1 Article Reprint Comparison between IRIS iq 200ELITE and microscopy for urinalysis and evaluation of performance in predicting outcome of urine cultures S. Ledru, J.P. Canonne Laboratoire de microbiologie, Centre hospitalier Schaffner, Lens Annales de Biologie Clinique, Volume 66, Number 5, September-October 2008 Article Received 2 June 2008, Accepted 4 September 2008

2 Comparison between IRIS iq 200ELITE and microscopy for urinalysis and evaluation of performance in predicting outcome of urine cultures S. Ledru, J.P. Canonne, Laboratoire de microbiologie, Centre hospitalier Schaffner, Lens Annales de Biologie Clinique, Volume 66, Number 5, September-October 2008, Article Received 2 June 2008, Accepted 4 September 2008 Key words: urinary tract infection, urine cell analyzer, urinalysis, comparative study Summary Since the marketing of the urinary strips, various automated instruments have appeared using the principle of flow cytometry for analysis of urinary elements; Sysmex UF-100, UF-50 and UF Recently a new analyzer, that uses an innovative method, the Iris iq 200ELITE, has been introduced that analyzes urine specimens by photographic analysis of cells passing in a laminar flow in front of a microscopic objective. Five hundred microscopic fields are photographed and analyzed. Each particle of the microscopic field is separated and classified using Auto- Particle Recognition software. The possibilities of visualization by the operator permits us to avoid confusion between various artifacts identified on flow cytometry analyzers (for example: yeasts confused with red blood cells, spermatozoa interpreted like bacilli, etc). Apart from any interpretation Iris iq200elite has compared to the culture a sensitivity of 68%, a specificity of 80%, a positive predictive value of 60% and a negative predictive value of 86%. By combining the various parameters provided by Iris iq200elite, we can eliminate 43% from the direct microscopic examinations and to predict the negativity of 32% of the cultures with a sensitivity and a negative predictive value of 100%, which quantitatively optimizes this most important work station. Introduction UTI s are the most frequent pathologies found in Private Laboratories and in hospitals. For its diagnosis, the gold standard remains a direct cytological examination followed by a culture 1. In our hospital, cytodiagnostic urinalysis is the number one analysis done in the laboratory between 20 and 25% of the occurrence which generates an important investment in materials and human resources. However, a majority of cytodiagnostic urinalysis are negative (70% in our series), but we cannot accept any tolerance compared to the bacteria culture because of the risks of chronic morbidity that a failure of diagnosis can involve. Various automated flow cytometry instruments (Sysmex UF-50, Sysmex UF-100, Sysmex UF-1000) were put on the market with the goal of screening urines to reduce the cost of those tests with more or less success 2-4. Here are the results of an analyzer which is comparable with the usual cytodiagnostic test: an analysis of pictures of cells passing in front of a microscope objective. Materials and Methods Fresh urine specimen samples arriving at the laboratory were analyzed in a consecutive way from February 18th to March 28th, of each sample was carried out on routine tests and an analysis was run on the Iris iq200elite (Iris Diagnotics France, Baillet-en-France, France). The routine analysis consisted of a cytological examination on Slide Fast-Read 102 (Biosigma, S.R.L., Italy) (same system as KOVA slides) with numeration of red blood cells and leucocytes, direct microscopic examination and a culture with AGAR plates. The direct microscopic examination was carried out in the following way: a gauged amount of 10 μl of urine was deposited and spread out in a circle traced with an engraver. The slide was dried and then it was colored by Gram stain and the full slide was observed under microscope at objective 100 with immersion oil. The results were reported in semi-quantitative grades according to Table 1. Presence of bacteria was performed using plates of BCP media (Bio-Rad ), CNA Colombia media (Oxoid ) and chromogene media for yeast (CAND BD ) if there was presence of yeasts with the microscopic examination. Cultures were interpreted according to the European recommendations 5. The urine was tested in parallel on the Iris iq200elite. The Iris iq200elite is an automated microscope. A known volume of urine passes in front of a microscope objective. A camera takes 500 fields per sample. Within each field, images of the particles are individualized and isolated. The particles are classified using a model of a neural network used by the software: Auto-Particle Recognition (APR ). In the majority of the tests, automatic classification can be realized in all confidence without any operator help. For samples with very abnormal particles, a user can visually confirm the identification of those particles according to their morphological details seen on the screen (Figure 1). The minimum size of the recognized particles is 3 μm; that is sufficient for bacilli and Gram positive cocci chains. Bacteria smaller than 3µm are classified in All Small Particles, which makes this parameter important to look at in correlation with leucocytes concentration and the presence of bacteria. Table 1: Interpretation of the direct microscopic examination. Observation 0 Absence < 1 in 10 fields Rare 1/10 fields to 1/field Occasional 1/field to 10/field 10/field to 100/field Semi-quantitative results Few Moderate > 100/field Many 1

3 Table 3: Diagnostic performance of the Iris iq200elite compared to direct microscopic examination (ED) and to bacteria cultures of the 400 urine samples. Comparison %FP 1 (n) %FN 2 (n) Se% 3 Spe 4 % PPV 5 % NPV 6 % ED Iris 7 /Culture 40 (56) 14 (38) ED 8 / Culture 30 (41) 9 (23) ED Iris 7 / ED 8 25 (34) 13 (34) Percentage of False Positive (number); 2 Percentage of False Negative (Number); 3 Sensitivity; 4 Specificity; 5 Positive Predictor Value; 6 Negative Predictor Value; 7 Result of the detection of bacteria by the analyzer; 8 Result of the detection of bacteria by the direct microscopic examination Figure 1: Particle images from iq 200ELITE. The user can visually confirm the identification of particles according to their morphological details seen on the screen. Table 2: Distribution of the various bacterial pathogens isolated from urine samples. Bacteria Escherichia coli 73 Enterococcus 23 Proteus mirabilis 10 Klebsiella pneumoniae 10 Number Table 4: Combinations of various values provided by Iris iq200elite and the direct microscopic examination (ED). Comparison of the parameters of Iris iq200elite to direct microscopic examination ED Iris Negative 7 and WBC 8 < 5 ED Iris Negative 7 and WBC 8 > 40 ED Iris Negative 7 and WBC 8 < 40 and ASP 9 < 10,000 %FP 1 (n) %FN 2 (n) Se 3 % Spe 4 % PPV 5 % NPV 6 % 58 (190) 0 (0) (58) 36 (115) (102) 8 (5) (a) Percentage of False Positive (number); 2 Percentage of False Negative (Number); 3 Sensitivity; 4 Specificity; 5 Positive Predictor Value; 6 Negative Predictor Value; 7 Absence of bacteria detected by the analyzer; 8 Leukocytes; 9 All Small Particles; (a) 8 cultures not significant (6 negative, 2 bacteria in insufficient concentration). Enterobacter aerogenes 6 Other Enterobacteriaceae 9 Candida albicans 8 Torulopsis glabrata 2 Pseudomonas aeruginosa 3 Staphylococcus aureus 3 Staphylococcus with negative coagulase enzyme 3 Streptococcus agalactiae 2 2

4 Leukocytes 40/µL 5/µL N = 54 ED Mandatory N = 98 ED Negative = 90 ED Positive = 8 8 Cultures not significant N = 70 ED = Negative Negative Culture 10,000/µL N = 26 ED Mandatory N = 12 ED Mandatory N = 2 ED = Negative Negative Culture All Small Particles Figure 2: Opportunity of doing a direct microscopic examination (ED) according to the parameters provided by Iris iq200elite in case of an absence of bacteria (N = 262). Direct on Analyzer (N = 400) Negative (N=262) Positive (N=138) Leukocytes Leukocytes 10/µL (N=124) > 10/µL (N=138) 100/µL (N=57) > 100/µL (N=81) All Small Particles Direct and Culture Mandatory Direct Direct and Culture Mandatory < 10,000/µL No Direct No Culture (Sterile) (N=119) Negative No Culture (Sterile) (N=3) 10,000/µL Direct Positive Culture Mandatory (N=2) Negative (N=16) Culture (Sterile) Positive (N=41) Culture Mandatory Figure 3: Decisional algorithm with a negative predictive value of 100% to reduce of the number of cultures and give faster results of negative cultures on patient s report. 3

5 Results During this period, 400 urines were consecutively analyzed. The specimen analyzed consists of 255 women and 145 men age 0 to 99 years (average age: 53). The distribution between hospital departments was: 161 patients from medical department, 88 from gynecology-obstetrics department, 41 of surgery department, 34 from pediatric department, 29 from geriatric department, 28 from emergencies, 13 from general-purpose and 6 from psychiatry department. The incidence of positive cultures was 30% (120 of 400). Only one pathogen was isolated in 88 from the 120 positive bacteria cultures, 2 pathogens in 27 bacteria cultures and finally 3 pathogens in 5 cultures. The distribution of the bacteria is presented in Table 2. The direct microscopic examination and culture remaining the reference method 1, were initially compared to results of the analyzer (Table 3). One notes a good correlation between the Iris iq200elite results and direct microscopic examination, and also between results on the Iris iq200elite and bacteria cultures. However, as for the microscopic direct examination, a negative result in bacteria and yeasts on Iris iq200elite does not make it possible to omit the culture (more sensitive and more specific). To improve the productivity of the cytodiagnostic urinalysis job, we searched for combinations of various parameters from the Iris iq200elite (numeration of the leucocytes, numeration of All Small Particles) in order to eliminate direct examinations (time-saver) in the case of absence of bacteria detected by the Iris iq200elite (Table 4). When leucocytes are lower than 5/μL, whatever the number of All Small Particles, the direct microscopic examination is always negative. When leucocytes are more than 40/μL, whatever the number of All Small Particles, the direct microscopic examination is mandatory because of the possibility of false negative (36%). When the leucocytes concentration is between 5/μL and 40/μL, we must take into consideration the All Small Particles count. If the number of All Small Particles is lower than 10,000/μL, the direct examination is negative in 92% of the cases and the 8% of the cases when it was positive, the cultures were not significant. However, if the number of All Small Particles is higher than 10,000/μL, one cannot bypass direct examination because of the possibility of presence of Gram positive cocci colonies. All of the possible situations are summarized in Figure 2. In our series, that allows us to eliminate 43% of the direct microscopic examinations with a sensitivity and a negative predictive value of 100% on direct microscopic examination. On the same principle, we searched for the combinations which could enable us to predict the negativity of cultures (Figure 3). In case of absence of bacteria and yeasts, with a number of leucocytes lower or equal to 10/µL and of a number of small particles less than 10,000/µL, the direct microscopic examination is always negative and culture is sterile (sensitivity = 100%, specificity = 43%, PPV = 43%, NPV = 100%). We can also predict the negativity of the cultures with a sensitivity and a NPV of 100% if the direct examination on the instrument is negative, the number of leucocytes lower or equal to 10/µL, the number of small particles equal to or higher than 10,000/µL and a negative direct microscopic examination or if the direct examination on the analyzer is positive, and the number of leucocytes is lower or equal to 100/μL and the direct microscopic examination is negative. This algorithm would allow us, if the French health reimbursement mode should change, to eliminate 32% of the cultures with a sensitivity and a negative predictive value of 100%. Discussion The goal of this study was to evaluate the performance of Iris iq200elite in order to decrease the number of microscopic examinations and to anticipate the negativity of the cultures with a sensitivity and a negative predictive value from 100% on one of the most important work stations in microbiology laboratory: cytodiagnostic and microbiological urinalysis. Other authors already studied the performances of the strips and analyzers which use the flow cytometry technology. Some authors are satisfied this analysis by accepting a percentage of false negatives. It is the case of Evans, et al. 6 and Kim, et al. 7 for the UF-100 (Sysmex, Japan). Other authors are not satisfied with those results and do cultures on all urines because of the possibility of false negatives 4. Iris iq200elite has a major advantage over others systems of detection: that is the possibility of pictures of urine particles. Compared to the other systems, that allows the avoidance of several pitfalls. According to Steinmetz, et al. 8, UF-100 over-estimates the presence of leucocytes and red blood cells. This over-estimation can be due to the presence of bacteria or yeasts which are counted like red blood cells 9, oxalate crystals 10, amorphous phosphate, yeast or spermatozoa 11,12. It is very easy on Iris iq200elite to eliminate those artifacts. Thanks to this innovative process, we can visualize the elements and thus eliminate artifacts (Figure 1). 4

6 Conclusion The Iris iq200elite enabled us to significantly increase the productivity of the work station of the cytodiagnostic and microbiology urinalysis. It allows us to eliminate 43% of direct examinations of urine specimens per day. This automation allows a standardization of results and precision of the direct examinations. More, the possibility of connection to LIS decreases the risks of transcription errors and optimizes the ergonomics of this work station. Automation enables us to predict during the current day 32% of the urinary cultures which will be negative with a sensitivity and a negative predictive value of 100%. References 1. Kouri T, Fogazzi G, Gant V, Hallander H, Hofmann W, Guder WG. European urinalysis guidelines. ECLM-European urinalysis group. Scand J Clin Lab Invest 2000 ; 60(Suppl.) : Boucaud-Maitre Y, Thoinet S, Fuhrmann C, Coronel B, Freney J. Evaluation des performances et de la praticabilité de l automate de cytobactériologie Sysmex UF-100TM pour la prédiction de l infection urinaire. Rev Fr Lab 2001 ; 338 : Okada H, Sakai Y, Miyazaki S, Arakawa S, Hamagushi Y, Kamidono S. Detection of significant bacteriuria by automated urinalysis using flow cytometry. J Clin Microbiol 2000 ; 38 : Zaman Z, Roggeman S, Vaerhaegen J. Unsatisfactory performance of flow cytometer UF-100 and urine strips in predicting outcome of urine cultures. J Clin Microbiol 2001 ; 39 : Aspevall O, Hallander H, Gant V, Kouri T. European guidelines for urinalysis : a collaborative document produced by European clinical microbiologists and clinical chemists under ECLM (European Confederation for Clinical Microbiology and Infectious Diseases). Clin Microbiol Infection 2001 ; 7 : Evans R, Davidson MM, Sim LW, Hay A. Testing by Sysmex UF-100 flow cytometer and with bacterial culture in a diagnostic laboratory : a comparison. J Clin Pathol 2006 ; 59 : Kim SY, Kim YJ, Hwang SH, Kim HH, Son HC, Lee E. Evaluation of the Sysmex UF-100 urine cell analyser as a screening test to reduce the need for urine cultures for community-acquired urinary tract infection. Am J Clin Pathol 2007 ; 128 :

7 8. Steinmetz J, Henny J, Gueguen R. Etablissement de limites de référence des sédiments urinaires mesurés sur UF-50TM Sysmex. Ann Biol Clin 2004 ; 62 : Fenili D, Pirovano B. The automation of sediment urinalysis using a new urine flow cytometer (UF-50TM). Clin Chem Lab Med 1998 ; 36 : Hannemen-Pohl K, Kampf SC. Automation of urine sediment examination : a comparison of the Sysmex UF-100 automated flow cytometer with routine manual diagnosis (microscopy, test strips, and bacterial culture). Clin Chem Lab Med 1999 ; 37 : Langlois MR, Delanghe JR, Steyaert SR, Everaert KC, De Buyzerz ML. Automated flow cytometry compared with an automated dipstick reader for urinalysis. Clin Chem 1999 ; 45 : Lun A, Zoebog R, Priem F, Filler G. Routine workflow for use of urine strip and urine cytometer UF-100 in the hospital laboratory. Clin Chem 2001 ; 45 :

8 9172 Eton Avenue Chatsworth, CA A Division of IRIS International, Inc. Copyright 2008 Iris Diagnostics. All rights reserved Rev. A

Cornelia Ottiger and Andreas R. Huber * Analytical Techniques. Clinical Chemistry 49: (2003) Automation and

Cornelia Ottiger and Andreas R. Huber * Analytical Techniques. Clinical Chemistry 49: (2003) Automation and Clinical Chemistry 49:4 617 623 (2003) Automation and Analytical Techniques Quantitative Urine Particle Analysis: Integrative Approach for the Optimal Combination of Automation with UF-100 and Microscopic

More information

Urinalysis: Comparison between Microscopic Analysis and a New Automated Microscopy Image-Based Urine Sediment Instrument

Urinalysis: Comparison between Microscopic Analysis and a New Automated Microscopy Image-Based Urine Sediment Instrument Clin. Lab. 2014;60:693-697 Copyright SHORT COMMUNICATION Urinalysis: Comparison between Microscopic Analysis and a New Automated Microscopy Image-Based Urine Sediment Instrument PAULA V. BOTTINI, MAYARA

More information

Screening of urine samples by flow cytometry reduces the need for culture

Screening of urine samples by flow cytometry reduces the need for culture JCM Accepts, published online ahead of print on 30 June 2010 J. Clin. Microbiol. doi:10.1128/jcm.00617-10 Copyright 2010, American Society for Microbiology and/or the Listed Authors/Institutions. All Rights

More information

True Walkaway Automated Urinalysis, Chemistry to Microscopy The CLINITEK AUWi System Answers for life.

True Walkaway Automated Urinalysis, Chemistry to Microscopy The CLINITEK AUWi System Answers for life. www.siemens.com/diagnostics True Walkaway Automated Urinalysis, Chemistry to Microscopy The CLINITEK AUWi System Answers for life. Can I achieve true unattended urinalysis testing in my laboratory? With

More information

Guidelines for Laboratory Verification of Performance of the FilmArray Blood Culture Identification (BCID) Panel

Guidelines for Laboratory Verification of Performance of the FilmArray Blood Culture Identification (BCID) Panel Guidelines for Laboratory Verification of Performance of the FilmArray Blood Culture Identification (BCID) Panel Purpose The Clinical Laboratory Improvement Amendments (CLIA), passed in 1988, establishes

More information

Protocols for Laboratory Verification of Performance of the FilmArray Blood Culture Identification (BCID) Panel

Protocols for Laboratory Verification of Performance of the FilmArray Blood Culture Identification (BCID) Panel Protocols for Laboratory Verification of Performance of the FilmArray Blood Culture Identification (BCID) Panel Purpose The Clinical Laboratory Improvement Amendments (CLIA), passed in 1988, establishes

More information

F or many years, the isolation of urinary pathogens has

F or many years, the isolation of urinary pathogens has 608 ORIGINAL ARTICLE A comparison of the performance of commercially available chromogenic agars for the isolation and presumptive identification of organisms from urine D Fallon, G Ackland, N Andrews,

More information

Protocols for Laboratory Verification of Performance of the BioFire FilmArray Blood Culture Identification (BCID) Panel

Protocols for Laboratory Verification of Performance of the BioFire FilmArray Blood Culture Identification (BCID) Panel Protocols for Laboratory Verification of Performance of the BioFire FilmArray Blood Culture Identification (BCID) Panel A Laboratory Protocol for Use with Live s Purpose The Clinical Laboratory Improvement

More information

Improving the efficiency and efficacy of pre-analytical and analytical work-flow of urine cultures with urinary flow cytometry

Improving the efficiency and efficacy of pre-analytical and analytical work-flow of urine cultures with urinary flow cytometry NEW MICROBIOLOGICA, 31, 501-505, 2008 Improving the efficiency and efficacy of pre-analytical and analytical work-flow of urine cultures with urinary flow cytometry Shamanta Grosso, Graziano Bruschetta,

More information

Detection of Significant Bacteriuria. by Use of the iq200 Automated Urine Microscope

Detection of Significant Bacteriuria. by Use of the iq200 Automated Urine Microscope JCM Accepts, published online ahead of print on 28 May 2014 J. Clin. Microbiol. doi:10.1128/jcm.00112-14 Copyright 2014, American Society for Microbiology. All Rights Reserved. 1 2 3 4 5 6 7 8 9 10 11

More information

Protocols for Laboratory Verification of Performance of the BioFire FilmArray Blood Culture Identification (BCID) Panel

Protocols for Laboratory Verification of Performance of the BioFire FilmArray Blood Culture Identification (BCID) Panel Protocols for Laboratory Verification of Performance of the BioFire FilmArray Blood Culture Identification (BCID) Panel Laboratory Protocols for Use with Microbiologics Helix Elite Molecular Standards

More information

Disruptive Technology to Guide Precision Antibiotic Therapy

Disruptive Technology to Guide Precision Antibiotic Therapy Disruptive Technology to Guide Precision Antibiotic Therapy Company Presentation, October 2018 Forward Looking Statement This presentation includes statements relating to the company's Acuitas AMR Gene

More information

Table of Contents. A Culture of Service

Table of Contents. A Culture of Service 102714tr Table of Contents 1 BlueEcoli 2 Candida 3 CRE 4 ECC 5 HurBi 6 ESBL 7 Listeria 8 MRSA 9 O157 10 Sakazakii 11 Salmonella 12 SS 13 Staph aureus 14 UTI 15 Vibrio Headquarters A Culture of Service

More information

Looking to the Future: How Automation will Grow the Value of Microbiology

Looking to the Future: How Automation will Grow the Value of Microbiology Looking to the Future: How Automation will Grow the Value of Microbiology Nathan A Ledeboer Associate Professor of Pathology Medical College of Wisconsin Medical Director, Microbiology and Molecular Pathology

More information

The Cat s Out of the Bag: Microbiological Investigations of Acute Transfusion Reactions.

The Cat s Out of the Bag: Microbiological Investigations of Acute Transfusion Reactions. The Cat s Out of the Bag: Microbiological Investigations of Acute Transfusion Reactions. Philippe Lagacé-Wiens, MD FRCPC, DTM&H plagacewiens@sharedhealthmb.ca COI declaration I have no conflicts, real

More information

Chromogenic Culture Media

Chromogenic Culture Media Color is my day-long obsession, joy, and torment. - Claude Monet Chromogenic Culture Media A Culture of Service Contents BluEcoli CRE Candida ECC E. coli O157 ESBL HUrBi Listeria MRSA SS (Salmonella Shigella)

More information

2015 EDITION. Selection. Publications. FAN Plus Media

2015 EDITION. Selection. Publications. FAN Plus Media 2015 EDITION of Selection Publications FAN Plus Media INTRODUCTION biomérieux solutions for diagnosing bloodstream infections and sepsis. Sepsis, a systemic inflammation due to infection, can lead to severe

More information

Automated Direct Antimicrobial Susceptibility Testing of

Automated Direct Antimicrobial Susceptibility Testing of JOURNAL OF CLINICAL MICROBIOLOGY, Feb. 1980, p. 157-161 0095-1137/80/02-0157/05$02.00/0 Vol. 11, No. 2 Automated Direct Antimicrobial Susceptibility Testing of Microscopically Screened Urine Cultures ROBYN

More information

Dependability of sensitivity tests in primary culture

Dependability of sensitivity tests in primary culture J. clin. Path., 1976, 29, 179-184 Dependability of sensitivity tests in primary culture PAMELA M. WATERWORTH AND M. DEL PIANO1 From the Department of Microbiology, University College Hospital, London WCJ

More information

detection of microbial infection of urine

detection of microbial infection of urine Journal of Clinical Pathology, 1979, 32, 386-390 Particle size distribution analysis for the rapid detection of microbial infection of urine C. S. DOW1, A. D. FRANCE', M. S. KHAN2, AND T. JOHNSON2 From

More information

Evaluation of The BIOMIC V3 Microbiology System for Identification of Selected Species on BBL CHROMagar Orientation and CHROMagar MRSA ACCEPTED

Evaluation of The BIOMIC V3 Microbiology System for Identification of Selected Species on BBL CHROMagar Orientation and CHROMagar MRSA ACCEPTED JCM Accepts, published online ahead of print on 13 August 2008 J. Clin. Microbiol. doi:10.1128/jcm.02460-07 Copyright 2008, American Society for Microbiology and/or the Listed Authors/Institutions. All

More information

CORPORATE PRESENTATION

CORPORATE PRESENTATION CORPORATE PRESENTATION March 2018 1 Forward Looking Statement This presentation includes statements relating to the company's Acuitas MDRO and Acuitas Lighthouse clinical laboratory services, FDA cleared

More information

Rapid Detection of Bacterial Growth in Blood Cultures by Bioluminescent Assay of Bacterial ATP

Rapid Detection of Bacterial Growth in Blood Cultures by Bioluminescent Assay of Bacterial ATP JOURNAL OF CLINICAL MICROBIOLOGY, Sept. 1983, p. 521-525 0095-1137/83/090521-05$02.00/O Copyright C 1983, American Society for Microbiology Vol. 18, No. 3 Rapid Detection of Bacterial Growth in Blood Cultures

More information

Multicenter Evaluation of an Image Analysis Device (APAS): Comparison Between Digital Image and Traditional Plate Reading Using Urine Cultures

Multicenter Evaluation of an Image Analysis Device (APAS): Comparison Between Digital Image and Traditional Plate Reading Using Urine Cultures Original Article Clinical Microbiology Ann Lab Med 2017;37:499-504 https://doi.org/10.3343/alm.2017.37.6.499 ISSN 2234-3806 eissn 2234-3814 Multicenter Evaluation of an Image Analysis Device (APAS): Comparison

More information

Microbiology for Oral and Topical Products - The basics Scott Colbourne Business Manager NSW ALS Food & Pharmaceutical

Microbiology for Oral and Topical Products - The basics Scott Colbourne Business Manager NSW ALS Food & Pharmaceutical Microbiology for Oral and Topical Products - The basics Scott Colbourne Business Manager NSW ALS Food & Pharmaceutical RIGHT S O L U T I O N S RIGHT PARTNER Contents TGO 77 - Introduction Tests Performed

More information

Firefly Luciferase ATP Assay as a Screening Method for Bacteriuria

Firefly Luciferase ATP Assay as a Screening Method for Bacteriuria JOURNAL OF CLINICAL MICROBIOLOGY, Feb. 1983, p. 218-224 0095-1137/83/020218-07$02.OO/O Copyright 1983, American Society for Microbiology Vol. 17, No. 2 Firefly Luciferase ATP Assay as a Screening Method

More information

Automated methods in the Microbiology Lab-Issues and Troubleshooting! Presented By: Dr B. Oboudi

Automated methods in the Microbiology Lab-Issues and Troubleshooting! Presented By: Dr B. Oboudi In the name of GOD Automated methods in the Microbiology Lab-Issues and Troubleshooting! Presented By: Dr B. Oboudi Overview Background/history Plate streakers and Gram stainers Blood cultures Automated

More information

Urinary tract infections (UTIs) are the second most common infections, only after respiratory tract tinfections.

Urinary tract infections (UTIs) are the second most common infections, only after respiratory tract tinfections. Detection of fu Uropathogens by Using Chromogenic Mdi Media Farzaneh Azizmohseni, Director of Persian Type Culture Collection ((PTCC) Introduction Urinary tract infections (UTIs) are the second most common

More information

A membrane filter technique for testing disinfectants

A membrane filter technique for testing disinfectants J. clin. Path., 1975, 28, 71-76 A membrane filter technique for testing disinfectants JEAN PRINCE', C. E. A. DEVERILL, AND G. A. J. AYLIFFE From the Hospital Infection Research Laboratory, Birmingham SYNOPSIS

More information

O ver recent years, the systematic identification of all

O ver recent years, the systematic identification of all 524 ORIGINAL ARTICLE A comparison of the performance of cystine lactose electrolyte deficient (CLED) agar with Oxoid chromogenic urinary tract infection (CUTI) medium for the isolation and presumptive

More information

Project 5: Urine Cultures and Identification

Project 5: Urine Cultures and Identification Project 5: Urine Cultures and Identification Readings: http://www.webmd.com/a-to-z-guides/urine-culture http://www.medscape.com/viewarticle/558845 (Listen to the two lectures by Dr. Robert A. Weinstein.)

More information

Automation, Artificial Intelligence and Molecular Diagnostics; What is the Future of Microbiology?

Automation, Artificial Intelligence and Molecular Diagnostics; What is the Future of Microbiology? Automation, Artificial Intelligence and Molecular Diagnostics; What is the Future of Microbiology? Nathan A Ledeboer Associate Professor of Pathology Medical College of Wisconsin Medical Director, Microbiology

More information

WP8 Branka Bedenic, University Hospital Zagreb, Croatia Valentino D Onofrio, Jessa hospital and Hasselt University, Belgium

WP8 Branka Bedenic, University Hospital Zagreb, Croatia Valentino D Onofrio, Jessa hospital and Hasselt University, Belgium WP8 Branka Bedenic, University Hospital Zagreb, Croatia Valentino D Onofrio, Jessa hospital and Hasselt University, Belgium Inge Gyssens, Jessa hospital and Hasselt University and Radboud University Medical

More information

Preparing and Gram-staining a bacteriological smear

Preparing and Gram-staining a bacteriological smear College of Life Sciences and Technology Medical Laboratory Science (Applied Learning) AP 84-205-00 (42) Module 4: Practical 2014-16 Preparing and Gram-staining a bacteriological smear Date Time Class Venue

More information

GUIDELINES FOR WRITING A LAB REPORT FOR BIOL 215L (MICROBIOLOGY FOR HEALTHCARE PROFESSIONALS)

GUIDELINES FOR WRITING A LAB REPORT FOR BIOL 215L (MICROBIOLOGY FOR HEALTHCARE PROFESSIONALS) GUIDELINES FOR WRITING A LAB REPORT FOR BIOL 215L (MICROBIOLOGY FOR HEALTHCARE PROFESSIONALS) Your lab report will focus only on your unknown bacteria, which you collected, cultured, isolated, analyzed,

More information

Rapid Bacterial Identification Using a Mass Spectrometry Based Molecular Diagnostics Approach: Evaluation of the Iridica Platform

Rapid Bacterial Identification Using a Mass Spectrometry Based Molecular Diagnostics Approach: Evaluation of the Iridica Platform Rapid Bacterial Identification Using a Mass Spectrometry Based Molecular Diagnostics Approach: Evaluation of the Iridica Platform Alec Saitman, PhD, Jane Y. Yang PhD, Sharon Reed, David Pride, Michele

More information

HardyCHROM HUrBi Biplate

HardyCHROM HUrBi Biplate HardyCHROM HUrBi Biplate Cat. no. J100 HUrBi, 15x100mm Biplate, 10ml/10ml 10 plates/bag INTENDED USE HardyCHROM HUrBi Biplate is a selective chromogenic medium recommended for the cultivation, differentiation

More information

Preparing and Gram-staining a bacteriological smear

Preparing and Gram-staining a bacteriological smear College of Life Sciences and Technology Medical Laboratory Science (Applied Learning) AP 84-205-00 (52) Module 4: Practical 2015-17 Preparing and Gram-staining a bacteriological smear Date Time Class Venue

More information

Athanassios Tsakris Medical School, University of Athens Greece

Athanassios Tsakris Medical School, University of Athens Greece Athanassios Tsakris Medical School, University of Athens Greece Each hour of delay (over the first 6 hours) resulting in a 7.6% decrease in survival «In septic shock, time is life» Kumar A, Crit Care Med

More information

Indigo-Clean White Paper: #1 Bactericidal Performance Testing of Indigo-Clean Upon Bacterial Species. Healthcare

Indigo-Clean White Paper: #1 Bactericidal Performance Testing of Indigo-Clean Upon Bacterial Species. Healthcare Indigo-Clean White Paper: #1 Bactericidal Performance Testing of Indigo-Clean Upon Bacterial Species Healthcare HC Bactericidal Performance Testing of Indigo-Clean Upon Bacterial Species Clifford J. Yahnke,

More information

RAPID DIAGNOSIS OF BACTERIURIA BY MEASURING BACTERIAL ADENOSINE TRIPHOSPHATE USING A BIOLUMINESCENCE ASSAY

RAPID DIAGNOSIS OF BACTERIURIA BY MEASURING BACTERIAL ADENOSINE TRIPHOSPHATE USING A BIOLUMINESCENCE ASSAY Rapid diagnosis of bacteriuria using a bioluminescence assay RAPID DIAGNOSIS OF BACTERIURIA BY MEASURING BACTERIAL ADENOSINE TRIPHOSPHATE USING A BIOLUMINESCENCE ASSAY SATOSHI ISHIHARA, HIDEJI HAYASHI,

More information

Published Standards ISO General instructions for microbiological examination ISO Enumeration and detection of aerobic mesophilic bacteri

Published Standards ISO General instructions for microbiological examination ISO Enumeration and detection of aerobic mesophilic bacteri ISO Standards Cosmetics Microbiology TC 217 WG1 Joyce Beauchamp L OREAL USA Products, Inc. 2010 PCPC Microbiology Workshop ISO TC 217 WG1 Microbiological standards and limits Secretariat IRISI Secretary

More information

Urine Monovette with Boric Acid

Urine Monovette with Boric Acid with Boric Acid Simulated use - Study showing stabilization of specific microorganisms in urine prepared by: Aktiengesellschaft & Co. SARSTEDT AG & Co. P.O. Box 1220 D-51582 Nümbrecht Phone (+49) 0 22

More information

Int.J.Curr.Microbiol.App.Sci (2014) 3(10)

Int.J.Curr.Microbiol.App.Sci (2014) 3(10) ISSN: 2319-7706 Volume 3 Number 10 (2014) pp. 810-815 http://www.ijcmas.com Original Research Article Comparison of Tissue Culture plate method, Tube Method and Congo Red Agar Method for the detection

More information

BD CHROMagar Orientation Medium / Columbia CNA Agar (Biplate)

BD CHROMagar Orientation Medium / Columbia CNA Agar (Biplate) INSTRUCTIONS FOR USE READY-TO-USE PLATED MEDIA PA-254489.03 Rev.: Sep 2011 BD CHROMagar Orientation Medium / Columbia CNA Agar (Biplate) INTENDED USE BD CHROMagar Orientation Medium / Columbia CNA Agar

More information

IN THIS SECTION MICROBIOLOGY TESTING EXPERT SOLUTIONS FOR PRODUCT DEVELOPMENT. Bacterial Endotoxin (LAL) Testing

IN THIS SECTION MICROBIOLOGY TESTING EXPERT SOLUTIONS FOR PRODUCT DEVELOPMENT. Bacterial Endotoxin (LAL) Testing EXPERT SOLUTIONS FOR PRODUCT DEVELOPMENT IN THIS SECTION MICROBIOLOGY TESTING Microbial assays involve a variety of tests, from the determination of the numbers and types of organisms naturally present

More information

CLINICAL MICROBIOLOGY HOSPITAL LABORATORY PRACTICE

CLINICAL MICROBIOLOGY HOSPITAL LABORATORY PRACTICE Jordan University of Science and Technology Faculty of Applied Medical Sciences Department of Medical Laboratory Sciences CLINICAL MICROBIOLOGY HOSPITAL LABORATORY PRACTICE Course Syllabus Course Title

More information

What have we learnt from clinical trials in invasive candidiasis?

What have we learnt from clinical trials in invasive candidiasis? What have we learnt from clinical trials in invasive candidiasis? Eva Mª Roselló Micology Unit Microbiology Department Vall d Hebron Hospital Barcelona Epidemiology Candida spp: most common cause of invasive

More information

COMPARISON OF PLATELET COUNTS BY SYSMEX XE 2100 AND LH 750 WITH THE INTERNATIONAL FLOW REFERENCE METHOD IN THROMBOCYTOPENIC PATIENTS

COMPARISON OF PLATELET COUNTS BY SYSMEX XE 2100 AND LH 750 WITH THE INTERNATIONAL FLOW REFERENCE METHOD IN THROMBOCYTOPENIC PATIENTS COMPARISON OF PLATELET COUNTS BY SYSMEX XE 2100 AND LH 750 WITH THE INTERNATIONAL FLOW REFERENCE METHOD IN THROMBOCYTOPENIC PATIENTS TINA DADU 1, KUNAL SEHGAL 2, ANJUM SHAIKH 2, SHANAZ KHODAIJI 2 1. BLK

More information

M. Ben-David 1, O. Hammer 1, A.Shinderman 1, Y. Gluckman- Yavo 1, M. Fridman 1, D. Gohman 1, G. Ingber 1 and E. Zahavy 2

M. Ben-David 1, O. Hammer 1, A.Shinderman 1, Y. Gluckman- Yavo 1, M. Fridman 1, D. Gohman 1, G. Ingber 1 and E. Zahavy 2 437 Fast Antibiotic Susceptibility Testing Utilizing a Unique Spectral Intensity Ratio Analysis via Single Fluorescence Membrane Dye Staining and Flow Cytometry M. Ben-David 1, O. Hammer 1, A.Shinderman

More information

Centrifuge. a horizontal tube in the chamber, through a hole in a

Centrifuge. a horizontal tube in the chamber, through a hole in a JOURNAL OF CLINICAL MICROBIOLOGY, Dec. 1982, p. 1052-1056 0095-1137/82/121052-05$02.00/0 Copyright 1982, American Society for Microbiology Vol. 16, No. 6 Concentrated Gram Stain Smears Prepared with a

More information

School of Allied Medical Sciences. Course Description Guide Associate in Medical Laboratory Sciences. Page 1 of 15

School of Allied Medical Sciences. Course Description Guide Associate in Medical Laboratory Sciences. Page 1 of 15 School of Allied Medical Sciences Course Description Guide Associate in Medical Laboratory Sciences Page 1 of 15 Program Name & Definition: Associate in Medical Laboratory Sciences Laboratory sciences

More information

AMP-6000 AUTOMATED MICROBIOLOGICAL PLATFORM

AMP-6000 AUTOMATED MICROBIOLOGICAL PLATFORM MICROBIOLOGY AMP-6000 AUTOMATED MICROBIOLOGICAL PLATFORM Microbiological Counting System rapid results simple documentation earlier product release reduced workload simple testing rapid process intervention

More information

COMPASS Ecc Agar. Microorganisms Typical phenotype Colony color. GUD + / -gal + GUD - / -gal + GUD - / -gal - 1/5

COMPASS Ecc Agar. Microorganisms Typical phenotype Colony color. GUD + / -gal + GUD - / -gal + GUD - / -gal - 1/5 COMPASS Ecc Agar INTENDED USE COMPASS Ecc Agar is a selective chromogenic medium used for the simultaneous and specific enumeration of all Escherichia coli and coliforms in food products and animal feeding

More information

Activity 5.1.4: Gram Staining

Activity 5.1.4: Gram Staining Activity 5.1.4: Gram Staining Introduction In the last activity, you isolated the bacteria that was responsible for Anna s illness and performed a gross examination of the resultant colonies. While gross

More information

Evaluation of Use of a New Chromogenic Agar in Detection of Urinary Tract Pathogens

Evaluation of Use of a New Chromogenic Agar in Detection of Urinary Tract Pathogens JOURNAL OF CLINICAL MICROBIOLOGY, Apr. 1998, p. 990 994 Vol. 36, No. 4 0095-1137/98/$04.00 0 Copyright 1998, American Society for Microbiology Evaluation of Use of a New Chromogenic Agar in Detection of

More information

Endoscope Microbial Surveillance Testing Made Easy

Endoscope Microbial Surveillance Testing Made Easy Endoscope Microbial Surveillance Testing Made Easy Alpa Patel, B.S.RM (NRCM) Senior Scientist Nelson Laboratories Kaumudi Kulkarni, M.S. Manager of Research and Development Healthmark Industries Disclosure

More information

Automated screening of blood cultures with the

Automated screening of blood cultures with the J Clin Pathol 1988;41:671-675 Automated screening of blood cultures with the Malthus microbiological growth analyser D F J BROWN, M WARNER, C E D TAYLOR, R E WARREN Clinical Microbiology and Public Health

More information

MORE SAMPLES, LESS MICROSCOPE

MORE SAMPLES, LESS MICROSCOPE MORE SAMPLES, LESS MICROSCOPE WE HAVE A SOLUTION TO YOUR NEEDS IN URINALYSIS IRIS: MOVING URINALYSIS FORWARD WE HAVE A SOLUTION TO YOUR NEEDS We can help solve your problems 1. Improved Workflow and Productivity

More information

Meets Standards AFTER REVIEW ( 2.5 pts) Fail to meet the standard safety precautions, counseled one time

Meets Standards AFTER REVIEW ( 2.5 pts) Fail to meet the standard safety precautions, counseled one time Morgan State University 2014 MDTC 411 CLINICAL CHEMISTRY AND URINALYSIS PRACTICUM SUGGESTED LABORATORY ROTATION SCORING RUBRIC Please attach the student s test record to this evaluation Clinical Chemistry

More information

INTRODUCTION water-soluble Figure 1.

INTRODUCTION water-soluble Figure 1. INTRODUCTION Natural waters contain bacteria. The aerobic gram negative bacillus of the genera Psedomonas, Alcalignes, and Flavobacterium are common in natural waters. Many of these bacteria are able to

More information

BD Columbia CNA Agar with 5% Sheep Blood, Improved II

BD Columbia CNA Agar with 5% Sheep Blood, Improved II INSTRUCTIONS FOR USE READY-TO-USE PLATED MEDIA PA-257303.04 Rev.: Sep 2011 BD Columbia CNA Agar with 5% Sheep Blood, Improved II INTENDED USE BD Columbia CNA Agar with 5% Sheep Blood, Improved II is a

More information

Controlled Clinical Comparison of the BacT/ALERT FN and the Standard Anaerobic SN Blood Culture Medium

Controlled Clinical Comparison of the BacT/ALERT FN and the Standard Anaerobic SN Blood Culture Medium JOURNAL OF CLINICAL MICROBIOLOGY, Oct. 2004, p. 4581 4585 Vol. 42, No. 10 0095-1137/04/$08.00 0 DOI: 10.1128/JCM.42.10.4581 4585.2004 Copyright 2004, American Society for Microbiology. All Rights Reserved.

More information

Inoculate: Media. Physical State of Media: Liquid. The Five I s: Basic Techniques to Culture Microbes Tools of the Microbiology Laboratory

Inoculate: Media. Physical State of Media: Liquid. The Five I s: Basic Techniques to Culture Microbes Tools of the Microbiology Laboratory The Five I s: Basic Techniques to Culture Microbes Tools of the Microbiology Laboratory 1. Inoculate 2. Incubate 3. Isolate 4. Inspect 5. Identify The Five I s: Inoculate Inoculate: Media Classified according

More information

Environmental Surveillance FIDSSA Dr Ben Prinsloo Medical Microbiologist

Environmental Surveillance FIDSSA Dr Ben Prinsloo Medical Microbiologist Environmental Surveillance FIDSSA 2015 Dr Ben Prinsloo Medical Microbiologist The Scientist A scientist, in a broad sense, is a person engaging in a systematic activity to acquire knowledge We require

More information

Adapted from Biology 15 Laboratory Manual Supplement: Wrightsman, Ininns and Cannon-Moloznic, Saddleback College, CA 92692

Adapted from Biology 15 Laboratory Manual Supplement: Wrightsman, Ininns and Cannon-Moloznic, Saddleback College, CA 92692 Biology 4B Laboratory Bacteriological Examination of Water Adapted from Biology 15 Laboratory Manual Supplement: Wrightsman, Ininns and Cannon-Moloznic, Saddleback College, CA 92692 Objectives Carry out

More information

University Medical Microbiology

University Medical Microbiology Madison College 20806273 University Medical Microbiology Outline of Instruction Course Information Description Instructional Level Total Credits 5.00 Total Hours 116.00 University Medical Microbiology

More information

JOHN DEMPSEY HOSPITAL Farmington, Connecticut ANTIBIOTIC SUSCEPTIBILITY PROFILES for INPATIENT Bacterial Isolates

JOHN DEMPSEY HOSPITAL Farmington, Connecticut ANTIBIOTIC SUSCEPTIBILITY PROFILES for INPATIENT Bacterial Isolates JOHN DEMPSEY HOSPITAL Farmington, Connecticut 2017 ANTIBIOTIC SUSCEPTIBILITY PROFILES for INPATIENT Bacterial Isolates **GROUPED BY CULTURE SOURCES** (data from 1/1/17 1/1/18) Prepared by: UCHC/JDH Antimicrobial

More information

the Simple, Efficient, Effective

the Simple, Efficient, Effective IVD solutions through partnership R the Simple, Efficient, Effective solution for urine microbiology Technical Information R The microbiological analysis of urine samples for urinary tract infections (UTI)

More information

Automated methods in the Microbiology Lab-Issues and Troubleshooting!

Automated methods in the Microbiology Lab-Issues and Troubleshooting! Automated methods in the Microbiology Lab-Issues and Troubleshooting! Overview Background/history Plate streakers and Gram stainers Blood cultures Automated Susceptibility Testing Automated ID testing

More information

Adaptation of a Bacterial Growth Detection Assay on the VICTOR Nivo Multimode Plate Reader for Measurement of Antibiotic Effects

Adaptation of a Bacterial Growth Detection Assay on the VICTOR Nivo Multimode Plate Reader for Measurement of Antibiotic Effects APPLICATION NOTE Multimode Detection Authors: Maria Kuzikov Dr. Bernhard Ellinger Fraunhofer IME ScreeningPort Hamburg, Germany Adaptation of a Bacterial Growth Detection Assay on the VICTOR Nivo Multimode

More information

Daily minimum Daily minimum. FiO What if the settings were as follows?

Daily minimum Daily minimum. FiO What if the settings were as follows? Ventilator - Associated Event Case Studies Cindy Gross, MT, SM (ASCP), CIC Division of Healthcare Quality Promotion Centers for Disease Control and Prevention October 4, 2012 The following examples are

More information

The way forward for fast microbiology

The way forward for fast microbiology REVIEW 10.1111/j.1469-0691.2011.03520.x The way forward for fast microbiology F. Mulatero, V. Bonnardel and C. Micolaud biomérieux SA Chemin de l Orme, 69280 Marcy l Etoile, France Abstract In microbiology,

More information

Inoculation efficiency is the first step in the bacteriology testing process and is the first opportunity for automation.

Inoculation efficiency is the first step in the bacteriology testing process and is the first opportunity for automation. 0000 Inoculation Inoculation efficiency is the first step in the bacteriology testing process and is the first opportunity for automation. SAMPLES ENABLING TIME OPTIMIZATION UP TO 2HR TRUE WALK-AWAY TIME

More information

Protocol Reference: Verification Protocol

Protocol Reference: Verification Protocol Protocol Reference: Verification Protocol Contents 1. About BioBall Multishot 550 3 2. About This Protocol 4 3. Safety Precautions 4 4. Protocol Pre-approval 4 5. Protocol Scope 5 5.1. Intended Location

More information

Female Genital Tract Bacterial Coisolates With Candida albicans in Patients Without Clinical Vaginitis

Female Genital Tract Bacterial Coisolates With Candida albicans in Patients Without Clinical Vaginitis Infectious Diseases in Obstetrics and Gynecology 6:52-56 (1998) (C) 1998 Wiley-Liss, Inc. Female Genital Tract Bacterial Coisolates With Candida albicans in Patients Without Clinical Vaginitis G.R.G. Monif*

More information

Prepared by Date Adopted Supersedes Procedure # Adapted from HPTN Policy. Review Date Revision Date Signature

Prepared by Date Adopted Supersedes Procedure # Adapted from HPTN Policy. Review Date Revision Date Signature APPENDIX IV: LABORATORY QUALITY CONTROL POLICY Prepared by Date Adopted Supersedes Procedure # Adapted from HPTN Policy N/A Review Date Revision Date Signature Distributed to # of Copies Distributed to

More information

MLT 116L Clinical Microbiology Lab

MLT 116L Clinical Microbiology Lab Page 1 of 5 MLT 116L Clinical Microbiology Lab Approval Date: Effective Term: Department: MEDICAL LABORATORY TECHNICIAN Division: Allied Health/Public Safety Units: 1.00 Grading Option: Letter Grade Transferability:

More information

Isolation and Characterization of Two Antibiotic-Producing Bacteria

Isolation and Characterization of Two Antibiotic-Producing Bacteria Isolation and Characterization of Two Antibiotic-Producing Bacteria Madeline Gibson Abstract The discovery of antibiotics with novel mechanisms has plateaued in the last twenty years. As antibiotics are

More information

Optimized Pathogen Detection with 30- Compared to 20-Milliliter Blood Culture Draws

Optimized Pathogen Detection with 30- Compared to 20-Milliliter Blood Culture Draws JOURNAL OF CLINICAL MICROBIOLOGY, Dec. 2011, p. 4047 4051 Vol. 49, No. 12 0095-1137/11/$12.00 doi:10.1128/jcm.01314-11 Copyright 2011, American Society for Microbiology. All Rights Reserved. Optimized

More information

Malaysian Journal of Microbiology

Malaysian Journal of Microbiology Malaysian Journal of Microbiology Vol 13(1) March 2017, pp. 6-12 Malaysian Journal of Microbiology Published by Malaysian Society for Microbiology (In since 2011) Combination of quantitative bacterial

More information

BD Mac Conkey II Agar / Columbia CNA Agar with 5% Sheep Blood (Biplate)

BD Mac Conkey II Agar / Columbia CNA Agar with 5% Sheep Blood (Biplate) PA-254447.02-1 - INSTRUCTIONS FOR USE READY-TO-USE PLATED MEDIA PA-254447.02 Rev.: Sep 2011 BD Mac Conkey II Agar / Columbia CNA Agar with 5% Sheep Blood (Biplate) INTENDED USE BD Mac Conkey II Agar /

More information

55 Industrial Park Road Boothbay, ME USA

55 Industrial Park Road Boothbay, ME USA Introduction Increasing the level of sanitation and sterility of health-care environments is critical in reducing the transmission of health-care associated infections. The Center for Disease Control estimates

More information

Bacterial load was measured using the method described previously (1). The necrotic tissue

Bacterial load was measured using the method described previously (1). The necrotic tissue Supplemental Methods Bacterial Load Quantification Bacterial load was measured using the method described previously (1). The necrotic tissue was excised, and then 2 mm (width) 2 mm (length) 1-5 mm (depth)

More information

University of Groningen

University of Groningen University of Groningen Fast and accurate prediction of positive and negative urine cultures by flow cytometry Moshaver, Bijan; de Boer, Foppie; van Egmond-Kreileman, Heidi; Kramer, Ellen; Stegeman, Coen;

More information

FAPIC Workshop Point of Care Device Commercialization. Michael G. Lorenz Molzym GmbH & Co. KG Bremen, Germany

FAPIC Workshop Point of Care Device Commercialization. Michael G. Lorenz Molzym GmbH & Co. KG Bremen, Germany FAPIC Workshop Point of Care Device Commercialization Michael G. Lorenz Molzym GmbH & Co. KG Bremen, Germany Development of a fast method for culture independent ID and characterisation of pathogens Automated

More information

Evaluation of Two Bioluminescence-Measuring Instruments, the Turner Design and Lumac Systems, for the Rapid Screening of

Evaluation of Two Bioluminescence-Measuring Instruments, the Turner Design and Lumac Systems, for the Rapid Screening of JOURNAL OF CLINICAL MICROBIOLOGY, Dec. 984, p. 65-70 0095-37/84/265-06$02.00/0 Copyright 984, American Society for Microbiology Vol. 20, No. 6 Evaluation of Two Bioluminescence-Measuring s, the and Systems,

More information

Evaluation of Two Bioluminescence-Measuring Instruments, the Turner Design and Lumac Systems, for the Rapid Screening of

Evaluation of Two Bioluminescence-Measuring Instruments, the Turner Design and Lumac Systems, for the Rapid Screening of JOURNAL OF CLINICAL MICROBIOLOGY, Dec. 984, p. 65-70 0095-37/84/265-06$02.00/0 Copyright 984, American Society for Microbiology Vol. 20, No. 6 Evaluation of Two Bioluminescence-Measuring s, the and Systems,

More information

Evaluation of Two Bioluminescence-Measuring Instruments, the Turner Design and Lumac Systems, for the Rapid Screening of

Evaluation of Two Bioluminescence-Measuring Instruments, the Turner Design and Lumac Systems, for the Rapid Screening of JOURNAL OF CLINICAL MICROBIOLOGY, Dec. 984, p. 65-70 0095-37/84/265-06$02.00/0 Copyright 984, American Society for Microbiology Vol. 20, No. 6 Evaluation of Two Bioluminescence-Measuring s, the and Systems,

More information

GRAM STAIN/CULTURE of CSF and OTHER BODY FLUIDS by HEMATOLOGY LAB (11:30pm 07:30am)

GRAM STAIN/CULTURE of CSF and OTHER BODY FLUIDS by HEMATOLOGY LAB (11:30pm 07:30am) Created Rebecca Riemann Clin. Lab 5/22/2012 Tech 1 Revised Eric Spitzer Chief of 5/22/2012 Micro/Assoc. Dir Labs Reviewed Jay Bock M.D. 5/22/2012 Approved Eric Spitzer Chief of Micro/Assoc. Dir Labs mic

More information

Ten Minute, Reagent-Free identification of Bacteria Containing Resistance Genes Using a Rapid Intrinsic Fluorescence Method

Ten Minute, Reagent-Free identification of Bacteria Containing Resistance Genes Using a Rapid Intrinsic Fluorescence Method 548 Ten Minute, Reagent-Free identification of Bacteria Containing Resistance Genes Using a Rapid Intrinsic Fluorescence Method R. Rozen-Sadowsky 1, A. Shinderman 1, D. Gohman 1, D. Shimonov 1, Y. Gluckman

More information

A Tsunami of Change, Automation in Clinical Microbiology

A Tsunami of Change, Automation in Clinical Microbiology A Tsunami of Change, Automation in Clinical Microbiology Nathan A Ledeboer Associate Professor of Pathology Medical College of Wisconsin Medical Director, Microbiology and Molecular Pathology Interim Co-Director,

More information

Control Issued by: LABORATORY MANAGER Original Date: April 11, 2001 Approved by: Laboratory Director Revision Date: February 27, 2004

Control Issued by: LABORATORY MANAGER Original Date: April 11, 2001 Approved by: Laboratory Director Revision Date: February 27, 2004 Section: Policy # MI\QC\05\v02 Page 1 of 5 Subject Title: Kirby Bauer Quality Control Issued by: LABORATORY MANAGER Original Date: April 11, 2001 Approved by: Laboratory Director Revision Date: February

More information

SAMPLE. Urinalysis; Approved Guideline Third Edition

SAMPLE. Urinalysis; Approved Guideline Third Edition February 2009 Urinalysis; Approved Guideline Third Edition This document addresses procedures for testing urine, including materials and equipment; macroscopic/physical evaluation; chemical analysis; and

More information

Lab Activity #14 - Bacteriological Examination Of Water and Milk (Adapted from Lab manual by Dr. Diehl)

Lab Activity #14 - Bacteriological Examination Of Water and Milk (Adapted from Lab manual by Dr. Diehl) Lab Activity #14 - Bacteriological Examination Of Water and Milk (Adapted from Lab manual by Dr. Diehl) Some of the diseases that humans can contract from drinking polluted water include typhoid, dysentery,

More information

M. Ben-David 1, O. Hammer 1, A.Scinderman 1, Y. Gluckman-Yavo 1, M. Fridman 1, D. Gohman 1, G. Ingber 1 and E. Zahavy 2

M. Ben-David 1, O. Hammer 1, A.Scinderman 1, Y. Gluckman-Yavo 1, M. Fridman 1, D. Gohman 1, G. Ingber 1 and E. Zahavy 2 441 Rapid Gram Negative Antimicrobial Susceptibility Testing Directly from Positive Blood Culture based on a Unique Spectral Intensity Ratio Analysis via Single Fluorescence Membrane Dye Staining and Flow

More information

Project 7: Wound Cultures and Identification

Project 7: Wound Cultures and Identification Project 7: Wound Cultures and Identification Readings: https://labtestsonline.org/understanding/analytes/wound-culture/tab/test Identification of Gram-Positive & Gram-Negative Bacteria Guide to laboratory

More information

Effect of Diuresis on Staphylococcus aureus Kidney

Effect of Diuresis on Staphylococcus aureus Kidney INFECTION AND IMMUNITY, Dec. 1971, p. 74-746 Copyright 1971 American Society for Microbiology Vol. 4, No. 6 Printed in U.S.A. Effect of Diuresis on Staphylococcus aureus Kidney Infections in Mice DOLORES

More information

EVALUATION OF LIQUID-PREP TM ENCAPSULATION METHOD FOR LIQUID-BASED CYTOLOGY: CELL LOSS ESTIMATES DURING PROCESSING ABSTRACT

EVALUATION OF LIQUID-PREP TM ENCAPSULATION METHOD FOR LIQUID-BASED CYTOLOGY: CELL LOSS ESTIMATES DURING PROCESSING ABSTRACT EVALUATION OF LIQUID-PREP TM ENCAPSULATION METHOD FOR LIQUID-BASED CYTOLOGY: CELL LOSS ESTIMATES DURING PROCESSING James Geyer, Ph.D., James Marino, CMIAC, Fort Lauderdale, FL, USA Seacoast Pathology,

More information

Diagnostic Microbiology

Diagnostic Microbiology Diagnostic Microbiology Identification of Microbes Lecture: 1 Out lines What is expected out of this course??? At the end of this course, you will be able to apply Conventional/ Molecular diagnostic methods

More information