks, BioTek Instruments, Inc., Winooski,

More information

Estimation of Relative Potency from Bioassay Data that Include Values below the Limit of Quantitation

Estimation of Relative Potency from Bioassay Data that Include Values below the Limit of Quantitation Estimation of Relative Potency from Bioassay Data that Include Values below the Limit of Quantitation FRANCIS BURSA, KELLY J. FLEETWOOD, KARIE J. HIRST, AND ANN YELLOWLEES Experimental data may include

More information

ENVIRONMENTAL PROTECTION AGENCY APTI 435: Atmospheric Sampling Course. Student Manual

ENVIRONMENTAL PROTECTION AGENCY APTI 435: Atmospheric Sampling Course. Student Manual ENVIRONMENTAL PROTECTION AGENCY APTI 435: Atmospheric Sampling Course Student Manual Table of Contents Chapter 1: An Introduction to Atmospheric Sampling... 1-1 1.1 Introduction... 1-2 1.2 Sampling Train

More information

reverse transcription! RT 1! RT 2! RT 3!

reverse transcription! RT 1! RT 2! RT 3! Supplementary Figure! Entire workflow repeated 3 times! mirna! stock! -fold dilution series! to yield - copies! per µl in PCR reaction! reverse transcription! RT! pre-pcr! mastermix!!!! 3!!! 3! ddpcr!

More information

Investigating the State of Acidification in the Southern California Bight

Investigating the State of Acidification in the Southern California Bight Investigating the State of Acidification in the Southern California Bight Karen McLaughlin SCCWRP Symposium February 26, 2015 What is Ocean Acidification? Ocean Acidification is the other CO 2 problem

More information

Simplicity. Accuracy. Durability. Lotix. Automated High Temperature Combustion TOC Analyzer

Simplicity. Accuracy. Durability. Lotix. Automated High Temperature Combustion TOC Analyzer Simplicity Accuracy Durability Lotix Automated High Temperature Combustion TOC Analyzer The Lotix TOC Combustion analyzer is designed to accurately measure carbon content in aqueous matrices down to the

More information

CLSI C60: Assay Validation & Post-Validation Monitoring

CLSI C60: Assay Validation & Post-Validation Monitoring CLSI C60: Assay Validation & Post-Validation Monitoring Ross J. Molinaro, MT(ASCP), PhD, DABCC, FACB Medical Director Core Laboratory, Emory University Hospital Midtown Emory Clinical Translational Research

More information

Elemental Scientific. prepfast. The Next Leap in Automation. Autodilution Autocalibration Intelligent QC Dilution

Elemental Scientific. prepfast. The Next Leap in Automation. Autodilution Autocalibration Intelligent QC Dilution Elemental Scientific prepfast The Next Leap in Automation Autodilution Autocalibration Intelligent QC Dilution I C P I C P M S A A 1 prepfast Inline prepfast Automation prepfast is an autodilution system

More information

Onsite quantitative FTIR analysis of water in turbine oil

Onsite quantitative FTIR analysis of water in turbine oil Onsite quantitative FTIR analysis of water in turbine oil Application Note Author Frank Higgins Agilent Technologies, Connecticut, USA Introduction The availability of the Agilent 5500t FTIR spectrometers,

More information

Uranium Testing and On Site Calibration for Water Testing Detection Range: 2 60ppb

Uranium Testing and On Site Calibration for Water Testing Detection Range: 2 60ppb Document: AND010 08 2017 Uranium Testing and On Site Calibration for Water Testing Detection Range: 2 60ppb August 2017 Edition 2 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Uranium Testing

More information

March 2011 Instrument Assessment Report. Hygiena International Ltd. Campden BRI. Assessment of a novel ATP monitoring device.

March 2011 Instrument Assessment Report. Hygiena International Ltd. Campden BRI. Assessment of a novel ATP monitoring device. March 2 Instrument Assessment Report Campden BRI food and drink innovation Hygiena International Ltd Assessment of a novel ATP monitoring device Executive summary The Hygiena EnSURE luminometer was very

More information

Attachment 1. Categorical Summary of BMP Performance Data for Solids (TSS, TDS, and Turbidity) Contained in the International Stormwater BMP Database

Attachment 1. Categorical Summary of BMP Performance Data for Solids (TSS, TDS, and Turbidity) Contained in the International Stormwater BMP Database Attachment 1 Categorical Summary of BMP Performance Data for Solids (TSS, TDS, and Turbidity) Contained in the International Stormwater BMP Database Prepared by Geosyntec Consultants, Inc. Wright Water

More information

Lead & Copper Water Quality Parameters

Lead & Copper Water Quality Parameters Lead & Copper Water Quality Parameters Routine Quality Control Requirements 141.89 Analytical methods. a) Analyses for lead, copper, ph, conductivity, calcium, alkalinity, orthophosphate, silica, and temperature

More information

INGOLD. Leading Process Analytics THORNTON. MT Service. Leading Pure Water Analytics. A Partner You Can Trust to Get It Right from the Start

INGOLD. Leading Process Analytics THORNTON. MT Service. Leading Pure Water Analytics. A Partner You Can Trust to Get It Right from the Start MT Service INGOLD Leading Process Analytics THORNTON Leading Pure Water Analytics A Partner You Can Trust to Get It Right from the Start 2MT Service Conductivity Sensor Calibrations Maintain Accuracy and

More information

Phosphorus, Reactive (Orthophosphate)

Phosphorus, Reactive (Orthophosphate) Phosphorus, Reactive (Orthophosphate) DOC316.53.01118 USEPA 1, 2 PhosVer 3 Method 3 Method 8048 0.06 to 5.00 mg/l PO 4 (0.02 to 1.60 mg/l P) Test N Tube Vials Scope and application: For drinking water,

More information