Analytical Instrument Qualification According to USP <1058>: Requirements and Examples for the Agilent 1290 Infinity LC System

Size: px
Start display at page:

Download "Analytical Instrument Qualification According to USP <1058>: Requirements and Examples for the Agilent 1290 Infinity LC System"

Transcription

1 Analytical Instrument Qualification According to USP <1058>: Requirements and Examples for the Agilent 1290 Infinity LC System 1) USP <1058>: Scope, approach, requirements 2) Qualification Examples for LC ) Selected performance tests Dr. Edgar Naegele Application Scientist Agilent Technologies, Germany 1

2 Design Qualification Selected HPLC specifications for design qualification. Selected specifications of the Agilent 1290 infinity it autosampler 2

3 Design Qualification Selected criteria i for vendor assessment. All requirements of this custom vendor assessment template are fulfilled by Agilent Technologies! X X X X X X X X 3

4 Installation Qualification 1) Template and examples to document completeness of shipment for IQ. 3) Equipment documentation for IQ. yes yes 2) Verification of construction requirements for IQ. 4

5 Installation Qualification Testing for Installation Qualification Verification of correct system installation for IQ. For this purpose p Agilent uses a certified standard! 5

6 Operation qualification - an example procedure HPLC - System Pump Injector Oven Detector Computer Control& Evaluation Flow rate Inj. Vol. Temperature Signal precision accuracy accuracy and/or noise Specifications of the Agilent 1290 infinity LC system Flow rate precision Inj. Vol. accuracy Temperature accuracy Signal and/or noise 0.07% 07% RSD or <025RSD /- 08 C 0.8 <+/ +/- 3x10-6 AU SD at Precision ml/min +/ C from 5 20 µl < 0.50 RSD from 2 5 µl < 0.70 RSD from 1 2 µl at 230/4 nm, slit width 4nm, TC 2 s ASTM 6

7 Operation qualification - an example procedure Computer Control & Evaluation Generated Automatically Header Signed and dated Summary Test/Limit/Result/Status Configuration Module/FW/Serial No. & Software Verification Log Book Test Reports Sequence/Log Method Chromatograms/Spectra/Reports Results/Sample Summary 7

8 Gradient test depending on the volume of the mixers (blue-35µl, µ, red-135µl, µ,g green-300µl) µ ) mau Gradient (stepwise): Time % A % B Influence of the mixer Gradient Testing: Mobile Phase A: Water Mobile Phase B: Water + 0,5% Aceton (v/v) min

9 1290 Infinity Binary Pump Degasser High resolution pump di drives Silicon Carbide piston Channel A Solvent selection valve Purge valve Jet Weaver Mixer Multi layer heat exchanger Channel B Next Generation Technology ml/min ml/min Minimum dwell volume: 10 µl / 45 µl Minimum Mixing Noise Maximum RT repeatability Automatic Purge Automatic correction of von micro leaks 9

10 1290 Infinity binary Pump Minimum dead volume and mixing noise 1200 Series Solvent A 1200 Series Solvent B Passive Damper µl Passive Mixer 400 µl Low Noise ~ 800 µl dead volume 1290 Inifinity Solvent A 1290 Inifinity Solvent B Active Damping: 10 µl Jet Weaver Mixer: 35 µl Firmware Hardware Agilent microfluidic, Multilayer Technologie + Pump Pump + = 10 µl / 45 µl DV: 80x 20x / smaller 20x kleiner Niedriges Lower Noise Noise < 50% 10

11 Example: Performance Qualification Documentation ti of PQ tests. t ChemStation Report with detailed statistical information Amount S.D. : RSD:

12 Performance qualification an example mau AU RSD % RT (n=10) RSD % Area (n=10) Mean min Sample: Check Check-out out sample with 9 phenones 12 Peak Column: 2.1 x 50 mm, 1.8 µm Gradient: 5 to 95 % B in 0.5 min Flow rate: 1.2 ml/min Injection volume: 1µL

13 Analytical Instrument Qualification According to USP <1058>: Requirements and Examples for the Agilent 1290 Infinity LC System 1) USP <1058>: Scope, approach, requirements 2) Qualification Examples for LC ) Selected performance tests and performance data 13

14 Application depended Performance Data and System Suitability - Experimental conditions for the Agilent 1290 infinity LC system Configuration of the 1290-System G4220A 1290 Infinity Binary pump with integrated vacuum degasser and different solvent mixers G4226A 1290 Infinity Autosampler G1316C 1290 Infinity Column Compartment G4212A 1290 Infinity Diode Array Detector Software: Chemstation B Reference samples A liquid formulation of metoclopramide hydrochloride (1 mg/ml) comprising 0.01% concentration of impurities was used for all tests. Setup for testing 1. Determination of pump performance depending on dwell volumes by gradient tests 2. Establishment of a chromatographic separation to achieve data for long time evaluations 3. Similar peak pattern and resolution according to selected column with respect to particle size and column dimension and adapted to gradient shape and flow 4. Precision of areas must be < 2 % RSD. 5. Precision of retention times must be < 0.5 % RSD. 6. Linearity should be given at least with R2>0.999 Sample setup for testing Sample Purpose Injections Blank solution Verify baseline stability and identify artifacts 3 Suitability sample Verify precision of areas and 10 retention times for reference solution Calibration Verify linearity 3 for each level Highest concentration ti Verify Carry over 3 of each sample and Blank solution 14 Confidentiality Label

15 Separation of Metoclopramide and its impurities on different columns. A B C D A) Agilent TC C18, B) Eclipse Plus C18, C) Eclipse Plus C18, D) Waters BEH C18, 250*4.6mm, 5µm 150*2.1mm, 3.5µm 150*2,1mm, 1.8µm 150*2.1mm, 1.7µm Flow rate 1.058ml/min 1.058ml/min 0.221ml/min 0.221ml/min Gradient 0-25min 5-57% B 0-15min 5-57% B 0-15min 5-57% B 0-15min 5-57% B Temperature 37 C 37 C 37 C 37 C Injection 8µl 2µl 1µl 1µl volume Detection DAD, Signal 275/4, Reference 400/60, standard cell Data rate 2 Hz 2 Hz 40 Hz 40 Hz Maximum 138 bar 145 bar 310 bar 435 bar pressure Mobile phase A : 0.25% w/w Ammonium acetate in water Mobile phase B : Acetonitrile

16 Determination of the precision of areas and retention times Impurities in Metoclopramide p Determination of the precision of areas and retention times Retention times Areas Linearity Mean RSD Mean RSD R 2 Impurity Impurity Metoclopramid Impurity Impurity Impurity Impurity

17 Agilent 1290 Infinity Autosampler Flow-through design Advantage of the flow-through design No sample loss Flexible injection volume with high precision from 0.1 to 20 µl No need to exchange the sample loop Compromise of the flow-through design higher dead volume of 85 µl For the 1290 Agilent infinity LC system 125 µl dead volume if loop is in main path 17

18 Detection of carry over. A) Injection of sample with highest concentration of metoclopramide. B) Injection of a blank solvent sample. C) Carry-over specification. A Metoclopramide mau DAD1 A, Sig=257,4 Ref=360,100 (F:\CHLORHEX100\WATER D) DAD1 A, Sig=257,4 Ref=360,100 (F:\CHLORHEX100\CHLOREX D) Chlorhexidine 0.75 B % carry over min Column: 2.1x50mm Eclipse Plus C18, 1.8µm Sample: Chlorhexidine (1200ng/µl) dissolved in Water+0.1%TFA Mobile Phases: water+0.05%tfa, Acetonitrile+0.045%TFA Isocratic: 64 % A + 33 % B. Flow rate: 0,5ml/min, Stop time: 2.5min Injection volume: 1µl, external needle wash, water + 0.1%TFA, 10sec Column temperature: 50 C DAD: 257/4nm, Ref 360/100nm, 20Hz

19 Speed optimized separation of Metoclopramide and its impurities Zorbax Eclipse Plus C18 150*2 1mm 18µm Zorbax Eclipse Plus C18, mm, 1.8µm Flow rate : 0.44ml/min Gradient : 10 minutes

20 1290 Infinity Diode Array Detector Optofluidic Waveguides Non-coated fiber (fused silica) High light transmission by Total-Internal Reflection (TIR) Principle Advantage High Sensitivity at Lowest Cell Volume No RI (Refractive Index) and Thermal Effects (Solvent Temperature) Data Rate up to160 Hz (Multi Wave Length + Spectra) One Flow Cell for all Applications Cartridge Design for easy Handling 20

21 Conclusion The new Agilent Infinity LC is designed to provide highest speed, resolution and sensitivity. A new power range allows you to operate with any particle type, any column dimensions, or any mobile and stationary phase. The Agilent 1290 Infinity HPLC system is can be qualified according to the USP <1058> chapter and will meet highest requirements for every LC application. A selection of tests that can be established to evaluate a LC system were shown: For instance gradient testing, the determination of precision of areas and retention and the determination of flinearity. it All results show explicitly the applicability of the 1290 Infinity LC system for quality control testing as well as in the FDA regulated environment. Agilent Services qualifies your system according to USP1058! 21

22 References/Further reading United States Pharmacopiea, Chapter <1058>, Analytical Instrument Qualification, USA, 2008 Analytical Instrument Qualification and System Validation, January 1, 2009, Agilent Publication Number EN Increasing productivity in the analysis of impurities iti in metoclopramide hydrochloride hl formulations using the Agilent 1290 Infinity LC system, Agilent Technologies Pub.-Number: EN Analytical instrument qualification and system validation according to USP Chapter <1058> for the Agilent 1290 Infinity LC system, Agilent Technologies Pub.-Number: EN 22

23 Thank you for your attention! Do you have Questions? 23

New Advances in UHPLC -Resolution, Speed & Sensitivity

New Advances in UHPLC -Resolution, Speed & Sensitivity New Advances in UHPLC -Resolution, Speed & Sensitivity Enhancing Productivity with Agilent s Newest Instrumentation Patrick Cronan LC Applications Scientist Boston, MA September 14, 2012 Page 1 The 1290

More information

Qualification of High-Performance Liquid Chromatography Systems

Qualification of High-Performance Liquid Chromatography Systems Qualification of High-Performance Liquid Chromatography Systems L. Huber Hewlett- Packard GmbH, Waldbronn December 1998 Manuscript of an article published in BioPharm, Vol 11 Number 11, November 1998,

More information

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany New Features of the Agilent Method Scouting Wizard for Automated Method Development of Complex Samples Analysis of Large Data Sets by Method Scouting Reports and Automated Adjustment of Flow Rates and

More information

Analysis of amoxicillin and five impurities on the Agilent 1220 Infinity LC System

Analysis of amoxicillin and five impurities on the Agilent 1220 Infinity LC System Analysis of amoxicillin and five impurities on the Agilent Infinity LC System LC analysis of impurities down to the.% level with long sub--µm columns, high flow rates and back pressure greater than bar

More information

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC Quality-by-Design-Based Method Development Using an Agilent 129 Infinity II LC An Efficient Method Development Workflow Combined with ISET-mediated Method Transfer Under Waters Empower 3 CDS Control Application

More information

Controlling Agilent 1200 Series Rapid Resolution LC systems through Waters Empower chromatography data software. Technical Overview.

Controlling Agilent 1200 Series Rapid Resolution LC systems through Waters Empower chromatography data software. Technical Overview. Controlling Agilent 1200 Series Rapid Resolution LC systems through Waters Empower chromatography data software Technical Overview Abstract The Waters Empower 2 chromatography data software with ICS 1.05

More information

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column Disulfide Linkage Analysis of IgG1 using an Agilent 126 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column Application Note Biotherapeutics & Biosimilars Author M. Sundaram

More information

Analysis of TCM injections using the Agilent 1290 Infinity LC system

Analysis of TCM injections using the Agilent 1290 Infinity LC system Analysis of TCM injections using the Agilent 1290 Infinity LC system Application Note Traditional Chinese Medicine Authors Zhixiu Xu Agilent Technologies Shanghai Shanghai 200131 China mau 400 350 300

More information

Application Note. Author. Abstract. Small Molecule Pharmaceuticals. Sonja Krieger Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Author. Abstract. Small Molecule Pharmaceuticals. Sonja Krieger Agilent Technologies, Inc. Waldbronn, Germany Method Transfer from an Agilent 1100 Series Quaternary LC to an Agilent 1260 Infinity II LC Proof of Equivalency for the Analysis of Antihistaminic Drugs Application Note Small Molecule Pharmaceuticals

More information

easy. HPLC has never been so Agilent 1120 Compact LC Our measure is your success.

easy. HPLC has never been so Agilent 1120 Compact LC Our measure is your success. easy. HPLC has never been so Agilent 1120 Compact LC Our measure is your success. Introducing the Agilent 1120 Compact LC Simply a better value. Is your lab looking for better, more reproducible results

More information

ACQUITY UPLC. I-Class PLUS WHAT SEPARATES YOU FROM EVERYONE ELSE

ACQUITY UPLC. I-Class PLUS WHAT SEPARATES YOU FROM EVERYONE ELSE ACQUITY UPLC I-Class PLUS WHAT SEPARATES YOU FROM EVERYONE ELSE COMPLEX CHALLENGES SOLVED Looking for answers to the most complex scientific challenges is what you do. Your research could lead to ground-breaking

More information

Application Note. Author. Abstract. Small Molecule Pharmaceuticals & Generics. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Author. Abstract. Small Molecule Pharmaceuticals & Generics. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany Analysis of Degradation Products of Doxycycline in Solution Exposed to Light and Elevated Temperatures Using the Agilent 1 Infinity Series High Dynamic Range Diode Array Detector Impurity Analyzer System

More information

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider Combining small-scale purification and analysis of monoclonal antibodies on one instrument Protein purification with high-volume injection using the Agilent 126 Infinity Bio-inert Quaternary LC System

More information

C196-E067A. Prominence UFLCXR. Shimadzu Ultra Fast Liquid Chromatograph

C196-E067A. Prominence UFLCXR. Shimadzu Ultra Fast Liquid Chromatograph C196-E067A Prominence UFLCXR Shimadzu Ultra Fast Liquid Chromatograph New Prominence UFLC Series Line-up Prominence UFLCXR In Pursuit of Higher Separation and Sensitivity The Prominence UFLCXR is an ultra-high-speed

More information

Method Translation in Liquid Chromatography

Method Translation in Liquid Chromatography Method Translation in Liquid Chromatography Technical Overview Abstract Ronald E. Majors Agilent Technologies, Inc. 2850 Centerville Rd Wilmington, DE 19808 USA With the recent emphasis on high performance

More information

Application Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany Quantification of trace-level herbicides in drinking water by online enrichment with the Agilent Infinity Series Online-SPE Solution and Triple Quadrupole MS Detection Application Note Environmental Author

More information

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany D-LC/MS Characterization of Charge Variants Using Ion Exchange and Reversed-Phase Chromatography Multiple Heart-Cutting D-LC Analysis of Innovator versus Biosimilar Monoclonal Antibodies Application Note

More information

Peptide Mapping: A Quality by Design (QbD) Approach

Peptide Mapping: A Quality by Design (QbD) Approach Peptide Mapping: A Quality by Design (QbD) Approach Application Note Bio-Pharmaceutical Authors Sreelakshmy Menon and Suresh babu C.V. Agilent Technologies, Inc. Richard Verseput S-Matrix Corporation Abstract

More information

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System Application Note Biologics & Biosimilars Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction Chromatography Agilent 126 Infinity II Bio-Inert System 7 6 5 4 5. 7.5

More information

Agilent 1220 Infinity II LC GET ON THE FAST TRACK TO HIGHEST EFFICIENCY AT A PRICE YOU CAN AFFORD

Agilent 1220 Infinity II LC GET ON THE FAST TRACK TO HIGHEST EFFICIENCY AT A PRICE YOU CAN AFFORD 1220 Agilent 1220 Infinity II LC GET ON THE FAST TRACK TO HIGHEST EFFICIENCY AT A PRICE YOU CAN AFFORD AGILENT 1220 INFINITY II LC AFFORDABLE EFFICIENCY WITH CONFIDENCE Take a close look at the Agilent

More information

Technical Note. Introduction

Technical Note. Introduction Online HPLC for Biotechnology: Groton Biosystems Automated Reactor Sampling System and Agilent 1200 Series LC System Technical Note Introduction Online analytical instrumentation facilitates productive,

More information

Validation Strategies for Equipment from Multiple Vendors

Validation Strategies for Equipment from Multiple Vendors Welcome to our E-Seminar: Validation Strategies for Equipment from Multiple Vendors Page 1 Content FDA guidelines and inspectional observations Validation & Qualification Approaches Instrument Qualification

More information

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study Application note Biotherapeutics and Biologics Authors Suresh Babu C.V., Anne

More information

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column Application Note Bio-Pharmaceutical Authors M. Sundaram Palaniswamy and Andrew Coffey Agilent Technologies,

More information

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility Technical Overview Introduction Agilent AdvanceBio SEC columns are a new family of size exclusion chromatography (SEC) columns

More information

HPLC to UPLC Method Migration: An Overview of Key Considerations and Available Tools

HPLC to UPLC Method Migration: An Overview of Key Considerations and Available Tools HPLC to UPLC Method Migration: An Overview of Key Considerations and Available Tools Dr. Michael Swartz, Ph. D. Principal Consulting Scientist Worldwide Pharmaceutical Business Operations Waters Corporation

More information

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions Screening and Confirmation Chemical Analysis Solution Unit November, 2009 Page 1 Agilent s New 6540 Ultra High

More information

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube Application Note Environmental Authors Don Noot, Ralph Hindle Vogon Laboratory Services Cochrane, AB

More information

BioHPLC columns. Tim Rice Biocolumn Technical Specialist

BioHPLC columns. Tim Rice Biocolumn Technical Specialist BioHPLC columns Tim Rice Biocolumn Technical Specialist AU Typical Application Areas Size Exclusion: Aggregation Analysis Ion Exchange: Charge Isoform Analysis 0.035 Monomer 0.030 0.025 0.020 0.015 Dimer

More information

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography Characterize Fab and Fc Fragments by Cation-Exchange Chromatography Application Note Biologics and Biosimilars Authors Isabel Vandenheede, Emmie Dumont, Pat Sandra, and Koen Sandra Research Institute for

More information

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC The Agilent 126 Infinity Bio-Inert Quaternary LC System and the AdvanceBio SEC 3Å, 2.7 µm Column Application Note Biologics

More information

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system Application Note Biopharmaceuticals Authors Arunkumar Padmanaban and Sreelakshmy Menon

More information

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Student Learning Objectives High Performance Liquid Chromatography With the advent of relatively inexpensive and reliable pumps, the development of

More information

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns Agilent AdvanceBio SEC 3 Å, 2.7 µm columns Application note Bio-Pharmaceutical Author

More information

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS Agilent AdvanceBio Columns ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS with Agilent ZORBAX RRHD 3Å 1.8 µm columns ns Rapid resolution high definition columns for UHPLC protein and peptide separations

More information

IC-D3160 Ion Chromatograph

IC-D3160 Ion Chromatograph IC-D3160 Ion Chromatograph IC-D3160 ion chromatograph is a new modular design of the high stability of ion chromatograph, combining Analytical Technologies Limited independent research and development

More information

Minimize Method Development Time for Pesticide Screening Using LC/MS Application Kit

Minimize Method Development Time for Pesticide Screening Using LC/MS Application Kit Minimize Method Development Time for Pesticide Screening Using LC/MS Application Kit Dr. Jerry Zweigenbaum Senior Applications Scientist Food Safety Page 1 Agilent 1260 LC SampliQ Extraction and Dispersive

More information

Agilent Infinity LC Instrument Component Software Version 1.0 for Empower Software

Agilent Infinity LC Instrument Component Software Version 1.0 for Empower Software R E L E A S E N O T E S Agilent Infinity LC Instrument Component Software Version 1.0 for Empower Software These release notes explain how to install the Waters Instrument Component Software (ICS), version

More information

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions An Automated System for At-Line cess Analysis of Biopharmaceutical Fermentation Reactions Frank Steiner, Susanne Fabel, Fraser McLeod, Dionex Corporation, Germering, Germany Abstract Monitoring amino acid

More information

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography Carl Fisher, Monika Verma, and Pranathi Perati; Thermo Fisher Scientific,

More information

Preparative Purification of Corticosteroids by HPLC; Scalability and Loadability Using Agilent Prep C18 HPLC Columns Application

Preparative Purification of Corticosteroids by HPLC; Scalability and Loadability Using Agilent Prep C18 HPLC Columns Application Preparative Purification of Corticosteroids by PLC; Scalability and Loadability Using Agilent Prep C18 PLC Columns Application Pharmaceuticals Authors Cliff Woodward and Ronald Majors Agilent Technologies,

More information

APPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology

APPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology TN-7 Determination of Impurities and Related Substances for Glibenclamide (EP Monograph 78). Increased Sensitivity, Improved Resolution and Faster Analysis Using Kinetex.6 µm Core-Shell LC Columns Elli

More information

SFC Säulen für analytische und preparative Anwendungen

SFC Säulen für analytische und preparative Anwendungen SFC Säulen für analytische und preparative Anwendungen Technology Symposium Vienna, 25 th October 2016 DI Verena Schmid 2016 Waters Corporation 1 CHIRAL SEPARATIONS ACQUITY UPC 2 Trefoil 2016 Waters Corporation

More information

Heart-cut 2D-LC/MS approach for pharmaceutical impurity identification using an Agilent 6540 Q-TOF LC/MS System

Heart-cut 2D-LC/MS approach for pharmaceutical impurity identification using an Agilent 6540 Q-TOF LC/MS System Heart-cut 2D-LC/M approach for pharmaceutical impurity identification using an Agilent 640 Q-TF LC/M ystem Application ote Pharmaceuticals Author iji Joseph Agilent Technologies, Inc. Bangalore, India

More information

Agilent 1260 Infinity II LC ASPIRE TO NEW LEVELS OF EFFICIENCY EVERY DAY IN EVERY WAY

Agilent 1260 Infinity II LC ASPIRE TO NEW LEVELS OF EFFICIENCY EVERY DAY IN EVERY WAY 1260 Agilent 1260 Infinity II LC ASPIRE TO NEW LEVELS OF EFFICIENCY EVERY DAY IN EVERY WAY AGILENT 1260 INFINITY II LC EVERYDAY EFFICIENCY WITH CONFIDENCE Put your trust in the LC platform that runs, and

More information

Update to USP <621> Chromatography: What does this mean? USP 37 NF 32 1S Current as of August 2014

Update to USP <621> Chromatography: What does this mean? USP 37 NF 32 1S Current as of August 2014 Update to USP Chromatography: What does this mean? USP 37 NF 32 1S Current as of August 2014 2015 Waters Corporation 1 What is the USP-NF? The United States Pharmacopeia National Formulary (USP-NF)

More information

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides High-pH-stable, superficially porous particle columns for LC/UV and LC/MS Application Note Biologics and Biosimilars Authors

More information

Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS

Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS Application Note Biopharmaceuticals and Biosimilars Authors Emmie Dumont, Isabel Vandenheede, Pat Sandra, Koen Sandra Research

More information

Analysis of Monoclonal Antibody Digests with the Agilent 1290 Infinity 2D-LC Solution

Analysis of Monoclonal Antibody Digests with the Agilent 1290 Infinity 2D-LC Solution Analysis of Monoclonal Antibody Digests with the Agilent 1290 Infinity 2D-LC Solution Application Note Biotherapeutics & Biosimilars Authors Gerd Vanhoenacker, Koen Sandra, Isabel Vandenheede, Frank David,

More information

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column mab Titer Analysis with the Agilent Bio-Monolith Protein A Column Application Note Biopharmaceuticals and Biosimilars Authors Emmie Dumont, Isabel Vandenheede, Pat Sandra, and Koen Sandra Research Institute

More information

Gas Chromatography Assignment Chem 543/443

Gas Chromatography Assignment Chem 543/443 Gas Chromatography Assignment Chem 543/443 1. Introduction Capillary gas chromatography (GC) is one of the most popular analytical techniques used in today s research. Its popularity is mainly due to efficient

More information

High Resolution Analysis of Heparin and Heparin-like Impurities on Glycomix TM an Anion Exchange Column

High Resolution Analysis of Heparin and Heparin-like Impurities on Glycomix TM an Anion Exchange Column High Resolution Analysis of and -like Impurities on Glycomix TM an Anion Exchange Column Authors Xiaomi Xu Yuna Li Aiyan Cui Haiying Chen Sepax Technologies, Inc. 5 Innovation Way Newark, DE 19711 USA

More information

Agilent 6430 Triple Quadrupole LC/MS System

Agilent 6430 Triple Quadrupole LC/MS System Agilent 6430 Triple Quadrupole LC/MS System Ideal Quantitative LC/MS for UHPLC with Dynamic MRM and Fast Polarity Switching, plus Unsurpassed Sensitivity with Chip LC Summary The Agilent 6430 Triple Quadrupole

More information

Overview of preparative HPLC. Analytical Technologies Limited

Overview of preparative HPLC. Analytical Technologies Limited Overview of preparative HPLC Analytical Technologies Limited 1.Review of liquid phase chromatography origin develop prospect Mobile phase Tewett 1903 Stationa ry phase origin Separation result Liquid-stationary

More information

BÜCHI Labortechnik AG

BÜCHI Labortechnik AG BÜCHI Labortechnik AG Purification BUCHI Chromatography History 1977: BUCHI Fraction Collector 1978: 1 st preparative chromatography system 2008: automated chromatography system 2014: automated high performance

More information

Method Development Considerations for Reversed-Phase Protein Separations

Method Development Considerations for Reversed-Phase Protein Separations Method Development Considerations for Reversed-Phase Protein Separations Hillary B. Hewitson, Thomas E. Wheat, Paula Hong, Kenneth J. Fountain APPLICATION BENEFITS n The BEH00 C 4 chemistry is available

More information

Protein-Pak Hi Res HIC Column and HIC Protein Standard

Protein-Pak Hi Res HIC Column and HIC Protein Standard Protein-Pak Hi Res HIC Column and HIC Protein Standard CONTENTS I. INTRODUCTION II. a. Mobile Phase b. Flow Direction CONNECTING COLUMN TO LC SYSTEM I. INTRODUCTION This offering contains non-porous, polymethacrylate-based

More information

Part II: Column Temperature and Refractive Index Detection

Part II: Column Temperature and Refractive Index Detection Part II: Column Temperature and Refractive Index Detection K. Choiket and G. Rozing, Agilent Technologies Deutschland GmbH, Waldbronn, Germany. A systematic investigation of baseline disturbances in chromatographic

More information

Forced Degradation of Ibuprofen in Bulk Drug and Tablets

Forced Degradation of Ibuprofen in Bulk Drug and Tablets Forced Degradation of Ibuprofen in Bulk Drug and Tablets and Deteration of Specificity, Selectivity, and the Stability-Indicating Nature of the USP Ibuprofen Assay Method Sherri Farmer, Pamela Anderson,

More information

Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES

Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES THE AGILENT 7820A GC SYSTEM FORWARD-THINKING INNOVATION, EVERYDAY RELIABILITY AND VALUE Is your lab looking for better, more reproducible results

More information

On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water

On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water application note On-line SPE-LC/MS/MS of Pesticides On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water API 3200 LC/MS/MS System Overview The feasibility of

More information

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis Shanhua Lin, Julia Baek, and Chris Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA Overview

More information

A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts

A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts Andrew Aubin and Ronan Cleary Waters Corporation, Milford, MA, USA APPLICATIN BENEFITS A modular preparative HPLC

More information

Preparative HPLC System

Preparative HPLC System High Performance Liquid Chromatograph Preparative System Preparative HPLC System C90-E84 From Preparative LC to Purification and Powderization Purification Enters the Next Era Shimadzu provides a wide

More information

Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns

Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns Hongwei Xie, Martin Gilar, and Jeff Mazzeo Waters Corporation, Milford, MA U.S. APPLICATION BENEFITS The ACQUITY UPLC

More information

Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal Antibody Charge Variants

Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal Antibody Charge Variants Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal ntibody Charge Variants Ken Cook, Frank Steiner, Mauro De Pra Thermo Fisher Scientific, Hemel Hempstead,

More information

Genotoxicity is the property of a compound

Genotoxicity is the property of a compound Impurities Analysis in Pharmaceuticals: Genotoxicity is the property of a compound known to have irreversible effects on the structure and functionality of the DNA in cells and cause DNA loss, DNA replication

More information

Combining High Temperature and Small Particles: The Advantages of Zirconia

Combining High Temperature and Small Particles: The Advantages of Zirconia Combining High Temperature and Small Particles: The Advantages of Zirconia EAS 9 Dan owlan 1, Bingwen Yan 1, Clayton V. Mceff 1, R.A. Henry 1 ZirChrom Separations, Inc. 617 Pierce St., Anoka, M 5533 Independent

More information

Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental

Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental Samples Michael S.Young, Jeremy C. Shia, Kim vantran, Kevin M. Jenkins and Masayo Yabo Waters Corporation 34 Maple Street,

More information

Quantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy

Quantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy Advances in Bioscience and Biotechnology, 2010, 1, 367-371 doi:10.4236/abb.2010.15049 Published Online December 2010 (http://www.scirp.org/journal/abb/). Quantitative determination of residual 2-(2-chloroethoxy)

More information

Agilent 6500 Series Q-TOF LC/MS System

Agilent 6500 Series Q-TOF LC/MS System Agilent 6500 Series Q-TOF LC/MS System Quick Start Guide What s New in Data Acquisition 3 Where to find information 7 Getting Started 9 Step 1. Start the Data Acquisition software 10 Step 2. Prepare the

More information

AZURA Bio purification

AZURA Bio purification Science Together AZURA Bio purification The new standard of FPLC 1 KNAUER protein purification: The flexible FPLC platform AZURA Bio purification systems AZURA Bio purification: You choose the method Size

More information

Enhancing your HPLC Separations using Temperature Programming

Enhancing your HPLC Separations using Temperature Programming Enhancing your HPLC Separations using Temperature Programming Brian Jones, Stephanie J. Marin, Jody Clark, and Dale Felix Selerity Technologies, Inc. 2484 W. Custer Road Salt Lake City, UT 84104 www.selerity.com

More information

2017 by Agilent Technologies Agilent CrossLab Compliance Services

2017 by Agilent Technologies Agilent CrossLab Compliance Services EQP Name: Service Type: OQ Company Name: Customer Name/Title: EQP Filename:.eqp EQP Release Date: January 2017 Print Date: Page 1 / 16 Table of Contents Section Page Page 2 / 16 Scope and Purpose Overview

More information

2017 by Agilent Technologies Agilent CrossLab Compliance Services

2017 by Agilent Technologies Agilent CrossLab Compliance Services EQP Name: Service Type: OQ Company Name: Customer Name/Title: EQP Filename:.eqp EQP Release Date: July 2017 Print Date: Page 1 / 17 Table of Contents Section Page Page 2 / 17 Scope and Purpose Overview

More information

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime EASY-Spray Columns Guidance for column set up and installation Tips to maximize column lifetime EASY-Spray Column Tips and Tricks This document provides guidance for Thermo Scientific EASY-Spray column

More information

Analysis of Monoclonal Antibody Digests with the Agilent 1290 Infinity 2D-LC Solution Part 2: HILIC RPLC-MS

Analysis of Monoclonal Antibody Digests with the Agilent 1290 Infinity 2D-LC Solution Part 2: HILIC RPLC-MS Analysis of Monoclonal Antibody Digests with the Agilent 90 Infinity D-LC Solution Part : HILIC RPLC-MS Application Note Biotherapeutics & Biosimilars Authors Gerd Vanhoenacker, Koen Sandra,, Isabel Vandenheede,,

More information

Prominence Preparative HPLC System

Prominence Preparative HPLC System High Performance Liquid Chromatograph Preparative System Prominence Preparative HPLC System C190-E110B Prominence Preparative System [Product Concept] The most important aspects of separation and purification

More information

mabs and ADCs analysis by RP

mabs and ADCs analysis by RP mabs and ADCs analysis by RP Shanhua Lin, Ph.D. The world leader in serving science Protein and mab Separation by HPLC Size difference? YES Size Exclusion Chromatography (SEC) MAbPac SEC-1 NO NO Charge

More information

Fast Preparative Column Liquid Chromatography (PCLC)

Fast Preparative Column Liquid Chromatography (PCLC) Fast Preparative Column Liquid Chromatography (PCLC) Application Note 224 Joan Stevens, Ph.D. and Gary Scharrer Introduction Preparative HPLC is recognized as a prime method for obtaining pure compounds

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis S(-)-Nicotine, Primary Standard N-008 Page 1 of 7 Catalog Number: N-008 H Lot: Expiration: October 2019 N Description: S(-)-Nicotine in Methanol.

More information

Puzzled About LCMS? Sample. Sensitivity in Mass Spec Analysis. Prep. Adapting LC-UV. Optimizing and Maintaining Your Mass Spec LCMS

Puzzled About LCMS? Sample. Sensitivity in Mass Spec Analysis. Prep. Adapting LC-UV. Optimizing and Maintaining Your Mass Spec LCMS Puzzled About LCMS? Sensitivity in Mass Spec Analysis Sample Prep Adapting LC-UV Optimizing and Maintaining Your Mass Spec To LCMS Paul Altiero Alex Ucci Applications Phone Support Chemistry and Supplies

More information

Agilent 1260 Infinity High Performance Autosampler

Agilent 1260 Infinity High Performance Autosampler Agilent 1260 Infinity High Performance Autosampler User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2010-2013, 2014 No part of this manual may be reproduced in any form or by any means

More information

Elute LC series. Innovation with Integrity. UHPLC Precision for MS applications LC-MS

Elute LC series. Innovation with Integrity. UHPLC Precision for MS applications LC-MS Elute LC series UHPLC Precision for MS applications Innovation with Integrity LC-MS Delivering Precision and Speed for Mass Spectrometry Designed for reproducibility in retention time and peak shape, Elute

More information

AZURA Bio LC. 2 FPLC and Purification Solutions

AZURA Bio LC. 2 FPLC and Purification Solutions AZURA Bio LC 2 FPLC and Purification Solutions Flexible and adjustable to your biopurification tasks Fast protein liquid chromatography (FPLC) of bio molecules can be a task. KNAUER offers with AZURA an

More information

Application Note. Abstract. Authors. Introduction. Environmental

Application Note. Abstract. Authors. Introduction. Environmental EPA Method 538: Determination of Selected Organic Contaminants in Drinking Water by Direct Aqueous Injection with the Agilent 646 Triple Quadrupole LC/MS System Application Note Environmental Authors Imma

More information

Fundamentals and Techniques of Preparative HPLC. Parto Zist Behboud Tel: 42108

Fundamentals and Techniques of Preparative HPLC. Parto Zist Behboud  Tel: 42108 Fundamentals and Techniques of Preparative HPLC Parto Zist Behboud www.partoz.com Tel: 42108 Today s Overview 1 Basic know-how for up-scaling 2 Practical examples of up-scaling 3 Recycle preparation 4

More information

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph Analysis of USP Residual Solvents using the Agilent 797A Headspace Sampler with the Agilent 79B Gas Chromatograph Application Note Pharmaceuticals Authors Bart Tienpont, Frank David, and Pat Sandra

More information

The Theory of HPLC Gradient HPLC

The Theory of HPLC Gradient HPLC The Theory of HPLC Gradient HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual. 1 Contents

More information

Identification, expression and characterization of a R- -transaminase from Capronia semiimmersa.

Identification, expression and characterization of a R- -transaminase from Capronia semiimmersa. Applied Microbiology and Biotechnology Supporting Information for Identification, expression and characterization of a R- -transaminase from Capronia semiimmersa. César Iglesias 1,2, Paola Panizza 1,2

More information

Preparative HPLC: Factors and Parameters that Directly Affect Recovery of Collected Fractions

Preparative HPLC: Factors and Parameters that Directly Affect Recovery of Collected Fractions Preparative HPLC: Factors and Parameters that Directly Affect Recovery of Collected Fractions Gary Scharrer and Joan Stevens, Ph.D. Gilson, Inc Middleton, WI www.gilson.com Abstract Purification of synthetic,

More information

Development of Analysis Methods for Therapeutic Monoclonal Antibodies Using Innovative Superficially Porous Particle Biocolumns

Development of Analysis Methods for Therapeutic Monoclonal Antibodies Using Innovative Superficially Porous Particle Biocolumns Development of Analysis Methods for Therapeutic Monoclonal Antibodies Using Innovative Superficially Porous Particle Biocolumns Anne Blackwell Bio Columns Product Support Scientist Suresh Babu Senior Application

More information

Everyone can be an Expert. Electrochemical Detector. Autosampler. Complete HPLC-ECD

Everyone can be an Expert. Electrochemical Detector. Autosampler. Complete HPLC-ECD Everyone can be an Expert Electrochemical Detector Autosampler Complete HPLC-ECD Eicom ECD-700 Eicom ECD-700 Electrochemical Detector for HPLC Eicom's electrochemical detector, the ECD-700, offers superior

More information

ISSN India; g,secunderabad. Abstractt. a flow rate. of 1ml/min. di hydrogen. which acts. and chronic. including minimize (5) Figure

ISSN India; g,secunderabad. Abstractt. a flow rate. of 1ml/min. di hydrogen. which acts. and chronic. including minimize (5) Figure B.Lakshmi et al SPJTS,2013,1(1),015-023 RP-HPLC METHOD FOR THE QUANTIFICATION OF ROFLUMILAST IN FORMULATIONS ISSN 2321-4597 B..Lakshmi 1, Prof. T.V.Reddy 2 1.Kallam Haranadha Reddy Institute of Technology,NH-5,

More information

Antibody Analysis by ESI-TOF LC/MS

Antibody Analysis by ESI-TOF LC/MS Application Note PROTEOMICS METABOLOMICS GENOMICS INFORMATICS GLYILEVALCYSGLUGLNALASERLEUASPARG CYSVALLYSPROLYSPHETYRTHRLEUHISLYS Antibody Analysis by ESI-TOF LC/MS Authors Lorenzo Chen, Merck & Company,

More information

MAbPac SCX-10 Columns

MAbPac SCX-10 Columns User Manual MAbPac SCX-10 Columns 065393 Revision 09 October 2015 For Research Use Only. Not for use in diagnostic procedures. Product Manual for MAbPac SCX-10, 10µm Analytical Column (10μm, 9 x 250 mm,

More information

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover columns ProSwift Reversed-Phase Monolith Columns for Protein Analysis ProSwift reversed-phase columns use a unique monolith technology for fast, high-resolution HPLC and LC/MS separations of proteins.

More information

A green HPLC technique with a 100% water mobile phase for detecting imidacloprid and its metabolite 6-chloronicotinic acid

A green HPLC technique with a 100% water mobile phase for detecting imidacloprid and its metabolite 6-chloronicotinic acid Current Chemistry Letters 5 (2016) 27 32 Contents lists available atgrowingscience Current Chemistry Letters homepage: www.growingscience.com/ccl A green HPLC technique with a 100% water mobile phase for

More information

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates Julia Baek, Shanhua Lin, Xiaodong Liu, Thermo Fisher Scientific, Sunnyvale, CA Application Note 2126 Key Words Size exclusion

More information