Validation of Nitrous Oxide Using SKC Passive Sampler SKC Inc. 863 Valley View Road Eighty Four, PA 15330

Size: px
Start display at page:

Download "Validation of Nitrous Oxide Using SKC Passive Sampler SKC Inc. 863 Valley View Road Eighty Four, PA 15330"

Transcription

1 Validation of Using SKC Passive Sampler SKC Inc. 863 Valley View Road Eighty Four, PA 15330

2 VALIDATION OF A DIFFUSIVE SAMPLER (SKC ) for NITROUS OXIDE Cynthia Kuhlman Linda Coyne SKC, Inc. 863 Valley View Road Eighty Four, PA ABSTRACT A partial validation was performed on the ability of a diffusive sampler (Cat. No , SKC Inc., Eighty Four, PA, U.S.A.) to accurately sample nitrous oxide in workplace air. Validation was conducted as outlined in a modified NIOSH protocol. A desorption efficiency (DE) study was performed using samplers spiked with nitrous oxide at 0.1, 0.25, 0.5, 1, and 2 times the PEL (50 ppm) for an 8-hour period. The desorption efficiency was determined to be 100% with a relative standard deviation (RSD) of 6.2%. The uptake rate was determined for samplers exposed to nitrous oxide levels ranging from to 890 ppm-hr and at 25 0 C and 20 to 80% relative humidity (RH). The mean uptake rate of these experiments was 0.81 ml/min and was based on the combined results of 66 samplers exposed and analyzed within 24 hours. A study of reverse diffusion showed that reverse diffusion does not take place. A storage study indicated that exposed samples could be stored for up to three weeks in a freezer and for one week at ambient temperatures. Samples were transferred from the passive sampler to a thermal desorption tube, desorbed, and analyzed using an electron capture detector (ECD). INTRODUCTION Nitrous oxide or N 2 O (CAS# ) is also known as dinitrogen monoxide or laughing gas. It is a colorless gas with a sweetish odor and taste and is very stable and inert chemically at room temperatures. Nitrous oxide is not classifiable as a human carcinogen; however, it is a narcotic at high concentrations. Although nitrous oxide is non-flammable it can support combustion and can form an explosive mixture with air. Nitrous oxide is used as an anesthetic in dentistry and surgery, as a propellant gas in food aerosols such as whipped cream, and as an inert gas to displace bacteria-inducing oxygen when filling packages of chips and other similar snack foods. Nitrous oxide is also used for leak detection and in motor racing as an oxidizer to increase the power output of engines. At high concentrations, nitrous oxide is an asphyxiant. Exposures to lower levels affect the central nervous system, the cardiovascular system, the liver, the hematopoietic system, and the reproductive system. Page 1 of 14

3 Epidemiological studies have shown increased risks of spontaneous abortion, premature delivery, and involuntary infertility among occupationally exposed populations. ACGIH used a guideline of 50 ppm based on an 8-hour TWA, NIOSH recommends 25 ppm as a time-weighted average, and OSHA does not currently regulate nitrous oxide. (1) Techniques available for sampling nitrous oxide include pumping ambient air through a spectrophotometer sample cell or filling a sample bag with a personal sampling pump at a calculated rate over the duration of the sampling period. The bag sample should be analyzed within two hours after completion of sampling to minimize compound loss. (2) These methods are highly accurate and reproducible. However, they are expensive, cumbersome, and analysis requires special equipment. The SKC Diffusive Samplers are inexpensive, simple to operate, unobtrusive, the sample can be stored up to three weeks in a freezer, and analysis is preformed with standard analytical equipment. OSHA has tested diffusive samplers for various methods and for monitoring exposures in the work environment. (3,4,5) Diffusive samplers have been evaluated recently for sampling lower levels commonly found in indoor air. (6,7) Detection of these lower levels is obtained easily when the diffusive sample is combined with thermal desorption. The validation of the diffusive sampler is important because the sampling can be affected by many environmental conditions such as temperature, humidity, chemical hazard concentrations, time of exposure, wind velocity, atmospheric pressure, sampler orientation, interfering compounds, and reverse diffusion of the compound of interest from the surface of the sorbent. (8,9) ASTM International (American Society for Testing and Materials) and ANSI (American National Standard Institute) have published guidelines for the validation of diffusive samplers. (10,11,12) ANSI has published a Standard Guide that describes testing to determine the effects of the factors listed above (13). Some of these are implemented in the SKC validation program. The validation of the SKC Diffusive Sampler for nitrous oxide was based on these guidelines and the factors tested qualify this study as a partial validation of the Diffusive Sampler for nitrous oxide. The disadvantages of diffusive samplers are that they cannot function in a static environment or at low wind levels. However, a study in the U.K. has shown typical workplace air velocities to be around a mean of 30 to 40 cm sec -1 (14), which is sufficient for the operation of diffusive samplers. Page 2 of 14

4 EXPERIMENTAL Reagents and Apparatus Nitrous oxide (Scott Specialty Gases [certified master grade], Plumsteadsville, PA,U.S.A.) was separated on a 30-meter, 0.32-mm ID CARBOPLT capillary column (Agilent Technologies, Wilmington, DE, U.S.A.). Pure nitrous oxide gas was used to make various concentration levels for a standard five-point curve. The standards were made daily and desorbed with the samples by thermal desorption and analyzed with an ECD. The Thermal Desorber was a Perkin Elmer Turbo Matrix and the GC was an Agilent 5890 with an electron capture detector. See Appendix A for GC and thermal desorber conditions. The diffusive sampler used was the SKC Diffusive Sampler containing 736 mg of washed Molecular Sieve 5A 60/80 (Alltech, State College, PA, U.S.A.). Standard atmospheres around the TLV (50 ppm, 25 o C, 80% RH) were generated using an atmospheric chamber. The atmosphere was calibrated using thermal desorption tubes packed with Molecular Sieve 5A 60/80. The analytical limit of detection for nitrous oxide was 2 ppm-hr per sampler. PROCEDURES The analytical recovery (desorption efficiency) was determined by using four samplers at each exposure level to be tested. The loading was based on equivalent 8-hour exposures to 0.1, 0.25, 0.5, 1, and 2 times the inhouse limit value (50 ppm). A known concentration of nitrous oxide was injected directly into the sampler for each expected exposure level based on the estimated sampling rate. The samplers were capped and stored at room temperature overnight to equilibrate, and were analyzed the following day. The amount recovered on analysis was determined as a percentage of the initial loading. The calculated uptake rate of the samplers for nitrous oxide was verified for concentrations in the range of to 890 ppm-hr at 25 0 C and relative humidity levels from 20 to 80%. Four samplers were exposed simultaneously to the atmosphere for each time period. Sampling times ranged from 15 minutes to 8 hours. After exposure, the molecular sieve was transferred from the samplers to a thermal desorption tube and analyzed. A set of 40 samplers was spiked with 41 µg of nitrous oxide for the storage study. After the samples equilibrated for two hours, 16 of the samples were run for Day-0 of the storage study. The remaining 24 samples were transferred to vials and divided into two sets. One set was stored at room temperature and the other set was stored at -8 0 C. Four samples from each set were analyzed weekly for three weeks. Page 3 of 14

5 A study of possible reverse diffusion in non-constant atmospheres was performed by exposing eight samplers to177.5 ppm-hr nitrous oxide at 25 0 C and 60% relative humidity for four hours. The samples were removed from the chamber and capped while the nitrous oxide was purged from the chamber. Then four of the samplers were uncapped and placed back into the chamber for another four hours. All eight samplers were analyzed the same day. Throughout the study, thermal desorption tubes packed with molecular sieve were used to determine the true concentration of the atmosphere. Samples were taken at 20 ml/min with a Pocket Pump (Cat. No MH, SKC Inc., Eighty Four, PA, U.S.A.). RESULTS and DISCUSSION The desorption efficiency results for nitrous oxide on the diffusive samplers are provided in Table 1. The mean recovery for nitrous oxide on the SKC Diffusive Samplers was 100% (6.2% RSD). The nitrous oxide results were consistent even at the low levels of concentration (Table 1). The results of the reverse diffusion study are shown in Table 2 and indicate that reverse diffusion of nitrous oxide does not take place on SKC Diffusive Samplers. Tables 3 through 5 show the uptake rate at different concentrations, temperatures, and relative humidities. The mean sampling rate of nitrous oxide at to 890 ppm-hr was determined to be 0.81 ml/min with an RSD of 12% and was based on the combined results of 76 samplers exposed and analyzed within 24 hours. The sampling rate is not affected by changes in concentration, temperature, or relative humidity. The storage stability results show that there was less than 10% change between the initial week and the third week period when the samplers were stored in the freezer. The samples can store for one week at room temperature. All storage stability results are shown in Table 6. QA RESULTS All analyses were run with a five-point calibration curve and check standards. The correlation coefficient for the calibration curves was 0.95 or better and all the check standards were within 10% of the original value. Thermal desorption tubes used to calculate the ppm concentration of the atmosphere showed that the appropriate nitrous oxide atmosphere was achieved. Page 4 of 14

6 CONCLUSION The SKC Diffusive Samplers have been partially validated for sampling nitrous oxide in occupational hygiene situations according to a modified NIOSH protocol. Nitrous oxide can be sampled accurately with the SKC Diffusive Samplers at an experimentally determined uptake rate of 0.81 ml/min (RSD ± 12%). The data obtained through this study shows that the diffusive samplers can be used at concentrations up to 890 ppm-hr and at various humidities. A desorption efficiency correction of 100% should be applied to Diffusive Samplers. The SKC Diffusive Samplers show at least a 91% recovery when stored up to three weeks in the freezer (-8 0 C). Page 5 of 14

7 REFERENCES 1) Occupational Safety and Health Guideline for, OSHA, 2) Burroughs, G.E. and Woebkenberg, M.L., Method 6600:, NIOSH Manual of Analytical Methods (NMAM), Fourth Edition, NIOSH, Aug ) Yost, C. and Harper, M., Validation of Parachlorobenzotrifluoride, Benzotrifluoride, and Monochlorotoluene on Diffusive Samplers, Applied Occup. and Environ. Hyg., Vol. 15, No. 1, Jan. 2000, pp ) Hendricks, W., Method 1004: 2-Butanone (MEK) Hexone (MIBK), OSHA Analytical Methods Manual, Sept ) Harper, M. and Fiore, A., Determination of Methyl tert-butyl Ether in Air Using a Diffusive Sampler, Applied Occup. and Environ. Hyg., Vol. 10, No. 7, July 1995, pp ) Hendricks, W., The Marines Project: A Laboratory Study of Diffusive Sampling/Thermal Desorption/Mass Spectrometry Techniques for Monitoring Personnel Exposure to Toxic Industrial Chemicals, OSHA Salt Lake Technical Center, Salt Lake City, UT, Apr ) Hendricks, W., Performance of SKC Ultra Passive Samplers Containing Carboxen 1016, Carbotrap Z, or Chromosorb 106 when Challenged with a Mixture Containing Twenty of OSHA SLTC s Top Solvent Analytes, OSHA Salt Lake Technical Center, Salt Lake City, UT, Feb ) Harper, M. and Purnell, C.J., Diffusive Sampling A Review, Am. Ind. Hyg. Assoc. Journal, Vol. 48, Issue 3, Mar. 1987, pp ) Guild, L. V., Dietrich, D. F., and Moore, G., Assessment of Sampling Accuracy in Diffusive Samplers, Am. Ind. Hyg. Assoc. Journal, Vol. 52, Issue 5, May 1991, pp ) Cassinelli, M. E., Hull, R. D., Crable, J. V., and Teass, A. W. Protocol for the Evaluation of Passive Monitors, In Diffusive Sampling: An Alternative Approach to Workplace Air Monitoring, A. Berlin, R. H. Brown, and K. J. Saunders, eds., Royal Society of Chemistry, London, 1987, pp ) Harper, M. and Guild, L.V., Experience in the Use of the NIOSH Diffusive Sampler Evaluation Protocol, Amer. Ind. Hyg Assoc. Journal, Vol. 57, Issue 12, Dec. 1996, pp ) ASTM Standard D : Standard Practice for Evaluating the Performance of Diffusive Samplers, ASTM Annual Book of Standards, Section 11, Vol , ASTM International, West Conshohocken, PA, ) ANSI/ISEA Standard : American National Standard for Air Sampling Devices- Diffusive Type for Gases and Vapors in Working Environments, ISEA, Arlington, VA, 1998 (standard reaffirmed in 2003) 14) Baldwin, P. E. J. and Maynard, A. D., A Survey of Wind Speeds in Indoor Workplaces, Annals of Occupational Hygiene, Vol. 42, No. 5, May 1998, pp Page 6 of 14

8 Table 1. Analytical Recovery on SKC Diffusive Samplers PEL Spike (µg) Recovered (µg) Recovery (%) Mean: 113% Std. Dev.: RSD: 6.2% Page 7 of 14

9 Table 2. Reverse Diffusion ppm-hr on SKC Diffusive Samplers (25 0 C, 60% RH) 4-hr Exposure ppm-hr 4-hr Exposure ppm-hr and 4-hr Exposure to 0 ppm-hr µg µg µg µg µg µg µg µg Mean: µg Mean: µg Std. Dev.: 1.11 Std. Dev.: 0.67 RSD: 7.2% RSD: 4.4% Page 8 of 14

10 Table 3. Sampling Rate and Capacity on SKC Diffusive Samplers (25 0 C, 60% RH) ppm-hr Uptake (µg) Sampling Rate (ml/min) Mean: ml/min Std. Dev.: RSD: 11.2% Page 9 of 14

11 Table 4. Sampling Rate and Capacity on SKC Diffusive Samplers (25 0 C, 80% RH) ppm-hr Uptake (µg) Sampling Rate (ml/min) Mean: ml/min Std. Dev.: RSD: 13.7% Page 10 of 14

12 Table 5. Sampling Rate and Capacity on SKC Diffusive Samplers (25 0 C, 20% RH) ppm-hr Uptake (µg) Sampling Rate (ml/min) Mean: ml/min Std. Dev.: RSD: 11.4% Page 11 of 14

13 Table 6. Storage Study on SKC Diffusive Samplers 40-µg Spike Recovery (%) Storage Period Freezer (-8 C) Ambient Week Week Week Week Page 12 of 14

14 Appendix A GC/ATD Conditions GC Conditions: Agilent 5890 Column: Oven: Agilent CARBOPLT, 30 m, 0.32-mm ID, 1.5-µm film C isothermal, hold 10 minutes Range: 4 Attn: 0 Detector: ECD, C Injector: None Transfer line connects directly to the GC column. Thermal Desorber: Perkin Elmer Turbo Matrix ATD Temp: Valve: C Transfer: C Tube: C Rate: 40 0 C/s Trap: HIGH C, LOW C Timing: Purge: 1 minute Desorb: 10 minutes Trap hold: 10 minutes Option: Injection/tube 1 Inlet split: disabled Outlet split: enabled Operating mode: 2-Stg Desorb PNU: Outlet split: 100 ccm Desorb: 100 ccm Column: 16 psi Page 13 of 14

Validation of Sulfur Dioxide Using. SKC UME x 200 Passive Sampler Cat. No Page 1 of 14 Publication 1781 Rev 1607

Validation of Sulfur Dioxide Using. SKC UME x 200 Passive Sampler Cat. No Page 1 of 14 Publication 1781 Rev 1607 Validation of Sulfur Dioxide Using SKC UME x 200 Passive Sampler Cat. No. 500-200 Page 1 of 14 Publication 1781 Rev 1607 Research Report Validation of Sulfur Dioxide Using SKC UME x 200 Passive Sampler

More information

Formulas: See Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1. METHOD: 2555, Issue 1 EVALUATION: PARTIAL Issue 1: 15 March 2003

Formulas: See Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1. METHOD: 2555, Issue 1 EVALUATION: PARTIAL Issue 1: 15 March 2003 KETONES I 2555 Formulas: See Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1 METHOD: 2555, Issue 1 EVALUATION: PARTIAL Issue 1: 15 March 2003 OSHA : See Table 2 NIOSH: See Table 2 ACGIH: See Table 2 PROPERTIES:

More information

ANILINE, o-toluidine, AND NITROBENZENE 2017

ANILINE, o-toluidine, AND NITROBENZENE 2017 ANILINE, o-toluidine, AND NITROBENZENE 2017 (1) Aniline C 6 H 7 N MW: (1) 93.13 CAS: (1) 62-53-3 RTECS: (1) BW6650000 (2) o-toluidine C 7 H 9 N (2) 107.2 (2) 95-53-4 (2) XU2975000 (3) Nitrobenzene C 6

More information

Principles, Practices, and Equipment for Performance Evaluation of Personal Monitoring Badges. Maria D.R. Peralta, PhD and C.R.

Principles, Practices, and Equipment for Performance Evaluation of Personal Monitoring Badges. Maria D.R. Peralta, PhD and C.R. Principles, Practices, and Equipment for Performance Evaluation of Personal Monitoring Badges Maria D.R. Peralta, PhD and C.R. Manning, PhD, CIH 1 Personal Sampling has arisen In response to the Occupational

More information

Research Report. Determination of Formaldehyde Using The SKC UMEx 100 (Cat. No ) Diffusive Sampler. Abstract

Research Report. Determination of Formaldehyde Using The SKC UMEx 100 (Cat. No ) Diffusive Sampler. Abstract Research Report Determination of Formaldehyde Using The SKC UMEx 100 (Cat. No. 500-100) Diffusive Sampler Abstract A sampling method for formaldehyde in air using a UMEx 100 Diffusive Sampler has been

More information

Laboratory Validation of AT575 Nitrous Oxide Vapor Sampler

Laboratory Validation of AT575 Nitrous Oxide Vapor Sampler These tests for the evaluation of diffusive air samplers were conducted within the guidelines described in ANSI 104-1998. AT575 Nitrous Oxide Vapor Sampler Assay Technology s 575 nitrous oxide (N 2 O)

More information

New Developments in Thermal Desorption (TD) Tube & Canister Technologies for Collection & Analysis of Soil Gas

New Developments in Thermal Desorption (TD) Tube & Canister Technologies for Collection & Analysis of Soil Gas New Developments in Thermal Desorption (TD) Tube & Canister Technologies for Collection & Analysis of Soil Gas Presented by Roy Smith, M.Sc., MBA, C.Chem. Why Use TD Tube or Canister Sampling For Collection

More information

Passive Monitoring: A Guide to Sorbent Tube Sampling for EPA Method 325. Nicola Watson July 2015

Passive Monitoring: A Guide to Sorbent Tube Sampling for EPA Method 325. Nicola Watson July 2015 Passive Monitoring: A Guide to Sorbent Tube Sampling for EPA Method 325 Nicola Watson July 2015 EPA 325 - Refinery perimeter monitoring Revised Federal regulation (CFR 40, part 60 and 63) to be implemented

More information

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools Application Note Abstract The Commonwealth of Massachusetts Department of Environmental Protection

More information

THIOPHANATE-METHYL in air 5606

THIOPHANATE-METHYL in air 5606 THIOPHANATE-METHYL in air 5606 C 12 H 14 N 4 O 4 S 2 MW: 342.40 CAS: 23564-05-8 RTECS: BA3675000 METHOD: 5606, Issue 1 EVALUATION: PARTIAL Issue 1: 15 March 2003 OSHA: N/A NIOSH: N/A ACGIH: N/A PROPERTIES:

More information

HYDROCARBONS, BP C 1500

HYDROCARBONS, BP C 1500 HYDROCARBONS, BP 36 216 C 1500 FORMULA: Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1 METHOD: 1500, Issue 3 EVALUATION: PARTIAL Issue 1: 15 August 1990 Issue 3: 15 March 2003 OSHA : Table 2 NIOSH: Table

More information

MERCAPTANS, METHYL-, ETHYL-, and n-butyl- 2542

MERCAPTANS, METHYL-, ETHYL-, and n-butyl- 2542 MERCAPTANS, METHYL-, ETHYL-, and n-butyl- 2542 SH MW: 48.11 CAS: 74-93-1 RTECS: PB4375000 SH 62.13 75-08-1 KI9625000 SH 90.19 109-79-5 EK6300000 METHOD: 2542, Issue 1 EVALUATION: PARTIAL Issue 1: 15 August

More information

Hydrocarbon Exposures During Tank Gauging and Sampling Operations

Hydrocarbon Exposures During Tank Gauging and Sampling Operations Hydrocarbon Exposures During Tank Gauging and Sampling Operations Todd Jordan, MSPH, CIH Director, USDOL/OSHA Health Response Team NORA Oil and Gas Sector Council Meeting March 19, 2015 Acknowledgements

More information

Low-level Detection of Ethanol, 1,4-Dioxane, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator

Low-level Detection of Ethanol, 1,4-Dioxane, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator Application Note 25370206 Low-level Detection of,, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator Keywords Eclipse Fuel Oxygenates Methanol Model 4552 Purge and Trap Sample Concentrator

More information

Test Sampling Matrix Location Treatment Samples

Test Sampling Matrix Location Treatment Samples Measurement of airborne volatile fatty acids emitted from dairy cows and their waste using sorbent tubes sampling and thermal desorption/gc-ms analysis (in collaboration with Dr. F. Mitloehner, UC Davis)

More information

Gas Chromatography SSI-GC-003

Gas Chromatography SSI-GC-003 Gas Chromatography SSI-GC-003 Trace Oxygenated Hydrocarbons in Liquid Hydrocarbon Streams by Multidimensional GC Introduction It has been widely accepted that oxygenates are related to corrosion and fouling

More information

AWS F1.5M:2003 An American National Standard. Methods for Sampling and Analyzing Gases from Welding and Allied Processes

AWS F1.5M:2003 An American National Standard. Methods for Sampling and Analyzing Gases from Welding and Allied Processes AWS F1.5M:2003 An American National Standard Methods for Sampling and Analyzing Gases from Welding and Allied Processes Key Words Welding gases, air sampling, analytical methods, ozone, carbon monoxide,

More information

Historical Review: Fast & micro Gas Chromatography

Historical Review: Fast & micro Gas Chromatography Historical Review: Fast & micro Gas Chromatography Dr. Ed Overton, Professor Emeritus Department of Environmental Sciences Louisiana State University School of the Coast & Environment Presenter: John Crandall

More information

Chambers and Instrumentation CHAMBERS. STORA: 22 m 3 chamber

Chambers and Instrumentation CHAMBERS. STORA: 22 m 3 chamber Chambers and Instrumentation CHAMBERS STORA: 22 m 3 chamber The chamber is made of electropolished stainless steel with inner dimensions 3.2 x 2.9 x 2.4 meters (breadth x length x height). The volume of

More information

Collection of Samples for Analysis of Volatile Organic Compounds (VOCs) in Indoor Air Using Sorbent Tubes. BAA Guide October 1, 2007

Collection of Samples for Analysis of Volatile Organic Compounds (VOCs) in Indoor Air Using Sorbent Tubes. BAA Guide October 1, 2007 BERKELEY ANALYTICAL ASSOCIATES 815 Harbour Way South, Unit 6 Richmond, CA 94804-3612 Tel: 510-236-2325 Fax: 510-236-2335 Email: baalab@berkeleyanalytical.com www.berkeleyanalytical.com Collection of Samples

More information

Corn Syrup Analysis E-1a- 1. ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

Corn Syrup Analysis E-1a- 1. ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) Corn Syrup Analysis E-1a- 1 ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) PRINCIPLE Isovaleraldehyde (IVA) and acetaldehyde are released from the syrup with the aid of dilute phosphoric acid and

More information

Determination of Volatile Organic Compounds (VOCs) present in the interiors of car by using GCMS/MS with static and dynamic headspace

Determination of Volatile Organic Compounds (VOCs) present in the interiors of car by using GCMS/MS with static and dynamic headspace PO-CON44E Determination of Volatile Organic Compounds (VOCs) present in the interiors of car by using GCMS/MS ASMS 20 ThP 138 Sanket Chiplunkar, Ankush Bhone, Durvesh Sawant, Dheeraj Handique, Prashant

More information

VOC-TD Application Manual

VOC-TD Application Manual VOCs Diffusive Sampling Device VOC-TD Application Manual Revision 1.0B Supelco Supelco Park Bellefonte, PA 16823 Thank you for purchasing the Supelco VOC-TD Diffusive Sampler. This unique diffusive sampler

More information

TurboMatrix Headspace and Headspace Trap Samplers

TurboMatrix Headspace and Headspace Trap Samplers S P E C I F I C A T I O N S Gas Chromatography TurboMatrix Headspace and Headspace Trap Samplers For added flexibility to meet your laboratory s changing requirements, five different PerkinElmer TurboMatrix

More information

2.1 This procedure provides steps and conditions for the determination of Vapor Space Organic Compounds in ground waters, and industrial waste waters.

2.1 This procedure provides steps and conditions for the determination of Vapor Space Organic Compounds in ground waters, and industrial waste waters. 1. TITLE: MSD Analytical Method for Vapor Space Organics 2. PURPOSE: 2.1 This procedure provides steps and conditions for the determination of Vapor Space Organic Compounds in ground waters, and industrial

More information

Neue Möglichkeiten zur Überwachung der Umgebungsluft mit Thermodesorption und GC/MS

Neue Möglichkeiten zur Überwachung der Umgebungsluft mit Thermodesorption und GC/MS Neue Möglichkeiten zur Überwachung der Umgebungsluft mit Thermodesorption und GC/MS Bernhard Rothweiler Senior Applications Chemist Agilent Technologies Europe, Middle East, Africa and India What is Thermal

More information

SCION GC Series. The Gas Chromatographers Choice for Separations

SCION GC Series. The Gas Chromatographers Choice for Separations SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Scion Instrument s tradition of innovation and product reliability have combined to create the next generation

More information

SCION GC Series. The Gas Chromatographers Choice for Separations

SCION GC Series. The Gas Chromatographers Choice for Separations SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Scion Instrument s tradition of innovation and product reliability have combined to create the next generation

More information

Title: Standard Operating Procedure for Sample Collection Using Tedlar Bags and Adsorbent Tubes

Title: Standard Operating Procedure for Sample Collection Using Tedlar Bags and Adsorbent Tubes Procedure No: SOP-015 Revision No: 1.0 (January 19, 2011) Page No.: 1 of 11 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting Tedlar

More information

Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D*

Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D* Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D* UCT Products: ECUNIPAH (ENVIRO-CLEAN Universal PAH/DRO 2000 mg/83 ml cartridge) Method Summary Method 8015D may be used

More information

C146-E096 TD-20. Shimadzu Thermal Desorption System

C146-E096 TD-20. Shimadzu Thermal Desorption System C146-E096 TD-20 Shimadzu Thermal Desorption System Thermal Desorption Thermal desorption systems trap target substances carried in mobile phase by adsorption in a sample tube packed with adsorbent. Then,

More information

Analysis of Semivolatiles by GC/FID using the Ultra Inert Inlet Liners with Wool

Analysis of Semivolatiles by GC/FID using the Ultra Inert Inlet Liners with Wool Analysis of Semivolatiles by GC/FID using the Ultra Inert Inlet Liners with Wool Application Note Environmental Authors Limian Zhao and Bill Wilson Agilent Technologies, Inc. 2850 Centerville Road Wilmington,

More information

Residual Solvents in medical cannabis analysis using the SRI 8610C FID GC June 2014

Residual Solvents in medical cannabis analysis using the SRI 8610C FID GC June 2014 June 2014 The SRI Medical Herb Potency 8610C GC is shown at right. This GC can also be used to test for residual solvents (i.e. butane, acetone, gasoline residue, etc.) in medical cannabis. These solvents

More information

A Spatial Investigation of Highly-Resolved Environmental Benzene Concentrations Using Passive Sampling

A Spatial Investigation of Highly-Resolved Environmental Benzene Concentrations Using Passive Sampling A Spatial Investigation of Highly-Resolved Environmental Benzene Concentrations Using Passive Sampling 2011-A-72-SP-AWMA Samantha C. Fridh, Amy L. Stuart University of South Florida, Department of Environmental

More information

Harmonising Analysis of VOCs from Spray Polyurethane Foam Insulation. Caroline Widdowson, BSc, PhD, MRSC

Harmonising Analysis of VOCs from Spray Polyurethane Foam Insulation. Caroline Widdowson, BSc, PhD, MRSC Harmonising Analysis of VOCs from Spray Polyurethane Foam Insulation Caroline Widdowson, BSc, PhD, MRSC Measuring Emissions from Spray Polyurethane Foam (SPF) Insulation ASTM D22.05 Indoor Air International

More information

Standard Operation Procedure (SOP) for the Extraction and Instrument Analysis of Siloxanes from Cartridges

Standard Operation Procedure (SOP) for the Extraction and Instrument Analysis of Siloxanes from Cartridges Standard Operation Procedure (SOP) for the Extraction and Instrument Analysis of Siloxanes from Cartridges Created March 17, 2016 Updated April 14, 2016 Dr. Rachel Marek (Asst. Research Scientist) 4105

More information

Performance Characteristics of the Agilent 1220 Infinity Gradient LC system

Performance Characteristics of the Agilent 1220 Infinity Gradient LC system Performance Characteristics of the Agilent 122 Infinity Gradient LC system An integrated LC system for conventional LC and UHPLC Technical Overview 7 5 4 3 2 1.5 1 1.5 2 2.5 3 Introduction The Agilent

More information

TECHNICAL MEMORANDUM. Pamela Moss, EA Team (EA Engineering, Science, and Technology, Inc., PBC/ CH2M HILL)

TECHNICAL MEMORANDUM. Pamela Moss, EA Team (EA Engineering, Science, and Technology, Inc., PBC/ CH2M HILL) EA Engineering, Science, and Technology, Inc., PBC 2363 N. Hill Field Road, Suite 104 Layton, Utah 84041 Telephone: 385-393-4982 Fax: 801-779-0294 www.eaest.com TECHNICAL MEMORANDUM TO: FROM: Lynne Welsh,

More information

Beyond Canisters (SUMMAs): Passive and Active Samplers and International Perspective

Beyond Canisters (SUMMAs): Passive and Active Samplers and International Perspective Beyond Canisters (SUMMAs): Passive and Active Samplers and International Perspective Joe Odencrantz, Ph.D., P.E. Tri-S Environmental Newport Beach, California December 1, 2010 Report Documentation Page

More information

Material Safety Data Sheet

Material Safety Data Sheet Material Safety Data Sheet 1. PRODUCT AND COMPANY IDENTIFICATION Product Name: 2P-10 Aerosol Activator Part Number: 2P-10 SOLO ACT 12OZ Revision Date: November 13, 2007 Company Name: FastCap LLC 3725 Irongate

More information

Ethylene Oxide. Fully validated method. This method has been subjected to the established evaluation procedures of the Methods Development Team.

Ethylene Oxide. Fully validated method. This method has been subjected to the established evaluation procedures of the Methods Development Team. Ethylene Oxide Method number: 11 Version: 2. Target concentration: 1. ppm (1.8 mg/m 3 ) OSHA PEL (TWA): 1. ppm (1.8 mg/m 3 ) Action Level:.5 ppm (.9 mg/m 3 ) Excursion Limit: 5. ppm (9 mg/m 3 ) (15 min)

More information

Environmental Air Monitoring

Environmental Air Monitoring Environmental Air Monitoring Dr Caroline Widdowson Material Emission Specialist cwiddowson@markes.com Dr Nicola Watson Environmental Specialist nwatson@markes.com What do we mean by air monitoring? Whichever

More information

Rapid Microchamber Tests for Screening Chemical Emissions from Car Trim in Accordance with ISO

Rapid Microchamber Tests for Screening Chemical Emissions from Car Trim in Accordance with ISO Application Note Rapid Microchamber Tests for Screening Chemical Emissions from Car Trim in Accordance with ISO 119- Abstract This Application Note demonstrates the efficiency and reliability of Markes

More information

CORESTA Recommended Method No. 57

CORESTA Recommended Method No. 57 Cooperation Centre for Scientific Research Relative to Tobacco Tobacco and Tobacco Products Analytes Sub-Group CORESTA Recommended Method No. 57 DETERMINATION OF WATER IN TOBACCO AND TOBACCO PRODUCTS BY

More information

Gas Chromatography Assignment Chem 543/443

Gas Chromatography Assignment Chem 543/443 Gas Chromatography Assignment Chem 543/443 1. Introduction Capillary gas chromatography (GC) is one of the most popular analytical techniques used in today s research. Its popularity is mainly due to efficient

More information

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water Application Note Environmental Authors Min Cai and Yun Zou Agilent Technologies Co. Ltd, 412 Ying Lun Road Waigaoqiao

More information

Advanced Monitoring Method for Air Environment by Ogawa Passive Sampler

Advanced Monitoring Method for Air Environment by Ogawa Passive Sampler Advanced Monitoring Method for Air Environment by Ogawa Passive Sampler Keiro Higuchi 1, D. R. Schaeffer 2 *, Koichiro Hirano 3 1 Ogawa & Co., Ltd. 3-1-25-501 Hio-cho, Nada-ku, Kobe 657-0029 Japan 2 Ogawa

More information

SAFETY DATA SHEET SDS NA-SC103. Section 1 Product and Supplier Identification. Product identifier used on the label: Carbon/Graphite Powder/Pellets

SAFETY DATA SHEET SDS NA-SC103. Section 1 Product and Supplier Identification. Product identifier used on the label: Carbon/Graphite Powder/Pellets SAFETY DATA SHEET Section 1 Product and Supplier Identification Product identifier used on the label: Carbon/Graphite Powder/Pellets Other means of identification: See list of products/grades in Section

More information

Validation Data for the Analysis of Chlorotalonil in Leek Using Mini-Luke Method Followed by GC-QqQ-MS/MS

Validation Data for the Analysis of Chlorotalonil in Leek Using Mini-Luke Method Followed by GC-QqQ-MS/MS Validation Data for the Analysis of Chlorotalonil in Leek Using Mini-Luke Method Followed by GC-QqQ-MS/MS CONTENTS 1. Aim and Scope 2 2. Short Description 2 3. Apparatus and Consumables 2 4. Chemicals

More information

EVALUATION OF THE USE OF LOW FLOW PASSIVE SAMPLING TECHNIQUE IN OFFSET PRINTING PLANTS

EVALUATION OF THE USE OF LOW FLOW PASSIVE SAMPLING TECHNIQUE IN OFFSET PRINTING PLANTS Nofer Institute of Occupational Medicine, Łódź, Poland ORIGINAL PAPERS International Journal of Occupational Medicine and Environmental Health 2006;19(4):228 34 DOI 10.2478/v10001-006-0031-8 EVALUATION

More information

SAFETY DATA SHEET SDS NA-MC101. Section 1 Product and Supplier Identification. Mechanical carbon products; grades that do not contain metal additives

SAFETY DATA SHEET SDS NA-MC101. Section 1 Product and Supplier Identification. Mechanical carbon products; grades that do not contain metal additives SAFETY DATA SHEET SDS NA-MC101 Section 1 Product and Supplier Identification Product identifier used on the label: Other means of identification: Uses (and restrictions): Supplier and contact information:

More information

Thermal Desorption Technical Support Note 87: A cryogen-free method for monitoring trace greenhouse gases in air. Key Words:

Thermal Desorption Technical Support Note 87: A cryogen-free method for monitoring trace greenhouse gases in air. Key Words: Thermal Desorption Technical Support Note 87: A cryogen-free method for monitoring trace greenhouse gases in air Key Words: Trace greenhouse gases, Kyoto Protocol, CFCs, HCFCs, UNITY 2, CIA 8, air toxics,

More information

Arc Carbons. Not applicable +1(864)

Arc Carbons. Not applicable +1(864) SAFETY DATA SHEET Section 1 Product and Supplier Identification Product identifier used on the label: Other means of identification: Uses (and restrictions): Supplier and contact information: Morgan Advanced

More information

METHOD 3 - GAS ANALYSIS FOR THE DETERMINATION OF DRY MOLECULAR WEIGHT. NOTE: This method does not include all of the

METHOD 3 - GAS ANALYSIS FOR THE DETERMINATION OF DRY MOLECULAR WEIGHT. NOTE: This method does not include all of the 312 METHOD 3 - GAS ANALYSIS FOR THE DETERMINATION OF DRY MOLECULAR WEIGHT NOTE: This method does not include all of the specifications (e.g., equipment and supplies) and procedures (e.g., sampling) essential

More information

Markes International Thermal Desorption G8125A, G8126A, G8127A, G8128A, G8131A, G8132A, G8136A, G8133A, MKI-MicroC Site Preparation Checklist

Markes International Thermal Desorption G8125A, G8126A, G8127A, G8128A, G8131A, G8132A, G8136A, G8133A, MKI-MicroC Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e 1

Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e 1 Application Note US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e 1 Abstract US EPA Method 8260 in conjunction

More information

Performance characteristics of the 1260 Infinity Quaternary LC system

Performance characteristics of the 1260 Infinity Quaternary LC system Performance characteristics of the 1260 Infinity Quaternary LC system The new standard in HPLC Technical Overview Introduction The Agilent 1260 Infinity LC system consists of modular units that operate

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis S(+)-Methamphetamine, Primary Standard M-020 Page 1 of 7 Catalog Number: M-020 Lot: Expiration: Description: Packaging: Storage: Shipping: Intended

More information

Analytical Method of Limit Test for Hexachlorobenzene. in Picloram TC and Method Validation Data*

Analytical Method of Limit Test for Hexachlorobenzene. in Picloram TC and Method Validation Data* Analytical Method of Limit Test for Hexachlorobenzene in Picloram TC and Method Validation Data* Test Method (ABCTM-2010-01-07) 1. Apparatus Electronic Balance GC/MS System with NCI Ultrasonic Water Bath

More information

Volatile Organic Compounds in Solids PBM

Volatile Organic Compounds in Solids PBM Organics Revision Date: Aug 15, 2014 Volatile Organic Compounds in Solids PBM Parameter Analytical Method Introduction Method Summary Parameter Applicability, MDLs Volatile Organic Compounds (VOCs) in

More information

SAFETY DATA SHEET SDS NA-MC112. Section 1 Product and Supplier Identification. Silicon carbide products, grades that contain boron carbide

SAFETY DATA SHEET SDS NA-MC112. Section 1 Product and Supplier Identification. Silicon carbide products, grades that contain boron carbide SAFETY DATA SHEET SDS NA-MC112 Section 1 Product and Supplier Identification Product identifier used on the label: Other means of identification: Uses (and restrictions): Supplier and contact information:

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis S(-)-Nicotine, Primary Standard N-008 Page 1 of 7 Catalog Number: N-008 H Lot: Expiration: October 2019 N Description: S(-)-Nicotine in Methanol.

More information

Application Note AN1501

Application Note AN1501 Application Note AN1501 Ultra-Trace Mercury Determination in Bottled Water, EPA Method 1631, Using the QuickTrace M-8000 CVAFS Introduction Mercury determination in water and foodstuffs is a crucial part

More information

Analysis of styrene leached from polystyrene cups using GCMS coupled with Headspace (HS) sampler

Analysis of styrene leached from polystyrene cups using GCMS coupled with Headspace (HS) sampler PO-CON464E Analysis of styrene leached from polystyrene cups using GCMS coupled with Headspace (HS) sampler ASMS 204 TP763 Ankush Bhone (), Dheeraj Handique (), Prashant Hase (), Sanket Chiplunkar (),

More information

Mechanical carbon products; grades that contain antimony

Mechanical carbon products; grades that contain antimony SAFETY DATA SHEET SDS NA-MC102 Section 1 Product and Supplier Identification Product identifier used on the label: Other means of identification: Uses (and restrictions): Supplier and contact information:

More information

DATE MSDS PRINTED & REVIEWED: 02/15/10 LAST FORMULA REVISION DATE: 06/17/09

DATE MSDS PRINTED & REVIEWED: 02/15/10 LAST FORMULA REVISION DATE: 06/17/09 Page 1 of 6 SECTION 1: DISTRIBUTOR IDENTIFICATION Distributor: FORMULA # 31A107N12A Uline Chemtrec 24-hour Emergency Phone: (800) 424-9300 2200 S. Lakeside Dr. Chemtrec Foreign Emergency Phone: (703) 527-3887

More information

Concentration % by weight Silicon Carbide % Antimony %

Concentration % by weight Silicon Carbide % Antimony % SAFETY DATA SHEET SDS NA-MC111 Section 1 Product and Supplier Identification Product identifier used on the label: Other means of identification: Uses (and restrictions): Supplier and contact information:

More information

A Presentation by OI Analytical. Oxygenate Analysis on the Eclipse Purge-and-Trap Sample Concentrator

A Presentation by OI Analytical. Oxygenate Analysis on the Eclipse Purge-and-Trap Sample Concentrator A Presentation by OI Analytical Oxygenate Analysis on the Eclipse Purge-and-Trap Sample Concentrator Introduction Oxygen-containing compounds are frequently used as fuel additives Boost octane rating Fuel

More information

Micro GCs: Current Capabilities and Future Trends

Micro GCs: Current Capabilities and Future Trends Micro GCs: Current Capabilities and Future Trends Analytical Rational, Data Quality Objectives, Cost, Time Sample Collection Sample Extraction Extract Analysis Data Analysis Decisions & Actions Detection

More information

Solutions for monitoring VOCs in air

Solutions for monitoring VOCs in air Solutions for monitoring VOCs in air AQE, Telford 14 th March Nicola Watson Environmental Specialist Markes International nwatson@markes.com Agenda Sampling VOCs in air focusing on: Online/Canister sampling

More information

Two Headspace Solutions for the Analysis of Dissolved Gases in Water by Method RSKSOP-175

Two Headspace Solutions for the Analysis of Dissolved Gases in Water by Method RSKSOP-175 Two Headspace Solutions for the Analysis of Dissolved Gases in Water by Method RSKSOP-175 Application Note Abstract Dissolved gas analysis has been performed using a simplified headspace method following

More information

Sampling and Analysis in Vapor Intrusion Investigations. Andy Rezendes, Technical Sales-Air Analysis

Sampling and Analysis in Vapor Intrusion Investigations. Andy Rezendes, Technical Sales-Air Analysis Sampling and Analysis in Vapor Intrusion Investigations Andy Rezendes, Technical Sales-Air Analysis Introduction General Principles Introduction to Measurement Techniques for VOCs in Air Principles of

More information

The Analysis of 4-Vinyl-Cyclo-Hexene (VCH) in Butadiene with the DVLS LGI Injector

The Analysis of 4-Vinyl-Cyclo-Hexene (VCH) in Butadiene with the DVLS LGI Injector The Analysis of 4-Vinyl-Cyclo-Hexene (VCH) in Butadiene with the DVLS LGI Injector Introduction In synthetic rubber manufacturing, 4-Vinylcyclohexene (VCH) can be found as a byproduct from the chemical

More information

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Student Learning Objectives High Performance Liquid Chromatography With the advent of relatively inexpensive and reliable pumps, the development of

More information

Measurement of Trihalomethanes in Water Using the Dielectric Barrier Discharge Electron Capture Detector and Headspace Extraction

Measurement of Trihalomethanes in Water Using the Dielectric Barrier Discharge Electron Capture Detector and Headspace Extraction Measurement of Trihalomethanes in Water Using the Dielectric Barrier Discharge Electron Capture Detector and Headspace Extraction Matthew Monagle AIC Corp. 5645B Paradise Blvd Albuquerque, NM 87114 Aicmm@flash.net

More information

ROGER LINDAHL 1 *, ANNA-SARA CLAESSON 2, MUHAMMAD AKHTAR KHAN 3 and JAN-OLOF LEVIN 1,4

ROGER LINDAHL 1 *, ANNA-SARA CLAESSON 2, MUHAMMAD AKHTAR KHAN 3 and JAN-OLOF LEVIN 1,4 Ann. Occup. Hyg., Vol. 55, No. 6, pp. 681 687, 2011 Ó The Author 2011. Published by Oxford University Press on behalf of the British Occupational Hygiene Society doi:10.1093/annhyg/mer039 Development of

More information

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE.

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE Authors: DANI Instruments Inc 257 Simarano Drive Marlborough, MA 01752 Ph: 508-281-6232

More information

Product Name: Color Guard(R) Tough Rubber Coating - Blue Page 1 of 5

Product Name: Color Guard(R) Tough Rubber Coating - Blue Page 1 of 5 Page 1 of 5 81810 1. CHEMICAL PRODUCT AND COMPANY IDENTIFICATION MATERIAL SAFETY DATA SHEET Page 01 of 05 Product Type: Solvent Based Coating Material. 2. COMPOSITION, INFORMATION ON INGREDIENTS Ingredients

More information

Fast On-Site Mine Safety Analysis by the Agilent 490 Micro GC

Fast On-Site Mine Safety Analysis by the Agilent 490 Micro GC Fast On-Site Mine Safety Analysis by the Agilent 490 Micro GC Application Note Micro Gas Chromatography, Mine gas analysis, Mine safety Authors Darren Brady Safety in Mines Testing and Research Station

More information

Available Laboratory Methods for Soil Gas Analysis

Available Laboratory Methods for Soil Gas Analysis Available Laboratory Methods for Soil Gas Analysis US EPA Soil Gas Sample Collection & Analysis Workshop San Diego, CA March 21-22, 2007 Gina M. Plantz NewFields Environmental Forensics Practice, LLC gplantz@newfields.com

More information

Project Summary Report

Project Summary Report Summary Report : Odour measurement and assessment, evaluation of odour treatment technologies 1. Background Gas phase samples are often collected from operating sewerage systems during investigations intended

More information

Liners and consumables for OPTIC Inlet. Liner Selection Guide

Liners and consumables for OPTIC Inlet. Liner Selection Guide Liners and consumables for OPTIC Inlet Liner Selection Guide Find the right liner Choosing a correct liner for a particular application must take several variables into account. The variables to consider

More information

Mechanical carbon products; grades that contain silver

Mechanical carbon products; grades that contain silver SAFETY DATA SHEET SDS NA-MC105 Section 1 Product and Supplier Identification Product identifier used on the label: Other means of identification: Uses (and restrictions): Supplier and contact information:

More information

Vapor Intrusion: The Conference An Analytical Perspective

Vapor Intrusion: The Conference An Analytical Perspective Vapor Intrusion: The Conference An Analytical Perspective January 23-24, 2014 Tod Kopyscinski TABLE OF CONTENTS What is Vapor Intrusion Analytical Methods Air Sampling Hardware and Procedure Analytical

More information

HENKEL CONSUMER ADHESIVES 02/05/02 AVON, OH TELEPHONE: (440) MATERIAL SAFETY DATA SHEET Page 01 of 05

HENKEL CONSUMER ADHESIVES 02/05/02 AVON, OH TELEPHONE: (440) MATERIAL SAFETY DATA SHEET Page 01 of 05 00-20273 00-20273 1. CHEMICAL PRODUCT AND COMPANY IDENTIFICATION Product Type: Bondfast Rubber Cement 2. COMPOSITION, INFORMATION ON INGREDIENTS MATERIAL SAFETY DATA SHEET Page 01 of 05 Ingredients CAS

More information

Modified Draft Indian Standard METHODS FOR CHEMICAL ANALYSIS OF STEELS

Modified Draft Indian Standard METHODS FOR CHEMICAL ANALYSIS OF STEELS Doc: MTD 4( 4850) For Comments Only Modified Draft Indian Standard METHODS FOR CHEMICAL ANALYSIS OF STEELS Part 20 DETERMINATION OF CARBON AND SULFUR BY INFRA RED ABSORPTION METHOD (FOR CARBON 0.0005 TO

More information

Analysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS

Analysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS Analysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS Authors: Conor Smith, Doug Fryer United Science Corp. 15911 Furuby Rd. Center City, MN 55012 USA Abstract

More information

The Use Of Automated Backflush on the 7890A/5975A GC-MS System. Application. Setup Example Using Biomarkers. Authors. Abstract.

The Use Of Automated Backflush on the 7890A/5975A GC-MS System. Application. Setup Example Using Biomarkers. Authors. Abstract. The Use Of Automated Backflush on the 7890A/5975A GC-MS System Setup Example Using Biomarkers Application Hydrocarbon Processing Authors Courtney Milner and Russell Kinghorn BST International 41 Greenaway

More information

Improving the Performance of Time Integrated Sampling of. Steel Canisters. Application Note: 902. Authors. Introduction. Abstract

Improving the Performance of Time Integrated Sampling of. Steel Canisters. Application Note: 902. Authors. Introduction. Abstract Improving the Performance of Time Integrated Sampling of TO14 Compounds into Stainless Steel Canisters Application Note: 902 Authors Dan Cardin Ken Langford Vinay Shetty Entech Instrurments, Inc Simi Valley,

More information

Material Safety Data Sheet

Material Safety Data Sheet Material Safety Data Sheet Section 1: Product and Company Identification U.S. Department of Labor Occupational Safety and Health Administration (Non-Mandatory Form). Format meets ANSI Z400.1-1998, OSHA

More information

Emergency Overview: DANGER! Extremely Flammable. Contents Under Pressure.

Emergency Overview: DANGER! Extremely Flammable. Contents Under Pressure. Page 1 of 6 SECTION 1: MANUFACTURER IDENTIFICATION MANUFACTURED BY: FORMULA # 38A798M12 Custom-Pak Products Inc. Chemtrec 24-hour Emergency Phone: (800) 424-9300 N118 W18981 Bunsen Drive Chemtrec Foreign

More information

MassPREP On-Line Desalting Cartridge

MassPREP On-Line Desalting Cartridge CONTENTS I. INTRODUCTION II. INSTALLING THE MASSPREP ON-LINE DESALTING CARTRIDGE INTO THE SENTRY.1 X 1 MM GUARD COLUMN HOLDER III. RECOMMENDED LC/MS SYSTEM CONFIGURATION TO MINIMIZE MS SOURCE CONTAMINATION

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 10 CONTENTS 1.0 SCOPE AND APPLICATION* 2.0 METHOD SUMMARY* 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS* 6.0 REAGENTS

More information

PAL SPME Fibers Optimized for Automation

PAL SPME Fibers Optimized for Automation PAL SPME Fibers Optimized for Automation PAL SPME Fiber Since its introduction by Pawliszyn et al. (ref. 1) Solid Phase Micro Extraction (SPME) has seen a tremendous development. SPME is a very effective

More information

US EPA Method Atomx XYZ and Agilent 7890B GC/5977A MS

US EPA Method Atomx XYZ and Agilent 7890B GC/5977A MS Application Note Environmental US EPA Method 524.3 Atomx XYZ and Agilent 7890B GC/5977A MS Author Amy Nutter Applications Chemist, Teledyne Tekmar Abstract US EPA Method 524.3 was used to determine the

More information

PAL SPME Fibers Optimized for Automation

PAL SPME Fibers Optimized for Automation PAL SPME Fibers Optimized for Automation PAL SPME Fiber Since its introduction by Pawliszyn et al. (ref. 1) Solid Phase Micro Extraction (SPME) has seen a tremendous development. SPME is a very effective

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis T-32 Page 1 of 7 (-)- 8 -THC, Primary Standard Catalog Number: T-32 Lot: Expiration: Description: October 22 (-)- 8 -THC in Methanol. Packaging: Solution

More information

Capillary GC Troubleshooting: a Practical Approach

Capillary GC Troubleshooting: a Practical Approach Capillary GC Troubleshooting: a Practical Approach Katherine Stenerson Gas Separations R&D Supelco, Supelco Park, Bellefonte, PA 16823 1999 Sigma-Aldrich Co. Outline Basic Troubleshooting Strategy Preventing

More information

Joette L. Steger, Joan T. Bursey, and David Epperson

Joette L. Steger, Joan T. Bursey, and David Epperson United States National Exposure Environmental Protection Research Laboratory Agency Research Triangle Park, NC 27711 Research and Development EPA/600/SR-97/140 April 1998 Project Summary Acetonitrile Field

More information

Evaluation of Volatile Organic Compounds and aldehydes emitted by a Anodized aluminium profile supply by ALUMINIOS CORTIZO according to ISO 16000

Evaluation of Volatile Organic Compounds and aldehydes emitted by a Anodized aluminium profile supply by ALUMINIOS CORTIZO according to ISO 16000 Saint Ouen l'aumone, the 22 th of June 2016 Evaluation of Volatile Organic Compounds and aldehydes emitted by a Anodized aluminium profile supply by ALUMINIOS CORTIZO according to ISO 16000 Report N D-310516-04389-002

More information