online measurement electrodes

Size: px
Start display at page:

Download "online measurement electrodes"

Transcription

1 online measurement electrodes ph Conductivity ORP Turbidity Optical Dissolved Oxygen Dissolved Ozone Dissolved Oxygen Free Chlorine

2 For three decades HF scientific has developed specialized expertise in measurements for water quality. Our original involvement in water-quality focused on Turbidity measurement. These products have positioned us as one of the worldwide leaders in Turbidity measurement and monitoring. Over the years our expertise expanded to include devices for monitoring other parameters such as Residual Chlorine, Streaming Current and UV% Transmission. We are now excited to expand our product offering with the addition of our new AdvantEDGE Series of measurement electrodes. These electrode systems have been designed to reduce equipment and installation costs while improving reliability. Measurement parameters include ph, ORP, Free Chlorine, Raw Water Turbidity, Dissolved Oxygen, Optical Dissolved Oxygen, Dissolved Ozone, Contacting Conductivity and Toroidal Conductivity. The AdvantEDGE Series systems feature factory pre-calibrated electrodes. The electrodes can subsequently be checked for calibration, diagnosed and configured directly from a PLC or other computer system (e.g., SCADA). Although they are available for use if desired, a local display/analyzer is not necessary. Numerous integrated communications protocols are supported, including Modbus RTU, RS-232, RS-485 ASCII, Profibus, CANOpen, DeviceNet, Ethernet IP, and USB. Contact us directly or call the HF scientific representative serving your area to learn how the new AdvantEDGE Series can bring your facility into a new generation of water quality monitoring. SOLTECH LTDA. Calle 76 #20B-24 Of Tel Fax: Bogotá D.C. Colombia. infotecnica@soltechltda.com HF scientific 3170 Metro Parkway Ft. Myers, FL Ph: Fax: Toll Free:

3 Table of Contents: AdvantEDGE series AdvantEDGE series Overview... 1 ph... 2 ORP... 2 Dissolved Ozone... 3 Dissolved Oxygen... 3 Optical Dissolved Oxygen... 4 Contacting Conductivity... 5 Toroidal Conductivity... 5 Free Chlorine Panel System... 6 Turbidity... 6 LD500 Local Display... 7 PA500 Profibus Gateway... 7 Mounting Hardware... 8 Accessories & Replacement Parts... 8 Specifications and Content Information Subject to change without notice

4 AdvantEDGE series The AdvantEDGE sensor assembly consists of three components designed for direct connection to control systems without the need for intermediate analyzer electronics. A Communications Module plugs into the back of the Interface Module and provides direct interactive communication of measurement, configuration, calibration and diagnostic information with computers and networks. A full family of sensor heads plug into the front of the Interface Module and automatically provides calibrated, high resolution measurements. Communications Adapter Provides power and direct interactive communication with control systems and computers. A selection of plug and play communication modules provide power and direct interactive communications with control systems and computers. The table below shows the available protocols. If a local display is desired, the LD500 can be used. Multiple sensor assemblies can be used with the LD500. Interface Module The heart of the sensor assembly is the Universal Interface Module. It s encapsulated measurement electronics automatically converts sensor signals into calibrated measurement units. It works with a variety of Communication Modules to operate directly with any computer system. Sensor Head Pre-calibrated plug and play sensor heads can be replaced at any time, making AdvantEDGE the most versatile and accurate analytical measurement system on the market. The table below shows the parameters available. PARAMETERS INTERFACE COMMUNICATIONS PROTOCOLS CONTROL ROOM ph Modbus RTU ORP Profibus Free Chlorine Turbidity Direct RS 485 ASCII CANOpen SCADA PLC Dissolved Oxygen Dissolved Ozone Local Display Device Net Ethernet IP DCS PC Conductivity USB Optional LD500 Local Display Typical AdvantEDGE System Assembly 1

5 AdvantEDGE ph Differential ph measurement Replaceable quad junction salt bridges Electrode protection options Connect this ph sensor to a PLC for seamless integration with industrial control systems. Use any computer to display data, calibrate and customize the measurement without an intermediate analyzer electronics box. A LD500 local display can be used if desired. Sensor heads are pre-calibrated and can be replaced or exchanged with any other type of sensor without taking the system down. Save space, time and money. Range: 0 to 14 ph Resolution: 0.01 ph Accuracy: ph Step Response Time: 90% in 30 seconds Temperature Range: -5 C to 95 C Maximum Pressure: C C Immunity & Emissions: CE (EN61326 and EN61326:1998) Process Electrodes: G, HF or Low Temperature Glass Ground Rod: Titanium, 316 Stainless Steel or Hastelloy C Measurement Units: ph, mv Automatic Buffer: 1 and 2 point Sample: 1 and 2 point Note: ph and temperature are precalibrated at the factory Temperature Compensation Options Linear: % per C Built-in tables: Ammonia or Morpholine Note: Temperature can be entered manually Auto Buffer Standards: (4, 7, 10) and DIN AdvantEDGE ORP Range: -2100mV to +2100mV Resolution: 0.1mV Accuracy: 0.1% of reading Step Response Time: 90% in 30 seconds Temperature Range: -5 C to 95 C Maximum Pressure: C C Process Electrodes: Platinum or Gold Ground Rod: Titanium, 316 Stainless Steel or Hastelloy C Measurement Units: mv Sample: 1 point Note: ORP and temperature are precalibrated at the factory Temperature Compensation Options Temperature measurement is provided as an independent measurement. Note: Temperature can be entered manually Immunity & Emissions: CE (EN61326 and EN61326:1998) Differential ORP measurement Replaceable quad junction salt bridges Electrode protection options Connect this ORP sensor to a PLC for seamless integration with industrial control systems. Use any computer to display data, calibrate and customize the measurement without an intermediate analyzer electronics box. A LD500 local display can be used if desired. Sensor heads are pre-calibrated and can be replaced or exchanged with any other type of sensor without taking the system down. Save space, time and money. 2

6 AdvantEDGE Ozone Clark Cell technology Rugged four resistant membrane Electrode protection options Simple membrane cartridge replacement Connect this Dissolved Ozone sensor to a PLC for seamless integration with industrial control systems. Use any computer to display data, calibrate and customize the measurement without an intermediate analyzer electronics box. A LD500 local display can be used if desired. Sensor heads are pre-calibrated and can be replaced or exchanged with any other type of sensor without taking the system down. Save space, time and money. Range: 0 to 40 ppm Resolution: 0.01 ppm Accuracy: 0.04 ppm Step Response Time: 90% in 90 seconds PEEK Sensor Head Temperature Range: -5 C to 50 C Maximum Pressure: C Process Electrodes: Gold Cathode, Silver Anode, Silver Reference (3 Electrode Polarographic Clark Cell) Membrane: FEP Teflon Sensor Head Material: PEEK Measurement Units: ppm, mg/l, % Air: Automatically adjusts for set pressure Sample: 1 point Zero: 1 point Note: D.O. and temperature are precalibrated at the factory Immunity & Emissions: CE (EN61326 and EN61326:1998) AdvantEDGE DO Clark Cell technology Rugged four resistant membrane Electrode protection options Simple membrane cartridge replacement Connect this Dissolved Oxygen sensor to a PLC for seamless integration with industrial control systems. Use any computer to display data, calibrate and customize the measurement without an intermediate analyzer electronics box. A LD500 local display can be used if desired. Sensor heads are pre-calibrated and can be replaced or exchanged with any other type of sensor without taking the system down. Save space, time and money. Range: 0 to 40 ppm Resolution: 0.01 ppm Accuracy: 0.04 ppm Step Response Time: 90% in 90 seconds PEEK Sensor Head Temperature Range: -5 C to 50 C Maximum Pressure: C Process Electrodes: Gold Cathode, Silver Anode, Silver Reference (3 Electrode Polarographic Clark Cell) Guard: Platinum Membrane: PFA Teflon Measurement Units: ppm, mg/l, % Air: Automatically adjusts for set pressure Sample: 1 point Zero: 1 point Note: D.O. and temperature are precalibrated at the factory Immunity & Emissions: CE (EN61326 and EN61326:1998) 3

7 AdvantEDGE Optical Dissolved Oxygen Robust Cost Effective Options No Hassles Low Maintenance Accurate The AdvantEDGE sensor uses the latest technology to measure dissolved oxygen (DO) in demanding environments. The AdvantEDGE series Optical Dissolved Oxygen RDO sensor o ffers several advantages for monitoring DO in process applications: Specifications Sensor Type Transmitter/local display Range Accuracy (DO) Response time, cap Resolution Usage life of cap Shelf life of cap Operating temperature IP rating Compliance Storage conditions, cap Storage conditions, sensor Salinity range ph range Barometric range Internal mounting thread Communications options Maximum cable length Warranty The AdvantEDGE sensor measures dissolved oxygen (DO) using the principle of dynamic luminescence quenching. The sensor (Figure 1) uses lifetime-based optical fluorescent technology to provide an extremely stable, accurate, lowphotodiode maintenance DO sensor. The sensor measures the phase (or delay) of the returned signal for Sensor Diagnostics compared to the excitation signal, and is thus based on the lifetime rather than the intensity of the luminescence. The presence of oxygen in the foil quenches the Figure 1: AdvantEDGE sensor design luminescence and causes a phase shift in the returned signal, detected by the photodiode. The phase difference between the blue excitation light and the return red light is measured, and the result is used to quantify DO. Luminescent dissolved oxygen sensor Optional, not required 0 to 20 mg/l concentration 0 to 200% saturation ±0.1 mg/l, 0 to 8 mg/l ±0.2 mg/l, 8 to 20 mg/l T90: 30 sec T95: C 0.01 mg/l 1 year from the first instrument reading 2 years from date of manufacture (install within 1 year from date of manufacture) 0 to 50 (32 to 122 F) IP-67 with cap off, IP-68 with cap installed Heavy industrial, IEC : to 60 (33 to 140 F), in factory container -5 to 60 (23 to 140 F) 0 to 42 PSU, fi xed or real-time capable 2-10 ph mbar, fi xed or real-time capable 1¼-11½ NPT Modbus RTU (RS485), 4-20mA, SDI-12 Up to 4000 ft (Modbus and 4-20 ma) or up to 200 ft (SDI-12) Sensor: 3 years from date of manufacture Cap: 2 years from date of manufacture 4

8 AdvantEDGE Contacting Conductivity High resolution measurement 0 to 2 S/cm range Connect this Conductivity sensor to a PLC for seamless integration with industrial control systems. Use any computer to display data, calibrate and customize the measurement without an intermediate analyzer electronics box. A LD500 local display can be used if desired. Sensor heads are pre-calibrated and can be replaced or exchanged with any other type of sensor without taking the system down. Save space, time and money. Range: 0 to 2,000,000 us/cm (2 S/cm) Resolution: 4.5 significant digits Accuracy: 0.1% of reading Step Response Time: 90% in 30 seconds CPVC Sensor Head Temperature Range: -5 C to 50 C Maximum Pressure: C Cell Constants: 0.1 for Resistivity 1.0 and 5.0 for Conductivity Sensor Head Material: CPVC Weight: 1.2 lbs Measurement Units: μs/cm, TDS, %, MΩ cm Zero: In dry air Span: 1 point Note: Conductivity and temperature are precalibrated at the factory Concentration Table: User 10-point Immunity & Emissions: CE (EN61326 and EN61326:1998) 5 AdvantEDGE Toroidal Conductivity High resolution measurement 0 to 2 S/cm range Connect this Toroidal Conductivity sensor to a PLC for seamless integration with industrial control systems. Use any computer to display data, calibrate and customize the measurement without an intermediate analyzer electronics box. A LD500 local display can be used if desired. Sensor heads are pre-calibrated and can be replaced or exchanged with any other type of sensor without taking the system down. Save space, time and money. Range: 0 to 2,000,000 us/cm (2 S/cm) Resolution: 4.5 significant digits Accuracy: 0.1% of reading Step Response Time: 90% in 30 seconds CPVC Sensor Head Temperature Range: -5 C to 50 C Maximum Pressure: C Sensor Head Material: CPVC Electrode Material: Titanium (Standard) or 316 Stainless Steel Weight: 1.2 lbs (Peek, Ryton or CPVC) 2.6 lbs (316 Stainless Steel) Measurement Units: μs/cm, TDS, %, MΩ cm Zero: In dry air Span: 1 point Note: Conductivity and temperature are precalibrated at the factory Concentration Table: User 10-point Immunity & Emissions: CE (EN61326 and EN61326:1998)

9 AdvantEDGE Free Chlorine Panel Free Chlorine Amperometric Measurement Temperature Compensated Optional ph Correction Pre-Calibrated Plug & Play Rugged Teflon Membrane Replaceable Membrane Cap Remote Measurement,, Configuration, and Diagnostics. This Free Chlorine sensor will measure accurately in processes that are between 4 and 8 ph. When used with the LD500 Local Display unit and a ph measurement system, hypochlorous acid (HOCL) and hypochlorite ion (OCL-) concentration can be used to determine free chlorine present. Measurement System Preformance: Ranges: 0 to 5 ppm and 0 to 20 ppm Resolution: ppm Accuracy: 2% or +/- 10 ppb (whichever is greater) Step Response Time: 90% in 90 seconds : Temperature Range: 0 C to 45 C (32 F to 113 F) Maximum Pressure: C System Flow in Chamber: 200 to 250ml/min : : Process Electrodes: Gold Cathode, Silver Anode. Membrane: Teflon O-Rings: Viton Sensor Head Material: Noryl AdvantEDGE Material: CPVC or PEEK Weight: 1.2 lbs : Measurement Units: ppm, mg/l free chlorine : Sample: 1 point Zero: 1 point Temperature: point Note: Precalibrated at the factory : : Immunity and Emissions: CE Certified 89/336/ EEC:CISPER 11, EN61000 (-4-2,-4-3,-4-4,-4-6, 4-8) Safety: culus Listed; 367G E Hazardous Locations: Haz Loc Class 1, Division 2,Groups A, B, C, D. Max Ambient 50 C 6 AdvantEDGE Turbidity Connect this Raw Water Turbidity sensor to a PLC for seamless integration with industrial control systems. Use any computer to display data, 0 to 4000 NTU measurement range Conforms to ISO degree and 180 degree light paths Rugged, foul resistant construction Local and Remote configuration and diagnostics Temperature measurement included calibrate and customize the measurement without an intermediate analyzer electronics box. A LD500 local display can be used if desired. Sensor has a second measurement mode to report sediment load. It does this using a spectral measurement to compensate for particle color, size and shape. : Range: 0 TO 4000 NTU Resolution: 0.1 NTU Accuracy: 0.2 NTU Step Response Time: 90% in 15 seconds Note: Typical at 25 C;. : Temperature Range: -5 to 50 C Maximum Pressure: C Maximum Flow Rate: 10 ft./sec. : Measurement Modes: Turbidity: 90 degree/ 180 degree, narrow beam, IR light Sediment Load: Blue to IR spectrum detected at 90 degrees and 180 degrees Temperature: Automatic from -5 to 50 C : Windows: Quartz or Sapphire Optical Topology: Multi-Color Source Sensor Head Material: Polycarbonate Weight: 1.2lbs : Measurement Units: Turbidity Mode: NTU Sediment mode: mg/l % Temperature Units: C, F : Sample: 1 point Zero: 1 point Note: Turbidity and temperature are precalibrated at the factory : : Immunity and Emissions: CE Certified 89/336/EEC:CISPER 11, EN61000 (-4-2,-4-3,-4-4,-4-6, 4-8) Safety: culus Listed; 367G E Hazardous Locations: Haz Loc Class 1, Division 2,Groups A, B, C, D. Max Ambient 50 C

10 AdvantEDGE LD500 Local Display AdvantEDGE PA500 Profibus Gateway Functions and Features Connects to any AdvantEDGE system. Convenient 1/4 DIN enclosure Display sensor and temperature data Menu Functions:, Configuration and Diagnostics Digital Protocol Options for remote measurement, calibrations, configurations, and diagnostics. One 4 to 20 ma loop output (Standard) Two 4 to 20 ma loop outputs (Optional) Two form C relays for alarm, control, and wash (Optional) PID for ph 7 Conduit knockouts for easy cabling Mounting Options Conduit Box, NEMA 4X (Standard) 1/2 DIN NEMA 4X Panel Mount (Optional) PA500 Profibus Gateway Adapter connects up to 30 Modbus Communications Modules to a Profibus DP (V0 or V1) network with M12 connector. Full-featured measurement, calibration, configuration and diagnostics. 24 VDC power. Interface Display: 2 line by 16 character LCD Backlight: high contrast green Keypad: 4 directional arrows, MENU, ESCAPE, and ENTER Enviromental Conditions Ambient Operating Temperature Range: -20 to 65 C Maximum Relative Humidity: 95% non-condensing Meets CE requirements for heavy industrial use. Nema 4X housing. AdvantEDGE Sensor Connections Supports 1 to 255 Modbus RTU AdvantEDGE sensors of any type. Supported Sensors: ph, ORP, Two Electrode Conductivity, Two electrode resistivity, Toroidal Conductivity, Dissolved Oxygen, Ozone, Turbidity, and Free Chlorine. 4 to 20 ma Current Loop Options One Isolated Current Loop (Standard) Two Isolated Current Loops (Optional) Assignable Parameters PID function when ph parameter selected Loop power provided by LD500 Relays/Optional Two 1 Amp form C relays with NO and NC contacts. Assignable Parameters Configured for Alarm, Control, & Timer Functions Digital Communication Options to Host Computer Network MODBUS RTU Device Net Profibus DP/V1 RS-232 ASCII RS-232 DF1 RS485 ACSII CANOpen Others: Consult Factory Power Supply Input: 10 to 30 VDC; 100 to 240 VAC optional Power: 2W at 25 C 7

11 Mounting Options 1 Inch Tee Mounting 1.5 Inch Union Mounting 1.5 Inch Ball Mounting Material CPVC 316 Stainless Steel Sanity Mount Material CPVC 316 Stainless Steel Handrail Mounting Assembly Material CPVC (Low) 316 Stainless Steel (Low) CPVC (High) 316 Stainless Steel (High) 1 Inch Immersion Mounting with Junction Box Material 316 SS (closed) 216 SS (open) ph Salt Bridge Replacements SBS01: PEEK Protected SBS02: PEEK Flat SBS03: CPVC Protected SBS04: CPVC Flat SBC01: Storage cap with sponge Material Immersion PVC Accessories & Replacements Parts Dissolved Ozone ph Solutions RCS03: storage solution, 60 ml PHSOL4: Buffer, 500 ml PHSOL7: Buffer, 500 ml PHSOL10: Buffer, 500 ml RCSO1-500: Reference Cell Solution 500 ml Ozone Membrane Replacements DMR05: Ryton Protected DMR06: Ryton Flat DMR07: PEEK Protected DMR08: PEEK Flat SBC01: Storage cap with sponge Material: PVC Ozone Solutions 500 ml bottles RDOK2: Ozone Electrolyte ZDOKO: Zero Powder Wash Adapters MH1142: air purge, CPVC MH1132: water purge, CPVC ORP Salt Bridge Replacements SBS01: PEEK Protected SBS02: PEEK Flat SBS03: CPVC Protected SBS04: CPVC Flat SBC01: Storage cap with sponge Dissolved Oxygen D.O. Membrane Replacements DMR01: Ryton Protected DMR02: Ryton Flat DMR03: PEEK Protected DMR04: PEEK Flat SBC01: Storage cap with sponge ORP Solutions RCS04: ORP storage solution, 60 ml ORPSOL200: 200mV, 500 ml ORPSOL600: 600 mv calibration, 500 ml RCS01: Reference cell solution, 500 ml Wash Adapters MH1142: air purge, CPVC MH1132: water purge, CPVC Turbidity Turbidity Solutions 500 ml bottles FOR4K: 4000 NTU Formazin Stock 500 ml FROO4O: 40 NTU Cal Kit Wash Adapters MH1142: air purge, CPVC MH1132: water purge, CPVC Conductivity Conductivity Solutions 500 ml bottles SOL1000: 1000 μs/cm Solution SOL2000: 2000 μs/cm Solution SOL5000: 5000 μs/cm Solution SOL2C: Conductivity Electrode Cleaner 8