Certificate of Accreditation

Size: px
Start display at page:

Download "Certificate of Accreditation"

Transcription

1 PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: San Nicolás de los Garza, Nuevo León, México. C.P (Hereinafter called the Organization) and hereby declares that Organization is accredited in accordance with the recognized International Standard: ISO/IEC 17025:2005 This accreditation demonstrates technical competence for a defined scope and the operation of a laboratory quality management system (as outlined by the joint ISO-ILAC-IAF Communiqué dated April 2017): Dimensional, Mechanical, Chemical, Thermodynamic, Mass, Force and Weighing Devices, Time and Frequency and Optical Calibration (As detailed in the supplement) Accreditation claims for such testing and/or calibration services shall only be made from addresses referenced within this certificate. This Accreditation is granted subject to the system rules governing the Accreditation referred to above, and the Organization hereby covenants with the Accreditation body s duty to observe and comply with the said rules. For PJLA: Initial Accreditation Date: Issue Date: Expiration Date: July 05, 2012 May 08, 2018 June 30, 2020 Tracy Szerszen President/Operations Manager Perry Johnson Laboratory Accreditation, Inc. (PJLA) 755 W. Big Beaver, Suite 1325 Troy, Michigan Accreditation No.: Certificate No.: L The validity of this certificate is maintained through ongoing assessments based on a continuous accreditation cycle. The validity of this certificate should be confirmed through the PJLA website: Page 1 of 5

2 Dimensional Certificate of Accreditation: Supplement San Nicolás de los Garza, Nuevo León, México CP RANGE OR NOMINAL DEVICE SIZE AS APPROPRIATE Caliper F 0.5 mm to 600 mm ( x 10-2 L) μm Gage Block Grade 0 Micrometer F 0.5 mm to 600 mm ( x 10-2 L) μm Indicator F 0.5 mm to 50.8 mm 2.8 μm Coating Thickness 23.2 µm to µm ( x 10-4 L) µm Elcometer Calibration Gage FO Standards Height Gage F 50 mm to 600 mm (5.04 x x 10-6 L) mm Gage Block Grade 1 Mechanical Pressure FO -14 psi to -0.2 psi psi Pressure Gauge Dwyer 10 psi to 300 psi 0.33 psi DPGW-00 DPGW psi to psi 5.8 psi DPG psi to psi 29 psi DPG-111 Viscosity Meters Dynamic Viscosity FO Ford cups 2, 3, 4,5 FO Zhan Cups 2,3,4,5 FO ISO Cups 3, 4, 6 FO DIN Cups 4 FO SHELL Cups 2, 2 ½, 3, 3 ½, 4, 5, 6 FO Chemical 100 mpa s to mpa s 2.5 % of reading Cannon Standard Oil 1.44 mm 2 /s to 367 mm 2 /s 2.5 % of reading 3.5 mm 2 /s to 400 mm 2 /s 2.5 % of reading mm 2 /s to 570 mm 2 /s 2.5 % of reading 4.57 mm 2 /s to 452 mm 2 /s 2.5 % of reading mm 2 /s to 452 mm 2 /s 2.5 % of reading ph Meters FO 4.01 ph ph CENAM Traceable 7.01 ph ph Standards Fermont Issue: 05/2018 This supplement is in conjunction with certificate #L Page 2 of 6

3 Chemical ph Meters FO ph ph CENAM Traceable Standards Fermont Conductivity Meters FO 84 µs/cm 1 μs/cm NIST Traceable Standards μs/cm 7 μs/cm HANNA μs/cm 20 μs/cm µs 64 μs/cm Thermodynamic Infrared Measuring Equipment FO 50 C to 500 C 1.2 C Hart Calibrator Black Body Thermal Chamber Thermal Oven Controllers FO -50 ºC to C 0.6 C Thermometer with TC TYPE K Digital Thermometer Brand: Fluke, Model: 52 II Glass Thermometer FO -30 ºC to 500 ºC 1.1 C Dry Block with Bimetallic -30 ºC to 500 ºC 1.1 C Thermocouple Thermometer FO -30 ºC to 500 ºC 0.6 C Refrigerated Bath Measurement with Circulator, Brand: Labtech, Model: LCB-R13 Thermocouple Type J FO Measurement with Thermocouple Type K FO Measurement with Thermocouple Type T FO -30 ºC to 500 ºC 0.6 C -30 ºC to 400 ºC 0.6 C -30 ºC to 500 ºC 0.6 C Measurement RTD Pt100 FO Relative Humidity O 11.3 % RH 0.7 % RH Satured Salt Solution, 32.9 % RH 0.6 % RH Vaisala 75.4 % RH 0.9 % RH 97.4 % RH 1.5 % RH Issue: 05/2018 This supplement is in conjunction with certificate #L Page 3 of 6

4 Mass, Force and Weighing Devices Analytical Balance O Precision Balance O Scale O 1 mg to g (Res.= 0.01 mg) 1 mg to 200 g (Res.= g) 0.1 g to g (Res.= 0.01 g) 5 kg to kg (Res.= 1 g) (2 x x 10-6 Wt) g OIML E2 Class Weights (2 x x 10-6 Wt) g OIML F1 Class Weights (1.16 x x 10-6 Wt) g ( x 10-4 Wt) g OIML M1 Weights Time and Frequency Frequency of Rotation FO 1.8 m/min to m/min 2 % of reading Tachometer Shimpo DT-105A Optical p(λ)spectral Reflectance FO P: 360 nm to 740 nm, 0 % to 100 % 2.5 % of reading White, Blue, Green, Red and Gray Standard Tile X: 0.3 to 95 Y: 0.3 to 100 Z: 0.3 to 100 Color Values: CIE L*: CIE a*: CIE b*: Spectrophotometers Transmittance Absorbance At these Wave Lengths λ: 230 nm to 700nm FO Transmittance Density (400 nm to 700 nm) FO Reflective Density (400 nm to 700 nm) FO 0 to to to 70 τ: 1 % to 95 % α: 0.03 to % of reading 0.05 % of reading Neutral Density Filters, Holmium Oxide Glass % OD to 4 % OD 0.02 % OD Density Filter Standard 0.06 to % OD Target Density Calibration Issue: 05/2018 This supplement is in conjunction with certificate #L Page 4 of 6

5 Optical Gloss / Specular Reflectance Angle of Incline FO p(e): 20 º 0.17 Gloss Units By Gardner Gloss and Semi-Gloss Standards p(e): 60 º 0.16 Gloss Units p(e): Ev Illuminance Ev Light Color 85 º 0 lux to lux 50 K to K 0..2 Gloss Units 1.3 % of reading 15 K Luxmeter Ev Light Meter FO 100 lux to lux 2 % of reading Refractometer FO 1 Brix to 80 Brix 0.55 % of reading Sucrose Standards, OIML R The CMC (Calibration and Measurement Capability) stated for calibrations included on this scope of accreditation represents the smallest measurement uncertainty attainable by the laboratory when performing a more or less routine calibration of a nearly ideal device under nearly ideal conditions. It is typically expressed at a confidence level of 95 % using a coverage factor k (usually equal to 2). The actual measurement uncertainty associated with a specific calibration performed by the laboratory will typically be larger than the CMC for the same calibration since capability and performance of the device being calibrated and the conditions related to the calibration may reasonably be expected to deviate from ideal to some degree. 2. The laboratories range of calibration capability for all disciplines for which they are accredited is the interval from the smallest calibrated standard to the largest calibrated standard used in performing the calibration. The low end of this range must be an attainable value for which the laboratory has or has access to the standard referenced. Verification of an indicated value of zero in the absence of a standard is common practice in the procedure for many calibrations but by its definition it does not constitute calibration of zero capacity. 3. The presence of a superscript F means that the laboratory performs calibration of the indicated parameter at its fixed location. Example: Outside Micrometer F would mean that the laboratory performs this calibration at its fixed location. 4. The presence of a superscript O means that the laboratory performs calibration of the indicated parameter onsite at customer locations. Example: Outside Micrometer O would mean that the laboratory performs this calibration onsite at the customer s location. 5. The presence of a superscript FO means that the laboratory performs calibration of the indicated parameter both at its fixed location and onsite at customer locations. Example: Outside Micrometer FO would mean that the laboratory performs this calibration at its fixed location and onsite at customer locations. Issue: 05/2018 This supplement is in conjunction with certificate #L Page 5 of 6

6 6. Measurement uncertainties obtained for calibrations performed at customer sites can be expected to be larger than the measurement uncertainties obtained at the laboratories fixed location for similar calibrations. This is due to the effects of transportation of the standards and equipment and upon environmental conditions at the customer site which are typically not controlled as closely as at the laboratories fixed location. 7. The term L represents length in inches or millimeters as appropriate to the uncertainty statement. 8. The term Wt represents weight in pounds or grams (including SI multiple and submultiple units) appropriate to the uncertainty statement. Issue: 05/2018 This supplement is in conjunction with certificate #L Page 6 of 6