TECHNICAL INFORMATION

Size: px
Start display at page:

Download "TECHNICAL INFORMATION"

Transcription

1 T-1860A TECHNICAL INFORMATION Capillary Electrophoresis PRECISION IN CAPILLARY ELECTROPHORESIS Harry Whatley and Jeff Chapman Beckman Coulter, Inc. Introduction Instrumentation for routine chemical analysis must have an acceptable degree of precision. The instrument must give the same result for a given sample from run to run and from day to day. A technology must be shown to be sufficiently reproducible for the task at hand before it can be accepted for routine use. Many advances in instrument system design have been made since capillary electrophoresis was commercialized in the late 1980s. As this note will demonstrate, P/ACE System MDQ is capable of achieving the degree of migration time and peak area precision necessary for routine chemical analysis. Reproducibility in CE Many factors are involved in reproducibility. Some of these, such as temperature control, voltage control, and sample injection precision, are inherent in the design of the instrument. Other factors, such as the quality of the reagents used and the manner in which the instrument is programmed and operated, are completely in the hands of the user. These two factors, the quality of the instrument and the quality of the operation, are both required in order to achieve reproducible results. To determine the typical reproducibility that can be achieved on the P/ACE System MDQ, we evaluated the separation of two very similar analytes in a commercially available test mixture (System Performance Test Mix B, Beckman Coulter, Inc.). This test mix contains two organic acids: p-hydroxybenzoic acid (0.72 mmol/l) and p-hydroxyphenylacetic acid (0.66 mmol/l). Figure 1 illustrates a typical separation of these two compounds. Figure 2 shows the peak area plotted against run number from 100 successive runs of a single sample of the test mixture. The peak area %RSD for the entire set of 100 runs is 1.52% for Peak A and 1.67% for Peak B. It is not usual for analyses to require this many replicates: six is a typical number. Figure 3 shows the moving %RSD of the area of Peak A for any six consecutive runs (1-6, 2-7, etc., giving 95 sets of data points). The average %RSD for the sets of six runs is 0.89% with a range from 0.42% to 1.48%. The above data represent peak areas which have not been corrected for velocity. In CE, peak area is not only a function of concentration but also of an analyte s velocity through the detection window. Any irreproducibility in migration time will compound the error in peak area determination. In capillary electrophoresis, it is usual to normalize peak areas for migration velocity. The resulting parameter is referred to as corrected peak area. The corrected area %RSD for the entire set of 100 runs is 1.43% for Peak A and 1.56% for Peak B. The average %RSD for the sets of six runs is 0.90% with a range from 0.44% to 1.46%. The fact that the %RSD values from both methods of area calculation are essentially the same indicates that the migration velocities were nearly constant throughout this experiment. Peak migration time is another factor for which precision is important. Very often the migration time of a given analyte is used to identify the species present in a mixture. Peaks that drift from run to run make identification difficult. In the present experiment the migration times of the two components in

2 Run # 20 of 100 Peak B %RSD, A Absorbance A = p-hydroxyphenylacetic acid B = p-hydroxybenzoic acid Peak A %RSD Minutes Figure 1. Typical separation from this experiment. Peak Area Area A Area B Run Number Figure 2. Peak area for 100 runs Set Number (n=6) Figure 3. %RSD values for sets of 6 consecutive runs, Peak A (Peak B is similar). mt, Minutes mt A mt B Run Number Figure 4. Migration time of Peaks A and B. the mixture were quite consistent (as shown in Figure 4). The migration time %RSD for the entire set of 100 runs is 0.19% for both Peaks A and B. Using the six-run-moving-average method for %RSD gives values ranging from 0.00% to 0.34% for Peak A and 0.00% to 0.38% for Peak B. Reproducibility of Mobility Separation of analytes in capillary zone electrophoresis is a function of charge and mass. The calculated mobility (µ, cm 2 /Volt-sec) of the analytes is an important factor to be considered. Mobility is useful in analyte identification strategies as it accounts for normal variations in such run factors as electroosmotic flow and applied voltage. Automatic mobility determination is one of the CE-specific functions of the P/ACE System MDQ software. Accurate determination of the voltage profile up to the point where the peak emerges is critical to the correct calculation of mobility. P/ACE System MDQ integrates the voltage profile on a point-by-point basis to account for variations such as voltage ramps. In the determination of mobility, a neutral reference marker such as acetone or benzyl alcohol is used to take into consideration the impact of electroosmotic flow (EOF). This is often inconvenient in routine analyses. Mobility values suitable for peak identification can be obtained by using a reference marker that is similar in chemical structure to the analytes of interest in place of the neutral marker. By way of example, we can consider Peak A as a reference marker and assign it a value of µ = cm 2 /Volt-sec (this value having been previously obtained using a neutral marker). The P/ACE System MDQ software can use Peak A as a mobility standard against which other peaks can be compared. Using this technique to calculate the mean mobility of Peak B for the entire set of 100 runs gives a value of µ = cm 2 /Volt-sec with a %RSD of 0.07%. Using the six-run-movingaverage method for %RSD gives values ranging from 0.01% to 0.10%. 2

3 How Are Results Like This Obtained? Table 1 describes the separation conditions employed in this study. The results of these experiments are summarized in Appendix A. The complete set of MDQ data files is available for review contact a Beckman Coulter, Inc., representative for details on how to obtain a copy. The key factors in obtaining the reproducible results shown in this report are: Control of buffer ion depletion Proper injection conditions Adequate capillary conditioning Buffer Ion Depletion: During any electrophoretic separation, buffer ions migrate from one buffer reservoir to another along with the analytes. Over large numbers of injections, the buffer composition in the reservoirs will change sufficiently that the migration of the analytes will be affected. This can be controlled by limiting the number of samples run with any one pair of buffer vials or by using larger buffer reservoirs. The separations shown here were performed with the large buffer reservoir accessory on the P/ACE System MDQ (Figure 5). Each buffer tray has two 25 ml reservoirs. This allows extended analyses without the effects of ion depletion. Injection Conditions: P/ACE capillary electrophoresis instruments utilize a pressure-time integral calculation to correct for variances during injection. With positive pressure injection, this is made necessary by the time required for pressure to ramp up and ramp down and by the response time of pressure-sensing devices. These variances are monitored and the injection time adjusted so that the product (pressure time) is constant from run to run. In theory, a 5-psi, 1-second injection and the 0.5-psi, 10- second injection used in this study should give the same injected volume. In practice, a longer, lower pressure injection gives the instrumentation more time to respond to variances and will give a better result. Another useful technique that was employed here is to dip the capillary into water between injection and separation. The parallel electrode-capillary design of P/ACE System MDQ makes a simple water dip effective for reducing sample transfer from vial to vial. This step is accomplished through the P/ACE System MDQ software by programming a wait function while sending the capillary to a water vial. Using a wait time of zero will result in a brief dip followed immediately by the next method step. Table 1. Separation Conditions Instrument: P/ACE System MDQ Capillary: 75 µm internal diameter 50 cm to the detector window 60 cm total length Run Buffer: P/ACE Capillary Performance Test Buffer A (P/N ) in Large Reservoir Tray (P/N ) Sample: P/ACE Capillary Test Mixture B (P/N ) (A mixture of p-hydroxybenzoic and p-hydroxyphenylacetic acids) Detection: UV absorbance at 214 nm Method: Inject Positive pressure, 0.5 psi for 10 seconds Wait Water vial dip, time = 0.0 Separate 30 kv constant voltage, 5 minutes, normal polarity, 25 C Rinse Run buffer, 20 psi for 1 minute Figure 5. Buffer Reservoir Tray as used in this study (P/N ). Injection precision is a prerequisite for peak area reproducibility. If injection volumes are not controlled from run to run, it will be impossible to construct a calibration curve that will be suitable for quantitative analysis. When using external standards to construct a calibration curve, the standards and the samples should be matched as closely as possible. If the viscosity of the sample and the standards is not matched (such as sampling blood serum against standards dissolved in water), the actual volume of sample injected may be different from the volume of standard injected. Similarly, if the conductivity of the standards and the samples is not matched, the migration times or peak widths of the analytes may differ between the standards and the 3

4 samples. If it is not possible to match the standards with the samples, spiking or co-injection of standards into samples will provide a better way to identify and quantitate unknowns than will using an external calibration curve. At very high analyte concentrations, it may be observed that peak area increases without a corresponding increase in peak height (i.e., the peaks become wider). With oncapillary detection, this is seldom a problem except with very closely spaced peaks. With extended-path detection schemes, this peak broadening may increase the probability of the co-detection of multiple components. Capillary Conditioning: Capillary conditioning is often underestimated as a source of reproducibility problems. A surface should be chosen that is compatible with the types of analytes that are being separated. When using a non-coated, fused-silica capillary, it is important to understand the nature of the charge at the walls. At ph values above 3.0, the capillary wall begins to be charged and electroosmotic flow becomes a significant factor in the separation. A used capillary surface will also exhibit a memory effect it remembers the last buffer system to which it was exposed. Time is required for equilibration with a new buffer system. A new, bare fused-silica capillary should be cleaned and regenerated prior to use. A suggested protocol for starting up is given in Table 2. After cleaning (and again after storage), the capillary must be equilibrated with the buffer system that will be used for analysis. A simple way to do this is to run five to ten blank runs (no sample injection) prior to running sample analyses. The need for this can be seen in Figure 4. For the first few runs, the peak migration times drift slightly, even though this capillary had been used a few days earlier for this same analysis. It is a good practice to dedicate a piece of capillary to one type of buffer and preferably to one analytical task. Alternating a single piece of capillary between different buffer systems (such as borate and phosphate) may necessitate very long equilibration times. Table 2. Regeneration Procedure for Uncoated Silica Capillary For Maximized EOF Methanol (HPLC grade or better) 10 minutes* HCl, 1 N, 10 minutes NaOH, 0.1 N, 10 minutes Run buffer, 10 minutes Use of 0.1 N NaOH between sample runs may be necessary to maintain consistent EOF flow. For Analyses with Low-pH Phosphate Buffer (EOF Minimized) Methanol (HPLC grade or better) 10 minutes HCl, 1 N, 10 minutes Phosphoric acid, 1 N, 10 minutes Run buffer, 10 minutes When using low-ph buffers, do not rinse with NaOH between runs. If clean-up is necessary, use 1 N acid. * All times are for a 75 µm capillary, 60 cm total length, 20 to 30 C, 20 psi. For other capillary dimensions, adjust time and/or pressure so that the wash solution is replaced at least 20 times. CE Expert (available from Beckman Coulter, Inc.) is a software package that can be used to calculate fluid delivery under various conditions. Conclusion Capillary electrophoresis can be a reproducible technique. The basic requirements are a quality instrument system that is properly operated, calibrated, and equilibrated. In addition to the factors covered here, reproducible results require the use of clean, fresh chemistries and good day-to-day system cleaning and maintenance. By taking care to recreate the run conditions, capillary electrophoresis can be a valuable tool for routine analytical tasks. 4

5 Appendix A. Results from 100 Runs Migration (min) Mobility Raw Peak Area Corrected Area Rep mt A mt B Peak B Area A Area B c Area A c Area B

6 All trademarks are the property of their respective owners Developing innovative solutions in genetic analysis, drug discovery, and instrument systems. Beckman Coulter, Inc N. Harbor Boulevard, Box 3100 Fullerton, California Sales: Service: Telex: Fax: Worldwide Bioresearch Division Offices: Australia (61) Canada (905) China (8610) Eastern Europe, Middle East, Africa (41) France (33) Germany (49) Hong Kong (852) Italy (39) Japan Mexico Netherlands (31) Singapore (65) South Africa (27) /5 Spain (34) Sweden 46 (0) Switzerland Taiwan (886) U.K. (44) U.S.A B CB Beckman Coulter, Inc. Printed in U.S.A. on recycled paper.

LIFluor EnhanCE. for use with dsdna 1000 and dsdna 20,000 Kits

LIFluor EnhanCE. for use with dsdna 1000 and dsdna 20,000 Kits LIFluor EnhanCE for use with dsdna 1000 and dsdna 20,000 Kits For the Separation of Double Stranded DNA by P/ACE Fluorescent Capillary Electrophoresis 725824 AB September 2006 Beckman Coulter, Inc. 4300

More information

Instructions for Capillary Electrophoresis Peptide Analysis Kit

Instructions for Capillary Electrophoresis Peptide Analysis Kit Instructions for Capillary Electrophoresis Peptide Analysis Kit Catalog Number 148-4110 For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800-4BIORAD (1-800-424-6723) Table of

More information

Application Note. Biopharm. Author. Abstract. Christian Wenz Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Biopharm. Author. Abstract. Christian Wenz Agilent Technologies, Inc. Waldbronn, Germany Performance of commercially available gels for protein characterization by capillary gel electrophoresis with UV detection on the Agilent 7100 CE System Application Note Biopharm Author Christian Wenz

More information

Optimization of cief to Resolve Challenging MAb Samples of Clinical Immunodiagnostic Relevance

Optimization of cief to Resolve Challenging MAb Samples of Clinical Immunodiagnostic Relevance Optimization of cief to Resolve Challenging MAb Samples of Clinical Immunodiagnostic Relevance Sushma Rampal, Sr. Application Scientist, Beckman Coulter Life Sciences, Brea, CA, USA Marcia Santos, Staff

More information

SCIEX Anion Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide

SCIEX Anion Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems A49108AE May 2015 AB Sciex Pte. Ltd and its affiliates disclaims all warranties with respect to this document, expressed or implied, including

More information

PA 800 plus Catalog. Systems/Software

PA 800 plus Catalog. Systems/Software PA 800 plus Catalog Built on 20 years of innovation and leadership in the field of capillary electrophoresis, Beckman Coulter s PA 800 plus provides solutions for protein purity analysis, charge heterogeneity

More information

ssdna 100-R Kit Care and Use Instructions AG June 2015

ssdna 100-R Kit Care and Use Instructions AG June 2015 Care and Use Instructions 726479 AG June 2015 Beckman Coulter, Inc. 250 S. Kraemer Blvd., Brea, CA 92821 USA 2015 Beckman Coulter, Inc. Copyright, Licenses and Trademarks Copyright Beckman Coulter, Inc.,

More information

Separation of Recombinant Human Erythropoietin (rhepo) using the European Pharmacopoeia Method on the PA 800 plus Pharmaceutical Analysis System

Separation of Recombinant Human Erythropoietin (rhepo) using the European Pharmacopoeia Method on the PA 800 plus Pharmaceutical Analysis System Separation of Recombinant Human Erythropoietin (rhepo) using the European Pharmacopoeia Method on the PA 800 plus Pharmaceutical Analysis System Marcia R Santos Global Tactical Marketing, Beckman Coulter

More information

IgG Purity/Heterogeneity and SDS-MW Assays with High- Speed Separation Method and High Throughput Tray Setup

IgG Purity/Heterogeneity and SDS-MW Assays with High- Speed Separation Method and High Throughput Tray Setup IgG Purity/Heterogeneity and SDS-MW Assays with High- Speed Separation Method and High Throughput Tray Setup High Throughput Methods to Maximize the Use of the PA 800 Plus system Jose-Luis Gallegos-Perez

More information

Hillary B. Hewitson, Thomas E. Wheat, Diane M. Diehl INTRODUCTION

Hillary B. Hewitson, Thomas E. Wheat, Diane M. Diehl INTRODUCTION Enhanc em ent o f t h e U P L C Am ino Ac i d Ana lysis So lu t io n w it h F l e x ibl e Detector Options Hillary B. Hewitson, Thomas E. Wheat, Diane M. Diehl INTRODUCTION The measurement of amino acids

More information

APPLICATION TECHNICAL INFORMATION

APPLICATION TECHNICAL INFORMATION T-1975A APPLICATION TECHNICAL INFORMATION Genetic Analysis METHOD FOR LONG SEQUENCING READ LENGTHS ON THE CEQ GENETIC ANALYSIS SYSTEM Doni Clark, Jim Thorn, and Keith Roby Beckman Coulter, Inc. Introduction

More information

Chromatography Column Performance and Data Analysis Success Guide. Hints and Tips for Better Purifications

Chromatography Column Performance and Data Analysis Success Guide. Hints and Tips for Better Purifications Chromatography Column Performance and Data Analysis Success Guide Hints and Tips for Better Purifications This Chromatography Success Guide provides practical advice on preparative chromatography and protein

More information

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime EASY-Spray Columns Guidance for column set up and installation Tips to maximize column lifetime EASY-Spray Column Tips and Tricks This document provides guidance for Thermo Scientific EASY-Spray column

More information

APPLICATION INFORMATION

APPLICATION INFORMATION A-1932A APPLICATION INFORMATION Genetic Analysis: CEQ Series AUTOMATED BINNING PROCESS FOR THE GENERATION OF LOCUS TAGS Heather Gull, Dana Campbell, and Mark Dobbs Beckman Coulter, Inc. Short tandem repeats

More information

Nuvia cprime Hydrophobic Cation Exchange Media

Nuvia cprime Hydrophobic Cation Exchange Media Nuvia cprime Hydrophobic Cation Exchange Media Instruction Manual Catalog numbers 156-3401 156-3404 156-3402 156-3405 156-3403 156-3406 Please read the instructions in this manual prior to using Nuvia

More information

New Core-Shell Technology

New Core-Shell Technology New Core-Shell Technology Fortis Speedcore columns are the very latest in core-shell technology. Incorporating our optimised bonding and packing practices with a core-shell particle provides the analyst

More information

ANALYTICAL STANDARDS AND REAGENTS

ANALYTICAL STANDARDS AND REAGENTS ANALYTICAL STANDARDS AND REAGENTS productivity future growth globalization compliance Waters, the industry leader in liquid chromatography and mass spectrometry, understands that standards and reagents

More information

CSE Separation of Cancer Cell Lysate

CSE Separation of Cancer Cell Lysate Application Note: CSE Separation of Cancer Cell Lysate No. 9.200411 Rev. A 2004 Target Discovery, Inc. Target Discovery, Inc. 4015 Fabian Way Palo Alto, CA 94303 Tel: 650-812-8100 www.targetdiscovery.com/eotrol

More information

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover columns ProSwift Reversed-Phase Monolith Columns for Protein Analysis ProSwift reversed-phase columns use a unique monolith technology for fast, high-resolution HPLC and LC/MS separations of proteins.

More information

CAPILLARY ELECTROPHORESIS SEPARATION OF SINGLE STRANDED DNA OLIGONUCLEOTIDES WITH AN UNCOATED BARE FUSED SILICA CAPILLARY. Corey J.

CAPILLARY ELECTROPHORESIS SEPARATION OF SINGLE STRANDED DNA OLIGONUCLEOTIDES WITH AN UNCOATED BARE FUSED SILICA CAPILLARY. Corey J. CAPILLARY ELECTROPHORESIS SEPARATION OF SINGLE STRANDED DNA OLIGONUCLEOTIDES WITH AN UNCOATED BARE FUSED SILICA CAPILLARY Corey J. Garman A Thesis Submitted to the University of North Carolina Wilmington

More information

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis Shanhua Lin, Julia Baek, and Chris Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA Overview

More information

electrophoresis tech Versatile Separation Capabilities of the PROTEAN i12 IEF System

electrophoresis tech Versatile Separation Capabilities of the PROTEAN i12 IEF System electrophoresis tech note 6139 Versatile Separation Capabilities of the PROTEAN i12 IEF System ChengJun Sun, Adriana Harbers, Tom Berkelman, and Katy McGirr, Life Science Group, Bio-Rad Laboratories Inc.,

More information

ALLIANCE SYSTEM FOR CARBAMATE ANALYSIS

ALLIANCE SYSTEM FOR CARBAMATE ANALYSIS ALLIANCE SYSTEM FOR CARBAMATE ANALYSIS Are you concerned with any of these issues? Is your method capable of meeting the precision and accuracy requirements for carbamates mandated by the Environmental

More information

Bio-Rad Laboratories BIOPLEX 2200 SYSTEM. BioPlex 2200 ANA Screen with MDSS. The first and only fully-automated, multiplexed ANA Screen

Bio-Rad Laboratories BIOPLEX 2200 SYSTEM. BioPlex 2200 ANA Screen with MDSS. The first and only fully-automated, multiplexed ANA Screen Bio-Rad Laboratories BIOPLEX 2200 SYSTEM BioPlex 2200 ANA Screen with MDSS The first and only fully-automated, multiplexed ANA Screen Bio-Rad Laboratories BIOPLEX 2200 SYSTEM Like no other The BioPlex

More information

Thermo Scientific EASY-nLC 1200 System. Leading in simplicity. and performance

Thermo Scientific EASY-nLC 1200 System. Leading in simplicity. and performance Thermo Scientific EASY-nLC 1200 System Leading in simplicity and performance Peak performance made EASY Effortless ultra high performance for everybody A straightforward LC-MS solution Optimized and integrated

More information

Affi-Gel Protein A MAPS II Kit Instruction Manual

Affi-Gel Protein A MAPS II Kit Instruction Manual Affi-Gel Protein A MAPS II Kit Instruction Manual Catalog Number 153-6159 For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800-4BIORAD (1-800-424-6723) Table of Contents Introduction...1

More information

Determination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography

Determination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography Application Note 206 Determination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography Introduction The Bayer Process is used to isolate alumina (Al 2 O ) from bauxite. In this process,

More information

ProteoMiner Protein Enrichment Technology

ProteoMiner Protein Enrichment Technology Sample Preparation ProteoMiner Protein Enrichment Technology Digging Deeper in the Proteome Detect More Proteins With ProteoMiner Technology Than With Immunodepletion ProteoMiner Technology vs. an Agilent

More information

ab83355 ATP Assay Kit (Colorimetric/ Fluorometric)

ab83355 ATP Assay Kit (Colorimetric/ Fluorometric) Version 18 Last updated 2 January 2018 ab83355 ATP Assay Kit (Colorimetric/ Fluorometric) For the rapid, sensitive and accurate measurement of ATP in a variety of samples. This product is for research

More information

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Student Learning Objectives High Performance Liquid Chromatography With the advent of relatively inexpensive and reliable pumps, the development of

More information

Imaging and Electrophoresis. Gel Doc EZ Imager. Accelerate your research. Push-button imaging in just 10 seconds.

Imaging and Electrophoresis. Gel Doc EZ Imager. Accelerate your research. Push-button imaging in just 10 seconds. Imaging and Electrophoresis Gel Doc EZ Imager Accelerate your research. Push-button imaging in just 10 seconds. ready imaging fault EZ imaging... Pick your tray and get your results 1LOAD TRAY Smart Tray

More information

Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES

Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES Marine Beauvir, Application Specialist, Thermo Fisher Scientific, Cambridge, UK Application Note 43161 Key Words High throughput,

More information

Capillary Electrophoresis of Proteins

Capillary Electrophoresis of Proteins Capillary Electrophoresis of Proteins SDS Capillary Gel Electrophoresis SDS-CGE Outline CE-SDS Gel Analysis Description of Technique Method Development Tips PA800 plus kits SDS-MW IgG Purity & Heterogeneity

More information

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions An Automated System for At-Line cess Analysis of Biopharmaceutical Fermentation Reactions Frank Steiner, Susanne Fabel, Fraser McLeod, Dionex Corporation, Germering, Germany Abstract Monitoring amino acid

More information

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility Technical Overview Introduction Agilent AdvanceBio SEC columns are a new family of size exclusion chromatography (SEC) columns

More information

SmartSpec Plus spectrophotometer

SmartSpec Plus spectrophotometer SmartSpec Plus spectrophotometer Simply Brilliant. The SmartSpec Plus has a more complete range of features and functions than many other benchtop spectrophotometers. It s a complete quantitation package

More information

BioPlex 2200 System. Like no other. The first and only fully-automated, random access, multiplex testing platform. Bio-Rad Laboratories

BioPlex 2200 System. Like no other. The first and only fully-automated, random access, multiplex testing platform. Bio-Rad Laboratories Bio-Rad Laboratories BIOPLEX 2200 SYSTEM BioPlex 2200 System Like no other. The first and only fully-automated, random access, multiplex testing platform Bio-Rad Laboratories BIOPLEX 2200 SYSTEM Fully

More information

Comparison of Different Brands of Carrier Ampholytes for Monoclonal Antibody Charge Heterogeneity Analysis by Capillary Isoelectric Focusing

Comparison of Different Brands of Carrier Ampholytes for Monoclonal Antibody Charge Heterogeneity Analysis by Capillary Isoelectric Focusing Comparison of Different Brands of Carrier Ampholytes for Monoclonal Antibody Charge Heterogeneity Analysis by Capillary Isoelectric Focusing Application Note Biotherapeutics & Biosimilars Author Christian

More information

protein interaction analysis tech note 5367

protein interaction analysis tech note 5367 protein interaction analysis tech note 5367 Rapid Optimization of Immobilization and Binding Conditions for Kinetic Analysis of Protein-Protein Interactions Using the ProteOn XPR36 Protein Interaction

More information

Determination of Sulfite in Food and Beverages by Ion Exclusion Chromatography with Pulsed Amperometric Detection

Determination of Sulfite in Food and Beverages by Ion Exclusion Chromatography with Pulsed Amperometric Detection Application Note 54 Determination of Sulfite in Food and Beverages by Ion Exclusion Chromatography with Pulsed Amperometric Detection INTRODUCTION Sulfite is a widely used food preservative and whitening

More information

Quick Manual to the Waters UPLC System

Quick Manual to the Waters UPLC System 1 Quick Manual to the Waters UPLC System 1. Application of Chiral HPLC Columns Warning! The HPLC columns are destroyed if the pressure overrides 2000 psi. The highest pressure applied for routine analysis

More information

Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal Antibody Charge Variants

Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal Antibody Charge Variants Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal ntibody Charge Variants Ken Cook, Frank Steiner, Mauro De Pra Thermo Fisher Scientific, Hemel Hempstead,

More information

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column mab Titer Analysis with the Agilent Bio-Monolith Protein A Column Application Note Biopharmaceuticals and Biosimilars Authors Emmie Dumont, Isabel Vandenheede, Pat Sandra, and Koen Sandra Research Institute

More information

QIAcube HT Your Purification Expert

QIAcube HT Your Purification Expert HT Your Purification Expert Fast and reliable 96-well nucleic acid purification Sample & Assay Technologies Economical, high-throughput nucleic acid purification from virtually all sample types enables

More information

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography Carl Fisher, Monika Verma, and Pranathi Perati; Thermo Fisher Scientific,

More information

Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis

Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis Atis Chakrabarti* and Roy Eksteen + Tosoh Bioscience LLC, King of Prussia, PA 19406 *Corresponding Author. + Current address:

More information

Protein-Pak Hi Res HIC Column and HIC Protein Standard

Protein-Pak Hi Res HIC Column and HIC Protein Standard Protein-Pak Hi Res HIC Column and HIC Protein Standard CONTENTS I. INTRODUCTION II. a. Mobile Phase b. Flow Direction CONNECTING COLUMN TO LC SYSTEM I. INTRODUCTION This offering contains non-porous, polymethacrylate-based

More information

Nuvia S and Q High-Capacity Ion Exchange Media Instruction Manual

Nuvia S and Q High-Capacity Ion Exchange Media Instruction Manual ... Nuvia S and Q High-Capacity Ion Exchange Media Instruction Manual Catalog numbers 156-0311 156-0411 156-0313 156-0413 156-0315 156-0415 156-0317 156-0417 Please read these instructions before you use

More information

Transwell Permeable Supports Including Snapwell and Netwell Inserts Instructions for Use

Transwell Permeable Supports Including Snapwell and Netwell Inserts Instructions for Use Transwell Permeable Supports Including Snapwell and Netwell Inserts Instructions for Use Life Sciences TABLE OF CONTENTS Introduction..................................... 1 Transwell Insert Product Descriptions...............

More information

SPRIworks Systems. Push button. Walk away. Fully automated library construction systems with built-in size selection and cleanup BR-15981B

SPRIworks Systems. Push button. Walk away. Fully automated library construction systems with built-in size selection and cleanup BR-15981B SPRIworks Systems Fully automated library construction systems with built-in size selection and cleanup Push button. Walk away. BR-15981B SPRIworks Technology Fully automated fragment library construction

More information

ab Porcine VEGF-A ELISA Kit

ab Porcine VEGF-A ELISA Kit Version 1 Last updated 27 January 2017 ab218298 Porcine VEGF-A ELISA Kit For the quantitative measurement of porcine VEGF-A in serum, plasma and cell culture supernatants. This product is for research

More information

PESTICIDE SCREENING APPLICATION SOLUTION

PESTICIDE SCREENING APPLICATION SOLUTION PESTICIDE SCREENING APPLICATION SOLUTION The most innovative pesticide screening solution ever built for food, beverage, and water testing. As the global food trade continues to expand, so does the challenge

More information

HPCE-5 2. Realising the potential of Capillary Electrophoresis. Why is the HPCE-512 so much better?

HPCE-5 2. Realising the potential of Capillary Electrophoresis. Why is the HPCE-512 so much better? HPCE-5 2 Realising the potential of Capillary Electrophoresis Why is the HPCE-512 so much better? A Major Advance in Capillary Electrophoresis The HPCE-512 system based on multi-point detection overcomes

More information

Thermo Scientific icap 7000 Plus Series ICP-OES. Gain more power. experience more performance

Thermo Scientific icap 7000 Plus Series ICP-OES. Gain more power. experience more performance Thermo Scientific icap 7000 Plus Series ICP-OES Gain more power experience more performance Thermo Scientific icap 7000 Plus Series ICP-OES Powerful, easy-to-use, solution for multi-element analysis Maximize

More information

rprotein A GraviTrap Protein G GraviTrap rprotein A/Protein G GraviTrap

rprotein A GraviTrap Protein G GraviTrap rprotein A/Protein G GraviTrap GE Healthcare Life Sciences Data file 28-9921-04 AA rprotein A GraviTrap Protein G GraviTrap rprotein A/Protein G GraviTrap Protein sample preparation rprotein A GraviTrap, Protein G GraviTrap, and rprotein

More information

Analysis of Illegal Dyes in Food Matrices Using Automated Online Sample Preparation with Liquid Chromatography-Mass Spectrometry

Analysis of Illegal Dyes in Food Matrices Using Automated Online Sample Preparation with Liquid Chromatography-Mass Spectrometry Analysis of Illegal Dyes in Food Matrices Using Automated Online Sample Preparation with Liquid Chromatography-Mass Spectrometry Yang Shi, Catherine Lafontaine, and François A. Espourteille Thermo Fisher

More information

Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass

Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass Application Note: 529 Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass Xiang He, Marta Kozak; Thermo Fisher Scientific, San Jose,

More information

NovaSeq Series. Custom Primers Guide

NovaSeq Series. Custom Primers Guide NovaSeq Series Custom Primers Guide Introduction 3 Primer Positions on the Cluster Cartridge 3 Compatible Primer Combinations 3 Prepare and Add Custom Primers 4 Set Up a Run for Custom Primers 5 Technical

More information

[ product solution ] Waters Oasis µ Elution PlateS. Patented Innovation. Elution volume as low as 25 μl. No evaporation and reconstitution

[ product solution ] Waters Oasis µ Elution PlateS. Patented Innovation. Elution volume as low as 25 μl. No evaporation and reconstitution [ product solution ] Waters µ Elution PlateS Elution volume as low as 25 μl No evaporation and reconstitution Ideal for small sample volumes Up to a 15x increase in concentration Patented Innovation Now

More information

Analysis of Trace Elements in Seawater Using the Thermo Scientific icap 7000 Series ICP-OES Duo

Analysis of Trace Elements in Seawater Using the Thermo Scientific icap 7000 Series ICP-OES Duo Analysis of Trace s in Seawater Using the Thermo Scientific icap 7000 Series ICP-OES Duo Christos Spanos, Applications Chemist, Thermo Fisher Scientific, Cambridge, UK Application Note 43185 Key Words

More information

TechnicalNOTE METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES INTRODUCTION

TechnicalNOTE METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES INTRODUCTION METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES Jose M. Castro-Perez 1, Jane Kirby 1, Laurent Leclercq 2, Claude Delatour 2, Jean-Michel Mineur 2,

More information

Chromatography Success Guide. Tips and Tricks for Successful Preparative Chromatography and Protein Purification

Chromatography Success Guide. Tips and Tricks for Successful Preparative Chromatography and Protein Purification Chromatography Success Guide Tips and Tricks for Successful Preparative Chromatography and Protein Purification This Chromatography Success Guide provides practical advice on preparative chromatography

More information

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website: Published in November 2013

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website:  Published in November 2013 5-100 Innovation Way Newark, DE 19711, USA Tel:302-3661101 Fax:302-3661151 Website: www.sepax-tech.com Published in November 2013 P/N SECKIT-7830 These Phases are developed based on innovative surface

More information

Bio-Rad Laboratories BIOPLEX 2200 SYSTEM. BioPlex 2200 System. Workflow: Redefined

Bio-Rad Laboratories BIOPLEX 2200 SYSTEM. BioPlex 2200 System. Workflow: Redefined BioPlex 2200 System Workflow: Redefined Workflow: Redefined Leveraging the power of multiplex testing, the BioPlex 2200 system brings flexible workflow solutions to your laboratory. Disease Focused Multiplex

More information

Application Note USD Purification of Mouse IgM from Cell Culture Supernatant by Cation Exchange Chromatography on CM Ceramic HyperD F Sorbent

Application Note USD Purification of Mouse IgM from Cell Culture Supernatant by Cation Exchange Chromatography on CM Ceramic HyperD F Sorbent Application Note USD 241 Purification of Mouse IgM from Cell Culture Supernatant by Cation Exchange Chromatography on CM Ceramic HyperD F Sorbent What this Study Demonstrates T h i s s t u d y o n C a

More information

METHCATHINONE Latest Revision: July 11, 2005

METHCATHINONE Latest Revision: July 11, 2005 METHCATHINONE Latest Revision: July 11, 2005 O CH 3 HN CH 3 1. SYNONYMS CFR: Methcathinone CAS #: Base: 5650-44-2 Hydrochloride: 49656-78-2 Other Names: N-methylcathinone Methylcathinone 2-Methylamino-1-phenyl-1-propanone

More information

ENHANCED MS WITH THE TURN OF A KEY

ENHANCED MS WITH THE TURN OF A KEY [ ionkey/ms ] ENHANCED MS WITH THE TURN OF A KEY EXPERIENCE THE FUTURE OF MS, EXPERIENCE ionkey/ms THE LC-MS YOU KNOW, LIKE YOU VE NEVER SEEN BEFORE ionkey/ms Systems integrate the UPLC separation into

More information

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC The Agilent 126 Infinity Bio-Inert Quaternary LC System and the AdvanceBio SEC 3Å, 2.7 µm Column Application Note Biologics

More information

Qualification of High-Performance Liquid Chromatography Systems

Qualification of High-Performance Liquid Chromatography Systems Qualification of High-Performance Liquid Chromatography Systems L. Huber Hewlett- Packard GmbH, Waldbronn December 1998 Manuscript of an article published in BioPharm, Vol 11 Number 11, November 1998,

More information

QIAcube Pure Efficiency

QIAcube Pure Efficiency QIAcube Pure Efficiency Sample & Assay Technologies Walkaway spin-column processing The QIAcube automates your spin preps The revolutionary QIAcube makes automated sample prep available to all labs. The

More information

Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application

Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application Food Authors Frank David Research Institute for Chromatography Pres. Kennedypark

More information

Automated and quantitative

Automated and quantitative Automated and quantitative analysis of biologics. PA 800 plus Pharmaceutical Analysis System Genomics Cell Analysis Particle Characterization Capillary Electrophoresis Lab Automation Centrifugation Bioseparation

More information

Bio-Monolith Protein G Column - More Options for mab Titer Determination

Bio-Monolith Protein G Column - More Options for mab Titer Determination Bio-Monolith Protein G Column - More Options for mab Titer Determination Application Note Biologics and Biosimilars Author Phu T. Duong Agilent Technologies, Inc. Introduction In recent years, monoclonal

More information

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus Robert Birdsall, Thomas Wheat, and Weibin Chen Waters Corporation,

More information

Simultaneous Determination of Inorganic Anions, Organic Acids, Amino Acids and Carbohydrates by Capillary Electrophoresis

Simultaneous Determination of Inorganic Anions, Organic Acids, Amino Acids and Carbohydrates by Capillary Electrophoresis Simultaneous Deteration of Inorganic Anions, Organic Acids, Ao Acids and Carbohydrates by Capillary Electrophoresis Tomoyoshi Soga Yokogawa Analytical Systems Inc., Tokyo, Japan Gordon A. Ross Hewlett-Packard

More information

Maximizing the Performance of Capillary Electrophoresis Systems The Maximizing Data Quality Series Part 4

Maximizing the Performance of Capillary Electrophoresis Systems The Maximizing Data Quality Series Part 4 Maximizing the Performance of Capillary Electrophoresis Systems The Maximizing Data Quality Series Part 4 Introduction Since the introduction of the ABI PRISM 310 Genetic Analyzer in 1996, capillary electrophoresis

More information

INTERNATIONAL PHARMACOPOEIA MONOGRAPH ON ARTEMETHER AND LUMEFANTRINE CAPSULES REVISED DRAFT FOR DISCUSSION

INTERNATIONAL PHARMACOPOEIA MONOGRAPH ON ARTEMETHER AND LUMEFANTRINE CAPSULES REVISED DRAFT FOR DISCUSSION October 2007 RESTRICTED ` INTERNATIONAL PHARMACOPOEIA MONOGRAPH ON ARTEMETHER AND LUMEFANTRINE CAPSULES REVISED DRAFT FOR DISCUSSION World Health Organization 2007 All rights reserved. This draft is intended

More information

QIAsymphony SP Protocol Sheet

QIAsymphony SP Protocol Sheet QIAsymphony SP Protocol Sheet Reference_200_V5 protocol General information These protocols are for purification of total DNA from samples encountered in forensic, human identity, and biosecurity applications

More information

determine optimum instrument settings for their own instruments and establish their own daily values.

determine optimum instrument settings for their own instruments and establish their own daily values. PC7 (770/488) SETUP KIT 6607121 PN 4299504-C FLOW CYTOMETER ALIGNMENT VERIFICATION FLUOROSPHERES FLOW CYTOMETER DETECTOR STANDARDIZATION FLUOROSPHERES INTENDED USE For Research Use Only. Not for use in

More information

PowerPlex 5C Matrix Standards, 310

PowerPlex 5C Matrix Standards, 310 TECHNICAL MANUAL PowerPlex 5C Matrix Standards, 310 Instruc ons for Use of Product DG5640 Printed 10/15 TMD050 PowerPlex 5C Matrix Standards, 310 All technical literature is available at: www.promega.com/protocols/

More information

Preparative HPLC is still the

Preparative HPLC is still the B I O P R O C E S S TECHNICAL Optimizing Sample Load Capacity and Separation Through a Series of Short Prep Columns Mark Crawford, Joan Stevens, and Luke Roenneburg Preparative HPLC is still the dominate

More information

C196-E067A. Prominence UFLCXR. Shimadzu Ultra Fast Liquid Chromatograph

C196-E067A. Prominence UFLCXR. Shimadzu Ultra Fast Liquid Chromatograph C196-E067A Prominence UFLCXR Shimadzu Ultra Fast Liquid Chromatograph New Prominence UFLC Series Line-up Prominence UFLCXR In Pursuit of Higher Separation and Sensitivity The Prominence UFLCXR is an ultra-high-speed

More information

Product Catalog # Description List Price (JPY) Primer Assays (desalted)

Product Catalog # Description List Price (JPY) Primer Assays (desalted) Assays and Controls Primer Assay Pricing Primer Assays (desalted) 10025636 Primer assay desalted, 200 reactions 16,000 (Wet-lab validated human, mouse, and rat) 10025637 Primer assay desalted, 1,000 reactions

More information

PowerPlex 5-Dye Matrix Standards, 310

PowerPlex 5-Dye Matrix Standards, 310 TECHNICAL BULLETIN PowerPlex 5-Dye Matrix Standards, 310 Instructions for Use of Product DG4600 Revised 3/16 TBD023 PowerPlex 5-Dye Matrix Standards, 310 All technical literature is available at: www.promega.com/protocols/

More information

Internal Comparative Evaluation Study of Alternative Enrichment Protocols for Detection of Salmonella in Untreated Spices

Internal Comparative Evaluation Study of Alternative Enrichment Protocols for Detection of Salmonella in Untreated Spices Internal Comparative Evaluation Study of Alternative Enrichment Protocols for Detection of Salmonella in Untreated Spices Tech Note Introduction Between 1973 and 2010, there were 14 outbreaks of foodborne

More information

AppNote 5/2006 ABSTRACT

AppNote 5/2006 ABSTRACT AppNote 5/26 Stir Bar Sorptive Extraction Based on Restricted Access Material for the Direct Extraction of Drugs and Metabolites in Biological Fluids Wayne M. Mullett and Elizabeth Kwong Merck Frosst,

More information

Pulsed Field Gel Electrophoresis

Pulsed Field Gel Electrophoresis Pulsed Field Gel Electrophoresis Protocol Bulletin 6225 Restriction Enzyme Digestion Restriction enzyme digestion is used to prepare DN for analysis or other processing steps. Guidelines about preparing

More information

QIAGEN Supplementary Protocol

QIAGEN Supplementary Protocol QIAGEN Supplementary Protocol PCR and data analysis using the cador T. equigenitalis PCR Kit on Rotor-Gene Q instruments This protocol is designed for using the cador T. equigenitalis PCR Kit for detection

More information

Gas Chromatography Assignment Chem 543/443

Gas Chromatography Assignment Chem 543/443 Gas Chromatography Assignment Chem 543/443 1. Introduction Capillary gas chromatography (GC) is one of the most popular analytical techniques used in today s research. Its popularity is mainly due to efficient

More information

hp hardware support onsite global next day response

hp hardware support onsite global next day response hp care pack hp hardware support onsite technical data hp hardware support onsite global next day response service overview benefits to you service highlights HP s global next day response service provides

More information

The Theory of HPLC Gradient HPLC

The Theory of HPLC Gradient HPLC The Theory of HPLC Gradient HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual. 1 Contents

More information

electrophoresis tech Performance Comparison of the Experion Automated Electrophoresis System and SDS-PAGE for Protein Analysis

electrophoresis tech Performance Comparison of the Experion Automated Electrophoresis System and SDS-PAGE for Protein Analysis electrophoresis tech note 5299 Performance Comparison of the Experion Automated Electrophoresis System and for Protein Analysis Karen Zhu, Marie Nguyen, William Strong, and Christina Whitman-Guliaev, Bio-Rad

More information

Keysight Technologies Laser Heated Tip and High Temperature Stage for G200 Nano Indenter. Data Sheet

Keysight Technologies Laser Heated Tip and High Temperature Stage for G200 Nano Indenter. Data Sheet Keysight Technologies Laser Heated Tip and High Temperature Stage for G200 Nano Indenter Data Sheet Overview The Nano Indenter G200 from Keysight Technologies, Inc. is the world s most flexible, and user-friendly

More information

protein interaction analysis

protein interaction analysis protein interaction analysis protocol guide 5821 Guide to Ligand Immobilization on the ProteOn XPR36 System Laura Moriarty, Bio-Rad Laboratories, Inc., 2000 Alfred Nobel Drive, Hercules, CA 94547 USA.

More information

A Practical User Guide for the Determination of Optimal Purification Gradients for the Gilson PLC 2020

A Practical User Guide for the Determination of Optimal Purification Gradients for the Gilson PLC 2020 A Practical User Guide for the Determination of Optimal Purification Gradients for the Gilson PLC 2020 Application Note PHA0211 Keywords Purification, Pharmaceuticals, Optimal Solvent Gradients, Isolation

More information

ENSURING QUALITY THROUGH COMPLIANCE [ COMPLIANCE ]

ENSURING QUALITY THROUGH COMPLIANCE [ COMPLIANCE ] ENSURING QUALITY THROUGH COMPLIANCE [ COMPLIANCE ] BEING IN COMPLIANCE HAS NEVER BEEN MORE IMPORTANT Quality encompasses more than meeting regulatory requirements. It extends throughout your organization

More information

BIOO RESEARCH PRODUCTS. Alkaline Phosphatase (AP) Color Endpoint Assay Kit Manual Catalog #:

BIOO RESEARCH PRODUCTS. Alkaline Phosphatase (AP) Color Endpoint Assay Kit Manual Catalog #: BIOO RESEARCH PRODUCTS Alkaline Phosphatase (AP) Color Endpoint Assay Kit Manual Catalog #: 3460-09 BIOO Scientific Corp. 2011 TABLE OF CONTENTS GENERAL INFORMATION... 1 Product Description... 1 Procedure

More information

Determination of Monocrotophos, Diazinon, Malathion, EPN, and Methamidaphos from Aqueous Samples Using Atlantic HLB SPE Disks

Determination of Monocrotophos, Diazinon, Malathion, EPN, and Methamidaphos from Aqueous Samples Using Atlantic HLB SPE Disks Determination of Monocrotophos, Diazinon, Malathion, EPN, and Methamidaphos from Aqueous Samples Using Jim Fenster, Julie McGettrick, Horizon Technology, Inc., Salem, NH Introduction Monocrotophos, Diazinon,

More information

Accuracy and Precision. Dilution Error in Error in Factor Accuracy Precision 200 <2% <2%

Accuracy and Precision. Dilution Error in Error in Factor Accuracy Precision 200 <2% <2% Specifications SDS-270 Sample Options Three sample racks, containing: 3 x 60 position x 16 mm OD (20 ml) (default configuration) or optionally: 3 x 90 position x 13 mm OD (13 ml) 3 x 40 position x 20 mm

More information

PAL SPME Fibers Optimized for Automation

PAL SPME Fibers Optimized for Automation PAL SPME Fibers Optimized for Automation PAL SPME Fiber Since its introduction by Pawliszyn et al. (ref. 1) Solid Phase Micro Extraction (SPME) has seen a tremendous development. SPME is a very effective

More information