1 Supporting Information for the Environmental Science and Technology article: Enhanced biodegradation of carbamazepine after UV/H 2 O 2 advanced oxidation Prepared for submission to Environmental Science and Technology September 2011 Olya S. Keen 1, Seungyun Baik 2, Karl G. Linden 1*, Diana S. Aga 2 and Nancy G. Love 3 1 Civil, Environmental and Architectural Engineering, University of Colorado Boulder 2 Chemistry Department, University at Buffalo, The State University of New York, Buffalo, NY 3 Civil and Environmental Engineering Department, University of Michigan Ann Arbor * Corresponding author: karl.linden@colorado.edu 10 pages, including 1 narrative, 1 table and 7 figures S1

2 Text S1: Liquid chromatography/mass spectrometry analysis details Analyses for CBZ and byproducts in samples were performed using liquid chromatographer with ion trap mass spectrometer (LC-ITMS) detection (LCQ Advantage, Thermo Finnigan, CA, USA) equipped with a UV 6000LP UV-vis diode array detector. Analysis was performed using an electrospray ionization (ESI) source in positive ion mode. Data were collected under full scan conditions, from 50 to 500 m/z. Nitrogen gas was used as sheath gas. Electrospray voltage was 4500 V and capillary temperature and voltage were 250 ºC and 41 V, respectively. Separation was achieved using a Betabasic-18 C 18 column, (100 X 2.1 mm i.d., 3 µm particle size), equipped with a guard cartridge (10 X 2.1 mm i.d., 3 µm particle size) (Thermo Fisher Scientific, Waltham, MA). The gradient mobile phase consisted of LC/MSgrade acetonitrile (A, Honeywell B & J, Muskegon, MI) and Nanopure water with 0.3% formic acid (B). The initial conditions were 2% A and 98% B; after 5 min, A was increased to 95% over 15 minutes; this was held for 4 more minutes. The initial mobile phase condition was then restored within 1 min, and run for 5 more min to maintain the condition. The flow rate was set at 200 µl/min, the injection volume was 10 µl, and the column oven temperature was held at 30ºC. The total run time was 30 min, including post-run protocols. For radioactive samples, analysis was performed using Surveyor HPLC (Thermo Finnigan, San Jose, CA) equipped with an online radiochromatographic detector (IN/US Systems, Inc., Tampa, FL), which uses a flow-through cell with a volume of 0.5 ml and a 3:1 scintillation fluid/eluent ratio (Ecoscint, National Diagnostics, Atlanta, GA). For qualitative analysis, 3 ml aliquots were taken from each sample and reduced to 0.2 ml under a gentle stream of nitrogen. Furthermore, selected samples were subjected to a clean-up procedure using solid phase extraction (SPE) to reduce the sample matrix that affects both S2

3 separation and the signals of the analytes in the LC-ITMS. The SPE eluates were reduced in volume to 0.2 ml prior to analysis by LC-radiochromatography, and the results were compared with the samples that were concentrated by direct evaporation (without SPE clean up). For the SPE preparation, Phenomenex Strata-X (Polymeric Reversed Phase, 200 mg, 3 ml) cartridges were pre-conditioned with 3 ml HPLC-grade methanol (Honeywell B & J, Muskegon, MI) and washed with 3 ml Nanopure water. Cartridges were then loaded with 1 ml sample and 2 ml Nanopure water mixture. Cartridges were then washed with 4 ml 5% (v/v) HPLC-grade methanol and the eluate from each cartridge was collected. After 1 min of drying, cartridges were eluted with 4 ml 50:50 of HPLC-grade methanol and acetonitrile (Sigma-Aldrich, St. Louis, MO). At each step in the SPE procedure, a small aliquot was taken from the eluate and from the waste for analysis by LSC in order to check for percent recovery. S3

4 Table S1: Wastewater Treatment Plant Effluent Quality During Two Sampling Events. Water quality parameter Method used Units Effluent 1 Effluent 2 Alkalinity Hach digital titrator mg/l as CaCO ph Beckman ph meter Ammonium Nitrite Nitrate Dissolved organic carbon Total nitrogen Hach colorimetric test kit Hach colorimetric test kit Hach colorimetric test kit Shimadzu TOC analyzer Shimadzu TN analyzer mg/l as N < mg/l as N < < mg/l as N mg/l as C 7.5 -* mg/l as N * * TOC/TN measurement was discarded due to analytical problems that were discovered later. S4

5 (4) (3) (2) (1) Figure S1: Experimental setup for assessing the biological stability of effluents treated with AOP versus those not treated with AOP. The building air delivered to the system passed through a regulator which assured constant flow (1), a 0.2 micron filter removed non-gaseous particles (2), a check valve prevented backflow (3), and a wash bottle hydrated the air to minimize sample evaporation (4). Air was delivered to the samples through a series of manifolds. Each bottle had an individual regulator valve allowing for the gaseous flow rate to be adjusted. S5

6 % Radioactivity % Radioactivity Retention time, min Figure S2: Oxidation products retained after rotary evaporation (top) compared to solid phase extraction (bottom). S6

7 Normalized SUVA without CBZ with CBZ Duration of biodegradation, days Figure S3: Change in SUVA 254 during degradation of effluent with activated sludge, both with and without CBZ. S7

8 Pre-AOP sample Pre-AOP control Post-AOP sample Post-AOP control Figure S4: Sample 1 mass balance. Light gray color fraction of radioactivity due to parent compound and products still in solution. Dark gray color fraction of radioactivity that migrated into the KOH traps as CO 2, i.e. fraction of the radiolabeled compound fully mineralized. The results are for one of the two replicates. Left - samples; right - controls (Sample + NaN 3 ); top before AOP; bottom after AOP. The values are the means of several measurements taken from the same sample with error bars representing 95% confidence intervals for the measurements. S8

9 Pre-AOP sample Pre-AOP control Post-AOP sample Post-AOP control Days of biodgeradation Figure S5: Sample 2 mass balance. Light gray color fraction of radioactivity due to parent compound and products still in solution. Dark gray color fraction of radioactivity that migrated into the KOH traps as CO 2, i.e. fraction of the radiolabeled compound fully mineralized. The results are for the second of the two replicates. Left - samples; right - controls (Sample + NaN 3 ); top before AOP; bottom after AOP. The values are the means of several measurements taken from the same sample with error bars representing 95% confidence intervals for the measurements. S9

10 Intensity Retention time, min Figure S6: UV 254 absorbance chromatogram of the sample pretreated with AOP after 25 days of biodegradation. The main peak at min retention time is the parent CBZ. All other minor peaks were overlapped with their full-scan mass spectra, but none of these peaks showed m/z peak that corresponds to a tricyclic by-product of CBZ devoid of the 14 CONH 2 moiety. Intensity Retention time, min Figure S7: Extracted ion chromatogram for tricyclic fragment of CBZ (m/z 194) of the sample pretreated with AOP after 25 days of biodegradation. S10

Application Note. Abstract. Authors. Introduction. Environmental

Application Note. Abstract. Authors. Introduction. Environmental EPA Method 538: Determination of Selected Organic Contaminants in Drinking Water by Direct Aqueous Injection with the Agilent 646 Triple Quadrupole LC/MS System Application Note Environmental Authors Imma

More information

Robustness of the Agilent Ultivo Triple Quadrupole LC/MS for Routine Analysis in Food Safety

Robustness of the Agilent Ultivo Triple Quadrupole LC/MS for Routine Analysis in Food Safety Technical Overview Robustness of the Agilent Ultivo Triple Quadrupole LC/MS for Routine Analysis in Food Safety Authors Mark Sartain, Theresa Sosienski, and Dan Hui Dorothy Yang Agilent Technologies, Inc.

More information

Supporting Information

Supporting Information Supporting Information A Simple Method to Quantify PC and PET Microplastics in the Environmental Samples by LC-MS/MS Lei Wang *, Junjie Zhang, Shaogang Hou, and Hongwen Sun Ministry of Education Key Laboratory

More information

Using HRAM Survey Analysis Combined with Rapid MS2 Data to Develop a Fragmentation Based Detection Workflow for Structure ID Acquisition

Using HRAM Survey Analysis Combined with Rapid MS2 Data to Develop a Fragmentation Based Detection Workflow for Structure ID Acquisition Using HRAM Survey Analysis Combined with Rapid MS Data to Develop a Fragmentation Based Detection Workflow for Structure ID Acquisition Tim Stratton, 1 Caroline Ding, 1 Christoph Henrich, Hans Grensemann,

More information

Quantitation of cyanotoxins in drinking water according to EPA 544 guidelines

Quantitation of cyanotoxins in drinking water according to EPA 544 guidelines APPLICATION NOTE 64968 Quantitation of cyanotoxins in drinking water according to EPA 544 guidelines Authors Ali Haghani, 1 Andy Eaton, 1 Neloni Wijeratne, Claudia Martins 1 Eurofins Eaton Analytical,

More information

Benefits of 2D-LC/MS/MS in Pharmaceutical Bioanalytics

Benefits of 2D-LC/MS/MS in Pharmaceutical Bioanalytics Application Note Small Molecule Pharmaceuticals Benefits of 2D-LC/MS/MS in Pharmaceutical Bioanalytics Avoiding Matrix Effects Increasing Detection Sensitivity Authors Jonas Dinser Daiichi-Sankyo Europe

More information

Developing Purification Strategies for the Agilent 1260 Infinity II Preparative LC/MSD System

Developing Purification Strategies for the Agilent 1260 Infinity II Preparative LC/MSD System Developing Purification Strategies for the Agilent 126 Infinity II Preparative LC/MSD System Technical Overview Author Florian Rieck Agilent Technologies, Inc. Waldbronn, Germany Abstract Peak-based fraction

More information

Analysis of Illegal Dyes in Food Matrices Using Automated Online Sample Preparation with Liquid Chromatography-Mass Spectrometry

Analysis of Illegal Dyes in Food Matrices Using Automated Online Sample Preparation with Liquid Chromatography-Mass Spectrometry Analysis of Illegal Dyes in Food Matrices Using Automated Online Sample Preparation with Liquid Chromatography-Mass Spectrometry Yang Shi, Catherine Lafontaine, and François A. Espourteille Thermo Fisher

More information

Introduction. Yan Liu, Victor Barreto, Rosanne Slingsby, Chris Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA

Introduction. Yan Liu, Victor Barreto, Rosanne Slingsby, Chris Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA Determination of Trace-Level and in Environmental Samples Using Two-Dimensional Ion Chromatography with Suppressed Conductivity and Mass Spectrometry Detection Yan Liu, Victor Barreto, Rosanne Slingsby,

More information

High-Throughput LC/MS Purification of Pharmaceutical Impurities

High-Throughput LC/MS Purification of Pharmaceutical Impurities High-Throughput LC/MS Purification of Pharmaceutical Impurities Application Note Small Molecule Pharmaceuticals Author Florian Rieck Agilent Technologies, Inc. Waldbronn, Germany Abstract Legal regulations

More information

Strategies for Phospholipid Removal using Polymer-based SPE

Strategies for Phospholipid Removal using Polymer-based SPE Strategies for Phospholipid emoval using Polymer-based SPE Lee Williams, Helen Lodder, Matthew Cleeve, Scott Merriman, Steve Jordan, ichard Calverley, Steve Plant, Joanna Smith Introduction Sample preparation

More information

Improving the Detection Limits for Highly Basic Neuropeptides Using CESI-MS

Improving the Detection Limits for Highly Basic Neuropeptides Using CESI-MS Improving the Detection Limits for Highly Basic Neuropeptides Using CESI-MS Stephen Lock SCIEX UK Overview Who Should Read This: Senior Scientists and Lab Directors with an Interest in Protein Analysis

More information

ENVIRONMENTAL ANALYSIS. Solutions for Your Analytical Business Markets and Applications Programs. Solution Note

ENVIRONMENTAL ANALYSIS. Solutions for Your Analytical Business Markets and Applications Programs. Solution Note ENVIRONMENTAL ANALYSIS A TURN-KEY GUARANTEED ANALYSER FOR THE ROUTINE MEASUREMENT OF GLYPHOSATE AND AMPA IN DRINKING WATER (EUROPEAN DRINKING WATER DIRECTIVE 98/83/EC) AND OTHER ENVIRONMENTAL WATER SAMPLES

More information

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Student Learning Objectives High Performance Liquid Chromatography With the advent of relatively inexpensive and reliable pumps, the development of

More information

Infrastructure. Demonstrating Advanced Oxidation Coupled with Biodegradation for Removal of Carbamazepine. Co-published by

Infrastructure. Demonstrating Advanced Oxidation Coupled with Biodegradation for Removal of Carbamazepine. Co-published by Infrastructure Demonstrating Advanced Oxidation Coupled with Biodegradation for Removal of Carbamazepine Co-published by INFR6SG09 DEMONSTRATING ADVANCED OXIDATION COUPLED WITH BIODEGRADATION FOR REMOVAL

More information

CHROMATOGRAPHY/MASS SPECTROMETRY

CHROMATOGRAPHY/MASS SPECTROMETRY DETERMINATION OF 1,4-DIOXANE IN DRINKING WATER BY SOLID PHASE EXTRACTION (SPE) AND GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS) WITH SELECTED ION MONITORING (SIM)* EPA Method 522 Part Number: EU52112M6

More information

A Novel Screening Method for Anthropogenic Sewage Pollutants in Waste Water, Ground Water and Drinking Water Samples by LC HRAM Analysis

A Novel Screening Method for Anthropogenic Sewage Pollutants in Waste Water, Ground Water and Drinking Water Samples by LC HRAM Analysis A Novel Screening Method for Anthropogenic Sewage Pollutants in Waste Water, Ground Water and Drinking Water Samples by LC HRAM Analysis Sebastian Westrup, Nick Duczak, and Michal Godula Thermo Fisher

More information

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B Stephan Meding, Aran Paulus, and Remco Swart ¹Thermo Fisher Scientific, Germering,

More information

Analysis of Nucleosides Using an Agilent Infinity II High Speed UHPLC with the 6130 Single Quadrupole Mass Selective Detector

Analysis of Nucleosides Using an Agilent Infinity II High Speed UHPLC with the 6130 Single Quadrupole Mass Selective Detector Analysis of Nucleosides Using an Agilent Infinity II High Speed UHPLC with the 6130 Single Quadrupole Mass Selective Detector Application Note Clinical Research Author Patrick Cronan Agilent Technologies,

More information

Screening for Emerging Chemical Contaminants in Water Using LC/Q-TOF and Mass Profiler Professional Software

Screening for Emerging Chemical Contaminants in Water Using LC/Q-TOF and Mass Profiler Professional Software Screening for Emerging Chemical Contaminants in Water Using LC/Q-TOF and Mass Profiler Professional Software Application Note Environmental Authors Sylvain Merel and Shane Snyder Department of Chemical

More information

for Acclaim Carbamate Column

for Acclaim Carbamate Column for Acclaim Carbamate Column Product Manual for Acclaim Carbamate Page 1 of 13 Product Manual for Acclaim Carbamate Analytical Columns 5 µm, 4.6 x 250 mm, P/N 072924 3 µm, 4.6 x 150 mm, P/N 072925 3 µm,

More information

Analysis of the Massachusetts Cannabis Pesticides List Using the SCIEX QTRAP System

Analysis of the Massachusetts Cannabis Pesticides List Using the SCIEX QTRAP System Analysis of the Massachusetts Cannabis Pesticides List Using the SCIEX QTRAP 6500+ System Evoking the power of MS3 to reduce matrix background in cannabis extracts Craig M. Butt 1, Robert Di Lorenzo 2,

More information

Improving Sensitivity for an Immunocapture LC-MS Assay of Infliximab in Rat Plasma Using Trap-and-Elute MicroLC-MS

Improving Sensitivity for an Immunocapture LC-MS Assay of Infliximab in Rat Plasma Using Trap-and-Elute MicroLC-MS Improving Sensitivity for an Immunocapture LC-MS Assay of in Rat Plasma Using Trap-and-Elute MicroLC-MS Using the SCIEX M3 MicroLC system for Increased Sensitivity in Antibody Quantitation Remco van Soest

More information

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column Rapid UHP Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 3SB-C8 Column Application Note BioPharma Authors James Martosella, Phu Duong, Susanne

More information

A Simple Rapid and Sensitive Method Development for Quantification of Quetiapine Fumarate in Bulk and Dosage Forms Using RP-HPLC

A Simple Rapid and Sensitive Method Development for Quantification of Quetiapine Fumarate in Bulk and Dosage Forms Using RP-HPLC Human Journals Research Article February 2018 Vol.:11, Issue:3 All rights are reserved by Priyanka Teepoju et al. A Simple Rapid and Sensitive Method Development for Quantification of Quetiapine Fumarate

More information

A Generic Approach to the Extraction of Multi-functional Drugs using Mixed-mode SPE with LC-MS/MS Analysis

A Generic Approach to the Extraction of Multi-functional Drugs using Mixed-mode SPE with LC-MS/MS Analysis A Generic Approach to the Extraction of Multi-functional Drugs using Mixed-mode SPE with LC-MS/MS Analysis Matthew Cleeve, Scott Merriman, Lee Williams, Steve Jordan, Richard Calverley, Joanna Smith &

More information

A highly sensitive and robust 150 µm column to enable high-throughput proteomics

A highly sensitive and robust 150 µm column to enable high-throughput proteomics APPLICATION NOTE 21744 Robust LC Separation Optimized MS Acquisition Comprehensive Data Informatics A highly sensitive and robust 15 µm column to enable high-throughput proteomics Authors Xin Zhang, 1

More information

Performance characteristics of the 1260 Infinity Quaternary LC system

Performance characteristics of the 1260 Infinity Quaternary LC system Performance characteristics of the 1260 Infinity Quaternary LC system The new standard in HPLC Technical Overview Introduction The Agilent 1260 Infinity LC system consists of modular units that operate

More information

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany New Features of the Agilent Method Scouting Wizard for Automated Method Development of Complex Samples Analysis of Large Data Sets by Method Scouting Reports and Automated Adjustment of Flow Rates and

More information

Determination of Table 3 Compounds by LC/MS/MS Chemours Fluoroproducts Analytical Method Revision date: 4/26/2018

Determination of Table 3 Compounds by LC/MS/MS Chemours Fluoroproducts Analytical Method Revision date: 4/26/2018 Determination of Table 3 Compounds by LC/MS/MS Chemours Fluoroproducts Analytical Method Revision date: 4/26/2018 Instrument Setup (Note: Trapping column installed on Line A. Delay column installed between

More information

Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS

Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS Using the SCIEX M3 MicroLC system for Increased Sensitivity in Antibody Quantitation Remco van Soest and Lei Xiong SCIEX, Redwood City,

More information

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions Screening and Confirmation Chemical Analysis Solution Unit November, 2009 Page 1 Agilent s New 6540 Ultra High

More information

Application Note. Abstract. Authors. Food Safety

Application Note. Abstract. Authors. Food Safety Determination of Chloramphenicol, Florfenicol, and Thiamphenicol in Honey Using Agilent SampliQ PT Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry Application Note

More information

Extraction of Vitamin B7 (Biotin) from Serum Using EVOLUTE

Extraction of Vitamin B7 (Biotin) from Serum Using EVOLUTE Application Note AN880 Extraction of Vitamin B7 (Biotin) from Serum Using EVOLUTE EXPRESS ABN Prior to LC-MS/MS Analysis Page 1 Extraction of Vitamin B7 (Biotin) from Serum Using EVOLUTE EXPRESS ABN Prior

More information

On-Line. User s Guide SPE CARTRIDGES. for Rapid Cleanup and Extraction of Analytes

On-Line. User s Guide SPE CARTRIDGES. for Rapid Cleanup and Extraction of Analytes On-Line SPE CARTRIDGES for Rapid Cleanup and Extraction of Analytes User s Guide What is the strata-x on-line SPE cartridge? The strata-x on-line cartridge combines the revolutionary benefits of the patent

More information

MS Certified Vials. Pre-cleaned and certified vials for mass spectrometry

MS Certified Vials. Pre-cleaned and certified vials for mass spectrometry Pre-cleaned and certified vials for mass spectrometry The FIRST and ONLY pre-cleaned, low particle, low background chromatography vial When your instrumentation, sample handling and methodology is pushing

More information

Ultrapure water: LC-MS suitability tests

Ultrapure water: LC-MS suitability tests Ultrapure water: LC-MS suitability tests Application details Contaminations such as ions and organic traces present in ultrapure water affect LC- MS analyses and decrease perfomance of the analytical instumentation.

More information

Illicit Drug Analysis in Urine Using 2D LC-MS/MS for Forensic Toxicology

Illicit Drug Analysis in Urine Using 2D LC-MS/MS for Forensic Toxicology Claude Mallet 1 and Sabra Botch-Jones 2 1 Waters Corporation, Milford, MA, USA 2 Boston University School of Medicine, Boston, MA, USA APPLICATION BENEFITS Fast extraction protocol (2 min) 1 ppt detection

More information

Chemicals Branch United Nations Environment Programme (UNEP) Division of Technology, Industry and Economics. Geneva

Chemicals Branch United Nations Environment Programme (UNEP) Division of Technology, Industry and Economics. Geneva Procedure for the Analysis of Persistent Organic Pollutants in Environmental and Human Matrices to Implement the Global Monitoring Plan under the Stockholm Convention Protocol 1: The Analysis of Perfluorooctane

More information

Techniques for Reducing the Effects of Sample Solvent on UHPLC Analyses

Techniques for Reducing the Effects of Sample Solvent on UHPLC Analyses PO-CON1404E Techniques for Reducing the Effects of Sample Solvent Pittcon 2014 2050-11P Kenichiro Tanaka, William Hedgepeth Shimadzu Scientific Instruments, Inc., Columbia MD Introduction For HPLC, sample

More information

Evaluation by Competent Authorities

Evaluation by Competent Authorities LANXESS Deutschland GmbH Chlorophene 07/2007 ACTIVE SUBSTANCE RESIDUES IN ANIMAL AND HUMAN BODY FLUIDS AND TISSUES JUSTIFICATION FOR NON-SUBMISSION OF DATA Official use only Other existing data [ ] Technically

More information

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study Application note Biotherapeutics and Biologics Authors Suresh Babu C.V., Anne

More information

LC/MS/MS Determination of PFOS and PFOA in Water and Soil Matrices

LC/MS/MS Determination of PFOS and PFOA in Water and Soil Matrices Application Note Environmental LC/MS/MS Determination of PFOS and PFOA in Water and Soil Matrices Using an Agilent 19 Infinity II LC with Ultivo Tandem Mass Spectrometry Authors Wenlong Yang, Jing Guo,

More information

On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water

On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water application note On-line SPE-LC/MS/MS of Pesticides On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water API 3200 LC/MS/MS System Overview The feasibility of

More information

Supporting Information I

Supporting Information I Single-shot top-down proteomics with capillary zone electrophoresiselectrospray ionization-tandem mass spectrometry for identification of nearly 600 Escherichia coli proteoforms Rachele A. Lubeckyj 1,,

More information

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808 Reversed-Phase Optimization for Ultra Fast Profiling of Intact and Reduced Monoclonal Antibodies using Agilent ZORBAX Rapid Resolution High Definition 3SB-C3 Column Application Note Biopharma Authors James

More information

Determination of Total Nitrogen and Phosphorus in Environmental Waters

Determination of Total Nitrogen and Phosphorus in Environmental Waters Determination of Total Nitrogen and Phosphorus in Environmental Waters Brian De Borba, Jeffrey Rohrer, and Richard Jack Thermo Fisher Scientific, Sunnyvale, CA, USA PITTCON Conference & Expo March 4, 2014

More information

Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma

Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma Yun Wang Alelyunas, Mark D. Wrona, Jim Murphy, Angela Doneanu, Gregory

More information

Supplementary Information. Active Sampling Device for Determining Pollutants in Surface and Pore Water the In

Supplementary Information. Active Sampling Device for Determining Pollutants in Surface and Pore Water the In 1 2 Supplementary Information 3 4 Active Sampling Device for Determining Pollutants in Surface and Pore Water the In Situ Sampler for Biphasic Water Monitoring 5 6 Samuel D. Supowit, Isaac B. Roll, Viet

More information

Genotoxicity is the property of a compound

Genotoxicity is the property of a compound Impurities Analysis in Pharmaceuticals: Genotoxicity is the property of a compound known to have irreversible effects on the structure and functionality of the DNA in cells and cause DNA loss, DNA replication

More information

Emulation of the Agilent 1100 Series LC Through Waters Empower Software Analysis of an Analgesic Mixture

Emulation of the Agilent 1100 Series LC Through Waters Empower Software Analysis of an Analgesic Mixture Agilent 129 Infinity II LC with ISET Emulation of the Agilent 11 Series LC Through Waters Empower Software Analysis of an Analgesic Mixture Application Note Small Molecule Pharmaceuticals Author Melanie

More information

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution Clone Selection Using the Agilent 9 Infinity Online D-LC/MS Solution Application Note Biopharmaceuticals Authors Abstract Ravindra Gudihal and This Application Note describes the Agilent 9 Infinity online

More information

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography Characterize Fab and Fc Fragments by Cation-Exchange Chromatography Application Note Biologics and Biosimilars Authors Isabel Vandenheede, Emmie Dumont, Pat Sandra, and Koen Sandra Research Institute for

More information

Supporting Information

Supporting Information Supporting Information Wiley-VCH 2005 69451 Weinheim, Germany Molecular and polymeric hybrids based on covalently linked polyoxometalates and transition metal complexes ** Jeonghee Kang, Bubin Xu, Zhonghua

More information

Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental

Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental Samples Michael S.Young, Jeremy C. Shia, Kim vantran, Kevin M. Jenkins and Masayo Yabo Waters Corporation 34 Maple Street,

More information

Rapid Determination of Drug Protein Binding Affinity using Solid Phase Microextraction

Rapid Determination of Drug Protein Binding Affinity using Solid Phase Microextraction Rapid Determination of Drug Protein Binding Affinity using Solid Phase Microextraction Len Sidisky, Craig Aurand, Emily Barrey, Candace Price, and David Bell MilliporeSigma, Bellefonte, PA, USA T416070

More information

A Highly Accurate Mass Profiling Approach to Protein Biomarker Discovery Using HPLC-Chip/ MS-Enabled ESI-TOF MS

A Highly Accurate Mass Profiling Approach to Protein Biomarker Discovery Using HPLC-Chip/ MS-Enabled ESI-TOF MS Application Note PROTEOMICS METABOLOMICS GENOMICS INFORMATICS GLYILEVALCYSGLUGLNALASERLEUASPARG CYSVALLYSPROLYSPHETYRTHRLEUHISLYS A Highly Accurate Mass Profiling Approach to Protein Biomarker Discovery

More information

Automated Scouting of Stationary and Mobile Phases Using the Agilent 1290 Infinity II Method Development Solution

Automated Scouting of Stationary and Mobile Phases Using the Agilent 1290 Infinity II Method Development Solution Automated Scouting of Stationary and Mobile Phases Using the Agilent 129 Infinity II Method Development Solution Technical Overview Authors Edgar Naegele and Sonja Schneider Agilent Technologies, Inc.

More information

Authors. Introduction. Experimental. Environmental Analysis

Authors. Introduction. Experimental. Environmental Analysis Determination of Phenyl Urea and Triazine Herbicides in Potable and Groundwater by LC/MS Using API-ESI Selective Ion Monitoring and Direct Large-Volume Aqueous Injection Application Environmental Analysis

More information

CHAPTER 1 : INTRODUCTION. Bisphenol A (BPA) had been used in the plastics industry since 1940s and 50s.

CHAPTER 1 : INTRODUCTION. Bisphenol A (BPA) had been used in the plastics industry since 1940s and 50s. CHAPTER 1 : INTRODUCTION 1.0 Introduction 1.1 Bisphenol A (BPA) Bisphenol A (BPA) had been used in the plastics industry since 1940s and 50s. It is a building block of polycarbonate plastics often used

More information

HepG2 Labeling and Cell Growth Protocol for IROA Phenotypic Metabolic Profiling

HepG2 Labeling and Cell Growth Protocol for IROA Phenotypic Metabolic Profiling HepG2 Labeling and Cell Growth Protocol for IROA Phenotypic Metabolic Profiling Candice Z. Ulmer 1, Felice A. de Jong 2, Chris Beecher 1, Timothy J. Garrett 3, Richard A. Yost 1,3 1 Department of Chemistry,

More information

Laboratory Water Quality Affects Protein Separation by 2D Gel Electrophoresis

Laboratory Water Quality Affects Protein Separation by 2D Gel Electrophoresis Laboratory Water Quality Affects Protein Separation by 2D Gel Electrophoresis 2D gel electrophoresis remains a dominant technique in proteomics. Obtaining high quality gels requires careful and tedious

More information

Minimizing Sample Carryover Using the Multiwash Function of the Agilent 1290 Infinity II Multisampler

Minimizing Sample Carryover Using the Multiwash Function of the Agilent 1290 Infinity II Multisampler Minimizing Sample Carryover Using the Multiwash Function of the Agilent 1290 Infinity II Multisampler Technical Overview Author Bettina Schuhn Agilent Technologies, Inc. Waldbronn, Germany Abstract The

More information

Biotherapeutic Method Development Guide

Biotherapeutic Method Development Guide Biotherapeutic Method Development Guide HALO BIOCLASS 1000 Å PROTEIN SELECTIVITY KIT Fused-Core Particle Technology BIOCLASS Strategy for Optimizing Protein Separations Using Reversed-phase Liquid Chromatography

More information

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube Application Note Environmental Authors Don Noot, Ralph Hindle Vogon Laboratory Services Cochrane, AB

More information

Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass

Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass Application Note: 529 Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass Xiang He, Marta Kozak; Thermo Fisher Scientific, San Jose,

More information

Agilent 1290 Infinity 2D-LC Solution. Selectable separation performance

Agilent 1290 Infinity 2D-LC Solution. Selectable separation performance Agilent 129 Infinity 2D-LC Solution Selectable separation performance AGILENT 129 INFINITY 2D-LC SOLUTION SELECTABLE SEPARATION PERFORMANCE The new Agilent 129 Infinity 2D-LC solution lets you match separation

More information

for water and beverage analysis

for water and beverage analysis Thermo Scientific EQuan MAX Plus Systems Automated, high-throughput LC-MS solutions for water and beverage analysis Pesticides Pharmaceuticals Personal care products Endocrine disruptors Perfluorinated

More information

Parallel LC with Capillary PS-DVB Monolithic Columns for High-Throughput Proteomics

Parallel LC with Capillary PS-DVB Monolithic Columns for High-Throughput Proteomics Parallel LC with Capillary PS-DVB Monolithic Columns for High-Throughput Proteomics INTRODUCTION Peptide sequencing by nano LC-ESI-MS/MS is a widespread technique used for protein identification in proteomics.

More information

Biotherapeutic Method Development Guide

Biotherapeutic Method Development Guide Biotherapeutic Method Development Guide HALO BIOCLASS 1000 Å PROTEIN SELECTIVITY KIT Fused-Core Particle Technology BIOCLASS Strategy for Optimizing Protein Separations Using Reversed-phase Liquid Chromatography

More information

Fast Analysis of Environmental Phenols with Agilent Poroshell 120 EC-C18 Columns

Fast Analysis of Environmental Phenols with Agilent Poroshell 120 EC-C18 Columns Fast Analysis of Environmental Phenols with Agilent Poroshell 120 EC-C18 Columns Application Note Environmental Authors William J. Long and Anne E. Mack Agilent Technologies, Inc. 2850 Centerville Road

More information

Comprehensive Single-Shot Proteomics with FAIMS on a Hybrid Orbitrap Mass Spectrometer

Comprehensive Single-Shot Proteomics with FAIMS on a Hybrid Orbitrap Mass Spectrometer Supporting Information Comprehensive Single-Shot Proteomics with FAIMS on a Hybrid Orbitrap Mass Spectrometer Alexander S Hebert 1,2, Satendra Prasad 3, Michael W Belford 3, Derek J Bailey 3, Graeme C

More information

Increased Throughput in the Determination of PPCPs in Water Using Optimized MS Cycle Times in a High Sensitivity UHPLC-Triple Quadrupole System

Increased Throughput in the Determination of PPCPs in Water Using Optimized MS Cycle Times in a High Sensitivity UHPLC-Triple Quadrupole System Increased Throughput in the Determination of PPCPs in Water Using Optimized MS Cycle Times in a High Sensitivity UHPLC-Triple Quadrupole System Application Note Environmental Authors Behrooz Zekavat, Thomas

More information

Key Words Pesticides analysis, food safety, TSQ Endura, TraceFinder, MRL, tsrm, residue analysis

Key Words Pesticides analysis, food safety, TSQ Endura, TraceFinder, MRL, tsrm, residue analysis Increased Productivity in Pesticide Residue nalysis Quantifying 440 Pesticides Following China GB 2763-2014: The Pesticide Explorer Collection Standard Quantitation Zeming Wu 1, Charles T. Yang 2, Zheng

More information

Monoclonal Antibody Analysis on a Reversed-Phase C4 Polymer Monolith Column

Monoclonal Antibody Analysis on a Reversed-Phase C4 Polymer Monolith Column Monoclonal Antibody Analysis on a Reversed-Phase C4 Polymer Monolith Column Shane Bechler 1, Ken Cook 2, and Kelly Flook 1 1 Thermo Fisher Scientific, Sunnyvale, CA, USA; 2 Thermo Fisher Scientific, Runcorn,

More information

Final Report. Characterisation of Sample Report. Job No 2016/11/12-34 AS No. 1234A. Client Example Contact Sample. Signed. Example report.

Final Report. Characterisation of Sample Report. Job No 2016/11/12-34 AS No. 1234A. Client Example Contact Sample. Signed. Example report. Final Report Title Characterisation of Job No 2016/11/12-34 AS No. 1234A Client Contact Sample Author report Signed Date 25 th November 2016 Easy Reach Report 20161124 1 of 30 Contents 1. Study Summary

More information

Mass Spectrometry Analysis of Liquid Chromatography Fractions using Ettan LC MS System

Mass Spectrometry Analysis of Liquid Chromatography Fractions using Ettan LC MS System GUIDE TO LC MS - December 21 1 Spectrometry Analysis of Liquid Chromatography Fractions using Ettan LC MS System Henrik Wadensten, Inger Salomonsson, Staffan Lindqvist, Staffan Renlund, Amersham Biosciences,

More information

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor 229 10. Validated Normal Phase HPLC Method for the Determination of Fulvestrant in Pharmaceutical Dosage Forms 10.1 Introduction Fulvestrant is primarily used in the treatment of hormone receptor positive

More information

Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D*

Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D* Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D* UCT Products: ECUNIPAH (ENVIRO-CLEAN Universal PAH/DRO 2000 mg/83 ml cartridge) Method Summary Method 8015D may be used

More information

Automated alternating column regeneration on the Agilent 1290 Infinity LC

Automated alternating column regeneration on the Agilent 1290 Infinity LC Automated alternating column regeneration on the Agilent 1290 Infinity LC Increasing throughput using two columns alternatively via an ultra-high pressure 2-position/10-port valve Application Note Environmental

More information

EFFECTIVE SPE STRATEGIES FOR LC/MS DETERMINATION OF SUDAN DYES IN CHILI PRODUCTS

EFFECTIVE SPE STRATEGIES FOR LC/MS DETERMINATION OF SUDAN DYES IN CHILI PRODUCTS EFFECTIVE SPE STRATEGIES FOR LC/MS DETERMINATION OF SUDAN DYES IN CHILI PRODUCTS KimVan Tran, Michael S. Young a Uwe Neue Waters Corporation, Milford, MA, USA. INTRODUCTION Recently there have been numerous

More information

DIDANOSINE ORAL POWDER Final text for addition to The International Pharmacopoeia

DIDANOSINE ORAL POWDER Final text for addition to The International Pharmacopoeia Document QAS/06.175(A)FINAL August 2007 DIDANOSINE ORAL POWDER Final text for addition to The International Pharmacopoeia This monograph was adopted at the 41 st WHO Expert Committee on Specifications

More information

Characterize mab Charge Variants by Cation-Exchange Chromatography

Characterize mab Charge Variants by Cation-Exchange Chromatography Characterize mab Charge Variants by Cation-Exchange Chromatography Application Note Biologics and Biosimilars Authors Isabel Vandenheede, Emmie Dumont, Pat Sandra, and Koen Sandra Research Institute for

More information

Characterization of a fusion protein under native and denaturing conditions with maxis II

Characterization of a fusion protein under native and denaturing conditions with maxis II Characterization of a fusion protein under native and denaturing conditions with maxis II Abstract The maxis II UHR-QTOF was used to characterize key attributes of a fusion protein in its native conformation

More information

Title: Non-absorbable apple procyanidins prevent obesity associated with gut microbial

Title: Non-absorbable apple procyanidins prevent obesity associated with gut microbial 1 2 3 SUPPORTING INFORMATION Title: Non-absorbable apple procyanidins prevent obesity associated with gut microbial and metabolomic changes 4 5 6 Authors: Saeko Masumoto, 1 Akari Terao, 2 Yuji Yamamoto,

More information

EPA Method 540: Selected Organic Contaminants Using Agilent Plexa Cartridges and the Agilent 6460 Triple Quadrupole LC/MS

EPA Method 540: Selected Organic Contaminants Using Agilent Plexa Cartridges and the Agilent 6460 Triple Quadrupole LC/MS EPA Method 540: Selected Organic Contaminants Using Agilent Plexa Cartridges and the Agilent 6460 Triple Quadrupole LC/MS Application Note Environmental Authors Don Noot and Ralph Hindle Vogon Laboratory

More information

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column Disulfide Linkage Analysis of IgG1 using an Agilent 126 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column Application Note Biotherapeutics & Biosimilars Author M. Sundaram

More information

Application Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany Quantification of trace-level herbicides in drinking water by online enrichment with the Agilent Infinity Series Online-SPE Solution and Triple Quadrupole MS Detection Application Note Environmental Author

More information

Method Development of a 2D LC-HRMS Extraction and Detection Method for Organophosphorus Flame Retardants in Environmental Water Samples

Method Development of a 2D LC-HRMS Extraction and Detection Method for Organophosphorus Flame Retardants in Environmental Water Samples Method Development of a 2D LC-HRMS Extraction and Detection Method for Organophosphorus Flame Retardants in Environmental Water Samples Ken Rosnack 1 Lauren Mullin 1,2, Malorie Mella 1, Claude Mallet 1,

More information

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography Carl Fisher, Monika Verma, and Pranathi Perati; Thermo Fisher Scientific,

More information

MassPREP On-Line Desalting Cartridge

MassPREP On-Line Desalting Cartridge CONTENTS I. INTRODUCTION II. INSTALLING THE MASSPREP ON-LINE DESALTING CARTRIDGE INTO THE SENTRY.1 X 1 MM GUARD COLUMN HOLDER III. RECOMMENDED LC/MS SYSTEM CONFIGURATION TO MINIMIZE MS SOURCE CONTAMINATION

More information

Preparative HPLC is still the

Preparative HPLC is still the B I O P R O C E S S TECHNICAL Optimizing Sample Load Capacity and Separation Through a Series of Short Prep Columns Mark Crawford, Joan Stevens, and Luke Roenneburg Preparative HPLC is still the dominate

More information

Analysis of Trace Organic Contaminants in Water by Direct Injection Using Agilent 6490 LC/MS/MS with Pos/Neg Switching

Analysis of Trace Organic Contaminants in Water by Direct Injection Using Agilent 6490 LC/MS/MS with Pos/Neg Switching Analysis of Trace Organic Contaminants in Water by Direct Injection Using Agilent 6490 LC/MS/MS with Pos/Neg Switching Application Note Environmental Authors Tarun Anumol, Shimin Wu, Sylvain Merel, and

More information

Product Guide for LudgerSep TM N2 High Resolution Amide HPLC Columns for Glycan Analysis

Product Guide for LudgerSep TM N2 High Resolution Amide HPLC Columns for Glycan Analysis Product Guide for LudgerSep TM N2 High Resolution Amide HPLC Columns for Glycan Analysis (Ludger Product Codes: LS-N2-4.6x150 and LS-N2-2.0x150) Ludger Document # LS-N2-Guide-v4.0 Ludger Ltd Culham Science

More information

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System Application Note Biologics & Biosimilars Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction Chromatography Agilent 126 Infinity II Bio-Inert System 7 6 5 4 5. 7.5

More information

Method Optimisation in Bottom-Up Analysis of Proteins

Method Optimisation in Bottom-Up Analysis of Proteins Method Optimisation in Bottom-Up Analysis of Proteins M. Styles, 1 D. Smith, 2 J. Griffiths, 2 L. Pereira, 3 T. Edge 3 1 University of Manchester, UK; 2 Patterson Institute, Manchester, UK; 3 Thermo Fisher

More information

Proteomics Training Kit

Proteomics Training Kit CONTENTS I. INTRODUCTION II. STANDARD PREPARATION III. PROTEIN EXPRESSION: MASSPREP MIX 1 VS. MIX 2 (75 µm) IV. PROTEIN EXPRESSION: MASSPREP MIX 1 VS. MIX 2 (300 µm) V. ORDERING INFORMATION I. INTRODUCTION

More information

Introduction Ron Majors is a Senior Scientist at Agilent. Bill Champion is a chemist in Agilent s HPLC Columns tech support group.

Introduction Ron Majors is a Senior Scientist at Agilent. Bill Champion is a chemist in Agilent s HPLC Columns tech support group. Video Notes LC Troubleshooting Series Ghost Peaks Introduction Ron Majors is a Senior Scientist at Agilent. Bill Champion is a chemist in Agilent s HPLC Columns tech support group. Ghost peaks can come

More information

Quantification of genotoxic "Impurity D" in Atenolol by LC/ESI/MS/MS with Agilent 1200 Series RRLC and 6410B Triple Quadrupole LC/MS

Quantification of genotoxic Impurity D in Atenolol by LC/ESI/MS/MS with Agilent 1200 Series RRLC and 6410B Triple Quadrupole LC/MS Quantification of genotoxic "Impurity D" in Atenolol by LC/ESI/MS/MS with Agilent 12 Series RRLC and 641B Triple Quadrupole LC/MS Application Note Manufacturing Process Development Author Siji Joseph Agilent

More information

Optimizing the Purification of a Standard Chiral Compound Utilizing a Benchtop, Multi-Purpose, Semi-Preparative to Preparative HPLC System

Optimizing the Purification of a Standard Chiral Compound Utilizing a Benchtop, Multi-Purpose, Semi-Preparative to Preparative HPLC System Optimizing the Purification of a Chiral Compound Utilizing a Benchtop, Multi-Purpose, Semi-Preparative to Preparative HPLC System Application Note PHA0111 Keywords Chiral, Enantiomer, HPLC, Normal Phase

More information