Environmental Express. Joe Boyd. Solids and BOD Analyses: Pitfalls and Troubleshooting

Size: px
Start display at page:

Download "Environmental Express. Joe Boyd. Solids and BOD Analyses: Pitfalls and Troubleshooting"

Transcription

1 Environmental Express Joe Boyd Solids and BOD Analyses: Pitfalls and Troubleshooting

2 Solids Analysis Total Dissolved Solids SM 2540C, ASTM D , USGS I Total Suspended Solids SM 2540D, USGS I Settleable Solids SM 2540F Total Solids SM 2540B

3 Total Dissolved Solids (TDS) Method Summary: An aliquot of sample is filtered through a glass fiber disk. The filtrate is collected in a pre-weighed evaporating dish. The liquid is evaporated at 104 C and the dish then placed in an oven at 180 C. The dish and residue are weighed with the mass of the residue determined by difference.

4 Total Dissolved Solids (TDS) Equipment Specification Filter Evaporating dish Binderless Glass Fiber, 1.5µm pore size Platinum, porcelain, or silica glass; 100mL vol. Vacuum filtration device Variety of options Oven 104 C and 180 C Weight stability +/- 0.5mg of previous weight

5 Total Dissolved Solids (TDS) Potential problems with evaporation dish The lighter the better. Platinum is the lightest material but the most expensive. Typical weight of porcelain dish is ~ 80g. Porcelain is porous. High humidity can affect weight stability. Allow the same cooling / desiccating time each time the sample is weighed to achieve consistent results. A 1-hour wait time in the desiccator is generally sufficient. Obtain the initial weight of the dish immediately prior to use.

6 Total Dissolved Solids (TDS) Why two temperatures and why are they important? The initial evaporation step at 104 C allows sample evaporation without any splattering. Drying at 180 C drives off all mechanically occluded water and most of the water of crystallization. Samples of high sulfur content may retain more water than normal The 180 C temperature also drives off carbon dioxide and converts bicarbonates to carbonates. Too high of a temp may cause further conversion to oxides. Some organic matter may be lost at 180 C but the temperature is a good compromise between complete water removal and minimal volatile loss.

7 Total Dissolved Solids (TDS) What is the right sample volume for this test? Standard Methods states Choose sample volume to yield between 2.5 and 200 mg dried residue. The lower limit is 5x the allowed variance between weights giving sufficient weight differences between initial and final weights. The upper limit allows for a timely and complete removal of water from the crystals in the sample. Amounts greater than 200mg may layer in the dish trapping water. Typically 100mL is a good volume to start with but the method does not dictate what you must use.

8 Total Suspended Solids (TSS) Method Summary An aliquot of sample (no more than 1-L) is passed through a pre-washed, pre-weighed filter. The filter is then dried at 104 C. After drying the filter is re-weighed and the TSS is calculated.

9 Total Suspended Solids (TSS) Equipment Specification Filter Binderless Glass Fiber, 1.5µm pore size, prewashed and weighed Vacuum filtration device Variety of options Oven 104 C Weight stability of filter +/- 0.5mg of previous weight

10 Total Suspended Solids (TSS) Potential problems with the filter paper Preparation is the key Wash, dry, desiccate, weigh. Repeat dry through weigh steps until a weight-stable filter is achieved. Prepared filters such as the ProWeigh can solve filter preparation problems. New documentation for DoubleWeigh filters. Shows weight stability of filter for audit trail. Many auditors and auditing bodies now requiring this type of documentation.

11 Total Suspended Solids (TSS) Date/Time Box ID Identifier Weight 1 Weight 2 Lot # Sep :38:26:743AM a00000 D Sep :38:34:493AM a00000 D Sep :38:43:243AM a00000 D Sep :39:12:760AM a00000 D Sep :39:21:260AM a00000 D Sep :39:50:777AM a00000 D Sep :39:59:277AM a00000 D Sep :40:29:793AM a00000 D Sep :40:38:543AM a00000 D Sep :41:10:560AM a00000 D

12 Total Suspended Solids (TSS) Filter Paper Continued Consistency in pore size 1.5µm is the industry standard pore size for filters used for TSS analysis. Once you analyze a sample for TSS using a filter with a particular pore size you MUST stick with that size. What filter has the edge? Frayed edges can cause loss of filter material during analysis resulting in biased results.

13 Total Suspended Solids (TSS) Filter Paper Continued Filters that stick to the pan indicate a high organic content. Wash the filter well with DI Water after sample filtration to remove any residue that is stuck to the bottom of the filter. Orientation of the filter There is an up and down side to the filters. Ensure they are oriented correctly prior to filtration.

14 Total Suspended Solids (TSS) Shake it up! The key to obtaining a representative aliquot of sample is to thoroughly mix the sample. Constant sample movement helps. Look for fine silt stuck to the bottom of containers. Remove debris that would not be considered part of a typical water sample such as sticks and leaves.

15 Total Suspended Solids (TSS) What is the right sample volume for this test? Try to limit the sample size to yield no more than 200mg of residue. Experience with the sample is your best guidance. Typical sample volumes are: Mixed Liquor: 1 10 ml Influents: ml Effluents: ml Storm water: ml Unlike TDS, Standard Methods dictates a sample volume of 1-L as the largest volume one can use.

16 Biochemical Oxygen Demand (BOD) Standard Methods 5210B A known volume of sample has its initial dissolved oxygen content recorded. After a five day incubation period at 20 C the sample is removed from the incubator and the final dissolved oxygen content is taken. The BOD value is calculated from the depletion and the amount of sample used.

17 Biochemical Oxygen Demand (BOD) Equipment / Reagents BOD bottles Incubator Dilution water Seed Standard DO Meter Specifications Hold 300mL, seal tight, no oxygen migration Hold temp of 20⁰C Contains the 4 required nutrients and is aerated Viable and active GGA with TV of 198mg/L Membrane electrode or optical luminescence

18 Biochemical Oxygen Demand (BOD) Bottles Glass or Plastic? Glass is traditional but can crack, chip, break, and are a pain to clean. If not cleaned properly can lead to poor results. Plastic bottles are designed for 1-time use no washing! Lighter, can fit more in an incubator. EPA approves both styles.

19 Biochemical Oxygen Demand (BOD) Dilution Water Choose your water type carefully. Distilled water is not always the best choice as certain metals and volatile organics may distill over. Deionized water is a popular choice but proper maintenance of your resin tanks is key. The oxygen depletion after 5 days must be < 0.2mg/L.

20 Biochemical Oxygen Demand (BOD) Seed Choosing your seed source can make or break your test. The safest route is generally to purchase a commercially available freeze dried seed. These are generally very consistent from one batch to the next and provide predictable results. If you have a source of seed material (i.e. wastewater treatment plant), research where you take your seed from. Should be a consistent part of the plant (NOT influent). Ahead of any disinfection area. Primary or Secondary Effluent are generally good places. A good Seed Correction Factor (SCF) should be between mg/l.

21 Biochemical Oxygen Demand (BOD) Seed Continued Too high of a SCF Check your water quality to ensure no contamination. Check other labware such as bottles, pipettes, tubing, etc for cleanliness. If you are providing your own seed look for changes in the seed source that could affect the strength of the seed and make adjustments. If the GGA is within range increase the amount of rehydration water used in the commercial freeze dried seed or check your seed source and potentially dilute it with DI Water. This process is approximately linear (especially with the commercial seeds) so if you need to reduce your SCF by 20% increase the rehydration water by 20%.

22 Biochemical Oxygen Demand (BOD) Seed Continued Too low of a SCF Check your water quality to ensure no contamination. Check other labware such as bottles, pipettes, tubing, etc for cleanliness. If you are providing your own seed look for changes in the seed source that could affect the strength of the seed and make adjustments. If purchasing a commercial freeze dried seed reduce the amount of rehydration water by the same factor you want to increase your SCF.

23 Biochemical Oxygen Demand (BOD) Standard The standard used for BOD analysis is Glucose / Glutamic Acid, better known as GGA. This is made by adding 150mg of glucose and 150mg of glutamic acid to distilled or DI Water and diluting to 1-L. The true value of the GGA standard is 198 mg/l of BOD. As this is a demand standard there is an acceptable range defined by the method of +/ mg/l. If making this standard up in the lab it is important to desiccate the glucose and glutamic acid prior to weighing out so it does not absorb moisture which can affect the weight. Commercially available, premade vials of GGA are available.

24 Biochemical Oxygen Demand (BOD) Standard Continued If your SCF is within range but your GGA results are high, decrease the amount of seed in your GGA and/or samples by 1-2 mls. If your SCF is within range but your GGA results are low, increase the amount of seed in your GGA and/or samples by 1-2 mls.

25 Biochemical Oxygen Demand (BOD) The biggest issue with the BOD test, however, is that the holding time for samples is 48 hours and the test takes 5 days. By the time you find out if your test is valid it is too late to do anything about it. One must give the utmost care to each step of your procedure to avoid the possibility of invalid data.

26 Questions? Joe Boyd Environmental Express December 26, 2012

ENGI Environmental Laboratory. Lab #2. Solids Determination. Faculty of Engineering & Applied Science

ENGI Environmental Laboratory. Lab #2. Solids Determination. Faculty of Engineering & Applied Science ENGI 9628 Environmental Laboratory Lab #2 Solids Determination Faculty of Engineering & Applied Science SOLIDS DETERMINATION PURPOSE Using Standard Methods for solids determination. THEORY Total solids

More information

EXPERIMENT 3 SOLIDS DETERMINATION

EXPERIMENT 3 SOLIDS DETERMINATION EXPERIMENT 3 SOIDS DETERMINATION A. PURPOSE To familiarize the students with various analytical operations such as weighing, filtration, evaporation and combustion which are commonly encountered in gravimetric

More information

SOP-C-130 Determination of Total Solids and Percent Solids

SOP-C-130 Determination of Total Solids and Percent Solids i. Identification of the method a. SM 2540 G & B (approved 2011) ii. Applicable matrix or matrices a. Nonpotable water (Not TNI accredited) b. Soils and other solids (not TNI accredited) iii. Limits of

More information

Total Suspended Solids by Gravimetric Determination

Total Suspended Solids by Gravimetric Determination SOP AMBL-105-D Page 1 of 5 Standard Operating Procedure AMBL-105-D Prepared: 12/27/2017 Revised: Prepared by: Terry E. Baxter Reviewed by: Total Suspended Solids by Gravimetric Determination METHOD SUMMARY

More information

Total Dissolved Solids Dried at 180ºC. Total Volatile Dissolved Solids at 550 C

Total Dissolved Solids Dried at 180ºC. Total Volatile Dissolved Solids at 550 C Title: Total Dissolved Solids Dried at 180ºC and Total Volatile Dissolved Solids at 500 C Page 1 of 9 Total Dissolved Solids Dried at 180ºC Total Volatile Dissolved Solids at 550 C References: SM 2540

More information

TREATMENT OF DAIRY WASTE WATER BY MORINGA OLEIFERA AS NATURAL COAGULANT

TREATMENT OF DAIRY WASTE WATER BY MORINGA OLEIFERA AS NATURAL COAGULANT . TREATMENT OF DAIRY WASTE WATER BY MORINGA OLEIFERA AS NATURAL COAGULANT Neethu.P 1, Navami.D 2, Anitha.K 3 1 P.G Student, Environmental Engineering, MCET, Kerala, India 2 P.G Student, Environmental Engineering,

More information

Gravimetric Analysis: Determination of % Sulfur in Fertilizer

Gravimetric Analysis: Determination of % Sulfur in Fertilizer Gravimetric Analysis: Determination % Sulfur in Fertilizer This is another "real world" sample experiment in this case we will analyze a fertilizer sample for the sulfate content and express the result

More information

Standard Methods for the Examination of Water and Wastewater

Standard Methods for the Examination of Water and Wastewater 2540 SOLIDS#(1)* 2540 A. Introduction Solids refer to matter suspended or dissolved in water or wastewater. Solids may affect water or effluent quality adversely in a number of ways. Waters with high dissolved

More information

3.0 DETECTION LIMIT 3.1. See Procedure ALQAP Procedure for the Determination of Uncertainty

3.0 DETECTION LIMIT 3.1. See Procedure ALQAP Procedure for the Determination of Uncertainty 1.0 IDENTIFICATION 1.1. SM 2540D Total Suspended Solids dried at 103-105 C. 2.0 APPLICABLE MATRIX 2.1. Drinking Water 3.0 DETECTION LIMIT 3.1. See Procedure ALQAP 011 - Procedure for the Determination

More information

Equipment Needs and Maintenance, What is Required in the Wet Chemistry Laboratory for Manual Method Testing

Equipment Needs and Maintenance, What is Required in the Wet Chemistry Laboratory for Manual Method Testing Equipment Needs and Maintenance, What is Required in the Wet Chemistry Laboratory for Manual Method Testing Debra Waller NJDEP-Office of Quality Assurance debra.waller@dep.nj.gov Evaluation of Suppliers

More information

Biochemical Oxygen Demand Do s and Don ts

Biochemical Oxygen Demand Do s and Don ts Biochemical Oxygen Demand Do s and Don ts Keith Chapman City of Salem-retired Willow Lake Laboratory ORWEF Short School March 29, 2017 Classic Historic Outline Definition and Overview Procedure Sampling

More information

Chapter 8. Gravimetric Analysis

Chapter 8. Gravimetric Analysis Chapter 8 Gravimetric Analysis Gravimetric analysis is the use of weighing to determine the amount of a component in your sample. Gravimetric analysis, or gravimetry is normally performed either as a :

More information

Fixed (Inorganic Ash) and Volatile Solids by Gravimetric Determination

Fixed (Inorganic Ash) and Volatile Solids by Gravimetric Determination SOP AMBL-105-E Page 1 of 7 Standard Operating Procedure AMBL-105-E Prepared: 12/28/2017 Revised: Prepared by: Terry E. Baxter Reviewed by: Fixed (Inorganic Ash) and Volatile Solids by Gravimetric Determination

More information

First Let s Define some Terms the most important terms used with these concepts are:

First Let s Define some Terms the most important terms used with these concepts are: Primer - Working With Total Solids, Fixed Solids and Volatile Solids in Sludge and Biosolids Michael D. Doran, DEE Professor of Practice Department of Civil & Environmental Engineering University of Wisconsin-Madison

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical without chloride removal, 10067 Manganese III Reactor Digestion Method (without chloride removal) 1 30 to 1000 mg/l COD Mn Scope and Application: For water and wastewater 1 U.S. Patent 5,556,787 DOC316.53.01102

More information

Northern Michigan Operators Spring Seminar. Lab Issues

Northern Michigan Operators Spring Seminar. Lab Issues Northern Michigan Operators Spring Seminar Lab Issues Wastewater Laboratory Issues Changes to List of EPA Approved Procedures Methods Update Rule (MUR) Changes to MDEQ-OTCU Wastewater Lab Training Manual

More information

Aflatoxin M1 ELISA Test Kit Manual Catalog #:

Aflatoxin M1 ELISA Test Kit Manual Catalog #: BIOO FOOD AND FEED SAFETY Aflatoxin M1 ELISA Test Kit Manual Catalog #: 1060-02 This kit is manufactured to the international quality standard ISO 9001:2008. ISO CI#: SARA-2009-CA-0114-01-B BIOO Scientific

More information

Experiment 1 Determination of ph

Experiment 1 Determination of ph Experiment 1 Determination of ph 1 Introduction: The term ph refers to the measure of hydrogen ion concentration in a solution and defined as the negative log of H + ions concentration in water and wastewater

More information

Approved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric)

Approved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric) METHOD #: 370.1 TITLE: Approved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric) ANALYTE: Silica, SiO 2 INSTRUMENTATION: Spectrophotometer STORET No. Dissolved 00955 1.0 Scope and Application

More information

Report No Waters Proficiency Testing Program. Round No Total Solids, Total Suspended Solids, Total Dissolved Solids - March 2017

Report No Waters Proficiency Testing Program. Round No Total Solids, Total Suspended Solids, Total Dissolved Solids - March 2017 Report No. 1017 Waters Proficiency Testing Program Round No. 06 - Total Solids, Total Suspended Solids, Total Dissolved Solids - March 017 Acknowledgments PTA wishes to gratefully acknowledge the technical

More information

Total Dissolved Solids

Total Dissolved Solids Total Dissolved Solids LabQuest 12 INTRODUCTION Solids are found in streams in two forms, suspended and dissolved. Suspended solids include silt, stirred-up bottom sediment, decaying plant matter, or sewage-treatment

More information

Measuring, sampling, and analyzing white waters

Measuring, sampling, and analyzing white waters T 656 cm-97 TENTATIVE STANDARD 1930 OFFICIAL STANDARD 1931 REVISED 1940 CORRECTED 1961 CLASSICAL METHOD 1983 REVISED 1997 1997 TAPPI The infmation and data contained in this document were prepared by a

More information

Top Tips for Conducting DMR-QA Analyses Waters Corporation 1

Top Tips for Conducting DMR-QA Analyses Waters Corporation 1 Top Tips for Conducting DMR-QA Analyses 2014 Waters Corporation 1 Presenter: Tom Widera Inorganics Product Line Manager ERA A Waters Company 2014 Waters Corporation 2 Founded in 1977 Golden, Colorado More

More information

Surface Water Samples: Containers, Preservation and Hold Times Table

Surface Water Samples: Containers, Preservation and Hold Times Table Surface Water Samples: Containers, Preservation and Hold Times Table North Carolina Division of Water Resources, Water Sciences Section Chemistry Laboratory Reference: 40 CFR Part 136.3 Table II Listed

More information

Porcine IgM (Immunoglobulin M) ELISA Kit

Porcine IgM (Immunoglobulin M) ELISA Kit Porcine IgM (Immunoglobulin M) ELISA Kit Catalogue No: EP0085 Size: 48T/96T Reactivity: Porcine Detection Range: 0.156-10ng/ml Sensitivity:

More information

DOWNSTREAM DEFENDER TESTING USING MAINE DEPARTMENT OF ENVIRONMENTAL PROTECTION PROTOCOL UTILIZING OK-110 FEED SAND

DOWNSTREAM DEFENDER TESTING USING MAINE DEPARTMENT OF ENVIRONMENTAL PROTECTION PROTOCOL UTILIZING OK-110 FEED SAND DOWNSTREAM DEFENDER TESTING USING MAINE DEPARTMENT OF ENVIRONMENTAL PROTECTION PROTOCOL UTILIZING OK-110 FEED SAND JULY 2002 Copyright 2001 H.I.L. Technology, Inc. All rights reserved. This report was

More information

MICROBIOLOGICAL BEST LABORATORY PRACTICES USP <1117>

MICROBIOLOGICAL BEST LABORATORY PRACTICES USP <1117> MICROBIOLOGICAL BEST LABORATORY PRACTICES USP MEDIA PREPARATION AND QUALITY CONTROL MAINTENANCE OF MICROBIOLOGICAL CULTURES LABORATORY EQUIPMENT LABORATORY LAYOUT AND OPERATIONS SAMPLE HANDLING

More information

2540 SOLIDS* 2540 A. Introduction

2540 SOLIDS* 2540 A. Introduction SOLIDS (2540)/Introduction 2-55 2540 SOLIDS* 2540 A. Introduction Solids refer to matter suspended or dissolved in water or wastewater. Solids may affect water or effluent quality adversely in a number

More information

Comparison of Water Quality Parameters

Comparison of Water Quality Parameters Name: Date: Comparison of Quality Parameters High School Environmental Science AP Module 2 Regional Reclamation Facility NGSSS Big Idea: Standard 1 Nature of Science Benchmark Code & Description: SC.912.N.1.1

More information

Isolation & Characterization of Bacteria

Isolation & Characterization of Bacteria PR025 G-Biosciences 1-800-628-7730 1-314-991-6034 technical@gbiosciences.com A Geno Technology, Inc. (USA) brand name Isolation & Characterization of Bacteria Teacher s Handbook (Cat. # BE 204) think proteins!

More information

Tour of Regional Water Reclamation Facility

Tour of Regional Water Reclamation Facility Name: Date: Tour of Regional Water Reclamation Facility High School Physical Science Fall Module 3 Regional Water Reclamation Facility NGSSS Big Idea: Standard 17 Interdependence Benchmark Code & Description:

More information

2. Crystallization. A. Background

2. Crystallization. A. Background 2. Crystallization A. Background Crystallization is one of several available techniques available to purify organic compounds. Unlike other techniques, however, crystallization is specific to the purification

More information

Aflatoxin B1 ELISA Test Kit Manual

Aflatoxin B1 ELISA Test Kit Manual BIOO FOOD AND FEED SAFETY Aflatoxin B1 ELISA Test Kit Manual Catalog #: 1055-04 This kit is manufactured to the international quality standard ISO 9001:2008. ISO CI#: SARA-2009-CA-0114--A BIOO Scientific

More information

Human TACI (TNFRSF13B) ELISA Kit

Human TACI (TNFRSF13B) ELISA Kit INSTRUCTIONS Human TACI (TNFRSF13B) ELISA Kit EHTNFRSF13B Number EHTNFRSF13B Description Human TACI ELISA Kit, sufficient reagents for 96 determinations Table of Contents Kit Contents Anti-Human TACI Precoated

More information

USEPA Membrane Filtration Method Method m-tec. Scope and Application: For potable water, nonpotable water, recreation water and wastewater.

USEPA Membrane Filtration Method Method m-tec. Scope and Application: For potable water, nonpotable water, recreation water and wastewater. , MF, m-tec, 8367 DOC316.53.001210 USEPA Membrane Filtration Method Method 8367 1 m-tec Scope and Application: For potable water, nonpotable water, recreation water and wastewater. 1 USEPA accepted. Test

More information

RayBio Rat IL-6 ELISA Kit (For Lysates)

RayBio Rat IL-6 ELISA Kit (For Lysates) RayBio Rat IL-6 ELISA Kit (For Lysates) Catalog #: ELR-IL6-CL User Manual Last revised April 15, 2016 Caution: Extraordinarily useful information enclosed ISO 13485 Certified 3607 Parkway Lane, Suite 100

More information

COLIFORM FILTERS. All filters listed on this page are white with black grids, 47 mm diameter, sterile.

COLIFORM FILTERS. All filters listed on this page are white with black grids, 47 mm diameter, sterile. COLIFORM FILTERS Which filters do I use for Fecal coliforms, and which do I use for Total coliforms? According to the 22 nd edition of Standard Methods, Procedure 9222B (Total Coliform MF-Procedure) and

More information

Evaluation of Conventional Activated Sludge Compared to Membrane Bioreactors

Evaluation of Conventional Activated Sludge Compared to Membrane Bioreactors Evaluation of Conventional Activated Sludge Compared to Membrane Bioreactors Short Course on Membrane Bioreactors 3/22/06 R. Shane Trussell, Ph.D., P.E. shane@trusselltech.com Outline Introduction Process

More information

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By the Authority Vested By Part 5 of the United States Code 552(a) and Part 1 of the Code of Regulations 51 the attached document has

More information

Assistant Lecturer. Sahar Mohammed Shakir Assistant Lecturer. Abdul Hafeedh Hameed Assistant Lecturer. Ali Basim

Assistant Lecturer. Sahar Mohammed Shakir Assistant Lecturer. Abdul Hafeedh Hameed Assistant Lecturer. Ali Basim Assistant Lecturer Sahar Mohammed Shakir Assistant Lecturer Abdul Hafeedh Hameed Assistant Lecturer Ali Basim Solid organic cpd.s when isolated from organic reaction are impure; they are contaminated with

More information

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856 , TNTplus856, 10220 DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and Application: For Water and Wastewater Test preparation Before starting the test: DR 2800:

More information

DAIRY WATERS. (Coliform Group and Escherichia coli) [E. coli verification required only on source water] IMS #24

DAIRY WATERS. (Coliform Group and Escherichia coli) [E. coli verification required only on source water] IMS #24 DAIRY WATERS (Coliform Group and Escherichia coli) [E. coli verification required only on source water] IMS #24 [Unless otherwise stated all tolerances are ±5%] 1. Laboratory Requirements a. Cultural Procedures

More information

Total Solids (TS) - material remaining after evaporation of sample liquid

Total Solids (TS) - material remaining after evaporation of sample liquid Page 1 of 8 Reference Two publications are widely used as the principal cookbooks for water and wastewater analysis: o Standard Methods for the Analysis of Water and Wastewater. American Water Works Association

More information

ort No. 964 Roun - Total Solids, Total Suspendedd Solids, Dissolved Solids - March 2016

ort No. 964 Roun - Total Solids, Total Suspendedd Solids, Dissolved Solids - March 2016 Repo ort No. 964 Waters Proficiency Testin ng Program Roun nd No. 192 - Total Solids, Total Suspendedd Solids, Total Dissolved Solids - March 2016 Acknowledgments PTA wishes to gratefully acknowledge the

More information

RayBio Human Caspase-3 ELISA Kit

RayBio Human Caspase-3 ELISA Kit RayBio Human Caspase-3 ELISA Kit Catalog #: ELH-CASP3 User Manual Last revised July 21, 2017 Caution: Extraordinarily useful information enclosed ISO 13485 Certified 3607 Parkway Lane, Suite 100 Norcross,

More information

Mouse ALP (Alkaline Phosphatase) ELISA Kit

Mouse ALP (Alkaline Phosphatase) ELISA Kit Mouse ALP (Alkaline Phosphatase) ELISA Kit Catalogue No.:MBS765655 Size:48T/96T Reactivity:Mouse Range:0.313-20ng/ml Sensitivity:

More information

ab Plant Chromatin Extraction Kit

ab Plant Chromatin Extraction Kit ab156906 Plant Chromatin Extraction Kit Instructions for Use For isolating chromatin or DNA-protein complex from plants in a simple and rapid format This product is for research use only and is not intended

More information

Module 20: Trickling Filters Answer Key

Module 20: Trickling Filters Answer Key Module 20: Trickling Filters Answer Key Calculation Capital City WWTF, which processes 2.0 MGD, is required to nitrify to meet the 2.0 mg/l ammonia discharge limit stated in their NPDES permit. A table

More information

RayBio Custom ELISA Kit

RayBio Custom ELISA Kit RayBio Custom ELISA Kit Catalog #: EL-PRELIM User Manual Last revised April 15, 2016 Caution: Extraordinarily useful information enclosed ISO 13485 Certified 3607 Parkway Lane, Suite 100 Norcross, GA 30092

More information

CORESTA RECOMMENDED METHOD N 61

CORESTA RECOMMENDED METHOD N 61 CORESTA RECOMMENDED METHOD N 61 DETERMINATION OF 1,2-PROPYLENE GLYCOL, GLYCEROL AND SORBITOL IN TOBACCO AND TOBACCO PRODUCTS BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC) (July 2015) 0. INTRODUCTION

More information

Mouse Heparan Sulfate(Heparan Sulfate) ELISA Kit

Mouse Heparan Sulfate(Heparan Sulfate) ELISA Kit Mouse Heparan Sulfate(Heparan Sulfate) ELISA Kit Catalogue No.:EM1594 Size:48T/96T Reactivity: Mouse Range:31.25-2000ng/ml Sensitivity:

More information

Fibrinogen ELISA. For the quantitative determination of fibrinogen in biological fluids, serum, and plasma.

Fibrinogen ELISA. For the quantitative determination of fibrinogen in biological fluids, serum, and plasma. Fibrinogen ELISA For the quantitative determination of fibrinogen in biological fluids, serum, and plasma. Please read carefully due to Critical Changes, e.g., Changes to preparation of standard volume

More information

Tetracycline ELISA Test Kit Manual Catalog #: 1016 Reference #: E

Tetracycline ELISA Test Kit Manual Catalog #: 1016 Reference #: E BIOO FOOD AND FEED SAFETY Tetracycline ELISA Test Kit Manual Catalog #: 1016 Reference #: 1016-04E This kit is manufactured to the international quality standard ISO 9001:2008. ISO CI#: SARA-2009-CA-0114-A

More information

Application Notes for COD Analysis DETERMINATION OF CHEMICAL OXYGEN DEMAND (COD) IN WATER AND WASTE WATER.

Application Notes for COD Analysis DETERMINATION OF CHEMICAL OXYGEN DEMAND (COD) IN WATER AND WASTE WATER. DETERMINATION OF CHEMICAL OXYGEN DEMAND (COD) IN WATER AND WASTE WATER. INTRODUCTION The chemical oxygen demand can be considered as an approximate measurement of the theoretical oxygen consumption, i.e.,

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical , TNTplus 820, 10211 DOC316.53.01103 USEPA Reactor Digestion Method Method 10211 ULR (1 60 mg/l COD) TNTplus 820 Scope and Application: For wastewater, process water, surface water, and cooling water;

More information

Guinea Pig IgA ELISA Kit

Guinea Pig IgA ELISA Kit CATALOG NO: IGPIGAKT Guinea Pig IgA ELISA Kit LOT NO: SAMPLE INTENDED USE The IgA test kits are a highly sensitive two-site enzyme linked immunoassay (ELISA) for measuring IgA in biological samples of

More information

THE USE OF MICROWAVE TECHNOLOGY FOR DRY ASHING PROCEDURES. by S. E. Carr and C. R. Moser. CEM Corporation, Matthews, North Carolina.

THE USE OF MICROWAVE TECHNOLOGY FOR DRY ASHING PROCEDURES. by S. E. Carr and C. R. Moser. CEM Corporation, Matthews, North Carolina. THE USE OF MICROWAVE TECHNOLOGY FOR DRY ASHING PROCEDURES by S. E. Carr and C. R. Moser CEM Corporation, Matthews, North Carolina March 1991 Presented at the 1991 Pittsburgh Conference and Exposition on

More information

ort No. 925 Proficiency Testin Roun - Total Solids, Total Suspendedd Solids, Dissolved Solids - August 2015

ort No. 925 Proficiency Testin Roun - Total Solids, Total Suspendedd Solids, Dissolved Solids - August 2015 Repo ort No. 95 Waters Proficiency Testin ng Roun nd No. 8 - Total Solids, Total Suspendedd Solids, Total Dissolved Solids - August 05 Acknowledgments PTA wishes to gratefully acknowledge the technical

More information

Human Bordetella Pertussis IgG ELISA kit

Human Bordetella Pertussis IgG ELISA kit Human Bordetella Pertussis IgG ELISA kit Catalog number: NR-R10157 (96 wells) The kit is designed to qualitatively detect Bordetella Pertussis IgG in Human serum or plasma. FOR RESEARCH USE ONLY. NOT FOR

More information

Experiment 2: Preparation of the Artificial Sweetener Dulcin

Experiment 2: Preparation of the Artificial Sweetener Dulcin Experiment 2: Preparation of the Artificial Sweetener Dulcin Organic compounds known as sugars are carbohydrates that occur widely in nature. For example, sucrose (aka table sugar) is found in sugar can,

More information

Oxygen. Oxygen is one of the fundamental resources required by life forms on Earth. Aquatic ecosystems have a wide assortment of life forms.

Oxygen. Oxygen is one of the fundamental resources required by life forms on Earth. Aquatic ecosystems have a wide assortment of life forms. Oxygen Oxygen is one of the fundamental resources required by life forms on Earth. Aquatic ecosystems have a wide assortment of life forms. Oxygen is also required for some natural chemical decays. What

More information

VICH Topic GL26 BIOLOGICALS: TESTING OF RESIDUAL MOISTURE

VICH Topic GL26 BIOLOGICALS: TESTING OF RESIDUAL MOISTURE The European Agency for the Evaluation of Medicinal Products Veterinary Medicines and Inspections CVMP/VICH/096/01-Final London, 30 April 2002 VICH Topic GL26 Step 7 BIOLOGICALS: TESTING OF RESIDUAL MOISTURE

More information

Standard Operation Procedure (SOP) Analytical methods in anaerobic digestion batch reactions

Standard Operation Procedure (SOP) Analytical methods in anaerobic digestion batch reactions Standard Operation Procedure (SOP) Analytical methods in anaerobic digestion batch reactions : 3 Pages: 7 Author: Lucien Biolley Date: 19.11.2015 Approved: Valid from: Changed: File: SOP_AnaerobicDigestion_Analysis_151119

More information

E.Z.N.A. Water DNA Kit. D preps D preps D preps

E.Z.N.A. Water DNA Kit. D preps D preps D preps E.Z.N.A. Water DNA Kit D5525-00 5 preps D5525-01 50 preps D5525-02 200 preps April 2017 E.Z.N.A. Water DNA Kit Table of Contents Introduction and Overview...2 Kit Contents/Storage and Stability...3 Preparing

More information

2. Sample dilution: Tissue lysate and cell lysate sample should be diluted at least 5-fold with 1x Sample Diluent Buffer.

2. Sample dilution: Tissue lysate and cell lysate sample should be diluted at least 5-fold with 1x Sample Diluent Buffer. Mouse IL-6 (Lysate) ELISA Kit Catalog No: CKM054 Size: 1 x 96 tests I. Introduction The Cell Sciences Mouse IL-6 ELISA (Enzyme-Linked Immunosorbent Assay) kit is an in vitro enzyme-linked immunosorbent

More information

Mouse Factor XII Total ELISA Kit

Mouse Factor XII Total ELISA Kit Mouse Factor XII Total ELISA Kit Catalog No: IMFXIIKT-TOT Lot No: SAMPLE INTENDED USE This mouse coagulation Factor XII antigen assay is intended for the quantitative determination of total Factor XII

More information

REPORT NUMBER REPORT DATE SEND TO ISSUE DATE Apr 18, Apr 18, 2017 RECEIVED DATE Apr 05, 2017

REPORT NUMBER REPORT DATE SEND TO ISSUE DATE Apr 18, Apr 18, 2017 RECEIVED DATE Apr 05, 2017 PAGE 1/7 Sample ID: EWA PELLETS Lab Number: 2653281 Date Sampled: 2017-04-04 0850 Carbon nitrogen ratio C/N 6 : 1 0.1 Calculation Auto-2017/04/12 Auto-2017/04/18 Carbon (total) 38.53 % 0.050 ASTM D 5373

More information

Fundamental Concepts: Overview of Water Quality

Fundamental Concepts: Overview of Water Quality Fundamental Concepts: Overview of Water Quality Ann Kenimer Texas A & M University University Curriculum Development for Decentralized Wastewater Management NDWRCDP Disclaimer This work was supported by

More information

AssayMax Mouse Transferrin ELISA Kit

AssayMax Mouse Transferrin ELISA Kit AssayMax Mouse Transferrin ELISA Kit Assaypro LLC 3400 Harry S Truman Blvd St. Charles, MO 63301 T (636) 447-9175 F (636) 395-7419 www.assaypro.com For any questions regarding troubleshooting or performing

More information

RayBio Human, Mouse and Rat Phospho-STAT3 (Tyr705) ELISA Kit

RayBio Human, Mouse and Rat Phospho-STAT3 (Tyr705) ELISA Kit RayBio Human, Mouse and Rat Phospho-STAT3 (Tyr705) ELISA Kit Catalog #: PEL-Stat3-Y705 User Manual Last revised August 10, 2016 Caution: Extraordinarily useful information enclosed ISO 13485 Certified

More information

Transferrin ELISA. For the quantitative determination of transferrin in human biological samples

Transferrin ELISA. For the quantitative determination of transferrin in human biological samples Transferrin ELISA For the quantitative determination of transferrin in human biological samples Please see Appendix A for Reference Serum information. For Research Use Only. Not For Use In Diagnostic Procedures.

More information

Evaluation copy. Total Dissolved Solids. Computer INTRODUCTION

Evaluation copy. Total Dissolved Solids. Computer INTRODUCTION Total Dissolved Solids Computer 12 INTRODUCTION Solids are found in streams in two forms, suspended and dissolved. Suspended solids include silt, stirred-up bottom sediment, decaying plant matter, or sewage-treatment

More information

RayBio Human HGF ELISA Kit

RayBio Human HGF ELISA Kit RayBio Human HGF ELISA Kit User Manual (for Cell Lysate and Tissue Lysate) (Revised Mar 1, 2012) RayBio Human HGF ELISA Kit Protocol (Cat#: ELH-HGF-001C) RayBiotech, Inc. We Provide You With Excellent

More information

ANCA Screen IgG ELISA Kit

ANCA Screen IgG ELISA Kit ANCA Screen IgG ELISA Kit Catalog number: NR-R10140 (96 wells) The kit is designed to qualitatively detect MPO and/or PR3 IgG in Human serum. FOR RESEARCH USE ONLY. NOT FOR DIAGNOSTIC OR THERAPEUTIC PURPOSES

More information

Tissue & Cell Genomic DNA Purification Kit. Cat. #:DP021/ DP Size:50/150 reactions Store at RT For research use only

Tissue & Cell Genomic DNA Purification Kit. Cat. #:DP021/ DP Size:50/150 reactions Store at RT For research use only Tissue & Cell Genomic DNA Purification Kit Cat. #:DP021/ DP021-150 Size:50/150 reactions Store at RT For research use only 1 Description: The Tissue & Cell Genomic DNA Purification Kit provides a rapid,

More information

Stream Ecology Lab Module 4/5 Determining Discharge and Sediment Yield in a Small Stream

Stream Ecology Lab Module 4/5 Determining Discharge and Sediment Yield in a Small Stream Stream Ecology Lab Module 4/5 Determining Discharge and Sediment Yield in a Small Stream Introduction: In this lab you will determine the total suspended solids yield per year for two streams a local stream

More information

THE USE OF PROCESS MODELING SOFTWARE IN INDUSTRIAL WASTEWATER TREATMENT - CASE STUDIES THAT ADDRESS AIR EMISSION, METALS REMOVAL AND WATER REUSE

THE USE OF PROCESS MODELING SOFTWARE IN INDUSTRIAL WASTEWATER TREATMENT - CASE STUDIES THAT ADDRESS AIR EMISSION, METALS REMOVAL AND WATER REUSE THE USE OF PROCESS MODELING SOFTWARE IN INDUSTRIAL WASTEWATER TREATMENT - CASE STUDIES THAT ADDRESS AIR EMISSION, METALS REMOVAL AND WATER REUSE Malcolm Fabiyi, PhD, MBA; Rajeev Goel, PhD, P.Eng., PE Hydromantis

More information

LDH-Cytotoxicity Assay Kit II

LDH-Cytotoxicity Assay Kit II LDH-Cytotoxicity Assay Kit II Catalog Number KA0786 500 assays Version: 08 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 General Information... 4 Materials

More information

SECTION D.3 CHLOROPHYLL AND PHEOPHYTIN

SECTION D.3 CHLOROPHYLL AND PHEOPHYTIN SECTION D.3 CHLOROPHYLL AND PHEOPHYTIN 1.1 Scope and Application 1.1.1 This spectrophotometric method is used in the determination of chlorophyll a, b, and c, and pheophytin a. This method can be used

More information

RayBio Human, Mouse and Rat Phospho-STAT3 (Tyr705) and Total STAT3 ELISA Kit

RayBio Human, Mouse and Rat Phospho-STAT3 (Tyr705) and Total STAT3 ELISA Kit RayBio Human, Mouse and Rat Phospho-STAT3 (Tyr705) and Total STAT3 ELISA Kit Catalog #: PEL-Stat3-Y705-T User Manual Last revised October 10, 2017 Caution: Extraordinarily useful information enclosed ISO

More information

Cat Toxoplasma IgG ELISA Kit

Cat Toxoplasma IgG ELISA Kit Cat Toxoplasma IgG ELISA Kit Catalog number: BG-CAT10299 (96 wells) The kit is designed to qualitatively detect Toxoplasma IgG in Cat serum. FOR RESEARCH USE ONLY. NOT FOR DIAGNOSTIC OR THERAPEUTIC PURPOSES

More information

Human IgG Antigen ELISA Kit

Human IgG Antigen ELISA Kit Human IgG Antigen ELISA Kit Catalog No: IHUIGGKT Lot No: SAMPLE INTENDED USE This human immunoglobulin G antigen assay is intended for the quantitative determination of total human IgG antigen in serum,

More information

DNA isolation from tissue DNA isolation from eukaryotic cells (max. 5 x 106 cells) DNA isolation from paraffin embedded tissue

DNA isolation from tissue DNA isolation from eukaryotic cells (max. 5 x 106 cells) DNA isolation from paraffin embedded tissue INDEX KIT COMPONENTS 3 STORAGE AND STABILITY 3 BINDING CAPACITY 3 INTRODUCTION 3 IMPORTANT NOTES 4 EUROGOLD TISSUE DNA MINI KIT PROTOCOLS 5 A. DNA isolation from tissue 5 B. DNA isolation from eukaryotic

More information

Rev Experiment 10

Rev Experiment 10 Experiment 10 SPECTROPHOTOMETRIC DETERMINATION OF IRON IN DRINKING WATER 2 lab periods Reading: 1) Chapter 17, pg 393-403, Quantitative Chemical Analysis, 8 h Edition, Daniel C. Harris (7 th Edition: Chapter

More information

Methods for determination of mercury in LP gas

Methods for determination of mercury in LP gas (JLPGA-S-07) Methods for determination of mercury in LP gas 1. Scope This standard specifies the methods for determination of gaseous mercury in vaporized liquid liquefied petroleum (LP) gas. Remarks 1

More information

Standard Operating Procedure Title: Operation and Calibration of Sievers 820 TOC Analyser

Standard Operating Procedure Title: Operation and Calibration of Sievers 820 TOC Analyser Table of Contents 1. 25 ppm IC as Sodium Carbonate - standard preparation procedure 2 2. 25 ppm TOC as KHP standard preparation procedure 3 3. Calibration Procedure 4 4. Performing Calibration - Prepare

More information

Principles and Practice of Agarose Gel Electrophoresis

Principles and Practice of Agarose Gel Electrophoresis Edvo-Kit #101 Principles and Practice of Agarose Gel Electrophoresis Experiment Objective: The objective of this experiment is to develop a basic understanding of electrophoretic theory, and to gain "hands-on"

More information

Human vascular endothelial cell growth factor A (VEGF-A) ELISA Kit

Human vascular endothelial cell growth factor A (VEGF-A) ELISA Kit Human vascular endothelial cell growth factor A (VEGF-A) ELISA Kit For the quantitative determination of human vascular endothelial cell growth factor A (VEGF-A) concentrations in serum, plasma, tissue

More information

RESOLUTION MEPC.227(64) Adopted on 5 October GUIDELINES ON IMPLEMENTATION OF EFFLUENT STANDARDS AND PERFORMANCE TESTS FOR SEWAGE TREATMENT

RESOLUTION MEPC.227(64) Adopted on 5 October GUIDELINES ON IMPLEMENTATION OF EFFLUENT STANDARDS AND PERFORMANCE TESTS FOR SEWAGE TREATMENT ANNEX 22 MEPC 64/23/Add.1 Annex 22, page 1 THE MARINE ENVIRONMENT PROTECTION COMMITTEE, RECALLING Article 38(a) of the Convention on the International Maritime Organization concerning the functions of

More information

Nadeem Shakir Qatar Petroleum. The 2nd Joint Qatar Japan Environmental Symposium, QP JCCP The 21st Joint GCC Japan Environmental Symposium

Nadeem Shakir Qatar Petroleum. The 2nd Joint Qatar Japan Environmental Symposium, QP JCCP The 21st Joint GCC Japan Environmental Symposium Nadeem Shakir Qatar Petroleum Scheme of Presentation General Overview of QP Refinery, Mesaieed. Challenges in Treatment Existing Waste Water Treatment Facilities Capacity Expansion and Upgradation of WWTP

More information

APOB (Human) ELISA Kit

APOB (Human) ELISA Kit APOB (Human) ELISA Kit Catalog Number KA1028 96 assays Version: 17 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 Principle of the Assay... 3 General

More information

Pretreatment of Industrial Wastewater for ZLD/Resources Recovery

Pretreatment of Industrial Wastewater for ZLD/Resources Recovery Pretreatment of Industrial Wastewater for ZLD/Resources Recovery An Application of Oxidative Evaporation Technology Philip Chen, PhD, P.E. O E Technology, LLC Info@oetech.com 1 Process Plant Bio-Unfit

More information

METHOD #: Approved for NPDES (Issued 1971) Oxygen, Dissolved (Modified Winkler, Full-Bottle Technique) ANALYTE: CAS # O Oxygen

METHOD #: Approved for NPDES (Issued 1971) Oxygen, Dissolved (Modified Winkler, Full-Bottle Technique) ANALYTE: CAS # O Oxygen METHOD #: 360.2 Approved for NPDES (Issued 1971) TITLE: Oxygen, Dissolved (Modified Winkler, Full-Bottle Technique) ANALYTE: CAS # O Oxygen 7782-44-7 INSTRUMENTATION: Titration, Probe STORET No. 00300

More information

Group Stormwater Compliance Program Sampling Instructions

Group Stormwater Compliance Program Sampling Instructions 2016-2017 Group Stormwater Compliance Program Sampling Instructions Read these instructions completely to ensure a thorough understanding of the sampling process before attempting to collect samples of

More information

Wastewater Pollutants & Treatment Processes. Dr. Deniz AKGÜL Marmara University Department of Environmental Engineering

Wastewater Pollutants & Treatment Processes. Dr. Deniz AKGÜL Marmara University Department of Environmental Engineering Wastewater Pollutants & Treatment Processes Dr. Deniz AKGÜL Marmara University Department of Environmental Engineering Wastewater combination of the liquid or water carried wastes removed from residences,

More information

Human anti-hepatitis A virus (HAV) antibody (IgM) ELISA Kit

Human anti-hepatitis A virus (HAV) antibody (IgM) ELISA Kit Human anti-hepatitis A virus (HAV) antibody (IgM) ELISA Kit For the qualitative determination of human anti-hepatitis A virus (HAV) antibody (IgM) concentrations in serum. This package insert must be read

More information

International Society of Beverage Technologists Beverage Production Seminar January 24-27, 2000 Wastewater Minimization and Treatment

International Society of Beverage Technologists Beverage Production Seminar January 24-27, 2000 Wastewater Minimization and Treatment International Society of Beverage Technologists Beverage Production Seminar January 24-27, 2000 Wastewater Minimization and Treatment Harry C. DeLonge USFilter Wastewater Minimization and Treatment Wastewater

More information

Manual innuprep Micro RNA Kit

Manual innuprep Micro RNA Kit Manual Order No.: 845-KS-2030010 10 reactions 845-KS-2030050 50 reactions 845-KS-2030250 250 reactions Publication No.: HB_KS-2030_e_120116 This documentation describes the state at the time of publishing.

More information

Human rotavirus (RV) antibody (IgG) ELISA kit

Human rotavirus (RV) antibody (IgG) ELISA kit Human rotavirus (RV) antibody (IgG) ELISA kit For the qualitative determination of human rotavirus (RV) antibody (IgG) concentrations in serum. This package insert must be read in its entirety before using

More information

Instructions for use

Instructions for use Instructions for use SARCOPTES-ELISA 2001 DOG Test to detect the presence of antibodies against the cause of mange of dogs, Sarcoptes scabiei var. canis In-vitro Diagnostic for Dog Contents of the Kit:

More information