Application Notes. MicroBeta INTRODUCTION.
|
|
- Michael Stone
- 6 years ago
- Views:
Transcription
1 Drug Discovery Research Clinical Screening MicroBeta Application Notes Colour quench correction in Scintillation Proximity Assays Using ParaLux Count Mode K.T. Hughes 1, J.C. Ireson 1, N.R.A. Jones 2 and P. Kivelä 2 ( 1 Amersham International plc., Forest Farm, Whitchurch, Cardiff. CF4 7YT, Wales UK, 2 PerkinElmer Life Sciences, P.O.Box 10, FIN-20101, Turku, Finland) INTRODUCTION The MicroBeta TriLux is a multi-detector instrument for liquid scintillation and luminescence counting. It can simultaneously count up to 12 samples at a time in microtitration plates or in tubes. There are three MicroBeta models; 16 and 32 shelf capacity instruments, and a version fitted with an external sample loading tray that can be part of a fully automatic robot system. Each detector in the MicroBeta comprises two photo-multiplier tubes (PMTs), one is positioned above the sample and one below. The PMTs are connected in coincidence circuitry, a robust and well established method for liquid scintillation counting. Scintillation Proximity Assay (SPA) represents an innovative approach to high throughput screening because it allows the rapid and sensitive assay of a wide variety of molecular interactions in a truly homogeneous system. SPA and MicroBeta counting systems can be readily automated. Much SPA involves the screening of natural products or colored compounds, and as a consequence of the non-separation method, this can result in many different and highly colored samples within one assay. For the purposes of high throughput screening, it is essential that an instrument can accurately assess the level of color quench (or counting efficiency) and correct the observed count rate (CPM) to the true activity (DPM). is bound to the bead, it is brought into close enough proximity to enable it to stimulate the scintillant to produce light. Unbound radioactivity is too distant from the scintillant and the energy released is dissipated before reaching the bead. In general, the beads sink towards the bottom of the microtitration plate well and are therefore much closer to the lower PMT in the MicroBeta. The magnitude of color quench is dependent on the path length of the light in the color and therefore the effect of color is much smaller in the lower PMT than in the upper one. In this sense, SPA represents a unique sample type for which a novel counting method has been developed. ParaLux Count Mode takes full advantage of the lower PMT in two ways. Firstly, the difference in count rates between coincidence and non-coincidence events in the lower PMT is used to determine the counting efficiency and secondly, considerably higher counting efficiency can be achieved by including non-coincident counts observed solely in the lower PMT in the overall count rate. All existing methods of quench correction ascertain counting efficiency by determining the position of the isotope spectrum. For ex- In SPA scintillant is incorporated into small fluomicrospheres. These microspheres or beads are designed in such a way as to bind specific molecules. If a radioactive molecule
2 No. of counts No. of counts No. of counts ample, the MicroBeta uses the Spectral Quench Parameter of the Isotope (SQP(I)) which is the multi-channel analyzer (MCA) channel number that equates to the mid point of the isotope spectrum. The shift in position of the isotope spectrum (in terms of MCA channel number or energy) is therefore a limiting factor. In the case of tritium, the movement of the isotope spectrum is quite small and therefore the SQP(I) has a limited sensitivity to changes in counting efficiency and also has a finite dynamic range. 50 SQP(I) 0 Fig. 1 The SQP(I) quench indicating parameter is based on the position of the isotope spectrum Fig. 2A Carbon-14 color quench series. SQP(I) provides a good indicator because there is dynamic movement of the spectra with change in counting efficiency Fig. 2B Tritium color quench series. Minimal shift in spectrum position results in the SQP(I) being relatively insensitive to changes in counting efficiency, especially with heavily color-quenched samples. PARALUX COUNT MODE ParaLux Count Mode represents a genuine improvement in counting efficiency and quench parameter. Twin MCA technology is applied to construct simultaneously two spectra of each sample (see figure 3.). MCA-1 is connected in conventional coincidence circuitry. This accepts all beta decay that is of sufficient energy to produce many photons that are in turn detected by both PMTs. MCA-2 stores the beta decay that is detected by the lower PMT only. Counts registered below channel 150 in MCA-2 are discarded because these include unwanted background counts that are a feature of all PMTs. It is possible to change this channel setting to optimize for efficiency or background. (See Operating Instructions.) ParaLux includes two different counting modes, high efficiency and low background. Both modes use a new quench parameter, the Asymmetric Quench Parameter of the Isotope (AQP(I)). This is calculated from the ratio of the counts in the two MCAs. Figure 4 shows that, for tritium, AQP(I) commands a much wider scale than spectrum position based quench parameters and is therefore more sensitive to changes in counting efficiency, especially with heavily quenched samples. In high efficiency mode, the contents of MCA-1 and MCA-2 are added together. With a default window setting, this results in an increase in counting efficiency of 10 % for uncolored samples and up to 300 % for colored samples. The count rate in low background mode is acquired from MCA-1 only and is therefore identical to conventional coincidence counting. In high throughput screening there may be a variety of different colours present in an assay and it is therefore necessary that a quench curve constructed from one color is suitable for quench correction of all other colors. Figure 5. shows that AQP(I) curves of various colors do overlap, and that therefore quench correction can be made with a single curve. 2
3 MCA-1, Coincidence PM Coincidence circuit PM MCA-2, Bottom PMT AQP(I) = MCA-1 / MCA-2 High Efficiency CPM = MCA-1 + MCA-2 Low Background CPM = MCA-1 Fig. 3 Principle of ParaLux count mode ParaLux Count Mode, AQP(I) H-3 Efficiency /% Coincidence, SQP(I) Fig.4 Comparison of tritium SPA color quench series counted with conventional coincidence circuitry and ParaLux high efficiency mode. Relative efficiency Yellow Red Orange Green Blue AQP(I) Fig.5 ParaLux low background count mode. Tritium SPA quench curves. AQP(I) provides overlapping color quench curves. This permits the use of a single quench series for DPM calculations. EXPERIMENTAL Two experiments were devized to compare ParaLux Count Mode with the traditional spectrum position, SQP(I), quench correction method. To simulate real assay conditions, two receptor binding assays using 3 H and 125 I labels were chosen. The activity in each was representative of count rates used in high throughput laboratories. The % quench figures quoted are based on coincidence counting. 3
4 [ 3 H] CHO M1 RECEPTOR ASSAY A quench curve was set up using a 3 H color quench kit (Amersham, TRKQ 7080) in a Wallac plate and allowed to settle overnight. The assay was twelve replicates of total binding wells from a 3 H receptor assay in the presence of various levels of quench. The assay consisted of 50 µl of buffer or tartrazine, 50 µl l-quinuclidinyl [phenyl-4-3 H] benzilate ([ 3 H]QNB, Amersham, TRK 604) (5.1 nm final concentration), 50 µl CHO cell membranes transfected with rat muscarinic M1 receptors (100 µg) and WGA- PVT SPA beads (Amersham) (2 mg) in 50 mm Tris-HCl buffer, ph 7.4, containing 0.1 % (w/ v) BSA and 5 mm MgCl 2. The plate was shaken for four hours and then left overnight for the beads to settle prior to counting. Table 1. Twelve replicates of total binding wells from [ 3 H] CHO M1 receptor assay 4
5 [ 125 I] ET-1 RECEPTOR ASSAY A quench curve was set up using an 125 I color quench kit (Amersham, RPAQ 4030) in a Wallac plate and allowed to settle overnight. The assay was twelve replicates of total binding wells from an 125 I receptor binding assay in the presence of various levels of quench. The assay consisted of 50 µl of buffer or tartrazine, 50 µl [ 125 I]endothelin-1 (Amersham, IM 223) (39 pm final concentration), 50 µl porcine lung membranes (5 µg), and WGA-PVT SPA beads (Amersham) (1 mg) in 50 mm Tris-HCl buffer, ph 7.4, containing 0.1 % (w/v) BSA, 10 mm MnCl 2, and 1mM EDTA. The plate was shaken for four hours and then left overnight for the beads to settle prior to counting. Table 2. Twelve replicates of total binding wells from [ 125 I] ET -1 receptor assay 5
6 RESULTS See tables 1 and 2 CONCLUSION Precision in QCCPM values are clearly improved in heavily colored samples in both Low Background and High Efficiency ParaLux Count mode. ParaLux Count Mode is an authentic advance in SPA counting. Compared to other techniques counting efficiency can be increased by several hundred percent. The new, ultrasensitive quench parameter, AQP(I), provides superior quench correction at considerably higher levels of color. Time-consuming and costly dilution and recounting of difficult samples can thereby be avoided. Improved quench correction precision could lead to overall shorter count times. 6
7 Counting Scintillation Proximity assay (SPA) on the Wallac MicroBeta TriLux Warning: For research only. Not recommended or intended for diagnosis of disease in humans or animals. Do not use internally or externally in humans or animals. Instructions for the safe handling and use of Scintillation Proximity Assay beads are provided in the safety warning and precautions section of the pack leaflet accompanying the product. 1. COUNTING WITHOUT COL- OUR QUENCH CORRECTION When performing SPAs that are not quenched due to the presence of colored compounds, the results are usually expressed as counts per minute (CPM) This is achieved by utilizing the CPM counting mode of the counter. The performance of photomultiplier tubes is never identical and so a calibration procedure is needed before counting unknown samples. This procedure is called Normalization and involves counting the same sample in each of the detectors. In this way the relative efficiency of each of the detectors can be determined. Once the normalization samples have been counted in the MicroBeta, all the relevant data are automatically stored for future use within a Counting Protocol. Any normalization protocol can be linked to any number of counting protocols. Notes Normalization data should be generated for different isotopes and plate types being used. Isotopic and optical crosstalk is insignificant for the radioisotopes employed in bead based assays ( 3 H, 125 I, 33 P and 35 S) performed in the recommended microtitration plates (part numbers and ), and therefore is required no correction) The observed count rate increases as the beads settle, therefore assays should be counted under stable conditions (either in suspension or settled). Yttrium silicate beads settle rapidly (within a few minutes), therefore these assays should be counted with the beads settled. PVT beads settle over a longer period of time (usually several hours). The rate of settling depends on the type of application and buffer components present. In some assays the beads may be prevented from settling by addition of, for example, glycerol, to adjust the density of the buffer. Assays employing 33 P and 35 S must be counted with the beads settled to maximize the signal and reduce the non-proximity effect; for assays employing PVT beads this can be achieved by settling for several hours or gentle centrifugation (typically ten minutes at 800 xg). ParaLux Count Mode normalization protocols must include background subtraction. The background sample should be a black plate that is counted before the plate containing the normalization samples. Instrument background will then be automatically subtracted from the unknown samples and therefore background subtraction should not be selected in the counting protocol. See SPA Spectra and Operating Instructions for window setting and protocol editing. 7
8 2. COUNTING WITH COLOR QUENCH CORRECTION SPA is an homogenous technique with no separation step. When performing assays with colored samples, the quenching effect of the color must be corrected for. This is achieved by utilizing the DPM counting mode of the instrument. This procedure is called Standardization and involves a series of samples containing different levels of color and a constant quantity of radioactivity. Simultaneously, the instrument carries out detector normalization. Once the standardization samples have been counted in the MicroBeta, all the relevant data are automatically stored for future use within a Counting Protocol. Any standardization protocol can be linked to any number of counting protocols. Notes Quench curves are set up using a series of coloured standards (for example, tartrazine) and the same quantity of radioactivity in each well of the curve. For assays involving the use of PVT beads and 3 H- or 125 I- labels, the use of the appropriate color quench kit from Amersham (TRKQ 7080 for 3 H and RPAQ 4030 for 125 I) is strongly recommended. Instruments equipped with ParaLux Count Mode (software version 4.3 or greater) will automatically select ParaLux when SPA and DPM are selected. The color quench curve should be set up with the same weight of beads, assay buffer and total volume as employed in the screening assay. A separate quench curve should be set up for each screen and on each instrument. A separate quench curve should be set up for each type of microtitration plate being used. It is essential to ensure that a smooth colour quench curve is obtained. A single outlying point may be edited out. Generally, the best precision of correction is obtained with the automatic smoothing spline curve fit. During programming of the counter, the counter will ask for the DPM value of the standards to be specified. In most screening assays the results are expressed as a relative measurement e.g. % inhibition or B/B 0, and therefore quench corrected CPM (QCCPM) values may be obtained by inserting the count rate of the unquenched standard. However, if there is a requirement for absolute DPM measurements, the unquenched standard should be counted in another appropriately standardized instrument and the absolute DPM value inserted. See Amersham s color quench kit pack leaflet for further details. MicroBeta reports QCCPM as DPM. A new header (for example QCCPM where QCCPM= DPM) can be created in the Output section of the protocol editor. ParaLux Count Mode standardization protocols must include background subtraction. The background sample is an empty plate and should be the same type that is used in the assay. This is counted before the standardization samples. Instrument background will then be automatically subtracted from the unknown samples and therefore background subtraction should not be selected in the counting protocol. Generally, using the SQP(I) quench indicator, the MicroBeta TriLux can correct with acceptable accuracy up to approximately 70% quench. (All replicates are corrected to within 10% of DPM in coincidence counting mode.) Improved precision with highly colored samples (up to 90% quench) can be obtained with ParaLux Count Mode. GENERAL PROCEDURE Instrument performance can be checked for counting efficiency and detector uniformity by following instructions detailed in the color quench kit pack leaflet. AQP(I) variation between detectors should be less than ± 8 units. Decide whether the quench curve will be installed using settled beads or beads in suspension. (See color quench kit pack for details.) See SPA Spectra and Operating Instructions for window setting and protocol editing. 8
9 9
10 10
11 OPERATING INSTRUCTIONS Illustrations are from Windows Workstation vers
12 MICROBETA TRILUX SPA SPECTRA Tritium YS Tritium PVT Default window Iodine-125 YS Iodine-125 PVT Default window Sulphur-35 PVT Phosphorus-33 PVT World Headquarters: PerkinElmer Life Sciences, 549 Albany Street, Boston, MA USA (800) In Europe: PerkinElmer Life Sciences, B-1930 Zaventem, Belgium, P.O.Box 10, FIN Turku, Finland, MicroBeta is a registered trademark of PerkinElmer, Inc. Amersham is a trademark of Amersham International, plc. Products for SPA (Scintillation Proximity Assay) are available from Amersham International, plc. Scintillation Proximity Assay Technology is covered by US Patent No , European Patent No and by Japanese Patent Application No. 84/ , Jan 2001 Printed in Finland by Offset House Oy Naantali 2001
Research. Image FlashPlate
Drug Discovery Research Image FlashPlate Clinical Screening Application Note Use of Image FlashPlate on the Wallac ViewLux ultrahts microplate imager The Wallac ViewLux is an ultra high throughput microplate
More informationApplication Note. Quench Correction for LANCE Time-Resolved Fluorescence Resonance Energy Transfer LANCE.
Drug Discovery Research Clinical Screening LANCE Application Note Quench Correction for LANCE Time-Resolved Fluorescence Resonance Energy Transfer INTRODUCTION In LANCE TR-FRET assays the sample interference
More informationQuench and Quench Correction
TCA-015 Quench and Quench Correction Abstract The TopCount Microplate Scintillation and Luminescence Counter is capable of analyzing radiolabeled microplate samples in a wide variety of formats using liquid
More informationDrug Discovery Research Clinical Screening. Comparison of ELISA and AlphaScreen Assay Technologies for Measurement of Protein Expression Levels
Drug Discovery Research Clinical Screening FLAG Expression Measurement Application Note Comparison of ELISA and AlphaScreen Assay Technologies for Measurement of Protein Expression Levels ALPHASCREEN APPLICATION
More informationHomogeneous GTPγS Assay for High Throughput Screening of GPCRs
Homogeneous GTPγS Assay for High Throughput Screening of GPCRs Gregory Warner, Ph.D.*, Patricia Kasila, and Harry Harney * Enanta Pharmaceuticals Inc., 750 Main Street, Cambridge, MA 02139, 549 Albany
More informationTime-Resolved Fluorescence Based GTP Binding Assay for G-Protein Coupled Receptors
Time-Resolved Fluorescence Based GTP Binding Assay for G-Protein Coupled Receptors Gregory Warner, Rita Syystö, Heini Frang, Christel Gripenberg-Lerche, Jurgen Vanhauwe, Tarja Ahola, and Satu Kovanen Abstract
More informationFlashPlate File #15. High Throughput Screening. J. Watson SmithKline Beecham Pharmaceuticals, UK.
Drug Discovery Research Clinical Screening High Throughput Screening FlashPlate File #15 Use of Novel FlashPlate Technology to Measure camp Accumulation in Chinese Hamster Ovary Cells Expressing Human
More informationDetermination of the 14 C content in bio-based products applications, techniques and tools
Application Note Liquid Scintillation Counters Determination of the 14 C content in bio-based products applications, techniques and tools In brief Various techniques can be used for the determination of
More informationComparison of camp Assay Technologies for High Throughput Screening
Comparison of camp Assay Technologies for High Throughput Screening Patricia Kasila and Harry Harney, 549 Albany Street, Boston, MA 02118 1 Abstract Homogeneous camp assays have been developed to allow
More informationCounting Aqueous Samples with the TopCount Microplate Scintillation Counter
TCA-005 Counting Aqueous Samples with the TopCount Microplate Scintillation Counter Abstract The TopCount TM Microplate Scintillation Counter quantitates radioisotopes by liquid scintillation counting
More informationFlashPlate File #13. A Homogeneous Protein Kinase Assay Performed on 384-Well Streptavidin-Coated FlashPlate HTS PLUS. High Throughput Screening
Drug Discovery Research Clinical Screening High Throughput Screening FlashPlate File #13 A Homogeneous Protein Kinase Assay Performed on 384-Well Streptavidin-Coated FlashPlate HTS PLUS Carol Geist, Sally
More informationUse and Preparation of Quench Curves in Liquid Scintillation Counting
APPLICATION NOTE Liquid Scintillation Authors: Jock Thomson PerkinElmer, Inc. Waltham, MA Use and Preparation of Quench Curves in Liquid Scintillation Counting Introduction The liquid scintillation process
More informationDOC Application note Version 1.0 HIDEX 300 SL
TDCR quench correction with the HIDEX 300 SL This application note provides an introduction to Quench correction with the Hidex 300 SL.This note focuses on Triple to Double Coincidence Ratio method or
More informationHighly Sensitive Detection of the Interaction Occurring Between Phage Displayed Peptides and their Target using AlphaScreen
Highly Sensitive Detection of the Interaction Occurring Between Phage Displayed Peptides and their Target using AlphaScreen N. Rouleau, P. Allard, P. Roby, D. Sexton*, F. Whelihan* and R. Bossé * DYAX
More informationFlashPlate File #11. Converting 96-Well Assays to 384-Well FlashPlate Assays. High Throughput Screening
Drug Discovery Research Clinical Screening High Throughput Screening FlashPlate File #11 Converting 96-Well Assays to 384-Well FlashPlate Assays Daniel L. Sissors, Ph.D. and Sally Casto PerkinElmer Life
More informationArmenian Hamster IgG Antigen ELISA Kit
Armenian Hamster IgG Antigen ELISA Kit Catalog No: IAHTIGGKT Lot No: SAMPLE INTENDED USE This Armenian hamster Immunoglobulin G (IgG) antigen assay is intended for the quantitative determination of total
More informationEvaluating Microplate Detection Instruments
AppNote Evaluating Microplate Detection Instruments Introduction Important features to consider when evaluating a microplate reader are the following: sensitivity, dynamic range, throughput, integration
More informationPathHunter β-arrestin Human and Ortholog GPCR Assays
PathHunter β-arrestin Human and Ortholog GPCR Assays Easy automation for cell-based assays on the Fluent laboratory automation solution Introduction The ability to carry out cell-based G-Protein coupled
More informationCyno Monkey IgG Antigen ELISA Kit
Cyno Monkey IgG Antigen ELISA Kit Catalog No: ICYIGGKT Lot No: SAMPLE INTENDED USE This cynomolgus macaque (Macaca fascicularis) monkey Immunoglobulin G (IgG) antigen assay is intended for the quantitative
More informationSPHK1 Assay Kit. Catalog Number KA assays Version: 05. Intended for research use only.
SPHK1 Assay Kit Catalog Number KA0906 400 assays Version: 05 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 General Information... 4 Materials Supplied...
More informationHuman IgG Antigen ELISA Kit
Human IgG Antigen ELISA Kit Catalog No: IHUIGGKT Lot No: SAMPLE INTENDED USE This human immunoglobulin G antigen assay is intended for the quantitative determination of total human IgG antigen in serum,
More informationHigh-throughput automation with the Attune NxT Autosampler: consistent results across all wells and across plates
APPLICATION NOTE Attune NxT Flow Cytometer with Autosampler High-throughput automation with the Attune NxT Autosampler: consistent results across all wells and across plates Introduction The emerging field
More informationMeasuring cell proliferation using the CyQUANT Cell Proliferation Assay with SpectraMax Microplate Readers
APPLICATION NOTE Measuring cell proliferation using the CyQUANT Cell Proliferation Assay with SpectraMax Microplate Readers Introduction Quantitation of cell proliferation using fluorescence allows one
More informationratmrp2 PREDIVEZ Protocol CAT. NO. SBPV03
ratmrp2 PREDIVEZ Protocol CAT. NO. SBPV03 Page 1 of 14 Determination of the interaction of drugs with rat Mrp2 using the rat Mrp2 Fluorescent PREDIVEZ Kit For the following membrane product: -HEK293-VT
More informationProductivity Barriers
LumiLux Cellular Screening Platform Break Through the Productivity Barriers in Cellular Screening Imagine an instrument that will cellular screening productivity Introducing a high impact instrument whose
More informationA Comparison of AlphaLISA and TR-FRET Homogeneous Immunoassays in Serum-Containing Samples
application Note A Comparison of and Homogeneous Immunoassays in Serum-Containing Samples Authors Anuradha Prasad, PhD, Catherine Lautenschlager, PhD, Stephen Hurt, PhD, David Titus, PhD and Stéphane Parent,
More informationRat Factor XII Total Antigen ELISA Kit
Rat Factor XII Total Antigen ELISA Kit Catalog No: IRFXIIKT-TOT Lot No: SAMPLE INTENDED USE This rat coagulation Factor XII antigen assay is intended for the quantitative determination of total Factor
More informationMST Starting Guide Monolith NT.115
MST Starting Guide Monolith NT.115 Contents 1. How to design an experiment 2. Before you start 3. Assay Setup Pretests 4. Assay Setup 5. MST-experiment using temperature control 6. Data Interpretation
More informationATPlite Assay Performance in Human Primary Cells
A P P L I C AT I O N N O T E Cell Viability Assays Author: Verena Brucklacher-Waldert Crescendo Biologics Cambridge, UK Assay Performance in Human Primary Cells Introduction In vitro assays using primary
More informationPI3 Kinase Assay Kit. Catalog Number KA assays Version: 05. Intended for research use only.
PI3 Kinase Assay Kit Catalog Number KA0904 400 assays Version: 05 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 General Information... 4 Materials Supplied...
More informationSTUDY OF THE PERFORMANCE OF EFFICIENCY TRACING TECHNIQUE ON A TRICARB 2100TR LIQUID SCINTILLATION ANALYZER
NUCLEAR PHYSICS RADIATION PROTECTION STUDY OF THE PERFORMANCE OF EFFICIENCY TRACING TECHNIQUE ON A TRICARB 2100TR LIQUID SCINTILLATION ANALYZER R. I. DOBRIN, C. N. DULAMA Institute for Nuclear Research,
More informationQuench & Quench Curves. Jock Thomson Global Product Manager, Cocktails & Vials PerkinElmer Life Sciences
Quench & Quench Curves Jock Thomson Global Product Manager, Cocktails & Vials PerkinElmer Life Sciences What is Quench? Basically, the liquid scintillation process is the conversion of the energy of a
More informationFaster Chemiluminescent camp detection by HitHunter camp XS+ on the SpectraMax L Luminescence Microplate Reader
SpectraMax Systems Faster Chemiluminescent camp detection by HitHunter camp XS+ on the SpectraMax L Luminescence Microplate Reader By Rajini Bompelli and Sunitha Ambikapathi, DiscoveRx Corporation, 42501
More informationCentroLIApc LB 962. Microplate Luminometer
detect and identify CentroLIApc LB 962 Microplate Luminometer CentroLIApc LB 962 Microplate Luminometer The Essence Over the past two decades RIA has, to a large extent, been replaced by none-radiometric
More informationTroponin I (Rat) ELISA Kit
Troponin I (Rat) ELISA Kit Catalog Number KA4341 96 assays Version: 01 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 General Information...
More informationMouse Factor XII Total ELISA Kit
Mouse Factor XII Total ELISA Kit Catalog No: IMFXIIKT-TOT Lot No: SAMPLE INTENDED USE This mouse coagulation Factor XII antigen assay is intended for the quantitative determination of total Factor XII
More informationGPCR induction followed by a fluorometric Ca 2+ assay in Thermo Scientific Varioskan LUX
GPCR induction followed by a fluorometric Ca 2+ assay in Thermo Scientific Varioskan LUX SP&A Application Laboratory, Thermo Fisher Scientific, Vantaa, Finland Application Notes Key Words Cellular signal,
More informationA Picofluor method for RNA quantitation using RiboGreen
A Picofluor method for RNA quantitation using 1. INTRODUCTION RNA Quantitation Reagent is an ultrasensitive fluorescent nucleic acid stain. It is a simple and rapid procedure for measuring RNA concentration
More informationHow to overcome the challenges presented by bioassays
How to overcome the challenges presented by bioassays Tony Mire-Sluis Vice President, North America, Singapore, Abingdon, Contract and Product Quality Amgen Inc Why are Bioassays Considered so Difficult
More informationBiological Applications of Microplate Scintillation Counting using the TopCount
Biological Applications of Microplate Scintillation Counting using the TopCount MICROPLATE SCINTILLATION & LUMINESCENCE COUNTER C A S E S T U D Y Abstract The microplate format has become increasingly
More informationPhospholipase D Assay Kit
Phospholipase D Assay Kit Catalog Number KA1636 100 assays Version: 04 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of the
More informationPolarScreen Estrogen Receptor-β Competitor Assay, Green. Table of Contents
PolarScreen Estrogen Receptor-β Competitor Assay, Green Catalog nos. P2615 and P2700 Storage: See page 2 Protocol part no. L0712 Revision date: 10 December 2009 Table of Contents Page 1. Introduction...
More informationImpact of Standardizing on Traceable QC for Volume Transfers Artel
Impact of Standardizing on Traceable QC for Volume Transfers 1 The challenges How is liquid handler performance evaluated in your lab? What information do you obtain (accuracy and/or precision)? How long
More informationMouse Factor X Total Antigen ELISA Kit
Mouse Factor X Total Antigen ELISA Kit CATALOG NO: IMFXKT-TOT LOT NO: SAMPLE INTENDED USE This mouse Factor X total assay is for the quantitative determination of total Factor X and Xa in biological fluids.
More informationratmdr1b PREDIVEZ Protocol CAT. NO. SBPV01
ratmdr1b PREDIVEZ Protocol CAT. NO. SBPV01 Page 1 of 12 Determination of the interaction of drugs with the rat Mdr1b transporter using the SB-ratMdr1b PREDIVEZ Kit For the following membrane product: -VT
More informationOptimizing Dual-Glo Luciferase Assays with the Synergy HT Multi-Detection Microplate Reader
Optimizing Dual-Glo Luciferase Assays with the Synergy HT Multi-Detection Microplate Reader Introduction Today s biological science and drug discovery research often involves the measurement of large numbers
More informationFlexStation 3 Microplate Reader
FlexStation 3 Microplate Reader A five-mode microplate reader with integrated fluid transfer > five-mode reader with wide range of applications > flexible liquid transfer enables more assay conditions
More informationFunctional GPCR Studies Using AlphaScreen camp Detection Kit
PPLICTION NOTE lpha Technology uthors: Jen Carlstrom Dawn Nida PerkinElmer, Inc. Hopkinton, M Functional GPCR Studies Using lphascreen cmp Detection Kit Introduction G protein-coupled receptors (GPCRs)
More informationFluo-4 NW Calcium Assay Kits (F36205, F36206)
Product Information Revised: 20 October 2006 Fluo-4 NW Calcium Assay Kits (F36205, F36206) F36205 Fluo-4 NW Calcium Assay Kit (high-throughput) *for 100 microplates* F36206 Fluo-4 NW Calcium Assay Kit
More informationRetinol Binding Protein Urinary EIA Kit
K-ASSAY KAMIYA BIOMEDICAL COMPANY KAMIYA BIOMEDICAL COMPANY Retinol Binding Protein Urinary EIA Kit For the quantitative determination of RBP in human, rat, dog and rhesus monkey urine Cat. No. KT-744
More informationBECAUSE CAN T WAIT BREAKTHROUGHS. Drug Discovery Screening Solutions SCREENING SOLUTIONS. Automated Liquid Handling. Assays and Reagents
BECAUSE BREAKTHROUGHS CAN T WAIT SCREENING SOLUTIONS Automated Liquid Handling Assays and Reagents Imaging and Detection Systems Informatics Drug Discovery Screening Solutions 2 As the premier provider
More informationHuman Prolactin Antigen ELISA Kit
Human Prolactin Antigen ELISA Kit Catalog No: IHPRLKT Lot No: SAMPLE INTENDED USE This human prolactin antigen assay is intended for the quantitative determination of prolactin antigen in human plasma
More informationLabcyte. Labcyte Consumables BROCHURE ECHO QUALIFIED WITH SUPERIOR ECHO PERFORMANCE. Version 3.0 APRIL 2017 LABCYTE INC.
Labcyte Consumables ECHO QUALIFIED WITH SUPERIOR ECHO PERFORMANCE Labcyte BROCHURE Version 3.0 APRIL 2017 LABCYTE INC. 170 Rose Orchard Way San Jose, CA 95134 USA Toll-free: +1 877 742 6548 Fax: +1 408
More informationA Bridging Immunogenicity Assay Using SPARCL TM Technology
A Bridging Immunogenicity Assay Using SPARCL TM Technology Wenhua F. Xie Lumigen Inc., a Beckman Coulter Company INTRODUCTION Evaluating the immune response in patients is an important aspect associated
More informationBoTest A/E Botulinum Neurotoxin Detection Kit- Detailed Description
Figure 1. BoTest A/E Reporter. (A) BioSentinel s BoTest A/E Reporter. CFP and YFP are connected by SNAP-25 (amino acids 141-206). The cleavage sites for each BoNT sero-type are indicated. (B) A FRET assay
More informationigem 2018 InterLab Study Protocol Before You Begin
igem 2018 InterLab Study Protocol Before You Begin Read through this entire protocol sheet carefully before you start your experiment and prepare any materials you may need. In order to improve reproducibility,
More informationMyBioSource.com. Rat Hydroxyproline ELISA USER INSTRUCTION
Rat Hydroxyproline ELISA USER INSTRUCTION Cat.No MBS162747 Standard Curve Range: 10ng/L - 4000ng/L Sensitivity: 4.99ng/L Size: 96 wells Storage: Store the reagents at 2-8 C. For over 6-month storage refer
More informationRat Cortisol(CORTISOL)ELISA Kit. Instruction. Sample Types Validated Serum, blood plasma,saliva, Urine, and other related tissue Liquid.
Rat Cortisol(CORTISOL)ELISA Kit Instruction Sample Types Validated Serum, blood plasma,saliva, Urine, and other related tissue Liquid. Please read this insert completely prior to using the product. FOR
More informationDynex DS2 2-Plate ELISA Processing System. Compact Easy to Use Innovative A Perfect Combination.
Dynex DS2 2-Plate ELISA Processing System Compact Easy to Use Innovative A Perfect Combination. Immunology Infectious Disease Autoimmune Allergy Food Safety Forensics Enterics Serology Immunology Infectious
More informationHigh Throughput Quantitation of Cytokine Biomarkers using LANCE Ultra TR-FRET Assays
APPLIATION NOTE LANE TR-FRET Authors: Jen arlstrom Stephen Hurt Roger osse PerkinElmer, Inc. Hopkinton, MA High Throughput Quantitation of ytokine iomarkers using LANE Ultra TR-FRET Assays Introduction
More informationPolarScreen Tyrosine Kinase Assay Kit, Far Red Protocol
PolarScreen Tyrosine Kinase Assay Kit, Far Red Protocol Part # PV3327 Lit. # 839-0408647 040804 TABLE OF CONTENTS 1.0 INTRODUCTION... 1 2.0 ASSAY THEORY... 2 3.0 KIT COMPONENTS... 3 3.1 Materials Provided...
More informationDELFIA, LANCE and TruPoint technologies Sensitivity and dynamic range using PerkinElmer instrumentation and microplates
DELFIA, LANCE and TruPoint technologies Sensitivity and dynamic range using PerkinElmer instrumentation and microplates Introduction DELFIA, LANCE and TruPoint are all assay technologies based on time-resolved
More informationHuman immunoglobulin G(IgG) ELISA Kit
Human immunoglobulin G(IgG) ELISA Kit For the quantitative determination of human immunoglobulin G (IgG) concentrations in serum, plasma, cell culture supernates, urine, tissue homogenates, cell lysates.
More informationQuench & Quench Curves. James Thomson Meridian Biotechnologies Ltd.
Quench & Quench Curves James Thomson Meridian Biotechnologies Ltd. What is Quench? Basically, the liquid scintillation process is the conversion of the energy of a radioactive decay event into photons
More informationFunctional Characterization and Screening of G αi -Coupled GPCRs using AlphaScreen camp Detection Technology
Functional Characterization and Screening of G αi -Coupled GPCRs using AlphaScreen camp Detection Technology Nathalie Bouchard, Erik Robitaille and Dean Wenham 1 Introduction G-protein coupled receptors
More informationQuantiFluor ONE dsdna System
TECHNICAL MANUAL QuantiFluor ONE dsdna System Instructions for Use of Products E4871 and E4870 Revised 6/17 TM405 QuantiFluor ONE dsdna System All technical literature is available at: www.promega.com/protocols/
More informationSynergy NEO HTS Multi-Mode Microplate Reader and Life Technologies Reagents
A p p l i c a t i o n G u i d e Synergy NEO HTS Multi-Mode Microplate Reader and Life Technologies Reagents High Performance Multi-Detection Capabilities that Span the Small Molecule Drug Discovery Process
More informationRayBio ApoSENSOR TM ATP Cell Viability Assay Kit
RayBio ApoSENSOR TM ATP Cell Viability Assay Kit User Manual Version 1.0 October 1 st, 2015 Cat#: 68CV-ATP-S200 Cat#: 68CV-ATP-S1000 RayBiotech, Inc. We Provide You With Excellent Support And Service Tel:(Toll
More informationChicken rudimental bovine serum albumin(bsa) check-up ELISA kit
Chicken rudimental bovine serum albumin(bsa) check-up ELISA kit Catalog Number. CSB-EQ027303CH For the quantitative determination of chicken rudimental bovine serum albumin(bsa) check-up concentrations
More informationAcetyl-p53 (K381) Cell-Based Colorimetric ELISA Kit
Acetyl-p53 (K381) Cell-Based Colorimetric ELISA Kit Catalog No. KA8015C Detection and Quantification of Acetyl-p53 (K381) Protein Concentration in Cell. Research Purposes Only. Not Intended for Diagnostic
More informationAMPURE PCR PURIFICATION PAGE 1 OF 7
PCR PURIFICATION PAGE 1 OF 7 Please refer to http://www.agencourt.com/technical/reagent_information/ for updated protocols. AMPure is a registered trademark of Agencourt Bioscience and is for laboratory
More informationMouse Alpha-2-Antiplasmin Total ELISA Kit
Mouse Alpha-2-Antiplasmin Total ELISA Kit Catalog No: IMA2APKT-TOT Lot No: 413 INTENDED USE This mouse α2-antiplasmin assay is for the quantitative determination of total α2-antiplasmin in mouse plasma.
More informationNew and Highly Sensitive AlphaScreen camp Assay to Measure Femtomol camp Variations in Cell and Membrane Based Assays
New and Highly Sensitive AlphaScreen camp Assay to Measure Femtomol camp Variations in Cell and Membrane Based Assays 1 Illy, C., 1 Boissonneault, M., 1 Bouchard, N., 2 Dimeo, J., 1 Hamann, G., 1 Legault,
More informationMyBioSource.com. Human Vitamin K2 ELISA USER INSTRUCTION
Human Vitamin K2 ELISA USER INSTRUCTION Cat.No MBS163021 Standard Curve Range: 5ng/ml - 1000ng/ml Sensitivity: 2.52ng/ml Size: 96 wells Storage: Store the reagents at 2-8 C. For over 6-month storage refer
More informationLyticBLAzer -FRET B/G Kit Protocol
LyticBLAzer -FRET B/G Kit Protocol Cat. nos. K1150 and K1151 O-13391-r1 US 0305 TABLE OF CONTENTS TABLE OF CONTENTS... 1 1.0 INTRODUCTION... 1 2.0 MATERIALS SUPPLIED... 2 3.0 MATERIALS AND EQUIPMENT REQUIRED,
More informationValidation and Comparison of Single-Step Flash and Dual-Spectral Luciferase Reporter Gene Assays using the Synergy Line of Microplate Readers
A p p l i c a t i o n N o t e Validation and Comparison of Single-Step Flash and Dual-Spectral Luciferase Reporter Gene Assays using the Synergy Line of Microplate Readers Peter J. Brescia and Peter Banks,
More informationRat α-melanocyte stimulating hormone (α-msh) ELISA Kit
Rat α-melanocyte stimulating hormone (α-msh) ELISA Kit For the quantitative determination of rat α-melanocyte stimulating hormone (α-msh) concentrations in serum, plasma, tissue homogenates. This package
More informationHuman connective tissue growth factor (CTGF) ELISA Kit. MyBioSource.com. This package insert must be read in its entirety before using this product.
Human connective tissue growth factor (CTGF) ELISA Kit Catalog Number. For the quantitative determination of human connective tissue growth factor (CTGF) concentrations in serum, plasma, tissue homogenates.
More informationMyBioSource.com. Human Osteoprotegerin ELISA USER INSTRUCTION
Human Osteoprotegerin ELISA USER INSTRUCTION Cat.No MBS162588 Standard Curve Range: 0.05ng/ml - 15ng/ml Sensitivity: 0.023ng/ml Size: 96 wells Storage: Store the reagents at 2-8 C. For over 6-month storage
More informationHigh Throughput Suspension Cell and Bead Analysis. Intellicyt ique Screener PLUS
High Throughput Suspension Cell and Bead Analysis Intellicyt ique Screener PLUS High Throughput Suspension Cell and Bead Analysis Intellicyt ique Screener PLUS the fastest path to actionable results Whether
More informationComparison of LANCE Ultra TR-FRET to PerkinElmer s Classical LANCE TR-FRET Platform for Kinase Applications
Comparison of LANCE Ultra TR-FRET to PerkinElmer s Classical LANCE TR-FRET Platform for Kinase Applications LANCE ULTRA TR-FRET TECHNOLOGY A P P L I C A T I O N N O T E Introduction Protein kinases play
More informationQuantiFluor dsdna System
TECHNICAL MANUAL QuantiFluor dsdna System Instructions for Use of Product E2670 Revised 5/15 TM346 QuantiFluor dsdna System All technical literature is available at: www.promega.com/protocols/ Visit the
More informationSialyltransferase Activity Kit Catalog Number: EA002
Sialyltransferase Activity Kit Catalog Number: EA002 The protocol presented is for a 96-well format. Materials provided are sufficient for two 96-well microplates or equivalent. This package insert must
More informationGel/PCR Extraction Kit
Gel/PCR Extraction Kit Item No: EX-GP200 (200rxns) Content Content Binding Buffer BD Wash Buffer PE Elution Buffer (10 mm Tris-HCl, ph 8.5) Spin Columns EX-GP200 80 ml 20 mlx3 10 ml 200 each Description
More informationValidation Using SDS Software Version on the Applied Biosystems 7500 Real-Time PCR System and the ABI PRISM 7000 Sequence Detection System
User Bulletin Quantifiler Kits April 2006 SUBJECT: Validation Using SDS Software Version 1.2.3 on the Applied Biosystems 7500 Real-Time PCR System and the ABI PRISM 7000 Sequence Detection System Note:
More informationbeta-secretase Activity Assay Kit
beta-secretase Activity Assay Kit Catalog Number KA0900 100 assays Version: 04 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 General Information... 4
More informationA Comparison of PerkinElmer s LANCE camp Assay Performance to that of State of the Art Competitive Technologies
A Comparison of PerkinElmer s LANCE camp Assay Performance to that of State of the Art Competitive Technologies GPCR ASSAYS A P P L I C A T I O N N O T E Introduction G-protein coupled receptors (GPCRs)
More informationMembrane Potential Assays Using the ValiScreen Human Kv1.3 Voltage-Gated K + Channel Cell Line on the EnVision Multilabel Plate Reader
TECHNICAL BRIEF ValiScreen Stable Recombinant Ion Channel Cell Line and the EnVision Multilabel Plate Reader Membrane Potential Assays Using the ValiScreen Human Kv1.3 Voltage-Gated K + Channel Cell Line
More informationDETERMINATION OF PROTEIN BINDING BY UPLC/MS/MS
DETERMINATION OF PROTEIN BINDING BY UPLC/MS/MS Darcy Shave and Pete Alden Waters Corporation, Milford, MA, U.S.A. INTRODUCTION Plasma protein binding (PPB) can significantly affect the therapeutic action
More informationTurboMass GC/MS Software
TurboMass GC/MS Software GAS CHROMATOGRAPHY / MASS SPECTROMETRY P R O D U C T N O T E Today s demanding laboratory requires software that is easy to learn yet offers sophisticated instrument control, robust
More informationab JC-10 Mitochondrial Membrane Potential Assay Kit Flow Cytometry
ab112133 JC-10 Mitochondrial Membrane Potential Assay Kit Flow Instructions for Use For detecting mitochondrial membrane potential changes in cells using our proprietary fluorescence probe. This product
More informationDevelopment of an AlphaLISA MEK1 Kinase Assay Using Full-Length ERK2 Substrate
TECHNICAL NOTE Development of an AlphaLISA MEK1 Kinase Assay Using Full-Length ERK2 Substrate AlphaLISA Technology Author: Jeanine Hinterneder, PhD PerkinElmer, Inc. Hopkinton, MA Introduction The mitogen-activated
More informationFlexStation 3 microplate reader
FlexStation 3 microplate reader A five-mode microplate reader with integrated fluid transfer > five-mode reader with wide range of applications > flexible liquid transfer enables more assay conditions
More informationHuman GTPase KRas (KRAS) ELISA Kit
Human GTPase KRas (KRAS) ELISA Kit For the quantitative determination of human GTPase KRas (KRAS) concentrations in serum, plasma, tissue homogenates. This package insert must be read in its entirety before
More informationRat gamma-aminobutyric acid (GABA) ELISA Kit
Rat gamma-aminobutyric acid (GABA) ELISA Kit Catalog Number. CSB-E08368r For the quantitative determination of endogenic rat gamma-aminobutyric acid (GABA) concentrations in serum, plasma, tissue homogenates.
More informationHuman ferritin(fe) ELISA Kit
Human ferritin(fe) ELISA Kit Catalog Number. CSB-E05187h For the quantitative determination of human ferritin(fe) concentrations in serum, plasma, tissue homogenates, cell lysates. This package insert
More informationEcoli 360 HCP ELISA ELISA Kit for the Determination of Ecoli-HCP in process-derived Samples
Ecoli 360 HCP ELISA ELISA Kit for the Determination of Ecoli-HCP in process-derived Samples FOR IN VITRO USE ONLY 1 INTENDED USE The supplied Ecoli 360 HCP ELISA kit is a sandwich ELISA to be performed
More informationMouse high density lipoprotein (HDL) ELISA Kit
Mouse high density lipoprotein (HDL) ELISA Kit For the quantitative determination of mouse high density lipoprotein (HDL) concentrations in serum, plasma, tissue homogenates. This package insert must be
More informationHuman vascular endothelial cell growth factor A (VEGF-A) ELISA Kit
Human vascular endothelial cell growth factor A (VEGF-A) ELISA Kit For the quantitative determination of human vascular endothelial cell growth factor A (VEGF-A) concentrations in serum, plasma, tissue
More informationA fluorogenic assay for identification of coagulation factor IIa inhibitors- screening for stroke treatment
A fluorogenic assay for identification of coagulation factor IIa inhibitors- screening for stroke treatment Easy automation of a biochemical assay on the Fluent laboratory automation solution Introduction
More information