Diffusion/Anneal SOP Page 1 of 12 Revision
|
|
- Leona Lawrence
- 6 years ago
- Views:
Transcription
1 Diffusion/Anneal SOP Diffusion/Anneal SOP Page 1 of 12 1 Scope 1.1 This document provides the procedures and requirements to introduce and/or drive dopants into silicon using the Canary furnace. 2 Table of Contents 1 Scope Table of Contents Reference Documents Referenced within this Document External Documents Equipment and/or Materials Safety Setup Procedures Enable Furnace Remove Boat Select Program and Temperature Set Gas Flows Load Program Set Diffusion/Anneal Time Load Sample Wafers General Procedures Load Furnace Start the Program Unload Process Measurement Place System in Standby Disable Furnace Process Notes and Characterization Solid Source Wafer Cure Gas Flow Setpoint Conversions Auto-Sequencer Input Assignments Temperature Controller Recipes Auto-Sequencer Programs Revision History Figure 1, Diffusion/Anneal Gas Controller... 4 Figure 2, Anneal Boat Loading Diagram... 5 Figure 3, Diffusion Boat Loading Diagram... 5 Figure 4, Push Rod Alignment... 6 Table 1, Diffusion/Anneal Programs... 3 Table 2, Solid Source Cure Loop Counter... 8 Table 3, Diffusion/Anneal Gas Flow Setpoint Conversions... 8
2 Diffusion/Anneal SOP Page 2 of 12 Table 4, Diffusion/Anneal Auto-Sequencer Output Assignments... 8 Table 5, Diffusion/Anneal Temperature Controller Setup... 9 Table 6, Diffusion/Anneal Auto-Sequence Controller Setup Reference Documents 3.1 Referenced within this Document Canary Furnace SOP 3.2 External Documents None 4 Equipment and/or Materials 4.1 Oxygen (O 2 ) 4.2 Hydrogen (H 2 ) 4.3 House Nitrogen (N 2 ) 4.4 Canary Diffusion Furnace 4.5 Quartz boats 4.6 Quartz push rod with mark at 26 from hook 4.7 Solid Source Wafers, BoronPlus GS245 (Techneglas) 4.8 Solid Source Wafers, PhosPlus TP470 (Techneglas) 4.9 Heat-resistant gloves 4.10 Monitor wafers (bare silicon <100>, n and p-type) 5 Safety 5.1 Follow all Nanofab safety procedures. 5.2 Do NOT open the element door (danger of high voltage). 5.3 This system uses hazardous gases. Refer to MSDS for more detail Do NOT adjust any regulators or valves for any gases. 5.4 Do NOT store flammable solvents (e.g., squeeze bottles, etc.) near the furnace or loading station. 5.5 Do NOT touch quartz boats with your hands. They may be hot. Use only a clean quartz boat holder or high temperature gloves. 5.6 Do NOT touch quartz or wafers with wipes, gloves, teflon, plastic, etc. 5.7 Place quartz only on dedicated quartz carriers, quartz surfaces, or stainless steel surfaces. 6 Setup Procedures All new, unused solid source wafers (BoronPlus or PhosPlus) must be cured before use. Follow the procedures in section 8.1 Solid Source Wafer Cure to cure the wafers.
3 6.1 Enable Furnace Diffusion/Anneal SOP Page 3 of In Coral, enable the furnace and complete all required data entry. 6.2 Remove Boat Enabling the furnace will release the interlock and provide access to the keypad Verify the furnace is in step If the furnace is NOT in step 69, contact Lab Staff before proceeding Open the furnace door Using the high temperature gloves, remove the end cap Using the high temperature gloves or the quartz handle, remove the boat from the furnace and place on the quartzware rack Using the high temperature gloves, replace the end cap Close the furnace door Allow the boat and wafers to cool at least 5 minutes. 6.3 Select Program and Temperature Review the pre-defined programs shown in Table 1, Diffusion/Anneal Programs. Table 1, Diffusion/Anneal Programs Loop Counter Step for Diff/Anneal Setpoint (V) Approx. Run Time (min) Program Description Time N2 O2 Pre Diff/Anneal Post 0 Dope without LTO (1050 C) n/a 50 User defined 80 1 Dope with LTO (1050 C) User defined 92 2 Custom Dope Program n/a n/a n/a 3 Anneal (1050 C) User defined Custom Anneal Program n/a n/a n/a 5 Solid Source Cure (1050 C) BoronPlus: 799 (13:19) PhosPlus: 1279 (21:19) Identify the desired program If the desired conditions (temperature and gas flows) are NOT shown in Table 1, Diffusion/Anneal Programs, contact the Lab Staff before proceeding. The Lab Staff can create custom programs, as needed. 6.4 Set Gas Flows Using Table 1, Diffusion/Anneal Programs, find the voltage setpoints for each of the gases On the Anneal Dry Oxidation controller located in the gas jungle (see Figure 1, Anneal Dry Oxidation Gas Controller), rotate the selection knob on the gas controller to select the desired gas.
4 6.4.3 While holding the toggle switch on SET, adjust the setpoint dial until the MFC Voltage Display shows the correct voltage value Repeat through for each process gas. 6.5 Load Program Following the procedures in Canary Furnace SOP, load the desired program. The controller will change to the first step of the program and immediately go on hold with 10 seconds on the time display. The READY light will be on Verify the number of the desired program is shown in the Program Display field of the Auto-Sequence Controller If the correct number is NOT displayed, repeat through Diffusion/Anneal SOP Page 4 of 12 Setpoint Dials Set/Read Switch Selection Knob Following the procedures in Canary Furnace SOP, release the hold. MFC Voltage Display Figure 1, Anneal Dry Oxidation Gas Controller The RUN light will be on. The program will count down to 0 on the time display and go on hold. The READY light will be on again. 6.6 Set Diffusion/Anneal Time Following the procedures in Canary Furnace SOP, set the value of the applicable loop counter step to the desired number. (Refer to Table 1, Diffusion/Anneal Programs to find the correct loop counter step for the program being used.) One loop is one minute of diffusion/anneal. For example, to get 45 minutes of diffusion/anneal, set the loop counter to If you need a diffusion/anneal step longer than 600 minutes, contact lab staff for approval before proceeding. 6.7 Load Sample Wafers Ensure all sample wafers are completely clean, dry, and free of any surface contamination.!! WARNING!! No photoresist, kapton tape, metals, or other foreign material are allowed in the furnace. Only quartz, silicon, oxide, nitride, and/or Polysilicon are allowed without lab staff approval.!! WARNING!! Using the vacuum wand, load the wafers as shown in Figure 2, Anneal Boat Loading Diagram or Figure 3, Diffusion Boat Loading Diagram.
5 Diffusion/Anneal SOP Page 5 of 12 Monitor Wafer Sample Wafer \ \ / / Test Run Polished Side Handle \ \ / / With Sample Wafers Figure 2, Anneal Boat Loading Diagram Monitor Wafer Dummy or Sample Wafer Solid Source Wafer Handle \ \ / / Figure 3, Diffusion Boat Loading Diagram Include a monitor wafer at the center of the boat Boron doping and anneal require an n-type wafer Phosphorous doping requires a p-type wafer Monitor wafers are provided in a labeled box located on the storage rack For Anneal: Wafer spacing may be every slot The polished side of all wafers must face the handle end of the boat (door of the furnace) For Diffusion: 7 General Procedures For Diffusion, the solid source wafers should always have a dummy wafer or sample wafer in adjacent slots even when stored. Remove the necessary dummy wafers (adjacent to the solid source wafers) and place them in empty slots at either end of the boat. Load sample and monitor wafers in the slot next to a solid source The polished side of all wafers must face the solid source 7.1 Load Furnace Open the furnace door Using the high temperature gloves, remove the end cap Place the boat at the end of the furnace tube Using the quartz push rod, push the boat into the furnace at 1 inch every 5 seconds (12 inches per minute) Align the mark on the push rod to the front face of the furnace scavenger (see Figure 4, Push Rod Alignment).
6 7.1.6 Remove the push rod and place in the storage cylinder Using the high temperature gloves, replace the end cap Close the furnace door. 7.2 Start the Program Following the procedures in Canary Furnace SOP, release the hold. The READY light will turn off and the RUN light will turn on. The controller will automatically execute the necessary steps to complete the diffusion/anneal process. A summary of the programs, steps, and outputs is given in Table 6, Diffusion/Anneal Auto-Sequence Controller Setup Wait until the controller is in step Unload Verify the program is complete. The controller will be on step Open the furnace door Using the high temperature gloves, remove the end cap. Diffusion/Anneal SOP Page 6 of Using the quartz push rod, pull the boat to the door of the furnace at 1 inch every 5 seconds (12 inches per minute) Place the push rod in the storage cylinder Allow the boat to cool a minimum of 2 minutes at the door of the furnace Using the high temperature gloves or the quartz handle, remove the boat from the furnace and place on the quartzware rack Using the high temperature gloves, replace the end cap Close the furnace door Allow the boat and wafers to cool at least 5 minutes Using the vacuum wand, remove the samples and monitor wafers from the quartz boat. 7.4 Process Measurement For Anneal, measure the oxide thickness of the monitor wafer. Figure 4, Push Rod Alignment Push Rod Mark Align Mark to Scavenger Face Scavenger Face For Diffusion without LTO, measure the oxide thickness and the resistivity of the monitor wafer.
7 Diffusion/Anneal SOP Page 7 of For Diffusion with LTO, do the following with the monitor wafer: Measure the oxide thickness Remove the oxide Measure the resistivity Place used monitor wafers in the labeled storage box located on the storage rack. 7.5 Place System in Standby This section is only necessary if the furnace will be idle for more than 1 hour Verify the furnace is in step If the furnace is NOT in step 69, contact Lab Staff before proceeding On the quartz boat, ensure each solid source wafer has a dummy wafer in the adjacent slots Open the furnace door Using the high temperature gloves, remove the end cap Place the boat at the end of the furnace tube Using the high temperature gloves, replace the end cap Close the furnace door Using the procedures in 6.4 Set Gas Flows, set the N2 flow at 1V (2 slpm). 7.6 Disable Furnace In Coral, disable the furnace and complete all data entry, including process measurement data. Disabling the furnace will activate the interlock and prevent access to the keypad. 8 Process Notes and Characterization 8.1 Solid Source Wafer Cure Load the solid source wafers into consecutive large slots in the quartz boat Load bare silicon wafers into all slots adjacent to the solid source wafers The polished side of the wafers must face the solid source wafer Set the gas flows per section 6.4 Set Gas Flows Set the cure time as per Table 2, Solid Source Cure Loop Counter and Canary Furnace SOP. One loop is equivalent to 10 minutes.
8 8.1.5 Run the cure recipe per sections Diffusion/Anneal SOP Page 8 of Gas Flow Setpoint Conversions Table 2, Solid Source Cure Loop Counter Source Wafer Type Loop Counter Value BoronPlus 60 PhosPlus Table 3, Diffusion/Anneal Gas Flow Setpoint Conversions shows the gas flow setpoints and corresponding voltage values for the MFC Voltage Display. 8.3 Auto-Sequencer Input Assignments Table 3, Diffusion/Anneal Gas Flow Setpoint Conversions Setpoint Display Setpoint Display N2 (slpm) O2 (slpm) Volts (V) N2 (slpm) O2 (slpm) Volts (V) Max Flow N 2 or O 2 : V = slpm/ Table 4, Diffusion/Anneal Auto-Sequencer Output Assignments shows the assignments of the input connections for the auto-sequencer. Table 4, Diffusion/Anneal Auto-Sequencer Output Assignments Output Description Output Description 1 N2 Upstream Valve 9 Temperature Control Bit1 2 O2 Upstream Valve 10 Temperature Control Bit2 3 Not used 11 Temperature Control Bit3 4 Not used 12 Temperature Control Bit4 5 Not used 13 Not used 6 Not used 14 Not used 7 Not used 15 Not used 8 Not used 16 Not used
9 8.4 Temperature Controller Recipes Diffusion/Anneal SOP Page 9 of Table 5, Diffusion/Anneal Temperature Controller Setup shows the setup of the temperature controller. Table 5, Diffusion/Anneal Temperature Controller Setup Recipe Setpoint Auto-Sequencer Outputs variable 1 variable variable 12 9 variable n/a n/a n/a n/a n/a n/a *Maximum Allowed Temperature: 1050 C Ramp Rate (Slope) for all recipes and zones: 7.00
10 Diffusion/Anneal SOP Page 10 of Auto-Sequencer Programs Table 6, Diffusion/Anneal Auto-Sequence Controller Setup shows the defined program steps. Table 6, Diffusion/Anneal Auto-Sequence Controller Setup Output LED s Program Step Instruction Program Alarm Change Branch Temp # Function # Description Start Enable Enable Step ( C) Time Push Load 6 HOLD AT END YES Ramp Up and Stabilize 4 HOLD WITH OUTPUTS none 0 No operation none 0 No operation Start Loop 1 SET LOOP Counter YES *** 0 5 Doping Cycle 4 HOLD WITH OUTPUTS End Loop 2 LOOP BACK Ramp Down 7 BRANCH AT END none 0 No operation none 0 No operation Push Load 6 HOLD AT END YES Ramp Up and Stabilize 4 HOLD WITH OUTPUTS none 0 No operation none 0 No operation Start Loop 1 SET LOOP Counter YES *** 1 15 Doping Cycle 4 HOLD WITH OUTPUTS End Loop 2 LOOP BACK Ramp Down/LTO 7 BRANCH AT END none 0 No operation none 0 No operation none 0 No operation none 0 No operation none 0 No operation none 0 No operation none 0 No operation none 0 No operation none 0 No operation none 0 No operation none 0 No operation none 0 No operation Push Load 6 HOLD AT END YES Ramp Up and Stabilize 4 HOLD WITH OUTPUTS Skin Oxidation 4 HOLD WITH OUTPUTS Ramp Up and Stabilize 4 HOLD WITH OUTPUTS Start Loop 1 SET LOOP Counter YES *** 3 35 Anneal 4 HOLD WITH OUTPUTS End Loop 2 LOOP BACK Ramp Down 7 BRANCH AT END none 0 No operation none 0 No operation 0.0
11 Diffusion/Anneal SOP Page 11 of 12 Table 6, Diffusion/Anneal Auto-Sequence Controller Setup Output LED s Program Step Instruction Program Alarm Change Branch Temp # Function # Description Start Enable Enable Step ( C) Time none 0 No operation none 0 No operation none 0 No operation none 0 No operation none 0 No operation none 0 No operation none 0 No operation none 0 No operation none 0 No operation none 0 No operation Push Load 6 HOLD AT END YES Ramp Up and Stabilize 4 HOLD WITH OUTPUTS none 0 No operation none 0 No operation Start Loop 1 SET LOOP Counter YES *** 5 55 Cure Sources 4 HOLD WITH OUTPUTS End Loop 2 LOOP BACK Ramp Down 7 BRANCH AT END none 0 No operation none 0 No operation none 0 No operation none 0 No operation none 0 No operation none 0 No operation Power out Program 0 6 HOLD AT END YES Power out Program 1 6 HOLD AT END YES Power out Program 2 6 HOLD AT END YES Power out Program HOLD AT END YES Abort 6 HOLD AT END YES Idle 6 HOLD AT END *** This value is adjusted as necessary between Time values with a decimal are in minutes; values without a decimal are in seconds.
12 Diffusion/Anneal SOP Page 12 of 12 9 Revision History Rev Date Orginator Description of Changes 5 12 May 2011 T. Olsen Update gas flow setting instructions (section 6.4) to reflect the replacement of portable multi-meters with a permanently mounted voltage display, including Figure 1 and Table Mar 2011 T. Olsen Replace references to log sheets with Enable/Disable in Coral and required data entry. Clarify use and availability of monitor wafers Jul 2010 T. Olsen Change length of push rod mark from 30 to 26. Update Table 2, Diffusion/Anneal Programs to include gas flows (move table to Table 1). Update Loading Diagram and instructions to keep dummy wafers next to solid source wafers. Updated Table 1, Diffusion/Anneal Programs and Table 6, Diffusion/Anneal Auto-Sequence Controller Setup to remove unused programs. Update Table 5 due to nonfunctioning outputs 9 and Mar 2010 T. Olsen Changed time of first step of each program from 0.1 minutes to 10 seconds. Updated/clarified procedures for program processing and loop counter usage. Added Handle and Polished Side to anneal wafer loading diagram. Added Pre and Post times to Table Feb 2010 T. Olsen Initial Release.
CTR LPCVD SOP Page 1 of 17 Revision
CTR LPCVD OP Page 1 of 17 CTR LPCVD OP 1 cope 1.1 This OP provides instructions to operate the CTR LPCVD urnaces. 2 Table of Contents 1 cope... 1 2 Table of Contents... 1 3 Reference Documents... 2 3.1
More informationReport 1. B. Starting Wafer Specs Number: 10 Total, 6 Device and 4 Test wafers
Aaron Pederson EE 432 Lab Dr. Meng Lu netid: abp250 Lab instructor: Yunfei Zhao Report 1 A. Overview The goal of this lab is to go through the semiconductor fabrication process from start to finish. This
More informationPOLYIMIDE BAKE PROCESSING SYSTEM YES-550PB6/8/12-2P-CP
POLYIMIDE BAKE PROCESSING SYSTEM YES-550PB6/8/12-2P-CP June 2013 The YES high temperature cure oven is designed to provide a controlled ramp curing process for temperatures up to 550 C in an oxygen-free
More informationLab 1: Field Oxide. Overview. Starting Wafers
Overview Lab 1: Field Oxide Brandon Baxter, Robert Buckley, Tara Mina, Quentin Vingerhoets Lab instructor: Liang Zhang Course Instructor: Dr. Gary Tuttle EE 432-532 January 23, 2017 In this lab we created
More informationFiji Thermal and Plasma Atomic Layer Deposition System (ALD) By Ultratech (Cambridge)
Fiji Thermal and Plasma Atomic Layer Deposition System (ALD) By Ultratech (Cambridge) PREPARED BY: You-Sheng (Wilson) Lin, Nanolab Staff 7-30-2013 Superusers: Steve Franz You-Sheng Lin Max Ho X68923 X68923
More informationUniversity of Minnesota Nano Center Standard Operating Procedure
Equipment Name: University of Minnesota Nano Center PECVD Coral Name: pecvd Revision Number: 2.0 Model: Plasmatherm 340 Revisionist: Mark Fisher Location: Bay 3 Date: 20 Sept 2013 1 Description The Plasmatherm
More informationTemperature Scales. Questions. Temperature Conversions 7/21/2010. EE580 Solar Cells Todd J. Kaiser. Thermally Activated Processes
7/1/010 EE80 Solar Cells Todd J. Kaiser Flow of Wafer in Fabrication Lecture 0 Microfabrication A combination of Applied Chemistry, Physics and ptics Thermal Processes Diffusion & xidation Photolithograpy
More informationMeasurement of Thermal Diffusivity Using a Laser Flash Thermal Property Analyzer
INL/EXT-08-14862 Measurement of Thermal Diffusivity Using a Laser Flash Thermal Property Analyzer J. E. Daw November 2008 The INL is a U.S. Department of Energy National Laboratory operated by Battelle
More informationPlasma-Enhanced Chemical Vapor Deposition
Plasma-Enhanced Chemical Vapor Deposition Steven Glenn July 8, 2009 Thin Films Lab 4 ABSTRACT The objective of this lab was to explore lab and the Applied Materials P5000 from a different point of view.
More informationLab #2 Wafer Cleaning (RCA cleaning)
Lab #2 Wafer Cleaning (RCA cleaning) RCA Cleaning System Used: Wet Bench 1, Bay1, Nanofabrication Center Chemicals Used: H 2 O : NH 4 OH : H 2 O 2 (5 : 1 : 1) H 2 O : HF (10 : 1) H 2 O : HCl : H 2 O 2
More informationCMOS LAB MANUAL CMOS LAB MANUAL
CMOS LAB MANUAL 2011 CMOS LAB MANUAL This manual was designed for use with the Montana Microfabrication Facility at MSU. The intention of the manual is to provide lab users and MSU students with a complete
More informationNon-contractual document, specifications subject to change without notice.
1 ANNEALSYS designs and manufactures Rapid Thermal Processing (RTA, RTCVD) Direct Liquid Injection (DLI-CVD, DLI-ALD) systems for research laboratories and companies for semiconductor, MEMS, nanotechnologies,
More informationSilicon Wafer Processing PAKAGING AND TEST
Silicon Wafer Processing PAKAGING AND TEST Parametrical test using test structures regularly distributed in the wafer Wafer die test marking defective dies dies separation die fixing (not marked as defective)
More informationNEVER REMOVE THE SAFETY GUARDS
Page 1 of 5 Name Recommended Use Designed for cutting stumps up to 200 mm above ground and 120 mm below ground. Maximum Stump diameter 300 mm. NOT suitable for stumps surrounded by rocks, stones, concrete
More informationHIGH TEMPERATURE FURNACES. High Temperature Chamber Series High Temperature Tube Series High Temperature Elevator Series
HIGH TEMPERATURE FURNACES High Chamber Series High Tube Series High Elevator Series Chamber Furnaces MoS Series MoS series have been used as trusted professional chamber furnaces for many years in laboratories.
More informationAutomatic hydraulic laboratory presses in two- and four-post models up to 30 ton clamp capacity
Auto Series compact automatic Automatic hydraulic laboratory presses in two- and four-post models up to 30 ton clamp capacity Access Door Clear, impact resistant polycarbonate door permits observation
More informationInstallation, Operation & Maintenance Manual
Original Instructions Installation, Operation & Maintenance Manual Sentry IC Indexing Cabinet Automatic Sampling Accessories S-AS-IOM-00462-2 11-17 Do not install, maintain, or operate this equipment without
More informationNext Review Date: Oct 2017 SWP Reference Number: Version: V3 Version Issue Date: Oct 2016
Faculty/School: Faculty of Pharmacy Next Review Date: Oct 2017 SWP Reference Number: Version: V3 Version Issue Date: Oct 2016 SWP Title: Operation of the autoclaves Prepared by: Padmaja Dhanvate and Dr
More informationELECTRICAL RESISTIVITY AS A FUNCTION OF TEMPERATURE
ELECTRICAL RESISTIVITY AS A FUNCTION OF TEMPERATURE Introduction The ability of materials to conduct electric charge gives us the means to invent an amazing array of electrical and electronic devices,
More informationY-HEAT THERMAL MANAGEMENT SYSTEM FOR OVEN, FURNACE, VACUUM FURNACE, AND CRYOGENIC CHAMBER SYSTEMS. Product Overview and Specifications
Y-HEAT THERMAL MANAGEMENT SYSTEM FOR OVEN, FURNACE, VACUUM FURNACE, AND CRYOGENIC CHAMBER SYSTEMS Product Overview and Specifications Y-HEAT Overview Throughout the thermal industry, ovens, furnaces, vacuum
More informationTVOC Heated FID VE 7
Rack Mount/Table Top THC Analyzer TVOC Heated FID VE 7 The VE7 19"/5PU rack mount and table top heated emission analyzer for the continuous determination of the mass concentration of total gaseous organic
More informationCompact Electro Plating Machine - CEPM-2A USER MANUAL
1 Compact Electro Plating Machine - CEPM-2A USER MANUAL PRODUCT INTRODUCTION: This compact electroplating machine is a high performance compact plating unit that is especially designed for coating all
More informationIndustrial flue gas probes. Instruction manual
Industrial flue gas probes Instruction manual 2 1 Contents 1 Contents 1 Contents... 3 2 Safety and the environment... 4 2.1. About this document... 4 2.2. Ensure safety... 4 2.3. Protecting the environment...
More informationInstructions for using the PerkinElmer DSC-8500
Instrument Summary: Instructions for using the PerkinElmer DSC-8500 Differential scanning calorimetry (DSC) measures the amount of energy required to heat or cool a sample to a given temperature. This
More informationOperations Manual. For Research Use Only. PH: EM: WEB:
Operations Manual For Research Use Only PH: 908-769-5555 EM: info@benchmarkscientific.com WEB: www.benchmarkscientific.com Thank you for choosing the BioClave Benchtop Sterilizer. Prior to operating this
More information// XPLORER-NS. Total Nitrogen and Sulfur Analyzer
// XPLORER-NS Total Nitrogen and Sulfur Analyzer Configuration: XPLORER with ARCHIE The XPLORER-NS was developed as a total nitrogen/total n/to tota tal sulfur ur combustion analyzer, alyz offering fast,
More informationVLSI INTRODUCTION P.VIDYA SAGAR ( ASSOCIATE PROFESSOR) Department of Electronics and Communication Engineering, VBIT
VLSI INTRODUCTION P.VIDYA SAGAR ( ASSOCIATE PROFESSOR) contents UNIT I INTRODUCTION: Introduction to IC Technology MOS, PMOS, NMOS, CMOS & BiCMOS technologies. BASIC ELECTRICAL PROPERTIES : Basic Electrical
More informationChapter 3 Silicon Device Fabrication Technology
Chapter 3 Silicon Device Fabrication Technology Over 10 15 transistors (or 100,000 for every person in the world) are manufactured every year. VLSI (Very Large Scale Integration) ULSI (Ultra Large Scale
More information4. Thermal Oxidation. a) Equipment Atmospheric Furnace
4. Thermal Oxidation a) Equipment Atmospheric Furnace Oxidation requires precise control of: temperature, T ambient gas, G time spent at any given T & G, t Vito Logiudice 34 4. Thermal Oxidation b) Mechanism
More informationIsolation Technology. Dr. Lynn Fuller
ROCHESTER INSTITUTE OF TECHNOLOGY MICROELECTRONIC ENGINEERING Isolation Technology Dr. Lynn Fuller Motorola Professor 82 Lomb Memorial Drive Rochester, NY 14623-5604 Tel (585) 475-2035 Fax (585) 475-5041
More informationFurnaces for Research and Laboratory. Testing Drying Ashing Tempering Preheating Sintering Firing Annealing
Furnaces for Research and Laboratory Testing Drying Ashing Tempering Preheating Sintering Firing Annealing Muffle Furnaces KL 03/11 KL 15/12 T max 1100 C and 1200 C Universal muffle furnaces for many laboratory
More informationPURDUE UNIVERSITY ELECTROPLATING GOLD PROCESS SPECIFICATION RF LAB. Professor in charge: William Chappell MSEE 289
RF LAB Professor in charge: William Chappell MSEE 289 Author: Bob Salisbury Date: 11/21/2007 Revised 6/16/2008 6/16/2008 page 1 of 11 SCOPE... 3 PURPOSE... 3 REFERENCE DOCUMENTS... 3 MATERIALS... 3 EQUIPMENT...
More informationFigure 2.3 (cont., p. 60) (e) Block diagram of Pentium 4 processor with 42 million transistors (2000). [Courtesy Intel Corporation.
Figure 2.1 (p. 58) Basic fabrication steps in the silicon planar process: (a) oxide formation, (b) selective oxide removal, (c) deposition of dopant atoms on wafer, (d) diffusion of dopant atoms into exposed
More information3.155J / 6.152J Micro/Nano Processing Technology TAKE-HOME QUIZ FALL TERM 2005
3.155J / 6.152J Micro/Nano Processing Technology TAKE-HOME QUIZ FALL TERM 2005 1) This is an open book, take-home quiz. You are not to consult with other class members or anyone else. You may discuss the
More informationMODELS 501, 501-N, 502, & 502-N
Grand Rapids, Michigan, U.S.A. 49504-5298 USER S OPERATING AND INSTRUCTION MANUAL MODELS 501, 501-N, 502, & 502-N PANINI GRILLS 0502S20000 INDEX Section Description Document No. Page No. SAFETY INSTRUCTIONS
More information!!!! WARNING!!!! WELDING FUMES AND GASES CAN BE DANGEROUS TO YOUR HEALTH.
CAREFULLY!!!! WARNING!!!! CAREFULLY WELDING FUMES AND GASES CAN BE DANGEROUS TO YOUR HEALTH. BEFORE USING THIS PRODUCT THE WELDER (END-USER) MUST READ AND UNDERSTAND THE COMPLETE PRODUCT WARNING LABEL
More informationPFBH room 341 Steris Autoclave Operation
University of California San Diego Dept. of Bioengineering Standard Operating Procedure SOP000026 V0.5 PFBH room 341 Steris Autoclave Operation Issuing Office: Dept. of Bioengineering, JSOE Responsible
More informationA Hydrogen Powered Bottle Rocket
A Hydrogen Powered Bottle Rocket Rockets are made by filling plastic water bottles with a mixture of hydrogen gas and oxygen gas. The rocket is launched by igniting the mixture with a flame. The bottle
More informationHigh Temperature Vacuum Tube Furnace GSL Operational Manual
High Temperature Vacuum Tube Furnace GSL-1100 Operational Manual MTI Corporation 860 South 19 th Street, Richmond, CA 94804, USA Tel: 510-525-3070 Fax: 510-525-4705 E-mail: info@mtixtl.com Web site: www.mtixtl.com
More informationMODEL P710 SERIES MACHINES AUTOMATIC STRAPPING MACHINES
MODEL P710 SERIES MACHINES AUTOMATIC STRAPPING MACHINES Reliable packaging solutions for all industries. Auto strap load for operator convenience and safety Loop eject avoids strap jamming in the machine
More informationContact Information: Laboratory Supervisor: Denise Kind Laboratory Manager: Mat Ashby
Biology and Wildlife STANDARD OPERATING PROCEDURE Autoclaving Location(s): Murie 215 Chemical(s): None Specific Hazards: o steam improper use of autoclave can expose user to dangerous steam burns o extremely
More informationModular Coldroom Installation Manual
Modular Coldroom Installation Manual May 2005 Disclaimer: The information contained in this installation guide is provided "as is". Bromic Pty Ltd does not warrant the accuracy, adequacy or completeness
More information... Targeted treatment for Rising Damp... F A S T C L E A N E F F E C T I V E
.... Targeted treatment for Rising Damp.... Dryzone has many advantages over conventional chemical injection systems: The Problem. Whenever Rising Damp is diagnosed it is important to have the condition
More information2015 EE410-LOCOS 0.5µm Poly CMOS Process Run Card Lot ID:
STEP 0.00 - PHOTOMASK #0- ZERO LEVEL MARKS Starting materials is n-type silicon (5-10 ohm-cm). Add four test wafers labeled T1-T4. T1 and T2 will travel with the device wafers and get all of the processing
More informationA Hydrogen Powered Bottle Rocket
A Hydrogen Powered Bottle Rocket Rockets are made by filling plastic water bottles with a mixture of hydrogen gas and oxygen gas. The rocket is launched by igniting the mixture with a flame. The bottle
More informationFabrication Technology
Fabrication Technology By B.G.Balagangadhar Department of Electronics and Communication Ghousia College of Engineering, Ramanagaram 1 OUTLINE Introduction Why Silicon The purity of Silicon Czochralski
More informationTitle: Standard Operating Procedure for the Collection of VOCs in Ambient Air Using Summa Canisters and Xontech Sampler
Procedure No: SOP-016 Revision No: 0 (new document) Page No.: 1 of 7 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to meet the
More informationW ater T alk. Field Corrosion Testing using Coupons
W ater T alk Field Corrosion Testing using Coupons Volume 12, Issue 1 April 2012 Most Cooling Water systems are continuously in use. Therefore it is rare to have access to inspect the actual system to
More informationSEB Autoclave Training
SEB Autoclave Training 1 Autoclave Details 3 Autoclaves located in SEB, one on each upper floor. Expensive pieces of equipment ($70,000 each). Essential for several groups research. Most malfunctions are
More informationSTAINLESS STEEL SHEETS
STAINLESS & NICKEL ALLOY CHATHAM STEEL CORPORATION 59 STAINLESS STEEL SHEETS TYPE 04, 04L, 6, 6L No. 2B Finish Cold Rolled, Annealed *No. 2D Finish No. Finish Polished One Side No. 4 Finish Polished One
More informationCH88 In Situ Flue Gas Oxygen Transmitter
CH88 Combustion Flue Gas Oxygen Analyzer System CMB-PDS-CH88 Product Data Sheet September 2017 CH88 In Situ Flue Gas Oxygen Transmitter World-Class TM performance Outstanding accuracy ±1% full scale 1056
More information!!!! WARNING!!!! WELDING FUMES AND GASES CAN BE DANGEROUS TO YOUR HEALTH.
!!!! WARNING!!!! WELDING FUMES AND GASES CAN BE DANGEROUS TO YOUR HEALTH. BEFORE USING THIS PRODUCT THE WELDER (END-USER) MUST READ AND UNDERSTAND THE COMPLETE PRODUCT WARNING LABEL AND MATERIAL SAFETY
More information1. Vertical reciprocating conveyor, Hydraulic straddle VRC.
PART 1: GENERAL 1.01 Section Includes: 1. Vertical reciprocating conveyor, Hydraulic straddle VRC. a. Machine, controller, platform, structural steel hoist frame. b. Wire mesh enclosure and gates. Related
More informationLABORATORY SAFETY EVALUATION. S = Satisfactory N = Needs Improvement* N/A = Not Applicable
LABORATORY SAFETY EVALUATION Principal Investigator: Department: PH# : Building: Date: Room(s) : S = Satisfactory N = Needs Improvement* N/A = Not Applicable *If an area is in need of improvement, please
More informationCooling Water System Inspection Process. Presented by Craig Nicks, City of Onkaparinga
Cooling Water System Inspection Process Presented by Craig Nicks, City of Onkaparinga What is the cooling water system used for? Air handling i.e. Commercial A/C Industrial / Commercial i.e. Plastics moulding
More informationMATERIALS OF CONSTRUCTION
HIGH PRESSURE PURIFIER (MODEL SG6140) Model SG6140 purifier is designed to remove water and/or oil from gas or liquid streams. In the laboratory, it protects delicate or sensitive instruments, prolongs
More informationImproving Production of Carbon Nanotube Composites
Improving Production of Carbon Nanotube Composites Research Experience for Teachers (RET Program) Brigham Young University Provo, Utah Dan Broadbent Physics Teacher Timpanogos High School Orem, Utah dbroadbent@alpinedistrict.org
More information8000 / 8100 ph Meters
8000 / 8100 ph Meters Operating Instructions GETTING STARTED To begin taking measurements, connect the probe(s) to the socket(s) on the instrument and remove the cap or bottle covering the electrode. Please
More informationPipe and Equipment Insulations. Zeston. PVC Insulated Fitting Covers Installation Instructions for Hot, Cold and USDA Piping Systems
Pipe and Equipment Insulations Zeston PVC Insulated Fitting Covers Installation Instructions for Hot, Cold and USDA Piping Systems Zeston PVC Insulated Fitting Covers Installation Instructions for Hot,
More informationEECS130 Integrated Circuit Devices
EECS130 Integrated Circuit Devices Professor Ali Javey 9/13/2007 Fabrication Technology Lecture 1 Silicon Device Fabrication Technology Over 10 15 transistors (or 100,000 for every person in the world)
More informationShur-Shot X-Proof Hydrogen Fluoride Alarm Operations Manual
Shur-Shot X-Proof Hydrogen Fluoride Alarm Operations Manual P/N 1000006053 Rev E $7%$QDO\WLFV// //& 733 Dairy Rd. Parkton, Md. 21120 www.atbanalytics.com (410) 733-6365 Table of Contents Chapter 1: Getting
More informationnual k Excella E-1 Open Air Shaker N) manual New Brunswick Excella E-1 Open Air Shaker Operating manual
nual k Excella E-1 Open Air Shaker N) manual New Brunswick Excella E-1 Open Air Shaker Operating manual Copyright 2014 Eppendorf AG, Germany. No part of this publication may be reproduced without the prior
More informationAEROSPACE MATERIAL SPECIFICATION
AEROSPACE MATERIAL SPECIFICATION AMS 2759/3D Issued OCT 1984 Revised FEB 2005 Superseding AMS 2759/3C Treatment Precipitation-Hardening Corrosion-Resistant and Maraging Steel Parts 1. SCOPE: This specification,
More informationSemiconductor Device Fabrication Study
Proceedings of The National Conference on Undergraduate Research (NCUR) 2003 University of Utah, Salt Lake City, Utah March 13-15, 2003 Semiconductor Device Fabrication Study Tsung-Ta Ho and Michael R.
More informationVLSI Digital Systems Design
VLSI Digital Systems Design CMOS Processing cmpe222_03process_ppt.ppt 1 Si Purification Chemical purification of Si Zone refined Induction furnace Si ingot melted in localized zone Molten zone moved from
More informationProduct Manual Release 05/17
Temporary Safety Barrier (TL-2 MASH) Product Manual VHD (v3) www.valmonthighway.com Table of Contents 1.0 ArmorZone Introduction... 3 2.0 Before Installation... 3 3.0 Limitations & Warnings... 3 4.0 System
More informationQUICK Guide. System start-up and shutdown 3. Operating panels. State colors. General process description. Device operating modes
QUICK Guide System start-up and shutdown 3 Operating panels 4 State colors 5 General process description 6 Device operating modes Recovering from an error situation 15 19 Tips 21 MMS5 rev. 2.0 / FMS Options
More information! CAUTION: Read and follow these instructions.
OmniLog SAFETY AND INSTALLATION INSTRUCTIONS! CAUTION: Read and follow these instructions. Avoid Contact with Hazardous Live Parts WARNING: Risk of electrical shock. Refer servicing to qualified personnel.
More informationNORTHWESTERN UNIVERSITY PROJECT NAME JOB # ISSUED: 03/29/2017
SECTION 12 2113 - LOUVER BLINDS PART 1 - GENERAL 1.1 RELATED DOCUMENTS A. Drawings and general provisions of the Contract, including General and Supplementary Conditions and Division 01 Specification Sections,
More informationMEMS LAB MANUAL. Matthew Leone Todd Kaiser Montana State University. Special Thanks to: Andy Lingley Brad Pierson Phil Himmer
MEMS LAB MANUAL 2007 A complete description of the fabrication sequence for piezoresistive MEMS sensors. This manual was designed for use with the Montana Microfabrication Facility at MSU. Special Thanks
More informationHeated FID 109A. Rack Mount/Table Top NMGOC Analyzer
Rack Mount/Table Top NMGOC Analyzer Heated FID 109A J.U.M. Engineering 2014/2016 The 109A is a rack mount and table top heated emission analyzer for the continuous / simultaneous determination of the mass
More informationHow To Write A Flowchart
1 Learning Objectives To learn how you transfer a device concept into a process flow to fabricate the device in the EKL labs You learn the different components that makes up a flowchart; process blocks,
More informationProduct data sheet DUMATHERM N Pro. for the automatic serial determination of NITROGEN
for the automatic serial determination of NITROGEN DUMATHERM N Pro Automated system for fast nitrogen determination acc. to Dumas Technical Data DUMATHERM N Pro Data DUMATHERM N Pro Sample size 0,5 mg
More information// XPLORER-NS. Full range Total Nitrogen, Total Sulfur instrument for the modern Laboratory.
Configuration: XPLORER with ARCHIE and GLS* // XPLORER-NS Full range Total Nitrogen, Total Sulfur instrument for the modern Laboratory. TE Instruments has developed the XPLORER-NS, a Total Nitrogen and
More informationPioneering ALD experience since The ALD Powerhouse PRODUCT CATALOGUE FOR SUNALE P-SERIES ALD SYSTEMS
Pioneering ALD experience since 1974 The ALD Powerhouse PRODUCT CATALOGUE FOR SUNALE P-SERIES ALD SYSTEMS SUNALE P-SERIES ALD SYSTEMS PRODUCT CATALOGUE Description SUNALE P-series ALD system SUNALE P-series
More informationFirmware Version:1.137 Display Version:2.06a. Xrd-384 Stacker Installation Manual
Firmware Version:1.137 Display Version:2.06a Xrd-384 Stacker Installation Manual 1 Content 1. Safety Precaution 3 2. To use instrument safety 3-4 3. Instruction for use 4 4. Specifications 5 5.Overview
More informationDeskDECT Feature Phone
DeskDECT Feature Phone Please read this manual carefully before using your DeskDECT. 1. 2. 2.1 2.2 INTRODUCTION -------------------------------------------------- GETTING STARTED ---------------------------------------------
More informationC. Up to Two and One-half (2-1/2) inches: Full port, 2-piece, bronze body ball valves with teflon seat conforming to MSS SP-110.
SECTION 220000 PLUMBING SYSTEMS 1.0 Valves A. ASME compliance: 1. ASME B1.20.1 2. ASME B16.1 3. ASME B16.10 4. ASME B16.34 5. ASME B16.18 6. ASME B31.9 B. All branch mains shall be provided with shutoff
More informationIon Implantation Most modern devices doped using ion implanters Ionize gas sources (single +, 2+ or 3+ ionization) Accelerate dopant ions to very
Ion Implantation Most modern devices doped using ion implanters Ionize gas sources (single +, 2+ or 3+ ionization) Accelerate dopant ions to very high voltages (10-600 KeV) Use analyzer to selection charge/mass
More informationFor Safe and Delicious Water
RESIDUAL CHLORINE ANALYZER 400G NON-REAGENT FREE CHLORINE ANALYZER FC400G 400G RESIDUAL CHLORINE ANALYZER FC400G NON-REAGENT FREE CHLORINE ANALYZER Bulletin 12F01A01-01E www.yokogawa.com/an/ For Safe and
More informationPriority 2 Access to Goods & Services
Based on the 2010 ADA Standards for Accessible Design Project Building Location Date Surveyors Contact Information The layout of the building should allow people with disabilities to obtain goods and services
More informationBULK BAG FILLING SYSTEMS
QUICKPAK SERIES 330E BULK BAGGING SYSTEM Unique Pin & Ladder frame adjusts quickly and safely without tools or outside lift device. Pneumatic carriage design can be easily programmed to lift, stretch and
More informationSOP: Applikon ez-control Bioreactor Controller Operation
Page 1 of 19 Approvals Preparer: John Buford Reviewer: Jason McMillan Reviewer: Dr. Margaret Bryans Date: 19DEC13 Date: 20DEC13 Date: 21DEC13 1. Purpose 1.1. Cultivate a cell culture using an Applikon
More informationLiquid layer sampler, cable operated
Liquid layer sampler, cable operated operating instructions Contents On these operating instructions...2 1. Description...2 2. Taking into operation and decontamination...3 3. Using the device...4 3.1
More informationFabrication and Layout
ECEN454 Digital Integrated Circuit Design Fabrication and Layout ECEN 454 3.1 A Glimpse at MOS Device Polysilicon Aluminum ECEN 475 4.2 1 Material Classification Insulators Glass, diamond, silicon oxide
More informationedna PROTOCOL SAMPLE COLLECTION Caren Goldberg and Katherine Strickler, Washington State University Revised January 2017
edna PROTOCOL SAMPLE COLLECTION Caren Goldberg and Katherine Strickler, Washington State University Revised January 2017 MATERIALS 1. Cellulose nitrate disposable filter funnels or other field-tested,
More informationHOMEWORK 4 and 5. March 15, Homework is due on Monday March 30, 2009 in Class. Answer the following questions from the Course Textbook:
HOMEWORK 4 and 5 March 15, 2009 Homework is due on Monday March 30, 2009 in Class. Chapter 7 Answer the following questions from the Course Textbook: 7.2, 7.3, 7.4, 7.5, 7.6*, 7.7, 7.9*, 7.10*, 7.16, 7.17*,
More informationUCI KCCAMS Facility. Swipe protocol April 28, 2011
UCI KCCAMS Facility Swipe protocol April 28, 2011 Contents I. Introduction page 2 II. Conducting a swipe page 2 III. Combustion of swipe samples page 4 IV. Sealing quartz tubes page 6 V. Placing the combustion
More informationBrian Izzio 5th Year Microelectronic Engineering Student. Rochester Institute of Technology
CAPACITAWZE-VOI~TAGE characterization FOR POLYSILICON GATE MOS CAPACITORS DJTRODUcTION Brian Izzio 5th Year Microelectronic Engineering Student Rochester Institute of Technology ABSTRACT The effects of
More informationChar-Gas Has Greater Value Than for Cooking: Example of C2P Char-Gasifiers
Char-Gas Has Greater Value Than for Cooking: Example of C2P Char-Gasifiers A Presentation at the 2017 ETHOS Conference in Seattle-Kirkland, WA, USA 27-29 January 2017 Gary Gilmore Forester in Pennsylvania
More informationSTEAG AWP-PR-STRIPPER
STEG WP-PR-STRIPPER Processes available : PRS 1000 IP/NMP Gripper Clean QDR MRNGONI-Dryer Wafer Transfer Station Load/Unload Wafer Size: 8 Main Body Size : W 1765 x L 4550 x H 2400 (mm) 54520-011 WP -
More informationSTANDARD OPERATING PROCEDURES
PAGE: 1 of 14 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY* 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS* 6.0 REAGENTS
More informationInstallation Instructions
Installation Instructions AquaSnap 010-055 Chilled Water Storage Tank 50/60 Hz Part No: -900---001; -900---009; -900---010 CONTENTS INTRODUCTION.................................. 1 SAFETY CONSIDERATIONS......................
More informationBulletin Working space about electrical equipment Rules 2-308, and Issued April 2010 Supersedes Bulletin 2-9-3
-9-4 Bulletin -9-4 Working space about electrical equipment Rules -308, -310 and -31 Issued April 010 Supersedes Bulletin -9-3 Scope (1) Minimum working space around electrical equipment () Questions and
More informationAP5K-C Precision AC Double-Pulse Spot Welding Machine User s Manual Shenzhen Will-Best Electronics Co., Ltd
AP5K-C Precision AC Double-Pulse Spot Welding Machine User s Manual Shenzhen Will-Best Electronics Co., Ltd 1 Content 1. Introduction...3 1.1 Functions...3 1.2 Units of AP5K-C...4 2. The Initial Installation
More informationHuman DNA Alu Amplification by Polymerase Chain Reaction (PCR)* Laboratory Procedure
Human DNA Alu Amplification by Polymerase Chain Reaction (PCR)* Laboratory Procedure *Polymerase Chain Reaction is covered by patents owned by Hoffmann-La Roche, Inc. This experiment was adapted from Laboratory
More informationAQUATERR I N S T R U M E N T S. M, T & EC -300 User s Manual
AQUATERR I N S T R U M E N T S M, T & EC -300 User s Manual Aquaterr Digital Soil Moisture, Temperature and Salinity Meters AQUATERR INSTRUMENTS Leader in Portable Soil Measuring Technology 2 It is important
More information3M Personal Safety Division. 3M TM EVM Series Environmental Monitor. EVM Series. Quick Start Guide
3M Personal Safety Division 3M TM EVM Series Environmental Monitor EVM Series Quick Start Guide EVM and User Interface Indicators Display Softkey menus Backlight Particulate pump Left arrow key Run/Stop
More informationGuidelines for QualiTru Line Sampling
Guidelines for QualiTru Line Sampling QUALITRU GUIDELINES FOR LINE SAMPLING 1. Purpose: Following these Standard Operating Procedures, farms and processing plants may use the QualiTru aseptic sampling
More informationL505 VISUAL LEVEL INDICATOR. Dip-Tape Visual Level Indicator INSTALLATION AND OPERATIONS MANUAL
L505 VISUAL LEVEL INDICATOR INSTALLATION AND OPERATIONS MANUAL Dip-Tape Visual Level Indicator READ THIS MANUAL PRIOR TO INSTALLATION This manual provides information on the L505 Dip-Tape Visual Level
More information