TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 2 (26)

Size: px
Start display at page:

Download "TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 2 (26)"

Transcription

1

2 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 2 (26) Document History Version Date of issue Revision(s) Issued by 01 20FEB2012 N/A (First version of document) David Graaf, Envirotainer

3 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 3 (26) Summary The verification testing of the RKN e1 container consists of basic tests and complementary tests at different container set temperatures (i.e. +5 C and +20 C). The basic tests include worst case load tests, specification tests and autonomy tests. This report describes the worst case load tests where two different load configurations, maximum and minimum load were studied in order to define worst case load for continued testing. Both loads were tested in warm and cold conditions with a container set temperature of +5 C. Both the minimum and maximum loads were loaded on a standard US-pallet, since this is the largest pallet that is commonly used in shipments. The RKN e1 can fit larger loads, but in order to generate generic material for all Envirotainer containers, the US-pallet(s), packed according to below has been chosen as standard loads. The performance and autonomy of the Envirotainer RKN e1 container were evaluated from the result of four (4) tests carried out on one (1) container, under the conditions specified below: Purpose Test 1 Load: Minimum Temperature profile: -15 C Test 2 Load: Minimum Temperature profile: +35 C Test 3 Load: Maximum Temperature profile: -15 C Test 4 Load: Maximum Temperature profile: +35 C The main purpose of these tests was to define the worst case load for the RKN e1 container through testing of minimum load and maximum load and comparing the results. The defined worst case load will then be used for the continued verification testing. The secondary purpose of these tests was to evaluate the performance and autonomy of the container at the different ambient temperatures. The performance is defined as the duration that the container manages to maintain the product temperature between +2 C and +8 C and the autonomy is the duration until the container battery level reaches 0 %. In both the performance and the autonomy cases the duration was calculated from test start. Acceptance Criteria In order for the test to be accepted, the following criteria need to be fulfilled. Product temperature within ±2 C of container set temperature at test start Battery level at 95% or above at test start Container alerts or alarms are allowed during the test, taken that they do not affect the function or the temperature in the container in any way Climate chamber temperature within ±3 C of chamber set temperature during the test after a 2 hour stabilization period from changing the set temperature

4 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 4 (26) Conclusion Maximum load obtained the largest values for the total average of absolute difference from set temperature for both the cold (-15 C) and warm profile (+35 C) and therefore maximum load is determined to be the worst case load and will therefore be used in further tests. The results show that at -15 C the container performance autonomy was approximately 27 hours. At +35 C the container performance was approximately 22.5 hours regarding both performance and autonomy. There was no major difference in performance and autonomy between the two loads for the different profiles. The tests were accepted since the acceptance criteria were fulfilled. Table 1-1. Summary of results. Test Load Profile Product temperature between +2.0 C and +8.0 C (performance) Battery level at 0 % (autonomy) Total average of absolute difference from set temperature 1 Min -15 C 27 h 26 h 50 min Min +35 C 23 h 22 h 20 min Max -15 C 27 h 10 min 26 h 50 min Max +35 C 22 h 40 min 22 h 30 min 0.66

5 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 5 (26) Table of Contents 1 Introduction Purpose Scope Acceptance Criteria Abbreviations and Definitions Evaluation Rationale for determination of worst case load Example Conditions Specific conditions General conditions Equipment Equipment list Envirotainer RKN e1 container Climate chamber Temperature measuring device Calibration Test setup Container load Minimum load Maximum load Temperature measuring Types of device Measuring frequency Positions (minimum load) Positions (maximum load) Container data Temperature profiles and charging Test procedure Preparations Preconditioning Container Product... 19

6 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 6 (26) 10.3 Test Post test work Results General Test 1 Minimum load at -15 C ambient Test 2 Minimum load at +35 C ambient Test 3 Maximum load at -15 C ambient Test 4 Maximum load at +35 C ambient Worst case load Minimum load at 15 C ambient Minimum load at +35 C ambient Maximum load at -15 C ambient Maximum load at +35 C ambient Conclusion... 26

7 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 7 (26) 1 Introduction The verification testing of the RKN e1 container consists of basic tests and complementary tests at different container set temperatures (i.e. +5 C and +20 C). The basic tests include worst case load tests, specification tests and autonomy tests. This report describes the worst case load tests where two different load configurations, maximum and minimum load were studied in order to define worst case load for continued testing. Both loads were tested in warm and cold conditions with a container set temperature of +5 C. Both the minimum and maximum loads were loaded on a standard US-pallet, since this is the largest pallet that is commonly used in shipments. The RKN e1 can fit larger loads, but in order to generate a generic material for all Envirotainer containers, the US-pallet, packed according to below has been chosen as standard maximum load. The performance and autonomy of the Envirotainer RKN e1 container were evaluated from the result of four (4) tests carried out on one (1) container, under the conditions specified below: Test 1 Load: Minimum Temperature profile: -15 C Test 2 Load: Minimum Temperature profile: +35 C Test 3 Load: Maximum Temperature profile: -15 C Test 4 Load: Maximum Temperature profile: +35 C The tests were carried out in the Envirotainer Laboratory in Lagga, Sweden. 2 Purpose The main purpose of these tests was to define the worst case load for the RKN e1 container through testing of minimum load and maximum load and comparing the results. The defined worst case load will then be used for the continued verification testing. The secondary purpose of these tests was to evaluate the performance and autonomy of the container at the different ambient temperatures. The performance is defined as the duration that the container manages to maintain the product temperature between +2 C and +8 C and the autonomy is the duration until the container battery level reaches 0 %. In both the performance and the autonomy cases the duration was calculated from test start. 3 Scope This report applies to the Envirotainer RKN e1 container with the test setup and test procedure described in sections 9 and 10.

8 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 8 (26) 4 Acceptance Criteria In order for the test to be accepted, the following criteria need to be fulfilled. Product temperature within ±2 C of container set temperature at test start Battery level at 95% or above at test start Container alerts or alarms are allowed during the test, taken that they do not affect the function or the temperature in the container in any way Climate chamber temperature within ±3 C of chamber set temperature during the test after a 2 hour stabilization period from changing the set temperature 5 Abbreviations and Definitions Product Empty boxes Worst case load See description in section 6 Failure In these tests defined as: product temperatures below +2 C or above +8 C recorded 6 Evaluation The worst case load is defined as the load with largest total average difference from set temperature +5 C and determined according to below: 1. Calculate the absolute difference from set temperature +5 C for each value on each sensor during a three (3) hour period. The three (3) hour period should start three (3) hours after test start (start of temperature profile) 2. Calculate the average difference from +5 C based on all values 3. The largest average is defined as the worst case load 6.1 Rationale for determination of worst case load By calculating the difference from set temperature in each measuring point for each sensor followed by calculating the total average deviation over a predetermined time range, an adequate measure of the difference from set temperature within the load is achieved. A larger total average indicates a larger deviation from the set temperature throughout the defined test period, which in turn indicates larger difficulties in maintaining +5 C. The larger total average therefore is defined as the worst case load.

9 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 9 (26) 6.2 Example In the case below the maximum load is considered to be the worst case load since the total average difference is larger for maximum load than for minimum load, see Table 6-1, Table 6-2, Table 6-3 and Table 6-4. Minimum load Table 6-1. Example temperature values for minimum load. Time Sensor 1 ( C) Sensor 2 ( C) Sensor 3 ( C) 00: : : : Table 6-2. Calculation for minimum load. Absolute difference from +5 ºC. Time Sensor 1 ( C) Sensor 2 ( C) Sensor 3 ( C) 00: : : : Total average 0.35 Maximum load Table 6-3. Example temperature values for maximum load. Time Sensor 1 ( C) Sensor 2 ( C) Sensor 3 ( C) 00: : : : Table 6-4. Calculation for maximum load. Absolute difference from +5 ºC. Time Sensor 1 ( C) Sensor 2 ( C) Sensor 3 ( C) 00: : : : Total average 0.46

10 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 10 (26) 7 Conditions The performance of the Envirotainer RKN e1 container was evaluated from the result of four (4) tests carried out on one (1) container, under the conditions specified below: 7.1 Specific conditions Test 1 Load: Minimum Temperature profile: -15 C Test 2 Load: Minimum Temperature profile: +35 C Test 3 Load: Maximum Temperature profile: -15 C Test 4 Load: Maximum Temperature profile: +35 C 7.2 General conditions Container set temperature: +5 C The same container was used for all tests The same climate chamber was used in all tests The load and container were preconditioned before each test The load was made up of empty cardboard boxes on a pallet The container batteries were fully loaded i.e. 95% or higher when test starts The container was running on batteries during the test The tests were running until failure (product temperatures below +2 C or above +8 C recorded)

11 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 11 (26) 8 Equipment 8.1 Equipment list The following equipment was used in the tests, see Table 8-1. Table 8-1. Equipment for the test. Container Climate chamber Equipment Temperature monitoring device Data acquisition hardware (thermocouples type T) Data acquisition software (thermocouples type T) DASYLab 6.0 Reference thermometer Calibration bath Product box Pallet Bubble wrap Stand Description Envirotainer RKN e1 container (RKN60714PC) Climate chamber 2 at Envirotainer, Lagga, Sweden Thermocouple type T sensor, model: N/N-24 (Pentronic) Datascan 7000 series (msl) 1521 with Pt-100 probe (Hart Scientific) CBN 8-30 (Heto Holten) containing a water/glycol mixture Corrugated cardboard box ES140 (Boxon) 360 x 250 x 179 mm, 14.2 x 9.8 x 7.0 in (L x W x H) Corrugated cardboard box modified ES140 (Boxon) 280 x 250 x 179 mm, 11.0 x 9.8 x 7.0 (L x W x H) US pallet Protech 4048 (TMF Corporation) 1219 x 1016 x 140 mm, 48 x 40 x 5.5 in (L x W x H) Plastic bubble wrap Plastic pole, PVC pipe 32 x 27.8 mm (Bauhaus) inserted in an axle stand, art no (Biltema)

12 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 12 (26) 8.2 Envirotainer RKN e1 container Information regarding the container that was used in the tests can be seen in Table 8-2 Figure 8-1. Envirotainer RKN e1 container. RKN60714PC was used in the tests. Table 8-2. Technical specifications Envirotainer RKN e1 container. Refrigerating system Thermostat-controlled air conditioning system with compressor cooling and electrical heating. Powered by rechargeable batteries. Recharging power supply Maximum power consumption during charging Maximum charging time V AC, Hz 1850 W Set temperature range 0 C to +20 C +32 F to +68 F Dimensions External cube (volume) 4.8 m 3 External dimensions (L x W x H) Loading space (L x W x H) Door opening (L x H) 8 h cu.ft Internal cube (volume) for refrigerated cargo 2.3 m 3 Tare weight Max net weight of cargo Weight Other information 2000 x 1530 x 1620 mm 78.7 x 60.2 x 63.8 in 1340 x 1319 x 1315 mm 52.7 x 51.9 x 51.7 in 1340 x 1315 mm 52.7 x 51.7 in 81.9 cu.ft 635 kg 1400 lbs 953 kg 2100 lbs Suitable for use on aircraft A300, A310, A330, A340, A380, B747, B767, B777, DC10, IL86, MD11, L1011 Forkliftable with a slot-height of 95 mm (3.7 in).

13 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 13 (26) 8.3 Climate chamber 2 Information regarding the climate chamber that was used in the tests can be seen in Table 8-3. Table 8-3. Technical specifications of Envirotainer climate chamber 2. Heating and cooling system The cooling system of the chamber (denomination: KA2) consists of 3 compressors connected in a cooling media circuit. There are 3 evaporators inside the chamber connected in parallel in the cooling media circuit. Electrical radiators are installed in front of the fans of the evaporators. The air flow from the fans is constant and each fan blows with a flow rate of 0.9 m 3 /s. The temperature in the climate chamber is regulated by a test control system provided by Hansson Kyl AB. Four sensors in the chamber (positioned on the walls) are used as input for the regulation system. Set temperature range Internal (L x W x H) Door opening (L x H) Dimensions Other information Installed in 2003 by Hansson Kyl i Uppsala AB (ref no 2219) -30 C to +50 C -22 F to +122 F 3.7 x 3.7 x 3.1 m x x 122 in 3.0 x 2.2 m x 86.6 in The accuracy of the climate chamber is ±3 C (±5.4 F) from set temperature. The climate chamber needs a stabilization period when a temperature change is made. 8.4 Temperature measuring device Calibration The thermocouples were calibrated in a calibration bath against a reference thermometer before and after the test series. The reference thermometer is calibrated yearly at Pentronic AB, accredited by Swedish authority to standard ISO/IEC The performance and properties of the used testing equipments are traceable, via Pentronic s accredited laboratory (SWEDAC 0076), to standards at the SP Swedish National Testing and Research Institute.

14 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 14 (26) 9 Test setup 9.1 Container load A minimum and a maximum container load were used in the tests. In both loads each layer of boxes on each pallet was arranged as seen in Figure 9-1. Figure 9-1. Arrangement of boxes on one pallet for RKN container (view from top). The front is the side towards the container doors Minimum load Since the airflow is affected by the load it was decided to put one layer of boxes inside the container instead of running it empty. The load was constituted of product boxes on a pallet, for details see Table 9-1 The boxes containing temperature sensors were filled with bubble wrap The remaining boxes were empty The load was not stretch wrapped The load was centered in the load area Dimensions of load (including pallet): 1219 x 1016 x 319 mm Table 9-1. Load details for minimum load in RKN container. Number of pallets per container 1 Number of layers of boxes per pallet: 1 Number of boxes per layer Number of boxes per pallet Amount 14 (12 ES140 and 2 modified ES140) 14 (12 ES140 and 2 modified ES140)

15 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 15 (26) Picture 9-1. Minimum load placed inside the RKN e Maximum load The load was constituted of product boxes on a pallet, for details see Table 9-2 The boxes containing temperature sensors were filled with bubble wrap The remaining boxes were empty The load was not stretch wrapped The load was centered in the load area Dimensions of load (including pallet): 1219 x 1016 x 1214 mm Table 9-2. Load details for maximum load in RKN container. Amount Number of pallets per container 1 Number of layers of boxes per pallet 6 Number of boxes per layer 14 (12 ES140 and 2 modified ES140) Number of boxes per pallet 84 (72 ES140 and 12 modified ES140) Picture 9-2. Maximum load placed inside the RKN e1

16 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 16 (26) 9.2 Temperature measuring Types of device One (1) type of temperature measuring device was used in the tests. Thermocouple sensors measured the product and ambient temperature Measuring frequency The thermocouples were set to log the temperature every 10 minutes, in sync with the logging of the container data, see section Positions (minimum load) To measure product temperature, thermocouples were placed inside the boxes, in the air. To measure ambient temperature, thermocouples were placed on a stand inside the climate chamber. Test ID number of the sensor in each position was noted in lab test worksheet FT For sensor positions, see Table 9-3 and Figure 9-3 Table 9-3. Sensor positions for minimum load. Positions Thermocouples Ambient 2 of box 4 Center of Load 1 of box 4 Total amount: 11 Figure 9-3. Sensor positions for minimum load.

17 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 17 (26) Positions (maximum load) To measure product temperature, thermocouples were placed inside the boxes, in the air. To measure ambient temperature, thermocouples were placed on a stand inside the climate chamber. Test ID number of the sensor in each position was noted in lab test worksheet FT For sensor positions, see Table 9-4 and Figure 9-4. Table 9-4. Sensor positions for maximum load. Positions Thermocouples Ambient 2 layer 4 Center of Load 1 layer 4 Total amount: 11 Figure 9-4. Sensor position for maximum load. 9.3 Container data The container ground handler (GH) data was downloaded after each test in order to see the container power steps and also to capture any alerts and alarms during the test.

18 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 18 (26) 9.4 Temperature profiles and charging For climate chamber set temperature during preconditioning, see Table 9-5. The temperature profiles in Table 9-6 and Table 9-7 were used in the tests. The climate chamber was programmed to the temperature profiles allowing the temperatures to automatically adjust at the times set. The container was fully charged (battery level of 95% or higher) at the start of each test. To achieve this level the batteries need to be drained to less than 80% before charging for 12 hours. The container was not charged during the profile. Table 9-5. Preconditioning. Step Climate chamber set temperature ( C) Duration (h) Charging Preconditioning +5 Until product temperature was +5 ºC ±2 C Charging Table C profile. Step Climate chamber set temperature ( C) Duration (h) Charging 1-15 Until failure Off Table C profile. Step Climate chamber set temperature ( C) Duration (h) Charging Until failure Off

19 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 19 (26) 10 Test procedure Test information including deviations or incidents (if any) was noted in lab test worksheet FT Preparations The thermocouples were calibrated at -30 C, +5 C and +50 C The container was placed in the climate chamber The container clock was checked to show the correct time The temperature sensors were positioned in the loads and in the climate chamber The load was placed in the container The climate chamber was programmed with the specific temperature profile 10.2 Preconditioning Container The container was preconditioned inside the climate chamber The container was preconditioned until product temperature was +5 ºC ±2 C The container was charged during the preconditioning The container was switched off during the preconditioning The container doors were open during the preconditioning Product The load was placed inside the container and it was preconditioned in the chamber at the same time as the container 10.3 Test After the preconditioning, each test was performed according to the following: The logging of temperature data was started The product temperature in the center of the load was checked to be within ±2 C of container set temperature The doors of the containers were closed The container was switched on and the battery level was checked to be 95% or higher The container temperature was set to +5 C The container was unplugged to run on batteries The temperature profile was started The test was run until failure The temperature data and GH data was downloaded and analyzed 10.4 Post test work The thermocouples were calibrated at -30 C, +5 C and +50 C

20 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 20 (26) 11 Results 11.1 General The performance and autonomy have been calculated from test start.

21 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 21 (26) 11.2 Test 1. Minimum load at -15 C ambient Figure Ambient and product temperatures for Test 1. Container ID: RKN60714PC Climate chamber: CC2 (small chamber) Profile start date: 27JAN2012 Profile start time: 10:00 Battery level at test start: 100 % The ambient temperature was ±3 C from the set temperature after 30 minutes. The product temperature was maintained between +2 C and +8 C for 27 hours and 10 minutes. The battery level reached 0 % after 26 hours and 50 minutes. Container was running on battery throughout the test. There were no alerts or alarms observed during the test. All acceptance criteria were fulfilled for the test.

22 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 22 (26) 11.3 Test 2. Minimum load at +35 C ambient Figure Ambient and product temperatures for Test 2. Container ID: RKN60714PC Climate chamber: CC2 (small chamber) Profile start date: 23JAN2012 Profile start time: 09:00 Battery level at test start: 100 % The ambient temperature was ±3 C from the set temperature after 1 hour. The product temperature was maintained between +2 C and +8 C for 23 hours. The battery level reached 0 % after 22 hour and 20 minutes. Container was running on battery throughout the test. There were no alerts or alarms observed during the test. All acceptance criteria were fulfilled for the test.

23 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 23 (26) 11.4 Test 3. Maximum load at -15 C ambient Figure Ambient and product temperatures for Test 3. Container ID: RKN60714PC Climate chamber: CC2 (small chamber) Profile start date: 18JAN2012 Profile start time: 09:00 Battery level at test start: 95 % The ambient temperature was ±3 C from the set temperature after 30 minutes. The product temperature was maintained between +2 C and +8 C for 27 hours and 10 minutes. The battery level reached 0 % after 26 hour and 50 minutes. Container was running on battery throughout the test. There were no alerts or alarms observed during the test. All acceptance criteria were fulfilled for the test.

24 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 24 (26) 11.5 Test 4. Maximum load at +35 C ambient Figure Ambient and product temperatures for Test 4. Container ID: RKN60714PC Climate chamber: CC2 (small chamber) Profile start date: 23JAN2012 Profile start time: 09:00 Battery level at test start: 100 % The ambient temperature was ±3 C from the set temperature after 40 minutes. The product temperature was maintained between +2 C and +8 C for 22 hours and 40 minutes. The battery level reached 0 % after 22 hours and 30 minutes. Container was running on battery throughout the test. There were no alerts or alarms observed during the test. All acceptance criteria were fulfilled for the test.

25 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 25 (26) 11.6 Worst case load Table 11-1, Table 11-2, Table 11-3 and Table 11-4 show the calculated average difference from +5 C (container set temperature) for each test Minimum load at 15 C ambient Table Minimum load at 15 C ambient. Position Center of Load Sensor Average of absolute difference from set temperature ( C) 1,37 1,43 1,10 0,78 1,25 1,08 1,01 0,90 0,73 Total average (ºC) 1, Minimum load at +35 C ambient Table Minimum load at +35 C ambient. Position Center of Load Sensor Average of absolute difference from set temperature ( C) 1,08 0,71 0,74 0,10 1,14 0,35 0,12 0,10 0,36 Total average (ºC) 0, Maximum load at -15 C ambient Table Maximum load at -15 C ambient. Position Center of Load Sensor Average of absolute difference from set temperature ( C) 1,79 1,62 1,44 0,93 1,02 1,09 0,72 0,84 0,59 Total average (ºC) 1, Maximum load at +35 C ambient Table Maximum load at +35 C ambient. Position Center of Load Sensor Average of absolute difference from set temperature ( C) 1,25 0,56 1,04 0,21 0,44 0,21 0,86 0,45 0,95 Total average (ºC) 0,66

26 TR Test Report Verification Testing RKN e1 Container (+5 C) Worst Case Load Version: 01 Page: 26 (26) 12 Conclusion The performance and autonomy of the Envirotainer RKN e1 container were evaluated from the results of four (4) tests carried out on one (1) container. Two different load configurations, maximum and minimum load, were studied in order to define worst case load for continued testing. Both loads were tested in warm and cold conditions with a container set temperature of +5 C. See Table 12-1 for a summary of the results. Maximum load obtained the largest values for the total average of absolute difference from set temperature for both the cold (-15 C) and warm profile (+35 C) and therefore maximum load is determined to be the worst case load and will therefore be used in further tests. The results show that at -15 C the container performance autonomy was approximately 27 hours. At +35 C the container performance was approximately 22.5 hours regarding both performance and autonomy. There was no major difference in performance and autonomy between the two loads for the different profiles. The tests were accepted since the acceptance criteria were fulfilled. Table Summary of results. Test Load Profile Product temperature between +2.0 C and +8.0 C (performance) Battery level at 0 % (autonomy) Total average of absolute difference from set temperature 1 Min -15 C 27 h 26 h 50 min Min +35 C 23 h 22 h 20 min Max -15 C 27 h 10 min 26 h 50 min Max +35 C 22 h 40 min 22 h 30 min 0.66