Quick Manual to the Waters UPLC System
|
|
- Samuel Barker
- 6 years ago
- Views:
Transcription
1 1 Quick Manual to the Waters UPLC System 1. Application of Chiral HPLC Columns Warning! The HPLC columns are destroyed if the pressure overrides 2000 psi. The highest pressure applied for routine analysis is 1400 psi. Before starting any operation check that the high pressure limit is set to 1400 psi. 1.1 Prime pumps and wash needle Prime the pumps from the Binary Solvent Manager under the Control menu. Prime and wash the needle before starting the run. This option can be found in the Sample Manager under the control menu Starting the analysis Open the Empower Pro and click on the Run Sample button. Specify the project and a new window (Run Sample) appears. Click on the Acquity Sample Manager symbol. In this new window, all system parameters are visualized (this window is recommended to be open under the entire analysis). The most important information involves the pressure profile of the pumps (pumps A and B). The window also shows Maximum pressure, the Minimum pressure and their difference Delta. The Delta value should be less than 10 psi for a certain solvent composition and flow rate. Be observant of any inconsistencies in the pressure and pump plots. Go back to the Run Sample window and switch on the lamp. This can be done in the Acquity PDA detector panel. When you click on the Lamp button a question appears. Choose Yes to switch on the lamp.
2 2 1.3 Changing the eluent composition and the flow-rate The Run sample window shows the eluent composition. Eluent A is usually isopropanol, while eluent B is usually hexane (at least for straight phase chiral HPLC analysis). Click on any of the components and a new window appears. Specify the composition of eluents A and B directly in % units. Usually the highest amount of isopropanol is 50% and the lowest amount is 0.2%. The maximum flow-rate is 1.4 ml/min and the lowest is 0.2 ml/min for routine analysis. Always change the eluent gradually. For example, changing 1% isopropanol/99% hexane to 30% isopropanol/70% hexane can be done in three steps: 10% isopropanol/90% hexane, wait 5 min; 20% isopropanol/80% hexane, wait 5 min and then 30% isopropanol/70% hexane. Apply the same caution when changing the flow rate. Notice that the viscosity of isopropanol is higher than that of hexane so increasing the amount of isopropanol leads to increased pressure with a constant flow-rate. When you increase the amount of isopropanol monitor the change of pressure, it cannot exceed 1400 psi! 1.4 Monitoring the base line After adjusting the eluent and the flow-rate you need to monitor the base line. Choose one of the existing instrument methods in the Instrument methods panel. For example take johannaluke, which can be used for the IC/OD-H columns (excellent for separation of esters, alcohols and sulfonylimines) Click on the Edit button of the Instrument methods panel. A new window opens. Click on the BSM (Binary solvent manager) button. Give the same flow-rate and eluent composition that you have adjusted above or define a gradient system. Choose the column by setting the switches according to the column position you intend to use. Save this method in the main menu and quit. You will return to the Run sample window. Click the monitor button. This starts the monitoring program. The Run button lights up on the menu panel. Check the base line and check the actual pressure, eluent composition and the flow-rate. The delta value has to be less than 10 psi after 5 min monitoring. If you are satisfied
3 3 with the base line, you have to stop the run by clicking the red-sphere button in the main-menu line. 1.5 Sample preparation Dissolve your silica chromatography (or crystallization) purified sample in a mixture of hexane and isopropanol. Use the same composition as in the eluent that you intend to use. Avoid using isopropanol, if you have ester groups in your analyte. You may improve the solubility by adding small amounts (less then 50%) of CHCl 3 to the hexane solution. Fill the sample vial to the upper mark (about 1 ml) and place the sample in the sample tray. Write up the tray position. 1.6 Injection of a sample After you have stopped running the instrument method monitoring, check the injection panel. In the Method Set exactly the same method has to appear as in the Instrument methods. For example you may use johannaluke method. Change the sample tray position (plate/well) and file name. Then inject your sample by clicking the injection button (left side). After the injection check the actual pressure, eluent composition and the flow-rate. If the eluent composition or the flow-rate suddenly changes, then you do not use the same Instrument method and Method Set. Wait until the analyte peak appears in the chromatogram, and stop the run only after this. Check that Instrument method and Method Set agrees and then start the next injection. 1.7 Review of the spectrum You have the possibility to review the spectrum under (or after) the analysis. Right-click in the spectrum window and choose Review. It may take up to 30 seconds to get the review, wait patiently Extracting the UV spectrum. When the Review window appears, click the 3Dchanels button in the bottom of the screen. Mark the entry in the appearing Table. Then
4 4 the UV spectral distribution appears in one of the windows. Right click in the sample spectrum and choose Extract XX nm. Then a cursor appears at the right side, which shows the actual wavelength of the analysis. You may draw the cursor to the required wavelength. In the spectrum window, the actual spectrum recorded at the adjusted wavelength will appear Overlaying the spectra of the enantiomers. Right-click the apex of the peak from your sample and choose the Extract spectrum at XX min menu. A new window appears with the UV spectrum of this peak. Note that the enantiomers of a certain compound give exactly the same UV spectrum. This can be used for identification of the enantiomers. Right-click on one of the two UV spectrums (which are believed to belong to the enantiomers of the same substance). Choose the Properties menu. A new window appears. Choose the overlay menu and click the Overlay in a single plot button, then choose the Scaling menu and choose the Normalization X and Normalization Y. This gives you the normalized overlays of the two UV spectra, which have to be completely identical for the enantiomers Getting integrals. On the main menu choose the Process button and click the Integrate, then choose the Peaks button at the bottom of the window. This gives the integrals of all detected peaks in a table form. Choose the rows of the appearing Table reporting uninteresting integrals (solvent peaks, impurities, etc) and let the rows of the desired peaks be unmarked. Right-click and choose remove. After the retention time column, a number (about 10) of unimportant columns appear. Mark them with the cursor, right-click and choose hide columns. After the areas and area% columns again a number of superfluous columns appear. Mark them and hide them. Then you need to print the table of integrals by right-click and Print, the spectrum (right-click and Print ) and the overlayed UV spectra (right-click and Print ). Do not forget to write up the solvent composition, the flow-rate, the filename and the date of the analysis.
5 5 1.8 Going to standby mode. Choose the Run Samples window. Decrease the flow rate to 0.01 ml/min and the eluent composition to 1% isopropanol/99% hexane. Follow the change of the pressure and wait until Delta is under 10 psi. Then, switch off the UV-lamp on the Lamp panel. The UV lamp has a finite lifetime and therefore it always has to be switched off after the analysis. You also have to check that the green light in the upper module of the instrument over Lamp is not on.
Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany
New Features of the Agilent Method Scouting Wizard for Automated Method Development of Complex Samples Analysis of Large Data Sets by Method Scouting Reports and Automated Adjustment of Flow Rates and
More informationOptimizing the Purification of a Standard Chiral Compound Utilizing a Benchtop, Multi-Purpose, Semi-Preparative to Preparative HPLC System
Optimizing the Purification of a Chiral Compound Utilizing a Benchtop, Multi-Purpose, Semi-Preparative to Preparative HPLC System Application Note PHA0111 Keywords Chiral, Enantiomer, HPLC, Normal Phase
More informationDCMS Link 2.10 for Compass / HyStar
DCMS Link 2.10 for Compass / HyStar DCMS Link Panel (Chromeleon Xpress) Window Compass / HyStar Window DCMS Link Event Log Window Switch between the Windows using the Taskbar Document version: 1.4 March
More informationFast Preparative Column Liquid Chromatography (PCLC)
Fast Preparative Column Liquid Chromatography (PCLC) Application Note 224 Joan Stevens, Ph.D. and Gary Scharrer Introduction Preparative HPLC is recognized as a prime method for obtaining pure compounds
More informationOverview of preparative HPLC. Analytical Technologies Limited
Overview of preparative HPLC Analytical Technologies Limited 1.Review of liquid phase chromatography origin develop prospect Mobile phase Tewett 1903 Stationa ry phase origin Separation result Liquid-stationary
More informationBÜCHI Labortechnik AG
BÜCHI Labortechnik AG Purification BUCHI Chromatography History 1977: BUCHI Fraction Collector 1978: 1 st preparative chromatography system 2008: automated chromatography system 2014: automated high performance
More informationDeveloping Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus
Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus Robert Birdsall, Thomas Wheat, and Weibin Chen Waters Corporation,
More informationAgilent 6500 Series Q-TOF LC/MS System
Agilent 6500 Series Q-TOF LC/MS System Quick Start Guide What s New in Data Acquisition 3 Where to find information 7 Getting Started 9 Step 1. Start the Data Acquisition software 10 Step 2. Prepare the
More informationFlexibility with. Preparative HPLC
Flexibility with Preparative HPLC Date: August 26 th, 2014 Time: 09.00 10.00 am CEST 1 Who are your Presenters? Dr. Christian Textor Product Manager Preparative HPLC Dipl. Biol. Hagen Schlicke Product
More informationChem 321 Lecture 23 - Liquid Chromatography 11/19/13
Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Student Learning Objectives High Performance Liquid Chromatography With the advent of relatively inexpensive and reliable pumps, the development of
More informationMethod Development Considerations for Reversed-Phase Protein Separations
Method Development Considerations for Reversed-Phase Protein Separations Hillary B. Hewitson, Thomas E. Wheat, Paula Hong, Kenneth J. Fountain APPLICATION BENEFITS n The BEH00 C 4 chemistry is available
More informationControlling Agilent 1200 Series Rapid Resolution LC systems through Waters Empower chromatography data software. Technical Overview.
Controlling Agilent 1200 Series Rapid Resolution LC systems through Waters Empower chromatography data software Technical Overview Abstract The Waters Empower 2 chromatography data software with ICS 1.05
More informationPurification Workflow Management
Ronan Cleary and Paul Lefebvre Waters Corporation, Milford, MA, USA INTRODUCTION A standard requirement for drug discovery screening of synthetic libraries is that the test compounds must have a minimum
More informationApplication of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study
Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study Application note Biotherapeutics and Biologics Authors Suresh Babu C.V., Anne
More informationFlash Chromatography & Preparative HPLC The right solution for every separation problem
The right solution for every separation problem Special Promotion until 30/11/2015 The right solution for every separation problem Every user has his own specific requirements in preparative chromatography:
More informationFraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System
Application Note Biologics & Biosimilars Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction Chromatography Agilent 126 Infinity II Bio-Inert System 7 6 5 4 5. 7.5
More informationREVIEW ON FLASH CHROMATOGRAPHY
International Journal of Pharmacy Review & Research www.ijprr.com REVIEW ON FLASH CHROMATOGRAPHY Dane Ganesh D *, Raka KC, Honde BS, Bhawal GS, Tajane PJ Department of Pharmaceutical Chemistry, S.V.N.H.T
More informationProtein-Pak Hi Res HIC Column and HIC Protein Standard
Protein-Pak Hi Res HIC Column and HIC Protein Standard CONTENTS I. INTRODUCTION II. a. Mobile Phase b. Flow Direction CONNECTING COLUMN TO LC SYSTEM I. INTRODUCTION This offering contains non-porous, polymethacrylate-based
More informationQualification of High-Performance Liquid Chromatography Systems
Qualification of High-Performance Liquid Chromatography Systems L. Huber Hewlett- Packard GmbH, Waldbronn December 1998 Manuscript of an article published in BioPharm, Vol 11 Number 11, November 1998,
More informationHillary B. Hewitson, Thomas E. Wheat, Diane M. Diehl INTRODUCTION
Enhanc em ent o f t h e U P L C Am ino Ac i d Ana lysis So lu t io n w it h F l e x ibl e Detector Options Hillary B. Hewitson, Thomas E. Wheat, Diane M. Diehl INTRODUCTION The measurement of amino acids
More informationPreparative HPLC is still the
B I O P R O C E S S TECHNICAL Optimizing Sample Load Capacity and Separation Through a Series of Short Prep Columns Mark Crawford, Joan Stevens, and Luke Roenneburg Preparative HPLC is still the dominate
More informationUFFLC of the Proteome Ultra High Pressure and Ultra Fast Flow Liquid Chromatography: A New Approach To Proteomics
Intact Proteins Mixed bed Ion exchange Fractions MLDI TOF/TOF Confirmation of Selected peaks Tryptic Digestion SCX This step could be eliminated due to pre-separations of Intact proteins RPC Informatics
More informationFundamentals and Techniques of Preparative HPLC. Parto Zist Behboud Tel: 42108
Fundamentals and Techniques of Preparative HPLC Parto Zist Behboud www.partoz.com Tel: 42108 Today s Overview 1 Basic know-how for up-scaling 2 Practical examples of up-scaling 3 Recycle preparation 4
More informationACQUITY UPC 2 BEH, CSH, and HSS Columns
ACQUITY UPC 2 BEH, CSH, and HSS Columns CONTENTS I. INTRODUCTION II. GETTING STARTED a. Safety Considerations b. Column Connectors and Installation c. ecord Installation d. Column Equilibration e. Initial
More informationSFC Säulen für analytische und preparative Anwendungen
SFC Säulen für analytische und preparative Anwendungen Technology Symposium Vienna, 25 th October 2016 DI Verena Schmid 2016 Waters Corporation 1 CHIRAL SEPARATIONS ACQUITY UPC 2 Trefoil 2016 Waters Corporation
More informationPAHs in Surface Water by PDA and Fluorescence Detection
A P P L I C AT I O N N O T E Liquid Chromatography Authors: Catharine Layton Wilhad M. Reuter PerkinElmer, Inc. Shelton, CT PAHs in Surface Water by PDA and Fluorescence Detection Introduction Heightened
More informationAgilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility
Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility Technical Overview Introduction Agilent AdvanceBio SEC columns are a new family of size exclusion chromatography (SEC) columns
More informationImproved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental
Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental Samples Michael S.Young, Jeremy C. Shia, Kim vantran, Kevin M. Jenkins and Masayo Yabo Waters Corporation 34 Maple Street,
More information2023 Reduction of D-(+)-camphor with lithium aluminium hydride to an isomeric mixture of (+)-borneol and ( )-isoborneol
223 Reduction of D-(+)-camphor with lithium aluminium hydride to an isomeric mixture of (+)-borneol and ( )-isoborneol LiAlH 4 tert-butyl-methyl-ether H + OH O OH H C 1 H 16 O (152.2) LiAlH 4 (38.) a C
More informationHeart-cut 2D-LC/MS approach for pharmaceutical impurity identification using an Agilent 6540 Q-TOF LC/MS System
Heart-cut 2D-LC/M approach for pharmaceutical impurity identification using an Agilent 640 Q-TF LC/M ystem Application ote Pharmaceuticals Author iji Joseph Agilent Technologies, Inc. Bangalore, India
More informationUpdate to USP <621> Chromatography: What does this mean? USP 37 NF 32 1S Current as of August 2014
Update to USP Chromatography: What does this mean? USP 37 NF 32 1S Current as of August 2014 2015 Waters Corporation 1 What is the USP-NF? The United States Pharmacopeia National Formulary (USP-NF)
More informationINTERNATIONAL PHARMACOPOEIA MONOGRAPH ON ARTEMETHER AND LUMEFANTRINE CAPSULES REVISED DRAFT FOR DISCUSSION
October 2007 RESTRICTED ` INTERNATIONAL PHARMACOPOEIA MONOGRAPH ON ARTEMETHER AND LUMEFANTRINE CAPSULES REVISED DRAFT FOR DISCUSSION World Health Organization 2007 All rights reserved. This draft is intended
More informationPreparative HPLC: Factors and Parameters that Directly Affect Recovery of Collected Fractions
Preparative HPLC: Factors and Parameters that Directly Affect Recovery of Collected Fractions Gary Scharrer and Joan Stevens, Ph.D. Gilson, Inc Middleton, WI www.gilson.com Abstract Purification of synthetic,
More informationImproving Resolution and Column Loading Systematically in Preparative Liquid Chromatography for Isolating a Minor Component from Peppermint Extract
Improving Resolution and Column Loading Systematically in Preparative Liquid Chromatography for Isolating a Minor Component from Peppermint Extract Jo-Ann M. Jablonski and Rui Chen Waters Corporation,
More informationQuality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC
Quality-by-Design-Based Method Development Using an Agilent 129 Infinity II LC An Efficient Method Development Workflow Combined with ISET-mediated Method Transfer Under Waters Empower 3 CDS Control Application
More informationFast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover
columns ProSwift Reversed-Phase Monolith Columns for Protein Analysis ProSwift reversed-phase columns use a unique monolith technology for fast, high-resolution HPLC and LC/MS separations of proteins.
More informationCharacterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system
Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system Application Note Biopharmaceuticals Authors Arunkumar Padmanaban and Sreelakshmy Menon
More informationTurboMass GC/MS Software
TurboMass GC/MS Software GAS CHROMATOGRAPHY / MASS SPECTROMETRY P R O D U C T N O T E Today s demanding laboratory requires software that is easy to learn yet offers sophisticated instrument control, robust
More informationPAT and UPLC Round Table Introducing PATROL Systems David Lascoux Jean-Michel Plankeele Biopharma LC Meeting Eschborn, October 17th
PAT and UPLC Round Table Introducing PATROL Systems David Lascoux Jean-Michel Plankeele Biopharma LC Meeting Eschborn, October 17th 2012 Waters Corporation 1 What Is PAT? What Are the PATROL Systems For?
More informationSize-Exclusion Chromatography (SEC) Optimization Guide
Size-Exclusion Chromatography (SEC) Optimization Guide 1 Table of Contents I. Introduction II. ACQUITY UPLC Instrumentation Considerations for Successful SEC Analysis III. Importance of Developing a Robust
More informationA Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts
A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts Andrew Aubin and Ronan Cleary Waters Corporation, Milford, MA, USA APPLICATIN BENEFITS A modular preparative HPLC
More informationACQUITY UPLC. I-Class PLUS WHAT SEPARATES YOU FROM EVERYONE ELSE
ACQUITY UPLC I-Class PLUS WHAT SEPARATES YOU FROM EVERYONE ELSE COMPLEX CHALLENGES SOLVED Looking for answers to the most complex scientific challenges is what you do. Your research could lead to ground-breaking
More informationPreparative HPLC System
High Performance Liquid Chromatograph Preparative System Preparative HPLC System C90-E84 From Preparative LC to Purification and Powderization Purification Enters the Next Era Shimadzu provides a wide
More informationGenotoxicity is the property of a compound
Impurities Analysis in Pharmaceuticals: Genotoxicity is the property of a compound known to have irreversible effects on the structure and functionality of the DNA in cells and cause DNA loss, DNA replication
More informationMaximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns
Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns Hongwei Xie, Martin Gilar, and Jeff Mazzeo Waters Corporation, Milford, MA U.S. APPLICATION BENEFITS The ACQUITY UPLC
More informationACQUITY UPLC Protein BEH SEC Columns and Standards
and Standards CONTENTS I. INTRODUCTION II. CONFIGURING AN ACQUITY UPLC SYSTEM FOR USE IN SEC PROTEIN SEPARATIONS a. Calibrators III. GETTING STARTED a. ecord installation b. Column connectors c. Column
More informationAZURA Bio purification
Science Together AZURA Bio purification The new standard of FPLC 1 KNAUER protein purification: The flexible FPLC platform AZURA Bio purification systems AZURA Bio purification: You choose the method Size
More informationSCIEX Anion Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide
For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems A49108AE May 2015 AB Sciex Pte. Ltd and its affiliates disclaims all warranties with respect to this document, expressed or implied, including
More informationAn Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions
An Automated System for At-Line cess Analysis of Biopharmaceutical Fermentation Reactions Frank Steiner, Susanne Fabel, Fraser McLeod, Dionex Corporation, Germering, Germany Abstract Monitoring amino acid
More informationCAMAG TLC-MS INTERFACE 2
CAMAG TLC-MS INTERFACE 2 WORLD LEADER IN PLANAR CHROMATOGRAPHY IDENTIFICATION AND CONFIRMATION OF UNKNOWN SUBSTANCES The elution-based CAMAG TLC-MS Interface 2 is a highly convenient and versatile instrument
More informationExploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma
Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma Yun Wang Alelyunas, Mark D. Wrona, Jim Murphy, Angela Doneanu, Gregory
More informationProminence Preparative HPLC System
High Performance Liquid Chromatograph Preparative System Prominence Preparative HPLC System C190-E110B Prominence Preparative System [Product Concept] The most important aspects of separation and purification
More informationA universal chromatography method for aggregate analysis of monoclonal antibodies
APPLICATION NOTE A universal chromatography method for aggregate analysis of monoclonal antibodies No. 2161 Amy Farrell 1, Jonathan Bones 1, and Ken Cook 2 1 NIBRT, Dublin, Ireland; 2 Thermo Fisher Scientific,
More informationHPLC/FLASH Purification ADVANCE THE PACE OF YOUR DISCOVERY TO ACHIEVE THE RESULTS OF TOMORROW
HPLC/FLASH Purification ADVANCE THE PACE OF YOUR DISCOVERY TO ACHIEVE THE RESULTS OF TOMORROW PLC Purification Systems Accelerate the pace of your discovery S ave time by minimizing the number of fractions
More informationMassPREP On-Line Desalting Cartridge
CONTENTS I. INTRODUCTION II. INSTALLING THE MASSPREP ON-LINE DESALTING CARTRIDGE INTO THE SENTRY.1 X 1 MM GUARD COLUMN HOLDER III. RECOMMENDED LC/MS SYSTEM CONFIGURATION TO MINIMIZE MS SOURCE CONTAMINATION
More informationAPPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology
TN-7 Determination of Impurities and Related Substances for Glibenclamide (EP Monograph 78). Increased Sensitivity, Improved Resolution and Faster Analysis Using Kinetex.6 µm Core-Shell LC Columns Elli
More informationGas Chromatography Assignment Chem 543/443
Gas Chromatography Assignment Chem 543/443 1. Introduction Capillary gas chromatography (GC) is one of the most popular analytical techniques used in today s research. Its popularity is mainly due to efficient
More informationDetermination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography
Application Note 206 Determination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography Introduction The Bayer Process is used to isolate alumina (Al 2 O ) from bauxite. In this process,
More informationConfident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions
Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions Screening and Confirmation Chemical Analysis Solution Unit November, 2009 Page 1 Agilent s New 6540 Ultra High
More informationMethod Translation in Liquid Chromatography
Method Translation in Liquid Chromatography Technical Overview Abstract Ronald E. Majors Agilent Technologies, Inc. 2850 Centerville Rd Wilmington, DE 19808 USA With the recent emphasis on high performance
More informationLiquid Chromatograph Liquid Chromatograph Mass Spectrometer. CL Series. for In Vitro Diagnostic Use
Liquid Chromatograph Liquid Chromatograph Mass Spectrometer CL Series for In Vitro Diagnostic Use Consistency. Performance. Reliability. Shimadzu s CL Series LC, LC-MS, and LC-MS/MS for in vitro diagnostic
More informationThe Theory of HPLC Gradient HPLC
The Theory of HPLC Gradient HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual. 1 Contents
More informationGenerating Automated and Efficient LC/MS Peptide Mapping Results with the Biopharmaceutical Platform Solution with UNIFI
with the Biopharmaceutical Platform Solution with UNIFI Vera B. Ivleva, Ying Qing Yu, Scott Berger, and Weibin Chen Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S The ability to
More informationAflatoxin & Ochratoxin Analysis. A Completely Automated Solution with the Freestyle, ThermELUTE, and SMART Immunoaffinity Cleanup Columns
Aflatoxin & Ochratoxin Analysis A Completely Automated Solution with the Freestyle, ThermELUTE, and SMART Immunoaffinity Cleanup Columns Visit us at www.lctech.de! Background Pickering Labs primarily makes
More informationApplication Note. Author. Abstract. Small Molecule Pharmaceuticals & Generics. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany
Analysis of Degradation Products of Doxycycline in Solution Exposed to Light and Elevated Temperatures Using the Agilent 1 Infinity Series High Dynamic Range Diode Array Detector Impurity Analyzer System
More informationDisulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column
Disulfide Linkage Analysis of IgG1 using an Agilent 126 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column Application Note Biotherapeutics & Biosimilars Author M. Sundaram
More informationAgilent MetaCarb Ca PLUS Carbohydrate Column Ca ++ Form User Manual Part No. A5205
Agilent MetaCarb Ca PLUS Carbohydrate Column Ca ++ Form User Manual Part No. A5205 WARNING: The Agilent MetaCarb Ca PLUS calcium column is packed with a polymeric material that requires special care. Introduction
More informationPerformance. Reliability. Productivity. Consumables
Performance Reliability Productivity Consumables Table of Contents n RediSep Rf Column Overview...2 n RediSep Rf Gold High Performance Flash Chromatography...3 & 4 n RediSep Rf Normal Phase Silica Columns
More informationSize Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column
Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column Application Note Bio-Pharmaceutical Authors M. Sundaram Palaniswamy and Andrew Coffey Agilent Technologies,
More informationSupplemental Information: Polyethylene bio-degradation by caterpillars of the wax moth Galleria mellonella
Supplemental Information: Polyethylene bio-degradation by caterpillars of the wax moth Galleria mellonella Paolo Bombelli, Christopher J. Howe and Federica Bertocchini Supplemental Experimental Procedures
More informationA Primer. Principles in preparative HPLC
A Primer Principles in preparative HPLC Principles in preparative HPLC Udo Huber Ronald E. Majors Principles in preparative HPLC Since the early days of synthetic chemistry, the generation of compounds
More informationChromatography Column Performance and Data Analysis Success Guide. Hints and Tips for Better Purifications
Chromatography Column Performance and Data Analysis Success Guide Hints and Tips for Better Purifications This Chromatography Success Guide provides practical advice on preparative chromatography and protein
More informationEPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube
EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube Application Note Environmental Authors Don Noot, Ralph Hindle Vogon Laboratory Services Cochrane, AB
More informationProSEC 300S. Protein Characterization columns
ProSEC 300S Protein Characterization columns Agilent s ProSEC 300S is a silica-based material specifically designed for the analysis of proteins by aqueous size exclusion chromatography. With a proprietary
More informationAnalysis and Purification of Polypeptides by Reversed-Phase HPLC
Analysis and Purification of Polypeptides by Reversed-Phase HPLC Reversed-phase HPLC is a valuable tool for the analysis and purification of proteins and peptides. It is effective in separating peptide
More informationHIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates
HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates Julia Baek, Shanhua Lin, Xiaodong Liu, Thermo Fisher Scientific, Sunnyvale, CA Application Note 2126 Key Words Size exclusion
More informationDetermination of Adenovirus pviii (31K) Concentration for Estimation of the Empty Capsid Concentration of the Adenovirus Reference Material (ARM)
Determination of Adenovirus pviii (31K) Concentration for Estimation of the Empty Capsid Concentration of the Adenovirus Reference Material (ARM) Sundari Ravindran, Mei Lin, Gary Vellekamp, (SPRI) Summary:
More informationICP Expert software. Technical Overview. Introduction
ICP Expert software Technical Overview Introduction The Agilent 5110 ICP-OES provides fast sample analysis, using less gas, without compromising on performance for tough samples. It has been designed for
More informationBivalirudin Purification:
Bivalirudin Purification: Sorbent Screening and Overload Experiments Marc Jacob, Joshua Heng, and Tivadar Farkas Phenomenex, Inc., 411 Madrid Ave., Torrance, CA 90501 USA PO94190412_W Abstract In this
More informationAZURA Bio LC. 2 FPLC and Purification Solutions
AZURA Bio LC 2 FPLC and Purification Solutions Flexible and adjustable to your biopurification tasks Fast protein liquid chromatography (FPLC) of bio molecules can be a task. KNAUER offers with AZURA an
More informationHPLC to UPLC Method Migration: An Overview of Key Considerations and Available Tools
HPLC to UPLC Method Migration: An Overview of Key Considerations and Available Tools Dr. Michael Swartz, Ph. D. Principal Consulting Scientist Worldwide Pharmaceutical Business Operations Waters Corporation
More informationAgilent 500 Ion Trap LC/MS with the Agilent 1200 Infinity Series
Agilent 500 Ion Trap LC/MS with the Agilent 1200 Infinity Series Quick Start Guide Step 1. Set up and learn about the system 3 Step 2. Create method and Sequence Sample List 4 Step 3. Start the Sample
More informationPractical guidance for switching between anion and cation analysis systems
TECHNICAL NOTE 729 Practical guidance for switching between anion and cation analysis systems Authors Introduction Manali Aggrawal and Jeffrey Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Ion chromatography
More informationThis User Bulletin does not apply to PROCISE clc instruments.
User Bulletin PROCISE Protein Sequencing System (491/492/494) 1 July 1997 (updated 06/2001) SUBJECT: PROCISE Feature Upgrade Kit Summary This User Bulletin describes the hardware, firmware, and software
More informationZwitterion Chromatography ZIC
Zwitterion Chromatography ZIC A novel technique, with unique selectivity, suitable for preparative scale separations? PhD Einar Pontén What is Zwitterion Chromatography? Our definition: Liquid chromatography
More informationThermo Scientific TSQ Quantum Access MAX
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Scientific TSQ Quantum Access MAX UNSURPASSED PRICE TO PERFORMANCE Superior
More informationPhysical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis
Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis Atis Chakrabarti* and Roy Eksteen + Tosoh Bioscience LLC, King of Prussia, PA 19406 *Corresponding Author. + Current address:
More informationC196-E067A. Prominence UFLCXR. Shimadzu Ultra Fast Liquid Chromatograph
C196-E067A Prominence UFLCXR Shimadzu Ultra Fast Liquid Chromatograph New Prominence UFLC Series Line-up Prominence UFLCXR In Pursuit of Higher Separation and Sensitivity The Prominence UFLCXR is an ultra-high-speed
More informationValidated Stability-indicating assay method for determination of Ilaprazole in bulk drug and tablets by high performance liquid chromatography
Validated Stability-indicating assay method for determination of Ilaprazole in bulk drug and tablets by high performance liquid chromatography Pradeep G. Shelke a*, Anil V. Chandewar a, Anil P. Dewani
More informationAnalytical Methods Development and Validation
Understanding and Implementing Efficient Analytical Methods Development and Validation Jay Breaux, Kevin Jones, and Pierre Boulas Analytical methods development and validation play important roles in the
More informationUltrahydrogel Columns
CONTENTS I. INTRODUCTION II. INSTALLATION a. Solvents b. Column c. Organic solvents d. Aqueous salt and buffer solution e. Flow rate f. Solvent and sample preparation g. Guard columns h. Column installation
More informationAgilent 7693A Series Automatic Liquid Sampler. Inject new performance into your gas chromotography
Agilent 7693A Series Automatic Liquid Sampler Inject new performance into your gas chromotography Agilent 7693A Series Automatic Liquid Sampler A powerful combination of productivity and flexibility How
More informationADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS
Agilent AdvanceBio Columns ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS with Agilent ZORBAX RRHD 3Å 1.8 µm columns ns Rapid resolution high definition columns for UHPLC protein and peptide separations
More informationMobileMapper Pro FAQ: Using Waypoints
MobileMapper Pro FAQ: Using Waypoints 13 September 2006 How do I create waypoint files in MobileMapper Office for uploading to the receiver? Click on Tools > Place Waypoints or on the Place Waypoints icon
More informationR R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website: Published in November 2013
5-100 Innovation Way Newark, DE 19711, USA Tel:302-3661101 Fax:302-3661151 Website: www.sepax-tech.com Published in November 2013 P/N SECKIT-7830 These Phases are developed based on innovative surface
More informationAPPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC
UOP 63-13: Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC Fast Analysis in
More informationA Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis
A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis Shanhua Lin, Julia Baek, and Chris Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA Overview
More informationIMPURITIES. Antony Fake API Focal Point, PQTm WHO PREQUALIFICATION TEAM MEDICINES
IMPURITIES Antony Fake API Focal Point, PQTm IMPURITIES PQTm Training Copenhagen, May 2016 1 Introduction This presentation is made with reference to the preparation of an API. This is because the API
More informationAgilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY
Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY AGILENT 7693A SERIES AUTOMATIC LIQUID SAMPLER A powerful combination of productivity and flexibility Autosampler
More informationAnalysis of biomolecules by SEC and Ion-Exchange UPLC
Analysis of biomolecules by SEC and Ion-Exchange UPLC Anders Feldthus, Waters Nordic 2011 Waters Corporation 1 Waters Commitment To develop, commercialize and market columns that when used on Waters ACQUITY
More information