1 Copyright (C) JCPDS International Centre for Diffraction Data Reduction of the Sample Size in the Analysis of Rock by EDXRF Riidiger Harmel, Ulrike Otto, Olaf Haupt, Clemens Sch$er and Walter Dannecker, University of Hamburg, Institute of Inorganic and Applied Chemistry, Martin-Luther-King-Platz 6, Hamburg, Germany Introduction The advantages of the XRF-technology, the easy preparation, the non-destructive measurement, the fast multi-element analysis, lead to a growing interest in this technology even in fields otherwise covered by ICP-MS or AAS because after the XRF-analysis it is possible to do further investigations with the same specimen. In the analysis of samples from ancient buildings and historic monuments, like churches, chapels and castles, the composition of stones from damaged and yet undamaged parts is of interest to find out if the weathered stones have lost salts by elution or get new ones by capillary effects, if the initial composition of the stone is still present or has changed. During the restoration of these buildings there is often only a small amount of material to analyze due to the protection of the monuments. Therefore it was necessary to make efforts to reduce the usual sample amount of 3-4 g which is needed for a pellet. Instrumentation In conventional operation, specimens with a diameter of 32 mm and an area of 8.06 cm2 were analyzed using an energy-dispersive XRF spectrometer (X-lab, Spectra A.I., Germany) containing secondary targets. The specimen was fixed in a rotating cup positioned on a plate for 18 cups. The primary source was a rhodium side-window tube, and the IS0 Debyeflex 300 x-ray generator (Seifert, Germany) produced high-voltage potentials between 8 and 60 kv. The emitted fluorescence radiation produced current pulses in a Si(Li)-semiconductor with an active area of 10 mm2 and an effective thickness of 3 mm. Reduction of the Sample Size Small Pellets In the instrument the radiation beam comes from underneath of the specimen with the detector at right angles to the incident radiation (figure 1): sample holder /,/I- -\,, / / x-ray \, / / I radiation \, x-ray tube / d&or secondary targets fig. 1: small pellet holder (side view) The radiation excites only a part of the usual 32 mm-pellet. Therefore a specimen rotation is used to cover the whole area of the pellet. But with this rotation turned off a smaller pellet can

2 This document was presented at the Denver X-ray Conference (DXC) on Applications of X-ray Analysis. Sponsored by the International Centre for Diffraction Data (ICDD). This document is provided by ICDD in cooperation with the authors and presenters of the DXC for the express purpose of educating the scientific community. All copyrights for the document are retained by ICDD. Usage is restricted for the purposes of education and scientific research. DXC Website ICDD Website -

3 874 be placed directly in the excited area (13 mm diameter). A special Teflon ring was constructed to minimize backscattering and the small pellet was arranged asymmetrically in the ring due to the geometry of the device (figure 2). teflon ring pelle fig. 2: small pellet holder (top view) The results in mg/kg obtained by this method with the usually used pellet calibration show approximately 60% of the concentration of a normal size pellet but vary dependent on the matrix. Further investigations on these effects didn t take place because the reduction from 4 g to 660 mg is not very remarkable. Dry Filter Preparation In our institute a calibration for quartz fiber [1,2] and membrane filters [3] was developed usually for use with aerosol sampling. However, due to the grain size difference of an aerosol and a finely ground stone is not very remarkable, the evaluation of a filter coated with powdered stone with this calibration is possible. To avoid in-deptheffects membrane filters were used (Sartorius, Germany, cellulose nitrate filter, 0.45 pm pore size, 50 mm diameter). The experimental set-up shows figure 3. The sample was placed at the bottom of the device, swirled up with purified compressed air and sucked through the filter. The big advantage of this method is the sample amount of only 4 mg. That is only the thousandth part of the original pellet. The disadvantage is the loose filter coating. It is necessary to handle the filters very carefully in order not to lose parts of the coating. Since the radiation in the spectrometer comes from beneath (figure 1) the specimen has to be placed upside down into it. Therefore a fixation of the coating is necessary. Several attempts with different methods have been tested. Up to now hair spray seems to deliver the best results but puyp / i., filter +.,..-..~_-- air admittance ---J fig. 3: dry filter preparation unit (25 I/min) -,: f is not an optimum. To avoid the loose coating another idea has been tried.

4 875 Wet Filter Preparation A suspension of the sample was prepared in a flask, stirred very well and a part of it was given into the instrumental set-up shown in figure 4. To obtain a homogeneous distribution over the whole filter area it is necessary to mix the sample suspension with the suspension liquid. The liquid was sucked through the filter and a stable coating of only 4 mg has been obtained. As liquid both, water and toluene have been used. With toluene the elution effects of the sample are reduced, but the handling is more difficult due to the smaller surface tension. In order to avoid in-depth-effects, to get a coating on top of the filter, a membrane filter with a pore size of 0.1 pm was used (Sartorius, Germany, cellulose nitrate filter, 50 mm diameter). The largest possible disadvantage of the wet filter preparation method, the elution of the sample by the liquid, is found to be below 3% with the samples tested yet. This error is lower than the complete mistake of the method. suspension liquid \ filter-2 ~ Pump waste I &%z?z?& Ii!I I fig. 4: wet filter preparation unit Weighing With an aerosol filter calibration the result is given in ng/cm2. The XRF-result of a stone is expected to be in mg/kg. Therefore the area weight of the filter has to be determined (equation 1). % (1) ;; = zg = ;; = resu~;af;;j;tion 1 _[ cm2 The area weight is calculated with the weight of the coating and the area of the coating. By weighing the empty filter and the coated filter five times each the coatings weight is obtained. The area is 32 mm in diameter (8.06 cm2). It is cut out from the whole 50 mm filter. This cut results in a mathematical correction which is described later. The membrane filters used have a weight of about 100 mg, but the coating is only 4 mg. For example a precision of 1% results in a repeatability of 40 pg. With a special micro-balance (Sartorius, Germany, MSP-OOOVOOl) including a faraday-cage it is possible to reach a precision of 2% for at least 5 measurements.

5 876 Results Dry preparation Figure 5 shows the measured concentration which is obtained by the filter calibration in ng/cm2 versus a reference concentration calculated by equation 2: [ 1 rock content z * coating s weight [ mg] [ ng 1 (2) reference concentration - = cm2 filter area [ cm* 1 The rock content was obtained by ICP-OES and ICP-MS. The correlation coefficient shows a relevant linear link between the measured concentration and the reference as expressed exemplary for Aluminum and Potassium. Other elements like Mg, Ca, Ti, Mn, Fe, Zn, Sr and Pb show a comparable behavior reference[ng/cm2] fig. 5: correlation curve - dry preparation Wet Preparation Nevertheless, figure 6 shows a similar correlation comparable to the dry preparation. Again, the reference concentration is calculated by equation 2.

6 Copyright (C) JCPDS International Centre for Diffraction Data reference [ng/cm*] Fig. 6: correlation curve - wet preparation There s a linear correlation between the measured and the calculated reference concentration. Hence it is possible to calculate a correction equation for each element. With these equations it is possible to calculate the concentration of other specimen prepared with the same method. The coated area of the wet preparation method is larger than the area of the dry preparation. This is due to the instrumental set-up. Therefore the sensitivity of the wet preparation method seems to be smaller than the sensitivity of the dry preparation method because the area which is cut out is the same (32 mm diameter),. Comparison with ICP-OES For the ICP-OES values shown in figure 7 the specimen was prepared by wet digestion. The XRF results were calculated by the concentration obtained from the former calibration curve (equation 3) and the weight of the coating as shown in equation 4. (3) cont. (cal curve) = cont. (measured) *a, + b, (x = element) (4) concentration (Cal. curve) Figure 7 presents only four elements with similar concentration ranges. But for other elements like Sr and Ti the results are in their range comparable.

7 x s g z$ T Mg Al K Fe ICP-OES digestion q XRF wet filter / Fig. 7: comparison ICP-OES / XRF The difference of the values ranges between 1 O-20% which is sufficient for a rough estimation of the main rock components. Conclusion The decrease from 4 g for the pellet method down to 4 mg with the filter method means a reduction by 99,9% in sample size. Nevertheless, due to the small sample size, trace element concentrations might be lower than the detection limit. Hence, for applications where a fingerprint of the analyzed stone from the main components is of interest the method is pretty good. And the differences of the ICP-OES and the XRF results shown in figure 7 might get lower if the calibration curve is enlarged with more samples. Further investigations have to take particle size distribution and different matrices into account. References 1 0. Haupt, B. Klaue, C. Schafer and W. Dannecker, Preparation of Quartz Fibre Filter Standards for X-Ray Fluorescence Analysis of Aerosol Samples, X-Ray Spectrometry 24, (1995) Haupt, C. Schafer, S. Strauss and W. Dannecker, Production of calibration standards for x-ray fluorescence analysis of aerosol particles precipitated on different filter materials, Fresenius J. Anal. Chem. 355,375 (1996) 3. C. Schafer, 0. Haupt and W. Dannecker, Comparison of two different approaches to calibrate an energy-dispersive x-ray fluorescence spectrometer for the analysis of aerosolloaded filters, Fresenius J. Anal. Chem. 355,379 (1996)

ENERGY-DISPERSIVE X-RAY FLUORESCENCE ANALYSIS OF MONO- AND POLYCRYSTALS OF SELENIDE SPINELS BY FUNDAMENTAL PARAMETER METHOD

ENERGY-DISPERSIVE X-RAY FLUORESCENCE ANALYSIS OF MONO- AND POLYCRYSTALS OF SELENIDE SPINELS BY FUNDAMENTAL PARAMETER METHOD 322 ENERGY-DISPERSIVE X-RAY FLUORESCENCE ANALYSIS OF MONO- AND POLYCRYSTALS OF SELENIDE SPINELS BY FUNDAMENTAL PARAMETER METHOD ABSTRACT Rafa Sitko, Beata Zawisza, Ewa Malicka Institute of Chemistry, Silesian

More information

HANDY WAVEGUIDE TXRF SPECTROMETER FOR NANOGRAM SENSITIVITY

HANDY WAVEGUIDE TXRF SPECTROMETER FOR NANOGRAM SENSITIVITY 213 HANDY WAVEGUIDE TXRF SPECTROMETER FOR NANOGRAM SENSITIVITY Shinsuke Kunimura and Jun Kawai Department of Materials Science and Engineering, Kyoto University, Sakyo-ku, Kyoto 606-8501, Japan ABSTRACT

More information

SPECTRAL INTERFERENCE IN X-RAY FLUORESCENCE ANALYSIS OF COMMON MATERIALS

SPECTRAL INTERFERENCE IN X-RAY FLUORESCENCE ANALYSIS OF COMMON MATERIALS Copyright JCPDS - International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Volume 46. 38 ISSN 097-0002 SPECTRAL INTERFERENCE IN X-RAY FLUORESCENCE ANALYSIS OF COMMON MATERIALS Frank

More information

Calibration Standard& for XRFA of Emitted Aerosol Particles Collected on a Newly Developed Filter Band with an Automated Sampler

Calibration Standard& for XRFA of Emitted Aerosol Particles Collected on a Newly Developed Filter Band with an Automated Sampler 760 Calibration Standard& for XRFA of Emitted Aerosol Particles Collected on a Newly Developed Filter Band with an Automated Sampler Olaf Haupt, Rtidiger Harmel and Walter Dannecker University of Hamburg,

More information

ANALYSIS OF TCLP EXTRACTS BY X-RAY FLUORESCENCE

ANALYSIS OF TCLP EXTRACTS BY X-RAY FLUORESCENCE 257 ANALYSIS OF TCLP EXTRACTS BY X-RAY FLUORESCENCE Douglas S. Kendall, Benjamin A. Burns, Jennifer A. Suggs and John J. Mackey National Enforcement Investigations Center, U.S. Environmental Protection

More information

APPLICATION OF A PORTABLE TXRF SPECTROMETER TO DETERMINE TRACE AMOUNTS OF TOXIC ELEMENTS

APPLICATION OF A PORTABLE TXRF SPECTROMETER TO DETERMINE TRACE AMOUNTS OF TOXIC ELEMENTS 18 APPLICATION OF A PORTABLE TXRF SPECTROMETER TO DETERMINE TRACE AMOUNTS OF TOXIC ELEMENTS Shinsuke Kunimura 1 and Jun Kawai 2 1 Materials Fabrication Laboratory, RIKEN (The Institute of Physical and

More information

Evaluation of Energy-Dispersive X-Ray Fluorescence for Mobile Analysis of Hazardous Metals in Transuranic Waste

Evaluation of Energy-Dispersive X-Ray Fluorescence for Mobile Analysis of Hazardous Metals in Transuranic Waste Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 146 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42

More information

PORTABLE XRF ANALYSIS OF JAPANESE HISTORICAL OBJECTS

PORTABLE XRF ANALYSIS OF JAPANESE HISTORICAL OBJECTS Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 36 PORTABLE XRF ANALYSIS OF JAPANESE HISTORICAL OBJECTS ABSTRACT Yasuhiro Hayakawa National Research

More information

REVISIT OF FORENSIC ANALYSIS OF ARSENIC POISONING CASE 1998

REVISIT OF FORENSIC ANALYSIS OF ARSENIC POISONING CASE 1998 Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN 1097-0002 177 REVISIT OF FORENSIC ANALYSIS OF ARSENIC POISONING CASE 1998 Jun Kawai Department of Materials Science and Engineering,

More information

Preconcentration of Environmental Waters by Agar for XRF Analysis

Preconcentration of Environmental Waters by Agar for XRF Analysis 96 Preconcentration of Environmental Waters by Agar for XRF Analysis Kazuhiko NAKANO 1, 2), Kenta OKUBO 1), Kouichi TSUJI 1) 1 ) Department of Applied Chemistry, Graduate School of Engineering, Osaka City

More information

ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS

ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 130 ABSTRACT ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS G Roach and J Tickner

More information

XRF S ROLE IN THE PRODUCTION OF MAGNESIUM METAL BY THE MAGNETHERMIC METHOD

XRF S ROLE IN THE PRODUCTION OF MAGNESIUM METAL BY THE MAGNETHERMIC METHOD Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 398 XRF S ROLE IN THE PRODUCTION OF MAGNESIUM METAL BY THE MAGNETHERMIC METHOD H. L. Baker Northwest Alloys,

More information

SYSTEMATIC ERRORS IN LINEAR PSD BASED HTXRD SYSTEMS

SYSTEMATIC ERRORS IN LINEAR PSD BASED HTXRD SYSTEMS Copyright(c)JCPDS-International Centre for Diffraction Data 2,Advances in X-ray Analysis,Vol.43 267 SYSTEMATIC ERRORS IN LINEAR PSD BASED HTXRD SYSTEMS E.A. Payzant and W.S. Harrison, III * Metals and

More information

Evaluation of Some Sample Support Films for Microsample X-Ray Analysis

Evaluation of Some Sample Support Films for Microsample X-Ray Analysis 301 Evaluation of Some Sample Support Films for Microsample X-Ray Analysis A.W. Wilson, D.C. Turner, A.A. Robbins, Process Analytics, Orem, UT Abstract Reduction of matrix and support film background counts

More information

2017 International Conference on Materials Science and Biological Engineering (ICMSBE 2017)

2017 International Conference on Materials Science and Biological Engineering (ICMSBE 2017) 2017 International Conference on Materials Science and Biological Engineering (ICMSBE 2017) Determination of strontium content in strontium nitrate used for fireworks and firecrackers based on Energy Dispersive

More information

AN INNOVATED LABORATORY XAFS APPARATUS

AN INNOVATED LABORATORY XAFS APPARATUS Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 397 AN INNOVATED LABORATORY XAFS APPARATUS TAGUCHI Takeyoshi XRD Division, Rigaku Corporation HARADA

More information

ElvaX Light for Petrochemical analysis

ElvaX Light for Petrochemical analysis ElvaX Light for Petrochemical analysis Introduction ElvaX Light is a powerful and productive instrument for analysis of petroleum products, such as automotive fuels (gasoline, diesel), lubricating oil,

More information

ANALYSIS OF PIGMENTS USED IN A JAPANESE PAINTING

ANALYSIS OF PIGMENTS USED IN A JAPANESE PAINTING ANALYSIS OF PIGMENTS USED IN A JAPANESE PAINTING 263 Yasuhiro Hayakawa, 1 Seiji Shirono, 1 Sadatoshi Miura, 1 and Tomohide Matsushima 2 1 National Research Institute for Cultural Properties, Tokyo, Japan

More information

Evaluation of methods for analysis of lead in air particulates: An intra-laboratory and interlaboratory

Evaluation of methods for analysis of lead in air particulates: An intra-laboratory and interlaboratory Evaluation of methods for analysis of lead in air particulates: An intra-laboratory and interlaboratory comparison Frank X. Weber, James M. Harrington, Andrea McWilliams, Keith E. Levine Trace Inorganics

More information

Expanded Coating Analysis Performance for electroless Ni Plating and Bath Analysis with Micro-XRF

Expanded Coating Analysis Performance for electroless Ni Plating and Bath Analysis with Micro-XRF Expanded Coating Analysis Performance for electroless Ni Plating and Bath Analysis with Micro-XRF Bruker Nano Analytics, Berlin, Germany Webinar, December 05, 2018 Innovation with Integrity Presenters

More information

OPTIMIZING THE BALANCE OF QUALITY AND TURNAROUND TIME FOR A PROCESS CONTROL XRF LABORATORY

OPTIMIZING THE BALANCE OF QUALITY AND TURNAROUND TIME FOR A PROCESS CONTROL XRF LABORATORY 104 OPTIMIZING THE BALANCE OF QUALITY AND TURNAROUND TIME FOR A PROCESS CONTROL XRF LABORATORY ABSTRACT Suzanne W. Bowe Kennecott Utah Copper, Magna, Utah Kennecott Utah Copper, outside Salt Lake City,

More information

S2 RANGER LE: Analysis of Light Elements in Cement, Slags and Feldspar

S2 RANGER LE: Analysis of Light Elements in Cement, Slags and Feldspar S2 RANGER LE: Analysis of Light Elements in Cement, Slags and Feldspar 1 Welcome Today s topics: S2 RANGER LE o o o Overview XFlash LE detector Improved X-ray tube Analysis of Light Elements in Cement

More information

Determination of iron content in iron oxide used for fireworks and firecrackers based on Energy Dispersive X-ray Fluorescence Spectrometry (EDXRF

Determination of iron content in iron oxide used for fireworks and firecrackers based on Energy Dispersive X-ray Fluorescence Spectrometry (EDXRF Advances in Engineering Research (AER), volume 148 3rd Workshop on Advanced Research and Technology in Industry Applications (WARTIA 2017) Determination of iron content in iron oxide used for fireworks

More information

Analysis of an OTC Multivitamin Tablet Using X-ray Fluorescence Spectroscopy and Atomic Absorption Spectroscopy

Analysis of an OTC Multivitamin Tablet Using X-ray Fluorescence Spectroscopy and Atomic Absorption Spectroscopy Analysis of an OTC Multivitamin Tablet Using X-ray Fluorescence Spectroscopy and Atomic Absorption Spectroscopy Anne Concepcion G. Pozon, Nathalie P. Wong, Dr. Julita C. Robles* and Dr. Wyona C. Patalinghug

More information

XRF ANALYSIS OF HIGH GAIN-ON-IGNITION SAMPLES BY FUSION METHOD USING FUNDAMENTAL PARAMETER METHOD

XRF ANALYSIS OF HIGH GAIN-ON-IGNITION SAMPLES BY FUSION METHOD USING FUNDAMENTAL PARAMETER METHOD Copyright JCPDS-International Centre for Diffraction Data 2013 ISSN 1097-0002 177 RF ANALYSIS OF HIGH GAIN-ON-IGNITION SAMPLES BY FUSION METHOD USING FUNDAMENTAL PARAMETER METHOD Mitsuru atanabe 1, Hisashi

More information

ZINC/IRON PHASE TRANSFORMATION STUDIES ON GALVANNEALED STEEL COATINGS BY X-RAY DIFFRACTION

ZINC/IRON PHASE TRANSFORMATION STUDIES ON GALVANNEALED STEEL COATINGS BY X-RAY DIFFRACTION Copyright JCPDS - International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Volume 46. 291 ZINC/IRON PHASE TRANSFORMATION STUDIES ON GALVANNEALED STEEL COATINGS BY X-RAY DIFFRACTION S.

More information

TXRF TO MONITOR FOR HIGH K DIELECTRIC MATERIAL CONTAMINATION IN A SEMICONDUCTOR FAB

TXRF TO MONITOR FOR HIGH K DIELECTRIC MATERIAL CONTAMINATION IN A SEMICONDUCTOR FAB Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 517 TXRF TO MONITOR FOR HIGH K DIELECTRIC MATERIAL CONTAMINATION IN A SEMICONDUCTOR FAB Chris Sparks

More information

Philips Analytical, Lelyweg 1, 7602 EA Almelo, The Netherlands

Philips Analytical, Lelyweg 1, 7602 EA Almelo, The Netherlands Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 284 MICRO-DIFFRACTION WITH MONO-CAPILLARIES M.J. Fransen, J.H.A. Vasterink and J. te Nijenhuis Philips

More information

Analysis of Pharmaceuticals by ICP-MS and WDXRF

Analysis of Pharmaceuticals by ICP-MS and WDXRF Analysis of Pharmaceuticals by ICP-MS and WDXRF Frank X. Weber, James M. Harrington, Andrea McWilliams, James Medlin, Keith E. Levine Trace Inorganics Department, RTI International, Research Triangle Park,

More information

F. J. Cadieu*, I. Vander, Y. Rong, and R. W. Zuneska, Physics Department, Queens College of CUNY, Flushing, NY

F. J. Cadieu*, I. Vander, Y. Rong, and R. W. Zuneska, Physics Department, Queens College of CUNY, Flushing, NY Copyright JCPDS-International Centre for Diffraction Data 2012 ISSN 1097-0002 1 X-Ray Measurements of Nanometer Thick Ta x O 1-x and Hf x O 1-x Films on Silicon Substrates for Thickness and Composition

More information

APPLYING THE RIETVELD METHOD TO MINERAL FILLED POLYPHENYLENE SULFIDE COMPOUNDS

APPLYING THE RIETVELD METHOD TO MINERAL FILLED POLYPHENYLENE SULFIDE COMPOUNDS Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 103 APPLYING THE RIETVELD METHOD TO MINERAL FILLED POLYPHENYLENE SULFIDE COMPOUNDS R. W. Morton, D. E.

More information

CHAPTER-6 ELEMENTAL ANALYSIS OF LUBRICATING OIL SAMPLES USING EDXRF TECHNIQUE

CHAPTER-6 ELEMENTAL ANALYSIS OF LUBRICATING OIL SAMPLES USING EDXRF TECHNIQUE CHAPTER-6 ELEMENTAL ANALYSIS OF LUBRICATING OIL SAMPLES USING EDXRF TECHNIQUE 6.1 INTRODUCTION 144 6.2 MATRIX EFFECTS 145 6.3 SAMPLE PREPARATION 146 6.4 EXPERIMENTAL METHODS 147 6.5 EVALUTION PROCEDURE

More information

CERTIFICATE OF ANALYSIS

CERTIFICATE OF ANALYSIS in cooperation with the WG Aluminium of the Committee of Chemists of GDMB CERTIFICATE OF ANALYSIS ERM -EB312a AlMgSi0,5 Certified value 1) Uncertainty 2) Element Mass fraction in % Si 0.403 0.008 Fe 0.198

More information

APPLICATIONS OF X-RAY STRESS MEASUREMENT FOR INTERFACE AREA OF Ni 3 Al SYSTEM INTERMETALLIC COMPOUND COATING

APPLICATIONS OF X-RAY STRESS MEASUREMENT FOR INTERFACE AREA OF Ni 3 Al SYSTEM INTERMETALLIC COMPOUND COATING Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 385 APPLICATIONS OF X-RAY STRESS MEASUREMENT FOR INTERFACE AREA OF Ni 3 Al SYSTEM INTERMETALLIC

More information

EDXRF APPLICATION NOTE IRON IN ALUMINUM # 1237

EDXRF APPLICATION NOTE IRON IN ALUMINUM # 1237 EDXRF APPLICATION NOTE IRON IN ALUMINUM # 1237 SCOPE The analysis of iron in aluminum extrussion billets is demonstrated. BACKGROUND Often, selection of a bauxite mine site is determined by the alumina,

More information

HIGH TEMPERATURE X-RAY DIFFRACTION STUDY OF REACTION RATES IN CERAMICS

HIGH TEMPERATURE X-RAY DIFFRACTION STUDY OF REACTION RATES IN CERAMICS Copyright JCPDS - International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Volume 46. 226 ABSTRACT HIGH TEMPERATURE X-RAY DIFFRACTION STUDY OF REACTION RATES IN CERAMICS M.S. Peterson,

More information

X-ray fluorescence analysis of zinc and zinc iron alloy coated steel sheet

X-ray fluorescence analysis of zinc and zinc iron alloy coated steel sheet Technical articles X-ray fluorescence analysis of zinc and zinc iron alloy coated steel sheet Takao Moriyama* and Kenji Kodama* 1. Introduction Since it is possible to apply zinc to the surface of steel

More information

ANALYTICAL PERFORMANCE OF A HANDHELD EDXRF SPECTROMETER WITH MINIATURE X-RAY TUBE EXCITATION

ANALYTICAL PERFORMANCE OF A HANDHELD EDXRF SPECTROMETER WITH MINIATURE X-RAY TUBE EXCITATION JPACSM 56 ANALYTICAL PERFORMANCE OF A HANDHELD EDXRF SPECTROMETER WITH MINIATURE X-RAY TUBE EXCITATION Volker Thomsen and Debbie Schatzlein NITON LLC Billerica, MA ABSTRACT The analytical performance of

More information

Electron Probe Micro-Analysis (EPMA)

Electron Probe Micro-Analysis (EPMA) Electron Probe Micro-Analysis (EPMA) Nilanjan Chatterjee, Ph.D. Principal Research Scientist 1 Electron Microprobe Facility Department of Earth, Atmospheric and Planetary Sciences Massachusetts Institute

More information

ElvaX ProSpector in Exploration & Mining

ElvaX ProSpector in Exploration & Mining ElvaX ProSpector in Exploration & Mining Introduction ElvaX ProSpector is a fast, accurate and easy tool for different mining applications. It provides onsite analysis of ore samples with minimal sample

More information

CHAPTER 7 MICRO STRUCTURAL PROPERTIES OF CONCRETE WITH MANUFACTURED SAND

CHAPTER 7 MICRO STRUCTURAL PROPERTIES OF CONCRETE WITH MANUFACTURED SAND 99 CHAPTER 7 MICRO STRUCTURAL PROPERTIES OF CONCRETE WITH MANUFACTURED SAND 7.1 GENERAL Characterizing the mineralogy of the samples can be done in several ways. The SEM identifies the morphology of the

More information

Lesson 1 Good Diffraction Data

Lesson 1 Good Diffraction Data Lesson 1 Good Diffraction Data Nicola Döbelin RMS Foundation, Bettlach, Switzerland Digital Diffractometers Transmission Geometry Debye-Scherrer Geometry Reflective Geometry Bragg-Brentano Geometry Glass

More information

X-RAY FLUORESCENCE SPECTROMETRY: DEVELOPMENT OF EFFECTIVE FUSION PROCEDURES FOR LIME PRODUCTS

X-RAY FLUORESCENCE SPECTROMETRY: DEVELOPMENT OF EFFECTIVE FUSION PROCEDURES FOR LIME PRODUCTS 76 X-RAY FLUORESCENCE SPECTROMETRY: DEVELOPMENT OF EFFECTIVE FUSION PROCEDURES FOR LIME PRODUCTS Jean-Philippe Gagnon Corporation Scientifique Claisse, 35 rue Franquet, Québec, QC, G1P 4P3, CANADA ABSTRACT

More information

X-Ray Fluorescence Lab Report

X-Ray Fluorescence Lab Report X-Ray Fluorescence Lab Report Arena Holguin, Sylvia Navidad, Gabriel Gonzalez MME5401/MASE 6390 11/14/11 1. Introduction X-Ray fluorescence (XRF) is a popular, nondestructive analytical method for elemental

More information

Thickness and composition analysis of thin film samples using FP method by XRF analysis

Thickness and composition analysis of thin film samples using FP method by XRF analysis Technical articles Thickness and composition analysis of thin film samples using FP method by XRF analysis Hikari Takahara* 1. Introduction X-ray fluorescence spectroscopy (XRF) is an elemental quantification

More information

APPROACHING A UNIVERSAL SAMPLE PREPARATION METHOD FOR XRF ANALYSIS OF POWDER MATERIALS

APPROACHING A UNIVERSAL SAMPLE PREPARATION METHOD FOR XRF ANALYSIS OF POWDER MATERIALS Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 368 APPROACHING A UNIVERSAL SAMPLE PREPARATION METHOD FOR XRF ANALYSIS OF POWDER MATERIALS Abstract John

More information

ANNEALING STUDIES OF PURE AND ALLOYED TANTALUM EMPLOYING ROCKING CURVES

ANNEALING STUDIES OF PURE AND ALLOYED TANTALUM EMPLOYING ROCKING CURVES Copyright JCPDS - International Centre for Diffraction Data 3, Advances in X-ray Analysis, Volume. 5 ISSN 197- Abstract ANNEALING STUDIES OF PURE AND ALLOYED TANTALUM EMPLOYING ROCKING CURVES David W.

More information

AN IN SITU HIGH-TEMPERATURE X-RAY DIFFRACTION STUDY OF PHASE TRANSFORMATIONS IN SILVER BEHENATE

AN IN SITU HIGH-TEMPERATURE X-RAY DIFFRACTION STUDY OF PHASE TRANSFORMATIONS IN SILVER BEHENATE Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 27 AN IN SITU HIGH-TEMPERATURE X-RAY DIFFRACTION STUDY OF PHASE TRANSFORMATIONS IN SILVER BEHENATE

More information

How to Increase Productivity in Mining with Automated Simultaneous XRF Analysis. Arkady Buman and Dr. Kai Behrens Bruker AXS Inc.

How to Increase Productivity in Mining with Automated Simultaneous XRF Analysis. Arkady Buman and Dr. Kai Behrens Bruker AXS Inc. How to Increase Productivity in Mining with Automated Simultaneous XRF Analysis Arkady Buman and Dr. Kai Behrens Bruker AXS Inc. Today s Speakers Arkady Buman Product Manager XRF Madison, WI, USA Dr. Kai

More information

S4 T STAR Biopharmaceutical production: Rapid nutrient and contaminant analysis of cell culture media by TXRF spectroscopy

S4 T STAR Biopharmaceutical production: Rapid nutrient and contaminant analysis of cell culture media by TXRF spectroscopy DMEM medium Homogenization S4 T-STAR Lab Report XRF 468 Low detection limits S4 T STAR Biopharmaceutical production: Rapid nutrient and contaminant analysis of cell culture media by TXRF spectroscopy Cell

More information

APPLICATION OF A PORTABLE TXRF SPECTROMETER TO DETERMINE TRACE AMOUNTS OF TOXIC ELEMENTS

APPLICATION OF A PORTABLE TXRF SPECTROMETER TO DETERMINE TRACE AMOUNTS OF TOXIC ELEMENTS 18 APPLICATION OF A PORTABLE TXRF SPECTROMETER TO DETERMINE TRACE AMOUNTS OF TOXIC ELEMENTS Shinsuke Kunimura 1 and Jun Kawai 2 1 Materials Fabrication Laboratory, RIKEN (The Institute of Physical and

More information

JEOL JXA-8200 Superprobe

JEOL JXA-8200 Superprobe JEOL JXA-8200 Superprobe Electron gun Liq N 2 for EDS Electron gun Condenser lens Optical Microscope Optical Microscope EDS WDS SE Liq N 2 EDS BSE Objective lens WDS Sample exchange chamber Stage High

More information

INVESTIGATION OF LOCAL TEXTURES IN EXTRUDED MAGNESIUM BY SYNCHROTRON RADIATION

INVESTIGATION OF LOCAL TEXTURES IN EXTRUDED MAGNESIUM BY SYNCHROTRON RADIATION Copyright JCPDS - International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Volume 46. 151 INVESTIGATION OF LOCAL TEXTURES IN EXTRUDED MAGNESIUM BY SYNCHROTRON RADIATION Heinz-Günter

More information

Analysis of inhomogeneous samples and trace element detection in alloys using QUANTAX Micro-XRF on SEM

Analysis of inhomogeneous samples and trace element detection in alloys using QUANTAX Micro-XRF on SEM Analysis of inhomogeneous samples and trace element detection in alloys using QUANTAX Micro-XRF on SEM Bruker Nano Analytics, Berlin, Germany Webinar, June 02, 2016 Innovation with Integrity Presenters

More information

IDENTIFICATION OF CuCl in Si by XRD-XRF ANALYSIS

IDENTIFICATION OF CuCl in Si by XRD-XRF ANALYSIS Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 366 IDENTIFICATION OF CuCl in Si by XRD-XRF ANALYSIS Chi-Tang Li and Leo Tarhay Dow Corning Corporation,

More information

A QUICK AND RELIABLE FUSION METHOD FOR SILICON AND FERROSILICON

A QUICK AND RELIABLE FUSION METHOD FOR SILICON AND FERROSILICON Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 415 A QUICK AND RELIABLE FUSION METHOD FOR SILICON AND FERROSILICON Jimmy Blanchette Corporation

More information

THE ANALYSIS OF STRESS DISTRIBUTION BASED ON MEASUREMENTS USING TWO METHODS OF X-RAY APPLICATION

THE ANALYSIS OF STRESS DISTRIBUTION BASED ON MEASUREMENTS USING TWO METHODS OF X-RAY APPLICATION 182 THE ANALYSIS OF STRESS DISTRIBUTION BASED ON MEASUREMENTS USING TWO METHODS OF X-RAY APPLICATION ABSTRACT Barbara Kucharska Institute of Materials Engineering, Czestochowa University of Technology,

More information

Wavelength-dispersive X-ray fluorescence spectrometer

Wavelength-dispersive X-ray fluorescence spectrometer New products Wavelength-dispersive X-ray fluorescence spectrometer 1. Introduction The newly-released sequential general-purpose wavelength-dispersive X-ray fluorescence spectrometer, ZSX PrimusIII is

More information

MEASUREMENT OF RESIDUAL PHASE STRESS OF THE METAL, MATRIX COMPOSITE MATERIAL USING SYNCHROTRON RADIATION

MEASUREMENT OF RESIDUAL PHASE STRESS OF THE METAL, MATRIX COMPOSITE MATERIAL USING SYNCHROTRON RADIATION Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 215 MEASUREMENT OF RESIDUAL PHASE STRESS OF THE METAL, MATRIX COMPOSITE MATERIAL USING SYNCHROTRON RADIATION

More information

Instrument Configuration for Powder Diffraction

Instrument Configuration for Powder Diffraction Instrument Configuration for Powder Diffraction Advanced X-ray Workshop S.N. Bose National Centre for Basic Sciences, 14-15/12/2011 Innovation with Integrity Overview What is the application? What are

More information

X-RAY STUDY OF THE INHOMOGENEITY OF SURFACE RESIDUAL STRESSES AFTER SHOT-PEENING TREATMENT

X-RAY STUDY OF THE INHOMOGENEITY OF SURFACE RESIDUAL STRESSES AFTER SHOT-PEENING TREATMENT Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 48 X-RAY STUDY OF THE INHOMOGENEITY OF SURFACE RESIDUAL STRESSES AFTER SHOT-PEENING TREATMENT V. Monin,

More information

On site heavy metal testing and monitoring using monochromatic XRF methods

On site heavy metal testing and monitoring using monochromatic XRF methods On site heavy metal testing and monitoring using monochromatic XRF methods X-Ray Fluorescence Analysis hn o hn i 8/5/2017 XOS Confidential File 2 Counts (100 seconds) Limitation of XRF High scattering

More information

Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume

Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 240 SIMULTANEOUS MEASUREMENTS OF X-RAY DIFFRACTION (XRD) AND DIFFERENTIAL SCANNING CALORIMETRY (DSC)

More information

Thermo Scientific ARL EQUINOX 100. X-ray Diffractometers

Thermo Scientific ARL EQUINOX 100. X-ray Diffractometers Thermo Scientific ARL EQUINOX 100 X-ray Diffractometers High performance in a compact size Thermo Scientific ARL EQUINOX 100 X-ray diffractometer (XRD) is designed to meet structural and phase analysis

More information

Analytical Techniques for Grade and Quality Control in Coal Mining. Dr Paul O Meara XRD Application Specialist

Analytical Techniques for Grade and Quality Control in Coal Mining. Dr Paul O Meara XRD Application Specialist Analytical Techniques for Grade and Quality Control in Coal Mining Dr Paul O Meara XRD Application Specialist Topics Pulsed Fast & Thermal Neutron Activation (PFTNA) Cross-belt Elemental Analysis of Coal

More information

Lesson 3 Sample Preparation

Lesson 3 Sample Preparation Lesson 3 Sample Preparation Nicola Döbelin RMS Foundation, Bettlach, Switzerland January 14 16, 2015, Bern, Switzerland Repetition: Bragg-Brentano Diffractometer Typical Configuration (with Kβ filter)

More information

XRF ANALYSIS OF AUTOMOTIVE CATALYSTS BY FLUX/FUSION

XRF ANALYSIS OF AUTOMOTIVE CATALYSTS BY FLUX/FUSION Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 472 XRF ANALYSIS OF AUTOMOTIVE CATALYSTS BY FLUX/FUSION A. R. Drews Physical and Environmental

More information

USING XRD ELASTIC AND PLASTIC STRAIN DATA TO EVALUATE THE EFFECTIVENESS OF DIFFERENT COLD-WORKING TECHNIQUES IN AEROSPACE MATERIALS

USING XRD ELASTIC AND PLASTIC STRAIN DATA TO EVALUATE THE EFFECTIVENESS OF DIFFERENT COLD-WORKING TECHNIQUES IN AEROSPACE MATERIALS 51 USING XRD ELASTIC AND PLASTIC STRAIN DATA TO EVALUATE THE EFFECTIVENESS OF DIFFERENT COLD-WORKING TECHNIQUES IN AEROSPACE MATERIALS Beth S. Matlock, Technology for Energy Corp., Knoxville, TN, USA Daniel

More information

Wavelength Dispersive XRF Spectrometer

Wavelength Dispersive XRF Spectrometer ISO9001 ISO14001 for Wafers, Media Disks, and Large Samples Wavelength Dispersive XRF Spectrometer ADVANCED XRF For Large and Irregularly Shaped Samples Is there any solution to analyze a large sample

More information

X-Ray Fluorescence Measurements of Molten Aluminum Elemental Composition

X-Ray Fluorescence Measurements of Molten Aluminum Elemental Composition B.2 Recycling Technologies X-Ray Fluorescence Measurements of Molten Aluminum Elemental Composition Leigh C. Duren (MS Candidate Industrial Intern) Advisors: D. Apelian & D. Backman Sponsor: wte Corporation

More information

XRD AND XAFS STUDIES OF CARBON SUPPORTED Pt-Ru ELECTROCATALYST IN A POLYMER-ELECTROLYTE-FUEL-CELL

XRD AND XAFS STUDIES OF CARBON SUPPORTED Pt-Ru ELECTROCATALYST IN A POLYMER-ELECTROLYTE-FUEL-CELL Copyright JCPDS - International Centre for Diffraction Data 4, Advances in X-ray Analysis, Volume 47. 56 XRD AND XAFS STUDIES OF CARBON SUPPORTED Pt-Ru ELECTROCATALYST IN A POLYMER-ELECTROLYTE-FUEL-CELL

More information

Guideline regarding analysis on Fujitsu Group specified chemical substances

Guideline regarding analysis on Fujitsu Group specified chemical substances Guideline regarding analysis on Fujitsu Group specified chemical substances June 20, 2009 (Edition 3.0) Fujitsu Limited 1. Objective 2. Reporting contents 3. Recommendation on analysis method 3.1 X-ray

More information

OPTIMIZING XRD DATA. By: Matthew Rayner

OPTIMIZING XRD DATA. By: Matthew Rayner OPTIMIZING XRD DATA By: Matthew Rayner 1 XRD Applications PANalytical classifies XRD applications in 4 groups 1. Powders 2. Nanomaterials 3. Solid objects 4. Thin films Many day-to-day samples cross these

More information

X-Ray Study of Soft and Hard Magnetic Thin Films

X-Ray Study of Soft and Hard Magnetic Thin Films Copyright (C) JCPDS-International Centre for Diffraction Data 1999 13 X-Ray Study of Soft and Hard Magnetic Thin Films Po-Wen Wang, 390 Reed St., Stormedia, Inc., Santa Clara CA. 95050 Abstract : This

More information

CHARACTERIZATION OF THE SOLIDS WASTE IN THE HANFORD WASTE TANKS USING A COMBINATION OF XRD, SEM AND PLM *

CHARACTERIZATION OF THE SOLIDS WASTE IN THE HANFORD WASTE TANKS USING A COMBINATION OF XRD, SEM AND PLM * Copyright JCPDS - International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Volume 46. 251 CHARACTERIZATION OF THE SOLIDS WASTE IN THE HANFORD WASTE TANKS USING A COMBINATION OF XRD,

More information

New Developments in Measuring Low Silicon in Process Piping Using Handheld X-ray Fluorescence (XRF)

New Developments in Measuring Low Silicon in Process Piping Using Handheld X-ray Fluorescence (XRF) New Developments in Measuring Low Silicon in Process Piping Using Handheld X-ray Fluorescence (XRF) Presented by: Mark Lessard Business Development Manager Thermo Fisher Scientific mark.lessard@thermofisher.com

More information

THE CLAISSE-CALCULATOR OF SAMPLES ACIDITY APPLICATIONS

THE CLAISSE-CALCULATOR OF SAMPLES ACIDITY APPLICATIONS 71 THE CLAISSE-CALCULATOR OF SAMPLES ACIDITY APPLICATIONS Fernand Claisse Fernand CLAISSE inc 2270, Léon Harmel, Suite 165 Québec, QC, G1N 4L2 ABSTRACT The CLAISSE-CALCULATOR is a new tool for XRF analysts

More information

METHODS FOR MOUNTING RADIOACTIVE POWDERS FOR XRD ANALYSIS

METHODS FOR MOUNTING RADIOACTIVE POWDERS FOR XRD ANALYSIS 54 METHODS FOR MOUNTING RADIOACTIVE POWDERS FOR XRD ANALYSIS Leah N. Squires, Robert D. Mariani, Thomas Hartmann* and J. Rory Kennedy Idaho National Laboratory, Fuel Fabrication and Characterization Department,

More information

Composition Analysis of Animal Feed by HR ICP-OES

Composition Analysis of Animal Feed by HR ICP-OES Composition Analysis of Animal Feed by HR ICP-OES Abstract The High-Resolution Array ICP-OES equipped with Standard-Kit was used for the analysis of Al, B, Ca, Co, Cr, Cu, Fe, I, K, Na, Mn, Mg, Mo, Ni,

More information

ISO Simplified Borate Fusion / WDXRF Analytical Method for Iron Ore Including Total Iron Analysis: Part 2

ISO Simplified Borate Fusion / WDXRF Analytical Method for Iron Ore Including Total Iron Analysis: Part 2 Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN 1097-0002 185 ISO 9516-1 Simplified Borate Fusion / WDXRF Analytical Method for Iron Ore Including Total Iron Analysis: Part 2 Mathieu

More information

Soil Sampling, Preparation and Analysis

Soil Sampling, Preparation and Analysis Soil Sampling, Preparation and Analysis Kim H. Tan The University of Georgia Athens, Georgia Marcel Dekker, Inc. New York* Basel Hong Kong CONTENTS Preface iii Chapter 1 PRINCIPLES OF SOIL SAMPLING 1 1.1

More information

ATTACHMENTES FOR APD 2000 PRO POWDER X-RAY DIFFRACTOMETER. Monochromators

ATTACHMENTES FOR APD 2000 PRO POWDER X-RAY DIFFRACTOMETER. Monochromators Monochromators Secondary graphite monochromator Johansson Ka 1 monochromator Parabolic monochromator Secondary flat and curved graphite monochromators suitable for Ag, Cr, Fe, Cu, Co and Mo radiations

More information

Identification of Crystal Structure and Lattice Parameter. for Metal Powders Using X-ray Diffraction. Eman Mousa Alhajji

Identification of Crystal Structure and Lattice Parameter. for Metal Powders Using X-ray Diffraction. Eman Mousa Alhajji Identification of Crystal Structure and Lattice Parameter for Metal Powders Using X-ray Diffraction Eman Mousa Alhajji North Carolina State University Department of Materials Science and Engineering MSE

More information

Global Cement and Raw Materials Fusion/XRF Analytical Solution

Global Cement and Raw Materials Fusion/XRF Analytical Solution 263 Global Cement and Raw Materials Fusion/XRF Analytical Solution Mathieu Bouchard, John Anzelmo, and Sebastien Rivard Corporation Scientifique Claisse, Quebec, Canada Alexander Seyfarth and Larry Arias

More information

JSM-7800F Field Emission Scanning Electron Microscope

JSM-7800F Field Emission Scanning Electron Microscope JSM-7800F catalogue JSM-7800F Field Emission Scanning Electron Microscope We provide high performance The Ultimate Research Tool for Multi-Disciplinary Research Institutions Extreme resolution The super

More information

DETERMINING OPERATING LIFETIME OF LANDING WHEEL HUBS WITH PRD PORTABLE DIFFRACTOMETER

DETERMINING OPERATING LIFETIME OF LANDING WHEEL HUBS WITH PRD PORTABLE DIFFRACTOMETER Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 345 DETERMINING OPERATING LIFETIME OF LANDING WHEEL HUBS WITH PRD PORTABLE DIFFRACTOMETER A.V. Lutzau

More information

Carnegie Mellon MRSEC

Carnegie Mellon MRSEC Carnegie Mellon MRSEC Texture, Microstructure & Anisotropy, Fall 2009 A.D. Rollett, P. Kalu 1 ELECTRONS SEM-based TEM-based Koseel ECP EBSD SADP Kikuchi Different types of microtexture techniques for obtaining

More information

Thermo Scientific X-ray product range. For chemical and phase analysis of solids, liquids and powders

Thermo Scientific X-ray product range. For chemical and phase analysis of solids, liquids and powders X-ray product range For chemical and phase analysis of solids, liquids and powders Metals, cement, mining, petrochemicals, environment, electronics, geology, glass, polymers, forensics, materials science,

More information

Thermo Scientific X-ray product range. For chemical and phase analysis of solids, liquids and powders

Thermo Scientific X-ray product range. For chemical and phase analysis of solids, liquids and powders X-ray product range For chemical and phase analysis of solids, liquids and powders Metals, cement, mining, petrochemicals, environment, electronics, geology, glass, polymers, forensics, materials science,

More information

Thermo Scientific ARL EQUINOX X-ray Diffractometers

Thermo Scientific ARL EQUINOX X-ray Diffractometers Thermo Scientific ARL EQUINOX 1000 X-ray Diffractometers High performance in a compact size Thermo Scientific ARL EQUINOX 1000 X-ray diffractometer (XRD) is designed to meet structural and phase analysis

More information

EFFECTS OF SUBSTRATE COATING WITH METAL SPRAYING ON RESIDUAL STRESSES IN SPUTTERED TITANIUM-NITRIDE FILMS

EFFECTS OF SUBSTRATE COATING WITH METAL SPRAYING ON RESIDUAL STRESSES IN SPUTTERED TITANIUM-NITRIDE FILMS Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 163 EFFECTS OF SUBSTRATE COATING WITH METAL SPRAYING ON RESIDUAL STRESSES IN SPUTTERED TITANIUM-NITRIDE

More information

The image part with relationship ID rid4 was not found in the file. Welcome

The image part with relationship ID rid4 was not found in the file. Welcome The image part with relationship ID rid4 was not found in the file. Welcome Trace Metals Analysis: Impurity Determinations By Thomas Kozikowski Chemist, R&D at Inorganic Ventures Key Considerations What

More information

EXPERIMENTAL STUDY OF X-RAY ENERGY SPECTRUM FORMED BY PLANAR WAVEGUIDE-RESONATOR WITH SPECIFIC ELEMENT COMPOSITION REFLECTORS

EXPERIMENTAL STUDY OF X-RAY ENERGY SPECTRUM FORMED BY PLANAR WAVEGUIDE-RESONATOR WITH SPECIFIC ELEMENT COMPOSITION REFLECTORS EXPERIMENTAL STUDY OF X-RAY ENERGY SPECTRUM FORMED BY PLANAR WAVEGUIDE-RESONATOR WITH SPECIFIC ELEMENT COMPOSITION REFLECTORS V.K. Egorov, E.V. Egorov IMT RAS, Chemogolovka, Moscow District, 142432 Russia

More information

January 12, Dr. Kevin Lackey TeckCominco Alaska Red Dog Mine Anchorage, AK Dear Dr. Lackey:

January 12, Dr. Kevin Lackey TeckCominco Alaska Red Dog Mine Anchorage, AK Dear Dr. Lackey: January 12, 2004 Dr. Kevin Lackey TeckCominco Alaska Red Dog Mine Anchorage, AK 99517 Dear Dr. Lackey: Thank you for submitting another set of samples for the final evaluation of our QuanX EC EDXRF system.

More information

CHARACTERIZATION OF X-RAY DIFFRACTION SYSTEM WITH A MICROFOCUS X-RAY SOURCE AND A POLYCAPILLARY

CHARACTERIZATION OF X-RAY DIFFRACTION SYSTEM WITH A MICROFOCUS X-RAY SOURCE AND A POLYCAPILLARY Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 278 CHARACTERIZATION OF X-RAY DIFFRACTION SYSTEM WITH A MICROFOCUS X-RAY SOURCE AND A POLYCAPILLARY OPTIC

More information

Applied Comparison in Coating Analysis Micro-XRF and Handheld XRF

Applied Comparison in Coating Analysis Micro-XRF and Handheld XRF Applied Comparison in Coating Analysis Micro-XRF and Handheld XRF Bruker Nano Analytics, Berlin, Germany Webinar, July 5 th 2017 Innovation with Integrity Bruker Webinar Outline Presenters Principles of

More information

ONLINE X-RAY ELEMENTAL ANALYSIS OF ORE AND OTHER BULKY MATERIALS

ONLINE X-RAY ELEMENTAL ANALYSIS OF ORE AND OTHER BULKY MATERIALS ONLINE X-RAY ELEMENTAL ANALYSIS OF ORE AND OTHER BULKY MATERIALS Dr. Albert Klein 1 Abstract Online elemental Analysis directly at the material stream or at a bypass is enabled by a modified X- ray fluorescence

More information

Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol

Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 407 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42

More information

Analysis of inorganic impurities in lubricating oils by ICP-MS

Analysis of inorganic impurities in lubricating oils by ICP-MS Analysis of inorganic impurities in lubricating oils by ICP-MS Application note Authors Energy and Fuels Junichi Takahashi, Kei Kasahara Agilent Technologies, Japan Introduction The most widely used techniques

More information

Appendix 11. Report by Becker et al on the BKA preliminary testing of the CIP entitled: analysis round robin.

Appendix 11. Report by Becker et al on the BKA preliminary testing of the CIP entitled: analysis round robin. Appendix 11. Report by Becker et al on the BKA preliminary testing of the CIP entitled: analysis round robin. Report on the analysis of PGM-bearing samples N1, N2, and N3 S. Becker KT 13 Forensic Science

More information