Instructions. Modular Bipolar Instruments 83930xxx 84930xxx. GA -B 241 /USA--3.0/ Index: / ÄM: PDZ

Size: px
Start display at page:

Download "Instructions. Modular Bipolar Instruments 83930xxx 84930xxx. GA -B 241 /USA--3.0/ Index: / ÄM: PDZ"

Transcription

1 Instructions Modular Bipolar Instruments 83930xxx 84930xxx GA -B 241 /USA--3.0/ Index: / ÄM: PDZ

2 Important general instructions for use Ensure that this product is used only as intended and described in this instruction manual, by adequately trained and qualified personnel, and that maintenance and repair is only carried out by authorized specialized technicians. Use this product only with the combinations and with the accessories and spare parts listed in this instruction manual. Use other combinations, accessories and replacement parts only if they are expressly intended for this use and if the performance and safety requirements are met. Reprocess the products before every application and before returning them for repair as required by the instruction manual in order to protect the patient, user or third parties. Subject to technical changes! Due to continuous development of our products, illustrations and technical data may deviate slightly from the data in this manual. Symbol CAUTION : Federal law restricts this device to sale by or on the order of a physician. Level of danger Safety instructions and levels of danger WARNING! Failure to observe can result in death or serious injury. CAUTION! Failure to observe can result in slight injury or damage to the product.. IMPORTANT! Failure to observe can result in damage to the product or surrounding.. NOTE! Tips for optimum use and other useful information. USA RICHARD WOLF Medical Instruments Corp. 353 Corporate Woods Parkway Vernon Hills, Illinois Telephone: Telefax: sales&marketing@richardwolfusa.com GERMANY RICHARD WOLF GmbH Knittlingen Pforzheimerstr. 32 Telephone: Telefax: MANUFACTURER info@richard--wolf.com UK RICHARD WOLF UK Ltd. Waterside Way Wimbledon SW17 0HB Telephone: Telefax: admin@richardwolf.uk.com BELGIUM / NETHERLANDS N.V. Endoscopie RICHARD WOLF Belgium S.A. Industriezone Drongen Landegemstraat Gent Drongen Telephone: Telefax: endoscopy@richard--wolf.be FRANCE RICHARD WOLF France S.A.R.L. Rue Daniel Berger Z.A.C. La Neuvillette Reims Telephone: Telefax: endoscopes@richardwolf.fr AUSTRIA RICHARD WOLF Austria Ges.m.b.H. Wilhelminenstraße 93 a 1160 Vienna Telephone: Telefax: info@richard--wolf.at Marketing Office U.A.E RICHARD WOLF Middle East P.O. Box AL Thuraya Tower 1 9 th Floor, Room 904, Dubai Telephone: Telefax: middle.east@richard--wolf.com INDIA RICHARD WOLF India Private Ltd. JMD Pacific Square No. 211 A, Second Floor Behind 32 nd Milestone Gurgaon National Capitol Region Telephone: Telefax: india@richard--wolf.com 0 GA--B 241

3 Contents 1 Technical description Intended use Indications and field of use Contraindications Combinations Illustration Legend and identification Use Preparation Disassembly Assembly Additional notes and instructions for use HF application Checks Visual check Function check Function test Reprocessing and maintenance Manual reprocessing Machine reprocessing Checks Sterilization Steam sterilization Gas sterilization Technical data and order data Accessories Operating, storage, transport and shipping conditions Disposal of product, packaging material and accessories Warranty and Customer Service GA--B 241 I

4 1 Technical description The Modular Bipolar Instruments consist of the following components: Z Jaw insert Z Sheath tube Z Handle 2 Intended use Under endoscopic view: Z Grasping, manipulating, cutting. Z Dissecting of soft tissue parts/organs. Z Coagulation, hemostasis, severing of tissue by means of bipolar HF current. 3 Indications and field of use For minimally invasive diagnosis and/or therapy under endoscopic control using suitable accessories in the various medical disciplines such as surgery, urology, thoracoscopy, gynecology, by adequately trained and qualified medical personnel. 4 Contraindications 5 Combinations Contraindications directly related to the product are presently unknown. On the basis of the patient s general condition the physician/surgeon in charge must decide whether the planned use is possible or not. For further information see the latest medical literature. The modular bipolar instruments are used in conjunction with HF devices/bipolar contact coagulation and the corresponding HF cables as well as endoscopic accessories such as operating laparoscopes, trocar sleeves etc.. CAUTION! Caution if products are incorrectly combined! Possible injury of the patient, user or others as well as possible damage to the product. Different products should only be used in combination if their intended uses and relevant technical data (working length, diameter, peak voltage, etc.) are the same. Follow the instruction manuals of the products used in conjunction with this product. Follow the Notes and instructions on HF applications, order no.: GA- S 002 as well as the HF device manufacturer s instructions. 1 GA--B 241

5 6 Illustration Consisting of : Legend and identification Item Designation Item Designation 1 Jaw insert 3 Handle 1.1 Jaw/scissor section 3.1 Bipolar H.F. connector 2 Sheath tube 3.2 Star wheel, rotatable 2.1 Sleeve 3.3 Movable grip GA--B 241 2

6 Symbols Meaning Attention, consult ACCOMPANYING DOCUMENTS REF Order number Lot identification number SN Series number Identification in conformity with Medical Devices Directive 93/42/EEC only valid if the product and/or packaging is marked with this symbol. Products of category IIa and above, as well as sterile products or products with measuring function of category I, are additionally marked with the code number of the notified body (0124). 3 GA--B 241

7 7 Use 7.1 Preparation 7.2 Disassembly CAUTION! The products have only limited strength! Exerting excessive force will cause damage, impair the function and therefore endanger the patient. Immediately before and after each use, check the products for damage, loose parts and completeness. Ensure that no missing instrument parts remain in the patient. Do not use products which are damaged, incomplete or have loose parts. Z Check assembly: section 7.3 Z Perform a visual check: section 7 and 7.1 Z Connect the instrument to the HF generator using the HF connection cable (see accessories). Z Check the power setting on the HF generator. Z Perform a functional check: section 7.2 Z Pull the sleeve (2.1) on the sheath tube (2) in the direction indicated by the arrow. ' The sheath tube (2) disengages from the handle (3). This causes the movable grip (3.3) to move upwards automatically.. NOTE! Do not hold the movable grip (3.3) during disassembly Z Pull the jaw insert (1) out of the sheath tube (2) while turning it slightly. 1 2 GA--B 241 4

8 7.3 Assembly. IMPORTANT! Before assembly make sure that the contacts (x) of the proximal end of the jaw insert (1) are clean, dry and undamaged. Z Insert the jaw insert (1) into the sheath tube (2) while turning it slightly. ' The twist lock mechanism locks in place. x 1 2 Z Insert the sheath tube (2) with jaw insert (1) in place into the handle (3) until the connection snaps into position with a click. ' During this the movable grip (3.3) moves down automatically.. NOTE! Do not hold the movable grip (3.3) during assembly Additional notes and instructions for use CAUTION! Exerting excessive force can cause breakage or damage to the product. Due to the small dimensions required the products have only limited strength. Additionally in the case of bipolar instruments, damage to the insulation caused by overloading can lead to electrical safety being compromised. If the coagulation effect suddenly ceases, do not increase the HF power indiscriminately, but carry out a function test (see chapter 8.2.1) as appropriate. We recommend keeping a replacement instrument available for therapeutic applications. Use the products only to grasp and remove small soft tissue parts or organs. Take particular care when handling shock sensitive ceramic components such as the scissors. Check the products for damage and completeness immediately after use. Make sure that no missing instrument parts remain in the patient. (metal and ceramic components are X- ray positive).. IMPORTANT! Insert and retract the modular bipolar instruments only with closed jaw insert/scissors (1). 5 GA--B 241

9 7.4.1 HF application Follow the Notes and instructions on HF applications, Order no.: GA--S 002 as well as the HF device manufacturer s instructions. CAUTION! Use the modular bipolar instruments in conjunction with high frequency current only up to a maximum recurring peak voltage of 300 V. The use of bipolar cutting current/cut mode is not permissible! The products are only suitable for short term coagulation of tissue and small vessels (tubular structures such as blood vessels). WARNING! Danger of injury. There is a danger of injury if the bipolar instruments are not visible through the scope! Unintentional tissue damage as well as damage to the distal end of the endoscope and other instrument parts are possible. Use the bipolar instruments only within the scope of their specifications (dielectric strength, mode etc.). Activate the bipolar instruments only if the part conducting live HF current is fully visible through the scope and is in contact with the target area. WARNING! Danger of injury! During coagulation the hinge area A must not touch any tissue to avoid inadvertent coagulation. A CAUTION! Danger of short circuit in grasping forceps! If the jaws touch each other during coagulation. The jaws must not touch each other during coagulation.. NOTE! Encrusted and soiled jaws/cutting edges impair the HF power during use. Z Wipe encrusted and soiled coagulation surfaces with 3% hydrogen peroxide solution (H 2 O 2 solution)and a pad. Z Then rinse with sterile water. GA--B 241 6

10 8 Checks CAUTION! Be careful if products are damaged or incomplete! Possible injury of patient, user or third persons. Run through the checks before and after each use. Do not use products which are damaged or incomplete or have loose parts. Return damaged products together with loose parts for repair. Do not attempt to do any repairs yourself. 8.1 Visual check Z Check the instruments, in particular the distal areas, and the accessories for -- damage -- sharp edges -- loose or missing parts -- rough surfaces. Z Check ceramic components (scissors) for cracks. ' Replace if the ceramic cutting edge is cracked. Z Any lettering, labeling or identification necessary for the safe intended use must be legible. ' Missing or illegible lettering, labeling or identification which may lead to wrong handling and reprocessing must be restored. 8.2 Function check Function test Z Check the jaw inserts for proper opening and closing. Z Check the jaw insert for easy rotation by means of the star wheel (3.2). Z Moisten a sterile swab with sterile saline. Z Grasp the swab with the bipolar instrument. Z Activate the HF surgical device. ' The liquid in the swab is heated and evaporates. Z If no heating or insufficient heating is observed: ' Check the connection between the cable and the instrument and the cable and the device. ' Replace the cable and / or instrument if necessary. 7 GA--B 241

11 9 Reprocessing and maintenance 9.1 Manual reprocessing 9.2 Machine reprocessing 9.3 Checks 9.4 Sterilization Steam sterilization Gas sterilization Z Wet preparation at the point of use Z Disassembly before cleaning: see section 7.2 Z Manual cleaning with approved enzymatic cleaner. Z Brush and flush all channels with enzymatic cleaner and deionized water. Z Dry preparation at the point of use Z Disassembly before cleaning: see section 7.2 Z Machine cleaning with approved enzymatic cleaner. Z Follow machines instructions for use. Z Perform a visual check (see section 8.1) Modular Bipolar Instruments are to be assembled before sterilization: see section 7.3. All jaws must be in the open position for sterilization. Z Steam sterilization at 132 C (270 F) using a Pre--Vac cycle at an exposure time of 4 minutes with a 20 minute dry time. Z Gas sterilization using ethylene oxide (EtO). GA--B 241 8

12 10 Technical data and order data Product no. Designation ERAGON, bipolar handle, rotatable without lock Sheath, WL = 330 mm Sheath, WL = 240 mm ERAGON jaw section, Metzenbaum scissors ERAGON Metzenbaum scissors, assembly, diameter 5.5 mm, WL = 330 mm, consisting of: Jaw section Sheath ERAGON jaw section, Metzenbaum scissors ERAGON Metzenbaum scissors, assembly, diameter 5.5 mm, WL = 240 mm, consisting of: Jaw section Sheath Jaw section of ERAGON grasping and preparation forceps ERAGON grasping and preparation forceps, assembly, diameter 5.5 mm, WL = 330 mm, consisting of: Jaw section Sheath Jaw section of ERAGON grasping and preparation forceps ERAGON grasping and preparation forceps, assembly, diameter 5.5 mm, WL = 240 mm, consisting of: Jaw section Sheath Jaw section of ERAGON grasping and dissection forceps Maryland dissector" ERAGON grasping and dissection forceps Maryland dissector", assembly, diameter 5.5 mm, WL = 330 mm, consisting of: Jaw section Sheath Jaw section of ERAGON grasping and dissection forceps Maryland dissector" ERAGON grasping and dissection forceps Maryland dissector", assembly, diameter 5.5 mm, WL = 240 mm, consisting of: Jaw section Sheath GA--B 241

13 Jaw section of ERAGON grasping and dissection forceps Dolphin" ERAGON grasping and dissection forceps Dolphin", diameter 5.5 mm, WL = 330 mm, consisting of: Jaw section Sheath Jaw section of ERAGON grasping and dissection forceps Dolphin" ERAGON grasping and dissection forceps Dolphin", diameter 5.5 mm, WL = 240 mm, consisting of: Jaw section Sheath Jaw section ERAGON grasping forceps, with window. ERAGON grasping forceps, with window, assembly, diameter 5.5 mm, WL = 330 mm, consisting of: Jaw section Sheath Jaw section of ERAGON "Hasskamp" preparation forceps ERAGON Hasskamp" preparation forceps, complete, diameter 5.5 mm, WL = 330 mm, consisting of Jaw section Sheath The products can be combined as required provided the relevant technical data and intended uses are observed. For the general overview please refer to the latest catalog sheets and brochures, or contact Richard Wolf or your Richard Wolf representative. 11 Accessories Illustration Product no. Device connection Technical data HF bipolar connection cable, 3 m Wolf/Martin HF bipolar connection cable, 5 m HF bipolar connection cable, 3 m Erbe HF bipolar connection cable, 5 m HF bipolar connection cable, 3 m 2 pin Valleylab HF bipolar connection cable, 5 m 12 Operating, storage, transport and shipping conditions +10 C to+40 C, 30% to 75% rel. humidity, Operating conditions atmospheric pressure 700 hpa to 1060 hpa Storage, transport and shipping conditions -20 C to+60 C, 10% to 90% rel. humidity, atmospheric pressure 700 hpa to 1060 hpa. NOTE! To prevent damage during transport or shipment of the products we recommend using the original packaging material Disposal of product, packaging material and accessories For the disposal observe the regulations and laws valid in your country. ' For further information please contact the manufacturer. GA--B

14 13 Warranty and Customer Service Richard Wolf guarantees our instruments to be free from any defects in materials and workmanship under normal use and service for one year. Richard Wolf general terms and conditions may be found on the back of our invoice. Parts delivered separately by Richard Wolf are subject to all of the same general terms and conditions for our products, including the limitations of warranty and liability. All products should be returned to Richard Wolf for any necessary or desired repair or part replacement. No product repair or part replacement should be done other than by Richard Wolf unless the care and instruction manual or other written information indicates that repair or part replacement is authorized. If authorized, parts must be replaced only by parts supplied or specified by Richard Wolf,and productrepair and part replacement must be done in strict conformance with Richard Wolf specifications and instructions for repair and part replacement, including post replacement testing and recalibration. Failure to follow this requirement in any way can be dangerous to you, your personnel and your patients and voids the warranty for the product repaired or the product in which the part was replaced and if the part was supplied by Richard Wolf, for that part. Delivery by Richard Wolf of technical documents such as circuit or other design diagrams does not constitute authorization for product repair or part replacement. Richard Wolf instruments and other products should never be modified or altered under any circumstances. Contact Richard Wolf if you have any question (1) whether replacement of a part or a repair is authorized by Richard Wolf, or (2) whether you have complete instructions and specifications for part replacement or repair. These instructions do not attempt to cover all details or variations in equipment, nor to provide for every possible contingency to be met in connection with installation, operation, or maintenance. Should further information be required or should problems arise which are not covered sufficiently for the purchaser s purpose, the matter should be referred to Richard Wolf Medical Instruments Corporation. Our national sales and service offices, as well as our manufacturing facility, are located in Illinois. Trained manufacturer s representatives are located throughout the U.S. to serve you. For any questions regarding these instruments, or to place an order, contact Richard Wolf customer service department at or WOLF (9653). INSTRUMENT ORDERING POLICY Richard Wolf reserves the right to make substitutions, if necessary, without prior notice. REPAIR POLICY Defective merchandise will be repaired or replaced at no charge to the customer, provided the customer delivers such defective merchandise prepaid. Any repairs, maintenance or servicing of Richard Wolf merchandise by anyone other than a factory authorized representative will render our warranty null and void. REPAIR SHIPMENTS When returning your instrument for repair, we suggest that you prevent shipping damage to the instrument by reusing the box that it was originally shipped in. Richard Wolf also recommends that the instrument be insured for an amount to cover the cost of replacement. IMPORTANT For general safety and health reasons, Richard Wolf requires that you clean and sterilize all instruments before returning them for repair. If instruments are received in an unsanitary condition, Richard Wolf will clean and sterilize each instrument and add a $ cleaning charge for each instrument requiring cleaning. 11 GA--B 241

Instruction Manual. Intrauterine Probe Set Uterus Catheter by Sellheim GA-E 074 / USA/ Index: V1.

Instruction Manual. Intrauterine Probe Set Uterus Catheter by Sellheim GA-E 074 / USA/ Index: V1. Instruction Manual Intrauterine Probe Set 8378001 Uterus Catheter by Sellheim 8368.00 GA-E 074 / USA/ Index: 2012-01 V1.0 / ECO 2011-0351 Important general instructions for use Ensure that this product

More information

Instruction Manual. Operating Laparoscopes xxx/8914.xxx/8915.xxx/

Instruction Manual. Operating Laparoscopes xxx/8914.xxx/8915.xxx/ Instruction Manual Operating Laparoscopes 8912.xxx/8914.xxx/8915.xxx/8931.31 GA-E 191 / USA / 2012-07 V2.0 / ECO 2012-0374 (RW: 11-06-5.0 / PDI 06-1397) Important general notes and instructions for use

More information

Instructions for Use Reprocessed Ethicon ENDOPATH XCEL TM Universal Trocar Stability Sleeve. Reprocessed Device for Single Use.

Instructions for Use Reprocessed Ethicon ENDOPATH XCEL TM Universal Trocar Stability Sleeve. Reprocessed Device for Single Use. Reprocessed by Instructions for Use Reprocessed Ethicon ENDOPATH XCEL TM Universal Trocar Stability Sleeve Reprocessed Device for Single Use Caution: Federal (U.S.A.) law restricts this device to sale

More information

RIGID ENDOSCOPE OPERATING INSTRUCTIONS

RIGID ENDOSCOPE OPERATING INSTRUCTIONS RIGID ENDOSCOPE OPERATING INSTRUCTIONS Note: Read this manual first. It contains important instructions and warnings concerning the proper assembly, use, and service of your new rigid endoscopy product(s).

More information

Instructions for Use Reprocessed Ethicon ENDOPATH XCEL TM Dilating Tip Trocar with OPTIVIEW Technology

Instructions for Use Reprocessed Ethicon ENDOPATH XCEL TM Dilating Tip Trocar with OPTIVIEW Technology Reprocessed by Instructions for Use Reprocessed Ethicon ENDOPATH XCEL TM Dilating Tip Trocar with OPTIVIEW Technology Reprocessed Device for Single Use Caution: Federal (U.S.A.) law restricts this device

More information

Entellus Medical Sinuscope INSTRUCTIONS FOR USE

Entellus Medical Sinuscope INSTRUCTIONS FOR USE Entellus Medical Sinuscope INSTRUCTIONS FOR USE Table of contents Entellus Medical Sinuscope Instructions for Use 1. About this document... 1 1.1 Purpose... 1 1.2 Symbols used... 1 2. Intended use... 2

More information

CORE Innovation Worldwide DESIGN CONSULTING SERVICE PRODUCT/SYSTEM INSTALLATION

CORE Innovation Worldwide DESIGN CONSULTING SERVICE PRODUCT/SYSTEM INSTALLATION CORE Innovation Worldwide DESIGN CONSULTING SERVICE PRODUCT/SYSTEM INSTALLATION DESIGN PRODUCT / SYSTEM CONSULTING SERVICE INSTALLATION 2 Contents OR -Endoscopy From the start of the 20th century to the

More information

ASAHI Masters PARKWAY SOFT. ASAHI Masters PARKWAY HF

ASAHI Masters PARKWAY SOFT. ASAHI Masters PARKWAY HF ASAHI Masters PARKWAY SOFT Microcatheter ASAHI Masters PARKWAY HF Microcatheter SYMBOLS 1 English 2 AMK-DT207 Ver.1.40 / 11TS028 SYMBOLS Legal manufacturer Consult instructions for use Do not use if package

More information

Instructions for Use Reprocessed LigaSure Blunt Tip Laparoscopic Sealer/Dividers (Model LF1637) Reprocessed Device for Single Use.

Instructions for Use Reprocessed LigaSure Blunt Tip Laparoscopic Sealer/Dividers (Model LF1637) Reprocessed Device for Single Use. Page 1 of 8 Reprocessed by Instructions for Use Reprocessed LigaSure Blunt Tip Laparoscopic Sealer/Dividers (Model LF1637) Reprocessed Device for Single Use Caution: Federal (U.S.A.) law restricts this

More information

Symphion 6.3 Endoscope

Symphion 6.3 Endoscope For single use only. Do not reuse. Federal (US) law restricts this device to sale by or on the order of a physician. PREFACE Symphion 6.3 Endoscope Instructions for Use This manual provides the information

More information

NeuroPort Array PN 4382, 4383, 6248, and 6249 Instructions for Use

NeuroPort Array PN 4382, 4383, 6248, and 6249 Instructions for Use 630 Komas Drive Suite 200 Salt Lake City UT 84108 USA P +1 801.582.5533 F +1 801.582.1509 www.blackrockmicro.com NeuroPort Array PN 4382, 4383, 6248, and 6249 Instructions for Use Revision 2.00 / LB-0612

More information

Control Your Energy. with the Right Settings. A guide to monopolar energy with the ERBE VIO dv. PN US Rev A 04/17

Control Your Energy. with the Right Settings. A guide to monopolar energy with the ERBE VIO dv. PN US Rev A 04/17 Control Your Energy with the Right Settings A guide to monopolar energy with the ERBE VIO dv A Versatile Energy Source at your fingertips The VIO dv generator offers a variety of monopolar settings so

More information

6 Closet light. Model Due to continuing improvements, actual product may differ slightly from the product described herein.

6 Closet light. Model Due to continuing improvements, actual product may differ slightly from the product described herein. 6 Closet light Model 95792 Assembly And Operation Instructions Due to continuing improvements, actual product may differ slightly from the product described herein. 3491 Mission Oaks Blvd., Camarillo,

More information

Competency Guide: Care and Handling of Rigid Endoscopes

Competency Guide: Care and Handling of Rigid Endoscopes Competency Guide: Care and Handling of Rigid Endoscopes This document details the basic steps to correctly process and maintain a rigid endoscope. This document can be used to perform a competency review

More information

KEEP FOR FUTURE REFERENCE FORKLIFT ADAPTER FOR MRTA6 AND MRTA8 LIFTERS READ ALL INSTRUCTIONS AND WARNINGS BEFORE USING THIS PRODUCT

KEEP FOR FUTURE REFERENCE FORKLIFT ADAPTER FOR MRTA6 AND MRTA8 LIFTERS READ ALL INSTRUCTIONS AND WARNINGS BEFORE USING THIS PRODUCT KEEP FOR FUTURE REFERENCE P.O. Box 368 908 West Main Laurel, MT USA 59044 phone 800-548-7341 phone 406-628-8231 fax 406-628-8354 INSTRUCTIONS International Version STOCK NUMBER: 95457 FORKLIFT ADAPTER

More information

Probe Manual EL Rev D. Handling, cleaning and sterilization of the Medistim EL Imaging Probe

Probe Manual EL Rev D. Handling, cleaning and sterilization of the Medistim EL Imaging Probe Handling, cleaning and sterilization of the Medistim EL 100015 Imaging Probe The purpose of this procedure is to guarantee that the probe performance is in accordance with the specifications during the

More information

KEEP FOR FUTURE REFERENCE MRT FORKLIFT ADAPTER READ ALL INSTRUCTIONS AND WARNINGS BEFORE USING THIS PRODUCT

KEEP FOR FUTURE REFERENCE MRT FORKLIFT ADAPTER READ ALL INSTRUCTIONS AND WARNINGS BEFORE USING THIS PRODUCT KEEP FOR FUTURE REFERENCE P.O. Box 368 908 West Main Laurel, MT USA 59044 phone 800-548-7341 phone 406-628-8231 fax 406-628-8354 INSTRUCTIONS International Version STOCK NUMBER: 95722 MRT FORKLIFT ADAPTER

More information

ASAHI Masters PARKWAY HF KIT

ASAHI Masters PARKWAY HF KIT ASAHI Masters PARKWAY HF KIT Guide Wire Pre-loaded Microcatheter SYMBOLS 1 English 2 AMK-DT202 Ver.1.30 / 11TS023 SYMBOLS Legal manufacturer GC Minimum guiding catheter I.D. Do not use if package is damaged

More information

Operator s Manual for Morse Heavy-Duty Kontrol-Karrier with 3-Piece Drum Holder

Operator s Manual for Morse Heavy-Duty Kontrol-Karrier with 3-Piece Drum Holder CONTENTS Page Receiving Procedures.................... 1 Warranty............................. 1 Safety Information..................... 1-2 Machine Description................... 3 Operating Instructions....................

More information

8 DIAL CALIPER. Model Diagrams within this manual may not be drawn proportionally.

8 DIAL CALIPER. Model Diagrams within this manual may not be drawn proportionally. 8 DIAL CALIPER Model 97389 Set up And Operating Instructions Diagrams within this manual may not be drawn proportionally. Due to continuing improvements, actual product may differ slightly from the product

More information

Instructions for Use Reprocessed Electrophysiology Catheter Cables

Instructions for Use Reprocessed Electrophysiology Catheter Cables Reprocessed by Instructions for Use Reprocessed Electrophysiology Catheter Cables Reprocessed Device for Single Use STERILE Explanation of Symbols Federal Law in the USA restricts this device to sale by

More information

GENERAL GUARANTEE CONDITIONS OF POLTRONIC S.A. (SUPPLIER) FOR A COMMERCIAL PARTNER (PARTNER) FOR NEXTEC BRAND PRODUCTS

GENERAL GUARANTEE CONDITIONS OF POLTRONIC S.A. (SUPPLIER) FOR A COMMERCIAL PARTNER (PARTNER) FOR NEXTEC BRAND PRODUCTS GENERAL GUARANTEE CONDITIONS OF POLTRONIC S.A. (SUPPLIER) FOR A COMMERCIAL PARTNER (PARTNER) FOR NEXTEC BRAND PRODUCTS 1. The SUPPLIER gives a 24-month guarantee for all of the SUPPLIER s products, counting

More information

1-300 SERIES Endover Rotator Operator s Manual for Morse End-Over-End Drum Rotators Series

1-300 SERIES Endover Rotator Operator s Manual for Morse End-Over-End Drum Rotators Series Contents Page Receiving Procedures.................... 1 Warranty............................. 1 Safety Information..................... 1-2 Machine Description................... 3 Operating Instructions....................

More information

Instructions for Use Reusable Electrophysiology Catheter Cables

Instructions for Use Reusable Electrophysiology Catheter Cables Processed by Instructions for Use Reusable Electrophysiology Catheter Cables STERILE Explanation of Symbols Federal Law in the USA restricts this device to sale by or on the order of a physician Sterilized

More information

Instructions for Use Reprocessed LigaSure TM Blunt Tip Laparoscopic Sealer/Dividers (Model LF1637) Manufactured JANUARY 02, 2018 or Later

Instructions for Use Reprocessed LigaSure TM Blunt Tip Laparoscopic Sealer/Dividers (Model LF1637) Manufactured JANUARY 02, 2018 or Later Reprocessed by Instructions for Use Reprocessed LigaSure TM Blunt Tip Laparoscopic Sealer/Dividers (Model LF1637) Manufactured JANUARY 02, 2018 or Later Reprocessed Device for Single Use Caution: Federal

More information

Instructions for Use Reusable Ablation Catheter Cables

Instructions for Use Reusable Ablation Catheter Cables Processed by Instructions for Use Reusable Ablation Catheter Cables STERILE Explanation of Symbols Federal Law in the USA restricts this device to sale by or on the order of a physician Sterilized by Ethylene

More information

Instructions for Use Reprocessed Balloon Inflation Devices

Instructions for Use Reprocessed Balloon Inflation Devices Reprocessed by Instructions for Use Reprocessed Balloon Inflation Devices Reprocessed Device for Single Use Caution: Federal (U.S.A.) law restricts this device to sale by or on the order of a physician.

More information

RePlay Hemostasis Clip

RePlay Hemostasis Clip RePlay Hemostasis Clip Instructions for use. Read carefully prior to use. Caution: Federal (U.S.A.) Law restricts this device to sale by or on the order of a physician. Diversatek Healthcare Corporate

More information

78-RK VARIABLE SPEED HEAVY DUTY VIBRATOR OPERATING MANUAL

78-RK VARIABLE SPEED HEAVY DUTY VIBRATOR OPERATING MANUAL 78-RK VARIABLE SPEED HEAVY DUTY VIBRATOR OPERATING MANUAL IMPORTANT INSTRUCTIONS WARNING! FOR YOUR SAFETY PLEASE READ INSTRUCTIONS BEFORE OPERATING TOOL & WEAR EYE PROTECTION FD.08/2009 THANK YOU FOR PURCHASING

More information

MODEL 185R Kontrol-Karrier Operator s Manual for Morse Kontrol-Karrier

MODEL 185R Kontrol-Karrier Operator s Manual for Morse Kontrol-Karrier Contents Page Receiving Procedures.................... 1 Warranty............................. 1 Safety Information..................... 1-2 Machine Description................... 3 Operating Instructions....................

More information

Cranio-caudal Fixation Unit CC-320-Plus

Cranio-caudal Fixation Unit CC-320-Plus Cranio-caudal Fixation Unit CC-320-Plus 118341 for immobilization combined with the NORAS Breast Biopsy 4-Ch Coil, the Siemens Breast Biopsy 4-Ch Coil, the Breast Biopsy 4-Ch Coil for GE or the NORAS Patient

More information

CM-95 ORDERCODE D1355 Highlite International B.V.

CM-95 ORDERCODE D1355 Highlite International B.V. CM-95 ORDERCODE D1355 Highlite International B.V. Vestastraat 2 6468 EX Kerkrade The Netherlands Congratulations! You have bought a great, innovative product from DAP Audio. The DAP Audio CM-95 brings

More information

July 13, 2017 URGENT: MEDICAL DEVICE RECALL PRODUCT: ReNew Reusable Tips (Scissors, Graspers and Dissectors)

July 13, 2017 URGENT: MEDICAL DEVICE RECALL PRODUCT: ReNew Reusable Tips (Scissors, Graspers and Dissectors) 50 Dunham Road, Suite 1500 July 13, 2017 URGENT: MEDICAL DEVICE RECALL PRODUCT: ReNew Reusable Tips (Scissors, Graspers and Dissectors) Attention: Customer Name Device Name Street Address City, State,

More information

PMA 31-G. English. Printed: Doc-Nr: PUB / / 000 / 00

PMA 31-G. English. Printed: Doc-Nr: PUB / / 000 / 00 PMA 31-G English 1 Information about the documentation 1.1 About this documentation Read this documentation before initial operation or use. This is a prerequisite for safe, trouble-free handling and

More information

DRYWALL DOLLY 900 LB. CAPACITY

DRYWALL DOLLY 900 LB. CAPACITY DRYWALL DOLLY 900 LB. CAPACITY 94883 ASSEMBLY & OPERATING INSTRUCTIONS Due to continuing improvement, actual product may differ slightly from the product described herein. 3491 Mission Oaks Blvd., Camarillo,

More information

PalletPal Low Profile Disc Turntable

PalletPal Low Profile Disc Turntable OWNER S MANUAL PalletPal Low Profile Disc Turntable Model # Serial # Date placed in Service SOUTHWORTH PRODUCTS CORP P.O. Box 1380 Portland, ME 04104-1380 Telephone: (207) 878-0700 or (800) 743-1000 Fax:

More information

UNIQUE HYBRID TECHNOLOGY. DOUBLE YOUR ENERGY THUNDERBEAT Type S Next Generation of Safety and Speed

UNIQUE HYBRID TECHNOLOGY. DOUBLE YOUR ENERGY THUNDERBEAT Type S Next Generation of Safety and Speed UNIQUE HYBRID TECHNOLOGY DOUBLE YOUR ENERGY THUNDERBEAT Type S Next Generation of Safety and Speed 15149 1 UNIQUE HYBRID TECHNOLOGY ADVANCED HEMOSTASIS Philosophy Olympus aims to provide innovative energy

More information

Instructions for Use Reprocessed LigaSure Impact without Nano-coating Large Jaw, Open Sealer/Dividers

Instructions for Use Reprocessed LigaSure Impact without Nano-coating Large Jaw, Open Sealer/Dividers Page 1 of 8 Reprocessed by Instructions for Use Reprocessed LigaSure Impact without Nano-coating Large Jaw, Open Sealer/Dividers Reprocessed Device for Single Use Caution: Federal (U.S.A.) law restricts

More information

Guide for Cleaning, Sterilization and Storage of Reusable Introducers

Guide for Cleaning, Sterilization and Storage of Reusable Introducers Guide for Cleaning, Sterilization and Storage of Reusable Introducers M Manufactured by: Caldera Medical, Inc. 5171 Clareton Drive Agoura Hills, CA 91301 U.S. Toll Free: 866-4-CALDERA Telephone: 818-879-6555

More information

A Sterile Processing User s Guide

A Sterile Processing User s Guide Product Manual A Sterile Processing User s Guide Safety, Assurance, Innovation, & Quality enabling perioperative professionals to achieve higher standards of performance and enhanced patient safety Phone:866-ONE-TRAY

More information

ENGLISH Compact Half Coupler 50 mm, SWL: 50 kg

ENGLISH Compact Half Coupler 50 mm, SWL: 50 kg MANUAL ENGLISH Compact Half Coupler 50 mm, SWL: 50 kg V3 Highlite International B.V. Vestastraat 2 6468 EX Kerkrade the Netherlands Table of contents Warning... 2 Safety Instructions... 2 Operating Determinations...

More information

MANUAL. ENGLISH Antenna Booster Ordercode: D1425. Highlite International B.V. Vestastraat EX Kerkrade the Netherlands

MANUAL. ENGLISH Antenna Booster Ordercode: D1425. Highlite International B.V. Vestastraat EX Kerkrade the Netherlands MANUAL ENGLISH Antenna Booster V1 Highlite International B.V. Vestastraat 2 6468 EX Kerkrade the Netherlands Table of contents Warning... 2 Unpacking Instructions... 2 Safety Instructions... 2 Operating

More information

RapidPort EZ Port Applier DIRECTIONS FOR USE (DFU)

RapidPort EZ Port Applier DIRECTIONS FOR USE (DFU) RapidPort EZ Port Applier DIRECTIONS FOR USE (DFU) RapidPort EZ Port Applier Ref. No. C-20390 RapidPort EZ Port Applier INTRODUCTION The RapidPort EZ Port Applier is an optional accessory for the LAP-BAND

More information

Urology and Women s Health

Urology and Women s Health Urology and Women s Health GYNECOLOGIC SURGERY PRODUCT CATALOG Hysteroscopic Tissue Removal Endometrial Ablation Introduction The all-in-one Symphion System is designed to transform the way you remove

More information

OPERATING MANUAL. Conveyor belts. CB-/B-Conveyor

OPERATING MANUAL. Conveyor belts. CB-/B-Conveyor OPERATING MANUAL Conveyor belts CB-/B-Conveyor Edition 1-5/2018 Contents Please note! -3 General Information -3 Validity of this manual and required compliance -3 How information is represented -4 For

More information

Operator s Manual for Morse Kontrol-Karrier with Spark Resistant Parts and Air Power Tilt

Operator s Manual for Morse Kontrol-Karrier with Spark Resistant Parts and Air Power Tilt Contents Page Receiving Procedures.................... 1 Warranty............................. 1 Safety Information..................... 1-2 Machine Description................... 2 Assembly Instructions..................

More information

Instructions for Use: Duodenoscope Sampling Kit

Instructions for Use: Duodenoscope Sampling Kit Instructions for Use: Duodenoscope Sampling Kit Brand Name of Product Generic Name of Product Product Code Number(s) Intended Use Duodenoscope Sampling Kit Endoscope sampling and culturing kit CK-250,

More information

ENGLISH WiFi Module for GIG-202 Tab

ENGLISH WiFi Module for GIG-202 Tab MANUAL ENGLISH WiFi Module for GIG-202 Tab V1 Highlite International B.V. Vestastraat 2 6468 EX Kerkrade the Netherlands Table of contents Warning... 2 Safety Instructions... 2 Operating Determinations...

More information

Rocker Tool Post ASSEMBLY AND OPERATING INSTRUCTIONS

Rocker Tool Post ASSEMBLY AND OPERATING INSTRUCTIONS Rocker Tool Post 94540 ASSEMBLY AND OPERATING INSTRUCTIONS Machine tool shown in photo not included. Due to continuing improvements, actual product may differ slightly from the product described herein.

More information

Universal Tactical Light

Universal Tactical Light M2 Universal Tactical Light Operator s Manual MADE IN THE USA WARNINGS 1. Before handling the firearm or the M2 UTL, read and understand the entire contents of the firearm and M2 UTL Operator s Manuals,

More information

MODEL GEK E3 Guard Enclosure Kit. Operator s Manual for Morse Guard Enclosure Kit Model GEK E3 for Can Tumbler Model E3

MODEL GEK E3 Guard Enclosure Kit. Operator s Manual for Morse Guard Enclosure Kit Model GEK E3 for Can Tumbler Model E3 Contents Page Receiving Procedures.................... 1 Warranty............................. 1 Safety Information..................... 1-2 Assembly Instructions................... 3-7 Parts List and

More information

MODEL GEK E1 Guard Enclosure Kit. Operator s Manual for Morse Guard Enclosure Kit Model GEK E1 for Can Tumbler Model E1

MODEL GEK E1 Guard Enclosure Kit. Operator s Manual for Morse Guard Enclosure Kit Model GEK E1 for Can Tumbler Model E1 Contents Page Receiving Procedures.................... 1 Warranty............................. 1 Safety Information..................... 1-2 Assembly Instructions................... 3-7 Parts List and

More information

INSTALLATION AND OPERATION MANUAL

INSTALLATION AND OPERATION MANUAL Pneumatic Wheel Lift Model: RWL-50T REV. A PLEASE READ THE ENTIRE CONTENTS OF THIS MANUAL PRIOR TO INSTALLATION AND OPERATION. BY PROCEEDING YOU AGREE THAT YOU FULLY UNDERSTAND AND COMPREHEND THE FULL

More information

QA-ES Automation Bundle

QA-ES Automation Bundle QA-ES Automation Bundle ESU Automation Kit Users Manual PN 3672225 March 211, Rev. 1 211 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice. All

More information

Installation- and Operating Instructions Interroll End stop RM 8811

Installation- and Operating Instructions Interroll End stop RM 8811 Installation- and Operating Instructions Interroll End stop Manufacturer's address Interroll Automation GmbH Dietmar-Hopp-Straße 3 74889 Sinsheim (Germany) Ph.: +49 7261 938 0 Fax: +49 7261 938 124 www.interroll.com

More information

FLARING MACHINE MAINTENANCE & INSTRUCTIONS MANUAL. Allswage UK. Roebuck Street, West Bromwich, B70 6RB

FLARING MACHINE MAINTENANCE & INSTRUCTIONS MANUAL. Allswage UK. Roebuck Street, West Bromwich, B70 6RB FLARING MACHINE MAINTENANCE & INSTRUCTIONS MANUAL A. WARRANTY AND RESPONSIBILITY Warranty: It's the supplier's responsibility to guarantee the conformity of the product, assuring that it's manufactured

More information

7 PIECE TUBE FLARING KIT

7 PIECE TUBE FLARING KIT 7 PIECE TUBE FLARING KIT Model 5969 Set up And Operating Instructions Diagrams within this manual may not be drawn proportionally. Due to continuing improvements, actual product may differ slightly from

More information

4 rotary milling table

4 rotary milling table 4 rotary milling table Model 97208 Set up And Operating Instructions Diagrams within this manual may not be drawn proportionally. Due to continuing improvements, actual product may differ slightly from

More information

Cranio-caudal Fixation Unit CC-GE MR10063-CC-GE

Cranio-caudal Fixation Unit CC-GE MR10063-CC-GE Cranio-caudal Fixation Unit CC-GE MR10063-CC-GE Cranio caudal Fixation Unit CC-GE for Immobilization with the GE 8 channel breast coil Operator s Manual Revision 04 Issued Date: August 2016 Technichal

More information

MANUAL ENGLISH. Ordercode: D3423/D3423W. Highlite International B.V. Vestastraat EX Kerkrade the Netherlands

MANUAL ENGLISH. Ordercode: D3423/D3423W. Highlite International B.V. Vestastraat EX Kerkrade the Netherlands MANUAL ENGLISH Xi-6 V1 Highlite International B.V. Vestastraat 2 6468 EX Kerkrade the Netherlands Table of contents Warning... 2 Safety Instructions... 2 Operating Determinations... 3 Return Procedure...

More information

Sby SR Instruments, Inc.

Sby SR Instruments, Inc. Part No.: MAN947IFS_170629 Page 1 of 12 Sby SR Instruments, Inc. SRV947IFS REMOTE DISPLAY PLATFORM SCALE Operating and Service Manual Part No.: MAN947IFS_170629 Page 2 of 12 TABLE OF CONTENTS TABLE OF

More information

Lotus. Reprocessing instructions for reusable instrumentation. French/Swiss protocol

Lotus. Reprocessing instructions for reusable instrumentation. French/Swiss protocol Lotus Reprocessing instructions for reusable instrumentation French/Swiss protocol Reprocessing instructions These instructions relate to the Lotus Transducers & Waveguides only. Following use remove the

More information

Dandy Leveler UDLV-150 & UDLV-500 (New Style)

Dandy Leveler UDLV-150 & UDLV-500 (New Style) Owner s Manual Dandy Leveler UDLV-150 & UDLV-500 (New Style) Model # Serial # Southworth Products Corp P.O. Box 1380/Portland, Maine 04104-1380 Phone 800-743-1000 FAX 207-797-4734 www.southworthproducts.com

More information

POWER PINNER BENCH HAND WELDER 7250 OPERATOR S MANUAL

POWER PINNER BENCH HAND WELDER 7250 OPERATOR S MANUAL POWER PINNER BENCH HAND WELDER 7250 OPERATOR S MANUAL Copyright: November 20, 2018 Revised: Serial No. Gripnail Corporation An Employee Owned Company 97 Dexter Road East Providence, Rhode Island 02914-2045

More information

Installation Guide Document No Transaction Windows Copyright 2017 Terra Universal Inc. All rights reserved. Revised October 2017

Installation Guide Document No Transaction Windows Copyright 2017 Terra Universal Inc. All rights reserved. Revised October 2017 Installation Guide Document No. 1788-01 Copyright 2017 Terra Universal Inc. All rights reserved. Revised October 2017 Terra Universal, Inc. TerraUniversal.com 800 S. Raymond Ave. Fullerton, CA 92831 TEL:

More information

ENDOSCOPY PRACTICE Manual cleaning and Disinfection of flexible endoscopes

ENDOSCOPY PRACTICE Manual cleaning and Disinfection of flexible endoscopes 1.0 Purpose Every patient undergoing endoscopy should be examined with clean, disinfected equipment. All endoscopic equipment will be cleaned and disinfected / sterilized according to current recommended

More information

C60 Standby UPS. 800VA Models. User & Installation Manual Xtreme Power Conversion Corporation. All rights reserved.

C60 Standby UPS. 800VA Models. User & Installation Manual Xtreme Power Conversion Corporation. All rights reserved. For more information, visit www.247technology.com/on-line-ups/c60-standby-ups C60 Standby UPS 800VA Models User & Installation Manual 2015. All rights reserved. (Rev 6/23/15) Table of Contents Package

More information

REF DOC # STANDARD IMAGING, INC Murphy Drive Middleton, WI TEL TEL FAX

REF DOC # STANDARD IMAGING, INC Murphy Drive Middleton, WI TEL TEL FAX WWW..COM REF 90086 STANDARD IMAGING, INC. 7601 Murphy Drive Middleton, WI 53562-2532 TEL 800.261.4446 TEL 608.831.0025 FAX 608.831.2202 Jul / 2006 2006 Standard Imaging, Inc. DOC #80173-04 General Precautions

More information

OWNER S MANUAL SERIES MDS-96-BK, SM, DR

OWNER S MANUAL SERIES MDS-96-BK, SM, DR VESTIL MANUFACTURING CORP. 2999 N. Wayne St., Angola, IN 46703 Ph: 260-665-7586 Fax: 260-665-1339 E-mail: sales@vestil.com Website: www.vestil.com SERIES MDS-96-BK, SM, DR OWNER S MANUAL Introduction...

More information

Pneumatic PalletPal Level Loader

Pneumatic PalletPal Level Loader Owner s Manual TM Pneumatic PalletPal Level Loader For products shipped prior to January 2014 Model # Serial # Placed in Service Southworth Products Corp P.O. Box 1380 Portland, Maine 04104-1380 Phone

More information

Agilent 16034E Test Fixture

Agilent 16034E Test Fixture Agilent 16034E Test Fixture Operation and Service Manual Third Edition Agilent PN 16034-90041 December 1999 Printed in: Japan Notices The information contained in this document is subject to change without

More information

Reusable 5x Holmium Laser Fibers Instructions for Use Model: LF0201X200, LF0201X272, LF0201X365, LF0201X550, LF Holmium and Nd:YAG Fiber

Reusable 5x Holmium Laser Fibers Instructions for Use Model: LF0201X200, LF0201X272, LF0201X365, LF0201X550, LF Holmium and Nd:YAG Fiber Instructions for Use (IFU) Intended Use Reusable 5x Holmium Laser Fibers Instructions for Use Model: LF0201X200, LF0201X272, LF0201X365, LF0201X550, LF0201940 Holmium and Nd:YAG Fiber The laser probes

More information

OWNER S MANUAL SERIES MDS-96-BK, SM, DR

OWNER S MANUAL SERIES MDS-96-BK, SM, DR SERIES MDS-96-BK, SM, DR OWNER S MANUAL Introduction... 1 Bill of Materials... 2 Assembly Instructions. 3 Warranty....12 IMPORTANT NOTES, WARNINGS AND SAFETY INSTRUCTIONS Ensure that all employees understand

More information

64C Amplification Block. Instruction Manual

64C Amplification Block. Instruction Manual 64C Amplification Block Instruction Manual Caution: All rights reserved. Quidel Corporation reserves the right to modify this manual at any time without notice. Any part of the manual shall not be duplicated,

More information

J60 Lithium Ion UPS 350VA Model User & Installation Manual Xtreme Power Conversion Corporation. All rights reserved.

J60 Lithium Ion UPS 350VA Model User & Installation Manual Xtreme Power Conversion Corporation. All rights reserved. J60 Lithium Ion UPS 350VA Model User & Installation Manual www.xpcc.com 2018. All rights reserved. (Rev 5/24/18) Table of Contents Package Contents...3 Product Introduction...3 Product Overview...3 Installation

More information

Wenger Corporation 2011 Printed in USA 02/11 Part #185A462-01

Wenger Corporation 2011 Printed in USA 02/11 Part #185A462-01 User Instructions Diva Air Transporter contents Introduction.......................................2 General......................................2 Manufacturer.................................2 European

More information

Industrial flue gas probes. Instruction manual

Industrial flue gas probes. Instruction manual Industrial flue gas probes Instruction manual 2 1 Contents 1 Contents 1 Contents... 3 2 Safety and the environment... 4 2.1. About this document... 4 2.2. Ensure safety... 4 2.3. Protecting the environment...

More information

MODEL Operator s Manual for Morse MORSpeed TM Fork Mount Drum Handler Model Fork Mount Drum Handler

MODEL Operator s Manual for Morse MORSpeed TM Fork Mount Drum Handler Model Fork Mount Drum Handler Contents Page Receiving Procedures.................... 1 Warranty............................. 1 Safety Information..................... 1-2 Machine Description................... 3 Operating Instructions....................

More information

LOAD RAMP 1000 LB. CAPACITY

LOAD RAMP 1000 LB. CAPACITY LOAD RAMP 1000 LB. CAPACITY 55424 ASSEMBLY AND OPERATING INSTRUCTIONS Safely roll Dirt Bikes and more into your Pickup! 3491 Mission Oaks Blvd., Camarillo, CA 93011 Visit our Web site at http://www.harborfreight.com

More information

Lotus. Reprocessing instructions for reusable instrumentation

Lotus. Reprocessing instructions for reusable instrumentation Lotus Reprocessing instructions for reusable instrumentation Reprocessing instructions These instructions relate to the Lotus Transducers & Waveguides only. Following use remove the Single Use Handpiece

More information

Owner s Manual & Safety Instructions

Owner s Manual & Safety Instructions Owner s Manual & Safety Instructions Save This Manual Keep this manual for the safety warnings and precautions, assembly, operating, inspection, maintenance and cleaning procedures. Write the product s

More information

Operating and Maintenance Manual. for. HADEF locking device for monorail trolleys

Operating and Maintenance Manual. for. HADEF locking device for monorail trolleys 5.52.653.99.0.1 Edition 08/2000/gr GB Operating and Maintenance Manual for HADEF locking device for monorail trolleys Subject to changes. 1 HADEF Contents page 1 General Information 3 2 Product Description

More information

GF6580BH-1 PATRIOT LIBERTY QUARTER RAILS:

GF6580BH-1 PATRIOT LIBERTY QUARTER RAILS: GF6580BH-1 PATRIOT LIBERTY QUARTER RAILS: INSTALLATION & OPERATION INSTRUCTIONS FOR PATRIOT BED PLEASE SAVE THESE INSTRUCTIONS FOR FUTURE USE Info: The most current version of these instructions can be

More information

MECHANICAL LEVELERS. P.O. Box 338 Spring Hill, Tennessee /486/ /251/3382 Fax 931/486/0316

MECHANICAL LEVELERS. P.O. Box 338 Spring Hill, Tennessee /486/ /251/3382 Fax 931/486/0316 EDE & BLE Owner s Manual MECHANICAL LEVELERS P.O. Box 338 Spring Hill, Tennessee 37174 931/486/2296 800/251/3382 Fax 931/486/0316 www.pioneerleveler.com SM SECTION 1 RECEIVING: INSTALLATION INSTRUCTIONS

More information

New Advanced Environmental Air Treatment System

New Advanced Environmental Air Treatment System Package Unit New Advanced Environmental Air Treatment System Thank you for purchasing the Package Unit by RGF Advanced Oxidation Systems. We are confident this system will provide you with years of Fresh,

More information

USER S MANUAL. Flexible Nasopharyngo-Laryngoscope

USER S MANUAL. Flexible Nasopharyngo-Laryngoscope USER S MANUAL COGENTIX MEDICAL Flexible Nasopharyngo-Laryngoscope ENT-4000/4500 NOTE: Federal (USA) law restricts this device to sale by, or on the order of, a physician or other appropriately licensed

More information

Guide for Cleaning, Sterilization and Storage of Introducers for Caldera Medical Implants

Guide for Cleaning, Sterilization and Storage of Introducers for Caldera Medical Implants Guide for Cleaning, Sterilization and Storage of Introducers for Caldera Medical Implants M Manufactured by: Caldera Medical, Inc. 5171 Clareton Drive Agoura Hills, CA 91301 U.S. Toll Free: 866-4-CALDERA

More information

Operator s Manual for Morse Mobile-Karrier for 55-Gallon Plastic or Steel Drum

Operator s Manual for Morse Mobile-Karrier for 55-Gallon Plastic or Steel Drum Contents Page Receiving Procedures.................... 1 Warranty............................. 1 Safety Information..................... 1-2 Machine Description................... 3 Operating Instructions....................

More information

GYNECOLOGIC SURGERY PRODUCT CATALOG

GYNECOLOGIC SURGERY PRODUCT CATALOG Urology and Women s Hlth 2018 GYNECOLOGIC SURGERY PRODUCT CATALOG Hysteroscopic Tissue Removal Endometrial Ablation ORDERING NUMBERS Boston Scientific Corporation uses a fully computerized order entry

More information

BESTMOVE STANDING TRANSFER AID

BESTMOVE STANDING TRANSFER AID BESTMOVE STANDING TRANSFER AID L.L.C. MODELS BestMove 400 BestMove 450 PRODUCT FEATURES *BestMove 450 shown Ensure the product has been assembled according to the instructions in this manual. OPERATING

More information

HST -B Interlocking device (Translation of Original Manual)

HST -B Interlocking device (Translation of Original Manual) Installation and Operating Manual for Components HST -B Interlocking device (Translation of Original Manual) HST-B Ident.-No.: 10201 HST-B Ident.-No.: 10207 HST-B Ident.-No.: 10202 HST-B Ident.-No.: 10208

More information

Introduction TMD Warnings and Cautions Caution: Caution: Caution: Caution: Caution: Caution: Modification

Introduction TMD Warnings and Cautions Caution: Caution: Caution: Caution: Caution: Caution: Modification Introduction This guide provides instruction for operation, maintenance, and troubleshooting procedures for the Tumescent Measuring Device - the TMD. All users should be versed in its use and application

More information

PLEASE SAVE THESE INSTRUCTIONS FOR FUTURE USE

PLEASE SAVE THESE INSTRUCTIONS FOR FUTURE USE GF6590BH-1 PATRIOT LIBERTY HALF NO-GAP HEAD RAILS: INSTALLATION & OPERATION INSTRUCTIONS FOR PATRIOT BED PLEASE SAVE THESE INSTRUCTIONS FOR FUTURE USE Info: The most current version of these instructions

More information

and Shield with Vertical Needle Holder REF 90072

and Shield with Vertical Needle Holder REF 90072 WWW..COM and Shield with Vertical Needle Holder REF 90072 STANDARD IMAGING, INC. 7601 Murphy Drive Middleton, WI 53562-2532 TEL 800.261.4446 TEL 608.831.0025 FAX 608.831.2202 Nov / 2006 2006 Standard Imaging,

More information

Dual Port isolated USB Charger

Dual Port isolated USB Charger User & Installation Manual EN Dual Port isolated USB Charger 3A R-net cable P010-61 3A DX cable P010-62 Contact & Product mo-vis BVBA Biebuyckstraat 15 D 9850 Nevele - Belgium Website: www.mo-vis.com E-mail:

More information

DA VINCI Xi SINGLE-SITE TECHNOLOGY SOLUTIONS FOR SINGLE-INCISION SURGERY

DA VINCI Xi SINGLE-SITE TECHNOLOGY SOLUTIONS FOR SINGLE-INCISION SURGERY DA VINCI Xi SINGLE-SITE SOLUTIONS FOR SINGLE-INCISION SURGERY TECHNOLOGY DA VINCI Xi SINGLE-SITE TECHNOLOGY AT INTUITIVE SURGICAL, WE STRIVE TO MAKE SURGERY MORE EFFECTIVE, LESS INVASIVE, AND EASIER ON

More information

C Series Mechanical Pit Levelers. P.O. Box 338 Spring Hill, Tennessee fax:

C Series Mechanical Pit Levelers. P.O. Box 338 Spring Hill, Tennessee fax: C Series Mechanical Pit Levelers SM P.O. Box 338 Spring Hill, Tennessee 37174 1-800-251-3382 fax: 931-486-0316 www.pioneerleveler.com DO NOT install, operate or service product unless you have read and

More information

BESTMOVE STANDING TRANSFER AID

BESTMOVE STANDING TRANSFER AID BESTMOVE STANDING TRANSFER AID MODELS BestMove 400 BestMove 450 PRODUCT FEATURES *BestMove 450 shown Ensure the product has been assembled according to the instructions in this manual. OPERATING ASSEMBLY

More information

Agilent 16048D Test Leads

Agilent 16048D Test Leads Agilent 16048D Test Leads Operation and Service Manual Fourth Edition Agilent Technologies Japan, Ltd. Agilent PN 16048-90031 May 2000 Printed in Japan Notices The information contained in this document

More information