AssayMAP Peptide Cleanup Protocol Guide

Size: px
Start display at page:

Download "AssayMAP Peptide Cleanup Protocol Guide"

Transcription

1 AssayMAP Peptide Cleanup Protocol Guide Original Instructions

2 Notices Agilent Technologies, Inc No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws. Manual Part Number G Edition Revision 01, July 2013 Contact Information Agilent Technologies Inc. Automation Solutions 5301 Stevens Creek Blvd. Santa Clara, CA USA Technical Support: (US only) or (rest of the world) service.automation@agilent.com Customer Service: or orders.automation@agilent.com European Service: +44 (0) euroservice.automation@agilent.com Documentation feedback: documentation.automation@agilent.com Web: automation Acknowledgements Microsoft and Windows are either registered trademarks or trademarks of the Microsoft Corporation in the United States and other countries. Warranty The material contained in this document is provided as is, and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control. Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license. Restricted Rights Legend If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as Commercial computer software as defined in DFAR (June 1995), or as a commercial item as defined in FAR 2.101(a) or as Restricted computer software as defined in FAR (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies standard commercial license terms, and non- DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR (c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR (June1987) or DFAR (b)(2) (November 1995), as applicable in any technical data. Safety Notices WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met. CAUTION A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met. AssayMAP Peptide Cleanup Protocol Guide ii

3 Contents Preface... 4 About this guide... 4 Reporting problems... 4 Before you begin... 5 Equipment... 5 Kits and labware requirements... 7 Reagent requirements... 7 Cartridge use and storage guidelines... 8 Preparing your samples... 9 Sample particulates... 9 Sample composition... 9 Sample load volume... 9 Sample plate setup Preparing your reagents Using the Peptide Cleanup Calculator Preparing reagents for the Peptide Cleanup protocol Setting up a Protocol Starting up Installing software and opening the application interface Opening the protocol and app Priming the wash station and initializing the Bravo platform Running a protocol Assay Development guidelines Automation Protocol Steps AssayMAP Peptide Cleanup Protocol Guide 3

4 Preface About this guide This guide is designed for both routine operators and assay developers. This guide does not cover how to set up and maintain the Bravo Platform, detailed safety guidelines, details about the Bravo 96AM Head, details about the Pump Module, or detailed editing of the protocol or application interface in the VWorks software. For information on these topics, go to the Automation Solutions knowledge base at and click the link for PDF user guides. Reporting problems If you find a problem with the AssayMAP Peptide Cleanup solution, contact Automation Solutions Technical Support. For contact information, see Notices on page ii. For reporting problems with... Have the following information ready... Hardware Software Protocol Guide Device serial number from the Bravo serial number label Short description of the problem Relevant software version number (for example, automation control software, diagnostics software, and firmware) Error message Relevant files, such as log files Send your comments in an to documentation@agilent.com AssayMAP Peptide Cleanup Protocol Guide 4

5 Before you begin Introduction The AssayMAP Peptide Cleanup solution enables automated cleanup of 8 to 96 peptide samples in a single run. The solution consists of the following components: Agilent Bravo Platform for AssayMAP (liquid handler) Agilent AssayMAP microchromatography cartridges packed with either a 20 μm diameter silica resin with 150 Å pores derivitized with C18 (C18) or a 20 μm diameter underivitized polystyrene polystyrene hydrophobic reversed-phase resin with 100 Å pores (RP-S). Agilent software and supporting files (VWorks software, Launch Pad, AssayMAP Peptide Cleanup protocol, and application interface) Equipment Table AssayMAP Bravo Platform requirements Component Part number Bravo Platform for AssayMAP technology Includes the base Bravo Platform, computer, and the following: Gripper Upgrade Bravo 96AM Head 96AM Wash Station Pump Module AM Cartridge & Tip Seating Station Custom Plate Nest Risers, 146 mm Peltier Thermal Station with STC controller Orbital Shaking Station with Control Unit Red PCR Plate Insert Two 10-L carboys G5542A G5498B#046 G5498B#057 G5498B#058 G G5498B#017 G5498B#055 G5498B#035 G5498B#033 G5498B#013 NA AssayMAP Peptide Cleanup Protocol Guide 5

6 Figure AssayMAP Bravo Platform and accessories Table Optional Equipment Centrifuge with microplate rotor SpeedVac PlateLoc Software This guide documents the following versions, which are required for the Bravo 96AM Head: VWorks Automation Control Bravo Diagnostics Bravo Firmware AssayMAP Launch Pad v3.0 NOTE You can download the most recent version of the installer from AssayMAP Peptide Cleanup Protocol Guide 6

7 Kits and labware requirements RP-S and C18 reverse phase cartridge types are currently available for performing peptide cleanup on the AssayMAP Bravo. Each of these cartridge types can either be purchased as a rack of 96 cartridges plus the labware required for peptide cleanup and in solution digestion or as an individual rack of 96 cartridges. Table AssayMAP Digestion and Cleanup Starter kits AssayMAP Digestion and RP-S Cleanup Starter Kit AssayMAP Digestion and C18 Cleanup Starter Kit G reversed-phase (RP-S) cartridges plus labware for In- Solution Digestion and Peptide Cleanup G C18 cartridges plus labware for In-Solution Digestion and Peptide Cleanup AssayMAP cartridge racks* G AssayMAP C18 cartridge rack (Qty 96) AssayMAP RP-S cartridge rack (Qty 96) * Labware is included in Starter Kits, but required for In-Solution Digestion A and Peptide Cleanup B when using individual racks mL Deep-Well Plates (AbGene AB-1127) A Well Round Bottom, White Plates (Greiner ) A µl Pipette Tips (Agilent ) A 4 12-Column Low-Profile Reservoirs (Seahorse ) B 2 96-Well Round-Bottom, Clear Plates (Greiner ) B 3 96-Well PCR Plates (Eppendorf ) A,B CAUTION The supplied protocol and app support a specific list of labware only. Using undefined or incorrectly identified labware on the Bravo deck can cause a crash and potential damage to the Bravo 96AM Head. For more details contact technical support. Reagent requirements The simplest deck configuration for the Peptide Cleanup has three solvents on the Bravo deck: 50% ACN:50% H2O:0.1%TFA, 100% H2O:0.1%TFA, and deionized water Table Required solvents to be supplied by customer Solution Quality Estimated Volume* acetonitrile (ACN) 100%, HPLC-grade ~100 ml trifluoroacetic acid (TFA) 100%, HPLC-grade ~0.3 ml purified deionized water (H2O) HPLC-grade, MilliQ, or equivalent ~5 L * Use provided Reagent Calculator for more accurate volumes and suggested overages. AssayMAP Peptide Cleanup Protocol Guide 7

8 Cartridge use and storage guidelines Follow these guidelines to get the best performance from the AssayMAP cartridges: Use only particulate-free samples, from a previous chromatographic run or by filtration (0.22 μm). Use only primed and equilibrated cartridges. We guarantee performance when you use AssayMAP cartridges on the same day you equilibrate them. Do not allow wetted cartridges to dry out. We do not guarantee performance of stored cartridges following equilibration. Store unused cartridges in the original shipping package at room temperature. Cartridges stored in this manner are stable for 6 months from the date of purchase. AssayMAP Peptide Cleanup Protocol Guide 8

9 Preparing your samples Sample particulates Samples must be free of macromolecular particulates, such as large protein aggregates and cellular debris. You may load any sample previously passed through another type of chromatography column or a 0.22 μm filter onto an AssayMAP cartridge. Sample composition What are optimal ph conditions? One of the most important considerations for highly efficient and unbiased binding of peptides to the reversed-phase resin is the ph of the sample, which should be acidic (ph<3) for standard, low-ph peptide cleanup with C18 or RP-S cartridges or basic (> ph 10) with RP-S cartridges. Peptides are amphoteric molecules with a diverse range of physiochemical properties. For some classes of peptides, cleanup under high-ph conditions may help to promote retention (e.g., highly basic peptides) or preserve acid-labile modifications (e.g., histidine phosphorylation). You should acidify samples before loading onto AssayMAP peptide cleanup cartridges, generally by adding TFA, formic acid, or acetic acid. You can use the Single Liquid Addition utility to perform this addition. What sample components cause concerns? Samples containing organic solvents or some types of detergents should be avoided as they may bias the peptides that bind the column. For example, loading samples in a buffer containing greater than 5% acetonitrile will inhibit binding of hydrophilic peptides. Silica-based C18 cartridges are at risk with a sample ph higher than 8. If you have concerns about a specific buffer component, you should examine scientific literature for the known effects of this type of molecule on reversed-phase resins. You can use the Single Liquid Addition utility to modify the composition of your samples with a dilution or ph adjustment, especially after using the In-Solution Digestion workflow. Sample load volume AssayMAP cartridges can accommodate volumes ranging from 10 to 245 μl. What is the binding capacity of the cartridge? For each AssayMAP peptide cleanup cartridge, the quantitative binding capacity depends on the relative hydrophobicity of the peptides in the sample and the sample complexity. When examining human insulin in isolation, the quantitative binding capacity is approximately 400 µg per peptide cleanup cartridge. When examining a tryptic digestion of BSA, the quantitative binding capacity is approximately 150 (without loss of hydrophilic peptides) to approximately 300 µg (with some loss of highly hydrophilic peptides) for this complex mixture. The total binding capacity (supersaturating amounts) is the maximum mass of peptide that can bind to the cartridge and is greater that the quantitative binding capacity. Binding mass can be increased if quantitative binding is not required. What is the concentration of the target in the sample? Use this formula to determine the volume of sample to load: μl sample to load = μg peptide desired μg/μl of peptide in sample AssayMAP Peptide Cleanup Protocol Guide 9

10 Sample plate setup How many samples can I process? You can process 8 to 96 samples in parallel. The AssayMAP Probe Syringes always move in unison. The position of samples in their microplate dictates the positions needed for cartridges in the 96AM Cartridge & Tip Seating Station, and the reservoirs for solutions at the other deck locations. How do I need to configure fewer than 96 samples in a microplate? When you use only a subset of wells in a sample microplate, fill the same subset of positions at all the deck locations. The default settings for the AssayMAP platform assume that samples will be arranged in multiples of 8 in a column-based configuration. The AssayMAP Bravo applies differential pressure to seat cartridges based upon the number of full columns of cartridges. Therefore, to achieve proper cartridge seating entire columns must be used to guarantee optimum performance. The kits supply either an Eppendorf PCR plate or a Greiner 96-Well Round Bottom, clear plate for sample set up. A full list of acceptable labware is available in the dropdown menu of the application interface. Figure Sample plate layout and reservoir recommendations If the number of samples you have is not a multiple of 8, use resin free AssayMAP cartridges (Agilent part number G ) to fill the empty column/row positions. This will prevent liquids from dripping on the deck or being dispensed on the deck during the cup washing step. How much excess sample do I need to load in a microplate well? Pipette an excess of at least 10 µl per well of extra volume for PCR plates, 20 µl for U-bottom and Deep Well plates and at least 3 mls excess per channel of a 12 channel reservoir to ensure that the sample does not fully deplete, which would result in aspiration of air into the cartridges. Modify this volume accordingly if evaporation due to volatile solvents is a concern. Samples and reagent plates should be prepared immediately before running the protocol to minimize evaporation. AssayMAP Peptide Cleanup Protocol Guide 10

11 Preparing your reagents Introduction to the Peptide Cleanup Calculator The Peptide Cleanup Calculator is an Excel worksheet located on the Workflow Navigator that uses formulas to calculate reagent needs and recipe volumes. Based on the number of samples to process, the calculator manages dead volume, pipetting overage, and evaporation concerns for your experiment. The Calculator is comprised of two panels for 1. Calculator Data Entry and 2. Reagent Preparation. Using the Peptide Cleanup Calculator To use the calculator, you enter values into the green cells for the number of samples, for the volumes used in the Equilibrate, Cup Wash, Internal Cartridge Wash, Stringent Syringe Wash, Elute, and Re-Equilibrate steps, and for the number of wash cycles used in the Cup Wash and Stringent Syringe Wash steps. Acceptable range values are shown for each of these items. A formula calculates the number of plate columns used from the number of samples. Make reagents according to the recipes provided in the Reagent Preparation panel of the calculator. AssayMAP Peptide Cleanup Protocol Guide 11

12 Table Calculator Data Entry fields Field Range Default Definition, Instructions & Notes Samples Enter the number of samples you intend to run. Note: The number of samples should be a multiple of 8. Columns of Cartridges used This field is calculated based on the number of samples. This is the number of cartridge columns and reagent columns that will be prepared to manage samples. Equilibrate This field is the volume of utility buffer used to equilibrate the cartridges. Cup Wash This field is the volume of utility buffer used to wash out the cartridge cup and the syringes to ensure that no sample contamination remains in the cup This field is the number of times to perform the cup wash step and its accompanying syringe wash cycle. Internal Cartridge Wash Stringent Syringe Wash This field is the volume of utility buffer used to minimize non-target element contaminants from the internal parts of the cartridge This field is the volume of syringe wash buffer used to comprehensively clean the syringes immediately before elution to ensure there is no carryover This field is the number of times to perform the stringent syringe wash of the bare syringe probes. Elute This field is the volume of elution buffer used to elute peptides from the cartridges. Re- Equilibrate This field is the volume of utility buffer used to displace organic elution buffer and return the cartridge ph to a neutral condition. NOTE The cells where you can enter data provide color cues to identify how the data meets with expectations. Green: is a cell that requires you to enter a value. As long as the cell stays green, the value meets all of the requirements, go. Yellow: The value meets some of the requirements, but use caution. Red: The value does not meet the requirements, stop. AssayMAP Peptide Cleanup Protocol Guide 12

13 Preparing reagents for the Peptide Cleanup protocol The simplest Bravo deck configuration for the Peptide Cleanup protocol uses two solvents in addition to the deionized water used at the wash station. Table Reagent preparation Reagent Composition, Instructions & Notes Syringe Wash Buffer 50% ACN:50% H20:0.1% TFA * Because Stringent Syringe Buffer at deck location 5 is used for both syringe washing and priming, its composition should not be altered without evaluation of how the change could affect the priming step. Elution Buffer 50% ACN:50% H20:0.1% TFA * The typical composition of the Elution Buffer is the same as the Syringe Wash Buffer. Utility Buffer 100% H20:0.1% TFA * Due to the volume change that occurs when mixing organic and aqueous solutions, the actual prepared volumes (and v/v proportions) may not be exactly equal to the values indicated. The Peptide Cleanup calculator is designed to prepare enough extra volume to account for volume changes due to mixing effects. The Calculator takes into consideration that 96-well plates (PCR, round-bottom, and deep-well versions) require about μl of dead volume per well and a 12-channel reservoir plate requires about 3 ml of dead volume per channel. Reagents should be dispensed immediately before running the protocol to avoid evaporation 1 Pipette Syringe Wash Buffer, Elution Buffer, and Utility Buffer into its appropriate reservoir. a b c Pipette the specified volume of Syringe Wash Buffer into each channel being used in the reservoir plate to be placed at deck location 5. Pipette the specified volume of Elution Buffer into each channel being used in the reservoir plate to be placed at deck location 6. Pipette the specified volume of Utility Buffer into each channel being used in the reservoir plate to be placed at deck location 8. 2 If bubbles are present in the reagent plates, it may be necessary to centrifuge to remove the bubbles. NOTE If you are using fewer than 96 cartridges, ensure that you fill the labware columns, rows, or wells to correspond with the samples at deck location 4 and cartridge positions on the 96AM Cartridge & Tip Seating Station. AssayMAP Peptide Cleanup Protocol Guide 13

14 Setting up a Protocol Starting up 1 Ensure that the main power cable and Ethernet or serial cables are plugged into the connection panel. 2 Clear the deck except for the 96AM Wash Station (Location 1) 3 Toggle on ( ) power switch on the side of the AssayMAP Bravo. The green light on the switch illuminates when the AssayMAP Bravo Platform is on. 4 Turn on any accessories, for example, Pump Modules. 5 Turn on the computer and the monitor, and start the Microsoft Windows operating system. Installing software and opening the application interface The AssayMAP Peptide Cleanup protocol and app work with the VWorks software installed on the computer that runs the AssayMAP Bravo Platform. For more information about the VWorks software, see the VWorks Automation Control Setup Guide or contact Automation Solutions Technical Support. To install the Launch Pad, application interfaces, protocols, and utilities: 1 Navigate to the downloaded copy of the installer on your computer, and double-click the file AgilentAssayMAPLaunchPad_build exe 2 Follow the instructions that appear in the installation wizard window. 3 Click Finish. NOTE You can download the most recent version of the installer from NOTE Warning messages may appear asking you how to handle existing labware definitions, device profiles, and liquid classes. Follow the onscreen instructions. Opening the protocol and app To open the protocol and application interface from Launch Pad: 1 Open Launch Pad by double-clicking the AssayMAP Launch Pad shortcut on your desktop or from the Windows Start menu For XP: Programs\Agilent Technologies\AssayMAP Launch Pad For Win7: All Programs\Agilent Technologies\AssayMAP Launch Pad The Launch Pad appears. 2 Click the Peptide Sample Prep tab. AssayMAP Peptide Cleanup Protocol Guide 14

15 Figure Launch Pad web interface showing the Peptide Sample Prep options 3 Click any of the 4 icons (In-Solution Digestion, Peptide Cleanup, Fractionation, or Utilities) to Launch the Peptide Sample Prep: Workflow Navigator. The VWorks software launches and the User Authentication dialog appears. 4 Insert your User name and Password and click OK to login. The Peptide Sample Prep: Workflow Navigator appears inside a VWorks window. 5 All of the necessary tools and protocols are accessible from the Workflow Navigator. AssayMAP Peptide Cleanup Protocol Guide 15

16 Priming the wash station and initializing the Bravo platform Priming ensures that the tubing from the source bottle and the chimneys in the wash station are filled. Use the following procedure to prime the wash station and initialize the platform. For additional information about the Pump Module, tubing connections, and setup of source and waste bottles, see Pump Module User Guide. WARNING When you initialize the Bravo Platform, the robot head and tie bar can move. To prevent potential injury, keep clear of the device while it is in motion. To Initialize the Bravo: Click the Initialize button in the Workflow Navigator. 1 If this is the first action performed by the Bravo platform for the day, an initialization process automatically starts. 2 A series of messages will appear during the initialization process. NOTE A microplate-in-gripper error message appears, regardless of whether the gripper is or is not holding a plate. If the gripper is not holding labware, click Ignore and Continue, leaving device in current state to continue the homing process. If the gripper is holding labware, click Abort to cancel the initialization, remove the labware from the gripper, and restart by clicking Initialize All Devices. NOTE A home the W-axis error message appears during every initialization process to alert you that liquid may be in the syringes if a run was aborted immediately before the current run. If a fluid-in-tips message appears, but no fluid is in the tips, click Retry to continue homing the w-axis. If you suspect that there is fluid in the tips, click Ignore and Continue, leaving device in current state. This ensures that the liquid in the syringes is not dispensed from the Bravo head to a position on the deck. The AssayMAP Bravo will continue the initialization process until the head reaches the wash station. The W-axis is not homed error message shown below will appear. NOTE If you clicked Ignore and Continue, leaving device in current state as a response to the message shown above, the W-axis is not homed message appears: Click Diagnostics to enter the VWorks diagnostic tools, click Home W, Enable Motor, and then OK. This will dispense any liquid in the syringes into the wash station to waste, complete the initialization process as well as the prime and wash procedure. To prime the wash station: 1 Fill the 10 L source bottle for the 96AM Wash Station with purified deionized water, ensure the waste bottle is empty, and verify that the tubing connections are intact. AssayMAP Peptide Cleanup Protocol Guide 16

17 3 Make sure that there is no labware on the deck except for the wash station. 4 Click Prime and Wash. To verify that the Bravo Platform is ready: 1 Carefully observe the top of the chimneys during the Prime and Wash procedure. Fluid should flow out of the top of each chimney and the shape of the fluid dome that comes out the top of each chimney should be consistent. If fluid does not appear to be flowing out of a chimney, a bubble may be blocking the flow. In this case, click Prime and Wash to repeat the procedure. If any chimney has a fluid dome shape that is different than the others, carefully examine the top of the chimney for an irregular shape. Chimney irregularities can result from head crashes into the wash station, may result in large droplets clinging to the syringe probes, and can introduce variability in the assay. 2 Ensure that the indicator lights on the front of the Bravo Platform are blue, and that the robot-disable button on the pendant is activated (red indicator light). NOTE If you change the composition of the solution in the source bottle between runs, ensure that the tubing between it and the chimneys in the wash station fills with purified deionized water. In some instances, this may require disconnecting tubing connections, draining and refilling tubing lines, and reconnecting tubing. Figure Indicator light locations on the Bravo Platform Figure Activating the robot-disable button NOTE You only need to activate the robot-disable button after it has been used to disable a run. AssayMAP Peptide Cleanup Protocol Guide 17

18 Running a protocol 1 Prepare sample plates. a b c d Use the Single Liquid Addition Utility on the Workflow Navigator to adjust the buffer composition of the samples for reversed phase cleanup (i.e. change ph or reduce organic solvent concentration). (Optional) Use the Sample Plate Transfer Utility on the Workflow Navigator to move samples into one of the labware options available in the dropdown list for Deck Location 4 of the Labware Table. A µl sample excess in each well is recommended for optimal sample loading onto cartridges. Samples should be prepared immediately before running the protocol to avoid evaporation. Ensure the samples do not contain bubbles. (optional) Centrifuge to remove bubbles. 2 Use the Cartridge Transfer Utility to move cartridges to the Cartridge and Tip Seating Station. Click Cartridge Transfer in the Peptide Cleanup Banner of the Workflow Navigator and click Restore Defaults. Fill in Application Settings. You will use the Number of Columns of Cartridges from the Calculator as the Number of Columns to Transfer in the Application Settings. a b c Confirm that the physical Deck Layout exactly matches the Deck Layout diagram (B) and Labware Table (C) section of the app and that that each piece of labware seats properly on its plate pad. See the Utilities Guide in the Workflow Navigator for more detailed instructions on Cartridge Transfer Utility Click Run Cartridge Transfer. The head transfers cartridges from the Source: Cartridge Rack to the Destination: Seating Station at deck location 2. Click OK when the Protocol Complete message appears. d Remove unused cartridges from Deck Position 6. e Click Return to Navigator and open the Peptide Cleanup Protocol 3 Set up the Peptide Cleanup protocol. a (optional) Click Reset Defaults to set all Application Settings to their defaults. AssayMAP Peptide Cleanup Protocol Guide 18

19 b Enter the number of columns, choose the steps to perform ( )or skip ( ), and enter setting values. The Number of Full Columns of Cartridges must match the value from the Peptide Cleanup Calculator for Columns of Cartridges used Application Settings Number of Full Columns of Cartridges Steps Notes Set up samples in columns; cartridge layout and reagent plate layouts must match this configuration. Use value for Columns of Cartridges used from the Peptide Cleanup Calculator. Conduct Step? Default Value (Range) 1 (1-12) Volume (µl) Flow Rate (µl/min) Wash Cycles Initial Syringe Rinses syringes with deionized water in the specified number of cycles. 5 (1-10) Wash Prime Prime the cartridges with syringe wash buffer. Equilibrate Defines the volume of utility buffer used to equilibrate the cartridges. 50 (0-250) 25 ( ) 0 (0-10) Load Samples Defines the volume of sample, the flow rate used to load a sample and the 50 (0-245) 5 ( ) 5 (0-10) number of syringe wash cycles. Ensure that each sample well has an excess of at least µl to ensure that air does not get introduced into the cartridge. A 5-µL sample chase with utility buffer occurs after sample loading. Cup Wash Defines the volume of utility buffer used to wash out each cartridge cup 50 (1-100) 2 (0-10) and the number of times to perform the cup wash/syringe wash cycle. Internal Cartridge Wash Defines the volume of utility buffer, the flow rate used to wash non-target elements from the cartridge resin, and the number of syringe wash cycles. 50 (0-250) 25 ( ) 3 (0-10) Stringent Syringe Wash Elute Re-Equilibrate Final Syringe Wash Defines the volume of syringe wash buffer and number of wash cycles used to wash each syringe immediately before elution. 50 (1-250) 1 (1-10) Defines the volume of elution buffer, the flow rate used to elute peptides 25 (0-250) 5 ( ) 0 (1-10) from the cartridges and the number of syringe wash cycles. Defines the volume of utility buffer and the flow rate used to displace 50 (0-250) 25 ( ) elution buffer from the cartridge. Rinses syringes with deionized water in the specified number of cycles. 3 (0-10) c Place Sample Plate at deck location 4. d Place the other reagent plates at their indicated positions AssayMAP Peptide Cleanup Protocol Guide 19

20 e Confirm that the physical Deck Layout exactly matches the Deck Layout diagram (B) and Labware Table (C) section of the app and that that each piece of labware seats properly on its plate pad. To prevent head crashes, ALWAYS use the specified labware at its designated location and ensure that labware seats properly on its plate pad 4 Run the Peptide Cleanup protocol. Click Run and follow the progress of your experiment in the Control Panel status bar. When you see the Protocol Complete message appear, remove each Sample Plate from the deck. NOTE When the protocol starts, you can walk away from the Bravo Platform for the duration of the protocol. The default protocol with 50 μl samples and 5 μl/min flow rate should take about 35 minutes to complete. 5 Analyze results Pausing or aborting the run While conducting a protocol run or simulation, click Pause. The protocol pauses and the VWorks Scheduler Paused dialog box opens. NOTE While the Schedule Paused dialog box will appear, the protocol may not pause immediately. When you click Pause, the platform will complete the task currently running, which could take minutes to an hour to finish, depending on the volume and flow rate of the task. Generally, Load Samples is the longest step in the protocol. You will know that the pause starts when the head lifts up above the plate and stops. In the Scheduler Paused dialog box, click one of the following: Button Continue Description Resumes the protocol run when you are ready to continue. AssayMAP Peptide Cleanup Protocol Guide 20

21 Abort process Aborts the run. Select Abort process if you have determined that the protocol run is not recoverable. For details on the other options, see the VWorks Automation Control User Guide. Cleaning up at the end of the day Because the Peptide Cleanup protocol uses purified, deionized water in the source bottle, there are no special cleanup procedures for the liquid handling platform other than disposal of liquids from the Organic Waste plate. CAUTION Discard waste according to waste disposal procedures and in compliance with all applicable regulations and personal safety protection practices. Shutting down 1 Remove the labware from the deck. 2 Discard solutions and used labware following appropriate waste disposal procedures. 3 (optional) Use the Syringe Wash utility to perform a stringent syringe wash to minimize the risk of sample carryover into the next protocol. 4 Empty the Waste Bottle. 5 Close VWorks and, shut down the computer. 6 Turn off any accessories (Pump Modules and temperature controller). 7 If using an Auto Filling Reservoir, disconnect the bottles to prevent siphoning. 8 Toggle the power switch to the off (o) position. AssayMAP Peptide Cleanup Protocol Guide 21

22 Assay Development guidelines Setup and equilibration The first stage in the Peptide Cleanup protocol is to set up the system and prepare the cartridges for sample loading. This involves washing the syringes in the Bravo 96AM head to ensure there is no carryover from previous experiments, priming the cartridges to condition the resin and remove entrapped air, and equilibrating the cartridges to create optimal conditions for binding. Cartridges ship dry and therefore contain air entrapped in the cartridge bed. Failure to prime cartridges can prevent access of the sample and buffers to parts of the bed, resulting in reduced capacity and poor reproducibility. Initial Syringe Wash The Initial Syringe Wash step uses the wash station to flush any potential contaminants from the syringes in the head. During each Initial Syringe Wash cycle, the head aspirates 250 μl into the syringes from the wash station chimneys and then moves by a fixed offset between the chimneys to dispense to waste. Select the Initial Syringe Wash step to clean the syringes. Skip this option if you are certain that the syringes are clean. Setting the Initial Syringe Wash step parameters Parameter Value Notes Wash cycles 1 10 Default: 5 Increase the number of wash cycles to better clean the syringes. Note that increasing the number of cycles increases the time of the assay and consumption of reagents, and affects wear on the syringes. Prime Cartridges The syringes fill with sufficient syringe wash buffer (100 μl) to perform the Prime step. After mounting the cartridges on the probes, the head moves to the Organic Waste plate to prime the cartridges by dispensing 100 μl at 300 μl/min through the cartridges to waste. The last task of the priming step ejects the cartridges into the seating station. The priming solution (syringe wash buffer) requires the solution contain at least 25% organic solvent. Higher levels of organic are also acceptable and maybe preferred if large samples are to be loaded onto the cartridge to avoid changes in the solution s composition while it is sitting on the deck. When to select this option Always perform the Prime step at the beginning of a run. In the default protocol, this step is selected. AssayMAP Peptide Cleanup Protocol Guide 22

23 Equilibrate Cartridges The syringes aspirate the specified volume of utility buffer and then dispense at the specified flow rate through the cartridges into the wash station. The wash station washes the outside of the cartridges. The head moves to the seating station and ejects the cartridges. The head moves the syringes to the wash station at deck location 1 and washes the syringes with 250 µl of H2O for the specified number of cycles. When to select this option This step exchanges the buffer used for priming with utility buffer within the resin bed. It is critical to equilibrate the cartridge bed with a buffer that establishes optimal chemical conditions for sample loading. In the default protocol, this step is selected. Setting the Equilibration step parameters Parameter Value Notes Volume Flow rate μl Default: 50 μl μl/min Default: 25 μl/min The default volume for this step is equal to 10 column volumes, which should be sufficient for complete buffer exchange. A rate slower than 25 μl/min will likely have no benefit, but will increase the total assay time. A rate faster than 25 μl/min might not equilibrate through the pores in the beads across the full length of the cartridge bed. Wash cycles 0 10 cycles Default: 0 This allows you to control the number of syringe washes after the equilibration aspiration step by setting the value up to 10 wash cycles. The default setting is 0 cycle. Load Samples During Load Samples, the AssayMAP Bravo mounts the peptide cleanup cartridges onto the syringes, aspirates the specified volume of samples through the cartridges at the specified flow rate, and performs an external cartridge tip wash at the wash station, where cartridge tips dip into the wash station chimneys to remove any sample on the outside of the cartridge. AssayMAP cartridges can accommodate volumes ranging from 10 to 245 μl. After sample loading, a 5 µl chase volume of Utility Buffer will be automatically aspirated through the cartridges to ensure that the entire sample volume has been fully exposed to the entire resin bed. Following sample loading and chase the cartridges ejected into the tip seating station and then the are the flow through plus chase are collected into the Flow through plate (deck location#7) What is the binding capacity of the cartridge? For each AssayMAP peptide cleanup cartridge, the quantitative binding capacity depends on the relative hydrophobicity of the peptides in the sample, the sample complexity and the chemistry. As an example, the choice of ion pair reagents will affect binding characteristics. When examining human insulin in isolation, the quantitative binding capacity is approximately 400 µg per peptide cleanup cartridge. AssayMAP Peptide Cleanup Protocol Guide 23

24 When examining a tryptic digestion of BSA, the quantitative binding capacity is approximately 150 (without loss of hydrophilic peptides) to approximately 300 µg (with some loss of highly hydrophilic peptides) for this complex mixture. The total binding capacity (supersaturating amounts) is the maximum mass of peptide that can bind to the cartridge. If quantitative binding is not required, greater mass can be loaded. What is the concentration of the target in the sample? Use this formula to determine the volume of sample to load: μl sample to load = μg peptide desired μg/μl of peptide in sample When to select this option Always select the Load Samples step during a normal run. In the default protocol, this step is selected. Setting the Load Samples step parameters Parameter Value Notes Volume μl Default: 50 The volume depends on the concentration and the mass of sample to capture. It is not possible to aspirate all of the liquid from a microplate well. Add at least μl extra sample liquid to each well to ensure that the sample does not deplete fully, which would result in aspiration of air into the cartridges. Large sample volume or slow load rates may require slightly more overage. Flow rate μl/min Default: 5 μl/min The value depends on a balance between the speed of the assay and desired recovery. Binding capacity is inversely proportional to flow rate. For many resins, a flow rate of 20 μl/min results in a low dynamic binding capacity. It also produces some loss of total binding capacity, as the sample does not have sufficient time to diffuse throughout the pores in the cartridge resin bed. As the rate increases above approximately 20 μl/min, the total binding capacity continues to decrease until you hit a failure point when cavitation occurs in the cartridge, which happens only in aspiration mode. At this point, the liquid no longer draws through the cartridge. Generally, this failure occurs at approximately μl/min. If the flow rate is too slow, the protocol can run at non-optimal speed without significant additional binding. Wash cycles 0 10 Default: 5 Generally, the syringes should be washed as part of the Load Sample step. Under certain circumstances, you can skip this wash step to minimize run time. AssayMAP Peptide Cleanup Protocol Guide 24

25 Cup Wash The Cup Wash addresses the small amount of sample liquid that may remain above the upper cartridge membrane after sample loading. This step removes this residual liquid by filling the cartridge cups with utility buffer and then removing it along with the residual sample solution. You control the number of Cup Wash cycles by changing the number of wash cycles in the app. Each Cup Wash is followed by a syringe wash at the wash station. When to select this option Select this step if the presence of unbound sample liquid in the cartridge cups could cause complications with analysis. Internal Cartridge Wash The AssayMAP Bravo aspirates utility buffer through the cartridges, removes the cartridges from the probes, and dispenses the contents of the syringes in the wash station. The head then washes syringes at the wash station with 250 µl of deionized water. When to select this option Use Internal Cartridge Wash to minimize non-target elements from the sample matrix in the cartridge, which could possibly contaminate the elution. In the default protocol, this step is selected. Setting the Internal Cartridge Wash step parameters Parameter Value Notes Volume μl Default: 50 μl This allows the user to define the volume of utility buffer to aspirate through the cartridges during the internal cartridge washing process. Flow rate μl/min Default: 25 μl/min A rate slower than 25 μl/min will likely have no benefit, but will increase the total assay time. A rate faster than 25 μl/min might not equilibrate through the pores in the beads across the full length of the cartridge bed, which leads to incomplete washing. Wash cycles 0 10 Default: 3 This parameter specifies the number of syringe wash cycles performed after an Internal Cartridge Wash step at the wash station. Stringent Syringe Wash This syringe wash step comprehensively cleans the syringes with the syringe wash buffer to ensure there is no carryover. Use the Wash Cycles setting for this step to control the number of times this step repeats. When to select this option Use a Stringent Syringe Wash when rigorous cleaning of the syringes is necessary to prevent carryover. In the default protocol, this step is selected. AssayMAP Peptide Cleanup Protocol Guide 25

26 Setting the Stringent Syringe Wash step parameters Parameter Value Notes Volume μl Default: 50 μl This allows you to define the volume to aspirate from the Stringent Syringe Wash and dispense through the syringes to Organic Waste. Wash cycles 1 10 Default: 1 This parameter specifies the number of syringe wash cycles performed with elution buffer. Elute The AssayMAP Bravo aspirates elution buffer using the bare probes. After mounting the cartridges, the head dispenses elution buffer through the mounted cartridges into the Eluate Collection plate. Under optional elution conditions, peptides fully elute 2 column volumes (10 ul). When to select this option Always select Elute to remove bound peptide from cartridges in a normal run. In the default protocol, this step is selected. Setting the Elute step parameters Parameter Value Notes Volume μl Default: 25 μl This allows you to define the volume of elution buffer aspirated into the syringe from deck location 6 and dispensed through the cartridges for Eluate Collection at deck location 9. Flow Rate μl/min Default: 5 μl/min A flow rate slower than 10 µl/min is unlikely to improve the elution yield. There is very little difference in yield using a flow rate from 2 to 20 µl/min Wash cycles 1 10 Default: 0 This parameter specifies the number of syringe wash cycles performed with elution buffer. Final Syringe Wash The Final Syringe Wash uses deionized water from the wash station to flush any potential contaminants from the syringes. The head performs these washes on the wash station. The AssayMAP Bravo aspirates 250 μl of fresh deionized water into the syringes using the bare probes and dispenses at an offset between the chimneys on the wash station. When to select this option Select Final Syringe Wash after each run to clean the syringes thoroughly before running a different assay. In the default protocol, this step is selected. AssayMAP Peptide Cleanup Protocol Guide 26

27 Setting the Final Syringe Wash step parameters Parameter Value Notes Wash cycles 1 10 Default: 3 Increase the number of wash cycles to better clean the syringes. Note that increasing the cycle number increases the duration of the assay and consumption of reagents, and affects wear on the syringes. NOTE The amount of carryover in the syringe (expressed as dilution of the starting sample) depends on the total number of syringe wash cycles. Five cycles give a dilution of up to In cases where carryover is a major concern, increasing the number of iterations also increases the dilution factor, but with a cost of increased assay time and reduced syringe lifetime. AssayMAP Peptide Cleanup Protocol Guide 27

28 Automation Protocol Steps This section describes the basic movements of the automation when a default protocol is selected. Changing the selections or deselecting options will alter the movement. Head moves to deck position: Action: Cartridge Transfer Protocol 6 Load the desired number of cartridges 2 Park cartridges on the Cartridge & Tip Seating Station and adjust their orientation Peptide Sample Cleanup Protocol 2 Head check Initial Syringe Wash 1 Initial Syringe Wash, repeat for the number of cycles Prime 5 Aspirate Syringe Wash for Prime 2 Load cartridges 3 Dispense Syringe Wash through cartridges into Organic Waste for priming step 2 Park cartridges 1 Wash Syringes Equilibrate 8 Aspirate Utility Buffer for equilibration 2 Load cartridges 1 Dispense Utility buffer through cartridges to equilibrate 2 Park Cartridge 1 Wash syringes, repeat for the number of cycles (default is 0) Load Samples 2 Load cartridges 4 Aspirate samples to load onto cartridges 1 External Cartridge Wash 8 Aspirate Utility buffer for sample chase 2 Park cartridges 7 Dispense flow through plus chase into flow through collection plate 1 Wash syringes, repeat for the number of cycles Cup Wash 8 Aspirate Utility buffer 2 Perform Cup Wash 1 Wash syringes, repeat for the number of cycles Internal Cartridge Wash 2 Load cartridges 8 Aspirate Utility buffer into cartridges for Internal Cartridge Wash 1 Dispense Internal Cartridge Wash 2 Park cartridges 1 Wash syringes, repeat for the number of cycles Stringent Syringe Wash 5 Aspirate Syringe Wash Buffer 3 Dispense into Organic Waste, repeat for the number of cycles 1 Wash syringes, repeat for the number of cycles AssayMAP Peptide Cleanup Protocol Guide 28

29 Elute 6 Aspirate Elution Buffer 1 Wash Syringe tips 2 Load cartridges 9 Elute Samples into Elution Collection Plate 1 Wash external cartridge 2 Park Cartridges Re-Equilibration 8 Aspirate Utility Buffer for re-equilibration 2 Load cartridges 1 Dispense Utility buffer through cartridges into wash station 2 Park cartridges Final Syringe Wash 1 Final Syringe Wash syringes, repeat for the number of cycles AssayMAP Peptide Cleanup Protocol Guide 29

Automated Plate Replication

Automated Plate Replication Automated Plate Replication Protocol Guide For Research Use Only. Not for use in diagnostic procedures. Original Instructions Notices Agilent Technologies, Inc. 2017 No part of this manual may be reproduced

More information

Normalization, Dilution and Transfer Methods for the Maxprep TM Liquid Handler Technical Manual

Normalization, Dilution and Transfer Methods for the Maxprep TM Liquid Handler Technical Manual Normalization, Dilution and Transfer Methods for the Maxprep TM Liquid Handler Technical Manual Instructions for use of Products AS9100, AS9101, AS9200, AS9201 TM530 11/17 Table of Contents 1 Description...2

More information

Agilent RNA 6000 Nano Kit Guide

Agilent RNA 6000 Nano Kit Guide Agilent RNA 6000 Nano Kit Guide Agilent Technologies Notices Agilent Technologies, Inc. 2000-2015, 2016 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime EASY-Spray Columns Guidance for column set up and installation Tips to maximize column lifetime EASY-Spray Column Tips and Tricks This document provides guidance for Thermo Scientific EASY-Spray column

More information

foodproof Magnetic Preparation Kit I

foodproof Magnetic Preparation Kit I For food testing purposes FOR IN VITRO USE ONLY foodproof Magnetic Preparation Kit I Version 1, August 2012 Application manual for the automated isolation of Gram-negative bacterial DNA from enrichment

More information

foodproof Magnetic Preparation Kit II

foodproof Magnetic Preparation Kit II For food testing purposes FOR IN VITRO USE ONLY foodproof Magnetic Preparation Kit II Version 3, November 2016 Application manual for the automated isolation of Gram-positive bacterial DNA from enrichment

More information

DRAFT. Apollo 324 System. Protocol. PrepX-32i DNA Library

DRAFT. Apollo 324 System. Protocol. PrepX-32i DNA Library Apollo 324 System PrepX-32i DNA Library Protocol Copyright 2012, IntegenX Inc. All rights reserved. Information in this document is subject to change without notice. IntegenX assumes no responsibility

More information

Applied Biosystems 7900HT Fast Real-Time PCR System. Performing Fast Gene Quantitation with 384-Well Plates DRAFT

Applied Biosystems 7900HT Fast Real-Time PCR System. Performing Fast Gene Quantitation with 384-Well Plates DRAFT User Bulletin Applied Biosystems 7900HT Fast Real-Time PCR System October 3, 2005 SUBJECT: Performing Fast Gene Quantitation with 384-Well Plates In This User Bulletin System Overview System Requirements

More information

Identity Automation DNA IQ System Protocol for the Tecan Freedom EVO Workstation Automated Protocol #EP039

Identity Automation DNA IQ System Protocol for the Tecan Freedom EVO Workstation Automated Protocol #EP039 Identity Automation DNA IQ System Protocol for the Tecan Freedom EVO Workstation Automated Protocol #EP039 DESCRIPTION OF THE DNA IQ SYSTEM METHOD FOR THE TECAN FREEDOM EVO WORKSTATION. All technical literature

More information

QIAGEN GeneRead QIAact Panel Cleanup Kit Handbook

QIAGEN GeneRead QIAact Panel Cleanup Kit Handbook QIAGEN GeneRead QIAact Panel Cleanup Kit Handbook June 2017 For cleanup of DNA during construction of targeted, molecularly bar-coded libraries for digital sequencing with next-generation sequencing (NGS)

More information

Library Loading Bead Kit (EXP-LLB001) Agencourt AMPure XP beads Vortex mixer. Freshly prepared 70% ethanol in nucleasefree

Library Loading Bead Kit (EXP-LLB001) Agencourt AMPure XP beads Vortex mixer. Freshly prepared 70% ethanol in nucleasefree Before start checklist Materials Consumables Equipment PCR Barcoding Kit (EXP-PBC001) NEBNext End repair / da-tailing Module (E7546) Thermal cycler at 20 C and 65 C Ligation Sequencing Kit 1D (SQK-LSK108)

More information

E.Z.N.A. mirna Kit. R preps R preps R preps

E.Z.N.A. mirna Kit. R preps R preps R preps E.Z.N.A. mirna Kit R7034-00 5 preps R7034-01 50 preps R7034-02 200 preps August 2011 E.Z.N.A. Micro RNA Kit Table of Contents Introduction and Overview...2 Kit Contents/Storage and Stability...3 Preparing

More information

Automated Protocol for GenElute HP 96 Well Plasmid Miniprep Kit Using the Biomek FX Workstation (Beckman Coulter)

Automated Protocol for GenElute HP 96 Well Plasmid Miniprep Kit Using the Biomek FX Workstation (Beckman Coulter) Automated Protocol for GenElute HP 96 Well Plasmid Miniprep Kit Using the Biomek FX Workstation (Beckman Coulter) Catalog Number NA9604 Automation Guide 2 I. Description 2 II. Product Components 2 III.

More information

Ligation Sequencing Kit 1D (SQK-LSK108) Ice bucket with ice NEBNext End repair / da-tailing Module (E7546)

Ligation Sequencing Kit 1D (SQK-LSK108) Ice bucket with ice NEBNext End repair / da-tailing Module (E7546) Before start checklist Ligation Sequencing Kit 1D (SQK-LSK108) Ice bucket with ice NEBNext End repair / da-tailing Module (E7546) Library Loading Bead Kit (EXP-LLB001) Timer NEB Blunt/TA Ligase Master

More information

Protein-Pak Hi Res HIC Column and HIC Protein Standard

Protein-Pak Hi Res HIC Column and HIC Protein Standard Protein-Pak Hi Res HIC Column and HIC Protein Standard CONTENTS I. INTRODUCTION II. a. Mobile Phase b. Flow Direction CONNECTING COLUMN TO LC SYSTEM I. INTRODUCTION This offering contains non-porous, polymethacrylate-based

More information

HighPrep Blood & Tissue DNA Kit

HighPrep Blood & Tissue DNA Kit MAGBIO ACCELERATING genomic research HighPrep Blood & Tissue DNA Kit Manual Revision v1.1 Catalog Nos. HPBTS-D16, HPBTS-D96, HPBTS-D96X4 Genomic DNA isolation from 20-250 μl of blood, lysate of tissues,

More information

This User Bulletin does not apply to PROCISE clc instruments.

This User Bulletin does not apply to PROCISE clc instruments. User Bulletin PROCISE Protein Sequencing System (491/492/494) 1 July 1997 (updated 06/2001) SUBJECT: PROCISE Feature Upgrade Kit Summary This User Bulletin describes the hardware, firmware, and software

More information

VERSANT kpcr Molecular System

VERSANT kpcr Molecular System VERSANT kpcr Molecular System Application Guide Addendum Running Dynamic Assay Protocols Using the VERSANT kpcr Molecular System SMN: 10814065 Rev. C Copyrights No part of this work covered by the copyrights

More information

Protocol. Path-ID Multiplex One-Step RT-PCR Kit Protocol TaqMan probe-based multiplex one-step real-time RT-PCR detection of RNA targets

Protocol. Path-ID Multiplex One-Step RT-PCR Kit Protocol TaqMan probe-based multiplex one-step real-time RT-PCR detection of RNA targets Protocol TaqMan probe-based multiplex one-step real-time RT-PCR detection of RNA targets For Research Use Only. Not for use in diagnostic procedures. Information in this document is subject to change without

More information

Maxwell RSC Whole Blood DNA Kit

Maxwell RSC Whole Blood DNA Kit TECHNICAL MANUAL Maxwell RSC Whole Blood DNA Kit Instructions for Use of Product AS1520 Note: To use the Maxwell RSC Whole Blood DNA Kit, you must have the Whole Blood DNA method loaded on the Maxwell

More information

Applied Biosystems TaqMan Array Microfluidic Cards

Applied Biosystems TaqMan Array Microfluidic Cards Performing Rapid Cycling Gene Quantitation on TaqMan Array Microfluidic Cards Applied Biosystems TaqMan Array Microfluidic Cards User Bulletin JULY 2010 SUBJECT: In this user bulletin Performing Rapid

More information

Agilent Protein 230. Kit Guide. Agilent Technologies

Agilent Protein 230. Kit Guide. Agilent Technologies Agilent Protein 230 Kit Guide Agilent Technologies Notices Agilent Technologies, Inc. 2000-2015, 2016 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

HT RNA Pico Sensitivity LabChip Kit LabChip GX/GXII User Guide

HT RNA Pico Sensitivity LabChip Kit LabChip GX/GXII User Guide Contents SPECIFICATIONS... 2 SAMPLE CONDITIONS... 2 KIT CONTENTS... 3 SAFETY WARNINGS AND PRECAUTIONS... 4 PREPARATION PROCEDURES... 5 ADDITIONAL ITEMS REQUIRED... 5 PREPARING LADDER ALIQUOTS... 5 PREPARING

More information

Application Note. Authors. Abstract. Clinical Research

Application Note. Authors. Abstract. Clinical Research Workflow Automation for LC/MS: In-Solution Protein Digestion, Peptide Cleanup, and Strong Cation-Exchange Fractionation of Peptides Enabled by AssayMAP Technology Application Note Clinical Research Authors

More information

1D^2 Sequencing Kit (SQK-LSK308) Pipettes P2, P10, P20, P100, P200, P1000 Freshly prepared 70% ethanol in nucleasefree

1D^2 Sequencing Kit (SQK-LSK308) Pipettes P2, P10, P20, P100, P200, P1000 Freshly prepared 70% ethanol in nucleasefree Before start checklist Library Loading Bead Kit (EXP-LLB001) Heating block at 37 C capable of taking 1.5 ml tubes Agencourt AMPure XP beads 1D^2 Sequencing Kit (SQK-LSK308) Pipettes P2, P10, P20, P100,

More information

E.Z.N.A. PF Micro RNA Kit. R preps R preps

E.Z.N.A. PF Micro RNA Kit. R preps R preps E.Z.N.A. PF Micro RNA Kit R7036-00 5 preps R7036-01 50 preps June 2013 E.Z.N.A. PF Micro RNA Kit Table of Contents Introduction and Overview...2 Kit Contents/Storage and Stability...3 Preparing Reagents...4

More information

User Guide BLF-7510 BUF High-Pass TM DNA Size Selection. Sage Science Part Nos:

User Guide BLF-7510 BUF High-Pass TM DNA Size Selection. Sage Science Part Nos: TM User Guide High-Pass TM DNA Size Selection Sage Science Part Nos: BLF-7510 BUF-7510 Sage Science, Inc. Suite 2400 500 Cummings Center Beverly, MA 01915 Using this Guide This guide is meant to provide

More information

HT RNA LabChip Kit, Version 2 LabChip GX/GXII User Guide

HT RNA LabChip Kit, Version 2 LabChip GX/GXII User Guide Contents Specifications 2 Sample Conditions 2 Kit Contents 2 Preparation Procedure 4 Additional Items Required 4 Preparing the Gel-Dye Solution 4 Preparing the RNA Samples and Ladder 4 Preparing the Buffer

More information

AMPURE PCR PURIFICATION PAGE 1 OF 7

AMPURE PCR PURIFICATION PAGE 1 OF 7 PCR PURIFICATION PAGE 1 OF 7 Please refer to http://www.agencourt.com/technical/reagent_information/ for updated protocols. AMPure is a registered trademark of Agencourt Bioscience and is for laboratory

More information

Procedure for DNA Quantitation with Quantifiler Duo

Procedure for DNA Quantitation with Quantifiler Duo Procedure for DNA Quantitation with Quantifiler Duo 1.0 Purpose This procedure specifies the steps for performing quantification of DNA extracts with the Quantifiler Duo kit on the ABI 7500 instruments.

More information

E.Z.N.A. Water DNA Kit. D preps D preps D preps

E.Z.N.A. Water DNA Kit. D preps D preps D preps E.Z.N.A. Water DNA Kit D5525-00 5 preps D5525-01 50 preps D5525-02 200 preps April 2017 E.Z.N.A. Water DNA Kit Table of Contents Introduction and Overview...2 Kit Contents/Storage and Stability...3 Preparing

More information

E.Z.N.A. Stool DNA Kit. D preps D preps D preps

E.Z.N.A. Stool DNA Kit. D preps D preps D preps E.Z.N.A. Stool DNA Kit D4015-00 5 preps D4015-01 50 preps D4015-02 200 preps April 2013 E.Z.N.A. Stool DNA Kit Table of Contents Introduction and Overview...2 Illustrated Protocol...3 Kit Contents/Storage

More information

Kit Specifications 45 g. 45 g of RNA 8 L

Kit Specifications 45 g. 45 g of RNA 8 L RNA Clean-Up and Concentration Micro-Elute Kit Product # 61000 Product Insert Norgen s RNA Clean-Up and Concentration Micro-Elution Kit provides a rapid method for the purification, cleanup and concentration

More information

RNA Clean-Up and Concentration Kit Product # 23600, 43200

RNA Clean-Up and Concentration Kit Product # 23600, 43200 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com RNA Clean-Up and Concentration Kit Product # 23600, 43200 Product

More information

E.Z.N.A. Blood DNA Mini Kit. D preps D preps D preps

E.Z.N.A. Blood DNA Mini Kit. D preps D preps D preps E.Z.N.A. Blood DNA Mini Kit D3392-00 5 preps D3392-01 50 preps D3392-02 200 preps January 2017 E.Z.N.A. Blood DNA Mini Kit Table of Contents Introduction and Overview...2 Illustrated Protocol...3 Kit Contents/Storage

More information

AU680 Chemistry Analyzer In-Lab Training Manual

AU680 Chemistry Analyzer In-Lab Training Manual AU680 Chemistry Analyzer In-Lab Training Manual AU680 In-Lab Training Manual Page 1 of 152 AU680 In-Lab Training Manual Page 2 of 152 AU680 Chemistry Analyzer WARNINGS AND PRECAUTIONS Read all product

More information

Guide to Trap Cartridge Care and Use

Guide to Trap Cartridge Care and Use Michrom BioResources, Inc. offers a wide range of innovative products for HPLC and LC/MS analysis of biological and pharmaceutical samples. Products include: HPLC instrumentation, MS instrumentation, LC/MS

More information

Gadus morhua Cod speciation kit. genesig Easy Kit. for use on the genesig q reactions. Primerdesign Ltd

Gadus morhua Cod speciation kit. genesig Easy Kit. for use on the genesig q reactions. Primerdesign Ltd TM Primerdesign Ltd Cod speciation kit genesig Easy Kit for use on the genesig q16 50 reactions For general laboratory and research use only 1 genesig Easy: at a glance guide For each DNA test Component

More information

Principles of Gel Filtration Chromatography

Principles of Gel Filtration Chromatography Edvo-Kit #108 Principles of Gel Filtration Chromatography Experiment Objective: The objective of this experiment is to introduce the principles of gel fi ltration chromatography as a method that separates

More information

E.Z.N.A. Stool DNA Kit. D preps D preps D preps

E.Z.N.A. Stool DNA Kit. D preps D preps D preps E.Z.N.A. Stool DNA Kit D4015-00 5 preps D4015-01 50 preps D4015-02 200 preps July 2017 E.Z.N.A. Stool DNA Kit Table of Contents Introduction and Overview...2 Illustrated Protocol...3 Kit Contents/Storage

More information

Guide to using the Bio Rad CFX96 Real Time PCR Machine

Guide to using the Bio Rad CFX96 Real Time PCR Machine Guide to using the Bio Rad CFX96 Real Time PCR Machine Kyle Dobbs and Peter Hansen Table of Contents Overview..3 Setup Reaction Guidelines 4 Starting up the Software 5 Setup Protocol on Software 6 Setup

More information

Performing the CTA Carryover PVT Procedure on the LXi or DXC 600i System

Performing the CTA Carryover PVT Procedure on the LXi or DXC 600i System TM Performing the CTA Carryover PVT Procedure on the LXi or DXC 600i System The CTA carryover procedure is intended to qualify the wash efficiency of the CTA to ensure that the carryover is within the

More information

01-08R-E02 TS. MediaScout RoboColumn

01-08R-E02 TS. MediaScout RoboColumn Instruction Manual 01-08R-E02 TS Automated Parallel Chromatography MediaScout RoboColumn Table of contents Table of contents...2 1 Introduction...2 2 Characteristics...3 3 Advice on handling...4 3.1 Equipment...4

More information

PureSpeed Tips. Superior Protein Purity and Concentration Protein Purification in a Pipette Tip

PureSpeed Tips. Superior Protein Purity and Concentration Protein Purification in a Pipette Tip PureSpeed Tips PureSpeed Protein Tips Highest purity and concentration Fast as little as 15 minutes Process many samples at once Superior Protein Purity and Concentration Protein Purification in a Pipette

More information

ACTH (Human) Extraction Free EIA Kit Protocol (Cat. No.:EKE ) (Range: 0-100ng/ml)

ACTH (Human) Extraction Free EIA Kit Protocol (Cat. No.:EKE ) (Range: 0-100ng/ml) ACTH (Human) Extraction Free EIA Kit Protocol (Cat. No.:EKE-001-01) (Range: 0-100ng/ml) PHOENIX PHARMACEUTICALS, INC. INTRODUCTION This Enzyme Immunoassay kit is designed to detect a specific peptide and

More information

FastPure Spin Columns (Mini and Midi)

FastPure Spin Columns (Mini and Midi) G-Biosciences 1-800-628-7730 1-314-991-6034 technical@gbiosciences.com A Geno Technology, Inc. (USA) brand name FastPure Spin Columns (Mini and Midi) (Cat. # 786-1088, 786-1089, 786-1090, 786-1091) think

More information

MicroElute Cycle-Pure Kit

MicroElute Cycle-Pure Kit MicroElute Cycle-Pure Kit D6293-00 5 preps D6293-01 50 preps D6293-02 200 preps MicroElute Gel Extraction Kit D6294-00 5 preps D6294-01 50 preps D6294-02 200 preps MicroElute DNA Clean Up Kit D6296-00

More information

General protocol for ENZYME IMMUNOASSAY KIT (range: 0-25ng/ml)

General protocol for ENZYME IMMUNOASSAY KIT (range: 0-25ng/ml) General protocol for ENZYME IMMUNOASSAY KIT (range: 0-25ng/ml) PHOENIX PHARMACEUTICALS, INC. INTRODUCTION This Enzyme Immunoassay kit is designed to detect a specific peptide and its related peptides based

More information

NOS terminator (GMO) genesig Easy Kit 2 Target Gene Kit for use on the genesig q16

NOS terminator (GMO) genesig Easy Kit 2 Target Gene Kit for use on the genesig q16 TM Primerdesign Ltd genesig Easy Kit 2 Target Gene Kit for use on the genesig q16 50 reaction For general laboratory and research use only 1 genesig Easy: at a glance guide For each DNA test AT-tNOS-GM

More information

Empore. Solid Phase Extraction Cartridges. Technical Information. Product Description. Product Characteristics. Sorbents Available

Empore. Solid Phase Extraction Cartridges. Technical Information. Product Description. Product Characteristics. Sorbents Available Technical Information Empore Solid Phase Extraction Cartridges Product Description Empore Solid Phase Extraction Cartridges are designed for sample pretreatment to remove or minimize sample matrix and

More information

Average Yields* Yeast DNA Yeast RNA Time to Complete 10 Purifications * Yield will vary depending on the type of sample processed

Average Yields* Yeast DNA Yeast RNA Time to Complete 10 Purifications * Yield will vary depending on the type of sample processed 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Fungi/Yeast RNA/DNA Purification Kit Product # 35800 Product Insert

More information

Automated Protocol for ANTI-FLAG High Sensitivity, M2 coated 96-well plate Using the Sciclone ALH 3000 Workstation (Caliper Life Sciences)

Automated Protocol for ANTI-FLAG High Sensitivity, M2 coated 96-well plate Using the Sciclone ALH 3000 Workstation (Caliper Life Sciences) Automated Protocol for ANTI-FLAG High Sensitivity, M2 coated 96-well plate Using the Sciclone ALH 3000 Workstation (Caliper Life Sciences) ANTI-FLAG High Sensitivity, M2 coated 96-well plate P 2983 Automation

More information

PowerPlex Matrix Standards, 3100/3130

PowerPlex Matrix Standards, 3100/3130 TECHNICAL BULLETIN PowerPlex Matrix Standards, 3100/3130 Instruc ons for use of Product DG4650 Note: The PowerPlex Matrix Standards, 3100/3130, can be used to perform spectral calibra on on the Applied

More information

Seeplex LabChip Kit LabChip Dx User Guide

Seeplex LabChip Kit LabChip Dx User Guide Contents SPECIFICATIONS.......2 SAMPLE CONDITIONS.........2 KIT CONTENTS.......2 SAFETY WARNINGS AND PRECAUTIONS... 4 ADDITIONAL ITEMS REQUIRED... 4 INSTRUMENTS... 4 CONSUMABLES... 4 REAGENTS... 4 PREPARATION

More information

Human IL-6 ELISA Set

Human IL-6 ELISA Set Human IL-6 ELISA Set Catalog No. CDK082B Quantity: 10 x 96 tests PRODUCT SPECIFICATIONS : Specificity: Recognizes both natural and recombinant human IL-6 Range: 6.25 pg / ml - 200 pg / ml Sensitivity:

More information

Cells and Tissue DNA Isolation Kit (Magnetic Bead System) 50 Preps Product # 59100

Cells and Tissue DNA Isolation Kit (Magnetic Bead System) 50 Preps Product # 59100 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Cells and Tissue DNA Isolation Kit (Magnetic Bead System) 50 Preps

More information

Protocol for Catalog # FEK BNP-45 (Mouse) FLUORESCENT IMMUNOASSAY KIT (range: 0-10,000pg/ml)

Protocol for Catalog # FEK BNP-45 (Mouse) FLUORESCENT IMMUNOASSAY KIT (range: 0-10,000pg/ml) Protocol for Catalog # FEK-011-23 BNP-45 (Mouse) FLUORESCENT IMMUNOASSAY KIT (range: 0-10,000) PHOENIX PHARMACEUTICALS, I NC. INTRODUCTION This Fluorescent Immunoassay kit is designed to detect a specific

More information

Human Myostatin, ELISA Kit (MSTN)

Human Myostatin, ELISA Kit (MSTN) Human Myostatin, ELISA Kit (MSTN) 96 Tests Catalog Number: MBS733837 Store all reagents at 2-8 C Valid Period: six months For samples: Cell culture fluid & body fluid & tissue homogenate Serum or blood

More information

UltraClean Midi Plasmid Prep Kit

UltraClean Midi Plasmid Prep Kit UltraClean Midi Plasmid Prep Kit Catalog No. Quantity 12700-20 20 Preps Instruction Manual Please recycle Version: 05232014 1 Table of Contents Introduction... 3 Protocol Overview... 3 Flow Chart... 4

More information

E.Z.N.A. Bacterial RNA Kit. R preps R preps

E.Z.N.A. Bacterial RNA Kit. R preps R preps E.Z.N.A. Bacterial RNA Kit R6950-00 5 preps R6950-01 50 preps July 2017 E.Z.N.A. Bacterial RNA Kit Table of Contents Introduction and Overview...2 Kit Contents/Storage and Stability...3 Before Beginning...4

More information

KAPA Library Amplification Kit Illumina Platforms

KAPA Library Amplification Kit Illumina Platforms KAPA Library Amplification Kit KR0408 v7.17 This provides product information and a detailed protocol for the KAPA Library Amplification Kits. This documents applies to KAPA Library Amplification Kits

More information

Formulating by Cost. Enginuity 5.0 by DS ENOVIA Dassault Systèmes

Formulating by Cost. Enginuity 5.0 by DS ENOVIA Dassault Systèmes Formulating by Cost Enginuity 5.0 by DS ENOVIA 2014 Dassault Systèmes Contents About This Guide... 1 Cost Formulation Workflow... 2 Define Ingredient Substitutions in the Formula... 4 Choose Ingredient

More information

sbeadex maxi plant kit

sbeadex maxi plant kit sbeadex maxi plant kit Catalogue numbers 41602 and 41620 (For research use only. Not for use in diagnostic procedures.) Contents 1. Introduction 2. Kit contents and storage conditions 2.1 sbeadex maxi

More information

Oracle. SCM Cloud Using Receiving. Release 13 (update 17D)

Oracle. SCM Cloud Using Receiving. Release 13 (update 17D) Oracle SCM Cloud Release 13 (update 17D) Release 13 (update 17D) Part Number E89219-01 Copyright 2011-2017, Oracle and/or its affiliates. All rights reserved. Author: Kristy Carlile This software and related

More information

Winzer Corporation 1 Revision: 4.0

Winzer Corporation 1 Revision: 4.0 Table of Content Section 1: Getting Started... 2 1.1 Overview... 2 Section 2: Installation Overview... 3 2.1 Install CD / Start.exe... 3 2.2 Welcome Screen... 4 2.3 Device Selection... 4 2.4 Auto or Industrial...

More information

Plexor HY System for the Applied Biosystems 7500 and 7500 FAST Real-Time PCR Systems

Plexor HY System for the Applied Biosystems 7500 and 7500 FAST Real-Time PCR Systems TECHNICAL MANUAL Plexor HY System for the Applied Biosystems 7500 and 7500 FAST Real-Time PCR Systems Instructions for Use of Products DC1000, DC1001 and DC1500 Technical Manuals with instructions for

More information

Hamilton Microlab STAR Reference Guide

Hamilton Microlab STAR Reference Guide Hamilton Microlab STAR Reference Guide Revision History 1 Introduction 2 ML STAR Components 3 System Software 7 Safety Precautions 8 Perform A Run 10 Deck Layout and Icon Legend 12 Maintenance 14 Troubleshooting

More information

E.Z.N.A. Blood DNA Maxi Kit. D preps D preps

E.Z.N.A. Blood DNA Maxi Kit. D preps D preps E.Z.N.A. Blood DNA Maxi Kit D2492-00 2 preps D2492-03 50 preps April 2014 E.Z.N.A. Blood DNA Maxi Kit Table of Contents Introduction and Overview...2 Kit Contents/Storage and Stability...3 Preparing Reagents...4

More information

Agilent MetaCarb Ca PLUS Carbohydrate Column Ca ++ Form User Manual Part No. A5205

Agilent MetaCarb Ca PLUS Carbohydrate Column Ca ++ Form User Manual Part No. A5205 Agilent MetaCarb Ca PLUS Carbohydrate Column Ca ++ Form User Manual Part No. A5205 WARNING: The Agilent MetaCarb Ca PLUS calcium column is packed with a polymeric material that requires special care. Introduction

More information

PathogINDICAtor qpcr Microbial Detection Assay on the AriaMX Real-Time PCR System Optional Decontamination Step Page 1 of 12.

PathogINDICAtor qpcr Microbial Detection Assay on the AriaMX Real-Time PCR System Optional Decontamination Step Page 1 of 12. Page 1 of 12 Please refer to http://www.medicinalgenomics.com/product-literature/ for updated protocols and Material Safety Data Sheets (MSDS). Consult MSDS before using any new product. PATHOGINDICATOR

More information

Microsoft Dynamics GP Business Portal. Project Time and Expense User s Guide Release 3.0

Microsoft Dynamics GP Business Portal. Project Time and Expense User s Guide Release 3.0 Microsoft Dynamics GP Business Portal Project Time and Expense User s Guide Release 3.0 Copyright Copyright 2005 Microsoft Corporation. All rights reserved. Complying with all applicable copyright laws

More information

QIAcube Pure Efficiency

QIAcube Pure Efficiency QIAcube Pure Efficiency Sample & Assay Technologies Walkaway spin-column processing The QIAcube automates your spin preps The revolutionary QIAcube makes automated sample prep available to all labs. The

More information

It All Starts Here. WEB WORK v.7 USER MANUAL Chapter 8 Preventive Maintenance (PM) Module Maintenance section of Azzier. Tero Consulting Ltd.

It All Starts Here. WEB WORK v.7 USER MANUAL Chapter 8 Preventive Maintenance (PM) Module Maintenance section of Azzier. Tero Consulting Ltd. It All Starts Here. WEB WORK v.7 USER MANUAL Chapter 8 Preventive Maintenance (PM) Module Maintenance section of Azzier Port Coquitlam BC V3C 6G5 T: 604-468-1401 F: 604-468-1408 E: sales@tero.ca W: www.tero.ca

More information

Agilent 2100 Bioanalyzer System

Agilent 2100 Bioanalyzer System Agilent 2100 Bioanalyzer System Maintenance and Troubleshooting Guide Agilent Technologies Notices Agilent Technologies, Inc. 2011-2016, 2017 No part of this manual may be reproduced in any form or by

More information

MD60002 MD MD62002

MD60002 MD MD62002 MagSi-DNA saliva Art.No. MD60002 MD61002 - MD62002 Product Manual Version 1.0 22/01/2015 Table of Contents 1. General Information...3 1.1 Intended Use...3 1.2 Kit specifications...3 1.3 Basic principle...3

More information

Peste-des-petits-ruminants virus. genesig Easy Kit for use on the genesig q reaction. Primerdesign Ltd

Peste-des-petits-ruminants virus. genesig Easy Kit for use on the genesig q reaction. Primerdesign Ltd TM Primerdesign Ltd Peste-des-petitsruminants virus genesig Easy Kit for use on the genesig q16 50 reaction For general laboratory and research use only 1 genesig Easy: at a glance guide For each RNA test

More information

Primerdesign TM Ltd. Eubacteria. genesig Easy Kit for use on the genesig q reactions. For general laboratory and research use only

Primerdesign TM Ltd. Eubacteria. genesig Easy Kit for use on the genesig q reactions. For general laboratory and research use only Primerdesign TM Ltd Eubacteria genesig Easy Kit for use on the genesig q16 50 reactions For general laboratory and research use only 1 genesig Easy: at a glance guide For each DNA test Component Volume

More information

Purification of mfp. from an Overnight Culture. Laboratory 17

Purification of mfp. from an Overnight Culture. Laboratory 17 Purification of mfp from an Overnight Culture When scientists at a therapeutics company, like Amgen, have successfully identified a promising therapeutic protein, two objectives would be to locate and

More information

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany New Features of the Agilent Method Scouting Wizard for Automated Method Development of Complex Samples Analysis of Large Data Sets by Method Scouting Reports and Automated Adjustment of Flow Rates and

More information

Application Note. Authors. Abstract

Application Note. Authors. Abstract Automated, High Precision Tryptic Digestion and SISCAPA-MS Quantification of Human Plasma Proteins Using the Agilent Bravo Automated Liquid Handling Platform Application Note Authors Morteza Razavi, N.

More information

Genome Reagent Kits v2 Quick Reference Cards

Genome Reagent Kits v2 Quick Reference Cards Chromium Genome Reagent Kits v2 Quick Reference Cards FOR USE WITH Chromium Genome Library Kit & Gel Bead Kit v2, 16 rxns PN-120258 Chromium Genome Chip Kit v2, 48 rxns PN-120257 Chromium i7 Multiplex

More information

E.Z.N.A. Tissue RNA Kit. R preps R preps

E.Z.N.A. Tissue RNA Kit. R preps R preps E.Z.N.A. Tissue RNA Kit R6688-00 5 preps R6688-01 50 preps May 2015 E.Z.N.A. Tissue RNA Kit Table of Contents Introduction and Overview...2 Kit Contents/Storage and Stability...3 Important Notes...4 Homogenization

More information

Note: for laboratory research use only. RNA High-purity Total RNA Rapid Extraction Kit (Spin-column) Signalway Biotechnology

Note: for laboratory research use only. RNA High-purity Total RNA Rapid Extraction Kit (Spin-column) Signalway Biotechnology Note: for laboratory research use only RNA High-purity Total RNA Rapid Extraction Kit (Spin-column) Cat. #: RP1202 (50preps) Signalway Biotechnology I. Kit Content, Storage Condition and Stability Content

More information

Human IL-1 alpha ELISA Kit

Human IL-1 alpha ELISA Kit Human IL-1 alpha ELISA Kit Catalog No: CDK025B Quantity: 2 x 96 tests SPECIFICITY : RANGE : SENSITIVITY : INCUBATION : SAMPLE TYPES : Recognizes both natural and recombinant human IL-1 alpha 31.2 pg /

More information

celldatasci.com/rnastorm RNAstorm RNA Isolation Kit for FFPE Tissue Samples Sample Kit (20 extractions)

celldatasci.com/rnastorm RNAstorm RNA Isolation Kit for FFPE Tissue Samples Sample Kit (20 extractions) celldatasci.com/rnastorm info@celldatasci.com RNAstorm RNA Isolation Kit for FFPE Tissue Samples Sample Kit (20 extractions) Support: Email: support@celldatasci.com Phone: 650.285.2376 (option 2) Toll-free:

More information

Human IgG ELISA Quantitation Set

Human IgG ELISA Quantitation Set Human IgG ELISA Quantitation Set Cat. No. E80-104 Components Supplied Affinity purified Goat anti-human IgG-Fc Coating Antibody A80-104A, 1 ml at 1 mg/ml Human Reference Serum, RS10-110-4, 0.1 ml HRP Conjugated

More information

Principles of Gel Filtration Chromatography

Principles of Gel Filtration Chromatography EDVOTEK P.O. Box 1232 West Bethesda, MD 20827-1232 The Biotechnology Principles of Gel Filtration Chromatography 108 EDVO-Kit # Storage: Store entire experiment at room temperature. Experiment Objective:

More information

EarlyTox Cell Integrity Kit

EarlyTox Cell Integrity Kit EarlyTox Cell Integrity Kit The EarlyTox Cell Integrity Kit from Molecular Devices is an optimized set of reagents that simplifies the measurement of live and dead cells in a single well. The assay uses

More information

NRAS Codon 61 Mutation Analysis Reagents

NRAS Codon 61 Mutation Analysis Reagents NRAS Codon 61 Mutation Analysis Reagents User Manual V1.0 Cat No. GP19 32 reactions 1 CONTENTS Introduction 4 Overview of Mutector TM Assay 5 Materials Provided 6 Materials Required 7 Equipment Required

More information

Urine DNA Isolation Maxi Kit (Slurry Format) Product # 50100

Urine DNA Isolation Maxi Kit (Slurry Format) Product # 50100 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Urine DNA Isolation Maxi Kit (Slurry Format) Product # 50100 Product

More information

****** Competition ELISA Kit Instruction

****** Competition ELISA Kit Instruction FOR RESEARCH USE ONLY. NOT FOR DIAGNOSTIC PURPOSES ****** Competition ELISA Kit Instruction Kit name and catalog number Your analyte ELISA Kit, Catalog#: ***** Intended use The kit is used to detect the

More information

QIAfilter Plasmid Midi Kit (Cat #: 12243)

QIAfilter Plasmid Midi Kit (Cat #: 12243) QIAfilter Plasmid Midi Kit (Cat #: 12243) Things to do before starting Add the provided RNase A solution to Buffer P1 before use. Use one vial of RNase A (centrifuge briefly before use) per bottle of Buffer

More information

Human Cancer Antigen 15-3 (CA 15-3) ELISA Kit

Human Cancer Antigen 15-3 (CA 15-3) ELISA Kit Product Manual Human Cancer Antigen 15-3 (CA 15-3) ELISA Kit Catalog Numbers PRB- 5069 PRB- 5069-5 96 assays 5 x 96 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures Introduction Breast

More information

Maxwell RSC Blood DNA Kit

Maxwell RSC Blood DNA Kit TECHNICAL MANUAL Maxwell RSC Blood DNA Kit Instructions for Use of Product AS1400 Caution: Handle cartridges with care; seal edges may be sharp. Revised 1/15 TM419 Maxwell RSC Blood DNA Kit All technical

More information

Apium graveolens genesig Easy Kit for use on the genesig q16

Apium graveolens genesig Easy Kit for use on the genesig q16 TM Primerdesign Ltd genesig Easy Kit for use on the genesig q16 50 reaction For general laboratory and research use only 1 genesig Easy: at a glance guide For each DNA test Component Volume Lab-in-a-box

More information

PowerSoil DNA Isolation Kit

PowerSoil DNA Isolation Kit PowerSoil DNA Isolation Kit Catalog No. Quantity 12888-50 50 Preps 12888-100 100 Preps Instruction Manual Introduction The PowerSoil DNA Isolation Kit* is comprised of a novel and proprietary method for

More information

TaKaRa MiniBEST Agarose Gel DNA Extraction Kit Ver.4.0

TaKaRa MiniBEST Agarose Gel DNA Extraction Kit Ver.4.0 Cat. # 9762 For Research Use TaKaRa MiniBEST Agarose Gel Product Manual Table of Contents I. Description... 3 II. Kit Components... 3 III. Shipping and Storage... 3 IV. Preparation before Use... 4 V. Protocol...

More information