+ - Conductivity. Conductivity Primer. Basic principles

Size: px
Start display at page:

Download "+ - Conductivity. Conductivity Primer. Basic principles"

Transcription

1 Conductivity Basic principles The specific electrical conductivity and the electrical conductance are a measure of the ability of a solution, a metal or a gas - in brief all materials - to conduct an electrical current. In solutions, the current is carried by cations and anions whereas in metals it is carried by electrons. If a substance has a high electrical conductance G, the electrical or ohmic resistance R is low. The electrical conductance G is the reciprocal of the resistance G = 1 R The unit of R is the Ohm and the unit of G is the Siemens. At this point, it would be useful to consider the measuring technique. To measure the electrical conductance, a voltage is applied to the electrode pairs and the current that flows is measured. During this process, the cations migrate to the negative electrode, the anions to the positive electrode and the solution acts as an electrical conductor. A conductor is defined by its length and crosssection. The smaller the electrode gap l and the larger the electrode area A, the larger the measurable current at the same electrolyte concentration and same voltage. + - A A - L The electrical conductance G is given by the equation: G = γ A l = 1 ρ A l where A is the electrode area, l the electrode gap, γ the specific conductivity and ρ the specific resistance. γ and ρ are material constants with the units S/m and Ω m. This equation also illustrates the relation between the specific conductivity γ and the conductance G. Page 1 of 18

2 As well as γ, σ and κ are also customary symbols used for specific conductivity. The quotient of the length and area is the cell constant K (resulting in the unit m -1 ). K = l A If the cell constant is known, the specific conductivity can be correspondingly determined from the measured conductance and depicts the result of a conductivity measurement. Conductivity measurement cells Basically, conductivity measuring cells consist of electrode pairs to which a voltage is applied. The current that flows is measured and the conductivity is calculated from it. This is a very rough approximation. The voltage applied is an alternating voltage to reduce polarization effects. Polarization of a conductivity measuring cell includes the effects that occur at the junction between the metal and liquid when a current flows and apparently causes the conductivity of the solution to change. If a voltage is applied to an electrode, a capacitor layer (double layer) is created because the electrode attracts inversely charged ions. e - e - e - e - e - e - e - e - voltage drop With increasing depth of the electrode in the solution, the effective voltage continues to drop further. Polarization effects can be reduced or prevented by applying an alternating voltage and by optimizing the electrode areas. In an alternating field, unequal charge distribution as shown in the diagram above cannot form so easily because the ions are alternately attracted by the two electrodes. Cations and anions oscillate about their location at the cycle of the applied frequency. This effect can be compared to a tug-of-war between two equally strong teams. Page 2 of 18

3 The higher the applied frequency, the lower the polarization effects that can be expected. Because the measuring frequency at high conductivity is restricted by instrument engineering, a suitable electrode material must be used, e.g. usually graphite or platinumplated platinum. The selection depends on the required measuring range of the conductivity measurement. The "classical" conductivity measuring cell consists of an electrode pair. A further development is provided by the four-electrode TetraCon measuring cell from WTW. It has a voltage electrode pair in addition to the current electrode pair. As a result, the ohmic voltage drop is determined in the solution with current flowing through it. The conductance of the solution results from the known current intensity and measured voltage drop. This measuring method does not detect polarization resistances. Moreover, a four-electrode cell is insensitive to measuring errors that result from contamination. + - The cell constant K cannot be determined by purely "measuring the electrode gap and area. Although the field lines in the area directly between the electrodes appear linear, at the edges scatter fields form and the electrode area relevant for the measurement is larger than the actual geometric area. This effect is encountered when the cells are calibrated. The cell constant is determined using a calibration solution with known conductivity, usually a 0.01 mol/l KCl solution. This is described in more detail later. Page 3 of 18

4 Temperature Compensation Conductivity is a parameter that is heavily dependent on temperature. A 0.01 molar potassium chloride solution is presented here as an example. The conductivity of this solution at 20 C is 1278 µs/cm whereas, at 25 C, it is 1413 µs/cm. As a result, values of the same sample that were measured at different temperatures cannot be compared with one another in practice. For this reason, the reference temperature was introduced. It is usually 20 C or 25 C. Conductivity measurements are normally conducted as follows. The measuring instrument records the actual conductivity and temperature, converts it to the reference temperature using a temperature compensation function and displays the conductivity at the reference temperature. Conductivity measurements and temperature measurement are intrinsically related and, thus, modern WTW conductivity measuring cells have temperature probes integrated in them. Depending on the type of sample used, different temperature compensation functions must be used: Linear function Non-linear function (nlf) for natural waters according to EN (DIN ) Non-linear function for purest water (additional consideration of the intrinsic conductivity of the water) Non-linear functions for special solutions No compensation The following example illustrates the measuring principle of conductivity measurements for a sample (e.g. waste water) using nlf temperature compensation. C onductivity nlf T ref =20 C T ref =25 C Actual Conductivity Temperature Linear functions are used, e.g. for saline solutions, acids and leaching solutions. However, the linear dependency is not suitable for many aqueous liquids under test. The temperature dependency can only be described by non-linear functions such as the nonlinear function for natural waters, i.e. for ground water, surface water, drinking water and waste water. High performance conductometers from WTW enable an automatic determination of non-linear functions for special samples by the user. To do this, the Page 4 of 18

5 sample to be tested is heated or cooled and the conductometer records the conductivity in dependency on the temperature and performs the corresponding curve adaptation. Selected values are listed in the following table as an example of typical conductivity: Conductivity at 25 C Purest water µs/cm De-ionized water 1 µs/cm Rainwater 50 µs/cm Drinking water 500 µs/cm Industrial wastewater 5 ms/cm Seawater 50 ms/cm 1 mol/l NaCl 85 ms/cm 1 mol/l HCl 332 ms/cm Even the purest water has a conductivity! This has its origin in the intrinsic dissociation of water that, according to the solubility product, forms oxonium and hydronium ions. H 2 O H 3 O + OH - H 2 O Page 5 of 18

6 Calibration and analytical quality assurance Calibration First a preliminary remark. The terms calibration and adjustment are often used synonymously. Strictly speaking, calibration describes a comparison with set points. The determined values must lie within specified tolerances. Adjustment means an active change of settings. As opposed to ph measurement and oxygen measurement, the measuring cell does not change and is not depleted by the measurement procedure in conductivity measurements. At least, this is the case if the cell is used correctly. The measuring cell is made of stainless steel, platinum, platinum-plated platinum or graphite electrode pairs that are chemically resistant and whose geometry is determined by the cell constant. The measuring cells are delivered from the factory with a specified and tested cell constant and the user can start measuring without any need for determine the cell constant. The production-dependent tolerances of the cell constant lie between 1.5% and 2% depending on the measuring range. If more precise measurements are required, the cell constant can be precisely adjusted using potassium chloride solutions. This basically applies to new cells or when a cell is changed. To do this, the use of commercially available KCl solutions is recommended as there is no fear of error as a result of dilution or contaminated substances (KCl or water). Calibration of the conductivity cell is to be understood in the sense of checking - and not in the sense of adjustment - as any change of the cell constant is usually caused by contamination. It is not logical to adjust the cell to the current contamination. It is better to clean the cell. Page 6 of 18

7 The EN recommends regular testing at least every six months. The standard calibration solution that corresponds to the cell constant must be used to do this. The solutions to be used are potassium chloride solutions of different concentrations. Measuring range Electrode material Recommended standard solution c (KCl) [mol/l] 200 µs/cm Steel, platinum µs/cm Plated platinum, graphite 2 ms/cm Plated platinum, platinum, graphite Electrical conductivity at 25 C [ms/cm] 300 ms/cm Platinum, graphite 3 approx µs/cm Platinum, graphite For the measuring range up to 2 µs/cm, (only the purest water such as boiler feed water or de-ionized water) can no contamination be expected. Standard solutions for this conductivity measuring range would be very unstable. It would also be problematic to use this type of standard calibration solution because contact with air must be completely excluded. Carbon dioxide found in the air dissolves in water to form hydrogen carbonate or carbonate ions. These ionic components cause dramatic changes to the measured conductivity and, consequently, produce meaningless results. CO 2 H 2 O CO 2 H 2 CO 3 H + HCO 3 - H + CO 3 2- Page 7 of 18

8 Monitoring the effect of carbon dioxide also applies to calibration as well as to the actual measurement. It must be performed with air excluded or using a protective gas. To check a purest water measurement cell, a simple functional test is the most useful. If the cell provides plausible values for de-ionized water, interference can be essentially ruled out. Checking cells for a conductivity range smaller than 200µS/cm A differentiation must be made according to the electrode material: Polished platinum or steel electrodes polarize at high conductivities so that a mol/l KCl solution must be used. A molar KCl solution can be produced by 1+9 dilution of commercially available 0.01 mol/l KCl solution (pay attention to the quality of the water!). Commercial 0.01 mol/l KCl solution available from WTW is recommended for platinumplated platinum or graphite electrodes because, as a result, any sources of error in the production can be excluded. For the measuring range up to 2 ms/cm The use of commercially available 0.01 mol/l KCl solution is also possible and additional sources of error in the production of the test solution can be excluded. In the measuring range up to 300 ms/cm Only a simple check can be made here where the 3 molar, silver chloride-free KCl solution for the ph measurement is used. The result should lie within the order of magnitude of 300 ms/cm. Page 8 of 18

9 Adjusting and checking the cell constant Conductivity measurements are extremely sensitive to ionic contamination. Therefore, the cleaned or new conductivity cell should be rinsed with standard solution before it is submersed in the standard solution. For laboratory measurement cells such as the TetraCon 325, it is recommended to submerse the sensor directly in the freshly opened calibration standard bottles to prevent contamination by the sample vessel. If this is not possible, the vessels to be used should be rinsed with standard solution. A special case is presented by stray field cells. The electrodes are attached to the front of the shaft (e.g. TetraCon 700 for online measurement). The cell constant is dependent on the size and type of the stray field. The stray field is changed in turn by limitations such as the base of the vessel or of the mounting tube. As a result, the adjustment or checking of the sensor in a mounting tube must be performed while it is installed. In other cases, a minimum gap to the vessel wall must be maintained. The measured value must be independent of the size of the vessel or the volume of standard solution. If the cell constant is adjusted to raise the measuring accuracy (not to compensate for contamination!), it is important to check whether the conductometer has an automatic calibration function as is the case in modern WTW conductometers. If it does, it is sufficient to submerse the cell in 0.01 mol/l KCl solution and to start the adjustment process. The background to this is that the temperature compensation function of the standard calibration solution is stored in the instrument. During this process, the instrument measures the conductivity and changes the calculated cell constant in such a way that the corresponding conductivity of the standard solution is reached at the selected reference temperature. After this is completed, the adjusted cell constant is displayed. In instruments without automatic calibration, this process is performed by the user. In this case, the user must ensure that adjustment takes place in the thermostatted state. The control solution and measuring cell must be at the reference temperature. The background to adjusting the cell constant is the same. The cell constant is changed so that, at the reference temperature, the corresponding conductivity is achieved as closely as possible. As closely as possible because the resolution of the cell constant is not often high enough to enable an adjustment to the exact value (e.g µs/cm for 0.01 molar KCl solution at 25 C). Automatic calibration function No automatic calibration function 0,01 n KCl 0,01 n KCl T = constant T = T Ref Page 9 of 18

10 Note: If the conductivity measuring cells continue to be submersed in the standard calibration solution after the adjustment is complete and the instrument is switched to the normal measuring mode using, e.g. an nlf temperature compensation function, this usually results in a different measured value. (Reason: A different temperature compensation function!) If a control measurement only is performed with the standard calibration solution and without adjustment of the cell constant, the correct temperature compensation function must be set. The value of the linear temperature compensation for the 0.01 molar KCl solution is 1.9%/K at a reference temperature of 25 C and 2.1% /K at a reference temperature of 20 C. Alternatively, the control solution can be thermostatted precisely to the reference temperature where adjustment of the corresponding temperature compensation function is no longer required. If the conductivity measuring cell is changed, the new cell constant must be transferred to the conductivity measuring instrument! Page 10 of 18

11 Testing the conductometer The primary process variable of the conductance is the resistance. The function of a conductometer can be tested using certified test resistances. To do this, WTW provides a set of six different resistances that are connected to the instrument instead of the measuring cell. If the measured values displayed lie within the tolerance specified in the certificate, the functional safety is met. If the displayed values lie outside the tolerance, the instrument must be sent for repair. The test set offered by WTW enables testing within the framework of the ISO It enables the restoring of the measured values to national standards. Page 11 of 18

12 Measurement and analytical quality assurance Which measuring cell is required for which application As in ph measurement, the selection of the conductivity cell used depends on the application. It is also true that the effort required is greater, the cleaner the sample to be inspected. A number of examples have been included to illustrate this but it also important to note that a great number of special cells are available: The standard TetraCon 325 conductivity measuring cells have graphite electrodes in four conductor technology and are, thus, insensitive to contamination and easy to clean; their application is in routine analysis where the measuring range lies between 1µS/cm and 2 S/cm. For partially aqueous and highly aggressive media, the use of the TetraCon 325 Pt with a measuring range between 1 µs/cm and 1 S/cm is recommended. The electrodes are made of platinum and are, thus, extremely resistant to aggressive media. Page 12 of 18

13 If a measurement is to be made in purest water such as boiler feed water or ion exchange water, it is necessary to use a flowthrough vessel. Contact with the air must be excluded. The reason for this is the carbon dioxide in the air that in turn forms carbonic acid in water and leads to a change in the conductivity. Two electrodes are available depending on the measuring range required. LR 325/01 (0.001 µs/cm 300 µs/cm) LR 325/001 ( µs/cm 30 µs/cm) Special tip and just to repeat: Two points are often overlooked in the measurement of conductivity: The result is usually referred to the reference temperature. Mostly, this is 25 C! If the reference temperature is set to 20 C in the instrument, a very different value is obtained that has nothing to do with incorrect measurement, but simply has a different reference. To calculate the conductivity at the reference temperature, the instrument converts from the current temperature of the material under test using a characteristic function. If this function does not correspond to the sample, an incorrect value is obtained. If, for example, a saline solution is to be tested, a linear function is required and not the non-linear function for natural waters. These two points should be taken into consideration in every measurement, set correctly and also documented according to the AQA. Page 13 of 18

14 Drift control Drift control, as in the ph and oxygen parameters, makes no sense in the case of conductivity measurement as the displayed measured value is the current measured value. No time delay results from the measurement technique. Cleaning the cells Warm water with some household detergent has proved itself as a cleaning agent for organic contamination. Alcohol may also be used. Calcium deposits are best cleaned using ten-percent citric acid. It is possible to clean graphite, steel or polished platinum electrodes mechanically using a soft brush. However, make sure that the surface of the electrodes is not scratched! Never use hard objects such as screwdrivers, etc. Even the use of a brush should be carefully considered. Platinum-plated platinum electrodes should only be cleaned chemically because even the softest brush would damage the platinum coating of the electrodes. It is possible to regenerate a damaged or heavily contaminated electrode surface by platinum plating the electrode surfaces. Information on how to do this is given in WTW application report 243. Storing the measuring cells In the same way as there are different methods of cleaning the measuring cells, there are various points that must be observed when storing the cells. Cells made of polished platinum, steel or graphite must be stored in a dry place. Platinum-plated platinum electrodes, however, must not be allowed to dry out. They must be stored in de-ionized water. Page 14 of 18

15 Practical exercises Preparation All practical experiments should be carried out in a suitable laboratory to guarantee working safety. While this is a general recommendation, however, in the trials of conductivity measurement described below no hazards arise from the chemicals used as KCl solution is non-toxic and can only leave a washable salt ring. Nevertheless, in the analytical field, the safety regulations should be routinely observed. Safety instructions General rules of conduct when handling chemical substances When working at a work place at which chemicals are handled, the following rules must be observed:! Follow the instructions on the chemical bottles! Always wear protective garments (goggles, gloves,...)! Never point open containers towards other persons! Do not eat, drink or smoke! Ensure the satisfactory disposal of chemicals! Carefully remove or clean up any spilled chemicals! Contact specialist personnel if any serious problems arise We provide these short instructions in the hope that they lead to successful and safe practical studies. Page 15 of 18

16 The following instruments and equipment must be available: Washable tables Resistant floor coverings Running water Eye bath Measuring station checklist 1 conductivity measuring instrument e.g. LF 330/340, inolab Cond 1 conductivity measuring cell Suitable for the measuring instrument 1 calibration standard 0.01 mol/l KCl solution 1 stand To hold the electrode 1 magnetic stirrer with agitator 2 beakers (150ml) Fresh distilled water Cleaning bottle with distilled water Protective goggles, gloves and apron Ballpoint pen and pad, calculator, lintfree cloths Standard calibration solutions must be beyond any doubt!! That is to say: Do not use any buffer solutions of unknown origin or history! Ideally, use freshly opened commercial standard calibration solutions that will be disposed of afterwards. Page 16 of 18

17 Practical exercises 1. Calibrating the measuring cell Rinse the clean sensor with standard calibration solution (0.01 mol/l KCl) and then submerse it in the solution Check the selected cell constant Set the reference temperature to 25 C and linear temperature compensation to 2.1%/K. Compare the displayed value with the nominal value of 1413mS/cm. Manual adjustment of the cell constant Change the cell constant while monitoring the change in the conductivity. (Caution: Here, thermostatting has not been used because the cell constant is automatically calibrated in the next step! If the cell constant can only be adjusted manually, it must be thermostatted to the reference temperature!) Automatic adjustment of the cell constant using the operating manual 2. Effect of the reference temperature Determine the conductivity of 0.01 mol/l KCl at a reference temperature of 20 C Linear temperature compensation with 1.9%/K Determine the conductivity of 0.01 mol/l KCl at a reference temperature of 25 C Linear temperature compensation with 2.1%/K 3. Effect of the adjusted temperature compensation function The values measured for tap water (reference temperature 25 C) should be determined using the nlf, a linear function (2.1 %/K) and no temperature compensation (0.0 %/K) and then compared. 4. Effect of carbon dioxide on the measurement of ion-deficient samples Freshly distilled water should be stirred using a magnetic agitator for 15 minutes. While doing so, the conductivity should be continuously monitored. (nlf function and 25 C reference temperature) Page 17 of 18

18 Bibliography [1] DIN EN 27888, Bestimmung der elektrischen Leitfähigkeit, 11/93 Page 18 of 18

Vernier Software & Technology

Vernier Software & Technology Vernier Software & Technology ORDER CODE U.S. EDUCATOR PRICE Product Page Download Manual PDF The Conductivity Probe can be used to measure either solution conductivity or total ion concentration of aqueous

More information

Instructions. accumet Conductivity Cells Instruction Manual THEORY

Instructions. accumet Conductivity Cells Instruction Manual THEORY Instructions THEORY Catalog Number accumet Conductivity Cells Instruction Manual Compatible with accumet Body Type Nominal Cell- - Constant cm Cell Configuration Ideal Range μs/cm 3-62-.. to 2, 3-62- XL.

More information

EC91346S Conductivity Cell Instruction Manual

EC91346S Conductivity Cell Instruction Manual EC91346S Conductivity Cell Instruction Manual Body Type Nominal Cell Constant Cell Configuration Housing Material Stainless Steel 0.3 4-Cell / Pt100 Ryton with 3/4 MNPT THEORY Conductivity - Conductivity

More information

User Guide. Conductivity Cell

User Guide. Conductivity Cell User Guide Conductivity Cell Introduction DuraProbe 4-Electrode Conductivity Cells Thermo Scientific Orion DuraProbe cells are durable and extremely accurate under a wide range of conditions. 1. DuraProbe

More information

Online Conductivity Measurement. Measuring Monitoring Controlling. On-line Conductivity Measurements. Municipal and Industrial Wastewater

Online Conductivity Measurement. Measuring Monitoring Controlling. On-line Conductivity Measurements. Municipal and Industrial Wastewater Online Conductivity Measurement Measuring Monitoring Controlling Conductivity is a well recognized, and sometimes indispensable, parameter of state-of-the-art water, wastewater and industrial process analysis.

More information

Standard Operating Procedure for the Determination of Specific Conductance CCAL 11A.2

Standard Operating Procedure for the Determination of Specific Conductance CCAL 11A.2 Standard Operating Procedure for the Determination of Specific Conductance CCAL 11A.2 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson Hall Corvallis,

More information

Module 2, Add on Lesson Conductivity Sensor. Student. 90 minutes

Module 2, Add on Lesson Conductivity Sensor. Student. 90 minutes Module 2, Add on Lesson Conductivity Sensor 90 minutes Student Purpose of this lesson Investigate the variation of resistance with conductivity Construct a sensor to measure the conductivity of water Graph

More information

Total Dissolved Solids

Total Dissolved Solids Total Dissolved Solids LabQuest 12 INTRODUCTION Solids are found in streams in two forms, suspended and dissolved. Suspended solids include silt, stirred-up bottom sediment, decaying plant matter, or sewage-treatment

More information

Module 2, Add on Lesson Conductivity Sensor. Teacher. 90 minutes

Module 2, Add on Lesson Conductivity Sensor. Teacher. 90 minutes Module 2, Add on Lesson Conductivity Sensor 90 minutes Teacher Purpose of this lesson Investigate the variation of resistance with conductivity Construct a sensor to measure the conductivity of water Graph

More information

Operating Manual for Digital ph-meter GPHR 1400

Operating Manual for Digital ph-meter GPHR 1400 Operating Manual for Digital ph-meter GPHR 1400 Specification Measuring range: Resolution: ph-electrode: conductivity Input resistance: Accuracy (unit only): Working temperature: Nominal temperature: Power

More information

Salinity Sensor PS Salinity Sensor Probe. Salinity Sensor Probe, 10 X, Conductivity/Temperature PASPORT Extension Cable

Salinity Sensor PS Salinity Sensor Probe. Salinity Sensor Probe, 10 X, Conductivity/Temperature PASPORT Extension Cable Instruction Sheet 012-10546A *012-10546* Salinity Sensor PS-2195 Six pin mini- DIN connector Salinity Sensor Temperature Input port Conductivity Input port Salinity Sensor Probe Temperature Input plug

More information

Advantages and Disadvantages of Conductivity Measurement

Advantages and Disadvantages of Conductivity Measurement Educat i onser i es Conduct i vi t ygui de What is Conductivity? Conductivity (or specifically electrolytic conductivity) is defined as the ability of a substance to conduct electric current. It is the

More information

Odyssey Conductivity & Temperature Logger.

Odyssey Conductivity & Temperature Logger. Odyssey Conductivity & Temperature Logger. Contents. Salinity Sensor & Software.... 2 Conductivity Cell Calibration.... 2 Standard Solutions.... 3 Temperature Coefficient... 5 Positioning the Cell....

More information

For products: ConeFET CupFET LanceFET LanceFET+H MicroFET MiniFET SI line ph PROBE MANUAL

For products: ConeFET CupFET LanceFET LanceFET+H MicroFET MiniFET SI line ph PROBE MANUAL For products: ConeFET 3280-010 CupFET 3200-010 LanceFET 2270-010 LanceFET+H 2274-010 MicroFET 9270-010 MiniFET 9202-010 SI line ph PROBE MANUAL Warning! There are no user replaceable parts in this probe.

More information

F U C H S L U B R I C A N T S C O.

F U C H S L U B R I C A N T S C O. F U C H S L U B R I C A N T S C O. SENSOR INFORMATION, CARE AND WARRANTY RM CONCENTRATION / BRIX SENSOR RANGE: 0-35 BRIX GENERAL RM CONCENTRATION SENSOR CARE AND MAINTENANCE RECOMMENDATIONS BY MANUFACTURER

More information

Evaluation copy. Total Dissolved Solids. Computer INTRODUCTION

Evaluation copy. Total Dissolved Solids. Computer INTRODUCTION Total Dissolved Solids Computer 12 INTRODUCTION Solids are found in streams in two forms, suspended and dissolved. Suspended solids include silt, stirred-up bottom sediment, decaying plant matter, or sewage-treatment

More information

Model 435A ph Monitor/Recorder Manual

Model 435A ph Monitor/Recorder Manual Model 435A ph Monitor/Recorder Manual A digital version of this manual is available online at www.krugerandeckels.com 2 Contents Description... 4 Operation... 4 Electrodes... 5 Electrode Temperature...

More information

Electrochemistry LEC Electrogravimetric determination of copper. What you need: What you can learn about. Principle and tasks

Electrochemistry LEC Electrogravimetric determination of copper. What you need: What you can learn about. Principle and tasks Electrochemistry LEC 06 What you can learn about Quantitative analysis Electrolysis Gravimetry Overpotential and electrode polarisation Principle and tasks Electrogravimetry is an important analytical

More information

Dissolved Oxygen Probe (Order Code: DO-BTA)

Dissolved Oxygen Probe (Order Code: DO-BTA) Dissolved Oxygen Probe (Order Code: DO-BTA) The Vernier Dissolved Oxygen Probe uses a Clark-type polarographic electrode to measure the concentration of dissolved oxygen in water. You can use this sensor

More information

PURE WATER ph MEASUREMENT IN LOW CONDUCTIVITY SAMPLES

PURE WATER ph MEASUREMENT IN LOW CONDUCTIVITY SAMPLES PURE WATER ph MEASUREMENT IN LOW CONDUCTIVITY SAMPLES Normally, ph measurements are made in water with 1,000 to 10,000 µs/cm conductivity. The conductivity comes from dissolved solids, typically, salts

More information

HYDROLOGY INVESTIGATION Electrical Conductivity. What is Conductivity?

HYDROLOGY INVESTIGATION Electrical Conductivity. What is Conductivity? What is Conductivity? Conductivity of a water sample is a measure of its ability to carry an electric current. The more impurities (total dissolved solids) in water, the greater its electrical conductivity.

More information

VAN LONDON COMPANY COMBINATION ph ELECTRODE INSTRUCTION MANUAL

VAN LONDON COMPANY COMBINATION ph ELECTRODE INSTRUCTION MANUAL VAN LONDON COMPANY COMBINATION ph ELECTRODE INSTRUCTION MANUAL GENERAL INFORMATION Van London combination electrodes offer the convenience of having the reference and measuring electrodes combined in a

More information

FIOLAX. Solutions. Buffer solutions in the unique double-end ampoules offer a particularly high degree of reliability and measuring accuracy.

FIOLAX. Solutions. Buffer solutions in the unique double-end ampoules offer a particularly high degree of reliability and measuring accuracy. Solutions Buffer solutions in the unique double-end ampoules offer a particularly high degree of reliability and measuring accuracy. The exactness of the ph measurement is mainly dependent on the accuracy

More information

Oyster ph - Conductivity - TDS - Temperature Meters

Oyster ph - Conductivity - TDS - Temperature Meters User's Guide Oyster ph - Conductivity - TDS - Temperature Meters Models 34135-A 341350 341350-P 34145A 341450 341450-P INTRODUCTION Congratulations on your purchase of the Extech's Oyster Series meter

More information

Potentiometric titration

Potentiometric titration Potentiometric titration TEC Related concept Potentiometric titration, redox reaction, Nernst equation, quantitative analysis. Principle In a potentiometric titration, the equivalence point is detected

More information

Cole Parmer Conductivity Meter

Cole Parmer Conductivity Meter Cole Parmer 01481-61 Conductivity Meter Printed in U.S.A. 102001-R3 Serial Number: Table of Contents Page 1. Introduction.. 1 2. Shipping Checklist.. 1 3. Specifications... 2 4. Display.. 2 5. Power Requirements..

More information

Introduction The Van London Co. Cadmium Ion Electrode is used to measure Cadmium ions in aqueous solutions.

Introduction The Van London Co. Cadmium Ion Electrode is used to measure Cadmium ions in aqueous solutions. GENERAL INSTRUCTIONS Introduction The Van London Co. Cadmium Ion Electrode is used to measure Cadmium ions in aqueous solutions. Required Equipment 1. An ion meter 2. Cadmium Ion Sensing Electrode 3. Polishing

More information

Electricity and Chemistry

Electricity and Chemistry Electricity and Chemistry Electrochemistry: It is a branch of chemistry that deals with the reactions involving the conversion of chemical energy into electrical energy and vice-versa. Electrochemical

More information

C300 Cond/TDS/Sal/T Meter. Manual

C300 Cond/TDS/Sal/T Meter. Manual C300 Cond/TDS/Sal/T Meter Manual Table of Contents 1. Brief Introduction ------------------------------------------------------------------------------------ 2 2. Technical Parameters---------------------------------------------------------------------------------

More information

Instruction Manual HI Portable ph/ec/tds/ C Meters

Instruction Manual HI Portable ph/ec/tds/ C Meters Instruction Manual HI 9811-5 Portable ph/ec/tds/ C Meters www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the meter.

More information

INTRODUCTION TO ELECTROCHEMISTRY: CURRENT, VOLTAGE, & BATTERIES. Introduction. Electrochemistry Revised 4/28/14

INTRODUCTION TO ELECTROCHEMISTRY: CURRENT, VOLTAGE, & BATTERIES. Introduction. Electrochemistry Revised 4/28/14 INTRODUCTION TO ELECTROCHEMISTRY: CURRENT, VOLTAGE, & BATTERIES Introduction Electrochemical Cells In this part of the experiment, four half cells are created by immersing metal strips of zinc, copper,

More information

Model DDS-11C Conductivity Meter Operation Manual Shanghai San-Xin Instrumentation,Inc.

Model DDS-11C Conductivity Meter Operation Manual Shanghai San-Xin Instrumentation,Inc. Model DDS-11C Conductivity Meter Operation Manual Shanghai San-Xin Instrumentation,Inc. Content 1. Summary --------------------------------------------------------------------------- 2 2. Technical Specifications

More information

Direct ISE Method Method to 10.0 mg/l F Fluoride ISE

Direct ISE Method Method to 10.0 mg/l F Fluoride ISE Fluoride DOC316.53.01237 Direct ISE Method Method 8323 0.1 to 10.0 mg/l F Fluoride ISE Scope and application: For drinking water 1 and wastewater 2. 1 The equipment that is specified in this method meets

More information

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By the Authority Vested By Part 5 of the United States Code 552(a) and Part 1 of the Code of Regulations 51 the attached document has

More information

Ion Concentration Controller General Information

Ion Concentration Controller General Information Document 40501, p. 1 of 8 Ion Concentration Controller General Information Common Questions Asked about Ion-Selective Electrodes 1. How often do you need to calibrate your ion-selective electrode (ISE)?

More information

Deruster 11 J Product Code: Revised Date: 01/26/2009. Deruster 11 J

Deruster 11 J Product Code: Revised Date: 01/26/2009. Deruster 11 J Deruster 11 J DESCRIPTION Deruster 11 J is a powdered, cyanide-free, highly alkaline water soluble product used to remove rust, heat scale, weld scale, smuts, certain fabrication oils, and to strip paint

More information

SUPER FRESHNESS METER KV-202. Instruction Manual

SUPER FRESHNESS METER KV-202. Instruction Manual 041109. SUPER FRESHNESS METER KV-202 Instruction Manual Rapid measurement of K value, the decisive factor in freshness, is now possible. Reliable instrumentation is also available, eliminating virtually

More information

Inductive Probe 8398 Instruction manual

Inductive Probe 8398 Instruction manual Inductive Probe 8398 Instruction manual 12/25 Hach Ultra Analytics 621=183=98 Issue D Table of contents 1. INTRODUCTION... 3 1.1 General...3 1.2 Probes...3 2. OPERATING PRINCIPLE... 4 2.1 Conductivity

More information

Measuring ORP : Tips, Cautions and Limitations

Measuring ORP : Tips, Cautions and Limitations Measuring ORP : Tips, Cautions and Limitations Introduction and Basic Theory As described in Standard Methods for the Examination of Water and Wastewater (Section 2580 B.), ORP is a potentiometric measurement

More information

(aq) + 5e - Mn 2+ (aq) + 4H 2

(aq) + 5e - Mn 2+ (aq) + 4H 2 EXPERIMENT 20 Titrimetric Determination of Iron INTRODUCTION Potassium permanganate is widely used as an oxidizing agent in titrimetric analysis. In acidic solution, a permanganate ion undergoes reduction

More information

Lenntech. sension + Meters & Probes. Electrochemistry. Tel Fax.

Lenntech. sension + Meters & Probes. Electrochemistry. Tel Fax. DATA SHEET sension + Meters & Probes Lenntech info@lenntech.com Tel. +31-152-610-900 www.lenntech.com Fax. +31-152-616-289 Electrochemistry Hach sension+, an all-in-one system with guided menu navigation

More information

WATER QUALITY CHECKER

WATER QUALITY CHECKER WATER QUALITY CHECKER U-10 ph. COND. TURB. D Simple, Handy Instrument Contains Powerful Sensor for S The Horiba U-10 water quality checker brings laboratory accuracy and the ease of push-button operation

More information

POLYMETRON Model 8398 Inductive probe

POLYMETRON Model 8398 Inductive probe Document Number 221=183=98 POLYMETRON Model 8398 Inductive probe USER MANUAL April 21, Version D Restriction of hazardous substances (RoHS) The European Union RoHS Directive and subsequent regulations

More information

Solutions. Standard buffer solutions according to DIN Hot steam sterilized for longer stability, no preservation agents used.

Solutions. Standard buffer solutions according to DIN Hot steam sterilized for longer stability, no preservation agents used. Solutions Standard buffer solutions according to DIN 19 266 Hot steam sterilized for longer stability, no preservation agents used. Order No. Type No. ph value at 25 C Contents 285137977 L 4791 1.68 60

More information

RANGE DIGISENS PHEHT : PH, REDOX & TEMPERATURE

RANGE DIGISENS PHEHT : PH, REDOX & TEMPERATURE RANGE DIGISENS PHEHT : PH, REDOX & TEMPERATURE Digital Technology for optimized measures Combination ph/redox/temp sensor Digital Sensor : Modbus RS 485 / SDI-12 Calibration data inside ph/orp Cartridge

More information

200 Series Benchtop ph/cond. Meter. Instruction Manual

200 Series Benchtop ph/cond. Meter. Instruction Manual 200 Series Benchtop ph/cond. Meter Instruction Manual P200-01 Benchtop ph Meter C200-01 Benchtop Cond. Meter PC200-01 Benchtop ph/cond. Meter P200-02 Benchtop ph Meter C200-02 Benchtop Cond. Meter PC200-02

More information

PORTABLE ph METER PRICE CODE NUMBER MODEL 100

PORTABLE ph METER PRICE CODE NUMBER MODEL 100 PORTABLE METER MODEL 100 PRICE CODE NUMBER 56-0032 Refer to Bulletin A-301. OPERATION AND SERVICE GUIDE O-670A DEC. 1997 GENERAL INFORMATION Model 100 meter is designed for measuring over the entire 0

More information

ppm Dissolved Oxygen Measurement

ppm Dissolved Oxygen Measurement ppm Dissolved Oxygen Measurement INTRODUCTION Dissolved oxygen (D.O.) levels are used as a general indicator of water quality. Oxygen is essential to life and vital for countless aquatic forms. D.O. level

More information

An Oxidation-Reduction Titration: The Reaction of Fe 2+ and Ce 4+

An Oxidation-Reduction Titration: The Reaction of Fe 2+ and Ce 4+ An Oxidation-Reduction Titration: The Reaction of Fe 2+ and Ce 4+ LAB ADV COMP 8 From Advanced Chemistry with Vernier, Vernier Software & Technology, 2004 INTRODUCTION A titration, as you recall, is a

More information

Typical Experiments. How it Works

Typical Experiments. How it Works ph Sensor DT016A The ph sensor can be connected to the Nova5000, MultiLogPRO or TriLink data loggers. The ph sensor is capable of measuring the entire range of 0-14 ph and is used for various experiments

More information

Drinking Water Monitoring. Ready-to-use systems for monitoring of quality relevant parameters in

Drinking Water Monitoring. Ready-to-use systems for monitoring of quality relevant parameters in Drinking Water Monitoring Ready-to-use systems for monitoring of quality relevant parameters in drinking water Single-parameter - Systems Single-parameter Systems Single-parameter systems for measurement

More information

FilterSorb SP3 Treatment Method: Part II

FilterSorb SP3 Treatment Method: Part II FilterSorb SP3 Treatment Method: Part II By Deepak Chopra WatchvWater GmbH Fahrlachstraße 14 68165 Mannheim Germany May 2013 Web: www.watchwater.de email: info@watchwater.de Telefon: +49 (0) 621 87951-0

More information

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment FAST SHIPPING AND DELIVERY TENS OF THOUSANDS OF IN-STOCK ITEMS EQUIPMENT DEMOS HUNDREDS OF MANUFACTURERS SUPPORTED

More information

TYPES OF CHEMICAL REACTIONS PART I INTRODUCTION

TYPES OF CHEMICAL REACTIONS PART I INTRODUCTION EXPERIMENT 10 (2 Weeks) Chemistry 100 Laboratory TYPES OF CHEMICAL REACTIONS PART I INTRODUCTION It is useful to classify reactions into different types, because products of reactions can be predicted.

More information

RANGE DIGISENS PHEHT:PH, REDOX & TEMPERATURE

RANGE DIGISENS PHEHT:PH, REDOX & TEMPERATURE RANGE DIGISENS PHEHT:PH, REDOX & TEMPERATURE Digital Technology for optimized measures Combination ph/redox/temp sensor Digital Sensor : Modbus RS 485 / SDI-12 Calibration data inside ph/orp Cartridge

More information

Revised Molar Mass Measurement Lab Purpose: Background:

Revised Molar Mass Measurement Lab Purpose: Background: Revised Molar Mass Measurement Lab (Based on prior Ch b experiment, incorporating revision suggestions from Tsze Tsang) Purpose: To use the colligative properties of cyclohexane solutions to determine

More information

Turbidity and Suspended Solids

Turbidity and Suspended Solids TURB/TSS and Turbidity and Turbidity Smell, taste and turbidity are the most important indicators for the quality of potable water. Effluent measurement of sewage treatment plants Total suspended solids

More information

Alchemy: A Cross-Curricular Activity Copper, Silver, and Gold Redox Reactions

Alchemy: A Cross-Curricular Activity Copper, Silver, and Gold Redox Reactions Alchemy: A Cross-Curricular Activity Copper, Silver, and Gold Redox Reactions SCIENTIFIC Introduction Turn an ordinary copper penny into silver and then into gold! Get rich quick by demonstrating this

More information

When this process occurs in water, the oxygen consumed is dissolved oxygen.

When this process occurs in water, the oxygen consumed is dissolved oxygen. Theory: BOD is Biochemical Oxygen Demand Referred to as a standardized laboratory procedure used to determine the relative oxygen requirements of wastewaters, effluents and polluted water. It is a measures

More information

Salinity in Seawater

Salinity in Seawater Salinity in Seawater Objective To familiarize students with the different methods used for measuring salinity of water. Introduction: Salinity exerts profound impacts on the marine environment. It controls

More information

CHAPTER 6: INSTRUMENTATION AND CONTROL OF A BIOREACTOR

CHAPTER 6: INSTRUMENTATION AND CONTROL OF A BIOREACTOR CHAPTER 6: INSTRUMENTATION AND CONTROL OF A BIOREACTOR Bioreactor Control Temperature Dissolved oxygen ph Foam Agitator speed Nutrient addition rate Level control Bioreactor Instrumentations Temperature

More information

Approved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric)

Approved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric) METHOD #: 370.1 TITLE: Approved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric) ANALYTE: Silica, SiO 2 INSTRUMENTATION: Spectrophotometer STORET No. Dissolved 00955 1.0 Scope and Application

More information

Lab-in-your-pocket. A Rainbow of Choices. Waterproof Pocket Water Quality Meter COND TDS 2 YEAR. Actual Size MADE IN JAPAN W A R R A N T Y

Lab-in-your-pocket. A Rainbow of Choices. Waterproof Pocket Water Quality Meter COND TDS 2 YEAR. Actual Size MADE IN JAPAN W A R R A N T Y Lab-in-your-pocket. A Rainbow of Choices. Actual Size COND TDS Waterproof Pocket Water Quality Meter 2 YEAR WARRANTY W A R R A N T Y MADE IN JAPAN Compact Water Quality Meter Get accurate direct measurement

More information

Guide to Optimal Steam Generation

Guide to Optimal Steam Generation Central Sterile Services HOW TO: Guide to Optimal Steam Generation Introduction Successful steam sterilization is much more than just achieving verification of sterility with biological indicators. Most

More information

Turbidity and Suspended Solids. Turbidity. Turbidity and Suspended Solids. Effluent measurement of sewage treatment plants. Sludge concentration

Turbidity and Suspended Solids. Turbidity. Turbidity and Suspended Solids. Effluent measurement of sewage treatment plants. Sludge concentration Turbidity and Turbidity Smell, taste and turbidity are the most important indicators for the quality of potable water. At the effluent of a waste water treatment plant, turbidity is a quantitative measure

More information

Portable Multi-Parameter Water Quality Meter LMPWM-A1 Series

Portable Multi-Parameter Water Quality Meter LMPWM-A1 Series Portable MultiParameter Water Quality Meter LMPWMA1 Series www.labtron.com info@labtron.com Portable Multiparameter Water Quality Meter LMPWMA1 series Portable Multiparameter Water Quality Meter LMPWMA1

More information

On-Site Generation Of Disinfectants

On-Site Generation Of Disinfectants On-Site Generation Of Disinfectants Content 1 SUMMARY... 3 2 WHAT IS ON-SITE GENERATION?... 3 3 HOW DO ON-SITE GENERATORS WORK?... 3 4 WHAT ARE THE BENEFITS OF USING OSGS?... 5 4.1 IMPROVED OPERATOR SAFETY...

More information

Water Analysis Kit for Home Brewers

Water Analysis Kit for Home Brewers Brew your best batch EVERY time! TM Water Analysis Kit for Home Brewers Code 7188-01 PLUS Scan for direct access to LaMotte Brewlab web page. For further tips and information, go to instructions/tips tab.

More information

USER MANUAL ON PROFESSIONAL ELECTROCHEMICAL STATION EPS-30A

USER MANUAL ON PROFESSIONAL ELECTROCHEMICAL STATION EPS-30A 1 USER MANUAL ON PROFESSIONAL ELECTROCHEMICAL STATION EPS-30A PRODUCT INTRODUCTION: This electrochemical station is a professional, high performance plating/forming equipment designed to plate, form, strip,

More information

EXPERIMENT 7A. Chemical Separation by Filtration and Recrystallization INTRODUCTION

EXPERIMENT 7A. Chemical Separation by Filtration and Recrystallization INTRODUCTION EXPERIMENT 7A Chemical Separation by Filtration and Recrystallization INTRODUCTION The solubilities of solid substances in different kinds of liquid solvents vary widely. Substances that we call salts

More information

REDOX AND REFERENCE ELECTRODE USER MANUAL

REDOX AND REFERENCE ELECTRODE USER MANUAL REDOX AND REFERENCE ELECTRODE USER MANUAL 3 Redox and reference electrode manual Copyright 2012 Unisense A/S Version October 2012 REDOX AND REFERENCE ELECTRODE MANUAL UNISENSE A/S TABLE OF CONTENTS WARRANTY

More information

PHS-3D-01 Lab ph Meter. User s Guide. Shanghai San-Xin Instrumentation, Inc.

PHS-3D-01 Lab ph Meter. User s Guide. Shanghai San-Xin Instrumentation, Inc. PHS-3D-01 Lab ph Meter User s Guide Shanghai San-Xin Instrumentation, Inc. Table of Contents 1. Brief Introduction----------------------------------------------------------------------- 2. Technical Specifications

More information

P300 ph/mv/t Meter. Manual

P300 ph/mv/t Meter. Manual P300 ph/mv/t Meter Manual Table of Contents 1. Brief Introduction-------------------------------------------------------------------------------------- 2. Technical Parameters -------------------------------------------------------------------------------

More information

2155 Cation exchange column

2155 Cation exchange column ABB MEASUREMENT & ANALYTICS USER GUIDE IM/2155 REV. G 2155 Cation exchange column Measurement made easy 2155 cation exchange column For more information Further publications are available for free download

More information

A Cycle of Copper Reactions

A Cycle of Copper Reactions EXPERIMENT A Cycle of Copper Reactions PURPOSE To demonstrate a series of copper reactions: starting with copper metal, oxidizing the metal to put it into solution and then, form a copper hydroxide, an

More information

Respirometric BOD 5 determination of waste water polluted with organic or inorganic toxins or inhibitors. OxiTop pressure measuring heads

Respirometric BOD 5 determination of waste water polluted with organic or inorganic toxins or inhibitors. OxiTop pressure measuring heads Application report Respirometric BOD 5 determination of waste water polluted with organic or inorganic toxins or inhibitors Foreword By reason of their origin, some wastewater can be polluted with both

More information

TN100 PortableTurbidity Meter. Instruction Manual.

TN100 PortableTurbidity Meter. Instruction Manual. TN100 PortableTurbidity Meter Instruction Manual. Table of Contents 1. Overview -------------------------------------------------------------------------------------- 1 2. Technical Specifications --------------------------------------------------------------

More information

Lab-in-your-pocket. A Rainbow of Choices. Waterproof Pocket Water Quality Meter COND TDS 2 YEAR MADE IN JAPAN W A R R A N T Y

Lab-in-your-pocket. A Rainbow of Choices. Waterproof Pocket Water Quality Meter COND TDS 2 YEAR MADE IN JAPAN W A R R A N T Y Lab-in-your-pocket. A Rainbow of Choices. COND TDS Waterproof Pocket Water Quality Meter WARRANTY 2 YEAR W A R R A N T Y MADE IN JAPAN Compact Water Quality Meter Get accurate direct measurement from a

More information

Instruction Manual. HI HI Portable ph/ec/tds/ C Meters

Instruction Manual. HI HI Portable ph/ec/tds/ C Meters Instruction Manual HI 9812-0 HI 9812-5 Portable ph/ec/tds/ C Meters w w w. h a n n a i n s t. c o m Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully

More information

REDOX AND REFERENCE ELECTRODE USER MANUAL

REDOX AND REFERENCE ELECTRODE USER MANUAL REDOX AND REFERENCE ELECTRODE USER MANUAL 1 Redox and reference electrode manual Copyright 2017 Unisense A/S Version June 2017 REDOX AND REFERENCE ELECTRODE MANUAL UNISENSE A/S TABLE OF CONTENTS WARRANTY

More information

442 ph/conductivity Meter

442 ph/conductivity Meter 442 ph/conductivity Meter Instruction Manual This manual contains complete instructions for setting up and using the 442 ph/conductivity meter. Applications information is also available. The information

More information

OPERATING MANUAL. ba77043e01 03/2014 LF413T IDS CONDUCTIVITY CELL

OPERATING MANUAL. ba77043e01 03/2014 LF413T IDS CONDUCTIVITY CELL OPERATING MANUAL ba77043e01 03/2014 CONDUCTIVITY CELL Copyright 2014, SI Analytics GmbH Reprinting - even as excerpts - is only allowed with the explicit written authorization of SI Analytics GmbH, Mainz.

More information

Demonstration of osmosis

Demonstration of osmosis Demonstration of osmosis We will try to carry out a classical experiment on demonstration of osmosis. The principle is shown in the figure. Water moves from the solution of lower osmolality, across the

More information

HACH TU 5 SERIES TURBIDIMETERS SOP: COMPARISON LAB AND PROCESS. The next standard in the evolution of turbidity

HACH TU 5 SERIES TURBIDIMETERS SOP: COMPARISON LAB AND PROCESS. The next standard in the evolution of turbidity HACH TU 5 SERIES TURBIDIMETERS SOP: COMPARISON LAB AND PROCESS The next standard in the evolution of turbidity COMPARE TU5 LAB AND PROCESS GUIDE & TROUBLESHOOT To guarantee that lab and process readings

More information

Elcometer 135C. Bresle Test Patch NEW. Determines the level of soluble salts on uncoated surfaces using the patch method in accordance with ISO

Elcometer 135C. Bresle Test Patch NEW. Determines the level of soluble salts on uncoated surfaces using the patch method in accordance with ISO NEW Elcometer 135C Key Features Bresle Test Patch Determines the level of soluble salts on uncoated surfaces using the patch method in accordance with ISO 8502-6 Easy to remove after test, leaving no foam

More information

HI GPS Multiparameter Meter. With great products, come great results HI 9829

HI GPS Multiparameter Meter. With great products, come great results HI 9829 HI 9829 HI 9829 GPS Multiparameter Meter ph/orp/ise, EC/TDS/Resistivity/Salinity/Seawater Specific Gravity, Turbidity, DO, Temperature and Atmospheric Pressure With great products, come great results an

More information

THE BRESLE METHOD. Presence of rust and mill scale; Presence of surface contaminants, such as salts, dust, oil and grease; Surface profile

THE BRESLE METHOD. Presence of rust and mill scale; Presence of surface contaminants, such as salts, dust, oil and grease; Surface profile THE BRESLE METHOD INTRODUCTION The performance of protective coatings applied to steel can be significantly affected by the condition of the steel surface immediately prior to painting. The key factors

More information

Experiment 3: Determination of an Empirical Formula

Experiment 3: Determination of an Empirical Formula Background Information The composition of a compound is defined by its chemical formula, which gives the number ratio of the different elements in the compound. For example, water has a fixed composition

More information

Evaluation copy. Chloride and Salinity. Computer INTRODUCTION

Evaluation copy. Chloride and Salinity. Computer INTRODUCTION Chloride and Computer 15 INTRODUCTION Chloride Chloride, in the form of the Cl ion, is one of the major inorganic anions, or negative ions, in saltwater and freshwater. It originates from the dissociation

More information

Rev Experiment 10

Rev Experiment 10 Experiment 10 SPECTROPHOTOMETRIC DETERMINATION OF IRON IN DRINKING WATER 2 lab periods Reading: 1) Chapter 17, pg 393-403, Quantitative Chemical Analysis, 8 h Edition, Daniel C. Harris (7 th Edition: Chapter

More information

1 analogue output. 4Programmable for measurement. 4 The limits of the outputs are freely programmable within the measuring range.

1 analogue output. 4Programmable for measurement. 4 The limits of the outputs are freely programmable within the measuring range. 35 series The 35 series developed by Chemitec is dedicated to the water treatment and industry sectors, and enables the following parameters to be measured: Process control device 4pH/REDOX 4Dissolved

More information

EXPERIMENT. The Reaction of Magnesium with Hydrochloric Acid; The Molar Volume of Hydrogen

EXPERIMENT. The Reaction of Magnesium with Hydrochloric Acid; The Molar Volume of Hydrogen EXPERIMENT The Reaction of Magnesium with Hydrochloric Acid; The Molar Volume of Hydrogen PURPOSE In this experiment you will determine the volume of the hydrogen gas which is produced when a sample of

More information

Process Electrodes and Probes

Process Electrodes and Probes Process Electrodes and Probes Table of Contents Page Introduction T2.2 Flat Tip Electrodes T2.8 AmpHel ph and ORP Electrodes T2.12 HI 1000 and HI 2000 Series T2.14 Easy ph and ORP Electrodes T2.17 Conductivity

More information

Technical Information Memosens CPS31D and Ceratex CPS31

Technical Information Memosens CPS31D and Ceratex CPS31 TI00030C/07/EN/13.13 71201979 Products Solutions Services Technical Information Memosens CPS31D and Ceratex CPS31 ph electrodes, analog and digital with Memosens technology For drinking water and swimming

More information

Uranium Testing and Site Calibration for Water Testing Detection Range: 2 60ppb

Uranium Testing and Site Calibration for Water Testing Detection Range: 2 60ppb Document: AND Uranium 100 10 2012 Uranium Testing and Site Calibration for Water Testing Detection Range: 2 60ppb October, 2012 Edition 1 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table

More information

CONDUCTIVITY INSTRUMENT

CONDUCTIVITY INSTRUMENT CONDUCTIVITY INSTRUMENT µs C YSI Model 3100 Conductivity Salinity Temperature Instrument Operations Manual Table of Contents 1. Introduction...1 1.1 Capabilities... 1 1.2 Controls... 2 Front Panel...

More information

Direct ISE Method Method to 4.00 mg/l NO 3 N Nitrate ISE

Direct ISE Method Method to 4.00 mg/l NO 3 N Nitrate ISE Nitrate DOC316.53.01239 Direct ISE Method Method 8359 0.04 to 4.00 mg/l NO 3 N Nitrate ISE Scope and application: For drinking water. Test preparation Instrument-specific information This procedure is

More information

Collecting a Surface Water Sample

Collecting a Surface Water Sample FIELD PROCEDURE Collecting a Surface Water Sample CAUTION: Never carry or lift the pole above your head, as touching power lines could cause electrocution. Notes: Students (

More information

Depletion. Applications and Procedures

Depletion. Applications and Procedures BSB BOD DBO OECD ISO EN DIN Depletion Depletion with OxiTop Control OC 110 The investigation and monitoring of biological cleaning treatment processes are becoming more and more important with respect

More information

SECTION D.3 CHLOROPHYLL AND PHEOPHYTIN

SECTION D.3 CHLOROPHYLL AND PHEOPHYTIN SECTION D.3 CHLOROPHYLL AND PHEOPHYTIN 1.1 Scope and Application 1.1.1 This spectrophotometric method is used in the determination of chlorophyll a, b, and c, and pheophytin a. This method can be used

More information