GUIDANCE NOTE. Performance of Electromagnetic Meters Designed for Use in Part-Filled Pipes.

Size: px
Start display at page:

Download "GUIDANCE NOTE. Performance of Electromagnetic Meters Designed for Use in Part-Filled Pipes."

Transcription

1 GUIDANCE NOTE Performance of Electromagnetic Meters

2 Performance of Electromagnetic Meters Introduction Electromagnetic flowmeters designed for use in part-filled pipes may be considered as a replacement for existing open channel flow metering systems (eg weirs and flumes) used in monitoring effluent discharges, storm water and sewer flows, and for industrial process control measurements. Such measurements are common in water and waste treatment, chemical and food industries. This Guidance Note summarises the results obtained from a DTI Flow Programme project in which the performances of a number of electromagnetic flowmeters designed for use in part-filled pipes were measured. These measurements were preceded by a literature search to identify the following: Manufacturers and suppliers of such meters The principles of operation of electromagnetic meters for use in part-filled pipes Technical details and installation requirements Previous experience and potential problems with the calibration of such meters. Manufacturers and Suppliers Four companies were identified as manufacturing electromagnetic flowmeters designed specifically for part-filled pipes. It was established in discussion with those companies that the most common meter sizes in use and consequently the most readily available were 200 mm and 400 mm nominal bore. Three of the companies agreed to supply one meter of each size to NEL for testing: Bailey-Fischer and Porter Ltd (Workington, Cumbria). Meters type PARTIMAG II range, supplied by J W Fairbairn Ltd, Glasgow. Krohne Altometer Ltd (Dordrecht, The Netherlands). Meters type TIDALFLUX IFM 4100 PF, supplied by Krohne Ltd, Wellingborough, Northamptonshire Marsh-McBirnie Europe (Belgium). Meter model 253 Flo-System TM, supplied by Flowline Manufacturing Ltd, Borehamwood. Flowmeter Operation General The flowmeters from the three manufacturers all operate on the principle of Faraday s law of magnetic induction. By that principle, a voltage will be generated across a conductor if that conductor moves at right angles through a magnetic field. Mains powered coils mounted in the meter body provide the magnetic field and the flowing liquid acts as the conductor. The voltage generated as the liquid moves through the magnetic field is sensed by electrodes mounted flush with the inner wall of the flowmeter. Alternatively, capacitive coupling may be used and this method avoids direct contact between the electrodes and the liquid. Bailey-Fischer & Porter flowmeters The meters are fitted with electromagnetic coils on a vertical axis to provide the required magnetic field across the meter section. Four pairs of electrodes are provided, one on the normal horizontal diameter location and the other three at lower positions. There is also a single electrode positioned near to the top of the meter bore. This single electrode, on sensing the presence of liquid, is used to switch the meter between part-filled and fully-filled operation. At each flow cross-section, full or part-full, the appropriate, weighting factor corrected, electrode pair is used for the voltage (flow signal) measurement. In addition to measuring the average flow velocity, the four pairs of electrodes detect a superimposed alternating current field for determination of the fill level. The signal voltage is corrected and converted to an output signal proportional to flowrate by using the fill level information and characteristic curves stored in the converter. The maximum measurement uncertainty claimed for the Bailey-Fischer & Porter meter in part-filled conditions (with flow velocity greater than 0.2 m/s and flow level above 0.15 x nominal meter diameter) is 3% of flowrate reading for flowrates above 0.02 of the range maximum. Below this rate and down to the minimum flowrate of range maximum, the quoted uncertainty is extended to 5% of reading. These values are for pulse output. In fully-filled conditions, the corresponding claimed uncertainty values are 0.04% of range maximum down to 0.04 of that maximum and 1% of rate below this value and down to the minimum flowrate. Krohne flowmeters One pair of electrodes is employed and these are positioned on a level of approximately 0.1 of the meter bore diameter on the lower part of that bore. Flow velocity is determined on the Faraday principle and the wetted cross-sectional area is computed from a capacitive level measuring system built into the meter tube liner. An additional feature of the Krohne flowmeter is that the ratio of flow level and flow rate is continuously measured and stored during normal operation. This data is evaluated for flow measurement at times when the electrodes are temporarily not in contact with the liquid, such as during the passage of wave motion. It is claimed that accurate measurements are therefore possible down to a flow level of 10% of meter tube bore diameter. The best measurement uncertainty claimed for the Krohne meter in part-filled conditions is 1% of the full scale flowrate for flow velocities 1 m/s. For fully-filled flowrates the uncertainty is quoted as 1% of the measured value for flow velocities 1 m/s. For fully filled flowrates less than 1 m/s the uncertainty is quoted as 0.5% of the measured value plus 5 mm/s.

3 June 1999 Marsh-McBirney Flo-System flowmeter This flowmeter (see Fig 1) is entirely different from conventional electromagnetic flowmeters (see Fig 1). It consists of a small velocity transducer which is installed in the bottom of a pipe bore. The transducer is held in position by a circular steel band sprung against the pipe bore wall. Calibration Methods Krohne Altometer at Dordrecht uses a calibration system involving two reservoirs which are connected by a low level pipe which always runs full and a high level pipe in which the liquid level can be adjusted by appropriate filling of the reservoirs. Flow is generated by a pump installed in the low level pipe and that flow is measured by a calibrated flowmeter also installed in that pipe. The flowmeter under test is installed in the high level pipe. Bailey-Fischer & Porter has a calibration facility at Göttingen in which full charged reference flowmeters are used at a low level in the facility while the meter under test is installed in a pipe at a higher level. The reference meters are calibrated against a gravimetric flow measurement system incorporated in the facility. The method used for the Marsh-McBirney meter is one of in situ calibration where the calibration coefficient is derived from a profiling (velocity traversing) technique using either the Flo-System sensor attached to the end of a rod or by some other means. Figure 1 - Installation of Marsh-McBirney meter in 200 mm pipe The meter signal is transmitted by a cable attached to the steel band and that cable emerges through an opening on the top of the pipe. Liquid level is measured by a piezoelectric pressure sensor located inside the velocity transducer. The velocity transducer generates a local electromagnetic field and the flowmeter instrumentation combines the transducer signals with level signals and uses this together with pre-established data concerning velocity profiles in a part-filled pipe to give the flowrate value. The claimed velocity measurement uncertainty of the Marsh- McBirney meter is 2% of reading over a defined velocity range. Liquid level measurement uncertainty is quoted as 6.4 mm and the conversion uncertainty (basic measurements converted to flowrate) is quoted as 2% of reading. NEL Calibration Facility and Programme The NEL large water calibration facility operates using flying-startand-stop diversion with static weighing. It was designed to operate with pipes fully charged. The 200 mm and 450 mm test lines of the facility were adapted to enable the flowmeter tests to be conducted under part-filled and fully-filled conditions. Conductivity of the facility water is between 3000 and 5000 µs/cm, this high value being caused by the sodium nitrite used as an anti-corrosion additive at NEL. The facility is NAMAS accredited with a best measurement uncertainty of 0.1% of flowrate with a coverage factor k = 2 giving a confidence level of 95% over a flowrate range of 3 l/s to 1.5 m 3 /s. Flowmeter Installation The Krohne and Bailey-Fischer & Porter meters have to be installed with the electrodes accurately positioned on a horizontal axis. There must be at least 10D of straight pipe upstream of each meter. Sudden changes in pipe section upstream of the meter are to be avoided and any flow control should be downstream of the meter; by at least 3D if a gate valve is used. The flow axis of the Krohne meter must be within 1% of the horizontal while the maximum pipe inclination with the Bailey- Fischer & Porter meter is 5%. The Marsh-McBirney meter can be installed in an inclined pipe (no inclination limits are quoted). The meter is programmed with a correction coefficient appropriate to the particular installation and determined in accordance with the manufacturer s instructions. Figure 2 - Installation of Krohne 200mm TIDALFLUX meter at NEL An example of one of the installations is on Fig 2 which shows the 200 mm Krohne meter installed in the 200 mm test line. Under part-filled flow conditions the test lines were vented to allow the contained air to remain at atmospheric pressure.

4 Performance of Electromagnetic Meters Control of flowrate and meter fill level were achieved by use of gate valves on the facility. Transparent Perspex viewing sections were installed to enable the fill level to be measured by a Vernier height gauge. Full details are given in Reference 5. Tests were conducted initially without controlling the water level at any particular value. The water flowed through the test meter and discharged freely into the end-of-line closed manifold. Under those conditions the maximum flowrate was limited by the resistance of the pipeline. The fill levels were then controlled using a downstream valve so that the test meter performance could be measured over a range of flowrates at each fill level. Three fill levels were selected and in the case of the Bailey-Fischer & Porter meter this allowed a different pair of electrodes to be brought into operation at each level. Krohne meters No satisfactory calibration could be obtained on the first 200 mm meter which was submitted. This was despite reprogramming of the meter by the manufacturer to suit the high conductivity of the water. Krohne offered to submit a new meter. This offer was accepted and the replacement meter performed normally. Results for the 200 mm, meter and the 400 mm one which performed normally, are shown on Fig 4 and 5 respectively. Measurements were also taken with the test meters operating under fully-filled pipe conditions. Results Bailey-Fischer & Porter meters The 200 mm meter did not operate correctly on initial installation nor after replacement of an electronic circuit. The manufacturer suspected the cause to be the high conductivity of the water. The meter was repaired and recalibrated in Germany. On its return to NEL it performed well at high flowrates but badly at low flows. The manufacturer again suspected that high water conductivity was the cause. The 400 mm meter initially gave erroneous results but after adjustment for high conductivity it operated normally. The results are shown on Fig 3. Figure 4 - Results from Krohne 200 mm TIDALFLUX meter Figure 3 - Results from 400 mm Bailey-Fischer & Porter meter Figure 5 - Results from Krohne 400 mm TIDALFLUX meter

5 June 1999 Marsh-McBirney meter Normal performance was obtained from this meter when it was installed in 200 mm and 400 mm pipes using the appropriate supporting band (see 3.4). Results are shown on Fig 6 and 7. Conclusions The NEL large water flow facility was adapted satisfactorily to allow the calibration of flowmeters operating under part-filled conditions. Of the five flowmeters tested in the project, two performed normally without adjustment or repair. The high conductivity of the NEL facility water was suspected by the manufacturers to be a cause of meter malfunctions and while this may have been the case it was not considered to be the only cause. Bailey-Fischer & Porter quotes a required minimum conductivity of 0.05 µs/cm but no maximum value. Krohne literature quotes a minimum value of greater than 50 µs/cm but, again, no maximum value. The Marsh McBirney literature does not appear to specify liquid conductivity requirements. As a result of the project, Bailey-Fischer & Porter and Krohne have both modified their meters to cope with higher conductivities. It would always be important, however, that a flowmeter manufacturer or supplier is fully informed regarding the specific application for a particular meter including the range of conductivities expected. Figure 6 - Results from Marsh-McBirney meter in 200 mm pipe The percentage deviations (see Fig 3-7) computed from the calibration data show that electromagnetic flowmeters for partfilled pipes are a viable alternative to the normal open channel flow structures. The electromagnetic meters also have the distinct advantage of being able to deal with fully-filled pipes and to measure the flow of liquids which would not be suitably transported in an open channel due to their hazardous nature. It is recognised that part-filled pipe flows may also be measured by ultrasonic means but this method was not included within the scope of the project. Figure 7 - Results from Marsh-McBirney meter in 400 mm pipe

6 Performance of Electromagnetic Meters References 1 Literature Survey on Electromagnetic Flowmeters for Use in Partially Filled Pipes (Manufacturers, General Principles of Operation and Methods of Calibration), Project WSDC54, Report No 027/97, National Engineering Laboratory, East Kilbride, November 1997 The work described in this Guidance Note was carried out as part of the Flow Programme under the sponsorship of the DTI s National Measurement System Policy Unit. NEL is grateful to the three companies who agreed to supply flowmeters to NEL for the project and for the assistance and information which they gave during it. 2 Performance Evaluation of a 200 mm Krohne Altometer Electromagnetic Flowmeter Installed in 200 mm Partially Filled Pipes, Project WSDC54, Report No 186/99, National Engineering Laboratory, East Kilbride, April Performance Evaluations of a Flowline (Marsh-McBirney) Electromagnetic Flowmeter Installed in 200 mm and 400 mm Partially Filled Pipelines, Project WSDC54, Report No 190/99, National Engineering Laboratory, East Kilbride, April Performance Evaluation of 200 mm and 400 mm PARTI-MAG II Bailey-Fischer & Porter Electromagnetic Flowmeters Installed in Partially Filled Pipes, Project WSDC54, Report No 195/99, National Engineering Laboratory, East Kilbride, April Evaluation of Electromagnetic Flowmeters Designed for Measuring Liquid Flow in Part-Filled and Fully-Charged Pipes, Project WSDC54, Report No 207/99, National Engineering Laboratory, East Kilbride, April 1999 The purpose of this Guidance Note is to provide, in condensed form, information on measurement methods and technologies. It was produced as part of the UK Government s National Measurement System. For further information, contact: TUV NEL, East Kilbride, GLASGOW, G75 0QF, UK Tel: + 44 (0) info@tuvnel.com TUV NEL Ltd 2010 Re-issued 2010 This publication is to provide outline information only which (unless agreed by TUV NEL in writing) may not be reproduced for any purpose or form part of any order or contract or be regarded as representation relating to products or services concerned.

Long-term Evaluation of Ultrasonic Flowmeters

Long-term Evaluation of Ultrasonic Flowmeters Long-term Evaluation of Ultrasonic Flowmeters A Report for National Measurement System Policy Unit DTI, London Project No: OR7 (OSDC57) Report No: 351/99 Date: 20 September 1999 Flow Centre Long-term Evaluation

More information

DIVA flowmeter. for saturated steam flow measurement

DIVA flowmeter. for saturated steam flow measurement DIVA flowmeter for saturated steam flow measurement DIVA the Direct In-line Variable Area flowmeter... Energy is often treated as an overhead but can represent a considerable part of the cost of production

More information

Erosion Effects in Flowmeters

Erosion Effects in Flowmeters Subsea EXPO 2015 Conference, Aberdeen 11-12th February 2015 Erosion Effects in Flowmeters John Peters, NEL Tariq Boussouara, NEL Slide 1 Presentation Contents Project Objective Flowmeter Details, Test

More information

The Paradigm Case Method of Flowmeter Selection

The Paradigm Case Method of Flowmeter Selection 1 By Jesse Yoder, Overview One of the most interesting developments in the flowmeter market today is the battle between the newer flow technologies and the traditional flowmeters. New-technology flowmeters

More information

FPI Mag Flow Meter. Next Generation Mag Meter: Solving WTP s Tight Installation Dilemma. McCrometer, Inc. (800)

FPI Mag Flow Meter. Next Generation Mag Meter: Solving WTP s Tight Installation Dilemma. McCrometer, Inc.  (800) FPI Mag Flow Meter Next Generation Mag Meter: Solving WTP s Tight Installation Dilemma McCrometer, Inc. www.mccrometer.com/fpimag (800) 220-2279 Lit Number 81000-07 Rev 2.1 / 3-16 TABLE OF CONTENTS: Retrofit

More information

An Intercomparison Between NEL, NRLM, Delft Hydraulics, Alden Research Laboratories, CENAM and NIST Using a 200 mm Twin Orifice Plate Package in Water

An Intercomparison Between NEL, NRLM, Delft Hydraulics, Alden Research Laboratories, CENAM and NIST Using a 200 mm Twin Orifice Plate Package in Water An Intercomparison Between NEL, NRLM, Delft Hydraulics, Alden Research Laboratories, CENAM and NIST Using a 200 mm Twin Orifice Plate Package in Water A Report for NMSPU DTI 151 Buckingham Palace Road

More information

Engineering Standard. Saudi Aramco DeskTop Standards

Engineering Standard. Saudi Aramco DeskTop Standards Engineering Standard 31 December, 2003 Instrumentation Standards Committee Members Al-Awami. L.H., Chairman Tuin, R.R., Vice Chairman Al-Dakhil, T.K. Al-Dhafeeri, F.T. Al-Khalifa, A.H. Al-Madhi, F.A. Alqaffas,

More information

Setting The Standard

Setting The Standard www.iacpe.com Setting The Standard Page : 1 of 91 Rev. 01- IACPE No 19, Jalan Bilal Mahmood 80100 Johor Bahru Malaysia MATERIAL SELECTION CPE LEVEL II The International of is providing the introduction

More information

Sentinel LNG. Panametrics Ultrasonic Flowmeter for Cryogenic Liquids. GE Sensing & Inspection Technologies. Benefits. Applications

Sentinel LNG. Panametrics Ultrasonic Flowmeter for Cryogenic Liquids. GE Sensing & Inspection Technologies. Benefits. Applications GE Sensing & Inspection Technologies Sentinel LNG Panametrics Ultrasonic Flowmeter for Cryogenic Liquids Benefits Improved performance, reduced maintenance and dynamic flow measurement is now available

More information

Venturi Tubes in Wet Gas Improved Models for the Over- Reading and the Pressure-Loss Ratio Method

Venturi Tubes in Wet Gas Improved Models for the Over- Reading and the Pressure-Loss Ratio Method Venturi Tubes in Wet Gas Improved Models for the Over- Reading and the Pressure-Loss Ratio Method Michael Reader-Harris, TUV NEL Emmelyn Graham, TUV NEL 1 INTRODUCTION Venturi tubes are one of the most

More information

PERFORMANCE EVALUATION OF EIGHT ELECTROMAGNETIC FLOWMETERS WITH FINE SAND SLURRIES. N I Heywood and K B Mehta

PERFORMANCE EVALUATION OF EIGHT ELECTROMAGNETIC FLOWMETERS WITH FINE SAND SLURRIES. N I Heywood and K B Mehta Paper presented at 14th International Conference on Slurry Handling and Pipeline Transport (Hydrotransport 14), Maastricht, The Netherlands. 8-10 September 1999, pp 413-431. PERFORMANCE EVALUATION OF EIGHT

More information

Data Sheet. Measuring System trikon Vortex Flow Meter EDZ 420 in Microprocessor Technology. Vortex meter trikon EDZ 420.

Data Sheet. Measuring System trikon Vortex Flow Meter EDZ 420 in Microprocessor Technology. Vortex meter trikon EDZ 420. Data Sheet Measuring System trikon Vortex Flow Meter EDZ 420 in Microprocessor Technology Application Metering and recording of volume or mass flow of gases, liquids and steam. Vortex meter trikon EDZ

More information

Industrial Flowmeters

Industrial Flowmeters spiraxsarco.com/global/us Industrial Flowmeters High-accuracy solutions for steam, gas and liquid flow applications 1 2 F l o w M e a s u r e m e n t S o l u t i o n s spiraxsarco.com/global/us Make Your

More information

DZR Commissioning Set. Flow Data and Installation Instructions

DZR Commissioning Set. Flow Data and Installation Instructions ART & 7 DZR Commissioning Set Flow Data and Installation Instructions This datasheet is designed as a guide and should not be regarded as wholly accurate in every detail.we reserve the right to amend the

More information

Insertion Flow Meters. The only hot tap Full Profile Insertion Flow Meter.

Insertion Flow Meters. The only hot tap Full Profile Insertion Flow Meter. TM FPI Mag TM SPI Mag Insertion Flow Meters The only hot tap Full Profile Insertion Flow Meter. Easy to install, rivals performance Economical for medium and large line sizes The total installed cost of

More information

F Series Flow Meter F300. User Manual

F Series Flow Meter F300. User Manual F Series Flow Meter F300 User Manual Contents 1. F300 Series Flow Meter ------------------------------------- 2 1.1 F300 Series Flow Meter Structure Drawing ---------------------------- 2 1.2 Introduction

More information

Flow Meter Calibration

Flow Meter Calibration fluid mechanics H4 A compact unit that compares and shows the accuracy, losses and use of fundamental flow meters Key Features Unique quick-change flow meter adaptors and pressure connections Four tube

More information

Glossary Irrigation Terminology

Glossary Irrigation Terminology Glossary Irrigation Terminology ANGLE VALVE A valve configured so its outlet is oriented 90 from its inlet. In irrigation, these valves are generally installed with the inlet at the bottom of the valve.

More information

February 5, 2013 Roy Ratcliffe, President RM Controls, Inc.

February 5, 2013 Roy Ratcliffe, President RM Controls, Inc. February 5, 2013 Roy Ratcliffe, President RM Controls, Inc. Volumetric Flow A velocity is a measure of the distance a particle of a substance moves in a given period of time The method of measurement used

More information

Moisture analyzers (moisture meters) FIZEPR-SW100

Moisture analyzers (moisture meters) FIZEPR-SW100 www.fizepr.ru Design Bureau FIZELEKTRONPRIBOR Moisture analyzers (moisture meters) FIZEPR-SW100 for liquid, paste-like and bulk materials including petroleum, fuel oil, sand, coal, ore, concrete mixture,

More information

Utility Flow Metering for Steam and Heated/Chilled Water: A Tutorial. Richard Gruskos Flow Products Manager Kessler Ellis Products

Utility Flow Metering for Steam and Heated/Chilled Water: A Tutorial. Richard Gruskos Flow Products Manager Kessler Ellis Products Utility Flow Metering for Steam and Heated/Chilled Water: A Tutorial Richard Gruskos Flow Products Manager Kessler Ellis Products Review Flow Metering Principles: Flow Meters and Flow Computers Utility

More information

Calculate the Costs of Piping System Elements

Calculate the Costs of Piping System Elements Calculate the Costs of Piping System Elements by Ray Hardee, Engineered Software, Inc. Last month s column described the process of creating an energy cost balance sheet for a piping system (see Figure

More information

ASME PTC Flow Measurement

ASME PTC Flow Measurement ASME PTC 19.5-2004 Flow Measurement Performance Test Codes A N A M E R I C A N N A T I O N A L S T A N D A R D ASME PTC 19.5-2004 Flow Measurement Performance Test Codes AN AMERICAN NATIONAL STANDARD Three

More information

CHAPTER 6 : GAS SAMPLING SYSTEMS

CHAPTER 6 : GAS SAMPLING SYSTEMS CHAPTER 6 : GAS SAMPLING SYSTEMS 1 Scope : 1.1 This Chapter describes two types of gas sampling systems in paragraphs 2.1 and 2.2 meeting the requirements specified in para 4.2 of Chapter 3 of this Part.

More information

Flow Test Data Book and Flow Handbook for 405 Compact Orifice Series and 1595 Conditioning Orifice Plate

Flow Test Data Book and Flow Handbook for 405 Compact Orifice Series and 1595 Conditioning Orifice Plate Test Book and Handbook for 405 Compact Orifice Series and 1595 Conditioning Orifice Plate Title Page 405 Compact Orifice Series and 1595 Conditioning Orifice Plate Test Book NOTICE Read this manual before

More information

SAMPLE SPECIFICATION for CROSSHOLE SONIC LOGGING (CSL) September 2015

SAMPLE SPECIFICATION for CROSSHOLE SONIC LOGGING (CSL) September 2015 SAMPLE SPECIFICATION for CROSSHOLE SONIC LOGGING (CSL) September 2015 Note: This sample specification contains recommended or typical quantities in parenthesis, in the format (quantity); the specifying

More information

Environment Agency Field Test Report TIENet 360 LaserFlow. Environment Agency Field Test Report TIENet 360 LaserFlow

Environment Agency Field Test Report TIENet 360 LaserFlow. Environment Agency Field Test Report TIENet 360 LaserFlow Richard Severn, Environment Agency, May 2014 Teledyne ISCO (USA) recently released the first non-contact open channel sub surface laser flowmeter. RS Hydro (www.rshydro.co.uk) is a UK distributor for the

More information

Gilflo flowmeters. for steam, liquids and gases

Gilflo flowmeters. for steam, liquids and gases Gilflo flowmeters for steam, liquids and gases The Gilflo spring loaded variable area flowmeter... offers an unrivalled 1:1 turndown and can be used to meter most industrial fluids including steam and

More information

Hydrology. Jürg M. Matter Lamont-Doherty Earth Observatory, The Earth Institute at Columbia University. May 9, 2008

Hydrology. Jürg M. Matter Lamont-Doherty Earth Observatory, The Earth Institute at Columbia University. May 9, 2008 Well Logging Principles and Applications Hydrology Jürg M. Matter Lamont-Doherty Earth Observatory, The Earth Institute at Columbia University May 9, 2008 Outline Background Flowmeter Logging Fluid Movement

More information

READ ONLY. Adopting Agency BSC SFM. Adopt Entire Chapter Adopt Entire Chapter as amended (amended sections listed below)

READ ONLY. Adopting Agency BSC SFM. Adopt Entire Chapter Adopt Entire Chapter as amended (amended sections listed below) CALIFORNIA PLUMBING CODE MATRIX ADOPTION TABLE APPENDIX C - ALTERNATE PLUMBING SYSTEMS (Matrix Adoption Tables are non-regulatory, intended only as an aid to the user. See Chapter 1 for state agency authority

More information

PRODUCT CONFORMITY CERTIFICATE

PRODUCT CONFORMITY CERTIFICATE PRODUCT CONFORMITY CERTIFICATE This is to certify that the FLOWSIC100 Gas Flowmeter Manufactured by: SICK Engineering GmbH Bergener Ring 27 01458 Ottendorf-Okrilla Germany Has been assessed by Sira Certification

More information

Thermal Imaging for Discharge and Velocity Measurements in Open Channels. PKI 205C, Omaha, NE ;

Thermal Imaging for Discharge and Velocity Measurements in Open Channels. PKI 205C, Omaha, NE ; Thermal Imaging for Discharge and Velocity Measurements in Open Channels E. Jackson 1, D. Admiraal 2, D. Alexander 3, J. Stansbury 4, J. Guo 5, D. Rundquist 6, and M. Drain 7 1 Dept. of Electrical Engineering,

More information

Implementation and Field Calibration Pipeline Doppler Meters in Northern California Stuart Styles 1, Lynn Groundwater 2, and Jim Weathers 3

Implementation and Field Calibration Pipeline Doppler Meters in Northern California Stuart Styles 1, Lynn Groundwater 2, and Jim Weathers 3 Implementation and Field Calibration Pipeline Doppler Meters in Northern California Stuart Styles 1, Lynn Groundwater 2, and Jim Weathers 3 1 Irrigation Training and Research Center, BioResource and Ag

More information

Water flow measurement in the oil and gas industry

Water flow measurement in the oil and gas industry Water flow measurement in the oil and gas industry The importance of selection Proper selection of the flow technology Why is it important In oil and gas we are seeing a drop in crude prices and the biggest

More information

FPI Mag Flow Meter. Next Generation Mag Meter Measuring Flow in a Pump Station. McCrometer, Inc. (800)

FPI Mag Flow Meter. Next Generation Mag Meter Measuring Flow in a Pump Station. McCrometer, Inc.  (800) FPI Mag Flow Meter Next Generation Mag Meter Measuring Flow in a Pump Station McCrometer, Inc. www.mccrometer.com/fpimag (800)220-2279 Michelle Pawlowicz Marketing Specialist Lit Number 81000-01 Rev 1.0/6-13

More information

Organisation Internationale de Métrologie Légale

Organisation Internationale de Métrologie Légale Organisation Internationale de Métrologie Légale OIML INTERNATIONAL RECOMMENDATION Testing procedures and test report format for pattern evaluation of fuel dispensers for motor vehicles Procédures d essai

More information

FLUXUS F601. Portable, quick and reliable. Clamp-on ultrasonic flow meter. Measure from outside what's flowing inside

FLUXUS F601. Portable, quick and reliable. Clamp-on ultrasonic flow meter. Measure from outside what's flowing inside Reliable flow measurement in less than 5 minutes. Wide application range. With long life marathon battery. Portable, quick and reliable FLUXUS F601 Clamp-on ultrasonic flow meter Accurate Flexible Quick

More information

Mark van der Zande April 2 nd 2015 Sense of Contact 17. Magnetic Resonance Technology A new concept for Multiphase Flow Measurement

Mark van der Zande April 2 nd 2015 Sense of Contact 17. Magnetic Resonance Technology A new concept for Multiphase Flow Measurement Mark van der Zande April 2 nd 2015 Sense of Contact 17 Magnetic Resonance Technology A new concept for Multiphase Flow Measurement 1. KROHNE 2. Multiphase flow in the oil industry 3. Magnetic Resonance

More information

Pulsed Eddy Current Testing (PECT) Inspection Technique

Pulsed Eddy Current Testing (PECT) Inspection Technique Introduction Pulsed Eddy Current Testing (PECT) is an inspection technique used for corrosion under insulation (CUI) screening on carbon steel structures as pipes, vessels, tanks and spherical tank legs

More information

Pulsed Eddy-Current (PEC) Inspection through Insulation

Pulsed Eddy-Current (PEC) Inspection through Insulation INDUSTRIAL SERVICES WHITE PAPER PULSED EDDY CURRENT Pulsed Eddy-Current (PEC) Inspection through Insulation Method statement of an advanced non-destructive testing technique for inspection through insulation.

More information

World s first LNG research and calibration facility

World s first LNG research and calibration facility World s first LNG research and calibration facility P. Lucas 1, O. Büker 2, A. Kenbar 3, H. Kolbjørnsen 4, K. Rasmussen 5, G. Rathwell 6, M. Safonova 7, K. Stolt 2, M. P. van der Beek 1 1 VSL Dutch Metrology

More information

Pressure Recorders: Low-Cost, Robust, Low-Battery-Use Field Measurement Techniques for Trending Water Pressure

Pressure Recorders: Low-Cost, Robust, Low-Battery-Use Field Measurement Techniques for Trending Water Pressure Pressure Recorders: Low-Cost, Robust, Low-Battery-Use Field Measurement Techniques for Trending Water Pressure Stuart Styles 1 and Kerilyn Ambrosini 2 1 Director, Irrigation Training and Research Center,

More information

TVA Target Variable Area Flowmeter

TVA Target Variable Area Flowmeter spiraxsarco.com/global/us TVA Target Variable Area meter for saturated steam 1 2 T V A f l o w m e t e r spiraxsarco.com/global/us Make Your Processes More Productive, Efficient and Sustainable Total Control

More information

KRAL Volumeter - OMG Series Universal Flowmeters

KRAL Volumeter - OMG Series Universal Flowmeters KRAL Volumeter - OMG Series Universal Flowmeters OMG - Universal OMG Flowmeters for a Wide Range of Applications Operating Conditions and Materials Flow Range: 0.03 to 2000 gpm Max. Pressure: 3600 psi

More information

Wastewater Flow Monitoring Services

Wastewater Flow Monitoring Services Wastewater Flow Monitoring Services For San Gabriel, CA July 13, 2015 through July 21, 2015 Revised October 9, 2015 Leaders in Sewer Flow Monitoring Services 601 N. Parkcenter Dr., Suite 209 Santa Ana,

More information

Discharge Measurement in Small Hydropower Stations using Acoustic Doppler Current Profiler

Discharge Measurement in Small Hydropower Stations using Acoustic Doppler Current Profiler Discharge Measurement in Small Hydropower Stations using Acoustic Doppler Current Profiler B.K. Gandhi 1, H.K. Verma 2 and S.S. Patnaik 3 1) Department of Mechanical & Industrial Engineering, Indian Institute

More information

Flanged Water Meters (non-continuous flow)

Flanged Water Meters (non-continuous flow) Flanged Water Meters (non-continuous flow) Features MW-F Suitable for use with domestic water Pulsed output Low head loss Specification Product Codes Meter: Cold water: Fluid temp. range: Cold 0 to 30

More information

introduced before the end of World War II. 2 They are less the focus of new product development than new-technology

introduced before the end of World War II. 2 They are less the focus of new product development than new-technology FLOW update Flow Measurement Part II: Trend Watch A look at recent developments in traditional-technology flowmeters The following is Part 2 in a series on recent developments in flowmeter technology and

More information

Flow Meters Demonstration Unit FMDU

Flow Meters Demonstration Unit FMDU Engineering and Technical Teaching Equipment Flow Meters Demonstration Unit FMDU PROCESS DIAGRAM AND UNIT ELEMENTS ALLOCATION ISO 9001: Quality Management (for Design, Manufacturing, Commercialization

More information

VENTURI FLOWMETERS. Your First Choice in Intelligent Flow Management...

VENTURI FLOWMETERS. Your First Choice in Intelligent Flow Management... VENTURI FLOWMETERS Your First Choice in Intelligent Flow Management... VENTURI FLOWMETERS Preso s complete line of Venturis offer reduced operating costs, proven accuracy, and great rangeability. Low permanent

More information

Vortex Flowmeter FV4000 (TRIO-WIRL V) Swirl Flowmeter FS4000 (TRIO-WIRL S)

Vortex Flowmeter FV4000 (TRIO-WIRL V) Swirl Flowmeter FS4000 (TRIO-WIRL S) Vortex Flowmeter FV4000 (TRIO-WIRL V) Swirl Flowmeter FS4000 (TRIO-WIRL S) For flow rate and volume measurement of liquids, gases and steam Fluid temperatures up to 400 C (752 F) High accuracy Short inlet

More information

ABB Kent-Taylor. Turbine Meter PTF. PTF meter for guaranteed reliability and high accuracy in a turbine meter. Wide flow range up to 40:1

ABB Kent-Taylor. Turbine Meter PTF. PTF meter for guaranteed reliability and high accuracy in a turbine meter. Wide flow range up to 40:1 Turbine Meter PTF Wide flow range up to :1 Intrinsically Safe and Flameproof versions available increases range of applications Excellent linearity and repeatability meets fiscal metering standards Linear

More information

Facilities Plan. Technical Memorandum No. TM-WW-7 Hydraulic Analysis and Effluent Pump Station

Facilities Plan. Technical Memorandum No. TM-WW-7 Hydraulic Analysis and Effluent Pump Station City of St. Joseph, Missouri Hydraulic Analysis and Effluent Pump Station By Work Order No. 09-001 B&V Project 163509 May 20, 2010 Table of Contents 1.0 Executive Summary...1 2.0 Purpose of Study...2 3.0

More information

ISO 3966 INTERNATIONAL STANDARD. Measurement of fluid flow in closed conduits Velocity area method using Pitot static tubes

ISO 3966 INTERNATIONAL STANDARD. Measurement of fluid flow in closed conduits Velocity area method using Pitot static tubes INTERNATIONAL STANDARD ISO 3966 Second edition 2008-07-15 Measurement of fluid flow in closed conduits Velocity area method using Pitot static tubes Mesurage du débit des fluides dans les conduites fermées

More information

Flow Characterization of a Piezo-Electric High Speed Valve

Flow Characterization of a Piezo-Electric High Speed Valve Flow Characterization of a Piezo-Electric High Speed Valve T. Takiya*, N. Fukuda*, M. Yaga** and M. Han*** *( Hitachi Zosen Corporation, Osaka 551-0022, Japan) ** (Department of Mechanical Systems Engineering,

More information

UFGS (August 2004) UNIFIED FACILITIES GUIDE SPECIFICATIONS

UFGS (August 2004) UNIFIED FACILITIES GUIDE SPECIFICATIONS USACE / NAVFAC / AFCEC / NASA UFGS-43 21 29 (April 2006) -------------------------- Preparing Activity: NAVFAC Replacing without change UFGS-13401 (August 2004) UNIFIED FACILITIES GUIDE SPECIFICATIONS

More information

Pulsed Eddy Current Corrosion Monitoring in Refineries and Oil Production Facilities Experience at Shell

Pulsed Eddy Current Corrosion Monitoring in Refineries and Oil Production Facilities Experience at Shell ECNDT 006 - Fr... Pulsed Eddy Current Corrosion Monitoring in Refineries and Oil Production Facilities Experience at Shell Paul CROUZEN and Ian MUNNS, Shell Global Solutions International, Amsterdam, The

More information

Types of Hydropower Facilities

Types of Hydropower Facilities Types of Hydropower Facilities 1 Impoundment Hydropower- uses a dam to store water. Water may be released either to meet changing electricity needs or to maintain a constant water level. 2 Run-of-River

More information

Handling Fluid Fertilizers Selecting the Right Equipment Components

Handling Fluid Fertilizers Selecting the Right Equipment Components Handling Fluid Fertilizers Selecting the Right Equipment Components 9/2/2010 1 Considerations Systematic approach Compatible components Short term vs. long term Plan ahead for future expansion 9/2/2010

More information

Bonded Neo Magnetization Guide

Bonded Neo Magnetization Guide Bonded Neo Magnetization Guide 1 Presentation Outline 1. Magnetizing Systems 2. Fixture Design 3. Construction 4. Testing 5. Design Study 2 Magnetizing Systems A typical magnetizing system consists of

More information

ANALYSIS OF APPLICABILITY OF FLOW AVERAGING PITOT TUBES IN THE AREAS OF FLOW DISTURBANCE

ANALYSIS OF APPLICABILITY OF FLOW AVERAGING PITOT TUBES IN THE AREAS OF FLOW DISTURBANCE Metrol. Meas. Syst., Vol. XXIII (2016), No. 1, pp. 71 84. METROLOGY AND MEASUREMENT SYSTEMS Index 330930, ISSN 0860-8229 www.metrology.pg.gda.pl ANALYSIS OF APPLICABILITY OF FLOW AVERAGING PITOT TUBES

More information

Inspection of water injection pipelines and future needs. By: Mikael Georgson and Roger Hunsbedt. Statoil ASA

Inspection of water injection pipelines and future needs. By: Mikael Georgson and Roger Hunsbedt. Statoil ASA Inspection of water injection pipelines and future needs By: Mikael Georgson and Roger Hunsbedt Statoil ASA Abstract Statoil ASA has recently inspected several water injection pipelines that have shown

More information

Liquid level switch BW 17

Liquid level switch BW 17 KROHNE 09/2001 7.02369.21.00 GR/PRINTO Liquid level switch Status: 10/99 Variable area flowmeters Vortex flowmeters Flow controllers Electromagnetic flowmeters Ultrasonic flowmeters Mass flowmeters Level

More information

Screening, Definition: The unit involved is called a screen.

Screening, Definition: The unit involved is called a screen. Screening, Definition: Screening is a unit operation that separates materials in and/or on water (found in different sizes) from water and from entering water treatment facilities and mains. The unit involved

More information

Dorot Irrigation Meter

Dorot Irrigation Meter DIM-A Irrigation Meter Application The DIM meter is designed for agriculture control applications. The turbine is located at the top of the pipe. Major section of the pipe is free for water flow. Features

More information

Understanding. Hydraulics. palgrave macmillan. Les Hamill. University of Plymouth THIRD EDITION

Understanding. Hydraulics. palgrave macmillan. Les Hamill. University of Plymouth THIRD EDITION Understanding Hydraulics THIRD EDITION Les Hamill Senior Lecturer in Civil Engineering, School of Marine Science and Engineering, University of Plymouth palgrave macmillan Preface to third edition Acknowledgements

More information

Fluid Flow Conditioning. August 2015

Fluid Flow Conditioning. August 2015 Fluid Flow Conditioning August 2015 Flow Conditioning There is no flow meter on the market that needs flow conditioning. All flow meters are effective without any type of flow conditioning. Flow Conditioning

More information

Vortex Mass Flowmeter

Vortex Mass Flowmeter Technology Solutions TE-VOR 1300C Vortex Mass Flowmeter FOW www.tek-trol.com Flow evel Temperature Pressure Valves Analyzers Accessories TekValSys Introduction A Tek-Vor 1300C Vortex Flowmeter (also called

More information

Intercomparison of Multiphase Flow Facilities: Comparison of NEL and Humble, USA (Texaco)

Intercomparison of Multiphase Flow Facilities: Comparison of NEL and Humble, USA (Texaco) Intercomparison of Multiphase Flow Facilities: Comparison of NEL and Humble, USA (Texaco) A Report for National Measurement System Policy Unit DTI, London Report No: 144/99 Date: 10 March 1999 Multiphase

More information

Water Measurement. Presentation by. L. Niel Allen Extension Irrigation Specialist Utah State University

Water Measurement. Presentation by. L. Niel Allen Extension Irrigation Specialist Utah State University Water Measurement Presentation by L. Niel Allen Extension Irrigation Specialist Utah State University to Utah State Engineer s Office Division of Water Rights Salt Lake City, Utah December 11, 2013 Introduction

More information

CHAPTER 10 PRELIMINARY TREATMENT

CHAPTER 10 PRELIMINARY TREATMENT CHAPTER 10 PRELIMINARY TREATMENT TM 5-814-3/AFM 88-11, Volume III 10-1. General considerations. Preliminary treatment of wastewater includes screening, grinding, grit removal, flotation, equilization,

More information

Instrumentation and Process Control

Instrumentation and Process Control Instrumentation and Process Control Topics Covered Critical Monitoring Leachate collection system performance Moisture measurement In-place specific weight Critical Monitoring - Gas Production Rate Gas

More information

Pumping System Vibration

Pumping System Vibration Pumping System Vibration Understanding the Fundamentals Eric J. Olson Mechanical Solutions, Inc. (MSI) ejo@mechsol.com Agenda 1. Vibration testing and analysis fundamentals 2. Structural and acoustic resonance

More information

OPTIMASS 7400 Technical Datasheet

OPTIMASS 7400 Technical Datasheet OPTIMASS 7400 Technical Datasheet Sensor for mass flow The optimum meter for demanding applications A single straight measuring tube Choice of 4 tube materials The documentation is only complete when used

More information

TYPICAL INSTALLATION INSTRUCTIONS

TYPICAL INSTALLATION INSTRUCTIONS TYPICAL INSTALLATION INSTRUCTIONS Environment One Grinder Pump Feature Identification 1. Grinder Pump Basin High density polyethylene (HDPE) 2. Accessway Cover Painted Steel 3. Electrical Quick Disconnect

More information

A REVIEW OF STEAM FLOWMETERING TECHNOLOGY

A REVIEW OF STEAM FLOWMETERING TECHNOLOGY A REVIEW OF STEAM FLOWMETERING TECHNOLOGY A Report for National Measurement System Directorate Department of Trade & Industry 151 Buckingham Palace Road London, SW1W 9SS Project No:FEUS04 Date: October

More information

Lifting Appliances of Ships and Offshore Installations

Lifting Appliances of Ships and Offshore Installations CCS Rule Change Notice For: Lifting Appliances of Ships and Offshore Installations Version: May, 2016. RCN No.1 Effective date: 19 May, 2016 Beijing CONTENTS Chapter 3 CRANES, LIFTS, RAMPS AND SELF-LOADING

More information

Flare Measurement Advanced Ultrasonics

Flare Measurement Advanced Ultrasonics Flare Measurement Advanced Ultrasonics John Chitty Regional Sales Manager SICK MAIHAK, Inc. 15415 International Plaza Drive, Suite 100 Houston, Texas 77032 1.0 Introduction There has been an increased

More information

Irrigation System. BWCDD Zanjero Training 2/13/2008

Irrigation System. BWCDD Zanjero Training 2/13/2008 Irrigation System BWCDD Zanjero Training Session #7 2/13/2008 Irrigation System The (main) intake structure t directs water from the source of supply, such as a reservoir or a river, into the irrigation

More information

195LiquidsCatalog.

195LiquidsCatalog. 195LiquidsCatalog www.energyu.org www.midwestenergy.org 195 Abnormal Operating Conditions... 3 195-0060 General Abnormal Operating Conditions (CT60)... 3 195-0000 Documentation & Reporting... 3 195-0403

More information

Performance Testing of The Robert Moses Niagara Power Plant and Sir Adam Beck Generating Station

Performance Testing of The Robert Moses Niagara Power Plant and Sir Adam Beck Generating Station Performance Testing of The Robert Moses Niagara Power Plant and Sir Adam Beck Generating Station By Albert F. Mikhail, P.Eng., Ontario Power Generation and Robert J. Knowlton, P.E., New York Power Authority

More information

The Application of Ultrasonic Meters in Multiphase Flow

The Application of Ultrasonic Meters in Multiphase Flow The Application of Ultrasonic Meters in Multiphase Flow A Report for NMSPU Department of Trade & Industry 151 Buckingham Palace Road London, SW1W 9SS Report No: 35/99 Date: 2 September 1999 Flow Centre

More information

Rosemount 405 Compact Orifice Primary Element

Rosemount 405 Compact Orifice Primary Element Rosemount 405 Compact Orifice Primary Element SPECIFICATIONS Performance Discharge Coefficient Uncertainty TABLE 20. 405 Compact Orifice Primary Element Type 405C (Conditioning Compact) 405P (1)(2) (Standard

More information

Fast Corrosion Screening Technique Saturation Low Frequency Eddy Current

Fast Corrosion Screening Technique Saturation Low Frequency Eddy Current Summary The inspection of ferromagnetic and non-ferromagnetic material components is an attractive and efficient method due to its fast screening of large areas with low surface preparations. Already qualified

More information

VERIS Accelabar. Superior Flow Measurement Accuracy

VERIS Accelabar. Superior Flow Measurement Accuracy VERIS Superior Flow Measurement Accuracy Designs, materials, weights and performance ratings are approximate and subject to change without notice. Visit for up-to-date information. ...A New Idea in Flow

More information

TSK S 3. Flap Flowmeter for liquids. Range: m 3 /h. Accuracy: ± 2.0 of reading. : PN 40, t max: C. p max

TSK S 3. Flap Flowmeter for liquids. Range: m 3 /h. Accuracy: ± 2.0 of reading. : PN 40, t max: C. p max Flap Flowmeter for liquids measuring monitoring analysing TSK Range: 0.5-3.5 200-1500 m 3 /h Accuracy: ± 2.0 of reading p max : PN 40, t max: -40 +300 C Connection: wafer flange DN 25 500 Material: stainless

More information

Technical Specifications and General Guidelines for Mass Flow Meters at Petroleum Oil Installations. Note:

Technical Specifications and General Guidelines for Mass Flow Meters at Petroleum Oil Installations. Note: Technical Specifications and General Guidelines for Mass Flow Meters at Petroleum Oil Installations Note: These are general specifications and guidelines for Mass Flow Meters for Petroleum Oil Installations.

More information

FLUXUS F/G601 and F/G608

FLUXUS F/G601 and F/G608 Portable, quick and reliable FLUXUS F/G01 and F/G0 Portable ultrasonic clamp-on gas and liquid flow meters Accurate Flexible Quick Rugged Ergonomic When measuring matters FLUXUS F/G0X Mobile flow measurement

More information

Design and Construction of a Mini Hydro Turbine Model

Design and Construction of a Mini Hydro Turbine Model American Journal of Modern Energy 2018; 4(1): 1-6 http://www.sciencepublishinggroup.com/j/ajme doi: 10.11648/j.ajme.20180401.11 ISSN: 2575-3908 (Print); ISSN: 2575-3797 (Online) Design and Construction

More information

HELIFLU TM TZN. The Dedicated Turbine Flowmeter for Custody Transfer Measurement. Proven Performance. Main Applications

HELIFLU TM TZN. The Dedicated Turbine Flowmeter for Custody Transfer Measurement. Proven Performance. Main Applications HELIFLU TM The Dedicated Turbine Flowmeter for Custody Transfer Measurement Proven Performance Main Applications Oil Production Allocation FSO, FPSO Metering Tanker Loading/Offloading Storage Management

More information

Index. Page numbers followed by f indicate figures.

Index. Page numbers followed by f indicate figures. Index Aerodynamic method, 103, 110 111 Algae, 131, 173, 175 Alternate depth, 88 Alternating block method, 132, 140 141 Attenuation, 106, 107f, 118, 120 Page numbers followed by f indicate figures. Baseflow

More information

Soil moisture measurements

Soil moisture measurements Soil moisture measurements Climatological and hydrological field work - FS2017 1 Introduction Soil moisture is a key variable in the climate system. By controlling evapotranspiration, soil moisture impacts

More information

Sanitary and Environmental Engineering I (4 th Year Civil)

Sanitary and Environmental Engineering I (4 th Year Civil) Sanitary and Environmental Engineering I (4 th Year Civil) Prepared by Dr.Khaled Zaher Assistant Professor, Public Works Engineering Department, Faculty of Engineering, Cairo University Wastewater Flow

More information

Offshore Measurement System for Relative Displacements of Grouted Joints

Offshore Measurement System for Relative Displacements of Grouted Joints Civil Structural Health Monitoring Workshop (CSHM-4) - Poster 08 Offshore Measurement System for Relative Displacements of Grouted Joints Niklas SCHOLLE *, Ludger LOHAUS * * Institute of Building Materials

More information

HYDRAULIC DEMONSTRATION CHANNEL

HYDRAULIC DEMONSTRATION CHANNEL HYDRAULIC DEMONSTRATION CHANNEL PACKING LIST Pitot-Static Tube and Manometer Tube Flow Meters 1. 48 Long Differential Manometer 2. Clear Tubing 3. Pitot-static Tube Clamp 4. Pitot-static Tube 1. Venturi

More information

The University of Liège Wind Tunnel Facility

The University of Liège Wind Tunnel Facility The University of Liège Wind Tunnel Facility Technical description The University of Liège Wind Tunnel Facility Technical description December 2009-1/27 The University of Liège Wind Tunnel Facility Technical

More information

tpb A Multiphase Flow Meter for the On-line Determination of the Flow Rates of Oil, Water and Gas

tpb A Multiphase Flow Meter for the On-line Determination of the Flow Rates of Oil, Water and Gas tpb AU9817323 A Multiphase Flow Meter for the On-line Determination of the Flow Rates of Oil, Water and Gas G. J. ROACH and J. S. WATT CSIRO Minerals, Private Mail Bag 5, Menai, NSW 2232, Australia SUMMARY.

More information

heliflu tm The Dedicated Turbine Flowmeter for Custody Transfer Measurement Proven Performance Main Applications

heliflu tm The Dedicated Turbine Flowmeter for Custody Transfer Measurement Proven Performance Main Applications heliflu tm tzn The Dedicated Turbine Flowmeter for Custody Transfer Measurement Proven Performance Main Applications Oil Production Allocation FSO, FPSO Metering Tanker Loading/Offloading Storage Management

More information

LNG Flow Calibration

LNG Flow Calibration LNG Flow Calibration Oswin Kerkhof OKerkhof@vsl.nl Gertjan Kok, Peter Lucas, Mijndert van der Beek, Erik Smits, Andrea Peruzzi, Gerard Blom, Inge van Andel, Jan van Geel Anders Andersson, Krister Stolt

More information

Impact of Pitot tube calibration on the uncertainty of water flow rate measurement

Impact of Pitot tube calibration on the uncertainty of water flow rate measurement Journal of Physics: Conference Series PAPER OPEN ACCESS Impact of Pitot tube calibration on the uncertainty of water flow rate measurement To cite this article: Icaro de Oliveira Buscarini et al 2015 J.

More information