Original Article. A Novel Automated Slide-Based Technology for Visualization, Counting, and Characterization of the Formed Elements of Blood

Size: px
Start display at page:

Download "Original Article. A Novel Automated Slide-Based Technology for Visualization, Counting, and Characterization of the Formed Elements of Blood"

Transcription

1 Original Article A Novel Automated Slide-Based Technology for Visualization, Counting, and Characterization of the Formed Elements of Blood A Proof of Concept Study James W. Winkelman, MD; Milenko J. Tanasijevic, MD, MBA; David J. Zahniser, PhD Context. A novel automated slide-based approach to the complete blood count and white blood cell differential count is introduced. Objective. To present proof of concept for an imagebased approach to complete blood count, based on a new slide preparation technique. A preliminary data comparison with the current flow-based technology is shown. Design. A prototype instrument uses a proprietary method and technology to deposit a precise volume of undiluted peripheral whole blood in a monolayer onto a glass microscope slide so that every cell can be distinguished, counted, and imaged. The slide is stained, and then multispectral image analysis is used to measure the complete blood count parameters. Images from a 600-cell white blood cell differential count, as well as 5000 red blood cells and a variable number of platelets, that are present in 600 high-power fields are made available for a technologist to view on a computer screen. An initial comparison of the basic complete blood count parameters was performed, comparing 1857 specimens on both the new instrument and a flow-based hematology analyzer. Results. Excellent correlations were obtained between the prototype instrument and a flow-based system. The primary parameters of white blood cell, red blood cell, and platelet counts resulted in correlation coefficients (r)of 0.99, 0.99, and 0.98, respectively. Other indices included hemoglobin (r ¼ 0.99), hematocrit (r ¼ 0.99), mean cellular volume (r ¼ 0.90), mean corpuscular hemoglobin (r ¼ 0.97), and mean platelet volume (r ¼ 0.87). For the automated white blood cell differential counts, r values were calculated for neutrophils (r ¼ 0.98), lymphocytes (r ¼ 0.97), monocytes (r ¼ 0.76), eosinophils (r ¼ 0.96), and basophils (r ¼ 0.63). Conclusions. Quantitative results for components of the complete blood count and automated white blood cell differential count can be developed by image analysis of a monolayer preparation of a known volume of peripheral blood. (Arch Pathol Lab Med. 2017;141: ; doi: /arpa OA) The most commonly performed clinical laboratory test is the complete blood count (CBC). Currently, laboratory Accepted for publication March 15, Published as an Early Online Release April 19, From the Department of Pathology, Brigham and Women s Hospital, and Harvard Medical School, Boston, Massachusetts (Drs Winkelman and Tanasijevic); and Roche Diagnostics Hematology, Boston (Dr Zahniser). Drs Winkelman, Tanasijevic, and Zahniser were founders of Cell Imaging LLC. The company received National Institutes of Health Small Business Innovation Research funding. Its technology was licensed to Constitution Medical Inc, which was subsequently acquired by Roche Holdings. Drs Winkelman and Tanasijevic received payments from Roche Holdings for the patent rights, development milestones, and consulting. A royalty agreement is in place. Dr Zahniser received payments from Roche Holdings for the patent rights and development milestones. A royalty agreement is in place. Dr Zahniser became an employee of Constitution Medical Inc and Roche Diagnostics Hematology Inc, and received, in addition to the above, salary and stock options. Reprints: Milenko Tanasijevic, MD, MBA, Brigham and Women s Hospital, 75 Francis St, Amory Bldg 215 A, Boston, MA ( mtanasijevic@bwh.harvard.edu). CBC systems obtain cell counts and morphologic information by maintaining the blood cells in a liquid flow and by using the Coulter impedance-based orifice or laser-based cytometry. In addition to providing basic CBC information, current instrumentation typically provides an automated white blood cell (WBC) differential count, and additional features, such as reticulocyte counts. This article describes a potential new approach to the CBC in which a known volume of undiluted anticoagulated whole blood is precisely applied onto a microscope slide as a monolayer. The method uses the standard morphology of an air-dried blood film stained with a Romanowsky stain. The analysis is accomplished by multispectral digital imaging that allows precise simultaneous counting of all of the formed elements, while their digital images are analyzed for their morphology and classification. An operator can readily review the digitized images on a computer screen. Thus, the prototype instrument provides quantitative and morphologic information from the same sample of blood simultaneously. Compared with the existing combination of a flow cytometer, slide maker, and staining device with a stand-alone cell imager, the device is Arch Pathol Lab Med Vol 141, August 2017 A Slide-Based Approach to Blood Cell Assessment Winkelman et al 1107

2 Figure 1. A microscope slide is moved under a blunt needle, providing a constant flow of whole blood. Typically, 1 ll of blood is deposited. a single integrated system that eliminates electrochemical and laser components, reduces the number and volumes of reagents and liquid consumables supplies, and features a much smaller footprint in the laboratory. The time to complete a report that requires direct visual assessment of cellular morphology from a glass slide could be greatly reduced. After the initial development work by the authors, 1 the new approach described here went into development to become a commercial instrument. Components of a prototype of that system are described here. BACKGROUND The CBC has evolved over time as new technologies have emerged. At present, more than 25 parameters appear in a typical CBC report, all of them derived from the quantitation and morphologic assessment of red blood cells (RBCs), WBCs, and platelets. Included are hematocrit and hemoglobin; the calculated RBC indices of mean cellular volume (MCV), mean corpuscular hemoglobin (MCH), and MCH concentration; and other derived values. Instrumentation has come a long way from the use of hemocytometer counting chambers for estimating cell numbers and the pulled, stained smear and microscope for evaluating normal and abnormal cells. From the introduction of single-channel Coulter Corporation particle counters in the 1950s to the current multichannel, multidetector flow cytometric systems, 2 accuracy has improved and costs have been reduced. Today s flow-based systems are capable of performing 5- or 7-part WBC differentials based on physical properties of blood cells, but they have no imaging component and still require addition specimen handling and microscopic manual review in many cases. 3 Because no images of cells are provided by a flow cytometric system, these systems require skilled operators with the ability to identify cell types and other parameters by interpreting their scattergram outputs. In terms of laboratory workflow, a significant problem with flow cytometry systems is that a large proportion of specimens give rise to flags, indicating that further testing is required. Flag rates typically vary between 10% and 50%, depending on the flagging criteria policy established by the laboratory and the patient population it serves, with a median rate of 27%. 4 The most common reasons for flagging and retesting are the presence of abnormal cell types or abnormal cell morphology, or a too-low or toohigh number of RBCs, WBCs, or platelets. In a recent survey by the College of American Pathologists, 35.7% of participants claimed to have learned additional information beyond what was available from the automated instrument s output by microscopic review of a peripheral smear. 4 A manual review is a labor-intensive and time-consuming step, requiring retrieval of the blood specimen tube, removal of blood from the tube, manual application of the sample onto a glass slide, and staining of the blood smear. The slide must then be examined under a microscope by a skilled hematology technologist or physician. Manual review delays the reporting time for the CBC, is labor intensive, and costs approximately 3 times the amount of the instrumental component of the CBC. Some flow cytometry blood analyzer systems provide slide maker and stainer devices separately or in tandem with their automated CBC system. There are also stand-alone computer image analysis instruments that can be used for morphologic analysis after the slide is produced. Such additions make current systems more expensive; add to the complexity, size, and logistics of the CBC workstation; and require integrating the results of the imaging back into the initial CBC. Most significantly, this process means that the slide making, staining, and slide review are each done in sequence, lengthening the overall turnaround time to provide a result to the clinician. Automated CBC instruments based on flow technology are considered to be mature, trustworthy, and efficient, and there have been no fundamentally different approaches introduced for several decades. Image analysis based systems for the WBC differential count were first introduced by 4 major biomedical device companies in the 1970s. 5 8 They required separate preparation and staining of conventional blood smears, did not measure any other CBC parameters besides the WBC differential count, and had other limitations to their image analysis capabilities. 9 In addition, they were expensive. As a result, they did not survive competition from flow cytometry based systems that came onto the market at that time; the flow cytometry based systems were capable of doing some portions of the WBC differential, albeit without producing images of the cells. Recent versions of conventional smear-based image analysis systems have led to successful stand-alone instruments for the WBC differential, but they do not perform any of the quantitative measurements required for a CBC. 10 NEW APPROACH TO INSTRUMENTATION FOR AN IMAGE-BASED CBC AND WBC DIFFERENTIAL The prototype for a new, integrated system for the automated CBC and WBC differential described here overcomes certain intrinsic limitations and fixed operating requirements of flow cytometry. We describe below how the system works. Racks of blood samples in standard ethylene diamine tetraacetic acid anticoagulated blood tubes are placed into the device for processing. For each tube in turn, a small volume (30 ll) of blood is first withdrawn from the tube, after automated tube handling and mixing. From this volume, a blunt stainless steel needle withdraws an aliquot and transfers a precise volume (a nominal 1 ll with 1% coefficient of variation) of blood directly onto a microscope slide (Figure 1). The blunt needle lays down adjacent rows of blood in a monolayer in which all cells can be distinguished and counted (Figure 2) Arch Pathol Lab Med Vol 141, August 2017 A Slide-Based Approach to Blood Cell Assessment Winkelman et al

3 Figure 2. A, The aliquot of whole blood is applied onto the microscope slide in adjacent rows, with all cells analyzable by computer imaging algorithms. B, Magnification of a section of the slide (original magnification 310). After the initial preparation, the slide is air-dried, fixed, and stained with a Romanowsky-type stain where the eosin solution and the methylene blue oxidized solution are applied sequentially, followed by rinsing. The stain formula was optimized to allow excellent visualization and assessment of the cell morphology from either the glass slide or from images on a computer screen, and to provide for consistent high-quality measurements by computerized analysis. An illustration of the staining mechanism used in the system is seen in Figure 3. The device stains the cells by infusing and then evacuating small amounts of the various reagents within a narrow space between the upside-down slide and a plastic surface. After staining, the slide is transferred to the first of 2 digital imaging stations. At the first station, the digital imaging is performed at low magnification to count RBCs, WBCs, platelets, and nucleated RBCs. A very fast linear stage is used to move the slide, and images are acquired by a black-and-white charge-coupled device (CCD) camera while the specimen is illuminated with different wavelengths of light using 4 light-emitting diodes (LEDs). An image for each color is acquired sequentially and at different focus levels to allow dynamic tracking of focus while the slide is being analyzed. The optical systems are designed to provide high-quality images without the use of oil immersion. A view of the slide stage of the imaging assembly is shown in Figure 4. After the low-magnification analysis is complete, the slide is transferred to the high-magnification imaging station. At this second imaging station at least 600 locations known to contain WBCs are digitized, again using LED illumination and a high-resolution, black-and-white charge-coupled device camera. From these images, a WBC differential of more than 600 cells is determined using image analysis techniques that take into consideration nuclear and cytoplasmic size, shape, texture, and optical absorption characteristics. For RBCs, cellular hemoglobin and volume are determined for approximately RBCs, and potential RBC inclusions are noted. From the approximately RBCs assessed, MCH is determined from optical absorption of specific wavelengths, and total hemoglobin is calculated from the MCH and total RBC count. The MCV is measured from the same RBCs by determining cellular volume using the size and optical absorption characteristics at multiple wavelengths of the individual RBCs. The MCV is determined from the average measurement of these cells, and hematocrit is calculated from the MCV and total RBC count. Distribution widths of the RBCs are calculated from the measured values of cellular volume. Additional analysis of the high-magnification images provides information on Figure 3. Rapid staining is enabled by using separate blue and red components, while applying small volumes of reagents. Figure 4. A high-speed servo motor stage moves the microscope slide under the imaging optics. Arch Pathol Lab Med Vol 141, August 2017 A Slide-Based Approach to Blood Cell Assessment Winkelman et al 1109

4 Figure 5. For white blood cell, red blood cell, and platelet review, individual cells are shown as a gallery of cells. A cell can be enlarged and displayed with its surrounding cells, as shown. individual platelet volumes, allowing mean platelet volume to be determined. The instrument is designed to move slides from station to station once a minute. The slide preparation and the 2 imaging stations each require 1 minute, whereas the staining mechanism requires 2 minutes of processing. For this reason, there are 2 stainers built in so that the entire system, with the 2 staining stations working in parallel on different patient samples, can provide results at the rate of 1 CBC with WBC differential per minute, for 60 patient samples per hour. In practice, the first sample in a batch requires 5 minutes for the preparation (1 minute), staining (2 minutes), low-magnification imaging (1 minute), and high-magnification imaging (1 minute). After that first sample, the remaining samples are processed and reported at a rate of 1 per minute. A display is used to both monitor the operation of the instrument and to review data and images from flagged specimens requiring further review. This viewing station resides on a state-of-the-art computer with proprietary software. The complete system is capable of measuring and reporting all standard CBC parameters and the WBC differential counts. The reportable data include a standard set of 23 parameters. Additionally, a reticulocyte count can be performed. When this is requested, an aliquot is mixed with a supravital stain before the slide is produced, stained, and imaged. Instrument messages or flags are similar to those found on flow systems, with the important distinction that when the operator clicks the mouse on a particular message on the computer screen, the program brings the operator immediately to the relevant images to allow review and resolution of the flag. White blood cells, RBCs, and platelets are presented in image galleries (Figure 5). The WBCs are partitioned by cell type, including the 5 standard normal categories as well as a sixth category of unclassified cells (ie, cells not of the 5 normal types) requiring review by an operator. Individual cells can be magnified on the computer screen and reclassified by the operator as appropriate. Double-clicking on any WBC image zooms in on a particular cell in a field of neighboring cells, providing the operator an opportunity to evaluate abnormalities in a broader context. The WBC differential values are updated automatically if any WBCs are reclassified during this review. Individual RBCs and platelets also may be shown isolated from their neighboring cells, which allows an entire gallery of those cells to be arranged and displayed. For example, Correlation of the Major Complete Blood Count (CBC) and Differential Parameters a Parameter Pearson r Value White blood cells 0.99 Red blood cells 0.99 Hemoglobin 0.99 Hematocrit 0.99 Mean cellular volume 0.90 Mean corpuscular hemoglobin 0.97 Platelets 0.98 Mean platelet volume 0.87 % Neutrophils 0.98 % Lymphocytes 0.97 % Monocytes 0.76 % Eosinophils 0.96 % Basophils 0.63 a A total of 1857 samples were included in the CBC correlation, and 1263 in the white blood cell differential correlation Arch Pathol Lab Med Vol 141, August 2017 A Slide-Based Approach to Blood Cell Assessment Winkelman et al

5 Figure 6. Comparability for the major parameters of (A) white blood cell (WBC), (B) red blood cell (RBC), (C) platelet (PLT), and (D) % neutrophil (NEUT) counts. The y-axis shows the new instrument s results; the x-axis shows the comparative flow-based system. RBCs can be sorted by size, shape, or hemoglobin content. In addition, the operator can sort platelets by volume. This allows the operator to quickly recognize the presence of unusual RBC or platelet morphology. COMPARISON STUDIES Initial proof of concept studies were performed to compare the CBC results obtained on the prototype instrument with those from the Sysmex XE-5000 Hematology Analyzer (Sysmex Inc, Kobe, Japan). Specimens for the comparison studies were obtained from the hematology laboratory of a large teaching hospital (Tufts Medical Center, Boston, Massachusetts) and tested per protocols for discarded specimens that were approved by that institution s review board (IRB No. 7492). As part of the subsequent commercial development phase, the initial studies presented here are being extensively enlarged and Arch Pathol Lab Med Vol 141, August 2017 A Slide-Based Approach to Blood Cell Assessment Winkelman et al 1111

6 all conventional measures validated for accuracy, precision, analytic and reporting ranges, interferences, and other characteristics. Documentation of those studies is beyond the scope of this initial presentation of the new technology. For this study, specimens were analyzed from a wide array of deidentified adult and pediatric inpatients and outpatients with a range of medical conditions, including hematologic and other malignancies. Specimens were sought and collected for comparison that had very high or low results for various parameters, without regard to diagnosis. The prototype instrument and the Sysmex XE-5000 Hematology Analyzer were first cocalibrated by being run for 2 consecutive days, processing 364 samples. During the 11 subsequent days, data were collected from a total of 1857 samples analyzed on both instruments, running first on the comparative analyzer, then on the new instrument. Both runs were performed within 4 hours of venipuncture. Of the 1857 samples thus analyzed, WBC differential data were available for 1263 samples. The Table shows the Pearson correlation coefficients for the major values of the CBC parameters and the percentages from the automated differential count. All parameters showed good to excellent correlation over a range of values spanning /ll to /ll; /ll to /ll; and /ll to /ll for RBCs, WBCs, and platelets, respectively. In Figure 6, scatter plots of the WBC, RBC, platelet, and neutrophil counts compare the new instrument to the comparative system. A high degree of correlation is noted in these plots. The full limits of the RBC, WBC, and platelet linearity range and reporting range are currently under investigation through extensive multicenter studies on the finalized version of the instrument. DISCUSSION A unique aspect of the new slide-based system is that its underlying methodology is based on cell identification and counting that are identical to those performed in conventional microscopy. The application of multispectral imaging and proprietary computer algorithms allows an accurate and precise assessment of all CBC and WBC differential parameters in a single instrument. In spite of the different methodology compared with flow-based instruments, all CBC and differential values appear to correlate well. Correlation coefficients and offsets between the new instrument and the comparative instrument show values very similar to those found when comparing 2 different flow-based analyzers. 11 In the new image-based system, both the actual slide and digitized images are available to operators and clinicians for real-time or subsequent review, resolution of abnormalities, archiving, and staff education. The system s throughput of 60 CBCs with automated WBC differentials per hour, after an initial dwell time of 5 minutes, is comparable to current flow-based systems. It provides the opportunity for realtime, more efficient resolution of abnormalities, compared with the current practice of postanalysis slide making and staining, and either manual or automated WBC differential counting. Our study has several limitations. We did not fully assess the effect of interfering substances on the CBC and automated differential count. These effects will be addressed fully in future reports employing the final commercial version of the product. However, it is intrinsic to our method that certain interferences that are important for flow-based CBC instruments be obviated with our imagebased system. For example, regardless of their nature, it is important that substances that can lead to clumping of formed elements that might affect flow cytometric measurements, such as cryoglobulins or other RBC agglutinins, elevated fibrinogen, macroglobulins, or naturally occurring factors that lead to platelet clumping, rouleaux, or other conditions such as polycythemia or thrombocytopenia, be recognized and corrected for by image analysis. Rouleaux formation, for example, can be observed directly. As mentioned previously, RBCs in contiguity or overlapping are detected and counted. Similarly, individual platelets within small clumps can be counted directly by the instrument software. The MCH, MCV, and inclusions are determined using single RBCs found in the more than 600 high-resolution images. The generally recognized RBC inclusion bodies can be identified by the image recognition software, with accuracy dependent on their prevalence and stain quality. Further, precision studies have been performed on a limited number of specimens. Results conformed to conventional standards (data not shown), with coefficients of variation similar to those of flow cytometry based systems. Large-scale multicenter studies designed to establish definitive precision as well as accuracy, linearity, and reportable ranges for each CBC parameter using the fully developed system are underway. Comprehensive evaluation of recognition and classification of various RBC shape abnormalities and how to quantify them are part of ongoing clinical studies. To our knowledge, we describe here, for the first time, a novel method for a monolayer preparation of a known volume of peripheral blood onto a microscope slide and an image analysis based approach for determining the CBC and WBC differential counts from the slide. References 1. Winkelman J, Tanasijevic M, Zahniser D, inventors; Roche Diagnostics Hematology Inc, assignee. Method for determining a complete blood count on a white blood cell differential count. US patent August 26, Hematology analyzers product guide. CAP Today. December 2014: Aller R. High volume hematology analyzers. CAP Today. December 2005: Novis DA. Laboratory productivity and the rate of manual peripheral blood smear review. Arch Pathol Lab Med. 2006;130(5): Miller MN. Design and clinical results of Hematrak: an automated differential counter. IEEE Trans Biomed Eng. 1976;23(5): Cotter DA, Sage BA. Performance of the LARC Classifier in clinical laboratories. J Histochem Cytochem. 1976;24(1): Daoust PR. The clinical detection of variations in the concentrations of normal leukocyte types: a laboratory comparison of 100-cell manual differential counts on wedge smears and 500-cell counts by the ADC-500. Blood Cells. 1980; 6(3): Marchand A, VanLente F, Galen RS. Automated differential leukocyte counters: a comparison of three systems. J Clin Lab Autom. 1983;3: Winkelman JW, Morris MN, O Leary M. Spuriously elevated band counts with an automated differential counter. J Clin Lab Autom. 1963;3(6): Cornet E, Perol JP, Troussard X. Performance evaluation and relevance of the CellaVision DM96 system in routine analysis and in patients with malignant hematological diseases. Int J Lab Hematol. 2008;30(6): Meintker L, Ringwald J, Rauh M, Krause SW. Comparison of automated differential blood cell counts from Abbott Sapphire, Siemens Advia 120, Beckman Coulter DxH 800, and Sysmex XE-2100 in normal and pathologic samples. Am J Clin Pathol. 2013;139(5): Arch Pathol Lab Med Vol 141, August 2017 A Slide-Based Approach to Blood Cell Assessment Winkelman et al

HST-330 TM and Total Laboratory Automation

HST-330 TM and Total Laboratory Automation HST-330 TM and Total Laboratory Automation Hyun-Sook CHI, MD* * Department of Clinical Pathology, University of Ulsan College of Medicine and Asan Medical Center, Seoul, Korea. Total laboratory automation

More information

Automated Hematology Analyzer. Small Footprint. Big Difference.

Automated Hematology Analyzer. Small Footprint. Big Difference. XS-1000i Automated Hematology Analyzer Small Footprint. Big Difference. Small But Powerful: The XS-1000i Even with these challenges, the need for hematology testing has remained steady or continues to

More information

See Brilliant Results

See Brilliant Results See Brilliant Results without all the reviews. * In Development Product not available in the U.S. First Pass Efficiency. Getting it right the first time. GOALS: More reportable WBC and WBC differential

More information

Ruby. A shining example of. Put science on your side. Intended Use:

Ruby. A shining example of. Put science on your side. Intended Use: High efficiency hematology A shining example of advanced technology Intended Use: The CELL-DYN Ruby is a multi-parameter automated Hematology analyzer designed for in vitro use in clinical laboratories.

More information

MANAGEMENT BRIEF. How Automation is Revolutionizing White Cell Differential Analysis.

MANAGEMENT BRIEF. How Automation is Revolutionizing White Cell Differential Analysis. MANAGEMENT BRIEF How Automation is Revolutionizing White Cell Differential Analysis Introduction Confidence building becomes a real issue when a laboratory technologist generalist must classify abnormal

More information

BEYOND A BETTER BOX DI-60 INTEGRATED SLIDE PROCESSING SYSTEM. Seamless Integration

BEYOND A BETTER BOX DI-60 INTEGRATED SLIDE PROCESSING SYSTEM. Seamless Integration BEYOND A BETTER BOX DI-60 INTEGRATED SLIDE PROCESSING SYSTEM Seamless Integration RESHAPING HEMATOLOGY Scalable Automation and On-board Slide Processing The Sysmex DI-60 Integrated Cell Image Analyzer

More information

Performance Evaluation of the Coulter LH 750 Hematology Analyzer

Performance Evaluation of the Coulter LH 750 Hematology Analyzer ISLH Laboratory Hematology 7:217 228 2001 Carden Jennings Publishing Co., Ltd. Official Publication Performance Evaluation of the Coulter LH 750 Hematology Analyzer TRACEY FERNANDEZ, LYNN BESSERT DOMACK,

More information

BC Auto Hematology Analyzer. A CUTE 5-part

BC Auto Hematology Analyzer. A CUTE 5-part BC-5000 Auto Hematology Analyzer A CUTE 5-part Why do we need 5-part hematology analyzers? WBC differential: 3-part WBC differential: 5-part Lymphocyte Mid-size cell? Eosinophil Basophil Neutrophil DIFF

More information

BC Auto Hematology Analyzer. A CUTE 5-part

BC Auto Hematology Analyzer. A CUTE 5-part BC-5150 Auto Hematology Analyzer A CUTE 5-part Why do we need 5-part hematology analyzers? WBC differential: 3-part WBC differential: 5-part Lymphocyte Mid-size cell? Eosinophil Basophil Neutrophil DIFF

More information

CONCURRENT 10: THE COMPLETE BLOOD COUNT AND BEYOND: QUALITY ISSUES AND REFERENCE METHODS

CONCURRENT 10: THE COMPLETE BLOOD COUNT AND BEYOND: QUALITY ISSUES AND REFERENCE METHODS CONCURRENT 10: THE COMPLETE BLOOD COUNT AND BEYOND: QUALITY ISSUES AND REFERENCE METHODS Saturday May 14 th 2016 1:30 3:00 PM (13.30 15.00hrs) International Society for Laboratory Hematology (ISLH) CONCURRENT

More information

Upon completion of the Clinical Hematology rotation, the MLS student will be able to:

Upon completion of the Clinical Hematology rotation, the MLS student will be able to: Clinical Performance Objectives in Clinical Hematology Department of Medical and Research Technology University of Maryland School of Medicine Spring 2015 Upon completion of the Clinical Hematology rotation,

More information

Evaluation of the Automated Immature Granulocyte Count (IG) on Sysmex XE-2100 Automated Haematology Analyser vs. Visual Microscopy (NCCLS H20-A)

Evaluation of the Automated Immature Granulocyte Count (IG) on Sysmex XE-2100 Automated Haematology Analyser vs. Visual Microscopy (NCCLS H20-A) Evaluation of the Automated Immature Granulocyte Count (IG) on Sysmex XE-21 Automated Haematology Analyser vs. Visual Microscopy (NCCLS H2-A) Th WEILAND *1, H KALKMAN *2, and H HEIHN *1 *1 Central Laboratory,

More information

Automated Hematology Analyzer. Differentiate with the XT-4000i

Automated Hematology Analyzer. Differentiate with the XT-4000i XT-4000i Automated Hematology Analyzer Differentiate with the XT-4000i Advanced Technology Solutions to Meet Your Lab s Needs Even with these challenges, the need for hematology testing has remained steady

More information

Pushing limits in routine laboratory haematology with the XT-4000i

Pushing limits in routine laboratory haematology with the XT-4000i Pushing limits in routine laboratory haematology with the XT-4000i Sysmex Xtra Online December 2013 Fig. 1 The XT-4000i haematology system With the launch of the XE-5000, Sysmex set new standards in haematology.

More information

SF Cube. Exclusive Technology, Inclusive Approach. BC-6800 Auto Hematology Analyzer. Cell Analysis Technology

SF Cube. Exclusive Technology, Inclusive Approach. BC-6800 Auto Hematology Analyzer. Cell Analysis Technology Cell Analysis Technology Exclusive Technology, Inclusive Approach BC-6800 Auto Hematology Analyzer Cell Analysis Technology SF Cube SF Cube is a pathbreaking technology for reliable blood cell analysis,

More information

Automated Hematology System. Your Complete Choice

Automated Hematology System. Your Complete Choice XE-5000 Automated Hematology System Your Complete Choice Advanced Technology Solutions to Meet Your Lab s Needs Even with these challenges, the need for hematology testing has remained steady or continued

More information

XN Series. What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems Sysmex America, Inc. All rights reserved.

XN Series. What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems Sysmex America, Inc. All rights reserved. XN Series What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems 2012 Sysmex America, Inc. All rights reserved. Agenda XN Series Product Overview A system to fit every lab

More information

Automation In Life Sciences and Diagnostics Applications

Automation In Life Sciences and Diagnostics Applications Automation In Life Sciences and Diagnostics Applications Pedro Diaz Director of Research Beckman Coulter Life Sciences February 20-22, 2013 Orlando World Marriott Center Orlando, Florida USA Beckman Coulter

More information

New Frontiers in the US Hematology and Flow Cytometry Market: Business Challenges, Emerging Technologies, Competitive Landscape.

New Frontiers in the US Hematology and Flow Cytometry Market: Business Challenges, Emerging Technologies, Competitive Landscape. New Frontiers in the US Hematology and Flow Cytometry Market: Business Challenges, Emerging Technologies, Competitive Landscape Table of Contents I. Introduction II. Worldwide Market and Technology Overview

More information

ABX Pentra DX120. Hematology analyzer. 45 parameters Integrated slide-maker Expert validation system

ABX Pentra DX120. Hematology analyzer. 45 parameters Integrated slide-maker Expert validation system ABX Pentra DX120 Hematology analyzer 45 parameters Integrated slide-maker Expert validation system ABX Pentra DX 120 In the heart of hematopoiesis Complete cytology platform 120 samples per hour Continuous

More information

CellaVision DM96. Automated Digital Cell Morphology

CellaVision DM96. Automated Digital Cell Morphology CellaVision DM96 Automated Digital Cell Morphology Advanced Technology Automated Digital Cell Morphology is the process by which cells are automatically located on a stained peripheral blood smear, pre-classified,

More information

Features & Benefits Five-part differential on cat, dog, horse, cow, alpaca and llama; three-part differential on the nine other species

Features & Benefits Five-part differential on cat, dog, horse, cow, alpaca and llama; three-part differential on the nine other species Telephone: Facsimile: e-mail: + 27 (0)11 791 0025 + 27 (0)86 551 0832 johan@anjotech.co.za VetScan HM5 Hematology System Advanced Hematology Five-Part Differential The VetScan HM5 is a fully automated

More information

New Frontiers in the UK Hematology and Flow Cytometry Market: Business Challenges, Emerging Technologies, Competitive Landscape.

New Frontiers in the UK Hematology and Flow Cytometry Market: Business Challenges, Emerging Technologies, Competitive Landscape. New Frontiers in the UK Hematology and Flow Cytometry Market: Business Challenges, Emerging Technologies, Competitive Landscape Table of Contents I. Introduction II. Worldwide Market and Technology Overview

More information

Verification of abnormal results from Coulter instrumentation

Verification of abnormal results from Coulter instrumentation Created Lisa Senzel Asst. Prof. - Clinical 5/9/2008 Path Revised Bruce Kube Day Heme Tech 1 10/24/2012 Reviewed Jay Bock M.D. 4/10/2012 Procedure: Approved Lisa Senzel Asst. Prof. - Clinical Path Replaces

More information

AUTOMATED HEMATOLOGY IN THE CLINICAL LAB

AUTOMATED HEMATOLOGY IN THE CLINICAL LAB AUTOMATED HEMATOLOGY IN THE CLINICAL LAB Diane Gaspari, SH(ASCP) Core Lab Manager York Hospital, York PA ggaspari@wellspan.org Objectives of Automation Increase efficiency; streamline workflow Increase

More information

The Simple Solution for CBC Testing

The Simple Solution for CBC Testing QBC STAR Centrifugal Hematology Analyzer The Simple Solution for CBC Testing Diagnostics Innovative Solutions for a Healthier World UNRIVALED SIMPLICITY The QBC STAR Centrifugal Hematology System is the

More information

Diagnostic Medical Equipment New Revenue for Your Medical Practice

Diagnostic Medical Equipment New Revenue for Your Medical Practice Diagnostic Medical Equipment New Revenue for Your Medical Practice Consultant Libby Knollmeyer s Lab Machine Primer Series Part 1: Hematology Analyzers When laboratories in physician offices move out of

More information

Reduction of the Platelet Review Rate Using the Two-Dimensional Platelet Method

Reduction of the Platelet Review Rate Using the Two-Dimensional Platelet Method Coagulation and Transfusion Medicine / REDUCTION OF THE PLTELET REVIEW RTE Reduction of the Platelet Review Rate Using the Two-Dimensional Platelet Method Deanne H. Chapman, MT(SCP)SH, 1 Jonn Hardin, MT,

More information

XS-1000i New Sysmex 5-part diff haematology analyser with fluorescence technology

XS-1000i New Sysmex 5-part diff haematology analyser with fluorescence technology XS-1000i New Sysmex 5-part diff haematology analyser Fig. 1: sysmex xs-1000i Since the introduction of fluorescence technology for leukocyte differentiation by sysmex in the year 1999, thousands of laboratories

More information

Novel POC analysis for determination of total and 5-part differential WBC count among a US population, in comparison to Beckman Coulter LH750

Novel POC analysis for determination of total and 5-part differential WBC count among a US population, in comparison to Beckman Coulter LH750 Novel POC analysis for determination of total and -part differential WBC count among a US population, in comparison to Beckman Coulter LH7 Reed, Jeremy*, Johnsson, Elin**, Hockum, Sandra** * Quest Diagnostics,

More information

Immature Granulocyte Enumeration Our Journey from Manual to Automated Reporting

Immature Granulocyte Enumeration Our Journey from Manual to Automated Reporting Immature Granulocyte Enumeration Our Journey from Manual to Automated Reporting Mary Capper MT(ASCP)SH Supervisor, Hematology/Hemostasis University of Iowa Hospitals and Clinics Objectives Describe the

More information

Evaluation of the Sysmex XN-550, a Novel Compact Haematology analyser from the XN-L series, compared to the XN-20 system.

Evaluation of the Sysmex XN-550, a Novel Compact Haematology analyser from the XN-L series, compared to the XN-20 system. Evaluation of the Sysmex XN-550, a Novel Compact Haematology analyser from the XN-L series, compared to the XN-20 system. H Tailor 1, I Mackie 2, A Mellick 3, S Machin 1, 2 1 Haematology Evaluations Unit,

More information

Feature. Hematology Analyzer: From Workhorse to Thoroughbred. Ellen Sullivan

Feature. Hematology Analyzer: From Workhorse to Thoroughbred. Ellen Sullivan Hematology Analyzer: From Workhorse to Thoroughbred Ellen Sullivan DOI: 10.1309/TMQ6T4CBCG408141 labmedicine.com May 2006 Volume 37 Number 5 LABMEDICINE 273 Hematology analyzers are the workhorses of the

More information

BEYOND A BETTER BOX XN-V SERIES TM MULTISPECIES HEMATOLOGY ANALYZERS. Accelerate Your Multispecies Research Laboratory XN-V IS FOR ANIMAL USE ONLY

BEYOND A BETTER BOX XN-V SERIES TM MULTISPECIES HEMATOLOGY ANALYZERS. Accelerate Your Multispecies Research Laboratory XN-V IS FOR ANIMAL USE ONLY BEYOND A BETTER BOX XN-V SERIES TM MULTISPECIES HEMATOLOGY ANALYZERS Accelerate Your Multispecies Research Laboratory XN-V IS FOR ANIMAL USE ONLY MOVE YOUR MULTI-SPECIES LABORATORY BEYOND A BETTER BOX*

More information

ABX Pentra XL 80. Hematology analyzer. Auto-sampling 5-part Differential Customized Dilution Ratio Integrated Validation Station

ABX Pentra XL 80. Hematology analyzer. Auto-sampling 5-part Differential Customized Dilution Ratio Integrated Validation Station ABX Pentra XL 80 Hematology analyzer Auto-sampling 5-part Differential Customized Dilution Ratio Integrated Validation Station Pentra XL 80 A new way to manage your hematology workflow With Pentra XL 80

More information

Sysmex Educational Enhancement and Development No

Sysmex Educational Enhancement and Development No SEED Haematology Sysmex Educational Enhancement and Development No 5 2012 The role of the peripheral blood smear in the modern haematology laboratory The purpose of this newsletter is to provide an overview

More information

Cell analysis and bioimaging technology illustrated

Cell analysis and bioimaging technology illustrated Cell analysis and bioimaging technology illustrated The Cell Analysis Center Scientific Bulletin Part 1 Sysmex has been studying and exploring principles of automated haematology analysers, making full

More information

Dry Hematology Analyzer Fast, Simple CBC Analysis Designed for the Point of Care

Dry Hematology Analyzer Fast, Simple CBC Analysis Designed for the Point of Care QBC STAR TM Dry Hematology Analyzer Fast, Simple CBC Analysis Designed for the Point of Care Reach for the STAR The QBC STAR hematology analyzer expands and advances the unique capabilities of previous

More information

Reshaping compact automation

Reshaping compact automation Automated Hematology Analyzer XN-SERIES Reshaping compact automation www.sysmex.com/us Optimize your laboratory As analyzers become increasingly sophisticated, the laboratory s contribution to the continuum

More information

Reshaping compact automation

Reshaping compact automation Automated Hematology Analyzer XN-SERIES Reshaping compact automation www.sysmex.com/us Optimize your laboratory As analyzers become increasingly sophisticated, the laboratory s contribution to the continuum

More information

UNRIVALED SIMPLICITY. Exceptional Ease of Use. The STAR brings new meaning to the word simple

UNRIVALED SIMPLICITY. Exceptional Ease of Use. The STAR brings new meaning to the word simple UNRIVALED SIMPLICITY The QBC STAR Centrifugal Hematology System is the simplest solution for in-office CBC testing. An innovative approach employing dry hematology reagents instead of the bulky liquid

More information

XN-SERIES AUTOMATED HEMATOLOGY SYSTEMS

XN-SERIES AUTOMATED HEMATOLOGY SYSTEMS XN-SERIES AUTOMATED HEMATOLOGY SYSTEMS Reshaping Compact Automation BEYOND A BETTER BOX MOVE YOUR HEMATOLOGY LABORATORY BEYOND A BETTER BOX It s what makes Sysmex the smarter choice and sets our integrated

More information

Automated Leukocyte Differentials: A Review and Prospectus

Automated Leukocyte Differentials: A Review and Prospectus Symposium Automated Leukocyte Differentials: A Review and Prospectus Thomas F. Dutcher, MD A utomation, or more correctly, instrumentation, that provides a high degree of accuracy and precision in blood

More information

Cell Imaging Analysis: Our journey to Hematology Productivity

Cell Imaging Analysis: Our journey to Hematology Productivity Cell Imaging Analysis: Our journey to Hematology Productivity Kelly T. Yoder MS, MT(ASCP)SH Administrative Lab Director Indiana Regional Medical Center Indiana, PA February 9, 2013 Agenda Introduction

More information

BC-6800 Auto Hematology Analyzer. Small Cube, Big Difference

BC-6800 Auto Hematology Analyzer. Small Cube, Big Difference BC-6800 Auto Hematology Analyzer Small Cube, Big Difference Small Cube, Big Difference 2D Forward scatter EOS NEU + Side scatter MON LYM 3D Forward scatter MON NEU LYM EOS + Side scatter + Fluorescence

More information

XN-9100 AUTOMATED HEMATOLOGY SYSTEMS

XN-9100 AUTOMATED HEMATOLOGY SYSTEMS XN-9100 AUTOMATED HEMATOLOGY SYSTEMS Reshaping Scalable Automation BEYOND A BETTER BOX MOVE YOUR HEMATOLOGY LABORATORY BEYOND A BETTER BOX It s what makes Sysmex the smarter choice and sets our integrated

More information

DxH part differential closed tube hematology analyzer

DxH part differential closed tube hematology analyzer DxH 520 5-part differential closed tube hematology analyzer DxH 520 HEMATOLOGY ANALYZER Minimum footprint. Maximum productivity. The new closed tube 5-part differential analyzer because great things come

More information

Manual For Blood Cells Atlas Morphology Download

Manual For Blood Cells Atlas Morphology Download We have made it easy for you to find a PDF Ebooks without any digging. And by having access to our ebooks online or by storing it on your computer, you have convenient answers with manual for blood cells

More information

Red Blood Cell Counter Using Embedded Image Processing Techniques

Red Blood Cell Counter Using Embedded Image Processing Techniques Red Blood Cell Counter Using Embedded Image Processing Techniques Mariya Yeldhos* and K.P. Peeyush 1 Department of Biomedical instrumentation engineering, Faculty of Engineering, Avinashilingam Institute

More information

COMPARISON OF PLATELET COUNTS BY SYSMEX XE 2100 AND LH 750 WITH THE INTERNATIONAL FLOW REFERENCE METHOD IN THROMBOCYTOPENIC PATIENTS

COMPARISON OF PLATELET COUNTS BY SYSMEX XE 2100 AND LH 750 WITH THE INTERNATIONAL FLOW REFERENCE METHOD IN THROMBOCYTOPENIC PATIENTS COMPARISON OF PLATELET COUNTS BY SYSMEX XE 2100 AND LH 750 WITH THE INTERNATIONAL FLOW REFERENCE METHOD IN THROMBOCYTOPENIC PATIENTS TINA DADU 1, KUNAL SEHGAL 2, ANJUM SHAIKH 2, SHANAZ KHODAIJI 2 1. BLK

More information

NEW PRODUCT DEVELOPMENT OPPORTUNITES AND BUSINESS EXPANSION STRATEGIES FOR INSTRUMENT AND REAGENT SUPPLIERS FRANCE GERMANY ITALY SPAIN UK USA

NEW PRODUCT DEVELOPMENT OPPORTUNITES AND BUSINESS EXPANSION STRATEGIES FOR INSTRUMENT AND REAGENT SUPPLIERS FRANCE GERMANY ITALY SPAIN UK USA MEDICAL DEVICES PHARMACEUTICALS CHEMICALS FOOD & BEVERAGE ELECTRONICS This study is being updated. Subscribers will receive complimentary copies of the previous edition while the new report is in preparation.

More information

Greater Possibilities By Design

Greater Possibilities By Design Greater Possibilities By Design January 23, 2013 Nilam Patel MT(ASCP)SH Sr. Product Manager, Automation Solutions Objectives Review current and new portfolio Technology and benefits Operational benefits

More information

True Walkaway Automated Urinalysis, Chemistry to Microscopy The CLINITEK AUWi System Answers for life.

True Walkaway Automated Urinalysis, Chemistry to Microscopy The CLINITEK AUWi System Answers for life. www.siemens.com/diagnostics True Walkaway Automated Urinalysis, Chemistry to Microscopy The CLINITEK AUWi System Answers for life. Can I achieve true unattended urinalysis testing in my laboratory? With

More information

Diagnostics solutions. hematology reagents and Analyzers

Diagnostics solutions. hematology reagents and Analyzers Diagnostics solutions hematology reagents and Analyzers Avantor Performance Materials is extending the reach of reliable diagnostic solutions to help improve healthcare outcomes and quality of life for

More information

A Division of Woodley Equipment Company Ltd

A Division of Woodley Equipment Company Ltd LEADING EDGE TECHNOLOGY FOR POINT OF CARE HAEMATOLOGY A Division of Woodley Equipment Company Ltd Contents The QBC STAR POC Haematology System...................4 Operating Ranges.......................................7

More information

11/2/2015. Describe the technologies available on the XN-Series and the reason(s) for selection

11/2/2015. Describe the technologies available on the XN-Series and the reason(s) for selection XN-Series Technology What s Inside the Box? OBJECTIVES Describe the technologies available on the XN-Series and the reason(s) for selection Utilize example case studies to illustrate the new capabilities

More information

Carl Zeiss MicroImaging AIS. Digital Pathology The Future. Andrew Lesniak Director, Product Management

Carl Zeiss MicroImaging AIS. Digital Pathology The Future. Andrew Lesniak Director, Product Management Carl Zeiss MicroImaging AIS Digital Pathology The Future Andrew Lesniak Director, Product Management Defined Industry Needs Computer Aided Imaging and Analysis will play a key role in the increase of Diagnostics

More information

norma Hematology family norma Access norma norma Vision Reagents

norma Hematology family norma Access norma norma Vision Reagents norma Hematology family norma Access norma Reagents norma Vision Smallest reagent consumption in the world Smallest processed sample volume, ideal for pediatric use Microfluidic technology 3 A smart hematology

More information

Examples of body fluid analysis with the XE-5000

Examples of body fluid analysis with the XE-5000 Examples of body fluid analysis with the XE-5000 Sysmex Xtra Online March 2011 If patients are to be treated rapidly and effectively, laboratory results must be both rapid and reliable if at all possible,

More information

norma Hematology family norma norma Vision Reagents

norma Hematology family norma norma Vision Reagents norma Hematology family norma Reagents norma Vision Smallest reagent consumption in the world Smallest processed sample volume, ideal for pediatric use Microfluidic technology 3 The Icon 3 is a smart,

More information

Auto-Validation Don t Rule It Out!

Auto-Validation Don t Rule It Out! Auto-Validation Don t Rule It Out! Elinore Craig Product Manager, IT Integration Agenda Barriers to auto-validation Success stories Overcoming the barriers 1 Barriers AACC Auto-validation Survey About

More information

Lab: Blood Smear and RBC Count

Lab: Blood Smear and RBC Count 2014-09- 01 Blood Page 1 of 5 Lab: Blood Smear and RBC Count Aim: Learn to count cells, observe and identify different blood cells in a smear, quantify their proportions and count RBCs per µl (mm^3) using

More information

ALL INFECTOLOGY. and now. Gucker, 1947 Wallace Coulter Coulter Orifice, Kamentsky, 1965 Mack Fulwyler sorter BLOOD:

ALL INFECTOLOGY. and now. Gucker, 1947 Wallace Coulter Coulter Orifice, Kamentsky, 1965 Mack Fulwyler sorter BLOOD: ALL BLOOD: RBC, PLT, WBC ( /-/ ) (+ LDH, carbamide, ESR, periphery blasts, liver function deviation, renal function dev.) Classification: Morphology + genetics+ immunology! B-cell: T-cell: Dedifferentiated:

More information

Beckman Coulter Spans the Biomedical Testing Continuum

Beckman Coulter Spans the Biomedical Testing Continuum Beckman Coulter Spans the Biomedical Testing Continuum What is Middleware? Software interface between analytical instruments/software and LIS/HIS Enables technologists to control multiple lab processes

More information

Simply stated, hematology analyzers determine the

Simply stated, hematology analyzers determine the Simply stated, hematology analyzers determine the quantity of each type of cell that makes up the patient s blood. Only recently has the role and importance of each blood component been understood. As

More information

Department of Pathology

Department of Pathology KASTURBA HEALTH SOCIETY'S MAHATMA GANDHI INSTITUTE OF MEDICAL SCIENCES, P.O. SEVAGRAM, WARDHA (DISTT.) MAHARASHTRA STATE - 442 102 Specifications for Equipment Department of Pathology 1) 3 Part Hematology

More information

ALL INFECTOLOGY COMMERCIALLY AVAILABLE DEVICES. and now BLOOD:

ALL INFECTOLOGY COMMERCIALLY AVAILABLE DEVICES. and now BLOOD: ALL BLOOD: RBC, PLT, WBC ( /-/ ) (+ LDH, carbamide, ESR, periphery blasts, liver function deviation, renal function dev.) Classification: Morphology + genetics+ immunology! B-cell: T-cell: Dedifferentiated:

More information

Business Case for Haematology Analyser Replacement Recommendation

Business Case for Haematology Analyser Replacement Recommendation Business Case for Haematology Analyser Replacement Recommendation Agenda Item 108/11 1. Executive Summary The Business Case [BC] identifying the need to replace the analytical platform in the Haematology

More information

XN Hematology Case Studies Every Picture Tells a Story

XN Hematology Case Studies Every Picture Tells a Story XN Hematology Case Studies Every Picture Tells a Story Barbara J Connell MS, MT(ASCP)SH Senior Product Manager Technical Marketing & Communications Sysmex America, Inc Objectives Describe the advantages

More information

CELL-DYN ImmunoPlt (CD61)

CELL-DYN ImmunoPlt (CD61) CELL-DYN ImmunoPlt (CD61) REF REAGENT 03H05-01 40 tests 1. NAME CELL-DYN ImmunoPlt (CD61) Reagent 2. INTENDED USE CELL-DYN ImmunoPlt (CD61) Reagent is a one-color direct immunofluorescence reagent for

More information

BEYOND A BETTER BOX XN-9000 AUTOMATED HEMATOLOGY SYSTEMS. Reshaping Scalable Automation

BEYOND A BETTER BOX XN-9000 AUTOMATED HEMATOLOGY SYSTEMS. Reshaping Scalable Automation BEYOND A BETTER BOX XN-9000 AUTOMATED HEMATOLOGY SYSTEMS Reshaping Scalable Automation 2 Sysmex XN-Series Automated Hematology Analyzers are designed to exceed the expectations of real people who work

More information

Landing Cyto System. Automated platform for GYN (compliant with The Bethesda System) and non-gyn cytology. Automated abnormal cell detection

Landing Cyto System. Automated platform for GYN (compliant with The Bethesda System) and non-gyn cytology. Automated abnormal cell detection Automated platform for GYN (compliant with The Bethesda System) and non-gyn cytology Landing Cyto System Automated abnormal cell detection Automation Intelligence Standardization FEATURE SUMMARY The Landing

More information

The Story of the Platelet Clump. Wayne Hall MTN Network Laboratory Magee-Womens Research Institute Pittsburgh, PA

The Story of the Platelet Clump. Wayne Hall MTN Network Laboratory Magee-Womens Research Institute Pittsburgh, PA The Story of the Platelet Clump Wayne Hall MTN Network Laboratory Magee-Womens Research Institute Pittsburgh, PA Reason for Alarm? Platelet Clumping was occurring in multiple African sites 1-2 times per

More information

ICSH guidelines for the verification and performance of automated cell counters for body fluids

ICSH guidelines for the verification and performance of automated cell counters for body fluids International Journal of Laboratory Hematology ORIGINAL ARTICLE The Official journal of the International Society for Laboratory Hematology INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY ICSH guidelines

More information

NovoCyte Flow Cytometer

NovoCyte Flow Cytometer NovoCyte Flow Cytometer The Flow Cytometer for Everyone 2 Experience the NovoCyte Advantage Focus on advancing your research. Let the flow cytometer do the rest. NovoCyte Flow Cytometer High Performance

More information

Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer

Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer Journal of Automated Methods & Management in Chemistry, 5 (5), no. 4, 35 39 c 5 Hindawi Publishing Corporation Validation of Platelet ing Accuracy With the Celltac F Automated Hematology Analyzer Yutaka

More information

Comparison of Two Methods for the Determination of the Effects of Ionizing Radiation on Blood Cell Counts in Mice

Comparison of Two Methods for the Determination of the Effects of Ionizing Radiation on Blood Cell Counts in Mice International journal of Biomedical science ORIGINAL ARTICLE Comparison of Two Methods for the Determination of the Effects of Ionizing Radiation on Blood Cell Counts in Mice Ana L. Romero-Weaver, Ann

More information

A previous ICCS Module entitled Instrument optimization - Adjusting PMT voltages and compensation 1 should be read as a prerequisite to this module.

A previous ICCS Module entitled Instrument optimization - Adjusting PMT voltages and compensation 1 should be read as a prerequisite to this module. Sponsored and reviewed by ICCS Quality and Standards Committee Title: Compensation Tips for Beckman Coulter 10-Color Navios Platform Written by: Salima Janmohamed-Anastasakis Ph.D., Applications Scientist,

More information

Case 1:17-cv DPG Document 1 Entered on FLSD Docket 11/03/2017 Page 1 of 10

Case 1:17-cv DPG Document 1 Entered on FLSD Docket 11/03/2017 Page 1 of 10 Case 1:17-cv-24049-DPG Document 1 Entered on FLSD Docket 11/03/2017 Page 1 of 10 BECKMAN COULTER, INC., IN THE UNITED STATES DISTRICT COURT FOR THE SOUTHERN DISTRICT OF FLORIDA MIAMI DIVISION v. Plaintiff,

More information

RDTs for use in areas with P.falciparum HRP2 deletions and P.vivax infections

RDTs for use in areas with P.falciparum HRP2 deletions and P.vivax infections RDTs for use in areas with P.falciparum HRP2 deletions and P.vivax infections Jane Cunningham XV Reunion Annual de Evaluacion de AMI/RAVREDA Bogota, Colombia 3 5 May 2016 Introduction High prevalence of

More information

BEYOND A BETTER BOX XN-SERIES TM AUTOMATED HEMATOLOGY SYSTEMS. Reshaping Compact Automation

BEYOND A BETTER BOX XN-SERIES TM AUTOMATED HEMATOLOGY SYSTEMS. Reshaping Compact Automation BEYOND A BETTER BOX XN-SERIES TM AUTOMATED HEMATOLOGY SYSTEMS Reshaping Compact Automation SMALL, SMART, COMPACT AUTOMATION It s an approach to hematology testing that s never been done before. An approach

More information

Tailored automation solutions that fit. Automation system solutions component overview

Tailored automation solutions that fit. Automation system solutions component overview Tailored automation solutions that fit Automation system solutions component overview Blood Banking Centrifugation Chemistry Flow Cytometry Hematology Hemostasis Immunoassay Information Systems Lab Automation

More information

Quintus 5-part hematology analyzer In-depth, quality-controlled 26-parameter results

Quintus 5-part hematology analyzer In-depth, quality-controlled 26-parameter results Outstanding user-interface Only three reagents needed 100-sample Autoloader option Quintus 5-part hematology analyzer In-depth, quality-controlled 26-parameter results Premium system solution boosted by

More information

Note: The taller 100 x 13 mm tube is not designed to run on the ACL TOP. Plasma after centrifugation of these tubes must be placed in sample cups.

Note: The taller 100 x 13 mm tube is not designed to run on the ACL TOP. Plasma after centrifugation of these tubes must be placed in sample cups. Created Revised Reviewed Approved Lupus Insensitive aptt on the ACL TOP HEM 4.15.1 7/13/2012 4/11/2013 4/11/2013 4/11/2013 I. PRINCIPLE The activated partial thromboplastin time is a global screening procedure

More information

Tailored automation solutions that fit. Automation system solutions component overview

Tailored automation solutions that fit. Automation system solutions component overview Tailored automation solutions that fit Automation system solutions component overview Blood Banking Centrifugation Chemistry Flow Cytometry Hematology Hemostasis Immunoassay Information Systems Lab Automation

More information

Transforming Cell Science

Transforming Cell Science Transforming Cell Science T E C H N OT E 1 1 / 2 0 1 1 Comparison of Moxi Z Cell Counter with an Imaging Cell Counter and a High-End Coulter System Introduction Cell counts are routinely performed in life

More information

CHAPTER-6 HISTOGRAM AND MORPHOLOGY BASED PAP SMEAR IMAGE SEGMENTATION

CHAPTER-6 HISTOGRAM AND MORPHOLOGY BASED PAP SMEAR IMAGE SEGMENTATION CHAPTER-6 HISTOGRAM AND MORPHOLOGY BASED PAP SMEAR IMAGE SEGMENTATION 6.1 Introduction to automated cell image segmentation The automated detection and segmentation of cell nuclei in Pap smear images is

More information

Human Cell Analysis The Technology Behind The World s Most Common Diagnos8c Test

Human Cell Analysis The Technology Behind The World s Most Common Diagnos8c Test Human Cell Analysis The Technology Behind The World s Most Common Diagnos8c Test IEEE- CNSV November 4, 2014 Giacomo Vacca, Ph.D. Founder & President Kine?c River Corp. Proprietary 1 OVERVIEW Blood Components

More information

Inoculation efficiency is the first step in the bacteriology testing process and is the first opportunity for automation.

Inoculation efficiency is the first step in the bacteriology testing process and is the first opportunity for automation. 0000 Inoculation Inoculation efficiency is the first step in the bacteriology testing process and is the first opportunity for automation. SAMPLES ENABLING TIME OPTIMIZATION UP TO 2HR TRUE WALK-AWAY TIME

More information

Schedule of Accreditation

Schedule of Accreditation Schedule of Accreditation Organisation Name INAB Reg No Contact Name Address West North West Hospitals Group Galway University Hospital Dept Haematology 239MT Mary Kilcooley Contact Phone No 091 544419

More information

9/29/2015. Describe the technologies available on the XN-Series and the reason(s) for selection

9/29/2015. Describe the technologies available on the XN-Series and the reason(s) for selection XN-Series Technology What s Inside the Box? OBJECTIVES Describe the technologies available on the XN-Series and the reason(s) for selection Utilize example case studies to illustrate the new capabilities

More information

SAMPLE. Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard Second Edition

SAMPLE. Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard Second Edition June 2010 Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard Second Edition This document provides guidance for the validation, verification, calibration,

More information

Biological Variations of Leukocyte Numerical and Morphologic Parameters Determined by UniCel DxH 800 Hematology Analyzer

Biological Variations of Leukocyte Numerical and Morphologic Parameters Determined by UniCel DxH 800 Hematology Analyzer Biological Variations of Leukocyte Numerical and Morphologic Parameters Determined by UniCel DxH 800 Hematology Analyzer Huqiang Tang, MS, MPH; Jiyong Jing, MS; Dandan Bo, BS; Dongsheng Xu, MD, PhD N Context.

More information

BioView. Karyotype Suite

BioView. Karyotype Suite Imaging and Analysis Solutions BioView Karyotype Suite BioView s wide range of automated Karyotyping imaging and analysis platforms ensure optimal solutions for various sample volumes and automation level

More information

Harmonization of red blood cell distribution width (RDW): an attainable target?

Harmonization of red blood cell distribution width (RDW): an attainable target? Original Article Page 1 of 5 Harmonization of red blood cell distribution width (RDW): an attainable target? Giuseppe Lippi 1, Silvia Pipitone 2, Emmanuel J. Favaloro 3,4 1 Section of Clinical Biochemistry,

More information

Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer

Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer Automated Methods & Management in Chemistry, 25 (25), no. 4, 235 239 c 25 Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer Yutaka Nagai, 1 Hiroshi Kondo, 2 and

More information

WADA Technical Document TD2019BAR. Blood Analytical Requirements for the Athlete Biological Passport

WADA Technical Document TD2019BAR. Blood Analytical Requirements for the Athlete Biological Passport 1. Introduction Blood Analytical Requirements for the Athlete Biological Passport This Technical Document (TD) has been established to harmonize the analysis of blood Samples collected, both In-Competition

More information

The Future of Hematological Diagnosis

The Future of Hematological Diagnosis Interview The Future of Hematological Diagnosis Bruce H. Davis, M.D. Bruce H. Davis, M.D. is the Director of Analytical Cytometry in the Department of Clinical Pathology at the William Beaumont Hospital

More information

Laboratory Accreditation Test Validation: A Brave New World for Anatomic Pathology

Laboratory Accreditation Test Validation: A Brave New World for Anatomic Pathology Laboratory Accreditation Test Validation: A Brave New World for Anatomic Pathology Francis E. Sharkey, MD, FCAP University of Texas Health Science Center, San Antonio, TX Richard W. Brown, MD, FCAP Memorial

More information

2111: ALL Post-HCT. Add/ Remove/ Modify. Manual Section. Date. Description. Comprehensive Disease- Specific Manuals

2111: ALL Post-HCT. Add/ Remove/ Modify. Manual Section. Date. Description. Comprehensive Disease- Specific Manuals 2111: ALL Post-HCT The Acute Lymphoblastic Leukemia Post-HCT Data Form is one of the Comprehensive Report Forms. This form captures ALL-specific post-hct data such as: planned treatments post-hct, the

More information