1 TECHNICAL ARTICLE MARKET FOCUSED - environmental ORGANOCHLORINE PESTICIDES With the widespread use of pesticides, especially over the last 40 years, pesticide analysis is one of the most universal methods used in environmental laboratories around the world. Their persistence in the environment, and in some cases, their non-specific mode of action, has led to bans in many countries. This persistence has also resulted in the continuing need for monitoring of this class of compounds. Figure. Organochlorine Pesticide Structures US EPA 808A Method THE ANALYSIS The volatile nature of chlorinated pesticides naturally leads to gas chromatographic analysis. However, resolution requirements and breakdown performance criteria, especially for sample extracts containing high-boiling contaminants, are usually quite stringent. The USEPA has targeted 20 common chlorinated pesticides. Together with two internal standards this leads to a requirement to resolve 22 compounds (Figure ). Because of the screening nature of most environmental methods, there is always pressure in a lab to reduce analysis time and hence increase sample throughput. 2,4,5,-tetrachloro-meta-xylene (surrogate) p,p -DDE Heptachlor Endrin Aldehyde Alpha BHC Dieldrin Aldrin p,p -DDT Beta BHC Endrin Heptachlor epoxide Endosulfan Sulfate Gamma BHC p,p -DDD Gamma Chlordane Methoxychlor Delta BHC Endosulfan B Alpha Chlordane Endrine Ketone Endosulfan A Decachlorobiphenyl Surrogate
2 CHOICE OF CAPILLARY COLUMN There is a wide choice of capillary columns from SGE for the analysis of pesticides. It is common practice in labs around the world to split the sample after GC injection to two different capillary columns. These columns are usually of differing polarity and must show different retention times of analytes, preferably with some peak switching. It is common to quantify pesticide response on a non-polar column such as the SGE 5% phenyl BPX5 column or the SGE 00% methyl SolGel-ms column. The latter column, although custom designed as a low-bleed mass spec column, can be used successfully with all GC detectors. The second column used is one of differing polarity such as higher percentage phenyl column such as the 35% or 50% phenyl columns (BPX35 and BPX50, respectively). Using any of the columns described above, it is possible to analyse the 22 targeted compounds with excellent baseline resolution in under 22 minutes. This allows for greater throughput of samples. BPX5 5% phenyl (equiv) polysilphenylene-siloxane USEPA 808 Organochlorine Pesticide mix analyzed on the BPX5 column (Figure 2), shows an excellent separation of the components in less than 24 minutes. DDT (peak 8) and Methoxychlor (peak 20) show no on-column breakdown indicating the high degree of inertness of BPX5. Also note the exceptional bleed profile at 300 C. Figure 2. USEPA 808 Organochlorine Pesticide BPX5 replaces... DB-5 DB-5MS DB-5.25 XTI-5 Rtx-5ms Ultra-2 HP-5 HP-5MS HP5-TA SPB-5 MDN-5S CP-Sil 8CB Rtx-5Sil MS AT-5 CP-Sil 8CB MS Phase: BPX5, 0.25mm film Sample: 00ppm in Dichloromethane 30m x 0.25 mm ID Initial Temp: 0 C, min. Rate : 25 C/min to 0 C, Rate 2: 2 C/min to 20 C Rate 3: C/min to 30 C Final Temp: 30 C, min. Detector Type: Mass Spectrometer Carrier Gas: He, 0.5 psi Carrier Gas Flow:.0 ml/min. Constant Flow: On Average Linear Velocity: 3 cm/sec at 0 C Injection Mode: Split Split Ratio: 50: Injection Volume: ml Injection Temperature: 250 C Autosampler: No Liner Type: 4 mm ID Single Taper Liner Liner Part Number: Column Part Number: 0540 COMPONENTS. 2,4,5,-Tetrachlorometa-xylene. Heptachlor 9. gamma-chlordane 0. alpha-chlordane. Endosulfan A 4. Endrin 7. Endrin aldehyde 8. p, p' - DDT 9. Endosulfan Sulfate 2. Endrin ketone 22. Decachlorobiphenyl Exceptioanlly low bleed at 300 C
3 THE ADVANTAGE OF HIGH MAXIMUM TEMPERATURE COLUMNS The composition of sample extracts in an environmental lab can be extremely variable. From relatively clean water samples to extremely complex dirty soil samples. It is extremely useful to use a capillary column with a high maximum operating temperature. The column can then be baked out following deposition of semi and non-volatile residue material on the front end of the column. Once peak shape deteriorates, it is common practice to condition the column at the upper maximum temperature of the column for at least 0 minutes to remove this residue. If this fails to re-establish good chromatography then the front 0.5 meter of the SolGel-ms column can be cut and discarded. The second advantage of a column with a high maximum operating temperature is lower bleed at the upper method temperature. Thus, for example, if the upper maximum temperature of the column is 370 C, the bleed level at 300 C will be lower compared with a column with an upper maximum level of 320 C. This leads to better sensitivity for the higher maximum temperature column. Thermal stability of SGE stationary phases is excellent because of advanced polymer technology. For example, up to 30 C for BPX5, 35, 50 and up to 380 C for a SolGel-ms column. The very low bleed level at the maximum operating temperatures of these columns is a particularly advantageous to the chromatographer. Figure 3 shows the USEPA 808 pesticide mix analyzed using a SolGel-ms column. Note the superb bleeding profile at 300 C and separation of the components with no coelution s in less than minutes allowing multiple samples to be run in the laboratory. Again, SolGel-ms, like BPX5, BPX35 and BPX50 shows superior inertness with no-column breakdown of the susceptible DDT and Methoxychlor. Also note, 4 different elution orders making SolGel-ms an exceptional conformational column to that of BPX5. Figure 3 - SolGel-ms Column Phase: SolGel-ms, 0.25µm film Sample: 00ppm in Dichloromethane 30m x 0.25mm ID Initial Temp: 00 C, min Rate : 25 C/min to 0 C Rate 2: 2 C/min to 20 C Rate 3: C/min to 300 C Final Temp: 300 C, 5 min Detector Type: Mass Spectrometer Carrier Gas: He, 3. psi Carrier Gas Flow:.7mL/min Constant Flow: On Average Linear Velocity: 35cm/sec at 0 C Injection Mode: Splitless Purge on Time: 0.5min Purge On (Split) Vent Flow: 0mL/min Injection Volume: µl Injection Temp.: 250 C Autosampler: No Liner Type: 4mm ID SIngle Taper Liner Liner Part No.: Column Part No.: ms-novent Part no.:3400 HP5973 restrictor: 3409 Full Scan/SIM: Full Scan COMPONENTS. 2,4,5,-Tetrachlorometa-xylene. Heptachlor 9. gamma-chlordane 0. alpha-chlordane. Endosulfan A 4. Endrin 7. Endrin aldehyde 8. p, p' - DDT 9. Endosulfan Sulfate 2. Endrin ketone 22. Decachlorobiphenyl
4 BPX35 35% phenyl (equiv) polysilphenylene-siloxane Figure 3 shows the 808 USEPA pesticide mix analyzed on a BPX35 column. Note the brilliant peak shape, separation and the high degree of inertness with respect to DDT and Methoxychlor components. Also note the different order of elution of the components. BPX35 has two differences in the order of elution making it a perfect confirmation column for BPX5. It also has a very low bleed profile at 300 C. BPX35 replaces... DB-35 DB-35MS Rtx-35ms HP-35 HP-35MS SPB-35 MDN-35 AT-35 Figure 3. USEPA 808 Organochlorine Pesticide Phase: BPX35, 0.25 µm 30m x 0.25mm ID Initial Temp: 40 C for min Rate : 30 C/min to 90 C for 3min Rate 2: 0 C/min to 300 C for 5min Detector Type: Mass Spectrometer Carrier Gas: Helium at 0.0psi Carrier Gas Flow:.3 ml/min. Constant Flow: OnAv. Linear Velocity: 3 cm/sec at 40 C Injection Mode: Splitless Purge on Time: min Purge on (Split) Vent: 0mL/min Injection Volume: µl Injection Temp: 250 C Autosampler: No Detector: MSD Liner Type: 4 mm ID Double Taper Liner Column Part No: Detector: Mass Spectrometer COMPONENTS:. 2,4,5,-Tetrachloro -meta-xylene. Heptachlor 9. gamma Chlordane 0. alpha Chlordane. Endosulfan A 4. Endrin 7. Endrin Aldehyde 8. p, p' - DDT 9. Endosulfan Sulfate 2. Endrin Ketone 22. Decachlorobipheny minutes
5 BPX50 50% phenyl (equiv) polysilphenylene-siloxane In Figure 4, the USEPA 808 pesticide mix is analyzed on a BPX50 column. Note the superb bleed profile at 300 C and separation of the components in less than 22 minutes allowing multiple samples to be run in the laboratory. Again BPX50, like BPX5, SolGel-ms and BPX35, shows superior inertness with no on-column breakdown of the susceptible DDT and Methoxychlor. Figure 4. USEPA 808 Organochlorine Pesticide BPX50 replaces... OV-7 SP-2250 DB-7 DB-7MS DB-7ht Rtx-50 SPB-50 HP-50+ AT Phase: BPX50, 0.25 µm 30m x 0.25mm ID Injector Temp: 240 C Injection Volume: µl Autosampler Syringe: 5µL Fixed Needle (p/n 0080) Septa: Auto-Sep T (p/n 04882) Injection Type: Splitless Purge On Time: 0.5min Purge On (Spilt) Vent: 0mL/min Liner Type: Single taper (p/n 09207) Carrier Gas: Helium Constant Flow: On Pressure: 4.psi Column Flow:.3mL/min Linear Velocity: 40 C Initial Temp: 40 C for min Rate : 30 C/min to 90 C for 3min Rate 2: 0 C/min to 300 C for 5min Run Time: 25min Detector Type: 30 C Column Part No: COMPONENTS:. 2,4,5,-Tetrachloro - meta-xylene. Heptachlor 9. gamma Chlordane 0. alpha Chlordane 2. Endosulfan A 4. Endrin 7. Endrin Aldehyde 8. p, p' - DDT 9. Endosulfan Sulfate 2. Endrin Ketone 22. Decachlorobiphenyl minutes THE CHROMATOGRAPHY SGE s BPX5, SolGel-ms BPX35 and BPX50 phases show excellent inertness and do not have problems such as on column breakdown of methoxychlor and DDT. Indeed these three columns are exceptional for the analysis of the 22 chlorinated pesticides in the U.S. EPA methods 808 and 08. With at least two changes in the elution order of each column to that of BPX5, SolGel-ms, BPX35 and BPX50 columns are excellent confirmation columns for this type of analysis.
6 Ordering Information MARKET FOCUSED - environmental ORGANOCHLORINE PESTICIDES US EPA 808A Method ID(mm) Length Film Thickness Temperature Limits Part No. (mm) (m) (µm) ( C) BPX to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 350/ to 350/ to 30/ to 30/ to 30/ to 350/ to 350/ to 30/ to 350/ BPX to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ BPX to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ to 30/ SolGel-ms to 370/ to 370/ to 370/ to 370/ Publication No. TA-0079-C Rev:02 05/0 SGE Office Details SGE International Pty. Ltd. Toll Free: Tel: + (0) Fax: + (0) support@sge.com Web: SGE Incorporated (USA) Toll Free: Fax: usa@sge.com SGE Europe Ltd. (UK) Tel: +44 (0) Fax: +44 (0) uk@sge.com SGE (France) Sarl Tel: +33 (0) Fax: +33 (0) france@sge.com SGE (Italia) Srl. Tel: Fax: sge.italia@tin.it SGE Deutschland GmbH Tel: +49 (0) 8048 Fax: +49 (0) germany@sge.com SGE China Service Centre Tel: +8 (0) Fax: +8 (0) sge@bj.col.com.cn SGE Japan Inc. Tel: +8 (45) Fax: +8 (45) japan@sge.com SGE India Tel: +9 (022) Fax: +9 (022) sgeindia@vsnl.com
Contract Laboratory Program (CLP) Pesticide Analysis with 0.18 mm ID High-Efficiency GC Columns Utilizing Helium Carrier Gas. Application.
Contract Laboratory Program (CLP) Pesticide Analysis with 0.8 mm ID High-Efficiency GC Columns Utilizing Helium Carrier Gas Application Environmental Authors Ken Lynam and William Long Agilent Technologies,
More informationA Direct Column-Performance Comparison for Rapid Contract Laboratory Program (CLP) Pesticide Analysis. Application. Authors. Abstract.
A Direct Column-Performance Comparison for Rapid Contract Laboratory Program (CLP) Pesticide Analysis Application Environmental Authors Ken Lynam and John W. Henderson, Jr. Agilent Technologies, Inc. 2850
More informationANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER
ENVIRONMENTAL ANALYSIS ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER Solutions for Your Analytical Business
More informationCapillary GC Troubleshooting: a Practical Approach
Capillary GC Troubleshooting: a Practical Approach Katherine Stenerson Gas Separations R&D Supelco, Supelco Park, Bellefonte, PA 16823 1999 Sigma-Aldrich Co. Outline Basic Troubleshooting Strategy Preventing
More informationSGE Environmental GC Columns The Suite for Environmental GC Analysis
The Suite for Environmental GC Analysis SGE Environmental GC Columns Performance Selectivity Delivery Promise SGE Environmental GC Columns The Suite for Environmental GC Analysis SGE GC Columns have a
More informationApplication Note: Multi-Residue Pesticides in Fruit and Vegetables (9/16/08)
Application Note: Multi-Residue Pesticides in Fruit and Vegetables (9/16/08) Product: DPX-Q (5 ml) INTRODUCTION The analysis of pesticides in fruit and vegetables is very important in the field of food
More informationAnalysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application
Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application Food Authors Frank David Research Institute for Chromatography Pres. Kennedypark
More informationAnalysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS
Analysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS Authors: Conor Smith, Doug Fryer United Science Corp. 15911 Furuby Rd. Center City, MN 55012 USA Abstract
More informationDetermination of Persistent Organic Pollutants (POPs) in Fish
Faculty of Veterinary Medicine Determination of Persistent Organic Pollutants (POPs) in Fish Dipartimento di Scienze Veterinarie e Sanità Pubblica (DIVET) Prof. L.Chiesa, Dr. S. Panseri -Prague, April
More informationGas Chromatography Assignment Chem 543/443
Gas Chromatography Assignment Chem 543/443 1. Introduction Capillary gas chromatography (GC) is one of the most popular analytical techniques used in today s research. Its popularity is mainly due to efficient
More informationPROFICIENCY TESTING PROGRAM. June 2013
Report No. 810 SOILS (CHEMICAL) (ROUND 7) PROFICIENCY TESTING PROGRAM June 2013 ACKNOWLEDGMENTS PTA wishes to gratefully acknowledge the technical assistance provided for this program by Mr C Wood from
More informationFULLY AUTOMATED SPE-DEX Horizon Technology. understands the. challenges today s. laboratories face. The automated. extraction systems
FULLY AUTOMATED SPE-DEX 4790 The Horizon Technology SPE-DEX 4790 Automated Extraction System is a programmable, multipurpose, automated SPE system, capable of processing aqueous samples directly from their
More informationEXPORT SOIL-MATERIAL QUALITY REQUIREMENTS
EXPORT SOIL-MATERIAL QUALITY REQUIREMENTS GENERAL Industrial operations have been conducted on Port of Long Beach property since the 1930s. Though it is unlikely that hazardous substances will be encountered
More informationOptimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools
Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools Application Note Abstract The Commonwealth of Massachusetts Department of Environmental Protection
More informationAnalysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph
Analysis of USP Residual Solvents using the Agilent 797A Headspace Sampler with the Agilent 79B Gas Chromatograph Application Note Pharmaceuticals Authors Bart Tienpont, Frank David, and Pat Sandra
More informationExtraction of POPs from Environmental Samples using Accelerated Solvent Extraction (ASE )
Extraction of POPs from Environmental Samples using Accelerated Solvent Extraction (ASE ) Frank Höfler* and Khalil Divan** *Dionex (Europe) Management AG Solothurner Strasse 259 4600 Olten Switzerland
More informationPROFICIENCY TESTING PROGRAM. June 2014
Report No. 859 SOILS (CHEMICAL) ROUND 9 PROFICIENCY TESTING PROGRAM June 2014 ACKNOWLEDGMENTS PTA wishes to gratefully acknowledge the technical assistance provided for this program by Mr M Blades from
More informationHYDROGEN AS A CARRIER GAS. Gas Chromatography new frontiers: can Hydrogen substitute Helium?
HYDROGEN AS A CARRIER GAS Gas Chromatography new frontiers: can Hydrogen substitute Helium? Manufacturing UHP gas generators for lab analysis since 1979, CLAIND is seating in Lenno, on Como Lake, where
More informationFast Detailed Hydrocarbon Analysis by Modified ASTM Method D6730
application Note Gas Chromatography Author Terry Osenbach PerkinElmer, Inc. Shelton, CT 06484 USA Fast Detailed Hydrocarbon Analysis by Modified ASTM Method D6730 Introduction Detailed hydrocarbon analysis
More informationSPE-DEX 4790 Automated Extraction System. A Fully Automated Sample Preparation System. votre distributeur
SPE-DEX 4790 Automated Extraction System A Fully Automated Sample Preparation System votre distributeur FULLY AUTOMATED SPE-DEX 4790 The Horizon Technology SPE-DEX 4790 Automated Extraction System is a
More informationImpact of GC Parameters on The Separation Part 4: Choice of Film Thickness
Cd The Chrom Doctor Impact of GC Parameters on The Separation Part 4: Choice of Film Thickness Jaap de Zeeuw, Restek Corporation, Middelburg, The Netherlands. In part 1, 2 and 3 of this series we focused
More informationSCION GC Series. The Gas Chromatographers Choice for Separations
SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Scion Instrument s tradition of innovation and product reliability have combined to create the next generation
More informationAccurately Identify and Quantify One Hundred Pesticides in a Single GC Run
Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run Application Note Author Jessica Westland Agilent Technologies, Inc. Abstract A selected target compound list of 195 various pesticides
More informationThe Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications
TECHNICAL BRIEF Gas Chromatography/ Mass Spectrometry The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications C 14 scans Introduction As gas chromatography
More informationCHROMATOGRAPHY/MASS SPECTROMETRY
DETERMINATION OF 1,4-DIOXANE IN DRINKING WATER BY SOLID PHASE EXTRACTION (SPE) AND GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS) WITH SELECTED ION MONITORING (SIM)* EPA Method 522 Part Number: EU52112M6
More informationSURPLUS SOIL-MATERIAL REUSE REQUIREMENTS
SURPLUS SOIL-MATERIAL REUSE REQUIREMENTS GENERAL Surplus soil-materials generated from most, if not all, Long Beach Harbor Department projects are unlikely to meet established Port Export Soil-material
More informationPerkinElmer Supplies/Parts. GC Accessories. Autosampler Syringes
Autosampler Syringes Volume, Code, Gauge, Length, Point Style QTY OEM No. PerkinElmer No. Part No. Hamilton Syringes 5μL, ASN, 23s, 70mm, #3 1 88035 N6101390 24-88035 5μL, ASN, 26s, 70mm, #3 1 88040 N6101380
More informationProtocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016
Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016 1.0 Method background: This method employs full evaporation technique headspace
More informationAgilent Ultimate Plus Fused Silica Tubing
Agilent Ultimate Plus Fused Silica Tubing Technical Overview Introduction Deactivated fused silica tubing is widely used for guard columns, transfer lines, or long retention gaps in GC and GC/MS applications
More informationSeparation and Quantitation of a Mixture of Alcohols by Gas Chromatography
Separation and Quantitation of a Mixture of Alcohols by Gas Chromatography 1. Purpose This procedure will determine the percentage of two alcohols in an unknown aqueous mixture. 2. Background Gas chromatography
More informationCONTENTS Alternate Retention Time Window Determination
PAGE: 1 of 40 CONTENTS 1.0 SCOPE AND APPLICATION* 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 3.1 Sample Storage 3.2 Holding Time 4.0 INTERFERENCES AND POTENTIAL PROBLEMS
More informationQuantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy
Advances in Bioscience and Biotechnology, 2010, 1, 367-371 doi:10.4236/abb.2010.15049 Published Online December 2010 (http://www.scirp.org/journal/abb/). Quantitative determination of residual 2-(2-chloroethoxy)
More informationFast GC Analyses of Volatiles
6 Fast GC Analyses of Volatiles Michael D. Buchanan mike.buchanan@sial.com Introduction The primary aim of Fast GC is to maintain (compared to conventional GC) sufficient resolving power in a shorter time.
More informationLow-level Detection of Ethanol, 1,4-Dioxane, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator
Application Note 25370206 Low-level Detection of,, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator Keywords Eclipse Fuel Oxygenates Methanol Model 4552 Purge and Trap Sample Concentrator
More informationSystem Overview. Valve oven. Restrictor. Valve. Sample in and out
System Overview The GC system is designed to provide a continous reading of total hydrocarbons in a gas stream, while periodically performing a chromatographic separation of the sample to determine the
More informationCo-ordination of an organochlorine pesticides in drinking water proficiency testing scheme in South(ern) Africa
Co-ordination of an organochlorine pesticides in drinking water proficiency testing scheme in South(ern) Africa D Prevoo-Franzsen 8 th Eurachem PT workshop, Berlin, Germany 09 Oct 2014 NMISA Our mandate:
More informationIt works as hard as you do.
It works as hard as you do. Agilent 6890N Network GC Rugged and reliable for unsurpassed uptime Advanced EPC ensures precision and accuracy of results Retention Time Locking for consistent results Accomplish
More informationAgilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES
Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES THE AGILENT 7820A GC SYSTEM FORWARD-THINKING INNOVATION, EVERYDAY RELIABILITY AND VALUE Is your lab looking for better, more reproducible results
More informationSTANDARD OPERATING PROCEDURES
PAGE: 1 of 5 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS
More informationDetermination of organochlorine pesticide residues in eggs by gel-permeation chromatography (GPC) cleanup and GC-ECD
Determination of organochlorine pesticide residues in eggs by gel-permeation chromatography (GPC) cleanup and GC-ECD 1. Experimental Section LabTech, Inc. 1.1 Instruments and reagents AutoClean automatic
More informationMS Certified Vials. Pre-cleaned and certified vials for mass spectrometry
Pre-cleaned and certified vials for mass spectrometry The FIRST and ONLY pre-cleaned, low particle, low background chromatography vial When your instrumentation, sample handling and methodology is pushing
More informationElectron Capture Detector for GC-2010 Plus
Electron Capture Detector for GC-2010 Plus C180-E080 The World's Highest Level of Durability and Performance for a Capillary ECD Electron Capture Detector Novel Contact Free Technology was incorporated
More informationSIMPLIFY AND ACCELERATE GAS ANALYSIS. Micro GC Fusion GAS ANALYZER
SIMPLIFY AND ACCELERATE GAS ANALYSIS Micro GC Fusion GAS ANALYZER Maximum throughput and easy operation for fast decisions when they matter most Micro GC Fusion offers significant throughput gain through
More informationThe Analysis of 4-Vinyl-Cyclo-Hexene (VCH) in Butadiene with the DVLS LGI Injector
The Analysis of 4-Vinyl-Cyclo-Hexene (VCH) in Butadiene with the DVLS LGI Injector Introduction In synthetic rubber manufacturing, 4-Vinylcyclohexene (VCH) can be found as a byproduct from the chemical
More informationUsing the Pulsed Flame Photometric Detector (PFPD) for Low-level Analysis of Organophosphorus Pesticides
Application Note 25390206 Using the Pulsed Flame Photometric Detector (PFPD) for Low-level Analysis of Organophosphorus Pesticides Keywords Model 5380 Organophosphorus Pesticide PFPD PFPDView Phosphorus
More informationGrace GC Capillary Column Families
Grace GC Capillary Column Families Alltech Heliflex High-Performance, Widest Selection of Phases Alltech Econo-Cap Best Value, Popular Phases Grace offers two lines of capillary columns to suit every application
More informationGas Chromatography (GC) Operation, Calibration, Troubleshooting and Practice Maintenance
Training Title Gas Chromatography (GC) Operation, Calibration, Troubleshooting and Practice Maintenance Training Duration 5 days Training Venue and Dates Gas Chromatography (GC) Operation, Calibration,
More informationFormulas: See Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1. METHOD: 2555, Issue 1 EVALUATION: PARTIAL Issue 1: 15 March 2003
KETONES I 2555 Formulas: See Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1 METHOD: 2555, Issue 1 EVALUATION: PARTIAL Issue 1: 15 March 2003 OSHA : See Table 2 NIOSH: See Table 2 ACGIH: See Table 2 PROPERTIES:
More informationC146-E096 TD-20. Shimadzu Thermal Desorption System
C146-E096 TD-20 Shimadzu Thermal Desorption System Thermal Desorption Thermal desorption systems trap target substances carried in mobile phase by adsorption in a sample tube packed with adsorbent. Then,
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares
More informationSCION GC Series. The Gas Chromatographers Choice for Separations
SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Bruker s Scion Instrument s long tradition long of tradition innovation of innovation and product and reliability
More informationTechnical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany
New Features of the Agilent Method Scouting Wizard for Automated Method Development of Complex Samples Analysis of Large Data Sets by Method Scouting Reports and Automated Adjustment of Flow Rates and
More informationCLEAN WATER Clean Water PT Scheme Schedule Clean Water PT Scheme Schedule. Schedule subject to change see ERA s website at
For International purchase only: Synthetic drinking water matrices with low concentrations of analytes for testing water supply, drinking water, or ground water. Standards are designed to meet the requirements
More informationPAL SPME Fibers Optimized for Automation
PAL SPME Fibers Optimized for Automation PAL SPME Fiber Since its introduction by Pawliszyn et al. (ref. 1) Solid Phase Micro Extraction (SPME) has seen a tremendous development. SPME is a very effective
More informationElute LC series. Innovation with Integrity. UHPLC Precision for MS applications LC-MS
Elute LC series UHPLC Precision for MS applications Innovation with Integrity LC-MS Delivering Precision and Speed for Mass Spectrometry Designed for reproducibility in retention time and peak shape, Elute
More informationDevelopment of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal Antibody Charge Variants
Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal ntibody Charge Variants Ken Cook, Frank Steiner, Mauro De Pra Thermo Fisher Scientific, Hemel Hempstead,
More informationHYDROCARBONS, BP C 1500
HYDROCARBONS, BP 36 216 C 1500 FORMULA: Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1 METHOD: 1500, Issue 3 EVALUATION: PARTIAL Issue 1: 15 August 1990 Issue 3: 15 March 2003 OSHA : Table 2 NIOSH: Table
More informationChem 321 Lecture 23 - Liquid Chromatography 11/19/13
Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Student Learning Objectives High Performance Liquid Chromatography With the advent of relatively inexpensive and reliable pumps, the development of
More informationMethod Translation in Liquid Chromatography
Method Translation in Liquid Chromatography Technical Overview Abstract Ronald E. Majors Agilent Technologies, Inc. 2850 Centerville Rd Wilmington, DE 19808 USA With the recent emphasis on high performance
More informationApplication Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany
Quantification of trace-level herbicides in drinking water by online enrichment with the Agilent Infinity Series Online-SPE Solution and Triple Quadrupole MS Detection Application Note Environmental Author
More informationSubstrate Selective Amide Coupling driven by Encapsulation of a Coupling Agent within a Self-Assembled Hexameric Capsule
Electronic Supplementary Material (ESI) for ChemComm. This journal is The Royal Society of Chemistry 14 Substrate Selective Amide Coupling driven by Encapsulation of a Coupling Agent within a Self-Assembled
More informationInternational Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories
International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories Vienna International Centre, P.O. Box 100, 1400 Vienna, Austria REFERENCE SHEET REFERENCE
More informationNEW? GC ACCESSORIES INSTRUMENT SUPPLIES. What s. look for this circle
GC ACCESSORIES INSTRUMENT SUPPLIES Flowmeter..............................................203 Leak Detector..........................................204 Septa...................................................205
More informationLower Mystic Fish Advisory Project Presentation to the Mystic River Watershed Initiative Science Forum May 5 th, 2017 Patrick Herron Rafael Mares
Lower Mystic Fish Advisory Project Presentation to the Mystic River Watershed Initiative Science Forum May 5 th, 2017 Patrick Herron Rafael Mares The Problem A fish advisory exists for the freshwater upper
More informationSuper Clean Gas Filters Simply the best
Super Clean Gas Filters TM Simply the best Cartridge Type rev. 1.4 Features EP-B-0606960 U.S. Patent 5.478.378 Ultra-high capacity High-purity output ensures 99.9999% Pure gas Quick connect fittings for
More informationAn Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions
An Automated System for At-Line cess Analysis of Biopharmaceutical Fermentation Reactions Frank Steiner, Susanne Fabel, Fraser McLeod, Dionex Corporation, Germering, Germany Abstract Monitoring amino acid
More informationCertified Reference Material - Certificate of Analysis
Certified Reference Material - Certificate of Analysis S(-)-Nicotine, Primary Standard N-008 Page 1 of 7 Catalog Number: N-008 H Lot: Expiration: October 2019 N Description: S(-)-Nicotine in Methanol.
More informationMassPREP On-Line Desalting Cartridge
CONTENTS I. INTRODUCTION II. INSTALLING THE MASSPREP ON-LINE DESALTING CARTRIDGE INTO THE SENTRY.1 X 1 MM GUARD COLUMN HOLDER III. RECOMMENDED LC/MS SYSTEM CONFIGURATION TO MINIMIZE MS SOURCE CONTAMINATION
More informationEPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube
EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube Application Note Environmental Authors Don Noot, Ralph Hindle Vogon Laboratory Services Cochrane, AB
More informationBÜCHI Labortechnik AG
BÜCHI Labortechnik AG Purification BUCHI Chromatography History 1977: BUCHI Fraction Collector 1978: 1 st preparative chromatography system 2008: automated chromatography system 2014: automated high performance
More informationSelect Al2O3 MAPD : a new Inert Al2O3 column for the trace analysis of polar reactive hydrocarbons.
GC Columns Select Al2O3 MAPD : a new Inert Al2O3 column for the trace analysis of polar reactive hydrocarbons. Jaap de Zeeuw and Coen Duvekot Varian, Inc. Middelburg, The Netherlands Varian PO Box 8033,
More informationTrihalomethane Formation Potential in Rainwater Harvested from Different Roofing Materials
Trihalomethane Formation Potential in Rainwater Harvested from Different Roofing Materials Project Number: 570652 Principal Investigator: Carolina Mendez, M.S., Department of Civil, Architectural and Environmental
More informationMaximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns
Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns Hongwei Xie, Martin Gilar, and Jeff Mazzeo Waters Corporation, Milford, MA U.S. APPLICATION BENEFITS The ACQUITY UPLC
More informationKEYWORDS ABSTRACT AppNote 7/2000
AppNote 7/0 Analysis of Wet Samples by Direct Ther mal Desorption GC Edward Pfannkoch, Jackie Whitecavage Gerstel, Inc., 701 Digital Drive, Suite J, Linthicum, MD 21090, USA KEYWORDS Thermal Desorption,
More informationGAS CHROMATOGRAPHY CLARUS 400/480 GC
GAS CHROMATOGRAPHY CLARUS 400/480 GC Customer Hardware and Service Guide Clarus 400/480 GC Customer Hardware and Service Guide Release History Part Number Release Publication Date 09936811 C January 2010
More informationAPPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC
UOP 63-13: Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC Fast Analysis in
More information1-(2-fluorophenyl)-2-(methylamino)propan-1-one
-luoromethcathinone Latest Revision: April 5, 0 CH CH. GENERAL INRMATIN IUPAC Name: -(-fluorophenyl)--(methylamino)propan--one CR: Not Scheduled (/0) CAS#: Synonyms: Source: Appearance: Kovat's Index:
More informationResidual Solvents in medical cannabis analysis using the SRI 8610C FID GC June 2014
June 2014 The SRI Medical Herb Potency 8610C GC is shown at right. This GC can also be used to test for residual solvents (i.e. butane, acetone, gasoline residue, etc.) in medical cannabis. These solvents
More informationTechnical Report. A Tool for Selecting an Adsorbent for Thermal Desorption Applications
Supelco Park 595 North Harrison Road Bellefonte, PA 16823-0048 USA Telephone 800-247-6628 814-359-3441 Fax 800-447-3044 814-359-3044 email: supelco@sial.com sigma-aldrich.com/supelco Technical Report A
More informationCOST SAVING SOLUTIONS
COST SAVING SOLUTIONS Customers often tell us that their lab s budget is tight and getting tighter, yet they re still expected to provide quality and timely laboratory results. A difficult challenge for
More informationZORBAX StableBond HPLC Columns
ZORBAX HPLC Columns Longest column lifetime and best reproducibility for low ph separations down to ph 1 Patented stable column chemistry allows use at high temperature and low ph without degradation Six
More informationA Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts
A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts Andrew Aubin and Ronan Cleary Waters Corporation, Milford, MA, USA APPLICATIN BENEFITS A modular preparative HPLC
More informationOverview of preparative HPLC. Analytical Technologies Limited
Overview of preparative HPLC Analytical Technologies Limited 1.Review of liquid phase chromatography origin develop prospect Mobile phase Tewett 1903 Stationa ry phase origin Separation result Liquid-stationary
More informationThermo Scientific EASY-nLC 1200 System. Leading in simplicity. and performance
Thermo Scientific EASY-nLC 1200 System Leading in simplicity and performance Peak performance made EASY Effortless ultra high performance for everybody A straightforward LC-MS solution Optimized and integrated
More informationNATURALLY BETTER CHROMATOGRAPHY SOLUTIONS. Clarus Gas Chromatographs Clarus 680/580/480
Clarus Gas Chromatographs Clarus 680/580/480 NATURALLY BETTER CHROMATOGRAPHY SOLUTIONS CLARUS DELIVERS POWER AND PERFORMANCE AT EVERY LEVEL The Clarus GC Family Performance you can rely on from the leader
More informationHIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates
HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates Julia Baek, Shanhua Lin, Xiaodong Liu, Thermo Fisher Scientific, Sunnyvale, CA Application Note 2126 Key Words Size exclusion
More informationOlga Shimelis, Katherine Stenerson, Michael Ye, Carmen T. Santasania, and Ed Mauney. Supelco, Div. of Sigma-Aldrich, Bellefonte, PA USA
Method for Analysis of rganophosphorous Pesticides in Produce using Acetonitrile Extraction, Dual-Layer Carbon-Aminopropylsilica Cleanup and GC-NPD Detection lga Shimelis, Katherine Stenerson, Michael
More informationNEW? GC COLUMNS FUSED SILICA COLUMNS. What s. look for this circle
GC CLUMNS FUSED SILICA CLUMNS Selecting a GC Column...............................23-27 GC Column Cross-References.........................28-29 GC Column Installation..................................30
More informationAgilent 8355 Sulfur Chemiluminescence Detector REIMAGINED. REDESIGNED.
Agilent 8355 Sulfur Chemiluminescence Detector REIMAGINED. REDESIGNED. AGILENT 8355 SULFUR CHEMILUMINESCENCE DETECTOR (SCD) THE ONLY INTEGRATED SYSTEM FOR LOW-LEVEL SULFUR ANALYSIS Why did Agilent reimagine
More informationVOC- AND SVOC-EMISSIONS FROM ADHESIVES, FLOOR COVERINGS AND COMPLETE FLOOR STRUCTURES
VOC- AND SVOC-EMISSIONS FROM ADHESIVES, FLOOR COVERINGS AND COMPLETE FLOOR STRUCTURES O Wilke *, O Jann and Brödner Federal Institute for Materials Research and Testing, 12200 Berlin, Germany ABSTRACT
More informationFast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover
columns ProSwift Reversed-Phase Monolith Columns for Protein Analysis ProSwift reversed-phase columns use a unique monolith technology for fast, high-resolution HPLC and LC/MS separations of proteins.
More informationInnovation with Purpose AGILENT INTUVO 9000 GC COLUMNS AND SUPPLIES
Innovation with Purpose AGILENT INTUVO 9000 GC COLUMNS AND SUPPLIES AGILENT INTUVO 9000 AGILENT SUPPLIES FOR PerkinElmer ICP-OES AGILENT INTUVO INNOVATION WITH PURPOSE Agilent has broken new ground in
More informationExploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma
Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma Yun Wang Alelyunas, Mark D. Wrona, Jim Murphy, Angela Doneanu, Gregory
More informationTotal Petroleum Hydrocarbons By Gas Chromatography in Colombian Waters and Soils
American Journal of Environmental Science, 2012, 8 (4), 396-402 ISSN: 1553-345X 2012 Science Publication doi:10.3844/ajessp.2012.396.402 Published Online 8 (4) 2012 (http://www.thescipub.com/ajes.toc)
More informationAgilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY
Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY AGILENT 7693A SERIES AUTOMATIC LIQUID SAMPLER A powerful combination of productivity and flexibility Autosampler
More informationApplication Note. Abstract. Authors. Introduction. Environmental
EPA Method 538: Determination of Selected Organic Contaminants in Drinking Water by Direct Aqueous Injection with the Agilent 646 Triple Quadrupole LC/MS System Application Note Environmental Authors Imma
More informationInvestigation of formalin and ethepon in some fruits of three local markets of Mymensingh district using gas chromatograph
J. Bangladesh Agril. Univ. 13(1): 7 12, 2015 ISSN 1810-3030 Investigation of formalin and ethepon in some fruits of three local markets of Mymensingh district using gas chromatograph M. A. Islam*, M. E.
More informationConfident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions
Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions Screening and Confirmation Chemical Analysis Solution Unit November, 2009 Page 1 Agilent s New 6540 Ultra High
More informationAgilent 7890B Gas Chromatograph RESOLVE YOUR SEARCH FOR RELIABILITY
Agilent 7890B Gas Chromatograph RESOLVE YOUR SEARCH FOR RELIABILITY THE AGILENT 7890B GC RESOLVE YOUR SEARCH FOR THE NEXT EVOLUTION IN GAS CHROMATOGRAPHY Building the world s most trusted GC system is
More informationAppendix A. Drinking Water Standards and Guidelines for Pesticides Sampled in Oregon Ground and Surface Water
Appendix A Drinking Water Standards and Guidelines for Pesticides Sampled in Oregon Ground and Surface Water Jeffrey J Jenkins Department of Environmental and Molecular Toxicology Oregon State University
More information