SIPAN 2EL Conductivity Sensor Manual

Size: px
Start display at page:

Download "SIPAN 2EL Conductivity Sensor Manual"

Transcription

1 SIPAN 2EL Conductivity Sensor Manual DOC Mar05

2 1 Introduction Using this manual Using the sensor Warranty information 2 2 Description 3 3 Installation Installing the sensor Installation in a pipe Installation in a continuous-flow vessel Connecting the cable 4 4 Setup and operation Checking / specifying the cell constant Calibrating for the first time Temperature compensation Conductivity Explosion Protection 5 5 Servicing and maintenance Measuring errors Cleaning 6 6 Technical data 7 2EL conductivity sensor 7MA2000-8P_0 1

3 Dear customer: Please read these instructions carefully before beginning operation of the sensor! These instructions provide you with important information and data you need to ensure proper functioning of the sensor and reduce maintenance costs. Following these instructions will help you to operate the sensor more easily and efficiently, allowing you to achieve reliable measuring results. 1 Introduction These instructions provide you with information on using, installing, operating, and maintaining the sensor. The sensor described here is usually operated using a control device / measuring transducer. For this reason, it is important to familiarize yourself with the proper operation of the control device / measuring transducer in order to ensure successful operation of the sensor. In particular, before using the device for new applications in the area of research and development, we recommend first contacting us to discuss the application in question. 1.1 Using this manual Pay particular attention to all special information and warnings. Information of this type is set apart from the rest of the text and labeled with the appropriate symbols (see example on left). This information provides you with useful tips and helps you to avoid improper use of the device. 1.2 Using the sensor Procedures such as proper transportation, storage, installation, operation and maintenance of the product are vital for ensuring correct and safe operation. Using the sensor in a way that is not described in these instructions can pose serious risks for the user. In addition, improper use of the equipment renders the device warranty null and void. 1.3 Warranty information The contents of this instruction manual do not constitute part of a past or existing agreement, promise, or legal relationship nor are they intended to modify these. All obligations of HACH LANGE GmbH originate from the purchase contract in question, the contents of which also include the full and exclusively valid warranty provisions. The contractual warranty provisions are neither extended nor in any way restricted by the information contained in this instruction manual. 2 2EL conductivity sensor 7MA2000-8P_0

4 2 Description The sensor is composed of two conductive electrodes. The electrodes are insulated from one another and arranged in a specific geometric pattern. The electrodes determine the cell constant. An alternating current is applied to the electrodes. The flow of the current is a measurement of the conductivity of the solution. Since the measuring results are affected by the temperature of the solution, the measurement must be made for a particular reference temperature (generally 25 C/77 F) so that the results can be compared. A Pt100 temperature probe is integrated into the sensor. 3 Installation 3.1 Installing the sensor Installation in a pipe When installing the sensor in a pipe, it is important to make sure that the inner electrode is completely immersed in the medium conducted through the pipe. The illustration on the left shows various examples of how the sensor can be installed in a pipe. The arrows indicate the direction in which the medium in the pipe is flowing. The letters in the illustration signify the following: A) Excellent positioning of the sensor, optimum contact with the flowing medium, and optimum rate of medium replenishment on the electrode surface. B) Possible positioning of the sensor, sufficient contact with the flowing medium, and sufficient rate of medium replenishment on the electrode surface. C) Poor positioning of the sensor. The electrodes have only partial contact with the flowing medium and the transmitted conductivity value is insufficient Installation in a continuous-flow vessel It is possible to install the sensor in a continuous-flow vessel (7MA2000-8PX1). For this type of installation, it is important to ensure a minimum flow rate of 20 l/h (5 ¼ USGal/h) (the optimum flow rate is 60 l/h, 15 ¾ USGal/h) in order to avoid the creation of bubbles on the surface of the probe. 2EL conductivity sensor 7MA2000-8P_0 3

5 The illustration shows various examples of how the sensor can be installed in a continuous-flow vessel. The letters in the illustration signify the following: A) Excellent positioning of the sensor, optimum contact with the flowing medium, and optimum rate of medium replenishment on the electrode surface. B) Possible positioning of the sensor, sufficient contact with the flowing medium, and sufficient rate of medium replenishment on the electrode surface. C) Poor positioning of the sensor. The electrodes have only partial contact with the flowing medium and the transmitted conductivity value is insufficient. 3.2 Connecting the cable When connecting the cable to the plug, it is important to make sure that there is no moisture in the plug and that no moisture can form in the plug. Connect the sensor cable to the measuring transducer as specified in the connection table. Connection on SIPAN Terminal number on measuring transducer 7MA204_-8A_ 7MA2034- _AA_ MA2034- _AB_0-0 0 Wire color Signal SIPAN 32 SIPAN 34 field housing SIPAN 34 housing insert Blue Pt Black Pt Red Outer electrode White Inner electrode (yellow end sleeve) White Inner shield 15 Inner plate Metal housing (orange end sleeve) White (red end sleeve) Outer shield Grounding plate shielding Inner plate shielding Metal housing 4 2EL conductivity sensor 7MA2000-8P_0

6 4 Setup and operation The sensor is always operated using a SIPAN 32 or SIPAN 34 measuring transducer. For this reason, it is important to follow the operating instructions provided for these devices to ensure proper functioning of the sensor. 4.1 Checking / specifying the cell constant The cell constant is a specific sensor value required for converting the measured resistance value into a conductivity value. To check the cell constant, you must use the value specified on the sensor. This value was calculated with an accuracy deviation of less than 2% according to ASTM N 1125 and ISO 7888 standards. The cell constant is specified in the measuring transducer menu. 4.2 Calibrating for the first time Temperature compensation For these sensors, the integrated temperature sensor is fitted as a two-wire connection. In this arrangement, the measured value is affected by the resistance of the sensor cable. To compensate for the cable resistance, the temperature must be calibrated before startup. To do this, proceed as follows: 1. Immerse the sensor for approx. ten minutes in a solution (e.g. distilled water) heated to room temperature. 2. Measure the temperature of the solution using a second calibrated thermometer having a resolution of 0.1 C/0.18 F or better. 3. Set the SIPAN measuring transducer to operating mode 'Calibrate' (CAL). 4. Enter the temperature shown on the thermometer as the calibration value on the SIPAN measuring transducer. For detailed information on temperature compensation, please refer to the instruction manual of the SIPAN measuring transducer you are using Conductivity To calibrate the conductivity, please read the relevant sections of the measuring transducer instruction manuals. 4.3 Explosion Protection The sensor is defined as standard electrical equipment as per EN 50020, Section 5.4. The only source of heat is the integrated Pt100 temperature probe, which is defined as a small component as per EN 50020, Table 3. As a result, the sensor can be operated under a maximum operating temperature of 125 C/257 F in temperature class T4. 2EL conductivity sensor 7MA2000-8P_0 5

7 5 Servicing and maintenance The conductivity sensors are extremely reliable and only require a minimum amount of calibration. 5.1 Measuring errors If you detect any measuring errors, we recommend first checking the following: A) Cable connections B) Measuring transducer settings C) Probe (Pt100 and electrodes). The following illustration shows the correct position of the connections: Electrodes Pt100: Measure the resistance of the Pt100 at the point of connection and compare the level of resistance with the (correct) values shown in the following table: Temperature in C Temperature in F Conductivity in Ω Electrodes: Check the insulation between the two electrodes (if the sensor is dry and exposed to air: resistance R = ). 5.2 Cleaning Insulating deposits (e.g. calcium deposits or oil film) formed on the electrodes during operation can affect the measuring results. If this occurs, you must clean the sensor. In most cases, you can clean the sensor using hot water and an acidic cleaning agent. Oil films can usually be removed using ethyl alcohol or propyl alcohol. Residue from solutions containing bacteria or algae should be removed using a cleaning agent containing chlorine. Deposits of metallic hydroxide must be treated with 20% nitric acid (in a ten-minute immersion bath). After cleaning the sensor, we recommend checking the calibration. 6 2EL conductivity sensor 7MA2000-8P_0

8 6 Technical data Sensor 7MA2000-8PA0 7MA2000-8PB0 7MA2000-8PC0 Cell constant 0.01 cm cm -1 1 cm -1 Smallest measured value 0.01 µs/cm 0.1 µs/cm 1 µs/cm Largest measured value 50 µs/cm 500 µs/cm 5000 µs/cm Electrode arrangement Concentric Concentric Plane parallel Vesel Factor 1,00 1,00 1,00 Electrode material and material contacting the medium Stainless steel (316 L SS) and PES (polyethersulfone) Stainless steel (316 L SS) and PES (polyethersulfone) 125 C/257 F Permitted operating temperature T B Permitted operating pressure P B 1000 kpa (10 bar, 145 PSI) Thermometer integrated Pt100 Connection cable Can be plugged in separately Process adaptation ¾" NPT Installation length 57 mm Diameter 16 mm Ø Protection class (with cable) IP65 Pressure equipment directive GAS 2/LIQUIDS 1 ART. 3.3 SEP Graphite and PES (polyethersulfone) 150 C/300 F Continuous-flow vessel 7MA2000-8PX1 Material Stainless steel (316 L SS) Permitted operating temperature T B Permitted operating pressure P B 2500 kpa (25 bar, 365 PSI) Sensor connection ¾" NPT Process connection ¼" NPT Total length 165 mm External diameter 34 mm Ø 2EL conductivity sensor 7MA2000-8P_0 7

9 Data and dimension sheets 7MA2000-8PA0 7MA2000-8PB0 7MA2000-8PC0 7MA2000-8PX1 8 2EL conductivity sensor 7MA2000-8P_0

10 Order Number Converting Table SIEMENS Order-Number Short Description HACH LANGE Order-Number 7MA1034* SIPAN 34 ph-analyzer 7MA1034* 7MA2034* SIPAN 34 LF-Analyzer 7MA2034* 7MA3034* SIPAN 34 O2-Analyzer 7MA3034* 7MA1040* SIPAN 32 ph-analyzer, ein Eingang 7MA1040* 7MA1041* SIPAN 32 X ph-analyzer, ein Eingang 7MA1041* 7MA1140* SIPAN 32 ph-analyzer, zwei Eingänge 7MA1140* 7MA1141* SIPAN 32 X ph-analyzer, zwei Eingänge 7MA1141* 7MA2040* SIPAN 32 LF-Analyzer, 2-EL 7MA2040* 7MA2041* SIPAN 32 X LF-Analyzer, 2-EL 7MA2041* 7MA2140* SIPAN 32 LF-Analyzer, 4-EL 7MA2140* 7MA2141* SIPAN 32 X LF-Analyzer, 4-EL 7MA2141* 7MA2240* SIPAN 32 LF-Analyzer, IND 7MA2240* 7MA2241* SIPAN 32 X LF-Analyzer, IND 7MA2241* 7MA3040* SIPAN 32 O2-Analyzer 7MA3040* 7MA3041* SIPAN 32 X O2-Analyzer 7MA3041* 7MA2000-8AA repl. Z38310=A=0000 (1) 7MA2000-8AB repl. Z38310=A=0000 (1) 7MA2000-8BA repl. Z38311=A=0000 (1) 7MA2000-8BB repl. Z38311=A=0000 (1) 7MA2000-8CA repl. Z38312=A=0000 (1) 7MA2000-8CB repl. Z38312=A=0000 (1) 7MA2000-8DD repl. Z38311=A=0000 (1) 7MA2000-8PA0 2 EL CONDUCTIVITY CELL K=0,01 Z38310=A=0000 7MA2000-8PB0 2 EL CONDUCTIVITY CELL K=0,1 Z38311=A=0000 7MA2000-8PC0 2 EL CONDUCTIVITY CELL K=1 Z38312=A=0000 7MA2000-8PX1 FLOW THROUGH CELL F. 2 EL CONDUCTIVITY CELL Z08318=A=0001 7MA2000-8PX2 CABLE WITH PLUG F. 2 EL CONDUCTIVITY CELL; 5M Z08319=A=0005 7MA2000-8PX3 CABLE WITH PLUG F. 2 EL CONDUCTIVITY CELL; 20M Z038319=A=0020 7MA2000-8PX4 CONNECTOR F. WELDING (STAINLES STEEL); 3/4 INCH FNPT THREAD LZY108 7MA2000-8PX5 DUMMY PLUG (STAINLES STEEL); 3/4 INCH NPT THREAD LZY109 7MA2100-8AB MA2100-8BB MA2100-8BC 4EL-SENSOR F. CONDUCTIVITY MEASUREMENT; MOUNTING FLANGE DN 50 7MA21008BC

11 7MA2100-8CA 4EL-SENSOR F. CONDUCTIVITY MEASUREMENT; FIXABLE 12 MM ELECTRODE 7MA21008CA 7MA2200-8BA repl. 7MA22008BF 7MA2200-8BD MA2200-8BF IND SENSOR F. CONDUCTIVITY MEASUREMENTS; MOUNTING FLANGE DN50 7MA22008BF 7MA2200-8CB IND-SENSOR F. CONDUCTIVITY MEASUREMENT; VARIVENT 60 MM 7MA22008CB 7MA2200-8DA IND-SENSOR F. CONDUCTIVITY MEASUREMENT; MOUNTING FLANGE DN50 7MA22008DA 7MA2200-8DD repl. 7MA22008CB 7MA2200-8EA repl. 7MA22008EB 7MA2200-8EB IND-SENSOR F. CONDUCTIVITY MEASUREMENT IN THE EX-AREA (EEX IB IIC T4); MOUNTING FLANGE DN50 7MA22008EB 7MA2200-8FA CALIBRATION SET F. INDUCTIVE CONDUCTIVITY SENSORS 7MA2200-8BA, -8CB, -8DA, -8DD, _8EA, AND -8EB LZY011 7MA3100-8CA repl. LZY072 7MA3100-8CC SENSOR F. O2 IN ULTRA-PURE WATER ONLY F. USE WITH SIPAN 34 LZY012 7MA3100-8CD MA3100-8CE (2) 7MA3100-8CF (2) 7MA3100-8EF SENSOR F. O2 IN ULTRA-PURE WATER INCL. FLOW FITTING ONLY F. USE WITH SIPAN 34 LZY013 7MA3100-8HA OXYGOLD G DO SENSOR EX-LABEL II 1/2G EEX IA II C T4/T5/T6;PG 13,5 THREAD LZY072 7MA3100-8HB OXYGOLD G MEMBRANE KIT F. OXYGOLD G DO SENSOR (ELECTROLYTE 7MA3100-8HC/LZY072 PLEASE ORDER SEPARATELY) LZY073 7MA3100-8HC OXYLYTE G-ELECTROLYTE F. OXYGOLD G DO SENSOR (50 ML) LZY074 7MA3100-8HD OXYFERM VP DO SENSOR EX-LABEL:. II 1/2G EEX IA II C T4/T5/T6; PG 13,5 TREAD CABLE LZY075 7MA3100-8HE OXYFERM MEMBRANE KIT F. OXYFERM VP DO SENSOR LZY076 7MA3100-8HF OXYLYTE-ELECTROLYTE F. OXYFERM DO SENSOR (50 ML) LZY077 7MA3100-8HG OXYSENS DO SENSOR F. WATER AND WASTEWATER EX-LABEL:. II 1/2G EEX IA II C T4/T5/T6; PG 13,5 THREAD LZY078 7MA3100-8HH CABLE WITH VP PLUG CONNECTOR F. OXYFERM VP AND OXYGOLD G DO SENSORS LZY079 7MA3100-8HJ FLOW FITTING F. DO SENSORS (STAINLES STEEL); PG 13,5 THREAD LZY080 7MA3100-8HK POLARIZATION MODULE G F. OXYFERM VP AND OXYGOLD G DO SENSORS LZY081 7MA3100-8HL 2EL-SENSOR WITH TWO GRAPHITE ELECTRODES F. CONDUCTIVITY MEASUREMENT; PG 13,5 THREAD LZY082

12 7MA55030AA MA55030AA MA8022-8AA MA8022-8BA MA8500-8AA PT100 COMPENSATION THERMOMETER 7MA85008AA 7MA8500-8AB FLOW FITTING (PTFE/TEFLON)/GF25 7MA85008AB 7MA8500-8AC VARIVENT CLAMPING RING INCL. SCREWS AND NUTS (STAINLESS STEEL) LZY083 7MA8500-8AD VARIVENT FITTING AS FLOW FITTING (STAINLESS STEEL), F. PIPELINES DN50 LZY084 7MA8500-8AE SEALING RING, VARIVENT (STAINLESS STEEL) LZY085 7MA8500-8AF MA8500-8AG VARIVENT IMMERSION FITTING WITH SCREW CONNECTOR F. WELDING ON F. IND CONDUCTIVITY SENSORS 7MA2200-8CD AND -8DD LZY086 7MA8500-8AH O-RINGS (EPOM) F. IND CONDUCTIVITY SENSORS 7MA2200-8CB AND -8DD (VARIVENT) LZY087 7MA8500-8AJ VARIVENT SPECIAL GASKET (VITON) F. IND CONDUCTIVITY SENSORS 7MA2200-8CB AND -8DD LZY088 7MA8500-8AK (2) 7MA8500-8AM FLOW FITTING, COMPLETE, F. DISSOLVED OXYGEN IN ULTRA- PURE WATER LZY202 7MA8500-8AS BUFFER PH 4,02 BUFFER PH 6,88 BUFFER PH 9,22 repl. Z363130,00500 repl. Z363131,00500 repl. Z363132, MA8500-8AU MA8500-8AV MA8500-8AW MA8500-8BG MA8500-8BH MA8500-8BJ MA8500-8BP (2) 7MA8500-8BQ (2) 7MA8500-8BS MA8500-8BT DUMMY SCREW SEAL (STAINLESS STEEL) LZY233 7MA8500-8BV MA8500-8BX DIFFERENTIAL-PH-SENSOR LZY015 7MA8500-8BY WELDING-TYPE CONNECTOR (STAINLESS STEEL) WITH UNION NUT LZY016 7MA8500-8CC CASE WITH ACCESSORIES: 1 SPARE DIAPHRAGM HEADS, ELY/A-AG/OXY ELECTROLYTE, RL-AG/OXI CLEANSING AGENT AND SF300 GRINDING FOIL LZY017

13 7MA8500-8CD MA8500-8CG (2) 7MA8500-8CH (2) 7MA8500-8CJ (2) 7MA8500-8CX DIFFERENTIAL PH SENSOR 7MA8500-8FM/LZY030 WITH PLUG CABLE COMBINATION 7MA8500-8DR/LZY022 MOUNTED ON A WELDING-TYPE CONNECTOR LZY018 7MA8500-8DG PIPE CLAMP (STAINLESS STEEL) LZY019 7MA8500-8DH VARIVENT SEALING UNIT (STAINLESS STEEL) LZY089 7MA8500-8DJ MA8500-8DK MA8500-8DL CLEANING ATTACHMENT F. O2 SENSORS 7MA3100-8CC/LZY022, -8CD AND -8CE 7MA85008DL 7MA8500-8DN repl. 7MA85008GC 7MA8500-8DP SPECIAL CABLE (SENSOR MONITORING PH) LZY020 7MA8500-8DQ PLUG-CABLE COMBINATION F. PH COMBINATION ELECTRODE, INCL. PT1000 LZY021 7MA8500-8DR SPECIAL PLUG/CABLE COMBINATION, F. DIFFERENTIAL PH SENSORS LZY022 7MA8500-8DS repl. LZY082 7MA8500-8DT DUMMY SEAL (STAINLESS STEEL) F. WELDING-TYPE CONNECTOR 7MA8500-8BY/LZY016 LZY090 7MA8500-8DU repl. C20C270 7MA8500-8EA MA8500-8EC WELDING-TYPE CONNECTOR (STAINLESS STEEL) 15 ANGLED LZY234 7MA8500-8ED MA8500-8EE MA8500-8EH WELDING-TYPE CONNECTOR (STAINLESS STEEL) STRAIGHT LZY235 7MA8500-8FA PH COMBINATION ELECTRODE F. SERVICE WASTE WATER LZY023 7MA8500-8FB repl. LZY023 7MA8500-8FC PH COMBINATION ELECTRODE F. BOILER FEEDWATER, ULTRA PURE WATER AND CONTAMINATED WATER LZY025 7MA8500-8FD repl. LZY023 7MA8500-8FE MA8500-8FF PH COMBINATION ELECTRODE F. SERVICE WASTE WATER WITH PG 13,5 SCREW PLUG CONNECT LZY027 7MA8500-8FG ORP COMBINATION ELECTRODE LZY028 7MA8500-8FH RESISTANCE THERMOMETER PT1000 F. COMBINATION WITH PH COMBINATION ELECTRODES 7MA8500-8FC/LZY025, -8FB, -8FE AND -8BF LZY029 7MA8500-8FJ

14 7MA8500-8FK MA8500-8FM DIFFERENTIAL PH SENSOR, INTEGRATED PT100 LZY030 7MA8500-8FR RECTRACTABLE FITTINGS F. IN-LINE APPLICATIONS LZY236 7MA8500-8FS RECTRACTABLE FITTINGS F. IN-LINE APPLICATIONS LZY237 7MA8500-8FT RECTRACTABLE FITTINGS F. IN-LINE APPLICATIONS LZY238 7MA8500-8FU MA8500-8FV MA8500-8FW MA8500-8FY MA8500-8GC PLUG/CABLE COMBINATION F. COMBINATION ELECTRODES, MEASURING AND REFERENCE ELECTRODES WITH AG/AGCL TAPPING SYSTEM AND F. PT1000 LZY031 7MA8500-8GD repl. 7MA85008DP 7MA8500-8GE BUFFER PH 4,02 7MA8500-8GH BUFFER PH 6,88 BUFFER PH 9,22 repl. Z363130,00500 repl. Z363131,00500 repl. Z363132,00500 A5E RETROFIT SET SIPAN 32 PROFIBUS PA, VERSION WITHOUT EXPLOSION PROOF LZY208 A5E RETROFIT SET SIPAN 32 X WITH PROFIBUS PA EX LZY209 A5E PROFIBUS PA M12 PLUG, ROUND, FEMALE, TO CONNECT ON CABLE F. SIPAN 32 LZY210 A5E A5E PROFIBUS PA M12 PLUG, ROUND MALE, MOUNTING ON SIPAN 32 HOUSING LZY211 C70153-A503-C C70211-A1959-A C71165-Z358-P C71451-Z500-L2 KCL IN PLASTIC BOTTLE (1 KG) LZY091 C74450-A184-A1 KCL RESERVOIR PRESSURE EQUIPMENT LZY092 C74450-A184-D1 CONNECTION HOSE F. KCL RESERVOIR LZY093 C74451-A1789-A1 FLOW CHAMBER WITH THREADED SOCKET (STAINLES STEEL) LZY032 C74451-A1789-A10 IMMERSION TUBE (POLYVINYL CHLORIDE) MAX. IMMERSION LENGTH 600 MM LZY094 C74451-A1789-A11 C74451-A1789-A12 C74451-A1789-A13 C74451-A1789-A14 C74451-A1789-A15 C74451-A1789-A16 IMMERSION TUBE (POLYVINYL CHLORIDE) MAX. IMMERSION LENGTH 1800 MM LZY095 C74451-A1789-A2 FLOW CHAMBER WITH FLANGE LZY033 repl. LZY094 / LZY095 repl. LZY094 / LZY095 repl. LZY094 / LZY095 repl. LZY094 / LZY095 repl. LZY094 / LZY095

15 C74451-A1789-A21 FLOW CHAMBER WITH THREADED SOCKET (STAINLESS STEEL) LZY034 C74451-A1789-A22 repl. LZY40 (3) C74451-A1789-A C74451-A1789-A3 FLOW CHAMBER WITH THREADED SOCKET (POLYPROPYLENE) LZY096 C74451-A1789-A C74451-A1789-B1 ELECTRODE HOLDER (POLYPROPYLENE) LZY097 C74451-A1789-B2 ELECTRODE HOLDER (STAINLESS STEEL), WITH UNION NUT LZY098 C74451-A1789-B3 ELECTRODE HOLDER (POLYVINYLIDENFLUORIDE), WITH UNION NUT LZY099 C74451-A1789-C2 repl. LZX041 C74451-A1789-D1 SET OF MOUNTING PARTS (MOUNTING BRACKET AND HOSE CLAMP) C74451A1789D1 C74451-A1789-D C74451-A1789-D C79000-B5400-C45 MANUAL SIPAN 32, COND, D repl. DOC C79000-B5400-C46 MANUAL SIPAN 32, ph, D repl. DOC C79000-B5400-C47 MANUAL SIPAN 32, O2, D repl. DOC C79000-B5472-C45 MANUAL SIPAN 32, COND, I repl. DOC C79000-B5472-C46 MANUAL SIPAN 32, ph, I repl. DOC C79000-B5472-C47 MANUAL SIPAN 32, O2, I repl. DOC C79000-B5476-C C79000-B5476-C C79000-B5476-C45 MANUAL SIPAN 32, COND, GB repl. DOC C79000-B5476-C46 MANUAL SIPAN 32, ph, GB repl. DOC C79000-B5476-C47 MANUAL SIPAN 32, O2, GB repl. DOC C79000-B5477-C45 MANUAL SIPAN 32, COND, F repl. DOC C79000-B5477-C46 MANUAL SIPAN 32, ph, F repl. DOC C79000-B5477-C47 MANUAL SIPAN 32, O2, F repl. DOC C79000-B5478-C45 MANUAL SIPAN 32, COND, E repl. DOC C79000-B5478-C46 MANUAL SIPAN 32, ph, E repl. DOC C79000-B5478-C47 MANUAL SIPAN 32, O2, E repl. DOC C79000-G5400-C41 MANUAL SIPAN 34, COND, D repl. DOC C79000-G5400-C42 MANUAL SIPAN 34, ph, D repl. DOC C79000-G5400-C43 MANUAL SIPAN 34, O2, D repl. DOC C79000-G5464-C50 MANUAL ON CD, ALL LANGUAGES repl. DOC C79000-G5472-C41 MANUAL SIPAN 34, COND, I repl. DOC C79000-G5472-C42 MANUAL SIPAN 34, ph, I repl. DOC C79000-G5472-C43 MANUAL SIPAN 34, O2, I repl. DOC

16 C79000-G5476-C41 MANUAL SIPAN 34, COND, GB repl. DOC C79000-G5476-C42 MANUAL SIPAN 34, ph, GB repl. DOC C79000-G5476-C43 MANUAL SIPAN 34, O2, GB repl. DOC C79000-G5477-C41 MANUAL SIPAN 34, COND, F repl. DOC C79000-G5477-C42 MANUAL SIPAN 34, ph, F repl. DOC C79000-G5477-C43 MANUAL SIPAN 34, O2, F repl. DOC C79000-G5478-C41 MANUAL SIPAN 34, COND, E repl. DOC C79000-G5478-C42 MANUAL SIPAN 34, ph, E repl. DOC C79000-G5478-C43 MANUAL SIPAN 34, O2, E repl. DOC C79195-A3453-N100 EXTENSION CABLE F. 4-EL SENSORS, LENGTH 10 M LZY239 C79195-A3453-N200 repl. LZY239 / LZY240 C79195-A3453-N300 EXTENSION CABLE F. 4-EL SENSORS, LENGTH 30 M LZY240 C79451-A3177-D C79451-A3177-D11 BASE PLATE LZY035 C79451-A3177-D12 PROTECTIVE HOOD WITH BASE PLATE C79451-A3177-D11/LZY035 (STAINLESS STEEL) LZY036 C79451-A3246-D C79451-A3298-N C79451-A3298-N C79451-A3298-N C79451-A3300-N100 EXTENSION CABLE F. IND CONDUCTIVITY SENSORS, LENGTH 10 M LZY101 C79451-A3300-N200 repl. LZY101 / LZY102 C79451-A3300-N300 EXTENSION CABLE F. IND CONDUCTIVITY SENSORS, LENGTH 30 M LZY102 C79451-A3302-B6 THERMOMETER PROTECTIVE TUBE (PTFE) LZY103 C79451-A3522-D M54145-A M54145-A15-A M54145-A15-A6 PLUG/CABLE COMBINATION F. COMBINATION AND REFERENCE ELECTRODES LZY037 M54145-A15-A11 repl. M54145A15A6 M54145-A M54145-A M54145-A M54145-A M54145-A M54145-A48 EARTHING PIN WITH SCREW PLUG CONNECTOR F. PH SENSOR MONITORING LZY104 M54145-A

17 M54145-A M54145-A M54145-A92 FLOW FITTING WITH THREADED COUPLING AND GASKET (POLYPROPYLEN) LZY038 M54145-A93 FLOW FITTING WITH THREADED COUPLING AND GASKET (PVDF) LZY039 M54145-A102 FLOW FITTING (POLYPROPYLENE) LZY040 M54445-A M54445-A M54445-A23 UNION NUT DN50 (STAINLESS STEEL) LZY041 M54445-A24 SET OF GASKETS (VITON) (5 PIECES) LZY042 M54445-A25 FITTING F. INSTALLATION IN PIPELINES (STAINLESS STEEL) LZY043 M54445-A M54445-A M54445-A M54445-A M54445-A M54445-A M54445-A M54445-A M54445-A S79610G51A S79610G53A S79610G54A S79610G55A S79610G56A obsolete, without replacement (1) Replacement is not mechanically compatible. An accordingly cable and flow through chamber might be requiciered (2) Compareable products within the HACH LANGE product portfolio. However not necessarily compatible. (3) Different specifications (4) No suitable waste water DO sensor for this instrument available