Stamolys CA71MN. Technical Information

Size: px
Start display at page:

Download "Stamolys CA71MN. Technical Information"

Transcription

1 Technical Information Manganese Analyser Compact photometric analysis system for manganese measurement in drinking water and wastewater Application In soil, manganese is often found as brownstone. With changing ground water level, water-soluble manganese compounds are formed due to microbial processes. Manganese monitoring in drinking water treatment plants Precipitant monitoring after aeration and filtration processes in drinking water and wastewater treatment plants Your benefits Stainless steel or glass-fibre reinforced carbon housing available Two channel version available Measured value storage using integrated data logger Automatic calibration and self-cleaning Free selectable measuring, cleaning and calibration intervals TI6C/07/en/0.0 06

2 Function and system design Measuring principle After sample conditioning, the analyser sample pump conveys a part of the filtrate to a mixing vessel. The reagent pump adds reagent at a specific ratio. As a result of the reaction, the sample turns a characteristic colour. The photometer determines the sample s absorption of an emitted light at a specific wavelength (s. Fig., Pos. ). The wavelength is parameter specific. The absorbance is proportional to the concentration of the specified parameter in the sample (Pos. ). Additionally, the absorption of a reference light is determined to receive a genuine measuring result. The reference signal is subtracted from the measuring signal to prevent any effects due to turbidity, contamination and ageing of the LEDs. The temperature in the photometer is controlled thermostatically so that the reaction is reproducible and takes place within a short period of time. Reference LED Emitter LED Sample Reference detector Measuring detector Photometric principle a0006 Manganese Photometric manganese determination Interferences Manganese is the second most common heavy metal in the earth's crust after iron. It is also contained in many iron ores. Manganese is often present in the ground as manganese ore. The manganese ore becomes water soluble through the action of air as the ground water level varies. Manganese, in normal concentrations as found in water, is not damaging to the health. It is an important trace element for plants. There are no limit values in wastewater while for drinking water there is the limit value of 0.0 mg/litre, since higher concentrations in water lead to odour, colouring, turbidity and fine grained build-up. Dairies require manganese-free water. Leuco malachite method In the presence of manganese, the reagent indicator turns the sample colour to green-blue. The absorption is determined at a wavelength of 6 nm. The absorption intensity is proportional to the manganese concentration in the sample. The reference wavelength is 880 nm. No interferences up to the given concentration: Concentration [mg/l] Interference 00,000 Mg + 70,000 Cl,000 SO 700 Ca + (as CaCO ) 0 Zn + Fe +, Fe + " Caution! Amines, aldehydes and alcohols interfer in any concentration. Strongly buffered samples or extreme ph values may exceed the buffer capacity of the reagent. Endress+Hauser

3 Sample conditioning Micro/ultrafiltration (Stamoclean CAT0, optional) A membrane filter element is suspended directly into the wastewater basin or channel. A hose pump is located in a pump box on the basin rim.the pump creates a vacuum between the membrane and the carrier plate of the filter element. This vacuum makes the filtrate pass through the filter membrane. Suspended materials, particles, algae and bacteria are collected on the surface of the membrane. Due to alternating pumping and pause, intervals of more than one month are achieved between cleaning cycles. Parallel connection of two or four filter elements increases the sampling quantity up to approx. l/h. The hose pump pressure transports the sample to a collecting vessel near the analyzer over a distance of 0 m. For distances up to 00 m the sample is transported to the collecting vessel by means of compressed air. The analyzers suck the needed sample volume from the collecting vessel. Membrane filtration (Stamoclean CAT, optional) A sample flow of 0.8 to.8 m /h is continuously conducted through the micro filter via a pressure pipe. A part of the sample passes the filter membrane and is then conveyed to the measuring device as filtrate. Sampling is based on the cross flow filtration principle. The PTFE filter membrane separates particles with sizes > 0. µm from the filtrate. These particles are collected in front of the membrane and are washed away with the sample flow. The medium is conducted in a meander-like channel through the filter element. This results in a constantly high flow rate. The high flow rate generates the self cleaning effect. Therefore, mechanical drives for the generation of a flow at the filter surface are not necessary. Backwash filter (Stamoclean CAT, optional) A sample flow of to. m /h is permanently conveyed through the backwash filter by means of a sampling pump or compressed air or rinse water. The filtrate passes through the wedge wire sieve and is then transported to the measuring device. Clogging is minimised by the flow at the wedge wire sieve. Automatic backwashing results in a filter operating time of several weeks. The automatic backwashing and a small compressor or compressed air resp. rinse water supply guarantee low-maintenance and low-energy operation. Customer specific solution Before analysis, the sample has to be conditioned and to be transported to an external or to the delivered collecting vessel. Measuring system A complete measuring system comprises: An analyser A sample conditioning system (optionally): Micro filtration / ultra filtration Stamoclean CAT0 or Stamoclean CAT Backwash filter Stamoclean CAT Customer specific solution Collecting vessel (see product structure) Micro / ultra filtration Control box Pump Control unit Collecting unit (optional) Collecting vessel Overflow Analyser Aeration basin Membrane filter 9 8 Measuring system with Stamoclean CAT0 a00066 Endress+Hauser

4 Stamoclean CAT Inlet Sample pump or hydraulic main Filtrate line Collecting vessel Overflow Analyser Analyser sample line Outlet Measuring system with Stamoclean CAT a0006 Backwash filter 6 7 Stamoclean CAT Compressor or compressed air Sample pump or hydraulic main Sample outlet Collecting vessel Overflow Analyser 7 6 Measuring system with Stamoclean CAT a00067 Standard applications Drinking water quality monitoring at the outlet of the water treatment plant Sampling from hydraulic main, pressureless collecting vessel with level monitoring: Analyser with collecting vessel, -AxCA Outlet monitoring in the water treatment plant Sampling from hydraulic main, pressureless collecting vessel with level monitoring: Backwash filter Stamoclean CAT (order no. CAT-Axxx) Compressor for CAT (order no. ) Analyser with collecting vessel, -BxBA Input Measured variable Measuring ranges Wavelength Reference wavelength Mn [µg/l]... 0 µg/l (MN-A) µg/l (MN-B) 6 nm (MN-A) 0 nm (MN-B) 880 nm Output Output signal Signal on alarm Load 0/... 0 ma Contacts: limit contacts (per channel), system alarm contact optional: end of measurement (with two channel version display of channel no. available) max. 00 Ω Endress+Hauser

5 Photometer +7V Photometer -7V Photometer 0V fx Analog Out Kanal Kanal MP MP6 Channel Channel BK 6 0/-0mA F M0.A F M0.A Shield COM 8 6 GN 6 0V BK 8 Shield 0 Shield NC 7 keine Probe 80 7 V Screen Shield BN -7 V YE No sample 9 I + NO 6 0 AC Motor F OR 79 7 V V 6 Reserve 8 I - COM BU 7 L WH 9 T0.A Mess BK 78 7 V 68 6 fx 7 YE 6 N YE Shield Meas NC 8 RD L Elektronik RD F V BU +V I + WH 0 NO 7 Electronics GN/YE PE T0.A YE V BN 66 GY 9 Status GN 0V I - 9 0V RELAIS / Relay: max. Last / load: A bei / at /0V AC, A bei / at 0V DC Klemmen führen auch bei ausgeschaltetem Gerät Spannung!!! Terminals voltage to ground even when the unit is switched off!!! NETZ MAINS L N PE Data interface Data logger Load capacity RS C 0 data pairs per channel with date, time and measured value 00 data pairs with date, time and measured value for calibration factor determination (diagnostic tool) 0 V / V AC max. A, 0 V DC max. A Electrical connection " Caution! Power supply The following figure ( å ) shows the connection department sticker as an example. Terminal assignment and cable core colours can be different to the originals. For connecting your analyser only use the terminal assignment of the connection department sticker in the device ( å )! Photometer +7V Photometer -7V Photometer 0V fx Analog Out Kanal Kanal MP MP6 Channel Channel BK 6 0/-0mA F M 0. A F M 0. A Shield COM 8 6 GN 6 0V BK 8 Shield 0 Shield NC 7 keine Probe 80 7 V Screen Shield BN -7 V YE No sample 9 I + NO 6 0 OR 79 7 V V 6 Reserve 8 I - COM WH 9 BK 78 7 V 68 Mess YE 6 fx 7 Shield Meas NC 8 RD 77 7 V BU V 6 I + NO WH 0 7 YE V BN 66 GY 9 Status GN 0V I - 9 0V RELAIS / Relay: max. Last / load: A bei / at /0V AC, A bei / at 0V DC SCHRITTMOTOR STEP MOTOR LECKF. LEAK S. Klemmen führen auch bei ausgeschaltetem Gerät Spannung!!! Terminals voltage to ground even when the unit is switched off!!! ALARM ALARM OPTION STÖRUNG ERROR ALARM ALARM BU 7 L AC Motor T 0.A F NETZ MAINS YE RD GN/YE 6 N L PE AC MOTOR Elektronik Electronics T 0.A F L N PE Fig. : Example of the connection sticker a00069 SCHRITTMOTOR STEP MOTOR STEP MOTOR LECKF. LEAK S. LEAK S. ALARM ALARM STÖRUNG ERROR ALARM ALARM AC MOTOR AC MOTOR ALARM ALARM OPTION ALARM ALARM Fig. : Analyser from top (open version resp. swung out) Connection department sticker Printed circuit board with terminal strip Backside of the analyser a0007 Supply voltage V AC / 0 V AC ±0%, 0/60 Hz Endress+Hauser

6 Power consumption Current consumption Fuses approx. 0 VA approx. 0. A at 0 V approx. 0. A at V x time-lag 0. A for electronics x medium time-lag 0. A for photometer x time-lag 0. A for motors Performance characteristics Time between two measurements Maximum measured error Measuring interval Reaction time Sample requirement Reagent requirement Calibration interval Rinse interval t mes = reaction time + rinse time + waiting time + rinse again time + filling time + sampling time + reagent refusal time (min. waiting time = 0 min) ± % of measuring range end t mes to 0 min minutes ml (0.00 US.gal.) per measurement MN-A: x 0.8 ml ( US.gal). l (0.9 US.gal) per reagent per month with 0 minute measuring interval MN-B: x 0. ml (0.000 US.gal).9 l (0. US.gal) per reagent per month with 0 minute measuring interval 0 to 70 h 0 to 70 h Rinse time selectable from 0 to 00 s (standard = 60 s) Rinse again time Filling time Maintenance interval Servicing requirement 0 s s (MN-A) s (MN-B) 6 months (typical) minutes per week (typical) Environment Ambient temperature Humidity... 0 C (... 0 F), avoid strong fluctuations below the condensation limit, installation in usual, clean rooms outdoor installation only possible with protective devices (customer supplied) Ingress protection IP 6 Endress+Hauser

7 Process Sample temperature Sample flow rate Consistence of the sample Sample inlet to 0 C ( to 0 F) min. ml (0.00 US.gal.) per min low solid content (< 0 ppm) pressureless Mechanical construction Design, dimensions Stainless steel housing Stainless steel version a0006-en Endress+Hauser 7

8 RS 0/.6 GFR housing 6 / 6. 0/9.8 Ø 6/.0 Ø 6/0.6 0 /.7 0 /.99 60/.6 60/6.0 0/. 00/.9 7-/ (M 6) 00/7.87 0/.7 /9.7 /.8 6/6.0 /.9 9/.69 Ø /0.8 78/.07 7/.8 Ø 6/.0 Ø / /. 6/.8 Ø 6/.0 Ø 6/0.6 60/.7 6/7.7 6/8. 00/9.69 0/.7 Ø 7/.06 mm / inch GFR version a000-en Open version 8/ /.8 9/0.7 97/7.76 6/. 7/6.89 Ø6/0. 9/7.6 /.80 9/0. Open version (without housing) 8/0. mm/inch a0006-en 8 Endress+Hauser

9 Collecting vessel 7 60 xø x Ø 8/0 Ø / 6 xø./0. 7 Ø6 Ø 0H7 * * 8 M0x. Ø * 70 Collecting vessel at analyser (optional) Ventilation Sample inlet from sampling Collecting vessel Electrical connections Analyser sample inlet a0006 Collecting vessel dimensions * variable, freely adjustable dimensions 6 Sampling for analyser 7 analyser outlet 8 Sample overflow a0006-en Weight GFR housing Stainless steel housing Without housing approx. 8 kg (6.7 lb) approx. kg (7.8 lb) approx. kg (0.6 lb) Material Housing: Front windows: Endless hose: Pump hose: Valves: Stainless steel.0 (AISI 0) or glass-fibre reinforced carbon(gfr) Polycarbonate C-Flex, Norprene Tygon, Viton Tygon, silicone Connecting the sample line One channel version Collecting vessel (at analyser, with or without level measurement) Connection hose ID. mm (0.") Customer collecting vessel Connection hose ID.6 mm (0.06") Max. distance from collecting vessel to analyser m (.8 ft) Max. height difference from collecting vessel to analyser 0. m (.6 ft) Two channel version Depending on the ordered version, one or two collecting vessels (with or without level measurement) are included in the scope of delivery. Level measurement is only possible for one channel. Only one collecting vessel can be mounted at the housing. The second is to be placed nearby the analyser. Sample outlet Connection Hose ID 6. mm (0.") Max. length of closed loop: m (.8 ft) Open outlet downgrade installed No combination of several devices to a closed-loop system Min. volume per measurement 0 ml (0.00 US.gal.) Endress+Hauser 9

10 Human Interface Display and operating elements M CE E K RS Display and operating elements LED (measured value) LC display (measured value and status) Serial interface RS Operating keys and control LEDs a Ordering information Product structure Measuring range A Measuring range... 0 µg/l Mn B Measuring range µg/l Mn Y Special version acc. to customer s specification Sample transfer Sample transfer from one measuring point (one-channel version) Sample transfer from two measuring points (two-channel version) Power supply 0 0 V AC / 0 Hz V AC / 60 Hz V AC / 0 Hz 0 V AC / 60 Hz Collecting vessel for up to analysers A Without collecting vessel B With collecting vessel without level measurement C With collecting vessel with level measurement (one-channel version only) D With two collecting vessels without level measurement (two-channel version) Housing version Without housing With GFK housing With stainless steel.0 (AISI 0) housing Communication A 0/... 0 ma, RS Additional equipment Quality certificate Quality certificate + set of inactive reagents Quality certificate + three sets of inactive reagents CA7MN - complete order code Scope of delivery The scope of delivery comprises: an analyser with mains plug a cleaning injector a tin of silicone spray a Norprene hose, length. m (8. ft), ID.6 mm (0.06") 0 Endress+Hauser

11 ! Note! a C-flex hose, length. m (8. ft), ID 6. mm (0.") a C-flex hose, length. m (8. ft), ID. mm (0.") two hose fittings of each size:.6 mm x.6 mm (0.06" x 0.06").6 mm x. mm (0.06" x 0.") 6. mm x. mm (0." x 0.") two T-hose fittings of each size:.6 mm x.6 mm x.6 mm (0.06" x 0.06" x 0.06"). mm x. mm x. mm (0." x 0." x 0.") an interference suppressor for the current output edge covers a quality certificate Operating Instructions (English). Please, order reagents separately with analyser version CA7XX-XXXXXX. With all other versions, inactive reagents are included in the scope of delivery. You have to mix the reagents before using them. Please, read the instructions attached to the reagents. Certificates and approvals approval Declaration of conformity The product meets the legal requirements of the harmonised European standards. The manufacturer confirms compliance with the standards by affixing the symbol. Test reports Quality certificate Depending on the order code, you receive a quality certificate. With the certificate the manufacturer confirms compliance with all technical regulations and the successful individual testing of your product. Accessories Reagents and standard solutions Cleaner for hoses Collecting vessel Reagent set, l MN+MN+MN each; order no. CAY8-V0AAE Reagent set, inactive, l MN+MN+MN each; order no. CAY8-V0AAH Cleaning agent, l; order no. CAY8-V0AAE Standard solution 0. mg/l Mn; order no. CAY8-V0C0AAE Standard solution 0. mg/l Mn; order no. CAY8-V0C0AAE Cleaning agent, alkaline, 00 ml; order no. CAY76-V0AAE Cleaning agent, acidic, 00 ml; order no. CAY77-V0AAE for sampling from pressurised systems results in an unpressurised continuous sample stream Collecting vessel without level measurement; order no. 088 Collecting vessel with level measurement (conductive); order no. 089 Maintenance kit Maintenance kit CAV 70: set pump hoses yellow/blue set pump hoses black/black set hose connectors per hose set order no. CAV 70-A Additional accessories Interference suppressor for control, power and signal lines order no. 800 Silicon spray order no. 0 Valve set, pieces, for two-channel version order no. Endress+Hauser

12 Upgrade kit for upgrading from one-channel to two-channel version order no. 60 Documentation Technical Information Stamoclean CAT0, TI 8C/07/en Technical Information Stamoclean CAT, TI 9C/07/en Technical Information Stamoclean CAT, TI 8C/07/en International Head Quarters Endress+Hauser GmbH+Co. KG Instruments International Colmarer Str Weil am Rhein Deutschland Tel Fax info@ii.endress.com TI6C/07/en/ Printed in Germany / FM+SGML 6.0 / DT