Carbon / Sulfur Analyzer ELEMENTRAC CS-i

Size: px
Start display at page:

Download "Carbon / Sulfur Analyzer ELEMENTRAC CS-i"

Transcription

1 General Information The new carbon / sulfur analyzer ELEMENTRAC CS-i was developed for the accurate and safe analysis of carbon and sulfur in inorganic samples. The analyzer is equipped with a powerful induction furnace for sample combustion. Up to four highly sensitive infrared (IR) cells allow the analyzer to determine both high and low carbon and sulfur concentrations in only one measurement run. The measuring range of each cell may be adapted to the user's specific requirements to ensure optimum measuring conditions for each application. The analyzer is supplied with the new comprehensive and convenient ELEMENTS software, which features statistics, groupings, reports, diagnosis tools and many additional functions for ELEMENTRAC analyzers. Application Examples alloys, carbides, cast iron, cement, ceramics, copper, glass, iron, minerals, ores, refractory metals, sand, steel, titanium,... Product Advantages simultaneous carbon and sulfur determination with minimal sample preparation wide range of inorganic materials can be analyzed freely selectable configuration of each IR cell optional gold IR path for analysis of halogen or acid containing samples control of induction performance provides more precise analysis of samples with low melting temperature powerful (2.2 kw) induction furnace for temperatures above 2,000 C NEW: automatic vacuum cleaning system ensures higher measurement precision and stability heated and easily accessible dust trap allows for improved sulfur detection optimized catalyst reactor permits more accurate carbon detection NEW: ELEMENTS software with comprehensive analysis and diagnosis tools (supporting data and application export, comment fields, and many more) single and multipoint calibration low maintenance robust design allows usage in production control and laboratory Features Measured elements Samples Furnace alignment Sample carrier carbon, sulfur inorganic vertical ceramic crucibles Field of application construction materials, engineering / electronics, geology / mining, glass / page 1/5

2 ceramics, steel / metallurgy Furnace induction furnace, above 2,000 C Detection method Number of IR cells 1-4 Material of IR path Typical analysis time Chemicals required Gas required Power requirements Dimensions (W x H x D) Weight Required equipment Optional accessories solid state infrared absorption aluminum (optional gold) 40 seconds (nominal) magnesium perchlorate, platinumbased catalyst, sodium hydroxide compressed air (4 bar/ 60 psi) oxygen 99.5 % pure (2-4 bar / psi) 230 V, 50/60 Hz, max. 15 A, 3450 W 52 x 84 x 75 cm ~ 150 kg balance (resolution g), monitor, PC Autoloader (for 36 or 130 crucibles), carrier gas purification, halogen trap, HTF-540 pre-heating furnace, voltage stabilizer 5 KVA Function Principle Carbon / Sulfur Analyzer CS-i: Operation Operating the CS-i is simple and safe. After weighing the sample in a ceramic crucible, the weight is transferred from the interfaced balance to the PC. If required, sample weights can also be entered manually. Then an accelerator (such as iron or tungsten) is added and, after having placed the crucible on the pedestal, the analysis starts. The analysis time is 40 to 50 seconds. The detector signals and instrument parameters are displayed during analysis. Evaluation of the signals and display of the results are done automatically; the data can be transferred to a laboratory information management system (LIMS). The CS-i requires minimum maintenance. The particle filters and chemicals which need to be maintained are easily accessible. Carbon / Sulfur Analyzer CS-i: Measuring Principle In the induction furnace the sample is melted in a pure oxygen atmosphere, causing sulfur to react to sulfur dioxide (SO2) and carbon to a mixture of carbon monoxide (CO) and carbon dioxide (CO2). The combustion gases pass through a dust filter and moisture absorber for purification. In the next step the sulfur dioxide is detected in infrared cells. In the CS-i infrared cells with different sensitivities (high/low) can be adapted according to the user's requirements. Oxidation of both, carbon monoxide to carbon dioxide and sulfur dioxide to sulfur trioxide follow the sulfur measurement. The SO3 gas is removed with cellulose wool; the carbon content is detected by infrared cells which can be individually customized. The ELTRA CS-i analyzer can be equipped with up to 4 independent infrared cells. Order Data page 2/5

3 ELEMENTRAC CS-i (Please order PC, monitor, balance and consumables (starter-kit, anhydrone, sodium hydroxide, Pt/Si catalyst) separately) Measuring ranges at 500 mg sample weight CS-i 1xC 0.1-7% C CS-i 2xC % C 0.1% - 7% C CS-i 1xS % S CS-i 2xS % S 0.3% - 4.5% S CS-i 1xC 0.1-7% C + 1xS % S CS-i 2xC % C 0.1% - 7% C + 1xS % S CS-i 1xC 0.1-7% C + 2xS % S 0.3% - 4.5% S CS-i 2xC % C 0.1-7% C + 2xS % S 0.3% - 4.5% S Further measuring range combinations on request PC, Monitor, Balance Computer with dual core processor, 300 GB HDD, 4 GB RAM, Windows operating system, DVD-ROM, keyboard, mouse Monitor, TFT Balance (resolution g) Printer Accessories HTF-540 crucible pre-heating furnace Oxygen regulator Crucible holder Consumables Required consumables Starter-kit for 1,000 analyses (1,000 crucibles, 2,500 g tungsten, 908 g pure iron accelerator, 50 g glass wool, 50 g cellulose, 50 g quartz wool) Anhydrone (magnesium perchlorate), 454 g Sodium hydroxide, 500 g Pt/Si catalyst,15 g Optional consumables Ceramic crucibles, premium, Ø 1, foil-wrapped, 1,000 pieces Ceramic lids, 250 pieces page 3/5

4 90220 Tungsten, premium, 2,500 g Glass wool, 454 g Cellulose, 100 g Cellulose, 50 g Quartz wool, 50 g Iron accelerator, premium, 908 g Iron accelerator, high purity, 454 g Tungsten-tin accelerator, 750 g Tube of high vacuum grease Spare and wear parts Metal dust filter Balston filter Paper filters, 10 pieces Combustion coil Cleaning mechanism complete Upper furnace closing Cleaning brush for combustion tube Combustion tube Oscillator tube Ceramic heat shield for brush Pedestal Cleaning brush for metal dust filter Metal brush Brush O-ring 35x O-ring 38x O-Ring 8x1.5 for catalyst furnace O-Ring 9x Quarz tube Reagent tube 240x Reagent tube 240x20/16.4 General Tools and Accessories Spatula, 1 piece, M size Spatula, 1 piece, L size Spoon, 1 piece, for dosing sample and accelerator in CS series Tongs for ceramic crucibles and boats, 220 mm, 1 piece for CS series Tweezers (160 mm), curved, 1 piece page 4/5

5 Powered by TCPDF ( Product Information Tweezers (145 mm), straight, 1 piece Set with 6 spatula and 1 tweezers for multiple weighing procedures page 5/5