1 CHEM 3281 Experiment Ten Determination of Phosphate by Flow Injection Analysis Objective: The aim of the experiment is to investigate the experimental variables of FIA for a model system and then to use the system to determine the phosphate content in an unknown sample. Text Reference: Skoog, Holler and Nieman, Principles of Instrumental Analysis, 5th edition, Saunders College Publishing, Fort Worth, TX 1998, Ch 33. Introduction: In most analytical measurement systems, the signal is obtained on a discrete or batchwise basis. In certain circumstances, however, it can be advantageous to obtain analytical measurements on a continuous basis. This can allow the following advantages: 1. greater speed of analysis; 2. the ability to automate the analysis; 3. better reproducibility; 4. smaller sample volume. In automated analysis systems, solution is continuously pumped through tubes (with a peristaltic pump), the sample is injected into the stream and various reactions are carried out by mixing streams of sample and reagent, buffer, etc. The concentration of the product(s) is determined by a detector or transducer. Initially analysis in a flowing stream was done using air bubbles introduced in regular time intervals (segmented-flow analysis). This was done to prevent excess sample dispersion, to insure mixing of sample and reagents and to prevent cross-contamination. Ruzicka and Hansen (the co-inventors of FIA) found that the air bubbles were unnecessary in properly designed systems and that sample throughput could be much higher and simpler equipment could be used. (Air bubbles cause problems in most detection systems!) Flow Injection Analysis is now the method of choice in automated analysis. To use FIA knowledgeably, you must understand the principles of dispersion as they apply to FIA. Theory A rectangular "plug" of sample injected into a flowing stream will initially experience no dispersion and will assume the concentration profile shown in Fig. 1.a. If this plug passes a detector, the recorder readout of the peak would have the shape shown underneath. In practice, however, between the points of injection and detection, some dispersion will always have occurred. Physical dispersion is brought on by convection and diffusion.
2 Initially convection caused by laminar flow gives the profile in Fig 1.b. This gives rise to a parabolic head and tail and concomitant detector output. Further broadening is due to diffusion. Two types of diffusion can occur: radial and longitudinal. In narrow bore tubing, however, longitudinal diffusion is not significant and can be ignored. Radial diffusion (diffusion perpendicular to flow direction) is important in FIA and serves to insure adequate mixing and keeping the walls clean. Most FIA experiments are run under conditions where both convection and radial diffusion are contributing to dispersion. (Fig 1.c.) The dispersion, D, is given by: D = C o /C= q H o /H where: C o is the concentration of the injected sample (mol/l), C is the concentration of sample at detector (mol/l), q is a constant, H is the recorded peak height, and H o is the peak height with the cell full of undispersed colored solution. Dispersion will take place in the injection port, D i, in the tubing, D t, and the detector, D d such that the total dispersion, D total is given by: D total = D i D t D d Dispersion is classified into three categories for FIA: limited (D t = 1-3); medium (D t = 3-10); and large (D t 10). Limited dispersion, D = 1-3 : The rise of curve will have the form: C = C o ( 1 - exp {- ks}) and the fall curve C = C o exp {-ks}, where k is a system constant and s is the sample volume (ul). Here dispersion is due mostly to convection and gives rise to the classic Poisson distribution or "tailing peak" seen in the detector output in Fig 1.b. Medium Dispersion, D = 3-10: Here, a combination of convection and radial diffusion are creating the dispersion. Detector output should look like Fig 1.c. Dispersion should (1) increase linearly with pumping rate; (2) increase as the square root of traveled distance or residence time. Large Dispersion, D 10: Here, dispersion is caused by radial diffusion only and the
3 distribution is gaussian as seen in Fig.1.d. Most FIA experiments employ narrow tubing ( mm) and are conducted using medium flow rates (10-2 ml/min). Under these conditions, the dispersion is generally medium and mixing is due to radial diffusion. Sample cross-contamination is minimized by adequate time delay between injections. The delay period is determined by the extent of sample dispersion. The analytical procedure is based upon the following reactions. 7H 3 PO (NH 4 ) 6 Mo 7 O H + 7(NH 4 ) 3 PO 4-12MoO NH H 2 O (1) (NH 4 ) 3 PO 4-12MoO 3 + Ascorbic Acid blue Mo(V) complex (2) The intermediate product (NH 4 ) 3 PO 4-12MoO 3 is a yellow colored Mo(VI) complex. When this species is treated with ascorbic acid, a blue colored Mo(V) complex results and is detected spectrophotometrically. Experimental Apparatus: For the colorimetric analysis of phosphate, the flow manifold should be set up as follows: where: A-- is distilled water B-- is the acidified ammonium heptamolybdate solution ( M (NH 4 ) 6 Mo 7 O 24 and 0.2 M HNO 3 ) C --is 0.25% (w/v) ascorbic acid solution (with glycerine & Sb catalyst) A phosphate standard (~100 ug/ml) and a phosphate unknown will be provided.
4 Experimental Procedure: The experimental apparatus will be set up when you arrive to the lab. To begin, place all three of the inlet tubes into a large beaker of water and turn on the pump. Place the waste end of the line into the nearest sink. Allow the water to run through the system until all air bubbles are eliminated in the lines and detector. Preparation: 1. PUMPING SYSTEM. Set the pump speed to "4". Once you are assured that only water is exiting the waste line is at this speed, place the line in a 10 ml graduated cylinder and measure the time required to deliver 1-2 ml of water. Calculate the flow rate in ml/min (remember there are three inlet lines, so the measured flow rate at waste needs to be divided by three). Repeat this procedure for pump speeds of "1", 2, "6" and "8". The aqueous waste generated can be poured down the sink. Place the waste line into the sink. 2. DETECTOR. Open the FIALAB software. GO to Spectrometer and take a DARK SPECTRUM. Turn on the visible source and click on VOLTAGE & REPEATED SCAN. Go to SPECTROMETER SETUP. Make sure there are NO AIR BUBBLES in the detector flow cell. Lower the integration time until guassian power spectrum of tungsten lamp has the highest point about 3500 with no flattening. (Will have to switch back and forth between SPECTROMETER SETUP and SPECTROMETER ). Wavelength should be set at ~660 nm. 3. STANDARDS. Prepare standards containing 100 ppm, 50 ppm, 25 ppm, 12.5 ppm of phosphate by successive dilution of the stock solution with water. Use 50 ml volumetric flasks. (Remember C 1 V 1 = C 2 V 2 )! Obtain an UNKNOWN and record its number. 4. INJECTOR. Practice filling the 50 ul injection loop with a syringe equipped with a Luer fitting and injecting this into the sample stream. A counter clockwise rotation of the injector knob is appropriate for filling the loop; a clockwise rotation is appropriate for flushing the loop. Make sure you can fill the loop without introducing any air bubbles. System Optimization: The following study will provide you with understanding of the experimental variables affecting the dispersion and overall efficiency of FIA. Both reaction (1) & (2) are NOT instantaneous so you must optimize the system to allow enough time for the blue Mo(V) complex to form and not allow too much dispersion so the peak shape doesn t get too broad. The LINE LENGTH STUDY will show you what coil lengths to use.
5 One person must inject while the other runs the spectrometer Now place the inlet lines into their designated container A, B, or C. Be sure you get it into the correct bottle. Set the flow rate to 4 and fill all lines with the correct chemical solution. 1. LINE LENGTH STUDY : Observe the influence of tube length (in the mixing coil) on the peak shape by varying the length of tubing for each coil over the range 50 cm, 100 cm. and 200 cm. (Hold the other coil constant at 100 cm.) Be sure each inlet line is full of the correct solution, no air bubbles in the lines. You must learn how to run the spectrometer so that each run can be compared to the other runs. This most easily accomplished by forcing the X and Y axes to always have the same scale each run so that you can easily make the plots called for later on. Unfortunately this software is autoscaling both Absorbance and Time continuously which is not useful for this optimization. Plan to run the spectrometer for say 60 secs each run. Load the 100 ppm phosphate standard into the sample loop Take a REFERENCE before each run (Make the injector person check detector for air bubbles Turn on REPEATED SCAN and tell lab partner to inject. Run for 60 secs. Go to ANALYSIS mode and call up data. Your most recent data should be in TIME SERIES BY CHANNEL. Click on AXIS icon and scale the Y axis to same scale each time (say 0.8 to 0.2 absorbance units). (Can click on STYLE and deselect data point tags if desired ) Note: LOCAL MAX. and PEAK AREA are random numbers only-not functional software! Click on PRINTER icon and then click on Print on system page to obtain hardcopy. While computer person is scaling the run the injector person should turn off pump, put another coil length in place and turn pump back on to remove air bubbles. Vary coil I then coil II. By using the reaction coils of different lengths, you are allowing for longer times spans in which the reactions can occur. Determine from the data, the optimum reaction coil lengths and install them for the FLOW RATE STUDY below. Note that if the flow rate is too fast the blue product will form too late (past the detector). Adjust the flow rate appropriately if needed. FLOW RATE STUDY. With your optimized coil lengths in place, inject 50 ul samples of 100 ppm Phosphate Standard into the stream at flow rates of "2", "3", "4", and "6" and 8. Run the lowest flow rate first so you can determine how long to run the spectrometer each time. Start with 30 secs and see how long it takes for the blue peak to come through the system.
6 Load the 100 ppm phosphate standard into the sample loop Take a REFERENCE before each run (Make the injector person check detector for air bubbles Turn on REPEATED SCAN and tell lab partner to inject. Run for 30 secs or until you are sure the peak has cleared the system! Go to ANALYSIS mode and call up data. Click on AXIS icon and scale the Y axis to same scale each time (say 0.8 to 0.2 absorbance units). Click on PRINTER icon and then click on Print on that page to obtain hardcopy. While computer person is scaling the run the injector person should turn be adjusting the flow rate and loading the sample loop again. 3. THROUGHPUT STUDY: At the flow rate that provides the highest absorbance reading in the shortest amount of time, determine how rapidly it is possible to make successive injections without causing the resultant peaks to overlap. You decide how to do this experimentally! Do this by timing the interval between injections and monitoring the output on the spectrometer. The smallest time interval that will "resolve" the peaks is what you are trying to find!! Decide if you want to use PEAK AREA or PEAK HEIGHT for an analysis. Unknown Determination: Now that you understand how varying the flow rate and line length varies your output, set the optimum parameters (flow rate and coil length) for running a calibration curve and unknown. 1. Prepare a calibration curve by injecting your standards containing 100, 50, 25 and 12.5 ppm of phosphate; starting with the highest concentration. Again run the spectrometer for the same time each run and scale the Y axis post-run to be the same. Run one standard in triplicate to obtain an estimate of the error. 2. Repeatedly inject your unknown (at least three times) and measure the peak absorbance. Calculations Make the following plots: Can measure peak height and width at half-height (W 1/2 ) with a good millimeter ruler. In the real world you would run each data set into EXCELL and have spreadsheet do these calculations! 1. Peak height vs. line length for coil I and II 2. W 1/2 vs.line length. 3. Peak height vs. flow rate 4. W 1/2 vs. flow rate 5. Peak height or peak area (1/2 base x height) vs. concentrations of standards Calculate the concentration of unknown from this plot. Report unknown + error
7 Discussion: Include a discussion of the specific procedure and criteria you used in selecting the optimum parameters. The discussion section should contain answers to the following questions: 1. What type of dispersion do you appear to have at the different flow rates? Look at the shape of each curve. 2. What is the maximum number of samples per hour that you estimate could be determined by this procedure? (Assume an autosampler, rather than manual injection) Explain your answer. 3. What effect does the tube length have on peak height and peak width. Why? 4. What is the concentration of your unknown. Did you use peak height or area? Is your calibration curve linear or non-linear? What is your uncertainty in the unknown measurement and how did you obtain it? 5. Do you think flow injection analysis could be used with detectors other than a spectrophotometer? Suggest a possible alternative detector and describe a suitable analysis.
SPECTROPHOTOMETRIC DETERMINATION OF IRON
SPECTROPHOTOMETRIC DETERMINATION OF IRON In this experiment you will determine trace amounts of iron using spectrophotometric methods. BACKGROUND In solution ferrous iron combines with 2,2 bipyridyl to
More informationRev Experiment 10
Experiment 10 SPECTROPHOTOMETRIC DETERMINATION OF IRON IN DRINKING WATER 2 lab periods Reading: 1) Chapter 17, pg 393-403, Quantitative Chemical Analysis, 8 h Edition, Daniel C. Harris (7 th Edition: Chapter
More informationEXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron with 1,10-Phenanthroline
EXPERIMENT 5 Molecular Absorption Spectroscopy: Determination of Iron with 1,10-Phenanthroline UNKNOWN Submit a clean, labeled 100-mL volumetric flask to the instructor so that your unknown iron solution
More informationChem 321 Lecture 23 - Liquid Chromatography 11/19/13
Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Student Learning Objectives High Performance Liquid Chromatography With the advent of relatively inexpensive and reliable pumps, the development of
More informationCH 149: Chemical Principles. Fall KP1019 Module TA Manual
CH 149: Chemical Principles Fall 2012 KP1019 Module TA Manual 2 Table of Contents KP1019 Synthesis 3 ORP Titrations...10 Electrochemistry 15 ph Kinetics 19 Temperature/NaCl Kinetics...24 3 Equipment (per
More informationCharacteristic and efficiency of PEM fuel cell and PEM electrolyser
Related topics Electrolysis, electrode polarisation, decomposition voltage, galvanic elements, Faraday s law. Principle and task In a PEM electrolyser, the electrolyte consists of a protonconducting membrane
More informationElectricity. Characteristic and efficiency of PEM fuel cell and PEM electrolyser Stationary currents. What you need:
Stationary currents Electricity Characteristic and efficiency of PEM fuel cell and PEM electrolyser What you can learn about Electrolysis Electrode polarisation Decomposition voltage Galvanic elements
More informationElectrolysis, electrode polarisation, decomposition voltage, galvanic elements, Faraday s law.
Characteristics and efficiency of PEM fuel cell TEP Related Topics Electrolysis, electrode polarisation, decomposition voltage, galvanic elements, Faraday s law. Principle In a PEM electrolyser, the electrolyte
More informationApproved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric)
METHOD #: 370.1 TITLE: Approved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric) ANALYTE: Silica, SiO 2 INSTRUMENTATION: Spectrophotometer STORET No. Dissolved 00955 1.0 Scope and Application
More informationMeasuring Manganese Concentration Using Spectrophotometry
Measuring Manganese Concentration Using Spectrophotometry Objectives To use spectroscopy to determine the amount of Manganese is an unknown sample. Scenario Your have just joined a "Green Team" at the
More informationPreparative HPLC: Factors and Parameters that Directly Affect Recovery of Collected Fractions
Preparative HPLC: Factors and Parameters that Directly Affect Recovery of Collected Fractions Gary Scharrer and Joan Stevens, Ph.D. Gilson, Inc Middleton, WI www.gilson.com Abstract Purification of synthetic,
More informationChem 2115 Experiment #9. Consumer Chemistry: Determining the Iron Content in Supplements
Chem 2115 Experiment #9 Consumer Chemistry: Determining the Iron Content in Supplements OBJECTIVE: The goal of this experiment is to use the quantitative technique of spectrophotometry to determine the
More informationOptimizing Purified Sample Recovery on Reverse-Phase HPLC Systems
Optimizing Purified Sample Recovery on Reverse-Phase HPLC Systems Application Note 209 Joan Stevens, PhD, Alan Hamstra, Luke Roenneburg, Tim Hegeman (Gilson, Inc.) Craig Esser, PhD, Regina Black, Derek
More informationCORESTA RECOMMENDED METHOD N 61
CORESTA RECOMMENDED METHOD N 61 DETERMINATION OF 1,2-PROPYLENE GLYCOL, GLYCEROL AND SORBITOL IN TOBACCO AND TOBACCO PRODUCTS BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC) (July 2015) 0. INTRODUCTION
More informationNATIONAL STANDARD OF THE PEOPLE S REPUBLIC OF CHINA (Zhenghua Renmin Gongheguo Guojia Biaozhun)
1/9 GB NATIONAL STANDARD OF THE PEOPLE S REPUBLIC OF CHINA (Zhenghua Renmin Gongheguo Guojia Biaozhun) GB 5009.15-2014 FOOD SAFETY NATIONAL STANDARD DETERMINATION OF CADMIUM IN FOODS Issued 28 January
More informationGenotoxicity is the property of a compound
Impurities Analysis in Pharmaceuticals: Genotoxicity is the property of a compound known to have irreversible effects on the structure and functionality of the DNA in cells and cause DNA loss, DNA replication
More informationCE3502. ENVIRONMENTAL MONITORING, MEASUREMENTS & DATA ANALYSIS. Inflow. Outflo
CE35. ENVIRONMENTAL MONITORING, MEASUREMENTS & DATA ANALYSIS Lab exercise: Statistics topic: Distributions Environmental Engineering Topic: Plug Flow Reactors Environmental Engineering Background There
More informationAnalysis of steel and its alloys using the GB/T standard and an Agilent 5100 ICP-OES in dual view mode
Analysis of steel and its alloys using the GB/T 20125-2006 standard and an Agilent 5100 ICP-OES in dual view mode Application note Metals analysis and production Authors John Cauduro Agilent Technologies
More informationSprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES
Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES Marine Beauvir, Application Specialist, Thermo Fisher Scientific, Cambridge, UK Application Note 43161 Key Words High throughput,
More informationLABORATORY 10 SOIL FERTILITY. Objectives
LABORATORY 10 SOIL FERTILITY I Objectives Learn a method for measuring extractable soil P. Make fertilizer recommendations for a crop. Gain experience with calculations involving inorganic fertilizer materials.
More informationEngineered Systems & Designs TURBIDITY METER
Engineered Systems & Designs TURBIDITY METER Operator s Manual for Model 800 Turbidimeter Model 800 with Optional Battery Pack and Carrying Case INDEX Specifications for Turbidity.................. 1 Introduction..............
More informationICP Expert software. Technical Overview. Introduction
ICP Expert software Technical Overview Introduction The Agilent 5110 ICP-OES provides fast sample analysis, using less gas, without compromising on performance for tough samples. It has been designed for
More informationBy Authority Of THE UNITED STATES OF AMERICA Legally Binding Document
By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By the Authority Vested By Part 5 of the United States Code 552(a) and Part 1 of the Code of Regulations 51 the attached document has
More informationSECTION D.3 CHLOROPHYLL AND PHEOPHYTIN
SECTION D.3 CHLOROPHYLL AND PHEOPHYTIN 1.1 Scope and Application 1.1.1 This spectrophotometric method is used in the determination of chlorophyll a, b, and c, and pheophytin a. This method can be used
More informationMethods for determination of mercury in LP gas
(JLPGA-S-07) Methods for determination of mercury in LP gas 1. Scope This standard specifies the methods for determination of gaseous mercury in vaporized liquid liquefied petroleum (LP) gas. Remarks 1
More informationCOMPUTER SIMULATION OF ENZYME KINETICS
COMPUTER SIMULATION OF ENZYME KINETICS I. Introduction. Enzymes are biological catalysts. A catalyst alters the speed at which a chemical reaction reaches its completion or equilibrium point. It does not
More informationModule 2, Add on Lesson Turbidity Sensor. Student. 90 minutes
Module 2, Add on Lesson Turbidity Sensor Student 90 minutes Purpose Construct a sensor to measure the turbidity of water Graph data and reason about curves and linear relationships Calibrate the turbidity
More informationQuantum Dot applications in Fluorescence Imaging for Calibration and Molecular Imaging
Quantum Dot applications in Fluorescence Imaging for Calibration and Molecular Imaging Introduction In this application note, we will discuss the application of quantum dots in fluorescence imaging, both
More informationCalcium and Magnesium; Chlorophosphonazo Rapid Liquid Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Pour-Thru Cell
Hardness, Total DOC316.53.01045 Calcium and Magnesium; Chlorophosphonazo Rapid Liquid Method Method 8374 4 to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Pour-Thru Cell Scope and application: For boiler and ultrapure
More informationSemiconductors. Lab. FCJJ 16 - Solar Hydrogen Science Kit. Next Generation Science Standards. Initial Prep Time. Lesson Time. Assembly Requirements
Next Generation Science Standards NGSS Science and Engineering Practices: Asking questions and defining problems Developing and using models Planning and carrying out investigations Analyzing and interpreting
More informationComposition Analysis of Animal Feed by HR ICP-OES
Composition Analysis of Animal Feed by HR ICP-OES Abstract The High-Resolution Array ICP-OES equipped with Standard-Kit was used for the analysis of Al, B, Ca, Co, Cr, Cu, Fe, I, K, Na, Mn, Mg, Mo, Ni,
More informationMouse Factor XII Total ELISA Kit
Mouse Factor XII Total ELISA Kit Catalog No: IMFXIIKT-TOT Lot No: SAMPLE INTENDED USE This mouse coagulation Factor XII antigen assay is intended for the quantitative determination of total Factor XII
More informationSMS 100 MERCURY ANALYZER SMS 100. Mercury analysis directly on solid samples from the undisputed leader in Atomic Spectroscopy
SMS 100 MERCURY ANALYZER SMS 100 Mercury analysis directly on solid samples from the undisputed leader in Atomic Spectroscopy SMS 100 Overview The SMS 100 is a dedicated mercury analyzer for the determination
More informationChem 355 Jasperse DISTILLATION
Chem 355 Jasperse DISTILLATION 1 Background Distillation is a widely used technique for purifying liquids. The basic distillation process involves heating a liquid such that liquid molecules vaporize.
More informationDetermination of Iron Content in Different Hemoglobin Samples from Some Patients by UV-Visible Spectrophotometer
Advances in Analytical Chemistry 2016, 6(2): 35-40 DOI: 10.5923/j.aac.20160602.01 Determination of Iron Content in Different Hemoglobin Samples from Some Patients by UV-Visible Spectrophotometer Isam Eldin
More informationThe image part with relationship ID rid4 was not found in the file. Welcome
The image part with relationship ID rid4 was not found in the file. Welcome Trace Metals Analysis: Impurity Determinations By Thomas Kozikowski Chemist, R&D at Inorganic Ventures Key Considerations What
More informationMethod 108C - Determination of Arsenic Content in Ore Samples From Nonferrous Smelters (Molybdenum Blue Photometric Procedure)
While we have taken steps to ensure the accuracy of this Internet version of the document, it is not the official version. Please refer to the official version in the FR publication, which appears on the
More informationLab Exercise 13: Growth Curve
Lab Exercise 13: Growth Curve OBJECTIVES 1. Know the different phases of a standard growth curve. 2. Understand and perform direct measurement of bacterial growth through serial dilutions and standard
More informationCOLORIMETRIC SANDWICH ELISA KIT INSTRUCTION MANUAL
Page 1 of 7 COLORIMETRIC SANDWICH ELISA KIT INSTRUCTION MANUAL This product is for research use ONLY and not for human or animal therapeutic or diagnostic use. Page 2 of 7 Contents Page 3 I. Supplied Materials:
More informationDETERMINATION OF ELEMENTS IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS 10500, & USING THE AGILENT 5100 ICP-OES
ENVIRONMENTAL ANALYSIS DETERMINATION OF ELEMENTS IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS 10500, 14543 & 13428 USING THE AGILENT 5100 ICP-OES Solutions for Your Analytical Business Markets and
More informationBIOO RESEARCH PRODUCTS. Alkaline Phosphatase (AP) Color Endpoint Assay Kit Manual Catalog #:
BIOO RESEARCH PRODUCTS Alkaline Phosphatase (AP) Color Endpoint Assay Kit Manual Catalog #: 3460-09 BIOO Scientific Corp. 2011 TABLE OF CONTENTS GENERAL INFORMATION... 1 Product Description... 1 Procedure
More informationGel Electrophoresis and Analysis
6/28/2016 Gel Electrophoresis and Analysis B3 Summer Science Camp at Olympic High School Dr. Jennifer Weller Lab Method: Gel electrophoresis 2 6/28/2016 Electrophoresis: separating molecules in a charged
More informationAn Oxidation-Reduction Titration: The Reaction of Fe 2+ and Ce 4+
An Oxidation-Reduction Titration: The Reaction of Fe 2+ and Ce 4+ LAB ADV COMP 8 From Advanced Chemistry with Vernier, Vernier Software & Technology, 2004 INTRODUCTION A titration, as you recall, is a
More informationFluorescence Quenching of Human Serum Albumin by Caffeine
CHEM 411L Instrumental Analysis Laboratory Revision 2.1 Fluorescence Quenching of Human Serum Albumin by Caffeine In this laboratory exercise we will examine the fluorescence of Human Serum Albumin (HSA)
More informationIDTox Lactate Dehydrogenase (LDH) Enzyme Cytotoxicity Kit For Cell Culture Supernatant Samples
Rev Date: Mar 26 2015 Rev: 1 1 of 1 IDTox Lactate Dehydrogenase (LDH) Enzyme Cytotoxicity Kit For Cell Culture Supernatant Samples 192 Wells Enzyme Immunoassay for the determination of the Lactate Dehydrogenase
More informationMeasurement of Enzyme Kinetics by UV-Visible Spectroscopy
UV-0002 UV-Visible Spectroscopy Introduction Enzyme activity is frequently investigated in the medicinal, biochemistry, and food science research fields to elucidate the rate of which reaction occurs and
More informationELISA IMMUNOASSAY FOR HUMAN
Unit 6/ Module 1 /Version A pg. 1 California Lutheran University s Enriched Science (Clues) and California State University Program for Education and Research in Biotechnology (C-SUPERB) ELISA IMMUNOASSAY
More informationAnalysis of Stainless Steel by Dual View Inductively Coupled Plasma Spectrometry
Application Note - AN1505 Analysis of Stainless Steel by Dual View Inductively Coupled Plasma Spectrometry Introduction Page 1 Stainless steels are a corrosion resistant family of iron alloys that have
More informationQualification of High-Performance Liquid Chromatography Systems
Qualification of High-Performance Liquid Chromatography Systems L. Huber Hewlett- Packard GmbH, Waldbronn December 1998 Manuscript of an article published in BioPharm, Vol 11 Number 11, November 1998,
More informationAttune TM Acoustic Focusing Cytometer Training. Manik Punj Attune Training
Attune TM Acoustic Focusing Cytometer Training Manik Punj Attune Training Attune Training Agenda Section 1 An Introduction to Flow Cytometry Section 2 An Introduction to Acoustic Focusing Hydrodynamic
More informationSummary of the Spectra Experiment
Summary of the Spectra Experiment This document summarises the Spectra Experiment. The full description of the experiment appears at: http://www.upscale.utoronto.ca/iyearlab/intros/spectra/spectra.html.
More informationHuman Myostatin, ELISA Kit (MSTN)
Human Myostatin, ELISA Kit (MSTN) 96 Tests Catalog Number: MBS733837 Store all reagents at 2-8 C Valid Period: six months For samples: Cell culture fluid & body fluid & tissue homogenate Serum or blood
More informationSCIEX Anion Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide
For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems A49108AE May 2015 AB Sciex Pte. Ltd and its affiliates disclaims all warranties with respect to this document, expressed or implied, including
More informationHuman Angiotensin 2 (Ang2) ELISA
Human Angiotensin 2 (Ang2) ELISA For the quantitative determination of human Ang2 in serum, plasma, cell culture fluid and other biological fluids Cat. No. KT-52748 For Research Use Only. Not for use in
More informationAccuracy and Precision. Dilution Error in Error in Factor Accuracy Precision 200 <2% <2%
Specifications SDS-270 Sample Options Three sample racks, containing: 3 x 60 position x 16 mm OD (20 ml) (default configuration) or optionally: 3 x 90 position x 13 mm OD (13 ml) 3 x 40 position x 20 mm
More informationGas Chromatography Assignment Chem 543/443
Gas Chromatography Assignment Chem 543/443 1. Introduction Capillary gas chromatography (GC) is one of the most popular analytical techniques used in today s research. Its popularity is mainly due to efficient
More informationStoichiometry. Lab. FCJJ 16 - Solar Hydrogen Science Kit. Next Generation Science Standards. Initial Prep Time. Lesson Time. Assembly Requirements
Next Generation Science Standards NGSS Science and Engineering Practices: Asking questions and defining problems Developing and using models Planning and carrying out investigations Analyzing and interpreting
More informationABSTRACT. This study investigates the use of inductively coupled plasma-optical emission spectrometry
Benefits of a Dual-View ICP-OES for the Determination of Boron, Phosphorus, and Sulfur in Low Alloy Steels Michael Duffy and Robert Thomas The Perkin-Elmer Corporation, 761 Main Avenue, Norwalk, CT 06859-0219
More informationAMAZINGLY CAPABLE REMARKABLY AFFORDABLE. Avio 200 ICP Optical Emission Spectrometer. For research use only. Not for use in diagnostic procedures.
AMAZINGLY CAPABLE REMARKABLY AFFORDABLE Avio 200 ICP Optical Emission Spectrometer For research use only. Not for use in diagnostic procedures. PERFORMANCE VALUE AND EASE OF USE IN ONE COMPACT PACKAGE
More informationDetermination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography
Application Note 206 Determination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography Introduction The Bayer Process is used to isolate alumina (Al 2 O ) from bauxite. In this process,
More informationApplication Note 6: On-line Reaction Monitoring of Alcoholic Fermentation
Application Note 6: On-line Reaction Monitoring of Alcoholic Fermentation NMR spectroscopy is an excellent technique for monitoring organic reactions due to its high degree of functional group specificity.
More informationMethod Development and Validation for Online UV-Dissolution Methods Using Fiber-Optic Technology
Technical Overview Method Development and Validation for Online UV-Dissolution Methods Using Fiber-Optic Technology Introduction Online fiber-optic and multicell UV-dissolution systems have become increasingly
More informationPowerPlex Matrix Standards, 3100/3130
TECHNICAL BULLETIN PowerPlex Matrix Standards, 3100/3130 Instruc ons for use of Product DG4650 Note: The PowerPlex Matrix Standards, 3100/3130, can be used to perform spectral calibra on on the Applied
More informationSCION GC Series. The Gas Chromatographers Choice for Separations
SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Scion Instrument s tradition of innovation and product reliability have combined to create the next generation
More informationProtein-Pak Hi Res HIC Column and HIC Protein Standard
Protein-Pak Hi Res HIC Column and HIC Protein Standard CONTENTS I. INTRODUCTION II. a. Mobile Phase b. Flow Direction CONNECTING COLUMN TO LC SYSTEM I. INTRODUCTION This offering contains non-porous, polymethacrylate-based
More informationPERFUSION TECHNOLOFISTS OF GREATER CHICAGO, INC. POLICY AND PROCEDURE MEDTRONIC HEMOTEC ACTIVATED CLOTTING TIME (ACT) OPERATING PROCEDURE
DATE OF REVISION Page 1 of 13 (ACT) OPERATING PROCEDURE PRINCIPLE: The ACT test is a definitive test used to monitor the anticoagulant effect of heparin. It measures the clotting time of fresh whole blood
More informationSeparation and Quantitation of a Mixture of Alcohols by Gas Chromatography
Separation and Quantitation of a Mixture of Alcohols by Gas Chromatography 1. Purpose This procedure will determine the percentage of two alcohols in an unknown aqueous mixture. 2. Background Gas chromatography
More informationThe determination of copper in brass
The determination of copper in brass Objective - To determine the amount of copper in a brass sample Background Brass is an alloy made of copper and zinc. Most brass contains about 60% copper. The proportions
More information5 - Chlorophyll Analysis Lab.
5 - Chlorophyll Analysis Lab. Objectives: To determine the chlorophyll a concentration. Introduction: Chlorophylls are the essential photosynthetic pigments present in all green plants and algae, and allow
More informationLAB 6: Agarose Gel Electrophoresis of Restriction Digested Plasmid DNA
LAB 6: Agarose Gel Electrophoresis of Restriction Digested Plasmid DNA I. Objectives The purpose of today s lab is to learn how to set up and run an agarose gel, separate DNA fragments on the gel, and
More informationThe Determination of Trace Elements in Stainless Steel by Forked Platform GFAAS
The Determination of Trace Elements in Stainless Steel by Forked Platform GFAAS Application Note Atomic Absorption Author Lucinda M. Beach Introduction The metallurgical behavior of stainless steel is
More informationIDENTIFYING UNKNOWN SUBSTANCES
IDENTIFYING UNKNOWN SUBSTANCES LAB 15 EXPERIMENT STUDENT BOOK Chapter 1, page 25 TOOLBOX Page 4 and 36 Goal Identify unknown substances with the help of different tests. 1. What is the independent variable
More informationDNA RESTRICTION ANALYSIS
DNA RESTRICTION ANALYSIS In this experiment, DNA from the bacteriophage Lambda (48,502 base pairs in length) is cut with a variety of restriction enzymes and the resulting fragments are separated using
More informationAPPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC
UOP 63-13: Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC Fast Analysis in
More informationBacterial Counts - Quantitative Analysis of Microbes
Bacterial Counts - Quantitative Analysis of Microbes Introduction: It is often important to know not only what types of bacteria are in a sample but also how many of them are present. Food manufacturers
More informationDesign of Experiment. Jill Williams and Adam Krinke. Fuel Cell Project
Design of Experiment Jill Williams and Adam Krinke Fuel Cell Project Spring 2005 Introduction The Proton Exchange Membrane (PEM) fuel cell requires a relatively stringent environment for operation. The
More informationINSTRUCTIONS & CALIBRATION PHOTOBOOK
INSTRUCTIONS & CALIBRATION PHOTOBOOK Copyright, Hanby Environmental, LLC 2012 Table Of Contents Included in your Kit / A Note on Soil Sampling..Page 4 Kit Instruc ons... Pages 5-7 Calibra on Photos Standards......
More informationSTANDARD OPERATING PROCEDURES
PAGE: 1 of 5 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS
More informationHuman Leukotriene E4 (LTE4) ELISA Kit
Human Leukotriene E4 (LTE4) ELISA Kit Catalog No. CSB-E05176h (96T) This immunoassay kit allows for the in vitro quantitative determination of human LTE4 concentrations in serum, plasma and other biological
More informationHuman Collagen Type III (COL3) ELISA
Human Collagen Type III (COL3) ELISA For the quantitative determination of human COL3 in serum, plasma, cell culture fluid and other biological fluids Cat. No. KT-61018 For Research Use Only. Not for use
More informationName Honors Chemistry / /
Name Honors Chemistry / / SOL Questions Chapter 1 Each of the following questions below appeared on an SOL Chemistry Exam. For each of the following bubble in the correct answer on your scantron. 1. The
More informationHuman connective tissue growth factor (CTGF) ELISA Kit. MyBioSource.com. This package insert must be read in its entirety before using this product.
Human connective tissue growth factor (CTGF) ELISA Kit Catalog Number. For the quantitative determination of human connective tissue growth factor (CTGF) concentrations in serum, plasma, tissue homogenates.
More informationRESTORE 1030TM Testing Protocol
RESTORE 1030TM Testing Protocol Test protocols will vary depending on the type of equipment being evaluated. The length of time required to do the testing, however, remains consistent no matter what the
More informationSystem Overview. Valve oven. Restrictor. Valve. Sample in and out
System Overview The GC system is designed to provide a continous reading of total hydrocarbons in a gas stream, while periodically performing a chromatographic separation of the sample to determine the
More informationPinAAcle 500. Flame Atomic Absorption Spectrometer
PinAAcle 500 Flame Atomic Absorption Spectrometer RELIABILITY SENSITIVITY AFFORDABILITY Engineered to deliver an uncompromising level of performance at an unbeatable price, the PinAAcle 500 puts the industry
More informationHuman IgG Antigen ELISA Kit
Human IgG Antigen ELISA Kit Catalog No: IHUIGGKT Lot No: SAMPLE INTENDED USE This human immunoglobulin G antigen assay is intended for the quantitative determination of total human IgG antigen in serum,
More informationSemiconductors. Lab. FCJJ 30 - Electric Mobility Science Kit. Next Generation Science Standards. Initial Prep Time. Lesson Time. Assembly Requirements
Next Generation Science Standards NGSS Science and Engineering Practices: Asking questions and defining problems Developing and using models Planning and carrying out investigations Analyzing and interpreting
More informationRecycled Materials as Insulation. Topic of the Research Paper- Examining the insulating capacity of recyclable materials
International Journal of Scientific & Engineering Research, Volume 7, Issue 12, December-2016 Recycled Materials as Insulation Topic of the Research Paper- Examining the insulating capacity of recyclable
More informationEvaluation Report of
Evaluation Report of InnerGeo WaterReco s Hydrocarbon Extraction Vessel Report Prepared for: Analysis Performed by: Report Prepared by: 6/6/2016 Hanby Environmental and Hanby PetroAnalysis Charles D. Fator,
More informationElectrophoresis 101 Student Worksheet
1 Electrophoresis 101 Student Worksheet Experiment Objective To develop an understanding of electrophoresis principles. To analyze results and to calculate the sizes of unknown charged molecules from given
More informationRayBio Human Caspase-3 ELISA Kit
RayBio Human Caspase-3 ELISA Kit Catalog #: ELH-CASP3 User Manual Last revised July 21, 2017 Caution: Extraordinarily useful information enclosed ISO 13485 Certified 3607 Parkway Lane, Suite 100 Norcross,
More informationBovine IgG ELISA Catalog #:
INTENDED USE The IMMUNO-TEK Bovine IgG ELISA Kit is a rapid, easy to use enzyme linked immunosorbent assay (ELISA) designed for the measurement of bovine IgG in bovine colostrum, milk, serum, plasma or
More informationAUTOMATED WATER ANALYSIS
APPLICATION NOTE - CFA AUTOMATED WATER ANALYSIS APPLICATIONS OF CONTINUOUS-FLOW ANALYZERS In laboratories around the world, the SEAL Analytical analyzers measure all types of water - quickly, accurately
More informationTHE EFFECT OF ph ON OXYGEN BINDING BY HEMOCYANIN. BACKGROUND READING Animal Physiology by Hill, Wyse & Anderson, 2004: pp ; Fig
THE EFFECT OF ph ON OXYGEN BINDING BY HEMOCYANIN BACKGROUND READING Animal Physiology by Hill, Wyse & Anderson, 2004: pp. 577 593; Fig. 23.22. PRE-LAB (Due at the start of lab) ** In your lab notebook,
More informationRayBio Custom ELISA Kit
RayBio Custom ELISA Kit Catalog #: EL-PRELIM User Manual Last revised April 15, 2016 Caution: Extraordinarily useful information enclosed ISO 13485 Certified 3607 Parkway Lane, Suite 100 Norcross, GA 30092
More information3. Add 0.4 ml of. 7. Use a TenSette
Method 10129 ORGANIC CARBON, TOTAL, Low Range (0.0 20.0 mg/l C) Direct Method * For water, drinking water, and wastewater HRS MIN SEC 1. Turn on the DRB 200 reactor. Heat to 103-105 C. Note: See DRB 200
More informationBovine prolactin/luteotropic hormone (PRL/LTH) ELISA Kit
Bovine prolactin/luteotropic hormone (PRL/LTH) ELISA Kit Catalog Number. MBS703224 For the quantitative determination of bovine prolactin/luteotropic hormone (PRL/LTH) concentrations in serum, plasma.
More informationCadmium Testing and On Site Calibration for Water Testing Detection Range: ppm
Document: AND016 04 2014 Cadmium Testing and On Site Calibration for Water Testing Detection Range: 0.1 1.0ppm April, 2014 Edition 3 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of
More informationFundamentals and Techniques of Preparative HPLC. Parto Zist Behboud Tel: 42108
Fundamentals and Techniques of Preparative HPLC Parto Zist Behboud www.partoz.com Tel: 42108 Today s Overview 1 Basic know-how for up-scaling 2 Practical examples of up-scaling 3 Recycle preparation 4
More informationTorch Combustion TOC Analyzer
Torch Combustion TOC Analyzer The Torch Combustion TOC Analyzer utilizes patent pending Static Pressure Concentration (SPC) for the analysis of TOC using high temperature combustion with an optional TN
More information